» Guided Surgery. by MEDENTiKA « MEDENTIGUIDE SURGERY MANUAL MICROCONE. IPS Implant systems

Similar documents
» Guided Surgery. by MEDENTiKA « MEDENTIGUIDE SURGERY MANUAL QUATTROCONE QUATTROCONE30. IPS Implant systems

» Prosthetic Options by MEDENTiKA «

Pilot guided surgery

MICROCONE. IPS Implant Systems

For technical assistance: Tel Working time: 8.30 am pm

» QuattroFix treatment concept by MEDENTIKA «

High-end solution STEP-BY-STEP USER GUIDE

Multi-Unit Abutment System SIC Safe on Four. Optimum use of available bone by angled placement of implants

FOR C1 & V3 IMPLANTS STEP-BY-STEP GUIDED SURGICAL PROCEDURE

4766 Research Dr. San Antonio, TX insightdentalsystems.com

NEODENT GUIDED SURGERY

» Passion IMPLANT PICK-UP. for Precision « MPS Multi Platform Systems. Closed tray

Guided surgery manual

» MEDENTIKA PreFace Abutments

Implant System MAKE IT SIMPLE

Basic Information on Straumann Guided Surgery. Straumann Dental Implant System

The LOCATOR concept. Simplicity and versatility for prosthesis fixation

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

straumann Guided surgery Access to the Digital Experience

TABLE OF CONTENTS P. 4-5 P. 6-7 P. 8-9 P P P P

Patella Planing System

IMPRESSION-TAKING, BITE REGISTRATION, AND TEMPORARY RESTORATION ON CAMLOG IMPLANTS. a perfect fit

NARROW DIAMETER implant

Technique Guide. Compact 2.0 LOCK Mandible. The locking system for the mandible.

TSLP Thoracolumbar Spine Locking Plate

Guided Surgery System. Place implants with greater confidence. hahnimplant.com

INSTRUCTIONS FOR THE DENTIST. synocta impression procedure, Screw-retained and Snap-on. Screw-retained (Open tray) Snap-on (Closed tray)

MICROCONE. IPS Implant Systems

ACETABULAR CUP SURGICAL TECHNIQUE

Astra Tech Implant System. Manual and product catalog. Guided surgery. Computer guided implant treatment with the Astra Tech Implant System EV

PROXIMAL TIBIAL PLATE

SICvantage max Swiss Cross Technology

GLENOID SURGICAL TECHNIQUE

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

GMI. MONOLITH implant system. Surgical procedures guide

Very small abutment head easy and secure handling. Ankylos. The SmartFix concept. Prosthetic solution on angled implants

Correction System. Surgical Technique

SURGICAL PROCEDURES FOR ROXOLID IMPLANTS

Straumann Guided Adapter. Basic information on. the Straumann Guided Adapter

A complete surgical kit for Premium and Shelta systems

Correction System. Surgical Technique

GMI FRONTIER implant system. Surgical procedures guide

TRI + Digital Solutions DENTAL WINGS MANUAL

isy indications and techniques

NEW. The SmartFix Concept Prosthetic solution on tilted implants. SmartFix

Biomimetic Product catalogue

NEW NEW. An Extended Solution for Narrow Ridges

Instructions for Use DENTAL IMPLANTS

TITANIUM BASES CAD/CAM FOR CROWN AND BRIDGE RESTORATIONS CAD/CAM PROSTHETICS ON CAMLOG AND CONELOG IMPLANTS

Locator retained mandibular complete prosthesis (isy Implant System)

ORTH easy. The Complete System for Cortical Anchorage. OrthoEasy Pins

RATCHETS RATCHET ACCESSORIES DENTAL PRODUCTS PRODUCT CATALOG A company of the KLINGEL GROUP

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

GenXT. NextGenerationImplantology. Product Catalogue 2014

Guided immediate loading implant surgery planned with Implant Studio D.D.S. Jae-min, Lee

Anchorage system. tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

NARROW DIAMETER implant

GUIDED SURGERY TECHNIQUE

A complete surgical kit for Premium and Shelta systems

GUIDED SURGERY TOOL KIT (GSTK) & MASTER SLEEVES

RibFix Blu. Thoracic Fixation System

Clinical Evaluation Surgical Technique

A self-tapping internal hex. implant by MAKE IT SIMPLE

The REAL Mini Implant

Osteotomy System. The reference lines on the plate help facilitate the creation of the osteotomy, when a free hand cut is preferred.

FLH183 04/08. Biomet UK Ltd Waterton Industrial Estate Bridgend, South Wales CF31 3XA, United Kingdom. Tel. +44 (0) Fax: +44 (0)

PRODUCT CATALOG. Since 1985» Simple. Predictable. Profitable.

WINSTA-C. Clavicle Plating System

DOUBLE/TRIPLE PELVIC OSTEOTOMY PLATES For Treating Coxofemoral Joint Instability and Subluxation in Immature Dogs

Utilizing Digital Treatment Planning and Guided Surgery in Conjunction with Narrow Body Implants. by Timothy F. Kosinski, DDS, MAGD

Guidance for. implant removal. Straumann Dental Implant System

EasyLineImplant. Guided Surgery

Cochlear Baha Attract Surgery. Aurélie Dewael Education Manager Cochlear Academy EMEA

Product Catalog. Since 1985» Simple. Predictable. Profitable.

with Ø Ø 1.5 mm Titanium Screws distributed by ARCS Alveolar Reconstruction System E

Smarter Thinking. Simpler Design. Prima Plus. Surgical Manual

Conventus CAGE PH Surgical Techniques

AxSOS Locking Plate System

2.0 mm Mandible Locking Plate System

Cost-effective immediate restoration in the anterior region of the mandible (isy by CAMLOG)

Guided surgery as a way to simplify surgical implant treatment in complex cases

Technique Guide. SureLock Distal Targeting Device. C-arm guided targeting for trochanteric fixation nail.

User Manual Complete Prosthetic Kit (CPK)

Trefoil Procedure manual

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

TELEFIX SURGICAL TECHNIQUE. Implant system for the anterior stabilization of the thoracolumbar spine

TRINON. Index. Index. Overview Page 4. Catalogue

Bone Level. Surgical Guideline. Dental Implant System

Fixed Restoration for Atrophic Ridges

Implant Studio Patient Case

PRIMARY POROUS PATELLA WITH TRABECULAR METAL. Surgical Technique

HammerFUZE HAMMERTOE COMPRESSION SYSTEM SURGICAL TECHNIQUE

PROSTHETIC COMPONENTS

Technique Guide Hammertoe Correction System

I M P L A N T O L O G Y

General prosthetic advice for ANKYLOS

NEW. An Extended Solution. for Narrow Ridges

Distal Cut First Femoral Preparation

Transcription:

» Guided Surgery by MEDENTiKA «IPS Implant systems MEDENTIGUIDE MICROCONE SURGERY MANUAL

» MedentiGuide «MedentiGuide drill sleeves support the surgeon in preparing the implant bed for MEDENTiKA implants. Their use must be planned with a specially designed 3D planning system and surgical drilling template. You can plant the surgery with standard planning programs. Treatment planning based on three dimensional imaging procedures (CT, DVT) enables high precision treatment planning and means that the treatment outcome can be accurately predicted. These software manufacturers* currently support the MedentiGuide System The advantages over conventional planning include: Precision three-dimensional planning and implantation, taking into account the desired restoration Automatic collision control that displays if the distances to the implants or nerves are too short Information on peri-implant bone quality so that conclusions can be drawn on the expected primary stability An individual drilling template can be produced on the basis of the digital planning data. This ensures the exact and precise transfer of the planning outcome to the patient s mouth. Note: MEDENTiKA GmbH accepts no liability for the correct planning, implementation and production of the drilling template. Sufficient knowledge of the 3D planning system being used and the MEDENTiKA implant system is essential. It is imperative that the user is very confident in the use of 3D planning systems before using the MedentiGuide drill sleeves. Furthermore, sufficient expertise in preoperative implant planning and dental implantology is required. * to some extent this depends on the availability of the updates of the specific manufacturer. 2 3

SYSTEM DESCRIPTION» Drilling template «An individual drilling template can be produced on the basis of the digital planning data. This ensures the exact and precise application of the planning outcome to the patient s mouth. MedentiGuide drill sleeves can be used in drilling templates of various designs. The templates may be produced using suitable milling or printing systems with CADCAM technology or using alternative procedures.» MedentiGuide Insertion tool «The appropriate insertion tool is used to ensure that the outer sleeve is inserted securely into the template. NB: The drilling template must be firmly and securely seated on the jaw with no gaps. Drilling template with inserted outer sleeves Drilling template with inserted outer and innner sleeves 4 5

SYSTEM DESCRIPTION» Surgical Tray Drill lengths «MICROCONE Easy to use: No additional surgical tray is required with the MedentiGuide drill sleeve system. The drill sleeves are adapted to the Microcone standard drill. This significantly simplifies the preparation of the implantation and considerably reduces the material and storage costs. The MedentiGuide insertion tools are optionally available to guide the implants through the drill sleeves. The MedentiGuide System supports two drill lengths: 20 mm and 25 mm. In the planning phase it is important to ensure that the correct drill length is selected. PLEASE NOTE: The MedentiGuide drill sleeves are only designed for standard drills. The defined drilling depths do not include the drill tip of 0.3 mm. Please note their lengths if only minimum space is available to anatomical structures. Example for Microcone D 3.5 mm: 15 13 11 9 8 6.5 0 Standard drill Ø 3.0 mm, bone quality D3/D4 TWO DRILL LENGTHS: Short drill 1 marking ring Long drill 2 marking rings Total length 35.3 mm 20 mm Total length 40.3 mm 25 mm 0.3 mm drill tip 0.3 mm drill tip 6 7

SYSTEM DESCRIPTION» Sleeve in sleeve «MICROCONE MEDENTIGUIDE DRILL SLEEVES Locks in the rotational direction of the drill The inner sleeves are colour-matched to the corresponding colour-coded implants for easy recognition. Sleeve Art. No. Description Outer diameter Inner diameter Implant diameter 0-32-06 Standard outer sleeve D 6.3 d 5.03 D 3.0-4.5 mm Drill 0-32-07 0-32-08 Large outer sleeve Adapter sleeve D 8.3 d 7.03 D 5.0 mm D 7.0 d 5.03 D 5.0 mm MedentiGuide Drill Sleeves are a sleeve in sleeve system, made up of various outer sleeves and matching inner sleeves. MedentiGuide drill sleeves can be used in drill sleeves of various designs. The templates may be produced using suitable milling or printing systems with CADCAM technology or using alternative procedures. The MedentiGuide sleeve system works with a bayonet lock. The lock engages clockwise and thus in the rotational direction of the drill. 0-32-09 Inner sleeve D 5.0 d 2.03 Pilot 0-32-10 Inner sleeve D 5.0 d 2.53 D 3.0 mm 0-32-11 Inner sleeve D 5.0 d 3.03 D 3.5 mm D 4.5-3.5 mm 0-32-12 Inner sleeve D 5.0 d 3.53 D 4.0 mm 0-32-13 Inner sleeve D 5.0 d 4.03 D 4.5 mm 0-32-14 Inner sleeve D 7.0 d 4.53 D 5.0 mm 1-14-01 1-14-04 1-14-02 1-14-05 2-14-01 2-14-09 2-14-03 2-14-11 2-14-05 2-14-13 2-14-07 2-14-15 8 9

SYSTEM DESCRIPTION» MICROCONE implants D 3.0 4.5 mm MICROCONE implants D 5.0 mm «The standard outer sleeve is used in conjunction with the Microcone Implants D 3.0 - D 4.5. The large outer sleeve must be used for the Microcone implants D 5.0. d 2.02 d 2.02 H 5.0 mm Inner sleeve e.g. pilot drills Inner sleeve e.g. pilot drills H 5.0 mm D 5.3 mm d 5.03 D 5.0 mm d 5.0 H 4.0 mm Adapter sleeve To hold the standard inner sleeve Standard outer sleeve For implants measuring D 3.0-4.5 H 4.0 mm D 7.0 mm d 7.03 D 6.3 mm H 3.0 mm Large outer sleeve For implants D 5.0 mm D 8.3 mm MICROCONE IMPLANTS D 3.0 - D 4.5 MM MICROCONE IMPLANTS D 5.0 MM Pilot 0-32-09 D 3.0 0-32-10 D 3.5 0-32-11 D 4.0 0-32-12 D 4.5 0-32-13 Pilot 0-32-09 D 3.0 0-32-10 D 3.5 0-32-11 D 4.0 0-32-12 D 4.5 0-32-13 Inner sleeve D 5.0 0-32-14 Adapter sleeve 0-32-08 Standard outer sleeve 0-32-06 Large outer sleeve 0-32-07 Large outer sleeve 0-32-07 10 11

SYSTEM DESCRIPTION» Selection of drill length «MICROCONE The MedentiGuide System software supports the standard drill with two drill lengths: 20 mm and 25 mm. The distance from the top edge of the inner sleeve to the bottom edge of the implant is thus defined by the selected drill length. The distance between the lower edge of the outer sleeve and the top edge of the implant is always defined by the selected implant and drill lengths. The distance between the lower edge of the sleeve and the upper edge of the implant can be defined, depending on the desired implant length, by the choice of the drill length. NB: The desired drill length must therefore be taken into account during planning. Long drill Short drill 1 mm 4 mm 1 mm 20 mm 4 mm 4 mm 25 mm 9 mm 11 mm 11 mm 12 13

SYSTEM DESCRIPTION» MedentiGuide Placement Instrument«MICROCONE The implants can optionally be guided through the template for positioning. This involves the use of the corresponding MedentiGuide placement instruments, either for manual and ratchet or the contra-angled handpiece. The drill length determines whether the short or long MedentiGuide placement instrument is used. Long drill Short drill 6.5 8 9 11 11 13 15 13 Short drill & MedentiGuide placement instrument, short >> 1 marking ring Long drill & MedentiGuide placement instrument, long >> 2 marking rings 14 15

SYSTEM DESCRIPTION» MedentiGuide Placement instrument Implant alignment «MICROCONE MedentiGuide placement instrument, short To ensure that the implant is inserted at the correct height, it is rotated through the outer sleeve until the corresponding depth marker is flush with the top edge of the outer sleeve. 1 mm 4 mm Depth marker for implant length 8 mm If required, the implants can be aligned to the indexing of the implant connection. The MedentiGuide placement instrument is used in this case. The vertical marking on the placement instrument is aligned in the direction of the area of the square in the implant. The marking/notch on the outer sleeve is used as a guide to correctly position the implant. 20 mm 7 mm 8 mm Implant length 8 mm MedentiGuide Placement Instrument, long 1 mm Depth marker for implant length 15 mm 4 mm 5 mm 25 mm 15 mm Implant length 15 mm 16 17

» MICROCONE «drilling protocol (For example for D 4.5 x 15 mm implant with the standard outer sleeve) Initial situation Pilot hole with the Ø 2.0 mm pilot drill Optional incremental drilling; deep drill with the Ø 3.0 mm standard drill Deep drilling with the final Ø 4.0 mm standard drill Insert the implant either by hand/ratchet or with the contraangled handpiece. Submerged or transgingival healing can be achieved 1 2 3 4 5 6 Cut a flap in the soft tissue to expose the bone around the implantation point or optionally expose the implantation point with the mucosal punch. NB: Check the drilling template at every step to ensure that it is correctly seated. It is important to observe the surgical manual for each implant system. The pilot hole is drilled with the Ø 2.0 mm pilot drill. Guided by the drill sleeve (inner sleeve), this defines the sagittal direction of the implant axis and the drilling depth. The recommended speed is 300-600 rpm and the maximum speed is 800 rpm. Incremental drilling is performed with the corresponding drill sleeve (inner sleeve). In this case with the D 3.0 mm standard drill. The recommended speed is 300-600 rpm and the maximum speed is 800 rpm. Incremental drilling is performed with the corresponding drill sleeve (inner sleeve), in this case with the D 4.0 mm standard drill. The recommended speed is 300-600 rpm and the maximum speed is 800 rpm. NB: Before inserting the implant, remove the inner sleeve. Now the implant is screwed in position through the outer sleeve until the depth marking of the placement instrument is flush with the upper edge of the outer sleeve. Check the vertical marking on the placement instrument to ensure the desired alignment of the implant connection. The implant can heal either with submerged or transgingival methods. 18 19

» MICROCONE «drilling protocol (For example for D 5.0 x 15 mm implant with the large outer sleeve) Initial situation Pilot hole with the Ø 2.0 mm pilot drill Incremental drilling with the Ø 4.0 mm standard drill Deep drilling with the final Ø 4.5 mm standard drill Insert the implant either by hand/ratchet or with the contraangled handpiece. Submerged or transgingival healing can be achieved 1 2 3 4 5 6 Cut a flap in the soft tissue to expose the bone around the implantation point or optionally expose the implantation point with the mucosal punch. NB: Check the drilling template at every step to ensure that it is correctly seated. It is important to observe the surgical manual for each implant system. The pilot hole is made with the Ø 2.0 mm pilot drill. Guided by the drill sleeve (adapter sleeve and the corresponding inner sleeve), this defines the sagittal direction of the implant axis and the drilling depth. The recommended speed is 300-600 rpm and the maximum speed is 800 rpm. Incremental drilling is performed with the corresponding drill sleeve (adapter sleeve and inner sleeve). In this case with the D 4.0 mm standard drill. The recommended speed is 300-600 rpm and the maximum speed is 800 rpm. The adapter sleeve is now removed with the inner sleeve from the previous drilling. The inner sleeve (0-32-14) for the final drilling is then inserted into the external sleeve. Drill to achieve the final depth is performed with the final drill, which in this case is the D 4.5 mm standard drill. The recommended speed is 300-600 rpm and the maximum speed is 800 rpm. NB: Before screwing the D 5.0 implant into position, the inner sleeve has to be replaced with the adapter sleeve. Now the implant is screwed through the through the outer sleeve until the depth marking of the placement instrument is flush with the upper edge of the adapter sleeve. Check the vertical marking on the placement instrument to ensure the desired alignment of the implant connection. The implant can heal either with submerged or transgingival methods. 20 21

» MedentiGuide products «IPS Implant systems

24» Surgery Surgery «25 MedentiGuide Outer sleeve standard Titanium Grade 5 CF MedentiGuide Placement instrument Implant Manual and ratchet Stainless steel Diameter (mm) D 6.3 / d 5.03 Article No. 0-32-06 Please note: This sleeve is used for implants D 3.0 - D 4.5. Implant connection NI NI RI RI Type Microcone Microcone Microcone Microcone Version short long short long Article No. 1-32-03 1-32-04 2-32-03 2-32-04 Please note: These insertion tools are used to insert implants when using MedentiGuide sleeves. MedentiGuide Outer sleeve large Titanium Grade 5 CF MedentiGuide Placement instrument Implant Contra-angle Stainless steel MedentiGuide Adapter sleeve Titanium Grade 5 CF Diameter (mm) D 8.3 / d 7.03 Article No. 0-32-07 Please note: This sleeve is used for implants D 5.0. Implant connection NI NI RI RI Type Microcone Microcone Microcone Microcone Version short long short long Article No. 1-32-01 1-32-02 2-32-01 2-32-02 Please note: These insertion tools are used to insert implants when using MedentiGuide sleeves. MedentiGuide Inner sleeve Microcone implant Titanium Grade 4 Diameter (mm) D 7.0 / d 5.03 Article No. 0-32-08 Please note: This sleeve is used as a connecting piece between the Outer sleeve large and the Inner sleeves for the drill diameter D 2.0 - D 4.0. Placement instrument MedentiGuide Stainless steel Diameter (mm) D 5.0 / d 2.03 D 5.0 / d 2.53 D 5.0 / d 3.03 D 5.0 / d 3.53 D 5.0 / d 4.03 D 7.0 / d 4.53 Colour code white orange green black yellow blue Drill diameter D 2.0 mm D 2.5 mm D 3.0 mm D 3.5 mm D 4.0 mm D 4.5 mm Article No. 0-32-09 0-32-10 0-32-11 0-32-12 0-32-13 0-32-14 Type Outer sleeve standard Outer sleeve large Article No. 0-32-19 0-32-20

Publisher: MEDENTiKA GmbH Hammweg 8 10 76549 Hügelsheim Tel: +49 (0)7229 69912-10 info@medentika.de www.medentika.com Design: The WeberFink GbR graphic design studio www.weberfink.de Date: August 2018 We are certified according to: DIN EN ISO13485:2012 Medical Devices Directive 93/42/EEC, Annex II Subject to modification without notice. Errors and omissions excepted. The instructions for use and guarantee conditions are available on the website www.medentika.com. Please contact the manufacturer directly for further information concerning the guarantee.

MEDENTiKA GmbH Hammweg 8-10 76549 Hügelsheim info@medentika.de www.medentika.com» MEDENTiKA always well-treated «