Pulse oximetry in the accident and emergency department

Similar documents
The role of pulse oximetry in the accident and emergency department

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

Biomedical Instrumentation D. The Photoplethysmogram

Serum potassium responses to nebulized salbutamol administered during an

Pulse oximetry use in small animals

Appendix E Choose the sign or symptom that best indicates severe respiratory distress.

ORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers

ALCO Regulations. Protocol pg. 47

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve

QOF indicator area: Chronic Obstructive Pulmonary disease (COPD)

Permanent City Research Online URL:

Interscalene brachial plexus blocks in the management of shoulder dislocations

Is Hypoxia common despite oxygen treatment in the Acute Medical ward?

CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

Non-Invasive Monitoring

COPD. Ebrahim Razi, Hossein Akbari

audit? Missed myocardial ischaemia in the accident & emergency department: E.C.G. a need for

Muscle OxyGen monitor. The Science Behind Moxy

CLINICAL PROCEDURE PAEDIATRIC OBSERVATIONS

Knowledge and views of paediatricians about pulse oximetry: a nationwide online survey in Nigeria

Carbon Monoxide Exposure: Dräger PAC 3500 & RAD-57

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values

Your anaesthetic for heart surgery

Aaron J. Katz, AEMT-P, CIC Outward signs of what is occurring inside the body

RESPONSE OF 10 PULSE OXIMETERS TO AN IN VITRO TEST SYSTEM

Analgesia for chest trauma - RVI

EUROANESTHESIA 2008 Copenhagen, Denmark, 31 May - 3 June 2008 CENTRAL VENOUS OXYGEN SATURATION (SCVO 2 ): INTEREST AND LIMITATIONS

Monitoring in Anesthesia

This work is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 3.0 License.

Pulse Oximetry Screening in Newborns to Enhance the Detection Of Critical Congenital Heart Disease. Frequently Asked Questions

Appendix D An unresponsive patient with shallow, gasping breaths at a rate of six per minute requires:

Pulse Oximeter Accuracy Study

Oxygen: Is there a problem? Tom Heaps Acute Physician

Pulse Oximeter Accuracy Study

BTS Guideline for Home Oxygen use in adults Appendix 9 (online only) Key Questions - PICO 10 December 2012

Resident At Risk. The National Early Warning Score (NEWS) and Monitoring Vital Signs

Assessing Global Patient Status with the Halo Index

Absolute Cerebral Oximeters for Cardiovascular Surgical Cases

Trust Nonin Oximeters, Even For Your Sickest Patients

Chapter 11. Objectives. Objectives 01/09/2013. Baseline Vital Signs, Monitoring Devices, and History Taking

Admission of patient CVICU and hemodynamic monitoring

Non Invasive methods of Hemoglobin

Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings

Hooray for Hypoxia (or don t give too much oxygen)

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

Physiological Measurements Training for Care/Nursing & LD Small Group Home Staff

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO. MD (ANAESTHESIOLOGY) FINAL EXAMINATION AUGUST 2011 Time : 1.00 p.m p.m.

The use of a muscle relaxant to supplement local anaesthetics for Bier's

Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2): Oxford University Press

Pulse. Oximetry: Uses. and. Limitations. Susan DeMeulenaere. Continuing Education

Audit of patients with chest pain presenting to an accident and emergency department over a 6-month period

Chapter 26. Assisting With Oxygen Needs. Elsevier items and derived items 2014, 2010 by Mosby, an imprint of Elsevier Inc. All rights reserved.

Experiment B-18 Heart Rate and Coughing

GUIDELINES ON CONSCIOUS SEDATION FOR DENTAL PROCEDURES

Anaesthetic considerations for laparoscopic surgery in canines

COPD Challenge CASE PRESENTATION

Arch Angiography. Exceptional healthcare, personally delivered

Sprain or fracture? An analysis of 2000

Non-invasive Ventilation protocol For COPD

Maria Tracey, Director-Perioperative and Elaine Warren, Directory-Surgery Level. III (Three)

Transcatheter Aortic Valve Implantation Procedure (TAVI)

Introduction. Peripheral arterial disease. Hospital inpatient data - 5,498 FCE (2009/10), & 530 deaths in England alone

I. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device

Emergency First Response (EFR) Assessment Sheets September 2011

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE

Chest pain in the accident and emergency department: is chest radiography worthwhile?

Air or oxygen as driving gas for nebulised salbutamol

City, University of London Institutional Repository

Epidural Infusions for Pain Relief Including Discharge Advice

Day care adenotonsillectomy in sleep apnoea

NIV - BI-LEVEL POSITIVE AIRWAY PRESSURE (BIPAP)

Anaesthesia for ECT. Session 2. Dr Richard Cree Consultant in Anaesthesia & ICU. Roseberry Park Hospital and The James Cook Hospital, Middlesbrough

Vital Signs in Children 2015/16

Vital Signs Monitoring Clinical Skills

Commissioning for Better Outcomes in COPD

NRP Raising the Bar for Providers and Instructors

Direct Current (DC) Cardioversion

EMT. Chapter 14 Review

EFFECT OF METHYLENE BLUE ON THE ABSORBANCE OF SOLUTIONS OF HAEMOGLOBIN

SUBJECT INFORMATION AND CONSENT FORM

Evaluation of Pulse Rate and Arterial Oxygen Saturation (Sa0 ) Levels in Children During Routine Dental Procedures

Thoracic anaesthesia. Simon May

, OR 8.73 (95% CI

1.1.2 CPAP therapy is used for patients who are suffering from an acute type 1 respiratory failure (Pa02 <8kPa with a normal or low Pac02).

International Journal of Health Sciences and Research ISSN:

Evaluation of oesophageal reflectance pulse oximetry in major burns patients

Transfusion-Related Acute Lung Injury (TRALI) and Strategies for Prevention. Khalid Abdulla Sharif, MD, MRCP (UK)*

Repair of Hydrocele. Patient Information. Day Surgery. Ward 3, Leigh Infirmary

*Corresponding Author:

Brachial Plexus Block

Ventilation/End of Life Neuromuscular Disorders. Dr Emma Husbands Consultant Palliative Medicine

Int J Physiother. Vol 2(5), , October (2015) ISSN:

Clinical guideline for acute wheeze & asthma in children 5 years and over Hospital care

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Oxygen prescription and administration at the Emergency Department and medical wards in Mater Dei Hospital

Care of the Deteriorating Patient in Recovery NADIA TICEHURST : CLINICAL NURSE EDUCATOR PERI ANAESTHETICS BENDIGO HEALTH

Treating your abdominal aortic aneurysm by open repair (surgery)

Facilitating EndotracheaL Intubation by Laryngoscopy technique and Apneic Oxygenation Within the Intensive Care Unit (FELLOW)

Transcription:

Archives of Emergency Medicine, 1989, 6, 137-142 Pulse oximetry in the accident and emergency department C. J. HOLBURN & M. J. ALLEN Accident and Emergency Department, Leicester Royal Infirmary, Leicester, England SUMMARY The early detection and treatment of hypoxia is important to reduce patient morbidity in the accident and emergency department. At present, the commonly used methods all have practical difficulties in the urgent situations that prevail in the department. Pulse oximetry has recently become a method of choice in many anaesthetic and intensive care units for the continuous monitoring of oxygenation and the early detection of hypoxia. As similar conditions prevail in the accident and emergency departnent, we have attempted to evaluate its ease of use and the quality of information obtained in our department. Patients presenting with chest pain to an accident and emergency department have their oxygen saturation measured by the pulse oximeter finger probe prior to the commencement of oxygen therapy. After 5 min of oxygen therapy, the measurement was repeated. Our result showed that although no patients could be judged as hypoxic on clinical grounds the pulse oximeter showed, within 30 s of admission, that a number had an abnormal oxygen saturation. Continuous monitoring with the probe after the application of low flow oxygen therapy also aided in monitoring their treatment and this improvement was again easily and quickly recorded. Our experience shows that the pulse oximeter may be a useful tool for evaluating hypoxia and oxygen therapy in the accident and emergency department and we feel that we should be one of the groups who should reply in the positive to Zorab's question, 'Who needs pulse oximetry?' (Zorab, 1988). INTRODUCTION Cellular hypoxia precedes cell death in a large number of insults to the body. In the accident and emergency department, changes in the patient's oxygenation must be rapidly assessed and corrected to prevent unnecessary morbidity and mortality. Correspondence: Mr C. J. Holburn, Senior Registrar, Accident and Emergency Department, Dudley Road Hospital, Dudley Road, Birmingham, England. Arch Emerg Med: first published as 10.1136/emj.6.2.137 on 1 June 1989. Downloaded from http://emj.bmj.com/ on 30 October 2018 by guest. Protected by

138 C. J. Holburn and M. J. Allen The clinical detection of hypoxia uses the presence or absence of cyanosis as a guide but this is unreliable (Comroe et al., 1947) especially in the artificial lighting of many departments (Taylor et al., 1986). The measurement of arterial P02 using blood gas analysis is reproducible but in an unstable patient requires multiple invasive procedures giving intermittent values (Lam, 1987). Transcutaneous oxygen monitoring does provide continuous information but the sensor requires a stabilization period of about 30 min before reading can be commenced. As the sensor also requires the skin to be heated to 41 C to arterialize the capilliary bed, burns of the sensor area have been reported (Yelderman et al., 1983). Measurement of oxygen saturation by detecting the difference in absorption of light by haemoglobin and oxy-haemoglobin had been developed for in vitro analysis of blood and a modification of this equipment using 8 wavelengths of light became the Hewlett Packard ear oximeter. While this equipment could give accurate and continuous information it was bulky, expensive and unsuitable for routine clinical use (Knill et al., 1982). With the advent of light emitting diodes (LEDs) the bulk of the detection system was reduced, but this unheated and simpler equipment measured the total absorbance from tissues, veins and capillaries in addition to the absorbance due to arterial oxygen saturation. The pulse oximeter analyses not only the absorbance of light but by using microprocessors the changes of the signal produced by arterial pulsations. By excluding the absorbance of the constant part of the signal the arterial saturation is recorded and has been shown to reflect simultaneous invasive measurements accurately (Yelderman et al., 1983; Mackenzie, 1985). The modern pulse oximeter is a compact and easily mobile piece of equipment (Figure 1). It has been shown to be accurate over a wide variety of temperatures, cardiac index and mean arterial pressure. If, however, the equipment cannot identify a pulse Arch Emerg Med: first published as 10.1136/emj.6.2.137 on 1 June 1989. Downloaded from http://emj.bmj.com/ Fig. 1 r _. - The Biox 4 pulse oximeter showing oxygen saturation (SaO2) and the pulse rate (BPM). I on 30 October 2018 by guest. Protected by

Pulse oximetry in the A&E department 139 due to low perfusion of the vascular bed it displays a 'low perfusion' message and not an inaccurate value. This instrument has now been introduced into a wide variety of clinical areas most notably the anaesthetic room and intensive care unit. A recent leader in the British Medical Journal posed the question, 'Who needs pulse oximetry?' (Zorab, 1988) but did not specifically mention the accident and emergency department. No reports of its use have been found in the British accident and emergency literature, and we therefore undertook a study to assess the potential of the pulse oximeter in the accident and emergency department. METHOD Patients presenting to the accident and emergency department with the complaint of 'chest pain' who required monitoring were included in the study. The BTI Biox 4 Oximeter was used in the study. This displays the oxygen saturation and the patient's pulse rate (Figure 1). Directly on admission to the accident and emergency department each patient had the finger probe placed on the index finger (Figure 2) while breathing room air. The machine was allowed to settle to a constant oxygen saturation figure and this was recorded if the pulse rate reading was within 5 bpm of the radial pulse rate. Immediately following this Fig. 2 The pulse oximetry finger probe in position. reading the patient was then given controlled oxygen therapy, 4 1/min by Hudson Mask, an approximate FiO2 of 35%, this being the recommended treatment for chest pain of cardiac origin (Yates & Redmond, 1986). After 5 min of oxygen the oxygen saturation displayed on the machine was recorded. Arch Emerg Med: first published as 10.1136/emj.6.2.137 on 1 June 1989. Downloaded from http://emj.bmj.com/ on 30 October 2018 by guest. Protected by

140 C. J. Holburn and M. J. Allen RESULTS Fifty patients (32 male, 18 female) average age 59 years (range 23-84) were studied. None had clinical evidence of cyanosis or respiratory distress in the accident and emergency department. All had stable cardiovascular systems, none were hypotensive and none had evidence of bleeding. The initial pulse oximetry reading varied between 88-100% saturation (normal > 94%) (Oh, 1985) and this was recorded within 1 min of the patient entering the department. After 5 min of oxygen therapy the second reading was obtained and as expected showed a higher value (Figure 3) with all values being within the normal range. All those with abnormal initial values were admitted to hospital most (12/13) with chest pain of cardiac origin. None of the patients with abnormal initial saturations could be identified by clinical means and indeed some were felt by the nursing staff not to require oxygen therapy. Arch Emerg Med: first published as 10.1136/emj.6.2.137 on 1 June 1989. Downloaded from http://emj.bmj.com/ Prereratment oxygen saturation (%) Fig. 3 The pre- and post-oxygen therapy saturation. The lines show the lower limit of normal. on 30 October 2018 by guest. Protected by

DISCUSSION Pulse oximetry in the A&E department 141 Unrecognized and uncorrected hypoxaemia can increase the morbidity associated with a large number of the conditions that present to the accident and emergency department. Many of the therapeutic interventions that are used in the acutely ill aim to improve the level of tissue oxygenation and it is important that the results of these interventions can be quickly assessed. The clinical detection of hypoxia is difficult (Comroe et al., 1947) and is affected by a wide variety of environmental conditions (Taylor et al., 1986). The rapidity with which a patient's oxygenation can change has been recorded in the anaesthetic literature and the pulse oximeter has been instrumental in alerting anaesthetists to the hypoxaemia even minor insults can cause (Mihn et al., 1985; Tyler et al., 1985). The same rapidly changing conditions exist in the accident and emergency department and a method of evaluating a patient's oxygenation which is reliable and responds quickly to changing conditions is needed. We have attempted to use the pulse oximeter for this purpose and have found it to be easy to use and in the group of patients studied gave valuable information both before and during oxygen therapy. It alerted us to the fact that clinical judgement of oxygenation is difficult and that within a group of patients presenting with 'chest pain' some may have deficient oxygen delivery. Although the total numbers in this sample are small, all of the patients with saturation values below 94% were admitted to hospital while some with higher values were discharged. We cannot determine whether the low saturation obtained on admission was as a result of pre-existing disease or due to the presenting condition but in patients with myocardial ischaemia any reduction in saturation from normal imposes a greater load on the heart which can further damage the myocardium. As the patients did not have blood gas analysis during this study we cannot be sure that the initial value represents just oxyhaemoglobin as other abnormal haemoglobin types can give a false value with this particular equipment but we can be sure that the increased value does represent an increase in the saturation. As the basic method of evaluation of oxygen saturation by this method relies on the presence of a pulse wave at the recording point the technique as described in this paper is not suitable for use in the hypotensive patient. This may preclude its use in trauma although we have no experience of its use in this field. As the technology of the instruments improves, the detection of oxygen saturation at lower blood pressures may become easier. The recent increase in litigation has prompted editorial comment (Mathias et al., 1987) on the need for early detection of iatrogenic complications of treatment. The author recommends the pulse oximeter as a useful tool for the anaesthetist. In accident and emergency departments the same vigilance is necessary and the pulse oximeter provides the information quickly and in an easily readable form. As the cost of investigations increases the use of reusable non-invasive techniques to reduce the number and frequency of blood gas analysis without replacing them completely will become an important consideration in clinical practice (Dautzenberg, 1986). Arch Emerg Med: first published as 10.1136/emj.6.2.137 on 1 June 1989. Downloaded from http://emj.bmj.com/ on 30 October 2018 by guest. Protected by

142 C. J. Holburn and M. J. Allen CONCLUSIONS In a small pilot study oxygen saturation was quickly and easily measured by the noninvasive pulse oximeter in a group of acutely ill patients presenting to an accident and emergency department. Further studies are required to establish the exact role of this instrument in the normal day-to-day practice of emergency medicine. REFERENCES Comroe J. H. & Botalho S. (1947) The unreliability of cyanosis in the recognition of arterial anoxemia. American Journal of the Medical Sciences 214, 1-6. Dautzenberg B., Gallinari C., Moreau A. & Sors C. (1986) The advantages of real-time oximetry over intermittent arterial blood gas analysis in a chest department In Pulse Oximetry, Payne J. P. & Severinghaus J. W.(eds), pp. 63-5. Springer-Verlag, Berlin. Knill R. L., Clement J. L., Kieraszewicz et al., (1982) Assessment of two noninavasive monitors of arteria oxygenation in anaesthetized man. Anaesthesia & Analgesia 61, 582-6. Lam A. M. (1987) Continuous arterial P02 monitoring, In: Monitoring of Oxygen: A Panel Summary. Canadian Anaesthetists' Society Journal 34, 56-63. Mackenzie N. (1985) Comparison of a pulse oximeter with an ear oximeter and an in-vitro oximeter. journal of Clinical Monitoring 1, 156-60. Mathias J. A. & Lunn J. N. (1987) Minimal monitoring and vigilance. Anaesthesia 42, 683-4. Mihn F. G. & Halperin B. D. (1985) Noninvasive detection of profound arterial desaturations using a pulse oximetry device. Anaesthesiology 62, 85-7. Oh T. E. (ed) (1985) Intensive Care Manual. Butterworth, London. Taylor M. B. & Whitwam J. G. (1986) The current status of pulse oximetry. Anaesthesia 41, 943-9. Tyler I. L., Tantisira B., Winter P. M., et al., (1985) Continuous monitoring of arterial oxygen saturation with pulse oximetry during transfer to the recovery room. Anaesthesia & Analgesia 64 1108-12. Yates D. W. & Redmond A. D. (1986) Lecture Notes on Accident and Emergency Medicine. Blackwell Scientific Publications, Oxford. Yelderman M. & New W. (1983) Evaluation of pulse oximetry. Anaesthesiology 59, 349-52. Zorab J. S. M. (1988) Who needs pulse oximetry? British Medical Journal 296, 658-9. Arch Emerg Med: first published as 10.1136/emj.6.2.137 on 1 June 1989. Downloaded from http://emj.bmj.com/ on 30 October 2018 by guest. Protected by