Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Similar documents
On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy


Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

International Association of Scientific Innovation and Research (IASIR) (An Association Unifying the Sciences, Engineering, and Applied Research)

Comparative Study on Homogeneous and Inhomogeneous Photon Dose Distribution from a High Dose Rate 192 Ir Brachytherapy Source using MCNP5

Validation of the MC platform GATE for radiation therapy applications Evaluation of dose deposited by electrons. Yann Perrot, LPC, CNRS/IN2P3

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

IN VIVO IMAGING Proton Beam Range Verification With PET/CT

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Review of TG-186 recommendations

Neutron Measurements for Intensity Modulated Radiation Therapy

[Setawati et. al., Vol.4 (Iss10): October, 2017] ISSN: DOI: /zenodo

Accuracy of the Small Field Dosimetry Using the Acuros XB Dose Calculation Algorithm within and beyond Heterogeneous Media for 6 MV Photon Beams *

Epithermal neutron beams from the 7 Li(p,n) reaction near the threshold for neutron capture therapy

Correction of Prompt Gamma Distribution for Improving Accuracy of Beam Range Determination in Inhomogeneous Phantom

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

Applications of Monte Carlo simulations to radiation dosimetry

Diode calibration for dose determination in total body irradiation

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments


Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry

A comparative dosimetric analysis of the effect of heterogeneity corrections used in three treatment planning algorithms

Imaging of Radiation Dose Using Cherenkov Light

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

STUDY OF DOSIMETRIC EFFECTS DUE TO METALLIC HETEROGENEITY OF TISSUE EXPANDERS IN POST-MASTECTOMY RADIOTHERAPY

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Transmission portal in vivo dosimetry by means of the Monte Carlo method and the mathematical programming language MATLAB

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe

Perturbation effects of the carbon fiber-peek screws on radiotherapy dose distribution

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

Effectiveness of compensating filters in the presence of tissue inhomogeneities

Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

Hampton University Proton Therapy Institute

Research Article Dose Distributions of an 192 Ir Brachytherapy Source in Different Media

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies

Calculation methods in Hermes Medical Solutions dosimetry software

Verification of micro-beam irradiation

Small field diode dosimetry

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

INTRODUCTION. Material and Methods

Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film

Limits of Precision and Accuracy of Radiation Delivery Systems

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

PROSTATE BRACHYTHERAPY DOSIMETRY USING VOXEL PHANTOMS

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Increased dose near the skin due to electromagnetic surface beacon transponder

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Basic dosimetric data for treatment planning system

Neutron dose evaluation in radiotherapy

Radiochromic film dosimetry in water phantoms

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Electron Beam ET - Reloaded Therapy - Reloaded

Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams.

Application of the Multi-Model Monte-Carlo Treatment Planning System Combined with PHITS to Proton Radiotherapy

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

Comparative evaluation of modern dosimetry techniques near low- and high-density heterogeneities

A COMPARISON BETWEEN GEANT4 AND MCNPX ON THE DOSIMETRY OF THE 192IR MICROSELECTRON V2 HDR BRACHYTHERAPY

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

A Monte Carlo Study for Photoneutron Dose Estimations around the High-Energy Linacs

Venue: IEEE NSS/MIC/RTSD Conference, Seoul, South Korea, 27 th October 2013 Workshop: NWK3/RD1 Radiation Protection and Dosimetry

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

V. 4. Design and Benchmark Experiment for Cyclotron-based Neutron Source for BNCT

Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

Evaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife

Transcription:

Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J. Phys.: Conf. Ser. 739 012117 Related content - Neutron contamination of Varian Clinac ix 10 MV photon beam using Monte Carlo simulation S Yani, R Tursinah, M F Rhani et al. - Evaluation of lens absorbed dose with Cone Beam IGRT procedures R Palomo, M C Pujades, J Gimeno-Olmos et al. - An empirical formula to obtain tissuephantom ratios from percentage depth-- dose curves for small fields George X Ding and Rob Krauss View the article online for updates and enhancements. This content was downloaded from IP address 148.251.232.83 on 30/10/2018 at 18:41

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system S Yani 1,2, M F Rhani 3, F Haryanto 1 and I Arif 1 1 Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung, Jalan Ganesa 10 Bandung 40132, Indonesia 2 Akademi Kebidanan Pelita Ibu Kendari, Jalan Syech Yusuf 12 Kendari 90132, Indonesia 3 Department of Radiation Oncology, National University Cancer Institute Singapore, Singapore/Singapore E-mail: sitti.yani@s.itb.ac.id Abstract. Treatment fields consist of tissue other than water equivalent tissue (soft tissue, bones, lungs, etc.). The inhomogeneity effect can be investigated by Monte Carlo (MC) simulation. MC simulation of the radiation transport in an absorbing medium is the most accurate method for dose calculation in radiotherapy. The aim of this work is to evaluate the effect of inhomogeneity phantom on dose calculations in photon beam radiotherapy obtained by different MC codes. MC code system EGSnrc and Geant4 was used in this study. Inhomogeneity phantom dimension is 39.5 30.5 30 cm 3 and made of 4 material slices (12.5 cm water, 10 cm aluminium, 5 cm lung and 12.5 cm water). Simulations were performed for field size 4 4 cm 2 at SSD 100 cm. The spectrum distribution Varian Trilogy Clinac ix 10 MV was used. Percent depth dose (PDD) and dose profile was investigated in this research. The effects of inhomogeneities on radiation dose distributions depend on the amount, density and atomic number of the inhomogeneity, as well as on the quality of the photon beam. Good agreement between dose distribution from EGSnrc and Geant4 code system in inhomogeneity phantom was observed, with dose differences around 5% and 7% for depth doses and dose profiles. 1. Introduction The accuracy of dose calculation in the treatment planning is important to ensure the appropriate dose delivered to patients undergoing radiation therapy. In the past 20 years, radiation therapy has become increasingly thrive. Uncertainty in dose calculation is very challenging today. Many researchers develop methods to reduce this in order to inhomogeneity inside treatment fields. This treatment fields consist of tissue other than water equivalent tissue (soft tissue, bones, lung etc). Where the tissue is not water, equivalent distortions to the beam profiles will occur. If the density is substantially less than that of water, as in lung or in air cavities, the attenuation will be reduced. Photon dose distribution using analytical methods are associated with large uncertainties, especially in irradiated volumes that contain inhomogeneity such as air cavities and bones. In clinically relevant dose calculations, errors of 10% or greater have been reported to be associated with analytic dose calculation methods. Monte Carlo (MC) techniques are becoming more widely used in all medical physics applications. MC simulation of radiation transport is one of the most accurate methods for predicting absorbed dose Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by Ltd 1

distributions in radiation therapy. The MC method is the most accurate method for dose calculation in photon beams and can potentially reduce these uncertainties to a few percent. MC algorithm would be acceptable for determining the absorbed dose in the presence of inhomogeneous tissue if the Monte Carlo code is tested sufficiently. In particular, MC simulation can handle backscatter from highdensity materials (such as aluminium) or scatter perturbations by low density material (such as lung) more accurately than any other current dose calculation method. Some studies have been conducted to investigate inhomogeneity effect using Monte Carlo simulation for external beam radiation therapy (EBRT) [1, 2] and brachytherapy [3, 4]. There are different MC codes for the simulation of photons, electrons and the coupled transport of electrons and photons e.g. MCNPX, Electron Gamma Shower (EGS) and Geant4. These code generally used in medical physics application [5, 6, 7]. Both of EGSnrc and Geant4 were used to model and simulate particle transport and interaction with matter across a wide range of energies using MC algorithm. Geant4 is a Monte Carlo-based code that is a successor of GEANT3 developed in two independent studies at CERN and KEK in 1993 [6]. This code have a wide range energy than EGSnrc code. Figure 1. Varian Trilogy Clinac ix linear accelerator (source: Tan Tock Seng Hospital (TTSH) Singapore). The aim of this study is to evaluate the effect of inhomogeneity phantom on dose calculations in Varian Trilogy Clinac ix 10 MV photon beam (figure 1) obtained by Geant4 and EGSnrc code system. Neutron can be emitted from high energy linear accelerator. For this study of inhomogeneity phantom on dose distribution used Geant4 code system to detect the existence of these particles. 2. Methods Monte Carlo simulations were performed using Geant4 [8] and DOSXYZnrc the user code of EGSnrc [9]. These codes use three-dimensional heterogeneous geometry and transports photons and electrons in the energy range from 1 KeV to some GeV. But, one of the limitation of EGSnrc is in order to calculate the neutron contamination and interaction in a simulation. The accelerator employed was a Varian Trilogy Clinac ix Varian Oncology Systems, Palo Alto, CA with the available nominal energies of 4, 6, 8, 10, 15, 18, and 20 MeV for photon beam. In this study we performed dosimetric verifications of the treatment planning system limited to 10 MeV photon beam. The dose profiles 2

curves and percent depth doses curves from Geant4 and EGSnrc were compared both of homogeneous and inhomogeneous phantom. 2.1. Homogeneous phantom A water phantom was used as a homogenous phantom for relative dosimetry. The phantom was a cube of dimensions 39.5 30 30 cm 3 were placed in the radiation field for scanning (figure 2a). 2.2. Inhomogeneous phantom The phantom were made of 4 material slices (3 cm water, 2 cm aluminium as high density material, 7 cm lung as low density material and 18 cm water) (figure 2b). The materials chosen had linear collision stopping powers and linear angular scattering powers close to body tissues. The aluminum and lung material were chosen to represent high and low density material, respectively. Sometimes, some patients have implants in their body that consist of metal or another high density materials. Inhomogeneity phantom dimension is 39.5 30.5 30 cm 3. This inhomogeneity phantom adopt from Jan et. al. (2011) [10]. beam axis beam axis X Y Water Z 30 cm 3 cm 2 cm 7 cm Water Aluminum Lung 30 cm 18 cm Water 30.5 30.5 cm 39.5 cm (a) Figure 2. Phantom design (a) Homogenous phantom (b) Inhomogeneous phantom. Simulations were performed for a square field size 4 4 cm 2 at a source-to-axis distance (SSD) of 120 cm. The source used in Geant4 and DOSXYZnrc simulation is point source from the front of rectangular collimation (source 3) using the spectrum distribution Varian Trilogy Clinac ix 10 MV photon beam from Tan Tock Seng Hospital (TTSH) Singapore (figure 3). About 10 9 histories are traced here to perform the simulation for evaluation of the homogenous and inhomogeneous phantom on dose distribution. The scoring voxel arrangement for percentage depth dose (PDD) and profile dose calculation is different to reduce simulation time and concern on high dose gradient and build-up region [11]. This simulation takes about 12-14 hours to complete depend on voxel sizes and the number of computer core. MC simulated and measured dose distributions in a water phantom. The output of this simulation i.e. the percent depth dose and dose profile in d m. For the calculations in this study, the global cut-off energies used in the simulations were electron cut-off energy (ECUT) = 521 KeV for electrons and photons global cut-off (PCUT) = 10 KeV for photons. The dose distribution from this simulation was analyzed and plotted using a STATDOSE code. The number of histories was selected in order to get the desired statistical uncertainty (standard deviation) on the dose calculation of about less than 2%. 39.5 cm (b) 3

3. Results and discussion Figure 3. Spectrum energy Varian Trilogy Clinac ix 10 MV photon beam. 3.1. Homogeneous phantom Figure 4 summarize the differences the calculated (simulated) dose in PDD and profile dose for field size 4 4 cm 2 and SSD 120 cm between Geant4 and EGSnrc code system. The discrepancies between measured and calculated dose data are within 5% of D max (D max is phantom depth with maximum dose). The dose profile simulate to verify the field size used in simulation in homogenous (water) phantom (figure 4). Dose profile scan the dose distribution along Y axis (in the plane perpendicular to the beam axis) in 10 cm depth. Figure 4. Comparison of 10 MV beam profile dose curve along the Y axis (crossline) between Geant4 and EGSnrc for a 4 4 cm 2 field in the homogeneous (water) phantom. The black line represents the dose distribution data obtained from EGSnrc, red line represents the data obtained from Geant4 and blue line represents the data obtained from measurement data From figure 4, there is a great deviation between Geant4 and EGSnrc dose distribution in the high dose gradient area. The full width at half maximum (FWHM) of each graph are 3.95, 4 and 4.3 cm for Geant4, EGSnrc and measurement data. The MC EGSnrc and Geant4 calculated dose functions show very good agreement for distances up to 5 cm with literature data. For distances greater than 5 cm, the relative difference between both calculated data and literature data. MC study of dosimetric parameters and dose distribution are slightly larger. A larger difference of dose distribution between measurement and MC calculation was in the penumbra region because of the changes in voxel sizes and source geometry. 4

The dose along the central axis calculated with EGSnrc and Geant4 and measurement data for water phantom with field size 4 4 cm 2 is plotted in figure 5. This data produce the depth dose profiles with more than 7% dose discrepancy for this inhomogenous water phantom. The depth maximum of PDD is shifted to the big depth around 0.1 and 0.2 cm for EGSnrc and Geant4, respectively. Figure 5. Comparison of 10 MV beam profile dose curve along the Y axis (crossline) between Geant4 and EGSnrc for a 4 4 cm 2 field in the homogeneous (water) phantom. The black line represents the dose distribution data obtained from EGSnrc, red line represents the data obtained from Geant4 and blue line represents the data obtained from measurement data 3.2. Inhomogeneous phantom An inhomogeneous phantom (water-aluminum-lung-water) with dimension of 39.5 30.5 30 cm 3 was employed. Percent depth dose (PDD) and dose profile were also investigated by Geant4 and EGSnrc code system. PDD and profile dose scan along Z axis (depth) and Y axis (depth 10 cm), respectively. Figure 6 shows comparisons of PDD in inhomogeneous phantom at depth 10 cm for the 10 MV photon beam with calculation using Geant4 and EGSnrc with field size 4 4 cm 2 and SSD 120 cm. The density of water, aluminum, and lung are about 1.0, 2.7, and 0.26 g/cm 3, respectively. For the inhomogeneous phantom, there is a great deviation in PDD and lateral profile between Geant4 and EGSnrc dose calculation. Water Al Lung Water Figure 6. Comparison of 10 MV beam PDD curve along the central axis between Geant4 and EGSnrc for a 4 4 cm 2 field in the inhomogeneous (water-aluminum-lung-water) phantom. The beam is incident on the top of phantom with its central axis on the center of the top water slab. Figure 6 shows that there is difference between dose of presence of aluminium and lung as inhomogeneous by EGSnrc and Geant4. Form the figure, it is easy to see by looking at the plotted that there is a big deviation between dose distribution produced by EGSnrc and Geant4 in the lung region. The EGSnrc result shows that behind aluminium material, the dose was found to be increased by 5

6.41%, and behind lung material, the dose was found decreased by 2.93%. Behind aluminium material, the dose was found to be increased by 4.73%, and behind lung material, the dose was found decreased by 1.54% in the Geant4 result. In the lung inhomogeneity, the dose was decreased, which may be due to increased attenuation of radiation, and its high density. Behind aluminium inhomogeneity (among water and aluminium slice), the dose was increased, which may be due to decreased attenuation of radiation, and its less density. There was dose increasing in the border area between lung and water (10-15 cm). The attenuation coefficient of water is bigger than lung material. Some photons passed by lung was forwarded to water. Most photons with various energy would accumulated in the boundary and caused the rose of dose in this area. PDD curve shows that there is dose reduced across the low density media both of EGSnrc and Geant4. Photon fields exposing low density material will undergo rose range of scattered electrons and decreased photon attenuation. Increased of electron range further restricts the scattering angle needed to keep the electrons within the field. The result of decreasing material density is small number of photons interacting and secondary electron depositing their energy further downstream if they deposit their energy in the field at all. A small field size (4 4 cm 2 ) used in this simulation results in more electrons leaving the volume of interest then entering it. This effect known as lateral electronic disequilibrium (LED) and produce a dose deficit in a low density region. In fact, photon and electron scatter change in the vicinity of the interface either reduced upon passing from high to low densities or rose when passing from low to high densities. Figure 7. Comparison of 10 MV beam lateral dose profile along the central axis between Geant4 and EGSnrc for a 4 4 cm 2 field in the inhomogeneous (water-aluminum-lung-water) phantom. The beam is incident on the top of phantom with its central axis on the center of the top water slab. Figure 7 shows the lateral dose profile a long Y axis (perpendicular with beam axis) in the inhomogeneous phantom by EGSnrc and Geant4 simulation. The big deviation founded on the penumbra region in the right side. The FWHM of EGSnrc and Geant4 curve are 4.4 and 4.05 cm, respectively. The effects of inhomogeneities on radiation dose distributions depend on the amount, density and atomic number of the inhomogeneity, as well as on the quality of the photon beam. 4. Conclusion Monte Carlo simulations were performed using Geant4 [8] and DOSXYZnrc the user code of EGSnrc [9]. These codes use three-dimensional heterogeneous geometry and transports photons and electrons in the energy range from 1 KeV to some GeV. But, one of the limitation of EGSnrc is in order to calculate the neutron. 6

Acknowledgements We are grateful to Tan Tock Seng Hospital, Singapore for the spectrum data for the Varian Clinac ix 10 MV photon beam beam. This study was partially supported by Penelitian Kerjasama Luar Negeri dan Publikasi Internasional Dikti, Ministry of Education Indonesia 2015 (312a/I1.C01/PL/2015). References [1] Cardoso S C, Alves V G L, da Rosa L A R, Campos L T, Batista D V S and Facure A 2010 PLoS ONE 5(5) e10466 [2] Chandola R M, Tiwari S, Painuly N K, Choudhary V, Azad S K and Beck M 2011 Journal of Cancer Research and Therapeutics 7(1) 29-34 [3] Lloyd S A M and Ansbacher W 2013 Med. Phys. 40(1) 011707-1-5 [4] Robinson D 2008 Journal of Applied Clinical Medical Physics 9(2) 1-11 [5] Nedaie H A, Mosleh Shirazi M A and Allahverdi M 2013 Journal of Medical Physics 38(1) 15-21 [6] Schiapparelli P, Zefiro D and Taccini G 2009 Med. Phys. 36(5) 1759-67 [7] Grevillot L, Frisson T, Maneval D, Zahra N, Badel J N and Sarrut D 2011 Phys. Med. Biol. 56 903 918 [8] Agostinelli S et al. 2003 Nuclear Instruments and Methods in Physics Research 506 250-303 [9] Kawrakow I and Rogers D W O 2013 The EGSnrc code system: Monte Carlo simulation of electron and photon transport (Ottawa: National Research Council of Canada) p. 16 [10] Jan S et al. 2011 Phys. Med. Biol. 56 881 901 [11] Yani S, Dirgayussa I G E, Rhani M F, Haryanto F and Arif I 2015 AIP Conference Proceeding 1677 040002-1-4 doi: 10.1063/1.4930646 doi: 10.1063/1.4930646 7