PATHWAY AVID SPINAL ELEMENTS PATHWAY AVID. CAUTION: Federal law (U.S.A.) restricts this device to sale by or on the order of a licensed physician.

Similar documents
PATHWAY PLIF and PATHWAY TLIF. SPINAL ELEMENTS PATHWAY PLIF And PATHWAY TLIF PATHWAY

SECURIS SPINAL Fixation System Package Insert

ISSYS LP Spinal Fixation Systems Package Insert

OPTIMUS CUSTOM SPINE OPTIMUS: STAND-ALONE ANTERIOR LUMBAR DEVICE

QIN4432TRICREV01. IMPORTANT PRODUCT INFORMATION FOR TRITANIUM C Cervical Cage. Version 1 1 STERILE PRODUCT

Features & Benefits. 1) Wide Bone Graft Space. Anterior Margin 3mm Posterior Margin 1.5mm. Marker Margin

INTELLIGENT SPINAL SYSTEM

STRYKER SPINE Spinal Fixation Systems XIA 3

PLIF/TPLIF Cage System

nvt Transforaminal Lumbar Interbody Fusion System

nvp Posterior Lumbar Interbody Fusion System

nva Anterior Lumbar Interbody Fusion System

A U X I L I A R Y C O N N E C T O R S Surgical Technique

LBL-014. Rev

Alamo T Transforaminal Lumbar Interbody System Surgical Technique

Alamo C. Cervical Interbody System Surgical Technique. An Alliance Partners Company

Aesculap CeSpace TM PEEK and CeSpace TM XP Spinal Implant System Instructions for Use

OPERATIVE TECHNIQUE. CONSTRUX Mini PTC. Mini PTC Spacer System

TABLE OF CONTENTS. ShurFit Anterior Cervical Interbody Fusion (ACIF) System Overview 2. Implant Specifications 3. Instrument Features 4

BONE GRAFT WASHER _Rev A IMPORTANT INFORMATION ON THE BONE GRAFT WASHER

Zeus Intervertebral Body Fusion Devices PACKAGE INSERT

Technique Guide KISSloc Suture System

Surgical Technique. Apache Anterior Lumbar Interbody Fusion

TM TM Surgical Technique

IBD PEEK C. Anterior Cervical Spacer System. Surgical Technique

Apache Cervical Interbody Fusion Device. Surgical Technique. Page of 13. LC-005 Rev F

2. Active systemic infection or infection localized to the site of the proposed implantation is contraindications to implantation.

ACP1 CERVICAL PLATE SPINAL SYSTEM SURGICAL TECHNIQUE GUIDE II.

DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY

Cervical Screw System

Surgical Technique. Apache Posterior Lumbar Interbody Fusion Apache Transforaminal Lumbar Interbody Fusion

C-THRU Anterior Spinal System

X-spine Systems, Inc. Axle Interspinous Fusion System

For the Attention of the Operating Surgeon: IMPORTANT INFORMATION ON THE MATRIXRIB FIXATION SYSTEM

HawkeyeTM Peek. surgical technique

D. The following factors are of extreme importance to the eventual success of the procedure.

product overview Implant heights range from 8mm-20mm in 2mm increments, with two lordocic angle options of 6 and 12.

X-spine Aranax Anterior Cervical Plating System Instructions for Use / Package Insert

Zimmer Anterior Buttress Plate System. Surgical Technique

Precautionary Statement ( )

For use by an Accredited Orthopaedic Surgeon only

Cervical Spacer System surgical technique

DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY

Pinit Plate Small Bone Fusion System Bone Plate & Screw System

X-spine Systems, Inc. Silex Sacroiliac Joint Fusion System

X-spine Systems, Inc. Calix PC Spinal Implant System

Lapidus Arthrodesis System Instructions for Use

Amendia Interbody Fusion Devices PACKAGE INSERT

Spartan S 3 Facet Screw System PACKAGE INSERT

EXACTECH SPINE. Operative Technique. Cervical Spacer System. Surgeon focused. Patient driven. TM

X-spine Systems, Inc. Irix-A Lumbar Integrated Fusion System

X-spine Systems, Inc. Certex Spinal Fixation System

X-spine Systems, Inc. Spider Cervical Plating System

Table of Contents.

Revision A Date:

VTI INTERLINK PEDICLE SCREW SYSTEM

Royal Oak IBFD System Surgical Technique Posterior Lumbar Interbody Fusion (PLIF)

Y o u r Id e a s En g i n e e r e d t o Li f e

X-spine Systems, Inc. Xpress Minimally Invasive Pedicle Screw System

Instructions for Use AccuFit Lateral Plate System Caution: Federal (USA) law restricts this device to sale by or on the order of a physician

X-spine Systems, Inc. Butrex Lumbar Buttress Plating System

TABLE OF CONTENTS. Vault C Anterior Cervical Discectomy 2 and Fusion (ACDF) System Overview. Implants 3. Instruments 5. Surgical Technique 10

Instructions for Use Reform POCT System Caution: Federal (USA) law restricts this device to sale by or on the order of a physician

DISTAL RADIUS. Instructions for Use

SURGICAL TECHNIQUE GUIDE BRECKENRIDGE. Intervertebral Body/VBR Fusion System. ACDF Anterior Cervical Discectomy and Fusion

Metallic Internal Fixation Devices The following languages are included in this packet:

It is intended that the implants be removed after successful fusion.

BENGAL System CONCORDE System CONCORDE Bullet System CONCORDE Inline System CONCORDE Curve System COUGAR System DEVEX System LEOPARD System

CROSS -FUSE P E E K V B R / I B F SYST E M

The Omega LIF System devices are made from Titanium alloy Ti6Al4V ELI, ASTM F136.

SURGICAL TECHNIQUE GUIDE TRESTLE. Anterior Cervical Plating System

3 Operative Technique U.S. EDITION

Instructions for use Prodisc -C Vivo Cervical disc prosthesis

SIDEKICK EZ FRAME EXTERNAL FIXATION SYSTEM The following languages are included in this packet:

Gallery Laminoplasty Spine System

PILLAR AL. Anterior Lumbar Interbody Fusion (ALIF) and Partial Vertebral Body Replacement (pvbr) PEEK Spacer System OPERATIVE TECHNIQUE

X-spine Systems, Inc. Fixcet Spinal Facet Screw System

M6-C Artificial Cervical Disc Surgical Instruments

ENDO-FUSE INTRA-OSSEOUS FUSION SYSTEM The following languages are included in this packet:

OPERATIVE TECHNIQUE. anterior cervical plating system

EVOLVE TRIAD BONE SCREWS

ACP. Anterior Cervical Plate System SURGICAL TECHNIQUE

Technique Guide Small Bone Fusion System

The A-Space SIBD Spinal System is a stand-alone device intended to be used with the four supplied bone screws if no supplemental fixation is used.

NEO PEDICLE SCREW SYSTEM V. 2.0

Instructions for use for Bone Plates, Bone Screws, Intramedullary Nails, Pins and Wires

Doc. No.: IU/8 for. Rev. No.: 07 ResTOR Prosthesis

Royal Oak Cervical Plate System

EXTERNAL FIXATION SYSTEMS The following languages are included in this packet:

Posterior Lumbar Interbody Fusion System

Asnis. Micro Cannulated screw system. Xpress operative technique

Customized Cranial and Craniofacial Implants. Instructions For Use. 7000revD_Instructions_For_Use. Page 1 of 8

TC-PLUS Primary Knee IMPORTANT MEDICAL INFORMATION SPECIAL NOTE. This Package Insert is for product distributed in the US only.

Salto Talaris Total Ankle Prosthesis

EVOLVE TRIAD SYSTEM

AVS Anchor -C Cervical Cage System Surgical Technique

Zimmer Facet Screw System Surgical Technique

T-PAL Spacer System. Transforaminal posterior atraumatic lumbar spacer system.

Transcription:

PATHWAY AVID SPINAL ELEMENTS PATHWAY AVID CAUTION: Federal law (U.S.A.) restricts this device to sale by or on the order of a licensed physician. NON STERILE PRODUCT MUST BE STERILIZED PRIOR TO USE SINGLE USE ONLY REQUIRES THE USE OF AN APPROVED SUPPLEMENTAL SPINAL FIXATION SYSTEM SINGLE USE DISPOSABLE TITANIUM OR MP35N IMPLANT INSERTION WIRE ATTACHED TO DEVICE MUST BE PROPERLY DISPOSED OF AFTER USE. DO NOT PLACE OR RETURN DISPOSABLE TITANIUM IMPLANT INSERTION WIRES IN THE STERILIZATION TRAY. Materials May Contain: Titanium (Ti-6AI-4V) Alloy or MP35N (Co-Cr-Mo-Ni). THE PATHWAY AVID HAS NOT BEEN TESTED FOR HEATING OR MIGRATION IN THE MR ENVIRONMENT Warning: The pins and linkages in this product contain titanium or MP35N (Co-Cr-Mo-Ni). Patients with known allergy or sensitivity to titanium, cobalt, chromium, nickel, or molybdenum should not have this device implanted as it may provoke an allergic reaction. Patients with suspected hypersensitivity to the materials listed above should undergo a skin test to assess hypersensitivity prior to a Pathway AVID placement procedure. The PATHWAY AVID Intervertebral Body Fusion Device is made from PEEK OPTIMA TM. The device consists of multiple cages that are attached by titanium (Ti-6AI-4V) or MP35N (Co-Cr-Mo-Ni) hinge pins. The hinge pins provide the means for the device to form into the final articulated shape. Also, the titanium hinge pins serve as a marker providing visual aid in determining the location of the implant both intra and postoperatively. There are serrated teeth located in the superior and inferior surfaces. The device operates by placing the unarticulated PATHWAY AVlD onto the inserter. The unarticulated device has a series of segments attached in a row and attached by the hinge pins. The device is then inserted into the prepared disc space and then articulated by contracting on the metal linkages between the PEEK cages to form the final U- shape between the two vertebrae. The grafting spaces in the cages are packed with autogenous bonegraft. The footprint of the articulated device can range from 8 mm to 14 mm in height, 20 mm in width, and 35 mm in length. The implants are provided non sterile. Instruments are provided clean and non-sterile. The devices and instruments are to be sterilized by the user facility. Intended Use The PATHWAY AVlD Intervertebral Body Fusion Device is intended for spinal fusion procedures at one or two contiguous levels (L2-S I) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). This device is intended to be used with autogenous bone graft.

Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the PATHWAY AVID device. The PATHYWAY AVID Intervertebral Body Fusion Device is to be inserted using a transforaminal approach only. Training The PATHWAY AVID is intended to be used by professionals that have been trained by an SPINAL ELEMENTS Representative. The Surgical Technique Guide should be reviewed prior to the use of the implant and instrumentation. Even experienced spine surgeons may require additional skills best acquired through a course of formal instruction with laboratory or bench training. Lack of experience or expertise with PATHWAY AVID may result in surgical difficulties, as explained below. Materials These devices are manufactured from Polyetheretherketone (PEEK) per ASTM F2026 and may contain Titanium Alloy (Ti-6AI-4V) per ASTM F-136 or MP35N (Co-Cr-Mo-Ni) per ASTM F562. CONTRAINDICATIONS Contraindications may be relative or absolute. The choice of a particular device must be carefully weighed against the patient's overall evaluation. Circumstances listed below may reduce the chances of a successful outcome: Any abnormality present which affects the normal process of bone remodeling including, but not limited to, severe osteoporosis involving the spine, active infection at the site or certain metabolic disorders affecting osteogenesis. Insufficient quality or quantity of bone which would inhibit rigid device fixation. Previous history of infection. Excessive local inflammation. Open wounds. Any neuromuscular deficit which places an unusually heavy load on the device during the healing period. Obesity. An overweight or obese patient can produce loads on the spinal system which can lead to failure of the fixation of the device or failure of the device itself. Patients having inadequate tissue coverage of the operative site. Pregnancy. A condition of senility, mental illness, or substance abuse. These conditions, among others, may cause the patient to ignore certain necessary limitations and precautions in the use of the implant, leading to failure or other complications. Foreign body sensitivity. Where material sensitivity is suspected, appropriate tests should be conducted prior to material selection or implantation. Other medical or surgical conditions which would preclude the potential benefit of spinal implant surgery, such as the presence of tumors, congenital abnormalities, elevation of sedimentation rate unexplained by other diseases, elevation of white blood cell count (WBC), or marked left shift in the WBC differential count. Any case not described in the indications for use. These contraindications can be relative or absolute and must be taken into account by the physician when making his or her decision. The above list is not exhaustive. WARNINGS The PATHWAY AVID is specifically intended to be implanted using a transforaminal ( TLIF ) approach. Other approaches may result in device malfunction, improper implantation, and/or injury to the patient. Implants and Instruments are provided non sterile and must be sterilized before use. Validated sterilization cycle parameter protocols are noted below in the Sterilization Section of this insert. The disposable titanium implant insertion wires used to articulate the PATHWAY AVID are intended to be a single use component. Following use, the insertion wires will be considered a biohazard and should be handled and treated appropriately. The insertion wires must be disposed of after use. DO NOT PLACE OR RETURN THE DISPOSABLE TITANIUM OR MP35N IMPLANT INSERTION WIRES IN THE STERILIZATION TRAY.

The information contained in this Package Insert is necessary but not sufficient for the use of this device. This information is in no sense intended as a substitute for the professional judgment, skill, and experience of the surgeon in careful patient selection, preoperative planning and device selection, knowledge of the anatomy and biomechanics of the spine, understanding of the materials and the mechanical characteristics of the implants used, training and skill in spinal surgery and the use of associated instruments for implantation, securing the patient's cooperation in following an appropriately defined post-operative management program, and conducting scheduled post-operative follow-up examinations. AVID can be provided with two materials linkage and hinge pin configurations: Titanium (TI-6AI-4V) alloy or MP35N (Co-Cr-Mo-Ni). Some patients may exhibit metal sensitivity to either or both of these materials. Appropriate metal allergy testing should be performed to determine if the patient reacts to either or both materials. If the patient shows allergy to either material, this product may not be used on the patient. INFORMATION FOR PATIENTS The surgeon should discuss all physical and psychological limitations inherent to the use of the device with the patient. This includes the rehabilitation regimen, physical therapy and wearing an appropriate orthosis as prescribed by the physician. Particular discussion should be directed to the issues of premature weight bearing, activity levels, and the necessity for periodic medical follow-up. The patient should be warned of the surgical risks and made aware of possible adverse effects. The patient should be warned that the device cannot and does not replicate the flexibility, strength, reliability or durability of normal healthy bone, that the implant can break or become damaged as a result of strenuous activity or trauma, and that the device may need to be replaced in the future. If the patient is involved in an occupation or activity which applies inordinate stress upon the implant (e.g., substantial walking, running, lifting, or muscle strain) resultant forces can cause failure of the device. Patients who smoke have been shown to have an increased incidence of non-unions. Such patients should be advised of this fact and warned of the potential consequences. INFECTION Transient bacteria can occur in daily life. Dental manipulation, endoscopic examination and other minor surgical procedures have been associated with transient bacteria. To help prevent infection at the implant site, it may be advisable to use antibiotic prophylaxis before and after such procedures. INSTRUMENTS Specialized instruments are provided by SPINAL ELEMENTS and must be used to assure accurate implantation of the device. While rare, intraoperative fracture or breakage of instruments can occur. Instruments that have experienced extensive use or extensive force are more susceptible to fracture depending on the operative precaution, number of procedures, and disposal attention. Instruments should always be examined for wear or damage prior to surgery. REUSE An implant should never be reused. While it may appear undamaged, a used implant may have acquired blemishes or latent compromise of its integrity which would reduce its service life. The disposable titanium implant insertion wires used to articulate the PATHWAY AVID are intended to be a single use component and should never be reused. It is recommended to verify that the instruments are in good condition and operating order prior to use during surgery. HANDLING Correct handling of the implant is extremely important. The operating surgeon should avoid notching or scratching the device. ALLERGY AND HYPERSENSITIVITY TO FOREIGN BODIES When hypersensitivity is suspected or proven, it is recommended that the tolerance of the skin to the material(s) that make up the implants be checked before they are implanted. IMPLANT SELECTION AND USE

The choice of proper shape, size, and design of the implant for each patient is crucial to the success of the surgery. The surgeon is responsible for this choice, which depends on each patient. Patients who are overweight may be responsible for additional stresses and strains on the device, which can speed up metal fatigue and/or lead to deformation or failure of the implants. The size and shape of a patient s bone structures determine the size, shape, and type of the implants. Once implanted, the implants are subjected to stresses and strains. These repeated stresses on the implants should be taken into consideration by the surgeon at the time of the choice of the implant, during implantation, as well as in the post-operative follow-up period. Indeed, the stresses and strains on the implants may cause metal fatigue or fracture or deformation of the implants, before the bone graft has become completely consolidated. This may result in further side effects or necessitate the early removal of the osteosynthesis device. Improper selection, placement, positioning and fixation of these devices may result in unusual stress conditions reducing the service life of the implant. Contouring or bending of rods or plates only is recommended if necessary according to the surgical technique of each system. Rods or plates should only be contoured with the proper contouring instruments. Incorrectly contoured rods/plates, or rods/plates which have been repeatedly or excessively contoured must not be implanted. The surgeon is to be thoroughly familiar with the surgical procedure, instruments and implant characteristics prior to performing surgery. Refer to the device s surgical protocols for additional procedural information. Periodic follow-up is recommended to monitor the position and state of the implants, as well as the condition of the adjoining bone. METAL COMPONENTS Questions have been raised in the scientific literature as to whether or not these alloys themselves may be carcinogenic in implant recipients. Studies conducted to evaluate this issue have not identified convincing evidence of such phenomena. SYSTEM COMPATIBILITY While some degree of corrosion occurs on all implanted metal and alloys, contact of dissimilar metals may accelerate this corrosion process. The presence of corrosion may accelerate fatigue fracture of implants, and the amount of metal compounds released into the body system will also increase. Internal fixation devices, such as rods, hooks, screws, wires, etc., which come into contact with other metal objects, must be made from like or compatible metals. This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). Care should be taken to ensure material compatibility. POSTOPERATIVE CARE Prior to adequate maturation of the fusion mass, implanted spinal instrumentation may need additional help to accommodate full load bearing. External support may be recommended by the physician from two to four months postoperatively or until x-rays or other procedures confirm adequate maturation of the fusion mass. External immobilization by bracing or casting may be employed. The patient should be instructed regarding appropriate and restricted activities during consolidation and maturation for the fusion mass in order to prevent placing excessive stress on the implants which may lead to fixation or implant failure and accompanying clinical problems. The patient should also be instructed to report any unusual changes of the operative site to his/her physician. The physician should closely monitor the patient if a change at the site has been detected. POTENTIAL ADVERSE EFFECTS Bending, disassembly or fracture of any or all implant components. Fatigue fracture of spinal fixation devices, including screws and rods, has occurred. Loosening of spinal fixation implants can occur resulting in loss of fixation and/or implant migration. Early mechanical loosening may result from inadequate initial fixation, latent infection, premature loading of the prosthesis or trauma. Late loosening may result from trauma, infection, biological complications or mechanical problems, with the substantial possibility of bone erosion, migration and/or pain. Reversion of the implant to its unarticulated shape. Intraoperative fissure, fracture, or perforation of the spine can occur due to implantation of the components.

Fracture of the vertebra during placement of the implant, or post-operatively. Postoperative fracture of the bone graft, the intervertebral body, pedicle, and/or sacrum above and/or below the level of surgery can occur due to trauma, the presence of defects, or poor bone stock. Dural leak requiring surgical repair. Inappropriate or improper surgical placement of this device may cause distraction or stress shielding of the graft or fusion mass. This may contribute to failure of an adequate fusion mass to form. Delayed Union or Nonunion: Internal fixation appliances are load sharing devices which are used to obtain alignment until normal healing occurs. In the event that healing is delayed, does not occur, or failure to immobilize the delayed/nonunion results, the implant will be subject to excessive and repeated stresses which can eventually cause loosening, bending or fatigue fracture. The degree or success of union, loads produced by weight bearing, and activity levels will, among other conditions, dictate the longevity of the implant. If a nonunion develops or if the implants loosen, bend or break, the device(s) should be revised or removed immediately before serious injury occurs. Pain, discomfort, or abnormal sensations due to the presence of the device. Pressure on skin from components where inadequate tissue coverage exists over the implant, with the potential extrusion through the skin. Loss of proper spinal curvature, correction, height and/or reduction. Peripheral neuropathies, nerve damage, heterotopic bone formation and neurovascular compromise, including paralysis, loss of bowel or bladder function, or foot-drop may occur. Serious complications may be associated with any spinal surgery. These complications include, but are not limited to: genitourinary disorders; gastrointestinal disorders; vascular disorders; including thrombus; bronchopulmonary disorders, including emboli; bursitis, hemorrhage, myocardial infarction, infection, paralysis or death. Neurological, vascular, or soft tissue damage. Decrease in bone density due to stress shielding. Failure to improve symptoms. While the expected life of spinal implant components is difficult to estimate, it is finite. These components are made of foreign materials which are placed within the body for the potential fusion of the spine and reduction of pain. However, due to the many biological, mechanical and physicochemical factors which affect these devices but cannot be evaluated in vivo, the components cannot be expected to indefinitely withstand the activity level and loads of normal healthy bone. Adverse effects may necessitate reoperation or revision. IMPLANT These devices are permanently implanted between the vertebrae. These devices are designed to stabilize the operative site during the normal healing process. If successful fusion occurs, the devices no longer serve any functional purpose; however, they remain part of the fusion mass. PACKAGING The implants are delivered in packages; these must be intact at the time of receipt. The systems are sometimes supplied as a complete set: implants and instruments are arranged on trays and placed in specially designed storage boxes. CLEANING, DECONTAMINATION AND STERILIZATION The PATHWAY AVID implant must be sterilized per the recommended instructions provided below. PATHWAY AVID surgical instruments that come into contact with blood and tissue must be cleaned between surgical cases per the recommended procedures below. Instruments must also be decontaminated per the instructions below. Following cleaning and decontamination, the surgical instruments must be sterilized per the instructions provided below. Warning: The disposable titanium or MP35N implant insertion wires used to articulate the PATHWAY AVID are intended to be a single use component. The insertion wires must be disposed of after use. DO NOT PLACE OR RETURN THE DISPOSABLE TITANIUM or MP35N IMPLANT INSERTION WIRES IN THE STERILIZATION TRAY.

RECOMMENDED CLEANING PROCEDURES FOR THE PATHWAY AVID INSTRUMENTATION PROVIDED NON STERILE Cleaning should be performed manually, ultrasonically, and in accordance with the specifications designated by the manufacturer of the hospital's equipment. This includes the use of neutral cleaners followed by deionized water rinses and a final sterile water rinse. Corrosive products or instruments should be avoided (abrasive sponges and metal brushes). Cleaning before sterilization: machine cleaning using a wide spectrum bactericide and fungicide. Cleaning products prohibited: Strong mineral acids (sulfuric, nitric, hydrochloric, etc... ) or strong Lewis acids such as Zinc chloride or boron trifluoride and strong oxidizing agents such as aqueous solutions containing sodium hypochlorite and NaOH (soda) or a high concentration of hypochlorite or permanganate ions. DO NOT USE SODA TO CLEAN INSTRUMENTS MADE OF ALUMINUM. DO NOT USE SODIUM CHLORITE TO CLEAN INSTRUMENTS MADE OF STAINLESS STEEL. Prolonged exposure to high temperatures and to most aggressive solvents (such as ethylene dichloride, phenolic and aniline solutions) must be avoided. It is recommended to use aqueous solutions that bave a ph above 4.0. Before using a cleaning product it is recommended to perform oxidation test. RECOMMEDED DECONTAMINATION PROCEDURE FOR THE PATHWAY AVID INSTRUMENTATION PROVIDED NON STERILE The instruments should be carefully cleaned before any decontamination. Cleaning should remove all dried infectious residual matter. The standard recommendations include the following methods of decontamination: o For devices made of Stainless Steel or Titanium: soak in IM soda for 1 hour. Note: this procedure is not appropriate for devices made of Aluminum as it will generate dangerous chemical reaction. o For devices made of Aluminum or Titanium: soak in sodium chlorite, at 2% active chlorine, for I hour. Note: this procedure is not appropriate for devices made of Stainless Steel as it will generate corrosion. o Or, autoclave treatment at 134*C (273*F) for 18 minutes in a porous load autoclave. For patients presenting a risk factor of transmittable subacute spongiform encephalopathy (previous histories of growth hormone treatment, genetic history, previous history of high-risk neurosurgical interventions), the World Health Organization recommends the following methods of decontamination: o For devices made of Stainless Steel or Titanium: autoclave treatment with soda, at a concentration 1M, at 121 *C (250*F) for 30 minutes in a gravity displacement autoclave, or soaking in 1M concentrate soda. Note: this procedure is not recommended for devices made of Aluminum as it will generate dangerous chemical reaction. o For devices made of Aluminum or Titanium: sodium chlorite at 2% active chlorine, for 1 hour, followed by autoclave treatment at 121*C (250*F) for 1 hour in a gravity displacement autoclave or by autoclave treatment at 134*C (273*F) for 1 hour in a porous load autoclave. Note: this procedure is not appropriate for devices made of Stainless Steel as it will generate corrosion. N.B.: To sterilize the material, it must then undergo the required sterilization process. RECOMMENDED STERILIZATION PROCEDURES FOR THE PATHWAY AVID IMPLANT PROVIDED NON STERILE AND FOR THE PATHWAY AVID INSTRUMENTATION FOLLOWING CLEANING AND DECONTAMINATION This implant and instrumentation are provided non sterile. The implant must be sterilized before use. All instruments must be cleaned and sterilized before use. Remove all implants and instruments from their packaging prior to sterilization. Sterilize the implants and instrumentation by steam autoclaving procedure regularly used in the hospital. The suggested sterilization method is:

Method Temperature Minimum Exposure Time Minimum Dry Time (Dry Cycle In Auto Clave) Wrapped Gravity 132 C (270 F) 15 Minutes 45 Minutes Pre-Vacuum 132 C (270 F) 4 Minutes 45 Minutes Other alternative sterilization methods and cycles are possible but must be validated. Autoclave must be validated to guarantee the recommended sterilization temperature is reached for the full duration of the exposure time. EtO sterilization, cold sterilization, irradiation, plasma techniques are NOT recommended for the inactivation of prions. If paper-filter sterilization boxes are used, it is advisable to verify whether the filters are intact before sterilization. If after sterilization water remains in the sterilization boxes, sterilization cannot be considered sufficient. NOTE: These parameters are validated to sterilize only the PATHWAY AVID implants and instruments. If other products are added to the sterilizer, the recommended parameters are not valid and new cycle parameters must be established by the user. The autoclave must be properly installed, maintained and calibrated. Ongoing test must be performed to confirm inactivation of all forms of viable microorganisms. FURTHER INFORMATION A Surgical Technique Guide is available on request through your SPINAL ELEMENTS agent or directly from SPINAL ELEMENTS. Users who have a Surgical Technique Guide that is over two years old at the time of surgery are advised to ask for an updated version. COMPLAINTS Any professional (e.g. Customer or user) who has experienced dissatisfaction in the services or who has any complaints about SPINAL ELEMENTS products referring to quality, identity, durability, reliability, safety, effectiveness, and/or performance, should report this to the representative or distributor. If the malfunction or deterioration of the device, or any inadequacy in instructions for use has led, or might lead, to the death or serious injury of a patient or user, please notify the representative or distributor immediately by telephone of fax. When filing a complaint form, please provide maximum information such as the identification of the product (name, reference, lot number), the nature of the complaint or the description of the incident, the consequences, and all technical elements that could be helpful in the future investigation, such as: the device, x-rays, etc... should be addressed. For further information or complaint, please contact: FEDERAL LAW (USA) RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN SINGLE USE DEVICE, DO NOT REUSE CONSULT ACCOMPANYING PACKAGE INSERT FOR LABELING INDICATIONS SPINAL ELEMENTS 1755 West Oak Parkway Marietta, GA 30062 USA TEL: 877-755-3329 FAX: 877-420-1213 DEVICES ARE SUPPLIED NON-STERILE CONSULT INSTRUCTIONS FOR USE PEEK OPTIMA is a registered trademark of Invibio