Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Similar documents
D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Diode calibration for dose determination in total body irradiation

INTRODUCTION. Material and Methods

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Small field diode dosimetry

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

Experimental Evaluation of Depth Dose by Exit Surface Diode Dosimeters for Off-Axis Wedged Fields in Radiation Therapy

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Protection of the contralateral breast during radiation therapy for breast cancer

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MP (CLINICAL ONCOLOGY) PART I EXAMINATION - AUGUST Time : p.m p.m.

Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution

Application(s) of Alanine

A new geometric and mechanical verification device for medical LINACs

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Limits of Precision and Accuracy of Radiation Delivery Systems

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Quality assurance in external radiotherapy

Neutron Measurements for Intensity Modulated Radiation Therapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Quality Assurance of Helical Tomotherapy Machines

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Characteristics Analysis of High Energy External Radiotherapy Beams in Water

Effectiveness of compensating filters in the presence of tissue inhomogeneities

Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

Dosimetry in digital mammography

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

Radiation Doses to Contralateral Breast During Irradiation of Breast Cancer

EPID dosimetry confi guration and pre-treatment IMRT verifi cation

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague

Absolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

IORT with mobile linacs: the Italian experience

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Detector comparison for small field output factor measurements with flattening filter free photon beams

Measuring radiation dose through the detection of radiation-induced acoustic waves

Patient dosimetry for total body irradiation using single-use MOSFET detectors

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Radiochromic film dosimetry in water phantoms

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

The determination of timer error and its role in the administration of specified doses

International Association of Scientific Innovation and Research (IASIR) (An Association Unifying the Sciences, Engineering, and Applied Research)

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS

Shielding Design Basis and its Calculation of High Energy Medical Linac Installed in Bangladesh Atomic Energy Commission, Bangladesh

Quality assurance program for prototype stereotactic system developed for Neptun 10 PC linac

Imaging of Radiation Dose Using Cherenkov Light

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

A translational couch technique for total body irradiation

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Helical Tomotherapy Experience. TomoTherapy Whole Brain Head & Neck Prostate Lung Summary. HI-ART TomoTherapy System. HI-ART TomoTherapy System

Portal dosimetry in wedged beams

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

MD 4300: Introduction to Medical Dosimetry (3semester hour) Course Syllabus for Fall 2009

Transcription:

Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri 2, H. Gholamhosseinian 3 1 Medical Physics Department, Mashhad University of Medical Sciences, Mashhad, Iran 2 Omid Oncology Hospital, Mashhad, Iran 3 Linac Department, Imam Reza Hospital, Mashhad, Iran ABSTRACT Background: In radiotherapy, wedge filters are used for optimizing the tumor dose distribution in patients. The attenuation in beam intensity due to the presence of wedge filter is compensated by means of a wedge factor measured at the central axis of the beam. The field size, depth and SSD dependence of wedge factor have been assessed for 9MV radiations of Neptun PC linear accelerator. Materials and Methods: Wedge factors (WF) at different SSD, field size (FS) and depth (d) in water were measured for 8 steel wedges with different sizes and angles of 15, 30, 45, and 60 degree. Experimental data were obtained using Neptun 10PC, Linac 9MV X-ray machine, a 3D water phantom, dosimeters and an electrometer. To study the effect of field size on WF, the wedge factor was measured for square field sizes from 5 5 to 20 20 cm, with 1 cm increment intervals for all wedges; and, at the depth of 10 cm, SSD of 100 cm with monitor unit (MU) of 80. Effects of depth on WF were studied by measurement in various depths from 3-19cm for all wedge angles at SSD of 100cm, field size of 10 10cm and 80 MU irradiation. Effects of SSD on WF were investigated by a variation of SSD from 90-110cm with 5cm increment intervals; while the dosimeter was set at depth of 10cm and field size of 10 10cm were irradiated for 80MU. Results: Linear dependence of WF with field size and depth of measurements were confirmed with 95% certainty. Shapiro-Wilk test, showed that the residual data of the regression tests have normal distributions (P>0.05). There was also found no linear relationship between WF and SSD (P>0.05). Conclusion: WF has linear dependence with field size and depth of measurements, but the rate of variations are less than 2.2% per 10cm variation in field size and less than 1.3% per 10 cm variation in depth of measurements, therefore, correction of WF for field size and depth of treatments in clinical trials is negligible. Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Keywords: Wedge factor, wedge filter, linear accelerator. INTRODUCTION I n radiotherapy, using wedge filter is a common technique for modifying iso-dose curves and it optimizes the tumor dose distribution in patients. There are generally two different methods for applying the filters: external mounting on the treatment head of machine (typically 15, 30, 45 * Corresponding author: Dr. M. Hajizadeh Saffar, Medical Physics Dept., Mashhad University of Medical Sciences, Mashhad, Iran Fax: +98 511 8517505 E-mail: hajizadeh@mums.ac.ir and 60 degree), and internal filter (65 degree), which is electronically moved in and out of the beam, and different wedge angles are obtained by weighted superposition of wedged and open beam. As the wedge filter reduces the beam intensity and can result in large errors in delivering dose; the precise wedge transmission factor or wedge factor (WF) needs to be determined and accounted for the computation of time setting or monitor unit to compensate the attenuation by extending the radiation time. It is important, therefore, to assess and generalize any variation to wedge factor resulting from field size, depth,

M. Hajizadeh Saffar et al. and SSD to be used in delivering the exact prescribed dose to patient. Palta et al. (1988) and Lee et al. (1989) have studied field dependence of wedge factors and concluded that the dependence of the WF on field size is attributed to the changes in phantom scatter, as well as to the change in collimator scattered photon that reaches the point of measurement. The depth dependence of the WF have been studied since many years ago, by Sewchand et al. (1978), and Wu et al. (1984) and was attributed to the hardening of the incident beam passing through the wedge filter which absorbs and/or scatter the low energy photons. Kalend et al. (1990) have shown that depth dependence of the WF caused by the dose gradient due to increase phantom scattering is as significant as the beam hardening, especially for low energy X-ray produced by Linac. The purpose of this study was to assess the dependence of WF on the most commonly used treatment depth, field size and SSD, to inform the physicist from the magnitude of the various effects of wedge filters used in their clinical treatment planning; and to provide a simple algorithm to predict the precise amount of corrected wedge factor needed for their clinical routines. MATERIALS AND METHODS Wedge factor at a depth (d) in water for a field size (FS), and at a certain SSD was defined as the ratio of the dose with the wedge beam, D w (FS, d), to the dose with open beam, D o (FS, d). In this study, experimental data were obtained using Neptun 10PC Linac 9MV X-ray machine, a 3D water phantom (Scanditronix 50 50 50 cm), two RK (0.12cc) ionizing chamber as a reference and field dosimeters, electrometer (Scanditronix model DPD510), and 8 steel wedges with different sizes and angles which are numerated based on their maximum field sizes as shown in table 1. To set the system, the phantom is placed horizontally in position at SSD of 100 cm and adjusted so that the phantom index overlapped with the light field index of the machine; then Wedge No. Table 1. Physical characteristics of the external wedges used for Neptun 10PC Linac. Wedge Angle (degree) Max. thickness (cm) Max. width (cm) Max. length (cm) Max. field size at isocenter (cm cm) 1 15 1.2 15.2 16 20 20 2 30 1.8 11.4 16 15 20 3 30 2.4 15.2 16 20 20 4 45 2.0 7.6 16 10 20 5 45 3.0 11.4 16 15 20 6 45 3.9 15.2 16 20 20 7 60 3.0 7.6 16 10 20 8 60 4.5 11.4 16 15 20 the field dosimeter is put at the centre of the field (overlapping light field index with white spot on the dosimeter) and reference dosimeter at the corner of light field, as is shown in figure 1. Accuracy of the obtained data were controlled by measuring the depth dose profiles, inline and cross line, prior to the measurements, with RFA- Plus software and performing all quality controls which might have affected to the measurements directly or indirectly. SSD=100 cm Reference dosime- w Water phantom To study the effect of field size on WF, the wedge factor were measured for square field sizes from 5 5 to 20 20 cm with 1 cm increment intervals at wedge angles of 15, 30, 45, Field dosimet Linac outl Wedge filter Figure 1. Measurement configuration diagram of the system for controlling the accuracy and measuring WFs. 54 Iran. J. Radiat. Res.; Vol. 2, No. 2, September 2004

Wedge factor dependence with depth, field size and SSD and 60 degree, and at the depth of 10 cm, SSD of 100 cm with monitor unit of 80. In this study, the entire field sizes were squared, but the rectangular field size allowable for each wedge could be used if the side of field sizes were derived from the equivalent Square as discussed by Popescu et al. (1999). Effects of depth on WF were studied by measurement in depths of 3, 5, 7, 10, 13, 16 and 19 cm, with wedges 15 o (No.1), 30 o (No.2), 45 o (No.4) and 60 o (No.7), SSD of 100cm, field size of 10 10cm, and 80 MU irradiation. Effects of SSD on WF were investigated by a variation of that from 90-110cm with 5cm increment interval, while the dosimeter was set at depth of 10cm and field size of 10 10cm, irradiated for 80MU. RESULTS AND DISCUSSION Measurements of the wedge factors were performed by studying the output charges of the field detector, alternatively for wedged and open beams, and for calculating the ratio of the wedged to open beam measurements. The results of WF, using various field size, depth and SSD are shown in tables 2, 3 and 4 respectively. Table 2. Wedge factor variations with field size (side of square field / cm) for Neptun 10PC Linac 9MV radiation. Side of 15º 30º 45º 60º F.S. No.1 No.2 No.3 No.4 No.5 No.6 No.7 No.8 cm 20 20 15 20 20 20 10 20 15 20 20 20 10 20 15 20 5 0.747 0.655 6 0.838 0.775 0.749 0.655 7 0.748 0.655 8 0.84 0.775 0.750 0.656 9 0.750 0.657 10 0.84 0.776 0.753 0.659 11 0.664 0.537 12 0.841 0.776 0.539 13 0.665 0.541 14 0.842 0.779 0.666 0.541 15 0.668 0.543 16 0.842 0.717 0.591 17 0.719 0.592 18 0.844 0.719 0.594 19 0.722 0.595 20 0.846 0.723 0.598 Table 3. Wedge factor variations with depth for Neptun 10PC Linac 9MV radiation. depth cm 15º 30º 45º 60º No.1 No.2 No.4 No.7 3 0.830 0.772 0.749 0.656 5 0.834 0.772 0.752 0.659 7 0.834 0.769 0.753 0.659 10 0.837 0.772 0.755 0.662 13 0.837 0.774 0.755 0.665 16 0.838 0.776 0.759 0.669 19 0.839 0.779 0.759 0.670 Table 4. Wedge factor variations with SSD for Neptun 10PC Linac 9MV radiation. SSD cm 15º 30º 45º 60º No.1 No.2 No.4 No.7 90 0.833 0.768 0.757 0.664 95 0.834 0.768 0.757 0.663 100 0.831 0.766 0.757 0.662 105 0.832 0.765 0.754 0.663 110 0.830 0.767 0.756 0.662 Iran. J. Radiat. Res.; Vol. 2, No. 2, September 2004 55

M. Hajizadeh Saffar et al. In the literature, wide variations of WF with FS has been reported, from no clear dependence reported by Dean (1991) and Cozzi et al. (1996) to more than 10% variation by Palta et al. (1988), Thomas (1990) and Podgorsak et al. (1993). In this study, various FS from 5 5 to 20 20 cm for angles of 30º, 45º and 60º, and 2 wedges with different maximum width were used respectively. As the central thickness of each wedge t/2, with angle α, is dependent on its maximum width (w), by t=w.tan α, therefore, measured WFs for a given angle and FS, by two wedges with maximum width w 2 and w 1, caused the wider wedge to attenuate the wedged beam more by a factor of e -µ t/2 ; where µ is the attenuation coefficient of the wedge and t is the difference between their central thicknesses. Hence, as shown in table 2, for given angles of 30º, 45º or 60º the wider the wedge, the lower the measured WF. Total variation of WF in the situation used in this study, varied from 0.008 to 0.006 (0.07% to 0.22% per cm cm variation in field size) from thinnest to thickest filter; therefore, any correction of WF in clinical trials (for maximum 10 cm variations in field size) seems to be negligible (maximum 2.2%) (Niroomand-Rad et al.1992, Heukelom et al.1994). Wide variation of WF with depth (from 2% to 10%) has been reported in the literature; however in this study WF dependence per 10cm variation in depth varied from 0.68% for 15º (no.1) thin filters to 1.3% for 60º (no.7) thick filters. The existence of linear relation between WF and different variable factors, FS, depth and SSD, were investigated by statistical regression test; and also to ensure of the normalization of the residual data distributions, Shapiro-Wilk test was used. To investigate the FS dependence of WF, first, the measured WFs were corrected for the maximum width of the wider available wedges, and then the statistical regression test, and also Shapiro-Wilk test were applied. The results and the regression lines are shown in figure 2. P-values of the regression test for all wedge filters were less than 0.05; therefore, linear dependence of WF with field size was confirmed with 95% certainty. Also, the p-values of the Shapiro-Wilk test were more than 0.05, which meant that the residual data of the regression tests had normal distributions. Equation of the regression lines and regression coefficients are shown in figure 2. Linear dependence of WF with depth and SSD for 15º (No.1), 30º (No.2), 45º (No.4) and 60º (No.7) wedge filters, shown in tables 3 and 4, were also investigated with regression and Shapiro-Wilk tests. The linear dependence with 95% of certainty, (p-values less than 0.05) was confirmed only for depth. P values for SSD variations for all wedge filters were more than 0.05; therefore, there is no linear relation between WF and SSD. In fact, an inverse linear dependence between WF and SSD has been found by the authors for 60 Co gamma radiations, which is attributed to attenuation of collimator scattered photons with increasing SSD. That was due to the fact that for 60 Co radiation, as reported by Kalend et al. (1990), there was no beam hardening effects due to presence of wedge filter, while it existed for linac x-rays and could therefore diminish the Figure 2. Variations of wedge factor with field size for 9MV radiation from Neptun 10PC Linac. 56 Iran. J. Radiat. Res.; Vol. 2, No. 2, September 2004

Wedge factor dependence with depth, field size and SSD dependence of WF to SSD. The lack of WF dependence with SSD was confirmed by Popescu et al. (1999) who studied different wedge (external and internal) filters with photon energies 4-24 MV. The linear dependence of WF with field size and depth and also lack of dependence with SSD were in agreement with the results reported by Popescu et al. (1999) and Niroomand-Rad et al. (1992). The regression line of WF with field size, FS, for wedge No.1 (15º) at the depth of 10 cm, field size of 10 10 cm, and SSD=100cm, can be expressed as: WF c = WF (10 10, 15º) [1 + a (FS-10)] Where WF c is the corrected WF for any field size, FS, and a=0.0005 is the slope of regression line shown in figure 2. In a similar way the WF at different depth can be calculated by the following formula: WF c = WF (10 10, 15º) [1 + b (d-10)] Where WF c is the corrected WF for any depth, d, and b=0.0005 is the slope of regression line shown in figure 3. To merge these equations and modify that for other filters, the difference between central thicknesses of the filters must be taken into account. This is performed by the following formula: WF c = WF (10 10, 15º) e -µ(t-1.2)/2 [1 + a (FS-10)] [1 + b (d-10)] Where, t, is the maximum thicknesses (from table 1) and µ (with the average of 0.27 cm -1 ) is the linear attenuation coefficient of the filters. The value of a, and b for different wedge angles are derived from figures 2 and 3, and shown in table 5. All the measured data shown in tables 2, 3 and 4 were obtained using this formula with maximum error of less than 1%. Therefore, this formula can be used by physicist to predict the precise amount of corrected wedge factor of any filter needed for their clinical routine treatments. Table 5. Slope of regression lines of wedge factor with field size and depth for different filters Used with Neptun 10PC linac 9MV radiation. Wedge angle a b 15º (No.1) 0.0005 0.0005 30 ס (No.2) 0.0007 0.0005 45 ס (No.4) 0.001 0.0006 60º (No.7) 0.001 0.0009 Figure 3. Variations of wedge factor with Iran. J. Radiat. Res.; Vol. 2, No. 2, September 2004 57

M. Hajizadeh Saffar et al. CONCLUSION Although the WFs have linear dependence with field size and depth of measurements, the rate of WF variations at the situation used in this study, are less than 1.72% per 10cm variation in field size, and less than 0.87% per 10 cm variation in depth of measurements. So any correction of WF for clinical trials is negligible. This is in agreement with the ICRU report 24 (1976) which neither specifies measurements of no depth or field size. ACKNOWLEDGEMENT We would like to thank Mr. Sh. Monadi for his helpful assistance and discussions and Mr. M. Sarmad for statistical analysis, and also all of linac department's staffs at Imam Reza hospital for their kindly cooperations. REFERENCES Cozzi F.A., Cozzi L., Garavaglia G. (1996). Wedge factors: dependence on depth and field size. Radiother. Oncol., 39: 31-34. Dean E.M. and Davis J.B. (1991).The variation of wedge factors with field size on a linear accelerator with wedge tray beneath secondary collimator. Br. J. Radiol., 64: 184-185. Heukelom S., Lanson J.H., Mijnheer B.J. (1994). Wedge factor constituents of high-energy photon beams: Head and phantom scatter dose components. Radiother. Oncol., 32: 73-83. ICRU Report (1976). Determination of absorbed dose in a patient irradiated by beams of X or gamma-ray radiotherapy procedures". ICRU Report No. 24. (ICRU Washington). Kalend A. M., Wu A., Yoder V., Maitz A. (1990). Separation of dose gradient effect from beam-hardening effect on wedge factors in photon fields. Med. Phys., 17: 701-704. Lee Y., Guru Prasad S., Parthasaradhi K., Satyanarayana H., Pepela M.,,Garces R. M., (1989). Study of wedge transmission factors. Med. Phys., 16: 481. Niroomand-Rad A., Haleem M., Rodgers J., obcemea C. (1992). Wedge factor dependence on depth and field size for various beam energies using symmetric and half-collimated asymmetric jaw settings. Med. Phys., 19: 1445-1450. Palta J.R., Daftari I., Suntharalingam N. (1988). Field size dependence of wedge factors. Med. Phys., 15: 624-626. Podgorsak M.B., Kubsand S.S., Paliwal B.R. (1993). Dosimetry of large wedged highenergy photon beams. Med. Phys., 20: 369-373. Popescu A., Lai K., Singer K., Phillips M. (1999). wedge factor dependence with depth, field size, and nominal distance-a general computational rule. Med. Phys. 26: 541-549. Sewchand W., Khan F.M., Williamson J. (1978). Variation in depth dose data between open and wedged fields for 4-MV X-rays. Radiology, 127: 789. Thomas J. (1990). The variation of wedge factors with field size on a linear accelerator. Br. J. Radiol., 63: 355-356. Wu A., Zwicher R.D., Krasin F., Sternick E.S (1984). Dosimetry characteristics of large wedges for 4- and 6- MV X-rays. Med. Phys., 11: 186-188. 58 Iran. J. Radiat. Res.; Vol. 2, No. 2, September 2004