LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

Similar documents
LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

LCP Metaphyseal Plates. For extra-articular fractures.

LCP DISTAL TIBIA PLATE

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

The Calcaneal Plate. The Synthes non-locking solution for the Calcaneus.

Distal Radius Plate 2.4/2.7 dorsal and volar

3.5 mm LCP Olecranon Plates

Surgical Technique. This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

VA-LCP Olecranon Plates 2.7/3.5. The fracture-specific low-profile fixation system with variable angle locking technology.

3.5 mm LCP Extra-articular Distal Humerus Plate

Low Bend Distal Tibia Plates

3.5 mm LCP Distal Humerus Plates

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

LCP Periarticular Proximal Humerus Plate 3.5. The anatomic fixation system with anterolateral shaft placement.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

VA LOCKING CALCANEAL PLATES 2.7

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Superior Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Button Plate. Reinforcement for transosseous fixations.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

3.5 mm LCP Clavicle Hook Plates

LCP Ulna Osteotomy System 2.7. Low profile angular stable fixation for ulna shortening osteotomies.

LCP Wrist Fusion Set. Anatomic plates for total wrist fusion.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

3.5 mm LCP Hook Plate

VA-LCP Distal Humerus Plates 2.7/3.5. The low-profile fixation system with variable angle locking technology.

ANGLED BLADE PLATES FOR ADULTS

Wrist Fusion Instrument and Implant Set.

Variable Angle LCP Volar Rim Distal Radius Plate 2.4. For fragment-specific fracture fixation with variable angle locking technology.

VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments

Femoral Neck System. Surgical Technique

2.7 mm/3.5 mm LCP Distal Fibula Plate

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

Small Fragment Locking Compression Plate (LCP ) System

3.0/3.5/4.0/4.5/6.5/7.0/7.3. Cannulated Screws. Surgical Technique

Technique Guide. PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

Part of the DePuy Synthes Locking Compression Plate (LCP ) System. 3.5 mm LCP Medial Proximal Tibia Plates

2.4 mm Variable Angle LCP Volar Rim Distal Radius Plates

LCP Proximal Tibial Plate 4.5/5.0 with Periarticular Aiming Arm Instruments

DOUBLE/TRIPLE PELVIC OSTEOTOMY PLATES For Treating Coxofemoral Joint Instability and Subluxation in Immature Dogs

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

The Locking Calcaneal Plate Instrument and Implant Sets

VA LCP MEDIAL COLUMN FUSION PLATES 3.5

3.5 mm Clavicle Hook Plates

3.5 mm LCP Superior Anterior Clavicle Plates

LCP Percutaneous Aiming System 3.5 for PHILOS. For less invasive surgery at the proximal humerus.

LCP Extra-articular Distal Humerus Plate.

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

2.4 mm Variable Angle LCP Dorsal Distal Radius Plate

VA-LCP Ankle Trauma System 2.7/3.5. Our most comprehensive ankle plating system.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

LCP Locking Compression Plate. Surgical Technique

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

Periarticular Aiming Arm Instruments for LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

Sterile Distal Radius Kit

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Technique Guide. LCP Posterior Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment LCP system.

3.5 mm LCP Anterolateral Distal Tibia Plates

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

4.5 mm LCP Medial Proximal Tibia Plates

VA-LCP Ankle Trauma System 2.7/3.5. Our most comprehensive ankle plating system.

Orthodontic Bone Anchor (OBA) System

Large Fragment LCP Instrument and Implant Set

VA-LCP Distal Humerus Plates 2.7/3.5. The low-profile fixation system with variable angle locking technology.

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

3.5 mm LCP Distal Tibia T-Plates

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system

3.5 mm Locking Attachment Plate

LCP Wrist Fusion Set. Anatomic plates for total wrist fusion.

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. VA-LCP Distal Humerus Plates 2.7/3.5. The low-profile fixation system with variable angle locking technology.

VECTRA-T SURGICAL TECHNIQUE. The Translational Anterior Cervical Palate System. This publication is not intended for distribution in the USA.

LCP Distal Humerus Plates

Technique Guide. Small Fragment Locking Compression Plate (LCP) System. Stainless steel and titanium.

LOW PROFILE NEURO. This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

Distal Radius Plate Instrument and Implant Set. Discontinued December 2017 DSUS/TRM/0916/1063(1)

Transcription:

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

This description alone does not provide sufficient background for direct use of the instrument set. Instruction by a surgeon experienced in handling these instruments is highly recommended. Processing, Reprocessing, Care and Maintenance For general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance For general information about reprocessing, care and maintenance of Synthes reusable devices, instrument trays and cases, as well as processing of Synthes non-sterile implants, please consult the Important Information leaflet (SE_023827) or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance

Table of Contents Introduction LCP Proximal Radius Plates 2.4 2 AO Principles 3 Intended Use and Indications 4 Surgical Technique Preparation 5 Reduction 7 Proximal Screw Insertion 8 Shaft Screw Insertion 9 General Notes on Technique 10 Postoperative Treatment and Implant Removal 13 Product Information Plates 14 Screws 15 Instruments 16 MRI Information 17 LCP Proximal Radius Plates 2.4 Surgical Technique DePuy Synthes 1

LCP Proximal Radius Plates 2.4 Plate features Nine LCP Proximal Radius Plates available to address various fracture patterns of the proximal radius Plates are precontoured for anatomical fit Combi holes allow fixation with locking screws in the threaded section for angular stability, and cortex screws in the Dynamic Compression Unit (DCU) section for distraction. A fixed-angle construct provides advantages in osteopenic bone or multifragment fractures, where traditional screw purchase is compromised. Carefully apply for osteoporotic bone Additional features Limited-contact design shaft with 2, 3, and 4 combi-holes The holes in the head of the plate accept 2.4 mm locking screws The shaft holes accept 2.4 mm locking screws in the threaded portion or 2.7 mm cortex screws and 2.4 mm cortex screws in the distraction portion Plates for radial head rim available in right and left plates with a 5º tilt to match the anatomy of the radial head Plates for radial head neck fit both the left and right side of the proximal radius 2 DePuy Synthes LCP Proximal Radius Plates 2.4 Surgical Technique

AO Principles AO PRINCIPLES In 1958, the AO formulated four basic principles, which have In become 1958, the guidelines AO formulated internal four basic fixation principles, 1,2. which have become the guidelines for internal fixation 1, 2. 4_Priciples_03.pdf 1 05.07.12 12:08 Anatomic reduction Fracture reduction and fixation to restore restore anatomical relationships. 1 2 Stable fixation Fracture fixation providing absolute absolute or relative stability, stability, as required as by the or relative patient, required the by injury, the patient, and the the personality injury, of and the the fracture. personality of the fracture. Early, active mobilization Early and safe mobilization and rehabilitation of the injured part and and the patient the patient as a whole. as a whole. 4 3 Preservation of blood supply Preservation of of the blood supply to soft tissues and bone by by gentle reduction gentle reduction techniques techniques and careful and handling. careful handling. 1 Müller ME, M Allgöwer, R Schneider, H Willenegger. Manual of Internal Fixation. 3rd ed. Berlin Heidelberg New York: Springer. 1991. 2 Rüedi TP, RE Buckley, CG Moran. AO Principles of Fracture Management. 2nd ed. Stuttgart, New York: Thieme. 2007. 1 Müller ME, Allgöwer M, Schneider R, Willenegger H. Manual of Internal Fixation. 3 rd ed. Berlin, Heidelberg, New York: Springer. 1991. 2 Rüedi TP, Buckley RE, Moran CG. AO Principles of Fracture Management. 2 nd ed. Stuttgart, New York: Thieme. 2007. 4 DePuy Synthes Expert Lateral Femoral Nail Surgical Technique LCP Proximal Radius Plates 2.4 Surgical Technique DePuy Synthes 3

Intended Use and Indications Intended Use The plate and screw implants included in the Radius Plate product family are intended for temporary fixation, correction or stabilization in the radius anatomical region. Indications Extra-articular and intra-articular fractures of the proximal radius and multifragmented radial neck fractures. 4 DePuy Synthes LCP Proximal Radius Plates 2.4 Surgical Technique

Preparation Patient position Position the patient supine on the table. Prepare the extremity from the axilla to the hand. This allows rotation of the forearm and flexion and extension of the elbow during the operative fixation. Approach A lateral approach is most commonly used. Take care to avoid the deep branch of the radial nerve, which runs anterior to the capsule and the radial head. To minimize the risk of operative disruption of the lateral collateral ligament, the capsular incision should remain in the front of the anterior margin of the anconeus muscle and parallel to the fascial limit of the extensor carpi ulnaris. The annular ligament, a true thickening of the capsule, is opened laterally or slightly anteriorly to allow full inspection of the fragments. In selected cases, an osteotomy of the lateral epicondyle will allow an extensile approach. Illustrations on this page are reproduced from Rüedi/Murphy, AO Principles of Fracture Management by permission of AO Publishing, Copyright 2000 AO Publishing, Switzerland. LCP Proximal Radius Plates 2.4 Surgical Technique DePuy Synthes 5

Preparation Safe zone Both plates fit within the Hotchkiss safe zone 2 which is described as an area of 105 on the radial head that is free of impingement between ulna and radius. The Hotchkiss safe zone is located on the opposite side of the radial tuberosity. Neutral Supination = Safe zone Pronation 2 Robert N. Hotchkiss, MD: Displaced Fractures of the Radial Head: Internal Fixation or Excision? AAOS, Vol. 5, No. 1, January/February 1997. 6 DePuy Synthes LCP Proximal Radius Plates 2.4 Surgical Technique

Reduction 1 Reduce fracture Perform preliminary fixation with Kirschner wires. Precaution: The plate reduction wires and Kirschner wires are single use items, do not re-use. 2 Apply Radial Head Plate Instruments 311.430 Handle with Quick Coupling, length 110 mm 314.467 Screwdriver Shaft, Stardrive, T8, self-holding After reducing the fracture, apply the plate and insert a preliminary cortex screw through the DCU portion of one of the elongated Combi holes. Note: The plates are precontoured to fit the anatomy of the radial head. There are no undercuts on the plates to protect the threaded holes form distortion during bending. If it is necessary to bend the plate, use bending pliers. Precaution: Reverse bending or use of the incorrect instrumentation for bending may weaken the plate and lead to premature plate failure (e.g. breakage). Do not bend the plate beyond what is required to match the anatomy. LCP Proximal Radius Plates 2.4 Surgical Technique DePuy Synthes 7

Proximal Screw Insertion 3 Insert proximal screws Instruments 311.430 Handle with Quick Coupling, length 110 mm 03.110.005 Handle for Torque Limiters 0.4/0.8/1.2 Nm 314.467 Screwdriver Shaft, Stardrive, T8, self-holding 511.776 Torque Limiter 0.8 Nm, with Quick Coupling 310.509 Drill Bit B 1.8 mm, with marking, length 110/85 mm, 2-flute, for Quick Coupling 323.029 LCP Drill Sleeve 2.4, with Scale up to 30 mm, for Drill Bits B 1.8 mm Use 2.4 mm locking screws in the head of the plate. Precaution: Careful drilling is necessary, as interference with screws in the proximal portion of the plate is possible. In case of interference, stop drilling and use a screw of appropriate length. See General Notes on Technique (page 10) for additional information. 8 DePuy Synthes LCP Proximal Radius Plates 2.4 Surgical Technique

Shaft Screw Insertion 4 Insert screws in the shaft of the plate Instruments 311.430 Handle with Quick Coupling, length 110 mm 314.467 Screwdriver Shaft, Stardrive, T8, self-holding For cortex screws 2.4 mm 310.509 Drill Bit B 1.8 mm, with marking, length 110/85 mm, 2-flute, for Quick Coupling 310.530 Drill Bit B 2.4 mm, length 100/75 mm, 2-flute, for Quick Coupling 323.202 Universal Drill Guide 2.4 For cortex screws 2.7 mm 310.534 Drill Bit B 2.0 mm, length 100/75 mm, 2-flute, for Quick Coupling 310.260 Drill Bit B 2.7 mm, length 100/75 mm, 2-flute, for Quick Coupling 323.260 Universal Drill Guide 2.7 For locking screws 2.4 mm 310.509 Drill Bit B 1.8 mm, with marking, length 110/85 mm, 2-flute, for Quick Coupling 323.029 LCP Drill Sleeve 2.4, with Scale up to 30 mm, for Drill Bits B 1.8 mm Use 2.4 mm and 2.7 mm cortex screws and 2.4 mm locking screws in the shaft of the plate. LCP Proximal Radius Plates 2.4 Surgical Technique DePuy Synthes 9

General Notes on Technique Determine screw choice If planning a combination of locking and cortex screws, first use a cortex screw to pull the plate to the bone. Precaution: If using a locking screw first, ensure that the plate is held securely to the bone, to avoid spinning of the plate as the screw locks into the plate. Insert cortex screws Instruments 310.534 Drill Bit B 2.0 mm, with marking, length 110/85 mm, 2-flute, for Quick Coupling 310.260 Drill Bit B 2.7 mm, length 100/75 mm, 2-flute, for Quick Coupling 310.509 Drill Bit B 1.8 mm, with marking, length 110/85 mm, 2-flute, for Quick Coupling 310.530 Drill Bit B 2.4 mm, length 100/75 mm, 2-flute, for Quick Coupling 319.005 Depth Gauge for Screws B 2.0 and 2.4 mm, measuring range up to 40 mm 323.202 Universal Drill Guide 2.4 323.260 Universal Drill Guide 2.7 Use the 2.4 or 2.7 universal drill guide for an eccentric (distraction) or neutral (buttress) insertion of cortex screws. For the 2.4 mm cortex screw, use the 1.8 mm drill bit for the threaded hole and the 2.4 mm drill bit for the gliding hole. Determine the length of the screw by using the depth gauge. For 2.7 mm cortex screws, use the 2.0 mm drill bit for the threaded hole and the 2.7 mm drill bit for the gliding hole. 11 DePuy Synthes LCP Proximal Radius Plates 2.4 Surgical Technique

Insert locking screws Instruments 310.509 Drill Bit B 1.8 mm, with marking, length 110/85 mm, 2-flute, for Quick Coupling 311.430 Handle with Quick Coupling, length 110 mm 03.110.005 Handle for Torque Limiters 0.4/0.8/1.2 Nm 314.467 Screwdriver Shaft, Stardrive, T8, self-holding 511.776 Torque Limiter 0.8 Nm, with Quick Coupling 319.005 Depth Gauge for Screws B 2.0 and 2.4 mm, measuring range up to 40 mm 323.029 LCP Drill Sleeve 2.4, with Scale up to 30 mm, for Drill Bits B 1.8 mm Screw the LCP drill guide into a hole until fully seated. For the 2.4 mm locking screws, use the 1.8 mm drill bit with marking, to drill to the desired depth. Determine the screw length directly from the mark on the drill bit and scale on the drill sleeve. This may be verified using a depth gauge for 2.0 mm and 2.4 mm screws. Insert the locking screw manually with the screwdriver shaft handle and Torque Limiter 0.8 Nm. Carefully tighten the locking screw. LCP Proximal Radius Plates 2.4 Surgical Technique DePuy Synthes 11

General Notes on Technique Alternative: Locking screw insertion with holding sleeve Instruments 311.430 Handle with Quick Coupling, length 110 mm 03.110.005 Handle for Torque Limiters 0.4/0.8/1.2 Nm 314.467 Screwdriver Shaft, Stardrive, T8, self-holding 511.776 Torque Limiter 0.8 Nm, with Quick Coupling 314.468 Holding Sleeve for Screws Stardrive B 2.4 mm, T8, for Screwdriver Shafts B 3.5 mm, for 314.467 An alternate method may be used for insertion of locking screws, using the locking screw to pull the plate to the bone. Place the holding sleeve onto the screwdriver shaft. Pick up the locking screw with the holding sleeve and insert the screw. With the locking screw still held by the holding sleeve, tighten the screw until the plate is drawn to the bone. Pull up on the holding sleeve to release the screw head, and tighten the screw until a click is heard. Precaution: Take care to hold the plate securely on the bone to avoid spinning the plate as the screw locks to the plate. 11 DePuy Synthes LCP Proximal Radius Plates 2.4 Surgical Technique

Postoperative Treatment and Implant Removal Postoperative treatment Postoperative treatment with locking compression plates does not differ from conventional internal fixation procedures. Implant removal To remove locking screws, first unlock all screws from the plate and then remove the screws completely from the bone. This avoids rotation of the plate when removing the last locking screw. LCP Proximal Radius Plates 2.4 Surgical Technique DePuy Synthes 11

Plates Rim plates 2-hole plates X41.680 (right) X41.681 (left) 3-hole plates X41.682 (right) X41.683 (left) 4-hole plates X41.684 (right) X41.685 (left) Neck plates 2-hole plates X41.690 3-hole plates X41.691 4-hole plates X41.692 All implants are available non-sterile or sterile packed. Add suffix S to article number to order sterile product. X = 2: Stainless Steel X = 4: Titan (TiCP) 11 DePuy Synthes LCP Proximal Radius Plates 2.4 Surgical Technique

Screws Cortex Screws 2.4 mm, self-tapping, Stardrive, T8 (X01.760 X01.780, length 10 30 mm) For use in Combi holes in plate shaft, to provide distraction or neutral fixation Low-profile head in the plate holes Stardrive recess mates with self-retaining screwdriver 10 mm to 30 mm lengths Locking Screws 2.4 mm, self-tapping, Stardrive, T8 (X12.806 X12.830, length 6 30 mm) Threaded, conical head locks securely into the plate to provide angular stability Locked screws allow unicortical screw fixation and load transfer to the near cortex Stardrive recess mates with self-retaining screwdriver 6 mm to 30 mm lengths Cortex Screws 2.7 mm, self-tapping, Stardrive, T8 (X02.870 X02.886, length 10 26 mm) For use in Combi holes in plate shaft, to provide distraction or neutral fixation Low-profile head in the plate holes Stardrive recess mates with self-retaining screwdriver 10 mm to 26 mm lengths Note: For information on fixation principles using conventional and locked plating techniques, please refer to the LCP Surgical Technique DSEM/TRM/0115/0278 (036.000.019). All implants are available non-sterile or sterile packed. Add suffix S to article number to order sterile product. X = 2: Stainless Steel X = 4: Titanium Alloy (TAN) LCP Proximal Radius Plates 2.4 Surgical Technique DePuy Synthes 11

Instruments 311.430 Handle with Quick Coupling, length 110 mm 314.467 Screwdriver Shaft Stardrive T8, self-holding 314.468 Holding Sleeve for Screws Stardrive 2.4, for Screwdriver Shaft 314.467 323.029 LCP Drill Sleeve 2.4, with scale up to 30 mm, for Drill Bit 1.8 mm 310.534 Drill Bit B 2.0 mm with marking, length 110/85 mm, 2-fluted, for Quick Coupling 310.509 Drill Bit B 1.8 mm with marking, length 110/85 mm, 2-fluted, for Quick Coupling 511.776 Torque Limiter 0.8 Nm, with Quick Coupling 03.110.005 Handle for Torque Limiters 0.4/0.8/1.2 Nm 11 DePuy Synthes LCP Proximal Radius Plates 2.4 Surgical Technique

MRI Information Torque, Displacement and Image Artifacts according to ASTM F 2213-06, ASTM F 2052-06e1 and ASTM F 2119-07 Non-clinical testing of a worst case scenario in a 3 T MRI system did not reveal any relevant torque or displacement of the construct for an experimentally measured local spatial gradient of the magnetic field of 3.69 T/m. The largest image artifact extended approximately 91 mm from the construct when scanned using the Gradient Echo (GE). Testing was conducted on a 3 T MRI system. Radio-Frequency-(RF-)induced heating according to ASTM F 2182-11a Non-clinical electromagnetic and thermal simulations of a worst case scenario lead to temperature rises of 7.5 C (1.5 T) and 6.1 C (3 T) under MRI Conditions using RF Coils (whole body averaged specific absorption rate [SAR] of 2 W/kg for 15 minutes). Precautions: The above mentioned test relies on nonclinical testing. The actual temperature rise in the patient will depend on a variety of factors beyond the SAR and time of RF application. Thus, it is recommended to pay particular attention to the following points: It is recommended to thoroughly monitor patients undergoing MR scanning for perceived temperature and/or pain sensations. Patients with impaired thermoregulation or temperature sensation should be excluded from MR scanning procedures. Generally, it is recommended to use an MRI system with low field strength in the presence of conductive implants. The employed specific absorption rate (SAR) should be reduced as far as possible. Using the ventilation system may further contribute to reduce temperature increase in the body. LCP Proximal Radius Plates 2.4 Surgical Technique DePuy Synthes 11

Synthes GmbH Eimattstrasse 3 Not all products are currently available in all markets. 4436 Oberdorf Switzerland This publication is not intended for distribution in the USA. Tel: +41 61 965 61 11 Fax: +41 61 965 66 00 All surgical techniques are available as PDF files at www.depuysynthes.com www.depuysynthes.com/ifu 0123 DePuy Synthes Trauma, a division of Synthes GmbH. 2016. All rights reserved. 036.000.681 DSEM/TRM/0615/0399(1) 05/16