Raju Vaishya, Amit Kumar Agarwal, Nishint Gupta & Vipul Vijay

Similar documents
Failed Subtrochanteric Fracture How I Decide What to Do?

Assessment of Prognosis of Patients with Intertrochanteric Fractures Undergoing Treatment with PFN: An Observational Study

HOW TO CITE THIS ARTICLE:

ORIGINAL ARTICLE. INTER TROCHANTERIC # NECK FEMUR FIXATION WITH TFN 250 CASES. Prasad Vijaykumar Joshi, Chandrashekar Yadav.

A comparative study of 30 cases of trochanteric fracture femur treated with dynamic hip screw and proximal femoral nailing

Comparitive Study between Proximal Femoral Nailing and Dynamic Hip Screw in Intertrochanteric Fracture of Femur *

Treatment of extracapsular hip fractures with the proximal femoral nail (PFN) : Long term results in 45 patients

PFN in trochanteric fractures: Clinico radiological outcome study

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

A Comparison of the Proximal Femoral Nail Antirotation Device and Dynamic Hip Screw in the Treatment of Unstable Pertrochanteric Fracture

Technique Guide. DHS Blade. For osteoporotic bone.

A Prospective Study to Evaluate the Management of Sub-trochanteric Femur Fractures with Long Proximal Femoral Nail

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia

Provision of Rotational Stability: Prevention of Collapse: Closed Fracture Reduction: Minimally Invasive Surgery with no Exposure of the Fracture:

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

JMSCR Vol. 03 Issue 08 Page August 2015

Journal of Orthopaedic Surgery and Research 2010, 5:62

Salvage of failed dynamic hip screw fixation of intertrochanteric fractures

Crossed Steinmann Pin Fixation In Supracondylar Femur Fractures In Adults A Case Series

Review of Proximal Nail Antirotation (PFNA) and PFNA-2 Our Local Experience

Distal femoral fracture with subsequent ipsilateral proximal femoral fracture

STUDY OF RESULTS OF ENDER NAILING AND CANNULATED CANCELLOUS SCREW IN THE TREATMENT OF INTERTROCHANTERIC FRACTURE FEMUR

Breakage of the Tail Portion of the Lag Screw during Removal of Proximal Femoral Zimmer Natural Nail: Report of Two Cases with Technical Notes

BRIDGE PLATING OF COMMINUTED SHAFT OF FEMUR FRACTURES

Distal Femur Fractures in The Elderly The Ideal Construct

Randomized comparative study to evaluate the role of proximal femoral nail and dynamic hip screw in unstable trochanteric fractures

PFNA-II. Proximal Femoral Nail Antirotation.

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

The Journal of the Korean Society of Fractures Vol.16, No.1, January, 2003

3.5 mm Locking Attachment Plate

Chronic Lymphedema of the Lower Limb: A Rare Cause of Dislocation of Total Hip Arthroplasty

9/24/2015. When Can I Use a SHS? When CAN T I Use a SHS? Sliding Hip Screw. Time proven. Technically simple. Cheap. Quick

MRIMS Journal of Health Sciences 2016;4(1) pissn: , eissn:

Case Report Late Occurring Medial Migration of a Lag Screw in Gamma Nailing

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Subtrochanteric femoral fracture is one of

Proximal femur locking compression plate in complex proximal femoral fractures: a retrospective analysis

Functional evaluation of proximal femoral fractures managed with cephalomedullary nailing by oxford hip score - A prospective study

The Lateral Trochanteric Wall A Key Element in the Reconstruction of Unstable Pertrochanteric Hip Fractures

Locked Plating: Clinical Indications

Technique Guide. DHS/DCS System. Including LCP DHS and DHS Blade.

Valgus subtrochanteric osteotomy for malunited intertrochanteric fractures : Our experience in 5 cases

ABSTRACT INTRODUCTION. Hui Xie 1,*, Zhan Wang 2,*, Junji Zhang 3, Langhai Xu 2 and Bao Chen 1. Research Paper

Results of Proximal Femoral Nail in Intertrochanteric Fracture of Femur

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual

Treatment of ipsilateral femoral neck and shaft fractures. Mohamed E. Habib, Yasser S. Hannout, and Ahmed F. Shams

A randomised comparison of AMBI, TGN and PFN for treatment of unstable trochanteric fractures

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

PFNA. With Augmentation Option.

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

PATENTED A-PFN. Antirotator Proximal Femoral Nail. Medical Devices

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

Closed reduction and internal fixation of fractures of the shaft of the femur by the Titanium Elastic Nailing System in children.

Is the biomechanical problem of trochanteric fracture solved?

Page Proof 1 of 5. Fig. E1-A The INTERTAN nail was short or long.

PFNA. With Augmentation Option.

Acu-Sinch Repair System. Technical Monograph

Intramedullary Nailing: History & Rationale

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

Techique. Results. Discussion. Materials & Methods. Vol. 2 - Year 1 - December 2005

Comparative study between proximal femoral nail and proximal femoral nail antirotation in management of unstable trochanteric fractures

Results of distal femur locking plate in communited distal femur fractures

AO / Synthes Proximal Posterior Medial Tibia Plate. Tibial Plateau Fx. Osteosynthesis

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

JMSCR Vol 05 Issue 07 Page July 2017

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee.

Functional outcome of proximal femoral nailing in inter trochanteric fractures of femur: A prospective study

Treatment of Femoral Intertrochanteric Fracture with Proximal Femoral Nail

A comparative study of less invasive stabilization system and titanium elastic nailing for subtrochanteric femur fractures in older children

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

designed to advance the treatment of hip fractures.

Technique Guide Small Bone Fusion System

Pediatric LCP Hip Plate. For osteotomy and trauma applications in the proximal femur.

AcUMEDr. LoCKING CLAVICLE PLATE SYSTEM

Locked Plating: Biomechanics and Biology

Use Of A Long Femoral Stem In The Treatment Of Proximal Femoral Fractures: A Report Of Four Cases

Reverse LISS plate in treating femoral intertrochanteric fractures: A review of 22 cases.

Augmentation of proximal femoral nail in unstable trochanteric fractures

Ipsilateral femoral neck and shaft fractures: a retrospective analysis of two treatment methods

.org. Tibia (Shinbone) Shaft Fractures. Anatomy. Types of Tibial Shaft Fractures

Minimally Invasive Plate Osteosynthesis (MIPO) for Proximal and Distal Fractures of The Tibia: A Biological Approach

Nonunion of the Femur Treated with Conventional Osteosynthesis Combined with Autogenous and Strut Allogeneic Bone Grafts

Locked plating constructs are creating a challenge for surgeons.

INTERTROCHANTERIC FEMORAL FRACTURES TREATED BY DYNAMIC HIP SCREW

Should the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)? A retrospective study

)23( COPYRIGHT 2016 BY THE ARCHIVES OF BONE AND JOINT SURGERY

PROXIMAL FEMORAL NAILING VS DYNAMIC HIP SCREW FOR INTERTROCHANTERIC FRACTURES A COMPARATIVE CLINICAL STUDY

Dynamic hip screw fixation of intertrochanteric fractures without using traction table

Peritroch Hip Fractures. Robert M Harris MD. Hip Fractures. Factors Influencing Construct Strength: Uncontrolled factors 4/28/2016

Lower risk of implant failure and reoperation

Biomechanics of Fractures and Fixation

Pelvis injuries Fractures of the femur (proximal,shaft) Dr Tamás Bodzay

A Clinical Study For Evaluation Of Results Of Closed Interlocking Nailing Of Fractures Of The Shaft Of The Tibia

Comparative Study between Locking Compression Plate vs. Supracondylar Nail for Supracondylar Femur Fractures

Type I : At the level of lesser trochanter Type II : Less than 2.5 cm below lesser trochanter. Type III : cm below lesser trochanter

Low Bend Distal Tibia Plates

EVOS MINI with IM Nailing

Aesculap Targon FN. Head Preserving Solution for Medial Femoral Neck Fractures. Aesculap Orthopaedics

Fractures of the tibia shaft treated with locked intramedullary nail Retrospective clinical and radiographic assesment

Transcription:

Reversed distal femoral locking plate for failed proximal femoral nail with nonunion of proximal femoral fractures Raju Vaishya, Amit Kumar Agarwal, Nishint Gupta & Vipul Vijay International Orthopaedics ISSN 0341-2695 DOI 10.1007/s00264-015-3002-5 1 23

Your article is protected by copyright and all rights are held exclusively by SICOT aisbl. This e-offprint is for personal use only and shall not be self-archived in electronic repositories. If you wish to self-archive your article, please use the accepted manuscript version for posting on your own website. You may further deposit the accepted manuscript version in any repository, provided it is only made publicly available 12 months after official publication or later and provided acknowledgement is given to the original source of publication and a link is inserted to the published article on Springer's website. The link must be accompanied by the following text: "The final publication is available at link.springer.com. 1 23

DOI 10.1007/s00264-015-3002-5 ORIGINAL PAPER Reversed distal femoral locking plate for failed proximal femoral nail with non-union of proximal femoral fractures Raju Vaishya 1 & Amit Kumar Agarwal 1 & Nishint Gupta 1 & Vipul Vijay 1 Received: 10 July 2015 /Accepted: 14 September 2015 # SICOT aisbl 2015 Abstract Introduction Failure of proximal femoral fracture managed by proximal femoral nail (PFN) leads to a very difficult situation to handle with conventional techniques, and reversed distal femoral locking compression plate (DF-LCP) is of great benefit in these selective cases. Methods Twelve patients with ununited proximal femoral fractures including subtrochanteric fractures with a failed PFN implant were included in the study. All patients with periprosthetic fractures and fractures treated by implants other than PFN were excluded from this study. Result All fractures went into union in an average time of nine months and 15 days with no implant failures. The mean time of re-osteosynthesis after the primary index surgery of PFN was one year eight months. Mean surgical time of reosteosynthesis was 110 minutes, and average blood loss during surgery was 550 ml. Discussion The PFM is one of the most commonly used implant for unstable proximal femur fractures. The use of PFN is technically demanding and is associated with high failure rates. Although dynamic compression screw (DCS), proximal femoral locking plate (PF-LCP) and other implants can be used in these failed situations, they are associated with a high * Amit Kumar Agarwal amitorthopgi@yahoo.co.in 1 Raju Vaishya raju.vaishya@gmail.com Nishint Gupta nishintgupta@gmail.com Vipul Vijay dr_vipulvijay@yahoo.com Department of Orthopaedics, Indraprastha Apollo Hospital, Sarita Vihar, Mathura Road, New Delhi, India 110076 complication rate. The reversed DF-LCP is a rescue implant for these complex situations. Apart from anatomical and biomechanical advantages, there are several other clinical benefits of using DF-LCP. Conclusion We conclude that DF-LCP is a potential and safe implant of choice for the management of nonunion associated with failed PFN. It may be considered an implant of choice as rescue from such a complex situation. It offers several anatomical, biomechanical and clinical advantages over other available conventional implants. Keywords Nonunion fracture. Proximal femur. Distal femoral locking plate. Re-osteosynthesis Introduction Despite the advancement of several fixation devices, treatment of unstable pertrochanteric fractures and subtrochanteric fractures of the femur remains a challenge [1]. Until the introduction of intramedullary nails (IM), these unstable pertrochanteric fractures were mostly treated with a sliding hip screw system, which was associated with high failure rates [2, 3]. To overcome the problems associated with the sliding hip screw fixation system, the proximal femoral nail (PFN) was introduced in 1997 by AO/ASIF [4, 5]. It combined the features of a femoral IM with hip screws to provide rotational stability of the proximal fracture fragment along with locking at the distal end. However, the use of PFN is technically demanding, and the margin of error is low. Failure to achieve precise reduction of fractured fragments and faulty insertion of this nail is often responsible for the failure of osteosynthesis [6] and non-union. It leaves behind a complex situation to deal with due to the disturbed anatomy of the proximal femur and poor bone stock left for fracture fixation by other implants

during re-osteosynthesis. Conventional implants used for proximal femoral fracture fixation, such as the dynamic hip screw (DHS), dynamic condylar screw (DCS), PFN, 950 angle blade plate etc., may not be able to fix these complex and ununited fractures. Hence, there is a need to devise a suitable implant to address this complex situation after a failed PFN. The proximal femoral locking compression plate (AO, Synthes)wasintroducedrecentlytodealwithsuchfractures andfailedconditionsbuthasbeenreportedtobeassociated with varus collapse of the fracture and screw breakage due to high axial bending forces, with its high failure rate of up to 27 % [7]. We present our experience with reversed anatomical distal femoral locking compression plate (DF-LCP) for unstable and ununited proximal femur fracture following a failed PFN. Materials and methods We had operated upon 12 patients who had a failed PFN and ununited fractures using a reverse DF-LCP (i.e. a right-side plate for a left proximal femoral fracture) classified using the AO/ASIF system [8]. We only included patients with ununited proximal femoral fractures (A2.2 A3.3), including subtrochanteric fractures with a failed PFN implant. All patients with periprosthetic fractures and fractures treated with implants other than PFN were excluded. Patients were primarily assessed clinically and radiologically for limb shortening, fracture union and implant failure. Fig. 1 Anatomical positioning of distal femoral locking compression plate (DF-LCP) in a reverse pattern over proximal femur (front and lateral views) long plate was used in all cases to allow at least eight to ten cortices to be fixed distal to the fracture. These patients were mobilised non-weight bearing using walking aids (crutches, frame, walker) two days postoperatively after removal of drains from the surgical site. Hip and knee movements were allowed and encouraged. Weight bearing was gradually increased depending upon union progression on serial radiographs during follow-up, which lasted until a sound union was achieved on radiographs. Surgical technique Surgery was performed with the patient in either a lateral decubitus or a supine position on a traction table. In lateral decubitus, anteroposterior (AP) view was taken by rotating the C-arm at 90 and the lateral view by both gently rotating the hip and adjusting the C-arm; in the supine position, the C-arm was position in between the two limbs [9]. The failed or broken implants were first removed (distal and proximal screws and then the nail) and the ununited proximal femoral fracture was freshened up and fixed using a reverse DF-LCP. In three cases, varus deformity of the proximal fragment was corrected using a Schanz pin in the femoral neck as a Bjoystick^ [10]. We used a DF-LCP in a reversed position to fix the proximal femoral fracture fragments (Fig. 1) employing a minimally invasive technique, which allowed extra periosteal introduction with minimal damage to soft tissues. The broad end of the plate allowed a sufficient number of 5-mm locking screws to be placed in the proximal femoral fragment in order to optimise adequate fixation. The shaft of the plate allowed fixation with both locking and nonlocking cortical screws, which provided the flexibility to achieve plate-to-bone opposition as well as axial compression or angular stability. A sufficiently Results There 12 patients (seven men, five women) with an average age of 64.6 (range 37 85) years. Mean time of reosteosynthesis after the primary index surgery of PFN was one year three months to two years nine months (average 1 year 8 months). All 12 cases had established nonunion of their fractures (Table 1) with implant failure (loosening, breakage, screw cutout) of PFN (Table 2). Mean surgical time of reosteosynthesis was 110 (range 90 150) min. In all patients, fracture fragments were first freshened and then fixed with a reversed DF-LCP after proper reduction of the fracture fragments. Corticocancellous bone grafts from the contralateral iliac crest were harvested simultaneously by another surgical team and used at the fracture site in all cases. Stability and implant positioning of the construct was checked intra-operatively using an image intensifier and checked post-operatively by X-ray (Figs. 2 and 3). Average blood loss during surgery was 550 (range 320 870) ml. No postoperative splinting was required in any of these cases. The average duration of followupwas 13 months 41 days(range 12 23 months). We achieved

Table 1 Study demographics Case no. Sex Age (years) Reason for failure Union status Mean time of re-osteo synthesis (months) Surgery duration (min) Blood loss (ml) Bone graft Av. follow-up (months) Preop shortening (cm) Postop shortening (cm) Status of union at last follow-up Av. union time (months) Complication 1 M 55 Screw cutout Nonunion 15 100 750 Done 12 3.6 1.5 United 10 Nil 2 M 57 Screw cutout Nonunion 22 120 870 Done 16 5 2 United 14 Wound dehiscence Nonunion 18 90 600 Done 12 4 1.6 United 8 Nil 3 F 61 Broken neck screw 4 M 37 Nail breakage Nonunion 33 105 350 Done 13 3.2 1.3 United 7 Nil 5 F 63 Screw cutout Nonunion 18 110 650 Done 22 5 1.5 United 12 Nil 6 F 72 Screw cutout Nonunion 24 120 410 Done 12 4.2 2 United 9 Nil Nonunion 20 150 500 Done 12 4.5 1.5 United 9 Nil 7 M 85 Broken neck screw 8 F 59 Nail breakage Nonunion 18 100 400 Done 12 6.5 5.1 United 9 Nil 9 M 72 Screw cutout Nonunion 15 90 320 Done 12 3.6 1.5 United 8 Nil Nonunion 24 110 500 Done 12 4 2 United 10 Nil 10 F 67 Broken neck screw 11 M 72 Screw cutout Nonunion 15 130 650 Done 14 5 1.4 United 9 Nil Nonunion 18 95 600 Done 12 4.4 1.5 United 9 Nil 12 M 76 Nail breakage Mean 64.6 years 20 (1 year 8 months) 110 min. 550 (ml) 13.41 4.416 1.9 9.5

Table 2 Patterns of proximal femoral nail failure Pattern of failure No. cases 1) Screw cutout 7 a) Z effect 5 b) Reverse Z effect 2 2) Broken screw 3 3) Broken nail 2 union in all cases with a mean time to radiographic union of nine months and 15 days (7 14 months) (Figs. 4 and 5). None of the DF-LCP implants showed failure on follow-up radiographs. Femoral shortening was improved from an average of 4.4 (range 3.2 6.5) cm preoperatively to 1.9 (range 1.3 5.1) cm post-operatively. No neurovascular complications were observed after the surgical procedure. One case had wound dehiscence and minor surgical-site infection, which was managed successfully by wound debridement and secondary suturing. Discussion Unstable pertrochanteric and subtrochanteric fractures are often seen in geriatric individuals due to the high prevalence of osteoporosis in this age group [11], whereas in younger patients, this type of injury is usually caused by high-energy trauma. Fixation of these fractures can be achieved with various modalities, such as PFN, DCS and 90 -angle blade plate [12], but these methods are often associated with complications such as nonunion, shortening, varus collapse, loss of fixation, implant loosening and Fig. 2 Anteroposterior (AP) and lateral radiographs of case 4 showing broken proximal femoral nail (PFN) at the subtrochanteric region, with nonunion of the fracture site Fig. 3 Thirteen months post-operative anteroposterior (AP) and lateral radiographs of case 4 showing union of subtrochanteric fracture treated with reversed DF-LCP and bone graft implant breakage. Fixation failure and nonunion in subtrochanteric fractures are typically high and are reported with an incidence of up to 28 % [13]. This is common due to various factors, such as weak, osteoporotic bone, thick cortical bone with scarce blood supply in the subtrochanteric region and biomechanical factors such as high varus stress on weight bearing at the fracture site, which ultimately lead to implant breakage and nonunion. Surgical factors such as poor reduction, wrong nail-entry site and improper positioning of screws leads to screw breakage; screw backout and varus collapse also contribute to failure [14, 15]. One of the most commonly used implants for unstable proximal femur fractures is PFN, which allow rotational stability of the proximal fracture fragment through proximal neck screws. However, use of the PFN is technically demanding and requires high technical expertise, precise reduction, correct entry site, proper positioning of neck screws, failure of which may lead to deficient osteosynthesis and nonunion. The various studies have reported high re-operation rate of 4 % to 28 % with PFN done primarily in unstable proximal femur fractures [16]. In our experience the complication rate of PFN done primarily in unstable proximal femur fractures is approximately 10 % and DF-LCP has turned out to be a good option in these failed cases. In our study, we found three main pattern of failure of PFN (Figs. 6 and 7) in the form of neck-screw cutout (lateral migration of proximal screws) due to the BZ^ [17] or reverse BZ^ effect, broken nail in subtrochanteric region and broken neck screws (Table 2). Cutout of the neck screw or lateral migration of the proximal screw shows a BZ^ effect phenomenon in which the superior screw penetrates the femoral head or pelvis while the inferior screw migrates

Fig. 4 Anteroposterior (AP) and lateral radiographs of case 1 showing failed proximal femoral nail (PFN) with lateral displacement of proximal screws, varus collapse and nonunion at both proximal and distal fracture sites laterally, which leads to varus collapse of the fracture [18]. The reverse BZ^ effect phenomenon occurs due to lateral sliding of the superior screw while the inferior screw remain impacted to the nail [4]. Causes of nonunion could be due to interfragmentary motion due to failure of rigid fixation, interposition of fibrous tissue at the fracture site, malposition of bony fragments by inappropriate reduction or fracture comminution [19]. A variety of implants has been used to fix proximal femoral nonunion. The DCS is a good alternative for failed pertrochanteric fractures but requires considerable technical expertise and leads to hampered periosteal blood supply due to excessive stripping in an already failed fracture with poor biological environment and bone stock [20]. The proximal femoral LCP (PF-LCP; AO Synthes) was introduced recently to deal with such fractures and failed conditions. It is, however, associated with varus collapse of the fracture and screw breakage due to high axial bending forces. Lee et al. reported a high PF- LCP failure rate of up to 27 %, with screw loosening being most common in their study of 26 cases [7]. We have had good outcomes after using reversed DF-LCP for complex proximal femoral fractures with failed PFN and nonunion. Some studies have shown that IMs have limitations in managing fractures types more severe than 31A2.2 [21, 22]. In these cases, a reversed DF-LCP plate was successfully used. In our experience, severe lateral wall comminution is one of the main indication for using a DF-LCP as a primary mode of fixation. A large displaced greater trochanteric fragment may be considered a relative indication due to difficulty in finding the correct entry point if a PFN is being used. In this study, we achieved union in all 12 ununited fractures with the use of reversed DF-LCP and bone grafting. Furthermore, stability of this construct allowed better rehabilitation, appropriate correction of limb-length discrepancy and minimal complications. This partial restoration of length was achieved by correcting varus deformity at the fracture site. We believe that the DF-LCP is anatomically suitable for the proximal femur using a plate from the contralateral side in a reverse pattern [23]. The head of the plate consists of five threaded 5.0-mm peripheral screw holes that accept locking Fig. 5 Twelve months post-operative anteroposterior (AP) and lateral radiograph of case 1 showing fracture union with good callus

Fig. 6 Broken neck screws of the proximal femoral nail (PFN) (case 3); screw cutout (BZ^ effect) (case 1) screws. The central 7.3-mm screw has an angle of 95 to the plate shaft. The shaft has Combi holes that combine a dynamic compression unit (DCU) hole with a locking screw hole, allowing insertion of standard bicortical screws or a locked screw. Studies have shown that DF-LCP bears more maximal axial load than the DCS by 34 % and of an IM by 13 %, thus advocating its biomechanical advantage. Apart from anatomical and biomechanical advantages, there are several other clinical benefits of using DF-LCP (Table 3) for proximal femoral nonunion, especially after a failed PFN. It has the advantage of multiple screw purchase in the proximal femur and permits the use of minimally invasive surgical techniques in selected cases. Moreover, it is easily available in all setups and can be fixed to the proximal femur even with the retained femoral implant. The locking system of plate and screws improves the stability of construct in an osteoporotic bone that allows early rehabilitation. DF-LCP can be used for periprosthetic fractures of the hip, as well. It causes less damage to the traumatised bone and soft tissue after failed PFN and is relatively affordable compared with other implants. Based on our experience and good outcomes, we propose that DF-LCP is a safe as a potential implant of choice for managing nonunion associated with failed PFN. It may be considered an rescue implant of choice in such a complex situation. We are, however, aware that a DF-LCP was not designed for use in proximal femoral fractures. We had used it in selected cases, since it adapts so well to the contour of the proximal femur and provides excellent fixation of complex fractures where other implants are not feasible. Further clinical and biomechanical studies are required to advocate its routine use for all proximal femoral fractures. We believe that due to the background of success using implant, more dedicated implants for fixation of complex fractures of the proximal femur will be available in the near future. Table 3 Clinical advantages of the reversed distal femoral locking compression plate (DF-LCP) Serial no. Advantages of reverse DF-LCP in failed proximal femoral fractures after proximal femoral nail Fig. 7 Anteroposterior (AP) radiograph of cases 3 and 1 showing fracture fixation with a reversed distal femoral locking compression plate (DF-LCP) 12 months post-operatively 1 Easy availability in all setups 2 Multiple screw fixations in proximal femur 3 Permits the use of minimally invasive surgical techniques 4 Can be fixed to proximal femur even with retained implant in the femur 5 Provide a stable construct which allow early rehabilitation 6 Can be used in periprosthetic fractures of the hip 7 Cost effective 8 Locking system of plate and screws improves stability of construct in osteoporotic bone 9 Less damage to traumatised bone and soft tissue after failed PFN

Compliance with Ethical Standards Conflict of interest interests. References The author(s) declare that they have no competing 1. Schipper IB, Steyerberg EW, Castelein RM et al (2004) Treatment of unstable trochanteric fractures: randomized comparison of the gamma nail and the proximal femoral nail. J Bone Joint Surg (Br) 86(1):86 94 2. Simpson AH, Varty K, Dodd CA (1989) Sliding hip screws: modes of failure. Injury 20:227 231 3. Davis TR, Sher JL, Horsman A et al (1990) Intertrochanteric femoral fractures: mechanical failure after internal fixation. J Bone Joint Surg (Br) 72(1):26 31 4. Boldin C, Seibert FJ, Fankhauser F et al (2003) The proximal femoral nail (PFN) a minimal invasive treatment of unstable proximal femoral fracturesa prospective study of 55 patients with a followup of 15 months. Acta Orthop Scand 74(1):53 58 5. Simmermacher RKJ, Bosch AM, Van der Werken C (1999) The AO/ASIF-proximal femoral nail (PFN): a new device for the treatment of unstable proximal femoral fractures. Injury 30:327 332 6. Kanthimathi B, Narayanan VL (2012) Early complications in proximal femoral nailing done for treatment of subtrochanteric fractures. Malays Orthop J 6(1):25 9 7. Lee WT, Murphy D, Kagda FH et al (2014) Proximal femoral locking compression plate for proximal femoral fractures. J Orthop Surg (Hong Kong) 22:287 293 8. Newey ML, Ricketts D, Roberts L (1993) The AO classification of long bone fractures: an early study of its use in clinical practice. Injury 24(5):309 12 9. Ehlingera M, Brinkerta D, Bessea J et al (2011) Reversed anatomic distal femur locking plate for periprosthetic hip fracture fixation. Orthop Traumatol: Surg Res 97:560 564 10. Pape HC, Tarkin IS (2009) Intraoperative reduction techniques for difficult femoral fractures. J Orthop Trauma 23(5):S6 11 11. Cummings SR, Melton LJ (2002) Epidemiology and outcomes of osteoporotic fractures. Lancet 359:1761 1767 12. Tencer AF, Johnson KD, Johnston D et al (1984) A biomechanical comparison of various methods of stabilization of subtrochanteric fractures of the femur. J Orthop Res 2:297 305 13. Kang SH, Han SK, Kim YS et al (2013) Treatment of subtrochanteric nonunion of the femur: whether to leave or to exchange the previous hardware. Acta Orthop Traumatol Turc 47(2): 91 95 14. Bredbenner TL, Snyder SA, Mazloomi FR et al (2005) Subtrochanteric fixation stability depends on discrete fracture surface points. Clin Orthop Relat Res 432:217 225 15. Parker MJ, Dutta BK, Sivaji C et al (1997) Subtrochanteric fractures of the femur. Injury 28:91 95 16. Zhou F, Zhang ZS, Yang H et al (2012) Less invasive stabilization system (LISS) versus proximal femoral nail anti-rotation (PFNA) in treating proximal femoral fractures: a prospective randomized study. J Orthop Trauma 26:155 162 17. Tyllianakis M, Panagopoulos A et al (2004) Treatment of extracapsular hip fractures with the proximal femoral nail (PFN): long-term results in 45 patients. Acta Orthop Belg 70:444 454 18. Werner-Tutschku W, Lajtai G, Schmiedhuber G et al (2002) Intraund perioperative Komplikationen bei der Stabilisierungvon perund subtrochantären Femur fracturen mittels. PFN. Unfallchirurg 105:881 885 19. De Vries JS, Kloen P, Borens O (2006) Treatment of subtrochanteric nonunions. Injury 37:203 211 20. Sanders R, Regazzoni P (1989) Treatment of subtrochanteric femur fractures using the dynamic condylar screw. J Orthop Trauma 3(3): 206 13 21. Sahin S, Ertürer E, Oztürk I et al (2010) Radiographic and functional results of osteosynthesis using the proximal femoral nail antirotation (PFNA) in the treatment of unstable intertrochanteric femoral fractures. Acta Orthop Traumatol Turc 44(2):127 34 22. Simmermacher RKJ, Ljungqvist J, Bail H et al (2008) The new proximal femoral nail anti-rotation (PFNA) in daily practice- results of a multicentre clinical study. Injury 39:932 939 23. Haq RU, Manhas V, Pankaj A et al (2014) Proximal femoral nails compared with reverse distal femoral locking plates in intertrochanteric fractures with a compromised lateral wall; a randomized controlled trial. Int Orthop 38(7):1443 1449