LabVIEW based Non-Invasive prototype device for cardiac diagnosis using Nadi Shastra

Similar documents
International Journal of Engineering Research-Online A Peer Reviewed International Journal Articles available online

Design & Development of a System for Nadi Pariksha

Diagnosis of Diabetes Based on Nadi Pariksha Using Tridosha Analysis and ANN

Optical sensor for Indian Siddha Diagnosis System

Wrist Energy Fuzzy Assist Cancer Engine WE-FACE

A Novel Blood Pressure-independent Arterial Wall Stiffness Parameter; Cardio-Ankle Vascular Index (CAVI)

LABVIEW based expert system for Detection of heart abnormalities

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.

Blood Pressure Determination Using Analysis of Biosignals

International Journal of Ayurveda and Pharma Research

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Keun-Ho Kim and Sang-Suk Lee* Department of Oriental Biomedical Engineering, Sangji University, Wonju , Korea

External Oscillatory Blood Pressure - EOBPTM

This article appeared in a journal published by Elsevier. The attached copy is furnished to the author for internal non-commercial research and

Department of Internal Medicine, Sakura Medical Center, School of Medicine, Toho University, Chiba, Japan

Clinical application of Arterial stiffness. pulse wave analysis pulse wave velocity

Coronary artery disease (CAD) risk factors

Special Lecture 10/28/2012

DETECTION OF HEART ABNORMALITIES USING LABVIEW

VLSI Implementation of the DWT based Arrhythmia Detection Architecture using Co- Simulation

NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE General practice Indicators for the NICE menu

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

THE IMPACT OF CCB AND RAS INHIBITOR COMBINATION THERAPY TO PREVENT CKD INCIDENCE IN HYPERTENSION AND ADVANCED ATHEROSCLEROSIS

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Keywords: Nadi parikshan, Tridosha signal, ECG electrodes, Data acquisition.

Managing cardiovascular risk with SphygmoCor XCEL

Blood Pressure Estimation Using Photoplethysmography (PPG)

Acute and Chronic Effects of Smoking on Arterial Stiffness

Two-point Method for Arterial Local Pulse Wave Velocity Measurement by Means of Ultrasonic RF Signal Processing

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG

Topic 6: Human Physiology

APPLICATION OF PHYSICAL METHODS FOR DETERMINATION OF FUNCTIONAL PARAMETERS OF ARTERIES IN RHEUMATIC PATIENTS

Arterial stiffness index: A new evaluation for arterial stiffness in elderly patients with essential hypertension

Introduction of diverse devices to measure arterial stiffness

HEART DISEASE PREDICTION BY ANALYSING VARIOUS PARAMETERS USING FUZZY LOGIC

Detection of Coronary Plaque and Sensing Risks Factors of Heart at early stages using various Image Processing and Segmentation Techniques

Temporal Analysis and Remote Monitoring of ECG Signal

Analysis of Fetal Stress Developed from Mother Stress and Classification of ECG Signals

EVALUATION OF NONINVASIVE PULSE TRANSIT TIME METHODOLOGIES FOR DIAGNOSIS OF HYPERTENSION DANIEL JOHN BADGER. A thesis submitted to the

Theoretical and practical questions in the evaluation of arterial function Miklós Illyés MD. Ph.D.

Modeling of a Ring for Blood Pressure Measurement in COMSOL Multiphysics for Medical Applications

ARTERIAL STIFFNESS ASSESSMENT COMMENTS:

ECG Signal Based Heart Disease Detection System for Telemedicine Application Using LabVIEW

Ultimate Solution. for Healthy Life. HANBYUL Overview

The Effect of Nitroglycerin on Arterial Stiffness of the Aorta and the Femoral-Tibial Arteries

Biomedical Instrumentation E. Blood Pressure

AS OCR PHYSICAL EDUCATION The Vascular System

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model

A breakthrough from Atreya Innovations

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Various Methods To Detect Respiration Rate From ECG Using LabVIEW

Association between arterial stiffness and cardiovascular risk factors in a pediatric population

ELEC ENG 4BD4 Lecture 1. Biomedical Instrumentation Instructor: Dr. Hubert de Bruin

Cardiovascular Diseases Detecting via Pulse Analysis

Research Paper Available online at: TRADITIONAL MEDICINES - KOREA, CHINA, INDIA

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

37 1 The Circulatory System

Design of Remote Monitoring System for the Detection of Vital Signs

Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable Device

Deprivation Study. The Freiburg Study

Impedance Cardiography (ICG) Application of ICG for Hypertension Management

Lab 4: Introduction to Physiological Measurements - Cardiovascular

An Approach To Develop Knowledge Modeling Framework Case Study Using Ayurvedic Medicine

Ch. 12 The Circulatory System. The heart. The heart is a double pump. A quick note on arteries vs. veins. = the muscular pump of the CV system

Automatic Detection of Heart Disease Using Discreet Wavelet Transform and Artificial Neural Network

BP-600 Noninvasive Blood Pressure Sensor

Which method is better to measure arterial stiffness; augmentation index, pulse wave velocity, carotid distensibility? 전북의대내과 김원호

pulmonary artery vasoreactivity in patients with idiopathic pulmonary arterial hypertension

Arterial Age and Shift Work

Heart Abnormality Detection Technique using PPG Signal

Article by Anette Marina Rodrigues Nursing, Texila American University, India Abstract

The Methods of Correction of Endothelial Dysfunction and Systemic Inflammation in Patients with Chronic Pyelonephritis and Hypertension

Original Contribution

boso Germany PAOD! The crucial Premium-quality risk marker for heart attack and stroke! Now PAOD screening with the boso ABI-system!

Know Your Number Aggregate Report Single Analysis Compared to National Averages

The effect of measurement position on brachial ankle pulse wave velocity

JAMES R. KNIGHT

Impedance Cardiography

The Relationship Between the Acute Changes of the Systolic Blood Pressure and the Brachial-Ankle Pulse Wave Velocity

(received 23 September 2004; accepted 18 October 2004)

PFIZER INC. These results are supplied for informational purposes only. Prescribing decisions should be made based on the approved package insert.

ECE 4600 Group Design Project PROPOSAL Right Heart Catheterization Simulator

Relationship of Atherosclerotic Risk Factors with Pulmonary Age and Vascular Age. Masao Shimizu, Asako Okano, Masaki Adachi, Yoshiaki Maruyama

Departments of Cardiology and Vascular Surgery Michaelidion Cardiac Center University of Ioannina, Greece

Lab 16. The Cardiovascular System Heart and Blood Vessels. Laboratory Objectives

Application of PPG Signal Processing to Patients undergoing Angioplasty Procedure

Cardiovascular system

Heart Rate, Blood Pressure, and Exercise. Evaluation copy

TONGUE BASED DIAGNOSIS SYSTEM USING NI-LABVIEW AND VISION ASSISTANT TOOL

Endothelial function is impaired in women who had pre-eclampsia

How Low Do We Go? Update on Hypertension

MYOCARDIALINFARCTION. By: Kendra Fischer

Conflict of Interest Slide

Blood Pressure 1 of 22 Boardworks Ltd 2011

Abstract. (Left Mouse-click to begin and move to next slide.)

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 10, April 2013

Arterial function and longevity Focus on the aorta

BIOL 219 Spring Chapters 14&15 Cardiovascular System

Cardiovascular Diseases in CKD

Transcription:

Original Article LabVIEW based Non-Invasive prototype device for cardiac diagnosis using Nadi Shastra Parth Shah 1, A.V. Leio 1, Siddaram Mahajan* 1, Shaik Jameel 2 1 Department of Biomedical Engineering, SBST, Vellore Institute of Technology, Vellore, India 2 Department of Medical Biotechnology, SBST, Vellore Institute of Technology, Vellore, India Keywords: Pulse Wave Velocity, Transit Time Delay, Cardio Ankle Vascular Index, LabVIEW. Abstract In a country like India where there is a huge population suffering from cardiovascular disorders, Atherosclerosis is one of the reasons leading to high chances of occurrence of cardiac related deaths. Hence there is a severe need for introduction of non invasive technology which can help in early diagnostics of Atherosclerosis. Hence the main aim of this research is to design a non invasive technique to detect Atherosclerosis using Nadi Shashtra. Nadi Shashtra is one of the ancient medical technologies, which was originated in India. It has been derived from Ayurveda. The instrument so designed to diagnose Atherosclerosis is based on two basic parameters i.e. Pulse Wave Velocity (PWV) and Cardio Ankle Vascular Index (CAVI). The basic workflow of this device would be non-invasive calculation of PWV and CAVI using software LabVIEW and then the values so obtained will be compared to the standard prescribed range and the chances of occurrence of Atherosclerosis would be displayed in LabVIEW. To get more accurate analysis the test is done twice, once on upper limb and once on lower limb and the mean values are considered for calculations. *Corresponding author: Siddaram Mahajan, Department of Biomedical Engineering, SBST, Vellore Institute of technology, Vellore, India. email: sddmhjn616@gmail.com Date of Submission: June 20, 2014 Date of Acceptance: July 7, 2014 Date of Publishing: Aug 1, 2014 How to cite this paper: Shah P, Leio AV, Mahajan S, Jameel S. Labview based non-invasive prototype device for cardiac diagnosis using nadi shastra. Annals of Applied Bio-Sciences. 2014;1: A16-22

A-17 Cardiac diagnosis using Nadi Shastra INTRODUCTION Ayurveda is one of the oldest forms of medical diagnosis and treatment which originated in India. The basic principle of diagnosis and treatment in it is Prakruti, and the method of diagnosis is known as Prakruti Nidana. For this, the technique used is called Nadi Parikshan. This is done on the basis of three doshas or Tridosha namely Vata, Pitta, and Kapha. [1] Nadi Parikshan is done to know about these parameters by Nadi Shashtra scholars. Nadi Parikshan involves checking the radial artery with three fingers as shown in Figure1. 120/80mmHg.Changes in blood pressure can act as a first indicator of cardiac related disorders. [3,4] Pulse Wave Velocity (PWV): Pulse Wave Velocity is one of the major parameters used to detect the extent of blockage in arteries. With the increase in deposition of cholesterol in the arteries PWV increases. [4,5] The technique used to measure PWV includes measurement of distance between two places in between which the PWV is calculated and the Transit Time Delay involved in the pressure wave to reach the second point from first point. Formula for Pulse Wave Velocity is given as: PWV = Distance/Transit Time Delay [6] A study was done to calculate the range of pulse wave velocity in both normal and diseased patient. 150 subjects were tested. The graph shown in Figure 2 shows the analysis of PWV performed. Figure 1 : Nadi- Parikshan measurement method in Ayurveda [2] To design a non-invasive device to detect Arthrosclerosis the first task would involve the study of basic blood lipid profile and blood pressure. Further there needs to be an analysis of Pulse Pressure waveform to calculate Pulse Wave Velocity which can in turn help in calculating Cardio Ankle Vascular Index. Also a study of a wide range of Indian subject both suffering from Atherosclerosis and normal subjects is required to be performed to generalize the ranges to be used for calibration of prototype device so designed. PARAMETERS TO BE CONSIDERED Blood Lipid Profile: Blood lipid profiling is used to detect the different levels of High Density Lipid (HDL), Low Density Lipid (LDL), triglycerides, cholesterol content whose increased concentrations are primary markers for Arthrosclerosis. [3,4] Blood Pressure: Blood pressure is the pressure exerted by the blood on the arteries. Normal blood pressure Figure2. Analysis of PWV study performed. The above analysis shows the PWV of normal subjects in red colour and the diseased patients in blue colour. It was found that the range of normal PWV is around 6.0 to 9.0 m/s and for diseased patients it is in the range of 10.0 to 12.0 m/s or more. Cardio-Ankle Vascular Index (CAVI): CAVI is the parameter used to measure the amount of stiffness of any artery. [4] CAVI basically depends on Pulse Wave Velocity and Blood Pressure vales. [7] CAVI is calculated by formula given by Bramwell-Hill s equation: [8] and is derived from the stiffness parameter β first proposed by a group of scientist working under Hayashi in 1980. [9,10] Annals of Applied Bio-Sciences, Vol. 1. 2014

Shah et al A-18 Figure3. Formula for CAVI [8] Where: Pulse Pressure = Systolic Blood Pressure Diastolic Blood Pressure. [4] A study was done to calculate the range of CAVI in both normal and diseased patient. 150 subjects were tested and there data for all of the above parameters were recorded and analysed. Figure 4 shows the data analysis of CAVI of the data acquired from the subjects. Time Delay is calculated by detecting the peak values of ECG waveform s R peak and the peaks from Pulse waveform. [11,12] The logic for PWV calculation is shown in Figure 5. CAVI is calculated by using the formula as stated above and LED indicators are used to check the chances of Atherosclerosis as per the ranges stated above. The logic for checking this is shown in Figure 6. The ECG and Pulse waveforms are displayed using waveform recorders the coding for which is shown in Figure 7 LabVIEW SIMULATION Due to pending approval from Human Ethic Committee the LabVIEW design was simulated by using a set of sample waveforms acquired from database available of a well known website. Case1. Normal Patient Systolic Blood Pressure: 124 mmhg Dystolic Blood Pressure: 86 mmhg Figure8. Shows the output acquired on simulation of ECG and pulse pressure waveform of the patient. Figure 4. Analysis of CAVI study performed The above analysis shows the CAVI of normal subjects in green colour and the diseased patients in blue colour. It was found that the range of normal CAVI is around 7-12 and for diseased patients it is in the range of 14-21. Hence it was decided in our studies to set 3 ranges for CAVI: a. No Atherosclerosis- CAVI below 12. b. Possible chances of Atherosclerosis- CAVI range between 12 to 14. c. Atherosclerosis- CAVI above 14. LabVIEW PROGRAMMING LOGIC LabVIEW is a graphical programming language by National Instruments. LabVIEW was selected to be used because of its real time application and for compatibility with the Data Acquisition Card of National Instruments. Electrocardiogram (ECG) and Pulse waveform signals are feed to the Lab- VIEW program which is used to calculate the PWV. The standard value as per International standards for distance between heart and radial pulse calculating site is used as 0.85m. Transit Case2. Diseased Patient Systolic Blood Pressure: 144mmHg Dystolic Blood Pressure: 92 mmhg Figure9. shows the output acquired on simulation of ECG and pulse pressure waveform of the patient. DISCUSSION: The study was done on 150 normal and diseased subjects from an age group of 20 to 60 to detect the normal and diseased range of PWV and CAVI and accordingly a virtual device was made in software LabVIEW to check for the chances of Atherosclerosis data acquired. It was found that normal range for PWV was 6-8m/sec and normal range for CAVI was around 7-12. The diseased range for PWV was found to be around 10-12m/s and CAVI was 14-21. Primary testing of the prototype design with waveforms from standard databases shows promising results. After getting clearance from Human Ethic Committee, clinical testing can be performed on human subjects, the device design can be accordingly modified to achieve more accurate results ISSN No. 2349-6991 http://www.pacificejournals.com/aabs

A-19 Cardiac diagnosis using Nadi Shastra Figure 5. Logic for calculation of Pulse Wave Velocity. Figure6. Logic for detecting chances or occurrence of Atherosclerosis. Annals of Applied Bio-Sciences, Vol. 1. 2014

Shah et al A-20 Figure 7: Logic to display the ECG and Pulse waveform Figure8. Simulation of the design for Normal Patient. ISSN No. 2349-6991 http://www.pacificejournals.com/aabs

A-21 Cardiac diagnosis using Nadi Shastra Figure 9. Simulation of the design for Diseased Patient. CONCLUSION: Only a small group of people have knowledge about Nadi Shashtra. Development of such a device to detect Atherosclerosis can form a bridge between traditional Ayurvedic knowledge and modern instrumentation techniques which can further start a new domain of possible research. Such a non-invasive device can help cardiac specialist and doctors to easily diagnose Atherosclerosis which plays an instrumental role in detection of occurrences of other severe cardiac diseases.. Acknowledgements We would like to thank Dr. Jameel Shaik for giving us an opportunity to work under him. He provided us timely excellent and continuous guidance throughout the project. He supported and suggested new, constructive ideas and the necessary changes whenever required. Without him the task of successfully completing the project would have been impossible. We would also like to thank Chief Medical Officer, Dr. Neelakantan, VIT Health Centre for helping us with data required for our project. Funding None. Competing Interests None declared. References 1. AE Kalange, BP Mahale, ST Aghav, SA Gangal, Nadi Parikshan Yantra and analysis of radial pulse, Physics and Technology of Sensors (ISPTS), 2012;165-8 2. Website: http://2.bp.blogspot.com/_osqu5ptyo6g/ SgwhZLnTvmI/AAAAAAAAASE/o9l9V4 Nvat0/s400/pulso.jpg 3. Article The surgery Article by: Amersham Health Centre, Website: http://www.amershamhealthcentre.co.uk/ re 4. Parth B. Shah, Siddaram Mahajan, A. V. Leio, Dr. Shaik Jameel, Non-Invasive Device for Cardiac Diagnosis using Nadi Shastra, National Conference on Emerging Trends in Medical Science, 7th February 2014. pp.39. 5. Article: A Clinical Guide: Pulse Wave Analysis (2005) Website: http://www.atcormedical.com/users_guide. html 6. J. McLaughlin, M. McNeill, B. Braun, and PD McCormack, Piezo-electric sensor determination of arterial pulse wave velocity, Phy- Annals of Applied Bio-Sciences, Vol. 1. 2014

Shah et al A-22 siological Measurement, vol. 24;2003: 693-702 7. Shirai K, Utino J, Otsuka K, Takata M., A novel blood pressure-independent arterial wall stiffness parameter; cardio-ankle vascular index (CAVI). J Atherosclerosis Thromb 2006; 13: 101 7. 8. Bramwell JC, Hill AV. Velocity of the pulse wave in man. Proc Roy Soc B. 1922; 93:298-306. 9. Hayashi K, Handa H, Nagasawa S, Okumura A, Moritake K. Stiffness and elastic behavior of human intracranial and extracranial arteries. J Biomech 1980;13: 175-184. 10. Nakamura K, Tomaru T, Yamamura S, Miyashita Y, Shirai K, Noike H. Cardio-Ankle Vascular Index is a Candidate Predictor of Coronary Atherosclerosis Circulation Journal. 2007;72:598-604 11. He W, Xiao H, Li S, Correia D. Instrument Design, Measurement and Analysis of Cardiovascular Dynamics Based on Lab- VIEW, Practical Applications and Solutions Using LabVIEW Software, Dr. Silviu Folea (Ed.), ISBN:2011; 978-953- 307-650-8. 12. McCarthy BM, O Flynn B, Mathewson A. An Investigation of Pulse Transit Time as a Non-Invasive Blood Pressure Measurement Method, Journal of Physics: Conference 2011;Series 307 ISSN No. 2349-6991 http://www.pacificejournals.com/aabs