Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at METAS,

Similar documents
Progress Report on Radiation Dosimetry at NPL

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

Dosimetry. Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

Absolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro

Recent Activities in Measurement Standards and Dosimetry at ARPANSA,

Recent advances in dosimetry in reference conditions for proton and light-ion beams

High-Dose Standardization Study For γ-rays radiation processing at NIM

Dosimetry and QA of proton and heavier ion beams

EDITORIAL NOTE A short CV

Applications of Monte Carlo simulations to radiation dosimetry

EDITORIAL NOTE. factor, which has been used instead of the chamber factor N D

Progress Report Department 'Dosimetry for Radiation Therapy'

The need for standardization of dosimetry in experimental radiation biology

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Dosimetry in digital mammography

Highlights from TC-IR

Protection of the contralateral breast during radiation therapy for breast cancer

Process Control Methods in Radiation Technologies

Active Scanning Beam 3 Checking Delivery/Dosimetry

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

MAY NOT BE SAME AS THE PUBLISHED VERSION

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

GEX Technical Information Report

Radiotherapy Standards Users Meeting, 1 Dec 2008 Abstracts. New calorimeters Simon Duane, NPL

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Catharine Clark NPL, Royal Surrey County Hospital and RTTQA

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Metrology Laboratory of Ionizing Radiation

3D Dosimetry with Polymer Gel

NRC Activities and Publications, Report to CCRI(I) Meeting, BIPM May

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions)

REPORT TO THE CCRI SECTION I ON THE ACTIVITY CARRIED OUT AT ENEA-INMRI ON PHOTON AND CHARGED PARTICLE DOSIMETRY IN THE PERIOD MAY MAY 2003

IORT with mobile linacs: the Italian experience

RPC s Credentialing Programs for Clinical Trials

COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE

Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom

Strengthening Activity Measurement Quality in Radiation Protection from Metrological Science to Reliable End-user Application

Publishable JRP Summary Report for JRP HLT11 MetroMRT Metrology for Molecular Radiotherapy Background

Venue: IEEE NSS/MIC/RTSD Conference, Seoul, South Korea, 27 th October 2013 Workshop: NWK3/RD1 Radiation Protection and Dosimetry

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

The determination of timer error and its role in the administration of specified doses

World-Wide Radiation Metrology: The BIPM, the CIPM MRA, and ICRU

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Progress Report on Radiation Dosimetry at NPL

Application of international standards to diagnostic radiology dosimetry

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control

The Journey of Cyberknife Commissioning

Monte Carlo for CyberKnife Incise TM MLC

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Dosimetric Consideration in Diagnostic Radiology

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10

2015 APMP Meeting TCRI Report

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

QA for Clinical Dosimetry with Emphasis on Clinical Trials

SSDL Newsletter. Contents. From the editor. No. 51 September IAEA/WHO Network of Secondary Standards Dosimetry Laboratories

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

Verification of Advanced Radiotherapy Techniques

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Application(s) of Alanine

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Slide 1. Slide 2. Slide 3. Introduction of INTRABEAM IORT. Disclosure. Contents. I have nothing to disclose.

EORTC Member Facility Questionnaire

5th ADAMAS Workshop at GSI December 15-16, 2016, Darmstadt, Germany

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects

Development of Graphite Calorimetry at the NPL for Proton Beam Therapy


EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Diode calibration for dose determination in total body irradiation

Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning

The APLMF Guide. Guide to the application of ionizing radiation metrology in. medicine under legal metrological system

High Precision Dose Delivery from Electron & X-ray Beam Lines

NRC Activities and Publications, Report to the CCRI(I) Meeting, BIPM March 2013

Transcription:

Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at, 2005-2007 1. Introduction Gerhard Stucki, Solange Gagnebin, Damian Twerenbold, Sándor Vörös - Federal Office of Metrology Lindenweg 50, 3003 Bern-Wabern SWITZERLAND The Thermometry and Ionising Radiation Section continues to maintain and develop the Swiss s of absorbed dose and air kerma and to provide calibration and verification services for therapy level dosimeters. Considerable staff changes have occurred since the previous meeting of the CCRI(I) in 2005. The former head of section, Dr Manfred Sassowsky, left on 30 April 2005. This position was taken over by Dr Damian Twerenbold on 1 June 2005. Dr Harald Quintel left in 2005 and Dr Sándor Vörös became a scientific collaborator on a permanent position the same year. Solange Gagnebin started to work on her PhD thesis at on 1 May 2006. 2. Facilities 2.1 Microtron Accelerator maintains a M22 microtron electron accelerator with two beam lines. One beam line is equipped with a conventional treatment head producing therapeutic photon and electron beams. Nine photon beam qualities in the range from TPR 20,10 = 0.639 to TPR 20,10 = 0.802 (4 MV to 21 MV) with dose rates from about 0.5 Gy/min to 4 Gy/min are regularly used for measurements with the primary sealed water calorimeter. Some other additional beam qualities are occasionally used for research purposes. Ten electron beam qualities in the range from R 50 = 1.75 gcm -2 to R 50 = 8.54 gcm -2 (5.3 MeV to 22.4 MeV) with dose rates from about 1 Gy/min to 4 Gy/min are regularly used for measurements with the primary chemical dosimeter. All beam qualities are also used for verification or calibration of reference dosimeters. The second beam line produces a narrow electron beam in the energy range from 5.3 MeV to 22.4 MeV with beam pulse currents from 100 μa to 50 ma. This beam line is mainly used for research experiments and for establishing the total absorption experiment using Fricke solution. In 2005/2006 the accelerator system was upgraded and the existing computer control system was replaced by a modern Siemens Simatic S7-400 SPS control system. Specifications and control procedures were developed by, the system itself was realized and installed by an external company. In parallel the existing analog field servo of the treatment head was replaced by a digital one (developed by ) which in turn is controlled by the new SPS system. The existing modulator which becomes more and more unstable should be replaced by two new solid sate modulators. The replacement is planned in the fall of 2007. This will also allow an extension of the energy range down to about 2.5 MeV and a variable pulse length from 1.5 to 4 μs. It should improve the stability of the accelerator system considerably. 1

2.2 60 Co treatment unit A commercially available irradiation unit of type General Electric Alcyon II is used to produce the 60 Co beam. In December 2004 the source was exchanged and the radiation field was characterized before and after changing the source by means of a set of transfer s of types NE2611A, NE2561 and NE2571. The new source has an activity of about 165 TBq (01.05.2007) resulting in a dose rate of about 0.84 Gy/min at an SCD of 1 m at a depth of 5 gcm -2 in water. The treatment unit is mainly used for the realization and dissemination of the quantity absorbed dose. 2.3 160 kv X-ray Generator A Pantak HF160c X-ray generator equipped with an X-ray tube of type Comet MXR-160 04-1.5 is used to produce therapeutic X-ray fields from 10 to 100 kv. Standard radiation qualities in the range from T10 to T100 (according to DIN 6817) are used for realization and dissemination of the quantities absorbed dose and air kerma. The 3D remotely controlled table used to support and locate the s and reference dosimeters in the radiation fields was recently upgraded. 3. and Verification Services The current range of radiation dosimetry calibration and verification services provided by is summarised in table 1, at the end of this report. The reference dosimetry systems used in the Swiss treatment centres to determine the applied dose to a patient must be regularly verified by an authorized authority. is the corresponding authority for high-energy photon (including 60 Co) and high-energy electron beams as well as for low-energy X-rays in the range from 10 to 100 kv. Reference dosimetry at the Swiss treatment centres is performed in terms of absorbed dose to water. The applied procedures for this dosimetry are described in the recommendations No 8, 9 and 10 published by the Swiss Society of Radiobiology and Medical Physics (SSRMP). The verifications provided by comply with these recommendations. Verifications provided by in the period from 01.05.2005 to 30.04.2007: High-energy photons (including 60 Co): 28 (total 66 beams, incl 18 60 Co) High-energy electrons 9 (total 47 beams) Low energy X-rays 13 (total 61 beams) The number of verifications per year is slowly increasing in all energy ranges. This is due to the increasing number of treatment centres. Recently opened centres even install low energy X-rays generators for treatment purposes and other existing centres resume the treatments in this energy range. The SSRMP is organizing a yearly TLD comparison among its members. In the frame of this comparison acts as a reference laboratory, irradiating a number of TLDs in its 60 Co beam as well as in the accelerator beams under reference conditions. 4. Dissemination of the Quantity Absorbed Dose to Water for International Customers In 2006 six dosimetry systems from the Austrian s laboratory BEV (2 systems with 2 Roos s, 2 sytems with NACP02 s and 2 systems with 2

PTW 30012 s, each system with an UNIDOS10001 electrometer) were calibrated against the primary s. The four plane parallel s were calibrated in six electron beams each and the 2 thimble s in five photon beams (including 60 Co). 5. Absorbed Dose Standards The absorbed dose primary s operated at cover the full range of the current range of radiation dosimetry calibration and verification services provided by according to table 1. 5.1 Photon Beam Dosimetry Standards The internal transfer s, a number of s of types NE2561/2611A and NE2571A, are regularly calibrated against the sealed water calorimeter in the entire energy range. The calorimeter itself was upgraded in 2006: The water tank was replaced by a new one with a Polystyrene entrance window with a thickness of 2.58 mm. A new support of the vessel and the s was built in. This new support improves the alignment of the vessel and the s. Work is also underway to replace the single AC bridge with 2 thermistors in opposite arms by 2 independent AC bridges, each equipped with a single themistor and a separate lock-in amplifier. The regularly calibrated thermistors have been proven to be stable within 0.1 % over a period of more than five years, irrespective of the accumulated dose. No changes in the calibration parameters due to an accumulated dose of more than 3000 Gray could be observed. 5.2 Electron Beam Dosimetry Standards The primary is a chemical dosimeter which is based on the total absorption of an electron beam in Fricke solution at a given well known energy. The thus calibrated Fricke solution is used to regularly calibrate plane parallel s of type NACP02 which in turn are used for the conservation and dissemination of the quantity absorbed dose. The total absorption experiment was repeated in 2007. The new results are in agreement with the results obtained in 1999/2000 to within 0.2 %. The pulse dose dependence of the radiation yield of Fricke solution containing 1 mmol/l NaCl was experimentally determined in the pulse dose range from 0.05 Gy/pulse to 600 Gy/pulse. The analysis of the experimental results is still ongoing. The internal transfer s are always calibrated against the Fricke solution just prior to using them for the yearly verifications. 5.3 Low Energy X-Rays Standards maintains five s (3 of type PTW 23342 and 2 of type PTW 23344) that are used to verificate customer dosimetry systems in the range from 10 kv to 100 kv. The qualities used are T10 to T100 according to DIN6817. In 2006 these five s were calibrated in terms of absorbed dose at the PTB against their primary. The calibration at the PTB is repeated every two years. Prior 3

to a verification of a customer dosimeter the stability of the s is checked at T30 using the 50 kv free air (FAC). 6. Air Kerma Standards Since the reference dosimetry procedures at the Swiss treatment centres are all based on absorbed dose measurements it has been decided to suspend developments of primary s for air kerma and to use s in cases where a calibration in terms of air kerma is required which is not covered by the 50 kv free air. 6.1 60 Co Standards maintains two s (1 of type NE2611A, 1 of type NE2571A). In 2006 these s were recalibrated by the BIPM. The new calibration coefficients agree within 0.3 % with the coefficients obtained in the previous calibration at BIPM in 1997. 6.2 Low Energy X-rays Standards On request offers calibrations in terms of air kerma at therapy level in the range from 10 kv to 100 kv. The qualities used are T10 to T100 according to DIN 6817. These qualities are established using the FAC (T10 to T50) or using the s traceable to PTB (T70 and T100). Prior to a calibration or verification of a customer dosimeter the stability of the s is checked at T30 using the FAC. 7. Comparisons and EUROMET Projects 7.1 EUROMET Project 605 acts as coordinator of the Project 605 Beam Quality Specification of High-Energy Photon Beams. The specification of the quality of a high-energy photon beam has been subject of several studies due to its relevance in radiation dosimetry. However, no beam quality specifier has been found that satisfies all the requirements of being unique for the entire range of photon energies used in radiotherapy and for all types of accelerators used in hospitals and laboratories, as well as being easy to measure. The aim of the project is to investigate the characteristics of the existing and widely used beam quality specifiers in more detail in order to determine the optimal beam quality specifier. A set of four cylindrical s belonging to (2 of type NE2571A, 1 of NE2561 and one of NE2611A) were calibrated in terms of absorbed dose in the 60 Co beams and in several high-energy X-ray beams at participating laboratories (ENEA, LNHB,, NPL, NRC and PTB). Beam quality measurements were performed in the accelerator beams in order to compare the efficacy of TPR 20,10 and %dd(10) x as beam quality specifiers. In total 28 accelerator beams were measured. The measurements were completed in 2006 and the analysis of the data is still ongoing. The final report is scheduled for June 2007. 7.2 EUROMET Project 813 participated in the EUROMET project 813 Comparison of air kerma and absorbed dose measurements of 60 Co radiation in radiotherapy with OMH acting as pilot laboratory. In January 2007 four s from OMH were calibrated in the 60 Co beam in terms of air kerma and absorbed dose against the Swiss s. 4

7.3 ARPANSA- Bilateral Comparison The analysis of the recent bilateral comparison of air kerma and absorbed in 60 Co radiation has been completed. The results indicate an agreement within 0.34 % (absorbed to water) and 0.48 % (air kerma) between the laboratories. The paper will be submitted for publication shortly. 7.4 -NPL High-Energy Electron Beam Comparison A comparison of absorbed dose primary s in high-energy electron beam radiation was started recently. Irradiations of four plane-parallel s of type NACP02 (2 from, 2 from NPL) in the energy range from 4 MeV to 22 MeV have been carried out at and at NPL. The experimental work is already completed and the analysis of the data is expected to be finished by June 2007. It is planned to submit the paper for publication in 2007. 8. Monte Carlo Simulations A Monte Carlo simulation project was started in the fall of 2003. Results have been obtained for several types of investigations, among which the following ones may be cited: Total absorption experiment in a Fricke solution: the energy losses due to escaping Bremsstrahlung photons, backscattering and absorption in the surrounding material have been calculated. Material corrections for the to primary comparison: small bags and ampoules of Fricke solution are used for measurement in parallel with ionization s, and small corrections for the presence of materials other than water are calculated by Monte Carlo using a full description of the treatment head. Beam quality specifiers: a detailed simulation of the treatment head was performed in order to calculate TPR 20,10 and %dd(10) for all photon beam qualities commonly used at, as well as R 50 for the electron beams. 9. Proton Beam Dosimetry Cancer therapy with proton beams has become increasingly important and is a well established method. In the framework of the ProScan project at the Paul Scherrer Institute (PSI), the spot-scanning technique is being characterized for subsequent application in hospitals. In a joint project between and PSI a modified version of the existing highenergy photon sealed water calorimeter which is in operation since 1999 is being tested for suitability as a primary for scanned proton radiation. In a first step a feasibility study has been conducted to investigate the linear energy transfer dependence of the heat defect and the influence of the time and space structure of the scanned proton beam on the homogeneity and stability of the temperature field in the water calorimeter. Simulations were validated against experimental data (taken in photon beams) of the existing calorimeter used in photon beams. In the second step a replicate of the existing calorimeter was experimentally tested in the ProScan proton test beam. The first experimental results were obtained in December 2006. Preliminary results indicate the feasibility of water calorimetry in scanned proton beams. In the third step the replicate of the existing calorimeter will be tested in the ProScan proton beam produced by the PSI gantry 2. 5

In order to use a sealed water calorimeter as a primary for scanned proton beams, the various correction factors will now be systematically investigated. 10. Publications and Reports (2003- April 2007) P J Allisy-Roberts, D Burns, G. Stucki, Comparison of the s for absorbed dose to water of the and the BIPM for 60 Co gamma radiation, Rapport BIPM 03/02 (2003) G Stucki, W Muench and H Quintel, The absorbed dose calibration service for high energy photon and electron beam radiotherapy, (Proc. Int. Symp., Vienna, 2002) IAEA Vienna (2003) IAEA-CN-96-8, Vol. I, 103-113 G Stucki, W Muench and H Quintel, The Photon-beam Primary Standard Sealed Water Calorimeter, Proceedings of a workshop on recent Advances in Absorbed Dose Standards (ARPANSA, Melbourne) http://www.arpansa.gov.au/absdos/proc.htm, 2003 J Medin, C K Ross, G Stucki, N Klassen and J P Seuntjens, Commissioning of an NRC-Type sealed water calorimeter at META using 60 Co γ-rays, Physics in Medicine and Biology 49 (2004) 4073-4086 M Sassowsky and E Pedroni, On the feasibility of water calorimetry with scanned proton radiation, Physics in Medicine and Biology 50 (2005) 5381-5400 S Gagnebin, D Twerenbold, C Bula, C Hilbes, D Meer, E Pedroni, S Zenklusen, Measurement of the absorbed dose of scanned protons with a water calorimeter, PSI Annual Report 2006 6

Table 1. Services in Photon and Electron Dosimetry or Measurement Service Quantity Instrument Type or Method Measurand Range [ Gy ] Minimum value Maximum value Measurement Conditions/Independent Variable Expanded Uncertain ty k=2 Parameter Specifications Value Standard Reference Standard used in calibration Source of traceability in a water phantom 1.0E-01 5.0E+01 Photons, high energy Tissue to phantom ratio (TPR) TPR = 0.634-0.795 1.4 Water calorimeter in a water phantom 1.0E-01 1.5E+02 Co-60 1 Water calorimeter in a water phantom 1.0E-01 1.5E+02 Electrons, high energy Half value depth R 50 =1.75-8.54 gcm -2 2 Total Absorption in Fricke solution in PMMA 2.0E+01 10 to 50 kv 0.03 HVL 1.0 3.1 PTB in PMMA 2.0E+01 50 to 100 kv 1.0 < HVL 4.7 3.1 PTB Air kerma free in air 1.0E-01 1.5E+02 Co-60 0.9 BIPM Air kerma primary free in air 5.0E-01 10 to 50 kv 0.03 < HVL 1.0 1.4 Parallel plate free air Air kerma free in air 2.0E+01 50 to 100 kv 1.0 < HVL 4.7 1.9 PTB 7