JOURNAL OF CLINICAL MICROBIOLOGY, Nov. 2000, p Vol. 38, No. 11. Copyright 2000, American Society for Microbiology. All Rights Reserved.

Size: px
Start display at page:

Download "JOURNAL OF CLINICAL MICROBIOLOGY, Nov. 2000, p Vol. 38, No. 11. Copyright 2000, American Society for Microbiology. All Rights Reserved."

Transcription

1 JOURNAL OF CLINICAL MICROBIOLOGY, Nov. 2000, p Vol. 38, No /00/$ Copyright 2000, American Society for Microbiology. All Rights Reserved. Improved Version 2.0 Qualitative and Quantitative AMPLICOR Reverse Transcription-PCR Tests for Hepatitis C Virus RNA: Calibration to International Units, Enhanced Genotype Reactivity, and Performance Characteristics SUNG C. LEE, ANISHA ANTONY, NITTA LEE, JEFF LEIBOW, JIAN Q. YANG, STEPHEN SOVIERO, KAREN GUTEKUNST, AND MAURICE ROSENSTRAUS* Roche Molecular Systems, Inc., Pleasanton, California Received 9 February 2000/Returned for modification 30 May 2000/Accepted 21 August 2000 Version 2.0 qualitative and quantitative AMPLICOR reverse transcription-pcr tests for HCV were designed to improve on the performance of first version of the hepatitis C virus (HCV) tests. The new tests were calibrated in international units, the new commonly accepted standard unit of measurement for HCV RNA. The sensitivity of the qualitative tests was enhanced by modifying the specimen processing procedure to achieve a limit of detection 50 IU/ml. The limit of detection for the quantitative tests was 600 IU/ml. Modifications to the amplification reaction mixture and thermal cycling conditions enabled all genotypes to be amplified with similar efficiency. The quantitative tests exhibited a linear range extending from 500 to 500,000 IU/ml and excellent reproducibility, with coefficients of variation ranging from 18 to 39%, within the linear range. These data indicate that the version 2.0 AMPLICOR HCV tests will improve diagnosis of HCV infection and will yield more-accurate titers for prognosis and for monitoring therapeutic efficacy, particularly at low viral loads. Furthermore, it will be possible to compare the performance characteristics and viral load measurements of AMPLICOR tests to those of other tests that adopt the international unit as the standard of measurement. Both qualitative and quantitative tests for hepatitis C virus (HCV) RNA have contributed to our understanding of the natural history of HCV infection. Quantitative tests are used to measure viral burden for patient identification and stratification (1, 4, 5, 9, 20, 25) and to provide real-time information for modifying treatment (3, 25, 33, 38). Qualitative tests are used to diagnose infection (6, 20, 22, 30) and may serve as a test of cure, where it may suffice to demonstrate that the viral RNA load falls below a critical threshold associated with long-term clinical benefit (14). The AMPLICOR products have been used successfully for diagnosis (10, 12), prognosis (11, 26, 31, 32), and monitoring therapy (13, 35). Furthermore, a proficiency study found that laboratories using the first version of the qualitative AMPLI- COR HCV and quantitative AMPLICOR HCV MONITOR tests obtained more-accurate results than laboratories using either in-house reverse transcription (RT)-PCR HCV tests or other commercially available HCV tests (16). Nevertheless, the AMPLICOR tests had two limitations. First, the limit of detection for the first version of the qualitative tests is only 1,000 copies/ml. Greater sensitivity is required, because some chronic HCV patients have viral loads below this level (11, 31) and because titers fall below this level in a substantial number of patients receiving interferon or combination interferon-ribavirin therapy (25, 36, 38). Also, as has proven true for treatment of human immunodeficiency virus type 1 (HIV-1), aggressive therapies may more effectively suppress viral load and thereby increase the likelihood of achieving a sustained response, making it important to monitor HCV-infected patients with the most-sensitive test available. The second limitation with the first version of the HCV RNA tests is that HCV genotypes 2 * Corresponding author. Present address: Roche Molecular Systems, 1080 Route 202, Somerville, NJ Phone: (908) Fax: (908) maurice.rosenstraus@roche.com. and 3 were detected less efficiently than genotype 1 (7, 19, 21). Equivalent detection and quantitation of all genotypes is required to reliably detect and monitor all HCV-positive patients and to determine whether viral load differences explain the relationship between genotype and response to interferon therapy (5, 9, 25). Version 2.0 (v2.0) AMPLICOR HCV tests were designed to overcome these two limitations. To enhance the sensitivity of the qualitative tests, the specimen processing was modified to increase the specimen volume tested 10-fold, from the equivalent of 5 to 50 l of plasma. To amplify all genotypes with equivalent efficiency, the composition of the PCR mixture and the thermal cycling parameters were modified. A limitation for all current HCV RNA assays is that it is impossible to compare assay performance and to determine which assays are most accurate and sensitive. The various assays employ different formats, have different sensitivities, and report results in different units of measurements (e.g., copies and genome equivalents). Additionally, assays that use the same unit of measurement give different results for the same sample; thus, copies measured by one assay format are not equivalent to copies measured by another assay format. Also, the viral load cut-off used to predict response to therapy is assay specific (11, 26, 31, 32). These limitations can be overcome by adopting a commonly accepted standard of measurement, the international unit. By international agreement, 100,000 IU is defined as the amount of virus in 1 ml of the First World Health Organization (WHO) International Standard for Nucleic Acid Amplification Technology Assays for HCV RNA (WHO Standard 96/790), which is distributed in vials containing 50,000 IU per vial (29). Recognizing the importance of a common unit of measurement, the version 2.0 AMPLI- COR tests are calibrated with standards containing a known quantity of international units. All results in the quantitative tests and the analytical sensitivity of the qualitative tests are 4171

2 4172 LEE ET AL. J. CLIN. MICROBIOL. now reported in international units per milliliter. As other test developers adopt the international unit as a unit of measurement, comparisons of assay sensitivity, dynamic range and other performance characteristics will be possible. Two qualitative version 2.0 AMPLICOR HCV tests are available: the fully automated COBAS AMPLICOR HCV test, v2.0, and the semiautomated AMPLICOR HCV Test, v2.0, which uses a microwell plate format. Similarly, there are two v2.0 quantitative tests: COBAS AMPLICOR HCV MONI- TOR test, v2.0, and AMPLICOR HCV MONITOR Test, v2.0. Here, we describe the procedure used to calibrate these v2.0 tests to the international unit. We evaluated the sensitivity, specificity and genotype reactivity of all four tests. In addition, we determined the linear range and precision of the quantitative tests. MATERIALS AND METHODS Clinical specimens. The high-titer genotype 1a clinical specimen used to construct the secondary standard and other clinical specimens used to evaluate the performance of the COBAS AMPLICOR and AMPLICOR v2.0 tests were obtained from various sources. For all samples, serum or plasma was separated from whole blood within 6hofcollection, aliquoted, and stored at 70 C or lower. Standardized, quantitated viral stocks. The WHO Standard 96/790, which was obtained from the National Institute for Biological Standards and Control (South Mimms, Potters Bar, Hertfordshire, England), served as the primary standard. Each vial of the WHO International Standard was reconstituted in 0.5 ml of distilled water as instructed by the accompanying insert from the National Institute for Biological Standards and Control to give a solution containing 10 5 IU/ml. Dilutions of the WHO International Standard were made with a pool of human plasma that had tested negative for HCV RNA by RT-PCR. A secondary standard was prepared by diluting a high-titer, HCV-positive (genotype 1a) plasma specimen 48-fold with a pool of human plasma that had tested negative for HCV RNA. Specimen preparation. (i) Qualitative tests. HCV RNA was extracted from 200 l of serum or plasma with 400 l of working HCV Lysis Buffer containing a known number of internal control (IC) RNA molecules. After a 10-min incubation at 60 C, the RNA was precipitated with isopropanol (600 l), recovered by microcentrifugation at maximum speed (at least 12,500 g) for 15 min at room temperature, washed with 1 ml of 70% ethanol, and resuspended in 200 l of HCV Specimen Diluent. Fifty microliters (50 l) of the processed specimen was added to Working HCV Master Mix (prepared by adding HCV manganese solution to HCV Master Mix) for the RT-PCR amplification reactions. The amplification reaction contained the HCV RNA recovered from 50 l of serum or plasma. (ii) Quantitative tests. The procedure for the quantitative tests was identical to that used for the qualitative tests with the following exceptions. The Lysis Buffer contained a known number of Quantitation Standard (QS) RNA molecules instead of IC molecules. The input sample volume was reduced to 100 l, the isopropanol volume was reduced to 500 l, and the resulting RNA pellet was resuspended in 1 ml of specimen diluent. Consequently, prepared samples for the quantitative tests were 10-fold less concentrated than prepared samples for the qualitative tests. As a result, the amplification reaction mixtures contained the HCV RNA recovered from 5 l of serum or plasma. RT-amplification. All HCV tests amplify and detect a 244-base target sequence located in a highly conserved 5 untranslated region of the HCV genome (37), defined by the primers KY78 and KY80. Only the downstream primer KY78 is biotinylated. The Working HCV master mix is a bicine-buffered solution containing glycerol, dimethyl sulfoxide, potassium acetate, deoxynucleoside triphosphates (datp, dctp, dgtp, and dutp), primer oligonucleotides, thermostable recombinant enzyme Thermus thermophilus DNA polymerase, manganese acetate, AMPERASE, and sodium azide. Processed specimens were amplified with a GeneAmp PCR system 9600 thermal cycler (Perkin-Elmer Corporation, Norwalk, Conn.) or with the integrated thermal cycler in the COBAS AMPLICOR analyzer. For the qualitative AM- PLICOR HCV test, v2.0, the thermal cycling profile consisted of 5 min at 50 C, 30 min at 62 C (for reverse transcription), 37 PCR cycles of 10 s at 95 C and 25 s at 58 C, and 2 h at 91 C. Samples may be removed for denaturation at any time during the final 2-h incubation at 91 C, which provides scheduling flexibility for laboratory personnel. The samples are held at elevated temperature to suppress any residual AMPERASE activity that could otherwise degrade amplification products; to avoid depurination of amplification products, the holding period must not exceed 2 h. The thermal cycling parameters for the quantitative AM- PLICOR HCV MONITOR Test, v2.0, were identical except that the number of PCR cycles was reduced from 37 to 32. For the COBAS AMPLICOR HCV test, v2.0, and COBAS AMPLICOR HCV MONITOR test, v2.0, the COBAS AM- PLICOR analyzer automatically performed the correct thermal cycling program after the user entered the correct test name. Hybridization and detection. The two amplification products, 244-bp sequences generated from HCV and IC (or QS) target RNA, were detected colorimetrically. Immediately upon completion of the amplification reaction, amplification products were denatured by adding 100 l of Denaturation Solution to each reaction mixture. For both the qualitative and MONITOR tests, this step was performed manually in the AMPLICOR test format or automatically in the COBAS AMPLICOR test format by the COBAS AMPLICOR analyzer. For the qualitative tests, denatured HCV and IC amplification products were detected by hybridizing 25- l aliquots of the denatured amplification product to HCV- and IC-specific probes in the presence of 100 l of Hybridization Buffer. The HCV-specific probe, KY150, is conserved among genotypes; its sequence is 5 CATAGTGGTCTGCGGAACCGGTGAGT 3. Probes were coated onto microwell plates and magnetic particles for the AMPLICOR and COBAS AM- PLICOR formats, respectively. For the MONITOR tests, HCV amplification products were quantitatively detected by hybridizing five (AMPLICOR) or four (COBAS AMPLICOR) serial dilutions of each amplification reaction mixture to the HCV-specific oligonucleotide probe. In the AMPLICOR format, these dilutions were prepared by adding 25 l of the denatured amplification product to 100 l of Hybridization Buffer and performing four, fivefold serial dilutions using Hybridization Buffer as the diluent. Similarly, QS amplification products were quantitatively detected by hybridizing three (AMPLICOR) or two (COBAS AMPLICOR) serial dilutions of each amplification reaction mixture to the QS-specific oligonucleotide probe. In the AMPLICOR format, these dilutions were prepared by adding 25 l ofthe denatured amplification product to 100 l of Hybridization Buffer and performing two fivefold dilutions using Hybridization Buffer as the diluent. In the AMPLI- COR format, the HCV- and QS-specific wells were assembled onto the same microwell plate frame. In the COBAS AMPLICOR format, the COBAS AM- PLICOR analyzer automatically performed the required dilutions and added the appropriate dilutions to tubes containing HCV- and QS-specific probe-coated magnetic particles. The qualitative and MONITOR tests used the same procedures for hybridization and colorimetric detection of the hybridized amplification products. In the AMPLICOR format, these procedures were performed as previously described (17). In the COBAS AMPLICOR format, the COBAS AMPLICOR analyzer automatically performed the hybridization and colorimetric detection steps. The absorbance of the reaction products was measured at 450 and 660 nm (single wavelength) in the AMPLICOR and COBAS AMPLICOR formats, respectively. Interpretation of results. (i) Qualitative tests. The results were interpreted by comparing the HCV and IC absorbance values to established cut-offs. The COBAS AMPLICOR analyzer performed this comparison automatically and reported the interpretation along with the actual absorbance values. For the AMPLICOR format, results were interpreted as follows. Specimens with an HCV A 450 equal to or greater than 1.0 were interpreted as positive for HCV RNA, regardless of the IC result; specimens with an HCV A 450 between 0.3 and 1.0 were interpreted as equivocal (see below). Specimens with an HCV A 450 less than 0.3 were interpreted as negative, provided the IC A 450 value was equal to or greater than 0.3. Specimens with an HCV A 450 less than 0.3 and an IC A 450 less than 0.3 were interpreted as invalid results, and a new aliquot of the specimen was tested. In the AMPLICOR format, specimens with an HCV A 450 equal to or greater than 0.3 but less than 1.0 were interpreted as equivocal for HCV RNA, regardless of the IC result. The COBAS AMPLICOR analyzer automatically interpreted specimens with weak signals (A 660 range of 0.15 to 1.0) as equivocal. To obtain an unequivocal result, these specimens were retested in duplicate. The final interpretation was positive if all three tests yielded an A 450 equal to or greater than 0.3 for AMPLICOR or an A 660 equal to or greater than 0.15 for COBAS AMPLICOR, regardless of the IC results. The final interpretation was negative if any of the three tests had A 450 values less than 0.3 for AMPLICOR or A 660 values less than 0.15 for COBAS AMPLICOR, provided that the corresponding IC absorbance values for these tests were equal to or greater than 0.3 (A 450 ) for AMPLICOR and 0.15 (A 660 ) for COBAS AMPLICOR. (ii) Quantitative tests. For the AMPLICOR HCV MONITOR Test, v2.0, the following calculations were performed manually. For both the target and QS, the dilution that yielded the lowest signal within the linear range of the colorimetric detection assay (i.e., A 450 between 0.15 and 2.0) was identified. The absorbance for this dilution was corrected by subtracting the background absorbance (A 450, 0.07), and the resulting corrected absorbance was multiplied by the dilution factor to determine the total signal generated. The input HCV RNA concentration was calculated by comparing the total HCV absorbance of the sample to the total QS absorbance for the sample, using the equation RNA (total target OD/total QS OD) QS per amplification reaction 200, where 200 is the conversion factor from international units per amplification reaction to international units/milliliter, RNA is in measured in international units per milliliter, OD is optical density, and QS is measured in international units. For the COBAS AMPLICOR HCV MONITOR test, v2.0, essentially the same calculations were performed automatically with one exception. Instead of choosing the dilution that gave the lowest A 450 between 0.15 and 2.0, the COBAS AMPLICOR analyzer identified the dilution with an A 660 between 0.10 and 2.0 that yielded the

3 VOL. 38, 2000 VERSION 2.0 AMPLICOR RT-PCR TESTS FOR HCV RNA 4173 highest total absorbance value when multiplied by its corresponding dilution factor. RESULTS Calibration to international units. In the COBAS AMPLI- COR HCV MONITOR and AMPLICOR HCV MONITOR tests, the amount of target RNA is calculated by dividing the total optical density [OD] generated from the target by the total OD generated from a known quantity of QS. This dimensionless ratio is then multiplied by the QS concentration and a factor of 200 that has units of reaction per milliliter (this is the inverse of the volume of specimen, ml, added to each reaction). Thus, the concentration of the QS must be expressed in international units per amplification reaction to report test results in international units per milliliter. To calibrate the QS in international units per reaction, we measured the WHO international standard with both the CO- BAS AMPLICOR and AMPLICOR HCV MONITOR tests. Two dilutions of the WHO international standard, which by definition contained 5,000 and 25,000 IU/ml, were tested. These dilutions were chosen for evaluation because they gave results that were well within the linear range of earlier versions of the COBAS AMPLICOR and AMPLICOR MONITOR tests, which were calibrated to report results in copies per milliliter but otherwise were identical to the tests being used. For each test format, the total OD generated from each diluted WHO standard was divided by the respective total OD generated from the QS. The amount of the WHO standard added to each reaction mixture (5,000 or 25,000 IU/ml ml/ amplification reaction) was divided by the total OD ratio to obtain the QS concentration expressed in international units per amplification reaction. QS concentration values (in international units per amplification reaction mixture) were determined by testing 5,000- and 25,000-IU/ml dilutions of the WHO international standard for HCV RNA with the COBAS AMPLICOR and AMPLICOR HCV MONITOR v2.0 tests. For each dilution, multiple sample extractions and multiple tests were performed using multiple lots of reagents. Within each test format, the results for the two dilutions were similar (geometric mean standard deviation log 10 and log 10 for 5,000 and 25,000 IU/ml, respectively, in the COBAS AMPLICOR format and log 10 and log 10 for 5,000 and 25,000 IU/ml, respectively, in the AMPLICOR format). The overall geometric mean for each format was determined by combining the results for both dilutions (n 223 and 222 for the COBAS AMPLICOR and AMPLICOR formats, respectively), and the anti-log 10 was calculated to transform the value to international units per amplification reaction mixture. The value for the AMPLICOR format, 108 IU/amplification reaction mixture, was approximately 3.3 times larger than the value for the COBAS AMPLICOR test format, 33 IU/amplification reaction mixture. The QS values (international units per amplification reaction for the two formats are different even though the quantity of QS RNA added to each amplification reaction is the same for both formats. The quantity of QS RNA was independently established by end point titration, which measures the number of amplifiable molecules (actually signal-generating units) through the use of end point dilution and Poisson statistics (27). A given number of amplifiable HCV QS RNA molecules generates an approximately fivefold higher total OD in the AMPLICOR HCV format than in the COBAS AMPLICOR format (data not shown). In contrast, the total OD generated from a given number of HCV international units is only 1.5- TABLE 1. Limit of detection for the COBAS AMPLICOR and AMPLICOR HCV version 2.0 tests a Actual input (IU/ml) % Positive (no. positive/total no. of tests) COBAS AMPLICOR format AMPLICOR format (6/6) 100 (12/12) (6/6) 100 (11/11) b (6/6) 100 (12/12) (111/111) c 100 (117/117) c (106/106) d 100 (108/108) (97/107) b 90 (96/107) b (8/12) 73 (8/11) b a Limit of detection was determined with dilutions of the WHO international standard for HCV (96/790) as described in Materials and Methods. b One reaction was excluded from the calculation because it gave a negative IC result. c Three reactions were excluded from the calculation because they gave negative IC results. d Two reactions were excluded from the calculation because they gave negative IC results. fold higher in the AMPLICOR HCV format (data not shown). Thus, the HCV target/qs total OD ratio is smaller in the AMPLICOR HCV format, which results in the larger QS international units per amplification reaction factor. By assigning format-specific international unit per amplification reaction values to the QS, we can ensure that a given test sample will yield the same titer (in international units per milliliter) in both the AMPLICOR HCV and COBAS AMPLICOR formats. Establishment of secondary standard. The supply of the WHO international standard, however, is limited. Thus, a secondary standard was created for performing routine calibrations. A high-titer, HCV-positive (genotype 1a) plasma specimen was diluted with a pool of negative plasma. Titers of the secondary standard (in international units per milliliter) were determined by testing two dilutions (40- and 200-fold) of the source plasma that were well within the linear range of the versions of the COBAS AMPLICOR and AMPLICOR MON- ITOR tests that were calibrated to report results in copies per milliliter. Multiple sample extractions and multiple tests were performed using multiple lots of reagents. For each test format, the 48 replicates at each dilution yielded approximately the same geometric mean for the source plasma. The overall geometric mean for the source plasma was determined in each test format by combining the 96 individual values for both dilutions. The corresponding concentrations of an intermediate dilution (48-fold) of the source plasma that was chosen to be the secondary standard were 26,000 and 28,000 IU/ml in the COBAS AMPLICOR and AMPLICOR formats, respectively. The titers from the COBAS AMPLICOR and AMPLICOR HCV MONITOR tests were averaged to assign a value of 27,000 IU/ml to the secondary standard. Limit of detection. The limit of detection was determined using a serial dilution series prepared from the WHO international standard for HCV (96/790). The COBAS AMPLICOR and AMPLICOR HCV v2.0 tests detected all of the samples with concentrations greater than or equal to 50 IU/ml but detected less than 95% of samples at 25 and 15 IU/ml (Table 1). Similarly, the quantitative COBAS AMPLICOR and AM- PLICOR HCV MONITOR v2.0 tests detected at least 95% of the samples with concentrations greater than or equal to 600 IU/ml (Table 2). Below 600 IU/ml, the detection rate was less than 95% and decreased as the concentration decreased (Ta-

4 4174 LEE ET AL. J. CLIN. MICROBIOL. TABLE 2. Limit of detection for the COBAS AMPLICOR and AMPLICOR HCV MONITOR version 2.0 tests a Actual input (IU/ml) % Positive (no. positive/total no. of tests) COBAS AMPLICOR AMPLICOR format format 1, (48/48) 100 (48/48) (91/96) 99 (94/95) b (95/96) 100 (96/96) (74/78) 96 (75/78) (92/102) 98 (106/108) (29/30) 89 (32/36) (17/30) 75 (27/36) a Limit of detection was determined with dilutions of the secondary standard that was calibrated against the WHO international standard for HCV (96/790) as described in Results. b One reaction was excluded from the calculation because it gave a negative IC result. ble 2). Because we define the detection limit of the test as the concentration required to yield positive results in at least 95% of replicate reactions, the detection limits for the qualitative and quantitative tests are 50 and 600 IU/ml, respectively. HCV genotype reactivity. To verify that the qualitative CO- BAS AMPLICOR and AMPLICOR HCV v2.0 tests exhibited similar sensitivity for all HCV genotypes, limiting-dilution experiments were performed for RNA transcripts representing genotypes 1a, 1b, 2a, 2b, 3a, 4, and 5. The transcripts were synthesized from cloned HCV DNA and purified by standard methods; the A 260 of each preparation was measured to provide an independent measure of the RNA concentration. The samples with 10 copies per reaction yielded positive results at least 95% of the time for all genotypes except genotype 5 (Table 3). The concentration required to yield positive results in at least 95% of replicate reactions was similar for genotypes 1a, 1b, 2a, 2b, 3a, and 4, with the AMPLICOR format exhibiting somewhat greater fluctuation between genotypes (Table 3). In both the COBAS AMPLICOR and AMPLICOR HCV formats, slightly higher concentrations of genotype 5 were required to obtain positive results 95% of the time (Table 3). Thus, the qualitative tests exhibited similar sensitivity for genotypes 1a, 1b, 2a, 2b, 3a, and 4 and were only approximately twofold less sensitive for genotype 5. To demonstrate that HCV genotypes 1a, 1b, 2a, 2b, and 3a are quantitated with equal accuracy, sets of serially diluted clinical specimens were tested by both the COBAS AMPLI- COR HCV MONITOR and AMPLICOR HCV MONITOR v2.0 tests. The concentrations tested spanned the linear range of the tests. The deviation from the expected concentration was calculated for each dilution of each genotype by subtracting the log 10 of the expected titer from the log 10 of the measured titer. At each concentration, the measured titers for each genotype were within 0.5 log 10 of the expected value; furthermore, the deviations from expected titer were similar for all genotypes (Table 4). For both the COBAS AMPLICOR and AMPLICOR HCV MONITOR v2.0 tests, the difference between the measured concentration and the expected concentration progressively decreased with increasing expected concentration (Table 4). These data indicate that the measured concentration increased somewhat less rapidly than the expected concentration, an observation consistent with results of experiments that showed the measured concentration increased linearly, but with a slope slightly less than 1.0 (see below and Fig. 1). Specificity. The COBAS AMPLICOR HCV, AMPLICOR HCV, COBAS AMPLICOR HCV MONITOR, and AMPLI- COR HCV MONITOR, v2.0, tests utilize the same primers and probe and thus should exhibit identical specificity. Analyt- Genotype TABLE 3. Genotype reactivity for the COBAS AMPLICOR and AMPLICOR HCV version 2.0 tests a Actual input (copies/pcr) a COBAS AMPLICOR format AMPLICOR HCV format % Positive b Concn 95% positive (copies/pcr) c % Positive b Concn 95% positive (copies/pcr) c 1a b a b a a Concentrated solutions of transcripts with an A 260 of 2.5, which is equivalent to copies/ml, were diluted with HCV specimen diluent to achieve final concentrations of 40, 100, and 200 copies/ml; 50 l of each dilution was added to each reaction mixture. b A total of 24 replicate tests were performed at each concentration. Three independent dilutions were prepared for each concentration, and eight replicate reactions were performed for each of the three dilutions. c The concentration that would yield positive results 95% of the time was determined by Probit analysis.

5 VOL. 38, 2000 VERSION 2.0 AMPLICOR RT-PCR TESTS FOR HCV RNA 4175 TABLE 4. Genotype reactivity of the COBAS AMPLICOR and AMPLICOR HCV MONITOR tests Format Genotype Log 10 measured concentration log 10 expected concentration a (IU/ml) b 3.5 log log log 10 COBAS AMPLICOR 1a b a b a AMPLICOR 1a b a b a a High-concentration clinical specimens were diluted to the indicated log 10 concentration (international units per milliliter). For each sample, results for six replicates were averaged and log 10 -transformed prior to calculating the difference between the measured and expected log 10 titers. The expected concentration was originally measured by performing both the COBAS AMPLICOR and AMPLICOR HCV MONITOR v2.0 tests on a sample of the original specimen that was diluted to achieve a concentration near the middle of the linear range. b For genotype 1b only, the input concentration was 5.2 log 10 IU/ml. ical specificity was evaluated for all four v2.0 tests by analyzing microbes and viruses that could potentially coinfect HCVpositive individuals as well as skin flora that could contaminate blood during collection. All of the following organisms gave positive IC (or QS) signals and HCV signals below the negative cut-off: adenovirus types 2, 3, and 7; Chlamydia trachomatis; coxsackievirus B1; cytomegalovirus; echovirus 1; Epstein-Barr virus; hepatitis A virus; hepatitis B virus; herpes simplex virus types 1 and 2; human herpes virus 6; human herpes virus 7; HIV-1, subtypes A through F; human T-cell leukemia virus type 1; human papilloma virus types 16 and 18; Propionibacterium acnes; Staphylococcus aureus; Staphylococcus epidermidis; and varicella-zoster virus (data not shown). In addition, cross-reactivity was assessed by testing a minimum of 495 samples obtained from HCV-seronegative blood donors. All of the samples gave positive IC (or QS) signals and HCV signals below the negative cut-off in the COBAS AM- PLICOR HCV, AMPLICOR HCV, COBAS AMPLICOR HCV MONITOR, and AMPLICOR HCV MONITOR, v2.0, tests (data not shown). Linear range. The linear range for quantitative tests was determined by analysis of a dilution series of an HCV-positive plasma specimen whose titer in international units per milliliter had been determined by the use of the HCV secondary standard. A set of 19 samples with concentrations ranging from 20 to 3,300,000 IU/ml of plasma was generated by diluting the HCV-positive specimen with various amounts of a pool of HCV RNA-negative human plasmas. Three individuals tested the panel in duplicate, for a total of six replicates at each virus concentration. The log 10 measured concentration for each replicate was compared to the log 10 expected concentration. For both the COBAS AMPLICOR and AMPLICOR HCV test formats, residual analysis showed that the concentration measured by either test was, on average, within 0.3 log 10 of the expected concentration for samples with measured titers ranging from 500 to 500,000 IU/ml (data not shown). Both formats yielded similar regression lines for samples within the linear range: the slopes were and 0.986, the intercepts were and 0.072, and the R 2 values were and for the COBAS AMPLICOR and AMPLICOR HCV formats, respectively (Fig. 1). Because the slopes of the regression lines were slightly less than 1.0 and the intercepts were not 0.0, the log 10 of the measured titers over the linear range differed slightly from the log 10 of the actual input titers. Agreement between test formats. A set of 51 HCV-positive clinical samples with titers ranging from 450 to 1,000,000 IU/ml were tested in parallel with the COBAS AMPLICOR and AMPLICOR HCV MONITOR v2.0 tests. Twenty-two of the 51 samples were tested undiluted, and the remaining 29 hightiter samples were diluted with HCV RNA-negative plasma prior to testing to bring their titers within the linear range of the assay. The set contained at least one sample for each of the following genotypes: 1a, 1b, 2a, 2b, 2c, 2d, 3a, 3b, 4a, 4c, 4f, 4g, 4h, 5a, and 6a (for simplicity, results for the subtypes of genotype 2, 3, and 4 are not distinguished in Fig. 2). To determine whether both tests yielded equivalent titers, we calculated the difference between the two measured titers for each sample in the set. For any one sample, the two titers may differ due to systematic error and test imprecision. If the two formats yield equivalent results, the average of these differences will approach zero (2). Furthermore, the difference between the titers will be independent of the actual titer. To determine if the data satisfied these conditions, we plotted the difference between the two titers versus the actual titer. Since the actual titer of the specimens was not determined by an independent reference test, we estimated the actual titer by averaging the results of the two formats. Because the titers of the samples varied over FIG. 1. Linear range for the COBAS AMPLICOR (A) and AMPLICOR HCV MONITOR (B) v2.0 tests. Twenty to 48 replicate tests were performed for each input viral RNA concentration. For expected RNA concentrations below 1,600 IU/mL, some of the replicates gave negative results and were excluded from the data analysis. At each expected RNA concentration, the mean of the log 10 measured RNA concentrations (circles) and the standard deviations (error bars) were determined. The closed circles, representing measured RNA concentrations between 500 and 500,000 IU/ml, show the results for concentrations used for linear regression analysis. The open circles show results for concentrations that were not included in the regression analysis. The regression analysis was performed using the averaged result for each expected RNA concentration.

6 4176 LEE ET AL. J. CLIN. MICROBIOL. 95% confidence interval were within a factor of 5 (0.7 log 10 )of the mean value (Fig. 3). The minimum and maximum values obtained for each sample were close to, or well within, the upper and lower limits of the 95% confidence interval (Fig. 3). DISCUSSION FIG. 2. Comparison of COBAS AMPLICOR HCV MONITOR test, v2.0, and AMPLICOR HCV MONITOR test (microwell plate [mwp]), v2.0, values for HCV RNA concentration in 51 clinical specimens of genotypes 1a ( ), 1b ( ), 2({),3( ),4(E), 5a ( ), and 6a ( ). For each specimen, the difference between the values reported by the COBAS AMPLICOR format and the AMPLICOR HCV (MWP) format is plotted against the average of the two values. several orders of magnitude, the measured titers were log transformed before calculating the differences and average titers. All of the individual differences were less than 0.5 log 10, with the exception of two low-titer specimens (Fig. 2). Linear regression analysis indicated that the difference between titers measured by the two formats varied only slightly with concentration; the slope of the regression line was 0.038, indicating that the difference decreased slightly as the viral titer increased (Fig. 2). On average, the difference in titers was small; regression analysis showed that the COBAS AMPLICOR format tended to yield values ranging from 0.02 to 0.13 log 10 higher than those obtained in the AMPLICOR format for samples having titers within the linear range of the assays (Fig. 2). Precision study. The total imprecision of the COBAS AM- PLICOR HCV MONITOR test, v2.0, was assessed by testing each of six samples 40 times over 10 days. Each sample was tested once each day by each of two analysts, each of whom used two different lots of reagents. The samples, which covered most of the linear range of the assay, consisted of serial dilutions prepared from two different clinical specimens. The total imprecision included between-analyst, between-reagent lot, and between-day variation. The estimated total imprecision decreased as the titer increased; the coefficient of variation (CV) ranged from 33% for the lowest-titer sample (8,600 copies/ml) to 18% for a high-titer sample (greater than 600,000 copies/ml [Fig. 3]). For each sample, the upper and lower limits of the 95% confidence interval were within a factor of 3 (0.5 log 10 ) of the mean value (Fig. 3). The minimum and maximum values obtained for each sample were close to the upper and lower limits of the 95% confidence interval (Fig. 3). The total imprecision of the AMPLICOR HCV MONITOR test, v2.0, was assessed by testing each of 16 samples 20 times over 10 days. Each sample was tested once each day by each of two analysts. The HCV RNA titers of the 16 samples ranged from 2,200 to 410,000 copies/ml, covering most of the linear range of the assay. The total imprecision included betweenanalyst and between-day variation. The imprecision for AM- PLICOR HCV MONITOR test, v2.0, was similar to that of the COBAS AMPLICOR HCV MONITOR test, v2.0, with CVs ranging from 20 to 39%, but did not vary with HCV RNA titer (Fig. 3). For each sample, the upper and lower limits of the The version 2.0 qualitative and quantitative AMPLICOR RT-PCR tests for HCV RNA were calibrated in international units, the new commonly accepted standard of measurement for HCV RNA (29). As other test developers adopt the international unit as a unit of measurement, comparisons of assay sensitivity, dynamic range, and other performance characteristics will be possible. For the COBAS AMPLICOR and AM- PLICOR HCV qualitative test formats, the limit of detection was calibrated in international units per milliliter by performing the test on dilutions of the WHO international standard, which contains a defined quantity of international units. To calibrate the two quantitative test formats to the international unit, format-specific calibration factors for the QS were calculated by performing both tests on diluted samples of the WHO international standard. The international unit value for the WHO sample was divided by the ratio of the sample total OD/QS total OD to calibrate the QS concentration in international units per amplification reaction. The WHO international standard is available in limited quantities. For routine use, a secondary standard was established by measuring the titer of an HCV-positive specimen using the QS that was calibrated against the WHO international standard. Once established, the secondary standard was then used to calibrate subsequent QS preparations in international units per amplification reaction. This value is supplied in each amplification kit, enabling the user to calculate the specimen concentration in international units per milliliter directly from the target and QS absorbance values. The assignment of format-specific values for the QS in international units per amplification reaction enables a given sample to yield similar HCV RNA titers in both test formats. Format-specific values FIG. 3. Precision profile of COBAS AMPLICOR and AMPLICOR HCV MONITOR v2.0 tests. For the COBAS AMPLICOR format, each of the six sets of values plotted represents results from 40 replicates for each of the six samples tested. For the AMPLICOR format, each of the 16 sets of values plotted represents results from 20 replicates for each of the 16 samples tested. The lower portion of the graph shows the CV ( ), and the upper portion of the graph shows the mean (E), the minimum and maximum values obtained ( ), and the 95% confidence interval (mean 2 standard deviations [ ]) for a single measurement.

7 VOL. 38, 2000 VERSION 2.0 AMPLICOR RT-PCR TESTS FOR HCV RNA 4177 are required because the QS yields a stronger signal (relative to the HCV signal) in the AMPLICOR HCV format than in the COBAS AMPLICOR format. Differences in hybridization kinetics for the HCV and QS in the two detection formats probably account for this phenomenon. Unlike the first version of the AMPLICOR HCV tests, the v2.0 tests amplified all genotypes with similar efficiency. End point dilution assays with RNA transcripts of each genotype demonstrated that the qualitative tests exhibit similar sensitivity for genotypes 1a, 1b, 2a, 2b, 3a, and 4 and slightly lower (twofold) sensitivity for genotype 5. This improved sensitivity for non-1 genotypes was confirmed in an independent end point dilution study (7). At concentrations throughout the linear range, the quantitative tests yielded titers within 0.5 log 10 of the actual titer for genotype 1a, genotype 1b, genotype 2a, genotype 2b, and genotype 3a clinical specimens. This improved genotype reactivity is attributed to a reformulated PCR mixture and modified thermal cycling parameters, which most likely promoted unfolding of secondary structures that could form at the relatively low annealing and extension temperatures employed in the first version of the tests. A recent independent evaluation of the AMPLICOR HCV MONITOR test, v2.0, also demonstrated that RNA transcripts of each genotype and a genotype 1 clinical specimen were accurately quantitated; the difference between the measured titer and the expected titer was nearly constant throughout the test s linear range (19). A genotype 2 clinical specimen and a genotype 3 clinical specimen, however, gave titers that were much lower than the expected titer when tested at concentrations near the top of the linear range (19). The difference between the measured titer and the expected titer decreased progressively and dramatically as the samples were diluted (19). The reason for this discrepancy is not clear. One possibility is that the efficiency of RT may be different for viral RNA and transcripts (19). The genotype 2 and genotype 3 specimens tested by Mellor et al. may have contained factors that reduced the efficiency of RT of the viral RNA without similarly reducing RT of the QS RNA transcript, which would cause the titer to be underestimated. This degree of underestimation would decrease as the samples were diluted, because the concentration of the hypothesized interfering factors would be reduced. Additional genotype 2 and genotype 3 specimens need to be studied to determine whether inaccurate quantitation is a specimen-specific phenomenon. The qualitative COBAS AMPLICOR and AMPLICOR HCV v2.0 tests were shown to be more sensitive than the first versions of the tests, exhibiting a limit of detection of 50 IU/ml. While these v2.0 tests were also able to detect samples with lower titers, a single result cannot reliably detect the presence of viral RNA when the titer is below 50 IU/ml; by performing replicate tests, we demonstrated that such specimens do not always yield positive results. The sensitivity of the first version of the tests is approximately 1,000 copies/ml (these tests were not evaluated using a standard calibrated in international units). Since an international unit is equivalent to approximately 0.93 to 3.1 copies as measured in the AMPLICOR HCV and COBAS AMPLICOR formats, respectively (data not shown), the v2.0 tests were approximately 6- to 22-fold more sensitive than the first versions of the tests. This is consistent with the results of a recent independent study (7) and agrees with the expected 10-fold difference given the modifications made to the specimen processing procedure: the input sample volume was increased by a factor of two and the final resuspension volume was decreased by a factor of five. The clinical performance of the qualitative version 2.0 tests has been evaluated in a recent study. Compared to enzyme immunoassay, sensitivity was 94 to 95% and specificity was 97 to 100% (Shiffer et al., unpublished data). The failure to detect HCV RNA in some serologically positive patients is expected; some patients may be infected but not viremic, and others may have resolved earlier infections. The IC showed that less than 3% of specimens were inhibitory when initially tested (Shiffer et al., unpublished data). The COBAS AMPLICOR and AMPLICOR HCV MONI- TOR v2.0 tests exhibited good reproducibility. The overall CVs ranged from 18 to 39%. For the COBAS AMPLICOR HCV MONITOR test, v2.0, lower titer specimens exhibited higher CVs, but for AMPLICOR HCV MONITOR test, v2.0, there was no apparent relationship between titer and CV. Given this level of imprecision, any one result will be within a factor of 3 to 5 (0.5 to 0.7 log 10 ) of the actual titer and a greater-thanthree (0.5 log 10, COBAS AMPLICOR)- or fivefold (0.7 log 10, AMPLICOR) change between successive samples can be considered significant. The COBAS AMPLICOR and AMPLICOR HCV MONI- TOR v2.0 tests produced a linear response for specimens with measured titers ranging from 500 to 500,000 IU of HCV RNA/ml of plasma. Measured titers at the upper end of the range were somewhat lower than, but within 0.3 log 10 of, the actual titer. Also, the measured titers leveled off rapidly above the upper end of the linear range, which increases the chance that out-of-range, high-titer specimens will be assigned a value within the linear range. Consequently, if precise titers need to be determined for samples yielding results near the top of range, the sample should be tested after it is diluted to give a titer that is well within the linear range. The two formats yielded virtually identical regression lines for specimens having titers within the linear range. This implies that the two tests can be used interchangeably, a prediction that was borne out by demonstrating that the tests gave very similar titers when individual specimens were tested in parallel. For any one specimen, the difference between the two results was generally small and within the limit predicted by the precision data; furthermore, the average difference was negligible (approximately 0.06 log 10, with COBAS AMPLICOR giving slightly higher values) for the group of specimens tested. The broad linear range of the v2.0 MONITOR tests makes them well-suited for predicting and monitoring the response to therapy. A large body of data has demonstrated that patients with lower baseline viral loads exhibit higher response rates when treated with interferon, although the positive and negative predictive values are not high enough for deciding whether to administer or withhold interferon treatment from an individual patient (5, 9, 20). Current European and U.S. guidelines recommend a cut-off of copies/ml, as determined by an in-house RT-PCR assay (8, 20). Other studies, however, have shown that there is little difference in prognosis for viral RNA titers above 10 5 to 10 6 copies/ml (11, 15, 18, 28, 31, 34). Thus, the fact that the HCV MONITOR tests underestimate viral loads above IU/ml may not be a limitation. Indeed, several studies have shown that the first version of the AM- PLICOR HCV MONITOR test and the Quantiplex HCV test, which does quantitate viral loads above 10 6 copies/ml, are equally accurate for predicting response to therapy from baseline viral load (26, 32). Future studies that use international units per milliliter as a standardized reporting format should identify a prognostic cut-off. It should be noted, however, that the prognostic value of pretreatment viral load may diminish with the availability of more potent antiviral treatment regimens since reasonably high sustained response rates can be achieved in patients with high viral loads (25). The decrease in viral load early after treatment initiation

8 4178 LEE ET AL. J. CLIN. MICROBIOL. appears to have more prognostic value than baseline viral load (11, 13, 25, 35, 36). Thus, measuring the early viral load response may prove useful for guiding changes in therapeutic regimens in individual patients (3, 25, 33, 38) and for comparing the efficacy of new treatment regimens. During interferontreatment, viral load often drops rapidly to low levels (25, 38). The ability of the HCV MONITOR tests to quantitate viral loads as low as 600 IU/ml make them ideal for measuring this early response. Less-sensitive tests such as the Quantiplex HCV test have limited value for monitoring the response to therapy, since the viral load quickly drops below their limit of detection (11, 23, 24). In summary, the v2.0 AMPLICOR HCV tests offer several advantages over their predecessors. The new qualitative CO- BAS AMPLICOR and AMPLICOR HCV v2.0 tests are approximately 10-fold more sensitive, with a detection limit of 50 IU/ml. The quantitative COBAS AMPLICOR and AMPLI- COR HCV MONITOR v2.0 tests accurately measure titers of all genotypes. Because they are calibrated in international units, the performance characteristics and viral load measurements of AMPLICOR tests can be compared to those of other tests that adopt the international unit as the standard of measurement. Given these features, the v2.0 AMPLICOR HCV tests will improve diagnosis of HCV infection and will yield more-accurate titers for prognosis and for monitoring therapeutic efficacy. ACKNOWLEDGMENTS We thank Minh Nguyen, Karen Mabee, and Kathy Mohr for their excellent technical assistance in the preparation and characterization of the secondary standard. REFERENCES 1. Berenguer, M., and T. L. Wright Hepatitis C and liver transplantation. Gut 45: Bland, J. M., and D. J. Altman Regression analysis. Lancet i: Bonetti, P., L. Chemello, C. Antona, A. Breda, P. Brosolo, P. Casarin, C. Crivellaro, G. Dona, S. Martinelli, R. Rinaldi, V. Zennaro, and M. Santonastaso, F. Urban, P. Pontisso, and A. Alberti Treatment of chronic hepatitis C with interferon-alpha by monitoring the response according to viraemia. J. Viral Hepat. 4: Charlton, M., E. Seaberg, R. Wiesner, J. Everhart, R. Zetterman, J. Lake, K. Detre, and J. Hoofnagle Predictors of patient and graft survival following liver transplantation for hepatitis C. Hepatology 28: Davis, G. L., and J. Y. N. Lau Factors predictive of a beneficial response to therapy of hepatitis C. Hepatology 26(Suppl. 1):122S 127S. 6. Dhumeaux, D., M. Doffoel, and J. P. Galmiche A French consensus conference on hepatitis C: screening and treatment. J. Hepatol. 27: Doglio, A., C. Laffont, F. X. Caroli-Bosc, P. Rochet, and J. Lefebvre Second generation of the automated Cobas Amplicor HCV assay improves sensitivity of hepatitis C virus RNA detection and yields results that are more clinically relevant. J. Clin. Microbiol. 37: EASL International Consensus Conference on Hepatitis C Consensus statement. J. Hepatol. 30: Fried, M. W Clinical applications of hepatitis C virus genotyping and quantitation. Clin. Liver Dis. 1: Gane, E. J., G. Maertens, A. Ducatteeuw, K. P. Qian, J. Y. Lau, H. Jones, E. Davies, N. V. Naoumov, and R. Williams Antibodies to hepatitis C virus envelope proteins correlate with hepatitis C viraemia after liver transplantation. Transplantation 67: Ichijo, T., A. Matsumoto, M. Kobayashi, K. Furihata, and E. Tanaka Quantitative measurement of HCV RNA in the serum: a comparison of three assays based on different principles. J. Gastroenterol. Hepatol. 12: John, M., J. Flexman, and M. A. French Hepatitis C virus-associated hepatitis following treatment of HIV-infected patients with HIV protease inhibitors: an immune restoration disease? AIDS 12: Kakumu, S., T. Aiyama, A. Okumura, K. Iwata, T. Ishikawa, and K. Yoshioka Earlier loss of hepatitis C virus RNA in interferon therapy can predict a long-term response in chronic hepatitis C. J. Gastroenterol. Hepatol. 12: Lau, D. T. Y., D. E. Kleiner, M. G. Ghany, Y. Park, P. Schmid, and J. H. Hoofnagle year follow-up after interferon-alpha therapy for chronic hepatitis C. Hepatology 28: Lau, J. Y., G. L. Davis, J. Kniffen, K. P. Qian, M. S. Urdea, C. S. Chan, M. Mizokami, P. D. Neuwald, and J. C. Wilber Significance of serum hepatitis C virus RNA levels in chronic hepatitis C. Lancet 341: Lelie, P. N., H. T. M. Cuypers, A. A. J. van Drimmelen, and W. G. V. Quint Quality assessment of hepatitis C virus nucleic acid amplification methods. Infusionsther. Transfusionsmed. 25: Loeffelholz, M. J., C. A. Lewinski, S. R. Silver, A. P. Purohit, S. A. Herman, D. A. Buonagurio, and E. A. Dragon Detection of Chlamydia trachomatis in endocervical specimens by polymerase chain reaction. J. Clin. Microbiol. 30: Manesis, E. K., C. Papaioannou, A. Gioustozi, G. Kafiri, J. Koskinas, and S. J. Hadziyannis Biochemical and virological outcome of patients with chronic hepatitis C treated with interferon alfa-2b for 6 or 12 months: a 4-year follow-up of 211 patients. Hepatology 26: Mellor, J., A. Hawkins, and P. Simmonds Genotype dependence of hepatitis C virus load measurement in commercially available quantitative assays. J. Clin. Microbiol. 37: National Institutes of Health Management of hepatitis C. NIH Consensus Statement Online 15(3):1 41. [Online.] Olmedo, E., J. Costa, F. X. Lopez-Labrador, X. Forns, S. Ampurdanes, M. D. Maluenda, M. Guilera, J. M. Sanchez-Tapias, J. Rodes, and M. T. Jimenez de Anta Comparative study of a modified competitive RT-PCR and Amplicor HCV monitor assays for quantitation of hepatitis C virus RNA in serum. J. Med. Virol. 58: Pawlotsky, J. M., I. Lonjon, C. Hezode, B. Raynard, F. Darthuy, J. Remire, C. J. Soussy, and D. Dhumeaux What strategy should be used for diagnosis of hepatitis C virus infection in clinical laboratories? Hepatology 27: Pawlotsky, J. M., M. Martinot-Peignoux, J. D. Poveda, A. Bastie, V. Le Breton, F. Darthuy, J. Remire, S. Erlinger, D. Dhumeaux, and P. Marcellin Quantification of hepatitis C virus RNA in serum by branched DNAbased signal amplification assays. J. Virol. Methods 79: Pockros, P. J., V. G. Bain, E. B. Hunter, A. Conrad, A. Balart, F. B. Hollinger, D. Albert, and the Consensus Interferon Study Group A comparison of reverse transcription-polymerase chain reaction and branched-chain DNA assays for hepatitis C virus RNA in patients receiving interferon treatment. J. Viral Hepat. 6: Poynard, T., J. McHutchison, Z. Goodman, M.-H. Ling, and J. Albrecht for the Algovirc Project Group Is an à la carte combination interferon alfa-2b plus ribavirin regimen possible for the first line treatment in patients with chronic hepatitis C? Hepatology 31: Reichard, O., G. Norkrans, A. Fryden, J. H. Braconier, A. Sonnerborg, and O. Weiland Comparison of 3 quantitative HCV RNA assays accuracy of baseline viral load to predict treatment outcome in chronic hepatitis C. Scand. J. Infect. Dis. 30: Rosenstraus, M., Z. Wang, S. Y. Chang, D. DeBonville, and J. P. Spadoro An internal control for routine diagnostic PCR: design, properties and effect on clinical performance. J. Clin. Microbiol. 36: Rumi, M., E. Del Ninno, M. L. Parravicini, R. Romeo, R. Soffredini, M. F. Donato, J. Wilber, A. Russo, and M. Colombo A prospective, randomized trial comparing lymphoblastoid to recombinant interferon alfa 2a as therapy for chronic hepatitis C. Hepatology 24: Saldanha, J., N. Lelie, and A. Heath Establishment of the first international standard for nucleic acid amplification technology (NAT) assays for HCV RNA. WHO Collaborative Study Group. Vox Sang. 76: Schneeberger, P. M., I. Keur, W. van der Vliet, K. van Hoek, H. Boswijk, A. M. van Loon, W. C. van Dijk, R. H. Kauffmann, W. Quint, and L. J. van Doorn Hepatitis C virus infections in dialysis centers in The Netherlands: a national survey by serological and molecular methods. J. Clin. Microbiol. 36: Shiratori, Y., N. Kato, O. Yokosuka, E. Hashimoto, N. Hayashi, A. Nakamura, M. Asada, H. Kuroda, H. Ohkubo, Y. Arakawa, A. Iwama, and M. Omata Quantitative assays for hepatitis C virus in serum as predictors of the long-term response to interferon. J. Hepatol. 27: Soffredini, R., M. G. Rumi, E. Del Ninno, M. L. Parravicini, A. Russo, and M. Colombo Serum levels of hepatitis C virus RNA predict nonresponse to interferon therapy: comparison of two commercial assays. J. Viral Hepat. 6: Tong, M. J., L. M. Blatt, L. T. Tong, K. Sayadzadeh, and A. Conrad Long-term retreatment in chronic hepatitis C patients who were non-responders to an initial course of interferon-alpha 2b. J. Viral Hepat. 5: Yamada, G., M. Takatani, F. Kishi, M. Takahashi, T. Doi, T. Tsuji, S. Shin, M. Tanno, M. S. Urdea, and J. A. Kolberg Efficacy of interferon alfa therapy in chronic hepatitis C patients depends primarily on hepatitis C virus RNA level. Hepatology 22: Yamakawa, Y., M. Sata, H. Suzuki, K. Tanaka, E. Tanaka, S. Noguchi, K.

Performance Characteristics of the COBAS AMPLICOR Hepatitis C Virus MONITOR Test, Version 2.0

Performance Characteristics of the COBAS AMPLICOR Hepatitis C Virus MONITOR Test, Version 2.0 Clinical Chemistry / EVALUATION OF QUANTITATIVE HEPATITIS C VIRUS ASSAY Performance Characteristics of the COBAS AMPLICOR Hepatitis C Virus MONITOR Test, Version 2.0 Maria Erali, MS, 1 Edward R. Ashwood,

More information

Performance of an automated system for quantification of hepatitis C virus RNA

Performance of an automated system for quantification of hepatitis C virus RNA Journal of Virological Methods 86 (2000) 55 60 www.elsevier.com/locate/jviromet Performance of an automated system for quantification of hepatitis C virus RNA Anne Marie Roque Afonso a, *, Josiane Didier

More information

Technical Bulletin No. 162

Technical Bulletin No. 162 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 162 cobas 6800 HCV Viral Load Assay - New Platform - June 1, 2017 Contact: Heather Habig, MLS (ASCP) CM, MB CM, 717-851-1422 Operations

More information

Standardization of Hepatitis C Virus RNA Quantification

Standardization of Hepatitis C Virus RNA Quantification Standardization of Hepatitis C Virus RNA Quantification JEAN-MICHEL PAWLOTSKY, 1,2 MAGALI BOUVIER-ALIAS, 1 CHRISTOPHE HEZODE, 3 FRANCOISE DARTHUY, 1 JOCELYNE REMIRE, 1 AND DANIEL DHUMEAUX 2,3 It was recently

More information

CALIBRATION OF ANALYTICAL STANDARDS FOR HBV-DNA, HCV-RNA AND HIV-1 RNA IN GENOME COPIES BY A REFERENCE METHOD

CALIBRATION OF ANALYTICAL STANDARDS FOR HBV-DNA, HCV-RNA AND HIV-1 RNA IN GENOME COPIES BY A REFERENCE METHOD CALIBRATION OF ANALYTICAL STANDARDS FOR HBV-DNA, HCV-RNA AND HIV-1 RNA IN GENOME COPIES BY A REFERENCE METHOD AAJ van Drimmelen, E.R. Bax and W.G.V. Quint, BioQControl (BQC), Delft Diagnostic Laboratories

More information

Technical Bulletin No. 161

Technical Bulletin No. 161 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 161 cobas 6800 HIV-1 Viral Load Assay - New Platform - June 1, 2017 Contact: Heather Habig, MLS (ASCP) CM, MB CM, 717-851-1422 Operations

More information

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR Real-Time (DNA) Real-time Exicycler 96 Rotor-Gene Q/6000 IU Mix1 Mix2 IU/μl IU/μl IU/μl IU/μl IU/μl IPC NTC C 1 Lot# 2 Freeze & thawing 1 MSDS: Material Safety Data Sheets (TaqMan Real-time ' FAM ' BHQ1

More information

CMV DNA Quantification Using an Automated Platform for Nucleic Acid Extraction and Real- time PCR Assay Set-up

CMV DNA Quantification Using an Automated Platform for Nucleic Acid Extraction and Real- time PCR Assay Set-up JCM Accepts, published online ahead of print on 11 May 2011 J. Clin. Microbiol. doi:10.1128/jcm.00721-11 Copyright 2011, American Society for Microbiology and/or the Listed Authors/Institutions. All Rights

More information

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: Alere q HIV-1/2 Detect WHO reference number: PQDx

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: Alere q HIV-1/2 Detect WHO reference number: PQDx WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: Alere q HIV-1/2 Detect WHO reference number: PQDx 0226-032-00 Alere q HIV-1/2 Detect with product codes 270110050, 270110010 and 270300001,

More information

Phosphate buffered saline (PBS) for washing the cells TE buffer (nuclease-free) ph 7.5 for use with the PrimePCR Reverse Transcription Control Assay

Phosphate buffered saline (PBS) for washing the cells TE buffer (nuclease-free) ph 7.5 for use with the PrimePCR Reverse Transcription Control Assay Catalog # Description 172-5080 SingleShot Cell Lysis Kit, 100 x 50 µl reactions 172-5081 SingleShot Cell Lysis Kit, 500 x 50 µl reactions For research purposes only. Introduction The SingleShot Cell Lysis

More information

Received 8 January 2007/Returned for modification 20 February 2007/Accepted 4 June 2007

Received 8 January 2007/Returned for modification 20 February 2007/Accepted 4 June 2007 JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 2007, p. 2529 2536 Vol. 45, No. 8 0095-1137/07/$08.00 0 doi:10.1128/jcm.00058-07 Copyright 2007, American Society for Microbiology. All Rights Reserved. Comparison

More information

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR Real-Time (DNA) Real-time Exicycler 96 Rotor-Gene Q/6000 IU Mix1 Mix2 IU/μl IU/μl IU/μl IU/μl IU/μl IPC NTC C 1 Lot# 2 Freeze & thawing 1 MSDS: Material Safety Data Sheets Real- (TaqMan time ' FAM ' BHQ1

More information

Molecular Diagnosis Future Directions

Molecular Diagnosis Future Directions Molecular Diagnosis Future Directions Philip Cunningham NSW State Reference Laboratory for HIV/AIDS & Molecular Diagnostic Medicine Laboratory, SydPath St Vincent s Hospital Sydney Update on Molecular

More information

RealLine HIV quantitative Str-Format

RealLine HIV quantitative Str-Format Instructions for use DETECTION AND QUANTIFICATION OF THE HUMAN IMMUNODEFICIENCY VIRUS RNA BY REAL TIME PCR Research Use Only (RUO) Attention! Please read the information about quantification process carefully!

More information

Roche Molecular Biochemicals Application Note No. HP 1/1999

Roche Molecular Biochemicals Application Note No. HP 1/1999 Roche Molecular Biochemicals Application Note No. HP 1/1999 Nucleic Acid Purification High Pure Viral Nucleic Acid Kit High Pure 16 System Viral Nucleic Acid Kit Efficiency of Hepatitis C Virus sample

More information

Laboratory for Clinical and Biological Studies, University of Miami Miller School of Medicine, Miami, FL, USA.

Laboratory for Clinical and Biological Studies, University of Miami Miller School of Medicine, Miami, FL, USA. 000000 00000 0000 000 00 0 bdna () 00000 0000 000 00 0 Nuclisens () 000 00 0 000000 00000 0000 000 00 0 Amplicor () Comparison of Amplicor HIV- monitor Test, NucliSens HIV- QT and bdna Versant HIV RNA

More information

Performance Characteristics of a Quantitative TaqMan Hepatitis C Virus RNA Analyte-Specific Reagent

Performance Characteristics of a Quantitative TaqMan Hepatitis C Virus RNA Analyte-Specific Reagent JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 2004, p. 3739 3746 Vol. 42, No. 8 0095-1137/04/$08.00 0 DOI: 10.1128/JCM.42.8.3739 3746.2004 Copyright 2004, American Society for Microbiology. All Rights Reserved.

More information

Prospective Multicenter Clinical Evaluation of AMPLICOR and COBAS AMPLICOR Hepatitis C Virus Tests

Prospective Multicenter Clinical Evaluation of AMPLICOR and COBAS AMPLICOR Hepatitis C Virus Tests JOURNAL OF CLINICAL MICROBIOLOGY, Nov. 2001, p. 4005 4012 Vol. 39, 11 0095-1137/01/$04.00 0 DOI: 10.1128/JCM.39.11.4005 4012.2001 Copyright 2001, American Society for Microbiology. All Rights Reserved.

More information

Quantification of HBV, HCV genotype and HIV subtype panels

Quantification of HBV, HCV genotype and HIV subtype panels Quantification of HBV, HCV genotype and HIV subtype panels Harry van Drimmelen 1,2, Wim Quint 2, Nico Lelie 3 and the international NAT study group 1. Biologicals Quality Control, 2. DDL Diagnostic Laboratory,

More information

Papers. Clinical application of the Quantiplex HCV RNA 2.0 and Amplicor HCV Monitor assays for quantifying serum hepatitis C virus RNA

Papers. Clinical application of the Quantiplex HCV RNA 2.0 and Amplicor HCV Monitor assays for quantifying serum hepatitis C virus RNA J Clin Pathol 1999;52:807 811 807 Papers Hepatobiliary Division, Department of Internal Medicine, Kaohsiung Medical College Hospital, No 100, Shih-Chuan 1st Rd, Kaohsiung, Taiwan, Republic of China M-L

More information

VQA Control SOP Version 4.0 Roche Amplicor HIV-1 DNA Test, v August 2007

VQA Control SOP Version 4.0 Roche Amplicor HIV-1 DNA Test, v August 2007 1. PRINCIPLE 1.1. The Virology Quality Assurance (VQA) Laboratory provides external cell pellet controls for use in the validation of assays that detect HIV proviral DNA. 1.2. HIV seronegative peripheral

More information

For in vitro Veterinary Diagnostics only. Kylt Rotavirus A. Real-Time RT-PCR Detection.

For in vitro Veterinary Diagnostics only. Kylt Rotavirus A. Real-Time RT-PCR Detection. For in vitro Veterinary Diagnostics only. Kylt Rotavirus A Real-Time RT-PCR Detection www.kylt.eu DIRECTION FOR USE Kylt Rotavirus A Real-Time RT-PCR Detection A. General Kylt Rotavirus A products are

More information

Introduction: Table/Figure Descriptions:

Introduction: Table/Figure Descriptions: Introduction: We have completed the analysis of your HIV RNA Validation Study. The validation plan was designed to verify the installation of an unmodified FDA-approved HIV RNA assay into your laboratory.

More information

Laboratory and Clinical Diagnosis of HCV Infection

Laboratory and Clinical Diagnosis of HCV Infection Laboratory and Clinical Diagnosis of HCV Infection Jean-Michel Pawlotsky,, MD, PhD Department of Virology (EA 3489) Henri Mondor Hospital University of Paris XII Créteil,, France I Nonspecific Liver Tests

More information

See external label 2 C-8 C 96 tests CHEMILUMINESCENCE. CMV IgG. Cat # Step (20-25 C Room temp.) Volume

See external label 2 C-8 C 96 tests CHEMILUMINESCENCE. CMV IgG. Cat # Step (20-25 C Room temp.) Volume DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: COBAS AmpliPrep/COBAS TaqMan HIV-1 Test, version 2.0 (TaqMan 48) Number: PQDx 0126-046-00 Abstract The COBAS AmpliPrep/COBAS TaqMan

More information

Monitoring of HCV RNA. Hepatitis C Requirements of Antiviral Therapy Monitoring

Monitoring of HCV RNA. Hepatitis C Requirements of Antiviral Therapy Monitoring Monitoring of HCV RNA Hepatitis C Requirements of Antiviral Therapy Monitoring Hepatitis C Requirements of Antiviral Therapy Monitoring Author: Ute Hofmann,, Konrad-Zuse-Str. 1, 07745 Jena, Germany Beatrix

More information

Trends in molecular diagnostics

Trends in molecular diagnostics Trends in molecular diagnostics Detection of target genes of interest Quantification Infectious diseases HIV Hepatitis C & B TB / MAC Cytomegalovirus Herpes simplex Varicella zoster CT/GC HPV Profiling

More information

Human Cytomegalovirus Virus (CMV) IgG ELISA Kit

Human Cytomegalovirus Virus (CMV) IgG ELISA Kit Human Cytomegalovirus Virus Catalog No: IRAPKT1410 (CMV) IgG ELISA Kit Lot No: SAMPLE INTENDED USE The CMV IgG ELISA is intended for use in evaluating a patient s serologic status to cytomegalovirus (CMV)

More information

AmpliX HBV Quantitative

AmpliX HBV Quantitative Instructions for use REAL TIME PCR DETECTION AND QUANTITATION KIT OF HEPATITIS B VIRUS DNA Research Use Only (RUO) (Lyo-format) VBD0595 96 rcs valid from May 2013 Explanation of symbols used in labeling

More information

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2 For detection and measurement of human interleukin 2 Catalog #02006 Catalog #02007 2 Plates 10 Plates Product Description The Human Interleukin 2 (IL-2) ELISA Kit is designed for the quantitative detection

More information

Strengths and Limitations of Commercial Tests for Hepatitis C Virus RNA Quantification

Strengths and Limitations of Commercial Tests for Hepatitis C Virus RNA Quantification JOURNAL OF CLINICAL MICROBIOLOGY, Jan. 2004, p. 421 425 Vol. 42, No. 1 0095-1137/04/$08.00 0 DOI: 10.1128/JCM.42.1.421 425.2004 Copyright 2004, American Society for Microbiology. All Rights Reserved. Strengths

More information

EXOTESTTM. ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids

EXOTESTTM. ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids DATA SHEET EXOTESTTM ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids INTRODUCTION Exosomes are small endosome-derived lipid nanoparticles

More information

VQA HIV DNA Control SOP Version 5.0 HIV DNA Testing 13 March 2012

VQA HIV DNA Control SOP Version 5.0 HIV DNA Testing 13 March 2012 1. PRINCIPLE 1.1. The Virology Quality Assurance (VQA) Laboratory provides external cell pellet controls for use in the validation of assays that detect HIV proviral DNA. 1.2. HIV seronegative peripheral

More information

For purification of viral DNA and RNA from a wide range of sample materials

For purification of viral DNA and RNA from a wide range of sample materials QIAamp virus kits For purification of viral DNA and RNA from a wide range of sample materials Automatable on QIAGEN s proven QIAamp Kits set the standard for purification of viral DNA and RNA. QIAamp virus

More information

Automated Quantitative Analysis of Hepatitis B Virus DNA by Using the Cobas Amplicor HBV Monitor Test

Automated Quantitative Analysis of Hepatitis B Virus DNA by Using the Cobas Amplicor HBV Monitor Test JOURNAL OF CLINICAL MICROBIOLOGY, Sept. 1999, p. 2793 2797 Vol. 37, No. 9 0095-1137/99/$04.00 0 Copyright 1999, American Society for Microbiology. All Rights Reserved. Automated Quantitative Analysis of

More information

Herpes Simplex Virus 2 IgM HSV 2 IgM

Herpes Simplex Virus 2 IgM HSV 2 IgM DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and 116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com

More information

CMV Diagnostic Strategies: Current and Future

CMV Diagnostic Strategies: Current and Future CMV Diagnostic Strategies: Current and Future Tony Mazzulli, MD, FRCPC, FACP Microbiologist-in-Chief Mount Sinai Hospital & University Health Network, Toronto Faculty/Presenter Disclosure Relationships

More information

Insulin (Porcine/Canine) ELISA

Insulin (Porcine/Canine) ELISA Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01

More information

Genotype Dependence of Hepatitis C Virus Load Measurement in Commercially Available Quantitative Assays

Genotype Dependence of Hepatitis C Virus Load Measurement in Commercially Available Quantitative Assays JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 1999, p. 2525 2532 Vol. 37, No. 8 0095-1137/99/$04.00 0 Copyright 1999, American Society for Microbiology. All Rights Reserved. Genotype Dependence of Hepatitis C

More information

See external label 96 tests HSV 2 IgA. Cat #

See external label 96 tests HSV 2 IgA. Cat # DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

RealLine HCV Qualitative Str-Format

RealLine HCV Qualitative Str-Format Instructions for use REAL TIME PCR DETECTION KIT FOR THE HEPATITIS C VIRUS RNA (HCV) Research Use Only (RUO) (Str-format) VBD0795 96 Tests valid from: October 2018 Rev05_1018_EN Page 1 of 8 Explanation

More information

Bovine Insulin ELISA

Bovine Insulin ELISA Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:

More information

RealLine HBV / HCV / HIV Str-Format

RealLine HBV / HCV / HIV Str-Format Instructions for use ASSAY KIT FOR THE QUALITATIVE AND DIFFERENTIAL DETECTION OF HEPATITIS B VIRUS DNA, HEPATITIS C VIRUS RNA, AND HUMAN IMMUNODEFICIENCY VIRUS TYPES 1 AND 2 RNA USING THE PCR/RT-PCR METHOD

More information

human Total Cathepsin B Catalog Number: DY2176

human Total Cathepsin B Catalog Number: DY2176 human Total Cathepsin B Catalog Number: DY2176 This DuoSet ELISA Development kit contains the basic components required for the development of sandwich ELISAs to measure natural and recombinant human Total

More information

Human HBcAb IgM ELISA kit

Human HBcAb IgM ELISA kit Human HBcAb IgM ELISA kit Catalog number: NR-R10163 (96 wells) The kit is designed to qualitatively detect HBcAb IgM in human serum or plasma. FOR RESEARCH USE ONLY. NOT FOR DIAGNOSTIC OR THERAPEUTIC PURPOSES

More information

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum Rat C-peptide ELISA For the quantitative determination of C-peptide in rat serum Please read carefully due to Critical Changes, e.g., see Calculation of Results. For Research Use Only. Not For Use In Diagnostic

More information

Rubella virus IgG ELISA Kit

Rubella virus IgG ELISA Kit Rubella virus IgG ELISA Kit Catalog Number KA0223 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of

More information

Bringing quantitative HBsAg to the US provider, drug development and patient network

Bringing quantitative HBsAg to the US provider, drug development and patient network Bringing quantitative HBsAg to the US provider, drug development and patient network Presenter Robert G Gish MD Adjunct Professor, Stanford University Medical Director HBV Foundation HBV Forum 3 October

More information

Diagnostic Methods of HBV and HDV infections

Diagnostic Methods of HBV and HDV infections Diagnostic Methods of HBV and HDV infections Zohreh Sharifi,ph.D Blood Transfusion Research Center, High Institute for Research and Education in Transfusion Medicine Hepatitis B-laboratory diagnosis Detection

More information

Experience with Standardisation of Blood Virology NAT. Clare Morris Division of Retrovirology National Institute for Biological Standards and Control

Experience with Standardisation of Blood Virology NAT. Clare Morris Division of Retrovirology National Institute for Biological Standards and Control Experience with Standardisation of Blood Virology NAT Clare Morris Division of Retrovirology National Institute for Biological Standards and Control Background of Blood Virology Standardisation In the

More information

HEPATITIS C VIRUS GENOTYPING IN CHRONIC HEPATITIS C PATIENTS

HEPATITIS C VIRUS GENOTYPING IN CHRONIC HEPATITIS C PATIENTS HEPATITIS C VIRUS GENOTYPING IN CHRONIC HEPATITIS C PATIENTS I. Qattan Centres for Hepatology, Royal Free & University College Medical School, London V. Emery Department of Virology, Royal Free & University

More information

Herpes Simplex Virus 2 IgG HSV 2 IgG

Herpes Simplex Virus 2 IgG HSV 2 IgG DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and 116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com

More information

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit Catalog No: E0442h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

H.Pylori IgG

H.Pylori IgG DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com

More information

H.Pylori IgG Cat # 1503Z

H.Pylori IgG Cat # 1503Z DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

DEVELOPMENT OF AN INTERNATIONAL REFERENCE PREPARATION FOR HEPATITIS D VIRUS RNA - UPDATE

DEVELOPMENT OF AN INTERNATIONAL REFERENCE PREPARATION FOR HEPATITIS D VIRUS RNA - UPDATE DEVELOPMENT OF AN INTERNATIONAL REFERENCE PREPARATION FOR HEPATITIS D VIRUS RNA - UPDATE SoGAT Clinical Diagnostics III 12-13 January 2011, London Michael Chudy Julia Kreß Micha Nübling Paul-Ehrlich-Institut

More information

Diagnostic Methods of HBV infection. Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO)

Diagnostic Methods of HBV infection. Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO) Diagnostic Methods of HBV infection Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO) Hepatitis B-laboratory diagnosis Detection of HBV infection involves

More information

Cotinine (Mouse/Rat) ELISA Kit

Cotinine (Mouse/Rat) ELISA Kit Cotinine (Mouse/Rat) ELISA Kit Catalog Number KA2264 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle

More information

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures.

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures. Rat C-peptide ELISA For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: Size: 80-CPTRT-E01 96 wells Version: May 26,

More information

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Rat Insulin ELISA For the quantitative determination of insulin in rat serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: Size: 80-INSRT-E01, E10 96 wells, 10

More information

Insulin ELISA. For the quantitative determination of insulin in serum and plasma

Insulin ELISA. For the quantitative determination of insulin in serum and plasma Insulin ELISA For the quantitative determination of insulin in serum and plasma For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United

More information

Performance of Version 2.0 of the Cobas AmpliPrep/Cobas TaqMan Real-Time PCR Assay for Hepatitis B Virus DNA Quantification

Performance of Version 2.0 of the Cobas AmpliPrep/Cobas TaqMan Real-Time PCR Assay for Hepatitis B Virus DNA Quantification JOURNAL OF CLINICAL MICROBIOLOGY, Oct. 2010, p. 3641 3647 Vol. 48, No. 10 0095-1137/10/$12.00 doi:10.1128/jcm.01306-10 Copyright 2010, American Society for Microbiology. All Rights Reserved. Performance

More information

Porcine/Canine Insulin ELISA

Porcine/Canine Insulin ELISA Porcine/Canine Insulin ELISA For the quantitative determination of insulin in porcine or canine serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research

More information

ACTG Laboratory Technologist Committee Version 2.0 ACTG Lab Man UltraSensitive Roche Monitor Test, v1.5- MWP 18 May 2004

ACTG Laboratory Technologist Committee Version 2.0 ACTG Lab Man UltraSensitive Roche Monitor Test, v1.5- MWP 18 May 2004 1. PRINCIPLE UltraSensitive Roche Monitor Test, v1.5- MWP 1.1 Name and Intended Use 1.1.1 The AMPLICOR HIV-1 MONITOR test version 1.5 is an in vitro nucleic acid amplification test for the quantitation

More information

Mel Krajden, 1 * Rainer Ziermann, 2 Asphani Khan, 1 Annie Mak, 1 Kimmy Leung, 3 David Hendricks, 2 and Lorraine Comanor 4

Mel Krajden, 1 * Rainer Ziermann, 2 Asphani Khan, 1 Annie Mak, 1 Kimmy Leung, 3 David Hendricks, 2 and Lorraine Comanor 4 JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 2002, p. 2903 2907 Vol. 40, No. 8 0095-1137/02/$04.00 0 DOI: 10.1128/JCM.40.8.2903 2907.2002 Copyright 2002, American Society for Microbiology. All Rights Reserved.

More information

ELEGANCE Chlamydia pneumoniae IgG ELISA KIT

ELEGANCE Chlamydia pneumoniae IgG ELISA KIT INTENDED USE The ELEGANCE Chlamydia pneumoniae IgG ELISA has been designed for the in vitro diagnostic measurement of anti- Chlamydia pneumoniae IgG in the screening of human serum. PRINCIPLES OF THE ELEGANCE

More information

altona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 always a drop ahead. 11/2012

altona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 always a drop ahead. 11/2012 altona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 11/2012 altona Diagnostics GmbH Moerkenstr. 12 22767 Hamburg Germany phone +49 40 548 06 76-0 fax +49 40 548 06 76-10 e-mail info@altona-diagnostics.com

More information

Mouse C-peptide ELISA

Mouse C-peptide ELISA Mouse C-peptide ELISA For the quantitative determination of C-peptide in mouse serum. For Research Use Only. Not for use in Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Preparation

More information

ACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man Coulter HIV-1 p24 ELISA May 21, 2004

ACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man Coulter HIV-1 p24 ELISA May 21, 2004 Coulter HIV p24 1. PRINCIPLE The Human Immunodeficiency Virus Type 1 (HIV-1) is recognized as the etiologic agent of acquired immunodeficiency syndrome (AIDS). The virus is transmitted by sexual contact,

More information

Rat Proinsulin ELISA

Rat Proinsulin ELISA Rat Proinsulin ELISA For the quantitative determination of proinsulin in rat serum For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-PINRT-E01 Size: 96 wells Version: June

More information

P0141 HBV 1000 copies/ml genotype reference panel

P0141 HBV 1000 copies/ml genotype reference panel P0141 HBV 1000 copies/ml genotype reference panel P0141 The kit insert contains a detailed protocol and should be read carefully before testing the run control to ensure optimal performance Table of contents

More information

HbA1c (Human) ELISA Kit

HbA1c (Human) ELISA Kit HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,

More information

Fully automated quantification of hepatitis C virus (HCV) RNA in human plasma and human serum by the COBAS AmpliPrep/COBAS TaqMan System,

Fully automated quantification of hepatitis C virus (HCV) RNA in human plasma and human serum by the COBAS AmpliPrep/COBAS TaqMan System, Journal of Clinical Virology 38 (2007) 326 333 Fully automated quantification of hepatitis C virus (HCV) RNA in human plasma and human serum by the COBAS AmpliPrep/COBAS TaqMan System, Dorothea Sizmann,

More information

Mouse C-peptide ELISA

Mouse C-peptide ELISA Mouse C-peptide ELISA For the quantitative determination of C-peptide in mouse serum. For Research Use Only. Not for use in Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Calculation

More information

TNF-alpha ELISA. For Research Use Only. Not For Use In Diagnostic Procedures.

TNF-alpha ELISA. For Research Use Only. Not For Use In Diagnostic Procedures. TNF-alpha ELISA For the quantitative determination of TNF-alpha in serum, plasma, buffered solution or cell culture medium. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number:

More information

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS altona DIAGNOSTICS Instructions for Use RealStar CMV PCR Kit 1.2 08/2017 EN RealStar RealStar CMV PCR Kit 1.2 For research use only! (RUO) 021202 INS-021200-EN-S01 48 08 2017 altona Diagnostics GmbH Mörkenstr.

More information

Micropathology Ltd. University of Warwick Science Park, Venture Centre, Sir William Lyons Road, Coventry CV4 7EZ

Micropathology Ltd. University of Warwick Science Park, Venture Centre, Sir William Lyons Road, Coventry CV4 7EZ www.micropathology.com info@micropathology.com Micropathology Ltd Tel 24hrs: +44 (0) 24-76 323222 Fax / Ans: +44 (0) 24-76 - 323333 University of Warwick Science Park, Venture Centre, Sir William Lyons

More information

Comparison of Quantitative HCV RNA Assays in Chronic Hepatitis C

Comparison of Quantitative HCV RNA Assays in Chronic Hepatitis C MICROBIOLOGY AND INFECTIOUS DISEASE Original Article Comparison of Quantitative HCV RNA Assays in Chronic Hepatitis C SHERAJ JACOB, MD, DEBORAH BAUDY, ELIZABETH JONES, LIZHE XU, PhD, ANDREW MASON, MBBS,

More information

DIAGNOSTIC AUTOMATION, INC.

DIAGNOSTIC AUTOMATION, INC. DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric)

EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric) EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric) Base Catalog # PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE Uses: The EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric)

More information

Mouse GLP-2 EIA. Cat. No. KT-374. For the quantitative determination of GLP-2 in mouse serum or plasma. For Research Use Only. 1 Rev.

Mouse GLP-2 EIA. Cat. No. KT-374. For the quantitative determination of GLP-2 in mouse serum or plasma. For Research Use Only. 1 Rev. Mouse GLP-2 EIA For the quantitative determination of GLP-2 in mouse serum or plasma. Cat. No. KT-374 For Research Use Only. 1 Rev. 11357374 PRODUCT INFORMATION Mouse GLP-2 EIA Cat. No. KT-374 INTENDED

More information

See external label 2 C-8 C 96 tests Chemiluminescence. CMV IgM. Cat # Diluted samples, controls & calibrator 100 µl 30 minutes

See external label 2 C-8 C 96 tests Chemiluminescence. CMV IgM. Cat # Diluted samples, controls & calibrator 100 µl 30 minutes DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

Non-competitive Internal Control Concept for PCR-based Qualitative Assays

Non-competitive Internal Control Concept for PCR-based Qualitative Assays Christian O. Simon Roche Molecular Diagnostics Rotkreuz, CH Non-competitive Internal Control Concept for PCR-based Qualitative Assays 23 rd SoGAT Meeting, 16 th 17 th April, 2012, Vilnius, Lithuania Design

More information

HIV-1 Viral Load Real Time (RG)

HIV-1 Viral Load Real Time (RG) -1 Viral Load Real Time (RG) Real Time RT-PCR type 1 RNA quantification assay MSP Reg. pending Valdense 3616. 11700. Montevideo. Uruguay. phone (598) 2 336 83 01. Fax (598) 2 336 71 60. Info@atgen.com.uy

More information

Insulin ELISA. For the quantitative determination of insulin in serum and plasma.

Insulin ELISA. For the quantitative determination of insulin in serum and plasma. Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United

More information

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet Interleukin5 (IL5) is a secreted glycoprotein that belongs to the α-helical group of cytokines (1, 2, 3). Unlike other family members, it is present as a covalently linked antiparallel dimer (4, 5). IL-5

More information

Hepatitis C Virus Genotype 2 may not be detected by the Cobas AmpliPrep/Cobas. TaqMan HCV Test, version 1.0

Hepatitis C Virus Genotype 2 may not be detected by the Cobas AmpliPrep/Cobas. TaqMan HCV Test, version 1.0 JCM Accepts, published online ahead of print on 25 September 2013 J. Clin. Microbiol. doi:10.1128/jcm.02102-13 Copyright 2013, American Society for Microbiology. All Rights Reserved. 1 2 Hepatitis C Virus

More information

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP)

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) European Medicines Agency Pre-authorisation Evaluation of Medicines for Human Use London, 21 September 2006 EMEA/CHMP/BWP/298390/2005 COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) GUIDELINE ON

More information

Mouse Ultrasensitive Insulin ELISA

Mouse Ultrasensitive Insulin ELISA Mouse Ultrasensitive Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use

More information

Human HIV (1+2) antigen&antibody ELISA Kit

Human HIV (1+2) antigen&antibody ELISA Kit Human HIV (1+2) antigen&antibody ELISA Kit Catalog Number. CSB-E18042h For the qualitative determination of human HIV (1+2) antibody and P24 antigen concentrations in serum, plasma. This package insert

More information

Triiodothyronine (T3) ELISA

Triiodothyronine (T3) ELISA For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. Triiodothyronine (T3) ELISA Kit is intended for the detection of total T3 in human serum or plasma. For research

More information

Procine sphingomyelin ELISA Kit

Procine sphingomyelin ELISA Kit Procine sphingomyelin ELISA Kit For the quantitative in vitro determination of Procine sphingomyelin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: VERSANT HIV-1 RNA 1.0 Assay (kpcr) Number: PQDx

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: VERSANT HIV-1 RNA 1.0 Assay (kpcr) Number: PQDx WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: VERSANT HIV-1 RNA 1.0 Assay (kpcr) Number: PQDx 0115-041-00 Abstract The VERSANT HIV-1 RNA 1.0 Assay (kpcr) with product codes 10375763,

More information

Comparison of Quantitative and Qualitative PCR Assays for Cytomegalovirus DNA in Plasma

Comparison of Quantitative and Qualitative PCR Assays for Cytomegalovirus DNA in Plasma JOURNAL OF CLINICAL MICROBIOLOGY, Apr. 2001, p. 1334 1338 Vol. 39, No. 4 0095-1137/01/$04.00 0 DOI: 10.1128/JCM.39.4.1334 1338.2001 Copyright 2001, American Society for Microbiology. All Rights Reserved.

More information

Apolipoprotein A1 (Apo A1) ELISA

Apolipoprotein A1 (Apo A1) ELISA Package Insert Apolipoprotein A1 (Apo A1) ELISA 1 x 96 Wells For Research Use Only (RUO). Not for use in clinical, diagnostic or therapeutic procedures. v. 1.0 Eagle Biosciences, Inc. 20A Northwest Blvd.,

More information

RealLine HCV quantitative Str-Format

RealLine HCV quantitative Str-Format Instructions for use QUANTITATIVE KIT FOR REAL TIME PCR DETECTION FOR RNA OF HEPATITIS C VIRUS Attention! Please read the information about quantification process carefully! Research Use Only (RUO) (Str

More information

Antiviral Therapy 2014; 19: (doi: /IMP2723)

Antiviral Therapy 2014; 19: (doi: /IMP2723) Antiviral Therapy 214; 19:449 454 (doi: 1.3851/IMP2723) Original article Performance characteristics of the COBAS Ampliprep/COBAS TaqMan v2. and the Abbott RealTime hepatitis C assays implications for

More information

RealLine HIV qualitative Str-Format

RealLine HIV qualitative Str-Format Instructions for Use REAL TIME PCR DETECTION KIT FOR HUMAN IMMUNODEFICIENCY VIRUS RNA Research Use Only (RUO) RealLine HIV Qualitative (Str-format) VBD0196 96 Tests valid from July 2016 Rev01072016_EN

More information

W ith an estimated million people worldwide

W ith an estimated million people worldwide 141 ORIGINAL ARTICLE Clinical evaluation of the COBAS Amplicor HBV monitor test for measuring serum HBV DNA and comparison with the Quantiplex branched DNA signal amplification assay in Taiwan C-Y Dai,

More information