1 Stability of Tumor Necrosis Factor, Interleukin 6, and Interleukin 8 in Blood Samples of Patients With Systemic Immune Activation Astrid Friebe, MD; Hans-Dieter Volk, MD Context. Tumor necrosis factor, interleukin 6, and interleukin 8 serum/plasma levels are frequently used for the monitoring of patients with systemic immune activation/ sepsis. This requires comparability of test results over time. However, cytokines are usually not considered to be very stable after blood collection, which might artificially interfere with test results. Objective. To obtain better knowledge about stability of these cytokines in blood samples for interpretation of test results. Design. Blood of patients with systemic immune activation was collected in EDTA, lithium heparin, ammonium heparin, and serum tubes. Aliquots were analyzed after storage at room temperature for 2 to 8 hours. Additionally, storage conditions for separated serum/plasma for 24 hours and the reproducibility of repeated cytokine measurements by an automated DPC Immulite analyzer were tested. Diagnostic analysis of cytokine levels in serum or plasma has become an important issue for several disease conditions. Especially in cases of systemic inflammation/sepsis, several cytokines, such as tumor necrosis factor (TNF), interleukin (IL) 8, and IL-6, proved to be of diagnostic and prognostic value and are frequently used for the monitoring of septic patients. 1 7 Serum/plasma concentrations are usually measured by enzymelinked immunosorbent assay and automated analyzers are best suited for routine diagnostics. Nevertheless, cytokine measurement can be associated with some problems. Generally, results are only comparable if obtained by the same enzyme-linked immunosorbent assay test resulting from different epitope specificities of utilized antibodies. However, even when the same test is used, cytokine levels are usually regarded as being very susceptible and results might easily be influenced by the period between blood Results. Tumor necrosis factor level was stable in EDTA plasma for 8 hours, while slightly increasing in heparin plasma and serum. Interleukin 6 concentrations were stable for 8 hours in all blood types, whereas interleukin 8 concentrations were stable only in EDTA plasma and were strongly increasing in heparin plasma and serum. Cytokine concentrations in separated serum/plasma were stable during 24 hours if stored at 4 C or frozen at 20 or 70 C. Reproducibility of repeated cytokine measurements revealed no significant differences for all blood types. Conclusions. Cytokine levels were most critically influenced by the period between blood collection and plasma separation, but its impact was strongly dependent on cytokine and anticoagulant. However, under appropriate conditions cytokine levels were surprisingly stable for up to 8 hours. (Arch Pathol Lab Med. 2008;132: ) collection and arrival in the laboratory. 8,9 Thus, separation of cells and plasma as quickly as possible is usually recommended, which is often difficult in clinical practice. Therefore, further knowledge about cytokine stability in blood samples is important for interpretation of test results, especially as a close cytokine monitoring requires comparability of values obtained at different time points. The aim of this study was to investigate the stability of TNF-, IL-8, and IL-6 in blood samples. We investigated the influence of several periods ranging from 2 to 8 hours between blood collection and sample analysis. Additionally, we tested different storage conditions for serum/ plasma for 24 hours, in case samples could not be analyzed the same day. Finally, we investigated the reproducibility of repeated cytokine measurements by an automated DPC Immulite analyzer. All tests were performed with blood anticoagulated by lithium heparin, ammonium heparin, EDTA, and serum. Accepted for publication May 8, From the Institute of Medical Immunology, Charité-Universitätsmedizin Berlin, Berlin, Germany. Dr Friebe is now at the Center for Internal Medicine and Dermatology, Biomedizinisches Forschungshaus, Charité-Universitätsmedizin Berlin, Berlin, Germany. The authors have no relevant financial interest in the products or companies described in this article. Reprints: Astrid Friebe, MD, Center for Internal Medicine and Dermatology, Charité-Universitätsmedizin Berlin, Biomedizinisches Forschungshaus, Augustenburger Platz 1, Berlin, Germany ( astrid.friebe@charite.de). MATERIALS AND METHODS Patients Blood was taken from 24 patients with mean age of years (range, years) consisting of 15 men and 9 women after getting informed consent. Patients were recruited from an intensive care unit and all patients suffered from systemic immune activation, either sepsis or noninfective postsurgical systemic immune response syndrome. The study was approved by the local ethics committee Arch Pathol Lab Med Vol 132, November 2008 Stability of TNF-, IL-6, and IL-8 in Blood Samples Friebe & Volk

2 Figure 1. Stability of tumor necrosis factor (TNF), interleukin (IL) 6, and IL-8 in whole blood. A, C, and E show mean cytokine levels measured 2, 4, and 8 hours after blood collection. Cytokines were measured in picograms per milliliter and are shown in percentage of baseline values, which were obtained immediately after blood collection. B, D, and F provide additional information concerning distribution of single patient values. Li. heparin indicates lithium heparin; Am. heparin, ammonium heparin; and Pat., patient. Sample Preparation Blood was collected in serum tubes and vacutainers containing EDTA, lithium heparin, and ammonium heparin as anticoagulants (Sarstedt, Nümbrecht, Germany). Testing of Cytokine Stability in Whole Blood Samples. Whole blood was transferred to Eppendorf tubes and 1 aliquot was immediately analyzed to obtain baseline values. Three aliquots remained at room temperature for 2, 4, and 8 hours. Cytokines were measured in serum/plasma obtained after centrifugation at 250g for 10 minutes at room temperature. Testing of Cytokine Stability in Separated Serum/Plasma. Serum/plasma was obtained immediately after blood collection. Samples were kept for 1 hour at room temperature, 24 hours at 4 C, 24 hours at 20 C, or 24 hours at 70 C, and cytokine concentrations were compared with baseline levels. Reproducibility of Repeated Cytokine Measurements by an Automated DPC Immulite Analyzer. Blood was collected from 5 patients in serum, lithium heparin, ammonium heparin, and EDTA tubes. Cytokine levels were determined in serum/ plasma 10 times for each blood type and for each patient. Arch Pathol Lab Med Vol 132, November 2008 Stability of TNF-, IL-6, and IL-8 in Blood Samples Friebe & Volk 1803

3 Figure 2. Stability of tumor necrosis factor (TNF), interleukin (IL) 6, and IL-8 in serum/plasma. A, C, and E show mean cytokine levels measured in serum/plasma, which was stored for 1 hour at room temperature (RT), 24 hours at 4 C, 24 hours at 20 C, and 24 hours at 70 C. Cytokines were measured in picograms per milliliter and are shown in percentage of baseline values, which were obtained immediately after blood collection. B, D, and F show distribution of single patient values. Li. heparin indicates lithium heparin; Am. heparin, ammonium heparin; and Pat., patient. Assays for Cytokine Detection Immulite TNF- assay ( pg/ml; DPC Biermann, Bad Nauheim, Germany), Immulite IL-6 assay ( pg/ml; DPC Biermann), and Immulite IL-8 assay ( pg/ml; DPC Biermann) were used. Tests were used with the IMMULITE Automated Analyzer (DPC Biermann). Statistical Methods All statistical analyses were performed by SPSS 12.0 for Windows (SPSS Inc, Chicago, Ill). The limit of significance was set at.05. Cytokine levels were normally distributed (Kolmogorov- Smirnov test). The effect of time for cytokine stability and the effect of storage conditions on cytokine levels were tested for significance by repeated-measures analysis of variance (ANOVA). For analysis of test reproducibility the coefficient of variation (CV) was calculated for each test series: CV (%) (SD 100)/ mean. Coefficients of variation were tested for significant differences between blood types by ANOVA. RESULTS Stability of TNF-, IL-6, and IL-8 in Whole Blood Baseline cytokine levels were determined in serum/ plasma immediately after blood collection. The remaining 1804 Arch Pathol Lab Med Vol 132, November 2008 Stability of TNF-, IL-6, and IL-8 in Blood Samples Friebe & Volk

4 Figure 3. Reproducibility of repeated tumor necrosis factor (TNF), interleukin (IL) 6, and IL-8 measurements. A, Mean coefficient of variation obtained by 5 test series for each blood type for TNF-, IL-6, and IL-8. B, Distribution of single coefficients of variation from each test series. Li. hep. indicates lithium heparin; Am. hep., ammonium heparin. blood was kept unchanged at room temperature for 2, 4, and 8 hours, respectively, before cytokine levels were measured. Results for TNF- showed a stable concentration if measured in EDTA plasma for up to 8 hours. In heparin plasma and especially in serum, values were slowly but continually increasing, although not reaching statistical significance, and exceeded the initial concentration after 8 hours by approximately 25% (Figure 1, A and B). Interleukin 6 concentrations turned out to be stable for all tested blood types and corresponded even after 8 hours with the initial values (Figure 1, C and D). In EDTA plasma, extreme outliers were repeatedly measured due to a disturbance of the test system caused by the presence of EDTA. Therefore, we excluded these results from further analysis. Meanwhile, this problem has been solved by the manufacturer and the IL-6 Immulite assay is now suitable for EDTA too. Interleukin 8 concentrations were stable only in EDTA plasma. In heparin plasma and even more in serum, values were continually and strongly increasing over time exceeding initial values by approximately 5 times for heparin plasma and 20 times for serum after 8 hours (Figure 1, E and F). This increase reached levels of significance as tested by ANOVA (lithium heparin: n 8, F , P.01; ammonium heparin: n 8, F , P.009; serum: n 8, F , P.01). Stability of TNF-, IL-6, and IL-8 in Serum/Plasma Under Different Storage Conditions We aimed to determine optimal storage conditions for serum/plasma during 24 hours in case an immediate cytokine analysis is not possible. Blood was centrifuged immediately after collection and cytokine baseline levels were determined. Remaining serum/plasma aliquots were stored for 1 hour at room temperature, 24 hours at 4 C, 24 hours at 20 C, and 24 hours at 70 C. Results showed that cytokine concentrations corresponded with the initial values obtained immediately after blood collection for all storage conditions, which was independent if cytokines were measured in EDTA, lithium heparin, ammonium heparin plasma, or serum (Figure 2, A through F). Reproducibility of Repeated TNF-, IL-6, and IL-8 Measurements Cytokines were measured 10 times in EDTA, lithium heparin, ammonium heparin plasma, and serum from 5 patients. The coefficient of variation (CV [SD 100]/ mean) was calculated for each test series and finally, the mean resulting from 5 test series was determined for each blood type. Repeated measurement of TNF- resulted in a comparable amount of variation between test results for EDTA (CV 5.7%), lithium heparin (CV 6.3%), ammonium heparin plasma (CV 6.2%), and serum (CV 6.1%). There were also no statistically significant differences between coefficients of variations for IL-6 measured in lithium heparin (CV 10%), ammonium heparin plasma (CV 6.3%), and serum (CV 7.1%) and no significant differences for IL-8 in EDTA (CV 4.0%), lithium heparin (CV 6.2%), ammonium heparin plasma (CV 4.4%), and serum (CV 5%) (Figure 3, A and B). COMMENT The aim of this study was to investigate the stability of TNF-, IL-6, and IL-8 in blood samples of patients with systemic immune activation. We investigated the influence of several time periods ranging from 2 to 8 hours between blood collection and sample analysis. TNF- was most stable in EDTA plasma, while slightly increasing in heparin plasma and even more in serum, which is probably due to spontaneous TNF- production by monocytes in response to syringe material. IL-6 concentrations were stable in all tested blood types. This is in line with the results of Kenis et al 8 who showed that IL-6 serum levels were stable for up to 4 days of clotting at 4 C and20 C. However, Thavasu et al 9 reported that TNF- as well as IL-6 concentrations were continually declining in EDTA, heparin plasma, and serum during 12 hours. These differences might result, in addition to a different enzyme-linked immunosorbent assay test that had been used, from the fact that tests were performed with blood from healthy volunteers supplemented with recombinant cytokines (as healthy subjects often do not have detectable levels of cytokines) and not from patients with an activated immune system, which might result in ongoing release of cytokines after blood collection. Interleukin 8 concentrations were stable only in EDTA plasma and were strongly increasing in heparin plasma Arch Pathol Lab Med Vol 132, November 2008 Stability of TNF-, IL-6, and IL-8 in Blood Samples Friebe & Volk 1805

5 and serum probably due to a release of IL-8 by erythrocytes. This resulted in a 5-fold increase of IL-8 levels in heparin plasma and a nearly 20-fold increase in serum during 8 hours. Thus, when comparing cytokine levels of healthy controls and systemic immune response syndrome patients as shown for example by Rodríguez-Gaspar et al, 10 IL-8 concentrations can be affected by specimen handling in a way that might lead to erroneous clinical impressions. In summary, EDTA plasma seems to be most suitable for stability reasons, but unexpected problems can occur due to an interference of EDTA with the test system. Serum seems to be most critical regarding stability and should be avoided for routine parameter analysis. If heparin plasma is used, plasma should be separated as quickly as possible within 4 hours after blood collection. Cytokine concentrations in separated serum/plasma seem to be very stable and can be stored for 24 hours at 4 C or can be frozen by 20 C or 70 C. Storage conditions for IL-6 were also investigated in the study by Kenis et al 8 and IL-6 was stable in serum for 21 days at room temperature (20 C and 30 C) and probably for several years when stored at 20 C. Regarding the reproducibility of repeated cytokine measurements by the DPC Immulite analyzer, there were no significant differences between blood types for TNF-, IL-6, and IL-8. Cytokine levels were most critically influenced by the period between blood collection and plasma separation, but its impact was strongly dependent on cytokine and anticoagulant. Send-out laboratories try to solve this problem by accepting only frozen serum/plasma samples and recommend in their user guidelines separation from cells within 30 minutes. However, because in daily routine this period might easily be exceeded, cytokine assay validation should particularly include careful attention to specimen type, either if samples are sent out or if readers wish to measure cytokines in their own laboratories. Nevertheless, if appropriate conditions are applied, cytokine levels were surprisingly stable for up to 8 hours. References 1. Volk H, Reinke P, Doecke WD. Immunological monitoring of the inflammatory process: which variables? When to assess? Eur J Surg Suppl. 1999;584: Kox W, Volk T, Kox S, Volk H. Immunomodulatory therapy in sepsis. Intensive Care Med. 2000;26: Adamik B, Zimecki M, Wlaszczyk A, Kubler A. Immunological status of septic and trauma patients: high tumor necrosis factor alpha serum levels in septic and trauma patients are not responsible for increased mortality; a prognostic value of serum interleukin 6. Arch Immunol Ther Exp. 1997;45(2 3): Oda S, Hirasawa H, Shiga H, Nakanishi K, Matsuda K, Nakamua M. Sequential measurement of IL-6 blood levels in patients with systemic inflammatory response syndrome (SIRS)/sepsis. Cytokine. 2005;29(4): Cojocaru IM, Musuroi C, Iacob S, Cojocaru M. Investigation of TNF-alpha, IL-6, IL-8 and of procalcitonin in patients with neurologic complications in sepsis. Rom J Intern Med. 2003;41(1): Perenvi A, Johann-Liang R, Stavola JJ. Assessment of cord blood IL-6 levels as an indicator of neonatal sepsis. Am J Perinatol. 1999;16: Horisberger T, Harbarth S, Nadal D, Baenziger O, Fischer JE. G-CSF and IL-8 for early diagnosis of sepsis in neonates and critically ill children safety and cost effectiveness of a new laboratory prediction model: study protocol of a randomized controlled trial. Crit Care. 2004;8(6):R443 R Kenis G, Teunissen C, De Jongh R, Bosmans E, Steinbusch H, Maes M. Stability of interleukin 6, soluble interleukin 6 receptor, interleukin 10 and CC16 in human serum. Cytokine. 2002;19(5): Thavasu PW, Longhurst S, Joel SP, Slevin ML, Balkwill FR. Measuring cytokine levels in blood: importance of anticoagulants, processing, and storage conditions. J Immunol Methods. 1992;153(1 2): Rodríguez-Gaspar M, Santolaria F, Jarque-Lopez A, et al. Prognostic value of cytokines in SIRS general medical patients. Cytokine. 2001;15(4): Arch Pathol Lab Med Vol 132, November 2008 Stability of TNF-, IL-6, and IL-8 in Blood Samples Friebe & Volk

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions

Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions Clinica Chimica Acta 300 (2000) 13 21 www.elsevier.com/ locate/ clinchim Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions L.V. Rao *, A.O.

More information

TNF-alpha ELISA. For Research Use Only. Not For Use In Diagnostic Procedures.

TNF-alpha ELISA. For Research Use Only. Not For Use In Diagnostic Procedures. TNF-alpha ELISA For the quantitative determination of TNF-alpha in serum, plasma, buffered solution or cell culture medium. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number:

More information

Research Article Multiplex Cytokine Concentration Measurement: How Much Do the Medium and Handling Matter?

Research Article Multiplex Cytokine Concentration Measurement: How Much Do the Medium and Handling Matter? Mediators of Inflammation Volume 2013, Article ID 890706, 13 pages http://dx.doi.org/10.1155/2013/890706 Research Article Multiplex Cytokine Concentration Measurement: How Much Do the Medium and Handling

More information

HbA1c (Human) ELISA Kit

HbA1c (Human) ELISA Kit HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,

More information

Pilot studies with reference specimens

Pilot studies with reference specimens Pilot studies with reference specimens HUPO: Plasma Proteome Project Workshop July 16-17, 17, 2003 Daniel W. Chan, Ph.D.,DABCC Professor of Pathology, Oncology, Radiology & Urology Director of Clinical

More information

Rat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit

Rat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit Rat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit Catalog No. E0228r (96 tests) Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS!

More information

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation.

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Sysmex America White Paper One Nelson C. White Parkway,

More information

DKK-1 ELISA, Cat.No. BI For the quantitative determination of DKK-1 in human serum

DKK-1 ELISA, Cat.No. BI For the quantitative determination of DKK-1 in human serum DKK-1 ELISA, Cat.No. BI-20413 For the quantitative determination of DKK-1 in human serum ASSAY CHARACTERISTICS Method Standard range Conversion factor Sample volume / well Incubation time, temp. Sensitivity

More information

Patient Preparation Unique patient preparation requirements are listed under each test in the Test Directory.

Patient Preparation Unique patient preparation requirements are listed under each test in the Test Directory. Specimen Collection Lab results are only as good as the specimen provided. Patient preparation, venipuncture technique, specimen handling, and transportation can all affect the quantity of results. Health

More information

Evaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers

Evaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Evaluation Report Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Location: Operators: Eurolyser Diagnostica GmbH Simone Wieser, Bettina Schönauer, Dr. Jürgen Berlanda Report created on 17

More information

EliKine Free Thyroxine (ft4) ELISA Kit

EliKine Free Thyroxine (ft4) ELISA Kit EliKine Free Thyroxine (ft4) ELISA Kit Booklet Item NO. KET0005 Product Name EliKine Free Thyroxine (ft4) ELISA Kit ATTENTION For laboratory research use only. Not for clinical or diagnostic use TABLE

More information

Human HBcAb IgM ELISA kit

Human HBcAb IgM ELISA kit Human HBcAb IgM ELISA kit Catalog number: NR-R10163 (96 wells) The kit is designed to qualitatively detect HBcAb IgM in human serum or plasma. FOR RESEARCH USE ONLY. NOT FOR DIAGNOSTIC OR THERAPEUTIC PURPOSES

More information

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE

More information

Supplementary Material 1. Statistical methods used to conduct power calculations.

Supplementary Material 1. Statistical methods used to conduct power calculations. Supplementary Material 1. Statistical methods used to conduct power calculations. Post-hoc power calculations and patient numbers needed to detect changes were conducted considering (i) the observed partial

More information

Human Cathepsin D ELISA Kit

Human Cathepsin D ELISA Kit GenWay Biotech, Inc. 6777 Nancy Ridge Drive San Diego, CA 92121 Phone: 858.458.0866 Fax: 858.458.0833 Email: techline@genwaybio.com http://www.genwaybio.com Human Cathepsin D ELISA Kit Catalog No. GWB-J4JVV9

More information

Changes in Cytokines at Extreme Surroundings in Antarctica

Changes in Cytokines at Extreme Surroundings in Antarctica Yonago Acta medica 2003;46:29 34 Changes in Cytokines at Extreme Surroundings in Antarctica Shinji Otani and Hiromitsu Kusagaya Division of Organ Surgical Oncology, Department of Surgery, School of Medicine,

More information

L utilizzo della Procalcitonina in Medicina d Urgenza

L utilizzo della Procalcitonina in Medicina d Urgenza L utilizzo della Procalcitonina in Medicina d Urgenza Stefania Battista Dirigente Medico S.C. Medicina d Urgenza Azienda Ospedaliero-Universitaria San Giovanni Battista di Torino Savona, 15 ottobre 2009

More information

LDL (Human) ELISA Kit

LDL (Human) ELISA Kit LDL (Human) ELISA Kit Cat. No.:DEIA3864 Pkg.Size:96T Intended use This immunoassay kit allows for the specific measurement of human low density lipoprotein, LDL concentrations in cell culture supernates,

More information

Procine sphingomyelin ELISA Kit

Procine sphingomyelin ELISA Kit Procine sphingomyelin ELISA Kit For the quantitative in vitro determination of Procine sphingomyelin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH

More information

The Usefulness of Sepsis Biomarkers. Dr Vineya Rai Department of Anesthesiology University of Malaya

The Usefulness of Sepsis Biomarkers. Dr Vineya Rai Department of Anesthesiology University of Malaya The Usefulness of Sepsis Biomarkers Dr Vineya Rai Department of Anesthesiology University of Malaya 1 What is Sepsis? Whole Body Inflammatory State + Infection 2 Incidence and Burden of Sepsis in US In

More information

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Gene L. Gulati, PhD; Lawrence J. Hyland, MD; William Kocher, MD;

More information

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100) Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November

More information

In-Common Laboratories

In-Common Laboratories Stability Information: Specimen stability information has been added or revised for the following tests: Vitamin B6, Dehydroepiandrosterone, Thyroid Antibodies (Thyroid Peroxidase Ab & Thyroglobulin Ab)

More information

Lab Guide 2019 Metabolic Section Lab Guide

Lab Guide 2019 Metabolic Section Lab Guide Lab Guide 2019 Metabolic Section Lab Guide Quantitative Amino acids Plasma Plasma. Container/Tube: Preferred EDTA, Place immediately in ice. Acceptable: lithium heparin, sodium heparin. Patient preparation:

More information

8.2 Principles of Quantitative Hematologic Determinations (1)

8.2 Principles of Quantitative Hematologic Determinations (1) 3 8. FIELD CENTER HEMATOLOGY SERVICES 8.1 Clinical Significance Quantitation of the formed elements of the blood (erythrocytes -RBCs, leukocytes - WBCs, and platelets) is important in the ARIC study primarily

More information

History of the Vacutainer Tube. Coagulant Blood Tests

History of the Vacutainer Tube. Coagulant Blood Tests History of the Vacutainer Tube The different tubes are known as a vacutainer and were developed by Joseph Kleiner in 1947. The vacutainer is currently being manufactured by Becton, Dickinson and Company

More information

Mouse Myeloperoxidase/MPO ELISA Kit

Mouse Myeloperoxidase/MPO ELISA Kit OriGene Technologies, Inc 9620 Medical Center Dr., Suite 200, Rockville, MD 20850 Phone: 1.888.267.4436 Fax: 301-340-9254 Email: techsupport@origene.com Web: Mouse Myeloperoxidase/MPO ELISA Kit Catalog

More information

Mouse Cathepsin B ELISA Kit

Mouse Cathepsin B ELISA Kit GenWay Biotech, Inc. 6777 Nancy Ridge Drive San Diego, CA 92121 Phone: 858.458.0866 Fax: 858.458.0833 Email: techline@genwaybio.com http://www.genwaybio.com Mouse Cathepsin B ELISA Kit Catalog No. GWB-ZZD154

More information

patients with gastrointestinal tract perforation

patients with gastrointestinal tract perforation Research Paper Mediators of Inflammation 1, 45-48 (1992) PLSM levels of endotoxin and various cytokines were assessed in 70 patients with gastrointestinal tract perforation. Sepsis developed in 29 of them,

More information

Blood Ammonia Stability Revisited

Blood Ammonia Stability Revisited Blood Ammonia Stability Revisited JACK HESTER, WILLIAM KORZUN, LARA MABRY ABSTRACT The instability of ammonia in whole blood has challenged clinical laboratory scientists to develop specimen collection

More information

Human HIV (1+2) antigen&antibody ELISA Kit

Human HIV (1+2) antigen&antibody ELISA Kit Human HIV (1+2) antigen&antibody ELISA Kit Catalog Number. CSB-E18042h For the qualitative determination of human HIV (1+2) antibody and P24 antigen concentrations in serum, plasma. This package insert

More information

Sample Type and Storage Conditions Affect Calprotectin Measurements in Blood

Sample Type and Storage Conditions Affect Calprotectin Measurements in Blood Sample Type and Storage Conditions Affect Calprotectin Measurements in Blood Lise Pedersen, 1 Erling Birkemose, 2 Charlotte Gils, 3 * Sara Safi, 2 and Mads Nybo 2 Background: Calprotectin, a complex of

More information

Mouse HBsAg(Hepatitis B Virus Surface Antigen) ELISA Kit

Mouse HBsAg(Hepatitis B Virus Surface Antigen) ELISA Kit Mouse HBsAg(Hepatitis B Virus Surface Antigen) ELISA Kit Catalogue No.: EM0002 Size: 96T Reactivity: Mouse Application: This immunoassay kit allows for the qualitative determination of HBsAg in Mouse serum

More information

BNP Fragment EIA (Cat.No. BI-20852W) For the Determination of BNP Fragment in Human Samples

BNP Fragment EIA (Cat.No. BI-20852W) For the Determination of BNP Fragment in Human Samples BNP Fragment EIA (Cat.No. BI-20852W) For the Determination of BNP Fragment in Human Samples ASSAY CHARACTERISTICS Method Competitive Enzyme Immunoassay, HRP/TMB. Microtiter plates are coated with a polyclonal

More information

Separation of Plasma and Serum and Their Proteins from Whole Blood

Separation of Plasma and Serum and Their Proteins from Whole Blood Separation of Plasma and Serum and Their Proteins from Whole Blood BCH 471 [Practical] BLOOD COMPOSITION Other names to blood cells Red blood cells (erythrocytes) White blood cells (leukocytes) Platelets

More information

BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION. F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME)

BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION. F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME) BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME) TOTAL VARIATION (CV (CV T ) T ) PREANALYTICAL VARIATION ANALYTICAL VARIATION

More information

What Makes a Satisfactory Newborn Screen Sample?

What Makes a Satisfactory Newborn Screen Sample? What Makes a Satisfactory Newborn Screen Sample? Feb 21, 2013 Christine McRoberts Newborn Screening Ontario One of the challenges the Newborn screening Ontario program (NSO) faces every day is the receipt

More information

Pro- and anti-inflammatory responses are regulated simultaneously from the first moments of septic shock

Pro- and anti-inflammatory responses are regulated simultaneously from the first moments of septic shock 82 Eur. Cytokine Netw. Vol. 22 n 2, June 2011, 82-7 RESEARCH ARTICLE Pro- and anti-inflammatory responses are regulated simultaneously from the first moments of septic shock Eduardo Tamayo 1,2,, Ana Fernández

More information

Erythrocyte and Plasma Volume Measurement

Erythrocyte and Plasma Volume Measurement Erythrocyte and Plasma Volume Measurement JP Esser, Meander Medical Centre, Amersfoort NOTE: no changes have been made since the version of 2007 Warning: I human serum albumin and 131 I human serum albumin

More information

Effects of analgesia methods on serum IL-6 and IL-10 levels after cesarean delivery

Effects of analgesia methods on serum IL-6 and IL-10 levels after cesarean delivery Effects of analgesia methods on serum IL-6 and IL-10 levels after cesarean delivery Z.-M. Xing*, Z.-Q. Zhang*, W.-S. Zhang and Y.-F. Liu Anesthesia Department, No. 1 People s Hospital of Shunde, Foshan,

More information

Human CD4+T Cell Care Manual

Human CD4+T Cell Care Manual Human CD4+T Cell Care Manual INSTRUCTION MANUAL ZBM0067.04 SHIPPING CONDITIONS Human CD4+T Cells, cryopreserved Cryopreserved human CD4+T cells are shipped on dry ice and should be stored in liquid nitrogen

More information

Human Alpha 1 microglobulin ELISA Kit

Human Alpha 1 microglobulin ELISA Kit Human Alpha 1 microglobulin ELISA Kit Catalogue No.: EH4144 Size: 48T/96T Reactivity: Human Range:0.625-40ng/ml Sensitivity:

More information

Fresh and Citrated Whole-Blood Specimens Can Produce Different Thromboelastography Results in Patients on Extracorporeal Membrane Oxygenation

Fresh and Citrated Whole-Blood Specimens Can Produce Different Thromboelastography Results in Patients on Extracorporeal Membrane Oxygenation Fresh and Citrated Whole-Blood Specimens Can Produce Different Thromboelastography Results in Patients on Extracorporeal Membrane Oxygenation Elizabeth A. Gilman, MD, 1 Christopher D. Koch, 1 Paula J.

More information

Rat Hemoglobin A1c (HbA1c) Kit Instructions

Rat Hemoglobin A1c (HbA1c) Kit Instructions V.3 Crystal Chem Rat Hemoglobin A1c (HbA1c) Kit Instructions For the quantitative determination of hemoglobin A1c (HbA1c) in rat whole blood Catalog #80300 96 Assays For research use only. Not for use

More information

Prothrombin (Human) ELISA Kit

Prothrombin (Human) ELISA Kit Prothrombin (Human) ELISA Kit Catalog Number KA0496 96 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay... 3 General

More information

Mouse Leptin ELISA Kit (mleptin-elisa)

Mouse Leptin ELISA Kit (mleptin-elisa) Mouse Leptin ELISA Kit (mleptin-elisa) Cat. No. EK0438 96 Tests in 8 x 12 divisible strips Background Leptin (or obese, OB) is a circulating hormone that is expressed abundantly and specifically in the

More information

REPRODUCIBILITY OF SYNOVIAL FLUID ANALYSES

REPRODUCIBILITY OF SYNOVIAL FLUID ANALYSES 770 REPRODUCIBILITY OF SYNOVIAL FLUID ANALYSES A Study Among Four Laboratories H. RALPH SCHUMACHER, JR., MARIE S. SIECK, SUSAN ROTHFUSS, GILDA M. CLAYBURNE, DOROTHY F. BAUMGARTEN, BONNIE S. MOCHAN, and

More information

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags.

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags. WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags www.sysmex.com/us Introduction For many decades, physicians have relied on

More information

The information contained here may be very important to your practice. Please take a moment to review this document.

The information contained here may be very important to your practice. Please take a moment to review this document. September 2017 Dear Healthcare Provider, The information contained here may be very important to your practice. Please take a moment to review this document. TEST BULLETIN T-SPOT TB TESTING Please see

More information

The Patient and Observer Scar Assessment Scale: A Reliable and Feasible Tool for Scar Evaluation

The Patient and Observer Scar Assessment Scale: A Reliable and Feasible Tool for Scar Evaluation The Patient and Observer Scar Assessment Scale: A Reliable and Feasible Tool for Scar Evaluation Lieneke J. Draaijers, M.D., Fenike R. H. Tempelman, M.D., Yvonne A. M. Botman, M.D., Wim E. Tuinebreijer,

More information

Fish follicle-stimulating hormone, FSH ELISA Kit

Fish follicle-stimulating hormone, FSH ELISA Kit Fish follicle-stimulating hormone, FSH ELISA Kit Catalog No: E0830f 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

Human Leptin Receptor ELISA Kit (hleptinr-elisa)

Human Leptin Receptor ELISA Kit (hleptinr-elisa) Human Leptin Receptor ELISA Kit (hleptinr-elisa) Cat. No. EK0439 96 Tests in 8 x 12 divisible strips Background Leptin receptor (or Obese receptor, OBR) is a single membrane-spanning receptor most related

More information

Standard Operation Procedure (SOP) for Biobanking Sampling Procedure Manual Use

Standard Operation Procedure (SOP) for Biobanking Sampling Procedure Manual Use Standard Operation Procedure (SOP) for Biobanking Sampling Procedure Manual Use 1. Plasma, Serum and PAXgene Blood Collection 1.1. Equipment and reagents to be supplied by user 1.1.1. For blood withdrawal

More information

E.Z.N.A. SQ Blood DNA Kit II. Table of Contents

E.Z.N.A. SQ Blood DNA Kit II. Table of Contents E.Z.N.A. SQ Blood DNA Kit II Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Blood Storage and DNA Yield...4 Preparing Reagents...5 100-500 μl Whole Blood Protocol...6

More information

March Immunoglobulins, Total (IgA, IgE, IgG, IgM) Interleukin-6 (IL6) Protein Electrophoresis, CSF. Thyrotropin Receptor Antibody

March Immunoglobulins, Total (IgA, IgE, IgG, IgM) Interleukin-6 (IL6) Protein Electrophoresis, CSF. Thyrotropin Receptor Antibody University of Washington March 2012 TEST CHANGE Alpha 1-Antitrypsin Anti-IgE Receptor Antibody Anti-Mullerian Hormone Aquaporin-4 Receptor Antibody ColoSeq - Polyposis Complement testing, C3, C4 Everolimus

More information

MANUAL IL-1alpha (mouse) ELISA Kit Cat. No. AG-45B-0003-KI01 [Interleukin-1 alpha (mouse) ELISA Kit]

MANUAL IL-1alpha (mouse) ELISA Kit Cat. No. AG-45B-0003-KI01 [Interleukin-1 alpha (mouse) ELISA Kit] MANUAL IL-1alpha (mouse) ELISA Kit [Interleukin-1 alpha (mouse) ELISA Kit] For research use only. Not for diagnostic use Version 1 (March-5-2013) Cat. No. AG-45B-0003-KI01 www.adipogen.com Table of Contents

More information

ab Cathepsin D Human ELISA Kit

ab Cathepsin D Human ELISA Kit ab119586 Cathepsin D Human ELISA Kit Instructions for Use For quantitative detection of Human Cathepsin D in cell culture supernatants, serum and plasma (heparin, EDTA). This product is for research use

More information

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit Catalog No: E0442h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

Study on diurnal variation in TSH and freet4 levels of healthy adults

Study on diurnal variation in TSH and freet4 levels of healthy adults Original Article Study on diurnal variation in TSH and freet4 levels of healthy adults Liyanage YSH 1, Siriwardhana ID 2, Dissanayake M 3, Dayanath BKPT 4 1 Allied Health Sciences Degree Program, Faculty

More information

SALEM HOSPITAL SALEM, OREGON 97309

SALEM HOSPITAL SALEM, OREGON 97309 SALEM HOSPITAL SALEM, OREGON 97309 Department: Phlebotomy TITLE: BLOOD COLLECTIONS, VENIPUNCTURE Area: Phlebotomy Effective Date: 04/01/96 Authored By: Cindy Humphrey, Diane Duncan Revised: 09/26/2000,

More information

AquaPreserve DNA/RNA/Protein Order # Preservation and Extraction Kit 8001MT, 8060MT

AquaPreserve DNA/RNA/Protein Order # Preservation and Extraction Kit 8001MT, 8060MT AquaPreserve DNA/RNA/Protein Order # Preservation and Extraction Kit 8001MT, 8060MT MoBiTec GmbH 2014 Page 2 Contents 1. Description... 3 2. Kit Contents... 3 3. Terms & Conditions... 3 4. AquaPreserve

More information

BIOCHEMISTRY OF BLOOD

BIOCHEMISTRY OF BLOOD BCH 471 BIOCHEMISTRY OF BLOOD Amal Alamri Experiment 1 Separation of Plasma and Serum from Whole Blood Whole Blood It is living tissue that circulates through the heart, arteries, veins, and capillaries

More information

Evaluation of the EUROLYSER-SOLO Total T4 assay. for veterinary samples

Evaluation of the EUROLYSER-SOLO Total T4 assay. for veterinary samples Evaluation of the EUROLYSER-SOLO Total T4 assay for veterinary samples Dr. rer.-nat. Volker Schlüter Zacherlweg 18 D-82061 Neuried Date: 2011 April 27 th Scope of the study Goal of this study was the validation

More information

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2 For detection and measurement of human interleukin 2 Catalog #02006 Catalog #02007 2 Plates 10 Plates Product Description The Human Interleukin 2 (IL-2) ELISA Kit is designed for the quantitative detection

More information

Technical Bulletin No. 172

Technical Bulletin No. 172 CPAL Central Pennsylvania Alliance Laboratory QuantiFERON -TB Gold Plus Assay Contact: J Matthew Groeller, MPA(HCM), MT(ASCP), 717-851-4516 Operations Manager, Clinical Pathology, CPAL Jennifer Thebo,

More information

RayBio Human Thyroglobulin ELISA Kit

RayBio Human Thyroglobulin ELISA Kit RayBio Human Thyroglobulin ELISA Kit Catalog #: ELH-Thyroglobulin User Manual Last revised July 6, 2017 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite

More information

The impact of preanalytical factors on glucose concentration measurement

The impact of preanalytical factors on glucose concentration measurement The impact of preanalytical factors on glucose concentration measurement Nora Nikolac, PhD University Department of Chemistry, Medical School University Hospital Sestre Milosrdnice, Zagreb, Croatia Glucose

More information

Human Leptin ELISA Kit

Human Leptin ELISA Kit Product Manual Human Leptin ELISA Kit Catalog Numbers MET-5057 MET-5057-5 96 assays 5 x 96 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Leptin is a polypeptide hormone

More information

Prognostic Value of Plasma Interleukin-6 Levels in Patients with Chronic Lymphocytic Leukemia

Prognostic Value of Plasma Interleukin-6 Levels in Patients with Chronic Lymphocytic Leukemia 1071 Prognostic Value of Plasma Interleukin-6 Levels in Patients with Chronic Lymphocytic Leukemia Raymond Lai, M.D, PhD. 1 Susan O Brien, M.D. 2 Taghi Maushouri, M.S. 1 Anna Rogers, 1 Hagop Kantarjian,

More information

EDUCATIONAL COMMENTARY DISSEMINATED INTRAVASCULAR COAGULATION

EDUCATIONAL COMMENTARY DISSEMINATED INTRAVASCULAR COAGULATION EDUCATIONAL COMMENTARY DISSEMINATED INTRAVASCULAR COAGULATION Educational commentary is provided through our affiliation with the American Society for Clinical Pathology (ASCP). To obtain FREE CME/CMLE

More information

Initial serum level of TNF-α as an outcome predictor in pediatric patient with sepsis.

Initial serum level of TNF-α as an outcome predictor in pediatric patient with sepsis. Curr Pediatr Res 2018; 22 (2): 146-151 ISSN 0971-9032 www.currentpediatrics.com Initial serum level of TNF-α as an outcome predictor in pediatric patient with sepsis. Idham Jaya Ganda*, Yustin Andriani,

More information

Acetylcholine Receptor Binding Antibody

Acetylcholine Receptor Binding Antibody Acetylcholine Receptor Binding Antibody Order Name: ACETY BND Test Number: 5500010 REV DATE:9/17/2008 Acetylcholine Receptor Binding Antibody RIA Preferred 1 ml (0.5) Serum Clot Activator (Red Top, No-Gel)

More information

Inflammatory Cytokines and Their Prognostic Ability in Cases of Major Burn Injury

Inflammatory Cytokines and Their Prognostic Ability in Cases of Major Burn Injury Original Article Diagnostic Immunology Ann Lab Med 2015;35:105-110 http://dx.doi.org/10.3343/alm.2015.35.1.105 ISSN 2234-3806 eissn 2234-3814 Inflammatory Cytokines and Their Prognostic Ability in Cases

More information

First Probable West Nile Virus Case in the Simcoe-Muskoka Area

First Probable West Nile Virus Case in the Simcoe-Muskoka Area Dr. Charles Gardner, Medical Officer of Health Dr. Colin Lee, Associate Medical Officer of Health Dr. Lisa Simon, Associate Medical Officer of Health First Probable West Nile Virus Case in the Simcoe-Muskoka

More information

Stapleton, Mary; Daly, Niamh; O'Kelly, Ruth; Turner, Michael.

Stapleton, Mary; Daly, Niamh; O'Kelly, Ruth; Turner, Michael. Provided by the author(s) and University College Dublin Library in accordance with publisher policies. Please cite the published version when available. Title Time And temperature affect glycolysis in

More information

Mohammadreza Tabatabaei IBTO COAG LAB

Mohammadreza Tabatabaei IBTO COAG LAB Tests for the Evaluation of Lupus Anticoagulants t Mohammadreza Tabatabaei MSc Hematology blood bank MSc Hematology blood bank IBTO COAG LAB Lupus Anticoagulants General Background Lupus anticoagulants

More information

Human Cathepsin V ELISA Kit

Human Cathepsin V ELISA Kit Human Cathepsin V ELISA Kit Catalog #: DIA-XYA118 Detection and Quantification of Human Cathepsin V Concentrations in Cell Lysates, Sera and Plasma. Please read the provided manual as suggested experimental

More information

Activated Partial Thromboplastin Time (aptt)

Activated Partial Thromboplastin Time (aptt) Activated Partial Thromboplastin Time (aptt) Order Name: PTT Test Number: 1500050 REV DATE:12/26/2008 Activated Partial Thromboplastin Time (aptt) CLOT Preferred 2.7 ml Whole Blood Sodium Citrate 3.2%

More information

SHOULD THERAPEUTIC AGENTS FOR SEPSIS TARGET THE GLYCOCALYX?

SHOULD THERAPEUTIC AGENTS FOR SEPSIS TARGET THE GLYCOCALYX? SHOULD THERAPEUTIC AGENTS FOR SEPSIS TARGET THE GLYCOCALYX? Sir Ganga Ram Hospital New Delhi, India Dr. Seema Bhargava Senior Consultant & Chairperson Department of Biochemistry & Professor, GRIPMER Sir

More information

Glutathione S-Transferase Assay Kit

Glutathione S-Transferase Assay Kit Glutathione S-Transferase Assay Kit Catalog Number KA1316 96 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay...

More information

Insulin Aspart ELISA Kit

Insulin Aspart ELISA Kit Insulin Aspart ELISA Kit Catalogue DEIABL215 For the qualitative determination of antibodies to insulin aspart in human serum and plasma FOR RESEARCH USE ONLY Creative Diagnostics. All rights reserved

More information

Use of Blood Lactate Measurements in the Critical Care Setting

Use of Blood Lactate Measurements in the Critical Care Setting Use of Blood Lactate Measurements in the Critical Care Setting John G Toffaletti, PhD Director of Blood Gas and Clinical Pediatric Labs Professor of Pathology Duke University Medical Center Chief, VAMC

More information

Blood Urea Nitrogen Detection Kit

Blood Urea Nitrogen Detection Kit Discovery through partnership Excellence through quality StressXpress Blood Urea Nitrogen Detection Kit Catalog# SKT-213 (2 Plate kit) Quantitative colorimetric measurement of urea nitrogen TABLE OF CONTENTS

More information

Page 1 of 2. Standard curve of Human IFN gamma ELISA Ready- SET-Go! Product Information Contents: Human IFN gamma ELISA Ready- SET-Go!

Page 1 of 2. Standard curve of Human IFN gamma ELISA Ready- SET-Go! Product Information Contents: Human IFN gamma ELISA Ready- SET-Go! Page 1 of 2 Human IFN gamma ELISA Ready-SET-Go! Catalog Number: 88-7316 Also known as: Interferon-gamma, IFN-g, IFNg RUO: For. Not for use in diagnostic procedures. Standard curve of Human IFN gamma ELISA

More information

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet Interleukin5 (IL5) is a secreted glycoprotein that belongs to the α-helical group of cytokines (1, 2, 3). Unlike other family members, it is present as a covalently linked antiparallel dimer (4, 5). IL-5

More information

New evidences. Biomarkers to explore immunoparalysis: what future for immunostimulation? B. François CHU Limoges (France)

New evidences. Biomarkers to explore immunoparalysis: what future for immunostimulation? B. François CHU Limoges (France) New evidences Biomarkers to explore immunoparalysis: what future for immunostimulation? B. François CHU Limoges (France) Background Hotchkiss, NEJM 2003 New understanding of response in sepsis PICS Persistent

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

Toxicology Blood Lead Comprehensive Testing

Toxicology Blood Lead Comprehensive Testing Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where

More information

Phlebotomy Top Gun Order of Draw: Do We Still Care?

Phlebotomy Top Gun Order of Draw: Do We Still Care? Phlebotomy Top Gun Order of Draw: Do We Still Care? 2018 MFMER 3793435-1 Presenter: Brad Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Division of Clinical Core Laboratory Services

More information

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Original: 5/02 Minor Revision: 9/15 Full Review: 1/15 Replaces: 1/15 Responsible Person: Point of Care Laboratory Specialist Approving Committee:

More information

Diagnostic Markers for Identifying Sepsis in Patients with Systemic Inflammatory Response Syndrome (SIRS): A Prospective Study

Diagnostic Markers for Identifying Sepsis in Patients with Systemic Inflammatory Response Syndrome (SIRS): A Prospective Study The Open Pediatric Medicine Journal, 29, 3, 1-7 1 Open Access Diagnostic Markers for Identifying Sepsis in Patients with Systemic Inflammatory Response Syndrome (SIRS): A Prospective Study Jana Pavare

More information

A. SAP is the D-Lab's name for a specific set of serum biochemical tests.

A. SAP is the D-Lab's name for a specific set of serum biochemical tests. Understanding CBC, SAP, UA/Laura J. Steadman, DVM I. CBC - Complete Blood Count A. Three major types of cells are counted 1. Red Blood Cells 2. White Blood Cells 3. Platelets B. Cells are counted at the

More information

Canine Thyroid Stimulating Hormone, TSH ELISA Kit

Canine Thyroid Stimulating Hormone, TSH ELISA Kit Canine Thyroid Stimulating Hormone, TSH ELISA Kit Catalog No: E0463c 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

Human Neurology 3-Plex A

Human Neurology 3-Plex A Human Neurology 3-Plex A SUMMARY AND EXPLANATION OF THE TEST The Human N3PA assay is a digital immunoassay for the quantitative determination of total Tau, Aβ42, and Aβ40 in human plasma and CSF. Determination

More information

The pages that follow contain information critical to protecting the health of your patients and the citizens of Colorado.

The pages that follow contain information critical to protecting the health of your patients and the citizens of Colorado. Health Alert Network Tri-County Health Department Serving Adams, Arapahoe and Douglas Counties Phone 303/220-9200 Fax 303/741-4173 www.tchd.org Follow us on Twitter @TCHDHealth and @TCHDEmergency John

More information

Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes

Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes Citation for published version (APA): Berkel, van, M., & Scharnhorst,

More information

HIV Diagnostic Testing

HIV Diagnostic Testing In The name of God HIV Diagnostic Testing By : Dr. Shahzamani PhD of Medical virology Purpose of HIV Testing To identify asymptomatic individuals To diagnose HIV infection in those who practice high risk

More information