Progressive Decline in Plethysmographic Lung Volumes in Infants Physiology or Technology?
|
|
- Robert Miller
- 6 years ago
- Views:
Transcription
1 Progressive Decline in Plethysmographic Lung Volumes in Infants Physiology or Technology? Georg Hülskamp, Ah-fong Hoo, Henrik Ljungberg, Sooky Lum, J. Jane Pillow, and Janet Stocks Portex Anaesthesia, Intensive Therapy and Respiratory Medicine Unit, Institute of Child Health and Great Ormond Street Hospital for Children NHS Trust; Neonatal Units, Homerton University Hospital and University College London Hospitals, London, United Kingdom; and Department of Pediatrics, University of Münster, Münster, Germany During the last 30 years, there has been an unexplained trend toward declining values for plethysmographic assessments of lung volume at functional residual capacity (FRC) in infants. The aim of this study was to compare data collected from healthy infants using contemporary equipment with published reference data and to explore reasons for discrepancies. Lung volumes were measured in 32 healthy infants (age, 4 93 weeks; weight, kg) using a new, commercially available infant plethysmograph. Mean (SD) FRC was 19.6 (3.4) ml/kg (within subject coefficient of variation 3.4 [2.3%]), which was on average 7.0 [3.5] ml/kg and 2.3 [1.2] SD (Z) scores lower than the recently collated reference data from an American Thoracic Society task force. A total of 66% of these healthy infants had a FRC that was below the predicted normal range. Comparison of equipment, software, and protocols with those from previous reports revealed the importance of minimization of dead space and of adequate subtraction of all compressible occluded volume when calculating FRC in infants. These findings emphasize the need to establish reference data for lung function tests in infants that are appropriate for the equipment and protocols in current use. Keywords: functional residual capacity; plethysmography; infant; respiratory function tests Estimates of lung volume are an important part of infant lung function testing, both when assessing normal growth and development of the lungs and when interpreting changes in respiratory mechanics in response to disease or therapeutic interventions. Functional residual capacity (FRC), the only lung volume that can readily be assessed in infants, is measured either by wholebody plethysmography (FRC p ) (1) or by gas dilution techniques (2). FRC p measurements in infants, as in adults, are rapid and reproducible but have generally been restricted to specialized centers as the technique involved was sophisticated and required considerable operator training. (Received in original form March 31, 2003; accepted in final form August 4, 2003) Supported by the Innovative Medizinische Forschung, University of Münster, and the Gesellschaft für Pädiatrische Pneumologie, Germany (G.H.), by Portex Ltd. (A.-F.H. and J.S.), by a European Respiratory Society Long Term Research Fellowship (H.L.), and by a Neil Hamilton Fairley NHMRC Postdoctoral Fellowship (J.P.). Research at the Institute of Child Health and Great Ormond Street Hospital for Children NHS Trust benefits from R&D funding received from the NHS Executive. Correspondence and requests for reprints should be addressed to Georg Hülskamp, M.D., Portex Anaesthesia, Intensive Therapy and Respiratory Medicine Unit, Institute of Child Health and Great Ormond Street Hospital for Children NHS Trust, 30 Guilford Street, London WC1N 1EH, UK. g.hulskamp@ich.ucl.ac.uk This article has an online supplement, which is accessible from this issue s table of contents online at Am J Respir Crit Care Med Vol 168. pp , 2003 Originally Published in Press as DOI: /rccm OC on August 6, 2003 Internet address: In 2001, a joint American Thoracic Society European Respiratory Society (ATS ERS) task force published recommendations on the methodology and the equipment for infant plethysmography (1, 3), and a new generation of infant body plethysmographs was developed that met most of these recommendations. Although these new developments could make infant plethysmography more applicable for routine physiologic and clinically orientated measurements, meaningful interpretation of results will still depend on the availability of appropriate reference data (4). A prediction equation with confidence limits for FRC p in infants was proposed in 1995 by Stocks and Quanjer (5) as part of an ATS task force on lung volumes (Clausen and Wagner, for the workshop participants. Consensus document on measurements of lung volume in humans. Developed from workshops sponsored by the American Thoracic Society and the National Heart, Lung, and Blood Institute. Manuscript submitted for posting on the American Thoracic Society website). When published data were collated as a basis for developing that equation, a seemingly secular trend toward lower values of FRC p became apparent in the more recent studies. Subsequent reports on FRC p from healthy infants have confirmed this trend (6 8). It is well recognized that prediction equations need to be constantly tested and validated and that investigators using such reference data should test a number of healthy subjects to determine whether selected prediction equations are appropriate for the specific populations, equipment, and protocols used (5, 9; Clausen and Wagner, submitted). This has rarely been done, and it remains unclear whether the observed decline in FRC p with time has a physiologic or epidemiologic basis or can be attributed to advances in measurement protocols and technology. The new Jaeger (J) MasterScreen BabyBodyplethysmograph (VIASYS Healthcare GmbH, Höchberg, Germany), which incorporates solid-state transducers with a much lower apparatus dead space than previously available, has been validated extensively in vitro (10) but, to our knowledge, values of FRC obtained in healthy infants using this equipment (FRC p J ) have not been compared with previously published normative data. The low absolute dead space of this system, together with a standardized and precise approach to correct for such apparatus dead space, might be expected to result in lower values of FRC than those previously reported. The aim of this study was to assess whether published reference data could be used to interpret results of FRC p obtained using current equipment and, if not, to investigate potential causes for discrepancies including those arising from recent changes in equipment, techniques, and protocols. We hypothesized that FRC p J from a cohort of healthy infants tested in adherence to recent ERS ATS recommendations (1) using contemporary equipment is significantly lower than previously published reference data (5; Clausen and Wagner, submitted). Some of the results of this study have been previously reported in the form of abstracts (11, 12).
2 1004 AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE VOL METHODS Study Population and Background Data Healthy white infants ( 36 gestational weeks) without congenital abnormalities or neonatal respiratory compromise were recruited as control subjects into ongoing epidemiologic studies. The local research Ethics Committees approved these studies, and written informed parental consent was obtained before lung function testing. Background information on clinical, medical, and social factors was obtained from obstetric and neonatal records and from interviewing parents on the day of the test. Body weight and crown heel length were measured as described previously (13). Equipment FRC p J was measured using the recently developed Jaeger MasterScreen BabyBodyplethysmograph. The system is based on a 98 L Plexiglas variable pressure plethysmograph and a low dead space pneumotachometer with solid-state transducers (see Table 1E in the online supplement). The combined dead space of the pneumotachometer and shutter was 4.3 ml. The total occluded dead space of the system, including the residual volume of the mask, was estimated to be 11.8 and 14.3 ml, respectively, for a Size 1 or Size 2 mask (14, 15) (see online supplement). The system was designed in adherence to the ERS ATS recommendations (1, 15, 16) and has been validated in vitro when it was found to measure volumes between 75 and 300 ml with an accuracy of plus or minus 2.5% (10). Data acquisition and analysis were performed on a Windows 98 based workstation (Jaeger Lab4 software version 4.53). Online data sampling and reanalysis were based on the structured Jaeger screen display (Figure 1). Results were displayed online instantaneously but could subsequently be checked for quality control, according to current ERS ATS recommendations and as defined in the online supplement, before acceptance. Protocol FRC was measured according to the manufacturer s instructions and ERS ATS recommendations (1) during quiet sleep, after light sedation with chloral hydrate. A minimum of three end-inspiratory occlusions was performed from which the mean (SD) FRC was calculated. Rigorous quality control features were implemented for data acquisition and analysis (see online supplement for details). Statistical Analysis Plethysmographic values of FRC obtained with the Jaeger equipment were compared with the prediction equation published by Stocks and Quanjer (5) using paired t tests and 95% confidence intervals according to the method of Bland and Altman (17). Based on collated reference data, Stocks and Quanjer (5) had proposed the following prediction equation: FRC p-pred 2.36L 0.75 W 0.63 (residual SD: 0.14; 95% limits of agreement: %), where FRC p-pred is the predicted FRC p ; L crown heel length (cm), and W body weight (kg). This equation can be transformed to: ln(frc p-pred ) ln(2.36) 0.75ln(L) 0.63ln(W) (1) where ln is the natural logarithm. The Z score is calculated as the difference between the natural logarithm of the predicted and the actual measured FRC (FRC p J ) divided by the residual SD (abbreviated as RSD in Equation 2): Z [ln(frc p-j ) ln(frc p-pred )]/RSD (2) RESULTS Z [ln(frc p-j ) ln(2.36) 0.75ln(L) 0.63ln(W)]/0.14. (3) Technically satisfactory measurements of FRC p J were obtained in 32 healthy infants on 35 occasions. Two studies in the youngest infants ( 6 weeks postnatal age) were performed in natural sleep without sedation. Table 1 summarizes subject details and results, with individual data documented in Table 2E in the online supplement. Reported results for each subject were based on a mean (SD; range) of 3.7 (1.1; 2 7) occlusions and were highly reproducible within each individual, the mean (SD) coefficient of variation for FRC p J being 3.4% (2.3). Individual values of FRC p J are plotted against length in Figure 2, together with the predicted regression line (5). Mean (SD) FRC p J was 172 (54) ml, which was on average (SD; 95% confidence interval), 59 (30; 49, 69) ml lower than predicted. The mean (SD) Z score for FRC p J from this group of healthy infants was 2.3 (1.2), which was significantly lower than predicted (p 0.001). Figure 1. Screen display showing: time-based plots; summary of results within one test occasion with the average values from the three individual acts (trials) shown in the best column; box volume (VB) versus changes of pressure at the airway opening (P ao ) during three respiratory efforts against the occlusion; zoomed volume versus time plot highlighting changes in end-expiratory level preand postocclusion. Displayed results are technically acceptable and demonstrate: (1) stable end-expiratory level before and after airway occlusion; (2) at least two respiratory efforts against each occlusion; (3) good phase relation between changes in P ao and box volume, i.e., no phase lag ( looping ) of the X Y plot; (4) three satisfactory maneuvers with low coefficient of variation (CV) (1.2%).
3 Hülskamp, Hoo, Ljungberg, et al.: Infant Plethysmographic Lung Volume 1005 TABLE 1. FRC p J IN HEALTHY INFANTS: BACKGROUND CHARACTERISTICS AND RESULTS Mean (SD) Mean Z Score (SD) n 32* % Male 37.5% Gestational age, wk 39.8 (1.2) Birth weight, kg 3.23 (0.46) 0.41 (1.0) Test age, wk 43.3 (21.7) Test weight, kg 8.72 (2.17) 0.21 (1.16) Test length, cm 72.7 (7.9) 0.7 (1.0) FRC p J, ml 172 (54) FRC p J, ml/kg 19.6 (3.4) 2.3 (1.2) * Thirty-five test occasions, 3 infants (2 female, 1 male) were tested on 2 occasions about 25 weeks apart. Freeman and coworkers (25). Corrected for gestational age. Stocks and Quanjer (5). DISCUSSION Plethysmographic measurements of FRC using new commercially available equipment were found to be significantly lower than predicted when based on the most recent and extensive compilation of reference data (5; Clausen and Wagner, submitted). When adjusted for weight, the mean (SD) FRC p J was 19.6 (3.4) ml/kg, with a range from 13.8 to 27.9 ml/kg. The mean (SD) within-subject difference between weight-adjusted FRC p J and predicted values was 7.0 (3.5) ml/kg, which is physiologically and clinically highly relevant and would preclude use of the ATS (1995) reference data when interpreting these results. Results from 23 of 35 (i.e., 66%) of these studies fell outside the 95% limits of agreement (i.e., less than 2 Z scores) of the prediction equation. The presented data need to be interpreted in the broader context of the declining plethysmographic lung volumes in healthy infants that have been reported over the last 30 years. During this period, the reported mean weight-corrected FRC p diminished from 35 to 23 ml/kg (see Table 2 for overview). This trend and the observed discrepancies between the presented data and those predicted could be related to: Differences in the background characteristics of the current cohort of healthy infants as compared with those contributing to the reference data and/or the accuracy with which anthropometric data were recorded. Figure 2. FRC p J in healthy infants plotted versus test length. Solid line indicates functional residual capacity measured by whole-body plethysmography (FRC p ) predicted for length and weight (5, 6) with 95% limits of agreement as dotted lines. Differences in equipment. Changes in protocols for data acquisition, analysis, and quality control. Infant Characteristics Subjects in this study were recruited from the community and did not include any hospital controls such as have been used in some previous reports (6, 7, 18). They should therefore represent a relatively unbiased sample of healthy, inner London infants. Recruitment of infants to the epidemiologic studies was limited to infants born of white mothers (19), whereas the ethnic background of the cohorts contributing to collated reference data was heterogeneous (5). Whereas anthropometric differences related to ethnic origin need to be taken into account for children and adults (5, 9, 20), ethnic differences in FRC have not been demonstrated in infants (21). The slightly higher proportion of girls included in this study should not have influenced the results because once body size has been taken into account no differences in lung volume have been observed in boys and girls in the first 2 years of life (5, 22, 23). Mean birth weight and gestational age of the present cohort were close to the predicted norms for British full-term infants, as was weight for postnatal age at the time of test (mean [SD] Z scores 0.21[1.16]) (24). In contrast, crown heel length in our cohort was somewhat higher than predicted (mean [SD] 0.7 [1.0] Z scores) (24). This may reflect differences in measurement protocols compared with those used by Freeman and coworkers (24). A similar elevation of length Z score in healthy infants has been reported in a recent epidemiologic study (19) that used an identical protocol when measuring infant length (including calibrated precision stadiometer and two trained operators) (13). Such details have not always been reported in previous studies (18, 25, 26), and it is possible that underestimation of infant length in at least some of the studies contributing to the collated reference data (5; Clausen and Wagner, submitted) could account for some of the observed discrepancies in this study. The magnitude of such errors would however need to be extremely large to make a substantial group difference. In this study, we estimated that a mean change in length Z score of 0.7 (on average equivalent to 1.7 cm in actual length for this cohort) would have resulted in an increase in mean Z score for FRC p J of 0.13 only (from 2.36 to 2.23), with the predicted weight-corrected FRC falling by 0.4 ml/kg only (from 26.6 to 26.2 ml/kg). Environmental tobacco smoke exposure is known to be relatively high for white infants living in London (22, 27), which could potentially have influenced our findings. However, in contrast to the adverse effect of nicotine exposure on airway function (8, 27 29), previous large epidemiologic studies have not shown any significant impact of smoke exposure on resting lung volume in newborns and young infants (8, 30, 31). Although the present study was not powered to verify any effect of nicotine exposure on FRC, mean [SD] Z scores for FRC p J in infants whose mothers smoked ( 2.2 [1.3]) was not different from those in nonexposed infants ( 2.3 [1.2]). Ten of the infants in this study had experienced their first lower respiratory infection before the test but were symptom free for at least 3 weeks before testing. Because wheezing illnesses occur in up to 40% healthy infants during the first year of life, exclusion of such infants from a cohort with a mean (SD) age at test of 43 (21) weeks is probably inappropriate. Previous studies have shown that, after recovery, there is no difference in FRC between infants with and without earlier uncomplicated episodes of wheezing (30, 32). Thus, it seems unlikely that there were any substantial differences in background characteristics of the infants that we studied to ex-
4 1006 AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE VOL TABLE 2. PLETHYSMOGRAPHICALLY DETERMINED FUNCTIONAL RESIDUAL CAPACITY IN INFANTS AND YOUNG CHILDREN Test Age Mean Weight FRC p (SD) DS Subtracted First Author, Year of Publication (Ref.) n (mo) (kg) (ml/kg) Mask DS Subtracted? (ml ) Doershuk, 1969 (27) (6) Y Not given Doershuk, 1970 (26) (5) Y 15 Stocks, 1977 (19) 48* (4) N 12 Gappa, 1993 (34) (5) N 7.6/26 Dundas, 1995 (31) (5) N 7.6/26 McCoy, 1995 (7) (5) Y 13/17 Dezateux, 1999 (23) (5) Y 36.1/54.5, Milner, 1999 (32) Y 25 Castile, 2000 (8) (4) Y 13/17 Dezateux, 2001 (33) Y 54.5/57, Current data, (3) Y 11.8/14.3 Definition of abbreviations: DS dead space; FRC p functional residual capacity measured by whole-body plethysmography. Mean results as reported in individual studies or calculated. FRC p is adjusted for weight. * Data from preterm infants ( 2.5 kg birth weight and 36 weeks gestational age) were excluded (according to Stocks and Quanjer [5]). Depending on pneumotachometer dead space. Excluding infants with history of wheezing. Data from end-inspiratory occlusion only. Total compressible dead space. SD cannot be calculated from available data. plain the magnitude of observed differences in FRC p J from predicted values. Equipment and Protocol One of the most relevant factors contributing to the observed differences may be the methods used to account for apparatus dead space and any additional occluded volume. During plethysmographic measurements of lung volume, it is crucial that all compressible volume between the infant s mouth and nose and the shutter is subtracted from the total occluded volume. This total compressible dead space includes not only the effective dead space through which the infant breathes (i.e., within the mask, pneumotachometer, shutter, and any connectors) but also the volume of any transducer tubing proximal to the shutter. Table 2 shows that the dead space accounted for in previous studies varied substantially from 12 to 57 ml. Although different equipment configurations will partially account for this discrepancy, previous underreporting of certain components of the total compressible dead space is likely to account for the remainder. Although dead space was mentioned in all studies represented in Table 2, it often remains unclear how exactly this was estimated and what volumes were included (i.e., volume of the pneumotachometer and connectors and/or mask) (26, 31). Details on estimation and subtraction of effective mask volume have not been consistently specified. In some studies, the dead space of the mask was not subtracted at all due to uncertainties about the volume remaining once in situ on the infant s face (18, 30, 33), whereas in others the method of estimating effective mask dead space is unclear (6, 7, 25, 26, 31). Recent consensus suggests subtraction of 50% of the water displacement of infant masks from any lung volume measurements (14, 34). Individual laboratory practice for the amount of putty used must be taken into account: current practice for the present cohort resulted in a residual mask dead space of 7.5 and 10 ml for Rendall-Baker Size 1 and 2 masks, respectively (14), which is slightly less than that suggested by recent ATS ERS guidelines (34). Overestimation of FRC p by up to 3 ml/kg could easily be attributed to a failure to account for residual mask volume accurately in the past, especially as relatively large masks were used in the early infant plethysmographic studies. None of the studies quoted in Table 2 commented on the volume within the tubings connecting the pneumotachometer and mouth pressure port to the transducers, which, although not representing any dead space for the infant, is part of the total compressible dead space. It is therefore difficult to know how other centers have corrected for this potential source of error. However, we do know that the volume of such tubing (which was subsequently found to be 21 ml [i.e., 3 7 ml]) was not subtracted from our own plethysmographic studies at the Hammersmith Hospital (35) or at the Institute of Child Health in London, until after 1995 (30, 33). Appropriate adjustments for mask dead space were introduced shortly afterward, thereby explaining the apparently large total compressible dead space accounted for by Dezateux and coworkers (8, 32) and the relatively lower values of FRC p when compared with previous studies from the same group (30, 33). Similar, and potentially larger, problems could occur if a bias flow is delivered to minimize rebreathing through a relatively large volume circuitry, unless the occlusion valve is close to the mask. To our knowledge, the total compressible dead space of the current Jaeger device (using solid-state transducers, with no associated tubing; flowsensor shutter mask 11.8 to 14.3 ml) is lower than previously reported equipment. As can be seen from Table 2, a similarly low volume has been subtracted from some other studies, but the details of these calculations are not mentioned. It is difficult to precisely account for a relatively large total compressible dead space as, in contrast to conditions within the lung, this volume will not be compressed under isothermal conditions during airway occlusion. Thus Boyle s law cannot be applied. It is unclear whether conditions within the total compressible dead space will be polytropic or adiabatic, but it can be assumed that the volume of the total compressible dead space, as assessed by water displacement, probably overestimates the effective total compressible dead space (3). Differences in data acquisition could also account for some of the observed discrepancies. Early plethysmographic studies routinely used airway occlusion at end expiration (18, 25, 26), whereas recent studies have used end-inspiratory occlusions (6 8, 30, 33). The latter protocol was recommended in the recent ERS ATS statement (1) as this approach disturbs infants to a
5 Hülskamp, Hoo, Ljungberg, et al.: Infant Plethysmographic Lung Volume 1007 lesser degree and results in less glottic activity during respiratory efforts against the occlusion. In addition, there may be better equilibrium of pressures throughout the respiratory system when occlusions are performed at a higher lung volume. Although there is some controversy regarding the precise effect (36), most authors (37 40) agree that in both healthy infants and in those with airway disease, FRC p is consistently higher when measured at end expiration rather than at end inspiration. In the most recent comparison, McCoy and coworkers (6) estimated the magnitude of this effect on FRC p in healthy infants to be on average (SD; range) 1.3 (1.7; 1.6 to 4.8) ml/kg. These differences may be explained by a lack of equilibration between changes in alveolar pressure and those measured at the airway opening in the presence of airway narrowing at lower lung volumes. In this situation, a relative underestimation of pressure changes at the airway opening would result in an overestimation of FRC p during end-expiratory occlusions. When performing end-inspiratory occlusions, the occluded (Vt) volume above FRC should be corrected to body temperature, pressure, and saturation conditions before subtracting it from the total occluded volume. Although this correction is incorporated into the current Jaeger software, no details are given in many papers (6, 7, 31), and to our knowledge, it has not been consistently used in the past (8, 30, 32, 33). Failure to apply this correction could result in an underestimation of the Vt occluded above FRC by up to 10% and hence in an overestimation of FRC of about 1 ml/kg in an infant with a Vt of 10 ml/kg. The very high values of thoracic gas volume reported during the 1970s will also have been biased by the fact that plethysmographic measurements of FRC were often preceded by assessments of airway resistance, which required the infant to breathe from a heated and humidified rebreathing bag (18). Although the significance of this was not appreciated at the time, this practice, together with the relatively large apparatus dead space common (though not always acknowledged) at the time, would inevitably have caused some hyperventilation and probably dynamic elevation of resting lung volume. During the last decade, most FRC p measurements (including those with the current Jaeger system) appear to have been made with the infants breathing either directly from the box or via a bias flow, so this should no longer be a problem. Nevertheless, a similar phenomenon could occur, albeit to a lesser extent, if the infant has to breathe through a relatively large dead space before FRC recordings and could account for some of the discrepancies between our current findings and previously reported FRC p values (8, 30, 32). With the exception of one report (31), the lowest published values of FRC p in infants have been reported from the most recent studies (6 8, 32). This may reflect gradual improvement in equipment and techniques and the precision of recent calculations and corrections for the total compressible dead space. Nevertheless, all reported FRC p values remain significantly higher than those found in this study. The remaining discrepancies between these studies and the data from the present study may be explained by subtle differences between the equipment (particularly the amount of dead space and resistance), software, and protocol, which have not been accounted for or are difficult to quantify. For example, when using the Jaeger system, FRC p J is calculated from the integrated slope of both inspiratory and expiratory efforts, according to recent recommendations (1), whereas data from the Columbus group has traditionally been calculated from the inspiratory portion (6, 7), a practice which could introduce subtle differences, especially in the presence of any substantial drift of the box signal due to lack of complete thermal equilibrium. In addition, caution is required when making comparisons of weight-corrected FRC p between different studies, as disproportionate growth and comparison of results from cohorts with different age ranges might influence the magnitude of differences. Ideally, when collating reference data from several centers (41), multiple regression analysis should be applied taking into account anthropometric and ethnic background details as well as measurement protocols. This Jaeger system was extensively tested in vitro (10) before this study, and we have subsequently undertaken regular checks of both equipment and software as described in the online supplement. In addition to the routine procedures recommended by the manufacturers, all sensors were intermittently checked for accuracy and precision, and results were always within the accepted limits (see Table 1E in the online supplement). The integrity of the algorithms used for calculating FRC p J was cross checked by calculating FRC p manually from the slope of pressure at the airway opening versus box volume changes during respiratory efforts against the occlusion (see online supplement) and was found to be accurate within 1.5% (equivalent to a reading error of 0.5 mm from the printed plot). Rigorous criteria for quality control were implemented throughout this study (see online supplement), and measures of FRC p J were highly reproducible within individuals. We are thus reasonably confident that FRC p results obtained using the Jaeger equipment represent correct and unbiased estimates of lung volume in healthy infants and that these results could be reproduced by other groups using the same equipment, protocols, and criteria for quality control. In summary, a combination of errors resulting from a failure to adequately subtract all relevant dead space, including that of the facemask and any compressible volume proximal to the occlusion, together with failure to correct the occluded Vt signal to body temperature, pressure, and saturation before subtraction from the total occluded gas volume could easily account for the observed discrepancy between published reference data and those presented in this study. Clinical and Physiologic Implications This study has shown that published reference data (5) cannot be used to interpret results of FRC p from the new Jaeger equipment. Indeed, attempts to do so would run the risk of missing significant hyperinflation in infants with airway obstruction whose results would simply fall within the expected range (42). There is therefore an urgent need for collaborative studies of healthy infants using the same equipment and carefully standardized protocols for data collection, analysis, and quality control, so that new reference data can be established as rapidly as possible. Until such data are available, it will be necessary for users of this new equipment to interpret results very cautiously, bearing in mind that the normal range suggested by this study is likely to be of the order of 13 to 26 ml/kg. In addition, the lower values of FRC p J now reported suggest that the relationship between FRC measurements from plethysmographic and gas dilution techniques in infants needs to be reevaluated. In contrast to findings in healthy adults and older children, in whom results are very similar, plethysmographic lung volumes in infants have been consistently higher than those obtained by gas dilution (6, 33, 43). Therefore, separate prediction equations for the two techniques were recommended by the 1995 ATS report (5; Clausen and Wagner, submitted). The reason for this discrepancy has traditionally been attributed to the increased tendency to airway closure and hence gas trapping during tidal breathing, even in healthy infants, as a result of developmental differences including the high compliance of the chest wall and airways in early life (44) but in light of the current findings may be more attributable to technologic factors.
6 1008 AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE VOL Conclusions In conclusion, FRC p in healthy infants measured using recently released Jaeger equipment is significantly lower than those predicted from published reference data. Consequently, these reference data should not be used to interpret FRC p J results. Careful comparison of the equipment, software, and protocol used in this study with those from previous reports demonstrates the importance of appropriate subtraction of the total compressible dead space (including that of the mask) and any additional occluded volume when calculating FRC p in infants. These findings also support the hypothesis that diminished values of plethysmographic lung volumes in healthy infants primarily reflect gradual technologic advances and refinements in protocols over the last 30 years rather than any physiologic changes in lung growth and development in early life or differences in background characteristics of the tested infants. Furthermore, previous assumptions regarding the physiologic presence of trapped gas in healthy infants might need to be reconsidered. Collation of appropriate reference data in healthy infants remains a challenge not only for lung volumes but also for all other parameters of lung function in infants, which should be facilitated by adherence to recent ERS ATS recommendations (1) and by use of the improved and readily available contemporary equipment. Conflict of Interest Statement : G.H. has no declared conflict of interest; A-F.H. has no declared conflict of interest; H.L. has no declared conflict of interest; S.L. has no declared conflict of interest; J.S. has been reimbursed (travel and accommodation only) by Jaeger/Viasys Healthcare to attend the ATS and ERS international conferences during the past three years, and the Portex Respiratory Physiology Unit received a 30,000 Educational grant ( ) from Viasys Healthcare; J.J.P. has no declared conflict of interest. Acknowledgment : The authors gratefully acknowledge the ongoing support of Jane Hawdon and Kate Costeloe in allowing measurements to be undertaken in infants delivered at University College London Hospital and Homerton University Hospital, London, UK. References 1. Stocks J, Godfrey S, Beardsmore C, Bar-Yishay E, Castile R. Standards for infant respiratory function testing: plethysmographic measurements of lung volume and airway resistance. Eur Respir J 2001;17: Morris MG, Gustafsson P, Tepper R, Gappa M, Stocks J. Standards for infant respiratory function testing: the bias flow nitrogen washout technique for measuring the functional residual capacity. Eur Respir J 2001;17: Stocks J, Marchal F, Kraemer R, Gutkowski P, Bar-Yishay E, Godfrey S. Plethysmographic assessment of functional residual capacity and airway resistance. In: Stocks J, Sly PD, Tepper RS, Morgan WJ, editors. Infant respiratory function testing, 1st ed. New York: John Wiley & Sons; p Stocks J, Sly P, Morris MG, Frey U. Standards for infant respiratory function testing: what (ever) next? Eur Respir J 2000;16: Stocks J, Quanjer PH. Reference values for residual volume, functional residual capacity and total lung capacity. Eur Respir J 1995;8: McCoy KS, Castile RG, Allen ED, Filbrun DA, Flucke RL, Bar-Yishay E. Functional residual capacity (FRC) measurements by plethysmography and helium dilution in normal infants. Pediatr Pulmonol 1995;19: Castile R, Filbrun D, Flucke R, Franklin W, McCoy K. Adult-type pulmonary function tests in infants without respiratory disease. Pediatr Pulmonol 2000;30: Dezateux CA, Stocks J, Dundas I, Fletcher ME. Impaired airway function and the risk of wheezing in infancy: the impact of maternal smoking and a genetic predisposition to asthma. Am J Respir Crit Care Med 1999;159: American Thoracic Society. Lung function testing: selection of reference values and interpretative strategies. Am Rev Respir Dis 1991;144: Reinmann B, Stocks J, Frey U. Assessment of an infant whole-body plethysmograph using an infant lung function model. Eur Respir J 2001;17: Hulskamp G, Hoo A-F, Ljungberg H, Lum S, Pillow JJ, Stocks J. Plethysmographisch bestimmte funktionelle Residualkapazität (FRC) bei Säuglingen. Aktuelle Ergebnisse versus Normwerte [abstract]. Atemweg- Lungenkrankh 2003;29: Hulskamp G, Hoo A-F, Ljungberg H, Lum S, Pillow JJ, Stocks J. FRCpleth in healthy infants using new low deadspace equipment: comparison with published reference data [abstract]. Am J Respir Crit Care Med 2003;167:A Gaultier C, Fletcher M, Beardsmore C, Motoyama E, Stocks J. Measurement conditions. In: Stocks J, Sly PD, Tepper RS, Morgan WJ, editors. Infant respiratory function testing, 1st ed. New York: John Wiley & Sons, Inc.; p Morris MG. A simple new technique to measure the effective dead space of the face mask with a water volumeter in infants. Eur Respir J 1999;14: Frey U, Stocks J, Coates A, Sly P, Bates J. Standards for infant respiratory function testing: specifications for equipment used for infant pulmonary function testing. Eur Respir J 2000;16: Frey U, Stocks J, Sly P, Bates J. Specifications for signal processing and data handling used for infant pulmonary function testing. Eur Respir J 2000;16: Bland JM, Altman DJ. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; Stocks J, Godfrey S. Specific airway conductance in relation to postconceptional age during infancy. J Appl Physiol 1977;43: Lum S, Hoo AF, Dezateux C, Goetz I, Wade A, DeRooy L, Costeloe K, Stocks J. The association between birthweight, sex, and airway function in infants of nonsmoking mothers. Am J Respir Crit Care Med 2001;164: Greenough A, Hird MF, Everett L, Price JF. Importance of using lung function regression equations appropriate for ethnic origin. Pediatr Pulmonol 1991;11: Yuksel B, Greenough A. Ethnic origin and lung function in infants born prematurely. Thorax 1995;50: Dezateux C, Stocks J, Dundas I, Fletcher ME. Impaired airway function and wheezing in infancy: the influence of maternal smoking and a genetic predisposition to asthma. Am J Respir Crit Care Med 1999;159: Jones M, Castile R, Davis S, Kisling J, Filbrun D, Flucke R, Goldstein A, Emsley C, Ambrosius W, Tepper RS. Forced expiratory flows and volumes in infants. Am J Respir Crit Care Med 2000;161: Freeman JV, Cole TJ, Chinn S, Jones PRM, White EM, Preece MA. Cross sectional stature and weight reference curves for the UK, Arch Dis Child 1995;73: Doershuk CF, Downs TD, Matthews LW, Lough MD. A method for ventilatory measurements in subjects 1 month-5 years of age: normal results and observations in disease. Pediatr Res 1970;4: Doershuk CF, Matthews LW. Airway resistance and lung volume in newborn infants. Pediatr Res 1969;3: Hoo A-F, Henschen M, Dezateux CA, Costeloe KC, Stocks J. Respiratory function among preterm infants whose mothers smoked during pregnancy. Am J Respir Crit Care Med 1998;158: Hanrahan JP, Tager IB, Segal MR, Tosteson TD, Castile RG, Van VH, Weiss ST, Speizer FE. The effect of maternal smoking during pregnancy on early infant lung function. Am Rev Respir Dis 1992;145: Dezateux C, Hoo A-F, Cole TJ, Stocks J, Hanrahan J. Influence of maternal smoking on peripheral airway function during the first 18 months of life [abstract]. Eur Respir J 2002;20:176s. 30. Dundas I, Dezateux CA, Fletcher ME, Jackson EA, Stocks J. Comparison of single breath and plethysmographic measurements of resistance in infancy. Am J Respir Crit Care Med 1995;151: Milner AD, Marsh MJ, Ingram DM, Fox GF, Susiva C. Effects of smoking in pregnancy on neonatal lung function. Arch Dis Child Fetal Neonatal Ed 1999;80:F8 F Dezateux C, Stocks J, Wade AM, Dundas I, Fletcher ME. Airway function at one year: association with premorbid airway function, wheezing and maternal smoking. Thorax 2001;56: Gappa M, Fletcher ME, Dezateux CA, Stocks J. Comparison of nitrogen washout and plethysmographic measurements of lung volume in healthy infants. Am Rev Respir Dis 1993;148: Frey U, Stocks J, Coates A, Sly P, Bates J. Standards for infant respiratory function testing: specifications for equipment used for infant pulmonary function testing. Eur Respir J 2000;16: Stocks J, Levy NM, Godfrey S. A new apparatus for the accurate measurement of airway resistance in infancy. J Appl Physiol 1977;43: Lanteri CJ, Raven JM, Sly PD. Should TGV be measured from end-
7 Hülskamp, Hoo, Ljungberg, et al.: Infant Plethysmographic Lung Volume 1009 inspiratory occlusions rather than end-expiratory occlusions in wheezy infants? Pediatr Pulmonol 1990;9: Helms P. Problems with plethysmographic estimation of lung volume in infants and young children. J Appl Physiol 1982;53: Beardsmore CS, Stocks J, Silverman M. Problems in measurement of thoracic gas volume in infancy. J Appl Physiol 1982;52: Godfrey S, Beardsmore CS, Maayan C, Bar-Yishay E. Can thoracic gas volume be measured in infants with airway obstruction? Am Rev Respir Dis 1986;133: Godfrey S. TGV or not TGV in WB? That is the question. Pediatr Pulmonol 1991;10: Hoo AF, Dezateux C, Hanrahan J, Cole TJ, Tepper R, Stocks J. Sexspecific prediction equations for V maxfrc in infancy: a multi-center collaborative study. Am J Respir Crit Care Med 2002;165: Ljungberg H, Hulskamp G, Hoo A-F, Pillow JJ, Cantarella A, Gustafsson P, Stocks J. Early detection of peripheral airway involvement in infants with cystic fibrosis [abstract]. Pediatr Pulmonol 2002;(Suppl 24): Wauer RR, Maurer T, Nowotny T, Schmalisch G. Assessment of functional residual capacity using nitrogen washout and plethysmographic techniques in infants with and without bronchopulmonary dysplasia. Intensive Care Med 1998;24: Stocks J. Developmental physiology and methodology. Am J Respir Crit Care Med 1995;151:S15 S17.
RVRTC standardization repository 03 March
Title: Raised volume forced expirations in infants: guidelines for current practice Author: ATS/ERS Consensus statement On-line data supplement 03 March 2004 1 1. DETAILED MEASUREMENT PROTOCOL 1) The infant
More informationObjectives After completing this article, readers should be able to:
Article pulmonology Measurement of Forced Expiratory Flows and Lung Volumes Myrza R. Perez, MD,* Daniel J. Weiner, MD* Objectives After completing this article, readers should be able to: 1. Describe the
More informationAssessment of passive respiratory mechanics in infants: double versus single occlusion?
Eur Respir J 2001; 17: 449 455 Printed in UK all rights reserved Copyright #ERS Journals Ltd 2001 European Respiratory Journal ISSN 0903-1936 Assessment of passive respiratory mechanics in infants: double
More informationA collaborative study of infant respiratory function testing
Eur Respir J 1998; 12: 944 953 DOI: 1.1183/931936.98.124944 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1998 European Respiratory Journal ISSN 93-1936 A collaborative study of infant
More informationEarly detection of cystic fibrosis lung disease: multiple-breath washout versus raised volume tests
341 CYSTIC FIBROSIS Early detection of cystic fibrosis lung disease: multiple-breath washout versus raised volume tests Sooky Lum, Per Gustafsson, Henrik Ljungberg, Georg Hülskamp, Andrew Bush, Siobhán
More informationK.C. Lødrup Carlsen*, P. Magnus, K-H. Carlsen**
Eur Respir J, 1994, 7, 166 1668 DOI: 1.1183/931936.94.79166 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1994 European Respiratory Journal ISSN 93-1936 Lung function by tidal breathing
More informationAirway function at one year: association with premorbid airway function, wheezing, and maternal smoking
680 Centre for Paediatric Epidemiology and Biostatistics, Institute of Child Health, 30 Guilford Street, London WC1N 1EH, UK C Dezateux AMWade Portex Unit of Anaesthesia, Intensive Therapy and Respiratory
More informationEarly detection of cystic fibrosis lung disease: multiple-breath washout vs. raised volume tests
Thorax Online First, published on November 22, 2006 as 10.1136/thx.2006.068262 LCI in infants with CF Early detection of cystic fibrosis lung disease: multiple-breath washout vs. raised volume tests 1
More informationReference equations for the interpretation of forced expiratory and plethysmographic measurements in infants
Received: 20 February 2018 Accepted: 8 May 2018 DOI: 10.1002/ppul.24063 ORIGINAL ARTICLE: PULMONARY PHYSIOLOGY Reference equations for the interpretation of forced expiratory and plethysmographic measurements
More informationEvaluation of the interrupter technique in healthy, unsedated infants
Eur Respir J 21; 18: 982 988 Copyright #ERS Journals Ltd 21 DOI: 1.1183/931936.1.22151 European Respiratory Journal Printed in UK all rights reserved ISSN 93-1936 Evaluation of the interrupter technique
More informationOriginal Contributions
Original Contributions Comparison of a New Desktop Spirometer (Spirospec) with a Laboratory Spirometer in a Respiratory Out-Patient Clinic François Swart, Macé M Schuurmans MD, Johannes C Heydenreich,
More informationP ast studies have shown lung function abnormalities in
388 ORIGINAL ARTICLE Lung function at one month of age as a risk factor for infant respiratory symptoms in a high risk population C S Murray, S D Pipis, E C McArdle, L A Lowe, A Custovic, A Woodcock, on
More informationLung function during infancy and preschool age
Lung function during infancy and preschool age Educational aims To analyse lung function techniques developed for preschool children and infants. To compare lung function techniques used in adults and
More informationPULMONARY FUNCTION TESTING. Purposes of Pulmonary Tests. General Categories of Lung Diseases. Types of PF Tests
PULMONARY FUNCTION TESTING Wyka Chapter 13 Various AARC Clinical Practice Guidelines Purposes of Pulmonary Tests Is lung disease present? If so, is it reversible? If so, what type of lung disease is present?
More informationComparison of patient spirometry and ventilator spirometry
GE Healthcare Comparison of patient spirometry and ventilator spirometry Test results are based on the Master s thesis, Comparison between patient spirometry and ventilator spirometry by Saana Jenu, 2011
More informationForced expiratory manoeuvres
CHAPTER 4 Forced expiratory manoeuvres S. Lum, J. Stocks Portex Respiratory Unit, University College London, Institute of Child Health, London, UK. Correspondence: S. Lum, Portex Respiratory Unit, Respiratory
More informationSmall Size at Birth and Greater Postnatal Weight Gain Relationships to Diminished Infant Lung Function
Small Size at Birth and Greater Postnatal Weight Gain Relationships to Diminished Infant Lung Function Jane S. Lucas, Hazel M. Inskip, Keith M. Godfrey, Claire T. Foreman, John O. Warner, Rachael K. Gregson,
More informationPREDICTION EQUATIONS FOR LUNG FUNCTION IN HEALTHY, LIFE TIME NEVER-SMOKING MALAYSIAN POPULATION
Prediction Equations for Lung Function in Healthy, Non-smoking Malaysian Population PREDICTION EQUATIONS FOR LUNG FUNCTION IN HEALTHY, LIFE TIME NEVER-SMOKING MALAYSIAN POPULATION Justin Gnanou, Brinnell
More informationTidal expired airflow patterns in adults with airway obstruction
Eur Respir J 1998; 12: 1118 1123 DOI: 1.1183/931936.98.1251118 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1998 European Respiratory Journal ISSN 93-1936 Tidal expired airflow patterns
More informationCHAPTER 3. Correction factors for oxygen and flow-rate effects on neonatal Fleisch and Lilly pneumotachometers
CHAPTER 3 Correction factors for oxygen and flow-rate effects on neonatal Fleisch and Lilly pneumotachometers Yvonne Snepvangers Peter de Winter Huibert Burger Hens Brouwers Jan Bogaard Kors van der Ent
More informationSubject Index. Carbon monoxide (CO) disease effects on levels 197, 198 measurement in exhaled air 197 sources in exhaled air 197
Subject Index Airway resistance airflow interruption measurement in preschoolers, see Forced oscillation technique; Interrupter technique plethysmography, see Plethysmography; Whole-body plethysmography
More informationbut did decrease TEF10 and 25/PT.
Critical Care April 2001 Vol 6 No 2 Totapally et al. Research Tidal breathing flow-volume loops in bronchiolitis in infancy: the effect of albuterol [ISRCTN47364493] Balagangadhar R Totapally*, Cem Demerci,
More informationThe role of Pulmonary function Testing In Interstitial lung disease in infants. [ ipft in child ]
The role of Pulmonary function Testing In Interstitial lung disease in infants [ ipft in child ] Introduction Managing infants with diffuse lung disease (DLD) suspected to have interstitial lung disease
More informationIn utero exposure to cigarette smoking influences lung function at birth
Eur Respir J 1997; 10: 1774 1779 DOI: 10.1183/09031936.97.10081774 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1997 European Respiratory Journal ISSN 0903-1936 In utero exposure to cigarette
More informationVariation in airways resistance when defined over different ranges of airflows
Thorax, 1978, 33, 401-405 Variation in airways resistance when defined over different ranges of airflows P. W. LORD AND J. M. EDWARDS From the MRC Toxicology Unit, Clinical Section, St. Bartholomew's Hlospital
More informationEffect of nasogastric tubes on nasal resistance during infancy
Archives of Disease in Childhood, 198, 55, 17-21 Effect of nasogastric tubes on nasal resistance during infancy JANET STOCKS Department ofpaediatrics and Neonatal Medicine, Institute of Child Health, Hammersmith
More informationLung function and clinical risk factors for asthma in infants and young children with recurrent wheeze
Thorax Online First, published on November 13, 2008 as 10.1136/thx.2008.099903 Lung function and clinical risk factors for asthma in infants and young children with recurrent wheeze Luis Miguel Borrego
More informationR egular lung function testing is an essential part of
1068 CYSTIC FIBROSIS Multiple breath inert gas washout as a measure of ventilation distribution in children with cystic fibrosis P Aurora, P Gustafsson, A Bush, A Lindblad, C Oliver, C E Wallis, J Stocks...
More informationAssessment of accuracy and applicability of a new electronic peak flow meter and asthma monitor
Eur Respir J 18; 12: 45 42 DOI: 18/1.8.5 Printed in UK - all rights reserved Copyright ERS Journals Ltd 18 European Respiratory Journal ISSN - 1 Assessment of accuracy and applicability of a new electronic
More informationHow should airways resistance be measured in young children: mask or mouthpiece?
Eur Respir J 2001; 17: 1244 1249 Printed in UK all rights reserved Copyright #ERS Journals Ltd 2001 European Respiratory Journal ISSN 0903-1936 How should airways resistance be measured in young children:
More informationAlterations in Airway Wall Properties in Infants with a History of Wheezing Disorders
Alterations in Airway Wall Properties in Infants with a History of Wheezing Disorders URS FREY, KATRI MAKKONEN, TERRY WELLMAN, CAROLINE BEARDSMORE, and MICHAEL SILVERMAN Department of Child Health, University
More informationUnderstanding the Basics of Spirometry It s not just about yelling blow
Understanding the Basics of Spirometry It s not just about yelling blow Carl D. Mottram, RRT RPFT FAARC Technical Director - Pulmonary Function Labs and Rehabilitation Associate Professor of Medicine -
More informationPULMONARY FUNCTION TESTS
Chapter 4 PULMONARY FUNCTION TESTS M.G.Rajanandh, Department of Pharmacy Practice, SRM College of Pharmacy, SRM University. OBJECTIVES Review basic pulmonary anatomy and physiology. Understand the reasons
More informationTeacher : Dorota Marczuk Krynicka, MD., PhD. Coll. Anatomicum, Święcicki Street no. 6, Dept. of Physiology
Title: Spirometry Teacher : Dorota Marczuk Krynicka, MD., PhD. Coll. Anatomicum, Święcicki Street no. 6, Dept. of Physiology I. Measurements of Ventilation Spirometry A. Pulmonary Volumes 1. The tidal
More informationSniff nasal inspiratory pressure in patients with chronic obstructive pulmonary disease
Eur Respir J 1997; 1: 1292 1296 DOI: 1.1183/931936.97.161292 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1997 European Respiratory Journal ISSN 93-1936 Sniff nasal inspiratory pressure
More informationbehaviour are out of scope of the present review.
explained about the test, a trial may be done before recording the results. The maneuver consists initially of normal tidal breathing. The subject then inhales to maximally fill the lungs. This is followed
More informationDifference Between The Slow Vital Capacity And Forced Vital Capacity: Predictor Of Hyperinflation In Patients With Airflow Obstruction
ISPUB.COM The Internet Journal of Pulmonary Medicine Volume 4 Number 2 Difference Between The Slow Vital Capacity And Forced Vital Capacity: Predictor Of Hyperinflation In Patients With Airflow Obstruction
More informationVALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS - A PROSPECTIVE COHORT STUDY
Diabetes Care Publish Ahead of Print, published online March 11, 2009 Validation of CGM in Children with CF VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS
More informationChapter 3. Pulmonary Function Study Assessments. Mosby items and derived items 2011, 2006 by Mosby, Inc., an affiliate of Elsevier Inc.
Chapter 3 Pulmonary Function Study Assessments 1 Introduction Pulmonary function studies are used to: Evaluate pulmonary causes of dyspnea Differentiate between obstructive and restrictive pulmonary disorders
More informationEndpoints for Clinical Trials in Young Children with Cystic Fibrosis
Endpoints for Clinical Trials in Young Children with Cystic Fibrosis Stephanie D. Davis 1, Alan S. Brody 2, Mary J. Emond 3, Lyndia C. Brumback 3, and Margaret Rosenfeld 4 1 Division of Pulmonology, Department
More informationLung Clearance Index and HRCT are complementary markers of lung abnormalities in young children with CF
< Additional tables and figures are published online only. To view these files please visit the journal online (http://thorax.bmj. com). 1 Department of Paediatric Radiology, Great Ormond Street Hospital
More informationK-H. Carlsen*, K.C.Lødrup Carlsen**
Eur Respir J, 1994, 7, 2154 2159 DOI: 10.1183/09031936.94.07122154 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1994 European Respiratory Journal ISSN 0903-1936 Tidal breathing analysis
More informationNew time-saving predictor algorithm for multiple breath washout in adolescents
nature publishing group Clinical Investigation New time-saving predictor algorithm for multiple breath washout in adolescents Jonathan Grønbæk 1, Henrik Wegener Hallas 1, Lambang Arianto 1, Knud Pedersen
More informationUNIT TWO: OVERVIEW OF SPIROMETRY. A. Definition of Spirometry
UNIT TWO: OVERVIEW OF SPIROMETRY A. Definition of Spirometry Spirometry is a medical screening test that measures various aspects of breathing and lung function. It is performed by using a spirometer,
More informationDiminished lung function, RSV infection and respiratory morbidity in. prematurely born infants
ADC Online First, published on September 27, 2005 as 10.1136/adc.2005.087270 Diminished lung function, RSV infection and respiratory morbidity in prematurely born infants Simon Broughton, Ravi Bhat, Alison
More informationTechnical Assessment of Spirometers Connected in Series
Technical Assessment of Spirometers Connected in Series Quentin Lefebvre, Thomas Vandergoten, Eric Derom MD PhD, Emilie Marchandise PhD, and Giuseppe Liistro MD PhD BACKGROUND: Office spirometers are now
More informationUp to 50% of continuous flow oxygen therapy patients experience clinically significant nocturnal desaturation. 1
Up to 50% of continuous flow oxygen therapy patients experience clinically significant nocturnal desaturation. 1 Continuous Flow Oxygen Delivery & Sleep A number of theories and studies are published surrounding
More informationNon-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis
Non-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis Gal S, Riskin A, Chistyakov I, Shifman N, Srugo I, and Kugelman A Pediatric Department and Pediatric Pulmonary Unit Bnai Zion
More informationHHS Public Access Author manuscript J Cyst Fibros. Author manuscript; available in PMC 2017 May 01.
INFANT LUNG FUNCTION TESTS AS ENDPOINTS IN THE ISIS MULTICENTER CLINICAL TRIAL IN CYSTIC FIBROSIS Stephanie D. Davis 1,*, Felix Ratjen 2, Lyndia C. Brumback 3, Robin C. Johnson 4, Amy G. Filbrun 5, Gwendolyn
More informationMETHODS. Subjects. Lung Function
Worsening of V maxfrc in Infants with Chronic Lung Disease in the First Year of Life A More Favorable Outcome after High-Frequency Oscillation Ventilation Ward Hofhuis, Marianne W. A. Huysman, Els C. van
More informationSpirometry and Flow Volume Measurements
Spirometry and Flow Volume Measurements Standards & Guidelines December 1998 To serve the public and guide the medical profession Revision Dates: December 1998 Approval Date: June 1998 Originating Committee:
More informationMai ElMallah,MD Updates in Pediatric Pulmonary Care XII: An Interdisciplinary Program April 13, 2012
Mai ElMallah,MD Updates in Pediatric Pulmonary Care XII: An Interdisciplinary Program April 13, 2012 Recognize the importance of Pulmonary Function Testing in Cystic Fibrosis Be aware of different types
More informationfirst three years of life
Journal of Epidemiology and Community Health, 1981, 35, 18-184 Parental smoking and lower respiratory illness in the first three years of life D. M. FERGUSSON, L. J. HORWOOD, F. T. SHANNON, AND BRENT TAYLOR
More informationImprovement of tidal breathing pattern analysis in children with asthma by on-line automatic data processing
Eur Respir J, 1996, 9, 136 1313 DOI: 1.1183/931936.96.96136 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1996 European Respiratory Journal ISSN 93-1936 TECHNICAL NOTE Improvement of tidal
More informationSPIROMETRY. Marijke Currie (CRFS) Care Medical Ltd Phone: Copyright CARE Medical ltd
SPIROMETRY Marijke Currie (CRFS) Care Medical Ltd Phone: 0800 333 808 Email: sales@caremed.co.nz What is spirometry Spirometry is a physiological test that measures the volume of air an individual can
More informationDifferences in airway wall compliance as a possible mechanism for wheezing disorders in infants
Eur Respir J 1; 1: 1 1 DOI:.11/1..1 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1 European Respiratory Journal ISSN - 1 Differences in airway wall compliance as a possible mechanism
More informationLung elastic recoil during breathing at increased lung volume
Lung elastic recoil during breathing at increased lung volume JOSEPH R. RODARTE, 1 GASSAN NOREDIN, 1 CHARLES MILLER, 1 VITO BRUSASCO, 2 AND RICCARDO PELLEGRINO 3 (With the Technical Assistance of Todd
More informationLung elastic recoil during breathing at increased lung volume
Lung elastic recoil during breathing at increased lung volume JOSEPH R. RODARTE, 1 GASSAN NOREDIN, 1 CHARLES MILLER, 1 VITO BRUSASCO, 2 AND RICCARDO PELLEGRINO 3 (With the Technical Assistance of Todd
More informationMicrorint pulmonary function testing in older adults with an intellectual disability
Respiratory Medicine (2009) 103, 1954e1959 available at www.sciencedirect.com journal homepage: www.elsevier.com/locate/rmed Microrint pulmonary function testing in older adults with an intellectual disability
More informationVariation in lung with normal, quiet breathing. Minimal lung volume (residual volume) at maximum deflation. Total lung capacity at maximum inflation
r Total lung capacity at maximum inflation Variation in lung with normal, quiet breathing Volume of lungs at end of normal inspiration (average 2,200 ml) Minimal lung volume (residual volume) at maximum
More informationLung function from infancy to school age in cystic fibrosis
Archives of Disease in Childhood 1995; 73: 519-523 Department of Child Health, University of Leicester, Robert Kilpatrick Clinical Sciences Building, Leicester Royal Infirmary, PO Box 65, Leicester LE2
More informationCochrane Pregnancy and Childbirth Group Methodological Guidelines
Cochrane Pregnancy and Childbirth Group Methodological Guidelines [Prepared by Simon Gates: July 2009, updated July 2012] These guidelines are intended to aid quality and consistency across the reviews
More informationThe recommended method for diagnosing sleep
reviews Measuring Agreement Between Diagnostic Devices* W. Ward Flemons, MD; and Michael R. Littner, MD, FCCP There is growing interest in using portable monitoring for investigating patients with suspected
More informationCopyright, 1995, by the Massachusetts Medical Society. Volume 332 JANUARY 19, 1995 Number 3
Copyright, 1995, by the Massachusetts Medical Society Volume 332 JANUARY 19, 1995 Number 3 ASTHMA AND WHEEZING IN THE FIRST SIX YEARS OF LIFE FERNANDO D. MARTINEZ, M.D., ANNE L. WRIGHT, PH.D., LYNN M.
More informationPFT Interpretation and Reference Values
PFT Interpretation and Reference Values September 21, 2018 Eric Wong Objectives Understand the components of PFT Interpretation of PFT Clinical Patterns How to choose Reference Values 3 Components Spirometry
More informationOffice Based Spirometry
Osteopathic Family Physician (2014)1, 14-18 Scott Klosterman, DO; Woodson Crenshaw, OMS4 Spartanburg Regional Family Medicine Residency Program; Edward Via College of Osteopathic Medicine - Virginia Campus
More informationCharisma High-flow CPAP solution
Charisma High-flow CPAP solution Homecare PNEUMOLOGY Neonatology Anaesthesia INTENSIVE CARE VENTILATION Sleep Diagnostics Service Patient Support charisma High-flow CPAP solution Evidence CPAP therapy
More informationPULMONARY FUNCTION TEST(PFT)
PULMONARY FUNCTION TEST(PFT) Objectives: By the end of the present lab, students should be able to: 1. Record lung volumes and capacities and compare them with those of a typical person of the same gender,
More informationVentilatory Mechanics in Patients with Cardio-Pulmonary Diseases. Part III. On Pulmonary Fibrosis
Ventilatory Mechanics in Patients with Cardio-Pulmonary Diseases Part III. On Pulmonary Fibrosis Kazuaki SERA, M.D. Pulmonary function studies have been undertaken on the pulmonary fibrosis as diagnosed
More informationRecognizing and Correcting Patient-Ventilator Dysynchrony
2019 KRCS Annual State Education Seminar Recognizing and Correcting Patient-Ventilator Dysynchrony Eric Kriner BS,RRT Pulmonary Critical Care Clinical Specialist MedStar Washington Hospital Center Washington,
More informationWhat do pulmonary function tests tell you?
Pulmonary Function Testing Michael Wert, MD Assistant Professor Clinical Department of Internal Medicine Division of Pulmonary, Critical Care, and Sleep Medicine The Ohio State University Wexner Medical
More informationAEROSURF Phase 2 Program Update Investor Conference Call
AEROSURF Phase 2 Program Update Investor Conference Call November 12, 2015 Forward Looking Statement To the extent that statements in this presentation are not strictly historical, including statements
More informationLUNG FUNCTION MEASUREMENT INSTRUMENT
LUNG FUNCTION MEASUREMENT INSTRUMENT THE WHISTLER IS THE FIRST HANDHELD, WIRELESS DEVICE IN THE WORLD FOR PERFORMING PULMONARY FUNCTION TESTS WITH BOTH BABIES AND PRESCHOOL CHILDREN 2 3 THE WHISTLER MAKES
More informationResistance-Volume Chart (R-V Chart) for simplified clinical and visual interpretation of body plethysmography
Resistance-Volume Chart (R-V Chart) for simplified clinical and visual interpretation of body plethysmography Hans-Jürgen Smith CareFusion, Höchberg Resistance-Volume Chart Innovation in bodyplethysmography
More informationDec Total Lung Capacity with the MiniBox TM : Clinical Results in Obstructed, Restrictive, and Healthy Adults. White Paper
Dec 01 Total Lung Capacity with the MiniBox TM : Clinical Results in Obstructed, Restrictive, and Healthy Adults White Paper Total Lung Capacity with the MiniBox TM : Clinical Results in Obstructed, Restrictive,
More informationKing s Research Portal
King s Research Portal DOI: 10.1007/s00431-015-2595-4 Document Version Peer reviewed version Link to publication record in King's Research Portal Citation for published version (APA): Shetty, S., Bhat,
More informationnormal and asthmatic males
Lung volume and its subdivisions in normal and asthmatic males MARGARET I. BLACKHALL and R. S. JONES1 Thorax (1973), 28, 89. Institute of Child Health, University of Liverpool, Alder Hey Children's Hospital,
More informationAddendum to the NRP Provider Textbook 6 th Edition Recommendations for specific modifications in the Canadian context
Addendum to the NRP Provider Textbook 6 th Edition Recommendations for specific modifications in the Canadian context A subcommittee of the Canadian Neonatal Resuscitation Program (NRP) Steering Committee
More informationAn Overview of Bronchopulmonary Dysplasia and Chronic Lung Disease in Infancy
An Overview of Bronchopulmonary Dysplasia and Chronic Lung Disease in Infancy Housekeeping: I have no financial disclosures Learning objectives: Develop an understanding of bronchopulmonary dysplasia (BPD)
More informationQuality Assurance Mapping Your QC Program Equipment and Test Quality. Susan Blonshine RRT, RPFT, FAARC, AE-C
Quality Assurance Mapping Your QC Program Equipment and Test Quality Susan Blonshine RRT, RPFT, FAARC, AE-C How to Begin Gather resources Define PF scope of service Procedures performed Equipment Describe
More informationAssessing Agreement Between Methods Of Clinical Measurement
University of York Department of Health Sciences Measuring Health and Disease Assessing Agreement Between Methods Of Clinical Measurement Based on Bland JM, Altman DG. (1986). Statistical methods for assessing
More informationGE Healthcare. Non Invasive Ventilation (NIV) For the Engström Ventilator. Relief, Relax, Recovery
GE Healthcare Non Invasive Ventilation (NIV) For the Engström Ventilator Relief, Relax, Recovery COPD is currently the fourth leading cause of death in the world, and further increases in the prevalence
More information1. When a patient fails to ventilate or oxygenate adequately, the problem is caused by pathophysiological factors such as hyperventilation.
Chapter 1: Principles of Mechanical Ventilation TRUE/FALSE 1. When a patient fails to ventilate or oxygenate adequately, the problem is caused by pathophysiological factors such as hyperventilation. F
More informationA comparison of global questions versus health status questionnaires as measures of the severity and impact of asthma
Eur Respir J 1999; 1: 591±596 Printed in UK ± all rights reserved Copyright #ERS Journals Ltd 1999 European Respiratory Journal ISSN 93-1936 A comparison of global questions versus health status questionnaires
More informationAbnormalities of plethysmographic lung volumes in asthmatic children
Respiratory Medicine (2010) 104, 966e971 available at www.sciencedirect.com journal homepage: www.elsevier.com/locate/rmed Abnormalities of plethysmographic lung volumes in asthmatic children Bruno Mahut
More informationThe Influence of Altered Pulmonarv
The Influence of Altered Pulmonarv J Mechanics on the Adequacy of Controlled Ventilation Peter Hutchin, M.D., and Richard M. Peters, M.D. W ' hereas during spontaneous respiration the individual determines
More informationLung Function Reference Values in Chinese Children and Adolescents in Hong Kong I. Spirometric Values and Comparison with Other Populations
Lung Function Reference Values in Chinese Children and Adolescents in Hong Kong I. Spirometric Values and Comparison with Other Populations MARY S. M. IP, EVA M. KARLBERG, JOHAN P. E. KARLBERG, KEITH D.
More informationAnnual lung function changes in young patients with chronic lung disease
Eur Respir J 22; 19: 886 891 DOI:.1183/931936.2.2492 Printed in UK all rights reserved Copyright #ERS Journals Ltd 22 European Respiratory Journal ISSN 93-1936 Annual lung function changes in young patients
More informationDevelopment of a self-reported Chronic Respiratory Questionnaire (CRQ-SR)
954 Department of Respiratory Medicine, University Hospitals of Leicester, Glenfield Hospital, Leicester LE3 9QP, UK J E A Williams S J Singh L Sewell M D L Morgan Department of Clinical Epidemiology and
More informationDependence of forced vital capacity manoeuvre on time course of preceding inspiration in patients with restrictive lung disease
Eur Respir J 1997; 1: 2366 237 DOI: 1.1183/931936.97.112366 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1997 European Respiratory Journal ISSN 93-1936 Dependence of forced vital capacity
More informationRelationship between parental lung function and their children s lung function early in life
Eur Respir J 2011; 38: 664 671 DOI: 10.1183/09031936.00034210 CopyrightßERS 2011 Relationship between parental lung function and their children s lung function early in life N. van Putte-Katier*, M. Koopmans*,
More informationSTOP SMOKING with Nicotine Replacement Therapy
STOP SMOKING with Nicotine Replacement Therapy Gita Anjali Shah, Clinical Pharmacist, UK and Guest Lecturer, Faculty of Pharmacy, Mahidol University Shocking Statistics According to the last Global Adult
More informationLung Pathophysiology & PFTs
Interpretation of Pulmonary Function Tests (PFTs) Course # 1612 2:00 5:00pm Friday February 22, 2013 Lung Pathophysiology & PFTs Mark F. Sands MD, FCCP, FAAAAI Division of Allergy, Immunology & Rheumatology
More informationOutline FEF Reduced FEF25-75 in asthma. What does it mean and what are the clinical implications?
Reduced FEF25-75 in asthma. What does it mean and what are the clinical implications? Fernando Holguin MD MPH Director, Asthma Clinical & Research Program Center for lungs and Breathing University of Colorado
More informationFaculty Disclosure. Off-Label Product Use
Faculty Disclosure X No, nothing to disclose Yes, please specify: Company Name Honoraria/ Expenses Consulting/ Advisory Board Funded Research Royalties/ Patent Stock Options Equity Position Ownership/
More information