1 Detection of Vertical Root Fractures by Using Cone Beam Computed Tomography with Variable Voxel Sizes in an In Vitro Model Senem Yig it Özer, DDS, PhD Abstract Introduction: The aim of this study was to compare the diagnostic accuracy of cone beam computed tomography (CBCT) scans with different voxel resolutions in the detection of simulated vertical root fracture (VRF). Methods: Sixty teeth, 30 with root fractures and 30 without, were examined through i-cat tomography at 4 different voxel resolutions (0.125, 0.2, 0.3, 0.4 mm). Three observers scored the presence of VRF in all modalities. Sensitivity, specificity, likelihood ratio, and accuracy were calculated for each modality. Results: Chi-square statistical analysis showed no significant difference among the resolutions or observers. However, positive likelihood ratio values of 24.5 for a mm voxel, for a 0.2-mm voxel, 13.2 for a 0.3-mm voxel, and 13 for a 0.4-mm voxel were found. Accuracy results confirmed the likelihood ratios as 0.97, 0.96, 0.93, and 0.92, respectively. Conclusions: CBCT scans were reliable in detecting simulated VRF, and a 0.2-mm voxel was the best protocol, considering the lower x-ray exposure and good diagnostic performance. (J Endod 2011;37:75 79) Key Words Cone-beam computed tomography, diagnosis, vertical root fracture, voxel From the Department of Operative Dentistry and Endodontics, Faculty of Dentistry, Dicle University, Diyarbakir, Turkey. Address requests for reprints to Dr Senem Yiğit Özer, Dicle University, Faculty of Dentistry, Department of Operative Dentistry and Endodontics, 21280, Diyarbakir, Turkey. address: senemygt@hotmail.com /$0 - see front matter Copyright ª 2011 American Association of Endodontists. doi: /j.joen Vertical root fractures (VRFs) are difficult to diagnose by conventional periapical radiography (CPR), and they can be overlooked if the x-ray beam does not pass along the fracture line. Thus, 2 or more radiographs taken at different angles are recommended (1, 2). Two-dimensional (2D) intraoral radiography systems fail to provide information regarding teeth and adjacent structures in the third dimension, and the superimposition of other structures further limits sensitivity (3). Thus, a need exists for an imaging system that provides 3-dimensional (3D) information in treatment assessment of root canal systems. Recently a 3D diagnostic imaging system, cone beam computed tomography (CBCT), has been reported to be more accurate than CPR for detecting longitudinal root fractures with high image quality, permitting direct visualization of fracture lines, which are masked in CPR (4, 5). CBCT scans have been found to be more accurate in detecting VRFs in many studies (5 10). CBCT has been found to perform better than CPR and other intraoral techniques in detecting VRFs; however, many studies have concluded that the radiation doses from CBCT were higher than from conventional dental radiography techniques (11). Briefly, digital images are composed of elements called pixels. A pixel is a picture element, a square in a 2D matrix. Each pixel has a specific size, intensity value, and location within the matrix. A voxel is a volume element, and it is the cubed form of a pixel, having a third dimension. Three-dimensional images are composed of voxels instead of the pixels in 2D digital images. The size of each voxel is determined by its height, width, and thickness, and a voxel is the smallest element of the 3D radiograph image volume (12, 13). CBCT images use isotropic voxels, meaning they have the same height, width, and thickness. The isotropic nature of the voxels affords the same quality as the original image in reconstructions. Image quality has been described as the visibility of diagnostically important structures in the CT image (14, 15), and a number of factors affect image quality such as the milliamperage settings, voxel size, field of view (FoV), and detector type of the scanner (15 17). Diagnosis with CBCT offers 3D images without superposition of adjacent structures, and it is also a noninvasive technique with high accuracy (18 23). However, the radiation dose is equivalent to that needed for 4 15 panoramic radiographs (11),andunfortunately,imagequalityisdirectlyrelatedtoradiationexposure (24). Voxel size has been reported to have a positive correlation with image quality (contrast and resolution) and exposure dose (6, 25). ALARA is the acronym for as low as reasonably achievable and is a basic principle for diagnostic radiology. Keeping the dose as low as possible while still obtaining the needed information constitutes the basis of this principle (26). It is known that with smaller voxel sizes, radiation exposure would be higher (12, 24, 27). Without sacrificing image quality and adopting the ALARA principle, the ability to select various voxel settings would be helpful in reducing the radiation dose to the patient. Thus, we compared the diagnostic accuracy of different voxel sizes for the detection of experimentally induced VRFs. Materials and Methods The experimental group consisted of 60 recently extracted human maxillary premolars, inspected with a stereomicroscope (MT4200 binocular; Meiji Techno JOE Volume 37, Number 1, January 2011 Detection of Vertical Root Fractures with CBCT with Variable Voxel Sizes 75
2 TABLE 1. VRFs Identified from a Total of 60 Teeth with Different Voxel Resolutions by 3 Observers Observer 1 Observer 2 Observer 3 Total Voxel size (mm) True False True False True False True False X 2 test P =.7 P =.06 Significance of c 2 test for observers voxels is defined as P value. (UK) Ltd, Axbridge, Somerset, UK) for the absence of VRFs. An access opening was made for each tooth, and the root canals were prepared with the ProTaper rotary system (Dentsply/Maillefer, Tulsa, OK) up to size F3. The teeth were divided into 2 groups of 30, with root fractures or intact. Root fractures were generated in the vertical plane according to a protocol reported in the literature (6). Five tooth roots broke into more than 3 fragments and were excluded from the study. Two fragments for 30 teeth were relocated with glue (Scotch Gel Universal; 3M, Leiden, The Netherlands). One human mandible was coated with 3 layers of dental wax buccally and lingually to simulate soft tissue and to reduce artifacts in the image (28). An observer who was not involved in preparation placed the teeth into sockets of the mandible at the most suitable position mesiodistally. They were held in place in the alveolus socket with boxing wax and were coded during imaging sessions. CBCT Scans and Data Treatment All images were evaluated by 3 observers (2 endodontists and 1 radiologist) at different sessions. CBCT images were scanned by using i-cat CBCT (120 KvP, 5 ma; Imaging Sciences, Hatfield, PA). Scans of the mandible were made with 4-cm FoV selection at 4 resolutions: mm voxel (4 seconds for acquisition), 0.2-mm voxel (4 seconds for acquisition), 0.3-mm voxel (7 seconds for acquisition), and 0.4- mm voxel (7 seconds for acquisition). Four-second and sevensecond acquisition periods with shorter scanning time were selected to reduce radiation dose, even though i-cat has longer scanning acquisition times with the same voxel sizes. After each scan with different voxel sizes, the scanner was calibrated. The mandible was placed on a horizontal plane and fitted to the chin support, with its occlusal plane parallel to the plate. By using the reference light lines in displaying the x and y coordinates, the chin support was moved upward, so that the y- line was placed between the central incisors sockets. Data sets were in DICOM 3 file format. The images were analyzed with the software for the CBCT machine (XoranCAT, version ; Xoran Technologies, Ann Arbor, MI). The images for each tooth in the 4 resolutions were reconstructed separately. All images were displayed on a 19-inch flat-panel screen (EIZO Flexscan S inch LCD monitor; Eizo Nanao Corp, Ishikawa, Japan). Detection of VRFs was decided by direct visualization of a radiolucent line on the root surface. The image enhancement mode of the software was not used for standardization of image quality. Observers had been trained and calibrated for reading CBCT in a previous pilot study. Each observer evaluated the 4 voxel resolutions (0.125, 0.2, 0.3, 0.4 mm) by using axial, sagittal, and coronal directions, and a single score was obtained for each tooth. Observers were not informed that some teeth were intact without fractures. Observation time was not restricted, and at a 1-week interval, the same observations were repeated. Weighted kappa coefficients were calculated to assess intraobserver and interobserver agreement. The 4 voxel resolutions (0.125, 0.2, 0.3, 0.4 mm) were analyzed independently by c 2 test at 5% significance. Sensitivity, specificity, positive likelihood ratio, and accuracy were evaluated for diagnostic performance according to the different voxel resolutions. The fracture/non-fracture assessment was scored binarily: 1 if a fracture was present or 0 if it was not. Fracture/non-fracture scores were evaluated as follows: correct identification of a non-fractured root (true negative [TN]), correct identification of fracture site in a fractured root (true positive [TP]), identification of a fracture in a nonfractured root (false positive [FP]), incorrect identification of a fracture site in a fractured root (false negative [FN]), and failure to identify a fracture in a fractured root (false negative [FN]). Sensitivity and specificity among the 4 modalities (0.125-, 0.2-, 0.3-, 0.4-mm voxel) were calculated as follows: Sensitivity = (TP/TP) + FN, Specificity = (TN/TN) + HP, Positive Predictive Value = (TP/TP) + FP, Negative Predictive Value = (TN/TN) + FN, Accuracy = (TP + TN)/(TP + TN + FP + FN), Positive Likelihood Ratio = Sensitivity/(1 Specificity). Results Intraobserver kappa values were above 0.86, indicating very good intraobserver agreement. Table 1 shows the number of true (TP + TN) and false (FP + FN) readings among the 3 observers and 4 different voxel resolutions. On the basis of the observed c 2 test results, no difference (P =.06) was found among voxel sizes or observers (P =.7). Table 2 shows the results of diagnostic performance tests according to voxel resolution. The results indicate that voxel sensitivity and specificity values were similar; the highest accuracy was attained with mm (97%) and in descending order, 0.2 mm (96%), 0.3 mm TABLE 2. Results of Diagnostic Performance Test with Different Variables at 4 Voxel Resolutions Voxel size (mm) Test Sensitivity Specificity Positive predictive value Negative predictive value Accuracy (%) Positive likehood ratio (%) Yiğit Özer JOE Volume 37, Number 1, January 2011
3 Figure 1. CBCT images of a fractured root with 4 different voxels in 3 planes. (1a) Axial plane, mm voxel; (1b) sagittal plane, mm voxel; (1c) coronal plane, mm voxel; (2a) axial plane, 0.2-mm voxel; (2b) sagittal plane, 0.2-mm voxel; (2c) coronal plane, 0.2-mm voxel; (3a) axial plane, 0.3-mm voxel; (3b) sagittal plane, 0.3-mm voxel; (3c) coronal plane, 0.3-mm voxel; (4a) axial plane, 0.4-mm voxel; (4b) sagittal plane, 0.4-mm voxel; (4c) coronal plane, 0.4-mm voxel. (93%), and 0.4 mm (92%). Furthermore, the positive likelihood ratio results are better with mm and 0.2-mm voxel sizes (24% 24.25%) than with 0.3-mm and 0.4-mm voxel sizes (13% 13.2%). Discussion The diagnosis of VRF is usually problematic because it often necessitates prediction rather than a definite identification (28). Because radiographs and CBCT scans acquired by using different imaging parameters cannot exactly visualize these fractures (29, 30), the location and defect size cannot always be objectively assessed until the tooth has been extracted or a simultaneous flap operation has been performed for direct visualization. Also, determining the position and extent of the fracture might be helpful to decide when to recommend extraction or treatment possibilities. The detection of VRF is usually not performed JOE Volume 37, Number 1, January 2011 Detection of Vertical Root Fractures with CBCT with Variable Voxel Sizes 77
4 during treatment, which leads to bone loss, pain, and malfunction of the involved area discomforting the patient maybe for years (31). Thus, identification of the problem in early stages as soon as the problem gets more complex is quite important. The present study compared the diagnostic accuracy of CBCT scans with different voxel resolutions in the detection of simulated VRFs. The dental CBCT, i-cat, chosen for this study has been shown to be more accurate than digital and conventional 2D intraoral techniques in detecting VRFs (5 7, 10). Hassan et al (6) compared 5 different CBCTs for the detection of VRFs and reported that the Next Generation i-cat was the most accurate system (with a voxel of 0.25 mm), followed by the Scanora 3D (voxel of 0.2 mm), Accuitomo-XYZ (voxel of 0.25 mm), Newtom 3G (voxel of 0.2 mm), and Galileos 3D (voxel of 0.3 mm). Generally, performance was better with 0.25-mm and 0.2-mm voxels than with the 0.3-mm voxel size. When Wenzel et al (7) evaluated the diagnostic accuracy of CBCT and a photostimulable storage phosphor plate system (PSP) to detect VRFs, they used an i-cat scanner with mm and 0.25-mm voxel sizes. Higher accuracy was reported with the smaller voxel sizes; mm voxels showed a specificity of 98 and sensitivity of 87. A 0.25-mm voxel resolution was more accurate than the periapical PSP system, and high-resolution scanning was recommended in cases of suspected VRF that could not be visualized in periapical images. Also, in an in vitro study to detect VRF by using CBCT, Kamburoğlu et al (10) compared 2 different CBCT units with varying resolutions of 0.19-, 0.1-, and 0.3-mm voxel sizes. The 0.19-mm and 0.1-mm voxel resolutions achieved better results than the 0.3-mm resolution, but with higher reconstruction times, meaning more radiation exposure. Liedke et al (25) also used i-cat in assessing diagnostic ability in the detection of simulated external root resorption with different voxel resolutions (0.2, 0.3, 0.4 mm); the best results were achieved with a voxel size of 0.2 or 0.3 mm. According to the ALARA principle, the 0.3-mm voxel size with a shorter scanning time was recommended to reduce the patient s x-ray exposure. Tanimoto et al (24) reported that choosing a small voxel size without changing the radiation dose increases the resolution. In the present study, no significant differences were found among observers or voxel sizes, and the sensitivity/specificity results were high for all groups (sensitivity/specificity for mm, 98/96; 0.2 mm, 97/ 96; 0.3 mm, 93/93; 0.4 mm, 91/93). The values revealed that all groups were similar in terms of sensitivity and specificity. The positive likelihood ratio showed that the mm and 0.2-mm voxel resolutions were better than the 0.3-mm and 0.4-mm voxel resolutions (24.5, 24.25, 13.2, and 13, respectively) (Fig. 1). Accuracy results also showed that the mm and 0.2-mm resolutions were more successful than the 0.3-mm and 0.4-mm resolutions in detecting VRFs, although the latter 2 voxel resolutions are also acceptable under clinical conditions (Fig. 1). However, there were cases identified as False-Negative in readings that failed to identify a fracture in a fractured root in all voxel sizes. Those might have resulted from low signal-to-noise ratio as a result of decreasing acquisition time to reduce the radiation dose. The 4 voxel resolutions showed similar results in detecting simulated VRFs. However, accuracy was higher and the decision was easier with mm and 0.2-mm voxels. The best voxel selection would be 0.2 mm, with a shorter scanning time and reduced radiation exposure to the patient. Conclusions CBCT is a reliable imaging system to detect VRFs, and considering the ALARA principle, 0.2-mm resolution is the best choice for diagnostic use. References 1. Andreasen FM, Andreasen JO. Root fractures. In: Andreasen JO, Andreasen FM, eds. Textbook and color atlas of traumatic injuries to the teeth. 3rd ed. Copenhagen, Denmark: Munsksgaard; 1994: Kositbowornchai S, Nuansakul R, Sikram S, Sinahawattana S, Saengmontri S. Root fracture detection: a comparison of direct digital radiography with conventional radiography. Dentomaxillofac Radiol 2001;30: Mora MA, Mol A, Tyndall DA, Rivera EM, Hill C. In vitro assessment of local computed tomography for the detection of longitudinal tooth fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;103: Bernardes RA, de Moraes IG, Duarte MAH, Azevedo BC, de Azevedo JR, Bramante CM. Use of cone-beam volumetric tomography in the diagnosis of root fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009;108: Hassan B, Metska ME, Ozok AR, van der Stelt P, Wesselink R. Detection of vertical root fractures in endodontically treated teeth by a cone beam computed tomography scan. J Endod 2009;35: Hassan B, Metska ME, Ozok AR, van der Stelt P, Wesselink PR. Comparison of five cone beam computed tomography systems for the detection of vertical root fractures. J Endod 2010;36: Wenzel A, Neto FH, Frydenberg M, Kirkevang LL. Variable-resolution cone-beam computerized tomography with enhancement filtration compared with intraoral photostimulable phosphor radiography in detection of transverse root fractures in an in vitro model. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009; 108: Hannig C, Dullin C, Hulsmann M, Heidrich G. Three-dimensional, non-destructive visualization of vertical root fractures using flat panel volume detector computer tomography: an ex vivo in vitro case report. Int Endod J 2005;38: Nair MK, Nair UDP, Gröndahl HG, Webber RL, Wallace JA. Detection of artificially induced vertical radicular fractures using tuned aperture computed tomography. Eur J Oral Sci 2001;109: Kamburoğlu K, Murat S, Yüksel SP, Cebeci AR, Horasan S. Detection of vertical root fracture using cone-beam computerized tomography: an in vitro assessment. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2010;109:e Scarfe WC, Farman AG, Sukovic P. Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc 2006;72: Farman A, Scarfe WC. The basics of maxillofacial cone beam computed tomography. Semin Orthod 2009;15: Hatcher DC, Aboudara CL. Diagnosis goes digital. Am J Orthod Dentofacial Orthop 2004;125: Sohaib SA, Peppercorn PD, Horrocks JA, Keene MH, Kenyon GS, Reznek RH. The effect of decreasing mas on image quality and patient dose in sinus CT. Br J Radiol 2001;74: Cohnen M, Fischer H, Hamacher J, Lins E, Kotter R, Modder U. CT of the head by use of reduced current and kilovoltage: relationship between image quality and dose reduction. AJNR Am J Neuroradiol 2000;21: Gündogdu S, Mahmutyazicioglu K, Ozdemir H, Savranlar A, Asil K. Assessment of image quality of a standard and three dose-reducing protocols in adult cranial CT. Eur Radiol 2005;15: Rustemeyer P, Streubühr U, Suttmoeller J. Low-dose dental computed tomography: significant dose reduction without loss of image quality. Acta Radiol 2004;45: Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shinoda K. Development of a compact computed tomographic apparatus for dental use. Dent Maxillofac Radiol 1999; 28: Cotton TP, Geisler TM, Holden DT, Schwartz SA, Schindler WG. Endodontic applications of cone beam volumetric tomography. J Endod 2007;33: Nair MK, Nair UP. Digital and advanced imaging in endodontics: a review. J Endod 2007;33: Trope M, Pettigrew J, Petras J, Barnett F, Tronstad L. Differentiation of radicular cyst and granulomas using computerized tomography. Endod Dent Traumatol 1989;5: Estrela C, Bueno MR, Leles CR, Azevedo BC, Azevedo JR. Accuracy of cone beam computed tomography, panoramic and periapical radiographic for the detection of apical periodontitis. J Endod 2008;34: Estrela C, Bueno MR, Azevedo B, Azevedo JR, Pecora JD. A new periapical index based on cone beam computed tomography. J Endod 2008;34: Tanimoto H, Arai Y. The effect of voxel size on image reconstruction in cone-beam computed tomography. Oral Radiol 2009;25: Liedke GS, da Silveira HED, da Silveira HLD, Dutra V, de Figueiredo JAP. Influence of voxel size in the diagnostic ability of cone beam tomography to evaluate simulated external root resorption. J Endod 2009;35: Farman AG. Alara still applies. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2005;4: Yiğit Özer JOE Volume 37, Number 1, January 2011
5 27. Ngan DC, Kharbanda OP, Geenty JP, Darendeliler MA. Comparison of radiation levels from computed tomography and conventional dental radiographs. Aust Orthod J 2003;19: Berman LH, Hartwell G. Diagnosis. In: Cohen S, Hargreaves KM, eds. Pathways of the pulp. 9th ed. St Louis, MO: Mosby; 2006: Patel S, Dawood A, Ford TP, et al. The potential applications of cone beam computed tomography in the management of endodontic problems. Int Endod J 2007;40: Nesari R, Rossman LE, Kratchman SI. Cone-beam computed tomography in endodontics: are we there yet? Compend Contin Educ Dent 2009;30: Fuss Z, Lustig J, Katz A, Tamse A. An evaluation of endodontically treated vertical root fractured teeth: impact of operative procedures. J Endod 2001; 27:46 8. JOE Volume 37, Number 1, January 2011 Detection of Vertical Root Fractures with CBCT with Variable Voxel Sizes 79
Peking University School and Hospital of Stomatology. c Professor, Department of Cariology, Endodontology, and Operative
The ability of cone-beam computerized tomography to detect vertical root fractures in endodontically treated and nonendodontically treated teeth: a report of 3 cases Xiaoying Zou, DDS, MD, a Denggao Liu,
More informationComparison of conventional radiography with cone beam computed tomography for detection of vertical root fractures: an in vitro study
593 Journal of Oral Science, Vol. 52, No. 4, 593-597, 2010 Original Comparison of conventional radiography with cone beam computed tomography for detection of vertical root fractures: an in vitro study
More informationJournal of Dental School 2014; 32(1): 23-29
Journal of Dental School 2014; 32(1): 23-29 Original Article Comparison of Charge-Coupled Devices and Photostimulable Phosphor Storage Plates for Detection of Vertical Root Fractures in Endodontically
More informationDiagnostic accuracy of conventional, digital and Cone Beam CT in vertical root fracture detection
ORIGINAL ARTICLE Solmaz Valizadeh1 DDS, MS, Maryam Khosravi2 DDS, MS, Zeynab Azizi3* DDS, MS Diagnostic accuracy of conventional, digital and Cone Beam CT in vertical root fracture detection 1.Assistant
More informationDiagnostic value of two modes of cone-beam computed tomography in evaluation of simulated external root resorption: an in vitro study
Imaging Science in Dentistry 2012; 42 : 19-24 http://dx.doi.org/10.5624/isd.2012.42.1.19 Diagnostic value of two modes of cone-beam computed tomography in evaluation of simulated external root resorption:
More informationPeriapical radiography and cone beam computed tomography for assessment of the periapical bone defect 1 week and 12 months after root-end resection
(2009) 38, 531 536 2009 The British Institute of Radiology http://dmfr.birjournals.org RESEARCH Periapical radiography and cone beam computed tomography for assessment of the periapical bone defect 1 week
More informationValue of cone beam CT in detection of dental root fractures
(2012) 41, 3 10 2012 The British Institute of Radiology http://dmfr.birjournals.org RESEARCH Value of cone beam CT in detection of dental root fractures Z Dalili Kajan* and M Taromsari Department of Maxillofacial
More informationA comparative study of cone-beam computed tomography and digital periapical radiography in detecting mandibular molars root perforations
Imaging Science in Dentistry 2014; 44: 115-9 http://dx.doi.org/10.5624/isd.2014.44.2.115 A comparative study of cone-beam computed tomography and digital periapical radiography in detecting mandibular
More informationCone-Beam Computed Tomography as a Tool in Endodontic Detection and Diagnosis
REVIEW ARTICLE Cone-Beam Computed Tomography as a Tool in Endodontic Detection and Diagnosis Krishnan Hari, G Anuradha ABSTRACT Radiographic imaging is essential in diagnosis, treatment planning and follow-up
More informationRoot resorption (RR) is either a physiologic or pathologic condition associated with
Method to Evaluate Inflammatory Root Resorption by Using Cone Beam Computed Tomography Carlos Estrela, DDS, MSc, PhD,* Mike Reis Bueno, DDS, MSc, Ana Helena Gonçalves De Alencar, DDS, MSc, PhD,* Rinaldo
More informationPeriapical bone defects of root filled teeth with persistent lesions evaluated by cone-beam computed tomography
doi:10.1111/j.1365-2591.2010.01814.x Periapical bone defects of root filled teeth with persistent lesions evaluated by cone-beam computed tomography T. Yoshioka, I. Kikuchi, C. G. Adorno & H. Suda Pulp
More informationAccuracy of chemically created periapical lesion measurements using limited cone beam computed tomography
(010) 39, 95 99 010 The British Institute of Radiology http://dmfr.birjournals.org RESEARCH Accuracy of chemically created periapical lesion measurements using limited cone beam computed tomography K Kamburoğlu*,1,,
More informationThe effect of metal artifacts on the identification of vertical root fractures using different fields of view in cone-beam computed tomography
Imaging Science in Dentistry 215; 45: 147-51 http://dx.doi.org/1.5624/isd.215.45.3.147 The effect of metal artifacts on the identification of vertical root fractures using different fields of view in cone-beam
More informationMehran Taramsari 1, Zahra Dalili Kajan 2, *, Parinaz Bashirzadeh 3, Fatemeh Salamat 4. Introduction ABSTRACT
Imaging Science in Dentistry 2013; 43: 171-7 http://dx.doi.org/10.5624/isd.2013.43.3.171 Comparison of high-resolution and standard zoom imaging modes in cone beam computed tomography for detection of
More informationAccuracy of digital periapical radiography and cone-beam computed tomography in detecting external root resorption
Imaging Science in Dentistry 2015; 45: 153-8 http://dx.doi.org/10.5624/isd.2015.45.3.153 Accuracy of digital periapical radiography and cone-beam computed tomography in detecting external root resorption
More informationDetection of horizontal root fracture using four different protocols of cone-beam computed tomography
Original Research Stomatology Detection of horizontal root fracture using four different protocols of cone-beam computed tomography Fernanda Cristina Sales SALINEIRO Lucas Rodrigues PINHEIRO Oséas dos
More informationAwareness and Attitude among Dental Professional towards CBCT
IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861. Volume 10, Issue 5 (Sep.- Oct. 2013), PP 55-59 Awareness and Attitude among Dental Professional towards CBCT
More informationAccuracy of Conventional and Digital Radiography in Detecting External Root Resorption
Accuracy of Conventional and Digital Radiography in Detecting External Root Resorption Abbas Mesgarani a, Sina Haghanifar b, Maryam Ehsani c*, Samereh Dokhte Yaghub d, Ali Bijani e a Department of Endodontics,
More informationThe detection and management of root resorption lesions using intraoral radiography and cone beam computed tomography an in vivo investigation
doi:10.1111/j.1365-2591.2009.01592.x The detection and management of root resorption lesions using intraoral radiography and cone beam computed tomography an in vivo investigation S. Patel 1,2, A. Dawood
More informationENDODONTICS: Colleagues for Excellence. aae.org/colleagues. Spring Published for the Dental Professional Community by the
The Impact of Cone Beam Computed Tomography in Endodontics: A New Era in Diagnosis and Treatment Planning Spring 2018 ENDODONTICS: Colleagues for Excellence Published for the Dental Professional Community
More informationComparison of intraoral radiography and limited cone beam computed tomography for the assessment of root-fractured permanent teeth
Dental Traumatology 2009; 25: 571 577; doi: 10.1111/j.1600-9657.2009.00833.x Comparison of intraoral radiography and limited cone beam computed tomography for the assessment of root-fractured permanent
More informationDiagnostic Accuracy of Cone-Beam Computed Tomography and Periapical Radiography in Internal Root Resorption
Diagnostic Accuracy of Cone-Beam Computed Tomography and Periapical Radiography in Internal Root Resorption Zahrasadat Madani a, Ehsan Moudi b, Ali Bijani c, Elham Mahmoudi a* a Department of Endodontics,
More informationEffect of Metal Artifacts on Detection of Vertical Root Fractures Using Two Cone-Beam Computed Tomography Systems
Effect of Metal Artifacts on Detection of Vertical Root Fractures Using Two Cone-Beam Computed Tomography Systems Yaser Safi a, Mohammad Mehdi Aghdasi a, Fatemeh Ezoddini-Ardakani b, Samira Beiraghi c,
More informationAccuracy of Cone Beam Computed Tomography and Panoramic and Periapical Radiography for Detection of Apical Periodontitis
Accuracy of Cone Beam Computed Tomography and Panoramic and Periapical Radiography for Detection of Apical Periodontitis Carlos Estrela, DDS, MSc, PhD,* Mike Reis Bueno, DDS, MSc, Cláudio Rodrigues Leles,
More informationApplication of Cone Beam Computed Tomography in Endodontics: A Review Article
Review Article Application of Cone Beam Computed Tomography in Endodontics: A Review Article N. Hoshyari 1, K. Nazarimoghaddam 2. 1 Assistant Professor, Department of Endodontics, School of Dentistry,
More informationHorizontal root fracture (HRF) implies a severe injury to cementum, dentin, and pulp
Contemporary Management of Horizontal Root Fractures to the Permanent Dentition: Diagnosis Radiologic Assessment to Include Cone-Beam Computed Tomography Jeremy J. May, BA,* Nestor Cohenca, DDS, and Ove
More informationEffect of different exposure settings on the diagnosis of vertical root fractures on cone-beam computed tomography images
Medical Communication Biosci. Biotech. Res. Comm. 11(1): 117-121 (2018) Effect of different exposure settings on the diagnosis of vertical root fractures on cone-beam computed tomography images Sandra
More informationInternational Journal of Basic and Clinical Studies (IJBCS) 2014;3(1): Uysal I and Oztekin F
Endodontic Treatment of Mandibular Incisors with Periapical Lesion by Using Single Instrument: Cone Beam Computed Tomography Imaging Ibrahim Uysal 1 Faruk Oztekin 2 1 PhD, Dicle University, Faculty of
More informationThe Evaluation of Root Fracture with Cone Beam Computed Tomography (CBCT): An Epidemiological Study
Journal section: Community and Preventive Dentistry Publication Types: Research doi:10.4317/jced.54009 http://dx.doi.org/10.4317/jced.54009 The Evaluation of Root Fracture with Cone Beam Computed Tomography
More informationEvaluation of Cone Beam Computed Tomography in Diagnosis and Treatment Plan of Impacted Maxillary Canines
Original Research Evaluation of Cone Beam Computed Tomography in Diagnosis and Treatment Plan of Impacted Maxillary Canines Seyed Hossein Hoseini Zarch 1, Farzin Heravi 2, Adineh Javadian Langaroodi 3,
More informationOriginal Research Article
DOI: 10.21276/sjds.2016.3.6.5 Scholars Journal of Dental Sciences (SJDS) Sch. J. Dent. Sci., 2016; 3(6):175-179 Scholars Academic and Scientific Publisher (An International Publisher for Academic and Scientific
More informationCone Beam Computed Tomography in Endodontics: An Overview.
e-issn:2320-7949 RESEARCH AND REVIEWS: JOURNAL OF DENTAL SCIENCES Cone Beam Computed Tomography in Endodontics: An Overview. Mithra N Hegde, Priyadarshini Hegde, Aditya Shetty, Lakshmi Nidhi Rao, and Litty
More informationTwo- and Three-dimensional Orthodontic Imaging Using Limited Cone Beam Computed Tomography
Original Article Two- and Three-dimensional Orthodontic Imaging Using Limited Cone Beam Computed Tomography Akira Nakajima a ; Glenn T. Sameshima b ; Yoshinori Arai c ; Yoshito Homme d ; Noriyoshi Shimizu
More informationEvidence-based decision-making in endodontics
Clin Dent Rev (2017) 1:6 https://doi.org/10.1007/s41894-017-0006-0 TREATMENT Evidence-based decision-making in endodontics Eyal Rosen 1 Igor Tsesis 1 Received: 15 June 2017 / Accepted: 9 July 2017 / Published
More informationDiagnostic accuracy of conventional radiography and digital radiography in detecting of root fractures (invitro study)
Diagnostic accuracy of conventional and digital in detecting of root fractures (invitro study) Amin-Tavakoli M. DDS, MS 1, Esmaili F. DDS, MS 2 1 Associate Prof., Dept. of Oral and Maxillofacial Radiology,
More informationThe detection of periapical pathosis using periapical radiography and cone beam computed tomography Part 1: pre-operative status
doi:10.1111/j.1365-2591.2011.01989.x The detection of periapical pathosis using periapical radiography and cone beam computed tomography Part 1: pre-operative status S. Patel 1,2, R. Wilson 3, A. Dawood
More informationComparative evaluation of shaping ability of two nickel-titanium rotary systems using cone beam computed tomography
Celikten et al. BMC Oral Health (2015) 15:32 DOI 10.1186/s12903-015-0019-5 RESEARCH ARTICLE Open Access Comparative evaluation of shaping ability of two nickel-titanium rotary systems using cone beam computed
More informationONLINE ONLY. e129. Oral Pathology and Maxillo-Facial Surgery. b Professor and head of the Oral Imaging Center, Department of Periodontology,
ONLINE ONLY Comparison of 6 cone-beam computed tomography systems for image quality and detection of simulated canine impaction-induced external root resorption in maxillary lateral incisors Ali Alqerban,
More informationDetection accuracy of root fractures in cone-beam computed tomography images: a systematic review and meta-analysis
doi:10.1111/iej.12490 REVIEW Detection accuracy of root fractures in cone-beam computed tomography images: a systematic review and meta-analysis R. H. Ma, Z. P. Ge & G. Li Department of Oral and Maxillofacial
More informationMonday Morning Pearls of Practice by Bobby Baig
Dec 19, 2016 Monday Morning Pearls of Practice by Bobby Baig baig@buildyoursmile.com Prosthodontic Associates 2300 Yonge St, suite 905 Toronto, M4P1E4 www.buildyoursmile.com CBCT and Implant Dentistry:
More informationCone Beam in the Modern Dental Office Better Imaging, Better Outcomes Incorporation of CBCT Data into Treatment Planning
1 Cone Beam in the Modern Dental Office Better Imaging, Better Outcomes Incorporation of CBCT Data into Treatment Planning Rick Ferguson DMD Clinical Assistant Professor University of Florida Diplomate
More informationCBCT Specific Guidelines for South African Practice as Indicated by Current Literature:
CBCT Specific Guidelines for South African Practice as Indicated by Current Literature: CF Hoogendijk Maxillo- facial and Oral surgery: Trauma: 1. Facial trauma for the confirmation or exclusion of fractures
More informationEvaluation of technical quality and periapical
www.scielo.br/jaos http://dx.doi.org/10.1590/1678-775720140110 Evaluation of technical quality and periapical Bilge Gülsüm NUR 1, Evren OK 2, Mustafa ALTUNSOY 1 3, Mehmet ÇOLAK 4, Enes GÜNGÖR 4 4- Department
More informationEvaluation of a Digital Measurement Tool to Estimate Working Length in Endodontics
Volume 2 Number 1 Winter Issue, 2001 Evaluation of a Digital Measurement Tool to Estimate Working Length in Endodontics Abstract The purpose of the study was to compare the segmental measurement tool from
More informationThe agony and ecstasy of buying cone beam technology Part 1: The Ecstasy
Clinical The agony and ecstasy of buying cone beam technology Part 1: The Ecstasy Dale A. Miles 1 Abstract Background: Since arriving in North America in 2001, cone beam computed tomography (CBCT) has
More informationDimensional stability in composite cone beam computed tomography
(2010) 39, 512 516 2010 The British Institute of Radiology http://dmfr.birjournals.org TECHNICAL REPORT Dimensional stability in composite cone beam computed tomography S Kopp* and P Ottl Department of
More informationEffect of Amperage and Field of View on Detection of Vertical Root Fracture in Teeth with Intracanal Posts
Effect of Amperage and Field of View on Detection of Vertical Root Fracture in Teeth with Intracanal Posts Yaser Safi a, Sepanta Hosseinpour b*, Alireza Aziz c, Masoud Bamedi d, Mahsa Malekashtari e, Zahra
More informationCone Beam Computed Tomography: A New Boon and a Ray of Hope to the Endodontist A Series of Cases
IJopRD IJOPRD Cone Beam Computed Tomography: A New Boon and a Ray of Hope 10.5005/jp-journals-10019-1139 to the Endodontist A Series of Cases Review Article Cone Beam Computed Tomography: A New Boon and
More informationUsing cone beam technology in orthodontics
PRACTICE US Using cone beam technology in orthodontics Continuing education earn 2 ce credits! Edward Lin explains the benefits of adding cone beam technology to the orthodontic armamentaria Since cone
More informationAssessment of the relationship between the maxillary molars and adjacent structures using cone beam computed tomography
Imaging Science in Dentistry 2012; 42 : 219-24 http://dx.doi.org/10.5624/isd.2012.42.4.219 Assessment of the relationship between the maxillary molars and adjacent structures using cone beam computed tomography
More information3-Dimensional Cone-Beam Computerized Tomographic Cephalometric Database on Jaw Dimensions in Chinese
The Open Anthropology Journal, 2010, 3, 2-7 2 Open Access 3-Dimensional Cone-Beam Computerized Tomographic Cephalometric Database on Jaw Dimensions in Chinese Anson C.M. Chau a, Ricky W.K. Wong*,b and
More informationInterpretation Basics of Cone Beam Computed Tomography COPYRIGHTED MATERIAL
Interpretation Basics of Cone Beam Computed Tomography COPYIGHTED MATEIA 1 Introduction to Cone Beam Computed Tomography Shawneen M. Gonzalez Introduction This chapter will cover basics of cone beam computed
More informationKarl Maloney, DDS,* Jairo Bastidas, DMD, Katherine Freeman, DrPH, Todd R. Olson, PhD, and Richard A. Kraut, DDS. Materials and Methods
J Oral Maxillofac Surg 69:1923-1929, 2011 Cone Beam Computed Tomography and SimPlant Materialize Dental Software Versus Direct of the Width and Height of the Posterior Mandible: An Anatomic Study Karl
More informationAccuracy of CBCT as modality to identify the presence of secondary mesiobuccal root canal in the maxillary first molar
Padjadjaran Journal of Dentistry 2017;29(3):143-148. Accuracy of CBCT as modality to identify the presence of secondary mesiobuccal root canal in the maxillary first molar Haikal Halil*, Mazlifa Mahidin*,
More informationTwo-dimensional (2D) images have provided. Cone Beam Computed Tomography in Dentistry: What Dental Educators and Learners Should Know
Cone Beam Computed Tomography in Dentistry: What Dental Educators and Learners Should Know Shawn Adibi, D.D.S., F.D.O.C.S., M.Ed.; Wenjian Zhang, D.D.S., M.S., Ph.D.; Tom Servos, D.D.S.; Paula N. O Neill,
More informationANALYSIS OF FRACTURED TEETH UTILIZING DIGITAL MICROSCOPY: A PILOT STUDY
ANALYSIS OF FRACTURED TEETH UTILIZING DIGITAL MICROSCOPY: A PILOT STUDY by Thomas Gene Cooper, D.M.D., M.P.H. Lieutenant Commander, Dental Corps United States Navy A thesis submitted to the Faculty of
More information3D Accuitomo XYZ Slice View Tomograph. Super High-Resolution Images of Region of Interest
3D Accuitomo XYZ Slice View Tomograph. Super High-Resolution Images of Region of Interest Thinking ahead. Focused on life. 2 Cone Beam X-Ray CT Imaging X-Ray Tube Imaging Intensifier Imaging Volume Voxel
More informationThe influence of sensor size and orientation on image quality in intra-oral periapical radiography
Clinical The influence of sensor size and orientation on image quality in intra-oral periapical radiography Tony Druttman 1 The periapical view is one of the standard intra-oral radiographs by which diagnostic
More informationLimited cone-beam CT and intraoral radiography for the diagnosis of periapical pathology
Limited cone-beam CT and intraoral radiography for the diagnosis of periapical pathology Sara Lofthag-Hansen, DDS, a Sisko Huumonen, DDS, PhD, b Kerstin Gröndahl, DDS, PhD, c and Hans-Göran Gröndahl, DDS,
More informationDiagnostic performance of cone-beam computed tomography on detection of mechanically-created artificial secondary caries
Imaging Science in Dentistry 2011; 41 : 143-50 http://dx.doi.org/10.5624/isd.2011.41.4.143 Diagnostic performance of cone-beam computed tomography on detection of mechanically-created artificial secondary
More informationReliability of Orthopantomography and Cone-beam Computed Tomography in Presurgical Implant Planning: A Clinical Study
JCDP Original Research Reliability of OPG and CBCT 10.5005/jp-journals-00000-0000 in Presurgical Implant Planning Reliability of Orthopantomography and Cone-beam Computed Tomography in Presurgical Implant
More informationClinical indications for digital imaging in dento-alveolar trauma. Part 2: root resorption
Dental Traumatology 2006; doi: 10.1111/j.1600-9657.2006.00546.x DENTAL TRAUMATOLOGY Clinical indications for digital imaging in dento-alveolar trauma. Part 2: root resorption Cohenca N, Simon JH, Mathur
More informationCourse File 243 DDS Physics of Diagnostic Radiology and Oral and Maxillofacial Radiology
King Saud University College of Dentistry Dept. of Oral Medicine & Diagnostic Sciences Division of Oral & Maxillofacial Radiology Course File 243 DDS Physics of Diagnostic Radiology and Oral and Maxillofacial
More informationSEDENTEXCT: Guidelines and evidence-based use of CBCT
: Guidelines and evidence-based use of CBCT Keith Horner John Dalton Rutherford Bragg UK largest single-site University 34,000 students University Dental Hospital of Manchester 90,000 patients attend
More informationInternational Journal of Current Medical and Pharmaceutical Research
ISSN: 2395-6429 International Journal of Current Medical and Pharmaceutical Research Available Online at http://www.journalcmpr.com DOI: http://dx.doi.org/10.24327/23956429.ijcmpr20170169 RESEARCH ARTICLE
More informationIn vitro cone beam computed tomography imaging of periodontal bone
(2008) 37, 319 324 2008 The British Institute of Radiology http://dmfr.birjournals.org RESEARCH In vitro cone beam computed tomography imaging of periodontal bone A Mol*,1 and A Balasundaram 2 1 Department
More informationNew dimensions in endodontic imaging: Part 2. Cone beam computed tomography
doi:10.1111/j.1365-2591.2008.01531.x REVIEW New dimensions in endodontic imaging: Part 2. Cone beam computed tomography S. Patel Endodontic Postgraduate Unit, King s College London Dental Institute, London,
More informationA Cone Beam Computed Tomography Study Among Dental Residents Based on a Questionnaire
A Cone Beam Computed Tomography Study Among Dental Residents Based on a Questionnaire Alexandra Mihaela Stoica, Assistant Lecturer, DMD Monica Monea, Associate Prof., DMD, PhD Dragos Dan Sita, Assistant
More informationScientific Affairs. Scientific Affairs
The use of cone-beam computed tomography in dentistry : An advisory statement from the American Dental Association Council on Scientific Affairs The American Dental Association Council on Scientific Affairs
More informationCBCT: A dentist perspective
SEGi Review ISSN: 1985.5672 Vol.9, December 2015 CBCT: A dentist perspective Dr. Ranjana Garg, Faculty of Dentistry SEGi University. Dr. Vivek Vijay Gupta Faculty of Dentistry SEGi University Abstract
More informationComparative assessment of subjective image quality of cross-sectional cone-beam computed tomography scans
501 Journal of Oral Science, Vol. 53, No. 4, 501-508, 2011 Original Comparative assessment of subjective image quality of cross-sectional cone-beam computed tomography scans Kıvanç Kamburoğlu 1), Sema
More informationA Comparison of 2D versus 3D Radiography in the Treatment Planning of Root Canal Treated Teeth with Periapical Lesions
A Comparison of 2D versus 3D Radiography in the Treatment Planning of Root Canal Treated Teeth with Periapical Lesions Heidi Kohltfarber, DDS A thesis submitted to the faculty of the University of North
More informationAndrea Fagundes Campello 1, Lucio Souza Gonçalves 1, Fábio Ribeiro Guedes 2, Fábio Vidal Marques 1, * Introduction ABSTRACT
Imaging Science in Dentistry 2017; 47: 25-31 https://doi.org/10.5624/isd.2017.47.1.25 Cone-beam computed tomography versus digital periapical radiography in the detection of artificially created periapical
More informationDental Implant Planning using Cone Beam CT imaging: a pictorial guide.
Dental Implant Planning using Cone Beam CT imaging: a pictorial guide. Poster No.: C-1970 Congress: ECR 2015 Type: Educational Exhibit Authors: S. R. Rice, G. Price, S. Morley, T. Beale; London/UK Keywords:
More informationOriginal Research THE USE OF REFORMATTED CONE BEAM CT IMAGES IN ASSESSING MID-FACE TRAUMA, WITH A FOCUS ON THE ORBITAL FLOOR FRACTURES
DOI: 10.15386/cjmed-601 Original Research THE USE OF REFORMATTED CONE BEAM CT IMAGES IN ASSESSING MID-FACE TRAUMA, WITH A FOCUS ON THE ORBITAL FLOOR FRACTURES RALUCA ROMAN 1, MIHAELA HEDEȘIU 1, FLOAREA
More informationIn-Office Cone Beam Computerized Tomography: Technology Review and Clinical Examples Michael Tischler, DDS
Page 1 of 8 Issue Date: June 2008, Posted On: 6/26/2008 In-Office Cone Beam Computerized Tomography: Technology Review and Clinical Examples Michael Tischler, DDS Electronic Medical Record Research EMR
More informationHead to new heights with your imaging SCANORA 3D
SCANORA 3D Head to new heights with your imaging Benefits at a glance The solution for dentomaxillofacial and ENT imaging Easy Patient seated for added stability during exposure. Clear, self-explinatory
More informationPeriodontal Disease. Radiology of Periodontal Disease. Periodontal Disease. The Role of Radiology in Assessment of Periodontal Disease
Radiology of Periodontal Disease Steven R. Singer, DDS srs2@columbia.edu 212.305.5674 Periodontal Disease! Includes several disorders of the periodontium! Gingivitis! Marginal Periodontitis! Localized
More informationDuring orthodontic treatment, the alignment of
TECHNO BYTES Comparison of mesiodistal root angulation with posttreatment panoramic radiographs and cone-beam computed tomography Daniel G. Bouwens, a Lucia Cevidanes, b John B. Ludlow, c and Ceib Phillips
More informationAssessment of the relationship between maxillary sinus floor and maxillary posterior teeth root apices using spiral CT scan
Assessment of the relationship between maxillary sinus floor and maxillary posterior teeth root apices using spiral CT scan Zainab Abdulhasan Hussein, B.D.S. (1) (2) Lamia H. Al-Nakib, B.D.S., M.Sc. ABSTRACT
More informationRumpa Ganguly 1, *, Aruna Ramesh 2, Sarah Pagni 3. Introduction
Imaging Science in Dentistry 2016; 46: 93-101 http://dx.doi.org/10.5624/isd.2016.46.2.93 The accuracy of linear measurements of maxillary and mandibular edentulous sites in conebeam computed tomography
More informationDigital radiography has
Better imaging The advantages of digital radiography Paul F. van der Stelt, DDS, PhD Digital radiography has been available in dentistry for more than 25 years, but digital imaging has not replaced conventional
More informationIs the diagnosis of calcified laryngeal cartilages on panoramic radiographs possible?
Imaging Science in Dentistry 2018; 48: 121-5 https://doi.org/10.5624/isd.2018.48.2.121 Is the diagnosis of calcified laryngeal cartilages on panoramic radiographs possible? Leyla Berna Çağırankaya 1, *,
More informationMaxillary Central Incisor Incisive Canal Relationship: A Cone Beam Computed Tomography Study
Maxillary Central Incisor Incisive Canal Relationship: A Cone Beam Computed Tomography Study Joseph Y.K. Kan, DDS, MS Professor, Department of Restorative Dentistry, Loma Linda University School of Dentistry,
More informationUse of cone-beam volumetric tomography in the diagnosis of root fractures
Use of cone-beam volumetric tomography in the diagnosis of root fractures Ricardo Affonso Bernardes, DDS, MSc PhD, a Ivaldo Gomes de Moraes, DDS, MSc, PhD, b Marco Antonio Húngaro Duarte, DDS, MSc, PhD,
More information3 Dimensional Diagnosis Unravelling Prognosis of Multiple Impacted Teeth A Case Report
3 Dimensional Diagnosis Unravelling Prognosis of Multiple Impacted Teeth A Case Report Adusumilli Gopinath 1, Naveen Admala Reddy 2, Mayur G Rohra 3 1 Professor, Department of Orthodontics, AME S Dental
More informationVariation in Mesiodistal Root Angulation of Panoramic Images Generated from Cone-beam Computed Tomography
Variation in Mesiodistal Root Angulation of Panoramic Images Generated from Cone-beam Computed Tomography Si Qi LIU 1, Fu Jia WEN 1, Hua CHEN 2, Yi LIU 1 Objective: To compare the mesiodistal root angulation
More informationEffect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography
Original Article Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography Mehrdad Panjnoush 1, Yasaman Kheirandish 2, Negar Zeini 3 1 Associate Professor,
More informationAnalysis of Root Resorption in Dentistry using Collaborative Tools & Strategies
International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 2, Issue 12 (August 2012), PP. 37-42 Analysis of Root Resorption in Dentistry using
More informationIt is essential to understand which factors positively or negatively influence the outcome
Endodontic Outcome Predictors Identified with Periapical Radiographs and Cone-beam Computed Tomography Scans Yu-Hong Liang, DDS,* Gang Li, DDS, PhD, Paul R. Wesselink, DDS, PhD, and Min-Kai Wu, MSD, PhD
More informationComparison of Periapical Radiography and Limited Cone-Beam Tomography in Posterior Maxillary Teeth Referred for Apical Surgery
Comparison of Periapical Radiography and Limited Cone-Beam Tomography in Posterior Maxillary Teeth Referred for Apical Surgery Kenneth M.T. Low, BDS,*, Karl Dula, PD DMD,* Walter Bürgin, MS, BE,* and Thomas
More informationApplication of cone-beam computed tomography (CBCT) for diagnosis and management of mandibular condyle fractures
Original Article Acharya et.al. for diagnosis and management of mandibular condyle fractures P Acharya 1, G Ren 2, MR Jaisani 1, A Dongol 1, RP Yadav 1, AK Mahat 1, F Dong 1 Oral and Maxillofacial Surgery
More informationKeywords: Mandible, Maxilla, Teeth, Cone Beam CT, Implants, Tumor and tumor-like conditions
Dental Cone Beam CT: a primer for clinical radiologists e-poster: EP-1 Congress: ESHNR 2014 Type: Educational Poster Topic: ESHNR 2014 Authors: C. Vanhoenacker 1, F. Vanhoenacker 1, L. Vannitsen 1, A.
More informationClinical details: Details of scan: CONE BEAM CT REPORT: Name: H. B. Gender: Reason for referral: Referred by:
Name: H. B. Gender: Male DOB: 11/12/1950 Age: 64 Date taken: 16/11/2015 Date reported: 19/11/2015 Clinical details: Reason for referral: Referred by: Investigate symptoms related to left TMJ. Reconstructed
More informationSubmitted: Revised: Published:
ORIGINAL ARTICLE ASIAN JOURNAL OF MEDICAL SCIENCES Correlation between maxillary sinus floor topography and relative root position of posterior teeth using Orthopantomograph and Digital Volumetric Tomography
More informationEvaluation of enamel pearls by cone-beam computed tomography (CBCT)
Journal section: Oral Medicine and Pathology Publication Types: Research doi:10.4317/medoral.17475 http://dx.doi.org/doi:10.4317/medoral.17475 Evaluation of enamel pearls by cone-beam computed tomography
More informationMædica - a Journal of Clinical Medicine. Interobserver variability of the diagnosis of apical periodontitis on panoramic radiography assessment
Mædica - a Journal of Clinical Medicine ORIGIN RIGINAL PAPERS Interobserver variability of the diagnosis of apical periodontitis on panoramic radiography assessment Mihaela HEDESIU, MD a ; Andrea SERBANESCU,
More informationPanoramic Radiograph and cone beam computed tomography- evaluation of Mandibular impacted third molar and Mandibular canal
Panoramic Radiograph and cone beam computed tomography- evaluation of Mandibular impacted third molar and Mandibular canal Authors & Affiliation: Correspondence To: Dr. S Jayachandran, MDS, PhD, Professor
More informationThe Effect of X Ray Vertical Angulation on Radiographic Assessment of Alveolar Bone Loss
1 The Effect of X Ray Vertical Angulation on Radiographic Assessment of Alveolar Bone Loss ABSTRACT M. Mehdizadeh MD*, M. Amintavakoli MD**, M. Allahverdi MD*** Introduction: Radiographs provide unique
More informationHow a Computerized Tomography Examination Changed the Treatment Plans of 80 Children with Retained and Ectopically Positioned Maxillary Canines
Original Article How a Computerized Tomography Examination Changed the Treatment Plans of 80 Children with Retained and Ectopically Positioned Maxillary Canines Krister Bjerklin a ; Sune Ericson b Abstract:
More information