1 Magnetic Resonance in Medicine 50: (2003) Tracer Arrival Timing-Insensitive Technique for Estimating Flow in MR Perfusion-Weighted Imaging Using Singular Value Decomposition With a Block-Circulant Deconvolution Matrix Ona Wu, 1 * Leif Østergaard, 2 Robert M. Weisskoff, 3 Thomas Benner, 1 Bruce R. Rosen, 1 and A. Gregory Sorensen 1 Relative cerebral blood flow (CBF) and tissue mean transit time (MTT) estimates from bolus-tracking MR perfusionweighted imaging (PWI) have been shown to be sensitive to delay and dispersion when using singular value decomposition (SVD) with a single measured arterial input function. This study proposes a technique that is made time-shift insensitive by the use of a block-circulant matrix for deconvolution with (osvd) and without (csvd) minimization of oscillation of the derived residue function. The performances of these methods are compared with standard SVD (ssvd) in both numerical simulations and in clinically acquired data. An additional index of disturbed hemodynamics (odelay) is proposed that represents the tracer arrival time difference between the AIF and tissue signal. Results show that PWI estimates from ssvd are weighted by tracer arrival time differences, while those from osvd and csvd are not. osvd also provides estimates that are less sensitive to blood volume compared to csvd. Using PWI data that can be routinely collected clinically, osvd shows promise in providing tracer arrival timing-insensitive flow estimates and hence a more specific indicator of ischemic injury. Shift maps can continue to provide a sensitive reflection of disturbed hemodynamics. Magn Reson Med 50: , Wiley-Liss, Inc. Key words: perfusion-weighted imaging; humans; dynamic susceptibility contrast-enhanced magnetic resonance imaging; cerebral blood flow; ischemia 1 Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital/Massachusetts Institute of Technology/Harvard Medical School, Boston, Massachusetts. 2 Department of Neuroradiology, Center for Functionally Integrative Neuroscience, Århus University Hospital, Århus, Denmark. 3 EPIX Medical, Cambridge, Massachusetts. Grant sponsor: NIH; Grant numbers: 5 R01 NS38477; 5 R01 HL39810; 5P41 RR14075; Grant sponsors: Danish Medical Research Council; Danish National Research Foundation. *Correspondence to: Ona Wu, Mailcode CNY , MGH/MIT/HMS Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Boston, MA ona@nmr.mgh.harvard.edu Received 26 November 2002; revised 27 February 2003; accepted 8 March DOI /mrm Published online in Wiley InterScience ( Wiley-Liss, Inc. 164 MR perfusion-weighted imaging (PWI) in acute stroke patients has been shown to be a sensitive indicator of tissue at risk of infarction (1,2). A prevalent technique for calculating cerebral blood flow (CBF) and mean transit time (MTT) in PWI uses singular value decomposition (SVD) for deconvolving the tissue signal from the first pass of a bolus of contrast agent with an arterial input function (AIF). Although this method is robust to noise and is independent of underlying vascular structure and volume (3), it has been shown to be sensitive to both delay and dispersion, leading to grossly underestimated CBF in these situations (3 6). Furthermore, it has been demonstrated that in some cases of major vasculopathy wherein the tracer arrives earlier in some tissue than in the chosen AIF, CBF can be either over- or underestimated, depending on the extent of the tracer arrival difference and underlying hemodynamics of the measured tissue (7). Although Fourier techniques are insensitive to delay in tracer arrival, they have been shown to provide poor estimates of high flow rates in the presence of noise (3,8). Maximum likelihood expectation maximization techniques (9) or model-dependent approaches that fit for delay and dispersion (10) are also promising, because they are both less sensitive to noise than Fourier techniques and less sensitive to tracer arrival time differences. However, the performances of these techniques have yet to be evaluated in cases of ischemic disease or major vasculopathy, where the assumptions and models upon which they are based may no longer be valid. Region-of-interest (ROI)- based methods to compensate for delay by selecting one AIF and shifting it in time until it is synchronized with the tissue signal curve, using parametric curve-fitting (11) or numerically estimated bolus arrival times (12), may not translate well on a voxel-by-voxel basis due to a lower signal-to-noise ratio (SNR) (6,8). This work proposes to improve flow estimates in two ways: 1) by performing deconvolution with a blockcirculant matrix to reduce sensitivity to tracer arrival differences between the AIF and tissue signal, and 2) by using local, rather than global, regularization. The performances of this method and the existing deconvolution approach using SVD (3) are evaluated in terms of flow estimation errors as a function of tracer arrival differences using numerical simulations where CBF values are known. In addition, techniques for estimating tracer arrival differences with respect to the AIF are evaluated. The performances of the methods are also compared qualitatively on clinically acquired human PW images. In particular, we examine two conditions that have been previously demonstrated to be problematic for SVD deconvolution: when internal carotid artery (ICA) stenosis (13) exists, and when the AIF regionally lags the tissue signal (7). We also examine two acute stroke cases in which initial PWI lesion volumes calculated using standard techniques are greater than lesion volumes detected on diffusion-weighted imaging (DWI).

2 Technique for Estimating CBF in PWI 165 THEORY Background Assuming a linear relationship between concentration of a high magnetic susceptibility contrast agent and change in transverse relaxation ( R 2 ) in dynamic susceptibility contrast-weighted (DSC) MR images (14), the change in concentration over time during the passage of a bolus of contrast agent in a volume of tissue C(t) can be characterized as: C t R 2 t 1 S t ln [1] TE S 0 where S 0 is the baseline MR intensity, and S(t) is the signal intensity over time. By defining the residue function R(t) as the fraction of tracer remaining in the system at a given time t, the concentration of tracer within the voxel of tissue as a function of time can be modeled as (3): C t F t C a t R t F t 0 t C a R t d [2] where C a (t) is the AIF and F t is the tissue blood flow. The AIF is typically measured directly from the MR images, which was shown in previous studies to correlate well with an AIF measured directly from arterial blood samples (15). R(t) can then be estimated using deconvolution with the measured R 2 (t), which is often performed using SVD (3). In SVD, by expressing Eq. [2] in discretized format: j c t j t F t C a t i R t j t i. [3] i 0 where t is the sampling interval, and expanding Eq. [3] into matrix notation, the deconvolution problem can be formulated as an inverse matrix problem: c t0 c t 1 c t N 1 t Ca t0 0 0 C a t 1 C a t 0 0 C a t N 1 C a t N 2 C a t 0 R t0 R t 1 R t N 1 Ft. [4] Simplifying the above equation to c A b, one can solve for b, the elements of R(t) scaled by F t. The measured AIF, C a (t), is typically prefiltered to reduce noise contributions and to compensate for discretization errors (3), resulting in A with elements: a i,j t C a t i j 1 4C a t i j C a t i j 1 /6 0 j i 0 otherwise. By decomposing A U S V T, where U and V are orthogonal matrices and S is a nonnegative square diagonal matrix, the inverse can be expressed as A 1 V W U T where W 1/S along the diagonals, and zero elsewhere. Values of W corresponding to values where S is less than a preset tolerance threshold, P SVD (usually a percentage of the maximum value of S), are set to zero. The residue function scaled by F t, b, can then be estimated by b F t V W U T c, and rcbf is estimated as b s maximum value. Deconvolution Using a Block-Circulant Matrix One of the assumptions built into Eq. [4] is causality, i.e., the voxel signal cannot arrive before the AIF. However, the AIF can lag C(t) by a time delay t d in practice, since the measured AIF, C a (t), is not necessarily the true AIF for that voxel, C a (t), thus resulting in C a (t) C a (t t d.). For example, this lag can occur when the chosen AIF comes from a highly diseased vessel. Therefore, the calculated R (t) should be R(t t d ) for C(t); but if causality is assumed, R (t) cannot be properly estimated by inversion of Eq. [4]. However, by using circular deconvolution instead of linear deconvolution, R (t) can be represented with R(t) circularly time shifted by t d. Circular convolution has been shown to be equivalent to linear convolution with time aliasing (16), and is also mathematically equivalent to a standard inverse Fourier method. By zero-padding the N- point time series C a (t) and C(t) to length L, where L 2N, time aliasing can be avoided. Therefore, replacing matrix A with a block-circulant matrix, D, whose elements are d i,j a ij for j i, and d i,j a L i-j,0 otherwise, Eq. [4] can be reformulated as g D f, where g is the zero-padded c, and f is the residue function scaled by F t. The inverse of D can be decomposed to D 1 V c W c U T c. One can again make use of SVD techniques to solve for f by f F t V c W c U T c g. When using circular deconvolution, however, due to the discontinuities at t 0 and t L, leakage frequencies may be amplified, giving rise to spurious oscillations dominating the deconvolved signal (17). Increasing P SVD reduces the oscillations (18). Using a modified oscillation index from that described by Gobbel and Fike (19): [5] O 1 L 1 1 L f max f k 2f k 1 f k 2 [6] k 2 where f is the scaled estimated residue function, f max is the maximum amplitude of f, and L is the number of sample points, P SVD can be varied until the estimated residue function s oscillation index falls below a user-specified value (OI) and CBF set to f max at that instance. METHODS Monte Carlo Simulations An AIF was simulated using a gamma-variate function, which was shown by previous studies to correlate well

3 166 Wu et al. with the shape of measured AIFs (3,6,9). The analytical expression for the AIF, C a (t) was: C a t 0 t t 0 C 0 t t 0 a e t t0 /b t t 0 [7] with a 3.0 and b 1.5 s, representative of data from normal adult volunteers (3,6,9). Simulations used C 0 1, t 0 20 s over a time range of 200 s to avoid truncation for the longest MTT simulated (24 s). To evaluate the techniques sensitivity to different underlying R(t), similarly to methods used in Refs. 3 and 4, three different models were examined: exponential, box-shaped, and linear. In all models, the MTT was calculated as MTT CBV/CBF from the central volume theorem (20). CBV was either 4% or 2%. These values were also used by other studies as representative for normal gray matter or white matter, respectively (3,6). For CBV 4%, flow values were varied between ml/100 g/min in 10 ml/100 g/min increments. For CBV 2%, flows were evaluated from 5 to 35 ml/100 g/min in 5 ml/100 g/min increments in order to maintain the same range of MTT values as for CBV 4%. Analytical expressions for C(t) were derived by convolving C a (t) with each R(t) (6). Signal curves were generated as S t S 0 e kc t TE, with baseline MR image intensity S 0 100, and TE 65 ms. For all simulations, a proportionality factor k was selected that resulted in a 40% peak signal drop at a flow rate of 60 ml/100 g/min and CBV 4%, corresponding to values typically found in human gray matter (3,6). The signal enhancement curve for the AIF, S a (t), was similarly modeled as S(t), except that C a (t) was substituted for C(t). The proportionality constant, k, in this case, was selected to generate a peak signal drop of 60%, which is a typical measured signal reduction for selected AIFs in our clinical PWI. Using previously described techniques (21), noise was added to S(t) and S a (t) to create signals with baseline SNRs of 20 and 100, respectively. To evaluate the sensitivity of flow estimates to differences in tracer arrival times between the AIF and tissue signal, S(t) was shifted up to 5s with respect to S a (t) in increments of 1 s, resulting in a total of 11 shifts. To simulate shifts that are not multiples of the sampling period, signals were created with t 100 ms, shifted after noise was added, and then resampled to TR 1 or 1.5 s. A TR 1 s was used primarily in our analysis to avoid confounds due to shifts that are not multiples of TR, and to be consistent with previous studies (3). Simulations were repeated at TR 1.5 s to investigate the effects of sampling interval on the algorithms performance. Using approaches similar to those described in Refs. 3 and 6, image data sets were created, resulting in a total of 1024 data points for each TR, SNR, shift, and flow. For ssvd and csvd, P SVD was varied between 0 and 95%. For osvd, OI was varied between 0 and 0.5. To determine absolute flow values, the calculated CBF values were rescaled by the k-factors used above for S(t) and S a (t). The following steps were repeated for each SNR and TR. For each P SVD and OI, the error at each iteration t (E t ) was calculated as E t 1/N f F F, where F is the true flow value, F is the calculated flow value, and N f is the number of simulated flow values (N f 7). The optimal P svd for ssvd and csvd, and optimal OI for osvd were determined as the values that minimized simultaneously over all assumed residue functions R(t) and N t 1024 iterations the average E t assuming zero time delays. The optimal P SVD and OI thresholds found in this step were then used to assess the performance of the techniques in terms of its mean error E t (E (D) 1/N t E t ) and standard deviation ( E (D)) over all N t 1024 iterations, as a function of tracer arrival time differences (D). Tracer arrival timing differences between tissue and the AIF were estimated as the sample point, m, where the maximum R(t) occurs. For ssvd, the estimated shift D m TR. For osvd, D m TR for m L/2 and D (L m) TR for L/2 m L, where L is the total number of points. The error in estimating tracer arrival time differences for each iteration t was calculated as ED t 1/N D D D, where D is the true time difference, D is the estimated difference, and N D is the number of simulated applied shifts (N D 11). The P SVD and OI used for estimating flows at each SNR and TR were used to estimate timing shifts. The mean delay error ED t (ED(F) 1/N t ED t ) over all N t 1024 iterations, as well as the SD ( ED (F)), were calculated at each flow rate, F. Clinical MRI Acquisition DSC MRI consisted of spin-echo, echo-planar images obtained during the first pass of 0.2 mmol/kg of a gadolinium-based contrast agent injected 10 s after start of imaging, at a rate of 5 ml/s, using an MRI-compatible power injector (Medrad, Pittsburgh, PA). Imaging studies were performed on 1.5 T GE Signa LX systems (GE Medical Systems, Milwaukee, WI). The parameters included TR/TE 1500/65 ms, field of view (FOV) cm 2 or cm 2, and acquisition matrix All studies consisted of 11 slices with a thickness of 6 mm and gap of 1 mm collected over 46 time points. All data analysis was performed retrospectively, with approval from our institution s committee for human subject research. Four patients were retrospectively examined. Patient demographics are shown in Table 1. Based on the simulation results, our analysis was limited to ssvd and osvd. Relative CBF (scbf and ocbf) and Delay (sdelay and odelay) maps were calculated using the same techniques as for the simulations. An AIF was selected from the ipsilateral hemisphere and used for analysis for both ssvd and osvd. Selection of P SVD for ssvd and OI for osvd were based on the optimal values found in the simulation section for SNR 20 and TR 1.5 s, which are typical for clinically acquired PW images at this institution. Relative cerebral blood volume (CBV) was calculated by numerically integrating the R 2 (t) curves (14,22). MTT values were calculated as smtt CBV/sCBF and omtt CBV/oCBF. RESULTS Monte Carlo Simulations For TR 1 s and SNR 100 and 20, the optimal P SVD s for ssvd were 4% and 20%, respectively; for csvd, they were

4 Technique for Estimating CBF in PWI 167 Table 1 Patient Demographics, Diagnosis, and Imaging Times Patient Age/sex Diagnosis Initial PWI Follow-up 1 67/F Transient ischemic attack due to left ICA stenosis 2 weeks T2 same day as PWI 2 56/F Right MCA stroke 4 hrs 22-day FLAIR 3 62/M Left MCA stroke 7 hrs 4 month FLAIR 4 52/F Right MCA stroke and complete right ICA occlusion 11 hrs 6 day FLAIR 5% and 10%; and for osvd, the optimal OIs were.065 and.035. The performances of ssvd, osvd, and csvd in estimating flow assuming no phase shift between the AIF and C(t) were found to be comparable. As expected, greater underestimation of flow was found at low SNR than at high SNR. Figure 1 shows the results for TR 1 s, and CBV 4% and underlying monoexponential R(t)atSNR 100 (a) and SNR 20 (b). Similar results were found for different R(t) (graphs not shown), except in the case of a box-shaped R(t) at SNR 100. In this condition, both ssvd and osvd tended to overestimate high flows while csvd did not. In addition, for CBV 2% and SNR 20 (graphs not shown), csvd performed slightly worse at low flow rates with greater overestimated flow, e.g., at 5 ml/100 g/min, csvd estimated flow as , while ssvd overestimated flow as and osvd estimated flow as ml/100 g/min. For all simulation conditions, E (D) for ssvd varied with timing shifts (D), especially when the AIF lagged the tissue signal (D 0), as shown in Fig. 1c d. In contrast, both osvd and csvd were less sensitive to tracer arrival differences. At D 0, osvd was found to perform comparably or better than ssvd (Table 2). However, the minimum E (D) for ssvd (E min ) often did not occur at D 0, but depended on simulation conditions. For SNR 100, E min occurred at D of 1, 1, 4, and 1 for monoexponential, linear, box-car R(t), and CBV 2%, respectively. For SNR 20, E min occurred at D of 2, 1, 0, and 1. The extent by which ssvd over- or underestimated flow was found to depend on MTT, tracer timing shift, and R(t). On the other hand, the ratios of estimated flow (F ) to true flow (F) for both osvd and csvd did not vary with shifts but depended on MTT. This is illustrated in Fig. 2, which clearly demonstrates the variability of the magnitude of FIG. 1. Comparison of performances of ssvd, osvd, and csvd in estimating flow for SNR 100 (left column) and 20 (right column) with TR 1 s, and CBV 4% and underlying monoexponential R(t). a and b: Estimated flow (F ) as a function of true flow (F) assuming no phase shift between AIF and C(t) shows comparable performances among the three techniques, with greater underestimation of flow at (b) low SNR. c and d: E (D) and E (D) as a function of tracer arrival timing shifts (D) show that ssvd s performance varies as a function of D, while osvd s and csvd s performances do not. For conspicuity, only the upper error bars (1 SD) are shown for ssvd and csvd, and the lower error bars (1 SD) for osvd.

5 168 Wu et al. Table 2 E (D) E (D) for D 0 and E min for ssvd Compared to Average E (D) E (D) of osvd and csvd for Different SNRs, Residue Functions and Blood Volume (CBV 4% vs CBV 2%) Simulation condition ssvd (D 0) SNR 100 SNR 20 ssvd osvd (E min ) (ml/100 g/min) csvd ssvd (D 0) ssvd osvd (E min ) (ml/100 g/min) Monoexponential R(t) Linear R(t) Box-car R(t) CBV 2% csvd F /F oscillations as functions of R(t) and MTT for ssvd. osvd and csvd produced results that were independent of tracer arrival differences for all R(t), i.e., producing straight lines such as that shown for the monoexponential R(t) (e and f). For osvd, the ratios were also independent of CBV, while csvd performed less accurately at lower CBV and longer MTT (Table 3). The effects of SNR and flow rate on the estimation of timing shifts are shown in Fig. 3a and b. Low flow rates produce the worst estimates. At high flow rates, osvd and csvd have lower ED(F) than ssvd. However, at low flow rates at SNR 20, ssvd estimates have both lower ED(F) and ED (F) than osvd and csvd. Similar behavior was found for the other simulation conditions (graphs not shown). However, there is a systematic error in ssvd in estimating negative shifts at both high and low SNR, demonstrated in Fig. 3c and d for the case F 60 ml/100 g/min. For positive delay values, ssvd performs comparably to osvd and csvd at high SNR. At SNR 20, ssvd is more accurate in determining small shifts, but less accurate for large shifts compared to osvd and csvd. However, when the shift is negative, ssvd has a systematic bias since the technique inherently cannot distinguish negative time shifts from zero time shifts, whereas both csvd and osvd can. Similar findings were obtained under other simulation conditions (graphs not shown). Repeating the above simulations with TR 1.5 s produced results similar to those obtained with TR 1s.For SNR 100 and 20, the optimal P SVD s for ssvd were 4% and 10%, respectively; and for csvd, they were 3% and 10%. For osvd, the optimal OIs were.085 and.095. Greater underestimation of flow than at TR 1 s was found for all techniques at high flow rates at both SNRs. The same sensitivity to time shifts for ssvd flow estimates were detected, while osvd and csvd were still shiftindependent. Oscillations in E (D) were detected for shifts that were not multiples of TR for osvd and csvd; however, they were of smaller magnitude than oscillations due to noise, and oscillations produced by ssvd. The accuracy of time shift estimates continued to be poorer at low flow rates. In addition, time shift estimates were quantized to units of 1.5 s for all three techniques, since shifts that are not multiples of TR cannot be characterized in this present implementation. Comparison of SVD and osvd on Clinically Acquired Human MRI From the results of the Monte Carlo simulations, we limited our clinical analysis to ssvd and osvd. Based on the calculated optimal thresholds for SNR 20 and TR 1.5 s, P SVD 10% was used for ssvd, and OI.095 was used for osvd perfusion analysis. In the case of patient 1 (Fig. 4), we found that ssvd overestimates the amount of tissue at risk of infarction (b and c), while osvd maps (f and g) show much smaller regions of abnormalities, even though the ssvd and osvd maps were created with the same AIF. Patient 1 s initial DWI (a) and CBV (e) show no abnormalities, while the regions of abnormalities on scbf and smtt (b and c) correspond with regions of the greatest phase shifts on the delay maps (d and h). The abnormalities on the osvd maps correspond with chronic regions of hyperintensity on the patient s T 2 fast spin-echo (FSE) (i) acquired at the same session. The patient was diagnosed with transient ischemic attack (TIA). Figure 5 demonstrates that for patient 2, ssvd and osvd can both acutely identify tissue that is ultimately infarcted on follow-up imaging (i) even though it presents with only minimal initial DWI (a) and CBV (e) abnormalities. Large perfusion abnormalities are evident in both ssvd (b and c) and osvd (f and g) maps, and correspond with regions of large shifts on the delay maps for both techniques (d and h). This demonstrates that osvd, although more specific as demonstrated by Fig. 4, can be still sensitive enough to detect tissue at risk of infarction that is initially normal appearing on CBV and DWI maps. Figure 6 demonstrates that osvd can provide a more specific indicator of tissue at risk of infarction than ssvd by showing an example in which there is mismatch in DWI and PWI lesion volumes using ssvd, and no mismatch when using osvd, as in the case of patient 3. The ssvd perfusion maps (b and c) show a much larger volume of tissue at risk of infarction than the amount of tissue shown to have infarcted on the 4-month follow-up T 2 fluid-attenuated inversion recovery (FLAIR) (i). In contrast, the osvd maps (f and g) do not demonstrate a mismatch in lesion volumes with the DWI (a) and CBV (e), and correspond well with the follow-up lesion volume (i). Figure 7 compares the sensitivity of both techniques to AIF selection in the case of patient 4. An AIF selected from the ipsilateral hemisphere along the expected course of the MCA (AIF 1) results in ssvd perfusion maps (b and c) that appear hypoperfused in regions that are inconsistent with the patient s initial DWI (a), initial CBV (b), and follow-up FLAIR study (c). From the delay maps, we see that regions identified as hypoperfused by scbf correspond to negative time differences on the odelay maps (i). In regions that appear hyperperfused (d, arrows), AIF 1 slightly lags the

6 Technique for Estimating CBF in PWI 169 FIG. 2. Comparison of the three techniques accuracy in estimating flow for different R(t) and MTTs. Shown are the ratios of F /F as a function of shifts for SNR 20, TR 1 s, and CBV 4% for ssvd and an R(t) that is (a) monoexponential, (b) linear, (c) box-shaped, or (d) monoexponential with CBV 2%. Results for (e) osvd and (f) csvd assuming monoexponential R(t) are also shown. The upper error bar (1 SD) is shown for MTT 24s(F 10 ml/100g/min for CBV 4%; F 5 ml/100 g/min for CBV 2%), and lower error bars (1 SD) for MTT 6.0s(F 40 ml/100 g/min for CBV 4%; F 20 ml/100 g/min for CBV 2%) and MTT 3.4s (F 70 ml/100 g/min for CBV 4%; F 35 ml/100 g/min for CBV 2%). Table 3 Mean SD of F /F for MTT Ranging From 3.4 to 24 s for osvd and csvd Techniques Assuming SNR 20 Simulation condition osvd (ml/100 g/min) csvd (ml/100 g/min) Monoexponential R(t) to to Linear R(t) to to Box-car R(t) to to CBV 2% to to

7 170 Wu et al. FIG. 3. Comparison of performances of ssvd, osvd, and csvd in estimating tracer arrival time differences for SNR 100 (left column) and 20 (right column) with TR 1 s, and CBV 4% and underlying monoexponential R(t). a and b: ED(F) and ED (F) plotted as a function of F demonstrate that all techniques produce poor estimates at low flow. ssvd produces less ED (F) atlowflow rates and SNR than osvd and csvd. c and d: Estimated shifts (D ) as a function of true shift (D) atf 60 ml/100 g/min show that for D 0, ssvd performs comparably to csvd and osvd at SNR 100, and overestimates large shifts to a greater extent at SNR 20. For D 0, ssvd has a systematic error, whereas osvd and csvd do not. For conspicuity, upper error bars (1 SD) are shown for ssvd and csvd, and lower error bars (1 SD) are shown for osvd. FIG. 4. Performances of ssvd and osvd techniques in patient 1, an example case of ICA stenosis in which there are large scbf and smtt abnormalities (arrowhead) that do not infarct, and minimal ocbf and omtt lesions (arrows). Imaging studies acquired 2 weeks after episodes of transient right-sided weakness: (a) normal-appearing DWI, (b) scbf, (c) smtt, (d) sdelay maps, (e) normal-appearing CBV, (f) ocbf, (g) omtt, and (h) odelay maps, and (i) T2 FSE with chronic regions of hyperintensities (arrow).

8 Technique for Estimating CBF in PWI 171 FIG. 5. Example case showing that both ssvd and osvd can detect regions of tissue at risk of infarction that is initially normal in the DWI. Acute imaging studies for patient 2: (a) DWI, (b) scbf, (c) smtt, (d) sdelay maps, (e) CBV, (f) ocbf, (g) omtt, and (h) odelay maps. i: A 22-day follow-up FLAIR shows that the infarct lesion matches with abnormalities on both ssvd and osvd perfusion maps (arrows). tissue signal, as indicated by the R 2 (t) curves for AIF 1 and ROI 2 (j). The results of osvd (g and h), on the other hand, show regions of hypoperfusion that correspond well with the patient s DWI (a) and CBV (b) lesions, although they underestimate the amount of tissue that is identified as infarcted on follow-up imaging (c). The odelay map (i), however, shows regions with large positive delays that correspond well with the patient s follow-up lesion (c). Choosing instead an AIF from the patient s right posterior cerebral artery (AIF 2) results in ssvd maps (k m) that are quite different from their counterparts using AIF 1 (d f), and that correspond better with the patient s follow-up infarct volume (c). osvd maps, on the other hand, are comparable to those created with AIF 1, with the exception that the odelay map in this case is positive. The relative scbf ratio of ROI 2/ROI 1 was 401% for AIF 1 and 35% for AIF 2. For osvd, the ocbf ratio of ROI 2/ROI 1 was 84% for AIF 1, and 88% for AIF 2. DISCUSSION Monte Carlo Simulations We have presented model-independent techniques (osvd and csvd) that are insensitive to tracer arrival time differences. Importantly, they also perform comparably to the standard SVD (ssvd) technique when there are no differences between the tracer arrival time of the AIF and the tissue signal (Fig. 1). In contrast, ssvd flow estimates have a high dependence on tracer arrival time differences (D), MTT, and the underlying R(t) (Fig. 2). This is especially true for negative shifts and at high SNR. We have extended previous studies that showed that ssvd was independent of underlying R(t) (3) by examining ssvd s performance as a function of varying R(t) and delayed tracer arrival. These techniques all depend on MTT because of insufficient temporal sampling with which to properly characterize high flow rates. FIG. 6. Example case of a mismatch in DWI and PWI lesion volumes when using ssvd, and no mismatch when using osvd. Acute imaging studies for patient 3: (a) DWI, (b) scbf, (c) smtt, (d) sdelay maps, (e) CBV, (f) ocbf, (g) omtt, and (h) odelay maps. i: A 4-month follow-up FLAIR shows an infarct that is well matched with the patient s acute DWI, CBV, ocbf, and omtt studies (arrows).

9 172 Wu et al. FIG. 7. Comparison of sensitivity of ssvd and osvd to AIF selection for patient 4. Lesions on initial (a) DWI and (b) CBV studies (small arrows) extend into larger infarct volume shown on (c) 6-day follow-up FLAIR. Shown are (d) scbf, (e) smtt (f) sdelay, (g) ocbf, (h) omtt, and (i) odelay maps using AIF1. Regions darker than background on the (i) odelay map represent negative values (small arrow). j: R 2 (t) for both AIFs and for two 5 5 ROIs. k: scbf. l: smtt. m: sdelay. n: ocbf. o: omtt. p: odelay maps using AIF2. Regions of hyperperfusion (arrows) or hypoperfusion (arrowheads) depend on AIF selection for ssvd (d and k) but not for osvd (g and n). Interestingly, for ssvd at low SNR, the effects of D on flow estimation have less variation than at high SNR due to the use of a higher P SVD (20% vs. 4%), demonstrating the importance of P SVD selection. Low P SVD ( 5%) has been shown to lead to greater overestimation of flow even when D 0 (18). Because the extent of overestimation increases for D 0 (Fig. 1), the use of a low P SVD resulted in even greater overestimation of flow. Because P SVD was determined assuming D 0, a low P SVD led to more accurate flow estimates (Fig. 1) when D 0, but led to larger E (D) for D 0 at high SNR. E min for SNR 100 was noted to occur at D 0, whereas for SNR 20, it occurred for D 0. Furthermore, E min for SNR 20 was sometimes lower than E min for SNR 100 (Fig. 2), since some flow values typically underestimated for D 0 (e.g., MTT 3.4 s) were accurately determined at D 0atSNR 20. In contrast, for SNR 100, flow values are overestimated for D 0. For D 0, E (D) for ssvd reaches steady-state values that fluctuate slightly about E (D) for osvd and csvd, as shown in Fig. 1c. Our findings suggest that even in the absence of tracer arrival time differences, ssvd CBF values are underestimated. This makes absolute quantification of CBF difficult (Figs. 1 and 2), consistent with results from previous simulation studies (3). However, our results show that if MTT is preserved, the accuracy of CBF estimates in gray and white matter would be comparable, e.g., estimates at CBV 4% and 2% (Fig. 2a and d), producing ratios of gray to white matter CBF values that correlate well with those from other quantitative imaging modalities. This is consistent with findings from studies performed on normal human volunteers in which MR CBF was compared with positron emission tomography (PET) (5) and single photon emission-computed tomography (SPECT) CBF (23) values. We also investigated the effects of changing the sampling rate, and the implications of these effects for flow

10 Technique for Estimating CBF in PWI 173 determination. As the TR increases, the performance of all of the algorithms degrades for estimating high flow rates. This is almost surely due to insufficient temporal sampling to properly characterize the hemodynamic properties of the measured tissue. Furthermore, the optimal thresholds for ssvd, osvd, and csvd techniques also change, becoming less stringent, with more data points in W and W c kept. In the presence of arrival time differences that are not multiples of TR, osvd and csvd also exhibit slight oscillations. Although this is the most likely situation in clinically acquired data, the fluctuations that result are smaller in magnitude than those due to noise and those measured in ssvd. Furthermore, by interpolating the signal to higher sampling rates, one can potentially compensate for flow errors due to these non-sample-unit shifts. Of the techniques investigated in this study that use block-circulant matrices, the osvd technique (which varied the SVD threshold on an individual voxel basis) showed improved estimation over csvd (which used a global SVD threshold). These findings are consistent with the results of Liu et al. (18), who showed that an adaptive technique with ssvd based on each voxel s concentration characteristic provided improved performance over ssvd using a global threshold. We did not separately compare improvements that may be due to voxel-based adaptive threshold techniques with those resulting from the use of a block-circular matrix (csvd), since ssvd did not include csvd s robustness to AIF lags to the tissue signal. In addition to the proposed technique (Eq. [6]), there may be other methods to adaptively minimize oscillations based on each individual voxel s tissue signal characteristics, which may further improve osvd flow estimates (17,18,24). Finally, in this study we also examined techniques for assessing the time difference between tracer arrival in tissue and the AIF by using the time of the peak of the deconvolved R(t). At high flow rates, all techniques performed comparably (Fig. 3). At low flow rates, all techniques had larger errors. However, ssvd cannot estimate negative shifts (Fig. 3c and d), and thus odelay maps are preferable. Comparison of SVD and osvd on Clinically Acquired Human MR Images We have presented examples of how the errors demonstrated by Monte Carlo simulations may translate in clinically acquired PW images, and how the proposed osvd technique may compensate for them. Furthermore, our findings suggest that scbf and smtt maps are weighted by delayed tracer arrival. In three of four cases (Figs. 4 6), the regions identified as hypoperfused with ssvd (b and c) correspond with regions of greatest delay (d and h). In regions of mismatch between ssvd and osvd in tissue that did not infarct (Figs. 4 and 6), the ratio of the flow with respect to the contralateral hemisphere was reduced by as much as 50% for phase shifts 2 s. This is consistent with our simulation results (Fig. 2), which showed that for time delays in this range, CBF in normal gray matter (F 60 ml/100 g/min) can be underestimated by as much as 50%, whereas tissue without tracer delay is estimated more accurately. This is most evident assuming a box-car residue function, which has been reported to correlate well with acquired human PWI (3). Furthermore, a 50% underestimation of CBF values has also been measured experimentally in previous studies of the effects of delays on acquired human PWI, in which single AIFs were artificially shifted in time (5,7). As a result, when taking relative flow ratios in ssvd, one may erroneously detect an overall 50% reduction that could be due to tracer arrival differences. Although osvd can also underestimate flow by close to 50% (Fig. 2, Table 3), it does so for tissue with and without tracer arrival differences. Therefore, when relative CBF analysis is performed, assuming equivalent flow rates, the ratio of tissue with delayed to nondelayed tracer arrival will be close to unity, resulting in less relative differences between the hemispheres in the osvdgenerated maps. Furthermore, osvd s use of an adaptive local threshold may provide additional accuracy over ssvd for different tissue types (18). Our findings also demonstrate that delay-insensitive CBF estimates using osvd may provide results that are robust not only to differences in tracer arrival times, but also to AIF selection (Fig. 7). This suggests that osvd maps will be less sensitive to user variability in AIF selection, which will perhaps make automated AIF selection algorithms (25) more feasible. To determine whether osvd provides more accurate quantitative relative flow estimates, correlation with PET flow values in humans and experimental animal stroke models should be performed in future studies. With less contamination of flow estimates by tracer arrival time differences between the AIF and tissue signals, it may be possible to obtain improved identification of salvageable tissue. Due to tracer arrival delay, the existing ssvd technique may produce estimates of flow that indicate ischemia in regions that are actually oligemic or normally perfused. This in turn may contribute to the large sensitivity and lack of specificity in traditional scbf and smtt maps (2). However, one can argue that, clinically, what is desired is a sensitive indicator of tissue with disturbed hemodynamics, and that therefore delay contamination in scbf and smtt maps is a clinically useful feature. The shift maps may represent tissue that is downstream from an occlusion or stenosis, but which may still receive sufficient flow at the time of imaging. The ocbf maps, on the other hand, represent the instantaneous flow at the acquisition time. Since CBF is a dynamic process, a single snapshot may not be a sensitive predictor of future infarction. Therefore, we speculate that the odelay maps presented here can continue to provide sensitivity in identifying tissue with disturbed hemodynamics, while the ocbf maps provide a more specific snapshot of the severity of ischemia in tissue at risk of infarction. The clinical utility of decoupling flow from delayed tracer arrival still remains to be proven in a larger study. The clinical findings presented here are clearly preliminary, and were meant to demonstrate the techniques in practice rather than prove their accuracy. Additional studies involving a larger cohort of patients are necessary to better determine the predictive power of osvd perfusion maps. Retrospective analysis of perfusion studies with follow-up MRIs should be performed in which lesion volumes identified by ocbf, odelay, CBV, scbf, and smtt

11 174 Wu et al. are correlated with lesion volumes on follow-up imaging studies. Alternatively, one may utilize predictive modeling algorithms that comprise the different sets of perfusion parameters, and evaluate which generated risk maps are more predictive of tissue infarction in terms of sensitivity and specificity (26). CONCLUSIONS We have extended previous studies that examined the sensitivity of the SVD technique to delay and dispersion (3 6,10) by investigating conditions in which the AIF lags the tissue signal, assuming different residue functions and CBV. We have proposed a technique that is less sensitive to tracer arrival timing differences than the current SVD technique, without sacrificing performance for different underlying residue functions and CBV. We have also demonstrated that the new technique may provide improved accuracy in the identification of tissue at risk of infarction in clinically acquired PWI. ACKNOWLEDGMENTS The authors express their gratitude to Joanie O Donnell, R.N., for chart review, and the Departments of Neuroradiology and Neurology, Massachusetts General Hospital, for their assistance in clinical data acquisition. REFERENCES 1. Baird AE, Warach S. Magnetic resonance imaging of acute stroke. J Cereb Blood Flow Metab 1998;18: Sorensen AG, Copen WA, Østergaard L, Buonanno FS, Gonzalez RG, Rordorf G, Rosen BR, Schwamm LH, Weisskoff RM, Koroshetz WJ. Hyperacute stroke: simultaneous measurement of relative cerebral blood volume, relative cerebral blood flow, and mean tissue transit time. Radiology 1999;210: Østergaard L, Weisskoff RM, Chesler DA, Gyldensted C, Rosen BR. High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part I: mathematical approach and statistical analysis. Magn Reson Med 1996;36: Østergaard L, Sorensen AG, Kwong KK, Weisskoff RM, Gyldensted C, Rosen BR. High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part II: experimental comparison and preliminary results. Magn Reson Med 1996;36: Østergaard L, Johannsen P, Host-Poulsen P, Vestergaard-Poulsen P, Asboe H, Gee AD, Hansen SB, Cold GE, Gjedde A, Gyldensted C. Cerebral blood flow measurements by magnetic resonance imaging bolus tracking: comparison with [(15)O]H2O positron emission tomography in humans. J Cereb Blood Flow Metab 1998;18: Calamante F, Gadian DG, Connelly A. Delay and dispersion effects in dynamic susceptibility contrast MRI: simulations using singular value decomposition. Magn Reson Med 2000;44: Wu O, Østergaard L, Benner T, Rosen BR, Weisskoff RM, Sorensen AG. Perfusion calculations using singular value decomposition are biased by tracer arrival timing [abstract]. In: Proceedings of the 10th Annual Meeting of ISMRM, Honolulu, Wirestam R, Andersson L, Østergaard L, Bolling M, Aunola JP, Lindgren A, Geijer B, Holtas S, Stahlberg F. Assessment of regional cerebral blood flow by dynamic susceptibility contrast MRI using different deconvolution techniques. Magn Reson Med 2000;43: Vonken EP, Beekman FJ, Bakker CJ, Viergever MA. Maximum likelihood estimation of cerebral blood flow in dynamic susceptibility contrast MRI. Magn Reson Med 1999;41: Østergaard L, Chesler DA, Weisskoff RM, Sorensen AG, Rosen BR. Modeling cerebral blood flow and flow heterogeneity from magnetic resonance residue data. J Cereb Blood Flow Metab 1999;19: Smith AM, Grandin CB, Duprez T, Mataigne F, Cosnard G. Whole brain quantitative CBF and CBV measurements using MRI bolus tracking: comparison of methodologies. Magn Reson Med 2000;43: Grandin CB, Duprez TP, Smith AM, Oppenheim C, Peeters A, Robert AR, Cosnard G. Which MR-derived perfusion parameters are the best predictors of infarct growth in hyperacute stroke? Comparative study between relative and quantitative measurements. Radiology 2002;223: Lythgoe DJ, Ostergaard L, William SC, Cluckie A, Buxton-Thomas M, Simmons A, Markus HS. Quantitative perfusion imaging in carotid artery stenosis using dynamic susceptibility contrast-enhanced magnetic resonance imaging. Magn Reson Imaging 2000;18: Rosen BR, Belliveau JW, Chien D. Perfusion imaging by nuclear magnetic resonance. Magn Reson Q 1989;5: Porkka L, Neuder M, Hunter G, Weisskoff RM, Belliveau JW, Rosen BR. Arterial input function measurement with MRI. In: Proceedings of the 10th Annual Meeting of SMRM, San Francisco, p Oppenheim AV, Schafer RW. The discrete Fourier transform. In: Discrete-time signal processing. Englewood Cliffs, NJ: Prentice Hall; p Lagendijk RL, Biemond J. Image formation models. In: Iterative identification and restoration of images. Boston: Kluwer Academic Publishers; p Liu HL, Pu Y, Liu Y, Nickerson L, Andrews T, Fox PT, Gao JH. Cerebral blood flow measurement by dynamic contrast MRI using singular value decomposition with an adaptive threshold. Magn Reson Med 1999;42: Gobbel GT, Fike JR. A deconvolution method for evaluating indicatordilution curves. Phys Med Biol 1994;39: Stewart GN. Researches on the circulation time in organs and on the influences which affect it (Parts I III). J Physiol 1894;15: Henkelman RM. Measurement of signal intensities in the presence of noise in MR images. Med Phys 1985;12: Weisskoff RM, Zuo CS, Boxerman JL, Rosen BR. Microscopic susceptibility variation and transverse relaxation: theory and experiment. Magn Reson Med 1994;31: Wirestam R, Ryding E, Lindgren A, Geijer B, Ostergaard L, Andersson L, Holtas S, Stahlberg F. Regional cerebral blood flow distributions in normal volunteers: dynamic susceptibility contrast MRI compared with 99mTc-HMPAO SPECT. J Comput Assist Tomogr 2000;24: Tikhonov AN, Arsenin VY. Solutions of ill-posed problems. New York: Halsted Press; p. 25. Alsop D, Wedmid A, Schlaug G. Defining a local input function for perfusion quantification with bolus contrast MRI [abstract]. In: Proceedings of the 10th Annual Meeting of ISMRM, Honolulu, Wu O, Koroshetz WJ, Østergaard L, Buonanno FS, Copen WA, Gonzalez RG, Rordorf G, Rosen BR, Schwamm LH, Weisskoff RM, Sorensen AG. Predicting tissue outcome in acute human cerebral ischemia using combined diffusion- and perfusion-weighted MR imaging. Stroke 2001; 32:

positron emission tomography PET

positron emission tomography PET 020-8505 19-1 1-3) positron emission tomography PET PET PET single photon emission computed tomography SPECT acetazolamide 4-10) magnetic resonance imaging MRI MRI 11,12) MRI perfusion weighted imaging

More information

The role of diffusion and perfusion MRI for the identification

The role of diffusion and perfusion MRI for the identification The Physiological Significance of the Time-to-Maximum (Tmax) Parameter in Perfusion MRI Fernando Calamante, PhD; Søren Christensen, PhD; Patricia M. Desmond, MD, FRANZCR; Leif Østergaard, MD, PhD; Stephen

More information

Tmax Determined Using a Bayesian Estimation Deconvolution Algorithm Applied to Bolus Tracking Perfusion Imaging: A Digital Phantom Validation Study

Tmax Determined Using a Bayesian Estimation Deconvolution Algorithm Applied to Bolus Tracking Perfusion Imaging: A Digital Phantom Validation Study Magn Reson Med Sci 2017; 16; 32 37 doi:10.2463/mrms.mp.2015-0167 Published Online: March 21, 2016 MAJOR PAPER Tmax Determined Using a Bayesian Estimation Deconvolution Algorithm Applied to Bolus Tracking

More information

Dynamic susceptibility contract-enhanced MRI is increasingly

Dynamic susceptibility contract-enhanced MRI is increasingly Influence of Arterial Input Function on Hypoperfusion Volumes Measured With Perfusion-Weighted Imaging Vincent N. Thijs, MD; Diederik M. Somford, MD; Roland Bammer, PhD; Wim Robberecht, MD, PhD; Michael

More information

MR Perfusion Imaging in Moyamoya Syndrome. Potential Implications for Clinical Evaluation of Occlusive Cerebrovascular Disease

MR Perfusion Imaging in Moyamoya Syndrome. Potential Implications for Clinical Evaluation of Occlusive Cerebrovascular Disease MR Perfusion Imaging in Moyamoya Syndrome Potential Implications for Clinical Evaluation of Occlusive Cerebrovascular Disease F. Calamante, PhD; V. Ganesan, MD; F.J. Kirkham, MB, BChir; W. Jan, MBBS; W.K.

More information

Detecting and rescuing the ischemic penumbra is the main

Detecting and rescuing the ischemic penumbra is the main Influence of the Arterial Input Function on Absolute and Relative Perfusion-Weighted Imaging Penumbral Flow Detection A Validation With 15 O-Water Positron Emission Tomography Olivier Zaro-Weber, MD; Walter

More information

The association between neurological deficit in acute ischemic stroke and mean transit time

The association between neurological deficit in acute ischemic stroke and mean transit time Neuroradiology (2006) 48: 69 77 DOI 10.1007/s00234-005-0012-9 DIAGNOSTIC NEURORADIOLOGY Peter D. Schellinger Lawrence L. Latour Chen-Sen Wu Julio A. Chalela Steven Warach The association between neurological

More information

PERFUSION MRI CONTRAST BASED TECHNIQUES

PERFUSION MRI CONTRAST BASED TECHNIQUES PERFUSION MRI CONTRAST BASED TECHNIQUES by Kenny K Israni Mar 28, 2006 PERFUSION - MRI Dynamic Susceptibility contrast Dynamic Relaxivity contrast STEADY-STATE STATE TECHNIQUES Steady-state Susceptibility

More information

Early and accurate detection of tissue at risk of infarction

Early and accurate detection of tissue at risk of infarction The Role of Bolus Delay and Dispersion in Predictor Models for Stroke Lisa Willats, PhD; Alan Connelly, PhD; Soren Christensen, PhD; Geoffrey A. Donnan, MD, FRACP; Stephen M. Davis, MD, FRACP; Fernando

More information

The advent of multidetector-row CT (MDCT) has facilitated

The advent of multidetector-row CT (MDCT) has facilitated Published September 3, 2008 as 10.3174/ajnr.A1274 ORIGINAL RESEARCH M. Sasaki K. Kudo K. Ogasawara S. Fujiwara Tracer Delay Insensitive Algorithm Can Improve Reliability of CT Perfusion Imaging for Cerebrovascular

More information

Acute stroke is the third leading cause of death and the

Acute stroke is the third leading cause of death and the Combined Diffusion-Weighted and Perfusion-Weighted Flow Heterogeneity Magnetic Resonance Imaging in Acute Stroke Leif Østergaard, MD, MSc; A. Gregory Sorensen, MD; David A. Chesler, PhD; Robert M. Weisskoff,

More information

occlusions. Cerebral perfusion is driven fundamentally by regional cerebral

occlusions. Cerebral perfusion is driven fundamentally by regional cerebral Appendix Figures Figure A1. Hemodynamic changes that may occur in major anterior circulation occlusions. Cerebral perfusion is driven fundamentally by regional cerebral perfusion pressure (CPP). In response

More information

Confounding Effect of Large Vessels on MR Perfusion Images Analyzed with Independent Component Analysis

Confounding Effect of Large Vessels on MR Perfusion Images Analyzed with Independent Component Analysis AJNR Am J Neuroradiol 23:1007 1012, June/July 2002 Confounding Effect of Large Vessels on MR Perfusion Images Analyzed with Independent Component Analysis Timothy J. Carroll, Victor M. Haughton, Howard

More information

AIF and deconvolution

AIF and deconvolution AIF and deconvolution Birgitte Fuglsang Kjølby, ISMRM May 2012 Center for Functionally Integrative Neuroscience, Aarhus University Hospital, Denmark Introduction Dynamic susceptibility contrast magnetic

More information

PHYSICS OF MRI ACQUISITION. Alternatives to BOLD for fmri

PHYSICS OF MRI ACQUISITION. Alternatives to BOLD for fmri PHYSICS OF MRI ACQUISITION Quick Review for fmri HST-583, Fall 2002 HST.583: Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Harvard-MIT Division of Health Sciences and Technology

More information

Perfusion MRI. Youngkyoo Jung, PhD Associate Professor Radiology, Biomedical Engineering, and Clinical & Translational Science Institute

Perfusion MRI. Youngkyoo Jung, PhD Associate Professor Radiology, Biomedical Engineering, and Clinical & Translational Science Institute Perfusion MRI Youngkyoo Jung, PhD Associate Professor Radiology, Biomedical Engineering, and Clinical & Translational Science Institute Perfusion The delivery of blood to a capillary bed in tissue Perfusion

More information

CT and MR perfusion (MRP) imaging are used frequently in. Exposing Hidden Truncation-Related Errors in Acute Stroke Perfusion Imaging

CT and MR perfusion (MRP) imaging are used frequently in. Exposing Hidden Truncation-Related Errors in Acute Stroke Perfusion Imaging ORIGINAL RESEARCH BRAIN Exposing Hidden Truncation-Related Errors in Acute Stroke Perfusion Imaging W.A. Copen, A.R. Deipolyi, P.W. Schaefer, L.H. Schwamm, R.G. González, and O. Wu ABSTRACT BACKGROUND

More information

Abstract. Introduction. Material and Methods. Results. Conclusion

Abstract. Introduction. Material and Methods. Results. Conclusion Dynamic susceptibility contrast MRI calibrated using T1-based steadystate CBV and vascular space occupancy (VASO): Comparison with model-free arterial spin labelling Emelie Lindgren Supervisors: Linda

More information

Dong Gyu Na, Vincent N. Thijs, Gregory W. Albers, Michael E. Moseley, and Michael P. Marks. AJNR Am J Neuroradiol 25: , September 2004

Dong Gyu Na, Vincent N. Thijs, Gregory W. Albers, Michael E. Moseley, and Michael P. Marks. AJNR Am J Neuroradiol 25: , September 2004 AJNR Am J Neuroradiol 25:1331 1336, September 2004 Diffusion-Weighted MR Imaging in Acute Ischemia: Value of Apparent Diffusion Coefficient and Signal Intensity Thresholds in Predicting Tissue at Risk

More information

Whole brain CT perfusion maps with paradoxical low mean transit time to predict infarct core

Whole brain CT perfusion maps with paradoxical low mean transit time to predict infarct core Whole brain CT perfusion maps with paradoxical low mean transit time to predict infarct core Poster No.: B-292 Congress: ECR 2011 Type: Scientific Paper Topic: Neuro Authors: S. Chakraborty, M. E. Ahmad,

More information

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling Investigations and research Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling J. Wang Z. Ying V. Yao L. Ciancibello S. Premraj S. Pohlman

More information

Diffusion-weighted Magnetic Resonance Imaging in the Emergency Department

Diffusion-weighted Magnetic Resonance Imaging in the Emergency Department 298 / = Abstract = Diffusion-weighted Magnetic Resonance Imaging in the Emergency Department Sung Pil Chung, M.D, Suk Woo Lee, M.D., Young Mo Yang, M.D., Young Rock Ha, M.D., Seung Whan Kim, M.D., and

More information

Dynamic susceptibility contrast MR imaging was performed to determine the local cerebral blood flow and blood volume.

Dynamic susceptibility contrast MR imaging was performed to determine the local cerebral blood flow and blood volume. RSNA, 2016 10.1148/radiol.2016152244 Appendix E1 Neuropsychological Tests All patients underwent a battery of neuropsychological tests. This included the Rey auditory verbal learning test (34), the digit

More information

Arterial Transit Time Imaging With Flow Encoding Arterial Spin Tagging (FEAST)

Arterial Transit Time Imaging With Flow Encoding Arterial Spin Tagging (FEAST) Magnetic Resonance in Medicine 50:599 607 (2003) Arterial Transit Time Imaging With Flow Encoding Arterial Spin Tagging (FEAST) Jiongjiong Wang, 1,2 * David C. Alsop, 3 Hee Kwon Song, 1 Joseph A. Maldjian,

More information

Diffusion- and Perfusion-Weighted MRI. The DWI/PWI Mismatch Region in Acute Stroke

Diffusion- and Perfusion-Weighted MRI. The DWI/PWI Mismatch Region in Acute Stroke Diffusion- and Perfusion-Weighted MRI The DWI/PWI Mismatch Region in Acute Stroke Tobias Neumann-Haefelin, MD; Hans-Jörg Wittsack, PhD; Frank Wenserski, MD; Mario Siebler, MD; Rüdiger J. Seitz, MD; Ulrich

More information

Reduced cerebral blood flow in white matter in ischaemic leukoaraiosis demonstrated using quantitative exogenous contrast based perfusion MRI

Reduced cerebral blood flow in white matter in ischaemic leukoaraiosis demonstrated using quantitative exogenous contrast based perfusion MRI 48 Department of Clinical Neurosciences, Guy s, King s, and St Thomas Hospital School of Medicine and the Institute of Psychiatry, London, UK H S Markus D J Lythgoe M O Sullivan S C R Williams Department

More information

Perfusion Physics. ICMRI2018 March 29-31, 2018 Grand Hilton Hotel, Seoul, Korea. Asian Forum Ⅱ: Perfusion MRI SY24-1.

Perfusion Physics. ICMRI2018 March 29-31, 2018 Grand Hilton Hotel, Seoul, Korea. Asian Forum Ⅱ: Perfusion MRI SY24-1. SY24-1 Perfusion Physics Hiroyuki Kabasawa MR Collaborations and Development, GE Healthcare, Tokyo, Japan Perfusion is referred as the blood supply to micro capillary in tissue. Perfusion parameter such

More information

Assessing Tissue Viability with MR Diffusion and Perfusion Imaging

Assessing Tissue Viability with MR Diffusion and Perfusion Imaging AJNR Am J Neuroradiol 24:436 443, March 2003 Assessing Tissue Viability with MR Diffusion and Perfusion Imaging Pamela W. Schaefer, Yelda Ozsunar, Julian He, Leena M. Hamberg, George J. Hunter, A. Gregory

More information

DIFFUSION AND DYNAMIC SUSCEPTIBILITY CONTRAST PERFUSION WEIGHTED MAGNETIC RESONANCE IMAGING IN HUMAN ACUTE ISCHEMIC STROKE

DIFFUSION AND DYNAMIC SUSCEPTIBILITY CONTRAST PERFUSION WEIGHTED MAGNETIC RESONANCE IMAGING IN HUMAN ACUTE ISCHEMIC STROKE UNIVERSITY OF HELSINKI REPORT SERIES IN PHYSICS HU-P-D120 DIFFUSION AND DYNAMIC SUSCEPTIBILITY CONTRAST PERFUSION WEIGHTED MAGNETIC RESONANCE IMAGING IN HUMAN ACUTE ISCHEMIC STROKE Jussi Perkiö Department

More information

Keiichi Kikuchi, Kenya Murase, Hitoshi Miki, Takanori Kikuchi, Yoshifumi Sugawara, Teruhito Mochizuki, Junpei Ikezoe, and Shiro Ohue

Keiichi Kikuchi, Kenya Murase, Hitoshi Miki, Takanori Kikuchi, Yoshifumi Sugawara, Teruhito Mochizuki, Junpei Ikezoe, and Shiro Ohue AJNR Am J Neuroradiol 22:248 254, February 2001 Measurement of Cerebral Hemodynamics with Perfusionweighted MR Imaging: Comparison with Pre- and Post-acetazolamide Xe-SPECT in Occlusive Carotid Disease

More information

Perfusion-weighted imaging (PWI) and diffusionweighted

Perfusion-weighted imaging (PWI) and diffusionweighted Apparent Diffusion Coefficient Thresholds Do Not Predict the Response to Acute Stroke Thrombolysis Poh-Sien Loh, FRACP; Ken S. Butcher, MD, PhD, FRCP(C), FRACP; Mark W. Parsons, PhD, FRACP; Lachlan MacGregor,

More information

Robust Dynamic Susceptibility Contrast MR Perfusion Using 4D Nonlinear Noise Filters

Robust Dynamic Susceptibility Contrast MR Perfusion Using 4D Nonlinear Noise Filters JOURNAL OF MAGNETIC RESONANCE IMAGING 26:1514 1522 (2007) Original Research Robust Dynamic Susceptibility Contrast MR Perfusion Using 4D Nonlinear Noise Filters Jayme Cameron Kosior, BSc, 1,2,5 Robert

More information

MRI based diffusion and perfusion predictive model to estimate stroke evolution

MRI based diffusion and perfusion predictive model to estimate stroke evolution Magnetic Resonance Imaging 19 (2001) 1043 1053 MRI based diffusion and perfusion predictive model to estimate stroke evolution Stephen E. Rose a, *, Jonathan B. Chalk a,b, Mark P. Griffin a, Andrew L.

More information

Related Symposia in AAPM 2007

Related Symposia in AAPM 2007 Related Symposia in AAPM 7 Functional and Physiological MR Imaging for Therapy Assessment Yue Cao,, Ph.D. Departments of Radiation Oncology and Radiology, University of Michigan President s s symposium:

More information

Imaging Acute Stroke and Cerebral Ischemia

Imaging Acute Stroke and Cerebral Ischemia Department of Radiology University of California San Diego Imaging Acute Stroke and Cerebral Ischemia John R. Hesselink, M.D. Causes of Stroke Arterial stenosis Thrombosis Embolism Dissection Hypotension

More information

Rapid MRI of the passage of a bolus of magnetic susceptibility

Rapid MRI of the passage of a bolus of magnetic susceptibility Cerebral Blood Flow and Blood Volume Measured by Magnetic Resonance Imaging Bolus Tracking After Acute Stroke in Pigs Comparison With [ 15 O]H 2 O Positron Emission Tomography Masaharu Sakoh, MD, PhD;

More information

How Reliable Is Perfusion MR in Acute Stroke? Validation and Determination of the Penumbra Threshold Against Quantitative PET

How Reliable Is Perfusion MR in Acute Stroke? Validation and Determination of the Penumbra Threshold Against Quantitative PET How Reliable Is Perfusion MR in Acute Stroke? Validation and Determination of the Penumbra Threshold Against Quantitative Masashi Takasawa, MD, PhD; P. Simon Jones, MSc; Joseph V. Guadagno, PhD, MRCP;

More information

Speed, Comfort and Quality with NeuroDrive

Speed, Comfort and Quality with NeuroDrive Speed, Comfort and Quality with NeuroDrive Echelon Oval provides a broad range of capabilities supporting fast, accurate diagnosis of brain conditions and injuries. From anatomical depiction to vascular

More information

Complete Recovery of Perfusion Abnormalities in a Cardiac Arrest Patient Treated with Hypothermia: Results of Cerebral Perfusion MR Imaging

Complete Recovery of Perfusion Abnormalities in a Cardiac Arrest Patient Treated with Hypothermia: Results of Cerebral Perfusion MR Imaging pissn 2384-1095 eissn 2384-1109 imri 2018;22:56-60 https://doi.org/10.13104/imri.2018.22.1.56 Complete Recovery of Perfusion Abnormalities in a Cardiac Arrest Patient Treated with Hypothermia: Results

More information

Assessing Tissue Oxygenation and Predicting Tissue Outcome in Acute Stroke

Assessing Tissue Oxygenation and Predicting Tissue Outcome in Acute Stroke The Danish National Research Foundation s Center of Functionally Integrative Neuroscience Aarhus University / Aarhus University Hospital - DENMARK Assessing Tissue Oxygenation and Predicting Tissue Outcome

More information

Perfusion-Based fmri. Thomas T. Liu Center for Functional MRI University of California San Diego May 7, Goal

Perfusion-Based fmri. Thomas T. Liu Center for Functional MRI University of California San Diego May 7, Goal Perfusion-Based fmri Thomas T. Liu Center for Functional MRI University of California San Diego May 7, 2006 Goal To provide a basic understanding of the theory and application of arterial spin labeling

More information

Turbo ASL: Arterial Spin Labeling With Higher SNR and Temporal Resolution

Turbo ASL: Arterial Spin Labeling With Higher SNR and Temporal Resolution COMMUNICATIONS Magnetic Resonance in Medicine 44:511 515 (2000) Turbo ASL: Arterial Spin Labeling With Higher SNR and Temporal Resolution Eric C. Wong,* Wen-Ming Luh, and Thomas T. Liu A modified pulsed

More information

Experimental Assessment of Infarct Lesion Growth in Mice using Time-Resolved T2* MR Image Sequences

Experimental Assessment of Infarct Lesion Growth in Mice using Time-Resolved T2* MR Image Sequences Experimental Assessment of Infarct Lesion Growth in Mice using Time-Resolved T2* MR Image Sequences Nils Daniel Forkert 1, Dennis Säring 1, Andrea Eisenbeis 2, Frank Leypoldt 3, Jens Fiehler 2, Heinz Handels

More information

CT Perfusion Parameter Values in Regions of Diffusion Abnormalities

CT Perfusion Parameter Values in Regions of Diffusion Abnormalities AJNR Am J Neuroradiol 25:1205 1210, August 2004 CT Perfusion Parameter Values in Regions of Diffusion Abnormalities Marcello Galvez, Gerald E. York, II, and James D. Eastwood BACKGROUND AND PURPOSE: Dynamic

More information

Title: Stability of Large Diffusion/Perfusion Mismatch in Anterior Circulation Strokes for 4 or More Hours

Title: Stability of Large Diffusion/Perfusion Mismatch in Anterior Circulation Strokes for 4 or More Hours Author's response to reviews Title: Stability of Large Diffusion/Perfusion Mismatch in Anterior Circulation Strokes for 4 or More Hours Authors: Ramon G. Gonzalez (rggonzalez@partners.org) Reza Hakimelahi

More information

Thrombolysis is the only proven pharmacological treatment

Thrombolysis is the only proven pharmacological treatment Perfusion Thresholds in Acute Stroke Thrombolysis K. Butcher, MD, PhD, FRCP(C); M. Parsons, PhD, FRACP; T. Baird, MBBS; A. Barber, PhD, FRANZ; G. Donnan, MD, FRACP; P. Desmond, FRACR; B. Tress, FRACR;

More information

NIH Public Access Author Manuscript Neuroradiology. Author manuscript; available in PMC 2009 September 1.

NIH Public Access Author Manuscript Neuroradiology. Author manuscript; available in PMC 2009 September 1. NIH Public Access Author Manuscript Published in final edited form as: Neuroradiology. 2008 September ; 50(9): 745 751. doi:10.1007/s00234-008-0403-9. Perfusion-CT compared to H 2 15O/O 15 O PET in Patients

More information

Reliable Perfusion Maps in Stroke MRI Using Arterial Input Functions Derived From Distal Middle Cerebral Artery Branches

Reliable Perfusion Maps in Stroke MRI Using Arterial Input Functions Derived From Distal Middle Cerebral Artery Branches Reliable Perfusion Maps in Stroke MRI Using Arterial Input Functions Derived From Distal Middle Cerebral Artery Branches Martin Ebinger, MD; Peter Brunecker, MSc; Gerhard J. Jungehülsing, MD; Uwe Malzahn,

More information

Diagnostic improvement from average image in acute ischemic stroke

Diagnostic improvement from average image in acute ischemic stroke Diagnostic improvement from average image in acute ischemic stroke N. Magne (1), E.Tollard (1), O. Ozkul- Wermester (2), V. Macaigne (1), J.-N. Dacher (1), E. Gerardin (1) (1) Department of Radiology,

More information

Cha pter 6. Perfusion MRI in Neuro-Psychiatric Systemic Lupus Erythemathosus

Cha pter 6. Perfusion MRI in Neuro-Psychiatric Systemic Lupus Erythemathosus Cha pter 6 Perfusion MRI in Neuro-Psychiatric Systemic Lupus Erythemathosus B. J. Emmer 1, M. J. van Osch 1, O. Wu 2, G. M. Steup-Beekman 3, S. C. Steens 1, T. W. Huizinga 3, M. A. van Buchem 1, J. van

More information

Assessment of Tissue Viability Using Diffusion- and Perfusion-Weighted MRI in Hyperacute Stroke

Assessment of Tissue Viability Using Diffusion- and Perfusion-Weighted MRI in Hyperacute Stroke Assessment of Tissue Viability Using Diffusion- and Perfusion-Weighted MRI in Hyperacute Stroke Won-Jin Moon, MD 1, 2 Dong Gyu Na, MD 1, 3 Jae Wook Ryoo, MD 1, 4 Hong Gee Roh, MD 1 Hong Sik Byun, MD 1

More information

BTB improvement after dexamethasone treatment. 11 This has led to the theory that dexamethasone acts

BTB improvement after dexamethasone treatment. 11 This has led to the theory that dexamethasone acts J Neurosurg 90:300 305, 1999 Early changes measured by magnetic resonance imaging in cerebral blood flow, blood volume, and blood-brain barrier permeability following dexamethasone treatment in patients

More information

CT/MRI 第 42 回日本脳卒中学会講演シンポジウム 総説 はじめに. MRI magnetic resonance imaging DWI diffusion-weighted. DWI/CBF malignant profile

CT/MRI 第 42 回日本脳卒中学会講演シンポジウム 総説 はじめに. MRI magnetic resonance imaging DWI diffusion-weighted. DWI/CBF malignant profile 第 42 回日本脳卒中学会講演シンポジウム 総説 CT/MRI 1 要旨 rt-pa MRI magnetic resonance imaging DWI diffusion-weighted image Tmax time-to-maximum DWI/PWI perfusion image mismatch CT computed tomography CBF cerebral blood flow

More information

A New Trend in Vascular Imaging: the Arterial Spin Labeling (ASL) Sequence

A New Trend in Vascular Imaging: the Arterial Spin Labeling (ASL) Sequence A New Trend in Vascular Imaging: the Arterial Spin Labeling (ASL) Sequence Poster No.: C-1347 Congress: ECR 2013 Type: Educational Exhibit Authors: J. Hodel, A. GUILLONNET, M. Rodallec, S. GERBER, R. 1

More information

MR Advance Techniques. Vascular Imaging. Class II

MR Advance Techniques. Vascular Imaging. Class II MR Advance Techniques Vascular Imaging Class II 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

The Egyptian Journal of Hospital Medicine (July 2018) Vol. 72 (10), Page

The Egyptian Journal of Hospital Medicine (July 2018) Vol. 72 (10), Page The Egyptian Journal of Hospital Medicine (July 2018) Vol. 72 (10), Page 5398-5402 The Role of Susceptibility Weighted Imaging (SWI) in Evaluation of Acute Stroke Maha Abdelhamed El Nouby*, Eman Ahmed

More information

Protocol variation analysis of whole brain CT perfusion in acute ischemic stroke

Protocol variation analysis of whole brain CT perfusion in acute ischemic stroke UNLV Theses, Dissertations, Professional Papers, and Capstones 12-2010 Protocol variation analysis of whole brain CT perfusion in acute ischemic stroke Peter T. Heiberger University of Nevada, Las Vegas

More information

CTP imaging has been used in the admission evaluation of

CTP imaging has been used in the admission evaluation of Published January 12, 2012 as 10.3174/ajnr.A2809 ORIGINAL RESEARCH Shervin Kamalian Shahmir Kamalian A.A. Konstas M.B. Maas S. Payabvash S.R. Pomerantz P.W. Schaefer K.L. Furie R.G. González M.H. Lev CT

More information

Reperfusion therapy improves outcomes of patients with acute

Reperfusion therapy improves outcomes of patients with acute Published September 25, 2014 as 10.3174/ajnr.A4103 ORIGINAL RESEARCH BRAIN Combining MRI with NIHSS Thresholds to Predict Outcome in Acute Ischemic Stroke: Value for Patient Selection P.W. Schaefer, B.

More information

Basics of Quantification of ASL

Basics of Quantification of ASL Basics of Matthias van Osch Contents Thoughts about quantification Some quantification models Bloch equations Buxton model Parameters that need to be estimated Labeling efficiency M Bolus arrival time

More information

Different CT perfusion algorithms in the detection of delayed. cerebral ischemia (DCI) after aneurysmal subarachnoid hemorrhage

Different CT perfusion algorithms in the detection of delayed. cerebral ischemia (DCI) after aneurysmal subarachnoid hemorrhage Neuroradiology (2015) 57:469 474 DOI 10.1007/s00234-015-1486-8 DIAGNOSTIC NEURORADIOLOGY Different CT perfusion algorithms in the detection of delayed cerebral ischemia after aneurysmal subarachnoid hemorrhage

More information

Predicting Final Infarct Size Using Acute and Subacute Multiparametric MRI Measurements in Patients With Ischemic Stroke

Predicting Final Infarct Size Using Acute and Subacute Multiparametric MRI Measurements in Patients With Ischemic Stroke JOURNAL OF MAGNETIC RESONANCE IMAGING 21:495 502 (2005) Original Research Predicting Final Infarct Size Using Acute and Subacute Multiparametric MRI Measurements in Patients With Ischemic Stroke Mei Lu,

More information

ASL Perfusion Imaging: Concepts and Applications

ASL Perfusion Imaging: Concepts and Applications ASL Perfusion Imaging: Concepts and Applications David C. Alsop, Ph.D. Beth Israel Deaconess Medical Center and Harvard Medical School, Boston USA INTRODUCTION Arterial Spin Labeling (ASL) perfusion imaging

More information

The Value of Apparent Diffusion Coefficient Maps in Early Cerebral Ischemia

The Value of Apparent Diffusion Coefficient Maps in Early Cerebral Ischemia AJNR Am J Neuroradiol 22:1260 1267, August 2001 The Value of Apparent Diffusion Coefficient Maps in Early Cerebral Ischemia Patricia M. Desmond, Amanda C. Lovell, Andrew A. Rawlinson, Mark W. Parsons,

More information

Predicting Cerebral Ischemic Infarct Volume with Diffusion and Perfusion MR Imaging

Predicting Cerebral Ischemic Infarct Volume with Diffusion and Perfusion MR Imaging AJNR Am J Neuroradiol 23:1785 1794, November/December 2002 Predicting Cerebral Ischemic Infarct Volume with Diffusion and Perfusion MR Imaging Pamela W. Schaefer, George J. Hunter, Julian He, Leena M.

More information

Early identification of tissue at risk of infarction after

Early identification of tissue at risk of infarction after Predicting Tissue Outcome From Acute Stroke Magnetic Resonance Imaging Improving Model Performance by Optimal Sampling of Training Data Kristjana Yr Jonsdottir, PhD, MSc; Leif Østergaard, DMSc, PhD, MD,

More information

Biases affecting tumor uptake measurements in FDG-PET

Biases affecting tumor uptake measurements in FDG-PET Biases affecting tumor uptake measurements in FDG-PET M. Soret, C. Riddell, S. Hapdey, and I. Buvat Abstract-- The influence of tumor diameter, tumor-tobackground activity ratio, attenuation, spatial resolution,

More information

CT perfusion (CTP) is an important diagnostic tool for imaging

CT perfusion (CTP) is an important diagnostic tool for imaging ORIGINAL RESEARCH I. van der Schaaf E.-J. Vonken A. Waaijer B. Velthuis M. Quist T. van Osch Influence of Partial Volume on Venous Output and Arterial Input Function BACKGROUND: CT perfusion (CTP) is an

More information

Diffusion-weighted MRI (DWI) and perfusion-weighted

Diffusion-weighted MRI (DWI) and perfusion-weighted Viability Thresholds of Ischemic Penumbra of Hyperacute Stroke Defined by Perfusion-Weighted MRI and Apparent Diffusion Coefficient L. Røhl, MD; L. Østergaard, MD, MSc, PhD; C.Z. Simonsen, MD; P. Vestergaard-Poulsen,

More information

Masayuki Maeda, Tatsuya Yamamoto, Shouichiro Daimon, Hajime Sakuma, and Kan Takeda

Masayuki Maeda, Tatsuya Yamamoto, Shouichiro Daimon, Hajime Sakuma, and Kan Takeda AJNR Am J Neuroradiol :6 66, April 001 Arterial Hyperintensity on Fast Fluid-attenuated Inversion Recovery Images: A Subtle Finding for Hyperacute Stroke Undetected by Diffusion-weighted MR Imaging Masayuki

More information

Correlative Assessment of Hemodynamic Parameters Obtained with T2*-weighted Perfusion MR Imaging and SPECT in Symptomatic Carotid Artery Occlusion

Correlative Assessment of Hemodynamic Parameters Obtained with T2*-weighted Perfusion MR Imaging and SPECT in Symptomatic Carotid Artery Occlusion AJNR Am J Neuroradiol 21:1450 1456, September 2000 Correlative Assessment of Hemodynamic Parameters Obtained with T2*-weighted Perfusion MR Imaging and SPECT in Symptomatic Carotid Artery Occlusion Jae

More information

Remission of diffusion lesions in acute stroke magnetic resonance imaging

Remission of diffusion lesions in acute stroke magnetic resonance imaging ORIGINAL RESEARCH Remission of diffusion lesions in acute stroke magnetic resonance imaging F. A. Fellner 1, M. R. Vosko 2, C. M. Fellner 1, D. Flöry 1 1. AKH Linz, Institute of Radiology, Austria. 2.

More information

Predicting Infarction Within the Diffusion-Weighted Imaging Lesion Does the Mean Transit Time Have Added Value?

Predicting Infarction Within the Diffusion-Weighted Imaging Lesion Does the Mean Transit Time Have Added Value? Predicting Infarction Within the Diffusion-Weighted Imaging Lesion Does the Mean Transit Time Have Added Value? Emmanuel Carrera, MD; P. Simon Jones, MSc; Josef A. Alawneh, MD; Irene Klærke Mikkelsen,

More information

Prediction of Hemorrhage in Acute Ischemic Stroke Using Permeability MR Imaging

Prediction of Hemorrhage in Acute Ischemic Stroke Using Permeability MR Imaging AJNR Am J Neuroradiol 26:2213 2217, October 2005 Technical Note Prediction of Hemorrhage in Acute Ischemic Stroke Using Permeability MR Imaging Andrea Kassner, Timothy Roberts, Keri Taylor, Frank Silver,

More information

Quantitative Assessment of the Time Course of Infarct Signal Intensity on Diffusion-Weighted Images

Quantitative Assessment of the Time Course of Infarct Signal Intensity on Diffusion-Weighted Images AJNR Am J Neuroradiol 24:680 687, April 2003 Quantitative Assessment of the Time Course of Infarct Signal Intensity on Diffusion-Weighted Images James D. Eastwood, Stefan T. Engelter, James F. MacFall,

More information

Quantification of diffusion-weighted images (DWI) and apparent diffusion coefficient maps (ADC) in the detection of acute stroke

Quantification of diffusion-weighted images (DWI) and apparent diffusion coefficient maps (ADC) in the detection of acute stroke Quantification of diffusion-weighted images (DWI) and apparent diffusion coefficient maps (ADC) in the detection of acute stroke P. Karina Tulipano a, William S. Millar b, Celina Imielinska c, Xin Liu

More information

Magnetic Resonance Angiography

Magnetic Resonance Angiography Magnetic Resonance Angiography 1 Magnetic Resonance Angiography exploits flow enhancement of GR sequences saturation of venous flow allows arterial visualization saturation of arterial flow allows venous

More information

Benign Oligemia Despite a Malignant MRI Profile in Acute Ischemic Stroke

Benign Oligemia Despite a Malignant MRI Profile in Acute Ischemic Stroke CASE REPORT J Clin Neurol 2010;6:41-45 Print ISSN 1738-6586 / On-line ISSN 2005-5013 10.3988/jcn.2010.6.1.41 Benign Oligemia Despite a Malignant MRI Profile in Acute Ischemic Stroke Oh Young Bang, MD a

More information

Naoaki Yamada, Satoshi Imakita, and Toshiharu Sakuma

Naoaki Yamada, Satoshi Imakita, and Toshiharu Sakuma AJNR Am J Neuroradiol 20:193 198, February 1999 Value of Diffusion-Weighted Imaging and Apparent Diffusion Coefficient in Recent Cerebral Infarctions: A Correlative Study with Contrast-Enhanced T1-Weighted

More information

Invited. Methodology of Brain Perfusion Imaging. Review. Emmanuel L. Barbier, PhD, 1,2 Laurent Lamalle, PhD, 1 and Michel Décorps, PhD 1 *

Invited. Methodology of Brain Perfusion Imaging. Review. Emmanuel L. Barbier, PhD, 1,2 Laurent Lamalle, PhD, 1 and Michel Décorps, PhD 1 * JOURNAL OF MAGNETIC RESONANCE IMAGING 13:496 520 (2001) Review Invited Methodology of Brain Perfusion Imaging Emmanuel L. Barbier, PhD, 1,2 Laurent Lamalle, PhD, 1 and Michel Décorps, PhD 1 * Numerous

More information

Supplementary information Detailed Materials and Methods

Supplementary information Detailed Materials and Methods Supplementary information Detailed Materials and Methods Subjects The experiment included twelve subjects: ten sighted subjects and two blind. Five of the ten sighted subjects were expert users of a visual-to-auditory

More information

Assessment of longer post-labeling delays to measure CBF using pseudo-continuous ASL with background suppression

Assessment of longer post-labeling delays to measure CBF using pseudo-continuous ASL with background suppression Assessment of longer post-labeling delays to measure CBF using pseudo-continuous ASL with background suppression Poster No.: C-2189 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit Y.

More information

T2, T2*, ute. Yeo Ju Kim. Radiology, Inha University Hospital, Incheon, Korea

T2, T2*, ute. Yeo Ju Kim. Radiology, Inha University Hospital, Incheon, Korea SY28-1 T2, T2*, ute Yeo Ju Kim Radiology, Inha University Hospital, Incheon, Korea T2 relaxation times relate to the rate of transverse magnetization decay, caused by the loss of phase coherence induced

More information

ORIGINAL CONTRIBUTION. Diffusion-Weighted Magnetic Resonance Imaging Identifies the Clinically Relevant Small-Penetrator Infarcts

ORIGINAL CONTRIBUTION. Diffusion-Weighted Magnetic Resonance Imaging Identifies the Clinically Relevant Small-Penetrator Infarcts ORIGINAL CONTRIBUTION Diffusion-Weighted Magnetic Resonance Imaging Identifies the Clinically Relevant Small-Penetrator Infarcts Jamary Oliveira-Filho, MD; Hakan Ay, MD; Pamela W. Schaefer, MD; Ferdinando

More information

Carotid artery stenting for long CTO and pseudo occlusion of carotid artery -2 case reports-

Carotid artery stenting for long CTO and pseudo occlusion of carotid artery -2 case reports- Carotid artery stenting for long CTO and pseudo occlusion of carotid artery -2 case reports- Katsutoshi Takayama, MD, Ph.D Department of Radiology and Interventional Neuroradiology Ishinkai Yao General

More information

CT perfusion in Moyamoya disease

CT perfusion in Moyamoya disease CT perfusion in Moyamoya disease Poster No.: C-1726 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. C. Lam, C. P. Tsang, K. K. Wong, R. LEE ; HK, Hong Kong/HK Keywords: Hemodynamics / Flow dynamics,

More information

Recent studies have demonstrated that multiparametric

Recent studies have demonstrated that multiparametric ORIGINAL RESEARCH D. Kaya A. Dinçer M.E. Yıldız M.O. Çizmeli C. Erzen Acute Ischemic Infarction Defined by a Region of Multiple Hypointense Vessels on Gradient-Echo T2* MR Imaging at 3T BACKGROUND AND

More information

ORIGINAL ARTICLE. Jack A Wells 1, Bernard Siow 1,2, Mark F Lythgoe 1,4 and David L Thomas 3,4

ORIGINAL ARTICLE. Jack A Wells 1, Bernard Siow 1,2, Mark F Lythgoe 1,4 and David L Thomas 3,4 Journal of Cerebral Blood Flow & Metabolism (2013) 33, 215 224 & 2013 ISCBFM All rights reserved 0271-678X/13 $32.00 www.jcbfm.com ORIGINAL ARTICLE Measuring biexponential transverse relaxation of the

More information

Stroke is a common and serious disorder, with an annual incidence

Stroke is a common and serious disorder, with an annual incidence ORIGINAL RESEARCH BRAIN Combined Low-Dose Contrast-Enhanced MR Angiography and Perfusion for Acute Ischemic Stroke at 3T: A More Efficient Stroke Protocol K. Nael, A. Meshksar, B. Ellingson, M. Pirastehfar,

More information

MINERVA MEDICA COPYRIGHT

MINERVA MEDICA COPYRIGHT EURA MEDICOPHYS 2007;43:271-84 Diffusion and perfusion-weighted magnetic resonance imaging techniques in stroke recovery Most of the functional recovery after stroke takes place during the first three

More information

SVD+ Dynamic Volume CT:

SVD+ Dynamic Volume CT: SVD+ Dynamic Volume CT: Delay Insensitive Brain Perfusion Analysis Erin Angel, PhD Manager, CT Clinical Science Toshiba America Medical Systems, Inc. SVD+ Dynamic Volume CT: Delay Insensitive Brain Perfusion

More information

Advanced Neuroimaging for Acute Stroke

Advanced Neuroimaging for Acute Stroke Advanced Neuroimaging for Acute Stroke E. Bradshaw Bunney, MD, FACEP Professor Department Of Emergency Medicine University of Illinois at Chicago Swedish American Belvidere Hospital Disclosures FERNE Board

More information

Comparison of 10 Perfusion MRI Parameters in 97 Sub-6-Hour Stroke Patients Using Voxel-Based Receiver Operating Characteristics Analysis

Comparison of 10 Perfusion MRI Parameters in 97 Sub-6-Hour Stroke Patients Using Voxel-Based Receiver Operating Characteristics Analysis Comparison of 10 Perfusion MRI Parameters in 97 Sub-6-Hour Stroke Patients Using Voxel-Based Receiver Operating Characteristics Analysis Søren Christensen, MSc; Kim Mouridsen, PhD; Ona Wu, PhD; Niels Hjort,

More information

Method Comparison for Interrater Reliability of an Image Processing Technique in Epilepsy Subjects

Method Comparison for Interrater Reliability of an Image Processing Technique in Epilepsy Subjects 22nd International Congress on Modelling and Simulation, Hobart, Tasmania, Australia, 3 to 8 December 2017 mssanz.org.au/modsim2017 Method Comparison for Interrater Reliability of an Image Processing Technique

More information

Acute Ischemic Stroke Imaging. Ronald L. Wolf, MD, PhD Associate Professor of Radiology

Acute Ischemic Stroke Imaging. Ronald L. Wolf, MD, PhD Associate Professor of Radiology Acute Ischemic Stroke Imaging Ronald L. Wolf, MD, PhD Associate Professor of Radiology Title of First Slide of Substance An Illustrative Case 2 Disclosures No financial disclosures Off-label uses of some

More information

The Low Sensitivity of Fluid-Attenuated Inversion-Recovery MR in the Detection of Multiple Sclerosis of the Spinal Cord

The Low Sensitivity of Fluid-Attenuated Inversion-Recovery MR in the Detection of Multiple Sclerosis of the Spinal Cord The Low Sensitivity of Fluid-Attenuated Inversion-Recovery MR in the Detection of Multiple Sclerosis of the Spinal Cord Mark D. Keiper, Robert I. Grossman, John C. Brunson, and Mitchell D. Schnall PURPOSE:

More information

Myocardial Blood Flow from Dynamic PET using Independent Component Analysis

Myocardial Blood Flow from Dynamic PET using Independent Component Analysis 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSSIMIC) M07-4 Myocardial Blood Flow from Dynamic PET using Independent Component Analysis Alexander Karpikov, Hemant Tagare,

More information

Background. Recommendations for Imaging of Acute Ischemic Stroke: A Scientific Statement From the American Heart Association

Background. Recommendations for Imaging of Acute Ischemic Stroke: A Scientific Statement From the American Heart Association for Imaging of Acute Ischemic Stroke: A Scientific Statement From the American Heart Association An Scientific Statement from the Stroke Council, American Heart Association and American Stroke Association

More information