1 I S S U E Inside this issue: Special Focus: Optimal testing, technologies to detect glaucoma damage and progression ` What s New: Assessing function in Glaucoma Clinical Issues: Optimizing Your Approach to Visual Field Testing Practical Tips: Minimizing artifacts and avoiding pitfalls in visual field interpretation PAGE 2 PAGE 7 PAGE 10 PAGE 13 LEARNING OBJECTIVES Special Focus A comprehensive overview on optimal testing, taking into account traditional and recent technologies to detect glaucoma damage and progression such as changes in optic disc and SAP, use of OCT, electrophysiology PERG/VEP and others. What s New A review of where are we going with the newest approaches such as OCT, PERG, VEP in the light of the newest clinical data. Clinical Issues Insight into an optimal approach to testing: What tests and strategies are to be used for whom and when? Practical Tips A practical approach to use of current test methods: Minimizing artifacts and avoiding pitfalls in interpretation. TARGET AUDIENCE Main topic: ASSESSING FUNCTION IN GLAUCOMA Glaucoma Now is a continuing medical education publication hosted by Glaucoma Australia Inc. Distributed worldwide, our goal is to educate and update general ophthalmologists, glaucoma specialists and ophthalmology residents. International leaders in the field of glaucoma are invited to contribute to the program, sharing the most recent insights. Supported by an unrestricted educational grant, the publication is non-promotional and has a fully independent Editorial Board. For each issue CME credits can be obtained by registering on our website and answering the uploaded questions. A newsletter is sent out to participants registered to the program. This educational program is aimed at general ophthalmologists, ophthalmology residents and glaucoma specialists. EDITORIAL BOARD Ivan Goldberg MBBS, FRANZCO, FRACS, Head of Glaucoma Unit, Sydney Eye Hospital & Clinical Professor, Discipline of Ophthalmology, University of Sydney, Australia. Remo Susanna MD, Professor and Head of Department of Ophthalmology, University of São Paulo, Brazil. Glaucoma Now is published and administrated by the editorial board and funded by Allergan International, Mundipharma PTY Limited, Novartis Pharmaceuticals Australia P/L and Pfizer Inc. Copyright 2010 Editorial Board. All rights reserved. No responsibility assumed for injury or damage to persons or propertyperty arising from the use of information or ideas contained in this publication. Editoral Manager of Glaucoma Now Program hosted by Glaucoma Australia: Patricia Buchholz RPh, PhD Karlsruhe, Germany Medical Responsible for CME accreditation: Andreas Buchholz MD, PhD, ROph Karlsruhe, Germany Production by Momentum Design / Caroline Herzog, Hamburg, Germany 1

2 Special Focus: Optimal testing, technologies to detect glaucoma damage and progression Linda Zangwill MD, Christopher Bowd, PhD Hamilton Glaucoma Center and the Shiley Eye Institute, University of California San Diego, California USA Core Concepts There are 3 global indices of visual field sensitivity: mean deviation (MD), pattern standard deviation (PSD) and Visual Field Index (VFI). Once glaucoma has been diagnosed MD, PSD and VFI provide data on disease severity. To be useful, tests must be reliable with few fixation losses and few false positive/ false negative responses. Data on disease severity with measured visual field deterioration, along with patient age and health status should be used to estimate likelihood of visual impairment. Progression detection is facilitated using the Guided Progression Analysis (GPA) and Visual Field Index (VFI). Frequency of follow-up testing should be dictated by the relative risk of progression. The higher the risk, the more frequently testing should take place. The World Glaucoma Association Consensus Group on Glaucomatous Progression suggests at least 2 reliable fields in the first 6 months after diagnosis and at least 2 more fields within the next 18 months. Increased testing from baseline up to 2 years identifies more rapidly progressing patients with a lower false positive rate. It is important to evaluate the central visual field to obtain 10-2 visual fields that use a 2-degree spacing as it may be compromised even in early glaucoma. 1) Technologies to Detect Glaucomatous Visual Field Damage There are several ways to identify glaucomatous defects from standard automated perimetry (SAP) print outs (Figure 1). Firstly, inspect sensitivity Figure. 1 - shows all of the information available on a standard automated perimetry print out. (Figure 1). Firstly, inspect sensitivity thresholds at each test point. Sensitivity thresholds are not particularly useful to identify defects, because these values are not age-corrected. In older patients, low thresholds could be attributable entirely to age. More meaningful information is available in the total deviation (TD) and pattern deviation (PD) probability plots (Figure 2). The TD probability plot shows the pointby-point probability of abnormality (in grey scale) compared with an age corrected normative database. The PD probability plot provides similar information adjusted for both age and generalized depression that could be caused by other pathologies (e.g. cataract). So in the PD plot, localized defects most likely from glaucoma are highlighted. Localized patterns of abnormality, identified on the PD probability plot, can be assessed to determine if they are what is considered characteristic for glaucoma (e.g. nasal step, superior or inferior arcuate patterns of defect). Odd-looking patterns might result from a testtaking artifact or patient inattention. 2 Glaucoma Now - Issue No 2,

3 Three global indices of visual field sensitivity are available. Mean deviation (MD) is the weighted average of the point-by-point decibel deviations from normal that are shown in the TD plot. Pattern standard deviation (PSD) represents the relative point-bypoint decibel deviations from normal across the visual field. PSD increases as the point-by-point deviations from normal across the visual field become more variable. PSD increases in the presence of localized defects. The Visual Field Index (VFI) is another summary parameter that is a modification of MD designed to be less affected by cataract and more sensitive to changes near the center of the visual field. The VFI is based on geographically weighted PD probability map values; it has been reported to represent the percentage of remaining useful vision. 1 Theoretically the VFI ranges from 100 in a healthy eye to zero in a seriously damaged eye. There is no suggested abnormality cut-off for this parameter. Although these global indices are not used for diagnosis, a PSD outside normal limits is likely to likely to detect local defects earlier. Once glaucoma has been diagnosed, MD, PSD and VFI provide a general idea of the severity of the disease although in advanced glaucoma PSD becomes less useful as it reverts to more normal values due to a generalized depression (i.e. relative reduction of local defects). Secondly, observed abnormalities should be repeatable and consecutive abnormal tests should have the same general patterns of defect. In the Ocular Hypertension Treatment Study (OHTS), 86% of 702 initial abnormal SAP tests did not show a subsequent abnormal test. In fact, 66% of these eyes had all indices within normal limits on follow-up testing 3, thus the importance of repeatable abnormal results to identify true glaucomatous defects. Finally, visual field results are quite variable and learning effects, artifact caused by drooping eyelids, small pupils and by inattention or trigger happiness can result in different patterns of defects that are not disease related: hence the need to check for similar and recognizable patterns of defect in consecutive tests (Figure 3). 2) Technologies to Detect Glaucomatous Progression Glaucoma management aims to preserve visual function for an individual s lifetime. To this end, clinicians need to monitor closelythe magnitude and rate of visual fields changes so that appropriate treatment can be initiated to preserve visual function. Information on the rate of visual field deterioration along with the severity of the disease at diagnosis and the patient s age and general health status (i.e. likely longevity) should be used to estimate the likelihood of visual impairment. 4 Similar to detection of disease using SAP, disease progression also can be assessed subjectively based on information available in TD and PD probability plots.to do this, a longitudinal series of TD or PD plots can be compared for increases in scotoma size and/or depth or the development of new scotomas. Because of the large variability in SAP results, this information is not always meaningful. Two useful progression detection algorithms currently are available in the Humphrey Statpac software: Guided Progression Analysis (GPA) 5 and linear regression of Visual Field Index (VFI). 1 would be damaging to the optic nerve. Guided Progression Analysis (GPA) GPA is a PD plot-based event analysis calculation that relies on the comparison of change at each test location to the variability observed between two baseline tests at the same location (Figure 4). According to the Early Manifest Glaucoma Trial (EMGT) criteria 5, Likely Progression is assigned to a visual field if at least three test points are flagged as significantly progressing The Glaucoma Hemifield Test (GHT) 2 is a summary parameter that describes the relative difference in sensitivity in five zones of the superior hemifield compared with five corresponding zones in the inferior hemifield. Results are either outside normal limits, borderline or within normal limits. This parameter is particularly useful to detect early to moderate disease because glaucomatous visual field defects usually manifest asymmetrically across the visual field horizontal midline. Consider other important items to identify glaucomatous defects. Firstly, tests must be reliable, with few fixation losses and few false positive and false negative responses. This last reliability criterion is less important in individuals with advanced disease because a reported false negative may be a truly unseen test point that is the result of a severe defect. Figure 2 - shows the Total Deviation probability plots from an eye with advanced glaucoma (MD = db, VFI = 34) using the 24-2 scan pattern (top) and the 10-2 scan pattern (bottom). Glaucoma Now - Issue No 2,

4 Figure 3 shows the Total Deviation and Pattern Deviation probability plots from an eye with a small pupil. This pattern of defect generally is not characteristic of glaucoma the 2011 World Glaucoma Association Consensus Group statements, 7 frequency of follow-up testing should be dictated by the relative risk of significant progression (based on extent of damage, life expectancy and any other relevant clinical observations, such as structural change and IOP control 14 ), with frequency increasing with increasing risk. Management needs to be individualized appropriately. 3) Optimal Testing at the same location over three consecutive tests ( Possible Progression is assigned if the same three points have progressed over two consecutive tests). GPA is attractive because it is objective, but it relies on the variability between baseline tests. Because of this, its sensitivity to change can be variable depending on baseline variability. Also, it does not require the progressing points to be contiguous, so the observed change may not represent the predicted increase in in scotoma size that is associatedwith disease progression. 6 Once a GPA result of Likely Progression has been assigned, new baseline examinations should be obtained to allow identification of a subsequent progression event. As suggested by the 2011 World Glaucoma Association Consensus Group, 7 new baseline examinations also should be obtained after any significant therapeutic intervention. Visual Field Index (VFI) Unlike the GPA, linear regression of the VFI can provide the rate of change over time (i.e. change in VFI value per year) and, because it is derived from PD and not MD, it likely is fairly resistant to the effects of developing/worsening cataracts. Knowing the rate of progression is more important for long-term disease management than simply knowing if a patient has progressed or not (Figure 5). This is in part because faster progression is a predictor for future progression and increases the likelihood of visual impairment in one s lifetime. 8 Like GPA, VFI is not perfect. For instance, using linear regression to suggest a predicted rate of change over time assumes that change over time is linear; this may not be true. 9 It also assumes a stable treatment regimen and it does not consider intervening surgical procedures. Further, assuming change over time is linear, estimates of change of VFI over time (i.e. estimates of the negative slope of VFI) will be more accurate when more tests are included in the regression equation. This is because adding more test points will result in a better ( tighter ) linear fit. Finally, linear regression of VFI is not a good indicator of progression in early disease because of a ceiling effect in early disease (VFI remains high as MD decreases) and in advanced disease (decrease in VFI can be highly variable after approximately -15 db MD). 10,11 Progressor Another available progression detection technique uses ordinary least squares linear regression and is implemented using Progressor software. 12 Based on five or more consecutive tests, Progressor software performs linear regression (like regression of VFI) of threshold sensitivity on a point-by-point basis using all test points in the 24-2 test pattern. Individual points are defined as progressed if the negative slope of the point-wise change in sensitivity meets or exceeds a user-selected cut-off. A global progression event can be assigned (similar to GPA) based on a user-defined global progression criterion (e.g. 3 of the same test points with a slope 1.0 db/year for inner points and 2.0 db for edge points, confirmed in three consecutive exams 13 ). No matter the technique used for progression detection, as suggested by Several critical clinical questions must be answered to determine how best to monitor closely visual function in glaucoma patients, using any or all of the techniques described above. How many visual fields are really needed? How often should visual field tests be obtained? Given the variability of visual field results, a sufficient number of visual field tests must be obtained in order confidently to identify changes that are greater than the variability in the measurements. More visual fields offer a better estimate of the rate of visual field progression. But how many visual field tests are needed over what period of time? According to Chauhan et al, 15 to detect relatively slow progression (-0.50 db/ year) in 2 years, 7 visual fields are needed each year; clearly more than is practical. On the other hand, to detect fast progressing glaucoma (-2.0 db/year), Chauhan et al 15 recommend that newly diagnosed patients complete visual field testing three times per year during the first two years after glaucoma diagnosis. 15 These recommendations were adopted by The European Glaucoma Society in The 2011 World Glaucoma Association Consensus Group 7 on Glaucomatous Progression developed similar guidelines on the optimal number of visual field tests required with newly diagnosed glaucoma patients. Specifically, during the first two years after a diagnosis of glaucoma at least two reliable visual fields is optimal in the first six months and at least two further visual fields should be performed within the next 18 months. Where the lifetime risk of visual disability is high, such as those who already 4 Glaucoma Now - Issue No 2,

5 have advanced damage, three baseline visual fields may be necessary. Wu and colleagues 17 recently showed that the time required to detect a significant decrease in MD slope decreased as the frequency of testing increased, although not proportionally. Based on their results, they suggested that obtaining two reliable visual fields at baseline followed by semi-annual testing, with confirmation of progression during follow-up, is a reasonable compromise for detecting glaucomatous progression while minimizing the burden on patients and clinicians associated with more frequent testing. After the initial two years following adiagnosis of glaucoma, the frequency of visual field testing should be based on the risk of visual impairment during the patient s lifetime. The 2011 World Glaucoma Association Consensus Group 7 recommends that, In low and moderate risk patients, subsequent visual field frequency should be one visual field per year and, as a rule, repeated sooner if other clinical observations are suggestive of possible progression or increased risk of progression. However, In high risk patients, subsequent visual field frequency should be two visual fields per year and repeated sooner if possible progression is identified on the basis of an event analysis, or if other clinical observations are suggestive of progression or increased risk of progression. Computer simulation has been used to determine the optimal spacing for the intervals between testing for detectionof rapidly progressing glaucoma (rate of loss: -2.0 db/year). Crabb and Garway-Heath 18 suggest that compared with evenly-spaced follow-up every six months, visual field testing 2 or 3 times at baseline and at the end of a 2 year period identifies more patients with rapidly progressing visual fields with a lower false positive rate. By aggregating the visual field testing earlier and later in the follow-up period, there is increased efficiency to reduce the false positive rate, and to estimate the rate of visual field loss. This wait and see approach, developed using moderate variability of the visual fields in the computer simulation to detect fast progression, is being tested in longitudinal studies. The number of fields and intervals between them can also influence the methods used to measure the rate of change. For example, Medeiros et al. 19 reported that a Bayesian-based slope estimate technique was more predictive of future impairment than ordinary least square regression, particularly in eyes with moderate to fast rates of change. Other statistical or machine learning based progression detection methods also have been proposed and compared to ordinary least square regression of VFI or MD with varying Figure 4 - shows the first incidence of possible and likely progression, based on GPA analysis, in an eye with a developing inferior temporal visual field defect. Half-filled triangles represent test points progressed, relative to baseline variability, in two consecutive tests (possible progression). Filled triangles represent progression in three consecutive tests (likely progression). Glaucoma Now - Issue No 2,

6 degrees of success but none have been incorporated into HFA software Most patients can be monitored using the 6-degree by 6-degree grid spacing of 24-2 visual field test pattern. However, when central vision is compromised or threatened, it may be useful to obtain 10-2 visual fields that use 2-degree by 2-degree spacing to document better how glaucoma is affecting central vision. 25,26 Several studies have reported evidence of visual field damage in the central visual field using 10-2 test pattern that is not detected in the peripheral field using 24-2 testing even in very early stages of the disease. Zhang et al. 25 documented that the 10-2 test pattern results are useful to make clinical management decisions; in about one third of visual fields considered to threaten fixation based on 30-2 or 24-2 test patters, the threat, based on 10-2 testing was not imminent. In some patients, 10-2 testing can be important to determine the level of visual impairment expected from glaucomatous visual field loss in the central and peripheral visual field. 4) Recent Advances Very recently two new test features have been introduced and will be incorporated into the Humphrey Visual Field Analyzer 3 (personal communication, Carl Zeiss Meditec Inc.). The first is a newfaster 24-2 test pattern called SITA Faster Results indicate that this algorithm requires a considerably shorter test time (half the time of SITA Standard 24-2) while providing very similar results compared with both SITA Fast 24-2 and SITA Standard 24-2 algorithms. A new combined 24-2 and 10-2 test pattern also has been introduced. This pattern, the SITA Faster 24-2C, is composed of 10 points from the 10-2 pattern tested at the end of the SITA Fast Results indicate that this test pattern provides an improved sensitivity to detect visual field loss in the central 10 degrees over the SITA Fast 24-2 pattern, which tests only 4 points within this region. 31 References: 1 Bengtsson B, Heijl A. A visual field index for calculation of glaucoma rate of progression. Am J Ophthalmol. 2008;145(2): Asman P, Heijl A. Glaucoma Hemifield Test. Automated visual field evaluation. Arch Ophthalmol. 1992;110(6): Keltner JL, Johnson CA, Cello KE, et al. Classification of visual field abnormalities in the ocular hypertension treatment study. Arch Ophthalmol. 2003;121(5): Saunders LJ, Medeiros FA, Weinreb RN, Zangwill LM. What rates of glaucoma progression are clinically significant? Expert Rev Ophthalmo. 2016;11(3): Leske MC, Heijl A, Hyman L, Bengtsson B. Early Manifest Glaucoma Trial: design and baseline data. Ophthalmology. 1999;106(11): Boden C, Blumenthal EZ, Pascual J, et al. Patterns of glaucomatous visual field progression identified by three progression criteria. Am J Ophthalmol. 2004;138(6): World Glaucoma Association Consensus Series 8. Progression of Glaucoma. Amsterdam: Kugler; Nouri-Mahdavi K, Hoffman D, Gaasterland D, Caprioli J. Prediction of visual field progression in glaucoma. Invest Ophthalmol Vis Sci. 2004;45(12): Caprioli J, Mock D, Bitrian E, et al. A method to measure and predict rates of regional visual field decay in glaucoma. Invest Ophthalmol Vis Sci. 2011;52(7): Rao HL, Senthil S, Choudhari NS, Mandal AK, Garudadri CS. Behavior of visual field index in advanced glaucoma. Invest Ophthalmol Vis Sci. 2013;54(1): Artes PH, O Leary N, Hutchison DM, et al. Properties of the statpac visual field index. Invest Ophthalmol Vis Sci. 2011;52(7): Fitzke FW, Hitchings RA, Poinoosawmy D, McNaught AI, Crabb DP. Analysis of visual field progression in glaucoma. Br J Ophthalmol. 1996;80(1): Krup in T, Liebmann JM, Greenfield DS, Ritch R, Gardiner S, Group L-PGS. A randomized trial of Brimonidine versus timolol in preserving visual function: Results from the Low-pressure Glaucoma Treatment Study. American Journal of Ophthalmology. 2011;151(4): De Moraes CG, Juthani VJ, Liebmann JM, et al. Risk factors for visual field progression in treated glaucoma. Arch Ophthalmol. 2011;129(5): Chauhan BC, Garway-Heath DF, Goni FJ, et al. Practical recommendations for measuring rates of visual field change in glaucoma. Br J Ophthalmol. 2008;92(4): Society EG. Terminology and Guidelines for Glaucoma. Savona, Italy Wu ZC, Saunders LJ, Daga FB, Diniz A, Medeiros FA. Frequency of Testing to Detect Visual Field Progression Derived Using a Longitudinal Cohort of Glaucoma Patients. Ophthalmology. 2017;124(6): Crabb DP, Garway-Heath DF. Intervals between visual field tests when monitoring the glaucomatous patient: Wait and See. Invest Ophthalmol Vis Sci Medeiros FA, Zangwill LM, Weinreb RN. Improved prediction of rates of visual field loss in glaucoma using empirical bayes estimates of slopes of change. J Glaucoma. 2012;21(3): Goldbaum MH, Lee I, Jang G, et al. Progression of patterns (POP): A machine classifier algorithm to identify glaucoma progression in visual fields. Invest Ophthalmol Vis Sci. 2012;53(10): O Leary N, Chauhan BC, Artes PH. Visual field progression in glaucoma: estimating the overall significance of deterioration with permutation analyses of pointwise linear regression (PoPLR). Invest Ophthalmol Vis Sci. 2012;53(11): Yousefi S, Balasubramanian M, Goldbaum MH, et al. Unsupervised Gaussian Mixture-Model With Expectation Maximization for Detecting Glaucomatous Progression in Standard Automated Perimetry Visual Fields. Transl Vis Sci Technol. 2016;5(3):2. 23 Zhu H, Crabb DP, Ho T, Garway-Heath DF. More Accurate Modeling of Visual Field Progression in Glaucoma: ANSWERS. Invest Ophthalmol Vis Sci. 2015;56(10): Zhu H, Russell RA, Saunders LJ, Ceccon S, Garway-Heath DF, Crabb DP. Detecting changes in retinal function: Analysis with non-stationary Weibull error regression and spatial enhancement (ANSWERS). PLoS ONE. 2014;9(1). 25 Zhang L, Drance SM, Douglas GR. Automated perimetry in detecting threats to fixation. Ophthalmology. 1997;104(11): Hood DC, Raza AS, de- Moraes CG, et al. Initial arcuate defects within the central 10 degrees in glaucoma. Invest Ophthalmol Vis Sci. 2011;52(2): Traynis I, De Moraes CG, Raza AS, Liebmann JM, Ritch R, Hood DC. Prevalence and nature of early glaucomatous defects in the central 10 degrees of the visual field. JAMA ophthalmology. 2014;132(3): De Moraes CG, Hood DC, Thenappan A, et al Visual Fields Miss Central Defects Shown on 10-2 Tests in Glaucoma Suspects, Ocular Hypertensives, and Early Glaucoma. Ophthalmology. 2017;124(10): Patella V, Bengtsson B, Lee G, et al. Clincal evaluation of a new perimetry testing algorithm, SITA Faster. Invest Ophthalmol Vis Sci. 2018;59:E-Abstract Monhart M, Lee G, Iwase A, Flanagan J. Selecting addtional test locations to enhance the 24-2 pattern using a scoring system. World Glaucoma Congress 2017; Helsinki, FIN. Figure 5 - shows the VFI regression analysis results from the same eye shown in Figure Callan T, Yu S, Lee G, Covita A, Severin T. Evaluation of the SITA Faster 24-2C vsual field test. Invest Ophthalmol Vis Sci. 2018;59:E-Abstract Glaucoma Now - Issue No 2,

7 ` What s new: Assessing function in Glaucoma Stuart Gardiner MD, PhD Devers Eye Institute, Legacy Research Institute, Portland, Oregon, USA Core Concepts Standard automated perimetry has been the clinical standard of care in glaucoma for decades. It is limited by excessively high variability, a limited dynamic range hampering both detection of early disease and the ability to monitor severe glaucoma, and poor patient acceptability. Objective functional testing can outperform perimetry in certain cases, and is useful when other results are ambiguous. However no such techniques have yet been shown to consistently outperform perimetry, and so for now they remain supplementary rather than primary diagnostic tools. New test stimuli are being developed that aim to improve the signal-to-noise ratio. In particular, flickering stimuli and alternative stimulus sizes show promise. New testing algorithms aim to reduce test duration, and/or increase testing of the central region of the visual field where defects are commonly missed. Portable visual field testing instruments are under development, in particular using tablet computers or using head-mounted screens. These would allow more frequent testing to be conducted, since they can be used outside clinical settings. 1) Why do we need something new? Clinicians and, often, patients are very familiar with standard automated perimetry (SAP), which has been the primary clinical standard of care for functional testing in glaucoma for many years. Even the testing algorithms that are most commonly used are now 20 years old. Why then, in the face of such stable technology, are advances in functional testing still needed? The answer, in part, is because SAP unfortunately continues to suffer from three major problems: variability, dynamic range, and patient acceptability. Variability Sensitivities measured using SAP can vary markedly from one day to the next. Variability also worsens with glaucoma. When a location is significantly damaged, the 95% interval for retesting that location covers over half of the presumed dynamic range of the instrument (see Figure 1). Indeed below db, sensitivities from SAP are no longer even significantly correlated with the true sensitivity. 1 Dynamic Range Due to this extremely high variability at low sensitivities, SAP cannot be relied upon to assess progression once sensitivity at a location has decreased below db. 1 Unfortunately, common structural measurements such as retinal nerve fiber layer thickness also reach their measurement floor at a similar level of damage. Therefore the unsatisfying conclusion is that for eyes with severe damage, the only way to assess progression at the moment is to look at the remaining areas of the visual field that have not yet reached this level of damage. At the other end of the scale, SAP is also suboptimal. Clear structural defects are often detected prior to functional loss (although functional loss is detected first in almost as many cases, with any time lag being principally the result of SAP having slightly higher test-retest variability). In particular, defects near fixation are often missed, despite their importance to the patient s visual performance and daily quality of life. 2 Patient Acceptability Patients commonly report disliking SAP, preferring structural imaging tests. 3 Figure 1 - shows how pointwise sensitivity varied from one test (x-axis) to the next (y-axis) in a longitudinal dataset. Boxes show the interquartile range, with whiskers showing 95% prediction interval. Historically, test duration and accompanying fatigue were seen as leading causes of this dissatisfaction. The most commonly-used testing algorithms now take around 5 minutes per eye, although with the just-released SITA- Faster program, it takes half this time. It takes an additional minutes to get the patient positioned and their eye sufficiently adapted to the background luminance to obtain reliable results, meaning that further reductions in test duration would have relatively little impact on the overall visit duration. Perhaps more importantly then, patients report frustration at deciding whether they actually see a stimulus or not; this is unsurprising when sensitivity is usually defined as the stimulus that the patient will respond to on 50% of presentations and not on the remaining 50%, and applies to all tests measuring threshold. 2) How can it be improved? Given these substantial problems with current testing, several avenues are being explored to improve functional Glaucoma Now - Issue No 2,

8 assessment. These can broadly be split into a four, often overlapping, areas of research. New Testing Paradigms SAP relies on the test subject responding to a visual stimulus by pressing a button. This can cause problems, especially in elderly patients who may have difficulty depressing the button sufficiently and reliably. Therefore objective functional testing is a natural avenue of research. For multifocal visual evoked potential (mfvep), the patient views a reversing checkerboard pattern while electrodes record the resultant activity in the visual cortex. Unfortunately the test is technically demanding to administer, and so to date it has remained a specialist test used in cases where other test results are not definitive, and/or to rule out non-visual sources of defects. 4 Multifocal pupillographic objective perimetry (mfpop) records pupillary changes in response to visual stimuli and appears to have similar performance to SAP, 5 and so may have promise as an objective measure of function, but this has not yet reached the point of clinical adoption. The pattern electroretinogram (PERG) records retinal ganglion cell function in response to stimuli. Since there are far more retinal ganglion cells centrally, PERG may perform better than SAP at detecting central functional loss, but less well in eyes with peripheral loss. 6 While each of these techniques can outperform SAP in certain cases, none can yet do so consistently in all cases. As such, they remain a supplementary tool in the clinician s arsenal, rather than something that could replace SAP in the foreseeable future. New Test Stimuli The majority of functional testing at the moment uses a static Goldmann Size III stimulus, with diameter 0.43º. This represents more of a historic accident than anything else. Researchers are examining the potential Figure. 2 - shows four of the different stimuli that have been used for perimetry. benefits of using larger Size V stimuli (diameter 1.72º), which is known to increase sensitivity and hence decrease variability, and so allow reliable measurements later into the disease process. 7 However, there is ongoing debate about whether this hampers the ability to detect early defects. 8 As it may be preferable for stimulus size to increase with eccentricity, 9 and/or with damage, 10 research in this area continues. Meanwhile, alternative stimulus types may provide better signal-to-noise ratios than the static circular stimuli currently used in most commercial perimeters for SAP. Previously, researchers tried using blueon-yellow stimuli (Short Wavelength Automated Perimetry, SWAP) to target certain subtypes of ganglion cell, but this increased variability so much that it negated any benefits for earlier detection of damage. 11 It was also adversely affected by nuclear sclerotic cataracts. More recent work has focused instead on optimizing stimuli for detection by wider ranges of ganglion cell types. Gabor patches, sinusoidally-modulated stimuli presented in a Gaussian envelope to blur the edges, may have lower variability and consequently better signal-to-noise ratio. 12 However, these cannot be presented using projectionbased perimeters like the Octopus or the Humphrey Field Analyzer (HFA). The most similar stimulus to have been implemented commercially is the Frequency- Doubling Technology (FDT) stimulus, which is a flickering sinusoidal stimulus, but with hard edges; while this is not optimal, it has still proven useful in particular for screening tests. 13 New Testing Algorithms An algorithm that aims to measure sensitivity at around 50 visual field locations in just 5 minutes can only afford to present 4-6 stimuli at each location. Within that limit, it has to determine sensitivity on a scale of 0-40 db, ideally to within 1-2 db. This severely limits the amount of information that can be obtained, and is probably the biggest cause of the high variability discussed above. Current testing algorithms actually perform impressively given this constraint, but various techniques have been proposed to further improve the reliability and accuracy of the results. Information from neighboring locations can be used to reduce variability, 14 although if taken too far this risks losing spatial information about small defects. Information from structural testing can also be used to inform the choice of stimulus contrasts to present. 15 Another approach is to stop testing locations that have progressed to the point where reliable sensitivities can no longer be measured, and instead spend the time on more accurate measurements elsewhere in the visual field. 16 Efficiency savings from new algorithms can be used to improve spatial information about defects. For example, targeting stimuli around the edge of defects has been shown to improve the ability to monitor disease progression. 17 The closest of these approaches to reaching clinical care is the Swedish Interactive Threshold Algorithm (SITA) Faster 24-2C test for the HFA3, which reduces redundancies in the SITA Fast algorithm and spends those gains not 8 Glaucoma Now - Issue No 2,

9 just on further reducing test time (which as mentioned above is of limited benefit at this point), but on adding testing at additional central locations in order to detect early defects that can be missed using the current 24-2 test grid. 2 This algorithm was presented by Carl- Zeiss Meditec at the 2018 ARVO meeting (Callan et al, E-Abstract #5111), and is becoming available as this article goes to print. New Test Equipment Current functional tests are performed using instruments that take up several cubic feet of space, without even including room for the test subject and operator. More portable options would allow for much easier testing. In particular, this would facilitate screening programs outside standard clinical settings; and/ or allow patients to test themselves at home, whereby any loss of accuracy caused by non-standardization of test conditions would be compensated by the ability to test patients far more frequently. 18 Two main technologies are currently under development that may address this issue. Perimetry could be performed on a tablet computer, subject to two main constraints. The stimulus intensity needs to remain constant on repeated testing, which is not feasible with many commercial tablets; and/or it needs to be automatically calibrated prior to each test. The Melbourne Rapid Fields (MRF) app for ipad has shown promise in this regard, and early results suggest similar performance to traditional perimeters. 19 An alternative portable approach is taken by the imo head mounted perimeter. 20 This requires more specialist equipment than just a tablet computer; but results in improved standardization of test conditions and hence of results. It remains to be seen which of these approaches will prove beneficial in more circumstances. 4) Summary While the gold standard for functional testing has not changed for many years, new techniques are in development that address some or all of the problems with SAP. Some, such as the upcoming SITA Faster 24-2C test, aim to refine SAP, and so have clear clinical utility. Other techniques such as objective testing are currently best seen as supplementary tests for use when SAP results are equivocal or unreliable, since none has yet proven to be sufficiently superior to SAP to justify switching completely. References: 1. Gardiner SK, Swanson WH, Goren D, Mansberger SL, Demirel S. Assessment of the reliability of standard automated perimetry in regions of glaucomatous damage. Ophthalmology. 2014; 121 (7): PMCID: PMC Hood DC, Raza AS, De Moraes CG, Liebmann JM, Ritch R. Glaucomatous damage of the macula. Prog Retin Eye Res. 2013; PMCID: PMC Gardiner SK, Demirel S. Assessment of patient opinions of different clinical tests used in the management of glaucoma. Ophthalmology. 2008; 115 (12): PMCID: PMC De Moraes CG, Liebmann JM, Ritch R, Hood DC. Clinical use of multifocal visualevoked potentials in a glaucoma practice: A prospective study. Documenta ophthalmologica Advances in ophthalmology. 2012; 125 (1): Carle CF, James AC, Kolic M, Loh YW, Maddess T. High-resolution multifocal pupillographic objective perimetry in glaucoma. Invest Ophthalmol Vis Sci. 2011; 52 (1): Ventura LM, Porciatti V. Pattern electroretinogram in glaucoma. Current opinion in ophthalmology. 2006; 17 (2): Gardiner SK, Demirel S, Goren D, Mansberger SL, Swanson WH. The effect of stimulus size on the reliable stimulus range of perimetry. Transl Vis Sci Technol. 2015; 4 (2): 10. PMCID: PMC Wall M, Doyle CK, Eden T, Zamba KD, Johnson CA. Size threshold perimetry performs as well as conventional automated perimetry with stimulus sizes iii, v, and vi for glaucomatous loss. Invest Ophthalmol Vis Sci. 2013; 54 (6): Phu J, Khuu ssk, Zangerl B, Kalloniatis M. A comparison of goldmann III, V and spatially equated test stimuli in visual field testing: The importance of complete and partial spatial summation. Ophthalmic & Physiological Optics. 2017; 37 (2): Rountree L, Mulholland PJ, Anderson RS, Garway-Heath DF, Morgan JE, Redmond T. Optimising the glaucoma signal/ noise ratio by mapping changes in spatial summation with area-modulated perimetric stimuli. Scientific reports. 2018; 8: Blumenthal E, Sample P, Berry C, Lee A, Girkin C, Zangwill L, Caprioli J, Weinreb R. Evaluating several sources of variability for standard and swap visual fields in glaucoma patients, suspects, and normals. Ophthalmology. 2003; 110 (10): Swanson WH, Malinovsky VE, Dul MW, Malik R, Torbit JK, Sutton BM, Horner DG. Contrast sensitivity perimetry and clinical measures of glaucomatous damage. Optometry and Vision Science. 2014; 91 (11): Cioffi GA, Mansberger S, Spry P, Johnson C, Van Buskirk EM. Frequency doubling perimetry and the detection of eye disease in the community. Trans Am Ophthalmol Soc. 2000; 98: 195-9; discussion PMCID: PMC Wild D, Kucur SS, Sznitman R. Spatial entropy pursuit for fast and accurate perimetry testing. Invest Ophthalmol Vis Sci. 2017; 58 (9): Rubinstein NJ, Mckendrick AM, Turpin A. Incorporating spatial models in visual field test procedures. Translational Vision Science & Technology. 2016; 5 (2): Gardiner SK, Mansberger SL. Effect of restricting perimetry testing algorithms to reliable sensitivities on test-retest variability. Invest Ophthalmol Vis Sci. 2016; 57 (13): PMCID: PMC Chong LX, Turpin A, Mckendrick AM. Targeted spatial sampling using goanna improves detection of visual field progression. Ophthalmic Physiol Opt. 2015; 35 (2): Anderson AJ, Bedggood PA, George Kong YX, Martin KR, Vingrys AJ. Can home monitoring allow earlier detection of rapid visual field progression in glaucoma? Ophthalmology. 2017; 124 (12): Prea SM, Kong YXG, Mehta A, He M, Crowston JG, Gupta V, Martin KR, Vingrys AJ. Six-month longitudinal comparison of a portable tablet perimeter with the humphrey field analyzer. Am J Ophthalmol. 2018; 190: Matsumoto C, Yamao S, Nomoto H, Takada S, Okuyama S, Kimura S, Yamanaka K, Aihara M, Shimomura Y. Visual field testing with head-mounted perimeter imo. PLoS One. 2016; 11 (8): e PMCID: PMC Glaucoma Now - Issue No 2,

10 Clinical Issues: Optimizing Your Approach to Visual Field Testing Brennan Eadie MD, PhD, FRCSC, Marcelo Nicolela MD, PhD, FRCSC Department of Ophthalmology & Visual Sciences, Dalhousie University Core Concepts The gold standard visual field test for most patients with glaucoma, or suspected of having glaucoma,in most situations remains Standard Automated Perimetry (SAP) with, for example, the SITA Standard testing strategy using the 24-2 program on the Humphrey Field Analyzer (HFA). The SITA FAST test strategy takes less time at the expense of greater variability making it acceptable for a first screening test, but suboptimal for routinely diagnosing and monitoring glaucoma. Frequency Doubling Technology (FDT) and Short Wavelength Auto mated Perimetry (SWAP) should not be used routinely instead of SAP. The 24-2 testing program remains preferable for most patients with glaucoma as it is faster than the 30-2 test pattern, and appears comparable for diagnosis and monitoring of glaucoma. A 10-2 testing program should be considered even in early or suspected glaucoma because a 24-2 testing pattern may miss small paracentral visual field defects that can be detected with this test. The 10-2 testing program is important for patients with visual field loss approaching fixation. Consider switching to a size V stimulus in cases of severe visual field damage, particularly with decreased visual acuity. In general, more frequent testing is recommended when patients are first diagnosed, in order to assess the rate of change within the first few years and to identify fast progressors (suggested six visual field tests in the first two years). Cluster-based trend analysis may be more sensitive to focal visual field progression while maintaining a similar sensitivity to analyzing global indices over time. Choose thoughtfully the best visual field test, or combination of tests, and the frequency of testing, at every visit. Visual field testing is critical for accurate diagnosis of glaucoma and monitoring for disease progression. All patients with glaucoma or suspected of having glaucoma, if able, should have periodic visual field testing. The optimal visual field test, or combination of tests, depends on the individual patient. The gold standard visual field test for most patients with glaucoma, or suspected of having glaucoma, remains Standard Automated Perimetry (SAP). Integral to SAP are various test strategies (e.g., Swedish Interactive Thresholding Algorithm [SITA] Standard or SITA Fast, Humphrey Field Analyzer) and various grids of points (e.g., 24-2, 30-2 or 10-2 programs) that can be employed to map the visual field. 1 Other hardware devices, such as Octopus, have similar fast thresholding strategies. The SITA FAST test strategy takes less time at the expense of greater variability. 1 As such, it may be acceptable as a tool to screen for glaucoma or monitor glaucoma suspects, but should not be used routinely for monitoring patients with glaucoma. A new strategy (SITA-faster) is currently becoming available, with claims of improved performance and even shorter test times, however no peer-reviewed publications are yet available at this time. The 24-2 testing program remains preferable for most patients with glaucoma as it is faster than the 30-2 test pattern, while maintaining similar performance for diagnosing and monitoring the disease. 2 In general, using the same test pattern increases one s chance of detecting progression. However, recent evidence suggests that a 24-2 test pattern may miss localized, paracentral visual field defects that can be detected with a 10-2 testing program. Thus, a 10-2 testing program should be considered even in early or suspected glaucoma. 3 (Figure 1) The 10-2 test pattern, in addition to its considered use in early or suspected glaucoma, is most useful for patients with visual field loss approaching fixation. It is sometimes reasonable to combine 10 Glaucoma Now - Issue No 2,

11 Figure. 1 - Examples of cases with 24-2 and 10-2 visual fields: (A) left eye with a paracentral defect shown both on 24-2 and 10-2 tests, but more pronounced in the latter; (b) left eye with a small localized paracentral defect missed with 24-2 test but identified with 10-2 test. Figure 2 - Comparison of Size III (A) vs Size V (B) testing stimulus in a patient with advanced visual field damage. Glaucoma Now - Issue No 2,

12 a 24-2 and 10-2 test pattern in the same eye at a single visit. One disadvantage of the 10-2 test is that there is currently no dedicated statistical progression analysis. One should consider using a size V stimulus in cases of severe visual field damage, particularly with decreased visual acuity, to preserve an adequate dynamic range to observe progression in the remaining visual field. (Figure 2) Other testing modalities, such as frequency doubling technology (FDT) and short-wavelength automated perimetry (SWAP) have been labeled as more sensitive than SAP for early defects; however, more recent evidence seems to suggest that this is not necessarily true, and these tests should not be routinely used at the expense of SAP. 4 There is considerable variation between individuals in their rates of progression. Performing more frequent visual field testing early after the initial diagnosis of glaucoma allows more timely detection of fast progressors, as these patients need to be treated aggressively. 5 After this initial period, the frequency of visual field testing can be reduced and depends on disease severity, age and general health, level of control of intraocular pressure, test tolerability and speed of change. Progression is commonly assessed by either event based or trend analysis. An example of event based analysis is the glaucoma progression analysis of the HFA, where points of each follow-up visual field are compared with a pair of baseline tests, and those points changing by a larger amount than expected by test-retest variability are flagged. Trend based analysis is usually performed on global parameters (such as MD or VFI). Although pointwise trend analysis can also be performed, this approach can lead to more false positive results. Another good strategy, available on the Octopus perimeter, is cluster-based trend analysis, which may have the added benefit of being more sensitive to focal visual field progression while maintaining a similar sensitivity to global indices analysis over time. 6 In summary, SAP using a program such as SITA 24-2 is the most utilized method for diagnosis and monitoring patients with glaucoma, and the frequency of testing should be adjusted depending on many factors. Clinicians should consider other approaches to visual field testing in special circumstances. References 1 Chauhan BC, Garway-Heath DF, Goni FJ, Rossetti L, Bengtsson B, Viswanathan AC, et al. Practical recommendations for measuring rates of visual field change in glaucoma. Br J Ophthalmol Apr;92(4): Artes PH, Iwase A, Ohno Y, Kitazawa Y, Chauhan BC. Properties of perimetric threshold estimates from full threshold, SITA standard, and SITA fast strategies. Invest Ophthalmol Vis Sci Aug;43(8): Khoury JM, Donahue SP, Lavin PJ, Tsai JC. Comparison of 24-2 and 30-2 perimetry in glaucomatous and nonglaucomatous optic neuropathies. J Neuroophthalmol Jun;19(2): De Moraes CG, Hood DC, Thenappan A, Girkin CA, Medeiros FA, Weinreb RN, Zangwill LM, Liebmann JM Visual fields miss central defects shown on 10-2 tests in glaucoma suspects, ocular hypertensives, and early glaucoma. Ophthalmology Oct;124(10): Aoki S, Murata H, Fujino Y, Matsuura M, Miki A, Tanito M, Mizoue S, Mori K, Suzuki K, Yamashita T, Kashiwagi K, Hirasawa K, Shoji N, Asaoka R. Investigating the usefulness of cluster-based trend analysis to detect visual field progression in patients with open-angle glaucoma. Br J Ophthalmol Apr;101(12): Nowomiejska K, Wrobel-Dudzinska D, Ksiazek K, Ksiazek P, Rejdak K, Maciejewski R, Juenemann AG, Rejdak R. Semi-automated kinetic perimetry provides additional information to static automated perimetry in the assessment of the remaining visual field in endstage glaucoma. Ophthalmic & Physiological Optics Mar; 35: Liu S, Lam S, Weinreb RN, Ye C, Cheung CY, Lai G, Lam DS, Leung CK. Comparison of standard automated perimetry, frequency-doubling technology perimetry, andnshort-wavelength automated perimetry for detection of glaucoma. Invest Ophthalmol Vis Sci Sep; 52(10): Glaucoma Now - Issue No 2,

13 Practical Tips: Minimizing artifacts and avoiding pitfalls in visual field interpretation C. Gustavo De Moraes, MD, MPH Columbia University Medical Center, New York, NY Core Concepts Make sure the correct date of birth has been entered. Keep in mind the effects of pupil size and do not perform visual field tests under pharmacologic mydriasis. Watch for ptosis and rim artifacts. Ensure the optimal reliability indices for each individual patient and confirm consistency across their visual field sequences. When assessing progression, reset the baseline tests based upon most recent changes in the treatment regimen. Do not rely solely on summary statistics. Perform 10-2 visual fields in patients with or suspected glaucoma. Standard automated perimetry (SAP) is a subjective, behavioral test. Its interpretation is therefore subject to patient-, disease-, and device-related sources of artifacts and variability. The following are some of the most relevant issues that need to be taken into account before, during, and after testing in all individuals undergoing SAP examinations. 1) Age There is a physiologic age-related decay in retinal sensitivity which is estimated to be approximately 0.07 to 0.08 db loss per year 1 and varies with eccentricity and hemifields. 2 The total deviation (TD) plot shows the differences of pointwise threshold sensitivities between the subject undergoing visual field testing and the corresponding average values of age-matched controls. The TD probabilities are based on the distribution of these normative values. An incorrect year of birth, particularly by more than a decade, may result in misclassification. If the tested patient is compared with younger subjects, for instance, the TD probability plot may falsely show clusters of significant abnormalities; if compared with older subjects, one may miss an early localized defect. Implications may be even more significant when assessing progression: even small changes (improvement or decay) in pointwise or global (mean deviation, MD) sensitivities even on a single visit from the entire sequence can affect their respective slopes and p-values, also leading to misclassification. 2) Pupil size Both mydriasis and miosis can lead to abnormal results on SAP. Miosis is more difficult to address before testing because it often results from the pharmacologic effects of intraocular pressure (IOP) lowering medications (pilocarpine) and their wash-out could be detrimental (e.g. from IOP elevation or angle closure). Mydriasis, on the other hand, is often induced pharmacologically during office visits for an enhanced fundus examination. Visual field testing with mydriasis can decrease foveal threshold and MD by 2.0 and 1.2 db on average, respectively. 3 This effect can have clinically significant implications both for the detection of visual field defects and for monitoring progressive changes. Therefore, clinicians should avoid performing visual field tests after pupil dilation. If the patient is on pilocarpine, clinicians should keep track of changes in pupil size between visits and consider its effect on sensitivities when assessing the overall level of functional loss. 3) Edge-artifacts Some arcuate defects in the superior hemifield can be the result of partial ptosis and an incorrect diagnosis of glaucomatous functional loss may ensue. Always check whether the upper lid is blocking even partially the visual axis. If so, consider elevating the eyelids (temporarily with tape or permanently). Moreover, if the corrective lens used during visual field testing is decentered or set too far from the eye, the lens rim may project into the central field and lead to arcuate-like defects that resemble glaucomatous defects. If the lens correction exceeds 5 diopters, a soft contact lens correction should be considered to minimize trial lens rim artifacts. 4 4) Reliability indices Unreliable tests are common even among experienced patients. For some patients, visual field test results meeting the standard reliability criteria may never be achieved. Glaucoma Now - Issue No 2,

14 To address this issue, a recent study investigated evidence-based criteria for assessment of visual field reliability. 5 Fixation losses (FL) had little impact on the MD variation and no level of FL produced more than 1 db variation at any disease stage. False negatives had a minimal impact up to 20% abnormal catch trials. More importantly, false positives yielded more positive MD values even when less than 20%, with greater effect above this level. If a patient is unable to perform tests with satisfactory reliability indices, clinicians should consider the variation within these indices over time when assessing progression. 5) Baseline tests Event- and trend-based progression analyses are performed using visual field tests performed within a time frame of interest. If progression is suspected or confirmed, treatment may be initiated or escalated. Clinicians should remember to re-set the baseline tests whenever a significant change in treatment occurs. If this is not done, the progression algorithms will continue to flag that sequence as progressing ( possible or likely in the case of event analysis, or statistically significant slopes in the case of trend analysis). Clinicians could then mistakenly assume the treatment change was not effective and subject the patient to unnecessary further interventions. 6) Summary statistics To increase objectivity when interpreting visual field results, computerized statistical analyses provide summary indices that may help to define the presence of functional loss, its severity, as well as tracking longitudinal changes. Most commonly used indices are the MD, pattern standard deviation (PSD), and glaucoma hemifield test (GHT). Based on how the patient s test compares with the normative database, these indices can be flagged with probabilities (MD and PSD) or whether they are outside normal limits (GHT). Based on how the patient s test compares with the normative database, these indices can be flagged with probabilities (MD and PSD) or whether they are outside normal limits (GHT). However, these indices should be used only to facilitate the interpretation of the test results, not to diagnose glaucoma. In particular, their results should be interpreted after a scrutinized evaluation of the TD and pattern deviation (PD) plots looking for patterns of loss that are typical of glaucoma (e.g. arcuate defects, nasal steps). More importantly, visual field test results should never be interpreted in isolation: they should always be correlated with other clinical information and (ideally) the results of structural tests (disc photos and optical coherence tomography). Figure 1 shows an example of an eye initially classified as preperimetric or glaucoma suspect because the 24-2 visual field was within normal limits. However, note that a cluster of test locations reached statistical significance at the 5% and 2% levels which matched the location flagged as abnormal on the OCT. 7) 10-2 visual fields 24-2 visual fields (or 30-2, or due to glaucoma. For the 24-2 and 30-2 grids, test location are 6 degrees apart from one another (3 degrees from the fovea). As a result, the central 10 degrees of the visual field (corresponding to the macular region) is tested with only 16 points (plus the fovea), which is insufficient for an accurate detection of abnormalities in an area that encompasses about 30% of all ganglion cells of the retina. 6 Figure 2 depicts a normal 24-2 visual field result while the 10-2 done on the same day shows a classic arcuate defect typically seen in glaucoma. About 52% of patients with early visual field loss have macular damage missed with the 24-2 but shown on the 10-2 and corroborated with OCT. 7 References 1 Jaffe GJ, Alvarado JA, Juster RP. Age-related changes of the normal visual field. Archives of ophthalmology (Chicago, Ill. : 1960). 1986;104(7): Haas A, Flammer J, Schneider U. Influence of age on the visual fields of normal subjects. American journal of ophthalmology. 1986;101(2): Kudrna GR, Stanley MA, Remington LA. Pupillary dilation and its effects on automated perimetry results. Journal of the American Optometric Association. 1995;66(11): Keltner JL, Johnson CA, Quigg JM, Cello KE, Kass MA, Gordon MO. Confirmation of visual field abnormalities in the Ocular Hypertension Treatment Study. Ocular Hypertension Treatment Study Group. Archives of ophthalmology (Chicago, Ill. : 1960). 2000;118(9): Yohannan J, Wang J, Brown J, et al. Evidence-based Criteria for Assessment of Visual Field Reliability. Ophthalmology. 2017;124(11): Curcio CA, Allen KA. Topography of ganglion cells in human retina. The Journal of comparative neurology. 1990;300(1): Grillo LM, Wang DL, Ramachandran R, et al. The 24-2 Visual Field Test Misses Central Macular Damage Confirmed by the 10-2 Visual Field Test and Optical Coherence Tomography. Translational vision science & technology. 2016;5(2): Glaucoma Now - Issue No 2,

15 Figure. 1 - Right panel: 24-2 visual field with summary indices (MD, PSD, and GHT) within normal limits. Note, however, that there are two points in the nasal area with P-values of 5 and 2%. Left panel: OCT report flipped to field view shows an arcuate-like defect in the nerve fiber layer (black arrow) that projects to the nasal area where the abnormal points are located (black shape). Figure. 2 - The importance of performing 10-2 visual fields. Top left: OCT scan of the macula, which includes 30% of the retinal ganglion cells (RGC). Below it, fundus photos with the test location of the central 10 degrees of the 24-2 (red crosses) and 10-2 (black circles) grids overlaid. Right panel, top: a normal 24-2 of with no abnormalities (red arrow) within the central 10 degrees (black square). Bottom: the 10-2 done on the same day shows an arcuate defect (red arrow). The red circles represent the 16 location of the 24-2 grid that fall within the central 10 degrees (macula). Glaucoma Now - Issue No 2,

16 For each issue 4 CME credits can be obtained by successfully passing the multiple choice test on our website! In order to maximize the learning effect, participants have the opportunity to register at our website and to answer a number of multiple choice questions for each of the four sections covering the key points of each section. Shortly after test completion, participants receive electronic feedback on successful accomplishment or failure. In case of failure the participant is encouraged to review articles and retake the test. A successful test will earn the participant valuable CME credits needed for their continuous medical education efforts. ACCREDITATIONS Individual issues of Glaucoma Now are accredited for Continuing Medical Education (CME) by the Physicians Chamber of Baden- Württemberg, Germany (Local Medical Responsible: Andreas Buchholz, MD, PhD, ROph). This accreditation automatically implies acceptance of credits throughout the European Union and associated countries. Glaucoma Now is also recognized by the Royal Australian and New Zealand College of Ophthalmologists as a valid Continuing Professional Development activity. Since 2013 the program is recognized by the Brazilian Council of Ophthalmology. Brazilian physicians successfully taking CME tests on our website are automatically awarded CME points by CBO. STATEMENT OF NEED AND PROGRAM DESCRIPTION Recent months and years have seen significant advances in our understanding of glaucoma. Much has been learned, not only about damage mechanisms and pathogenesis, but also about diagnosis and management. Treatment options both medical and surgical continue to expand. This program will review this new knowledge with an emphasis on incorporating recent insights into day-to-day practice. Date of original release: October 2018 Approved for a period of 12 months. This issue is accredited for Continuing Medical Education (CME) by the Physicians Chamber of Baden-Württemberg, Germany (Local Medical Responsible: Andreas Buchholz MD, PhD, ROph) DISCLAIMER Participants have an implied responsibility to use newly acquired information to enhance patient outcomes and professional development. The information presented in this activity is not meant to serve as a guideline for patient care. Any procedures, medications, or other courses of diagnosis or treatment discussed or suggested in this activity should not be used by clinicians without evaluation of their patient s conditions and possible contraindications or dangers in use, applicable manufacturer s product information, and comparison with recommendations of other authorities. Contributors Linda Zangwill, MD, PhD is Professor at Hamilton Glaucoma Center and the Shiley Eye Institute, University of California San Diego, California USA. She has received financial support from Heidelberg Engineering, Carl Zeiss Meditec, Topcon Medical Systems and Optovue Inc., in the form of research funding or research materials or services at no cost. She is an inventor/developer designated on the patent of Carl Zeiss Meditec. Christopher Bowd, MD, PhD is a researcher at Hamilton Glaucoma Center and the Shiley Eye Institute, University of California San Diego, California USA and a coworker of Linda Zangwill. He has no commercial relationships to disclose. Stuart Gardiner, MD, PhD is Professor at Devers Eye Institute, Legacy Research Institute, Portland, Oregon, USA. He has no commercial relationships to disclose. Marcelo Nicolela, MD, PhD, FRCSC is Professor at the Department of Ophthalmology & Visual Sciences, Dalhousie University, Halifax, Canada. He has no commercial relationships to disclose. Brennan Eadie MD, PhD, FRCSC, is a researcher at the Department of Ophthalmology & Visual Sciences, Dalhousie University, Canada and a coworker of Marcelo Nicolela. He has no commercial relationships to disclose. Gustavo de Moraes MD, PhD, is Associate Professor of Ophthalmic Sciences, Medical Director of Clinical Trials at Columbia University Medical Center, New York Presbyterian Hospital, Edward S. Harkness Eye Institute, New York, USA. Within the last 12 months he has received financial support from Carl Zeiss Meditec, Inc. and Topcon, Inc. in the form of research funding or research materials or services at no cost. DISCLOSURE STATEMENT EDITORIAL BOARD Ivan Goldberg serves on the Faculty and Advisory Boards of the following companies: Alcon, Allergan, Merck and Pfizer. Remo Susanna served on the Faculty and Advisory Boards of the following companies: Alcon, Allergan, Merck and Pfizer. 16 Glaucoma Now - Issue No 2,

The overall rate of decrease in visual field sensitivity in

The overall rate of decrease in visual field sensitivity in Glaucoma Estimating the True Distribution of Visual Field Progression Rates in Glaucoma Andrew J. Anderson Department of Optometry and Vision Sciences, The University of Melbourne, Parkville, Australia

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Artes, P. H., O'Leary, N., Nicolela, M. T., Chauhan, B. C. and Crabb, D. P. (2014). Visual Field Progression in Glaucoma

More information

21st Century Visual Field Testing

21st Century Visual Field Testing Supplement to Supported by an educational grant from Carl Zeiss Meditec, Inc. Winter 2011 21st Century Visual Field Testing the Evolution Continues 21st Century Visual Field Testing 21st Century Visual

More information

Financial Disclosure. Visual Field Interpretation RELIABILITY VISUAL FIELD INTERPRETATION IN GLAUCOMA METHODS OF DATA PRESENTATION

Financial Disclosure. Visual Field Interpretation RELIABILITY VISUAL FIELD INTERPRETATION IN GLAUCOMA METHODS OF DATA PRESENTATION VISUAL FIELD INTERPRETATION IN GLAUCOMA Danica J. Marrelli, OD, FAAO University of Houston College of Optometry Financial Disclosure I have received speaking and/or consulting fees from: Aerie Pharmaceutical

More information

The determination of glaucoma progression is based on a. Refinement of Pointwise Linear Regression Criteria for Determining Glaucoma Progression

The determination of glaucoma progression is based on a. Refinement of Pointwise Linear Regression Criteria for Determining Glaucoma Progression Glaucoma Refinement of Pointwise Linear Regression Criteria for Determining Glaucoma Progression Colleen M. Kummet, 1 K. D. Zamba, 2 Carrie K. Doyle, 3 5 Chris A. Johnson, 3,4 and Michael Wall 3 5 1 Department

More information

CLINICAL SCIENCES. Glaucoma Monitoring in a Clinical Setting

CLINICAL SCIENCES. Glaucoma Monitoring in a Clinical Setting CLINICAL SCIENCES Glaucoma Monitoring in a Clinical Setting Glaucoma Progression Analysis vs Nonparametric Progression Analysis in the Groningen Longitudinal Glaucoma Study Christiaan Wesselink, MD; Govert

More information

Citation for published version (APA): Wesselink, C. (2017). Glaucoma care optimised in an ageing population [Groningen]: Rijksuniversiteit Groningen

Citation for published version (APA): Wesselink, C. (2017). Glaucoma care optimised in an ageing population [Groningen]: Rijksuniversiteit Groningen University of Groningen Glaucoma care optimised in an ageing population Wesselink, Christiaan IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from

More information

FORUM Glaucoma Workplace from ZEISS Clinical Interpretation Guide

FORUM Glaucoma Workplace from ZEISS Clinical Interpretation Guide FORUM Glaucoma Workplace from ZEISS Clinical Interpretation Guide ZEISS FORUM Glaucoma Workplace.0 For years, doctors have asked for the operational capability to analyze data from their Humphrey Field

More information

3/16/2018. Perimetry

3/16/2018. Perimetry Perimetry The normal visual field extends further away from fixation temporally and inferiorly than superiorly and nasally. From the center of the retina this sensitivity decreases towards the periphery,

More information

MOVE IT OR LOSE IT: THE ROLE OF KINETIC VISUAL FIELDS

MOVE IT OR LOSE IT: THE ROLE OF KINETIC VISUAL FIELDS MOVE IT OR LOSE IT: THE ROLE OF KINETIC VISUAL FIELDS Course Objectives Review the visual field Review types of perimetry Discuss advantages and disadvantages of different types of visual field testing

More information

T he retinal ganglion cells of different sizes have distinct

T he retinal ganglion cells of different sizes have distinct 604 CLINICAL SCIENCE Agreement between frequency doubling perimetry and static perimetry in eyes with high tension glaucoma and normal tension glaucoma S Kogure, Y Toda, D Crabb, K Kashiwagi, F W Fitzke,

More information

VISUAL FIELD TESTING OCCUpies

VISUAL FIELD TESTING OCCUpies CLINICAL SCIENCES Prediction of Glaucomatous Visual Field Loss by Extrapolation of Linear Trends Boel Bengtsson, PhD; Vincent Michael Patella, OD; Anders Heijl, MD, PhD Objective: To investigate how well

More information

CORRELATING OF THE VISUAL FIELD INDEX WITH MEAN DEVIATION AND PATTERN STANDARD DEVIATION IN GLAUCOMA PATIENTS

CORRELATING OF THE VISUAL FIELD INDEX WITH MEAN DEVIATION AND PATTERN STANDARD DEVIATION IN GLAUCOMA PATIENTS CORRELATING OF THE VISUAL FIELD INDEX WITH MEAN DEVIATION AND PATTERN STANDARD DEVIATION IN GLAUCOMA PATIENTS Bui Thi Huong Giang, Pham Thi Kim Thanh Department of Ophthamology, Hanoi Medical University

More information

Access to the published version may require journal subscription. Published with permission from: Elsevier

Access to the published version may require journal subscription. Published with permission from: Elsevier This is an author produced version of a paper published in Ophthalmology. This paper has been peer-reviewed but does not include the final publisher proof-corrections or journal pagination. Citation for

More information

STANDARD AUTOMATED PERIMETRY IS A GENERALLY

STANDARD AUTOMATED PERIMETRY IS A GENERALLY Comparison of Long-term Variability for Standard and Short-wavelength Automated Perimetry in Stable Glaucoma Patients EYTAN Z. BLUMENTHAL, MD, PAMELA A. SAMPLE, PHD, LINDA ZANGWILL, PHD, ALEXANDER C. LEE,

More information

Expanding your field of vision. Visual Field Analyzers from Carl Zeiss

Expanding your field of vision. Visual Field Analyzers from Carl Zeiss Expanding your field of vision Visual Field Analyzers from Carl Zeiss Vision in focus Visual Field Analyzers Humphrey Field Analyzer/HFA II-i Series Humphrey Matrix Humphrey FDT Perimeter Software Guided

More information

LABORATORY SCIENCES. Spatial and Temporal Processing of Threshold Data for Detection of Progressive Glaucomatous Visual Field Loss

LABORATORY SCIENCES. Spatial and Temporal Processing of Threshold Data for Detection of Progressive Glaucomatous Visual Field Loss LABORATORY SCIENCES Spatial and Temporal Processing of Threshold Data for Detection of Progressive Glaucomatous Visual Field Loss Paul G. D. Spry, PhD; Chris A. Johnson, PhD; Alex B. Bates; Andrew Turpin,

More information

Misleading Statistical Calculations in Faradvanced Glaucomatous Visual Field Loss

Misleading Statistical Calculations in Faradvanced Glaucomatous Visual Field Loss Misleading Statistical Calculations in Faradvanced Glaucomatous Visual Field Loss Eytan Z. Blumenthal, MD, Ruthy Sapir-Pichhadze, MD Objective: In this study, the capability of statistical analysis indices

More information

VISUAL FIELDS. Visual Fields. Getting the Terminology Sorted Out 7/27/2018. Speaker: Michael Patrick Coleman, COT & ABOC

VISUAL FIELDS. Visual Fields. Getting the Terminology Sorted Out 7/27/2018. Speaker: Michael Patrick Coleman, COT & ABOC VISUAL FIELDS Speaker: Michael Patrick Coleman, COT & ABOC Visual Fields OBJECTIVES: 1. Explain what is meant by 30-2 in regards to the Humphrey Visual Field test 2. Identify the difference between a kinetic

More information

Description of new EyeSuite visual field and trend analysis functions

Description of new EyeSuite visual field and trend analysis functions Description of new EyeSuite visual field and trend analysis functions Global Trend Graph The trend graph indicates the normality range (95%) on top as grey band. Falling below that area means falling out

More information

New Concepts in Glaucoma Ben Gaddie, OD Moderator Murray Fingeret, OD Louis Pasquale, MD

New Concepts in Glaucoma Ben Gaddie, OD Moderator Murray Fingeret, OD Louis Pasquale, MD New Concepts in Glaucoma Ben Gaddie, OD Moderator Murray Fingeret, OD Louis Pasquale, MD New Concepts in Glaucoma Optical Coherence Tomography: Is it necessary and needed to diagnose and monitor glaucoma?

More information

Test-retest variability in visual field testing using frequency doubling technology

Test-retest variability in visual field testing using frequency doubling technology European Journal of Ophthalmology / Vol. 17 no. 2, 2007 / pp. 203-207 Test-retest variability in visual field testing using frequency doubling technology A. HORANI, S. FRENKEL, E.Z. BLUMENTHAL Department

More information

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Chiharu Matsumoto, Shiroaki Shirato, Mai Haneda, Hiroko Yamashiro

More information

CLINICAL SCIENCES. (FDP) was designed to emphasize the response characteristics of the parasol

CLINICAL SCIENCES. (FDP) was designed to emphasize the response characteristics of the parasol CLINICAL SCIENCES Detecting Visual Function Abnormalities Using the Swedish Interactive Threshold Algorithm and Matrix Perimetry in Eyes With Glaucomatous Appearance of the Optic Disc Lisandro M. Sakata,

More information

Evolving glaucoma management True diagnostic integration for the preservation of vision

Evolving glaucoma management True diagnostic integration for the preservation of vision Evolving glaucoma management True diagnostic integration for the preservation of vision // GLAUCOMA MANAGEMENT MADE BY ZEISS The moment you are certain it is glaucoma. This is the moment we work for. There

More information

CHAPTER 10 NON-CONVENTIONAL PERIMETRY

CHAPTER 10 NON-CONVENTIONAL PERIMETRY 193 CHAPTER 10 NON-CONVENTIONAL PERIMETRY INTRODUCTION Static Standard Automated Perimetry (SAP, alternatively called white-on-white perimetry), which uses a white Goldmann size III stimulus presented

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Russell, R. A., Garway-Heath, D. F. & Crabb, D. P. (2013). New insights into measurement variability in glaucomatous visual

More information

Glaucoma Progression. Murray Fingeret, OD

Glaucoma Progression. Murray Fingeret, OD Glaucoma Progression Murray Fingeret, OD Managing glaucoma or glaucoma suspects Summary For patients diagnosed or at risk of glaucoma Stage disease Risk assessment Asses for progression Treatment plan

More information

Main topic: Glaucoma Concepts Rate of progression

Main topic: Glaucoma Concepts Rate of progression ISSUE 2 2014 Inside this issue: Special Focus: Concept of glaucoma progression, importance of measurement, functional versus structural measurement, evidence from clinical trials What s New: New technologies

More information

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma.

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Donald C. Hood 1,2,* and Ali S. Raza 1 1 Department of Psychology, Columbia

More information

Visual Fields Shawn L. Cohen, M.D. Part 2 of 4. Definitions / Tables (Part 2 of 2) Static Perimetry (Humphrey, Octopus)

Visual Fields Shawn L. Cohen, M.D. Part 2 of 4. Definitions / Tables (Part 2 of 2) Static Perimetry (Humphrey, Octopus) Visual Fields Shawn L. Cohen, M.D. Part 2 of 4 Definitions / Tables (Part 2 of 2) Static Perimetry (Humphrey, Octopus) Normal Visual Field: Components: General Information Reliability Indices Raw Data

More information

Diagnosing open-angle glaucoma may be particularly

Diagnosing open-angle glaucoma may be particularly A New Look at Selective Perimetry What is its role in clinical practice? BY MURRAY FINGERET, OD Diagnosing open-angle glaucoma may be particularly challenging when damage is mild or early. Although most

More information

F requency doubling technology (FDT) perimetry was

F requency doubling technology (FDT) perimetry was 131 EXTENDED REPORT Clinical evaluation of frequency doubling technology perimetry using the Humphrey Matrix 24-2 threshold strategy P G D Spry, H M Hussin, J M Sparrow... See end of article for authors

More information

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography Original articles in Indian eyes with optical coherence tomography Malik A, Singh M, Arya SK, Sood S, Ichhpujani P Department of Ophthalmology Government Medical College and Hospital, Sector 32, Chandigarh,

More information

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects Romanian Journal of Ophthalmology, Volume 60, Issue 3, July-September 2016. pp:158-164 GENERAL ARTICLE Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

More information

PATTERNS OF GLAUCOMATOUS VISUAL FIELD LOSS IN SITA FIELDS AUTOMATICALLY IDENTIFIED USING INDEPENDENT COMPONENT ANALYSIS

PATTERNS OF GLAUCOMATOUS VISUAL FIELD LOSS IN SITA FIELDS AUTOMATICALLY IDENTIFIED USING INDEPENDENT COMPONENT ANALYSIS PATTERNS OF GLAUCOMATOUS VISUAL FIELD LOSS IN SITA FIELDS AUTOMATICALLY IDENTIFIED USING INDEPENDENT COMPONENT ANALYSIS BY Michael H. Goldbaum MD,* Gil-Jin Jang PhD, Chris Bowd PhD, Jiucang Hao PhD, Linda

More information

Assessing the GOANNA Visual Field Algorithm Using Artificial Scotoma Generation on Human Observers

Assessing the GOANNA Visual Field Algorithm Using Artificial Scotoma Generation on Human Observers Article Assessing the GOANNA Visual Field Algorithm Using Artificial Scotoma Generation on Human Observers Luke X. Chong 1,2, Andrew Turpin 3, and Allison M. McKendrick 1 1 Department of Optometry & Vision

More information

Definitions. Indications for Perimetry. Indications for Perimetry. Purposes of Perimetry. Indications for Perimetry 3/4/2015

Definitions. Indications for Perimetry. Indications for Perimetry. Purposes of Perimetry. Indications for Perimetry 3/4/2015 DISCLOSURE STATEMENT No disclosure statement. 2 Definitions Lecturer: UNDERSTANDING VISUAL FIELD TESTING Caroline B. Pate, OD, FAAO Visual Field: The portion of space which is visible when gaze is fixed

More information

Citation for published version (APA): Stoutenbeek, R. (2010). Population based glaucoma screening Groningen: s.n.

Citation for published version (APA): Stoutenbeek, R. (2010). Population based glaucoma screening Groningen: s.n. University of Groningen Population based glaucoma screening Stoutenbeek, Remco IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check

More information

How Useful Is Population Data for Informing Visual Field Progression Rate Estimation?

How Useful Is Population Data for Informing Visual Field Progression Rate Estimation? Glaucoma How Useful Is Population Data for Informing Visual Field Progression Rate Estimation? Andrew J. Anderson 1 and Chris A. Johnson 2 PURPOSE. Bayesian estimators allow the frequency of visual field

More information

Persistence, Spatial Distribution and Implications for Progression Detection of Blind Parts of the Visual Field in Glaucoma: A Clinical Cohort Study

Persistence, Spatial Distribution and Implications for Progression Detection of Blind Parts of the Visual Field in Glaucoma: A Clinical Cohort Study Persistence, Spatial Distribution and Implications for Progression Detection of Blind Parts of the Visual Field in Glaucoma: A Clinical Cohort Study Francisco G. Junoy Montolio 1., Christiaan Wesselink

More information

Correlating Structure With Function in End-Stage Glaucoma

Correlating Structure With Function in End-Stage Glaucoma C L I N I C A L S C I E N C E Correlating Structure With Function in End-Stage Glaucoma Eytan Z. Blumenthal, MD; Amjad Horani, MD; Rajesh Sasikumar, MD; Chandrasekhar Garudadri, MD; Addepalli Udaykumar,

More information

Ganglion cell complex scan in the early prediction of glaucoma

Ganglion cell complex scan in the early prediction of glaucoma Original article in the early prediction of glaucoma Ganekal S Nayana Super Specialty Eye Hospital and Research Center, Davangere, Karnataka, India Abstract Objective: To compare the macular ganglion cell

More information

TUMBLING E RESOLUTION PERIMETRY IN GLAUCOMA

TUMBLING E RESOLUTION PERIMETRY IN GLAUCOMA Tumbling E resolution perimetry in glaucoma 179 TUMBLING E RESOLUTION PERIMETRY IN GLAUCOMA FERGAL A. ENNIS 1, ROGER S. ANDERSON 1, WINSTON S. McCLEAN 1 and SIMON J.A. RANKIN 2 1 Vision Science Research

More information

As methods become available to measure the structure of

As methods become available to measure the structure of Combining Ganglion Cell Topology and Data of Patients with Glaucoma to Determine a Structure Function Map Andrew Turpin, 1 Geoff P. Sampson, 2 and Allison M. McKendrick 2 PURPOSE. To introduce techniques

More information

CHAPTER 6 SELECTING A TEST STRATEGY INTRODUCTION. accurate but shorter test may yield more useful visual

CHAPTER 6 SELECTING A TEST STRATEGY INTRODUCTION. accurate but shorter test may yield more useful visual 81 CHAPTER 6 SELECTING A TEST STRATEGY INTRODUCTION As illustrated in Chapter 4, determining sensitivity thresholds by assessing all levels of stimulus intensity (e.g., stimulus luminance) is not practical

More information

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS Nerve fiber layer thickness in normals and glaucoma patients 403 NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS HIROTAKA SUZUMURA, KAYOKO HARASAWA, AKIKO KOBAYASHI and NARIYOSHI ENDO Department

More information

Glaucoma is defined as a progressive optic neuropathy that. A Comparison of Functional and Structural Measures for Identifying Progression of Glaucoma

Glaucoma is defined as a progressive optic neuropathy that. A Comparison of Functional and Structural Measures for Identifying Progression of Glaucoma Glaucoma A Comparison of Functional and Structural Measures for Identifying Progression of Glaucoma Daiyan Xin, 1 Vivienne C. Greenstein, 2 Robert Ritch, 3,4 Jeffrey M. Liebmann, 3 Carlos Gustavo De Moraes,

More information

CHAPTER 11 KINETIC PERIMETRY WHAT IS KINETIC PERIMETRY? LIMITATIONS OF STATIC PERIMETRY LOW SPATIAL RESOLUTION

CHAPTER 11 KINETIC PERIMETRY WHAT IS KINETIC PERIMETRY? LIMITATIONS OF STATIC PERIMETRY LOW SPATIAL RESOLUTION 205 CHAPTER 11 KINETIC PERIMETRY WHAT IS KINETIC PERIMETRY? LIMITATIONS OF STATIC PERIMETRY LOW SPATIAL RESOLUTION Static perimetry is currently the most commonly used type of perimetry. With static perimetry,

More information

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma Module 3a GDx Educational Slide Deck Carl Zeiss Meditec, Inc. November 2005 1 Structure & Function Modules Module

More information

Receiver operating characteristic (ROC) curves are a wellaccepted

Receiver operating characteristic (ROC) curves are a wellaccepted A Statistical Approach to the Evaluation of Covariate Effects on the Receiver Operating Characteristic Curves of Diagnostic Tests in Glaucoma Felipe A. Medeiros, 1 Pamela A. Sample, 1 Linda M. Zangwill,

More information

Comparison of Visual Field Measurement with Heidelberg Edge Perimeter and Humphrey Visual Field Analyzer in Patients with Ocular Hypertension

Comparison of Visual Field Measurement with Heidelberg Edge Perimeter and Humphrey Visual Field Analyzer in Patients with Ocular Hypertension ORIGINAL PAPERS Adv Clin Exp Med 2016, 25, 5, 937 944 DOI: 10.17219/acem/62142 Copyright by Wroclaw Medical University ISSN 1899 5276 Kamil Kaczorowski 1, B E, Małgorzata Mulak 1, A, C, E, Dorota Szumny

More information

Learn Connect Succeed. JCAHPO Regional Meetings 2017

Learn Connect Succeed. JCAHPO Regional Meetings 2017 Learn Connect Succeed JCAHPO Regional Meetings 2017 Visual Field Testing Suzanne Hansen, M.Ed., COMT, OSC Why are these tests ordered? Visual field testing is ordered to help the physician diagnose and

More information

The Role of the RNFL in the Diagnosis of Glaucoma

The Role of the RNFL in the Diagnosis of Glaucoma Chapter 1. The Role of the RNFL in the Diagnosis of Glaucoma Introduction Glaucoma is an optic neuropathy characterized by a loss of of retinal ganglion cells and their axons, the Retinal Nerve Fiber Layer

More information

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation C L I N I C A L S C I E N C E The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation Amjad Horani, MD; Shahar Frenkel, MD, PhD; Eytan Z. Blumenthal, MD BACKGROUND

More information

OtticaFisiopatologica

OtticaFisiopatologica Anno quindicesimo dicembre 2010 How to assess the retinal nerve fiber layer thickness Antonio Ferreras Miguel Servet University Hospital, Zaragoza. Aragón Health Sciences Institute University of Zaragoza

More information

A Critical Reevaluation of Current Glaucoma Management

A Critical Reevaluation of Current Glaucoma Management A Critical Reevaluation of Current Glaucoma Management International Glaucoma Think Tank, July 27 29, 2006, Taormina, Sicily J. Caprioli, MD, 1 D. F. Garway-Heath, MD, FRCOphth 2 Think Tank* Co-chairs

More information

Investigative Ophthalmology & Vision Sciences MSc Course. Glaucoma Module. Visual Field Reliability Indices. David Henson 2014.

Investigative Ophthalmology & Vision Sciences MSc Course. Glaucoma Module. Visual Field Reliability Indices. David Henson 2014. Investigative Ophthalmology & Vision Sciences MSc Course Glaucoma Module Visual Field Reliability Indices David Henson 214 Variability 1 Variability Clinical problem How can we judge whether the change

More information

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS The Relationship between Standard Automated Perimetry and GDx VCC Measurements Nicolaas J. Reus and Hans G. Lemij PURPOSE. To investigate the relationship between retinal light sensitivity measured with

More information

C linical visual field tests are designed to provide. Measurement error of visual field tests in glaucoma LABORATORY SCIENCE

C linical visual field tests are designed to provide. Measurement error of visual field tests in glaucoma LABORATORY SCIENCE 107 LABORATORY SCIENCE Measurement error of visual field tests in glaucoma P G D Spry, C A Johnson, A M McKendrick, A Turpin... Br J Ophthalmol 2003;87:107 112 See end of article for authors affiliations...

More information

Incorporating Risk Factors to Improve the Assessment of Rates of Glaucomatous Progression

Incorporating Risk Factors to Improve the Assessment of Rates of Glaucomatous Progression Glaucoma Incorporating Risk Factors to Improve the Assessment of Rates of Glaucomatous Progression Felipe A. Medeiros, Linda M. Zangwill, Kaweh Mansouri, Renato Lisboa, Ali Tafreshi, and Robert N. Weinreb

More information

Glaucoma: a disease of the macula?

Glaucoma: a disease of the macula? Glaucoma: a disease of the macula? Derek MacDonald, OD, FAAO Waterloo, Ontario, Canada Please silence all mobile devices and remove items from chairs so others can sit. Unauthorized recording of this session

More information

Perimetry Phobia: Don t fear the field Savory Turman, COMT, CPSS

Perimetry Phobia: Don t fear the field Savory Turman, COMT, CPSS Perimetry Phobia: Don t fear the field Savory Turman, COMT, CPSS I have no financial interest in this presentation. Who am I? Where am I? What am I? The anatomy of the visual field Purpose of Visual Field

More information

PRESCRIBING IN GLAUCOMA: GUIDELINES FOR NZ OPTOMETRISTS

PRESCRIBING IN GLAUCOMA: GUIDELINES FOR NZ OPTOMETRISTS PRESCRIBING IN GLAUCOMA: GUIDELINES FOR NZ OPTOMETRISTS Introduction Independent prescribing relates to the capacity to use clinical judgement in respect of diagnosis and treatment. It does not mean working

More information

Perimetric testing is used clinically to detect visual field

Perimetric testing is used clinically to detect visual field Quantifying Effects of Retinal Illuminance on Frequency Doubling Perimetry William H. Swanson, Mitchell W. Dul, and Susan E. Fischer PURPOSE. To measure and quantify effects of variation in retinal illuminance

More information

Development of Visual Field Screening Procedures: A Case Study of the Octopus Perimeter

Development of Visual Field Screening Procedures: A Case Study of the Octopus Perimeter Article DOI: 10.1167/tvst.5.3.3 Development of Visual Field Screening Procedures: A Case Study of the Octopus Perimeter Andrew Turpin 1, Jonathan S. Myers 2, and Allison M. McKendrick 3 1 Department of

More information

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2013 October 01.

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2013 October 01. NIH Public Access Author Manuscript Published in final edited form as: Arch Ophthalmol. 2012 September ; 130(9): 1107 1116. doi:10.1001/archophthalmol.2012.827. A Combined Index of Structure and Function

More information

OCT in the Diagnosis and Follow-up of Glaucoma

OCT in the Diagnosis and Follow-up of Glaucoma OCT in the Diagnosis and Follow-up of Glaucoma Karim A Raafat MD. Professor Of Ophthalmology Cairo University Hmmmm! Do I have Glaucoma or not?! 1 Visual Function 100% - N Gl Structure : - 5000 axon /

More information

A Validated Risk Calculator to Assess Risk and Rate of Visual Field Progression in Treated Glaucoma Patients

A Validated Risk Calculator to Assess Risk and Rate of Visual Field Progression in Treated Glaucoma Patients Glaucoma A Validated Risk Calculator to Assess Risk and Rate of Visual Field Progression in Treated Glaucoma Patients Carlos Gustavo De Moraes, 1,2 Mitra Sehi, 3 David S. Greenfield, 3 Yun S. Chung, 3

More information

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Original article Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Dewang Angmo, 1 Shibal Bhartiya, 1 Sanjay K Mishra,

More information

Trans Am Ophthalmol Soc 2015;113:T4[1-10] by the American Ophthalmological Society.

Trans Am Ophthalmol Soc 2015;113:T4[1-10] by the American Ophthalmological Society. CAN VISUAL FIELD PROGRESSION BE PREDICTED BY CONFOCAL SCANNING LASER OPHTHALMOSCOPIC IMAGING OF THE OPTIC NERVE HEAD IN GLAUCOMA? (AN AMERICAN OPHTHALMOLOGICAL SOCIETY THESIS) By John Danias MD PhD and

More information

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma Med. J. Cairo Univ., Vol. 83, No. 2, September: 67-72, 2015 www.medicaljournalofcairouniversity.net Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

More information

A Visual Field Index for Calculation of Glaucoma Rate of Progression

A Visual Field Index for Calculation of Glaucoma Rate of Progression A Visual Field Index for Calculation of Glaucoma Rate of Progression BOEL BENGTSSON AND ANDERS HEIJL PURPOSE: To present a new perimetric index forglaucoma IS A PROGRESSIVE DISEASE THAT calculating the

More information

Characteristics of Frequency-of-Seeing Curves in Normal Subjects, Patients With Suspected Glaucoma, and Patients With Glaucoma

Characteristics of Frequency-of-Seeing Curves in Normal Subjects, Patients With Suspected Glaucoma, and Patients With Glaucoma Characteristics of Frequency-of-Seeing Curves in Normal Subjects, Patients With Suspected Glaucoma, and Patients With Glaucoma Balwantray C. Chauhan* James D. Tompkins,-\ Raymond P. LeBlanc* and Terry

More information

The Effect of Stimulus Size on the Reliable Stimulus Range of Perimetry

The Effect of Stimulus Size on the Reliable Stimulus Range of Perimetry Article DOI: 10.1167/tvst.4.2.10 The Effect of Stimulus Size on the Reliable Stimulus Range of Perimetry Stuart K. Gardiner 1, Shaban Demirel 1, Deborah Goren 1, Steven L. Mansberger 1, and William H.

More information

Fluctuation of Intraocular Pressure and Glaucoma Progression in the Early Manifest Glaucoma Trial

Fluctuation of Intraocular Pressure and Glaucoma Progression in the Early Manifest Glaucoma Trial Fluctuation of Intraocular Pressure and Glaucoma Progression in the Early Manifest Glaucoma Trial Boel Bengtsson, PhD, 1 M. Cristina Leske, MD, MPH, 2 Leslie Hyman, PhD, 2 Anders Heijl, MD, PhD, 1 Early

More information

Visual Field Interpretation Anthony B. Litwak, OD, FAAO VA Medical Center Baltimore, MD

Visual Field Interpretation Anthony B. Litwak, OD, FAAO VA Medical Center Baltimore, MD Visual Field Interpretation Anthony B. Litwak, OD, FAAO VA Medical Center Baltimore, MD Dr. Litwak is on the speaker bureau and advisory panel for Alcon and Zeiss Meditek o Confirms Glaucoma Diagnosis!

More information

Available online at Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6):

Available online at   Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6): Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(6):201-206 ISSN: 0976-8610 CODEN (USA): AASRFC Comparison of glaucoma diagnostic ability of retinal nerve

More information

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Yamanashi Med. J. 18(1), 15~ 20, 2003 Original Article Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Satoshi KOGURE, Yoshiki TODA, Hiroyuki IIJIMA and Shigeo TSUKAHARA Department of

More information

The Evolution of Fundus Perimetry

The Evolution of Fundus Perimetry The Evolution of Fundus Perimetry Company Profile CenterVue designs and manufactures highly automated medical devices for the diagnosis and management of ocular pathologies, including those that represent

More information

Visual Fields: Back to the Future

Visual Fields: Back to the Future Optometry in Practice 2011 Volume 12 Issue 1 11 20 Visual Fields: Back to the Future Tony Redmond 1 BSc PhD MCOptom, and Roger S Anderson 1, 2 BSc MPhil PhD MCOptom 1 NIHR Biomedical Research Centre for

More information

HFA3 with SITA Faster Frequently Asked Questions

HFA3 with SITA Faster Frequently Asked Questions HFA3 with SITA Faster Frequently Asked Questions 1) What is SITA Faster? a. SITA Faster is the newest addition to the SITA family of testing strategies for the Humphrey Field Analyzer 3 (HFA3) perimeter.

More information

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS Glaucoma Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study Jin Wook Jeoung, 1,2 Seok Hwan Kim, 1,3 Ki Ho Park, 1,2

More information

Scanning laser polarimetry (SLP) provides real-time, objective

Scanning laser polarimetry (SLP) provides real-time, objective The Effect of Atypical Birefringence Patterns on Glaucoma Detection Using Scanning Laser Polarimetry with Variable Corneal Compensation Christopher Bowd, Felipe A. Medeiros, Robert N. Weinreb, and Linda

More information

Advances in the Structural Evaluation of Glaucoma with Optical Coherence Tomography

Advances in the Structural Evaluation of Glaucoma with Optical Coherence Tomography Curr Ophthalmol Rep (2013) 1:98 105 DOI 10.1007/s40135-013-0014-4 DIAGNOSIS AND MONITORING OF GLAUCOMA (S SMITH, SECTION EDITOR) Advances in the Structural Evaluation of Glaucoma with Optical Coherence

More information

Fluctuations on the Humphrey and Octopus Perimeters

Fluctuations on the Humphrey and Octopus Perimeters May 987 Vol. 28/ Investigative Ophthalmology & Visual Science A Journal of Dosic and Clinical Research Articles Fluctuations on the and Perimeters Randall S. Drenton and William A. Argus Fluctuation of

More information

Electrodiagnostics Alphabet Soup

Electrodiagnostics Alphabet Soup Nathan Lighthizer, O.D., F.A.A.O Assistant Professor, NSUOCO Chief of Specialty Care Clinics Chief of Electrodiagnostics Clinic What is electrodiagnostics testing? Visual Pathway Basic Understanding VEP

More information

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 1: Categorical Outcomes Based on a Normative Database

More information

Rarebit perimetry and frequency doubling technology in patients with ocular hypertension

Rarebit perimetry and frequency doubling technology in patients with ocular hypertension European Journal of Ophthalmology / Vol. 18 no. 2, 2008 / pp. 205-211 Rarebit perimetry and frequency doubling technology in patients with ocular hypertension G. CORALLO, M. IESTER, R. SCOTTO, G. CALABRIA,

More information

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements 586 Glaucoma Service, Department of Ophthalmology, University of Campinas, Campinas, Brazil R Lauande-Pimentel R A Carvalho H C Oliveira D C Gonçalves L M Silva V P Costa Glaucoma Service, Department of

More information

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note.

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note. Case 1 Normal Tension Glaucoma 53 year old woman attends your practice for routine exam. She has no past medical history or family history of note. Table 1. Right Eye Left Eye Visual acuity 6/6 6/6 Ishihara

More information

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Original Article Philippine Journal of OPHTHALMOLOGY Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Dennis L. del Rosario, MD and Mario M. Yatco, MD University of Santo

More information

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS Quantitative Assessment of Atypical Birefringence Images Using Scanning Laser Polarimetry With Variable Corneal Compensation HARMOHINA BAGGA, MD, DAVID S. GREENFIELD, MD, AND WILLIAM J. FEUER, MS PURPOSE:

More information

eye as a camera Kandel, Schwartz & Jessel (KSJ), Fig 27-3

eye as a camera Kandel, Schwartz & Jessel (KSJ), Fig 27-3 eye as a camera Kandel, Schwartz & Jessel (KSJ), Fig 27-3 retinal specialization fovea: highest density of photoreceptors, aimed at where you are looking -> highest acuity optic disk: cell-free area, where

More information

Introduction. Paulo de Tarso Ponte Pierre-Filho, 1 Rui Barroso Schimiti, 1 Jose Paulo Cabral de Vasconcellos 1 and Vital Paulino Costa 1,2

Introduction. Paulo de Tarso Ponte Pierre-Filho, 1 Rui Barroso Schimiti, 1 Jose Paulo Cabral de Vasconcellos 1 and Vital Paulino Costa 1,2 ACTA OPHTHALMOLOGICA SCANDINAVICA 2006 Sensitivity and specificity of frequency-doubling technology, tendency-oriented perimetry, SITA Standard and SITA Fast perimetry in perimetrically inexperienced individuals

More information

University of Groningen. Glaucoma care optimised in an ageing population Wesselink, Christiaan

University of Groningen. Glaucoma care optimised in an ageing population Wesselink, Christiaan University of Groningen Glaucoma care optimised in an ageing population Wesselink, Christiaan IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from

More information

VISUAL EVOKED POTENTIAL PATTERN ELECTRORETINOGRAM ASSESSMENT OF NEURO VISUAL FUNCTION. Lee Shettle, D.O. Lee Shettle Eye & Hearing

VISUAL EVOKED POTENTIAL PATTERN ELECTRORETINOGRAM ASSESSMENT OF NEURO VISUAL FUNCTION. Lee Shettle, D.O. Lee Shettle Eye & Hearing VISUAL EVOKED POTENTIAL PATTERN ELECTRORETINOGRAM ASSESSMENT OF NEURO VISUAL FUNCTION Lee Shettle, D.O. Lee Shettle Eye & Hearing Overview of VEP / PERG Why should I Use this Test? How does it work? Clinical

More information

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography American Journal of Engineering Research (AJER) e-issn : 2320-0847 p-issn : 2320-0936 Volume-02, Issue-11, pp-104-110 www.ajer.org Research Paper Open Access To assess the glaucoma diagnostic ability of

More information

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Felipe A. Medeiros, Christopher Bowd, Linda M. Zangwill, Chirag Patel, and Robert N. Weinreb From the Hamilton

More information