A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study

Size: px
Start display at page:

Download "A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study"

Transcription

1 A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study ANEESH V. TOLAT, M.D.,* MELISSA WOICIECHOWSKI, M.S.N.,* ROSEMARIE KAHR, R.C.I.S.,* JOSEPH DELL ORFANO, M.D.,* ELLISON BERNS, M.D.,* BRUCE BERNSTEIN, PH.D., and NEAL LIPPMAN, M.D.* From the *Saint Francis Hospital and Medical Center, Section of Electrophysiology, University of Connecticut School of Medicine, Hartford, Connecticut; and St. Christopher s Hospital for Children, Section of General Pediatrics, Drexel University College of Medicine, Philadelphia, Pennsylvania Background: Progress in implantable cardiac defibrillator (ICD) technology has allowed for switching the sensing polarity for the detection of ventricular fibrillation (VF). However, whether one sensing polarity confers additional advantage over the other is not known. Objectives: To determine whether one sensing polarity is superior to the other for the detection of VF. Methods: Patients were enrolled into a prospective randomized study of sensing of VF and R waves in normal rhythm. Sensing of VF was determined by number of under sensed beats (USB), and time to detection of VF (TDVF). Each patient underwent ICD implantation followed by testing of the ICD. At each induction, patients were randomized to sensing in extended bipolar (EBP) or true bipolar (TBP) configuration. Additionally, R waves were compared at implant and at 1-month follow-up. Results: A total of 50 patients were enrolled into the study. When evaluating the primary endpoint, no difference was found between USB in EBP or TBP configuration; 1.1 ± 1.2 beats versus 1.3 ± 1.3 beats; P = NS. Also, no difference was found between TDVF in EBP or TBP configurations; 5.9 ± 0.6 seconds versus 5.9 ± 0.6 seconds; P = NS. With regard to the secondary endpoints, there was no difference between R waves in EBP or TBP configurations at the time of implant 10.9 ± 4.8 mv versus 10.9 ± 4.8 mv P = NS; or at 1-month follow-up 12.4 ± 4.7 mv versus 12.0 ± 5.4 mv P = NS. Conclusions: There is no difference in the detection of VF between EBP or TBP configurations in patients undergoing ICD implantation. (PACE 2013; 36: ) defibrillation ICD, clinical trials Introduction Since the introduction of the implantable cardiac defibrillator (ICD), technology has evolved rapidly to allow additional features to be added. More recent efforts have focused on avoiding inappropriate therapy using supraventricular tachycardia discriminators 1 and other algorithms. 2 Recently, Medtronic (Minneapolis, MN, USA) has provided the option of changing the sensing configuration using either true bipolar sensing (Vtip-Vring) or extended bipolar sensing (Vtip- RVcoil) on their defibrillator platform. This has Dr. Lippman has reported being a speaker for Medtronic. None of the other authors have reported any disclosures relevant to this study. This study was supported in part by an investigator initiated grant from Medtronic. Address for reprints: Aneesh V. Tolat, M.D., F.A.C.C., F.H.R.S., 1000 Asylum Ave., Suite 3206, Hartford, CT Fax: ; atolat@ctheartbeat.com Received September 12, 2012; revised November 10, 2012; accepted December 10, doi: /pace the potential for changing the way defibrillator implants and testing are performed. In addition, it may provide an opportunity to avoid oversensing of T waves and noise by changing polarity, while still maintaining sensitivity. Until recently, sensing in ventricular fibrillation (VF) could only be evaluated in one configuration. It is not clear whether sensing in VF in one configuration is necessarily the same, better, or worse than in another configuration. It is also not known whether the practice of using a longer detection interval to reduce inappropriate shocks may favor one sensing configuration over another. Finally, whether the amplitude of the R wave in sinus rhythm is more likely to change in one configuration versus the other over time is also not known. We performed a prospective randomized study comparing the true bipolar (TBP) and extended bipolar (EBP) sensing configurations for the detection of VF in each patient, and also compared the amplitude of R waves in normal sinus rhythm at implant and at 1-month followup. 2013, The Authors. Journal compilation 2013 Wiley Periodicals, Inc. PACE, Vol. 36 May

2 TOLAT, ET AL. Table I. Baseline Characteristics Demographic and Clinical Data N = 50 Patients Age-mean ± SD 63 ± 13 Male 35 (70%) History of CHF 45 (90%) History of MI 24 (48%) Cardiomyopathy Ischemic 29 (58%) Nonischemic 21 (42%) Primary prevention 41(82%) CHF = congestive heart failure; MI = myocardial infarction; SD = standard deviation. Methods Patient Population Our study enrolled 50 consecutive patients who were referred to the electrophysiology lab for implantation of a Medtronic ICD (Medtronic Inc., Minneapolis, MN, USA). All patients provided informed consent and participated in an institutional review board-approved protocol. Demographic and key clinical data are shown in Table I. Patients underwent implant of Medtronic ICD generator models Virtuoso (VR or DR) or Concerto (cardiac resynchronization therapydefibrillator), with lead model 6947 (Sprint Quattro, quadripolar, true bipolar, dual coil). All leads were placed at the right ventricular (RV) apical septum as prespecified in the protocol. Inclusion criteria included: age >18 with ability to give informed consent, subjects expected to live more than 1 year, with indication for defibrillator implant, and who were not pacemaker dependent. Patients were required to have either ischemic or nonischemic cardiomyopathy. Exclusion criteria included: the presence of preexisting leads, any condition which would preclude ICD testing at the end of the implant, and patients with inherited arrhythmias or ion channel-related arrhythmias. Study Protocol Patients who were enrolled into the study underwent standard ICD implantation followed by ICD testing at the time of implant. Testing was required in all patients with two tests performed for each patient unless in the opinion of the operator testing was not deemed to be safe for the patient. If only one test was performed, data for comparison between sensing in EBP and TBP configurations was not available and excluded from analysis. Patients were randomized to testing in one of two ways: (1) sensing of first induced VF episode in TBP followed by second induced VF in EBP, or (2) sensing of first induced VF episode in EBP, followed by second induced VF in TBP. VF was induced by either T wave shock or 50-Hz burst pacing, or both. Sensitivity on the RV ICD lead was set to 1.2 mv during testing, and the number of intervals to detect (NID) was programmed to 24/32 for all patients and both induction sequences. Final ICD programming for sensing vector (EBP or TBP) was required by protocol to be set the same as the first induction configuration unless the operator deemed the other configuration necessary for patient safety reasons. The number of beats undersensed by the device was counted from the first sensed beat to the declaration of VF detection by the device (Fig. 1). The number of undersensed beats was counted by two independent readers blinded to the sensing polarity. When disagreement existed about the number of undersensed beats, a third (blinded) experienced electrophysiologist then served to adjudicate the final number. R wave amplitudes were measured as reported by the device in both configurations at the time of implant prior to ICD testing, and at 1-month follow-up through the device reported sensing test. The primary endpoints of the study were prespecified as number of undersensed beats in VF, and time to detection of VF. Secondary endpoints of the study were sensed R wave amplitude in normal sinus rhythm or atrial fibrillation at the time of implant and at 1-month follow-up. Statistical Methods Data were evaluated and analyzed using SPSS software (IBM Corp., Armonk, NY, USA). Values of P < 0.05 were considered statistically significant. The paired T test was used to evaluate variables such as number of undersensed beats, time to VF detection, and R wave amplitudes while Pearson correlation was used to assess the relationship between decreasing ejection fraction (EF), and increased numbers of undersensed beats. Results Fifty patients were enrolled into the study, while 48 patients completed intraoperative testing, and 42 patients completed 1-month followup. Devices implanted by type included 26 single chamber, seven dual chamber, and 17 biventricular ICDs. Baseline characteristics of the study population can be found in Table I. With regard to the primary endpoints of the study, no significant difference was found between sensing of VF as measured by number of undersensed beats or time to detection of VF (Table II). During testing 542 May 2013 PACE, Vol. 36

3 THE PROPR STUDY Figure 1. Examples of differences in sensing of VF in true bipolar and extended bipolar configurations. In panel A, study patient #13 is shown undergoing defibrillation threshold (DFT) testing with undersensing seen in the true bipolar configuration (TBP). The same patient is seen in panel B undergoing a separate DFT test in extended bipolar configuration (EBP) with no undersensing. Panel C shows DFT testing in patient #27 showing fewer undersensed beats in the TBP configuration, as compared to a greater number of undersensed beats in the EBP configuration during a separate DFT test in the same patient. of VF detection at a ventricular sensitivity of 1.2 mv, the number of undersensed beats in EBP was Table II. Results of Testing of R Wave Sensing in VF and Sinus Rhythm Number of drop outs (beats) Time to VF detection (second) Mean R wave implant Mean R wave follow-up TBP EBP 1.3 ± ± 1.2 P = NS 5.9 ± ± 0.6 P = NS 10.9 ± ± 4.8 P = NS 12.0 ± ± 4.7 P = NS EBP = extended bipolar; TBP = true bipolar; VF = ventricular fibrillation. 1.1 ± 1.2 beats, while in TBP was 1.3 ± 1.3 beats; P = NS. The time to detection of VF in EBP configuration was 5.9 ± 0.6 second, and in TBP configuration was 5.9 ± 0.6 second; P = NS. With regard to secondary endpoints, no significant difference was found between the mean sensed R wave amplitude in TBP 10.9 ± 4.8 mv as compared to EBP 10.9 ± 4.8 mv configurations at the time of implant, or at 1-month follow-up; 12.0 ± 5.4 mv and 12.4 ± 4.7 mv, respectively (P = NS). There was a significant increase in mean R wave amplitude between EBP implant 10.5 mv and EBP follow-up 12.4 mv (P = 0.004) and between TBP implant 10.6 mv and TBP follow-up 12.0 mv (P = 0.033). We also evaluated the association between time to detection of VF and EF, and found a significant correlation between decreasing EF and increased time to detection of VF in TBP configuration; P = (Fig. 2). This trend was also seen in EBP configuration but did not reach statistical significance; P = PACE,Vol.36 May

4 TOLAT, ET AL. Figure 2. Shown above is the time to detection of VF as a function of EF. A significant (P=0.014) increase in time to detection is observed as EF decreases in the TBP configuration. Finally, we evaluated patients with a drop in R waves between implant and 1-month follow-up in TBP or EBP configurations to assess whether the alternative configuration had a similar effect. Of those patients who had a drop in R wave at followup, 26.2% (11 patients) occurred in the TBP configuration, while 26.8% (10 patients) occurred in the EBP configuration. A strong correlation of (P = 0.001) shown in Fig. 3, was observed between change in R waves (implant to follow-up) in TBP and EBP configurations. Discussion In our study of EBP versus TBP sensing of VF at the time of implant, we did not find a significant difference between either configuration as measured by the endpoints of mean number of undersensed beats or time to detection of VF (Table II). This finding has several important clinical implications. First, ICD testing can be performed in either configuration with an equal probability that the device will detect VF without undersensing or increased time to detection of VF. One recent smaller study of 28 patients reported a similar evaluation of EBP and TBP sensing in VF. 3 In the study, the investigators did not find a significant difference between EBP and TBP for the sensing of VF. However, the authors only measured time to detection of VF, and used an NID of 12/16. In contrast, our study used time to detection of VF and number of undersensed beats with an increased NID of 24/32, which still did not favor one configuration over another. We believe our results can be applied to patients with Figure 3. Relationship between change in R waves from implant to follow-up and sensing configuration. EBP FU initial indicates the extended bipolar R wave in follow-up minus implanted R wave; and TBP FU initial indicates true bipolar R wave in follow-up minus implanted R wave. A significant correlation of (P = 0.001) was found between change in R waves (implant to follow-up) and the two configurations. 544 May 2013 PACE, Vol. 36

5 THE PROPR STUDY a shorter NID because any difference in sensing of VF is less likely to be seen with a shorter NID. One important caveat however is that our testing was performed at 1.2 mv sensitivity favoring undersensing, while final programming is usually performed at 0.3 mv. Finally, while the PREPARE study 1 did show a reduction in arrhythmia-related shock therapy with an NID of 30/40 in primary prevention patients we chose to use a NID of 24/32 in our study because of the involvement of both primary and secondary prevention patients. At 1-month follow-up, we also did not see any oversensing occurring in either TBP or EBP sensing polarity. While the EBP sensing polarity may theoretically allow for more sensing of diaphragmatic myopotentials or T waves because of an increased electrode distance, this was not seen clinically. This may have been related to the small sample size and infrequent nature of this type of event. Other older studies with older leads and pulse generators have also shown a lack of difference between sensing of VF in TBP and EBP configurations. In a study by Goldberger et al., 4 the investigators compared simultaneous sensing in VF of both configurations at the time of ICD implant using a Transvene 6936 lead and found no significant difference between time to detection of VF or the number of undersensed beats, consistent with our findings. Interestingly, another older study did appear to find a difference between TBP and EBP sensing during redetection, after a failed first shock, favoring TBP sensing. 5 We were not able to confirm this observation in our study because we did not record simultaneously both TBP and EBP configurations, and the majority of our shocks were successful on first attempt. Another clinical implication of our study is that implanting physicians should not expect to see a difference in R waves between the two configurations in normal rhythm at the time of implant or at 1-month follow-up. This is in keeping with previous studies which also have not seen a difference in R waves. 3,5 There was, however, a significant increase in R waves between the time of implant and at follow-up, which is consistent with the generally accepted observation that the sensed electrogram usually becomes larger in follow-up. Time to Detection and EF There was a significant correlation between increased time to detection of VF, and reduced EF in the TBP configuration (Fig. 2). A similar trend was seen in the EBP configuration but was not statistically significant. To our knowledge, this observation has not been reported previously and deserves further investigation. There may be several explanations for this including that patients with lower EFs may have finer lower amplitude VF, making it more difficult for the ICD to detect; however, this would need to be evaluated in future studies. Recommendations for ICD Programming Given the results of our study, and more recent trends in ICD implantation, including the foregoing of ICD testing in primary prevention implantations, it would seem reasonable to program patients to either configuration if R waves were within acceptable range (>5 mv) in both configurations. In addition, if a patient experiences a drop in R waves in follow-up in one configuration, it may seem reasonable to reprogram to the other configuration, if the other configuration shows more robust R waves. Followup testing of the ICD may or may not need to be performed depending on the discretion of the operator, and clinical history of the patient, including whether the patient requires the ICD for primary or secondary prevention. In the future, ICD algorithms will continue to evolve which have the ability to switch sensing polarity automatically depending on the size of the R waves, presence of noise, T wave oversensing, and other factors. Our study would suggest that this could be performed since there was no significant difference overall between sensing in VF in either configuration. However, individual cases may vary and further studies would be required to confirm this. Finally, for those implanters who prefer to test defibrillators at the time of implant, testing both configurations may be a novel way of leaving the option of switching polarity without having to return to the EP lab. Limitations There are several limitations to our study that should be noted. First, our study was relatively small. In addition, the study only included patients without previous implants, and with either an ischemic or nonischemic cardiomyopathy. We did not record both TBP and EBP configurations simultaneously, but rather sequentially after testing in the opposite configuration. To reduce the chance of bias, however, we did randomize the sensing configuration tested during VF induction. Conclusions In conclusion, there was no difference in the detection of VF between EBP or TBP configurations as measured by undersensed beats, and time to detection of VF in patients with cardiomyopathy undergoing primary or secondary prevention ICD implantation. PACE,Vol.36 May

6 TOLAT, ET AL. References 1. Wilkoff BL, Williamson BD, Stern RS, Moore SL, Lu F, Lee SW, Birgersdotter-Green UM, et al. PREPARE Study Investigators. Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients: Results from the PREPARE (Primary Prevention Parameters Evaluation) study. J Am Coll Cardiol 2008; 52: Cao J, Gillberg JM, Swerdlow CD. A fully automatic, implantable cardioverter-defibrillator algorithm to prevent inappropriate detection of ventricular tachycardia or fibrillation due to T-wave oversensing in spontaneous rhythm. Heart Rhythm 2012; 9: Verga TA, Gillberg JM, Greenberg RM, Deger FT. True bipolar and integrated bipolar sensing and detection by implantable defibrillators. Pacing Clin Electrophysiol 2011; 34: Goldberger J, Horvath G, Donovan D, Johnson D, Challapalli R, Kadish AH. Detection of ventricular fibrillation by transvenous defibrillating leads: Integrated versus dedicated bipolar Ssensing. J Cardiovasc Electrophysiol 1998; 9: Frain B, Ellison K, Michaud G Koo CH, Buxton AE, Kirk MM. True bipolar defibrillator leads have increased sensing latency and threshold compared with the integrated bipolar configuration. J Cardiovasc Electrophysiol 2007; 18: May 2013 PACE, Vol. 36

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

Shock Reduction Strategies Michael Geist E. Wolfson MC

Shock Reduction Strategies Michael Geist E. Wolfson MC Shock Reduction Strategies Michael Geist E. Wolfson MC Shock Therapy Thanks, I needed that! Why Do We Need To Reduce Shocks Long-term outcome after ICD and CRT implantation and influence of remote device

More information

Dual-Chamber Implantable Cardioverter-Defibrillator

Dual-Chamber Implantable Cardioverter-Defibrillator February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary

More information

Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention

Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Dr Ngai-Yin Chan, MBBS, FRCP(Lond), FRCP(Edin), FRCP(Glasg), FACC, FHRS, Consultant Physician, Department of Medicine & Geriatrics, Princess

More information

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators 22 March 2002 Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators F. HINTRINGER, O. PACHINGER Division of Cardiology, Department for Internal Medicine, University

More information

Teaching Rounds in Cardiac Electrophysiology

Teaching Rounds in Cardiac Electrophysiology Teaching Rounds in Cardiac Electrophysiology Sustained Multiple Railroad Tracks on Implantable Cardiac Defibrillator Interval Plots Mechanisms and Management Alex Y. Tan, MD; Kenneth A. Ellenbogen, MD;

More information

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Programming of Bradycardic Parameters C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Programming of ICD Brady Parameters Conflict of Interest Biotronik

More information

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Device Interrogation- Pacemakers, ICD and Loop Recorders Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Disclosures Consultant: Medtronic Speaker s Bureau: St. Jude Medical

More information

GDT1000 SENSING ACUTE STUDY

GDT1000 SENSING ACUTE STUDY CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY GDT1000 SENSING ACUTE STUDY Boston

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Tseng ZH, Hayward RM, Clark NM, et al. Sudden death in patients with cardiac implantable electronic devices. JAMA Intern Med. Published online June 22, 2015. doi:10.1001/jamainternmed.2015.2641.

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Implantable cardioverter defibrillators for the treatment of arrhythmias and cardiac resynchronisation therapy for the treatment of heart failure (review

More information

Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm

Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm Guy Amit, MD, MPH Soroka University Medical Center Ben-Gurion University of the Negev Beer-Sheva, Israel Disclosures Consultant:

More information

The Nuts and Bolts of ICD Therapy

The Nuts and Bolts of ICD Therapy Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction

More information

INNOVATIONS IN DEVICE THERAPY:

INNOVATIONS IN DEVICE THERAPY: INNOVATIONS IN DEVICE THERAPY: Subcutaneous ICDs, Leadless Pacemakers, CRT Indications David J Wilber MD Loyola University Medical Center Disclosures: ACC Foundation: Consultant; Biosense / Webster: Consultant,

More information

Defibrillation testing should be routinely performed at the time of implantable cardioverter-defibrillator implantation Contra

Defibrillation testing should be routinely performed at the time of implantable cardioverter-defibrillator implantation Contra Defibrillation testing should be routinely performed at the time of implantable cardioverter-defibrillator implantation Contra Fernando Arribas Cardiology Service Hospital 12 de Octubre Madrid Spain D.O.I.

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

MEDICAL POLICY Cardioverter Defibrillators

MEDICAL POLICY Cardioverter Defibrillators POLICY........ PG-0224 EFFECTIVE......06/01/09 LAST REVIEW... 01/27/17 MEDICAL POLICY Cardioverter Defibrillators GUIDELINES This policy does not certify benefits or authorization of benefits, which is

More information

Tachycardia Devices Indications and Basic Trouble Shooting

Tachycardia Devices Indications and Basic Trouble Shooting Tachycardia Devices Indications and Basic Trouble Shooting Peter A. Brady, MD., FRCP Cardiology Review Course London, March 6 th, 2014 2011 MFMER 3134946-1 Tachycardia Devices ICD Indications Primary and

More information

A 'shocking' case rectified

A 'shocking' case rectified A 'shocking' case rectified Journal: Pacing and Clinical Electrophysiology Manuscript ID: PACE-13-00.R1 Wiley - Manuscript type: Device Rounds Date Submitted by the Author: n/a Complete List of Authors:

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Defibrillation threshold testing should no longer be performed: contra

Defibrillation threshold testing should no longer be performed: contra Defibrillation threshold testing should no longer be performed: contra Andreas Goette St. Vincenz-Hospital Paderborn Dept. of Cardiology and Intensive Care Medicine Germany No conflict of interest to disclose

More information

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012 IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT201203 JANUARY 24, 2012 The IHCP to reimburse implantable cardioverter defibrillators separately from outpatient implantation Effective March 1, 2012, the

More information

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm April 2000 107 Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm B. MERKELY Semmelweis University, Dept. of Cardiovascular Surgery,

More information

Implantable Cardioverter Defibrillator Therapy in MADIT II Patients with Signs and Symptoms of Heart Failure

Implantable Cardioverter Defibrillator Therapy in MADIT II Patients with Signs and Symptoms of Heart Failure Implantable Cardioverter Defibrillator Therapy in MADIT II Patients with Signs and Symptoms of Heart Failure Wojciech Zareba Postinfarction patients with left ventricular dysfunction are at increased risk

More information

Continuous ECG telemonitoring with implantable devices: the expected clinical benefits

Continuous ECG telemonitoring with implantable devices: the expected clinical benefits Continuous ECG telemonitoring with implantable devices: the expected clinical benefits C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Declaration of

More information

Advances in Arrhythmia and Electrophysiology

Advances in Arrhythmia and Electrophysiology Advances in Arrhythmia and Electrophysiology Troubleshooting Implanted Cardioverter Defibrillator Sensing Problems I Charles D. Swerdlow, MD; Samuel J. Asirvatham, MD; Kenneth A. Ellenbogen, MD; Paul A.

More information

The Immediate Reproducibility of T Wave Alternans During Bicycle Exercise

The Immediate Reproducibility of T Wave Alternans During Bicycle Exercise Reprinted with permission from JOURNAL OF PACING AND CLINICAL ELECTROPHYSIOLOGY, Volume 25, No. 8, August 2002 Copyright 2002 by Futura Publishing Company, Inc., Armonk, NY 10504-0418. The Immediate Reproducibility

More information

Strategic Programming of Detection and Therapy Parameters in Implantable Cardioverter-Defibrillators Reduces Shocks in Primary Prevention Patients

Strategic Programming of Detection and Therapy Parameters in Implantable Cardioverter-Defibrillators Reduces Shocks in Primary Prevention Patients Journal of the American College of Cardiology Vol. 52, No. 7, 2008 2008 by the American College of Cardiology Foundation ISSN 0735-1097/08/$34.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2008.05.011

More information

PARAD/PARAD+ : P and R Based Arrhythmia Detection

PARAD/PARAD+ : P and R Based Arrhythmia Detection Tech Corner PARAD/PARAD+ : P and R Based Arrhythmia Detection NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN

More information

Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis

Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis Investigator: Keiko Saito, MD Mentor: Yuji Saito, MD, PhD, FACP, FACC Department

More information

Diagnostic Value of Single Versus Dual Chamber Electrograms Recorded from an Implantable Defibrillator

Diagnostic Value of Single Versus Dual Chamber Electrograms Recorded from an Implantable Defibrillator Journal of Interventional Cardiac Electrophysiology 9, 49 53, 2003 C 2003 Kluwer Academic Publishers. Manufactured in The Netherlands. Diagnostic Value of Single Versus Dual Chamber Electrograms Recorded

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

Cardiac Devices CRT,ICD: Who is and is not a Candidate? Who Decides

Cardiac Devices CRT,ICD: Who is and is not a Candidate? Who Decides Cardiac Devices CRT,ICD: Who is and is not a Candidate? Who Decides Colette Seifer MB(Hons) FRCP(UK) Associate Professor, University of Manitoba, Cardiologist, Cardiac Sciences Program, St Boniface Hospital

More information

Summary, conclusions and future perspectives

Summary, conclusions and future perspectives Summary, conclusions and future perspectives Summary The general introduction (Chapter 1) of this thesis describes aspects of sudden cardiac death (SCD), ventricular arrhythmias, substrates for ventricular

More information

Critical Analysis of Dual-Chamber Implantable Cardioverter-Defibrillator Arrhythmia Detection

Critical Analysis of Dual-Chamber Implantable Cardioverter-Defibrillator Arrhythmia Detection Critical Analysis of Dual-Chamber Implantable Cardioverter-Defibrillator Arrhythmia Detection Results and Technical Considerations Bruce L. Wilkoff, MD; Volker Kühlkamp, MD; Kent Volosin, MD; Kenneth Ellenbogen,

More information

Automatic External Defibrillators

Automatic External Defibrillators Last Review Date: April 21, 2017 Number: MG.MM.DM.10dC3v4 Medical Guideline Disclaimer Property of EmblemHealth. All rights reserved. The treating physician or primary care provider must submit to EmblemHealth

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

Diaphragmatic myopotential oversensing in pacemaker-dependent patients with CRT-D devices

Diaphragmatic myopotential oversensing in pacemaker-dependent patients with CRT-D devices Europace (2008) 10, 1381 1386 doi:10.1093/europace/eun241 Diaphragmatic myopotential oversensing in pacemaker-dependent patients with CRT-D devices Katya Reis Santos*, Pedro Adragão, Diogo Cavaco, Francisco

More information

Inappropriate electrical shocks: Tackling the beast

Inappropriate electrical shocks: Tackling the beast ESC Paris 2011 Inappropriate electrical shocks: Tackling the beast Gerhard Hindricks University of Leipzig Heart Center Dept. of Electrophysiology ESC Paris 2011 Inappropriate electrical shocks: Tackling

More information

Early Recurrence of Atrial Fibrillation After Ambulatory Shock Conversion

Early Recurrence of Atrial Fibrillation After Ambulatory Shock Conversion Journal of the American College of Cardiology Vol. 40, No. 1, 2002 2002 by the American College of Cardiology Foundation ISSN 0735-1097/02/$22.00 Published by Elsevier Science Inc. PII S0735-1097(02)01912-5

More information

ICD in a young patient with syncope

ICD in a young patient with syncope ICD in a young patient with syncope Konstantinos P. Letsas, MD, FESC Second Department of Cardiology Evangelismos General Hospital of Athens Athens, Greece Case presentation A 17-year-old apparently healthy

More information

Subcutaneous Implantable Cardioverter Defibrillator (S-ICD)

Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) D. D. MANOLATOS, MD, PhD, FESC Electrophysiology and Device Lab General Hospital Evangelismos, Athens The Problem: 300,000 people die each year

More information

Title: Automatic External Defibrillators Division: Medical Management Department: Utilization Management

Title: Automatic External Defibrillators Division: Medical Management Department: Utilization Management Retired Date: Page 1 of 7 1. POLICY DESCRIPTION: Automatic External Defibrillators 2. RESPONSIBLE PARTIES: Medical Management Administration, Utilization Management, Integrated Care Management, Pharmacy,

More information

Introduction. CLINICAL RESEARCH Clinical Trial Design. Mohammad Saeed 1 *, Mehdi Razavi 1, Curtis G. Neason 2, and Simona Petrutiu 2. Aims.

Introduction. CLINICAL RESEARCH Clinical Trial Design. Mohammad Saeed 1 *, Mehdi Razavi 1, Curtis G. Neason 2, and Simona Petrutiu 2. Aims. Europace (2011) 13, 1648 1652 doi:10.1093/europace/eur195 CLINICAL RESEARCH Clinical Trial Design Rationale and design for programming implantable cardioverter defibrillators in patients with primary prevention

More information

Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report

Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report Cardiol Ther (2017) 6:139 143 DOI 10.1007/s40119-017-0086-6 CASE REPORT Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report Abhishek Shenoy. Abhishek

More information

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features Definition of SVT Discrimination Identify which patient populations might benefit from these features Understand the 4 types of SVT discriminators used by St Jude Medical Be aware of programmable parameters

More information

MADIT Studies: CRT in the Non-LBBB Patient and Other Findings. Arthur J. Moss, MD

MADIT Studies: CRT in the Non-LBBB Patient and Other Findings. Arthur J. Moss, MD MADIT Studies: CRT in the Non-LBBB Patient and Other Findings Arthur J. Moss, MD University of Rochester Medical Cntr. Rochester, NY 13 th Annual New Frontiers in Heart Failure Rx NYC, NY January 26, 2013

More information

Advanced ICD Troubleshooting: Part I REVIEW

Advanced ICD Troubleshooting: Part I REVIEW REVIEW Advanced ICD Troubleshooting: Part I CHARLES D. SWERDLOW* and PAUL A. FRIEDMAN *From the Cedars-Sinai Medical Center, Los Angeles, California, and Mayo Clinic, Rochester, Minnesota (PACE 2005;28:1322

More information

New generations pacemakers and ICDs: an update

New generations pacemakers and ICDs: an update Advances in Cardiac Arrhythmias and Great Innovations in Cardiology XXVII Giornate Cardiologiche Torinesi New generations pacemakers and ICDs: an update Prof. Fiorenzo Gaita, MD Division of Cardiology

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support

A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support 100 April 2000 A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support A. SH. REVISHVILI A.N. Bakoulev Institute of Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow,

More information

Ventricular tachycardia Ventricular fibrillation and ICD

Ventricular tachycardia Ventricular fibrillation and ICD EKG Conference Ventricular tachycardia Ventricular fibrillation and ICD Samsung Medical Center CCU D.I. Hur Ji Won 2006.05.20 Ventricular tachyarrhythmia ventricular tachycardia ventricular fibrillation

More information

Novel Approaches to VT Management Glenn M Polin MD

Novel Approaches to VT Management Glenn M Polin MD Novel Approaches to VT Management Glenn M Polin MD Medical Director, Electrophysiology Laboratory John Ochsner Heart and Vascular Institute New Orleans, LA Disclosures Pfizer Speaker Bureau Bristol Myers

More information

Panagiotis N. Margos MD, Rolf Schomburg MD, Jorg Kynast MD, Ahmed A. Khattab MD, Gert Richardt MD.

Panagiotis N. Margos MD, Rolf Schomburg MD, Jorg Kynast MD, Ahmed A. Khattab MD, Gert Richardt MD. www.ipej.org 64 Case Report Termination of Ventricular Tachycardia with Antitachycardia Pacing after Ineffective Shock Therapy in an ICD Recipient with Hypertrophic Cardiomyopathy Panagiotis N. Margos

More information

Biphasic Clinical Summaries

Biphasic Clinical Summaries Biphasic Clinical Summaries Defibrillation of Ventricular Fibrillation and Ventricular Tachycardia Background Physio-Control conducted a multi-centered, prospective, randomized and blinded clinical trial

More information

A Closer Look Product Education at a glance

A Closer Look Product Education at a glance A Closer Look Product Education at a glance an ICD or CRT-D When Atrial Information is Not Used SUMMARY Boston Scientific dual-chamber ICDs and multi-chamber CRT-Ds will respond to atrial sensing regardless

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

POLICIES AND PROCEDURE MANUAL

POLICIES AND PROCEDURE MANUAL POLICIES AND PROCEDURE MANUAL Policy: MP140 Section: Medical Benefit Policy Subject: Automatic Implantable Cardioverter-Defibrillator I. Policy: Automatic Implantable Cardioverter-Defibrillator II. Purpose/Objective:

More information

ENABLE MRI CLINICAL STUDY SUMMARY

ENABLE MRI CLINICAL STUDY SUMMARY CLINICAL STUDY SUMMARY ENABLE MRI CAUTION: Federal Law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. TABLE OF CONTENTS

More information

HRC Carole Joyce. Bradford Royal Infirmary. Senior Chief Cardiac Physiologist. Pacing & Invasive Services.

HRC Carole Joyce. Bradford Royal Infirmary. Senior Chief Cardiac Physiologist. Pacing & Invasive Services. HRC 2010 Carole Joyce Carole.joyce@bradfordhospitals.nhs.uk Pacing & Invasive Services Carole Joyce Introduction Basic Protocol Troubleshooting Case Studies Questions & Answer Carole Joyce ICD Troubleshooting

More information

Primary prevention ICD recipients: the need for defibrillator back-up after an event-free first battery service-life

Primary prevention ICD recipients: the need for defibrillator back-up after an event-free first battery service-life Chapter 3 Primary prevention ICD recipients: the need for defibrillator back-up after an event-free first battery service-life Guido H. van Welsenes, MS, Johannes B. van Rees, MD, Joep Thijssen, MD, Serge

More information

SUBCUTANEOUS IMPLANTABLE CARDIOVERTER DEFIBRILLATORS

SUBCUTANEOUS IMPLANTABLE CARDIOVERTER DEFIBRILLATORS SUBCUTANEOUS IMPLANTABLE CARDIOVERTER DEFIBRILLATORS Protocol: CAR044 Effective Date: December 1, 2017 Table of Contents Page COMMERCIAL & MEDICAID COVERAGE RATIONALE... 1 MEDICARE COVERAGE RATIONALE...

More information

Troubleshooting ICD. NASPE Training Lancashire & South Cumbria Cardiac Network

Troubleshooting ICD. NASPE Training Lancashire & South Cumbria Cardiac Network Troubleshooting ICD NASPE Training Lancashire & South Cumbria Cardiac Network Bradycardia Pacing by ICD Isolated ICD discharge Multiple discharges Appropriate Inappropriate No ICD Therapy or delay in ICD

More information

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART December 2015 This chart encompasses all Medtronic cardiac devices FDA-Approved as MR Conditional and included in the MRI SureScan portfolio. If a device

More information

Need to Know: Implantable Devices. Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia

Need to Know: Implantable Devices. Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia Need to Know: Implantable Devices Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia Disclosure Statement I have no relationships to disclose. Objectives Discuss the most

More information

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia By Sandeep Joshi, MD and Jonathan S. Steinberg, MD Arrhythmia Service, Division of Cardiology

More information

Selecting the Best ICD for your Patient-SICD v. TV. Martin C. Burke DO, FACOI CorVita Science Foundation & Academic Medical Center, Amsterdam

Selecting the Best ICD for your Patient-SICD v. TV. Martin C. Burke DO, FACOI CorVita Science Foundation & Academic Medical Center, Amsterdam Selecting the Best ICD for your Patient-SICD v. TV Martin C. Burke DO, FACOI CorVita Science Foundation & Academic Medical Center, Amsterdam COI DISCLOSURES I have received lecture and proctoring honoraria

More information

Tachyarrhythmia Suspicion and Detection

Tachyarrhythmia Suspicion and Detection Tech Corner Tachyarrhythmia Suspicion and Detection NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN THE ARTICLE.

More information

CONTAK RENEWAL 3 AVT CLINICAL SUMMARY

CONTAK RENEWAL 3 AVT CLINICAL SUMMARY CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY CONTAK RENEWAL 3 AVT Boston Scientific

More information

ICD leads survival and troubles in the last 10 years

ICD leads survival and troubles in the last 10 years XVI Symposium on Progress in Clinical Pacing, Rome, Italy - Dec 2 to 7, 2014 ICD leads survival and troubles in the last 10 years Antonio Raviele, MD, FESC, FHRS President ALFA Alliance to Fight Atrial

More information

Clinical and Electrocardiographic Characteristics of Patients with Brugada Syndrome: Report of Five Cases of Documented Ventricular Fibrillation

Clinical and Electrocardiographic Characteristics of Patients with Brugada Syndrome: Report of Five Cases of Documented Ventricular Fibrillation J Arrhythmia Vol 25 No 1 2009 Original Article Clinical and Electrocardiographic Characteristics of Patients with Brugada Syndrome: Report of Five Cases of Documented Ventricular Fibrillation Seiji Takashio

More information

Troubleshooting pacemakers and implantable cardioverter-defibrillators David L. Scher

Troubleshooting pacemakers and implantable cardioverter-defibrillators David L. Scher Troubleshooting pacemakers and implantable cardioverter-defibrillators David L. Scher Purpose of review The purpose of this review is to provide an update on stored diagnostic information furnished by

More information

Device Diagnostics and Stroke Prevention: State of the Art

Device Diagnostics and Stroke Prevention: State of the Art Device Diagnostics and Stroke Prevention: State of the Art Ass. Prof. Dr. Carsten W. Israel Dept. of Cardiology Evangelical Hospital Bielefeld Germany C.W.Israel@em.uni-frankfurt.de Disclosure of Relationships

More information

Journal of Arrhythmia

Journal of Arrhythmia Journal of Arrhythmia 28 (2012) 91 95 Contents lists available at SciVerse ScienceDirect Journal of Arrhythmia journal homepage: www.elsevier.com/locate/joa Review Unresolved matters related to implantable

More information

CLINICAL SUMMARY COGENT-4 FIELD FOLLOWING STUDY

CLINICAL SUMMARY COGENT-4 FIELD FOLLOWING STUDY CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY COGENT-4 FIELD FOLLOWING STUDY

More information

How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France

How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France Presenter Disclosure Information Christophe Leclercq, MD,

More information

Practice Questions.

Practice Questions. IBHRE Prep Practice Questions Question 1 The relative refractory yperiod of the ventricular myocardium corresponds to which of the following phases of the action potential? A. (0) B. (1) C. (2) D. (3)

More information

Upgrade to Resynchronization Therapy. Saeed Oraii MD, Cardiologist Interventional Electrophysiologist Tehran Arrhythmia Clinic May 2016

Upgrade to Resynchronization Therapy. Saeed Oraii MD, Cardiologist Interventional Electrophysiologist Tehran Arrhythmia Clinic May 2016 Upgrade to Resynchronization Therapy Saeed Oraii MD, Cardiologist Interventional Electrophysiologist Tehran Arrhythmia Clinic May 2016 Event Free Survival (%) CRT Cardiac resynchronization therapy (CRT)

More information

CONSULTA CRT-P Model C4TR01

CONSULTA CRT-P Model C4TR01 CONSULTA CRT-P Model C4TR01 Physical characteristics Volume a 15 cm 3 Mass H x W x D Surface area of titanium device can Surface area of each titanium nitride LECG electrode Radiopaque ID Materials in

More information

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds

More information

Case Description A 25 years old male served in a combat unit for 3 years (GOLANY) Implantation of Dual chamber ICD in 7/2010 due to Severe Non ischemi

Case Description A 25 years old male served in a combat unit for 3 years (GOLANY) Implantation of Dual chamber ICD in 7/2010 due to Severe Non ischemi ICD Case Study tracing your own images Prof Glikson, Sandra Schor-Wider Nadav Hayman Case Description A 25 years old male served in a combat unit for 3 years (GOLANY) Implantation of Dual chamber ICD in

More information

Advanced ICD Troubleshooting: Part II REVIEW

Advanced ICD Troubleshooting: Part II REVIEW REVIEW Advanced ICD Troubleshooting: Part II CHARLES D. SWERDLOW* and PAUL A. FRIEDMAN From the *Cedars-Sinai Medical Center, Los Angeles, California, and Mayo Clinic, Rochester, Minnesota, USA Failure

More information

Review of Pacemakers and ICD Therapy: Overview and Patient Management

Review of Pacemakers and ICD Therapy: Overview and Patient Management Review of Pacemakers and ICD Therapy: Overview and Patient Management Pacing Systems Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services OSU Division

More information

12 YEARS PROVEN PERFORMANCE. Sprint Quattro Defibrillation Leads

12 YEARS PROVEN PERFORMANCE. Sprint Quattro Defibrillation Leads 12 YEARS PROVEN PERFORMANCE Sprint Quattro Defibrillation Leads PROVEN PERFORMANCE. PROVEN RELIABLE. We made intentional choices and took proactive steps from lead design to active product monitoring to

More information

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Pacing and Device Jargon Made Simple Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Disclosure Dr Jonathan Timperley Honorarium from Bayer Arne Larson (1915

More information

A Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators

A Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators Journal of the American College of Cardiology Vol. 48, No. 2, 2006 2006 by the American College of Cardiology Foundation ISSN 0735-1097/06/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.03.037

More information

Is This Thing Working?

Is This Thing Working? Is This *#@!* Thing Working? Pacemaker (and ICD) ECG and Telemetry Pitfalls Wayne O. Adkisson, MD adki0004@umn.edu Disclosures I currently receive research support from Medtronic, Inc. I have been compensated

More information

ICD THERAPIES: are they harmful or just high risk markers?

ICD THERAPIES: are they harmful or just high risk markers? ICD THERAPIES: are they harmful or just high risk markers? Konstantinos P. Letsas, MD, PhD, FESC LAB OF CARDIAC ELECTROPHYSIOLOGY EVANGELISMOS GENERAL HOSPITAL ATHENS ICD therapies are common In a meta-analysis

More information

Biventricular Pacemakers Sensing

Biventricular Pacemakers Sensing Biventricular Pacemakers Sensing Sergio Dubner Director Arrhythmias and Electrophisiology Service, Clinica y Maternidad Suizo-Argentina, Buenos Aires, Argentina Left ventricular pacemaker lead technology

More information

MEDICAL POLICY SUBJECT: IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

MEDICAL POLICY SUBJECT: IMPLANTABLE CARDIOVERTER DEFIBRILLATOR MEDICAL POLICY SUBJECT: IMPLANTABLE CARDIOVERTER PAGE: 1 OF: 12 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial product, including

More information

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Case Report Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Taku Nishida MD, Tamio Nakajima MD, PhD, Yutaka Goryo MD, Ken-ichi Ishigami MD, PhD, Hiroyuki Kawata

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

Dialysis-Dependent Cardiomyopathy Patients Demonstrate Poor Survival Despite Reverse Remodeling With Cardiac Resynchronization Therapy

Dialysis-Dependent Cardiomyopathy Patients Demonstrate Poor Survival Despite Reverse Remodeling With Cardiac Resynchronization Therapy Dialysis-Dependent Cardiomyopathy Patients Demonstrate Poor Survival Despite Reverse Remodeling With Cardiac Resynchronization Therapy Evan Adelstein, MD, FHRS John Gorcsan III, MD Samir Saba, MD, FHRS

More information