1 Original Contribution A Novel Flow-Based Parameter of Collateral Function Assessed by Intracoronary Thermodilution Markus Lindner, Stephan B. Felix, MD, Klaus Empen, MD, Thorsten Reffelmann, MD Abstract: Background. Currently, many methods for quantitation of coronary collateral function are based on intracoronary pressure measurements distal of an occluded balloon, which do not fully account for the dynamic nature of collateral flow. Therefore, a flow-based parameter of coronary collateral function based upon principles of thermodilution was evaluated. Methods. In 26 patients with a high-grade coronary artery stenosis, intracoronary hemodynamics were analyzed by the RadiAnalyzer system (St Jude Medical), including fractional flow reserve (FFR), index of microcirculatory resistance (IMR), and the pressure-based collateral flow index (CFI) during balloon occlusion and hyperemia (intravenous adenosine). Moreover, immediately after an intracoronary bolus of room-temperature saline, the balloon was occluded and the intracoronary temperature distal to the balloon was analyzed over time. The slope of the temperaturetime curve was calculated after logarithmic transformation as an index of collateral blood flow (CBFI). Results. The coefficient of variation between two measurements of CBFI amounted to 11 ± 2%. In patients with CFI 0.25, CBFI amounted to 0.55 ± 0.09, whereas in those with CFI <0.25, CBFI was 0.37 ± CBFI correlated significantly with CFI (r = 0.65; P<.001). Interestingly, in the subgroup with IMR below the median (<14.2 mm Hg s), the slope of the linear regression for CBFI vs CFI was steeper than in individuals with higher IMR, which indicates more effective collateral flow for any given intracoronary pressure distal to the occluded balloon in the group with lower microvascular resistance. Conclusions. This novel index might be useful as a flow-based index of collateral function, and should be evaluated in further studies. the dynamic aspect of coronary collateral flow: apart from pressure differences between the three compartments (aortic pressure, distal occlusion pressure, venous pressure), the resistance of the collateral circulation and the resistance of the myocardium supplied by the coronary arterial segment, which is occluded by the balloon, determine flow through the collateral network. 2 In situations where distal myocardial resistance is increased (eg, due to vasoconstriction), CFI appears to be higher than expected, which led to the above-mentioned recommendation to measure CFI during maximum vasodilation (eg, 140 µg adenosine/kg/min intravenously). 3-7,9,10 Even with fixed elevations of myocardial resistance, as it may occur downstream to a stenosis due to the no-reflow phenomenon resulting from previous myocardial infarction or coronary microembolization, 11 apparently high CFI values may not only reflect low collateral resistance, but impeded outflow toward the venous vasculature. Therefore, parameters estimating effective collateral flow rather than distal coronary pressure at steady state during balloon occlusion may be more suitable for quantifying the functional collateral capacity. In this pilot study, we present a novel flow-based parameter of collateral function in patients with coronary artery disease based upon principles of intracoronary thermodilution. J INVASIVE CARDIOL 2014;26(4): Key words: coronary collateral function, fractional flow reserve To date, various angiographic indices as well as flow- and pressure-derived parameters are used for quantitation of coronary collateral function. 1-8 Collateral flow index (CFI), originally introduced as collateral fractional flow reserve in 1993, 2 measured as the difference between distal coronary occlusion pressure and central venous pressure during intracoronary balloon inflation divided by the difference between aortic pressure and central venous pressure, is presumed to be the gold standard, in particular when determined during maximum coronary vasodilation (eg, systemic adenosine infusion). 2,4,6,7,9,10 However, this index does not fully take into account From the Klinik und Poliklinik für Innere Medizin B, Universitätsmedizin Greifswald, Ernst-Moritz-Arndt-Universität Greifswald, Ferdinand-Sauerbruch-Strasse, Greifswald, Germany. Part of this work was presented as an abstract at the 78th scientific session of the German Cardiac Society in Mannheim in April Disclosure: The authors have completed and returned the ICMJE Form for Disclosure of Potential Conflicts of Interest. The authors report no conflicts of interest regarding the content herein. Manuscript submitted August 7, 2013, provisional acceptance given August 27, 2013, final version accepted September 13, Address for correspondence: Prof Dr Thorsten Reffelmann, Klinik und Poliklinik für Innere Medizin B, Universitätsmedizin Greifswald, Ernst-Moritz-Arndt-Universität Greifswald, Ferdinand-Sauerbruch-Str, Greifswald, Germany. ThorstenReffelmann@web.de Methods The study protocols are consistent with the principles of the Declaration of Helsinki and were approved by the Ethics Committee of the University of Greifswald. Patient population and inclusion/exclusion criteria. Prior to angioplasty, informed consent was obtained from all patients. Patients years-old with a high-grade coronary artery atherosclerotic stenosis scheduled for percutaneous coronary intervention (PCI) were eligible for this study. Only stenoses within the proximal parts of the epicardial artery were included in order to ensure that the tip of the guidewire could be safely positioned in the distal artery at a site of sufficient luminal diameter. Exclusion criteria were: acute ST-segment elevation and non-st segment elevation myocardial infarction (within 48 hours), glomerular filtration rate <30 ml/min (estimated by the Modification of Diet in Renal Disease [MDRD] formula), 12 stenoses with involvement of the left main stem, bifurcational stenosis, diffuse atherosclerotic lesions, lesions located in the distal third of the epicardial artery, and expectedly complex PCI procedures according to the physician s discretion, eg, those requiring special guidewires or additional devices. Study protocol. After taking the patient s history (including atherosclerotic risk factors), blood pressure was measured, a 12-lead electrocardiogram (ECG) was performed, and standard laboratory parameters including blood count, electrolytes, creatinine, urea, thyroid stimulating hormone, cholesterol, low-density lipoprotein, 148 The Journal of Invasive Cardiology

2 Figure 1. Typical temperature curve during collateral blood flow index (CBFI) measurement, as recorded by the RADI-analyzer. This figure shows the interface of the RADI-analyzer in temperature mode. At the top, mean pressure distal to the stenosis (Pd) and mean aortic pressure (Pa), as well as their ratio (FFR) are shown. In the bottom window, temperature is constantly recorded. After a flush of room-temperature saline, the temperature decreases, the balloon is occluded, and simultaneously, distal intracoronary pressure decreases. During occlusion, temperature returned to baseline. At the bottom right, the temperature curve is seen after logarithmic transformation. The slope of the linear regression equals CBFI. r=0.903 P<.001 high-density lipoprotein, triglycerides, glucose, and coagulation parameters were taken. All patients scheduled for PCI received oral aspirin and clopidogrel prior to the intervention. Coronary intervention and intracoronary measurements. All coronary interventions were carried out using standard procedures. r=0.754 P<.001 Figure 2. Repeatability of collateral flow index (CFI; left) and collateral blood flow index (CBFI; right) between the first and second registration of the intracoronary temperature curve during balloon occlusion after a room-temperature bolus injection of saline (regression line and 95% confidence interval). Flow-Based Parameter of Collateral Function Femoral arterial and venous access was obtained using the Seldinger technique. A pigtail catheter was advanced into the vena cava inferior to the right atrium junction to measure central venous pressure. After positioning the coronary guiding catheter a calibrated Radi PressureWire Certus (St Jude Medical) was advanced into the distal coronary artery. Only patients with a side branch of the coronary artery adjacent to the temperature sensor were included. Fractional flow reserve (FFR; mean distal intracoronary pressure / mean aortic pressure [pressure at the tip of the guiding catheter]) and coronary flow reserve (CFR; ratio of mean transit time of an intracoronary cold bolus under baseline conditions to hyperemia [3 ml room temperature saline, three measurements each]) were determined. 2,13-15 Hyperemia was induced by intravenous adenosine (140 µg/kg/ min). The index of microcirculatory resistance (IMR) was calculated as the pressure distal to the stenosis multiplied by mean transit time In patients with a chronic total occlusion, mean distal intracoronary pressure / mean aortic pressure during the first balloon inflation within the reopened coronary artery measured under adenosine was taken as an estimate of FFR. According to the manufacturer, the pressure sensor of the Radi PressureWire Certus measures with an accuracy of ±1 mm Hg from -30 to 300 mm Hg with a sampling frequency of 0-25 Hz. In temperature mode, the sensor measures from C with an accuracy of at least 0.05 C or 10% of T, whichever is greatest, with a sampling rate of 0-7 Hz. 19 Measurement of the novel collateral blood flow index (CBFI). During the angioplasty procedure, the intracoronary temperature distal to the occluded balloon was registered via the Radi PressureWire Certus during each balloon inflation with a simultaneous intravenous infusion of adenosine (140 µg/kg/min). To ensure that the temperature sensor of the wire was positioned distally to the balloon, the tip of the wire was advanced more than 3 cm distally to the distal end of the balloon. During the first balloon inflation (without administration of room-temperature saline), distal temperature was recorded to evaluate whether the balloon inflation itself had any influence on measured distal temperature. Prior to the second balloon inflation, an intracoronary flush of room-temperature saline (approximately 3-5 ml) was given until intracoronary temperature Vol. 26, No. 4, April

3 LINDNER, et al. r=0.650 P<.001 r=0.688 P<.05 r=0.407 P<.15 Figure 3. Top: Correlation between collateral flow index (CFI) and collateral blood flow index (CBFI) for the entire study population. Bottom: Linear regression between CFI and CBFI in the subgroup of patients with index of microcirculatory resistance (IMR) below (left) and above (right) the median (14.2 mm Hg s). The slope of the regression line was steeper in patients with evidence for lower microvascular resistance within the myocardial territory supplied by the coronary stenosis (IMR below median). had dropped by at least 1-3 C. Immediately after the temperature drop had occurred, the balloon was inflated and the flush of saline was stopped. The change in intracoronary temperature over time toward baseline temperature was recorded. To assess repeatability, the procedure was repeated during the next balloon inflation. When it was safe to advance the wire more distally, the measuring procedure was repeated with the wire tip more distal to estimate the influence of the wire position on the values of CBFI. During these measurements, temperature asymptotically increased from the initial temperature drop back to baseline (Figure 1). Temperature-time curve analysis started once complete balloon occlusion was reached, as indicated by the reduction of intracoronary pressure. After logarithmic transformation, this curve was approximated by a linear equation. By the method of least squares, CBFI was calculated as the slope of its linear regression and multiplied by 1000 (Appendix 1). Simultaneously, during each balloon occlusion, collateral flow index (CFI) was assessed as a ratio of pressure distal to the stenosis minus central venous pressure to mean aortic pressure minus central venous pressure. 2-4 To calculate CBFI, recorded data were exported from the RADI analyzer into ASCII format with 100 time points per second. Further conversion by Microsoft Excel resulted in a temperature listing in C to the third decimal place. Quantitative coronary angiography (QCA). Vessel diameters, diameter stenosis, and the distance between the tip of the PressureWire and the distal end of the balloon were analyzed offline, using the QCA tool provided by the program Horizon Cardiology 11.1, with client TCS 3.2 service pack 6 (McKesson). The diameter and length of the stenosis were measured three times (renewed calibration each) and then averaged. Statistical analysis. The statistical analysis was carried out using SigmaPlot 11.0 (Systat Software). For comparison between groups, unpaired t-test or one-way analysis of variance (ANOVA) was used. Correlation analysis was performed using Pearson s model or Spearman rank order. Multiple linear regressions (least square method) were compared by analysis of covariance (ANCOVA). Results are expressed as mean ± standard error of the mean or percentages. The coefficient of variation was calculated as standard deviation divided by the mean (as percentage). A P-value <.05 was considered statistically significant. Results Exclusions. Out of 32 patients, 6 were excluded from final analysis due to the following reasons: 5 patients were not included because QCA showed that the wire was positioned in a small side branch and the vessel diameter was too small (<1.3 mm). In 1 patient, the tip of the wire was <3 cm distal to the balloon. Thus, the final study population consisted of 26 patients. Patient characteristics. The risk profile and clinical data of the patients are presented in Table 1. Thirteen out of 26 patients had prior myocardial infarction (MI); 7 MIs had occurred in the area that was treated by angioplasty in this study. Four patients had a CTO of the treated coronary artery. Ten out of 26 patients had a CFI >0.25, which is assumed to be a marker for relevant collateral flow. According to the Rentrop s angiographic grading of collaterals, 3 patients were classified as Rentrop III, 1 patient as Rentrop II, and 22 as Rentrop 0. Table 2 illustrates hemodynamics as assessed by intracoronary pressure measurements and thermodilution. Measurement of CBFI, feasibility, and repeatability. During the first balloon inflation without the initial flush of saline, measured 150 The Journal of Invasive Cardiology

4 temperature during balloon occlusion in the distal coronary artery did not substantially change over time; it varied by 0.06 C ± C over 1 minute without a consistent trend over time. Measurements with an initial flush of room-temperature saline and concurrent balloon inflation were feasible in all 32 patients. Figure 1 shows a typical registration of distal coronary temperature after an initial flush of room-temperature saline with simultaneous balloon occlusion. After balloon occlusion, temperature increased asymptotically from the initial temperature drop back to baseline. This increase was logarithmically transformed, including all instances of time from complete balloon occlusion to deflation. Average measuring time was 29 ± 2 s. Subsequent linear regression led to a coefficient of determination of R² = 0.97 ± 0.01, validating the logarithmic approach. We performed two consecutive measurements of CBFI and CFI under identical conditions. Comparison of the two measurements exhibited a coefficient of variation of 11 ± 2% without evidence for collateral recruitment or other temporal effects. As shown in Figure 2, there is a statistically high level of correlation between the respective values of CBFI and CFI, which were measured under identical conditions (r = and P<.001 for CBFI; r = and P<.001 for CFI). Influence of heart rate, blood pressure, initial temperature drop, wire position, and vessel diameter. Absolute differences in heart rate, aortic pressure, and initial temperature drop between the two measurements under identical conditions were moderate. Within the observed differences, there was no consistent influence of heart rate, pressure, and temperature drop on CBFI. Furthermore, there was no apparent difference between measurements in the three coronary arteries. However, CBFI appeared to depend on the position of the sensor in the vessel: values increased slightly, yet statistically significantly, when the wire was further advanced into the vessel. In very small vessels (diameter of 1.3 mm), CBFI values appeared to be unproportionally high (0.72 ± 0.09 for CBFI 1.3 mm vs 0.44 ± 0.04 for CBFI >1.3 mm; P<.05). Therefore, measurements in vessels 1.3 mm were excluded. In contrast, there was no significant relation between vessel diameter and CBFI in coronary arteries with a diameter >1.3 mm (r = and P=.22). The average vessel diameter measured at the sensor position was 2.33 ± 0.07 mm. Correlation between CBFI and CFI. Overall, CBFI correlated significantly with CFI (Figure 3, top). CBFI was 0.55 ± 0.09 in patients with CFI 0.25 and 0.37 ± 0.03 in patients with CFI <0.25. Correlation tended to be better in subgroups with FFR and IMR below the median, as well as in patients with stenoses of higher degree by angiography. Interestingly, in patients with IMR below the median (14.2 mm Hg s), the slope of the linear regression line between CBFI and CFI was steeper, indicating more effective flow for any given distal intracoronary pressure during balloon occlusion, albeit not to the point of statistical significance by ANCOVA (Figure 3, bottom). Discussion This pilot study evaluated a novel parameter of coronary collateral function as assessed by intracoronary thermodilution. The measurements, based on the time course of intracoronary temperature during balloon occlusion during percutaneous coronary angioplasty after an initial flush of room-temperature saline, were easy to perform and feasible, as long as the wire tip could be safely positioned more than 3 cm distal to the balloon, and a side branch of the coronary artery existed near the stenosis. Flow-Based Parameter of Collateral Function In coronary arteries with a diameter >1.3 mm at the sensor position, CBFI correlated significantly with CFI. Regarding patients with increased myocardial resistance within the coronary territory supplied by the occluded part of the coronary artery, an interesting observation could be made, ie, the flow-based parameter of collateral function indicated less flow for any given value of CFI in comparison to coronary arteries with lower myocardial resistance. Hemodynamics of collateral function. Currently, CFI, originally introduced as collateral fractional flow reserve in 1993, 2 is presumed to be the gold standard, in particular when determined during maximum coronary vasodilation (eg, intravenous adenosine infusion). 3-5,8-10 However, collateral function is not only characterized by pressure differences between the three coronary compartments (aortic pressure, distal occlusion pressure, venous pressure), but also by the resistance of the collateral network as well as microvascular resistance of the myocardium supplied by the coronary arterial segment that is occluded by the balloon. Since flow over the collaterals to the coronary artery and flow from the coronary over the microvasculature toward the venous system are in series, both are equal during balloon occlusion. Therefore, provided there is no hemodynamically significant stenosis proximal to the origin of the collateral network, flow can be related to collateral and myocardial resistance (according to Ohm s law), using the following equations where Pa = aortic pressure, Pd = distal pressure, ZVD = central venous pressure, and R = resistance: Therefore, CFI equals: As a consequence, pressure-derived CFI is influenced by the resistance of the collateral circulation (with high collateral resistance [Rcollateral >> Rmyocardial] resulting in low CFI, and by the resistance of the myocardial circulation with high myocardial resistance [Rmyocardial >> Rcollateral] resulting in a CFI near 1, even if collateral flow is low). In situations where distal myocardial resistance is high (eg, due to vasoconstriction), CFI appears to be higher than expected, which led to the above-mentioned recommendation to measure CFI during maximal vasodilation (eg, 140 µg adenosine/kg/min intravenously). By the same token, this may also be the case with fixed elevations of myocardial resistance, as it may occur downstream to a stenosis, eg, due to the no-reflow phenomenon. 12,20 Apparently, high CFI values may not reflect low collateral resistance, but rather impeded outflow toward the venous vasculature. Therefore, measurements of parameters related to effective collateral flow rather than resulting coronary pressure at steady state may be more suitable for characterizing the functional capacity of the collateral network. Another frequently used index for characterizing collateral flow is derived from measurements of blood flow velocities using an intracoronary Doppler flow wire. For this purpose, coronary flow velocities during complete balloon occlusion (measured distally to the occluded balloon) were divided by flow velocities at the same site after successful balloon angioplasty. 3,6,8 Hereby, the average peak velocity, as assessed by intracoronary Doppler spectra during Vol. 26, No. 4, April

5 LINDNER, et al. maximum vasodilation distal to the occluded balloon was divided by the average peak velocity at the same site after balloon dilatation. This approach assumes that flow after angioplasty is immediately restored to normal; even more importantly, it cannot account for the unpredictable direction of coronary flow during balloon occlusion (as the main collaterals may insert more distally or proximally to the tip of the flow wire). A novel flow-based index of collateral function. Assuming that distal temperature (after coronary balloon occlusion) is predominately determined by convection, ie, collateral flow, and direct thermic conduction is negligible, which is presumed to be the case with sufficient vessel diameters, the temperature in the distal coronary segment should asymptotically increase back to zero after the initial temperature drop caused by the intracoronary saline flush. This equation is transformed logarithmically. The slope of the resulting linear equation is approximated by the method of least squares and taken for the amount of relative collateral blood flow, called CBFI. In order to standardize the procedure, the measurements should be undertaken during 140 µg/kg/min adenosine intravenously to ensure maximum coronary vasodilation. 4,5,10 The distal tip of the wire should be at least >3 cm distal to the balloon to make sure the sensor is reliably outside of the balloon. In addition, a side branch of the coronary artery should exist adjacent to the temperature sensor. The results of this pilot study indicate a good correlation between CBFI and CFI, as long as the vessel diameter is sufficient. Most importantly, this parameter appeared to reflect dynamic flow rather than static pressures as a marker of collateral function. This ties in well with the observation that the relation between CBFI and CFI tended to be higher when peripheral resistance was low in comparison to higher resistance in the myocardium supplied by the segment of the coronary artery treated by PCI. However, in the presence of a severe stenosis, IMR might overestimate myocardial resistance to a certain degree, which could slightly weaken this correlation. 16,17 Moreover, given further specification of accuracy and validity of CBFI, the relationship between intracoronary pressure distal to the occluded balloon, central venous pressure, and CBFI might also be used as an index of microvascular resistance in the supplied myocardial territory. Notably, as shown in Figure 2, the two measurements of collateral function under identical conditions did not point to the recruitment of collateral function during two subsequent balloon occlusions in our setting, neither when CFI nor when CBFI was used as a measure of collateral function. This is in contrast to a 1999 study by Billinger et al, where subsequent balloon inflations over 2 minutes resulted in an increase in CFI from 0.13 ± 0.11 during the first inflation to 0.19 ± 0.10 during the third inflation. 21 Apart from different study populations with different cardiovascular risk profile in Billinger s study, there might be two reasons for this discrepancy between the studies. First, in our study an initial balloon inflation was performed to exclude any influence of balloon occlusion on distal temperature measurements. This first balloon occlusion might have had influence on subsequent measurements in our study. Second, the balloon inflations in Billinger s study were carefully standardized to 2 minutes of occlusion each. In our study, the length of our balloon occlusions were left to the physician s discretion because it was judged necessary to achieve an adequate PCI result. In our setting, with an average measuring time of 29 ± 2 s, no evidence for collateral recruitment was demonstrated. Table 1. Patient characteristics, medications, and angiographic findings. Patients 26 (100%) Sex (male) 19 (73%) Age 70 ± 1.6 Current smokers 11 (42%) Hypertensives 26 (100%) Dyslipoproteinemia 23 (88%) Diabetes mellitus 8 (31%) Body mass index (kg/m 2 ) 30 ± 1.1 Prior myocardial infarction 13 (50%) Prior infarction in treated coronary artery 7 (27%) Prior PCI 14 (54%) 1-vessel coronary heart disease 3 (12%) 2-vessel coronary heart disease 8 (31%) 3-vessel coronary heart disease 15 (58%) Aspirin 26 (100%) Clopidogrel 26 (100%) Statin 25 (96%) Beta-blocker 24 (92%) ACE inhibitor/at-ii receptor antagonist 25 (96%) Data given as number (percentage) or mean ± standard error of the mean. Table 2. Hemodynamics assessed by intracoronary pressure measurement and thermodilution in patients with collateral flow index (CFI) 0.25 and patients with CFI <0.25. CFI 0.25 (n = 10) CFI <0.25 (n = 16) P- Value CFI 0.39 ± ± 0.01 <.001 CBFI 0.55 ± ± 0.03 <.05 CFR 1.35 ± ± 0.07 NS FFR 0.62 ± ± 0.03 <.05 IMR (mm Hg s) ± ± 1.60 NS Stenosis (%) * 86.5 ± ± 2.6 NS CFI = collateral flow index (pressure-derived); CBFI = collateral blood flow index (thermodilution); CFR = coronary flow reserve; FFR = fractional flow reserve; IMR = index of microvascular resistance. P-values between groups; NS = not significant. * Four total coronary occlusions. Study limitations. As demonstrated above, the measured CBFI value increased when the wire was further advanced into the coronary artery. Therefore, careful standardization of the measuring procedure is mandatory with the sensor position outside the occluded balloon and a coronary vessel of sufficient size. Only coronary arteries of sufficient size (>1.3 mm) and stenoses in the proximal to mid part of the arteries are eligible for these measurements. In smaller arteries, CBFI appeared to be inadequately high, probably due to direct thermal conduction. Furthermore, a marginal location of the wire within the coronary artery is likely to result in altered temperature measurements due to the proximity of the vessel wall. Repeatability of two subsequent measurements under identical conditions 152 The Journal of Invasive Cardiology

6 Flow-Based Parameter of Collateral Function was relatively good (coefficient of variation 11 ± 2%). Most importantly, the initial temperature drop, which substantially differed between two intracoronary injections of saline, had no consistent effect on CBFI, which was measured as the slope of the temperature time curve after logarithmic transformation. This is in agreement with the high coefficient of determination (R² = 0.97) of the linear regression of these curves. Moreover, the time point, when complete balloon occlusion was achieved, which was the starting point for the measurements, could be easily defined from the intracoronary pressure measurements (compare Figure 1). Even if this parameter does not represent absolute collateral flow, it does appear to be closely related to it, since indices of myocardial resistance strongly influenced the relation between CFI and CBFI. Appendix 1. Mathematical equations for calculation of collateral blood flow index (CBFI). During the performed measurements temperature asymptotically increased from the initial temperature drop back to baseline (Figure 1). After logarithmic transformation, this curve was approximated by a linear equation. By the method of least squares, CBFI was calculated from the slope of its linear regression: the approxima- Starting with a linear equation like tion by the method of least squares provides: Conclusion This novel parameter of collateral function might be validated in larger studies including other parameters of collateral flow (eg, scintigraphic techniques in total coronary occlusions). Our results suggest that this parameter may in part account for changes in distal myocardial microvascular resistance, therefore representing a flow-based parameter. If further validation is possible, and limitations of accuracy can be precisely defined, this parameter in conjunction with CFI might also be useful to estimate myocardial viability within the supplied coronary territory, as fixed elevation of microvascular resistance, for example in zones of microvascular obstruction was shown to be closely associated with myocardial necrosis. 10,11,20,22,23 Furthermore, collateral flow and myocardial viability may be closely linked in patients with recent myocardial infarction. 23 Studies on temporal development of collateral function and pathophysiological determinants may also include a flow-based parameter to fully account for the dynamic nature of collateral flow. 8,22 References 1. Rentrop KP, Cohen M, Blanke H, Phillips RA. Changes in collateral channel filling immediately after controlled coronary artery occlusion by an angioplasty balloon in human subjects. J Am Coll Cardiol. 1985;5(3): Pijls NH, van Son JA, Kirkeeide RL, et al. Experimental basis of determining maximum coronary, myocardial, and collateral blood flow by pressure measurements for assessing functional stenosis severity before and after percutaneous transluminal coronary angioplasty. Circulation. 1993;87(4): Seiler C, Fleisch M, Garachemani A, Meier B. Coronary collateral quantitation in patients with coronary artery disease using intravascular flow velocity or pressure measurements. J Am Coll Cardiol. 1998;32(5): Perera D, Patel S, Blows L, et al. Pharmacological vasodilatation in the assessment of pressurederived collateral flow index. Heart. 2006;92(8): Jensen LO, Thayssen P, Lassen JF, et al. Recruitable collateral blood flow index predicts coronary instent restenosis after percutaneous coronary intervention. Eur Heart J. 2007;28(15): Doucette JW, Corl PD, Payne HM, et al. Validation of a Doppler guide wire for intravascular measurement of coronary artery flow velocity. Circulation. 1992;85(5): Berry C, Balachandran KP, L Allier PL, et al. Importance of collateral circulation in coronary heart disease. Eur Heart J. 2007;28(3): van Royen N, Piek JJ, Schaper W, Fulton WF. A critical review of clinical arteriogenesis research. J Am Coll Cardiol. 2010;55(1): Perera D, Kanaganayagam GS, Saha M, et al. Coronary collaterals remain recruitable after percutaneous intervention. Circulation. 2007;115(15): Wustmann K, Zbinden S, Windecker S, et al. Is there functional collateral flow during vascular occlusion in angiographically normal coronary arteries? Circulation. 2003;107(17): Reffelmann T, Hale SL, Dow JS, Kloner RA. No-reflow phenomenon persists longterm after ischemia/reperfusion in the rat and predicts infarct expansion. Circulation. 2003;108(23): Levey AS, Bosch JP, Lewis JB, et al. A more accurate method to estimate glomerular filtration rate from serum creatinine: a new prediction equation. Modification of diet in renal disease study group. Ann Intern Med. 1999;130(6): Barbato E, Aarnoudse W, Aengevaeren WR, et al. Validation of coronary flow reserve measurements by thermodilution in clinical practice. Eur Heart J. 2004;25(3): Pijls NH, De Bruyne B, Peels K, et al. Measurement of fractional flow reserve to assess the functional severity of coronary-artery stenoses. N Engl J Med. 1996;334(26): Pijls NH, De Bruyne B, Smith L, et al. Coronary thermodilution to assess flow reserve: validation in humans. Circulation. 2002;105(21): Aaroundse W, Fearon WF, Manoharan G, et al. Epicardial stenosis severity does not affect minimal microcirculatory resistance. Circulation. 2004;110(15): Fearon WF, Aarnoudse W, Pijls NHJ, et al. Microvascular resistance is not influenced by epicardial coronary artery stenosis severity. Circulation. 2004;109(19): Fearon WF, Balsam LB, Farouque HM, et al. Novel index for invasively assessing the coronary microcirculation. Circulation. 2003;107(25): RadiMedical Systems. Radi Certus wire technical sheet Reffelmann T, Hale SL, Li G, Kloner RA. Relationship between no reflow and infarct size as influenced by the duration of ischemia and reperfusion. Am J Physiol Heart Circ Physiol. 2002;282(2):H766-H Billinger M, Fleisch M, Eberli FR, Garachemani A, Meier B, Seiler C. Is the development of myocardial tolerance to repeated ischemia in humans due to preconditioning or to collateral recruitment? J Am Coll Cardiol. 1999;33(4): Teunissen PFA, Horrevoets AJG, van Royen N. The coronary collateral circulation: Genetic and environmental determinants in experimental models and humans. J Mol Cell Cardiol. 2011;52(4): Sabia PJ, Powers ER, Ragosta M, et al. An association between collateral blood flow and myocardial viability in patients with recent myocardial infarction. N Engl J Med. 1992; 327(26): Vol. 26, No. 4, April

ABSOLUTE BLOOD FLOW MEASUREMENTS: PRINCIPLES

ABSOLUTE BLOOD FLOW MEASUREMENTS: PRINCIPLES CORONARY PHYSIOLOGY IN THE CATHLAB: ABSOLUTE BLOOD FLOW MEASUREMENTS: PRINCIPLES Educational Training Program ESC European Heart House april 23rd - 25th 2015 Nico H. J. Pijls, MD, PhD Catharina Hospital,

More information

Calculation of the Index of Microcirculatory Resistance without Coronary Wedge Pressure Measurement in the Presence of Epicardial Stenosis

Calculation of the Index of Microcirculatory Resistance without Coronary Wedge Pressure Measurement in the Presence of Epicardial Stenosis Calculation of the Index of Microcirculatory Resistance without Coronary Wedge Pressure Measurement in the Presence of Epicardial Stenosis ASC Yong 1,2, M Ho 3, MG Shah 3, D Daniels 3, C Chawantanpipat

More information

FRACTIONAL FLOW RESERVE Step-by-step measurement, Practical tips & Pitfalls

FRACTIONAL FLOW RESERVE Step-by-step measurement, Practical tips & Pitfalls FRACTIONAL FLOW RESERVE Step-by-step measurement, Practical tips & Pitfalls Ahmed M ElGuindy, MSc, MRCP(UK) Division of Cardiology Aswan Heart Centre 2013 Fractional Flow Reserve Essential diagnostic tool

More information

Value of Index of Microvascular Resistance (IMR) in Microvascular Integrity

Value of Index of Microvascular Resistance (IMR) in Microvascular Integrity Value of Index of Microvascular Resistance (IMR) in Microvascular Integrity Seung-Woon Rha, Korea University Guro Hospital, Myeong-Ho Yoon, Ajou University Hospital Imaging & Physiology Summit 2009 Nov

More information

Fractional Flow Reserve. A physiological approach to guide complex interventions

Fractional Flow Reserve. A physiological approach to guide complex interventions Fractional Flow Reserve A physiological approach to guide complex interventions What is FFR? Fractional Flow Reserve (FFR) is a lesion specific, physiological index determining the hemodynamic severity

More information

Journal of the American College of Cardiology Vol. 38, No. 7, by the American College of Cardiology ISSN /01/$22.

Journal of the American College of Cardiology Vol. 38, No. 7, by the American College of Cardiology ISSN /01/$22. Journal of the American College of Cardiology Vol. 38, No. 7, 2001 2001 by the American College of Cardiology ISSN 0735-1097/01/$22.00 Published by Elsevier Science Inc. PII S0735-1097(01)01664-3 Coronary

More information

The fluid, dynamic interaction of multiple sequential

The fluid, dynamic interaction of multiple sequential Coronary Pressure Measurement to Assess the Hemodynamic Significance of Serial Stenoses Within One Coronary Artery Validation in Humans Nico H.J. Pijls, MD, PhD; Bernard De Bruyne, MD, PhD; G. Jan Willem

More information

Case report. Resistance in the cath lab : the utility of hyperemic stenosis resistance in the functional assessment of coronary artery disease

Case report. Resistance in the cath lab : the utility of hyperemic stenosis resistance in the functional assessment of coronary artery disease Resistance in the cath lab : the utility of hyperemic stenosis resistance in the functional assessment of coronary artery disease Kalpa De Silva, Divaka Perera Cardiovascular Division, The Rayne Institute,

More information

Technical Aspects and Clinical Indications of FFR

Technical Aspects and Clinical Indications of FFR Technical Aspects and Clinical Indications of FFR Emanuele Barbato, MD, PhD, FESC Cardiovascular Center Aalst - OLV Clinic Aalst, Belgium Potential conflicts of interest Consulting fees and honoraria on

More information

Fractional Flow Reserve: Basics, FAME 1, FAME 2. William F. Fearon, MD Associate Professor Stanford University Medical Center

Fractional Flow Reserve: Basics, FAME 1, FAME 2. William F. Fearon, MD Associate Professor Stanford University Medical Center Fractional Flow Reserve: Basics, FAME 1, FAME 2 William F. Fearon, MD Associate Professor Stanford University Medical Center Conflict of Interest Advisory Board for HeartFlow Research grant from St. Jude

More information

Fractional Flow Reserve (FFR) --Practical Set Up Pressure Measurement --

Fractional Flow Reserve (FFR) --Practical Set Up Pressure Measurement -- Fractional Flow Reserve (FFR) --Practical Set Up Pressure Measurement -- JoonHyung Doh, MD, PhD Assistant Professor, Vision21 Cardiac and Vascular Center Inje University Ilsan Paik Hospital Goyang, Korea

More information

Novel insights into the complexity of ischaemic heart disease derived from combined coronary pressure and flow velocity measurements van de Hoef, T.P.

Novel insights into the complexity of ischaemic heart disease derived from combined coronary pressure and flow velocity measurements van de Hoef, T.P. UvA-DARE (Digital Academic Repository) Novel insights into the complexity of ischaemic heart disease derived from combined coronary pressure and flow velocity measurements van de Hoef, T.P. Link to publication

More information

Fractional Flow Reserve (FFR) --Practical Set Up Pressure Measurement --

Fractional Flow Reserve (FFR) --Practical Set Up Pressure Measurement -- Fractional Flow Reserve (FFR) --Practical Set Up Pressure Measurement -- Joon Hyung Doh, MD, PhD Associate Professor, Division of Cardiology Inje University Ilsan Paik Hospital Goyang, Korea 목차 Fractional

More information

Evaluation of Intermediate Coronary lesions: Can You Handle the Pressure? Jeffrey A Southard, MD May 4, 2013

Evaluation of Intermediate Coronary lesions: Can You Handle the Pressure? Jeffrey A Southard, MD May 4, 2013 Evaluation of Intermediate Coronary lesions: Can You Handle the Pressure? Jeffrey A Southard, MD May 4, 2013 Disclosures Consultant- St Jude Medical Boston Scientific Speaker- Volcano Corporation Heart

More information

PCIs on Intermediate Lesions NCDR Cath-PCI Registry

PCIs on Intermediate Lesions NCDR Cath-PCI Registry Practical Application Of Coronary Physiology in The Cath Lab Talal T Attar, MD, MBA, FACC PCIs on Intermediate Lesions NCDR Cath-PCI Registry Fraction of stenoses 50-70% treated with PCI without further

More information

Diffuse Disease and Serial Stenoses. Bernard De Bruyne Cardiovascular Center Aalst Belgium

Diffuse Disease and Serial Stenoses. Bernard De Bruyne Cardiovascular Center Aalst Belgium Diffuse Disease and Serial Stenoses Bernard De Bruyne Cardiovascular Center Aalst Belgium Atherosclerosis is a Diffuse Disease Serial Stenoses A B P a P m P d When A is isolated, hyperemic flow through

More information

FRACTIONAL FLOW RESERVE: CONCEPT, EXPERIMENTAL BASIS, CUT-OFF VALUES

FRACTIONAL FLOW RESERVE: CONCEPT, EXPERIMENTAL BASIS, CUT-OFF VALUES CORONARY PHYSIOLOGY IN THE CATHLAB: FRACTIONAL FLOW RESERVE: CONCEPT, EXPERIMENTAL BASIS, CUT-OFF VALUES Educational Training Program ESC European Heart House april 25th - 27th 2013 Nico H. J. Pijls, MD,

More information

Dave Kettles, St Dominics Hospital East London.

Dave Kettles, St Dominics Hospital East London. Dave Kettles, St Dominics Hospital East London. 110 x 150 Angina for a couple of months Trop T negative T wave inversion across the chest leads Not wanting to risk radial Huge struggle with femoral

More information

Physiologically Assessed Coronary Collateral Flow and Adverse Cardiac Ischemic Events: A Follow-Up Study in 403 Patients With Coronary Artery Disease

Physiologically Assessed Coronary Collateral Flow and Adverse Cardiac Ischemic Events: A Follow-Up Study in 403 Patients With Coronary Artery Disease Journal of the American College of Cardiology Vol. 40, No. 9, 2002 2002 by the American College of Cardiology Foundation ISSN 0735-1097/02/$22.00 Published by Elsevier Science Inc. PII S0735-1097(02)02378-1

More information

FRACTIONAL FLOW RESERVE USE IN THE CATH LAB BECAUSE ANGIOGRAPHY ALONE IS NOT ENOUGH!!!!!!!!

FRACTIONAL FLOW RESERVE USE IN THE CATH LAB BECAUSE ANGIOGRAPHY ALONE IS NOT ENOUGH!!!!!!!! FRACTIONAL FLOW RESERVE USE IN THE CATH LAB BECAUSE ANGIOGRAPHY ALONE IS NOT ENOUGH!!!!!!!! Juan Antonio Pastor-Cervantes,M.D FSCAI, FACC Cardiovascular Institute Memorial Regional Hospital Hollywood Florida

More information

Fractional Flow Reserve and instantaneous wave -free Ratio. Λάμπρος Κ. Μόσιαλος Επεμβατικός Καρδιολόγος ΓΝ Παπαγεωργίου

Fractional Flow Reserve and instantaneous wave -free Ratio. Λάμπρος Κ. Μόσιαλος Επεμβατικός Καρδιολόγος ΓΝ Παπαγεωργίου Fractional Flow Reserve and instantaneous wave -free Ratio Λάμπρος Κ. Μόσιαλος Επεμβατικός Καρδιολόγος ΓΝ Παπαγεωργίου DISCLOSURES There are no financial conflicts of interest relevant to this presentation

More information

Pressure Wire Study. Fractional Flow Reserve (FFR) Nishat Jahagirdar Principal Clinical Cardiac Physiologist Kings College Hospital

Pressure Wire Study. Fractional Flow Reserve (FFR) Nishat Jahagirdar Principal Clinical Cardiac Physiologist Kings College Hospital Pressure Wire Study Fractional Flow Reserve (FFR) Nishat Jahagirdar Principal Clinical Cardiac Physiologist Kings College Hospital History Andreas Gruentzig s paper on Non operative dilation of coronaryartery

More information

A Novel Simple Technique Using Hyperemia to Enhance Pressure Gradient Measurement of the Lower Extremity During Peripheral Intervention

A Novel Simple Technique Using Hyperemia to Enhance Pressure Gradient Measurement of the Lower Extremity During Peripheral Intervention A Novel Simple Technique Using Hyperemia to Enhance Pressure Gradient Measurement of the Lower Extremity During Peripheral Intervention Issam Koleilat, MD; Bruce Gray, MD From Greenville Health System,

More information

The Relationship between Microcirculatory Resistance and Fractional Flow Reserve in Patients with Acute Myocardial Infarction

The Relationship between Microcirculatory Resistance and Fractional Flow Reserve in Patients with Acute Myocardial Infarction Original Article Print ISSN 1738-5520 On-line ISSN 1738-5555 Korean Circulation Journal The Relationship between Microcirculatory Resistance and Fractional Flow Reserve in Patients with Acute Myocardial

More information

Microvasculature Clinical Importance. Keith G Oldroyd Golden Jubilee National Hospital Glasgow, Scotland

Microvasculature Clinical Importance. Keith G Oldroyd Golden Jubilee National Hospital Glasgow, Scotland Microvasculature Clinical Importance Keith G Oldroyd Golden Jubilee National Hospital Glasgow, Scotland William Fulton, MD Scottish Medical Journal, 1963 Body text Fresh explanted human hearts Physiological

More information

Coronary Physiology and FFR. David H. Sibley MD FACC, FSCAI, FACP

Coronary Physiology and FFR. David H. Sibley MD FACC, FSCAI, FACP Coronary Physiology and FFR David H. Sibley MD FACC, FSCAI, FACP Braunwald s Heart Disease, 7 th Edition Control of Coronary Blood Flow 1. CFR = max flow/basal flow and decreases with increasing stenosis

More information

Original Article. Introduction. Korean Circulation Journal

Original Article. Introduction. Korean Circulation Journal Original Article http://dx.doi.org/.47/kcj.14.44.3.141 Print ISSN 1738- On-line ISSN 1738- Korean Circulation Journal Association between Cardiac Troponin Level and Coronary Flow Reserve in Patients without

More information

NEW TECHNOLOGY FOR CARDIAC CHARACTERIZATION

NEW TECHNOLOGY FOR CARDIAC CHARACTERIZATION NEW TECHNOLOGY FOR CARDIAC CHARACTERIZATION COROVENTIS COROFLOW CARDIOVASCULAR SYSTEM Coroventis CoroFlow Cardiovascular System is an advanced platform for assessment of coronary physiology. Designed to

More information

Angiographic Assessment of Collateral Connections in Comparison With Invasively Determined Collateral Function in Chronic Coronary Occlusions

Angiographic Assessment of Collateral Connections in Comparison With Invasively Determined Collateral Function in Chronic Coronary Occlusions Angiographic Assessment of Collateral Connections in Comparison With Invasively Determined Collateral Function in Chronic Coronary Occlusions Gerald S. Werner, MD; Markus Ferrari, MD; Stephan Heinke, MD;

More information

FFR Incorporating & Expanding it s use in Clinical Practice

FFR Incorporating & Expanding it s use in Clinical Practice FFR Incorporating & Expanding it s use in Clinical Practice Suleiman Kharabsheh, MD Consultant Invasive Cardiology Assistant professor, Alfaisal Univ. KFHI - KFSHRC Concept of FFR Maximum flow down a vessel

More information

Functional and Anatomical Diagnosis of Coronary Artery Stenoses

Functional and Anatomical Diagnosis of Coronary Artery Stenoses Journal of Surgical Research 150, 24 33 (2008) doi:10.1016/j.jss.2007.10.018 Functional and Anatomical Diagnosis of Coronary Artery Stenoses Abhijit Sinha Roy, Ph.D.,* Martin R. Back, M.D.,** Saeb F. Khoury,

More information

Effects of Increasing Doses of Intracoronary Adenosine on the Assessment of Fractional Flow Reserve

Effects of Increasing Doses of Intracoronary Adenosine on the Assessment of Fractional Flow Reserve JACC: CARDIOVASCULAR INTERVENTIONS VOL. 4, NO. 10, 2011 2011 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ISSN 1936-8798/$36.00 PUBLISHED BY ELSEVIER INC. DOI: 10.1016/j.jcin.2011.08.004 Effects of

More information

FFR-guided Jailed Side Branch Intervention

FFR-guided Jailed Side Branch Intervention FFR-guided Jailed Side Branch Intervention - Pressure wire in Bifurcation lesions - Bon-Kwon Koo, MD, PhD Seoul National University Hospital, Seoul, Korea Bifurcation Lesions Bifurcation Lesions Still

More information

Debate Should we use FFR? I will say NO.

Debate Should we use FFR? I will say NO. Debate Should we use FFR? I will say NO. Hyeon-Cheol Gwon Cardiac and Vascular Center Samsung Medical Center Sungkyunkwan University School of Medicine Dr. Hyeon-Cheol Gwon Research fund from Abbott Korea

More information

T he coronary collateral circulation is an alternative source

T he coronary collateral circulation is an alternative source 1289 CARDIOVASCULAR MEDICINE Influence of diabetes mellitus on coronary collateral flow: an answer to an old controversy R Zbinden, S Zbinden, M Billinger, S Windecker, B Meier, C Seiler... See end of

More information

IMR in acute STEMI and clinical outcomes

IMR in acute STEMI and clinical outcomes IMR in acute STEMI and clinical outcomes Professor Colin Berry Golden Jubilee National Hospital, University of Glasgow. ETP Coronary Physiology 24 April 2015 Disclosures Speaker - Shire Pharmaceuticals,

More information

Diffuse Disease and Serial Stenoses

Diffuse Disease and Serial Stenoses Diffuse Disease and Serial Stenoses Bernard De Bruyne Cardiovascular Center Aalst Belgium Atherosclerosis is a Diffuse Disease Serial Stenoses A B P a P m P d When A is isolated, hyperemic flow through

More information

FFR vs icecg in Coronary Bifurcations FIESTA ClinicalTrials.gov Identifier: NCT

FFR vs icecg in Coronary Bifurcations FIESTA ClinicalTrials.gov Identifier: NCT FFR vs icecg in Coronary Bifurcations FIESTA ClinicalTrials.gov Identifier: NCT01724957 Dobrin Vassilev MD, PhD Assoc. Prof. in Cardiology Head Cardiology Clinic, Alexandrovska University Hospital Medical

More information

Coronary Artery Disease

Coronary Artery Disease Coronary Artery Disease Fractional Flow Reserve Assessment of Left Main Stenosis in the Presence of Downstream Coronary Stenoses Andy S.C. Yong, MBBS, PhD*; David Daniels, MD*; Bernard De Bruyne, MD, PhD;

More information

The coronary flow reserve assesses the physiological

The coronary flow reserve assesses the physiological Microvascular Dysfunction in Chronic Total Coronary Occlusions Gerald S. Werner, MD; Markus Ferrari, MD; Barbara M. Richartz, MD; Oliver Gastmann, MD; Hans R. Figulla, MD Background Microvascular dysfunction

More information

Effect of Intravascular Ultrasound- Guided vs. Angiography-Guided Everolimus-Eluting Stent Implantation: the IVUS-XPL Randomized Clinical Trial

Effect of Intravascular Ultrasound- Guided vs. Angiography-Guided Everolimus-Eluting Stent Implantation: the IVUS-XPL Randomized Clinical Trial Effect of Intravascular Ultrasound- Guided vs. Angiography-Guided Everolimus-Eluting Stent Implantation: the IVUS-XPL Randomized Clinical Trial Myeong-Ki Hong, MD. PhD on behalf of the IVUS-XPL trial investigators

More information

In the presence of obstructive or occlusive coronary artery

In the presence of obstructive or occlusive coronary artery Is There Functional Collateral Flow During Vascular Occlusion in Normal? Kerstin Wustmann, MD; Stephan Zbinden, MD; Stephan Windecker, MD; Bernhard Meier, MD; Christian Seiler, MD Background Thus far,

More information

Focus on Acute Coronary Syndromes

Focus on Acute Coronary Syndromes Focus on Acute Coronary Syndromes Emanuele Barbato, MD, PhD, FESC Cardiovascular Center Aalst, Belgium Potential conflicts of interest Consulting fees and honoraria on my behalf go to the Cardiovascular

More information

PressureWire Aeris with Agile Tip Technology. Wireless FFR Functionality and Handles like a Workhorse PCI Guidewire 1

PressureWire Aeris with Agile Tip Technology. Wireless FFR Functionality and Handles like a Workhorse PCI Guidewire 1 Home» Products» All International Products» PressureWire Aeris with Agile Tip Technology PressureWire Aeris with Agile Tip Technology This device is commercially available for use in select international

More information

FFR: Tips and Tricks. A/Prof (Adj) Yeo Khung Keong, MBBS, FAMS, FACC, FSCAI National Heart Centre Singapore

FFR: Tips and Tricks. A/Prof (Adj) Yeo Khung Keong, MBBS, FAMS, FACC, FSCAI National Heart Centre Singapore FFR: Tips and Tricks A/Prof (Adj) Yeo Khung Keong, MBBS, FAMS, FACC, FSCAI National Heart Centre Singapore Disclosures Abbott Vascular: Speaker, Proctor (MitraClip) Boston Scientific: Consultant, honorarium

More information

Several intracoronary physiological parameters have been

Several intracoronary physiological parameters have been Hyperemic Stenosis Resistance Index for Evaluation of Functional Coronary Lesion Severity Martijn Meuwissen, MD; Maria Siebes, PhD; Steven A.J. Chamuleau, MD; Berthe L.F. van Eck-Smit, MD; Karel T. Koch,

More information

Metabolic Syndrome Is A Key Determinant Of Coronary Microvascular Function In Patients With Stable Coronary Disease Undergoing PCI

Metabolic Syndrome Is A Key Determinant Of Coronary Microvascular Function In Patients With Stable Coronary Disease Undergoing PCI Metabolic Syndrome Is A Key Determinant Of Coronary Microvascular Function In Patients With Stable Coronary Disease Undergoing PCI Narbeh Melikian*, Ajay M Shah*, Martyn R Thomas, Roy Sherwood, Mark T

More information

Low fractional diastolic pressure in the ascending aorta increased the risk of coronary heart disease

Low fractional diastolic pressure in the ascending aorta increased the risk of coronary heart disease (2002) 16, 837 841 & 2002 Nature Publishing Group All rights reserved 0950-9240/02 $25.00 www.nature.com/jhh ORIGINAL ARTICLE Low fractional diastolic pressure in the ascending aorta increased the risk

More information

How to Evaluate Microvascular Function and Angina. Myeong-Ho Yoon Ajou University Hospital

How to Evaluate Microvascular Function and Angina. Myeong-Ho Yoon Ajou University Hospital How to Evaluate Microvascular Function and Angina Myeong-Ho Yoon Ajou University Hospital Angina without Coronary Artery Disease (CAD) Prevalence: 20-30% going c-angiography, with a higher prevalence (almost

More information

CPT Code Details

CPT Code Details CPT Code 93572 Details Code Descriptor Intravascular Doppler velocity and/or pressure derived coronary flow reserve measurement (coronary vessel or graft) during coronary angiography including pharmacologically

More information

ORIGINAL ARTICLE. Abstract. Introduction

ORIGINAL ARTICLE. Abstract. Introduction ORIGINAL ARTICLE A Comparison between the Instantaneous Wave-free Ratio and Resting Distal Coronary Artery Pressure/Aortic Pressure and the Fractional Flow Reserve: The Diagnostic Accuracy CanBeImprovedbytheUseofbothIndices

More information

Impact of coronary atherosclerotic burden on clinical presentation and prognosis of patients with coronary artery disease

Impact of coronary atherosclerotic burden on clinical presentation and prognosis of patients with coronary artery disease Impact of coronary atherosclerotic burden on clinical presentation and prognosis of patients with coronary artery disease Gjin Ndrepepa, Tomohisa Tada, Massimiliano Fusaro, Lamin King, Martin Hadamitzky,

More information

Solving the Dilemma of Ostial Stenting: A Case Series Illustrating the Flash Ostial System

Solving the Dilemma of Ostial Stenting: A Case Series Illustrating the Flash Ostial System Volume 1, Issue 1 Case Report Solving the Dilemma of Ostial Stenting: A Case Series Illustrating the Flash Ostial System Robert F. Riley * and Bill Lombardi University of Washington Medical Center, Division

More information

FFR= Qs/Qn. Ohm s law R= P/Q Q=P/R

FFR= Qs/Qn. Ohm s law R= P/Q Q=P/R 32 ο Πανελλήνιο Καρδιολογικό Συνζδριο, Θεσσαλονίκη 20/10/2011 Gould KL et al, JACC CARDIOVASC IMAG 2009 Gould KL et al AM J CARDIOL 1974 & JACC CARDIOVASC IMAG 2009 Under maximal hyperemia: Rs=Rn FFR=

More information

Σεμινάριο Ομάδων Εργασίας Fractional Flow Reserve (FFR) Σε ποιούς ασθενείς; ΔΗΜΗΤΡΗΣ ΑΥΖΩΤΗΣ Επιστ. υπεύθυνος Αιμοδυναμικού Τμήματος, Βιοκλινική

Σεμινάριο Ομάδων Εργασίας Fractional Flow Reserve (FFR) Σε ποιούς ασθενείς; ΔΗΜΗΤΡΗΣ ΑΥΖΩΤΗΣ Επιστ. υπεύθυνος Αιμοδυναμικού Τμήματος, Βιοκλινική ΕΛΛΗΝΙΚΗΚΑΡΔΙΟΛΟΓΙΚΗΕΤΑΙΡΕΙΑ Σεμινάριο Ομάδων Εργασίας 2011 Fractional Flow Reserve (FFR) Σε ποιούς ασθενείς; ΔΗΜΗΤΡΗΣ ΑΥΖΩΤΗΣ Επιστ. υπεύθυνος Αιμοδυναμικού Τμήματος, Βιοκλινική GUIDELINES ON MYOCARDIAL

More information

Between Coronary Angiography and Fractional Flow Reserve

Between Coronary Angiography and Fractional Flow Reserve Visual-Functional Mismatch Between Coronary Angiography and Fractional Flow Reserve Seung-Jung Park, MD., PhD. University of Ulsan, College of Medicine Asan Medical Center, Seoul, Korea Visual - Functional

More information

Z S Kyriakides, A Antoniadis, T M Kolettis, D T Kremastinos

Z S Kyriakides, A Antoniadis, T M Kolettis, D T Kremastinos Heart 1998;80:493 498 493 Second Department of Cardiology, Onassis Cardiac Surgery Centre, Athens, Greece Z S Kyriakides A Antoniadis T M Kolettis D T Kremastinos Correspondence to: Dr Z S Kyriakides,

More information

Coronary Physiology the current state of play

Coronary Physiology the current state of play Coronary Physiology the current state of play Background The concept of using the trans-stenotic pressure gradient in a diseased coronary artery as a measure to guide percutaneous coronary intervention

More information

Contrast Induced Nephropathy

Contrast Induced Nephropathy Contrast Induced Nephropathy O CIAKI refers to an abrupt deterioration in renal function associated with the administration of iodinated contrast media O CIAKI is characterized by an acute (within 48 hours)

More information

Anatomy is Destiny, But Physiology is Here Today

Anatomy is Destiny, But Physiology is Here Today Published on Journal of Invasive Cardiology (http://www.invasivecardiology.com) September, 2010 [1] Anatomy is Destiny, But Physiology is Here Today Thu, 9/9/10-10:54am 0 Comments Section: Commentary Issue

More information

Fractional Flow Reserve (FFR)

Fractional Flow Reserve (FFR) Non-invasive FFR using coronary CT angiography and computational fluid dyn amics predicts the hemodynamic signifi cance of coronary lesions First in man experience with CT-Flow Andrejs Erglis, Sanda Jegere,

More information

For Personal Use. Copyright HMP 2013

For Personal Use. Copyright HMP 2013 Case Report J INVASIVE CARDIOL 2013;25(2):E39-E41 A Case With Successful Retrograde Stent Delivery via AC Branch for Tortuous Right Coronary Artery Yoshiki Uehara, MD, PhD, Mitsuyuki Shimizu, MD, PhD,

More information

Cardiac transplant related arteriopathy remains a leading

Cardiac transplant related arteriopathy remains a leading Simultaneous Assessment of Fractional and Coronary Flow Reserves in Cardiac Transplant Recipients Physiologic Investigation for Transplant Arteriopathy (PITA Study) William F. Fearon, MD; Mamoo Nakamura,

More information

CLINICAL CONSEQUENCES OF THE

CLINICAL CONSEQUENCES OF THE CLINICAL CONSEQUENCES OF THE FAME STUDY TCT ASIA Seoul, Korea, april 26 th, 2012 Nico H. J. Pijls, MD, PhD Catharina Hospital, Eindhoven, The Netherlands GUIDELINES ESC SEPTEMBER 2010 FFR UPGRADED TO LEVEL

More information

Intracoronary Continuous Adenosine Infusion

Intracoronary Continuous Adenosine Infusion Circ J 2005; 69: 908 912 Intracoronary Continuous Adenosine Infusion A Novel and Effective Way of Inducing Maximal Hyperemia for Fractional Flow Reserve Measurement Bon-Kwon Koo, MD, PhD; Cheol-Ho Kim,

More information

Novel monorail infusion catheter for volumetric coronary blood flow measurement in humans: in vitro validation

Novel monorail infusion catheter for volumetric coronary blood flow measurement in humans: in vitro validation CLINICAL RESEARCH CORONARY INTERVENTIONS Novel monorail infusion catheter for volumetric coronary blood flow measurement in humans: in vitro validation Marcel van t Veer 1,2, MSc, PhD; Julien Adjedj 3,

More information

Many studies have emphasized the importance of collateral

Many studies have emphasized the importance of collateral Validation of Collateral Fractional Flow Reserve by Myocardial Perfusion Imaging Hitoshi Matsuo, MD; Sachiro Watanabe, MD; Tohru Kadosaki, MD; Takahiko Yamaki, MD; Shinichiro Tanaka, MD; Shuusaku Miyata,

More information

Coronary Flow Pressure Relationship Distal to Epicardial Stenosis

Coronary Flow Pressure Relationship Distal to Epicardial Stenosis Circ J 2003; 67: 525 529 Coronary Flow Pressure Relationship Distal to Epicardial Stenosis Nobuhiro Tanaka, MD; Kenji Takazawa, MD; Kazuhiro Takeda, MD; Masaru Aikawa, MD; Naohisa Shindo, MD; Kazutaka

More information

Declaration of conflict of interest. Nothing to disclose

Declaration of conflict of interest. Nothing to disclose Declaration of conflict of interest Nothing to disclose Hong-Seok Lim, Seung-Jea Tahk, Hyoung-Mo Yang, Jin-Woo Kim, Kyoung- Woo Seo, Byoung-Joo Choi, So-Yeon Choi, Myeong-Ho Yoon, Gyo-Seung Hwang, Joon-Han

More information

ARTICLE IN PRESS. Journal of Biomechanics

ARTICLE IN PRESS. Journal of Biomechanics Journal of Biomechanics 42 (2009) 2753 2759 Contents lists available at ScienceDirect Journal of Biomechanics journal homepage: www.elsevier.com/locate/jbiomech www.jbiomech.com Influence of coronary collateral

More information

Side Branch Occlusion

Side Branch Occlusion Side Branch Occlusion Mechanism, Outcome, and How to avoid it From COBIS II Registry Hyeon-Cheol Gwon Cardiac&Vascular Center, Samsung Medical Center Sungkyunkwan University School of Medicine SB occlusion

More information

Acetylcholine coronary spasm provocation testing: Revaluation in the real clinical practice

Acetylcholine coronary spasm provocation testing: Revaluation in the real clinical practice Research article Interventional Cardiology Acetylcholine coronary spasm provocation testing: Revaluation in the real clinical practice Background: Japanese Circulation Society guidelines for coronary spastic

More information

Am J Physiol Heart Circ Physiol 300: H382 H387, First published October 8, 2010; doi: /ajpheart

Am J Physiol Heart Circ Physiol 300: H382 H387, First published October 8, 2010; doi: /ajpheart Am J Physiol Heart Circ Physiol 300: H382 H387, 2011. First published October 8, 2010; doi:10.1152/ajpheart.00412.2010. Influence of heart rate on fractional flow reserve, pressure drop coefficient, and

More information

PressureWire Agile Tip Technology

PressureWire Agile Tip Technology PressureWire Agile Tip Technology FINALLY. INTRODUCING PRESSUREWIRE WITH AGILE TIP St. Jude Medical continues to set new standards in fractional flow reserve (FFR) measurement. With the next generation

More information

Coronary pressure derived fractional flow reserve (FFR)

Coronary pressure derived fractional flow reserve (FFR) Pressure-Derived Fractional Flow Reserve to Assess Serial Epicardial Stenoses Theoretical Basis and Animal Validation Bernard De Bruyne MD PhD; Nico HJ Pijls MD PhD; Guy R Heyndrickx MD PhD; Dominique

More information

Comparison between Myocardial Ischemia Evaluation by Fractional Flow Reserve and Myocardial Perfusion Scintigraphy

Comparison between Myocardial Ischemia Evaluation by Fractional Flow Reserve and Myocardial Perfusion Scintigraphy 1 International Journal of Cardiovascular Sciences. 2018; [online].ahead print, PP.0-0 ORIGINAL ARTICLE Comparison between Myocardial Ischemia Evaluation by Fractional Flow Reserve and Myocardial Perfusion

More information

Coronary Artery Dissection Following Angioplasty Evaluated by Fractional Flow Reserve: Report of Three Cases

Coronary Artery Dissection Following Angioplasty Evaluated by Fractional Flow Reserve: Report of Three Cases J Cardiol 2001 Dec; 386: 337 342 : 3 Coronary Artery Dissection Following Angioplasty Evaluated by Fractional Flow Reserve: Report of Three Cases Kazutaka AMAYA, MD Kenji TAKAZAWA, MD, FJCC Nobuhiro TANAKA,

More information

Introducing. Integrated FFR Platform

Introducing. Integrated FFR Platform Introducing Integrated FFR Platform Fractional Flow Reserve (FFR) Definition of FFR Maximum achievable blood flow in stenotic coronary artery divided by maximum blood flow in the same artery without stenosis.

More information

Women and Coronary Artery Disease:

Women and Coronary Artery Disease: Women and Coronary Artery Disease: Less is More? Interventional Cardiology 2017 32 th Annual International Symposium Snow Mass March 5-10 Yolande Appelman MD, PhD,FESC EAPCI-Women Interventional Cardiologist

More information

The Future of Coronary Physiology

The Future of Coronary Physiology The Future of Coronary Physiology Morton J. Kern, MD Chief of Medicine, VA Long Beach HCS Professor of Medicine University California Irvine Orange, California Disclosure: Morton J. Kern, MD Within the

More information

DEB experience in Gachon Universtiy Gil Hospital (in ISR) Soon Yong Suh MD., PhD. Heart Center Gachon University Gil Hospital Seoul, Korea.

DEB experience in Gachon Universtiy Gil Hospital (in ISR) Soon Yong Suh MD., PhD. Heart Center Gachon University Gil Hospital Seoul, Korea. DEB experience in Gachon Universtiy Gil Hospital (in ISR) Soon Yong Suh MD., PhD. Heart Center Gachon University Gil Hospital Seoul, Korea. In-stent restenosis (ISR) Remains important issue even in the

More information

Fractional Flow Reserve and the 1 Year Results of the FAME Study

Fractional Flow Reserve and the 1 Year Results of the FAME Study Imaging and Physiology Summit Seoul, Korea, November 22, 2008 Fractional Flow Reserve and the 1 Year Results of the FAME Study William F. Fearon, M.D. Assistant Professor Division of Cardiovascular Medicine

More information

Effectiveness of IVUS in Complex Cases

Effectiveness of IVUS in Complex Cases Effectiveness of IVUS in Complex Cases Satoru Sumituji,M.D. Rinku General Medical Center IVUS is can provide images of the vessel wall and the tissue around the vessel which cannot be viewed by angiography.

More information

Safety and Efficacy of Adenosine 5 -Triphosphate as a Hyperemic Agent for the Assessment of Peripheral Fractional Flow Reserve

Safety and Efficacy of Adenosine 5 -Triphosphate as a Hyperemic Agent for the Assessment of Peripheral Fractional Flow Reserve Showa Univ J Med Sci 30 2, 151 158, June 2018 Original Safety and Efficacy of Adenosine 5 -Triphosphate as a Hyperemic Agent for the Assessment of Peripheral Fractional Flow Reserve Atsuo MAEDA 1 2, Kazuma

More information

Qualitative and Quantitative Assessment of Perfusion

Qualitative and Quantitative Assessment of Perfusion APCDE 2011 Qualitative and Quantitative Assessment of Perfusion Hyun Ju Yoon Chonnam National University Hospital Gwangju, Korea ISCHEMIC CASCADE Blood flow mismatch Perfusion defects on nuclear imaging,

More information

The Impact of Sex Differences on Fractional Flow Reserve Guided Percutaneous Coronary Intervention

The Impact of Sex Differences on Fractional Flow Reserve Guided Percutaneous Coronary Intervention JACC: CARDIOVASCULAR INTERVENTIONS VOL. 5, NO. 10, 2012 2012 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ISSN 1936-8798/$36.00 PUBLISHED BY ELSEVIER INC. http://dx.doi.org/10.1016/j.jcin.2012.06.016

More information

FFR in Left Main Disease

FFR in Left Main Disease FFR in Left Main Disease William F. Fearon, MD Associate Professor of Medicine Director, Interventional Cardiology Stanford University Medical Center Why FFR instead of IVUS? Physiologic versus anatomic

More information

Fractional Flow Reserve: Review of the latest data

Fractional Flow Reserve: Review of the latest data Fractional Flow Reserve: Review of the latest data Michalis Hamilos, MD, PhD, FESC University Hospital of Heraklion Fractional Flow Reserve (FFR) Coronary angiography does not always tell the truth Most

More information

Imaging ischemic heart disease: role of SPECT and PET. Focus on Patients with Known CAD

Imaging ischemic heart disease: role of SPECT and PET. Focus on Patients with Known CAD Imaging ischemic heart disease: role of SPECT and PET. Focus on Patients with Known CAD Hein J. Verberne Academic Medical Center, University of Amsterdam, Amsterdam, Netherlands International Conference

More information

No-reflow Phenomenon in Patients with Acute Myocardial Infarction: Its Pathophysiology and Clinical Implications

No-reflow Phenomenon in Patients with Acute Myocardial Infarction: Its Pathophysiology and Clinical Implications No-reflow Phenomenon in Patients with Acute Myocardial Infarction: Its Pathophysiology and Clinical Implications * 164 Ito Acta Med. Okayama Vol. 63, No. 4 Normal case Anterior MI Fig. 3 Myocardial contrast

More information

Georgios Pavlakis. Consultant Interventional Cardiologist. K.A.T. General Hospital of Athens, GREECE

Georgios Pavlakis. Consultant Interventional Cardiologist. K.A.T. General Hospital of Athens, GREECE Georgios Pavlakis Consultant Interventional Cardiologist K.A.T. General Hospital of Athens, GREECE Male patient, 72 years-old was admitted because of STEMI of the Inferior wall. Pre-cathlab ECG : ST Elevation

More information

TCT mdbuyline.com Clinical Trial Results Summary

TCT mdbuyline.com Clinical Trial Results Summary TCT 2012 Clinical Trial Results Summary FAME2 Trial: FFR (fractional flow reserve) guided PCI in all target lesions Patients with significant ischemia, randomized 1:1 Control arm: not hemodynamically significant

More information

High-dose atorvastatin pretreatment could diminishes microvascular impairment in patients undergoing elective percutaneous coronary intervention

High-dose atorvastatin pretreatment could diminishes microvascular impairment in patients undergoing elective percutaneous coronary intervention Journal of Geriatric Cardiology (2013) 10: 355 360 2013 JGC All rights reserved; www.jgc301.com Research Article Open Access High-dose atorvastatin pretreatment could diminishes microvascular impairment

More information

FFR vs. icecg in Coronary Bifurcations (FIESTA) - preliminary results. Dobrin Vassilev MD, PhD National Heart Hospital Sofia, Bulgaria

FFR vs. icecg in Coronary Bifurcations (FIESTA) - preliminary results. Dobrin Vassilev MD, PhD National Heart Hospital Sofia, Bulgaria FFR vs. icecg in Coronary Bifurcations (FIESTA) - preliminary results Dobrin Vassilev MD, PhD National Heart Hospital Sofia, Bulgaria I would like to express my personal gratitude to Dr. BK Koo for opening

More information

Diagnostic Accuracy of Fractional Flow Reserve from Anatomic Computed TOmographic Angiography: The DeFACTO Study

Diagnostic Accuracy of Fractional Flow Reserve from Anatomic Computed TOmographic Angiography: The DeFACTO Study Diagnostic Accuracy of Fractional Flow Reserve from Anatomic Computed TOmographic Angiography: The DeFACTO Study James K. Min 1 ; Jonathon Leipsic 2 ; Michael J. Pencina 3 ; Daniel S. Berman 1 ; Bon-Kwon

More information

Ischemic Postconditioning During Primary Percutaneous Coronary Intervention Mechanisms and Clinical Application Jian Liu, MD FACC FESC FSCAI Chief Phy

Ischemic Postconditioning During Primary Percutaneous Coronary Intervention Mechanisms and Clinical Application Jian Liu, MD FACC FESC FSCAI Chief Phy Ischemic Postconditioning During Primary Percutaneous Coronary Intervention Mechanisms and Clinical Application Jian Liu, MD FACC FESC FSCAI Chief Physician, Professor of Medicine Department of Cardiology,

More information

PCI for Chronic Total Occlusions

PCI for Chronic Total Occlusions PCI for Chronic Total Occlusions Chronic Total Occlusions Why not Medical Treatment? Medical Treatment CTO in 891 pts over 24 years High 10% Mortality Low 2 % 1 year 10 years Puma JA, et al. JACC 1994;23:390A

More information

Ventricular Arrhythmias in Acute MI Patients Undergoing Primary PCI

Ventricular Arrhythmias in Acute MI Patients Undergoing Primary PCI Ventricular Arrhythmias in Acute MI Patients Undergoing Primary PCI Bulent Gorenek MD FACC FESC Eskişehir Osmangazi University Cardiology Department Eskisehir-Turkey I do not have any potential conflict

More information

Zachary I. Hodes, M.D., Ph.D., F.A.C.C.

Zachary I. Hodes, M.D., Ph.D., F.A.C.C. Zachary I. Hodes, M.D., Ph.D., F.A.C.C. Disclamer: I personally have no financial relationship with any company mentioned today. The Care Group, LLC does have a contract with Cardium to participate in

More information