The 1st transvascular patent ductus arteriosus
|
|
- Katrina Diana Randall
- 5 years ago
- Views:
Transcription
1 J Vet Intern Med 2012;26:85 92 Occlusion Devices and Approaches in Canine Patent Ductus Arteriosus: Comparison of Outcomes M. K. Singh, M. D. Kittleson, P. H. Kass, and L. G. Griffiths Background: A comparison of transvascular occlusion methods for closing patent ductus arteriosus (PDA) in dogs has not been done. Objectives: To determine if clinically important differences exist between the approaches and devices currently used. Animals: A total of 112 client-owned dogs with left-to-right shunting PDA. Methods: Retrospective study. Records from dogs that underwent attempted transvascular PDA occlusion from January 2006 to December 2009 were examined. Dogs were placed into 4 groups: Group 1: Amplatz Canine Duct Occluder (ACDO) (transarterial) 36 dogs; Group 2: Gianturco or MReye Flipper Detachable Embolization (Flipper) coil (transarterial) 38 dogs; Group 3: Amplatzer Vascular Plug (AVP) (transarterial) 23 dogs; Group 4: Flipper coil (transvenous) 15 dogs. Results: The overall success rate of the procedures was high (92%) with comparable success rates among groups (87 97%). There were significantly fewer complications (P <.0001) in dogs receiving an ACDO than in the remaining groups (3% for ACDO versus 26 33% for the other groups). Fluoroscopy time for the transvenous method was significantly longer (median, 13 minutes) than for the other groups (median, 6 minutes) (P <.0001). Severity of residual flow 24 hours postprocedure was significantly less in the ACDO group than in the remaining groups (P = ). Conclusions: The ACDO appears superior in ease of use, complication rate, and completeness of occlusion. The remaining limiting factor with this device is patient size. Until a smaller ACDO device is marketed, coils remain the only choice for interventional closure in very small dogs. Key words: Congenital heart defects; Ductal occluder; Interventional procedure; Thrombogenic coil. The 1st transvascular patent ductus arteriosus (PDA) closure was performed in a dog in a Since then, numerous reports have described minimally invasive transvascular PDA occlusion in dogs, and it now has become the treatment method of choice for most veterinary cardiologists The procedure historically has encompassed a variety of occlusion devices intended for human use, including thrombogenic coils, 1,4,10 12,14,15 the AVP 7,9 and the Amplatzer Duct Occluder. 2,3 Both transvenous and transarterial routes for deployment of occlusion devices have been used. 1,3,4,9 12,15 All techniques and devices have potential limitations and complications and no single technique appears suitable for all dogs. 1,3,4,7 9 In recent years, the Amplatz canine duct occluder (ACDO), 6,8 a nitinol mesh device specifically developed to fit the size and shape of the canine PDA, appears to have surpassed previous devices with regard to ease of use, degree of closure, and decreased complication rate. 6,8,13 Although there are individual reports on the From the William R. Pritchard Veterinary Medical Teaching Hospital (Singh); the Department of Medicine and Epidemiology (Kittleson, Griffiths); and the Department of Population Health and Reproduction (Kass), University of California, Davis, CA. This abstract was presented at the 2011 ACVIM Forum, Denver, Colorado. This study was done at the William R. Pritchard Veterinary Medical Teaching Hospital, University of California, Davis, Dr Singh s present address is the Small Animal Specialist Hospital, Level 1, 1 Richardson Place, North Ryde, Australia, Corresponding author: Mark D. Kittleson, Department of Medicine and Epidemiology, One Shields Avenue, University of California, Davis, CA; mdkittleson@ucdavis.edu. Submitted March 13, 2011; Revised October 13, 2011; Accepted November 17, Copyright 2011 by the American College of Veterinary Internal Medicine /j x Abbreviations: ACDO Amp Ao AVP LA MDD PDA VMTH PA Amplatz canine duct occluder ampulla aorta Amplatzer vascular plug left atrium minimal ductal diameter patent ductus arteriosus Veterinary Medical Teaching Hospital pulmonary artery use of these devices in dogs, the indications, complications, and long-term outcomes of the various devices has not been compared in 1 study. This study aims to compare the signalment, fluoroscopy time, outcomes, and complications among the 4 devices placed by 2 different approaches that were used to noninvasively close PDAs in dogs at our institution from January 2006 (when the ACDO became available at our facility) to December Materials and Methods Initial and follow-up records from all dogs that underwent attempted transvascular occlusion to correct PDA at the University of California, Davis, William R. Pritchard Veterinary Medical Teaching Hospital (VMTH) from January 2006 to December 2009 were retrospectively reviewed. Information extracted from the records included age, sex, body weight, presence of congestive heart failure, presence of other cardiac abnormalities, PDA morphology, minimal ductal diameter (MDD), ampulla diameter (amp), aortic diameter (Ao), type of occlusion device, vascular approach, fluoroscopy time, residual flow at time 0 (within minutes after device deployment) and 24 hours, left atrial: aorta ratio (LA:Ao), left ventricular internal diameter in diastole: aorta ratio (LVIDd:Ao) before occlusion and at 24 hours,
2 86 Singh et al complications, and success rate. In addition, the MDD:Ao and amp:ao ratios were calculated. Absence of clinically relevant pulmonary hypertension was verified by applying the modified Bernoulli equation to the ductal jet velocity. The presence of a PDA was confirmed and severity assessed by echocardiography, b,c which was performed in all dogs as previously described. 16 Assessment of severity was made based on the extent of LA enlargement, left ventricular volume overload, and presence or absence of systolic dysfunction. General anesthesia was induced using a variety of sedative and anesthetic agents as deemed appropriate by the Anesthesia Service at the VMTH. Transesophageal echocardiography d was performed if the patient was large enough in order to visualize the PDA and to obtain maximum ampulla and MDD for subsequent device selection. If the patient was not large enough, these variables were measured transthoracically and, in some, by angiography. The transvascular occlusion procedure was performed by the transarterial or transvenous route depending on the preference of the faculty clinician at the time. These procedures have been described previously in detail. 1,4,6,8,12,15 Briefly, dogs were positioned in right lateral recumbency and the right inguinal region prepared for isolation of the proximal right femoral artery or vein. For the transarterial method, the artery was surgically isolated and the largest end-hole catheter (3 8 F) e that could be inserted into the artery was advanced to the transverse or proximal descending aorta under fluoroscopic guidance. f Iodinated contrast material, g at a dosage of approximately 1 ml/kg, was rapidly injected manually to identify the position of the ductal ampulla and to measure ductal diameter in some instances. For the transvenous method, vascular access was attempted percutaneously using a modified Seldinger technique. 17 If this was not successful after multiple attempts, the femoral vein was surgically isolated. A 4 or 5 French introducer sheath then was inserted. h A balloon wedge catheter i was inserted under fluoroscopic guidance into the main pulmonary artery (PA) and then passed retrograde through the PDA and into the aorta. If passage of the balloon wedge catheter retrograde through the PDA was not successful, this was facilitated by first passing a guide wire. k The balloon wedge catheter then was fed over the guide wire and the wire removed. An exchange guide wire j was passed through the balloon catheter into the descending aorta and the balloon catheter exchanged for a 4 or 5 Fr angiographic catheter. l An angiogram was then performed as described previously. Angiograms were reviewed and ductal shape identified as type I, IIa, IIb, or III as previously described. 18 Occlusion devices used included detachable and nondetachable Gianturco or MReye Flipper Detachable Embolization coils, m AVP, n and ACDO. o The device used was chosen based on the largest gauge catheter that could be inserted in the vessel or introducer, the size of the MDD and ampulla and faculty clinician preference. The ACDO waist size was chosen to be the MDD, 6,8 coil diameters were generally chosen to be at least twice the MDD 1,4,15 and the AVP diameter was chosen to be approximately 1.39 the maximal ampulla diameter. 7,9 Coils were deposited within the ampulla of the PDA. If the transvenous method was used, most of the coil was placed in the ampulla with 1 2 loops spanning the MDD into the PA. 10,12 If the transarterial method was used, the spanning method was not performed and all coil loops were placed in the ductal ampulla. The AVP was deposited in the ampulla of the PDA using the transarterial method. 9 The ACDO was always deployed by the transarterial method as previously described. 6,8 Devices that did not appear stable or resulted in ongoing marked continuous flow were removed and replaced with a larger device. For the majority of cases, direct systolic and diastolic blood pressures were monitored during and immediately after deployment of the device and the dog was auscultated for a continuous murmur. If direct blood pressure monitoring was not possible, systolic blood pressure was measured indirectly by Doppler methodology. When closure was performed by the transarterial route, an angiogram, and occasionally transthoracic or transesophageal echocardiography, was performed to identify residual flow. For the transvenous approach, residual flow after deployment was assessed using transthoracic echocardiography. Cefazolin p was administered at a dosage of 30 mg/kg IV q2h during the procedure in some patients depending on clinician preference. Dogs were monitored continuously until complete recovery from anesthesia. Within 24 hours of recovery, thoracic radiographs (lateral and dorsoventral) were taken to confirm device location. Echocardiography also was performed the day after occlusion to identify any residual flow and to obtain LA and left ventricular measurements. A recheck echocardiogram was recommended 3 months later to look for delayed device embolization and residual flow, and to again obtain LA and left ventricular measurements. Data Analysis Patient records were retrospectively reviewed. When objective data were absent or subjectively reported, the original echocardiographic examination was reviewed by a single investigator (MKS) and objective measurements obtained. LA, LVIDd, MDD, and ampulla were indexed to the aortic diameter (LA:Ao, LVIDd:Ao, MDD:Ao, Amp:Ao) to take into account the various body sizes. 19,20 Residual ductal flow was classified using a modified scale of 0 4 by color flow Doppler echocardiography as previously described. 21 Briefly this was as follows: grade 0 no residual flow in the PA; grade 1 trivial flow at the entrance of the PDA to the main PA; grade 2 mild flow, where the jet of continuous flow passes through the PDA along the main PA but does not reach the PV; grade 3 moderate flow, defined as a broad turbulent jet in the main PA that reaches the PV. 21 In addition, dogs in which no closure was obtained by the device (ie, the jet of turbulent flow was similar in size to before the procedure), were classified as grade 4 (severe). Dogs were arranged into 4 groups depending on the device and type of vascular access. Group 1: ACDO (transarterial), Group 2: thrombogenic coil (transarterial), Group 3: AVP (transarterial), and Group 4: thrombogenic coil (transvenous). Statistical Analysis The data were not normally distributed, and the Kruskal Wallis test was used to identify differences in clinical measurements among the 4 groups. Dogs that had transvascular closure attempted using more than 1 device or route were grouped according to the initial procedure attempted for the purposes of statistical analysis. Average degree of residual flow was compared among groups using the previously described grading scheme. For categories that were significantly different, posthoc analysis using a Mann Whitney test was done to determine which groups within each category caused the significant difference. Holm s sequentially rejective method of multiple comparison adjustment was used to control Type I errors in posthoc analyses. 22 For paired data (ie, comparison of LA:Ao and LVIDd:Ao before the procedure and 24 hours later within each group), an exact Wilcoxon signed-ranks test was used. To look for differences in the distribution of PDA morphology among groups an exact chi-square test was used. For all tests, P <.05 was considered significant. Exact logistic regression models initially were created to perform univariate analysis of putative risk factors for complications, and variables with P-values <.15 were candidates for inclusion into a multiple logistic regression model. Group (with the ACDO group
3 Occlusion Devices in Patent Ductus Arteriosus 87 as reference) was included in all models as the variable of interest, and age, weight, and ductal diameter were included in all models as confounders. Only variables with improvement-in-fit likelihood ratio P-values <.05 were considered for inclusion in the final model. Results are presented as odds ratios (OR) and 95% confidence intervals (95% CI). Any individual with missing data for a particular variable was excluded from analysis for that variable. Statistical software was used. q Results Transvascular occlusion was attempted in 112 dogs over the 4-year period. All dogs had a continuous murmur best heard in the left axillary region. Grade of heart murmur ranged from II to VI and there was no significant difference among groups. The number of dogs in each group were Group 1 (ACDO) 36 dogs, Group 2 (transarterial thrombogenic coil) 38 dogs, Group 3 (AVP) 23 dogs, Group 4 (transvenous thrombogenic coil) 15 dogs. Of the 15 dogs in Group 4, 7 (47%) had the femoral vein surgically isolated for catheter introduction due to unsuccessful attempts to use the modified Seldinger technique. These dogs all weighed <3.5 kg. Transesophageal echocardiography was performed in dogs large enough to accommodate an adult-sized probe. Five dogs weighed <1.4 kg and could not have transesophageal echocardiography performed. Age ranged from 2 84 months (median, 7 months). The median age for group 1 was 10 months (range, 2 84 months), for group 2 was 6.5 months (range, 2 84 months), for group 3 was 10 months (range, 2 48 months), and for group 4 was 6 months (range, 3 36 months). Body weight ranged from 1 to 34 kg (median, 4.7 kg). The median body weight for group 1 was 12.5 kg (range, kg), group 2 was 2.2 kg (range, kg), group 3 was 6.1 kg (range, kg), and group 4 was 2 kg (range, kg). There was no significant difference in age among the 4 groups (Table 1). There was a significant difference in body weight among the 4 groups (P <.0001). Posthoc analysis showed that groups 1 and 3 had significantly higher body weight than groups 2 and 4 (P <.0001) but that groups 1 and 3 were not significantly different from each other and groups 2 and 4 were not significantly different from each other (Table 1). Eleven dogs had evidence of congestive heart failure before the occlusion, 6 (17%) from group 1, 2 (5%) from group 2, 3 (13%) from group 3, and 1 (7%) from group 4. Ten dogs had evidence of other structural heart disease based on echocardiography. These included 4 dogs in group 1 (2 with mild subaortic stenosis, 1 with mild mitral valve dysplasia, and 1 with both mild mitral and mild tricuspid valve dysplasia), 3 in group 2 (1 with severe valvular pulmonic stenosis, 1 with mild supravalvular pulmonic stenosis, and 1 with mild mitral regurgitation due to myxomatous mitral valve degeneration), 2 in group 3 (both with mild to moderate mitral valve dysplasia), and 1 in group 4 (with mild subaortic stenosis). Five dogs had a concurrent arrhythmia on the resting electrocardiogram before the occlusion procedure. All 5 dogs were in group 1 and consisted of 2 with atrial fibrillation, 1 with atrial fibrillation and ventricular premature complexes, 1 with ventricular premature complexes only, and 1 with both atrial and ventricular premature complexes. There was no evidence of clinically relevant pulmonary arterial hypertension in any dog, with pressure gradients across the PDA ranging from 80 to 140 mmhg (calculated from ductal jet velocity using the modified Bernoulli equation). Morphologically, the PDA in 3 dogs was classified as type I, in 86 as type IIa, in 20 as type IIb, and in none as type III. Three dogs could not be classified due to a lack of an angiogram in 2 (the femoral artery was too small to pass the catheter in 1 whereas the femoral artery tore in the other) and inability to visualize the PDA clearly in the 3rd. The morphology was evenly distributed among groups (P =.16). The PDA MDD ranged from 1.0 to 8 mm (median, 2.2 mm). The median PDA MDD for group 1 dogs was 2.9 mm (range, mm), for group 2 was 1.8 mm (range, mm), for group 3 was 2.3 mm (range, mm), and for group 4 was 1.7 mm Table 1. Demographic, echocardiographic, and fluoroscopic data (median and range). Entire population (n = 112) Group 1 (n = 36) Group 2 (n = 38) Group 3 (n = 23) Group 4 (n = 15) P Age (months) 7 (2 84) 10 (2 84) 6.5 (2 84) 10 (2 48) 6 (3 36).30 Body weight (kg) 4.7 (1 34) 12.5 a (2.6 34) 2.2 b ( ) 6.1 a (2.3 27) 2.0 b ( ) <.0001 PDA MDD (mm) 2.2 (1 8) 2.9 a (1.2 8) 1.8 b (1 3.1) 2.3 a (1.1 5) 1.7 b (1.1 4) <.0001 MDD/Ao 0.16 ( ) 0.16 ( ) 0.16 ( ) 0.14 ( ) 0.16 ( ).35 PDA amp (mm) 6.2 (2.5 23) 8 a (3 23) 4.6 b (3 8) 8.1 a (3 12.1) 4.4 b (3 12) <.0001 Amp/Ao 0.44 ( ) 0.44 ( ) 0.42 ( ) 0.44 ( ) 0.39 ( ).9 LA:Ao (prior) 1.5 ( ) 1.6 ( ) 1.5 ( ) 1.5 ( ) 1.4 ( ).25 LA:Ao (24 hours) 1.3 ( ) 1.3 ( ) 1.3 ( ) 1.3 ( ) 1.3 ( ).98 LVIDd:Ao (prior) 2.2 ( ) 2.2 ( ) 2.2 ( ) 2.1 ( ) 2.3 ( ).4 LVIDd/Ao (24 hours) 1.9 ( ) 1.8 ( ) 1.9 ( ) 1.8 ( ) 2.1 ( ).36 Fluoroscopy time (minutes) 8.0 (3 78) 5.1 a ( ) 8.0 c ( ) 5.3 a ( ) 13.0 b ( ) <.0001 a significantly different from b and c. MDD, minimal ductal diameter; LVIDd/Ao, left ventricular internal diameter in diastole:aorta ratio; LA:Ao, left atrial:aorta ratio; PDA, patent ductus arteriosus.
4 88 Singh et al (range, mm). There was a significant difference in MDD among the 4 groups (P <.0001) (Table 1). Posthoc analysis showed that groups 1 and 3 were significantly different from groups 2 and 4 (P <.0001), but that groups 1 and 3 were not significantly different from each other (P =.036) and groups 2 and 4 were not significantly different from each other (P =.77). When the MDD was indexed to the aorta to take into account the various body sizes (MDD/Ao) this value ranged from 0.02 to 0.32 (median 0.16). For group 1, the median MDD/Ao was 0.16 (range, ), for group 2 was 0.16 (range, ), for group 3 was 0.14 (range, ), and for group 4 was 0.16 (range, ). There was no significant difference among groups (Table 1). The PDA ampulla diameter (amp) ranged from 2.5 to 23 mm (median, 6.2 mm). For group 1, the median amp was 8 mm (range, 3 23 mm), for group 2 was 4.6 mm (range, 3 8 mm), for group 3 was 8.1 mm (range, mm) and for group 4 was 4.4 mm (range, 3 12 mm). There was a significant difference in amp diameter among groups (P <.0001). Posthoc analysis showed that groups 1 and 3 were significantly different from group 2 and 4 (P <.0001) but that groups 1 and 3 were not significantly different from each other (P =.45) and groups 2 and 4 were not significantly different from each other (P =.35). When the ampulla was indexed to the aorta in each dog (amp:ao) to take into account the various body sizes, the range was (median, 0.44). For group 1, the median amp:ao was 0.44 (range, ), for group 2 was 0.42 (range, ), for group 3 was 0.44 (range, ), and for group 4 was 0.39 (range, ). There was no significant difference among groups (Table 1). The LA:Ao ratio before the procedure ranged from 1 to 3.6 (median, 1.5). For group 1, the median LA: Ao ratio was 1.6 (range, ), for group 2 was 1.5 (range, ), for group 3 was 1.5 (range, ), and for group 4 was 1.4 (range, ). There was no significant difference in LA:Ao ratio among groups before the procedure. The LA:Ao ratio 24 hours after the procedure ranged from 1 to 2.4 (median, 1.3). For group 1, the median LA:Ao at 24 hours was 1.3 (range, ), for group 2 was 1.3 (range, ), for group 3 was 1.3 (range, ), and for group 4 was 1.3 (range, ). There was no significant difference in LA:Ao ratio among groups 24 hours after the procedure. The LVIDd:Ao ratio ranged from 1.4 to 3.9 (median, 2.2). For group 1, the median LVIDd: Ao ratio before to the procedure was 2.2 (range, ), for group 2 was 2.2 (range, ), for group 3 was 2.1 (range, ), and for group 4 was 2.3 (range, ). There was no significant difference in LVIDd:Ao ratio among groups before the procedure. The LVIDd:Ao 24 hours after the procedure ranged from 1.5 to 3.5 (median, 1.9). The median LVIDd:Ao 24 hours after the procedure for group 1 was 1.8 (range, ), for group 2 was 1.9 (range, ), for group 3 was 1.8 (range, ), and for group 4 was 2.1 (range, ). There was no significant difference in LVIDd:Ao among groups 24 hours after the procedure (Table 1). When the LA:Ao and LVIDd:Ao before the procedure was compared to that at 24 hours within each group, there was a significant reduction present for all groups and both parameters (P <.0001). Outcomes Complications occurred in 22/112 dogs (20%). Only 1 dog (3%) in group 1 (n = 36; ACDO) suffered a complication (no device large enough to close the PDA). Ten dogs (26%) in group 2 (n = 38; transarterial coil) suffered a complication. Five of these had pulmonary arterial embolization of a coil. In 4, the femoral artery tore, necessitating surgical closure of the PDA in 1 whereas the catheter procedure could be completed in the remaining 3. None of these dogs suffered clinically relevant bleeding. All 4 of these dogs were small ( 3 kg). In another dog, the artery was too small to advance a 4 Fr catheter necessitating surgical closure of the PDA. This dog weighed 0.8 kg. Six dogs (26%) in group 3 (n = 23; vascular plug) suffered a complication. These included immediate rotation of the device resulting in a marked increase in flow in 2 dogs, delayed recanalization or movement resulting in clinically relevant recrudescence of flow in 3 dogs and lack of thrombosis of the device due to Von Willebrand disease in 1 dog. Five dogs (33%) in group 4 (n = 15; transvenous coil) suffered complications. These included severe femoral vein hemorrhage requiring overnight monitoring in the intensive care unit in 1 dog in which the procedure was performed percutaneously, arterial laceration in 1 dog (this occurred during surgical isolation of the vein because the modified Seldinger technique was not successful), caudal vena caval perforation with bleeding into the retroperitoneal space in another dog in which the procedure was performed percutaneously, and inability to close the PDA in 2 dogs. There was no significant difference in complication rates among groups 2, 3, and 4 after controlling for age, weight, and ductal diameter. However, compared to group 1, complication proportions were significantly higher in group 2 (adjusted OR = 59.3, 95% CI = 2.1 1,704, P =.017), group 3 (adjusted OR = 38.8, 95% CI = , P =.013), and group 4 (adjusted OR = 92.2, 95% CI = 2.9 2,925, P =.010). Success for transvascular PDA occlusion was defined as a reduction of flow through the PDA to the point where it was considered hemodynamically insignificant. This was assessed by grade of residual flow using angiography in the minutes after deployment, an increase in diastolic blood pressure in those dogs that had direct arterial blood pressure monitoring performed, disappearance of the continuous heart murmur, and echocardiographic measurements at 24 hours and 3 months. If further intervention was deemed unnecessary, the procedure was considered successful. Transvascular PDA occlusion was successful in 103/112 dogs (92%). It was not successful in 1/ 36 group 1 dogs (3%). This was due to the PDA MDD being too big for the largest available ACDO
5 Occlusion Devices in Patent Ductus Arteriosus 89 occlusion device (14 mm). The ACDO device in this dog was unstable, and with gentle predetachment manipulation, continued to pass through the MDD into the PDA ampulla. In this dog, a 16 mm AVP also was unstable. This dog s PDA was successfully closed by surgery. PDA closure was not successful in 3/38 (8%) dogs in group 2. In 1 dog multiple coils embolized into the PA. In the 2 remaining dogs, very small femoral arteries precluded placement of a sufficiently sized catheter in 1, whereas the artery tore in the other. The PDA in all 3 dogs was successfully closed by surgery. Transvascular occlusion was not successful in 3/23 (13%) dogs in group 3. In 1 dog, the device recanalized 2 months postplacement. A 2nd larger AVP was placed that also resulted in ongoing marked residual flow. This dog s PDA eventually was surgically ligated. In another dog, the AVP was found to have rotated within the ductus, resulting in moderate residual flow with recurrence of volume overload to the left heart at the 3-month recheck. This dog s PDA also was successfully surgically ligated. The 3rd dog was a Doberman Pinscher that was later discovered to have Von Willebrand disease. This dog s PDA was not closed completely by the AVP resulting in ongoing moderate flow and volume overload. This dog was lost to follow-up. The procedure was not successful in 2/15 (13%) dogs in group 4. In both dogs, the Flipper coils were unstable in the PDA and resulted in severe ongoing residual flow or migration into the PA and were removed. One dog s PDA was successfully closed 1 month later using an ACDO. In the other dog, 10 mm and 12 mm AVPs were used but also were unsuccessful. The dog eventually was euthanized due to financial constraints. Total fluoroscopy time ranged from 3 to 78 minutes (median, 8 minutes). Median fluoroscopy time in group 4 (transvenous coil; 13 minutes), was significantly longer than in group 1 (ACDO; 5.1 minutes), group 2 (transarterial coil; 8 minutes), and group 3 (AVP; 5.3 minutes) (P <.0001). Fluoroscopy time in group 1 was significantly shorter than in group 2 (P =.0013) and group 4 (P <.0001) but not group 3 (P =.43) (Table 1). Residual flow was evaluated within minutes after device deployment by angiography (transarterial route) or echocardiography (transvenous route). All dogs had residual flow reevaluated the next morning by echocardiography. Overall, 48/110 (44%) of dogs had some residual flow immediately after deployment (dogs in which device deployment was never attempted were excluded). In group 1, 9/36 dogs (25%) had some immediate residual flow. Of these, 7 had grade 1 flow, 1 had grade 2 flow, and 1 had grade 4 flow (the device was removed and the dog underwent surgery). After 24 hours, 1/35 dogs (2.8%) had ongoing grade 1 residual flow. In group 2, 23/36 (64%) dogs had residual flow immediately after deployment. Of these, 10 had grade 1 flow, 10 had grade 2 flow, and 3 had grade 3 flow. After 24 hours, 16/35 (45%) of group 2 dogs had ongoing residual flow. Of these, 7 had grade 1 flow, 6 had grade 2 flow, 2 had grade 3 flow, and 1 had grade 4 flow. The dog with grade 4 residual flow suffered embolization of both coils to the lungs sometime during the recovery period. Its PDA was closed successfully by surgical ligation. In group 3, 11/23 (48%) dogs had some residual flow in the immediate period after deployment. Of these, 7 had grade 1, 3 grade 2, and 1 grade 3 flow. After 24 hours 9/23 (39%) had ongoing residual flow with 5 having grade 1, 3 grade 2, and 1 grade 3. Of the dogs in group 4, 5/15 (33%) had residual flow in the immediate post-deployment period. Of these 1 was grade 1, 2 were grade 2, and 2 were grade 4. The 2 dogs with grade 4 flow had the coils removed and other devices attempted (ACDO in one and AVP in the other). After 24 hours, 3/13 (23%) continued to have ongoing residual flow with 2 being grade 1 and 1 grade 2. Inadequate numbers of dogs returned for the 3-month recheck to assess residual flow at this time. Statistical analysis of residual flow data at 24 hours showed that group 1 had significantly less ongoing residual flow at this time than the other 3 groups (P = ). There was no significant difference in average degree of residual flow at 24 hours among groups 2, 3 and 4 (P =.15.58). Discussion Transvascular occlusion was successful in most dogs with an overall success rate of 92%. This result is similar to previous reports with success rates of 86% being reported for transarterial coil methods, 1, % reported for AVP methods, 7, % for the ACDO, 8,13 and % for transvenous occlusion methods. 3,10,12 Since the development of the ACDO and commercial availability in 2007, this device has gained popularity among veterinary cardiologists for its ease of use, low potential for device migration, and completeness of ductal occlusion. 6,8,13 The present study confirmed previous preliminary findings in support of the ACDO as the current device of choice for closure of PDA. There were no instances of ACDO migration in this study. In contrast, coil migration into the PA when deployed by the transarterial route occurred in 13% of cases in our study. In a previous retrospective study examining the use of thrombogenic coils in 125 dogs with PDA at the same facility there was a 22% incidence of coil migration. 4 Thus, pulmonary embolization appears to be a common complication that can occur with the use of thrombogenic coils. Previous studies have reported angiographic or echocardiographic persistence of ductal flow at conclusion of the procedure in 0 31% of dogs receiving an ACDO, 8, % of dogs receiving a thrombogenic coil 1,4,11,12,15 and in 25 33% after placement of an AVP or Amplatzer duct occluder. 3,7,9 Color flow Doppler done 24 hours postprocedure showed 29 dogs (27%) with residual flow, but the percentage of cases varied widely depending on the procedure. Only 3% (1 dog) had residual flow after placement of an ACDO, whereas 45% of dogs occluded transarterially with a thrombogenic coil, 39% with a vascular plug and 23% with a transvenously placed thrombogenic coil had
6 90 Singh et al residual flow. Although most dogs had trivial to mild persistent flow, 2 dogs from group 2 and 3 dogs from group 3 had moderate persistent flow and 1 dog from group 2 had severe persistent flow. These findings support previous reports that the ACDO is not only easy to use and highly successful, it also more frequently produces complete ductal occlusion than other devices. However, in all groups of dogs, there was a significant reduction in LA and left ventricular size (as shown by the significant reduction in LA:Ao and LVIDd:Ao), indicating considerable improvement in the severity of volume overload to the left heart. Unfortunately, not enough dogs returned for the recommended 3-month recheck to statistically assess the presence of persistent residual flow among groups at this time. Previous studies indicated delayed complete closure in a substantial number of cases that have undergone coil occlusion with reported cumulative closure rates ranging from 60 90%. 4,10,11 Complications that have been described for transvascular occlusion of PDA include pulmonary or systemic arterial embolism, hemodynamically relevant residual shunt, femoral artery tear, mechanical hemolysis as a result of incomplete ductal closure, left PA branch stenosis, and death. Such limitations have contributed to procedure abandonment in 11 27% of coil embolization attempts. 1,4 The complication rate for ACDO placement was lower (3%) than the other 3 methods (26 33%) in the present study. This finding was similar to that of other reports where complication rates of 5 7% with the ACDO were seen. 8,13 Although the AVP has a lower risk of embolization and may provide more complete ductal occlusion than thrombogenic coils, movement (rotation) appears to be a complicating factor 5 and occurred in 3 dogs in this study. Postplacement movement did not occur with any ACDO in the present study. The presence of congestive heart failure associated with a large PDA (large MDD resulting in high flow) has been shown to confer a 3-fold increase in risk of coil embolization. 4 Embolization did not occur with the use of ACDO in this study and procedural abandonment was rare even in the presence of a large PDA and congestive heart failure. One German Shepherd did have the ACDO procedure abandoned in favor of surgery due to a very large PDA (MDD, 7.5 mm; ampulla, 18 mm) being too large for the largest ACDO (14 mm) or AVP device (16 mm). Fluoroscopy time is a consideration when contemplating any interventional procedure because limiting excessive radiation exposure is an important workplace health consideration. In this study, the transvenous route was associated with significantly longer fluoroscopy time with a median of 13 minutes versus 5 8 minutes for transarterial methods. A recent study assessing the transvenous method of transvascular closure also suggested that fluoroscopy times were prolonged, but the actual numbers were not recorded. 10 One study in humans comparing a small number of babies who underwent transarterial closure to the transvenous approach at different institutions also found increased fluoroscopy times associated with the transvenous approach. 23 However, this finding did not reach statistical significance. Nevertheless, this may be a important workplace health consideration when considering the type of transvascular occlusion to be performed. The transarterial approach has been the most common method used for transvascular closure of PDA in veterinary medicine. In human medicine, the transvenous approach usually is preferred due to apparent lower complication rates. 24 However, this has not been proven to be the case in several previous studies comparing the different approaches in humans. 23,25 The transarterial approach avoids right heart catheterization that can be associated with difficult retrograde PDA access, kinking of the device sheath and cardiac arrhythmias. 3,10,15 In 1 previous study, the author encountered several difficulties during cardiac catheterization and deployment of the Amplatzer duct occluder transvenously. 3 In some dogs, it was difficult to pass the catheter or guide wire across the pulmonic orifice of the ductus arteriosus. This increased fluoroscopic and procedural time. In small dogs, the introducing sheath was reported to have a tendency to kink as it crossed the pulmonic valve. 3 In addition, if it is found during the procedure that more than 1 coil is required occlusion would be difficult to accomplish by the transvenous approach and the arterial approach then would be required. 12 Hemorrhage from the catheter site was a substantial problem in previous studies utilizing the percutaneous transvenous method, 3,12 and mild bruising and bleeding also were noted to occur in this study in some dogs that had the modified Seldinger venous approach despite prolonged digital occlusion. In 1 dog, this complication was severe, resulted in a 30% decrease in packed cell volume and required transfer to the intensive care unit for monitoring. The shape of the hind limb and close proximity to the inguinal region precludes application of an effective occlusive bandage. Thus, it previously has been suggested that one needs to anticipate the need to ligate the femoral vein if substantial bleeding occurs. 3 On the other hand, the transarterial approach resulted in femoral artery laceration in 4 dogs in this study. All dogs were small ( 3 kg). Although no dog suffering femoral artery laceration experienced clinically relevant hemorrhage, this is a potential life-threatening complication that needs to be considered when contemplating this approach in very small dogs. There was a significant difference in MDD and amp among the 4 groups with groups 1 and 3 being significantly larger than groups 2 and 4 but with groups 1 and 3 not significantly different from each other and groups 2 and 4 not significantly different from each other. However, when MDD was indexed to the aorta to take into account the various body sizes, there was no significant difference among groups. 19,26 This finding suggested that the larger MMD and amp in groups 1 and 3 was due to the larger body size of these dogs when compared to groups 2 and 4 and, when taking into account body size, there was no significant difference in the size of the PDA among groups.
7 Occlusion Devices in Patent Ductus Arteriosus 91 Both the ACDO and the AVP were used more successfully than coil methods in larger dogs and those with larger MDD, but the ACDO was associated with a lower complication rate and more complete closure than the AVP. However, the smallest dog in which the decision to place an ACDO was made and was successful was 2.6 kg whereas the PDA of dogs 2.5 kg were closed successfully with thrombogenic coils by both the transarterial and transvenous routes. The smallest dog successfully receiving a coil in the transarterial group was 1.0 kg and in the transvenous group was 1.2 kg. Despite recent studies suggesting that the transvenous approach may be more useful in very small dogs, there was no significant difference between body weights in dogs receiving transarterial or transvenous coils in this study. However we did find that very small dogs (<3 kg) were more likely to have complications such as a femoral arterial laceration or inability to place a catheter in the smaller diameter and less distensible femoral artery (5 dogs in this study) warranting consideration of the transvenous approach in very small dogs. If the currently developed low profile ACDO for dogs 3kg r becomes commercially available, this limiting factor may be circumvented. Several limitations need to be considered when interpreting the results of this study. Being a retrospective study, patients were not randomized to treatment groups introducing substantial bias in comparisons. In addition, results are limited by the accuracy of patient records. All procedures in this study were performed by a resident supervised by a faculty member. Device selection was not randomized but assigned by faculty clinician preference. It was not possible to perform clinician analysis to look for bias involving clinician preference because the faculty member involved usually was not recorded. Despite recommendations for 3- month recheck, a substantial number of dogs were not returned, resulting in inability to perform meaningful statistical analysis at this time. It is highly likely that a number of coil cases would have had more complete closure by this recheck. 4,10,11 Determining appropriate treatment is important in assuring optimal outcome. Successful PDA closure depends on device selection and typically is based on ductal morphology and dimensions as well as patient size. The results of this study support the ACDO as the device of choice for the majority of PDA occlusions although, currently, a small subset of dogs ( 2.5 kg) still may require the transvenous or transarterial coil approach for transvascular closure because the currently available ACDO device is too large for transarterial delivery in these dogs. Footnotes a Miller MW, Meurs KM, Boswood A. Echocardiographic assessment of PDA after occlusion. In: Proceedings of the 14th Annual Veterinary Forum, ACVIM, San Francisco, CA; 1994: (abstract) b Phillips ie33. Phillips Medical Systems, NA, Bothell, WA c Acuson 128 XP/10. HP sonos 5500, Hewlett Packard, Andover, MA d Phillips transesophageal ultrasound transducer. Phillips Medical Systems e Cook MPA non tapered end hole catheter, Cook Veterinary Products Inc, Bloomingdale, IN f GE OEC 9800, OEC Medical Systems Inc, Salt Lake City, UT g Hypaque, Aversham Health Inc, Princeton, NJ h Fast-Cath, St. Jude Medical, Minnetonka, MN, or Cook Mullins introducer set, Cook Veterinary Products Inc i Balloon Wedge Pressure Catheter, Arrow International Inc, Reading, PA k Fixed Core Wire Guide, Cook Medical Inc j Glidewire, Angled Flexible Tip, Terumo Medical Corporation, Somerset, NJ l Beacon, Straight Tip Royal Flush Plus Angiographic Catheter, Cook Medical Inc. m Cook embolization coils or MReye Flipper Detachable Embolization Coil, Cook Medical Inc. n Amplatzer Vascular Plug, AGA Medical Corp., Plymouth, MN o Amplatz Canine Duct Occluder, Infiniti Medical, LLC, West Hollywood, CA p Ancef, Cefazolin Sodium, Bristol-Myers Squibb Co, Princeton, NJ q StatXact 8.0, Cytel Software Corp., Cambridge, MA r Olson JLC, Tobias AH, Stauthammer CD, et al. Minimally invasive per-catheter patent ductus arteriosus occlusion in small dogs (</=3 kg): Preliminary results. In: ACVIM Forum Proceedings, Anaheim 2010 (abstract) Acknowledgment This study was not funded by a grant. References 1. Stokhof AA, Sreeram N, Wolvekamp WT. Transcatheter closure of patent ductus arteriosus using occluding spring coils. J Vet Intern Med 2000;14: Glaus TM, Berger F, Ammann FW, et al. Closure of large patent ductus arteriosus with a self-expanding duct occluder in two dogs. J Small Anim Pract 2002;43: Sisson D. Use of a self-expanding occluding stent for nonsurgical closure of patent ductus arteriosus in dogs. J Am Vet Med Assoc 2003;223: Campbell FE, Thomas WP, Miller SJ, et al. Immediate and late outcomes of transarterial coil occlusion of patent ductus arteriosus in dogs. J Vet Intern Med 2006;20: Hogan DF, Green HW, Sanders RA. Transcatheter closure of patent ductus arteriosus in a dog with a peripheral vascular occlusion device. J Vet Cardiol 2006;8: Nguyenba TP, Tobias AH. The Amplatz canine duct occluder: A novel device for patent ductus arteriosus occlusion. J Vet Cardiol 2007;9: Smith PJ, Martin MW. Transcatheter embolisation of patent ductus arteriosus using an Amplatzer vascular plug in six dogs. J Small Anim Pract 2007;48: Nguyenba TP, Tobias AH. Minimally invasive per-catheter patent ductus arteriosus occlusion in dogs using a prototype duct occluder. J Vet Intern Med 2008;22: Achen SE, Miller MW, Gordon SG, et al. Transarterial ductal occlusion with the Amplatzer vascular plug in 31 dogs. J Vet Intern Med 2008;22:
8 92 Singh et al 10. Henrich E, Hildebrandt N, Schneider C, et al. Transvenous coil embolization of patent ductus arteriosus in small (</=3.0 kg) dogs. J Vet Intern Med 2010;25: Hildebrandt N, Schneider C, Schweigl T, et al. Longterm follow-up after transvenous single coil embolization of patent ductus arteriosus in dogs. J Vet Intern Med 2010;24: Blossom JE, Bright JM, Griffiths LG. Transvenous occlusion of patent ductus arteriosus in 56 consecutive dogs. J Vet Cardiol 2010;12: Gordon SG, Saunders AB, Achen SE, et al. Transarterial ductal occlusion using the Amplatz Canine Duct Occluder in 40 dogs. J Vet Cardiol 2010;12: Schneider M, Hildebrandt N, Schweigl T, et al. Transthoracic echocardiographic measurement of patent ductus arteriosus in dogs. J Vet Intern Med 2007;21: Schneider M, Hildebrandt N, Schweigl T, et al. Transvenous embolization of small patent ductus arteriosus with single detachable coils in dogs. J Vet Intern Med 2001;15: Thomas WP, Gaber CE, Jacobs GJ, et al. Recommendations for standards in transthoracic two-dimensional echocardiography in the dog and cat. Echocardiography Committee of the Specialty of Cardiology, American College of Veterinary Internal Medicine. J Vet Intern Med 1993;7: Seldinger SI. Catheter replacement of the needle in percutaneous arteriography; a new technique. Acta Radiol 1953;39: Miller MW, Gordon SG, Saunders AB, et al. Angiographic classification of patent ductus arteriosus morphology in the dog. J Vet Cardiol 2006;8: Brown DJ, Rush JE, MacGregor J, et al. M-mode echocardiographic ratio indices in normal dogs, cats, and horses: A novel quantitative method. J Vet Intern Med 2003;17: Cunningham SM, Rush JE, Freeman LM, et al. Echocardiographic ratio indices in overtly healthy Boxer dogs screened for heart disease. J Vet Intern Med 2008;22: Swensson RE, Valdes-Cruz LM, Sahn DJ, et al. Real-time Doppler color flow mapping for detection of patent ductus arteriosus. J Am Coll Cardiol 1986;8: Holm S. A simple sequentially rejective multiple test procedure. Scand J Stat 1979;6: Hijazi ZM, Lloyd TR, Beekman RH, 3rd, et al. Transcatheter closure with single or multiple Gianturco coils of patent ductus arteriosus in infants weighing < or = 8 kg: Retrograde versus antegrade approach. Am Heart J 1996;132: Kannan BR, Anil SR, Padhi SS, et al. Transcatheter management of patent ductus arteriosus in sick ventilated small infants. Indian Heart J 2004;56: Lee ML, Chaou WT, Wang JK. Transarterial occlusion of patent ductus arteriosus with Gianturco coils in pediatric patients: A preliminary result in central Taiwan. Int J Cardiol 1999;69: Cornell CC, Kittleson MD, Della Torre P, et al. Allometric scaling of M-mode cardiac measurements in normal adult dogs. J Vet Intern Med 2004;18:
1. Definition/Overview
VCAwestlaspecialty.com VCA West Los Angeles recently performed closure of a patent ductus arteriosus via percutaneous transarterial embolization with an Amplatz canine ductal occluder on a female, four-month-old
More informationTranscatheter closure of the patent ductus arteriosus using the new Amplatzer duct occluder: Initial clinical applications in children
Transcatheter closure of the patent ductus arteriosus using the new Amplatzer duct occluder: Initial clinical applications in children Basil Thanopoulos, MD, PhD, a Nikolaos Eleftherakis, MD, a Konstantinos
More informationPatent ductus arteriosus (PDA) is the most common congenital
J Vet Intern Med 2004;18:325 329 Transarterial Coil Embolization of Patent Ductus Arteriosus in Small Dogs with 0.025-Inch Vascular Occlusion Coils: 10 Cases Daniel F. Hogan, Henry W. Green III, Sonya
More informationPaediatrica Indonesiana
Paediatrica Indonesiana VOLUME 53 July NUMBER 4 Original Article Transcatheter vs. surgical closure of patent ductus arteriosus: outcomes and cost analysis Mulyadi M Djer, Mochammading, Mardjanis Said
More informationPatent Ductus Arteriosus (PDA)
Patent Ductus Arteriosus (PDA) How does the heart work? The heart is the organ responsible for pumping blood to and from all tissues of the body. The heart is divided into right and left sides. The job
More informationEvaluation of Gianturco Coils for Closure of Large (> 3.5 mm) Patent Ductus Arteriosus
1856 JACC Vol. 30, No. 7 Evaluation of Gianturco Coils for Closure of Large (> 3.5 mm) Patent Ductus Arteriosus CARL Y. OWADA, MD, DAVID F. TEITEL, MD, PHILLIP MOORE, MD San Francisco, California Objectives.
More informationINTEGRATING ECHOCARDIOGRAPHY WITH CATHETER INTERVENTIONS FOR CONGENITAL HEART DISEASE. Krishna Kumar SevenHills Hospital, Mumbai, India
INTEGRATING ECHOCARDIOGRAPHY WITH CATHETER INTERVENTIONS FOR CONGENITAL HEART DISEASE Krishna Kumar SevenHills Hospital, Mumbai, India Why talk about it? What is the big deal? Are we not stating the obvious?
More informationType Size AP/PS RP/RS Qp/Qs. Ia Resistive <0.3 < Ib Resistive <0.3 < IIa Resistive <0.5 >2
Transcatheter closure of VSD using Duct Occluder device Nguyen Lan Hieu, MD, PhD Hanoi Medical University Vietnam Heart Institute Anatomy of VSD 1. Perimembranous VSD: Aneurysm septal membranous(tv or
More informationPatient guide: pfm Nit-Occlud PDA coil occlusion system. Catheter occlusion of. Patent Ductus Arteriosus. with the
Patient guide: Catheter occlusion of Patent Ductus Arteriosus with the pfm Nit-Occlud PDA coil occlusion system pfm Produkte für die Medizin - AG Wankelstr. 60 D - 50996 Cologne Phone: +49 (0) 2236 96
More informationLeft-to-right shunting patent ductus arteriosus
J Vet Intern Med 2013 Long-Term Outcome in Dogs with Patent Ductus Arteriosus: 520 Cases (1994 2009) A.B. Saunders, S.G. Gordon, M.M. Boggess, and M.W. Miller Background: Published information regarding
More informationFor Personal Use. Copyright HMP 2013
Case Report J INVASIVE CARDIOL 2013;25(2):E42-E44 Percutaneous Closure of Iatrogenic Ventricular Septal Defect Following Surgical Aortic Valve Replacement Using Two Different Approaches Takashi Matsumoto,
More informationOriginally Posted: November 15, 2014 BRUIT IN THE GROIN
Originally Posted: November 15, 2014 BRUIT IN THE GROIN Resident(s): Donald ML Tse, MD Attending(s): KT Tan, MD Program/Dept(s): University Health Network/Mount Sinai Hospital, Toronto, ON, Canada CHIEF
More informationUncommon Doppler Echocardiographic Findings of Severe Pulmonic Insufficiency
Uncommon Doppler Echocardiographic Findings of Severe Pulmonic Insufficiency Rahul R. Jhaveri, MD, Muhamed Saric, MD, PhD, FASE, and Itzhak Kronzon, MD, FASE, New York, New York Background: Two-dimensional
More informationPDA: Closure Using Coils and Devices: Indications, Technique & Outcome
PDA: Closure Using Coils and Devices: Indications, Technique & Outcome Daniel Levi, MD, FSCAI Associate Professor of Pediatrics UCLA Biomedical Engineering Mattel Children s Hospital at UCLA, Division
More informationOriginal Policy Date
MP 7.01.46 Transcatheter Closure of Patent Ductus Arteriosus Medical Policy Section Surgery Issue 12:2013 Original Policy Date 12:2013 Last Review Status/Date Reviewed with literature search/12:2013 Return
More informationTRANSCATHETER CLOSURE OF PERSISTENT DUCTUS ARTERIOSUS USING THE AMPLATZER DUCT OCCLUDER IN INFANTS
TRANSCATHETER CLOSURE OF PERSISTENT DUCTUS ARTERIOSUS USING THE AMPLATZER DUCT OCCLUDER IN INFANTS Al-Hakim FA 1, Harahsheh AS 2 & Hijazi IS 3 ABSTRACT Background: Despite technical improvements over the
More informationCase submission for CSI Asia-Pacific Case 2
Case submission for CSI Asia-Pacific 2018- Case 2 Title Page Case category: Coarctation and ducts, valves Title: Simultaneous balloon aortic valvuloplasty with transcatheter closure of large hypertensive
More informationReopening After Successful Coil Occlusion for Patent Ductus Arteriosus
444 JACC Vol. 31, No. 2 PEDIATRIC CARDIOLOGY Reopening After Successful Coil Occlusion for Patent Ductus Arteriosus CURT J. DANIELS, MD, STEVEN C. CASSIDY, MD, FACC, DOUGLAS W. TESKE, MD, FACC, JOHN J.
More informationHybrid Muscular VSD Closure in Small Weight Children
Hybrid Muscular VSD Closure in Small Weight Children Shakeel A Qureshi, on behalf of: John P. Cheatham, MD George H. Dunlap Endowed Chair in Interventional Cardiology Director Cardiac Catheterization &
More informationPDA in Pets There s Nothing Affectionate About Patent Ductus Arteriosus
PDA in Pets There s Nothing Affectionate About Patent Ductus Arteriosus People are frequently born with heart defects. They often require intricate and specialized surgeries to help correct them. So too
More informationTCTAP C-217 Bilateral Pulmonary Arterio Venus Fistula Managed by Vascular Plugs and Coil Manotosh Panja 1 1 BelleVue Clinic, India
S450 JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, VOL. 65, NO. 17, SUPPL S, 2015 Case Summary. Aneurysm at LVOT is a rare complication after Bentall operation and the aneurysm may increase in size. In
More information5.8 Congenital Heart Disease
5.8 Congenital Heart Disease Congenital heart diseases (CHD) refer to structural or functional heart diseases, which are present at birth. Some of these lesions may be discovered later. prevalence of Chd
More informationTreatment of Inter-MitraClip Regurgitation Due to Posterior Leaflet Cleft by Use of The Amplatzer Vascular Plug II device
Volume 1, Issue 3 Case Report ISSN: 2572-9292 Treatment of Inter-MitraClip Regurgitation Due to Posterior Leaflet Cleft by Use of The Amplatzer Vascular Plug II device Neha M. Mantri, Gagan D. Singh, Thomas
More informationPercutaneous VSD closure
Percutaneous VSD closure Gianfranco Butera San Donato Milanese - Italy Patients selection Pts having hemodynamically significant VSD Left ventricular enlargement (left ventricular overload),defined as
More informationCutting Edge Cardiology
Peer Reviewed Cutting Edge Cardiology Five State-of-the-Art Developments Ashley B. Saunders, DVM, Diplomate ACVIM (Cardiology) Texas A&M University Veterinary practitioners strive to provide patients and
More informationAtrial Septal Defect Closure. Stephen Brecker Director, Cardiac Catheterisation Labs
Stephen Brecker Director, Cardiac Catheterisation Labs ADVANCED ANGIOPLASTY Incorporating The Left Main 5 Plus Course Conflicts of Interest The following companies have supported educational courses held
More informationIMAGES. in PAEDIATRIC CARDIOLOGY. Abstract
IMAGES in PAEDIATRIC CARDIOLOGY Images Paediatr Cardiol. 2008 Apr-Jun; 10(2): 11 17. PMCID: PMC3232589 Transcatheter closure of symptomatic aortopulmonary window in an infant F Pillekamp, 1 T Hannes, 1
More informationNit-Occlud. Coil System for PDA Closure IMPLANTATION POCKET GUIDE. Rx only CV / B. Braun Interventional Systems Inc.
Refer to the Nit-Occlud PDA Instructions for Use for relevant warnings, precautions, complications and contraindications. This device has been designed for single use only. Nit-Occlud Coil System for PDA
More informationAdult Echocardiography Examination Content Outline
Adult Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 2 3 4 5 Anatomy and Physiology Pathology Clinical Care and Safety Measurement Techniques, Maneuvers,
More informationCardiolog procedure availabl a Davie Veterinar Specialist
Cardiolog Procedure An overview of the procedures and techniques available to our Cardiology patients. Cardiolog procedure availabl a Davie Veterinar Specialist - Patent Ductus Arteriosus Occlusion - Pulmonic
More informationCorrection of a Canine Left-to-Right Shunting
Correction of a Canine Left-to-Right Shunting By Connie K. Varnhagen, PhD, RAHT Patent Ductus Arteriosus Scrappy, a 5-year-old, 27-kg, neutered Labrador retriever, presented to the Calgary Animal Referral
More informationCOMPLEMENTARY TECHNIQUES OF PERCUTANEOUS CLOSURE OF DUCTUS ARTERIOSUS USING DETACHABLE COOK COILS AND AMPLATZER DEVICES
214 East African Medical Journal July 2013 East African Medical Journal Vol. 90 No. 7 July 2013 COMPLEMENTARY TECHNIQUES OF PERCUTANEOUS CLOSURE OF DUCTUS ARTERIOSUS USING DETACHABLE COOK COILS AND AMPLATZER
More informationCase Report Hemostasis of Left Atrial Appendage Bleed With Lariat Device
273 Case Report Hemostasis of Left Atrial Appendage Bleed With Lariat Device Amena Hussain MD, Muhamed Saric MD, Scott Bernstein MD, Douglas Holmes MD, Larry Chinitz MD NYU Langone Medical Center, United
More informationINTRAHEPATIC PSS: AN INTERVENTIONALIST S PERSPECTIVE
INTRAHEPATIC PSS: AN INTERVENTIONALIST S PERSPECTIVE Matthew W. Beal, DVM, DACVECC College of Veterinary Medicine, Michigan State University, East Lansing, MI Key Points: Percutaneous transjugular coil
More informationIMAGES. in PAEDIATRIC CARDIOLOGY
IMAGES in PAEDIATRIC CARDIOLOGY Images Paediatr Cardiol. 2005 Jan-Mar; 7(1): 12 17. PMCID: PMC3232568 Stent implantation for coarctation facilitated by the anterograde trans-septal approach N Sreeram and
More informationDepartment of Pediatrics and Child Health, Kurume University School of Medicine, Kurume, 830 Japan. Received for publication October 26, 1992
THE KURUME MEDICAL JOURNAL Vol.39, p.291-296, 1992 Jon-Invasive Evaluation of Pulmonary Arterial and Right Ventricular Pressures with Contrast Enhanced Doppler Signals of Tricuspid Regurgitation Flow Using
More informationAppendix II: ECHOCARDIOGRAPHY ANALYSIS
Appendix II: ECHOCARDIOGRAPHY ANALYSIS Two-Dimensional (2D) imaging was performed using the Vivid 7 Advantage cardiovascular ultrasound system (GE Medical Systems, Milwaukee) with a frame rate of 400 frames
More informationEchocardiographic Guidance During Placement of the Buttoned Double-Disk Device for Atrial Septa1 Defect Closure
Echocardiographic Guidance During Placement of the Buttoned Double-Disk Device for Atrial Septa1 Defect Closure L. LUANN MINICH, M.D., and A. REBECCA SNIDER, M.D. Department of Pediatrics, C.S. Mott Children
More informationComparative, multidimensional imaging of patent ductus arteriosus and a proposed update to the morphology classification system for dogs
Received: 7 July 2017 Revised: 8 November 2017 Accepted: 18 January 2018 DOI: 10.1111/jvim.15068 STANDARD ARTICLE Journal of Veterinary Internal Medicine Comparative, multidimensional imaging of patent
More informationTranscatheter closure of interatrial
372 Br HeartJf 1994;72:372-377 PRACTICE REVIEWED Department of Paediatric Cardiology, Royal Brompton Hospital, London A N Redington M L Rigby Correspondence to: Dr A N Redington, Department of Paediatric
More informationPART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING
PART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING STANDARD - Primary Instrumentation 1.1 Cardiac Ultrasound Systems SECTION 1 Instrumentation Ultrasound instruments
More informationInitial Human Experience. Objectives: To evaluate the feasibility, safety, and efficacy of the Occlutech VR
Catheterization and Cardiovascular Interventions 00:00 00 (2015) Original Studies The New Occlutech V R PDA Occluder: Initial Human Experience M. A. Elbashier Abdelbasit, 1 * MBBS, MD, Mazeni Alwi, 1 MBBS,
More informationReliability of Echocardiography Measurement of Patent Ductus Arteriosus Minimum Diameter: A Meta-analysis
International Journal of Cardiovascular and Cerebrovascular Disease 4(): 15-19, 016 DOI: 10.13189/ijccd.016.04001 http://www.hrpub.org Reliability of Echocardiography Measurement of Patent Ductus Arteriosus
More informationPDA Device Closure with and without Arterial Access
Research Article imedpub Journals www.imedpub.com Interventional Cardiology Journal DOI: 10.21767/2471-8157.100061 Abstract PDA Device Closure with and without Arterial Access Background: PDA device closure
More informationCONGENITAL HEART DEFECTS IN ADULTS
CONGENITAL HEART DEFECTS IN ADULTS THE ROLE OF CATHETER INTERVENTIONS Mario Carminati CONGENITAL HEART DEFECTS IN ADULTS CHD in natural history CHD with post-surgical sequelae PULMONARY VALVE STENOSIS
More informationRadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.
Interventional Radiology Coding Case Studies Prepared by Stacie L. Buck, RHIA, CCS-P, RCC, CIRCC, AAPC Fellow President & Senior Consultant INDICATION: Abdominal aortic aneurysm. INTERVENTIONAL RADIOLOGIST:
More informationThis is a provisional PDF comprising this cover note and the manuscript as it was upon acceptance for publication.
This is a provisional PDF comprising this cover note and the manuscript as it was upon acceptance for publication. A typeset PDF article will be published soon. Surgical Removal of a Knotted Intravascular
More informationDrs. Rottman, Salloum, Campbell, Muldowney, Hong, Bagai, Kronenberg
Rotation: or: Faculty: Coronary Care Unit (CVICU) Dr. Jeff Rottman Drs. Rottman, Salloum, Campbell, Muldowney, Hong, Bagai, Kronenberg Duty Hours: Mon Fri, 7 AM to 7 PM, weekend call shared with consult
More informationPediatric Echocardiography Examination Content Outline
Pediatric Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 Anatomy and Physiology Normal Anatomy and Physiology 10% 2 Abnormal Pathology and Pathophysiology
More informationTranscatheter device occlusion of patent ductus arteriosus
CONGENITAL HEART DISEASE The New Occlutech Duct Occluder: Immediate Results, Procedural Challenges, and Short-Term Follow-Up Vikram Kudumula, MRCPCH; Demetris Taliotis, MRCPCH; Christopher Duke, FRCP ABSTRACT:
More informationThe earliest application of intracardiac imaging
Intracardiac Echocardiography for Structural Heart Defects A review of ICE innovations, devices, and techniques. BY ZAHID AMIN, MD; QI-LING CAO, MD; AND ZIYAD M. HIJAZI, MD The earliest application of
More informationIndex. interventional.theclinics.com. Note: Page numbers of article titles are in boldface type.
Index Note: Page numbers of article titles are in boldface type. A Alagille syndrome, pulmonary artery stenosis in, 143 145, 148 149 Amplatz devices for atrial septal defect closure, 42 46 for coronary
More information1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER
Page 1 of 5 ASPIRATION CATHETER Carefully read all instructions prior to use, observe all warnings and precautions noted throughout these instructions. Failure to do so may result in complications. STERILE.
More informationCY2017 Hospital Outpatient: Vascular Procedure APCs and Complexity Adjustments
CY2017 Hospital Outpatient: Vascular Procedure APCs and Complexity Adjustments Comprehensive Ambulatory Payment Classifications (c-apcs) In CY2015 and in an effort to help pay providers for quality, not
More informationPathophysiology: Left To Right Shunts
Pathophysiology: Left To Right Shunts Daphne T. Hsu, MD dh17@columbia.edu Learning Objectives Learn the relationships between pressure, blood flow, and resistance Review the transition from fetal to mature
More informationIndications of Coronary Angiography Dr. Shaheer K. George, M.D Faculty of Medicine, Mansoura University 2014
Indications of Coronary Angiography Dr. Shaheer K. George, M.D Faculty of Medicine, Mansoura University 2014 Indications for cardiac catheterization Before a decision to perform an invasive procedure such
More informationNational Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION. Original Date: October 2015 Page 1 of 5
National Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION CPT Codes: 93451, 93452, 93453, 93454, 93455, 93456, 93457, 93458, 93459, 93460, 93461 LCD ID Number:
More informationBILLING BULLETIN. Re: Interventional Cardiology. Bulletin #: 1. Date Issued: November 10, Background
BILLING BULLETIN Re: Interventional Cardiology Bulletin #: 1 Date Issued: November 10, 2016 Background This Billing Bulletin provides billing guidance when submitting claims to Manitoba Health, Seniors
More informationRadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.
Interventional Radiology Coding Case Studies Prepared by Stacie L. Buck, RHIA, CCS-P, RCC, CIRCC, AAPC Fellow President & Senior Consultant Week of June 4, 2018 Thrombolysis, Thrombectomy & Angioplasty
More informationCATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018
WAVES CATCH A WAVE.. W I S C O N S I N P A R A M E D I C S E M I N A R A P R I L 2 0 1 8 K E R I W Y D N E R K R A U S E R N, C C R N, E M T - P Have you considered that if you don't make waves, nobody
More informationTranscatheter closure of right coronary artery fistula to the right ventricle
Case Report Transcatheter closure of right coronary artery fistula to the right ventricle Abstract Coronary artery fistula (CAF) is an uncommon anomaly usually congenital but can be acquired. Although,
More informationRepair or Replacement
Surgical intervention post MitraClip Device: Repair or Replacement Saudi Heart Association, February 21-24 Rüdiger Lange, MD, PhD Nicolo Piazza, MD, FRCPC, FESC German Heart Center, Munich, Germany Division
More informationΔιαδερμική θεραπεία των παραβαλβιδικών διαφυγών Σ.Ράμμος
ΔΜΙΝΑΡΙΟ ΔΠΙΣΗΜΟΝΙΚΩΝ ΔΝΩΔΩΝ ΔΛΛΗΝΙΚΟΤ ΚΟΛΛΔΓΙΟΤ ΚΑΡ ΙΟΛΟΓΙΑ 2-4.10.2015 Δπιζηημονική Ένωζη Δπεμβαηικής Καρδιολογίας Βαλβιδοπάθειες, Γιαδερμικές παρεμβάζεις Διαδερμική θεραπεία των παραβαλβιδικών διαφυγών
More informationAMPLATZER Septal Occluder
AMPLATZER Septal Occluder A Patient s Guide to the Non-Surgical Closure of the Atrial Septal Defect Using the AMPLATZER Septal Occluder System leadership through innovation TM AGA Medical Corporation 5050
More informationHEART MURMURS: DECIPHERING THEIR CAUSE AND SIGNIFICANCE
Vet Times The website for the veterinary profession https://www.vettimes.co.uk HEART MURMURS: DECIPHERING THEIR CAUSE AND SIGNIFICANCE Author : Pedro Oliveira Categories : Vets Date : May 27, 2013 PEDRO
More informationA case report: Percutaneous transcatheter treatment of Lutembacher syndrome
A case report: Percutaneous transcatheter treatment of Lutembacher syndrome Dr. Sainath Hegde* Dr. Tukaram Aute** *Registrar, Dept of Cardiology, M.G.M Hospital, Aurangabad. **Asst. Prof, Dept of Cardiology,
More informationNit-Occlud. Coil System for PDA Closure PATIENT EDUCATIONAL BROCHURE
Nit-Occlud PDA Coil System for PDA Closure PATIENT EDUCATIONAL BROCHURE A Patient s Educational Brochure to Non-Surgical Closure of the Patent Ductus Arteriosus (PDA) Using the Nit-Occlud PDA Inside This
More informationDiversion of the inferior vena cava following repair of atrial septal defect causing hypoxemia
Marshall University Marshall Digital Scholar Internal Medicine Faculty Research Spring 5-2004 Diversion of the inferior vena cava following repair of atrial septal defect causing hypoxemia Ellen A. Thompson
More informationTranscatheter Closure of Patent Ductus Arteriosus Using ADO Device: Retrospective Study of 149 Patients
Original Article Transcatheter Closure of Patent Ductus Arteriosus Using ADO Device: Retrospective Study of 149 Patients ABSTRACT Sadiq M. Al-Hamash, Hussein Abdul Wahab 1, Zayir H. Khalid 1, Isam V. Nasser
More informationCPT Code Details
CPT Code 93572 Details Code Descriptor Intravascular Doppler velocity and/or pressure derived coronary flow reserve measurement (coronary vessel or graft) during coronary angiography including pharmacologically
More informationCase Report International Journal of Basic and Clinical Studies (IJBCS) 2013;1(1): Elbey MA et al.
Treatment of Interrupted Aorta in Adult Patients; a Challenge Both in Surgery and Transcatheter Intervention Mehmet Ali Elbey MD 1, Ahmet Caliskan MD 2, Ferhat Isık MD 1, Faruk Ertas MD 1, Mehmet Serdar
More informationImplantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect
The Ochsner Journal 10:27 31, 2010 f Academic Division of Ochsner Clinic Foundation Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect Anas Bitar, MD, Maria Malaya
More informationPercutaneous atrial septal defect closure with the Occlutech Figulla Flex ASD Occluder.
Percutaneous atrial septal defect closure with the Occlutech Figulla Flex ASD Occluder. First case with a novel delivery system. Werner Budts, Md, PhD, FESC Congenital and Structural Cardiology University
More informationStandardising echocardiography and images. Version 2, 13/04/15
Standardising echocardiography and images 1. Review of ECHO eligibility criteria - trial entry - rescue treatment 2. Assessments - personnel - timing 3. Technical aspects of ECHO examination 1. Trial entry
More informationHOW IMPORTANT ARE THESE ECHO MEASUREMENTS ANYWAY?
HOW IMPORTANT ARE THESE ECHO MEASUREMENTS ANYWAY? John D. Carroll, MD Professor, Director of Interventional Cardiology and Co-Medical Director of the Cardiac and Vascular Center, University of Colorado
More informationHow to manage device embolization?
How to manage device embolization? Swiss Fellow Course Interventional Cardiology Amir-Ali Fassa Hôpital de La Tour Meyrin, Geneva Coronary interventions: Definitions Stent loss: stent dislodgment from
More informationPathophysiology: Left To Right Shunts
Pathophysiology: Left To Right Shunts Daphne T. Hsu, MD dh17@columbia.edu Learning Objectives Learn the relationships between pressure, blood flow, and resistance Review the transition from fetal to mature
More informationFabien Praz, Andreas Wahl, Sophie Beney, Stephan Windecker, Heinrich P. Mattle*, Bernhard Meier
Procedural Outcome after Percutaneous Closure of Patent Foramen Ovale using the Amplatzer PFO Occluder Without Intra-Procedural Echocardiography in 1,000 Patients Fabien Praz, Andreas Wahl, Sophie Beney,
More informationStraub Endovascular System &
Straub Endovascular System & S t r a u b E n d o v a s c u l a r To o l s Straub Endovascular System Effective debulking in occluded arteries and veins Effective debulking in many indications Rotarex
More informationCongenital tricuspid valve stenosis (TVS) is rare in
Standard Article J Vet Intern Med 2017 Balloon Valvuloplasty of Tricuspid Stenosis: A Retrospective Study of 5 Labrador Retriever Dogs G.A. Lake-Bakaar, L.G. Griffiths, and M.D. Kittleson Background: There
More informationAn Overview of Post-EVAR Endoleaks: Imaging Findings and Management. Ravi Shergill BSc Sean A. Kennedy MD Mark O. Baerlocher MD FRCPC
An Overview of Post-EVAR Endoleaks: Imaging Findings and Management Ravi Shergill BSc Sean A. Kennedy MD Mark O. Baerlocher MD FRCPC Disclosure Slide Mark O. Baerlocher: Current: Consultant for Boston
More informationPulmonary hypertension
Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich www.zora.uzh.ch Year: 2012 Pulmonary hypertension Glaus, T M Posted at the Zurich Open Repository
More informationPulmonic Stenosis. How does the heart work?
Pulmonic Stenosis How does the heart work? The heart is the organ responsible for pumping blood to and from all tissues of the body. The heart is divided into right and left sides. The job of the right
More informationIMAGES. in PAEDIATRIC CARDIOLOGY
IMAGES in PAEDIATRIC CARDIOLOGY Images Paediatr Cardiol. 2006 Jul-Sep; 8(3): 1 6. PMCID: PMC3232564 A large, single pulmonary arteriovenous fistula presenting hours after birth AH McBrien, 1 AJ Sands,
More informationTranscatheter Atrial Septa1 Defect Closure: Preliminary Experience with the Rashkind Occluder Device
Transcatheter Atrial Septa1 Defect Closure: Preliminary Experience with the Rashkind Occluder Device ROBERT H. BEEKMAN, M.D., ALBERT P. ROCCHINI, M.D., A. REBECCA SNIDER, M.D., and AMNON ROSENTHAL, M.D.
More informationPercutaneous Transapical Access for Thoracic Endovascular Repair
Percutaneous Transapical Access for Thoracic Endovascular Repair Atman P. Shah MD FACC FSCAI Co-Director, Hans Hecht Cardiac Catheterization Laboratory Clinical Director, Section of Cardiology Associate
More informationResponsibilities. Teaching Methods
Cognitive and Procedural Skills for Cardiothoracic Surgery Residents in Diagnostic and Therapeutic Cardiovascular Disease: Echocardiography, and Diagnostic and Therapeutic Catheterization The length of
More informationAnatomy & Physiology
1 Anatomy & Physiology Heart is divided into four chambers, two atrias & two ventricles. Atrioventricular valves (tricuspid & mitral) separate the atria from ventricles. they open & close to control flow
More informationAbsent Pulmonary Valve Syndrome
Absent Pulmonary Valve Syndrome Fact sheet on Absent Pulmonary Valve Syndrome In this condition, which has some similarities to Fallot's Tetralogy, there is a VSD with narrowing at the pulmonary valve.
More informationPercutaneous Coronary Intervention and Pulmonary Balloon Valvuloplasty in a Patient With Severe Valvular Pulmonary Stenosis
Case Report Percutaneous Coronary Intervention and Pulmonary Balloon Valvuloplasty in a 56-Year-Old Woman With Severe Valvular Pulmonary Stenosis: A Case Report Ata Firouzi 1, MD; Omid Shafe* 1, MD; Farzad
More informationJian Fang 1, Shaobo Xie 2, Lunchao Ma 2, Chao Yang 2. Original Article
Original Article Anatomic and surgical factors affecting the switch from minimally invasive transthoracic occlusion to open surgery during ventricular septal defect repair Jian Fang 1, Shaobo Xie 2, Lunchao
More information1. CARDIOLOGY. These listings cannot be correctly interpreted without reference to the Preamble. Anes. $ Level
1. CARDIOLOGY These listings cannot be correctly interpreted without reference to the Preamble. Anes. Referred Cases 33010 Consultation: To consist of examination, review of history, laboratory, X-ray
More informationDoppler-echocardiographic findings in a patient with persisting right ventricular sinusoids
Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich www.zora.uzh.ch Year: 1990 Doppler-echocardiographic findings in a patient with persisting right
More informationCY2015 Hospital Outpatient: Endovascular Procedure APCs and Complexity Adjustments
CY2015 Hospital Outpatient: Endovascular Procedure APCs Complexity Adjustments Comprehensive Ambulatory Payment Classifications (c-apcs) CMS finalized the implementation of 25 Comprehensive APC to further
More informationDebate in Management of native COA; Balloon Versus Surgery
Debate in Management of native COA; Balloon Versus Surgery Dr. Amira Esmat, El Tantawy, MD Professor of Pediatrics Consultant Pediatric Cardiac Interventionist Faculty of Medicine Cairo University 23/2/2017
More informationAtrioventricular Valve Dysplasia
Atrioventricular Valve Dysplasia How does the heart work? The heart is the organ responsible for pumping blood to and from all tissues of the body. The heart is divided into right and left sides. The job
More informationColor Doppler Echocardiographic Assessment of Valvular Regurgitation in Normal Infants
ORIGINAL ARTICLE Color Doppler Echocardiographic Assessment of Valvular Regurgitation in Normal Infants Shu-Ting Lee, Meng-Hsun Lin* Background/Purpose: Despite valvular regurgitation being a common finding
More informationBifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully
Physician Training Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully supported by self-expanding z-stents H&L-B
More informationIncidental Discovery of a Patent Ductus Arteriosus in Adults
BRIEF REPORT Incidental Discovery of a Patent Ductus Arteriosus in Adults Harvey D. Cassidy, MD, Lynsey A. Cassidy, MD, and Joseph L. Blackshear, MD Patent ductus arteriosus (PDA) is an uncommon clinical
More informationDoppler Basic & Hemodynamic Calculations
Doppler Basic & Hemodynamic Calculations August 19, 2017 Smonporn Boonyaratavej MD Division of Cardiology, Department of Medicine Chulalongkorn University Cardiac Center, King Chulalongkorn Memorial Hospital
More information