Arrhythmia/Electrophysiology

Size: px
Start display at page:

Download "Arrhythmia/Electrophysiology"

Transcription

1 Arrhythmia/Electrophysiology Comparison of Electrode Cooling Between Internal and Open Irrigation in Radiofrequency Ablation Lesion Depth and Incidence of Thrombus and Steam Pop Katsuaki Yokoyama, MD, PhD; Hiroshi Nakagawa, MD, PhD; Fred H.M. Wittkampf, PhD; Jan V. Pitha, MD, PhD; Ralph Lazzara, MD; Warren M. Jackman, MD Background Electrode cooling by circulating fluid within the electrode (closed loop) or open irrigation facilitates radiofrequency (RF) ablation. This study compared lesion parameters between closed loop and open irrigation with the use of a canine model. Methods and Results In 8 anesthetized dogs, the skin over the thigh muscle was incised and raised, forming a cradle superfused with heparinized blood (activated clotting time 350 seconds) at 37 C. A 7F 4-mm closed loop electrode (irrigation 36 ml/min) and 7.5F 3.5-mm open irrigation electrode (irrigation 17 ml/min) were positioned perpendicular to the thigh muscle at 10 g contact weight. RF was applied (n 121) at 20 or 30 W for 60 seconds in low (0.1 m/s) or high (0.5 m/s) pulsatile blood flow. Temperatures were measured in the electrode, electrode-tissue interface, and within the tissue at 3- and 7-mm depths. After each RF, the cradle was emptied to examine the electrode and interface for thrombus. There was no difference between closed loop and open irrigation in impedance, lesion depth, or tissue temperature at 20 or 30 W. Interface temperature and electrode temperature were greater in the closed loop application. Thrombus occurred in 32 of 63 closed loop versus 0 of 58 open irrigation RF applications (P 0.05) with interface temperature 80 C in all 32 (electrode temperature 40 C in 1, 40 C to 50 C in 26, and 50 C in 5). With closed loop, interface temperature and thrombus incidence were greater at 30 W and low blood flow. With open irrigation, interface temperature remained low ( 71 C) with no difference between 20 and 30 W or between low and high blood flow. Steam pop occurred at 20 W in 4 of 35 closed loop and 0 of 30 open irrigation and at 30 W in 15 of 28 closed loop and 4 of 28 open irrigation applications (P 0.05). Conclusions Lower interface temperature, thrombus, and steam pop, especially in low blood flow, indicate greater interface cooling with open irrigation. (Circulation. 2006;113:11-19.) Key Words: catheter ablation radiofrequency tachyarrhythmias thrombus Radiofrequency (RF) catheter ablation is most often performed in the temperature control mode. 1 7 Under temperature control, RF power is regulated to maintain a constant electrode temperature (commonly 55 C or 60 C), which is generally thought to limit the temperature at the electrodetissue interface to a level that will prevent an impedance rise, the formation of soft thrombus, and a steam pop. The ablation electrode temperature is dependent on the opposing effects of heating from the tissue and cooling by the blood flowing around the electrode At any given electrode temperature, the RF power delivered to the tissue is significantly reduced in areas of low blood flow. 10 The reduced cooling associated with low blood flow causes the electrode to reach the target temperature at lower power levels. Because lesion size is primarily dependent on the RF power delivered to the tissue, 12,14 lesion size will vary with the magnitude of local blood flow. Areas where lesion size is adversely affected because of low local blood flow (poor electrode cooling) include a deep pouch in the subeustachian isthmus (ablation of atrial flutter); dilated and poorly contracting atria (ablation of chronic atrial fibrillation or macroreentrant atrial tachycardia in patients after surgically repaired congenital heart disease or mitral valve disease); and dilated and poorly contracting ventricles (ablation of ventricular tachycardia associated with a postinfarction left ventricular aneurysm). 6,15 In these locations, increasing electrode temperature to 65 C or 70 C only minimally increases RF power and increases the risk of thrombus formation and impedance rise. 6,10,16 18 Clinical Perspective p 19 Two approaches have been developed to increase electrode cooling to allow RF power to be maintained in a desirable Received December 30, 2004; de novo received February 22, 2005; revision received October 10, 2005; accepted October 13, From the Cardiac Arrhythmia Research Institute, Departments of Medicine (K.Y., H.N., R.L., W.J.M.) and Pathology (J.V.P.), University of Oklahoma Health Sciences Center and Veterans Administration Medical Center, Oklahoma City, Okla; and Heart Lung Center, Utrecht, the Netherlands (F.H.M.W.). Correspondence to Hiroshi Nakagawa, MD, PhD, Cardiac Arrhythmia Research Institute, University of Oklahoma Health Sciences Center, 1200 Everett Dr (TUH-6E-103), Oklahoma City, OK hiroshi-nakagawa@ouhsc.edu 2006 American Heart Association, Inc. Circulation is available at DOI: /CIRCULATIONAHA

2 12 Circulation January 3/10, 2006 Figure 1. Schematic representation of the irrigated electrode catheters. A, Closed loop irrigation catheter has 7F, 4-mm tip electrode with an internal thermocouple. A 5% dextrose solution at room temperature was circulated continuously through the tip electrode at a flow rate of 36 ml/min, cooling the ablation electrode internally. B, Open irrigation catheter has 7.5F, 3.5-mm tip electrode with an internal thermocouple and 6 irrigation holes (0.4-mm diameter) located around the electrode, 1.0 mm from the tip. Heparinized (1 U/min) normal saline at room temperature was irrigated through the electrode and 6 irrigation holes at a flow rate of 17 ml/min during each RF application, providing internal and external electrode cooling. range in areas of low blood flow. One approach is to increase the electrode surface area exposed to the blood by increasing electrode circumference and length (8 to 10 mm). 19,20 The greater surface area increases convective cooling in the same level of blood flow. The second, and potentially more effective, approach is fluid irrigation of the electrode either by circulating fluid within the electrode (closed loop system) or flushing saline through openings in the electrode (open irrigation system). 8,9,14,24 The active electrode cooling by irrigation allows sustained RF power, even at sites with low blood flow, to produce deeper lesions. We hypothesized that, at the same constant RF power, lesion depth would be similar between closed loop and open irrigation electrodes, but the effect of flushing around the electrode at the surface of the tissue by the open irrigation system (external cooling) would decrease the formation of thrombus during ablation. The purpose of this study was to use an established canine model to compare between closed loop and open irrigation electrodes the RF lesion depth, incidence of thrombus formation, and steam pop at the same RF power in both low blood flow and high blood flow conditions. Methods Irrigation Ablation Catheters The closed loop irrigation catheter has a 7F, 4-mm length tip electrode (Chilli, Boston Scientific; Figure 1A). A solution of 5% dextrose at room temperature was circulated continuously through the tip electrode at a flow rate of 36 ml/min (during and between ablation) The open irrigation catheter has a 7.5F, 3.5-mm length tip with 6 small irrigation holes (0.4-mm diameter) around the circumference of the electrode, 1 mm from the tip (ThermoCool, Biosense Webster; Figure 1B). Between RF applications, the electrode was irrigated with heparinized (1 U/mL) normal saline (0.9% NaCl) at room temperature through the central lumen and out the 6 irrigation holes at a flow rate of 1 ml/min. During ablation, the irrigation flow rate was increased to 17 ml/min (beginning 3 seconds before and continuing until 3 seconds after each RF Figure 2. Schematic representation of the canine thigh muscle preparation. A, The skin over the thigh muscle was incised and raised to form a cradle. The cradle was filled with heparinized blood (activated clotting time [ACT] 350 seconds) from the same animal at 37 C to 38 C. B, Each of the 2 irrigated electrodes was held perpendicular to the thigh muscle at 10 g contact weight. Two fluoroptic temperature probes were positioned on opposite sides of the electrode-tissue interface. Two other fluoroptic temperature probes were positioned within the tissue, adjacent to the electrode at depths of 3 and 7 mm. Pulsatile blood flow was directed around the electrode at either low (0.1 m/s) or high (0.5 m/s) velocity. application). Previous studies demonstrated that an irrigation flow rate of 17 ml/min was adequate to prevent thrombus formation during RF applications at power 30 W. 25,26 The surface area of the closed loop and open irrigation electrodes is comparable (33.6 versus 32.4 mm 2 ). Therefore, at any given power, the power delivered to the tissue is expected to be similar. Both catheters contain a thermocouple embedded within the tip electrode for monitoring electrode temperature during ablation. Experimental Preparation The experimental protocol was approved by the University of Oklahoma Committee on the Use and Care of Animals. Eight mongrel dogs weighing 25 to 32 kg were anesthetized with sodium pentobarbital (25 mg/kg) and ventilated mechanically with room air. The right carotid artery was cannulated for monitoring arterial pressure. The canine thigh muscle preparation has been described previously. 9,10,14 Each dog was placed on its right side. A 20-cm skin incision was made over the left thigh muscle. The skin on each side of the incision was dissected free of the connective tissue. The skin edges were raised to form a cradle that was filled with heparinized blood (activated clotting time 350 seconds) from the same animal at 37 C to 38 C (Figure 2A).

3 Yokoyama et al Internal vs Open Irrigation for RF Ablation 13 the pulsatile blood flow (Figure 2B). The electrode-tissue interface temperature was taken as the higher of the 2 temperatures. Two additional fluoroptic temperature probes were bundled together with shrink tubing and inserted into the thigh muscle adjacent to the ablation electrode at 3 and 7 mm below the muscle surface. The electrode-tissue interface and 2 tissue temperatures were measured at 125-ms intervals (Figure 3). Figure 3. Electric parameters and temperatures during ablation. A, Closed loop electrode. RF energy was delivered at constant power of 20 W in low blood flow (0.1 m/s). The RF application was maintained for the full 60 seconds without an impedance rise. Impedance decreased from 98 to 80. The electrode temperature was 26 C before ablation (because of continuous internal irrigation) and remained low (maximum 42 C) throughout the RF application. However, the 2 electrode-tissue interface temperatures (Temp) reached 90 C and 95 C, associated with formation of soft thrombus (shown in Figure 4). Tissue temperature at depths of 3.0 and 7.0 mm increased progressively, reaching 70 C and 41 C, respectively, associated with lesion depth of 6.2 mm. B, Open irrigation electrode. RF energy was delivered at constant power of 30 W in low blood flow (0.1 m/s). Electrode temperature decreased to 29 C before the RF application because of an increase in saline irrigation flow rate from 1 to 17 ml/min. The RF application was maintained for the full 60 seconds without an impedance rise. Electrode temperature increased to a maximum of 41 C. The interface temperature increased to only 52 C and did not produce a thrombus. The tissue temperatures at depths of 3 and 7 mm increased to 101 C and 59 C, respectively, associated with lesion depth of 7.8 mm. The closed loop or open irrigation catheter was held perpendicular to the thigh muscle at 10 g contact weight. A contact weight of 10 g was chosen to simulate typical catheter force. Earlier studies in this model showed minimal contact and tiny RF lesions at 5 g and excessive temperature and lesion size at 20 g. 9,10,14 Pulsatile blood flow (Master Flex, Cole-Parmer Instrument) was directed at the ablation electrode from a plastic tube located 1.5 cm from the ablation electrode. The device produced a peak flow velocity at the electrode of either 0.1 or 0.5 m/s, measured by pulsed Doppler(Figure 2B). A peak flow velocity of 0.1 m/s simulates ablation sites with low local blood flow. The velocity of 0.5 m/s simulates the flow at pulmonary vein ostia during sinus rhythm. Electrode temperature was measured from the thermocouple within the tip electrode at 20-ms intervals. Two 0.3-mm diameter fluoroptic temperature probes (Luxtron, model 3100; accuracy 0.1 C at 0 C to 160 C) were positioned on opposite sides of the electrode-tissue interface, with variable orientation to the direction of Ablation Protocol Radiofrequency energy (500 khz) was delivered between the ablation electrode and an adhesive electrosurgical dispersive pad (20 10 cm) applied to the shaved skin of the opposite thigh. A custom RF generator (Radionics, model RFG-3DJ) was used that allowed the recording of power, impedance, and electrode temperature at 20-ms intervals (Figure 3). The root-mean-square power and impedance and the electrode temperature, interface temperature, and tissue temperature were monitored continuously and recorded (Bard Lab- System). RF applications were delivered at constant power of 20 or 30 W for 60 seconds but were terminated immediately in the event of an impedance rise of 10 above the lowest impedance during the RF application. After each RF application, the catheter was left in place, and the cradle was emptied of blood. The ablation electrode and the electrode-tissue interface were examined for thrombus (Figure 4). The electrode was then removed, cleaned, and positioned at a new site for another RF application. RF current was delivered at 6 to 8 sites along the surface of the left thigh muscle. The skin incision was then closed, and the animal was rotated to expose the right thigh muscle. The right thigh muscle was prepared with the same procedure, and 6 to 8 applications of RF current were delivered. Two hours after the ablation procedure, 20 ml of 10% triphenyl tetrazolium chloride was administered intravenously. Triphenyl tetrazolium chloride stains intracellular dehydrogenase a deep red color, distinguishing viable (red) and necrotic (pale) tissue. The animals were euthanized, and the thigh muscles were excised and fixed in 10% formalin. The thigh muscles were sectioned to measure lesion size. Measurements of Lesion Size The maximum depth, maximum diameter, depth at the maximum diameter, and lesion surface diameter were measured (Figure 5). Statistical Analysis Values are presented as mean SD. The factorial design of the study involved 2 catheter types (closed loop and open irrigation), 2 blood flow levels (0.1 and 0.5 m/s), and 2 power levels (20 and 30 W) for a total of 8 conditions. The mean electrode temperature, interface temperature, and tissue temperature at 3.0- and 7.0-mm depths and lesion dimensions were calculated for each of the 8 conditions in each dog. The means among the 8 dogs of electrode temperature, interface temperature, and tissue temperature at 3.0- and 7.0-mm depths and lesion dimensions for RF applications at constant power (20 and 30 W) in low and high blood flow (0.1 and 0.5 m/s) were compared between the closed loop catheter and open irrigation catheter by repeated-measures ANOVA. The significance of the difference between the closed loop catheter and open irrigation catheter in incidence of thrombus formation and steam pop was assessed by McNemar tests for paired observations. The significance of the relationship between electrode temperature and interface temperature and between electrode temperature and lesion depth among the 8 dogs was assessed by a simple linear regression analysis for the closed loop and open irrigation catheters at constant power (20 and 30 W). A probability value 0.05 was considered statistically significant. Results A total of 121 RF applications were delivered in the 8 dogs with the use of either the closed loop electrode (n 63) or open irrigation electrode (n 58), at power of 20 or 30 W and in either low (0.1 m/s) or high (0.5 m/s) pulsatile blood

4 14 Circulation January 3/10, 2006 Figure 4. Thrombus at the electrodetissue interface produced by the RF application shown in Figure 3A, delivered with the closed loop electrode at 20 W in low blood flow. The thrombus formed on the electrode-tissue interface surrounding the ablation electrode despite the absence of an impedance rise. The size of soft thrombus was 7 4 mm. flow(table). The impedance at the onset of the RF applications was similar for the closed loop and open irrigation electrodes (94 7 versus 94 6, respectively), and the surface areas of the electrodes were similar, suggesting that a similar ratio of RF power was delivered between the tissue and blood for the 2 electrodes. As a result, for RF applications at the same constant power (20 or 30 W), there was no difference between the closed loop and open irrigation electrodes in tissue temperatures at depths of 3 and 7 mm or in maximum lesion depth in the setting of either low (0.1 m/s) or high (0.5 m/s) blood flow (Table and Figure 5). The maximum diameter was slightly smaller with open irrigation. The diameter at the surface of the lesion was smaller and the depth at the maximum lesion diameter was greater in lesions produced by the open irrigation electrode (Figure 5). Figure 5. A, Lesion dimensions for RF applications in low blood flow (0.1 m/s). Values are expressed in millimeters (mean SD). A, Maximum lesion depth; B, maximum lesion diameter; C, depth at maximum lesion diameter; and D, lesion surface diameter. B, Lesion dimensions for RF applications in high blood flow (0.5 m/s). *P 0.05, comparison between the closed loop electrode and open irrigation electrode at the same RF power (20 or 30 W) and same blood flow (0.1 or 0.5 m/s). P 0.05, comparison between blood flow of 0.1 and 0.5 m/s at the same RF power (20 or 30 W).

5 Yokoyama et al Internal vs Open Irrigation for RF Ablation 15 Comparison of Ablation Parameters Between Closed Loop and Open Irrigation Electrodes Blood Flow, m/s RF Power, W Irrigation Mode Electrode Temperature, C Interface Temperature, C Tissue Temperatures, C, at Depths 3 mm 7 mm Thrombus Formation Steam Pop Impedance Rise 0.1 m/s 20 Closed loop 43 4* 86 9* /18* 2/18 0/18 Open irrigation /15 0/15 0/15 30 Closed loop 50 5* * /13* 7/13* 4/13 Open irrigation /14 0/14 0/ m/s 20 Closed loop 41 3* 66 7* /17 2/17 0/17 Open irrigation /15 0/15 0/15 30 Closed loop 48 4* 80 14* /15* 8/15 2/15 Open irrigation /14 4/14 0/14 *P 0.05, comparison between closed loop electrode and open irrigation electrode at the same RF power (20 or 30 W). P 0.05, comparison between blood flow of 0.1 and 0.5 m/s at the same RF power (20 or 30 W). Thrombus Formation and Steam Pop: Relationship to Electrode and Electrode-Tissue Interface Temperatures Closed Loop Electrode For the closed loop electrode, there was no difference in the peak electrode temperature between RF applications in low blood flow and high blood flow at the same power (20 or 30 W). The peak interface temperature was significantly greater in low blood flow than high blood flow (Table). There was no significant relationship between the peak electrode temperature and the peak interface temperature for RF applications at 20 or 30 W in low blood flow or high blood flow among the 8 dogs. Thrombus occurred in 32 of the 63 RF applications (51%) with the closed loop electrode (Table, Figures 4 and 6). The peak interface temperature was 80 C in all 32 RF applications with thrombus. All but 1 of the 33 RF applications with peak interface temperature 80 C produced thrombus (Figures 3A and 6A). Only 8 of the 32 RF applications with thrombus had a peak interface temperature of 100 C. An impedance rise ( 10 ) occurred in only 6 (19%) of the 32 RF applications with thrombus. In high blood flow, thrombus occurred in RF applications with electrode temperature 43 C. In low blood flow, thrombus occurred at any electrode temperature (Figure 6). For low and high blood flow combined, thrombus occurred in 1 of 6 RF applications (17%) with peak electrode temperature 40 C, 26 of 50 RF applications (52%) with peak electrode temperature 40 C to 50 C, and 5 of 7 RF applications (71%) with peak electrode temperature 50 C (Figure 6B). Steam pops were identified by an audible pop or small (3 to 5 ), brief (usually in the range of 50 ms) increase in impedance. A steam pop occurred in 19 of the 63 RF applications (30%) with the use of the closed loop electrode: 4 of 35 RF applications (11%) at 20 W and 15 of 28 RF applications (54%) at 30 W (P 0.05; Table, Figure 6). At 30 W, RF applications with and without a steam pop were similar in electrode temperature (49 5 C versus 48 5 C) or interface temperature (86 18 C versus C), but there was a trend toward higher tissue temperature at depths of 3mm(94 19 C versus C) and 7 mm (58 16 C versus C). There was no significant difference between low blood flow and high blood flow in the incidence of a steam pop. An impedance rise of 10 occurred in only 5 of 19 RF applications with a steam pop (Figure 6). Open Irrigation Electrode The peak electrode temperature and interface temperature were lower at each RF power and blood flow rate for the open irrigation electrode than for the closed loop electrode (Table). The peak interface temperature was 71 C in all 58 RF applications with the open irrigation electrode, with no thrombus or impedance rise (Figure 6 and Table). The range of peak electrode temperature was narrow (30 C to 44 C), and there was no significant relationship between peak electrode temperature and peak interface temperature in RF applications at 20 or 30 W in low or high blood flow among the 8 dogs. There was no significant difference in electrode temperature or interface temperature between RF applications in low and high blood flow conditions (Table). A steam pop occurred in 4 of 58 RF applications (7%) with the open irrigation electrode (none of 30 RF applications at 20 W and 4 of 28 RF applications at 30 W; Figure 6 and Table). The incidence of steam pop at 30 W was significantly lower with the open irrigation electrode than with the closed loop electrode (14% versus 54% of applications; P 0.05) (Table). The 4 open irrigation RF applications associated with a steam pop had peak electrode temperature ranging from only 36 C to 40 C and peak interface temperature ranging from only 38 C to 50 C, with tissue temperatures at 3-mm depth ranging from 91 C to 100 C and at 7-mm depth ranging from 47 C to 72 C. As demonstrated for the closed loop electrode, electrode temperature or interface temperature was similar between RF applications with and without steam pop (Figure 6). Relationship Between Electrode Temperature and Lesion Depth Among the 8 dogs, there was no significant relationship between peak electrode temperature and lesion depth, except for RF ablations with the closed loop electrode at 20 W in high blood flow.

6 16 Circulation January 3/10, 2006 Discussion In this study RF ablation with closed loop and open irrigation electrodes produced lesions of comparable depth at the power range tested, in both low and high blood flow conditions. This is due to the similar electrode surface area, contact force, and overall impedance resulting in similar power delivered to the tissue. 14 Open irrigation was more effective in cooling the electrode-tissue interface, reflected by (1) lower interface temperature; (2) lower incidence of thrombus; and (3) and smaller lesion diameter at the surface (with the maximum diameter produced deeper in the tissue). These differences between the 2 electrodes were greater in low blood flow, Figure 6. A, The occurrence of thrombus, steam pop, and impedance rise in relation to peak electrode-tissue interface temperature with the use of the closed loop electrode and open irrigation electrode for RF applications in both low and high blood flow. Data are combined for 20- and 30-W applications. Thrombus occurred in 32 of 63 RF applications with the closed loop electrode; all had peak electrode-tissue interface temperature 80 C. An impedance rise ( 10 ) occurred in only 6 of 32 RF applications with thrombus. Thrombus occurred in 24 of 31 RF applications (77%) in low blood flow (peak electrode temperature at 32 C to 59 C; B) and 8 of 32 RF applications (25%) in high blood flow (peak electrode temperature at 43 C to 55 C; B). A steam pop occurred in 19 of 63 RF applications with the use of the closed loop electrode (9 of 31 in low blood flow and 10 of 32 in high blood flow). For the open irrigation electrode, there was no thrombus or impedance rise in any of 58 RF applications in low or high blood flow. A steam pop occurred in only 4 of 58 applications (7%) with the open irrigation electrode, all at 30 W. B, The occurrence of thrombus, steam pop, and impedance rise in relation to peak electrode temperature with the use of the closed loop electrode and open irrigation electrode for RF applications in both low and high blood flow. Data are combined for 20- and 30-W applications. *RF application with an impedance rise ( 10 ). presumably because the flow of saline irrigation out of the electrode provides additional cooling of the electrode-tissue interface ( external cooling ). Ablation with the closed loop electrode (with irrigation providing only internal cooling) in low blood flow frequently resulted in high electrode-tissue interface temperature (despite low electrode temperature) and thrombus formation. The lower interface temperature at the same electrode temperature in higher blood flow with the closed loop electrode suggests that much of the interface cooling is derived from surrounding blood flow (Figure 1A). In contrast, with the open irrigation electrode, the interface temperature was similar in low and high blood flow, suggest-

7 Yokoyama et al Internal vs Open Irrigation for RF Ablation 17 ing that interface cooling was independent of the local blood flow. This study provides strong evidence that interface temperature is the major determinant of thrombus formation. Thrombus occurred consistently at an interface temperature of 80 C (Figure 6). Although it is frequently thought that thrombus occurs at an interface temperature of 100 C, 1 only 8 of the 32 RF applications producing thrombus had an interface temperature 100 C. The critical interface temperature of 80 C for thrombus formation is the same as previously found for 4- and 8-mm length nonirrigated electrodes. 10 Thrombus formation was not dependent on an impedance rise. 10,27 An impedance rise of 10 occurred in only 6 of 32 RF applications (19%) producing thrombus (Figure 6). Therefore, monitoring electrode temperature and impedance during ablation may not predict the presence or absence of thrombus formation. None of the RF applications using the open irrigation electrode reached an interface temperature of 80 C, and none produced thrombus. Despite the lower interface temperature, tissue temperatures at depths of 3 and 7 mm and lesion depth were similar to RF applications in which the closed loop electrode with higher interface temperature was used. This demonstrates that interface temperature is not a determinant of lesion size. A recent study from Demolin et al 27 found that delivering RF current within a beaker of freshly drawn heparinized human blood consistently produced a soft thrombus around the ablation electrode at an electrode temperature of 80 C. This thrombus was similar in appearance to the thrombus occurring in the thigh muscle preparation in the present study. Scanning electron microscopy of the thrombi in human blood revealed a dense mesh of denaturized blood protein with entrapped and damaged red blood cells. This study also included RF ablation in heparinized human serum. Thrombi also occurred in human serum, and scanning electron microscopy images revealed a similar structure, without entrapped red blood cells. These observations suggest that the thrombus during RF ablation is related to heatinduced denaturization of blood proteins. Importantly, human blood developed thrombus at the same temperature (80 C) as canine blood in the present study. The thermal effects on the thermocouple embedded in the electrode (electrode temperature) are dependent on electrode heating from the tissue, internal cooling by the irrigation fluid, and external cooling from blood flow or open irrigation. The difference between the electrode temperature and interface temperature was greater with the closed loop electrode than with the open irrigation electrode, and the difference increased with higher power and lower blood flow (Table). There was no significant correlation between the peak electrode temperature and the peak interface temperature with the closed loop catheter in low or high blood flow, which limits use of the electrode temperature as a guide of interface temperature. With open irrigation, the additional cooling of the electrode-tissue interface by the fluid exiting the irrigation holes narrows the range of electrode temperature and further decreases the correlation between electrode temperature and interface temperature. A steam pop is the result of rupture of the tissue surface by steam produced within the tissue. 9,11,14,23 This study found a lower incidence of steam pop with open irrigation compared with the closed loop system for RF applications at 30 W. The reduction in steam pop with open irrigation may be due to greater cooling of the tissue at shallow sites. There was a trend toward lower tissue temperature at 3-mm depth with the open irrigation electrode (Table). This would suggest that most of the steam pops with the closed loop system result from steam formation close to the surface. The incidence of steam pop would be expected to increase at higher power (such as 50 W) for both electrodes. A recent study in which the closed loop electrode was used in swine hearts found that only 3 of 21 steam pops occurred at electrode temperature 40 C. 23 The authors suggested that maintaining electrode temperature 40 C would reduce the risk of steam pop. However, in the present study 56% of RF applications maintaining just 20 W in low blood flow resulted in an electrode temperature 40 C, limiting the use of electrode temperature monitoring to prevent steam pops in low blood flow areas. There was no significant relationship between lesion depth and electrode temperature except for RF applications in which the closed loop electrode at 20 W in high blood flow was used. These findings indicate the limited value of using electrode temperature to predict lesion depth during clinical ablation procedures. The lesion geometry differed between the open irrigation and closed loop systems. With the closed loop system, the diameter of the lesion was greatest close to the surface, especially in low blood flow. With open irrigation, the diameter of the lesion at the surface was significantly smaller than with the closed loop system, and the maximum diameter was displaced 1.6 to 4.2 mm (median 2.8 mm) below the surface, consistent with greater surface cooling (lower electrode-tissue interface temperature). Clinical Implications The most important clinical implication of this study is the difference in the potential for thrombus formation between the closed loop and open irrigation electrodes, especially in conditions of low blood flow and at higher power. The potentially higher risk of a thromboembolic event with the closed loop electrode is suggested in the following 2 studies. The incidence of stroke or transient ischemic attack was 2.7% in a series of patients undergoing RF ablation of ventricular tachycardia associated with structural heart diseases 21 and 3.2% in another series of patients undergoing ablation of atrial fibrillation. 22 In the second study, reducing RF power on detection of microbubbles eliminated embolic events, suggesting that microbubble formation was a reliable guide to thrombus formation. Importantly, there was no correlation between microbubble formation and electrode temperature. In agreement with our findings, electrode temperature was not a reliable guide to thrombus formation with the closed loop electrode system (Figure 6). In contrast, there was no occurrence of cerebrovascular embolism in a study of 240 patients undergoing ablation of ventricular tachycardia associated with prior myocardial in-

8 18 Circulation January 3/10, 2006 farction with the use of the open irrigation catheter. 28 Definite conclusions cannot be derived from comparisons between different series, but these clinical observations are in accordance with the findings in our experimental model. Study Limitations The primary limitation of this study is that it was performed in a canine thigh muscle preparation instead of a beating heart. This preparation was chosen to control electrode orientation, contact pressure, and local blood flow and allowed the correlation of electrode temperature with interface temperature and temperatures within the tissue. This correlation would be very difficult in a beating heart because of limitations in measuring interface temperature and tissue temperature. Most importantly, it allowed the identification of thrombus on the electrode-tissue interface after each RF application. This preparation also provided the conditions of low blood flow, simulating commonly encountered clinical conditions in which sustained RF power is required to create adequate lesions and either of the 2 types of irrigation catheters would be used (such as ablation of chronic atrial fibrillation and ventricular tachycardia associated with prior myocardial infarction). 6,7,15 The findings in previous studies in which the thigh muscle preparation was used have correlated with clinical experience. With the use of a 4-mm nonirrigated electrode, thrombus occurred at interface temperature 80 C. Maintaining electrode temperature 55 C was required to keep interface temperature 80 C and prevent thrombus. 10 Clinical studies have suggested a reduction in electrode temperature to 55 C for ablation in the left atrium to minimize the risk of thrombus formation. 6 Conclusions RF lesion parameters were compared between 2 methods of electrode irrigation (closed loop and open irrigation electrodes) at constant power of 20 and 30 W and in low and high blood flow conditions. The similar electrode surface area produced similar tissue power, resulting in similar tissue temperatures (3- and 7-mm depths) and lesion depth. The occurrence of thrombus at the electrode-tissue interface correlated highly with interface temperature ( 80 C) and not with electrode temperature. Thrombus occurred frequently with the closed loop electrode (interface temperature 80 C), despite low electrode temperature. The interface temperature and the incidence of thrombus were greater in low blood flow, suggesting that local blood flow significantly contributes to electrode-tissue interface cooling with the closed loop electrode. With the open irrigation electrode, interface temperature was low (all 71 C) in low blood flow as well as in high blood flow, with no occurrence of thrombus. The low interface temperature indicates that open irrigation provides additional external cooling of the electrode-tissue interface, which is independent of local blood flow. Acknowledgments This study was supported in part by a grant from Biosense Webster, Inc (Diamond Bar, Calif). Disclosures Dr Nakagawa is a consultant for Biosense Webster, Inc and Boston Scientific Japan. Dr Jackman is a consultant for Biosense Webster, Inc. Dr Wittkampf has served as an advisor to Endocardial Solutions, Inc., which is now a part of St. Jude Medical. The other authors report no conflicts. References 1. Haines DE, Verow AF. Observation on electrode-tissue interface temperature and effect on electrical impedance during radiofrequency ablation of ventricular myocardium. Circulation. 1990;82: Langberg JJ, Calkins H, El-Atassi R, Borganelli M, Leon A, Kalbfleish SJ, Morady F. Temperature monitoring during radiofrequency catheter ablation of accessory pathways. Circulation. 1992;86: Nath S, Dimarco JP, Haines DE. Basic aspects of radiofrequency catheter ablation. J Cardiovasc Electrophysiol. 1994;5: Calkins H, Prystowsky E, Carlson M, Klein LS, Saul JP, Gillette P. Temperature monitoring during radiofrequency catheter ablation procedures using closed loop control. Circulation. 1994;90: Wen Z, Chen S, Tai C, Chiang C, Lee S, Chen Y, Yu W, Huang J, Chan M. Temperature monitoring in radiofrequency catheter ablation of atrial flutter using the linear ablation technique. J Cardiovasc Electrophysiol. 1996;7: Jais P, Shah DC, Haissaguerre M, Takahashi A, Lavergne T, Hocini M, Garrigue S, Barold SS, Metayer PL, Clementy J. Efficacy and safety of septal and left-atrial linear ablation for atrial fibrillation. Am J Cardiol. 1999;84:139R 146R. 7. Nakagawa H, Aoyama H, Beckman KJ, Po SS, Wu R, Lockwood D, Spector P, Calame JD, Lustgarten DL, Herring L, Hasdemir C, Singh D, Lazzara R, Jackman WM. Relation between pulmonary vein firing and extent of left atrial-pulmonary vein connection in patients with atrial fibrillation. Circulation. 2004;109: Budde T, Borggrefe M, Podczeck A, Jacob B, Langwasser J, Frenzel H, Breithard G. Radiofrequency ablation: an improvement of ablation techniques in comparison to direct-current delivery. In: Breithardt G, Borggrefe M, Zipes DP, eds. Nonpharmacological Therapy of Tachyarrhythmias. Mount Kisco, NY: Futura Publishing Company Inc; 1987: Nakagawa H, Yamanashi WS, Pitha JV, Arruda M, Wang X, Ohtomo K, Beckman KJ, McClelland JH, Lazzara R, Jackman WM. Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation. 1995;91: Matsudaira K, Nakagawa H, Yamanashi SW, Wittkampf FHM, Pitha JV, Imai S, Lazzara R, Jackman WM. High incidence of thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control. Pacing Clin Electrophysiol. 2003;26: Kongsgaard E, Steen T, Jensen O, Aass H, Amlie JP. Temperature guided radiofrequency catheter ablation of myocardium: comparison of catheter tip and tissue temperatures in vitro. Pacing Clin Electrophysiol. 1997;20: Wittkampf FHM, Hauer RNW, Robles de Medina EO. Control of radiofrequency lesion size by power regulation. Circulation. 1989;80: Grumbrecht S, Neuzner J, Pitschner HF. Interrelation of tissue temperature versus flow velocity in two different kinds of temperature controlled catheter radiofrequency energy application. J Interv Card Electrophysiol. 1998;2: Nakagawa H, Wittkampf FHM, Yamanashi WS, Pitha JV, Imai S, Campbell B, Arruda M, Lazzara R, Jackman WM. Inverse relationship between electrode size and lesion size during radiofrequency ablation with active electrode cooling. Circulation. 1998;98: Nakagawa H, Shah N, Matsudaira K, Overholt E, Chandrasekaran K, Beckman KJ, Specter P, Calame JD, Rao A, Hasdemir C, Otomo K, Wang Z, Lazzara R, Jackman WM. Characterization of reentrant circuit in macroreentrant right atrial tachycardia after surgical repair of congenital heart disease: isolated channels between scars allow focal ablation. Circulation. 2001;103: Thaker RK, Klein GJ, Yee R, Zardini M. Embolic complications after radiofrequency catheter ablation. Am J Cardiol. 1994;74: Epstein MR, Knapp LD, Martindill M, Lulu JA, Triedman JK, Calkins H, Stephen Huang SK, Walsh EP, Saul JP. Embolic complications associated with radiofrequency catheter ablation. Am J Cardiol. 1996;77:

9 Yokoyama et al Internal vs Open Irrigation for RF Ablation Zhou L, Keane D, Reed G, Ruskin J. Thromboembolic complications of radiofrequency catheter ablation. J Cardiovasc Electrophysiol. 1999;10: Langberg JJ, Gallagher M, Strickberger SA, Amirana O. Temperatureguided radiofrequency catheter ablation with very large distal electrodes. Circulation. 1993;88: Otomo K, Yamanashi WS, Tondo C, Antz M, Bussey J, Pitha JV, Arruda M, Nakagawa H, Wittkampf FHM, Lazzara R, Jackman WM. Why a large tip electrode makes a deeper radiofrequency lesion: effects of increase in electrode cooling and electrode-tissue interface area. J Cardiovasc Electrophysiol. 1998;9: Calkins H, Epstein A, Packer D, Arria AM, Hummel J, Gilligan DM, Trusso J, Carlson M, Luceri R, Kopelman H, Wilber D, Wharton JM, Stevenson W. Catheter ablation of Ventricular tachycardia in patients with structural heart disease using cooled radiofrequency energy. JAm Coll Cardiol. 2000;35: Marrouche NF, Martin DO, Wazni O, Gillinov AM, Klein A, Bhargava M, Saad E, Bash D, Yamada H, Jaber W. Schweikert R, Tchou P, Abdul-Karim A, Saliba W, Natale A. Phased-array intracardiac echocardiography monitoring during pulmonary vein isolation in patients with atrial fibrillation: impact on outcome and complication. Circulation. 2003;107: Cooper JM, Sapp JL, Tedrow U, Pellegrini CP, Robinson D, Epstein LM, Stevenson WG. Ablation with an internally irrigated radiofrequency catheter: learning how to avoid steam pops. Heart Rhythm. 2004;3: Yamane T, Jais P, Shah DC, Hocini M, Peng JT, Deisenhofer I, Clementy J, Haissaguerre M. Efficacy and safety of an irrigated-tip catheter for the ablation of accessory pathways resistant to conventional radiofrequency ablation. Circulation. 2000;102: Matsudaira K, Nakagawa H, Yamanashi SW, Wittkampf FHM, Pitha JV, Otomo K, Lazzara R, Jackman WM. Effects of saline irrigation in preventing soft thrombus formation during radiofrequency ablation. Pacing Clin Electrophysiol. 1999;22:821. Abstract. 26. Matsudaira K, Nakagawa H, Yamanashi SW, Wittkampf FHM, Pitha JV, Wang Z, Lazzara R, Jackman WM. Differential effects of saline irrigation flow rate on lesion depth and lesion volume. Pacing Clin Electrophysiol. 1999;22:800. Abstract. 27. Demolin JM, Eick OJ, Munch K, Koukkick E, Nakagawa H, Wittkampf FHM. Soft thrombus formation in radiofrequency catheter ablation. Pacing Clin Electrophysiol. 2003;25: Stevenson WG, Wilber DJ, Natale A, Marchlinski FE, Soejima K, Nakagawa H, Jackman MW. Multicenter trial of irrigated RF ablation with electroanatomic mapping for ventricular tachycardia after myocardial infarction. Circulation. 2004;110:394. Abstract. CLINICAL PERSPECTIVE Thromboembolism is one of the most feared complications of catheter ablation in the left atrium and ventricle. With conventional radiofrequency (RF) ablation, this risk is small when electrode temperature is limited to 60 C. Two RF ablation systems that use irrigation to cool the electrode to allow greater power application and increased lesion size are now in clinical use: a closed loop system and an open system in which saline washes over the electrode into the blood stream. Parameters to guide use for optimal safety and efficacy are still being defined. Although the operator reads temperature measured from the electrode, this study again demonstrates that tissue temperature can be substantially greater than temperature measured from the electrode with either closed loop or open irrigation. Thrombus formation consistently occurs when the electrode-tissue interface temperature reaches 80 C and unfortunately cannot be predicted from the measured electrode temperature or the change in impedance measured during ablation. In contrast to the closed irrigation system, the electrode-tissue interface temperature never reached 80 C with open irrigation, and therefore no thrombus was produced. We also show that local blood flow over the electrode significantly contributes to electrode-tissue interface cooling with the closed loop electrode. Open irrigation provides additional external cooling of the electrode-tissue interface that is hoped to reduce the risk of thromboembolism during catheter ablation.

Basic concepts, techniques and safety issues of arrhythmia ablation

Basic concepts, techniques and safety issues of arrhythmia ablation EHRA educational review and preparatory course for accreditation examination Invasive Cardiac Electrophysiology examination preparatory course Basic concepts, techniques and safety issues of arrhythmia

More information

Just over a decade ago, the advent of tip temperature monitoring during radiofrequency (RF) energy

Just over a decade ago, the advent of tip temperature monitoring during radiofrequency (RF) energy EP Lab Digest - ISSN: 1535-2226 - Volume 3 - Issue 2:2003 (March/April) - March 2003 - Pages: 1, 8-11 From: http://www.eplabdigest.com/article/1382# Features: Cooled Tip Catheter Ablation - Linda Moulton

More information

Catheter Ablation of Atypical Atrial Flutter after Cardiac Surgery Using a 3-D Mapping System

Catheter Ablation of Atypical Atrial Flutter after Cardiac Surgery Using a 3-D Mapping System Catheter Ablation of Atypical Atrial Flutter after Cardiac Surgery Using a 3-D Mapping System Myung-Jin Cha Seil Oh ECG & EP CASES Myung-Jin Cha, MD, Seil Oh, MD, PhD, FHRS Department of Internal Medicine,

More information

Radiofrequency Ablation at Low Irrigation Flow Rates Using a Novel 12-Hole Gold Open-Irrigation Catheter

Radiofrequency Ablation at Low Irrigation Flow Rates Using a Novel 12-Hole Gold Open-Irrigation Catheter Radiofrequency Ablation at Low Irrigation Flow Rates Using a Novel 12-Hole Gold Open-Irrigation Catheter FERDI AKCA,* ENDRE ZIMA, M.D., PH.D., ESZTER M. VÉGH, PH.D., GÁBOR SZÉPLAKI, M.D., JUDIT SKOPÁL,

More information

Case Report Figure-8 Tachycardia Confined to the Anterior Wall of the Left Atrium

Case Report Figure-8 Tachycardia Confined to the Anterior Wall of the Left Atrium www.ipej.org 146 Case Report Figure-8 Tachycardia Confined to the Anterior Wall of the Left Atrium Ioan Liuba, M.D., Anders Jönsson, M.D., Håkan Walfridsson, M.D. Department of Cardiology, Heartcenter,

More information

Lesion Dimensions During Temperature-Controlled Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium

Lesion Dimensions During Temperature-Controlled Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Lesion Dimensions During Temperature-Controlled Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site, Electrode Size, and Convective Cooling Helen Høgh Petersen,

More information

Radiofrequency Energy: Irrigation and Alternate Catheters. Andreas Pflaumer

Radiofrequency Energy: Irrigation and Alternate Catheters. Andreas Pflaumer Radiofrequency Energy: Irrigation and Alternate Catheters Andreas Pflaumer Irrigated tip RF ablation Irrigated tip How does it work? Potential benefits? Potential risks? How is this relevant to pediatric

More information

Ablation Lesion Assessment

Ablation Lesion Assessment HRC 2016 Ablation Lesion Assessment The creation of effective and permanent lesions Ian Wright Imperial College Healthcare Wed 09:00-09:30 Hall 11 Objective Examine the role of existing strategies and

More information

Atrial ectopic beats within the pulmonary veins (PVs) are

Atrial ectopic beats within the pulmonary veins (PVs) are Phased-Array Intracardiac Echocardiography Monitoring During Pulmonary Vein Isolation in Patients With Atrial Fibrillation Impact on Outcome and Complications Nassir F. Marrouche, MD; David O. Martin,

More information

Characteristics of Rapid Rhythms Recorded Within Pulmonary Veins During Atrial Fibrillation

Characteristics of Rapid Rhythms Recorded Within Pulmonary Veins During Atrial Fibrillation Characteristics of Rapid Rhythms Recorded Within Pulmonary Veins During Atrial Fibrillation HIROSHI TADA, MEHMET ÖZAYDIN, HAKAN ORAL, BRADLEY P. KNIGHT, AMAN CHUGH, CHRISTOPH SCHARF, FRANK PELOSI, Jr.,

More information

Methods. Utrecht and Amsterdam, The Netherlands

Methods. Utrecht and Amsterdam, The Netherlands 231 Effects of Intracavitary Blood Flow and Electrode Target Distance on Radiofrequency Power Required for Transient Conduction Block in a Langendorff-Perfused Canine Model TIMOTHY A. SIMMERS, MD, JACQUES

More information

Analysis of Catheter-Tip (8-mm) and Actual Tissue Temperatures Achieved During Radiofrequency Ablation at the Orifice of the Pulmonary Vein

Analysis of Catheter-Tip (8-mm) and Actual Tissue Temperatures Achieved During Radiofrequency Ablation at the Orifice of the Pulmonary Vein Analysis of Catheter-Tip (8-mm) and Actual Tissue Temperatures Achieved During Radiofrequency Ablation at the Orifice of the Pulmonary Vein T. Jared Bunch, MD; G. Keith Bruce, MD; Susan B. Johnson, BS;

More information

Review Article Irrigated Tip Catheters for Radiofrequency Ablation in Ventricular Tachycardia

Review Article Irrigated Tip Catheters for Radiofrequency Ablation in Ventricular Tachycardia BioMed Research International Volume 2015, Article ID 389294, 6 pages http://dx.doi.org/10.1155/2015/389294 Review Article Irrigated Tip Catheters for Radiofrequency Ablation in Ventricular Tachycardia

More information

From the Division of Cardiology, Department of Internal Medicine, University of Michigan Medical Center, Ann Arbor, Michigan

From the Division of Cardiology, Department of Internal Medicine, University of Michigan Medical Center, Ann Arbor, Michigan 118 Reprinted with permission from JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Volume 13, No. 2, February 2002 Copyright 2002 by Futura Publishing Company, Inc., Armonk, NY 10504-0418 Differentiation

More information

480 April 2004 PACE, Vol. 27

480 April 2004 PACE, Vol. 27 Incremental Value of Isolating the Right Inferior Pulmonary Vein During Pulmonary Vein Isolation Procedures in Patients With Paroxysmal Atrial Fibrillation HAKAN ORAL, AMAN CHUGH, CHRISTOPH SCHARF, BURR

More information

CATHETER ABLATION FOR TACHYCARDIAS

CATHETER ABLATION FOR TACHYCARDIAS 190 CATHETER ABLATION FOR TACHYCARDIAS MASOOD AKHTAR, M.D. T ACHY ARRHYTHMIAS constitute a major cause of mortality and morbidity. The most serious manifestation of cardiac arrhythmia is sudden cardiac

More information

Several experimental studies have shown that electrode-tissue. Original Article

Several experimental studies have shown that electrode-tissue. Original Article Original rticle Relationship etween Catheter Contact Force and Radiofrequency Size and Incidence of Steam Pop in the eating Canine Heart Electrogram mplitude, Impedance, and Electrode Temperature re Poor

More information

Catheter Ablation: Atrial fibrillation (AF) is the most common. Another Option for AF FAQ. Who performs ablation for treatment of AF?

Catheter Ablation: Atrial fibrillation (AF) is the most common. Another Option for AF FAQ. Who performs ablation for treatment of AF? : Another Option for AF Atrial fibrillation (AF) is a highly common cardiac arrhythmia and a major risk factor for stroke. In this article, Dr. Khan and Dr. Skanes detail how catheter ablation significantly

More information

How to Distinguish Focal Atrial Tachycardia from Small Circuits and Reentry

How to Distinguish Focal Atrial Tachycardia from Small Circuits and Reentry How to Distinguish Focal Atrial Tachycardia from Small Circuits and Reentry Pierre Jaïs; Bordeaux, France IHU LIRYC ANR-10-IAHU-04 Equipex MUSIC imaging platform ANR-11-EQPX-0030 Eutraf HEALTH-F2-2010-261057

More information

Comparison of Irrigated Electrode Designs for Radiofrequency Ablation of Myocardium

Comparison of Irrigated Electrode Designs for Radiofrequency Ablation of Myocardium Journal of Interventional Cardiac Electrophysiology 5, 391 400, 2001 #2001 Kluwer Academic Publishers. Manufactured in The Netherlands. Comparison of Irrigated Electrode Designs for Radiofrequency Ablation

More information

Catheter mapping and ablation are increasingly performed

Catheter mapping and ablation are increasingly performed Teaching Rounds in Cardiac Electrophysiology Three-Dimensional Mapping of Cardiac Arrhythmias What Do the Colors Really Mean? Freddy Del Carpio Munoz, MD; Traci L. Buescher, RN; Samuel J. Asirvatham, MD

More information

Effects of Partial and Complete Ablation of the Slow Pathway on Fast Pathway Properties in Patients with Atrioventricular Nodal Reentrant Tachycardia

Effects of Partial and Complete Ablation of the Slow Pathway on Fast Pathway Properties in Patients with Atrioventricular Nodal Reentrant Tachycardia 645 Effects of Partial and Complete Ablation of the Slow Pathway on Fast Pathway Properties in Patients with Atrioventricular dal Reentrant Tachycardia S. ADAM STRICKBERGER, M.D., EMILE DAOUD, M.D., MARK

More information

Since pulmonary veins (PVs) have

Since pulmonary veins (PVs) have Case Report Hellenic J Cardiol 2011; 52: 371-376 Left Atrial-Pulmonary Vein Reentrant Tachycardia Following Pulmonary Vein Isolation Dionyssios Leftheriotis, Feifan Ouyang, Karl-Heinz Kuck II. Med. Abteilung,

More information

JACC: CLINICAL ELECTROPHYSIOLOGY VOL. 3, NO. 10, 2017 ª 2017 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION

JACC: CLINICAL ELECTROPHYSIOLOGY VOL. 3, NO. 10, 2017 ª 2017 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ª 2017 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION PUBLISHED BY ELSEVIER ISSN 2405-500X/$36.00 http://dx.doi.org/10.1016/j.jacep.2017.08.016 In Vitro Validation of the Lesion Size Index to Predict

More information

Peri-Mitral Atrial Flutter with Partial Conduction Block between Left Atrium and Coronary Sinus

Peri-Mitral Atrial Flutter with Partial Conduction Block between Left Atrium and Coronary Sinus Accepted Manuscript Peri-Mitral Atrial Flutter with Partial Conduction Block between Left Atrium and Coronary Sinus Ryota Isogai, MD, Seiichiro Matsuo, MD, Ryohsuke Narui, MD, Shingo Seki, MD;, Michihiro

More information

Kadlec Regional Medical Center Cardiac Electrophysiology

Kadlec Regional Medical Center Cardiac Electrophysiology Definition of electrophysiology study and ablation Kadlec Regional Medical Center Cardiac Electrophysiology Electrophysiology Study and Ablation An electrophysiology, or EP, study is a test of the heart

More information

Pulmonary Vein Antral Isolation Using an Open Irrigation Ablation Catheter for the Treatment of Atrial Fibrillation

Pulmonary Vein Antral Isolation Using an Open Irrigation Ablation Catheter for the Treatment of Atrial Fibrillation Journal of the American College of Cardiology Vol. 49, No. 15, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.12.041

More information

Where to draw the mitral isthmus line in catheter ablation of atrial fibrillation: histological analysis

Where to draw the mitral isthmus line in catheter ablation of atrial fibrillation: histological analysis European Heart Journal (2005) 26, 689 695 doi:10.1093/eurheartj/ehi095 Clinical research Where to draw the mitral isthmus line in catheter ablation of atrial fibrillation: histological analysis Fred H.M.

More information

How to ablate typical atrial flutter

How to ablate typical atrial flutter Europace (1999) 1, 151 155 HOW TO... SERIES How to ablate typical atrial flutter A. Takahashi, D. C. Shah, P. Jaïs and M. Haïssaguerre Electrophysiologie Cardiaque, Hopital Cardiologique du Haut-Lévêque,

More information

Catheter ablation is one of the primary treatments for

Catheter ablation is one of the primary treatments for Assessment of Catheter Tip Contact Force Resulting in Cardiac Perforation in Swine Atria Using Force Sensing Technology Francesco Perna, MD; E. Kevin Heist, MD, PhD; Stephan B. Danik, MD; Conor D. Barrett,

More information

II V 1 HRA 3 4 HB 5 6 HB 3 4 HB 1 2 CS 7 8 CS 5 6 CS 3 4 CS 1 2 ABL 3 4 ABL 1 2 RVA 3 4. T. Suga et al.

II V 1 HRA 3 4 HB 5 6 HB 3 4 HB 1 2 CS 7 8 CS 5 6 CS 3 4 CS 1 2 ABL 3 4 ABL 1 2 RVA 3 4. T. Suga et al. Table 1. The distance between the positions of the catheter tip Patient Location Before elimination of After elimination of Number of the the accessory pathway the accessory pathway accessory pathway RAO

More information

Reentry in a Pulmonary Vein as a Possible Mechanism of Focal Atrial Fibrillation

Reentry in a Pulmonary Vein as a Possible Mechanism of Focal Atrial Fibrillation 824 Reentry in a Pulmonary Vein as a Possible Mechanism of Focal Atrial Fibrillation BERNARD BELHASSEN, M.D., AHARON GLICK, M.D., and SAMI VISKIN, M.D. From the Department of Cardiology, Tel-Aviv Sourasky

More information

Raphael Rosso MD, Yuval Levi Med. Eng., Sami Viskin MD Tel Aviv Sourasky Medical Center

Raphael Rosso MD, Yuval Levi Med. Eng., Sami Viskin MD Tel Aviv Sourasky Medical Center Radiofrequency Ablation of Atrial Fibrillation: Comparison of Success Rate of Circular Ablation vs Point-by-Point Ablation with Contact Force Assessment in Paroxysmal and Persistent Atrial Fibrillation

More information

Several groups have reported the use of linear lesions

Several groups have reported the use of linear lesions Comparison of Epicardial and Endocardial Linear Ablation Using Handheld Probes Stuart P. Thomas, PhD, Duncan J. R. Guy, MBBS, Anita C. Boyd, BMedSc, Vicki E. Eipper, David L. Ross, MBBS, and Richard B.

More information

Circulation: Arrhythmia and Electrophysiology CHALLENGE OF THE WEEK

Circulation: Arrhythmia and Electrophysiology CHALLENGE OF THE WEEK A 14-year-old girl with Wolff-Parkinson-White syndrome and recurrent paroxysmal palpitations due to atrioventricular reentry tachycardia had undergone two prior failed left lateral accessory pathway ablations

More information

Utility of Atrial and Ventricular Cycle Length Variability in Determining the Mechanism of Paroxysmal Supraventricular Tachycardia

Utility of Atrial and Ventricular Cycle Length Variability in Determining the Mechanism of Paroxysmal Supraventricular Tachycardia 698 Utility of Atrial and Ventricular Cycle Length Variability in Determining the Mechanism of Paroxysmal Supraventricular Tachycardia THOMAS C. CRAWFORD, M.D., SIDDHARTH MUKERJI, M.D., ERIC GOOD, D.O.,

More information

Use of Catheter Ablation in the Treatment of Ventricular Tachycardia Triggered by Premature Ventricular Contraction

Use of Catheter Ablation in the Treatment of Ventricular Tachycardia Triggered by Premature Ventricular Contraction J Arrhythmia Vol 22 No 3 2006 Case Report Use of Catheter Ablation in the Treatment of Ventricular Tachycardia Triggered by Premature Ventricular Contraction sao Kato MD, Toru wa MD, Yasushi Suzuki MD,

More information

Mechanism of Immediate Recurrences of Atrial Fibrillation After Restoration of Sinus Rhythm

Mechanism of Immediate Recurrences of Atrial Fibrillation After Restoration of Sinus Rhythm Mechanism of Immediate Recurrences of Atrial Fibrillation After Restoration of Sinus Rhythm AMAN CHUGH, MEHMET OZAYDIN, CHRISTOPH SCHARF, STEVE W.K. LAI, BURR HALL, PETER CHEUNG, FRANK PELOSI, JR, BRADLEY

More information

Catheter ablation of atrial fibrillation: Indications and tools for improvement of the success rate of the method. Konstantinos P.

Catheter ablation of atrial fibrillation: Indications and tools for improvement of the success rate of the method. Konstantinos P. Ioannina 2015 Catheter ablation of atrial fibrillation: Indications and tools for improvement of the success rate of the method Konstantinos P. Letsas, MD, FESC SECOND DEPARTMENT OF CARDIOLOGY LABORATORY

More information

Electrical isolation of the pulmonary veins (PVs) to treat

Electrical isolation of the pulmonary veins (PVs) to treat Mechanisms of Organized Left Atrial Tachycardias Occurring After Pulmonary Vein Isolation Edward P. Gerstenfeld, MD; David J. Callans, MD; Sanjay Dixit, MD; Andrea M. Russo, MD; Hemal Nayak, MD; David

More information

Role of Transisthmus Conduction Intervals in Predicting Bidirectional Block after Ablation of Typical Atrial Flutter

Role of Transisthmus Conduction Intervals in Predicting Bidirectional Block after Ablation of Typical Atrial Flutter Reprinted with permission from JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Volume 12, No. 2, February 2001 Copyright 2001 by Futura Publishing Company, Inc., Armonk, NY 10504-0418 169 Role of Transisthmus

More information

Adult Echocardiography Examination Content Outline

Adult Echocardiography Examination Content Outline Adult Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 2 3 4 5 Anatomy and Physiology Pathology Clinical Care and Safety Measurement Techniques, Maneuvers,

More information

Atrial fibrillation (AF) is the most common sustained arrhythmia in adults and

Atrial fibrillation (AF) is the most common sustained arrhythmia in adults and 12 Chapter Atrial Fibrillation Ablation Atrial Fibrillation Ablation George Thomas, MD Amin Al-Ahmad, MD Executive Summary Atrial fibrillation (AF) is the most common sustained arrhythmia in adults and

More information

Author s Accepted Manuscript

Author s Accepted Manuscript Author s Accepted Manuscript Cavo-tricuspid isthmus high density mapping Antonio De Simone, Vincenzo La Rocca, Francesco Solimene, Francesco Maddaluno, Maurizio Malacrida, Giuseppe Stabile www.elsevier.com/locate/buildenv

More information

Prevalence of Asymptomatic Recurrences of Atrial Fibrillation After Successful Radiofrequency Catheter Ablation

Prevalence of Asymptomatic Recurrences of Atrial Fibrillation After Successful Radiofrequency Catheter Ablation 920 Prevalence of Asymptomatic Recurrences of Atrial Fibrillation After Successful Radiofrequency Catheter Ablation HAKAN ORAL, M.D., SRIKAR VEERAREDDY, M.D., ERIC GOOD, D.O., BURR HALL, M.D., PETER CHEUNG,

More information

Novel ECG Predictor of Difficult Cases of Outflow Tract Ventricular Tachycardia:

Novel ECG Predictor of Difficult Cases of Outflow Tract Ventricular Tachycardia: Circulation Journal Official Journal of the Japanese Circulation Society http://www.j-circ.or.jp ORIGINAL ARTICLE Arrhythmia/Electrophysiology Novel ECG Predictor of Difficult Cases of Outflow Tract Ventricular

More information

Catheter ablation procedures have evolved in recent years to

Catheter ablation procedures have evolved in recent years to Atrial Linear Ablations in Pigs Chronic Effects on Atrial Electrophysiology and Pathology Lior Gepstein, MD, PhD; Gal Hayam, BSc; Shlomo Shpun, DSc; David Cohen, MD; Shlomo A. Ben-Haim, MD, DSc Background

More information

Contemporary Strategies for Catheter Ablation of Atrial Fibrillation

Contemporary Strategies for Catheter Ablation of Atrial Fibrillation Contemporary Strategies for Catheter Ablation of Atrial Fibrillation Suneet Mittal, MD Director, Electrophysiology Medical Director, Snyder Center for Atrial Fibrillation The Arrhythmia Institute at The

More information

AF :RHYTHM CONTROL BY DR-MOHAMMED SALAH ASSISSTANT LECTURER CARDIOLOGY DEPARTMENT

AF :RHYTHM CONTROL BY DR-MOHAMMED SALAH ASSISSTANT LECTURER CARDIOLOGY DEPARTMENT AF :RHYTHM CONTROL BY DR-MOHAMMED SALAH ASSISSTANT LECTURER CARDIOLOGY DEPARTMENT 5-2014 Atrial Fibrillation therapeutic Approach Rhythm Control Thromboembolism Prevention: Recommendations Direct-Current

More information

CARDIOINSIGHT TM NONINVASIVE 3D MAPPING SYSTEM CLINICAL EVIDENCE SUMMARY

CARDIOINSIGHT TM NONINVASIVE 3D MAPPING SYSTEM CLINICAL EVIDENCE SUMMARY CARDIOINSIGHT TM NONINVASIVE 3D MAPPING SYSTEM CLINICAL EVIDENCE SUMMARY April 2017 SUPPORTING EVIDENCE RHYTHM AF VT PUBLICATIONS Driver Domains in Persistent Atrial Fibrillation (Haissaiguerre, et al)

More information

The major thoracic veins, with their specific electrical

The major thoracic veins, with their specific electrical Atrial Fibrillation Originating From Persistent Left Superior Vena Cava Li-Fern Hsu, MBBS; Pierre Jaïs, MD; David Keane, MD; J. Marcus Wharton, MD; Isabel Deisenhofer, MD; Mélèze Hocini, MD; Dipen C. Shah,

More information

Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter

Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter Caution: Federal (USA) law restricts this device to sale by or on the order of a physician. DEVICE DESCRIPTION

More information

Case Report Substrate Based Ablation of Ventricular Tachycardia Through An Epicardial Approach

Case Report Substrate Based Ablation of Ventricular Tachycardia Through An Epicardial Approach www.ipej.org 364 Case Report Substrate Based Ablation of Ventricular Tachycardia Through An Epicardial Approach Aman Makhija DM 1, Ajit Thachil DM 1, C Sridevi, DNB 2, B Hygriv Rao, DM 2, S Jaishankar

More information

Case Report. Sumito Narita MD 1;3, Takeshi Tsuchiya MD, PhD 1, Hiroya Ushinohama MD, PhD 2, Shin-ichi Ando MD, PhD 3

Case Report. Sumito Narita MD 1;3, Takeshi Tsuchiya MD, PhD 1, Hiroya Ushinohama MD, PhD 2, Shin-ichi Ando MD, PhD 3 Case Report Identification and Radiofrequency Catheter Ablation of a Nonsustained Atrial Tachycardia at the Septal Mitral Annulus with the Use of a Noncontact Mapping System: A Case Report Sumito Narita

More information

Case Report Catheter ablation of ventricular tachycardia related to a septo-apical left ventricular aneurysm

Case Report Catheter ablation of ventricular tachycardia related to a septo-apical left ventricular aneurysm Int J Clin Exp Med 2015;8(10):19576-19580 www.ijcem.com /ISSN:1940-5901/IJCEM0014701 Case Report Catheter ablation of ventricular tachycardia related to a septo-apical left ventricular aneurysm Radu Rosu

More information

Integration of CT and fluoroscopy images in the ablative treatment of atrial fibrillation

Integration of CT and fluoroscopy images in the ablative treatment of atrial fibrillation Clinical applications Integration of CT and fluoroscopy images in the ablative treatment of atrial fibrillation C. Kriatselis M. Tang M. Roser J-H. erds-li E. leck Department of Internal Medicine/Cardiology,

More information

Bipolar Radiofrequency Energy

Bipolar Radiofrequency Energy Bipolar Radiofrequency Energy Ralph J. Damiano, Jr., MD Evarts A. Graham Professor of Surgery Chief, Division of Cardiothoracic Surgery Barnes-Jewish Hospital Washington University School of Medicine St.

More information

Chapter 16: Arrhythmias and Conduction Disturbances

Chapter 16: Arrhythmias and Conduction Disturbances Complete the following. Chapter 16: Arrhythmias and Conduction Disturbances 1. Cardiac arrhythmias result from abnormal impulse, abnormal impulse, or both mechanisms together. 2. is the ability of certain

More information

Transcoronary Chemical Ablation of Atrioventricular Conduction

Transcoronary Chemical Ablation of Atrioventricular Conduction 757 Transcoronary Chemical Ablation of Atrioventricular Conduction Pedro Brugada, MD, Hans de Swart, MD, Joep Smeets, MD, and Hein J.J. Wellens, MD In seven patients with symptomatic atrial fibrillation

More information

Ventricular Tachycardia in Structurally Normal Hearts (Idiopathic VT) Patient Information

Ventricular Tachycardia in Structurally Normal Hearts (Idiopathic VT) Patient Information Melbourne Heart Rhythm Ventricular Tachycardia in Structurally Normal Hearts (Idiopathic VT) Patient Information What is Ventricular Tachycardia? Ventricular tachycardia (VT) is an abnormal rapid heart

More information

Evidence for Longitudinal and Transverse Fiber Conduction in Human Pulmonary Veins

Evidence for Longitudinal and Transverse Fiber Conduction in Human Pulmonary Veins Evidence for Longitudinal and Transverse Fiber Conduction in Human Pulmonary Veins Relevance for Catheter Ablation Javier E. Sanchez, MD; Vance J. Plumb, MD; Andrew E. Epstein, MD; G. Neal Kay, MD Background

More information

Case Report Catheter ablation of Atrial Incisional Tachycardia mistaken for Atrial flutter

Case Report Catheter ablation of Atrial Incisional Tachycardia mistaken for Atrial flutter www.ipej.org 134 Case Report Catheter ablation of Atrial Incisional Tachycardia mistaken for Atrial flutter Ottaviano L 1,2, Muto C 1, Carreras G 1, Canciello M 1, Tuccillo B 1. 1 Department of Cardiology,

More information

Determination of Inadvertent Atrial Capture During Para-Hisian Pacing

Determination of Inadvertent Atrial Capture During Para-Hisian Pacing Determination of Inadvertent Atrial Capture During Para-Hisian Pacing Manoj Obeyesekere, MBBS; Peter Leong-Sit, MD; Allan Skanes, MD; Andrew Krahn, MD; Raymond Yee, MD; Lorne J. Gula, MD; Matthew Bennett,

More information

Complex Left Atrial (CLA) Best Practices for Navigant

Complex Left Atrial (CLA) Best Practices for Navigant Complex Left Atrial (CLA) Best Practices for Navigant Ischemic Ventricular Tachycardia (VT) Best Practices for Navigant Ischemic VT Ablation Procedures Set Up In the absence of a septal defect or other

More information

Coronary sinus (CS) diverticula 1 5 have been associated

Coronary sinus (CS) diverticula 1 5 have been associated Coronary Sinus-Ventricular Accessory Connections Producing Posteroseptal and Left Posterior Accessory Pathways Incidence and Electrophysiological Identification Yingxian Sun, MD; Mauricio Arruda, MD; Kenichiro

More information

More Musing About the Interrelationships of Atrial Fibrillation and Atrial. Flutter and Their Clinical Implications

More Musing About the Interrelationships of Atrial Fibrillation and Atrial. Flutter and Their Clinical Implications More Musing About the Interrelationships of Atrial Fibrillation and Atrial Flutter and Their Clinical Implications Running title: Waldo; AF-AFL Interrelationships Albert L. Waldo, MD Case Western Reserve

More information

Ablation of Typical Atrial Flutte. Background: Large-tip (10 mm) catheters (LTCs) and openirrigation-tip

Ablation of Typical Atrial Flutte. Background: Large-tip (10 mm) catheters (LTCs) and openirrigation-tip Randomized Comparison of Cavotricuspid Isthmus Ablation for Atrial Flutter Using an Open Irrigation-Tip versus a Large-Tip Radiofrequency Ablation Catheter KARL J. ILG, M.D., MICHAEL KÜHNE, M.D., THOMAS

More information

Peri-mitral atrial flutter in patients with atrial fibrillation ablation

Peri-mitral atrial flutter in patients with atrial fibrillation ablation Peri-mitral atrial flutter in patients with atrial fibrillation ablation Seiichiro Matsuo, MD,* Matthew Wright, MBBS, PhD,* Sébastien Knecht, MD,* Isabelle Nault, MD,* Nicolas Lellouche, MD,* Kang-Teng

More information

Case Report Electroanatomical Mapping and Ablation of Upper Loop Reentry Atrial Flutter

Case Report Electroanatomical Mapping and Ablation of Upper Loop Reentry Atrial Flutter Hellenic J Cardiol 46: 74-78, 2005 Case Report Electroanatomical Mapping and blation of Upper Loop Reentry trial Flutter POSTOLOS KTSIVS, PNGIOTIS IONNIDIS, CHRLMOS VSSILOPOULOS, THIN GIOTOPOULOU, THNSIOS

More information

Role of LAA isolation in AF cure

Role of LAA isolation in AF cure MAM 2017, Zurich Role of LAA isolation in AF cure Sakis Themistoclakis, MD Director, Unit of Electrophysiology and Cardiac Pacing Department of Cardiothoracic & Vascular Medicine Ospedale dell Angelo,

More information

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment Karen L. Booth, MD, Lucile Packard Children s Hospital Arrhythmias are common after congenital heart surgery [1]. Postoperative electrolyte

More information

The presence of epicardial circuits may adversely affect

The presence of epicardial circuits may adversely affect Catheter Ablation of Ventricular Epicardial Tissue A Comparison of Standard and Cooled-Tip Radiofrequency Energy André d Avila, MD; Christopher Houghtaling, BS, MS; Paulo Gutierrez, MD; Olivera Vragovic,

More information

Radiofrequency ablation for atrial fibrillation

Radiofrequency ablation for atrial fibrillation European Heart Journal Supplements (2003) 5 (Supplement H), H34 H39 Radiofrequency ablation for atrial fibrillation P. Jaïs 1, D.C. Shah 2, M. Hocini 1, L. Macle 1, K. J. Choi 1, M. Haïssaguerre 1 and

More information

Mechanism of Ventricular Tachycardia Termination by Pacing at Left Ventricular Sites in Patients with Coronary Artery Disease

Mechanism of Ventricular Tachycardia Termination by Pacing at Left Ventricular Sites in Patients with Coronary Artery Disease Journal of Interventional Cardiac Electrophysiology 6, 35 41, 2002 C 2002 Kluwer Academic Publishers. Manufactured in The Netherlands. Mechanism of Ventricular Tachycardia Termination by Pacing at Left

More information

Ablation of Post-Infarct VTs Unmasking VT Isthmuses Using Pace-Mapping

Ablation of Post-Infarct VTs Unmasking VT Isthmuses Using Pace-Mapping Ablation of Post-Infarct VTs Unmasking VT Isthmuses Using Pace-Mapping Christian de Chillou, MD, PhD Department of Cardiology University Hospital Nancy, France 29/05/2015 Post-infarct mappable VT Isthmus

More information

Ankara, Turkey 2 Department of Cardiology, Division of Arrhythmia and Electrophysiology, Yuksek Ihtisas

Ankara, Turkey 2 Department of Cardiology, Division of Arrhythmia and Electrophysiology, Yuksek Ihtisas 258 Case Report Electroanatomic Mapping-Guided Radiofrequency Ablation of Adenosine Sensitive Incessant Focal Atrial Tachycardia Originating from the Non-Coronary Aortic Cusp in a Child Serhat Koca, MD

More information

The pulmonary veins have been demonstrated to often

The pulmonary veins have been demonstrated to often Pulmonary Vein Isolation for Paroxysmal and Persistent Atrial Fibrillation Hakan Oral, MD; Bradley P. Knight, MD; Hiroshi Tada, MD; Mehmet Özaydın, MD; Aman Chugh, MD; Sohail Hassan, MD; Christoph Scharf,

More information

Case Report Pulmonary Vein Compression After Implantation of a Left Atrial Appendage Occluder: Presentation and Discussion of a Case

Case Report Pulmonary Vein Compression After Implantation of a Left Atrial Appendage Occluder: Presentation and Discussion of a Case 194 Case Report Pulmonary Vein Compression After Implantation of a Left Atrial Appendage Occluder: Presentation and Discussion of a Case Maryam Ayati MD, Feifan Ouyang MD, KH Kuck MD Department of cardiology,

More information

Arrhythmia/Electrophysiology

Arrhythmia/Electrophysiology Arrhythmia/Electrophysiology Electrophysiological Findings During Ablation of Persistent Atrial Fibrillation With Electroanatomic Mapping and Double Lasso Catheter Technique Feifan Ouyang, MD; Sabine Ernst,

More information

Flow Effect on Lesion Formation in RF Cardiac Catheter Ablation

Flow Effect on Lesion Formation in RF Cardiac Catheter Ablation IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 48, NO. 4, APRIL 2001 425 Flow Effect on Lesion Formation in RF Cardiac Catheter Ablation Hong Cao, Student Member, IEEE, Vicken R. Vorperian, Supan Tungjitkusolmun,

More information

Impedance Monitoring During Radiofrequency Catheter Ablation in Humans

Impedance Monitoring During Radiofrequency Catheter Ablation in Humans Impedance onitoring During Radiofrequency Catheter Ablation in Humans ARK HARVEY, YOON-NYUN KI, JOAO SOUSA, RAEL EL-ATASSI, RED ORADY, HUGH CALKINS, and JONATHAN J. LANGBERG rom the Department of Intemal

More information

Should hybrid ablation be the standard of care instead of transcatheter ablation techniques?

Should hybrid ablation be the standard of care instead of transcatheter ablation techniques? Should hybrid ablation be the standard of care instead of transcatheter ablation techniques? Christian Shults, MD Assistant Professor, Georgetown University School of Medicine Cardiac Surgeon, Medstar

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Verma A, Champagne J, Sapp J, et al. Asymptomatic episodes of atrial fibrillation before and after catheter ablation: a prospective, multicenter study. JAMA Intern Med. Published

More information

Pulmonary vein isolation (PVI) has become the cornerstone. Original Article

Pulmonary vein isolation (PVI) has become the cornerstone. Original Article Original Article In Vivo Contact Force Analysis and Correlation With Tissue Impedance During Left Atrial Mapping and Catheter Ablation of Atrial Fibrillation Hisaki Makimoto, MD, PhD; Tina Lin, MD; Andreas

More information

A Narrow QRS Complex Tachycardia With An Apparently Concentric Retrograde Atrial Activation Sequence

A Narrow QRS Complex Tachycardia With An Apparently Concentric Retrograde Atrial Activation Sequence www.ipej.org 125 Case Report A Narrow QRS Complex Tachycardia With An Apparently Concentric Retrograde Atrial Activation Sequence Miguel A. Arias MD, PhD; Eduardo Castellanos MD, PhD; Alberto Puchol MD;

More information

Atrial fibrillation (AF) is associated with increased morbidity

Atrial fibrillation (AF) is associated with increased morbidity Ablation of Atrial Fibrillation with Concomitant Surgery Edward G. Soltesz, MD, MPH, and A. Marc Gillinov, MD Atrial fibrillation (AF) is associated with increased morbidity and mortality in coronary artery

More information

Catheter Ablation of Atrial Tachycardia Originating from the Tip of Right Atrial Appendage

Catheter Ablation of Atrial Tachycardia Originating from the Tip of Right Atrial Appendage Case Report Catheter Ablation of Atrial Tachycardia Originating from the Tip of Right Atrial Appendage Masaru Inoue MD, Takao Matsubara MD, Toshihiko Yasuda MD, Kenji Miwa MD, Tadatsugu Gamou MD, Hounin

More information

Electrophysiological Characteristics of Atrial Tachycardia After Pulmonary Vein Isolation of Atrial Fibrillation

Electrophysiological Characteristics of Atrial Tachycardia After Pulmonary Vein Isolation of Atrial Fibrillation Circulation Journal Official Journal of the Japanese Circulation Society http://www.j-circ.or.jp Advance Publication by J-STAGE REVIEW Electrophysiological Characteristics of Atrial Tachycardia After Pulmonary

More information

Atrial fibrillation ablation in patients with known sludge in the left atrial appendage

Atrial fibrillation ablation in patients with known sludge in the left atrial appendage J Interv Card Electrophysiol (2014) 40:147 151 DOI 10.1007/s10840-014-9892-0 Atrial fibrillation ablation in patients with known sludge in the left atrial appendage Mohammed Hajjiri & Scott Bernstein &

More information

Creating Transmural, Linear Epicardial Ablation Lesions via a Non-surgical, Percutaneous, Subxyphoid, Electrogram-guided Approach

Creating Transmural, Linear Epicardial Ablation Lesions via a Non-surgical, Percutaneous, Subxyphoid, Electrogram-guided Approach The Journal of Innovations in Cardiac Rhythm Management, 3 (2012), 647 653 ATRIAL FIBRILLATION RESEARCH ARTICLE Creating Transmural, Linear Epicardial Ablation Lesions via a Non-surgical, Percutaneous,

More information

Recent Advances in the Treatment of Arrhythmias

Recent Advances in the Treatment of Arrhythmias SPECIAL ARTICLE Circ J 2003; 67: 651 655 Recent Advances in the Treatment of Arrhythmias Augustus O Grant, MD Advances in endocardial mapping techniques and ablation have greatly increased the indications

More information

Catheter Ablation of a Complex Atrial Tachycardia after Surgical Repair of Tetralogy of Fallot Guided by Combined Noncontact and Contact Mapping

Catheter Ablation of a Complex Atrial Tachycardia after Surgical Repair of Tetralogy of Fallot Guided by Combined Noncontact and Contact Mapping J Arrhythmia Vol 26 No 1 2010 Case Report Catheter Ablation of a Complex Atrial Tachycardia after Surgical Repair of Tetralogy of Fallot Guided by Combined Noncontact and Contact Mapping Eitaro Fujii MD,

More information

Jurnal Teknologi FREQUENCY ESTIMATION FOR ATRIAL FIBRILLATION FROM HUMAN STUDIES. Full Paper. Anita Ahmad *

Jurnal Teknologi FREQUENCY ESTIMATION FOR ATRIAL FIBRILLATION FROM HUMAN STUDIES. Full Paper. Anita Ahmad * Jurnal Teknologi FREQUENCY ESTIMATION FOR ATRIAL FIBRILLATION FROM HUMAN STUDIES Anita Ahmad * Department of Control and Mechatronics Engineering, Faculty of Electrical Engineering, Universiti Teknologi

More information

Right and Left Atrial Flutter: How To Differentiate Them on the Basis of Surface Electrocardiogram?

Right and Left Atrial Flutter: How To Differentiate Them on the Basis of Surface Electrocardiogram? Right and Left Atrial Flutter: How To Differentiate Them on the Basis of Surface Electrocardiogram? G. INAMA, C. PEDRINAZZI,O.DURIN,P.GAZZANIGA,P.AGRICOLA Introduction Atrial flutter is a common arrhythmia

More information

AF ABLATION Concepts and Techniques

AF ABLATION Concepts and Techniques AF ABLATION Concepts and Techniques Antony F Chu, M.D. Director of Complex Ablation Arrhythmia Services Section Division of Cardiology at the Rhode Island and Miriam Hospital HIGHLIGHTS The main indications

More information

Electrical disconnection of pulmonary vein (PV) myocardium

Electrical disconnection of pulmonary vein (PV) myocardium Left Atrial Appendage Activity Masquerading as Pulmonary Vein Potentials Dipen Shah, MD; Michel Haissaguerre, MD; Pierre Jais, MD; Meleze Hocini, MD; Teiichi Yamane, MD; Laurent Macle, MD; Kee Joon Choi,

More information

2018 CODING AND REIMBURSEMENT FOR. Cardiac Surgical Ablation and Left Atrial Appendage Management

2018 CODING AND REIMBURSEMENT FOR. Cardiac Surgical Ablation and Left Atrial Appendage Management 2018 CODING AND REIMBURSEMENT FOR Cardiac Surgical Ablation and Left Atrial Appendage Management Introduction This information is shared for educational purposes and current as of January 2018. Healthcare

More information

Analysis During Sinus Rhythm of Critical Sites in Reentry Circuits of Postinfarction Ventricular Tachycardia

Analysis During Sinus Rhythm of Critical Sites in Reentry Circuits of Postinfarction Ventricular Tachycardia Journal of Interventional Cardiac Electrophysiology 7, 95 103, 2002 C 2002 Kluwer Academic Publishers. Manufactured in The Netherlands. Analysis During Sinus Rhythm of Critical Sites in Reentry Circuits

More information