Analysis of Catheter-Tip (8-mm) and Actual Tissue Temperatures Achieved During Radiofrequency Ablation at the Orifice of the Pulmonary Vein

Size: px
Start display at page:

Download "Analysis of Catheter-Tip (8-mm) and Actual Tissue Temperatures Achieved During Radiofrequency Ablation at the Orifice of the Pulmonary Vein"

Transcription

1 Analysis of Catheter-Tip (8-mm) and Actual Tissue Temperatures Achieved During Radiofrequency Ablation at the Orifice of the Pulmonary Vein T. Jared Bunch, MD; G. Keith Bruce, MD; Susan B. Johnson, BS; Alvaro Sarabanda, MD, PhD; Mark A. Milton, MD; Douglas L. Packer, MD Background Many ablative approaches in or near the orifice of the pulmonary vein (PV) have demonstrated success in eliminating atrial fibrillation. Despite current practice, there are no data regarding the in vivo efficacy and safety of an 8-mm catheter tip for ablation at the PV orifice. Methods and Results Ten mongrel dogs were studied. Thermocouples were implanted in the atrial muscle of the PV orifice. Intracardiac echocardiography monitored catheter position, tip/tissue orientation, and microbubble formation. Ninety-four ablations were performed for 120 seconds. A temperature discrepancy 10 C between the catheter tip and tissue occurred during 47 (50%) of the ablations. Despite termination of energy delivery, the average tissue temperature remained within 1 C of the achieved steady state for 9 seconds. A temperature discrepancy 10 C was more common in the right superior PV, with oblique catheter positioning, when tissue temperatures were 60 C or 80 C, and with type 1 or type 2 microbubble formation. However, microbubbles were not present in 7 (13%, type 1) and 10 (40%, type 2) ablations with tissue temperatures 80 C. The maximum tissue temperature achieved with non full-thickness lesions was C vs C (P ) for full-thickness lesions. Conclusions Marked discrepancies between catheter-tip and tissue temperatures occurred with higher temperatures, prolonged ablation times, and unfavorable catheter thermistor tissue contact. Also, these data suggest a conservative approach to atrial ablation, because full-thickness lesions were obtained when tissue temperatures reached 50 C to 60 C and the tissue retained high heat levels despite termination of radiofrequency energy. Finally, microbubbles are inconsistent markers of tissue overheating. (Circulation. 2004;110: ) Key Words: catheter ablation myocardium temperature veins fibrillation Predicated on pulmonary vein (PV) involvement in the pathophysiology of atrial fibrillation, many different ablative approaches at or near the orifice of the PV have demonstrated substantial therapeutic success. Nevertheless, a variety of complications, including PV stenosis, char and thrombus formation, and stroke and other embolic events remain a significant source of morbidity. The causes of these complications are incompletely understood, although several groups have suggested excess thermal injury or other components of the ablative strategy, such as repeated energy deliveries, as contributors. 1 6 However, recent data suggest that the catheter-tip temperatures used to titrate energy delivery may be unreliable. 7,8 These ongoing problems have provided the incentive for developing other means of grading energy titration, such as using the formation of microbubbles as a marker of tissue heating. 4,5 Optimism for such an approach is premature without studies to correlate microbubble formation with actual tissue temperatures. Alternatively, 8-mm and irrigatedtip technologies have been developed to optimize ablation efficacy while minimizing complications. Despite current clinical practice, few data exist regarding the in vivo efficacy and safety of using an 8-mm catheter tip for ablation at the PV orifice or in left atrial tissue. It is unclear whether the temperature recorded at the catheter tip accurately reflects tissue temperature at times or if an extensive cooled-tip effect is present near the PV orifice. 9 Therefore, the purposes of this study were to (1) determine the catheter-tip versus tissue temperature profile during ablation with an 8-mm tipped catheter, (2) determine the clinical factors that may result in inaccurate catheter-tip reflection of actual tissue temperatures, and (3) evaluate the impact of repetitive burns at a single site on the dynamics of tissue heating. Methods Study Protocol The study was approved by the Mayo Foundation Institutional Animal Care and Use Committee. Ten mongrel dogs weighing Received March 31, 2004; revision received June 16, 2004; accepted June 21, From the Division of Cardiovascular Disease, Department of Internal Medicine, Mayo Clinic, Rochester, Minn. Correspondence to Douglas L. Packer, MD, Heart Rhythm Services, Mayo Clinic, St. Marys Hospital, nd St SW, AL 2-416, Rochester, MN packer.douglas@mayo.edu 2004 American Heart Association, Inc. Circulation is available at DOI: /01.CIR

2 Bunch et al Atrial Ablation With an 8-mm Catheter 2989 TABLE 1. Baseline Ablation Characteristics Variable n (%) Ablation No. First 36 (38.3%) Second 32 (34.0%) Third 26 (27.7%) PV Right superior 38 (40.4%) Left superior 28 (29.8%) Left inferior 28 (29.8%) Catheter orientation Oblique 53 (56.4%) Perpendicular 26 (27.7%) Parallel 16 (17.0%) 60 C 72 (76.6%) 80 C 32 (34.0%) Tissue/catheter-tip temperature discrepancy 10 C 47 (50%) 20 C 28 (30%) Microbubbles Type 1 31 (33.0%) Type 2 25 (26.6%) between 30 and 40 kg were used. The animals were anesthetized with intravenous ketamine (10 mg/kg) and diazepam (0.5 mg/kg), intubated, and maintained on 1% to 3% isoflurane with positivepressure ventilation. The hair at both groins, jugular regions, and chest wall was shaved for bilateral thoracotomies and ablation patch positioning. The body temperature was maintained at 37 C by water-flow heating pads and warmed intravenous solutions. The surface ECG, temperature, and blood pressure were recorded before catheterization and every 30 minutes from the time of catheterization. Catheterization By a percutaneous approach, 5 sheaths were positioned: an 8F into the left external jugular vein, a 12F into the right external jugular Figure 2. Tissue (gray dots) and respective multisensor catheter temperatures are displayed. Solid line is exponential linear fit, which is labeled with respective R 2 value. Dotted lines are reference to 95% CIs of data. vein, an 8F into the right femoral vein and artery, and a 12F into the left femoral vein. A 6F coronary sinus catheter was introduced through the left external jugular vein and advanced to the distal coronary sinus for anatomic guidance and pacing. Thermocouple Placement After establishment of anesthesia, a small left-sided, posterolateral thoracotomy was performed. Careful dissection exposed the external aspects of each PV orifice. For each PV, 2 to 4 thermocouples were implanted in the atrial muscle orifice of the PV. A radiopaque marker was attached to the tip of the thermocouple for fluoroscopic localization. After completion of the left-sided implantation and ablation, the same procedure was followed to dissect out the right superior PV via a right thoracotomy. The PV orifice tissue temperature was measured directly by embedded IT-21 copper/constantan T-type thermocouples (Physitemp Instrument Inc) with a time constant of 0.08 second. The thermocouples were 0.23 mm in diameter. The sensor signals were amplified and filtered by low-pass filters to minimize the powersource 60-Hz and radiofrequency (RF)-source 500-kHz interference. An analog-to-digital converter (National Instruments) sampled the data at 2.8 samples per channel when acquiring 4 simultaneous channels. The accuracy of the temperature acquisition system was tested in vitro during RF ablation as previously reported. 8 In brief, postabla- Figure 1. Examples of ablations for which there was minimal discrepancy (A) and marked discrepancy (B) between tissue and cathetertip temperatures. Tip Temp indicates maximum catheter-tip temperature, and Tissue Temp, maximum tissue temperature.

3 2990 Circulation November 9, 2004 Figure 3. Average tissue and catheter-tip temperatures with corresponding tissue difference per 20-second intervals. Error bars are indicative of 1 SD from 20-second average. tion temperature decay was examined and compared with that of water-heated tissue to examine the effects of the thermocouples acting as RF antennas. Sampling frequency was set at 2.8 samples per second. The temperature measurement system was calibrated before use against a digital thermometer, with an accuracy of 0.42 C. Intracardiac Echocardiography A 10F, 64-element, 5.5- to 10.0-MHz, phased-array intracardiac echocardiography (ICE) catheter was introduced via the right external jugular vein and advanced to the level of the tricuspid annulus. ICE-guided transseptal catheterization to the left atrium was performed with an 8F Mullins sheath and a Brockenbrough needle, as previously described. 10 The ICE catheter was advanced through the left femoral vein to the level of the tricuspid annulus to monitor the catheter positioning, tip/tissue orientation, the development of microbubbles, and/or tissue changes during ablation. Catheterization and Ablation A temperature-controlled, multisensor catheter with an 8-mm-tip electrode was inserted through the right femoral sheath. The catheter system referred to as the tip in this article consisted of 4 temperature sensors embedded within the distal catheter, with 1 at the most distal tip location, and 3 located 120 axially from each other in a more proximal position. Power output was limited by the highest of the 4 thermistor temperatures. The maximum temperature from the 4 thermistor recordings during a 1-second interval was considered the maximum catheter-tip temperature for the subsequent analysis. Under guidance by fluoroscopy and ICE, the ablation catheter tip was placed endocardially, directly across from the epicardially implanted thermocouple. The catheter position and orientation were verified by fluoroscopy, ICE, and the radiopaque thermocouple marker. Ablation energy was first titrated to achieve a steady-state temperature of 50 C at the catheter tip. Once this temperature was reached, the RF power was increased to increment the tip temperature by 5 C. When the steady-state temperature was achieved, the process of increasing the temperature was repeated over the course of 120 seconds. The rate of temperature increase was 5 to 15 seconds, depending on the time required to reach a steady-state catheter-tip temperature. Simultaneous data acquisition of catheter-tip and PV tissue temperatures were made, and power and impedance were recorded. Occurrence of microbubbles was noted and characterized as scattered microbubbles (type 1) and a brisk shower of dense microbubbles (type 2). 4 The catheter temperature was increased throughout the ablation period, independent of microbubble formation. Pathology At the end of the study, ventricular fibrillation was induced with high-rate burst pacing, and the animal was exsanguinated. The entire heart-lung preparation was removed with the pericardium intact. A gross examination was performed at this time, during which both endocardial and epicardial tissue surfaces of ablation lesions were reviewed. Evidence of device-related trauma to any cardiac structure was documented. Ablation lesions were characterized as full or partial thickness. Statistical Analysis Continuous variables were reported as mean SD, and comparisons between groups were based on a 2-sample t test (parametric) or the Wilcoxon rank-sum test (nonparametric). Categorical variables were summarized as percentages, and group comparisons were based on the 2 test. Subsequent multivariate analyses were based on logisticregression models in which a temperature discrepancy of 10 C and 20 C and the development of type 1and type 2 microbubbles were dependent variables. In each analysis, odds ratios (ORs) with 95% confidence intervals (CIs) were used to characterize the association of ablation characteristics (catheter orientation, PV, ablation number, and temperature 60 C and 80 C) with the likelihood of developing the dependent variable. A fit of the in vivo PV orifice temperature was examined by plotting this versus the catheter-tip temperature. The data correlation and trendline were assessed by the highest R 2 result without evidence of lack of fit. Results General Ablation Characteristics A total of 94 ablations were performed at the superior and inferior PV orifices; the right superior PV in 38 (40%), left Figure 4. A, Tissue and catheter-tip temperature discrepancy for all ablations up to 120 seconds. Lines refer to percentages of ablation attempts with temperature 10 C (black) or 20 C (white). Numbers within lines represent actual number of ablations with temperature discrepancy per ablation attempt number. B, Subgroup of all ablations analyzed up to 60 seconds.

4 Bunch et al Atrial Ablation With an 8-mm Catheter 2991 TABLE 2. Characteristics of Ablation Attempts Averaged for 10-Second Intervals When a Steady-State Tissue Temperature Was Reached at 30, 60, and 90 Seconds Variable Ablation 1 Ablation 2 Ablation 3 P, C 30 Seconds Seconds Seconds Tip temperature, C 30 Seconds Seconds Seconds Tissue/tip discrepancy, C 30 Seconds Seconds Seconds Power, J 30 Seconds Seconds Seconds Impedence, 30 Seconds Seconds Seconds superior PV in 28 (30%), and left inferior PV in 28 (30%). The catheter position was oblique during 53 (56%) ablations, perpendicular in 26 (38%) ablations, and parallel in 15 (16%) ablations, as established by ICE and fluoroscopy. Details of the ablations are listed in Table 1. Examples showing the correlation between catheter-tip and tissue temperatures are shown in Figure 1 (A, minimal tissue/tip temperature discrepancy; B, marked [ 20 C] tissue/tip temperature discrepancy). Figure 2 displays the actual tissue temperatures achieved per individual catheter-tip temperature recording. Figure 3 displays the average tissue and catheter-tip temperatures per 20-second intervals. Impact of Ablation Number The ablation attempts were examined at 30, 60, and 90 seconds, for 10 seconds, after a tissue steady-state temperature was reached (Table 2). Despite similar power, there was a trend toward increased tissue temperatures with more ablations at 30, 60, and 90 seconds. The temperature discrepancy was 10 C during 18 (50%, ablation 1), 18 (56%, ablation 2), and 11 (42%, ablation 3) ablations. The discrepancy was 20 C during 9 (25%, ablation 1), 12 (38%, ablation 2), and 11 (42%, ablation 3) ablations (Figure 4A). Both the number of discrepancies 10 C and 20 C were reduced when assessing the first 60 seconds of energy delivery, although the incidence increased with ablation number (Figure 4B). Predictors of Temperature Discrepancy Although the overall average tissue and catheter-tip temperatures were similar, there were periods when there was a marked discrepancy. A temperature discrepancy 10 C was more common in the right superior PV, with an oblique catheter position, with tissue temperatures 60 C, and when type 1 or type 2 microbubbles were present (Table 3). In multivariate analysis, the ablation number and type 1 microbubbles were predictive of an increased risk of a 10 C tissue discrepancy (Figure 5A). Similarly, a marked temperature discrepancy of 20 C was more likely with ablation in the right superior PV, with an oblique catheter position, when tissue temperatures were 60 C, and when type 1 or type 2 microbubbles were present (Table 4). Multivariate predictors of a 20 C temperature discrepancy included ablation in the right superior PV, with a strong risk trend with type 2 microbubble appearance (Figure 5B). Microbubble Formation Type 1 microbubbles occurred at C versus type 2 microbubbles at C (P 0.016). Type 1 microbubble formation occurred more frequently with ablation of the right superior PV and with tissue temperatures 60 C (Table 5). However, in 7 (13%) ablations with tissue temperatures in excess of 80 C, no type 1 microbubbles were present. Multivariate predictors of type 1 microbubble formation included ablation in the right superior PV (OR 4.25; 95% CI, 2.98 to 5.52; P 0.039) and temperatures 60 C (OR 5.64; 95% CI, 3.91 to 7.40; P 0.018). Similarly, type 2 microbubble formation occurred more frequently with ablation of the right superior PV, with temperatures 60 C, and when type 1 microbubbles were present (Table 6). In multivariate analysis, the appearance of type 1 microbubbles

5 2992 Circulation November 9, 2004 TABLE 3. Characteristics of Ablation Attempts With Tissue and Catheter-Tip Temperature Discrepancies >10 C Variable (n) Temp 10 C (n 47) Temp 10 C (n 47) P Ablation No. (94) 0.572* First (36) 18 (38.3%) 18 (38.3%) Second (32) 14 (29.8%) 18 (38.3%) Third (26) 15 (31.9%) 11 (23.4%) PV (94) 0.034* Right superior (38) 13 (27.7%) 25 (53.2%) Left superior (28) 18 (38.3%) 10 (21.3%) Left inferior (28) 16 (34.0%) 12 (25.5%) Catheter orientation (94) 0.090* Oblique (53) 26 (55.3%) 27 (57.5%) Perpendicular (26) 10 (21.3%) 16 (34.0%) Parallel (16) 11 (23.4%) 4 (8.5%) 60 C (72) 25 (53.2%) 47 (100.0%) C (32) 4 (8.5%) 28 (59.6%) Microbubbles Type 1 (31) 7 (14.9%) 24 (51.1%) Type 2 (25) 5 (10.6%) 20 (42.6%) *P values are the significance of ablation No., PV, and catheter orientation relative to tissue/tip temperature discrepancy 10 C. TABLE 4. Characteristics of Ablation Attempts With Tissue and Catheter-Tip Temperature Discrepancies >20 C Variable (n) Temp 20 C (n 66) Temp 20 C (n 28) P Ablation No. (94) 0.500* First (36) 27 (40.9%) 9 (32.1%) Second (32) 20 (30.3%) 12 (42.9%) Third (26) 19 (28.8%) 7 (25.0%) PV (94) 0.007* Right superior (38) 20 (30.3%) 18 (64.3%) Left superior (28) 24 (36.4%) 4 (14.3%) Left inferior (28) 22 (33.3%) 6 (21.4%) Catheter orientation (94) 0.050* Oblique (53) 36 (54.6%) 17 (60.7%) Perpendicular (26) 16 (24.2%) 10 (35.7%) Parallel (15) 14 (21.2%) 1 (3.6%) 60 C (72) 44 (66.7%) 28 (100.0%) C (32) 9 (13.6%) 23 (82.1%) Microbubbles Type 1 (31) 16 (24.2%) 15 (53.6%) Type 2 (25) 10 (15.2%) 15 (53.6%) *P values are the significance of ablation No., PV, and catheter orientation relative to tissue/tip temperature discrepancy 20 C. predicted the occurrence of type 2 microbubble development (OR 3.35; 95% CI, 2.23 to 4.47; P 0.067). However, in 10 (40%) ablations, type 1 microbubbles did not precede the formation of type 2 microbubbles. In addition, type 2 microbubbles were not present in 10 (40%) ablations when temperatures were 80 C. Postablation Cooling Despite termination of energy delivery at 120 seconds, temperatures remained within 1 C of the achieved steady state for 9 seconds before decreasing (Figure 6). In comparison, tissue temperatures remained constant after 60- second ablations for only 6 seconds. The subsequent temperature loss was less dramatic after 60 seconds versus 120 seconds (maximum temperature decay for 60 seconds, 6 to 9 seconds: ablation 1, 0.53 C/s; ablation 2, 0.80 C/s; for 120 seconds, 10 to 13 seconds: ablation 1, 1.51 C/s; ablation 2, 1.82 C/s). Pathology At autopsy, no device-related trauma was noticed. Fullthickness lesions were achieved in all but 3 (3.2%) of the PV orifices. A total of 9 energy deliveries were delivered to these PV orifices. Figure 7 contains a histogram of ablation lesion Figure 5. Multivariate analysis of predictors associated with catheter-tip and tissue temperature discrepancy 10 C (A) and 20 C (B). Black circle at center is OR, with horizontal line representing 95% CI. Associated probability values are displayed below or alongside OR. MB indicates microbubbles.

6 Bunch et al Atrial Ablation With an 8-mm Catheter 2993 TABLE 5. Ablation Characteristics During Type 1 Microbubble (MB) Formation Variable (n) No MBs (n 53) Type 1 MBs (n 31) P Ablation No. (84)* First (31) 23 (43.4%) 8 (25.8%) Second (28) 16 (30.2%) 12 (38.7%) Third (25) 14 (24.6%) 11 (35.5%) PV (84)* Right superior (31) 13 (24.5%) 18 (58.1%) Left superior (27) 20 (37.7%) 7 (22.6%) Left inferior (26) 20 (37.7%) 6 (19.4%) Catheter orientation (84)* Oblique (47) 31 (58.5%) 16 (51.6%) Perpendicular (22) 11 (20.8%) 11 (35.5%) Parallel (15) 11 (20.8%) 4 (12.9%) 60 C (62) 33 (62.3%) 29 (93.6%) C (25) 7 (13.2%) 18 (58.1%) *In 10 ablations, type 2 microbubbles developed without precedent type 1 microbubbles. P values are the significance of ablation No., PV, and catheter orientation relative to type 1 microbubble formation. characteristics based on the maximum tissue temperature achieved. The maximum tissue temperature achieved during non full-thickness ablation attempts was C in comparison with a maximum temperature of C TABLE 6. Ablation Characteristics During Type 2 Microbubble (MB) Formation Variable (n) No Type 2 MBs (n 69) Type 2 MBs (n 25) P Ablation No. (94) 0.730* First (36) 28 (40.6%) 8 (32.0%) Second (32) 23 (33.3%) 9 (36.0%) Third (26) 18 (26.1%) 8 (32.0%) PV (94) 0.019* Right superior (38) 22 (31.8%) 16 (64.0%) Left superior (28) 23 (33.3%) 5 (20.0%) Left inferior (28) 24 (34.8%) 4 (16.0%) Catheter orientation (94) 0.395* Oblique (53) 38 (55.1%) 15 (60.0%) Perpendicular (26) 18 (26.1%) 8 (32.0%) Parallel (15) 13 (18.8%) 2 (8.0%) 60 C (72) 47 (68.1%) 25 (100.0%) C (32) 17 (24.6%) 15 (60.0%) Microbubbles Type 1 (31) 16 (23.2%) 15 (60.0%) *P values are the significance of ablation No., PV, and catheter orientation relative to type 2 microbubble formation. Figure 6. s after cessation of ablation (t 0). Positive delta temperature represents gain in temperature after ablation cessation, whereas negative delta temperature refers to loss. Each line is labeled with number corresponding to ablation attempt. (P ) achieved when full-thickness lesions were obtained. Discussion This study has several important findings relevant to the clinical practice of RF ablation of atrial arrhythmias. First, the 8-mm catheter-tip temperatures were correlated, in general, with tissue temperatures during 60-second energy deliveries. Second, after completion of the ablation, the steady-state temperature achieved persisted. In addition, the subsequent temperature decay was dependent on the initial ablation time. Third, large tissue and catheter-tip temperature discrepancies were more common with higher temperatures, an oblique catheter orientation, and prolonged ablation times. Fourth, estimation of tissue heating with either type 1 or type 2 microbubble formation as a guide was inaccurate, even at very high tissue temperatures ( 80 C). Although catheter-tip temperatures have been traditionally used to guide RF energy delivery, very little is known about the accompanying in vivo tissue temperatures. 11,12 Even less is known regarding tissue temperatures generated at the PV orifice. In general, in non-pv tissue, tissue temperatures are correlated with lesion size With excessive temperatures, tissue charring, thrombus formation, and impedance rises occur, and the potential for collateral tissue injury is increased. 13 These surface changes may provide the substrate for embolic phenomena. Furthermore, PVs subject to high temperatures ( 60 C) lose structural integrity of the underlying collagen and elastin matrix, with resultant circumferential narrowing. 3 These data underscore the need for accurate catheter-tip measurement of tissue temperatures. Newer catheter designs incorporating 8-mm and irrigatedtip technologies have been developed to optimize efficacy while minimizing complications With larger electrode sizes, the tip surface area exposed to the circulating blood increases, resulting in convective heat loss and tip cooling. 22 The process is similar to cooling produced by saline-irrigated tips. 15 Despite the use of these catheter designs in clinical practice, few data exist regarding the in vivo efficacy and safety. Furthermore, it is unclear whether temperatures recorded with an 8-mm tip accurately reflect tissue temperature

7 2994 Circulation November 9, 2004 Figure 7. Histogram characterizing 120- second ablation lesion characteristics based on maximum target temperature obtained. Lesions are characterized as full thickness (white) and partial thickness (black). at times or if an extensive cooled-tip effect is present near the PV orifice. This study provides in vivo data generated at the PV orifice with use of an 8-mm tip and suggests a good correlation between catheter-tip and tissue temperatures during the first 60 seconds of ablation. A time-dependent augmentation in catheter-tip and tissue temperature discrepancy may reflect a thermal latency of the tissue in vivo. As previously reported and demonstrated in this study, although tip temperatures dropped instantaneously with termination of RF energy, tissue temperatures may continue to be elevated. 15,23 25 Mechanistically, this is thought to result from 2 different heating processes. First, there is rapid resistive heating of a thin layer of tissue around the catheter tip during ablation. Second, there is a slower conductive heating of tissue at a distance. 24 With time, the overall impact of conductive heating is augmented, resulting in higher tissue temperatures despite consistent local resistive heating, as evidenced by the higher initial and latter steady-state temperatures with subsequent ablations in the same area. Furthermore, conductive energy persists after termination of energy delivery, which is directly related to ablation duration. These data explain why shorter ablation attempts may really reflect longer heating periods. These data suggest that catheter-tip orientation influences the accuracy of predicted tissue temperatures. In particular, the oblique position was most associated with large temperature discrepancies. This may, in part, reflect incomplete contact between the catheter-tip temperature sensor and the tissue. With this catheter design, the parallel/perpendicular positions allow a thermistor to be in direct contact with the tissue. If the temperature sensor is exposed to a mixed blood/tissue interface, as with the oblique orientation, convective heat loss into the circulating blood may result in lower registered temperatures at the catheter tip. In turn, the catheter will increase the power erroneously, resulting in higher tissue temperatures. As demonstrated previously, in high-blood-flow states, there is an increased tissue-to catheter-tip temperature discrepancy in a tissue-bath model in comparison with low-blood-flow states. 9 In the canine model, a perpendicular tip/tissue orientation in the right superior PV is difficult to achieve because of the angle created in the small left atrium. Furthermore, a parallel position is difficult to achieve in the right superior PV without entering the PV. Anatomic obstacles that prevent close contact of sensing thermistors to the tissue may, in part, account for the temperature discrepancies seen with the right superior PV in this study. These data reinforce the need to verify catheter position by ICE and fluoroscopy to ensure thermistor contact with the tissue. Clinical Implications This study suggests that a conservative approach to temperature titration be pursued. Marked catheter-tip and tissue temperature discrepancies were more likely to occur at higher temperatures. In addition, all lesions that reached a temperature of 60 C were transmural. Previous reports have demonstrated that irreversible electrophysiologic changes to myocytes occur early at 50 C. 26 Because only 1 ablation 55 C did not result in a transmural lesion, these data suggest that the target tissue temperature at the PV orifice should be 55 C. Furthermore, tissue temperatures remain elevated even after termination of RF energy. These data suggest that shorter ablation times may achieve therapeutic outcomes while minimizing complications associated with overheating. These data also demonstrate the fallibility of titration of energy to achieve microbubbles as a surrogate marker of tissue overheating. This finding is of important clinical relevance because of the increasing practice of titrating ablation energy to type 1 microbubble formation. 4,5 Although, on average, microbubble formation was correlated with excessive temperatures, there were ablation attempts with temperatures 80 C without microbubbles. Furthermore, type 2 microbubbles did not always follow type 1 microbubble formation. Limitations Data from an in vivo canine model may not be similar to human tissue characteristics and responses; however, this model represents a close approximation. Second, the thermocouples were implanted in a single area and may not fully represent 3-dimensional, conductive tissue heat transfer. Third, the data were generated from an 8-mm temperature-

8 Bunch et al Atrial Ablation With an 8-mm Catheter 2995 controlled catheter, and results may be different with other ablation tip and system designs. Fourth, the pathologic findings in this study must be viewed in the context of multiple regional ablation attempts. However, the interpretation of tissue temperatures in which nontransmural lesions occurred is more straightforward, thus allowing a general tissue temperature target. Fifth, these data were generated in left atrial tissue at the PV orifice. Linear ablation at other locations within the left atrium was not examined. Conclusions Current ablation approaches are limited because of the imprecise means of assessing actual in vivo tissue temperatures. Without accurate means to determine tissue temperatures, ablative complications that result from overheating will persist. This is a novel study revealing the actual in vivo tissue temperatures while using an 8-mm temperaturecontrolled catheter. In general, the catheter-tip temperatures were similar to those in tissue up to 60 seconds. Prolonged ablation times, tissue/tip contact, and catheter orientation influenced the accuracy of tissue temperature prediction. However, even after cessation of ablation, tissue temperatures remain elevated. To minimize excessive tissue heating, these findings should be considered when additional ablation attempts are made in a targeted area or when ablation delivery times are long. Additional studies of other catheter types in current ablative practice need to be performed to determine their in vivo accuracy to prevent complications associated with tissue overheating. Acknowledgment This study was supported by an unrestricted research grant from Boston Scientific/EPT Technologies. References 1. Calkins H, Hall J, Ellenbogen K, et al. A new system for catheter ablation of atrial fibrillation. Am J Cardiol. 1999;83:227D 236D. 2. Taylor GW, Kay GN, Zheng X, et al. Pathological effects of extensive radiofrequency energy applications in the pulmonary veins in dogs. Circulation. 2000;101: Kok LC, Everett THT, Akar JG, et al. Effect of heating on pulmonary veins: how to avoid pulmonary vein stenosis. J Cardiovasc Electrophysiol. 2003;14: Marrouche NF, Martin DO, Wazni O, et al. Phased-array intracardiac echocardiography monitoring during pulmonary vein isolation in patients with atrial fibrillation: impact on outcome and complications. Circulation. 2003;107: Saad EB, Marrouche NF, Saad CP, et al. Pulmonary vein stenosis after catheter ablation of atrial fibrillation: emergence of a new clinical syndrome. Ann Intern Med. 2003;138: Saad EB, Rossillo A, Saad CP, et al. Pulmonary vein stenosis after radiofrequency ablation of atrial fibrillation: functional characterization, evolution, and influence of the ablation strategy. Circulation. 2003;108: Farah A, Khan F, Machado C. Thrombus formation at the site of radiofrequency catheter ablation. Pacing Clin Electrophysiol. 2000;23: Su WW, Johnson SB, Packer DL. Marked discrepancies between catheter-tip and actual tissue temperatures achieved during RF ablation of pulmonary veins. Circulation. 2002;106:3116. Abstract. 9. Matsudaira K, Nakagawa H, Wittkampf FH, et al. High incidence of thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control. Pacing Clin Electrophysiol. 2003;26: Johnson SB, Seward JB, Packer DL. Phased-array intracardiac echocardiography for guiding transseptal catheter placement: utility and learning curve. Pacing Clin Electrophysiol. 2002;25: McRury ID, Panescu D, Mitchell MA, et al. Nonuniform heating during radiofrequency catheter ablation with long electrodes: monitoring the edge effect. Circulation. 1997;96: McRury ID, Whayne JG, Haines DE. Temperature measurement as a determinant of tissue heating during radiofrequency catheter ablation: an examination of electrode thermistor positioning for measurement accuracy. J Cardiovasc Electrophysiol. 1995;6: Petersen HH, Svendsen JH. Can lesion size during radiofrequency ablation be predicted by the temperature rise to a low power test pulse in vitro? Pacing Clin Electrophysiol. 2003;26: Petersen HH, Chen X, Pietersen A, et al. s and lesion size during irrigated tip catheter radiofrequency ablation: an in vitro comparison of temperature-controlled irrigated tip ablation, powercontrolled irrigated tip ablation, and standard temperature-controlled ablation. Pacing Clin Electrophysiol. 2000;23: Nakagawa H, Yamanashi WS, Pitha JV, et al. Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation. 1995;91: Thiagalingam A, Campbell CR, Boyd A, et al. Catheter intramural needle radiofrequency ablation creates deeper lesions than irrigated tip catheter ablation. Pacing Clin Electrophysiol. 2003;26: Dorwarth U, Fiek M, Remp T, et al. Radiofrequency catheter ablation: different cooled and noncooled electrode systems induce specific lesion geometries and adverse effects profiles. Pacing Clin Electrophysiol. 2003;26: Macle L, Jais P, Weerasooriya R, et al. Irrigated-tip catheter ablation of pulmonary veins for treatment of atrial fibrillation. J Cardiovasc Electrophysiol. 2002;13: Khargi K, Deneke T, Haardt H, et al. Saline-irrigated, cooled-tip radiofrequency ablation is an effective technique to perform the maze procedure. Ann Thorac Surg. 2001;72:S1090 S Nakagawa H, Wittkampf FH, Yamanashi WS, et al. Inverse relationship between electrode size and lesion size during radiofrequency ablation with active electrode cooling. Circulation. 1998;98: Petersen HH, Chen X, Pietersen A, et al. Temperature-controlled irrigated tip radiofrequency catheter ablation: comparison of in vivo and in vitro lesion dimensions for standard catheter and irrigated tip catheter with minimal infusion rate. J Cardiovasc Electrophysiol. 1998;9: Chan RC, Johnson SB, Seward JB, et al. The effect of ablation electrode length and catheter tip to endocardial orientation on radiofrequency lesion size in the canine right atrium. Pacing Clin Electrophysiol. 2002; 25: Nath S, Whayne JG, Kaul S, et al. Effects of radiofrequency catheter ablation on regional myocardial blood flow: possible mechanism for late electrophysiological outcome. Circulation. 1994;89: Haines DE. The biophysics of radiofrequency catheter ablation in the heart: the importance of temperature monitoring. Pacing Clin Electrophysiol. 1993;16: Wittkampf FH, Nakagawa H, Yamanashi WS, et al. Thermal latency in radiofrequency ablation. Circulation. 1996;93: Nath S, Lynch C 3rd, Whayne JG, et al. Cellular electrophysiological effects of hyperthermia on isolated guinea pig papillary muscle: implications for catheter ablation. Circulation. 1993;88:

Basic concepts, techniques and safety issues of arrhythmia ablation

Basic concepts, techniques and safety issues of arrhythmia ablation EHRA educational review and preparatory course for accreditation examination Invasive Cardiac Electrophysiology examination preparatory course Basic concepts, techniques and safety issues of arrhythmia

More information

Atrial ectopic beats within the pulmonary veins (PVs) are

Atrial ectopic beats within the pulmonary veins (PVs) are Phased-Array Intracardiac Echocardiography Monitoring During Pulmonary Vein Isolation in Patients With Atrial Fibrillation Impact on Outcome and Complications Nassir F. Marrouche, MD; David O. Martin,

More information

Ablation Lesion Assessment

Ablation Lesion Assessment HRC 2016 Ablation Lesion Assessment The creation of effective and permanent lesions Ian Wright Imperial College Healthcare Wed 09:00-09:30 Hall 11 Objective Examine the role of existing strategies and

More information

480 April 2004 PACE, Vol. 27

480 April 2004 PACE, Vol. 27 Incremental Value of Isolating the Right Inferior Pulmonary Vein During Pulmonary Vein Isolation Procedures in Patients With Paroxysmal Atrial Fibrillation HAKAN ORAL, AMAN CHUGH, CHRISTOPH SCHARF, BURR

More information

Radiofrequency Energy: Irrigation and Alternate Catheters. Andreas Pflaumer

Radiofrequency Energy: Irrigation and Alternate Catheters. Andreas Pflaumer Radiofrequency Energy: Irrigation and Alternate Catheters Andreas Pflaumer Irrigated tip RF ablation Irrigated tip How does it work? Potential benefits? Potential risks? How is this relevant to pediatric

More information

Bipolar Radiofrequency Energy

Bipolar Radiofrequency Energy Bipolar Radiofrequency Energy Ralph J. Damiano, Jr., MD Evarts A. Graham Professor of Surgery Chief, Division of Cardiothoracic Surgery Barnes-Jewish Hospital Washington University School of Medicine St.

More information

Evidence for Longitudinal and Transverse Fiber Conduction in Human Pulmonary Veins

Evidence for Longitudinal and Transverse Fiber Conduction in Human Pulmonary Veins Evidence for Longitudinal and Transverse Fiber Conduction in Human Pulmonary Veins Relevance for Catheter Ablation Javier E. Sanchez, MD; Vance J. Plumb, MD; Andrew E. Epstein, MD; G. Neal Kay, MD Background

More information

Just over a decade ago, the advent of tip temperature monitoring during radiofrequency (RF) energy

Just over a decade ago, the advent of tip temperature monitoring during radiofrequency (RF) energy EP Lab Digest - ISSN: 1535-2226 - Volume 3 - Issue 2:2003 (March/April) - March 2003 - Pages: 1, 8-11 From: http://www.eplabdigest.com/article/1382# Features: Cooled Tip Catheter Ablation - Linda Moulton

More information

How to ablate typical atrial flutter

How to ablate typical atrial flutter Europace (1999) 1, 151 155 HOW TO... SERIES How to ablate typical atrial flutter A. Takahashi, D. C. Shah, P. Jaïs and M. Haïssaguerre Electrophysiologie Cardiaque, Hopital Cardiologique du Haut-Lévêque,

More information

Intra-atrial Temperatures in Radiofrequency Endocardial Ablation: Histologic Evaluation of Lesions

Intra-atrial Temperatures in Radiofrequency Endocardial Ablation: Histologic Evaluation of Lesions Intra-atrial Temperatures in Radiofrequency Endocardial Ablation: Histologic Evaluation of Lesions Teresa Santiago, MS, João Q. Melo, MD, PhD, Rosa H. Gouveia, MD, and Ana P. Martins, MD Instituto do Coração,

More information

Cardiac Imaging in abnormal rhythm Role of MDCT

Cardiac Imaging in abnormal rhythm Role of MDCT Cardiac Imaging in abnormal rhythm Role of MDCT Cardiac Imaging in abnormal rhythm Role of MDCT Scope of the problem CT in Atrial Fibrillation CT and pacing Ventricular arrhythmia Other applications 1

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Verma A, Champagne J, Sapp J, et al. Asymptomatic episodes of atrial fibrillation before and after catheter ablation: a prospective, multicenter study. JAMA Intern Med. Published

More information

Επιπλοκές κατάλυσης πνευµονικών φλεβών

Επιπλοκές κατάλυσης πνευµονικών φλεβών Επιπλοκές κατάλυσης πνευµονικών φλεβών Παναγιώτης Ιωαννίδης Διευθυντής Τµήµατος Αρρυθµιών & Επεµβατικής Ηλεκτροφυσιολογίας Βιοκλινικής Αθηνών ΣΕΜΙΝΑΡΙΑ ΟΜΑΔΩΝ ΕΡΓΑΣΙΑΣ Ιωάννινα, 27-2-2015 Solving an equation

More information

Characteristics of Rapid Rhythms Recorded Within Pulmonary Veins During Atrial Fibrillation

Characteristics of Rapid Rhythms Recorded Within Pulmonary Veins During Atrial Fibrillation Characteristics of Rapid Rhythms Recorded Within Pulmonary Veins During Atrial Fibrillation HIROSHI TADA, MEHMET ÖZAYDIN, HAKAN ORAL, BRADLEY P. KNIGHT, AMAN CHUGH, CHRISTOPH SCHARF, FRANK PELOSI, Jr.,

More information

Radiofrequency Ablation at Low Irrigation Flow Rates Using a Novel 12-Hole Gold Open-Irrigation Catheter

Radiofrequency Ablation at Low Irrigation Flow Rates Using a Novel 12-Hole Gold Open-Irrigation Catheter Radiofrequency Ablation at Low Irrigation Flow Rates Using a Novel 12-Hole Gold Open-Irrigation Catheter FERDI AKCA,* ENDRE ZIMA, M.D., PH.D., ESZTER M. VÉGH, PH.D., GÁBOR SZÉPLAKI, M.D., JUDIT SKOPÁL,

More information

Integration of CT and fluoroscopy images in the ablative treatment of atrial fibrillation

Integration of CT and fluoroscopy images in the ablative treatment of atrial fibrillation Clinical applications Integration of CT and fluoroscopy images in the ablative treatment of atrial fibrillation C. Kriatselis M. Tang M. Roser J-H. erds-li E. leck Department of Internal Medicine/Cardiology,

More information

Mechanism of Immediate Recurrences of Atrial Fibrillation After Restoration of Sinus Rhythm

Mechanism of Immediate Recurrences of Atrial Fibrillation After Restoration of Sinus Rhythm Mechanism of Immediate Recurrences of Atrial Fibrillation After Restoration of Sinus Rhythm AMAN CHUGH, MEHMET OZAYDIN, CHRISTOPH SCHARF, STEVE W.K. LAI, BURR HALL, PETER CHEUNG, FRANK PELOSI, JR, BRADLEY

More information

Ventricular Tachycardia in Structurally Normal Hearts (Idiopathic VT) Patient Information

Ventricular Tachycardia in Structurally Normal Hearts (Idiopathic VT) Patient Information Melbourne Heart Rhythm Ventricular Tachycardia in Structurally Normal Hearts (Idiopathic VT) Patient Information What is Ventricular Tachycardia? Ventricular tachycardia (VT) is an abnormal rapid heart

More information

JACC: CLINICAL ELECTROPHYSIOLOGY VOL. 3, NO. 10, 2017 ª 2017 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION

JACC: CLINICAL ELECTROPHYSIOLOGY VOL. 3, NO. 10, 2017 ª 2017 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ª 2017 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION PUBLISHED BY ELSEVIER ISSN 2405-500X/$36.00 http://dx.doi.org/10.1016/j.jacep.2017.08.016 In Vitro Validation of the Lesion Size Index to Predict

More information

METHODS. Journal of the American College of Cardiology Vol. 37, No. 6, by the American College of Cardiology ISSN /01/$20.

METHODS. Journal of the American College of Cardiology Vol. 37, No. 6, by the American College of Cardiology ISSN /01/$20. Journal of the American College of Cardiology Vol. 37, No. 6, 2001 2001 by the American College of Cardiology ISSN 0735-1097/01/$20.00 Published by Elsevier Science Inc. PII S0735-1097(01)01208-6 Ablation

More information

Catheter Ablation: Atrial fibrillation (AF) is the most common. Another Option for AF FAQ. Who performs ablation for treatment of AF?

Catheter Ablation: Atrial fibrillation (AF) is the most common. Another Option for AF FAQ. Who performs ablation for treatment of AF? : Another Option for AF Atrial fibrillation (AF) is a highly common cardiac arrhythmia and a major risk factor for stroke. In this article, Dr. Khan and Dr. Skanes detail how catheter ablation significantly

More information

Peri-Mitral Atrial Flutter with Partial Conduction Block between Left Atrium and Coronary Sinus

Peri-Mitral Atrial Flutter with Partial Conduction Block between Left Atrium and Coronary Sinus Accepted Manuscript Peri-Mitral Atrial Flutter with Partial Conduction Block between Left Atrium and Coronary Sinus Ryota Isogai, MD, Seiichiro Matsuo, MD, Ryohsuke Narui, MD, Shingo Seki, MD;, Michihiro

More information

AF ABLATION Concepts and Techniques

AF ABLATION Concepts and Techniques AF ABLATION Concepts and Techniques Antony F Chu, M.D. Director of Complex Ablation Arrhythmia Services Section Division of Cardiology at the Rhode Island and Miriam Hospital HIGHLIGHTS The main indications

More information

Since pulmonary veins (PVs) have

Since pulmonary veins (PVs) have Case Report Hellenic J Cardiol 2011; 52: 371-376 Left Atrial-Pulmonary Vein Reentrant Tachycardia Following Pulmonary Vein Isolation Dionyssios Leftheriotis, Feifan Ouyang, Karl-Heinz Kuck II. Med. Abteilung,

More information

Contemporary Strategies for Catheter Ablation of Atrial Fibrillation

Contemporary Strategies for Catheter Ablation of Atrial Fibrillation Contemporary Strategies for Catheter Ablation of Atrial Fibrillation Suneet Mittal, MD Director, Electrophysiology Medical Director, Snyder Center for Atrial Fibrillation The Arrhythmia Institute at The

More information

Kadlec Regional Medical Center Cardiac Electrophysiology

Kadlec Regional Medical Center Cardiac Electrophysiology Definition of electrophysiology study and ablation Kadlec Regional Medical Center Cardiac Electrophysiology Electrophysiology Study and Ablation An electrophysiology, or EP, study is a test of the heart

More information

Electrical disconnection of pulmonary vein (PV) myocardium

Electrical disconnection of pulmonary vein (PV) myocardium Left Atrial Appendage Activity Masquerading as Pulmonary Vein Potentials Dipen Shah, MD; Michel Haissaguerre, MD; Pierre Jais, MD; Meleze Hocini, MD; Teiichi Yamane, MD; Laurent Macle, MD; Kee Joon Choi,

More information

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network INTERNAL CARDIOVERSION Lancashire & South Cumbria Cardiac Network DC Cardioversion AF & AFL safe 1,2 efficacious 3,4 (60-94%) 5 SR - Increases exercise tolerance 6 Maintainence SR unlikely in patients

More information

2013 PHYSICIAN PROCEDURE CODE CHANGES

2013 PHYSICIAN PROCEDURE CODE CHANGES 2013 PHYSICIAN PROCEDURE CODE CHANGES Page 1 of 7 Effective for dates of service on or after 1/1/2013, refer to the New Codes listed below for billing. The discontinued codes are not valid for billing

More information

Pulmonary Vein Antral Isolation Using an Open Irrigation Ablation Catheter for the Treatment of Atrial Fibrillation

Pulmonary Vein Antral Isolation Using an Open Irrigation Ablation Catheter for the Treatment of Atrial Fibrillation Journal of the American College of Cardiology Vol. 49, No. 15, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.12.041

More information

CATHETER ABLATION CODING & REIMBURSEMENT GUIDE. Updated September 2018

CATHETER ABLATION CODING & REIMBURSEMENT GUIDE. Updated September 2018 CATHETER ABLATION CODING & REIMBURSEMENT GUIDE Updated September 2018 TABLE OF CONTENTS Diagnosis Codes...3 ICD-10-CM Diagnosis Codes Coverage for Catheter Ablation Procedures....4 Medicare Other Payers

More information

Several groups have reported the use of linear lesions

Several groups have reported the use of linear lesions Comparison of Epicardial and Endocardial Linear Ablation Using Handheld Probes Stuart P. Thomas, PhD, Duncan J. R. Guy, MBBS, Anita C. Boyd, BMedSc, Vicki E. Eipper, David L. Ross, MBBS, and Richard B.

More information

2018 CODING AND REIMBURSEMENT FOR. Cardiac Surgical Ablation and Left Atrial Appendage Management

2018 CODING AND REIMBURSEMENT FOR. Cardiac Surgical Ablation and Left Atrial Appendage Management 2018 CODING AND REIMBURSEMENT FOR Cardiac Surgical Ablation and Left Atrial Appendage Management Introduction This information is shared for educational purposes and current as of January 2018. Healthcare

More information

Balloon and Mesh Catheter Ablation of Pulmonary Veins

Balloon and Mesh Catheter Ablation of Pulmonary Veins Balloon and Mesh Catheter Ablation of Pulmonary Veins ISHNE 2009 Atul Verma, MD FRCPC Cardiology/Electrophysiology Heart Rhythm Program, Division of Cardiology Southlake Regional Health Centre Newmarket,

More information

National Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION. Original Date: October 2015 Page 1 of 5

National Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION. Original Date: October 2015 Page 1 of 5 National Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION CPT Codes: 93451, 93452, 93453, 93454, 93455, 93456, 93457, 93458, 93459, 93460, 93461 LCD ID Number:

More information

2004 3 32 3 Chin J Cardiol, March 2004, Vol. 32 No. 3 211 4 ( ) 4 (HRA) (CS), 10 (Lasso ),, 4 (3 ) (1 ), 118,3,1, 417, ; ; The electrophysiological characteristics and ablation treatment of patients with

More information

Comparison of Irrigated Electrode Designs for Radiofrequency Ablation of Myocardium

Comparison of Irrigated Electrode Designs for Radiofrequency Ablation of Myocardium Journal of Interventional Cardiac Electrophysiology 5, 391 400, 2001 #2001 Kluwer Academic Publishers. Manufactured in The Netherlands. Comparison of Irrigated Electrode Designs for Radiofrequency Ablation

More information

The radial procedure was developed as an outgrowth

The radial procedure was developed as an outgrowth The Radial Procedure for Atrial Fibrillation Takashi Nitta, MD The radial procedure was developed as an outgrowth of an alternative to the maze procedure. The atrial incisions are designed to radiate from

More information

Catheter Ablation of Atrial Tachycardia Originating from the Tip of Right Atrial Appendage

Catheter Ablation of Atrial Tachycardia Originating from the Tip of Right Atrial Appendage Case Report Catheter Ablation of Atrial Tachycardia Originating from the Tip of Right Atrial Appendage Masaru Inoue MD, Takao Matsubara MD, Toshihiko Yasuda MD, Kenji Miwa MD, Tadatsugu Gamou MD, Hounin

More information

Case Report Pulmonary Vein Compression After Implantation of a Left Atrial Appendage Occluder: Presentation and Discussion of a Case

Case Report Pulmonary Vein Compression After Implantation of a Left Atrial Appendage Occluder: Presentation and Discussion of a Case 194 Case Report Pulmonary Vein Compression After Implantation of a Left Atrial Appendage Occluder: Presentation and Discussion of a Case Maryam Ayati MD, Feifan Ouyang MD, KH Kuck MD Department of cardiology,

More information

A Narrow QRS Complex Tachycardia With An Apparently Concentric Retrograde Atrial Activation Sequence

A Narrow QRS Complex Tachycardia With An Apparently Concentric Retrograde Atrial Activation Sequence www.ipej.org 125 Case Report A Narrow QRS Complex Tachycardia With An Apparently Concentric Retrograde Atrial Activation Sequence Miguel A. Arias MD, PhD; Eduardo Castellanos MD, PhD; Alberto Puchol MD;

More information

Effects of myocardial infarction on catheter defibrillation threshold

Effects of myocardial infarction on catheter defibrillation threshold Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1983 Effects of myocardial infarction on catheter defibrillation threshold

More information

AF :RHYTHM CONTROL BY DR-MOHAMMED SALAH ASSISSTANT LECTURER CARDIOLOGY DEPARTMENT

AF :RHYTHM CONTROL BY DR-MOHAMMED SALAH ASSISSTANT LECTURER CARDIOLOGY DEPARTMENT AF :RHYTHM CONTROL BY DR-MOHAMMED SALAH ASSISSTANT LECTURER CARDIOLOGY DEPARTMENT 5-2014 Atrial Fibrillation therapeutic Approach Rhythm Control Thromboembolism Prevention: Recommendations Direct-Current

More information

Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter

Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter Caution: Federal (USA) law restricts this device to sale by or on the order of a physician. DEVICE DESCRIPTION

More information

Histopathologic Characterization of Chronic Radiofrequency Ablation Lesions for Pulmonary Vein Isolation

Histopathologic Characterization of Chronic Radiofrequency Ablation Lesions for Pulmonary Vein Isolation Journal of the American College of Cardiology Vol. 59, No. 10, 2012 2012 by the American College of Cardiology Foundation ISSN 0735-1097/$36.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2011.09.076

More information

CRT Implantation Techniques 부천세종병원순환기내과박상원

CRT Implantation Techniques 부천세종병원순환기내과박상원 Cardiac Venous System and CRT Implantation Techniques 부천세종병원순환기내과박상원 Cardiac Resynchronization Therapy (CRT) Goal: Atrial synchronous biventricular pacing Transvenous approach for left ventricular lead

More information

Hybrid Ablation of AF in the Operating Room: Is There a Need? MAZE III Procedure. Spectrum of Atrial Fibrillation

Hybrid Ablation of AF in the Operating Room: Is There a Need? MAZE III Procedure. Spectrum of Atrial Fibrillation Hybrid Ablation of AF in the Operating Room: Is There a Need? MAZE III Procedure Paul J. Wang, MD Amin Al-Ahmad, MD Gan Dunnington, MD Stanford University Cox J, et al. Ann Thorac Surg. 1993;55:578-580.

More information

Linear Ablation Should Not Be a Standard Part of Ablation in Persistent AF. Disclosures. LA Ablation vs. Segmental Ostial Ablation With PVI for PAF

Linear Ablation Should Not Be a Standard Part of Ablation in Persistent AF. Disclosures. LA Ablation vs. Segmental Ostial Ablation With PVI for PAF Linear Ablation Should Not Be a Standard Part of Ablation in Persistent AF The CA Heart Rhythm Symposium September 7, 2012 Gregory K. Feld, MD Professor of Medicine Director, Cardiac EP Program University

More information

Case Report. Sumito Narita MD 1;3, Takeshi Tsuchiya MD, PhD 1, Hiroya Ushinohama MD, PhD 2, Shin-ichi Ando MD, PhD 3

Case Report. Sumito Narita MD 1;3, Takeshi Tsuchiya MD, PhD 1, Hiroya Ushinohama MD, PhD 2, Shin-ichi Ando MD, PhD 3 Case Report Identification and Radiofrequency Catheter Ablation of a Nonsustained Atrial Tachycardia at the Septal Mitral Annulus with the Use of a Noncontact Mapping System: A Case Report Sumito Narita

More information

CIPG Transcatheter Aortic Valve Replacement- When Is Less, More?

CIPG Transcatheter Aortic Valve Replacement- When Is Less, More? CIPG 2013 Transcatheter Aortic Valve Replacement- When Is Less, More? James D. Rossen, M.D. Professor of Medicine and Neurosurgery Director, Cardiac Catheterization Laboratory and Interventional Cardiology

More information

The pulmonary veins have been demonstrated to often

The pulmonary veins have been demonstrated to often Pulmonary Vein Isolation for Paroxysmal and Persistent Atrial Fibrillation Hakan Oral, MD; Bradley P. Knight, MD; Hiroshi Tada, MD; Mehmet Özaydın, MD; Aman Chugh, MD; Sohail Hassan, MD; Christoph Scharf,

More information

Economic Analysis of RF Transseptal Puncture. NRG Transseptal Needle USA

Economic Analysis of RF Transseptal Puncture. NRG Transseptal Needle USA Economic Analysis of RF Transseptal Puncture NRG Transseptal Needle USA Save up to $1,856 per case in savings through improved clinical outcomes with use of Baylis Medical's RF transseptal puncture technology.

More information

A atrial rate of 250 to 350 beats per minute that usually

A atrial rate of 250 to 350 beats per minute that usually Use of Intraoperative Mapping to Optimize Surgical Ablation of Atrial Flutter Shigeo Yamauchi, MD, Richard B. Schuessler, PhD, Tomohide Kawamoto, MD, Todd A. Shuman, MD, John P. Boineau, MD, and James

More information

The presence of epicardial circuits may adversely affect

The presence of epicardial circuits may adversely affect Catheter Ablation of Ventricular Epicardial Tissue A Comparison of Standard and Cooled-Tip Radiofrequency Energy André d Avila, MD; Christopher Houghtaling, BS, MS; Paulo Gutierrez, MD; Olivera Vragovic,

More information

Case Report Epicardial Ablation: Prevention of Phrenic Nerve Damage by Pericardial Injection of Saline and the Use of a Steerable Sheath

Case Report Epicardial Ablation: Prevention of Phrenic Nerve Damage by Pericardial Injection of Saline and the Use of a Steerable Sheath 87 Case Report Epicardial Ablation: Prevention of Phrenic Nerve Damage by Pericardial Injection of Saline and the Use of a Steerable Sheath Kars Neven, MD 1, Juan Fernández-Armenta, MD 2, David Andreu,

More information

Journal of the American College of Cardiology Vol. 33, No. 4, by the American College of Cardiology ISSN /99/$20.

Journal of the American College of Cardiology Vol. 33, No. 4, by the American College of Cardiology ISSN /99/$20. Journal of the American College of Cardiology Vol. 33, No. 4, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(98)00686-X The Creation

More information

Physiological Consequences of Bipolar Radiofrequency Energy on the Atria and Pulmonary Veins: A Chronic Animal Study

Physiological Consequences of Bipolar Radiofrequency Energy on the Atria and Pulmonary Veins: A Chronic Animal Study Physiological Consequences of Bipolar Radiofrequency Energy on the Atria and Pulmonary Veins: A Chronic Animal Study Sunil M. Prasad, MD, Hersh S. Maniar, MD, Michael D. Diodato, MD, Richard B. Schuessler,

More information

Effect of an increase in coronary perfusion on transmural. ventricular repolarization

Effect of an increase in coronary perfusion on transmural. ventricular repolarization Effect of an increase in coronary perfusion on transmural ventricular repolarization Yan-Zhou Zhang 1, MD, PhD, Ben He 1, MD, Le-Xin Wang 2, MD, PhD. From: 1 Department of Cardiology, Renji Hospital, Medical

More information

EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology

EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important

More information

Stand alone maze: when and how?

Stand alone maze: when and how? Stand alone maze: when and how? Dong Seop Jeong Department of Thoracic and Cardiovascular Surgery, HVSI Samsung Medical Center Type of atrial fibrillation First diagnose AF Paroxysmal AF: self-terminating

More information

Circulation: Arrhythmia and Electrophysiology CHALLENGE OF THE WEEK

Circulation: Arrhythmia and Electrophysiology CHALLENGE OF THE WEEK A 14-year-old girl with Wolff-Parkinson-White syndrome and recurrent paroxysmal palpitations due to atrioventricular reentry tachycardia had undergone two prior failed left lateral accessory pathway ablations

More information

Determination of Inadvertent Atrial Capture During Para-Hisian Pacing

Determination of Inadvertent Atrial Capture During Para-Hisian Pacing Determination of Inadvertent Atrial Capture During Para-Hisian Pacing Manoj Obeyesekere, MBBS; Peter Leong-Sit, MD; Allan Skanes, MD; Andrew Krahn, MD; Raymond Yee, MD; Lorne J. Gula, MD; Matthew Bennett,

More information

Atrial Fibrillation: Electrophysiological Mechanisms and the Results of Interventional Therapy

Atrial Fibrillation: Electrophysiological Mechanisms and the Results of Interventional Therapy Vol. 8, No. 3, September 2003 185 Atrial Fibrillation: Electrophysiological Mechanisms and the Results of Interventional Therapy A.SH. REVISHVILI Bakoulev Research Centre for Cardiovascular Surgery, Russian

More information

Case 1 Left Atrial Tachycardia

Case 1 Left Atrial Tachycardia Case 1 Left Atrial Tachycardia A 16 years old woman was referred to our institution because of recurrent episodes of palpitations and dizziness despite previous ablation procedure( 13 years ago) of postero-septal

More information

Atrial fibrillation (AF) is associated with increased morbidity

Atrial fibrillation (AF) is associated with increased morbidity Ablation of Atrial Fibrillation with Concomitant Surgery Edward G. Soltesz, MD, MPH, and A. Marc Gillinov, MD Atrial fibrillation (AF) is associated with increased morbidity and mortality in coronary artery

More information

II V 1 HRA 3 4 HB 5 6 HB 3 4 HB 1 2 CS 7 8 CS 5 6 CS 3 4 CS 1 2 ABL 3 4 ABL 1 2 RVA 3 4. T. Suga et al.

II V 1 HRA 3 4 HB 5 6 HB 3 4 HB 1 2 CS 7 8 CS 5 6 CS 3 4 CS 1 2 ABL 3 4 ABL 1 2 RVA 3 4. T. Suga et al. Table 1. The distance between the positions of the catheter tip Patient Location Before elimination of After elimination of Number of the the accessory pathway the accessory pathway accessory pathway RAO

More information

Mapping and Ablation of Challenging Outflow Tract VTs: Pulmonary Artery, LVOT, Epicardial

Mapping and Ablation of Challenging Outflow Tract VTs: Pulmonary Artery, LVOT, Epicardial Mapping and Ablation of Challenging Outflow Tract VTs: Pulmonary Artery, LVOT, Epicardial Samuel J. Asirvatham, MD Mayo Clinic Rochester California Heart Rhythm Symposium San Francisco, CA September 8,

More information

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie 268 Case Report Cardiac Resynchronization Therapy in a Patient with Persistent Left Superior Vena Cava Draining into the Coronary Sinus and Absent Innominate Vein: A Case Report and Review of Literature

More information

Medicine. Dynamic Changes of QRS Morphology of Premature Ventricular Contractions During Ablation in the Right Ventricular Outflow Tract

Medicine. Dynamic Changes of QRS Morphology of Premature Ventricular Contractions During Ablation in the Right Ventricular Outflow Tract Medicine CLINICAL CASE REPORT Dynamic Changes of QRS Morphology of Premature Ventricular Contractions During Ablation in the Right Ventricular Outflow Tract A Case Report Li Yue-Chun, MD, Lin Jia-Feng,

More information

Is cardioversion old hat? What is new in interventional treatment of AF symptoms?

Is cardioversion old hat? What is new in interventional treatment of AF symptoms? Is cardioversion old hat? What is new in interventional treatment of AF symptoms? Joseph de Bono Consultant Electrophysiologist University Hospitals Birmingham Atrial Fibrillation (AF) Affects 2% of the

More information

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics Hemodynamic Assessment Matt M. Umland, RDCS, FASE Aurora Medical Group Milwaukee, WI Assessment of Systolic Function Doppler Hemodynamics Stroke Volume Cardiac Output Cardiac Index Tei Index/Index of myocardial

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL Table S1. Patient Selection Codes, CIED Generator Procedures Code Type Code Description ICD9 Proc 00.51 Implantation of cardiac resynchronization defibrillator, total system [CRT-D]

More information

What s new in my specialty?

What s new in my specialty? What s new in my specialty? Jon Melman, MD Heart Rhythm Specialists McKay-Dee Hospital some would say some would say my specialty 1 some would say my specialty First pacemaker 1958 some would say my specialty

More information

Procurement Support to Ministry of Health, Ukraine. Invitation to Bid for:

Procurement Support to Ministry of Health, Ukraine. Invitation to Bid for: Procurement Support to Ministry of Health, Ukraine Invitation to Bid for: PROGRAMME XIX: MEDICAL PRODUCTS FOR HEALTHCARE INSTITUTIONS FOR TREATMENT OF PATIENTS WITH CARDIOVASCULAR AND CEREBROVASCULAR DISEASES

More information

Complex Left Atrial (CLA) Best Practices for Navigant

Complex Left Atrial (CLA) Best Practices for Navigant Complex Left Atrial (CLA) Best Practices for Navigant Ischemic Ventricular Tachycardia (VT) Best Practices for Navigant Ischemic VT Ablation Procedures Set Up In the absence of a septal defect or other

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

Ventricular Arrhythmias

Ventricular Arrhythmias Presenting your most challenging cases Venice Arrhythmias Ventricular Arrhythmias Gioia Turitto, MD Presenter Disclosure Information A questionable indication for CRT-D in a patient with VT after successful

More information

Chronic transmural atrial ablation by using bipolar radiofrequency energy on the beating heart

Chronic transmural atrial ablation by using bipolar radiofrequency energy on the beating heart Surgery for Acquired Cardiovascular Disease Prasad et al Chronic transmural atrial ablation by using bipolar radiofrequency energy on the beating heart Sunil M. Prasad, MD Hersh S. Maniar, MD Richard B.

More information

Intracardiac EchoCardiography (ICE) Common Views

Intracardiac EchoCardiography (ICE) Common Views Intracardiac EchoCardiography (ICE) Common Views Introduction What is ICE? Catheter with microscopic ultrasound transducer tip and doppler capabilities inserted into the heart via the IVC (typically) or

More information

Catheter Ablation of Atrial Fibrillation Strategy and Outcome Predictors Shih-Ann Chen MD

Catheter Ablation of Atrial Fibrillation Strategy and Outcome Predictors Shih-Ann Chen MD Catheter Ablation of Atrial Fibrillation Strategy and Outcome Predictors Shih-Ann Chen MD Taipei Veterans General Hospital, Taiwan Outline of AF Ablation 1. Strategy for Catheter Ablation of AF 2. Substrate

More information

VT Ablation in Structural Heart Disease Patient Information

VT Ablation in Structural Heart Disease Patient Information Melbourne Heart Rhythm VT Ablation in Structural Heart Disease Patient Information Ventricular Tachycardia in Structural Heart Disease (VT-SHD) Ventricular tachycardia (VT) is an abnormal rapid heart rhythm

More information

Raphael Rosso MD, Yuval Levi Med. Eng., Sami Viskin MD Tel Aviv Sourasky Medical Center

Raphael Rosso MD, Yuval Levi Med. Eng., Sami Viskin MD Tel Aviv Sourasky Medical Center Radiofrequency Ablation of Atrial Fibrillation: Comparison of Success Rate of Circular Ablation vs Point-by-Point Ablation with Contact Force Assessment in Paroxysmal and Persistent Atrial Fibrillation

More information

Case Report Successful Multi-chamber Catheter Ablation of Persistent Atrial Fibrillation in Cor Triatriatum Sinister

Case Report Successful Multi-chamber Catheter Ablation of Persistent Atrial Fibrillation in Cor Triatriatum Sinister www.ipej.org 50 Case Report Successful Multi-chamber Catheter Ablation of Persistent Atrial Fibrillation in Cor Triatriatum Sinister Andrew Gavin, MBChB 1, Cameron B Singleton, MD 1, Andrew D McGavigan,

More information

Pulmonary vein isolation (PVI) is

Pulmonary vein isolation (PVI) is Case Report Hellenic J Cardiol 2012; 53: 163-167 The Use of the Multi-Electrode Duty-Cycled Radiofrequency Ablation Catheter PVAC for the Ablation of a Left Atrial Tachycardia Dionyssios Leftheriotis 1,

More information

Computational Modeling to Study the Treatment of Cardiac Arrhythmias using Radiofrequency Ablation

Computational Modeling to Study the Treatment of Cardiac Arrhythmias using Radiofrequency Ablation Computational Modeling to Study the Treatment of Cardiac Arrhythmias using Radiofrequency Ablation A. González-Suárez* 1, M. Trujillo 1, J. Koruth 2, A. d'avila 2, and E. Berjano 1 1 Universitat Politècnica

More information

The major thoracic veins, with their specific electrical

The major thoracic veins, with their specific electrical Atrial Fibrillation Originating From Persistent Left Superior Vena Cava Li-Fern Hsu, MBBS; Pierre Jaïs, MD; David Keane, MD; J. Marcus Wharton, MD; Isabel Deisenhofer, MD; Mélèze Hocini, MD; Dipen C. Shah,

More information

Epicardial infrared ablation to create a linear conduction block on a beating right atrium

Epicardial infrared ablation to create a linear conduction block on a beating right atrium Kubota et al. Journal of Cardiothoracic Surgery (2018) 13:116 https://doi.org/10.1186/s13019-018-0801-y RESEARCH ARTICLE Epicardial infrared ablation to create a linear conduction block on a beating right

More information

Primary closed cooled tip ablation of typical atrial flutter in comparison to conventional radiofrequency ablation

Primary closed cooled tip ablation of typical atrial flutter in comparison to conventional radiofrequency ablation Europace (2002) 4, 265 271 doi:10.1053/eupc.2002.0240, available online at http://www.idealibrary.com on Primary closed cooled tip ablation of typical atrial flutter in comparison to conventional radiofrequency

More information

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION 1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,

More information

Catheter ablation of atrial fibrillation: Indications and tools for improvement of the success rate of the method. Konstantinos P.

Catheter ablation of atrial fibrillation: Indications and tools for improvement of the success rate of the method. Konstantinos P. Ioannina 2015 Catheter ablation of atrial fibrillation: Indications and tools for improvement of the success rate of the method Konstantinos P. Letsas, MD, FESC SECOND DEPARTMENT OF CARDIOLOGY LABORATORY

More information

Atrial Fibrillation Procedures Data Summary. Participant STS Period Ending 12/31/2016

Atrial Fibrillation Procedures Data Summary. Participant STS Period Ending 12/31/2016 Period Ending 12/31/2016 Number of Cases Preoperative Predominant Atrial Arrhythmia Type Paroxysmal Atrial Fibrillation... - - Persistent Atrial Fibrillation... - - Longstanding Persistent Atrial Fibrillation...

More information

Hybrid Surgical Ablation in South America: Lesson Learned. Joao R. Breda

Hybrid Surgical Ablation in South America: Lesson Learned. Joao R. Breda Hybrid Surgical Ablation in South America: Lesson Learned Joao R. Breda DISCLOSURES NONE 2 How to treat Atrial Fibrillation (AF) Understanding of pathophysiology mechanisms Ablation approach Choice of

More information

Μηνύματα από τις σπουδαιότερες μελέτες στην ηλεκτροφυσιολογία το διάστημα

Μηνύματα από τις σπουδαιότερες μελέτες στην ηλεκτροφυσιολογία το διάστημα Μηνύματα από τις σπουδαιότερες μελέτες στην ηλεκτροφυσιολογία το διάστημα 2015-2016 Δημήτριος M. Κωνσταντίνου Ειδικός Καρδιολόγος, MD, MSc, PhD, CCDS Πανεπιστημιακός Υπότροφος Dr. Konstantinou has received

More information

Ankara, Turkey 2 Department of Cardiology, Division of Arrhythmia and Electrophysiology, Yuksek Ihtisas

Ankara, Turkey 2 Department of Cardiology, Division of Arrhythmia and Electrophysiology, Yuksek Ihtisas 258 Case Report Electroanatomic Mapping-Guided Radiofrequency Ablation of Adenosine Sensitive Incessant Focal Atrial Tachycardia Originating from the Non-Coronary Aortic Cusp in a Child Serhat Koca, MD

More information

PEDIATRIC SVT MANAGEMENT

PEDIATRIC SVT MANAGEMENT PEDIATRIC SVT MANAGEMENT 1 INTRODUCTION Supraventricular tachycardia (SVT) can be defined as an abnormally rapid heart rhythm originating above the ventricles, often (but not always) with a narrow QRS

More information

Should hybrid ablation be the standard of care instead of transcatheter ablation techniques?

Should hybrid ablation be the standard of care instead of transcatheter ablation techniques? Should hybrid ablation be the standard of care instead of transcatheter ablation techniques? Christian Shults, MD Assistant Professor, Georgetown University School of Medicine Cardiac Surgeon, Medstar

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

The multi purse string maze procedure: A new surgical technique to perform the full maze procedure without atriotomies

The multi purse string maze procedure: A new surgical technique to perform the full maze procedure without atriotomies Ad Evolving Technology The multi purse string maze procedure: A new surgical technique to perform the full maze procedure without atriotomies Niv Ad, MD Objective: The maze procedure is the most effective

More information

ESSA HEART AND VASCULAR INSTITUTE APR/MAY/JUNE 2009 CLINICAL LETTER

ESSA HEART AND VASCULAR INSTITUTE APR/MAY/JUNE 2009 CLINICAL LETTER CLINICAL LETTER Exciting things are happening at the ESSA Heart and Vascular Institute and the Pocono Medical Center! We are all proud of the stellar team of professionals who are working very hard to

More information