Validation Study of Multi-Frequency Bioelectrical Impedance with Dual-Energy X-ray Absorptiometry Among Obese Patients

Size: px
Start display at page:

Download "Validation Study of Multi-Frequency Bioelectrical Impedance with Dual-Energy X-ray Absorptiometry Among Obese Patients"

Transcription

1 OBES SURG (2014) 24: DOI /s OTHER Validation Study of Multi-Frequency Bioelectrical Impedance with Dual-Energy X-ray Absorptiometry Among Obese Patients Silvia L. Faria & Orlando P. Faria & Mariane D. A. Cardeal & Marina Kiyomi Ito Published online: 26 January 2014 # Springer Science+Business Media New York 2014 Abstract Background Body mass index (BMI) is the most common parameter for classifying nutritional status. However, body composition (BC) may vary considerably among individuals with identical BMIs; consequently, we need to assess BC efficiently. Bariatric surgery is the most effective method for treating obesity. To improve quality assessment of postoperative weight loss, it is essential to assess BC. Multi-frequency bioelectrical impedance analysis (BIA) is a practical assessment instrument, though limited when applied among the obese population. Despite dual-energy X-ray absorptiometry (DXA) being the current reference standard, it has physical limitations which restrict its practical application. This study, therefore, sought to correlate the results of BC assessments of same patient population using BIA and DXA. Methods This was a cross-sectional validation study with patients invited to undergo a multi-frequency BIA (Inbody 720 ) and afterwards a DXA examination Statistical analyses were done using the intraclass correlation coefficient (ICC), paired t-test and the Bland Altman plot analysis. Results A total of 108 patients were randomly selected, with 73 meeting the criteria for study inclusion. Most were female (89 %) and had an average BMI of 40.17±4.08 kg/m 2.An almost perfect correlation of fat (kg) and fat-free mass (kg) was found in results from the BIA and DXA examination (ICC=0.832 and ICC=0.899, respectively). A substantial correlation was also found between the percentage of body fat (%BF) and the percentage of fat-free mass (%FFM). The comparison made between the BIA and DXA using the t-test showed significant differences between all parameters. The S. L. Faria: O. P. Faria : M. D. A. Cardeal Gastrocirurgia, Brasilia, Brazil S. L. Faria (*): M. D. A. Cardeal : M. K. Ito University of Brasília, Campus Universidade Darcy Ribeiro, Brasilia, Distrito Federal , Brazil silvialeitefaria@brturbo.com.br Bland Altman plot showed that the BIA method tends to underestimate the FM and overestimate the LM measurements when compared with DXA. Conclusion BIA proved to be a safe alternative for assessing BC in clinically severely obese patients and thus provides a more accessible evaluation tool for this population. But, consideration should be given to the formula added to the BIA measurement, adjusting the values to differences observed in order to reduce errors when compared with the DXA measurements. Keywords Dual-energy X-ray absorptiometry. Bioelectrical impedance analysis. Obesity. Bariatric surgery Introduction Obesity, defined as a body mass index (BMI) above 30 kg/m 2, increases the risk of death, whatever the actual cause may be [1]. In 2010, obesity among adult men exceeded 20 % in ten countries of the Western Hemisphere [2]. However, the definition of obesity must consider not only weight increase, but even more precisely any increase in body fat. The accumulation of the latter is associated with an increased risk of several diseases, such as type 2 diabetes and cardiovascular diseases [3]. Thus, there is a need for improvement in the clinical evaluation of patients with severe obesity (BMI 35 kg/m 2 with co-morbidities or BMI 40 kg/m 2, independent of comorbidities). Currently, BMI (weight/(height) 2 ) is the most common parameter used to classify the nutritional status of individuals. However, there may be considerable variation in body composition (BC) even between individuals with the same BMI [4]. In this sense, a serious need arises for an adequate assessment of BC of individuals, which can serve as a basis to better evaluate the effectiveness of various treatments of obesity [5]. At present, bariatric surgery is considered the best and most effective treatment for obesity, due to weight loss achieved

2 OBES SURG (2014) 24: and, principally, the long-term maintenance of this weight loss [6]. To improve the assessment of the quality of weight loss after this surgical procedure, it is essential that the assessment of BC be part of the clinical evaluation of these patients preoperatively and postoperatively, i.e., to assess whether the weight loss post-surgery was fat loss and/or loss of fat-free mass (FFM), with consequent malnutrition. Bioelectrical impedance analysis (BIA) is a rapid, practical, minimally invasive and relatively inexpensive instrument used to analyze BC of individuals based on the electrical conductivity of tissues. The method brings results in the form of body fat percentage (%BF), FFM percentage (%FFM) and distribution of body water. Studies have showed that BIA is a valid instrument for patients with a BMI up to 34 kg/m 2 [7]. The Multi-frequency Bioelectrical Impedance Analysis (Inbody 720, Biospace) uses an exclusive direct, targeted and multi-frequency system that is not based on statistical data of any specific population. It is capable of accurately assessing people with very different physical types, whether obese, elderly or athletic. Equations presented in recognized scientific literature were all developed using mono-frequency equipment and impedance measurement of the whole body. Thus, they are not comparable to results obtained from a segmental multi-frequency system. The trunk of the body provides about 50 % of conductivity, but only 10 % of the body impedance. As a result, studies show that there is a disproportion between body mass and body conductivity. This situation may decrease the accuracy in the results when using this method in obese patients. There is also a change in intra- and extracellular water in these patients, which may also interfere with the results [8]. Thus, validation studies are needed that measure BIA in patients with a BMI greater than 34 kg/m 2. Yet, despite the technological advances and the presence of different electric pulses of low intensity at different locations, which potentially increase the accuracy of the method, the formulas embedded in BIA machines do not offer specific considerations for the obese population. The current reference standard for analysis of BC is the dual-energy X-ray absorptiometry (DXA) examination. However, this examination has its own limitations related to the weight limit of the machine (in most cases, 120 kg) [5]and the cost involved. In view of the need to assess the BC of obese patients starting from the preoperative stage, and considering the cost and difficulty of performing DXA on all patients on a regular and frequent basis, this study sought to correlate the BC results of the same individuals using BIA and DXA. Material and Methods This was a cross-sectional validation study, where patients were first asked to undergo an examination of multi-frequency BIA (Inbody 720 ) to assess BC. Shortly after this examination, patients underwent a DXA examination (the gold standard for BC assessment) to evaluate the same BC data. Inclusion criteria for participation in this study were as follows: person having clinically severe obesity (BMI 35 kg/m 2 with co-morbidities or BMI 40 kg/m 2 ), participant in the preoperative treatment for bariatric surgery in the Gastrocirurgia Clinic of Brasilia (a private practice), weigh less than 120 kg (limitation of the DXA machine) and be between 18 and 65 years of age. Exclusion criteria were: being pregnant and/or with walking difficulties. Patients followed the following protocol for BIA assessment of BC: 8-h fast, as well as avoidance of strenuous physical activity, alcohol, along with food and drinks containing caffeine on the previous day. The intraclass correlation coefficient (ICC) combines random and systematic errors and, therefore, is preferable to the linear correlation coefficient of Pearson. Although confounding results between linear association and agreement does not merit concern, the main problem related to the correlation coefficients resides in the dependence on the amplitude of the measurement. In other words, the more heterogeneous the set of individuals (greater variability intra-individuals) studied is, the greater the value of the ICC. This indicator is a measure of corrected agreement due to chance and ranges from 1 and +1 (plus one). The maximum value indicates perfect reproducibility between measurements. The joint roughness coefficient (JRC) is equivalent to the kappa statistic for continuous variables and their interpretation can be given according to Landis and Koch [9]. The average of the differences between the two values of the pairs under observation is a measure of the degree of systematic differences between the two sets of readings. A paired t-test statistical analysis was performed to calculate the statistical significance of the calculated difference. The Bland Altman plot was used to assess the magnitude of disagreement (including systematic differences), to spot outliers, and to observe any trends. Results Out of a possible total of 104 patients, 73 met criteria for inclusion in this study and agreed to participate. Of these, the majority was female (89 %). The average age of the group was 37.31±10.49 years and their BMI was 40.17±4.08 kg/m 2. Mean values of %BF and %FFM, fat mass (FM) and FFM in kg, using the two methods, compared through paired t-test, are shown in Table 1. From the test results, we have: 1. The FM value (kg) obtained by the BIA method is significantly inferior (2.05 kg), on average, to the FM (kg) value obtained by the DXA method.

3 1478 OBES SURG (2014) 24: Table 1 Results of weight, height, % body fat, % lean mass, fat mass and lean mass in kg using two methods multi-frequency bioimpedance and DXA Variable DXA BIA Average (±SD) Average (±SD) Difference (BIA-DXA) Paired t value p value Weight (kg) (±10.83) (±11.45) Height (m) 1.63 (±0.08) 1.62 (±0.07) BMI (kg/m 2 ) (±4.09) (±3.97) FM (kg) (±7.46) (±7.37) < %BF (±4.80) (±4.04) < Lean mass (kg) (±7.55) (±7.05) Lean mass (%) (±4.80) (±4.04) < %BF percentage of body fat, BMI body mass index, DXA dual-energy X-ray absorptiometry, BIA multi-frequency bioimpedance, FM fat mass 2. The %BF value obtained by the BIA method is significantly inferior (1.61 %), on average, to the %BF obtained by the DXA method. 3. The FFM value (kg), obtained by the BIA method, is significantly superior (1.28 kg), on average, to the FFM value (kg) obtained by the DXA method. 4. The FFM value (%), obtained by the BIA method, is significantly superior (1.61 %), on average, to the FFM value (%) obtained by the DXA method. Applying the Bland Altman plot analysis, a statistically significant bias can be seen in Figs. 1 and 2 (p<0.0001), showing that the BIA method tends to underestimate the FM measurements (kg and %BF) when compared with DXA. A relation is seen between the bias (difference between the measurements) and the mean of the measurements. The bias appears to be greater for mean values larger than 55 kg and 48 %. For mean values smaller or equal to 55 kg and 48 %, the FM (kg and %BF) calculated from the BIA, adding a bias of 1.29 kg and 0.61 %, could be used (excluding the outlier point with bias close to 20 kg and 10 %) as an estimate of FM (kg and %BF) obtained by DXA. For mean values greater than 55 kg and 48 %, the FM (kg and %BF) obtained from BIA, adding a bias of 2.40 kg and 1.71 %, as an estimate of the FM (kg and %BF) obtained by DXA. Three points outside the limit of agreement can be seen, one of which showed a bias difference of nearly 20 kg and 10 % (Figs. 1 and 2). A statistically significant bias is seen in the graph, showing that the BIA method tends to overestimate the FFM measurements (kg and %FFM) when compared with DXA (p< and p<0.0001). The spatial distribution of points tends to be homogeneous, in the sense of not indicating relation between the bias (difference between measurements) and the mean of the measurements, leading to the conclusion that the bias is systematic. Therefore, despite the measurements not being consistent, FFM (kg) calculated by BIA, removing a bias of 1.28kgand1.61%,couldbeusedasanestimateofthe FFM (kg and %FFM) obtained by DXA. Three points outside the limit of agreement to FFM in kg can be seen (Fig. 3). But, for the %FFM, there are six points outside of the limit of agreement, one of which showed a bias difference of nearly 10 % (Fig. 4). Fig. 1 Bland-Altman plot analysis showing the Fat Mass (FM) difference using DXA and BIA Fig. 2 Bland-Altman plot analysis showing the %Fat Mass (%FM)

4 OBES SURG (2014) 24: Table 2 Correlation between parameters of body composition measuredbydxaandbia %BF percentage of body fat, BMIbody mass index a Intra-class coefficient (ICC), with a confidence interval of 95 % (in parentheses) ICC a Weight ( ) Height ( BMI ( ) Lean mass (kg) (0, ) Lean mass (%) ( ) Fat mass (kg) (0, ) %BF ( ) Fig. 3 Bland-Altman plot analysis showing Lean Mass (LM) in Kg With the application of the two methods, the ICC showed an almost perfect correlation between fat and FFM in kg. In relation to %BF and the amount of fat in kg, a substantial correlation was found (Table 2). Discussion The importance of assessing BC in obese patients can be seen in the evaluation of the actual effectiveness of existing treatments for obesity in favor of quality weight loss, principally of fat. But there are a number of factors that complicate the measurement of BC in these patients who often have significant changes in their body compartments when compared with eutrophic individuals. One major obstacle is the increase in total body water, especially extracellular water in relation to intracellular water [5], which impedes an accurate BC measurement when using various methods. All this makes for scant data on BC and reduced information on the methods Fig. 4 Bland-Altman plot analysis showing % Lean Mass (%LM) that can safely be used for the measurement of this variable among the obese population [2]. The DXA examination is the most valid technique currently used to measure BC in eutrophic patients who are overweight or moderately obese. Despite its detailed and accurate assessment, this method is expensive and cannot be repeated often due to the level of radiation emitted during the examination. Besides this, obese patients often exceed the capacity of the DXA unit [10]. For this reason, an alternative is to use multi-frequency BIA, a non-invasive and relatively inexpensive method that has been used for measuring total body water, FFM and extracellular water. However, studies show that in cases of clinically severe obesity, most prediction equations used with Bioelectral Impedance Analysis are unable to ascertain the static BC and over time are not reproducible for individuals [11]. The disproportion between body mass and conductivity of the body of obese patients seems to reduce the accuracy of BIA [7]. The method of bioelectrical impedance is based on the ability of hydrated tissues to conduct electricity. The measurement of total body impedance allows for the assessment of total body water, assuming a constant state of hydration. From the total body water (in a healthy adult, 60 % of body weight is attributable to water), assuming that water makes up 73 % of muscles and using validated equations and reference values, it is possible to estimate FFM and, by calculating the difference, body fat. With these assumptions, questions are being raised about using the method with the obese population, considered to be hyper-hydrated [4]. However, in this study, BIA presented highly reliable results in the measurement of BC among clinically severe obese patients when compared with the reference method DXA. Strain et al. [12], comparing a single frequency BIA examination (Tanita TBF-310) with the deuterium dilution method, observed a reasonable correlation between the methods (ICC=0.91), and affirmed that the use of bioelectrical impedance is safe for clinically severe obese patients. Donini et al. [4] studied over 100 overweight patients using a model of a single frequency bioelectrical impedance, found a reasonable correlation between body fat measured by DXA (r=0.91) and body fat measured by bioelectrical impedance (r=0.86). The

5 1480 OBES SURG (2014) 24: study of Donini et al. [4], however, was done with patients who had a BMI 25 and 35 kg/m 2. For this reason, they may have presented better results than those found in the present study. The increase in BMI and in fat distribution tend to decrease the accuracy of the BIA test [5]. The apparatus used in these previous studies was not the multi-frequency modality used in the present study, yet a reasonable correlation was presented. A multi-frequency bioelectrical impedance examination also uses empirical linear regression models. However, unlike the single-frequency bioelectrical impedance, it includes impedances at multiple frequencies (0, 1, 5, 50, 100, 200 to 500 khz) in order to evaluate FFM, the total body water, intracellular water and extracellular water, which can further improve the accuracy of the results [13]. The basic principle of DXA is that the attenuation of X-rays, using high and low energy photons, is measurable and depends on the thickness, density and chemical composition of underlying tissue. Thus, the attenuation of X-ray energies through the fat, bones and muscles varies due to differences in density and chemical composition within these tissues. In this way, the moisture level of the individual can change the result. However, studies show that changes in body hydration of up to 1 kg do not seem to have significant affect on the accuracy of DXA assessment [14]. Lohman et al. [15], after extensive review of the literature, theorized that a 5 % change in water present in FFM could affect the measurement of the % of body fat (%BF) in only 2 % of cases considered [15]. The limitations of this present study relate to external validity of results obtained. This study was undertaken mainly with female patients weighing up to 120 kg. Thus, for individuals with an altered hydration state, as possibly in the case of the obese individuals studied here, DXA appears, in fact, to be a reference for BC measurement, with which the multi-frequency bioelectrical impedance (Inbody 720 ) presented almost perfect correlation, proving itself to be a good method for assessing BC in clinically severe obese patients. Conclusion The multi-frequency BIA proved to be a safe alternative to assess BC in clinically severe obese patients who weigh up to 120 kg, thus providing easier access to this important measurement tool for this population, considering the facilities of use. Consideration should be given to the formula added to the BIA measurement, adjusting the values to differences observed in order to reduce errors when compared with the DXA measurements. Acknowledgments This study received financial support from Gastrocirurgia, Brasília, Brazil. SEPS 710/910, Centro Clínico Via Brasil, Salas 314, 334, 336, 337, 348, Brasília, Brazil. Conflict of interest interest. References All the authors declare that they have no conflict of 1. Wandell PE, Carlsson AC, Theobald H. The association between BMI value and long-term mortality. Int J Obes (Lond). 2009;33(5): World Health Organization (WHO). Accessed July Must A, Spadano J, Coakley EH, et al. The disease burden associated with overweight and obesity. Am J Med Assoc. 1999;282: Donini LM, Poggiogalle E, Del Balzo V, et al. How to estimate fat mass in overweight and obese subjects. Int J Endocrinol. 2013;2013: Das SK. Body composition measurement in severe obesity. Curr Opin Clin Nutr Metab Care. 2005;8(6): Valezi AC, Junior JM, de Menezes MA. Weight loss outcome after silastic ring Roux-en-Y gastric bypass: 8 years of follow-up. Obes Surg. 2010;20: Kyle UG, Bosaeus I, De Lorenzo AD, et al. Bioelectrical impedance analysis Part II. Utilization in clinical practice. Clin Nutr. 2004;23(6): Bracco D, Thiebaud D, Chiolero R, et al. Segmental body composition assessed by bioelectrical impedance analysis and DEXA in humans. J Appl Physiol. 1996;81: Szklo M, Nieto, JF. Epidemiology beyond the basics. 2000; Kehayias JJ, Heymsfield SB, Dilmanian FA, et al. Measurement of body fat by neutron inelastic scattering: comments on installation, operation and error analysis. Basic Life Sci. 1990;55: Hendel HW, Gotfredsen A, Hojgaard L, et al. Change in fat-free mass assessed by bioelectrical impedance, total body potassium and dual energy X-ray absorptiometry during prolonged weight loss. Scand J Clin Lab Invest. 1996;56(8): Strain GW, Wang J, Gagner M, et al. bioelectral impedance for severe obesity: comparing research methods for total body water and resting energy expenditure. Obesity (Silver Spring). 2008;16(8): Heyward VH, Wagner DR. Applied body composition assessment. Champaign: Human Kinetics Pub Incorporated; p. 14. Pietrobelli A, Wang Z, Formica C, et al. Dual-energy X-ray absorptiometry: fat estimation errors due to variation in soft tissue hydratation. Am J Physiol. 1998;274: Tg L. Harris M, Teixeira P J, et al. Assessing body composition and changes in body composition: another look at dual-energy X-ray absorptiometry. Ann N Y Acad Sci. 2000;904:45 54.

Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes

Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes Susumu Sato 1), Shinichi Demura 2), Tamotsu Kitabayashi

More information

Understanding Body Composition

Understanding Body Composition PowerPoint Lecture Outlines 7 Understanding Body Composition Objectives Define body composition. Explain why the assessment of body size, shape, and composition is useful. Explain how to perform assessments

More information

ESPEN Congress Prague 2007

ESPEN Congress Prague 2007 ESPEN Congress Prague 2007 Nutrition implication of obesity and Type II Diabetes Nutrition support in obese patient Claude Pichard Nutrition Support in Obese Patients Prague, 2007 C. Pichard, MD, PhD,

More information

Bioelectrical Impedance versus Body Mass Index for Predicting Body Composition Parameters in Sedentary Job Women

Bioelectrical Impedance versus Body Mass Index for Predicting Body Composition Parameters in Sedentary Job Women ORIGINAL ARTICLE Bioelectrical Impedance versus Body Mass Index for Predicting Body Composition Parameters in Sedentary Job Women Mohammad Javad Shekari-Ardekani 1, Mohammad Afkhami-Ardekani 2*, Mehrshad

More information

ESPEN Congress The Hague 2017

ESPEN Congress The Hague 2017 ESPEN Congress The Hague 2017 Paediatric specificities of nutritional assessment Body composition measurement in children N. Mehta (US) 39 th ESPEN Congress The Hague, Netherlands Body Composition Measurement

More information

The Assessment of Body Composition in Health and Disease

The Assessment of Body Composition in Health and Disease The Assessment of Body Composition in Health and Disease Giorgio Bedogni, Paolo Brambilla, Stefano Bellentani and Claudio Tiribelli CHAPTER 3 BODY AND NUTRITIONAL STATUS Nutritional status can be operationally

More information

Body Composition Analysis by Air Displacement Plethysmography in Normal Weight to Extremely Obese Adults

Body Composition Analysis by Air Displacement Plethysmography in Normal Weight to Extremely Obese Adults Body Composition Analysis by Air Displacement Plethysmography in Normal Weight to Extremely Obese Adults Kazanna C. Hames 1,2, Steven J. Anthony 2, John C. Thornton 3, Dympna Gallagher 3 and Bret H. Goodpaster

More information

Validation of bioimpedance body composition assessment by TANITA BC-418 in 7 years-

Validation of bioimpedance body composition assessment by TANITA BC-418 in 7 years- Validation of bioimpedance body composition assessment by TANITA BC-418 in 7 years- old children vs. Dual X-ray absorptiometry Veronica Luque on behalf to URV Team 18th November 2012 Milano, 10th Bi-annual

More information

Chapter 17: Body Composition Status and Assessment

Chapter 17: Body Composition Status and Assessment Chapter 17: Body Composition Status and Assessment American College of Sports Medicine. (2010). ACSM's resource manual for guidelines for exercise testing and prescription (6th ed.). New York: Lippincott,

More information

Body composition A tool for nutritional assessment

Body composition A tool for nutritional assessment Body composition A tool for nutritional assessment Ingvar Bosaeus Clinical Nutrition Unit Sahlgrenska University Hospital NSKE Oslo 2012-01-18 Outline What is body composition? What is nutritional assessment?

More information

COMPARISON OF BODY COMPOSITION ASSESSMENT IN WOMEN USING SKINFOLD THICKNESS EQUATIONS, BIOELECTRICAL IMPEDANCE ANALYSIS AND UNDERWATER WEIGHING

COMPARISON OF BODY COMPOSITION ASSESSMENT IN WOMEN USING SKINFOLD THICKNESS EQUATIONS, BIOELECTRICAL IMPEDANCE ANALYSIS AND UNDERWATER WEIGHING STUDIES IN PHYSICAL CULTURE AND TOURISM Vol. 17, No. 3, 2010 EFTEKHAR MOHAMMADI, SAEID SHAKERIAN Faculty of Physical Education & Sports Science, Shahid Chamran University, Iran COMPARISON OF BODY COMPOSITION

More information

BODY MASS INDEX AND BODY FAT CONTENT IN ELITE ATHLETES. Abstract. Introduction. Volume 3, No. 2, 2011, UDC :572.

BODY MASS INDEX AND BODY FAT CONTENT IN ELITE ATHLETES. Abstract. Introduction. Volume 3, No. 2, 2011, UDC :572. EXERCISE AND QUALITY OF LIFE Volume 3, No. 2, 2011, 43-48 UDC 796.034.6-051:572.087 Research article BODY MASS INDEX AND BODY FAT CONTENT IN ELITE ATHLETES Jelena Popadiã Gaãeša *, Otto Barak, Dea Karaba

More information

The effect of a meal on measures of impedance and percent body fat estimated using contact-electrode bioelectrical impedance technology

The effect of a meal on measures of impedance and percent body fat estimated using contact-electrode bioelectrical impedance technology European Journal of Clinical Nutrition (2013) 67, 950 955 & 2013 Macmillan Publishers Limited All rights reserved 0954-3007/13 www.nature.com/ejcn ORIGINAL ARTICLE The effect of a meal on measures of impedance

More information

In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in

In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in 1 2 In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in adults: the influence of body mass index 3 4 Running title: Precision of the idxa for the measurement of visceral

More information

Adult BMI Calculator

Adult BMI Calculator For more information go to Center for Disease Control http://search.cdc.gov/search?query=bmi+adult&utf8=%e2%9c%93&affiliate=cdc-main\ About BMI for Adults Adult BMI Calculator On this page: What is BMI?

More information

bioelectrical impedance measurements. Aviation Space and Environmental Medicine, 59,

bioelectrical impedance measurements. Aviation Space and Environmental Medicine, 59, MS ID: 9761874491197639 Authors: Macias et al Title: Body fat measurement by bioelectrical impedance versus air displacement plethysmography: a validation to design bioelectrical impedance equations in

More information

Development of Bio-impedance Analyzer (BIA) for Body Fat Calculation

Development of Bio-impedance Analyzer (BIA) for Body Fat Calculation IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Development of Bio-impedance Analyzer (BIA) for Body Fat Calculation Recent citations - Munawar A Riyadi et al To cite this article:

More information

COMPARISON OF AIR DISPLACEMENT PLETHYSMOGRAPHY TO HYDROSTATIC WEIGHING FOR ESTIMATING TOTAL BODY DENSITY IN CHILDREN

COMPARISON OF AIR DISPLACEMENT PLETHYSMOGRAPHY TO HYDROSTATIC WEIGHING FOR ESTIMATING TOTAL BODY DENSITY IN CHILDREN COMPARISON OF AIR DISPLACEMENT PLETHYSMOGRAPHY TO HYDROSTATIC WEIGHING FOR ESTIMATING TOTAL BODY DENSITY IN CHILDREN Geo Claros 1, Holly R Hull 1, and David A Fields 1,2 * 1 Department of Health and Exercise

More information

Body Composition. Sport Books Publisher 1

Body Composition. Sport Books Publisher 1 Body Composition Sport Books Publisher 1 The body composition The body composition is affected by the proportions of the body component (bones, muscles, and other tissues) It can be seen that the major

More information

Health Care & Human Care

Health Care & Human Care Health Care & Human Care The most ideal and convenient system for health care The revolutionary technology in BIA has created a new standard Leading novel technology provides the accurate results you can

More information

Bedside methods versus dual energy X-ray absorptiometry for body composition measurement in COPD

Bedside methods versus dual energy X-ray absorptiometry for body composition measurement in COPD Eur Respir J ; 19: 66 631 DOI: 1.1183/931936..796 Printed in UK all rights reserved Copyright #ERS Journals Ltd European Respiratory Journal ISSN 93-1936 Bedside methods versus dual energy X-ray absorptiometry

More information

Alexandra M. Johnstone a, Peter Faber b, Eileen R. Gibney c, Gerald E. Lobley a, R. James Stubbs a, Mario Siervo d, ORIGINAL ARTICLE

Alexandra M. Johnstone a, Peter Faber b, Eileen R. Gibney c, Gerald E. Lobley a, R. James Stubbs a, Mario Siervo d, ORIGINAL ARTICLE Obesity Research & Clinical Practice (2014) 8, e46 e54 ORIGINAL ARTICLE Measurement of body composition changes during weight loss in obese men using multifrequency bioelectrical impedance analysis and

More information

Energy Adaptations Persist 2 Years After Sleeve Gastrectomy and Gastric Bypass

Energy Adaptations Persist 2 Years After Sleeve Gastrectomy and Gastric Bypass OBES SURG (2016) 26:459 463 DOI 10.1007/s11695-015-1972-4 BRIEF COMMUNICATION Energy Adaptations Persist 2 Years After Sleeve Gastrectomy and Gastric Bypass Charmaine S. Tam 1 & Georgia Rigas 2 & Leonie

More information

Assessment of body composition of Sri Lankan Australian children using ethnic specific equations

Assessment of body composition of Sri Lankan Australian children using ethnic specific equations J.Natn.Sci.Foundation Sri Lanka 2015 43 (2): 111-118 DOI: http://dx.doi.org/10.4038/jnsfsr.v43i2.7938 RESEARCH ARTICLE Assessment of body composition of Sri Lankan Australian children using ethnic specific

More information

Luís B Sardinha, Timothy G Lohman, Pedro J Teixeira, Dartagnan P Guedes, and Scott B Going

Luís B Sardinha, Timothy G Lohman, Pedro J Teixeira, Dartagnan P Guedes, and Scott B Going Comparison of air displacement plethysmography with dual-energy X-ray absorptiometry and 3 field methods for estimating body composition in middle-aged men 1,2 Luís B Sardinha, Timothy G Lohman, Pedro

More information

Bioimpedance in medicine: Measuring hydration influence

Bioimpedance in medicine: Measuring hydration influence Journal of Physics: Conference Series Bioimpedance in medicine: Measuring hydration influence To cite this article: J Hlubik et al 0 J. Phys.: Conf. Ser. 0 View the article online for updates and enhancements.

More information

Premium solution for your health

Premium solution for your health 770 Premium solution for your health See What You re Made of Reveal the efficiency of your consultation throughout the InBody Test The InBody Test clearly visualize internal change of the body. Weight

More information

PROFESSIONAL FITNESS & WELLBEING CATALOGUE

PROFESSIONAL FITNESS & WELLBEING CATALOGUE PROFESSIONAL FITNESS & WELLBEING CATALOGUE CONTENTS BENEFITS OF BODY COMPOSITION GOOD, BETTER, BEST TANITA BIA TECHNOLOGY MULTI FREQUENCY BODY COMPOSITION ANALYSER BODY COMPOSITION ANALYSERS WEIGHING SCALES

More information

Whole Body Dual X-Ray Absorptiometry to Determine Body Composition

Whole Body Dual X-Ray Absorptiometry to Determine Body Composition Page: 1 of 6 Last Review Status/Date: March 2015 Determine Body Composition Description Using low dose x-rays of two different energy levels, whole body dual x-ray absorptiometry (DXA) measures lean tissue

More information

Assessing nutritional status after spinal cord injury

Assessing nutritional status after spinal cord injury Assessing nutritional status after spinal cord injury Prof Mary Galea Date: 30 October 2015 Research report #: 078-1015-R01 Further information For further information on this report, please email info@iscrr.com.au

More information

Validity of the Body Mass Index for Estimating Body Composition for Young Adults with. Intellectual Disabilities. Mary Ware

Validity of the Body Mass Index for Estimating Body Composition for Young Adults with. Intellectual Disabilities. Mary Ware Running Head: VALIDITY OF A NEW TEST ITEM 1 Validity of the Body Mass Index for Estimating Body Composition for Young Adults with Intellectual Disabilities Mary Ware Western Michigan University Mary Ware

More information

For Convenient Use

For Convenient Use www.inbody.com For Convenient Use www.inbody.com The most ideal and convenient system for health care The revolutionary technology in BIA has created a new standard Leading novel technology provides the

More information

Simopoulos AP (ed): Nutrition and Fitness: Obesity, the Metabolic Syndrome, Cardiovascular Disease and Cancer. Basel, Karger, 2005, vol 94, pp 60 67

Simopoulos AP (ed): Nutrition and Fitness: Obesity, the Metabolic Syndrome, Cardiovascular Disease and Cancer. Basel, Karger, 2005, vol 94, pp 60 67 WRN94060.qxd 4/11/05 4:57 PM Page 60 Simopoulos AP (ed): Nutrition and Fitness: Obesity, the Metabolic Syndrome, Cardiovascular Disease and Cancer. Basel, Karger, 2005, vol 94, pp 60 67 Physical Activity

More information

International Journal of Gerontology

International Journal of Gerontology International Journal of Gerontology 6 (2012) 273e277 Contents lists available at SciVerse ScienceDirect International Journal of Gerontology journal homepage: www.ijge-online.com Original Article Bioelectrical

More information

Victor Babes University of Medicine and Pharmacy, Timisoara, Romania b

Victor Babes University of Medicine and Pharmacy, Timisoara, Romania b Mædica - a Journal of Clinical Medicine MAEDICA a Journal of Clinical Medicine 2015; 10(3): 204-209 ORIGINAL PAPERS Reliability of Measuring Subcutaneous Fat Tissue Thickness Using Ultrasound in Non-Athletic

More information

Comparison of Bioelectrical Impedance Analysis with Dual Energy X-ray Absorptiometry in Obese Women. Abstract

Comparison of Bioelectrical Impedance Analysis with Dual Energy X-ray Absorptiometry in Obese Women. Abstract Comparison of Bioelectrical Impedance Analysis with Dual Energy X-ray Absorptiometry in Obese Women Yound Gyu Cho, Jae Heon Kand, Kyung A Kim Department of Family Medicine, Seoul Paik Hospital, College

More information

FAT MASS ESTIMATION BY BIOELECTRICAL IMPEDANCE ANALYSIS

FAT MASS ESTIMATION BY BIOELECTRICAL IMPEDANCE ANALYSIS FAT MASS ESTIMATION BY BIOELECTRICAL IMPEDANCE ANALYSIS FAT MASS ESTIMATION BY BIOELECTRICAL IMPEDANCE ANALYSIS Epidemiological studies have shown that relative risk of diabetes and cardiovascular diseases

More information

Title:Body adiposity index performance in estimating body fat in a sample of severely obese Brazilian patients

Title:Body adiposity index performance in estimating body fat in a sample of severely obese Brazilian patients Author's response to reviews Title:Body adiposity index performance in estimating body fat in a sample of severely obese Brazilian patients Authors: Giliane Belarmino (giliane85@hotmail.com) Lilian M Horie

More information

A Comparison of Multiple Frequency versus Single Frequency Bioelectrical Impedance Techniques for the Assessment of Body Composition

A Comparison of Multiple Frequency versus Single Frequency Bioelectrical Impedance Techniques for the Assessment of Body Composition University of Miami Scholarly Repository Open Access Theses Electronic Theses and Dissertations 2014-07-08 A Comparison of Multiple Frequency versus Single Frequency Bioelectrical Impedance Techniques

More information

Fitness Nutrition Coach. Part IV - Assessing Nutritional Needs

Fitness Nutrition Coach. Part IV - Assessing Nutritional Needs Part IV - Assessing Nutritional Needs 62 For the FNC, a nutrition assessment provides information on your client s diet quality and awareness about nutrient information. Clients should be encouraged to

More information

Spinal Cord Rehab Program, Toronto Rehabilitation Institute, University Health Network, 520 Sutherland Drive, Toronto, ON, Canada M4G 3V9 2

Spinal Cord Rehab Program, Toronto Rehabilitation Institute, University Health Network, 520 Sutherland Drive, Toronto, ON, Canada M4G 3V9 2 Journal of Obesity Volume 212, Article ID 936, 9 pages doi:1.11/212/936 Clinical Study Bioelectrical Impedance and Dual-Energy X-Ray Absorptiometry Assessments of Changes in Body Composition Following

More information

There are 4 user settings that store the memory of personal parameters for ease of continued use.

There are 4 user settings that store the memory of personal parameters for ease of continued use. INTRODUCTION Thank you for choosing the Healthy Balance Electronic Body Fat Scale. For proper usage of this device please read the following instructions carefully. Using the latest BIA (Bioelectrical

More information

Lighten-up your Child s Growth

Lighten-up your Child s Growth Specifications Key Specifications Bioelectrical Bioelectrical Impedance Analysis (BIA) Impedance (Z) Measurement Items 15 Impedance Measurements by Using 3 Different Frequencies (5kHz, 50kHz, 250kHz) at

More information

As of the wrestling season, the National

As of the wrestling season, the National Evaluation of Air Displacement for Assessing Body Composition of Collegiate Wrestlers ALAN C. UTTER 1, FREDRIC L. GOSS 2, PAMELA D. SWAN 3, GREGORY S. HARRIS 1, ROBERT J. ROBERTSON 2, and GREGORY A. TRONE

More information

Applied Physiology, Nutrition, and Metabolism

Applied Physiology, Nutrition, and Metabolism Applied Physiology, Nutrition, and Metabolism Predicted Versus Measured Thoracic Gas Volumes of Collegiate Athletes Made by the BOD POD Air Displacement Plethysmography System Journal: Applied Physiology,

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Birth Date: 40.2 years Height / Weight: 158.0 cm 52.0 kg Sex / Ethnic: Female Patient ID: Total Body Tissue Quantitation Composition Reference: Total Tissue 50% 40% 30% 20% 20 30 40 50 60 70 80 90 100

More information

Bioelectrical impedance analysis to assess body composition in obese adult women: The effect of ethnicity

Bioelectrical impedance analysis to assess body composition in obese adult women: The effect of ethnicity International Journal of Obesity (1998) 22, 243±249 ß 1998 Stockton Press All rights reserved 0307±0565/98 $12.00 Bioelectrical impedance analysis to assess body composition in obese adult women: The effect

More information

EXTRACELLULAR WATER REFERENCE VALUES. Extracellular Water: Reference values for Adults

EXTRACELLULAR WATER REFERENCE VALUES. Extracellular Water: Reference values for Adults CHAPTER 5 Extracellular Water: Reference values for Adults Analiza M. Silva, Jack Wang, Richard N. Pierson Jr., ZiMian Wang, David B. Allison Steven B. Heymsfield, Luis B. Sardinha, Stanley Heshka ABSTRACT

More information

DIALYSIS OUTCOMES Quality Initiative

DIALYSIS OUTCOMES Quality Initiative Bipedal Bioelectrical Impedance Analysis Reproducibly Estimates Total Body Water in Hemodialysis Patients Robert F. Kushner, MD, and David M. Roxe, MD Formal kinetic modeling for hemodialysis patients

More information

Accuracy of skinfold and bioelectrical impedance assessments of body fat percentage in ambulatory individuals with cerebral palsy

Accuracy of skinfold and bioelectrical impedance assessments of body fat percentage in ambulatory individuals with cerebral palsy DEVELOPMENTAL MEDICINE & CHILD NEUROLOGY ORIGINAL ARTICLE Accuracy of skinfold and bioelectrical impedance assessments of body fat percentage in ambulatory individuals with cerebral palsy DONNA J OEFFINGER

More information

Body fat prediction equations for skinfold and bioelectrical impedance analysis using dual-energy x-ray absorptiometry data as the criterion.

Body fat prediction equations for skinfold and bioelectrical impedance analysis using dual-energy x-ray absorptiometry data as the criterion. Research Article http://www.alliedacademies.org/journal-physical-therapy-sports-medicine/ Body fat prediction equations for skinfold and bioelectrical impedance analysis using dual-energy x-ray absorptiometry

More information

The Bone Wellness Centre - Specialists in Dexa Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in Dexa Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Birth Date: 24.7 years Height / Weight: 8.0 cm 79.0 kg Sex / Ethnic: Male Patient ID: Total Body Tissue Quantitation Composition Reference: Total Tissue 40% 30% 20% 0% 20 30 40 50 60 70 80 90 00 Centile

More information

Evaluation of DXA against the four-component model of body composition in obese children and adolescents aged 5 to 21 years

Evaluation of DXA against the four-component model of body composition in obese children and adolescents aged 5 to 21 years Europe PMC Funders Group Author Manuscript Published in final edited form as: Int J Obes (Lond). 2010 April ; 34(4): 649 655. doi:10.1038/ijo.2009.249. Evaluation of DXA against the four-component model

More information

BODY COMPOSITION COMPARISON: BIOELECTRIC IMPEDANCE ANALYSIS WITH DXA IN ADULT ATHLETES

BODY COMPOSITION COMPARISON: BIOELECTRIC IMPEDANCE ANALYSIS WITH DXA IN ADULT ATHLETES BODY COMPOSITION COMPARISON: BIOELECTRIC IMPEDANCE ANALYSIS WITH DXA IN ADULT ATHLETES A Thesis Presented to the Faculty of the Graduate School University of Missouri In Partial Fulfillment of the Requirements

More information

C H A P T E R 14 BODY WEIGHT, BODY COMPOSITION, AND SPORT

C H A P T E R 14 BODY WEIGHT, BODY COMPOSITION, AND SPORT C H A P T E R 14 BODY WEIGHT, BODY COMPOSITION, AND SPORT Learning Objectives Differentiate among body build, body size, and body composition. Find out what tissues of the body constitute fat-free mass.

More information

Note that metric units are used in the calculation of BMI. The following imperial-metric conversions are required:

Note that metric units are used in the calculation of BMI. The following imperial-metric conversions are required: Body Composition Body Composition: Assessment and Interpretation Body composition has great practical and functional significance for many of us: scientists, clinicians and the general population. It can

More information

TRANSFORM PARTNER CHALLENGE RULES, LIABILITY & PUBLICITY RELEASE

TRANSFORM PARTNER CHALLENGE RULES, LIABILITY & PUBLICITY RELEASE TRANSFORM PARTNER CHALLENGE RULES, LIABILITY & PUBLICITY RELEASE CHALLENGE INFO: Participants will enjoy 8 weeks full access to the gym and unlimited Elite Group Training, IQFit and QuickFit classes, as

More information

LOSE TO WIN CHALLENGE RULES, LIABILITY & PUBLICITY RELEASE

LOSE TO WIN CHALLENGE RULES, LIABILITY & PUBLICITY RELEASE LOSE TO WIN CHALLENGE RULES, LIABILITY & PUBLICITY RELEASE CHALLENGE INFO: Participants will enjoy 8 weeks of unlimited access to the gym AND group fitness classes (Elite Group Training and Quick Fit classes),

More information

TRANSFORM PARTNER CHALLENGE RULES, LIABILITY & PUBLICITY RELEASE

TRANSFORM PARTNER CHALLENGE RULES, LIABILITY & PUBLICITY RELEASE TRANSFORM PARTNER CHALLENGE RULES, LIABILITY & PUBLICITY RELEASE CHALLENGE INFO: Participants will enjoy 8 weeks of full access to the gym and unlimited Elite Group Training, IQFit and QuickFit classes,

More information

Body Composition Changes Following a Very Low Calorie Pre-Operative Diet in Patients Undergoing Bariatric Surgery

Body Composition Changes Following a Very Low Calorie Pre-Operative Diet in Patients Undergoing Bariatric Surgery Body Composition Changes Following a Very Low Calorie Pre-Operative Diet in Patients Undergoing Bariatric Surgery 03 September 2018 Jonathan Sivakumar, Lynn Chong, Salena Ward, Tom R Sutherland, Michael

More information

Rapid weight gain in early infancy is associated with adult body fat percentage in young women

Rapid weight gain in early infancy is associated with adult body fat percentage in young women Environ Health Prev Med (2010) 15:381 385 DOI 10.1007/s12199-010-0152-3 SHORT COMMUNICATION Rapid weight gain in early infancy is associated with adult body fat percentage in young women Mari Oyama Toshiko

More information

Body Composition. Lecture Overview. Measuring of Body Composition. Powers & Howely pp Methods of measuring body composition

Body Composition. Lecture Overview. Measuring of Body Composition. Powers & Howely pp Methods of measuring body composition Body Composition Powers & Howely pp 344-356 Lecture Overview Methods of measuring body composition Two-component system Body fatness for health & fitness Obesity and weight control Diet, exercise, and

More information

Sports Performance 15. Section 3.2: Body Composition

Sports Performance 15. Section 3.2: Body Composition Sports Performance 15 Section 3.2: Body Composition The relative percentage of muscle, fat, bone and other tissue in the body Our primary concern in this unit is body fatness and how it pertains to athletic

More information

This is a digitised version of a dissertation submitted to the University of Bedfordshire. It is available to view only.

This is a digitised version of a dissertation submitted to the University of Bedfordshire. It is available to view only. Title The Validity of Two Compartment Model Methods of Body Composition as Compared to Magnetic Resonance Imaging in Asian Indian Versus Caucasian Males Name Ben Davies This is a digitised version of a

More information

Body composition. Body composition models Fluid-metabolism ECF. Body composition models Elemental. Body composition models Anatomic. Molnár Dénes.

Body composition. Body composition models Fluid-metabolism ECF. Body composition models Elemental. Body composition models Anatomic. Molnár Dénes. Body composition models Fluid-metabolism Fat Body composition ECF Molnár Dénes BCM ICF ICS ECS FFM Body composition models Anatomic Fat NSMST SM Body composition models Elemental Miscellaneous Calcium

More information

Whole Body Dual X-Ray Absorptiometry (DXA) to Determine Body Composition

Whole Body Dual X-Ray Absorptiometry (DXA) to Determine Body Composition Whole Body Dual X-Ray Absorptiometry (DXA) to Determine Body Composition Policy Number: 6.01.40 Last Review: 4/2018 Origination: 4/2005 Next Review: 4/2019 Policy Blue Cross and Blue Shield of Kansas City

More information

Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and obese adolescents

Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and obese adolescents British Journal of Nutrition (2003), 90, 987 992 q The Authors 2003 DOI: 10.1079/BJN2003983 Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and

More information

Norland Densitometry A Tradition of Excellence

Norland Densitometry A Tradition of Excellence Norland Densitometry A Tradition of Excellence Norland DXA Bone Density Measurement Osteoporosis is a disease marked by reduced bone strength leading to an increased risk of fractures. About 54 million

More information

BODY COMPOSITION: AN ANALYSIS BETWEEN THE FOOTBALLER AND THANG-TA PRACTITIONER OF MANIPUR

BODY COMPOSITION: AN ANALYSIS BETWEEN THE FOOTBALLER AND THANG-TA PRACTITIONER OF MANIPUR BODY COMPOSITION: AN ANALYSIS BETWEEN THE FOOTBALLER AND THANG-TA PRACTITIONER OF MANIPUR T. INAOBI SINGH 1, MAIBAM CHOURJIT SINGH 2, CHETAN MAIBAM 3 1 Department of Physical Education & Sports Science,

More information

How Well Are We Predicting the Resting Energy Expenditure in Underweight to Obese Brazilian Adults?

How Well Are We Predicting the Resting Energy Expenditure in Underweight to Obese Brazilian Adults? Journal of Food and Nutrition Research, 2019, Vol. 7, No. 1, 19-32 Available online at http://pubs.sciepub.com/jfnr/7/1/4 Published by Science and Education Publishing DOI:10.12691/jfnr-7-1-4 How Well

More information

Bioimpedance Spectroscopy for the Estimation of Fat-Free Mass In End-Stage Renal Disease

Bioimpedance Spectroscopy for the Estimation of Fat-Free Mass In End-Stage Renal Disease Bioimpedance Spectroscopy for the Estimation of Fat-Free Mass In End-Stage Renal Disease A THESIS SUBMITTED TO THE FACULTY OF THE GRADUATE SCHOOL OF THE UNIVERSITY OF MINNESOTA BY Sara M. Vine IN PARTIAL

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Patient: Birth Date: 48.2 years Height / Weight: 150.0 cm 72.0 kg Sex / Ethnic: Female

More information

Prediction of extracellular water and total body water by multifrequency bio-electrical impedance in a Southeast Asian population

Prediction of extracellular water and total body water by multifrequency bio-electrical impedance in a Southeast Asian population Asia Pacific J Clin Nutr (1999) 8(2): 155 159 155 Original Article OA 88 EN Prediction of extracellular water and total body water by multifrequency bio-electrical impedance in a Southeast Asian population

More information

special communication

special communication special communication Determining bone and total body mineral content from body density and bioelectrical response spectroscopy STEVEN F. SICONOLFI, 1 RANDAL J. GRETEBECK, 2 WILLIAM W. WONG, 3 SHERIL S.

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Patient: Birth Date: 43.4 years Height / Weight: 170.0 cm 66.0 kg Sex / Ethnic: Female

More information

Duncan Macfarlane IHP, HKU Parts of this lecture were based on lecture notes provided by the Lindsay Carter Anthropometric Archive, AUT, NZ

Duncan Macfarlane IHP, HKU Parts of this lecture were based on lecture notes provided by the Lindsay Carter Anthropometric Archive, AUT, NZ Body composition assessment issues in athletes 1 Duncan Macfarlane IHP, HKU Parts of this lecture were based on lecture notes provided by the Lindsay Carter Anthropometric Archive, AUT, NZ LEARNING OUTCOMES:

More information

Phase Angle What is it & why is it Important?

Phase Angle What is it & why is it Important? RJL Systems is the Industry Innovator and First Company to introduce Bio-electrical Impedance Analysis ( BIA ) to the market in 1979. Phase Angle What is it & why is it Important? Phase Angle Basics What

More information

Competitor Series. Body Composition Analyzers

Competitor Series. Body Composition Analyzers Competitor Series Body Composition Analyzers Right Arm Bringing it All Together. Giving You the Advantage. Right Leg Trunk Left Arm Rice Lake s Competitor Series body composition analyzers (BCA) are designed

More information

Scientific Research. The Scientific Method. Scientific Explanation

Scientific Research. The Scientific Method. Scientific Explanation Scientific Research The Scientific Method Make systematic observations. Develop a testable explanation. Submit the explanation to empirical test. If explanation fails the test, then Revise the explanation

More information

The Ins and Outs of Body Composition Research

The Ins and Outs of Body Composition Research The Ins and Outs of Body Composition Research Jordan R. Moon, PhD, CSCS*D, HFS NSCA State Director: Colorado Research Director: MusclePharm Sports Science Institute Faculty: United States Sports Academy

More information

ORIGINAL ARTICLE. SA Jebb 1, M Siervo 1, PR Murgatroyd 2, S Evans 1, G Frühbeck 3 and AM Prentice 4. Introduction

ORIGINAL ARTICLE. SA Jebb 1, M Siervo 1, PR Murgatroyd 2, S Evans 1, G Frühbeck 3 and AM Prentice 4. Introduction (2006) 1 7 & 2006 Nature Publishing Group All rights reserved 0307-0565/06 $30.00 www.nature.com/ijo ORIGINAL ARTICLE Validity of the leg-to-leg bioimpedance to estimate changes in body fat during weight

More information

1. Frequently asked questions

1. Frequently asked questions 1. Frequently asked questions Q. How is BMR calculated by boditrax composition monitors? The electrical signals do not directly measure BMR. They measure electrical impedance, and from this, the weight

More information

Resting metabolic rate in Italians: relation with body composition and anthropometric parameters

Resting metabolic rate in Italians: relation with body composition and anthropometric parameters Acta Diabetol (2000) 37:77-81 Springer-Verlag 2000 ORIGINAL A. De Lorenzo A. Andreoli S. Bertoli G. Testolin G. Oriani P. Deurenberg Resting metabolic rate in Italians: relation with body composition and

More information

J EisenkoÈlbl 1, M Kartasurya 1 and K Widhalm 1 * Introduction

J EisenkoÈlbl 1, M Kartasurya 1 and K Widhalm 1 * Introduction (2001) 55, 423±429 ß 2001 Nature Publishing Group All rights reserved 0954±3007/01 $15.00 www.nature.com/ejcn Original Communication Underestimation of percentage fat mass measured by bioelectrical impedance

More information

InBody R20 Body composition Analyzer

InBody R20 Body composition Analyzer Claret Wine www.e-inbody.com InBody R20 Body composition Analyzer Specifications Model Data Transfer Type Electrode Method Frequency InBodyR20 Specification is subject to change without prior notice. USB

More information

The physical educator s role is evolving into that of a teacher who is well educated

The physical educator s role is evolving into that of a teacher who is well educated Assessment and Interpretation of Body Composition in Physical Education PAT VEHRS RON HAGER Should you use skinfolds, bio-electrical impedence, or BMI? The physical educator s role is evolving into that

More information

Fitness Concepts + Principles Packet. Name: Period:

Fitness Concepts + Principles Packet. Name: Period: Fitness Concepts + Principles Packet Name: Period: Content Pages + Directions 5 Components of Fitness.................................................... Slides 3-12 FITT Principle.........................................................

More information

InBody-the product of great technology Experience its speciality

InBody-the product of great technology Experience its speciality InBody-the product of great technology Experience its speciality 80 60 40 InBody is chosen by experts InBody has been praised by the world s medical professionals with its power to analyze and its clinical

More information

Development of a Viable Bedside Ultrasound Protocol to Accurately Predict Appendicular Lean Tissue Mass

Development of a Viable Bedside Ultrasound Protocol to Accurately Predict Appendicular Lean Tissue Mass Development of a Viable Bedside Ultrasound Protocol to Accurately Predict Appendicular Lean Tissue Mass by Michael Paris A thesis presented to the University of Waterloo in fulfillment of the thesis requirement

More information

JEPonline Journal of Exercise Physiologyonline

JEPonline Journal of Exercise Physiologyonline 26 JEPonline Journal of Exercise Physiologyonline Official Journal of The American Society of Exercise Physiologists (ASEP) ISSN 1097-9751 An International Electronic Journal Volume 6 Number 2 May 2003

More information

Single Frequency Bio-Impedance Analyse. Standard Operating Procedures

Single Frequency Bio-Impedance Analyse. Standard Operating Procedures Single Frequency Bio-Impedance Analyse Standard Operating Procedures English version 1.3 15/03/2018 Authors: Heidi Zweers, Hinke Kruizenga, Anneke van den Berg, Nel Reijven en Paul Hulshof Acknowledgment:

More information

Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual- Energy X-Ray Absorptiometers*

Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual- Energy X-Ray Absorptiometers* JOURNAL OF BONE AND MINERAL RESEARCH Volume 13, Number 10, 1998 Blackwell Science, Inc. 1998 American Society for Bone and Mineral Research Bone Mineral and Body Composition Measurements: Cross-Calibration

More information

Manipulating Pretesting Conditions for Measuring Body Composition Using the BodPod and Bioelectrical Impedance AN HONORS THESIS (HONR 499)

Manipulating Pretesting Conditions for Measuring Body Composition Using the BodPod and Bioelectrical Impedance AN HONORS THESIS (HONR 499) Manipulating Pretesting Conditions for Measuring Body Composition Using the BodPod and Bioelectrical Impedance AN HONORS THESIS (HONR 499) Jenna Elwert Thesis Advisor Paul Nagelkirk Signed Ball State University

More information

New bioelectrical impedance formula for patients with respiratory insufficiency: comparison to dual-energy X-ray absorptiometry

New bioelectrical impedance formula for patients with respiratory insufficiency: comparison to dual-energy X-ray absorptiometry Eur Respir J 1998; 12: 960 966 DOI: 10.1183/09031936.98.12040960 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1998 European Respiratory Journal ISSN 0903-1936 New bioelectrical impedance

More information

BODY COMPOSITION ASSESSMENT WITH SEGMENTAL MULTIFREQUENCY BIOIMPEDANCE METHOD *

BODY COMPOSITION ASSESSMENT WITH SEGMENTAL MULTIFREQUENCY BIOIMPEDANCE METHOD * http://www.jssm.org JOURNAL OF SPORTS SCIENCE & MEDICINE VOL. 2 SUPPLEMENTUM 3 2003 BODY COMPOSITION ASSESSMENT WITH SEGMENTAL MULTIFREQUENCY BIOIMPEDANCE METHOD * Jukka A. Salmi Department of Physiology,

More information

What Is Body Composition?

What Is Body Composition? Chapter Six What Is Body Composition? Body composition is the body s relative amounts of fat mass and fat-free mass Body fat includes two categories: Essential fat is crucial for normal body functioning

More information

Prediction of Fatness by Standing 8-Electrode Bioimpedance: A Multiethnic Adolescent Population

Prediction of Fatness by Standing 8-Electrode Bioimpedance: A Multiethnic Adolescent Population nature publishing group articles Prediction of Fatness by Standing 8-Electrode Bioimpedance: A Multiethnic Adolescent Population John D. Sluyter 1, David Schaaf 1, Robert K.R. Scragg 2 and Lindsay D. Plank

More information

Body composition in children and adults by air displacement plethysmography

Body composition in children and adults by air displacement plethysmography European Journal of Clinical Nutrition (1999) 53, 382±387 ß 1999 Stockton Press. All rights reserved 0954±3007/99 $12.00 http://www.stockton-press.co.uk/ejcn Body composition in children and adults by

More information

Accepted Manuscript. S (14) DOI: /j.nut Reference: NUT To appear in: Nutrition

Accepted Manuscript. S (14) DOI: /j.nut Reference: NUT To appear in: Nutrition Accepted Manuscript Body composition following stem cell transplant: comparison of bioimpedance and airdisplacement plethysmography Yun-Chi Hung, BHlthSci Judith D. Bauer, PhD Pamela Horsely, GradDipDietetics

More information

Suprailiac or Abdominal Skinfold Thickness Measured with a Skinfold Caliper as a Predictor of Body Density in Japanese Adults

Suprailiac or Abdominal Skinfold Thickness Measured with a Skinfold Caliper as a Predictor of Body Density in Japanese Adults Tohoku J. Exp. Med., 2007, Measurement 213, 51-61Error Characteristics of Skinfold Caliper 51 Suprailiac or Abdominal Skinfold Thickness Measured with a Skinfold Caliper as a Predictor of Body Density

More information