Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty

Size: px
Start display at page:

Download "Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty"

Transcription

1 Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty A. J. Price, A. Short, C. Kellett, D. Beard, H. Gill, H. Pandit, C. A. F. Dodd, D. W. Murray From the Nuffield Orthopaedic Centre, Oxford, England Polyethylene particulate wear debris continues to be implicated in the aetiology of aseptic loosening following knee arthroplasty. The Oxford unicompartmental knee arthroplasty employs a spherical femoral component and a fully congruous meniscal bearing to increase contact area and theoretically reduce the potential for polyethylene wear. This study measures the in vivo ten-year linear wear of the device, using a roentgenstereophotogrammetric technique. In this in vivo study, seven medial Oxford unicompartmental prostheses, which had been implanted ten years previously were studied. Stereo pairs of radiographs were acquired for each patient and the films were analysed using a roentgen stereophotogrammetric analysis calibration and a computer-aided design model silhouette-fitting technique. Penetration of the femoral component into the original volume of the bearing was our estimate of linear wear. In addition, eight control patients were examined less than three weeks postinsertion of an Oxford prosthesis, where no wear would be expected. The control group showed no measured wear and suggested a system accuracy of 0.1 mm. At ten years, the mean linear wear rate was 0.02 mm/year. The results from this in vivo study confirm that the device has low ten-year linear wear in clinical practice. This may offer the device a survival advantage in the long term. " A. J. Price, DPhil, FRCS(Orth), Clinical Lecturer & Honorary Consultant Orthopaedic Surgeon " A. Short, DPhil, Research Engineer " C. Kellett, FRCS, Specialist Registrar " D. Beard, DPhil, University Research Lecturer " H. Gill, DPhil, University Research Lecturer " H. Pandit, FRCS(Orth), Research Fellow " C. A. F. Dodd, FRCS, Consultant Orthopaedic Surgeon " D. W. Murray, MD, FRCS(Orth), Professor Orthopaedic Surgery Nuffield Orthopaedic Centre, Windmill Road, Headington, Oxford OX3 7LD, UK. Correspondence should be sent to Mr A. J. Price; andrew.price@ndos.ox.ac.uk 2005 British Editorial Society of Bone and Joint Surgery doi: / x.87b $2.00 J Bone Joint Surg [Br] 2005;87-B: Received 24 January 2005; Accepted after revision 14 June 2005 Polyethylene wear remains a major factor in the failure of knee replacements. 1 In the shorter term, catastrophic failure can result in early revision. 2 In the medium- to long-term wear particles are implicated in the pathogenesis of osteolysis and the production of aseptic loosening. 3 Theoretically reducing contact stress at the bearing surface can decrease wear. 4 The use of a fully congruent mobile bearing increases contact area, decreases contact stress and therefore reduces the potential for formation of wear particles. 5 The Oxford unicompartmental knee arthroplasty (UKA; Biomet, Swindon, UK) employs a meniscal bearing and was designed in an attempt to reduce wear. 6 In vitro measurement of polyethylene wear using laboratory simulators and retrieval studies have suggested that the device can produce low wear rates, when functioning normally. 7,8 However, these findings have not been confirmed in vivo. This study combines roentgenstereophotogrammetric analysis (RSA) and computer-aided design models to produce a system for calculating in vivo polyethylene wear as previously applied to assess a total knee replacement. 9,10 Our aim was to measure linear wear occurring in meniscal-bearing unicompartmental knee replacements at a minimum follow-up of ten years. Patients and Methods Measuring wear. A stereo pair of radiographs was simultaneously taken of each knee studied. The technique required the subject to stand on both legs, with the study knee placed within a calibration frame, which is a rigid, triangular open-ended structure. The three planes of the frame contained tantalum marker balls (0.8 mm in diameter) in accurately-known positions relative to a coordinate system embedded in the calibration frame. The marker balls gave rise to image points on the stereo radiographs. The relationship between the three-dimensional coordinate system of the frame and the twodimensional coordinate system of each radiograph was then determined. The process of calculating this relationship is called calibration. Once both images were calibrated, the threedimensional coordinates of any object visible in both images could be calculated with a high degree of accuracy. The images were then processed to detect the silhouettes of the metal components of the VOL. 87-B, No. 11, NOVEMBER

2 1494 A. J. PRICE, A. SHORT, C. KELLETT, D. BEARD, H. GILL, H. PANDIT, C. A. F. DODD, D. W. MURRAY Table I. Details of the control group Case Gender* Side (yrs) Age at surgery Time since surgery (days) 1 F R M R F L F R M L F L F L F L Mean (SD) 66.5 (7.8) 10.1 (5.6) * F, female; M, male all patients had unilateral procedures with phase 3 prostheses; R, right; L, left implant. These were separated into the femoral and tibial component silhouettes. For each component, a virtual silhouette could be generated using the photogrammetric relationship determined by calibration and the appropriate computer-aided design (CAD) model of the component. The model was considered to be placed within a virtual representation of the calibration frame, the projected image giving rise to the virtual silhouettes on each radiograph. The position of the CAD model was systematically adjusted to minimise the mismatch between the actual and virtual silhouettes. This allowed accurate calculation of the threedimensional location of each component. Validation of this method has been described previously. 10 The Oxford UKA has a flat tibial component and a spherical femoral component. Determination of the threedimensional positions of each allows the smallest perpendicular distance between the articulating surfaces of these components to be calculated. This represents the minimum of the bearing, which separates the two metal components. Minor differences in extra-articular bearing shape do not affect the method. By comparing the RSAcalculated bearing with the known of the bearing at time of insertion, the linear wear can be estimated. Estimating bearing at insertion. The nominal size of each bearing studied was known, but manufacturing tolerance introduces some variation to this value. To assess the degree of variation the minimum of 20 unused bearings was measured with a dial gauge. The mean difference in measured to nominal was mm (SD 0.07). Therefore, for all other calculations the of the polyethylene bearing before implantation was taken to be the nominal mm. Control group. As a control, eight patients who were undergoing a medial Oxford UKA were recruited and each had RSA radiographs taken after their operation. The implants were all phase 3 Oxford UKA. This prosthesis is manufactured using a CAD model, which had been made available for the experiment by the manufacturer. Using the system described above, the linear penetration of the femoral component into the bearing was measured, within three weeks of insertion, when it could be assumed no wear had occurred. The measured of polyethylene was compared with the of the bearing at insertion. The details of these patients are given in Table I. Study group. Patients were recruited from a series of medial Oxford UKAs performed by John Goodfellow, one of the designers of the prosthesis. 4,11 An accurate database of the 144 knees in 114 patients was available. All patients had undergone a medial Oxford UKA for anteromedial osteoarthritis, with a functioning anterior cruciate ligament and correctable varus deformity of the knee at the time of surgery. The recruitment of suitable patients from this database involved the following criteria: 1) the patients had to be alive, 2) have received the phase 2 prosthesis to enable the use of a single CAD model, 3) be living in the Oxfordshire region, and therefore not precluded by distance from reattending and 4) have received their prosthesis approximately ten years earlier. From the original 114 patients, seven were identified of whom five (seven knees) agreed to take part in the study. All patients completed an Oxford Knee Score questionnaire 12 and their knees were evaluated using the American Knee Society (knee and function) score. 13 The demographic and clinical details of the patients are given in Table II. All bearings assessed were made using the same manufacturing process and tolerance. There were minor differences in the extra-articular shape of the phase 2 and 3 Table II. Details of the study group Case Gender* Side unilateral Bilateral/ Phase Age at surgery (yrs) Time since surgery (yrs) OKS (0 to 48) AKS score (knee) (0 to 100) AKS score (function) (0 to 100) 1 F R Unilateral F R Bilateral F L Bilateral F R Unilateral F L Unilateral F R Bilateral F L Bilateral Mean (SD) 59.3 (9.61) 11.3 (1.44) 42.6 (2.5) 93.0 (5.6) 94.3 (5.3) * F, female R, right; L, left OKS, Oxford Knee Score 12 AKS, American Knee Society 13 THE JOURNAL OF BONE AND JOINT SURGERY

3 TEN-YEAR IN VIVO WEAR MEASUREMENT OF A FULLY CONGRUENT MOBILE BEARING UNICOMPARTMENTAL KNEE ARTHROPLASTY 1495 Penetration Mean Specimen number Fig. 1 The difference in measured from estimated at insertion for the control bearings. The mean value (95% confidence interval) is also displayed. Penetration Controls Ten-year bearings Time since insertion (yrs) Fig. 2 The measured penetration for each control and study bearing plotted against time. The gradient of the line plotted represents the mean wear rate. bearings used in this study, but importantly the intra-articular shape remained the same. The patients who agreed to enter the study had RSA radiographs taken and linear wear was calculated. For six knees, two sets of radiographs were taken, one in full extension and the other at 30 of flexion. One patient (one knee) declined the second set of radiographs. The full congruency of the articulation should not affect the measurement of bearing regardless of the degree of knee flexion. Therefore, comparison of the two measurements provided a test of reproducibility. Ethical approval for our radiographic study was given by the local ethics committee. Results The penetration measurements of the control bearings are displayed in Figure 1. The mean difference was 0.05 mm (95% confidence interval (CI) 0.01). The details of all linear wear measurements made for the study group are shown in Table III. From the six knees within the study group where two different pairs of radiographs had been taken, the mean difference between penetration measurements was 0.01 mm (95% CI 0.07). In one bearing, the mean penetration measurement was negative, but positive values were recorded for the others. Individual linear wear rates for each bearing have been calculated by dividing measured penetration by the time since insertion (years). The highest linear wear rate was 0.05 mm/year, but four bearings were wearing at approximately 0.02 mm/year. The mean penetration rate was 0.02 mm/year (95% CI 0.01). The results are summarised in Figure 2, which displays the linear wear measurements for each of the study and control bearings plotted against time. The error bars around the study bearings represent the error of the system (SD 0.01), estimated from the confidence interval around the control and repeated measurements. The mean linear wear rate of the study bearings is represented by the gradient of the line. Table III. Polyethylene penetration and calculated wear rates for study bearings. Where two pairs of radiographs were obtained the difference in penetration measurements is also displayed Case Years since surgery Nominal bearing size Nominal Adjusted Nominal Measured 0 flexion 30 flexion Measured difference (0 to 30 ) Penetration 0 flexion 30 flexion Penetration mean ( )/2 Linear wear rate (mm/yr) Mean % CI* * CI, confidence interval VOL. 87-B, No. 11, NOVEMBER 2005

4 1496 A. J. PRICE, A. SHORT, C. KELLETT, D. BEARD, H. GILL, H. PANDIT, C. A. F. DODD, D. W. MURRAY Discussion A combination of RSA techniques and CAD technology provided the methodological basis for measuring wear. The reported control experiments describe a system error of approximately 0.10 mm. The repeated measurements from the study group confirm that the method was repeatable. The radiation exposure from a stereo pair of RSA radiographs represents a very small relative risk to the patient, with the total dose equivalent to approximately three weeks of background radiation. It is concluded that the in vivo method for measuring polyethylene penetration is accurate, reliable and presents minimal risk to patients. The value for penetration represents a combination of creep and linear wear. Delineating the relative contribution of the two processes is a limitation of this method and the measurements represent an overestimation of wear. However the majority of creep is believed to occur within three months of implantation. 14,15 Therefore at ten years post-operation the proportion of true wear to creep will increase, reducing the contribution of creep to the penetration value. The mean in vivo wear rate for the group of ten-year Oxford UKAs was 0.02 mm/year. This figure represents a low wear rate and is similar in magnitude to both the 0.01 mm/year measured with in vitro wear testing, and the value estimated from retrieved bearings with normal macroscopic wear patterns. 7,8 The patients within the study group had good clinical outcomes and the results provide evidence that low wear rates are maintained for up to ten years in patients in whom the device works well. A low wear rate in itself would not be expected to produce any functional outcome benefit for the patient in the shortterm, but should offer the prosthesis a survival advantage, enhancing its longevity. The range of wear rates was to mm/yr (Table III). The apparent increase in bearing suggested by the negative value probably reflects the error introduced in estimating the original bearing. The mean wear rate of 0.02 mm/year measured in the in vivo study compares favourably with the published results of polyethylene penetration for other forms of arthroplasty which use a metal-on-polyethylene bearing. The value is approximately ten times less than the penetration rates of 0.1 to 0.2 mm/year reported for total hip arthroplasty Despite many published retrieval studies there are few reports of the linear penetration rate in total knee arthroplasty. Plante-Bordeneuve and Freeman 19 reported a mean wear rate of mm/year for a conforming total knee arthroplasty (TKA), with a large contact area. However, this device did not afford axial rotation. Contemporary, unconstrained TKAs have penetration rates of around 0.2 mm/year, reflecting their lower contact area. 9,20 Ashraf et al 21 reported a penetration rate of 0.15 mm/year for the St. Georg Sledge fixed-bearing UKA prosthesis. There are no other published penetration rates for fixed-bearing devices, but early catastrophic polyethylene wear has been a problem with many designs and it is safe to assume that linear penetration rates were high in these cases. 2,22,23 Despite high linear wear rates measured in some retrieved specimens, catastrophic wear was not a feature with the Oxford device. The use of in-air gamma-irradiated polyethylene may have been a factor in these findings, but low conformity and high contact stress may predispose fixedbearing devices to more aggressive wear patterns. The RSA system reported in this paper has a number of possible applications. It can be used as an outcome tool for monitoring the wear of established prostheses in the long term and provides a complimentary method to the other forms of assessing polyethylene wear. Additionally, it could be used for studying wear in new designs of prosthesis as they are introduced. The current RSA system can measure wear of approximately 0.1 mm. Wear rates in TKA are approximately 0.1 mm/year, therefore, the system could potentially be used to measure wear at two to three years post-implantation. The system could be employed to identify accelerated wear, which may predispose to early failure. The Oxford meniscal-bearing arthroplasty was designed to reduce wear, a factor that has adversely affected the results of fixed-bearing unicompartmental replacements. The results from this in vivo study confirm that low wear rates can be achieved and maintained in the long term. The author or one or more of the authors have received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this article. In addition, benefits have been or will be directed to a research fund, foundation, educational institution, or other nonprofit organisation with which one or more of the authors are associated. References 1. Goodman S, Lidgren L. Polyethylene wear in knee arthroplasty: a review. Acta Orthop Scand 1992;63: Palmer SH, Morrison PJ, Ross AC. Early catastrophic tibial component wear after unicompartmental knee arthroplasty. Clin Orthop 1998;350: Engh GA, Ammeen DJ. Periprosthetic osteolysis with total knee arthroplasty. Instr Course Lect 2001;50: Goodfellow J, O Connor J. The mechanics of the knee and prosthesis design. J Bone Joint Surg [Br] 1978;60-B: O Connor JJ, Goodfellow JW. Theory and practice of meniscal knee replacement: design against wear. Proc Inst Mech Eng [H] 1996;210: Goodfellow J, O Connor JJ, Murray D. Principles of meniscal bearing arthroplasty for unicompartmental knee replacement. In: Cartier P, Epinette JA, Deschamps G, Hernnigou P, eds. Unicompartmental arthroplasty. First ed. Paris: Expansion Scientific Francaise, Psychoyios V, Crawford RW, O Connor JJ, Murray DW. Wear of congruent meniscal bearings in unicompartmental knee arthroplasty: a retrieval study of 16 specimens. J Bone Joint Surg [Br] 1998;80-B: Scott R, Schroeder D. Correlation of knee similar to in vivo use: evaluating the Oxford unicompartmental knee [abstract]. Trans Orthopaedic Research Society 2000: Kellett K, Short A, Price A, Gill H, Murray D. In vivo measurement of total knee replacement wear. The Knee 2004;11: Short A, Gill H, Marks B, et al. A novel method for in vivo knee prosthesis wear measurement. J Biomech 2005;38: Murray DW, Goodfellow JW, O Connor JJ. The Oxford medial unicompartmental arthroplasty: a ten-year survival study. J Bone Joint Surg [Br] 1998;80-B: Dawson J, Fitzpatrick R, Murray D, Carr A. Questionnaire on the perceptions of patients about total knee replacement. J Bone Joint Surg [Br] 1998;80-B: Insall JN, Ranawat CS, Aglietti P, Shine J. A comparison of four models of total knee-replacement prostheses. J Bone Joint Surg [Am] 1976;58-A: Deng M, Latour RA, Ogale AA, Shalaby SW. Study of creep behavior of ultra-highmolecular-weight polyethylene systems. J Biomed Mater Res 1998;40: THE JOURNAL OF BONE AND JOINT SURGERY

5 TEN-YEAR IN VIVO WEAR MEASUREMENT OF A FULLY CONGRUENT MOBILE BEARING UNICOMPARTMENTAL KNEE ARTHROPLASTY Lee KY, Pienkowski D. Compressive creep characteristics of extruded ultra-highmolecular-weight polyethylene. J Biomed Mater Res 1998;39: Kabo JM, Gebhard JS, Loren G, Amstutz HC. In vivo wear of polyethylene acetabular components. J Bone Joint Surg [Br] 1993;75-B: Wroblewski BM. Direction and rate of socket wear in Charnley low-friction arthroplasty. J Bone Joint Surg [Br] 1985;67-B: Schmalzried TP, Callaghan JJ. Wear in total hip and knee replacements. J Bone Joint Surg [Am] 1999;81-A: Plante-Bordeneuve P, Freeman MA. Tibial high-density polyethylene wear in conforming tibiofemoral prostheses. J Bone Joint Surg [Br] 1993;75-B: Hoshino A, Fukouka Y, Ishida A. Accurate in vivo measurement of polyethylene wear in total knee arthroplasty. J Arthroplasty 2002;17: Ashraf T, Newman JH, Desai VV, Beard D, Nevelos JE. Polyethylene wear in a non-congruous unicompartmental knee replacement: a retrieval analysis. The Knee 2004;11: Engh GA, Dwyer KA, Hanes CK. Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prostheses. J Bone Joint Surg [Br] 1992;74-B: Lindstrand A, Stenstrom A, Lewold S. Multicenter study of unicompartmental knee revision: PCA, Marmor, and St Georg compared in 3,777 cases of arthrosis. Acta Orthop Scand 1992;63: VOL. 87-B, No. 11, NOVEMBER 2005

Polyethylene wear of mobile-bearing unicompartmental knee replacement at 20 years

Polyethylene wear of mobile-bearing unicompartmental knee replacement at 20 years KNEE Polyethylene wear of mobile-bearing unicompartmental knee replacement at 20 years B. J. L. Kendrick, D. J. Simpson, B. L. Kaptein, E. R. Valstar, H. S. Gill, D. W. Murray, A. J. Price From the Nuffield

More information

Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year

Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year H. Pandit, C. Jenkins, D. J. Beard, J. Gallagher, A. J. Price, C. A. F. Dodd, J. W. Goodfellow, D. W. Murray From

More information

Assessment of radiolucent lines around the Oxford unicompartmental knee replacement

Assessment of radiolucent lines around the Oxford unicompartmental knee replacement KNEE Assessment of radiolucent lines around the Oxford unicompartmental knee replacement SENSITIVITY AND SPECIFICITY FOR LOOSENING S. Kalra, T. O. Smith, B. Berko, N. P. Walton From Norfolk and Norwich

More information

Pre-operative clinical and radiological

Pre-operative clinical and radiological Pre-operative clinical and radiological assessment of the patellofemoral joint in unicompartmental knee replacement and its influence on outcome D. J. Beard, H. Pandit, S. Ostlere, C. Jenkins, C. A. F.

More information

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing Journal of Orthopaedic Surgery 2001, 9(1): 45 50 Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing KY Chiu, TP Ng, WM Tang and P Lam Department of Orthopaedic Surgery, The University

More information

Alignment of the femoral component in a mobile-bearing unicompartmental knee arthroplasty Kort, N. P.; van Raay, J. J. A. M.; Thomassen, B. J. W.

Alignment of the femoral component in a mobile-bearing unicompartmental knee arthroplasty Kort, N. P.; van Raay, J. J. A. M.; Thomassen, B. J. W. University of Groningen Alignment of the femoral component in a mobile-bearing unicompartmental knee arthroplasty Kort, N. P.; van Raay, J. J. A. M.; Thomassen, B. J. W. Published in: Knee DOI: 10.1016/j.knee.2007.04.007

More information

With over 40 years clinical experience, the Oxford Partial Knee is the most widely used, 1 clinically proven 2 partial knee system in the world.

With over 40 years clinical experience, the Oxford Partial Knee is the most widely used, 1 clinically proven 2 partial knee system in the world. Oxford Partial Knee A Definitive Implant With over 40 years clinical experience, the Oxford Partial Knee is the most widely used, 1 clinically proven 2 partial knee system in the world. Compared with total

More information

The effect of leg alignment on the outcome of unicompartmental knee replacement

The effect of leg alignment on the outcome of unicompartmental knee replacement The effect of leg alignment on the outcome of unicompartmental knee replacement A. Gulati, H. Pandit, C. Jenkins, R. Chau, C. A. F. Dodd, D. W. Murray From Nuffield Orthopaedic Centre, Oxford, England

More information

Oxford. Partial Knee

Oxford. Partial Knee Oxford Partial Knee Oxford Partial Knee A Definitive Implant The Oxford Partial Knee is the most widely used 38 and clinically proven 22 partial knee replacement (PKR) system in the world, offering industry

More information

Activity Level and Wear in Total Knee Arthroplasty

Activity Level and Wear in Total Knee Arthroplasty The Journal of Arthroplasty Vol. 16 No. 4 2001 Activity Level and Wear in Total Knee Arthroplasty A Study of Autopsy Retrieved Specimens Carlos J. Lavernia, MD,* Rafael J. Sierra, MD,* David S. Hungerford,

More information

Mid-Term Results of Oxford Medial Unicompartmental Knee Arthroplasty

Mid-Term Results of Oxford Medial Unicompartmental Knee Arthroplasty Original Article Clinics in Orthopedic Surgery 2011;3:178-183 http://dx.doi.org/10.4055/cios.2011.3.3.178 Mid-Term Results of Oxford Medial Unicompartmental Knee Arthroplasty Won-Sik Choy, MD, Kap Jung

More information

The Knee. The Twin Peg Oxford partial knee replacement: The first 100 cases. Stephen H. White a,, Sharon Roberts a, Peter W.

The Knee. The Twin Peg Oxford partial knee replacement: The first 100 cases. Stephen H. White a,, Sharon Roberts a, Peter W. The Knee 19 (2012) 36 40 Contents lists available at ScienceDirect The Knee The Twin Peg Oxford partial knee replacement: The first 100 cases Stephen H. White a,, Sharon Roberts a, Peter W. Jones b a Department

More information

Conversion of unicompartmental knee arthroplasty to total knee arthroplasty : The challenges and need for augments

Conversion of unicompartmental knee arthroplasty to total knee arthroplasty : The challenges and need for augments Acta Orthop. Belg., 2013, 79, 699-705 ORIGINAL STUDY Conversion of unicompartmental knee arthroplasty to total knee arthroplasty : The challenges and need for augments Zeeshan Khan, Syed Z. Nawaz, Steven

More information

Evolution. Medial-Pivot Knee System The Bi-Cruciate-Substituting Knee. Key Aspects

Evolution. Medial-Pivot Knee System The Bi-Cruciate-Substituting Knee. Key Aspects Evolution Medial-Pivot Knee System The Bi-Cruciate-Substituting Knee Key Aspects MicroPort s EVOLUTION Medial-Pivot Knee System was designed to recreate the natural anatomy that is lost during a total

More information

TOTAL KNEE ARTHROPLASTY (TKA)

TOTAL KNEE ARTHROPLASTY (TKA) TOTAL KNEE ARTHROPLASTY (TKA) 1 Anatomy, Biomechanics, and Design 2 Femur Medial and lateral condyles Convex, asymmetric Medial larger than lateral 3 Tibia Tibial plateau Medial tibial condyle: concave

More information

Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval

Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval J. P. Collier,* M. B. Mayor,+ V. A. Surprenant,* H. P. Surprenant,* and R.E. Jensen* *Dartmouth Biomedical

More information

BIOMECHANICAL MECHANISMS FOR DAMAGE: RETRIEVAL ANALYSIS AND COMPUTATIONAL WEAR PREDICTIONS IN TOTAL KNEE REPLACEMENTS

BIOMECHANICAL MECHANISMS FOR DAMAGE: RETRIEVAL ANALYSIS AND COMPUTATIONAL WEAR PREDICTIONS IN TOTAL KNEE REPLACEMENTS Journal of Mechanics in Medicine and Biology Vol. 5, No. 3 (2005) 469 475 c World Scientific Publishing Company BIOMECHANICAL MECHANISMS FOR DAMAGE: RETRIEVAL ANALYSIS AND COMPUTATIONAL WEAR PREDICTIONS

More information

Does a minimally invasive approach affect positioning of components in unicompartmental knee arthroplasty? Early results with survivorship analysis

Does a minimally invasive approach affect positioning of components in unicompartmental knee arthroplasty? Early results with survivorship analysis Acta Orthop. Belg., 2006, 72, 709-715 ORIGINAL STUDY Does a minimally invasive approach affect positioning of components in unicompartmental knee arthroplasty? Early results with survivorship analysis

More information

Case Report Meniscal Bearing Dislocation of Unicompartmental Knee Arthroplasty with Faint Symptom

Case Report Meniscal Bearing Dislocation of Unicompartmental Knee Arthroplasty with Faint Symptom Case Reports in Orthopedics Volume 2015, Article ID 217842, 5 pages http://dx.doi.org/10.1155/2015/217842 Case Report Meniscal Bearing Dislocation of Unicompartmental Knee Arthroplasty with Faint Symptom

More information

Management of Periprosthetic Fracture in Unicompartmental Knee Arthroplasty Patients: A Case Series

Management of Periprosthetic Fracture in Unicompartmental Knee Arthroplasty Patients: A Case Series Management of Periprosthetic Fracture in Unicompartmental Knee Arthroplasty Patients: A Case Series Yew Lok Woo, MBBS, Pak Lin Chin, MBBS, FRCS, Ngai Nung Lo, MBBS, FRCS, Shi-Lu Chia, PhD, Darren Keng

More information

Limb alignment in computer-assisted minimally-invasive unicompartmental knee replacement

Limb alignment in computer-assisted minimally-invasive unicompartmental knee replacement Limb alignment in computer-assisted minimally-invasive unicompartmental knee replacement G. Keene, D. Simpson, Y. Kalairajah From Sportsmed-SA Orthopaedic Clinic, Adelaide, South Australia Twenty patients

More information

Estimating Total Knee Arthroplasty Joint Loads from Kinematics

Estimating Total Knee Arthroplasty Joint Loads from Kinematics Estimating Total Knee Arthroplasty Joint Loads from Kinematics Clare K. Fitzpatrick, Paul Rullkoetter. University of Denver, Denver, CO, USA. Disclosures: C.K. Fitzpatrick: None. P. Rullkoetter: 5; DePuy

More information

Frontal Plane Kinematics After Mobile- Bearing Total Knee Arthroplasty

Frontal Plane Kinematics After Mobile- Bearing Total Knee Arthroplasty CLINICAL ORTHOPAEDICS AND RELATED RESEARCH Number 392, pp. 56 61 2001 Lippincott Williams & Wilkins, Inc. Frontal Plane Kinematics After Mobile- Bearing Total Knee Arthroplasty James B. Stiehl, MD*; Richard

More information

The young patient and the medial unicompartmental knee replacement

The young patient and the medial unicompartmental knee replacement Acta Orthop. Belg., 2015, 81, 283-288 ORIGINAL STUDY The young patient and the medial unicompartmental knee replacement Omar Faour Martín, Jose Antonio Valverde García, Miguel Ángel Martín Ferrero, Aurelio

More information

SEVERE VARUS AND VALGUS DEFORMITIES TREATED BY TOTAL KNEE ARTHROPLASTY

SEVERE VARUS AND VALGUS DEFORMITIES TREATED BY TOTAL KNEE ARTHROPLASTY SEVERE VARUS AND VALGUS DEFORMITIES TREATED BY TOTAL KNEE ARTHROPLASTY Th. KARACHALIOS, P. P. SARANGI, J. H. NEWMAN From Winford Orthopaedic Hospital, Bristol, England We report a prospective case-controlled

More information

Gold standard of a TKA. Conflicting goals? POLYETHYLENE WEAR THE SOLUTION: MOBILE BEARING KNEES. MOBILE BEARING A totally new approach (1977)

Gold standard of a TKA. Conflicting goals? POLYETHYLENE WEAR THE SOLUTION: MOBILE BEARING KNEES. MOBILE BEARING A totally new approach (1977) Changing designs : the case against mobile bearing? Gold standard of a TKA Goal of a TKA: 1. Pain 2. Motion 3. Longevity Stress Guy BELLIER M.D. PARIS France Conformity = Durability w/o constraints = non

More information

Unicompartmental or total knee replacement

Unicompartmental or total knee replacement Unicompartmental or total knee replacement THE 15-YEAR RESULTS OF A PROSPECTIVE RANDOMISED CONTROLLED TRIAL J. Newman, R. V. Pydisetty, C. Ackroyd From Avon Orthopaedic Centre, Bristol, England Between

More information

A study of functional outcome after Primary Total Knee Arthroplasty in elderly patients

A study of functional outcome after Primary Total Knee Arthroplasty in elderly patients Original Research Article A study of functional outcome after Primary Total Knee Arthroplasty in elderly patients Ragesh Chandran 1*, Sanath K Shetty 2, Ashwin Shetty 3, Bijith Balan 1, Lawrence J Mathias

More information

Midterm results of cemented Press Fit Condylar Sigma total knee arthroplasty system

Midterm results of cemented Press Fit Condylar Sigma total knee arthroplasty system Journal of Orthopaedic Surgery 2005:13(3):280-284 Midterm results of cemented Press Fit Condylar Sigma total knee arthroplasty system S Asif, DSK Choon Department of Orthopaedic Surgery, University of

More information

Revision surgery for failed unicompartmental knee replacement : Technical aspects and clinical outcome

Revision surgery for failed unicompartmental knee replacement : Technical aspects and clinical outcome Acta Orthop. Belg., 2013, 79, 312-317 ORIGINAL STUDY Revision surgery for failed unicompartmental knee replacement : Technical aspects and clinical outcome Curtis A. Robb, Gulraj S. Matharu, Khalid Baloch,

More information

The Survivorship and Clinical Results of Minimally Invasive Unicompartmental Knee Arthroplasty at 10-Year Follow-up

The Survivorship and Clinical Results of Minimally Invasive Unicompartmental Knee Arthroplasty at 10-Year Follow-up Original Article Clinics in Orthopedic Surgery 2015;7:199-206 http://dx.doi.org/10.4055/cios.2015.7.2.199 The Survivorship and Clinical Results of Minimally Invasive Unicompartmental Knee Arthroplasty

More information

The influence of degeneration of the lateral patellofemoral joint on outcome of medial unicompartmental knee replacement

The influence of degeneration of the lateral patellofemoral joint on outcome of medial unicompartmental knee replacement 2015 4 ORTHOPAEDIC BIOMECHANICS MATERIALS AND CLINICAL STUDY.29. doi: 10.3969/j.issn. 1672-5972.2015.02.008 swgk2015-03-0050 * 1 1 2 1 * 1 1 [ ] Oxford 2009 8 2013 12 126 (104 ) Ahlback Weidow 5 1 Hospital

More information

Implant Overhang after Unicompartmental Knee Arthroplasty: Oxford Prosthesis versus Miller-Galante II Prosthesis

Implant Overhang after Unicompartmental Knee Arthroplasty: Oxford Prosthesis versus Miller-Galante II Prosthesis Original Article Knee Surg Relat Res 2014;26(2):82-87 http://dx.doi.org/10.5792/ksrr.2014.26.2.82 pissn 2234-0726 eissn 2234-2451 Knee Surgery & Related Research Implant Overhang after Unicompartmental

More information

Minimally invasive unicompartmental knee arthroplasty in treatment of osteonecrosis versus osteoarthritis : a matched-pair comparison

Minimally invasive unicompartmental knee arthroplasty in treatment of osteonecrosis versus osteoarthritis : a matched-pair comparison Acta Orthop. Belg., 2015, 81, 333-339 ORIGINAL STUDY Minimally invasive unicompartmental knee arthroplasty in treatment of osteonecrosis versus osteoarthritis : a matched-pair comparison Qidong Zhang,

More information

FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY

FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY Fixation First FIXATION FIRST 1958 Charnley Hip Over 1,000,000 implanted 1 96.2 percent survivorship at 32 years

More information

Section of total knee replacement. Total Knee Replacement System. Knieendoprothesen System. Système de prothèse totale de genou

Section of total knee replacement. Total Knee Replacement System. Knieendoprothesen System. Système de prothèse totale de genou Section of total knee replacement Total Knee Replacement System Knieendoprothesen System Système de prothèse totale de genou Introduction: This knee system features great versality with its modular component

More information

The Oxford phase III unicompartmental knee replacement in patients less than 60 years of age Kort, Nanne P.; van Raay, Jos J. A. M.; van Horn, Jim J.

The Oxford phase III unicompartmental knee replacement in patients less than 60 years of age Kort, Nanne P.; van Raay, Jos J. A. M.; van Horn, Jim J. University of Groningen The Oxford phase III unicompartmental knee replacement in patients less than 60 years of age Kort, Nanne P.; van Raay, Jos J. A. M.; van Horn, Jim J. Published in: Knee surgery

More information

ATTUNE KNEE SYSTEM: SOFCAM CONTACT

ATTUNE KNEE SYSTEM: SOFCAM CONTACT ATTUNE KNEE SYSTEM: SOFCAM CONTACT Douglas A. Dennis, MD Medical Director at Porter Center for Joint Replacement Denver, Colorado Historically, sagittal plane instability following Total Knee Arthroplasty

More information

Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists

Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists : 31 35 doi 10.1308/1478708051450 Audit Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists Jac Ciampolini, Matthew JW Hubble Princess Elizabeth Orthopaedic Centre,

More information

MARATHON Cross-Linked Polyethylene

MARATHON Cross-Linked Polyethylene MARATHON Cross-Linked Polyethylene Fixation First... 1 Wear Reduction... 2 Oxidative Stability... 3 Mechanical Integrity... 4 Proven Technology...5-6 References... 7 Fixation First FIXATION FIRST Prodigy

More information

Three-Dimensional Analysis of Alignment Error in Using Femoral Intramedullary Guides in Unicompartmental Knee Arthroplasty

Three-Dimensional Analysis of Alignment Error in Using Femoral Intramedullary Guides in Unicompartmental Knee Arthroplasty The Journal of Arthroplasty Vol. 21 No. 2 2006 Three-Dimensional Analysis of Alignment Error in Using Femoral Intramedullary Guides in Unicompartmental Knee Arthroplasty Burton Ma, MSc,* William Long,

More information

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner Acta Orthop. Belg., 2008, 74, 337-342 ORIGINAL STUDY Poor mid-term results of total hip arthroplasty with use of a Hylamer liner Adrian SKWARA, Simone STRACKE, Carsten O. TIBESKU, Susanne FUCHS-WINKELMANN

More information

We studied the migration of 58 cemented Hinek

We studied the migration of 58 cemented Hinek MIGRATION OF CEMENTED FEMORAL COMPONENTS AFTER THR ROENTGEN STEREOPHOTOGRAMMETRIC ANALYSIS J. KISS, D. W. MURRAY, A. R. TURNER-SMITH, J. BITHELL, C. J. BULSTRODE From the Nuffield Orthopaedic Centre, Oxford,

More information

Simplifying the Most Clinically Proven 1 Partial Knee in the World. Oxford Partial Knee with Microplasty Instrumentation

Simplifying the Most Clinically Proven 1 Partial Knee in the World. Oxford Partial Knee with Microplasty Instrumentation Simplifying the Most Clinically Proven 1 Partial Knee in the World Oxford Partial Knee with Microplasty Instrumentation 1 Microplasty Instrumentation Innovative, Accurate, Reproducible Microplasty Instrumentation

More information

Rotaglide+ TM. Total Knee System Product overview

Rotaglide+ TM. Total Knee System Product overview Rotaglide+ TM Total Knee System Product overview Rotaglide+ TM Originality Stability History Originally implanted in 1988, Rotaglide+ was the first total knee design to adopt a true mobile bearing philosophy.

More information

39 year old, 260 lb. park ranger. TKA in situ: 4 Years. First reported Metal-backed, non-modular PCA knee. 16% incidence. Modular metalbacked

39 year old, 260 lb. park ranger. TKA in situ: 4 Years. First reported Metal-backed, non-modular PCA knee. 16% incidence. Modular metalbacked Designer (Royalty income) - DePuy A Johnson & Johnson Company Consultant on Knee Products for Smith & Nephew Orthopaedics General Research Funding Inova Health System and the U.S. Army Medical Research

More information

Complications of Medial Unicompartmental Knee Arthroplasty

Complications of Medial Unicompartmental Knee Arthroplasty Original Article Clinics in Orthopedic Surgery 2014;6:365-372 http://dx.doi.org/10.4055/cios.2014.6.4.365 Complications of Medial Unicompartmental Knee Arthroplasty Jong Hun Ji, MD, Sang Eun Park, MD,

More information

Design of a Novel Total Knee Prosthesis Using TRIZ

Design of a Novel Total Knee Prosthesis Using TRIZ Author: Yeh-Liang Hsu, Yung-Chieh Hung, Ju-Zhong Yin (2006-10-28); recommended: Yeh-Liang Hsu (2007-08-02). Note: This paper is published in Journal of Medical and Biological Engineering, Vol. 26, No.

More information

Stability without compromise. Epsilon Durasul Constrained Insert

Stability without compromise. Epsilon Durasul Constrained Insert Stability without compromise Epsilon Durasul Constrained Insert History of Dislocation and Constrained Acetabular Inserts The Need for a Better Solution The Problem Dislocation is the second most common

More information

Unicompartmental Knee Replacement

Unicompartmental Knee Replacement Unicompartmental Knee Replacement Results and Techniques Alexander P. Sah, MD California Orthopaedic Association Meeting Laguna Niguel, CA May 20th, 2011 Overview Why partial knee replacement? - versus

More information

Bilateral Total Knee Replacements using Two Different Implant Designs - Preliminary Report

Bilateral Total Knee Replacements using Two Different Implant Designs - Preliminary Report Bahrain Medical Bulletin, Vol. 27, No. 1, March 2005 Bilateral Total Knee Replacements using Two Different Implant Designs - Preliminary Report Ali Redha, MD* Abdalla A Malki FRCS Ed, MD Orth** Faisal

More information

CLINICAL AND OPERATIVE APPROACH FOR TOTAL KNEE REPLACEMENT DR.VINMAIE ORTHOPAEDICS PG 2 ND YEAR

CLINICAL AND OPERATIVE APPROACH FOR TOTAL KNEE REPLACEMENT DR.VINMAIE ORTHOPAEDICS PG 2 ND YEAR CLINICAL AND OPERATIVE APPROACH FOR TOTAL KNEE REPLACEMENT DR.VINMAIE ORTHOPAEDICS PG 2 ND YEAR Evolution of TKR In 1860, Verneuil proposed interposition arthroplasty, involving the insertion of soft tissue

More information

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head Enhancing stability and increasing range of motion Metasul LDH Large Diameter Head The Metasul large diameter head technology is the result of in-depth research, development and clinical experience that

More information

EMPOWERING LIFE THROUGH NATURAL MOTION

EMPOWERING LIFE THROUGH NATURAL MOTION EMPOWERING LIFE THROUGH NATURAL MOTION em pow r əm pou(ə)r/ verb make (someone) stronger and more confident, especially in controlling their life and claiming their rights. WHY NATURAL MOTION? Today s

More information

Fixed or Mobile bearing in Unicompartmental Knee Arthroplasty

Fixed or Mobile bearing in Unicompartmental Knee Arthroplasty Fixed or Mobile bearing in Unicompartmental Knee Arthroplasty Jean-Noel Argenson, Matthieu Ollivier, Xavier Flecher, Sebastien Parratte Institute for Locomotion Sainte Marguerite Hospital, Marseille, France

More information

UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD

UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS 66th Annual Meeting February 4-9, 1999 Anaheim, California COMMITTEE ON BIOMEDICAL ENGINEERING COMMITTEE ON

More information

Radiological Decision Aid to determine suitability for medial unicompartmental knee arthroplasty

Radiological Decision Aid to determine suitability for medial unicompartmental knee arthroplasty KNEE Radiological Decision Aid to determine suitability for medial unicompartmental knee arthroplasty DEVELOPMENT AND PRELIMINARY VALIDATION T. W. Hamilton, H. G. Pandit, A. V. Lombardi, J. B. Adams, C.

More information

HIGH FLEXION IN CONTEMPORARY TOTAL KNEE DESIGN: A PRECURSOR OF UHMWPE DAMAGE? A FINITE ELEMENT STUDY

HIGH FLEXION IN CONTEMPORARY TOTAL KNEE DESIGN: A PRECURSOR OF UHMWPE DAMAGE? A FINITE ELEMENT STUDY HIGH FLEXION IN CONTEMPORARY TOTAL KNEE DESIGN: A PRECURSOR OF UHMWPE DAMAGE? A FINITE ELEMENT STUDY Orthopaedic Research Laboratories Cleveland, Ohio Edward A. Morra, M.S.M.E. A. Seth Greenwald, D.Phil.(Oxon)

More information

Oxford Phase 3 unicompartmental knee arthroplasty: medium-term results of a minimally invasive surgical procedure

Oxford Phase 3 unicompartmental knee arthroplasty: medium-term results of a minimally invasive surgical procedure Knee Surg Sports Traumatol Arthrosc (2011) 19:277 284 DOI 10.1007/s00167-010-1213-2 KNEE Oxford Phase 3 unicompartmental knee arthroplasty: medium-term results of a minimally invasive surgical procedure

More information

Developing techniques for the dimensional assessment of true clinical wear in retrieved total joint replacements.

Developing techniques for the dimensional assessment of true clinical wear in retrieved total joint replacements. Developing techniques for the dimensional assessment of true clinical wear in retrieved total joint replacements. P. Bills*, L. Blunt*, G. Chakrabarty** *University of Huddersfield, Huddersfield, UK **West

More information

W. (2016) : - 98 (8) ISSN X,

W. (2016) : - 98 (8) ISSN X, Bell, Stuart W. and Anthony, Iain and Jones, Bryn and MacLean, Angus and Rowe, Philip and Blyth, Mark (2016) Improved accuracy of component positioning with robotic assisted unicompartmental knee arthroplasty

More information

Observations on retrieved articular surfaces of total knee prostheses. Department of Orthopaedics, Leiden University Medical Center, The Netherlands

Observations on retrieved articular surfaces of total knee prostheses. Department of Orthopaedics, Leiden University Medical Center, The Netherlands Chapter 11 Observations on retrieved articular surfaces of total knee prostheses Eric H. Garling 1, Henk K. Koerten 2, Edward R. Valstar 1,3, Rob G.H.H. Nelissen 1 1 Department of Orthopaedics, Leiden

More information

why bicompartmental? A REVOLUTIONARY ALTERNATIVE TO TOTAL KNEE REPLACEMENTS

why bicompartmental? A REVOLUTIONARY ALTERNATIVE TO TOTAL KNEE REPLACEMENTS why bicompartmental? A REVOLUTIONARY ALTERNATIVE TO TOTAL KNEE REPLACEMENTS TKR is not always the answer Today, many patients with medial or lateral disease and patellofemoral involvement receive a Total

More information

Triathlon Tritanium Knee System

Triathlon Tritanium Knee System Triathlon Tritanium Knee System Table of Contents Cementless TKA... 4 Why Cementless TKA?... 4 Cementless TKA - Clinical History... 4 Published RSA Results Comparing Fixation Methods... 5 Biologic Fixation...

More information

Medial unicompartmental knee replacement in the under-50s

Medial unicompartmental knee replacement in the under-50s Medial unicompartmental knee replacement in the under-50s S. Parratte, J.-N.A.Argenson, O. Pearce, V. Pauly, P. Auquier, J.-M. Aubaniac From Aix-Marseille University, Marseille, France We retrospectively

More information

The arc of pelvic motion has been shown to be as great as 70 through sit to stand activities 1 DiGioia CORR 2006

The arc of pelvic motion has been shown to be as great as 70 through sit to stand activities 1 DiGioia CORR 2006 The arc of pelvic motion has been shown to be as great as 70 through sit to stand activities 1 DiGioia CORR 2006 Overview Every patient moves differently 1 and their total hip replacement should be optimised

More information

Instrumentation Options

Instrumentation Options Instrumentation Options With the introduction of Microplasty and Premier Instrumentation platforms and advancements such as our patented Slidex Technology, the Vanguard Complete Knee System is the surgeon

More information

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY by Timothy McTighe, Exec. Dir., JISRF, Chagrin Falls, Ohio Ying Ko, Ph.D., Cincinnati, Ohio Russell B. Bennett, Ph.D., Cincinnati, Ohio James

More information

Original Article. Introduction. Materials and Methods

Original Article. Introduction. Materials and Methods Original Article Knee Surg Relat Res 2012;24(1):7-13 http://dx.doi.org/10.5792/ksrr.2012.24.1.7 pissn 2234-0726 eissn 2234-2451 Knee Surgery & Related Research A Comparison of the Clinical and Radiographic

More information

The Role of Partial Knee Arthroplasty for Medial OA of the Knee: It s Time has come, Again 34 th Annual Verne T. Inman Lecture.

The Role of Partial Knee Arthroplasty for Medial OA of the Knee: It s Time has come, Again 34 th Annual Verne T. Inman Lecture. The Role of Partial Knee Arthroplasty for Medial OA of the Knee: It s Time has come, Again 34 th Annual Verne T. Inman Lecture David R. Mauerhan, MD Department of Orthopaedic Surgery Carolinas Medical

More information

CLASSIFICATION OF MOBILE BEARING KNEE DESIGN: MOBILITY AND CONSTRAINT

CLASSIFICATION OF MOBILE BEARING KNEE DESIGN: MOBILITY AND CONSTRAINT CLASSIFICATION OF MOBILE BEAING KNEE DESIGN: MOBILITY AND CONSTAINT Orthopaedic esearch Laboratories Lutheran Hospital Cleveland Clinic Health System Christine S. Heim, B.Sc. Paul. D. Postak, B.Sc. Nicholas

More information

Unicondylar Knee Vs Total Knee Replacement: Is Less Better In the Middle Aged Athlete

Unicondylar Knee Vs Total Knee Replacement: Is Less Better In the Middle Aged Athlete Unicondylar Knee Vs Total Knee Replacement: Is Less Better In the Middle Aged Athlete Chair: Maurilio Marcacci, MD Alois Franz "Basic principles and considerations of the Unis" Joao M. Barretto "Sport

More information

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code Patrick Chaboche, MD patrick.chaboche@wanadoo.fr Former Head of Department of Orthopaedic Surgery. Centre

More information

1. White SH, Nazarian NA, McEwen Smith AM, Balfour TW. Periampullary adenoma causing pancreatitis. Br Med J 1981; 283:527.

1. White SH, Nazarian NA, McEwen Smith AM, Balfour TW. Periampullary adenoma causing pancreatitis. Br Med J 1981; 283:527. Mr Steve White, Consultant Orthopaedic Surgeon Peer-reviewed original papers: 1. White SH, Nazarian NA, McEwen Smith AM, Balfour TW. Periampullary adenoma causing pancreatitis. Br Med J 1981; 283:527.

More information

ATTUNE Knee System: Stability in Total Knee Replacement

ATTUNE Knee System: Stability in Total Knee Replacement ATTUNE Knee System: Stability in Total Knee Replacement Chadd Clary, PhD Staff Engineer DePuy Synthes Joint Reconstruction Young and active total knee replacement (TKR) patients demand a knee that feels

More information

Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo?

Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo? Journal of Orthopaedic Surgery 2005;13(1):46-51 Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo? H Tsumura, N Kaku, T Torisu Department of Orthopedic Surgery, Oita University,

More information

Bicruciate-Retaining or Medial Pivot Total Knee Prosthesis Pritchett 225 Fig. 3. The MP total knee prosthesis. Fig. 1. An anteroposterior radiograph o

Bicruciate-Retaining or Medial Pivot Total Knee Prosthesis Pritchett 225 Fig. 3. The MP total knee prosthesis. Fig. 1. An anteroposterior radiograph o The Journal of Arthroplasty Vol. 26 No. 2 2011 Patients Prefer A Bicruciate-Retaining or the Medial Pivot Total Knee Prosthesis James W. Pritchett, MD, FACS Abstract: Four-hundred forty patients underwent

More information

Total hip replacement with a zirconium oxide ceramic femoral head

Total hip replacement with a zirconium oxide ceramic femoral head Total hip replacement with a zirconium oxide ceramic femoral head A RANDOMISED ROENTGEN STEREOPHOTOGRAMMETRIC STUDY T. von Schewelov, L. Sanzén, I. Önsten, Å. Carlsson, J. Besjakov From Lund University

More information

Mid-Term Results of Oxford Medial Unicompartmental Knee Arthroplasty in Young Asian Patients Less than 60 Years of Age: A Minimum 5-Year Follow-up

Mid-Term Results of Oxford Medial Unicompartmental Knee Arthroplasty in Young Asian Patients Less than 60 Years of Age: A Minimum 5-Year Follow-up Original Article Knee Surg Relat Res 2017;29(2):122-128 https://doi.org/10.5792/ksrr.16.045 pissn 2234-0726 eissn 2234-2451 Knee Surgery & Related Research Mid-Term Results of Oxford Medial Unicompartmental

More information

National Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement

National Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement National Joint Replacement Registry Lay Summary 2015 Annual Report Hip and Knee Replacement SUPPLEMENTARY REPORT 2015 TABLE OF CONTENTS Introduction... 1 A brief history of the Registry origins... 1 The

More information

Simplifying the Most Clinically Proven Partial Knee in the World. Oxford Partial Knee with Microplasty Instrumentation

Simplifying the Most Clinically Proven Partial Knee in the World. Oxford Partial Knee with Microplasty Instrumentation Simplifying the Most Clinically Proven Partial Knee in the World Oxford Partial Knee with Microplasty Instrumentation 1 Microplasty Instrumentation Innovative, Accurate, Reproducible Microplasty instrumentation

More information

Advantages of Mobile-bearing Total Knee Arthroplasty

Advantages of Mobile-bearing Total Knee Arthroplasty Advantages of Mobile-bearing Total Knee Arthroplasty a report by Douglas A Dennis, MD Clinical Director, Rocky Mountain Musculoskeletal Research Laboratory, Denver Early historical failure of total knee

More information

Zimmer Longevity Highly Cross-linked Polyethylene. Clinical Value Dossier - APAC

Zimmer Longevity Highly Cross-linked Polyethylene. Clinical Value Dossier - APAC Zimmer Longevity Highly Cross-linked Polyethylene Clinical Value Dossier - APAC Table of Contents 1 Executive Summary...................................................... 1 2 Burden of Illness........................................................

More information

EXPERIENCE GPS FOR TOTAL KNEE ARTHROPLASTY DETERMINE YOUR OWN COURSE.

EXPERIENCE GPS FOR TOTAL KNEE ARTHROPLASTY DETERMINE YOUR OWN COURSE. EXPERIENCE GPS FOR TOTAL KNEE ARTHROPLASTY DETERMINE YOUR OWN COURSE. The year 2010 marked Exactech s silver aiversary and 25 years of mobility. As a company founded by an orthopaedic surgeon and a biomedical

More information

Range of Motion of Standard and High-Flexion Posterior Stabilized Total Knee Prostheses A PROSPECTIVE, RANDOMIZED STUDY

Range of Motion of Standard and High-Flexion Posterior Stabilized Total Knee Prostheses A PROSPECTIVE, RANDOMIZED STUDY 1470 COPYRIGHT 2005 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED Range of Motion of Standard and High-Flexion Posterior Stabilized Total Knee Prostheses A PROSPECTIVE, RANDOMIZED STUDY BY YOUNG-HOO

More information

(c) B. Ravi, IIT Bombay 1

(c) B. Ravi, IIT Bombay 1 Collaborative Engineering Evaluation, Testing TKP Safety, TKP System Mechanism FEA Physical testing Fatigue Wear What How Evaluate Testing Judgement Criteria's Testing Conditions Parameters Measured OrthoCAD

More information

Gender Solutions Patello-Femoral Joint System

Gender Solutions Patello-Femoral Joint System Zimmer Biomet is the leading company in partial knee arthroplasty (PKA) 1 with over 40 years experience, offering a comprehensive range of anatomic and innovative solutions. Research shows that surgeons

More information

Kinematic vs. mechanical alignment: What is the difference?

Kinematic vs. mechanical alignment: What is the difference? Kinematic vs. mechanical alignment: What is the difference? In this 4 Questions interview, Stephen M. Howell, MD, explains the potential benefits of 3D alignment during total knee replacement. Introduction

More information

Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008)

Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008) Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008) 8% at 5yrs FU indicated by preliminary study from same

More information

Stefan Rahm MD University Hospital Balgrist

Stefan Rahm MD University Hospital Balgrist Knee Prosthesis Models & Materials, Surgical Techniques and Approaches Stefan Rahm MD University Hospital Balgrist INTRODUCTION VARUS VALGUS 46 Y OLD MALE Pain in the medial compartment left more than

More information

Functional Outcome of Uni-Knee Arthroplasty in Asians with six-year Follow-up

Functional Outcome of Uni-Knee Arthroplasty in Asians with six-year Follow-up Functional Outcome of Uni-Knee Arthroplasty in Asians with six-year Follow-up Ching-Jen Wang, M.D. Department of Orthopedic Surgery Kaohsiung Chang Gung Memorial Hospital Chang Gung University College

More information

The UniSpacer : correcting varus malalignment in medial gonarthrosis

The UniSpacer : correcting varus malalignment in medial gonarthrosis DOI 10.1007/s00264-009-0908-9 ORIGINAL PAPER The UniSpacer : correcting varus malalignment in medial gonarthrosis Michael Clarius & Justus F. Becker & Holger Schmitt & Joern B. Seeger Received: 20 October

More information

Clinical Results of Lateral Unicompartmental Knee Arthroplasty: Minimum 2-Year Follow-up

Clinical Results of Lateral Unicompartmental Knee Arthroplasty: Minimum 2-Year Follow-up Original Article Clinics in Orthopedic Surgery 2016;8:386-392 https://doi.org/10.4055/cios.2016.8.4.386 Clinical Results of Lateral Unicompartmental Knee Arthroplasty: Minimum 2-Year Follow-up Kyung Tae

More information

We carried out weight-bearing video radiological

We carried out weight-bearing video radiological Femoral condylar lift-off in vivo in total knee arthroplasty D. A. Dennis, R. D. Komistek, S. A. Walker, E. J. Cheal, J. B. Stiehl From the Rocky Mountain Musculoskeletal Research Laboratory, Denver, USA

More information

A Kinematic Assessment of Knee Prosthesis from Fluoroscopy Images

A Kinematic Assessment of Knee Prosthesis from Fluoroscopy Images Memoirs of the Faculty of Engineering, Kyushu University, Vol. 68, No. 1, March 2008 A Kinematic Assessment of Knee Prosthesis from Fluoroscopy Images by Mohammad Abrar HOSSAIN *, Michihiko FUKUNAGA and

More information

Polyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data

Polyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data Skeletal Radiol (2003) 32:679 686 DOI 10.1007/s00256-003-0685-2 ARTICLE R. Košak V. Antolič V. Pavlovčič V. Kralj-Iglič I. Milošev G. Vidmar A. Iglič Polyethylene wear in total hip prostheses: the influence

More information

)301( COPYRIGHT 2018 BY THE ARCHIVES OF BONE AND JOINT SURGERY

)301( COPYRIGHT 2018 BY THE ARCHIVES OF BONE AND JOINT SURGERY )301( COPYRIGHT 2018 BY THE ARCHIVES OF BONE AND JOINT SURGERY RESEARCH ARTICLE Early Results of Oxford Mobile Bearing Medial Unicompartmental Knee Replacement (UKR) with the Microplasty Instrumentation:

More information

For Commercial products, please refer to the following policy: Preauthorization via Web-Based Tool for Procedures

For Commercial products, please refer to the following policy: Preauthorization via Web-Based Tool for Procedures Medical Coverage Policy Total Joint Arthroplasty Hip and Knee EFFECTIVE DATE: 08/01/2017 POLICY LAST UPDATED: 06/06/2017 OVERVIEW Joint replacement surgery, also known as arthroplasty, has proved to be

More information

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5.

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5. Surgical Technique Contents Templating and Pre Operative Planning 2 Preparation of the Acetabulum 4 Trial Sizing and Impaction of the Shell 5 Cup Positioning 6 Joint Stability 7 Trial sizing and Impaction

More information