DEVELOPMENT AND CONTROL OF A 3DOF UPPER-LIMB ROBOTIC DEVICE FOR PATIENTS WITH PARETIC LIMB IMPAIRMENT

Size: px
Start display at page:

Download "DEVELOPMENT AND CONTROL OF A 3DOF UPPER-LIMB ROBOTIC DEVICE FOR PATIENTS WITH PARETIC LIMB IMPAIRMENT"

Transcription

1 DEVELOPMENT AND CONTROL OF A 3DOF UPPER-LIMB ROBOTIC DEVICE FOR PATIENTS WITH PARETIC LIMB IMPAIRMENT Sado Fatai, Shahrul N. Sidek, Yusof M. Hazlina, Latif M. Hafiz and Adamu Y. Babawuro Department of Mechatronics Engineering, International Islamic University Malaysia, Malaysia abdfsado1@gmail.com ABSTRACT Rehabilitation therapy after stroke is crucial to helping patients regain as much as possible the use of their paretic limbs. The major challenges, however, in conventional post-stroke rehabilitation therapy is that the therapy is labourintensive, time demanding, and therefore, expensive with consequent reduction in the amount of training session required for optimal therapeutic outcome. In recent times, the use of robotic devices for rehabilitation training has been widely favoured. Robot-assisted rehabilitation therapy is cost-effective, fatigue-free, and has the potential to improve the efficiency of the rehabilitation process. More so, positive outcome of improved motor control abilities for patients undergoing robot-assisted therapy have been widely recorded. This paper presents the development and control of a portable three degree of freedom (3-DOF) end-effector-type robotic device for upper-extremity rehabilitation of paretic stroke patients. The device has three active DOFs consisting of two revolute joints and one prismatic joint (R-R-P) designed to allow 3-dimensional range of motion (ROM) exercise for elbow and shoulder rehabilitation. A novel adaptive hybrid impedance control framework has been developed for the device to allow safe robot-patient dynamic interaction during planned repetitive range of motion exercises and to keep track of patients motor recovery based on an embedded Modified Ashworth Scale assessment criteria. Experimental results performed, using a healthy subject, to test and evaluate the ability of the device to track a planned simple flexion/extension range of motion exercise for the elbow joint showed position and force tracking accuracy to a maximum root mean square error (RMSE) of 0.024m and 0.343N respectively which indicate the possibility of use of the device for real patients. Keywords: upper-extremity, robot-assisted, rehabilitation, hybrid impedance. INTRODUCTION In the last few decades, the increasing cases of upper-extremity disabilities resulting from stroke, spinal cord injuries (SCI), and other related illness have favoured the use of robotic devices to provide assistance and support to patients undergoing rehabilitation therapy[1, 2]. The effectiveness of the devices in extending the therapy session and in providing repetitive exercises aimed at inducing motor plasticity have been widely reported [2, 3]. Robotic devices have the potentials of allowing repeatability and automation of therapeutic procedures with significant improvement on the efficiency of the rehabilitation process [4]. With the increasing demand, however, for robotic devices, the need for portable, lightweight, and safe devices have created new challenges ranging from the design of compact robot mechanical/actuation systems to the development of effective control algorithms. A survey on robotic devices in the recent past, showed that most of the earlier robotic devices and some of the recent commercially available ones are mechanically bulky and suited for use only in rehabilitation centres [5-7]. Besides, most are less autonomous to the extent that they require constant assistance of a human therapist for effective usage, for task specification and for adjustment of therapy [7]. This implies a considerable amount of valuable therapist time in programming the robot, monitoring the patients, analysing data from the robot and also assessing the progress of the patients Krebs et al.[8] and Lum et al.[9] earlier proposed a 3-DOF robotic device that allowed horizontal motion and a 6-DOF Mirror Image Movement Enabler (MIME) robotic device that allowed 3-dimensional spatial motion, respectively, for upper-limb rehabilitation which were generally bulky and suited only for rehabilitation centres. The devices were less autonomous and require the manual monitoring of patients progress by an expert therapist. A 9-DOF cable-driven robotic device was also proposed by [10] based on GENTLE/S system. The device was effectively used for reach and grasp therapy for poststroke patients, but was equally bulky and non-adaptive to patients recovery. A more recent commercially available robotic device referred to as ArmeoPower [11] with 7- DOF for elbow and shoulder rehabilitation and additional DOFs for wrist and fingers flexion/extension rehabilitation exercise which has been successfully used in many rehabilitation centres is also seen to be largely stationary and not lightweight. As development of robotic devices for rehabilitation exercises continue, the need for portable and lightweight devices capable of autonomous guidance and monitoring of patients progress therefore remains a critical driving factor of consideration [7]. In this study, we report the development and control of a 3-DOF end-effector robotic device for autonomous rehabilitation of the shoulder and elbow region of patients with paretic upper-limb impairment. The device is made portable and lightweight to allow flexibility of usage and for possible use at homes. In addition, a novel adaptive control framework reported in [12] is developed for the device to allow safe robot

2 patients dynamic interaction and effective tracking of a planned range of motion exercise, and independent monitoring of patients physical recovery progress. The rest of the paper is organized as follows: section 2 reports on the kinematics and dynamic modelling of the robotic device, section 3 presents the features of the developed novel robot control framework, section 4 presents some experimental results and discussion and section 5 concludes the paper with proposed future works. ROBOT MODELING Forward kinematics The robotic device has three active DOF arranged in a revolute-revolute-prismatic (R-R-P) configuration with only two link lengths, l_2 and l_(3 ) parameters (to simplify the inverse kinematics) as shown in Figure-1 and Figure-2. The three joints allow for the rehabilitation (flexion, extension, abduction, and adduction) of the elbow and shoulder of the patients in three dimensional space. The relationship between the joints variables and the position and orientation of the robot s end-effector is derived from the four Denavit-Hartenberg (DH) parameters given in Table-1. The four parameters a_i, _i, d_i, θ_i are generally known as the link length, link twist, link offset, and joint angles respectively[13]. The link transformation matrices which relates any two successive frames attached to the joints are derived from the homogenous transformation matrice [13] given by Equation (1). where s and c stands for sine and cosine of the angles. The homogenous transformation matrix relating the end-effector frame (,, to the base frame (,,, see Figure-2, is therefore obtained by multiplying the individual link transformation matrices as shown in Equation (2). Finally, the forward kinematics equation derived from Equation (2) is expressed as, (1) (2) (3) Table-1. The DH-parameters. Figure-1. The Mechanical structure of the 3-DOF robotic device. Inverse kinematics We derive the inverse kinematics of the rehabilitation robot in order to determine the joint angles or variables in terms of the end-effector position and orientation. It is however more computationally demanding than the forward kinematics analysis due to the non-linearity in the forward kinematics equation. Since the robotic device is designed with two axis intersecting, the inverse kinematic analysis becomes fairly easy with a possible closed form solution. Observing that the forward kinematics equations are transcendental, the following substitutions are made where denotes the tangent of any half angle, and =,. By substituting Equation (4) in Equation (3) and solving for the joint variables = [ ], we obtain the inverse kinematics equations for the robotic system as (4) Figure-2. The robot kinematic model

3 (10) The solution is however not unique since there is the possibility of different configurations (some unreachable) for a given solution. We therefore avoids the use of the inverse kinematics for the controller development as will be seen in the next section. Velocity kinematics: The Jacobian The velocity kinematics equation relates the linear velocity v and angular velocity q of the end-effector to the joint velocities of the robotic device. Similar to the forward kinematics equation, it defines a function which maps between the Cartesian space and joint space. This relationship is expressed by the Jacobian time-varying linear transformation given by Equation (6). Taking the first derivative of the forward kinematics equations (refer to Equation (3)) and using the techniques described in [13], we obtain the Jacobian given as Where, The robot dynamics To obtain the robot dynamic model, we use the Lagrange equation of motion [14] for a conservative system given by where q is the 3x1 vector of generalized joint coordinates, τ is the 3x1 vector of generalized input actuator forces/torque, and L the Lagrangian. The Lagrangian is given by the expression, where is the kinematic energy of the robotic system, and is the potential energy of the system. Using Equation (8)-(9), we derive the overall dynamic model of the robotic device expressed by Equation (10) which is a requisite for developing the control algorithm. (5) (6) (7) (8) (9) represents the robot mass matrix, represents the actuator inertia,, represents the Coriolis and centrifugal terms, and represents the gravity term. Since the joint frictions are small for the robotic device, we assumed these components to be zero in the robot dynamic model. CONTROL ARCHITECTURE Sensors and actuation system A DC motor coupled at the prismatic joint, at frame (,,, provide linear actuation by means of a lead screw mechanism to allow flexion and extension of the elbow joint along 90 o to the supine (other configurations are possible when actuation is done in conjunction with other joints actuation), see Figure-1 and Figure-2. The DC motor is equipped with incremental encoder which gives precise measurement of linear position, velocity and acceleration of the end-effector. The other two revolute joints at frame (,, : denoted base frame, and frame (,, are actuated separately by means of brushless DC motors with coupled incremental encoders to allow precise measurement of joint angular positions, velocities, and accelerations along their respective frames. Actuation of the base frame allows for rotational motion along the vertical axis ( ) which assist shoulder reshabilitation, and in conjunction with the second joint actuation allows abduction and adduction ROM exercises of the shoulder. The actuation of the second joint at frame (,, allows tilting or rotational motion along the horizontal axis (. In addition, contact force measurement is achieved at the end-effector location by means of a 3-axis load cell custom fabricated into the end-effector (see Figure-1). Control hardware A hardware control unit designed for the robotic device consists of a central microcontroller (Arduino Mega), see Figure-3. The microcontroller provides interface between the Matlab/Simulink control model on the PC side and the robotic device. The unit also consists of motor drivers which provide interface between the microcontroller and the motors, including the PWM to analog voltage converters which outputs analog voltage corresponding to the control torque on the PC side for tuning of the motor speed. Figure-3. Control hardware

4 Feedback control linearization strategy Since the robot interacts with an environment (patients or human subjects in this case) with 3x1 joint space reaction torques given by τ_e, the robot-patient dynamic motion given by Equation (10) is modified as (11) The vector of environment reaction torque in joint space is related to the task space contact force by the Jacobian matrix given by Equation (11) can thus be rewritten as (12) (13) where represents the sum of the total robot mass matrix and the actuators inertia ( + ). We achieved the control objective by linearizing the non-linear dynamic equation given by Equation (13) using the concept of inner/outer loop control strategy [15] as shown in Figure-4. The inner loop represents the nonlinear feedback linearization or inverse dynamics control (given by Equation 14) and the outer-loop represent a hybrid impedance control strategies (given by Equation 15). (14) (15) where,, and represent the robot virtual impedance parameters (mass, damping and stiffness factors), represents the selector matrix for force and position control, represents the reference force, and represents the reference position. autoregressive exogenous recursive polynomial model estimator (ARX-RPME) as reported in earlier works [12], see Figure-5 for the adaptive framework. The RPME implements an ARX model of the upper-limb to estimate the mechanical impedance parameters of the limb as a measure of muscle tone. The estimated parameters serve as the inputs to the HA framework by means of which the automata accomplish two tasks. First, allows transition between its seven-state ( 6 ) automaton (including a default transition state) based on the level of muscle tone (impairment) as classified by the Modified Ashworth Scale (MAS) assessment criteria [12, 16], and second, allows switching, concurrently, of task space between force and position control, based on the changing upperlimb mechanical impedance and the interpretation of the duality principle[17, 18]. Figure-6 gives the overall control framework. Figure-5. The ARX-RPME and the HA model. RESULTS As a preliminary experimentation on the device, we tested its ability to follow a planned reference ROM trajectory of the end-effector, and also on its ability to maintain a certain reference interactive force at the endeffector/human contact point. Figure-4. Inner and outer loop control strategy. Adaptive control We developed an adaptive control framework (to allow task space switching of position and force control as well as monitoring of patients physical recovery progress) using a novel hybrid automata (HA) framework and an Figure-6. The overall adaptive control framework

5 Trajectory generation and position control To actualize the position control, we recorded (using a healthy subject) a simple elbow flexion range of motion exercise of the end-effector along the positive,, direction and derive the reference trajectory (from the data) using a second order polynomial (see Equation 16). Figure-7 shows the position tracking results for the actual and desired position trajectory of the end-effector along the x,y,z directions with RMSE of 0.023, 0.024, and 0.013m respectively. (16) Force tracking To test the robotic device/controller on the ability to maintain certain amount of contact force (exerted by the end-effector on the environment/subject s limb), we planned a reference force trajectory using a 4th order polynomial such that the robot end-effector is made to exert force (by actuating the individual joints) on the subject s upper-limb steadily from zero to a maximum of 50N, and thereafter reduce the exerted force gradually to zero over a computer time of 21s. The generated reference trajectory along the positive z-direction of the end-effector is given by Equation (17). Figure-8 shows the force tracking results by the controller along the z-direction with RMSE of 0.343N. (17) Discussion As can be seen from the Figure-8, there is fairly good tracking of the reference force trajectory due to noise from the custom-made force sensor. As a modification, for future studies, a signal conditioning circuit will be adopted therefore to eliminate noisy signals. Regarding stability, no analysis for controller stability is given in this paper since the stability of control frameworks based on the standard computed torque control, or inverse dynamic control, have been generally established. However, mechanical stability is guaranteed using worm gears and a firm tripod support. Figure-7. Experimental results for position tracking in the (a) z-direction (b) y-direction and (z) x-direction

6 biological engineering & computing, Vol. 45, pp [5] R. Bogue "Exoskeletons and robotic prosthetics: a review of recent developments," Industrial Robot: An International Journal, Vol. 36, pp , [6] B. Dellon and Y. Matsuoka "Prosthetics, exoskeletons, and rehabilitation," IEEE Robotics and Automation magazine, Vol. 14, p. 30. [7] I. Díaz, J. J. Gil and E. Sánchez "Lower-limb robotic rehabilitation: literature review and challenges," Journal of Robotics, Vol Figure-8. Experimental results for force tracking along the z-direction. CONCLUSIONS A new portable 3-DOF end-effector robotic device for autonomous rehabilitation of the shoulder and elbow region of patients with paretic upper-limb impairment has been presented. A novel adaptive control framework developed in previous study, by the Author, is adopted to allow safe robot-patients dynamic interaction and effective tracking of a planned range of motion exercise, as well as independent monitoring of patients physical recovery progress. Experimental results have shown the feasibility of use of the device for trajectory motion tracking of a reference ROM exercise. Future works however includes experimentation to test the adaptability of the robot control scheme to changing upper-limb mechanical impedance and final experimentation on actual patients. REFERENCES [1] M. J. O'Donnell, D. Xavier, L. Liu, H. Zhang, S. L. Chin and P. Rao-Melacini et al "Risk factors for ischaemic and intracerebral haemorrhagic stroke in 22 countries (the INTERSTROKE study): a casecontrol study," The Lancet, Vol. 376, pp [2] P. Maciejasz, J. Eschweiler, K. Gerlach-Hahn, A. Jansen-Troy, and S. Leonhardt "A survey on robotic devices for upper limb rehabilitation," Journal of neuroengineering and rehabilitation, Vol. 11. [3] H. S. Lo and S. Q. Xie "Exoskeleton robots for upper-limb rehabilitation: state of the art and future prospects," Medical engineering & physics, Vol. 34, pp [4] T. Nef, M. Mihelj and R. Riener "ARMin: a robot for patient-cooperative arm therapy," Medical & [8] H. I. Krebs, M. Ferraro, S. P. Buerger, M. J. Newbery, A. Makiyama, M. Sandmann. et al "Rehabilitation robotics: pilot trial of a spatial extension for MIT-Manus," Journal of NeuroEngineering and Rehabilitation, Vol. 1, pp. 5, [9] P. S. Lum, C. G. Burgar, M. Van der Loos, P. C. Shor, M. Majmundar and R. Yap "MIME robotic device for upper-limb neurorehabilitation in subacute stroke subjects: A follow-up study," Journal of rehabilitation research and development, Vol. 43, p [10] R. C. Loureiro and W. S. Harwin "Reach & grasp therapy: design and control of a 9-DOF robotic neuro-rehabilitation system," in Rehabilitation Robotics, ICORR IEEE 10th International Conference on, pp [11] T. Nef, M. Guidali, V. Klamroth-Marganska and R. Riener "ARMin-exoskeleton robot for stroke rehabilitation," in World Congress on Medical Physics and Biomedical Engineering, September 7-12, 2009, Munich, Germany, pp [12] F. Sado, S. N. Sidek and H. M. Yusof "Adaptive hybrid impedance control for a 3DOF upper limb rehabilitation robot using hybrid automata," in Biomedical Engineering and Sciences (IECBES), 2014 IEEE Conference on, pp [13] M. W. Spong, S. Hutchinson and M. Vidyasagar Robot modeling and control: John Wiley & Sons New York. [14] S. B. Niku Introduction to robotics: analysis, systems, applications Vol. 7: Prentice Hall New Jersey. [15] R. Anderson and M. W. Spong "Hybrid impedance control of robotic manipulators," Robotics and Automation, IEEE Journal of, vol. 4, pp

7 [16] R. W. Bohannon and M. B. Smith "Interrater reliability of a modified Ashworth scale of muscle spasticity," Physical therapy, Vol. 67, pp [17] F. L. Lewis D. M. Dawson and C. T. Abdallah Robot manipulator control: theory and practice: CRC Press. [18] M. W. Spong, S. Hutchinson and M. Vidyasagar Robot modeling and control: John Wiley & Sons New York

Navigator: 2 Degree of Freedom Robotics Hand Rehabilitation Device

Navigator: 2 Degree of Freedom Robotics Hand Rehabilitation Device Navigator: 2 Degree of Freedom Robotics Hand Rehabilitation Device Design Team Ray Adler, Katherine Bausemer, Joseph Gonsalves, Patrick Murphy, Kevin Thompson Design Advisors Prof. Constantinos Mavroidis,

More information

Design and Control of a Robotic Assistive Device for Hand Rehabilitation (RAD-HR)

Design and Control of a Robotic Assistive Device for Hand Rehabilitation (RAD-HR) Proceedings of the 3 rd International Conference on Control, Dynamic Systems, and Robotics (CDSR 16) Ottawa, Canada May 9 10, 2016 Paper No. 109 DOI: 10.11159/cdsr16.109 Design and Control of a Robotic

More information

Using the Kinect to Limit Abnormal Kinematics and Compensation Strategies During Therapy with End Effector Robots

Using the Kinect to Limit Abnormal Kinematics and Compensation Strategies During Therapy with End Effector Robots 2013 IEEE International Conference on Rehabilitation Robotics June 24-26, 2013 Seattle, Washington USA Using the Kinect to Limit Abnormal Kinematics and Compensation Strategies During Therapy with End

More information

Electromyogram-Assisted Upper Limb Rehabilitation Device

Electromyogram-Assisted Upper Limb Rehabilitation Device Electromyogram-Assisted Upper Limb Rehabilitation Device Mikhail C. Carag, Adrian Joseph M. Garcia, Kathleen Mae S. Iniguez, Mikki Mariah C. Tan, Arthur Pius P. Santiago* Manufacturing Engineering and

More information

ABSTRACT: Stroke is one of the prominent causes of disability in the

ABSTRACT: Stroke is one of the prominent causes of disability in the Design, Fabrication and Control of Upper Limb Rehabilitation Robot Prototype for Stroke Patients DESIGN, FABRICATION AND CONTROL OF UPPER LIMB REHABILITATION ROBOT PROTOTYPE FOR STROKE PATIENTS S. Mohamaddan

More information

Mechanical Design of a Distal Arm Exoskeleton for Stroke and Spinal Cord Injury Rehabilitation

Mechanical Design of a Distal Arm Exoskeleton for Stroke and Spinal Cord Injury Rehabilitation 2011 IEEE International Conference on Rehabilitation Robotics Rehab Week Zurich, ETH Zurich Science City, Switzerland, June 29 - July 1, 2011 Mechanical Design of a Distal Arm Exoskeleton for Stroke and

More information

Design and Control of an Exoskeletal Rehabilitation Device for Stroke and Spinal Cord Injury Patients

Design and Control of an Exoskeletal Rehabilitation Device for Stroke and Spinal Cord Injury Patients RICE UNIVERSITY Design and Control of an Exoskeletal Rehabilitation Device for Stroke and Spinal Cord Injury Patients by Ali Utku Pehlivan A THESIS SUBMITTED IN PARTIAL FULFILLMENT OF THE REQUIREMENTS

More information

Design and Modeling of an Upper Extremity Exoskeleton

Design and Modeling of an Upper Extremity Exoskeleton Design and Modeling of an Upper Extremity Exoskeleton Salam Moubarak, Minh Tu Pham, Thomas Pajdla, Tanneguy Redarce To cite this version: Salam Moubarak, Minh Tu Pham, Thomas Pajdla, Tanneguy Redarce.

More information

Right-Arm Robotic-Aided-Therapy with the Light-Exoskeleton: A General Overview

Right-Arm Robotic-Aided-Therapy with the Light-Exoskeleton: A General Overview Right-Arm Robotic-Aided-Therapy with the Light-Exoskeleton: A General Overview Luis I. Lugo-Villeda, Antonio Frisoli, Edoardo Sotgiu, Giovanni Greco, and Massimo Bergamasco Perceptual Robotics, Scuola

More information

NEUROLOGICAL injury is a leading cause of permanent

NEUROLOGICAL injury is a leading cause of permanent 286 IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, VOL. 16, NO. 3, JUNE 2008 Optimizing Compliant, Model-Based Robotic Assistance to Promote Neurorehabilitation Eric T. Wolbrecht,

More information

AN EXOSKELETAL ROBOT MANIPULATOR FOR LOWER LIMBS REHABILITATION

AN EXOSKELETAL ROBOT MANIPULATOR FOR LOWER LIMBS REHABILITATION The 9th Mechatronics Forum International Conference Aug. 30 - Sept. 1, 2004, Ankara, Turkey AN EXOSKELETAL ROBOT MANIPULATOR FOR LOWER LIMBS REHABILITATION Erhan AKDOĞAN 1, M. Arif ADLI 2 1 Marmara University,

More information

A NOVEL MULTI-DOF EXOSKELETON ROBOT FOR UPPER LIMB REHABILITATION

A NOVEL MULTI-DOF EXOSKELETON ROBOT FOR UPPER LIMB REHABILITATION Journal of Mechanics in Medicine and Biology Vol. 16, No. 8 (2016) 1640023 (11 pages) c World Scientific Publishing Company DOI: 10.1142/S0219519416400236 A NOVEL MULTI-DOF EXOSKELETON ROBOT FOR UPPER

More information

Rehabilitation (Movement Therapy) Robots

Rehabilitation (Movement Therapy) Robots Rehabilitation (Movement Therapy) Robots Allison M. Okamura Associate Professor Department of Mechanical Engineering Stanford University Portions of this material provided by H. F. Machiel Van der Loos

More information

Fast Simulation of Arm Dynamics for Real-time, Userin-the-loop. Ed Chadwick Keele University Staffordshire, UK.

Fast Simulation of Arm Dynamics for Real-time, Userin-the-loop. Ed Chadwick Keele University Staffordshire, UK. Fast Simulation of Arm Dynamics for Real-time, Userin-the-loop Control Applications Ed Chadwick Keele University Staffordshire, UK. Acknowledgements Dimitra Blana, Keele University, Staffordshire, UK.

More information

AT89C51 Microcontroller Based Control Model for Hybrid Assistive Limb (Knee )

AT89C51 Microcontroller Based Control Model for Hybrid Assistive Limb (Knee ) From the SelectedWorks of suresh L Winter January 10, 2013 AT89C51 Microcontroller Based Control Model for Hybrid Assistive Limb (Knee ) suresh L Available at: https://works.bepress.com/suresh_l/2/ AT89C51

More information

Mechanical Design of RiceWrist-S: a Forearm-Wrist Exoskeleton for Stroke and Spinal Cord Injury Rehabilitation

Mechanical Design of RiceWrist-S: a Forearm-Wrist Exoskeleton for Stroke and Spinal Cord Injury Rehabilitation The Fourth IEEE RAS/EMBS International Conference on Biomedical Robotics and Biomechatronics Roma, Italy. June 24-27, 2012 Mechanical Design of RiceWrist-S: a Forearm-Wrist Exoskeleton for Stroke and Spinal

More information

Study on the control of variable resistance for isokinetic muscle training system

Study on the control of variable resistance for isokinetic muscle training system Technology and Health Care 25 (2017) S45 S52 DOI 10.3233/THC-171305 IOS Press S45 Study on the control of variable resistance for isokinetic muscle training system Lan Wang, Zhenyuan Zhang, Yi Yu and Guangwei

More information

Comparison of Robot-Assisted Reaching to Free Reaching in Promoting Recovery From Chronic Stroke

Comparison of Robot-Assisted Reaching to Free Reaching in Promoting Recovery From Chronic Stroke Comparison of Robot-Assisted Reaching to Free Reaching in Promoting Recovery From Chronic Stroke Leonard E. Kahn, M.S. 1,2, Michele Averbuch, P.T. 1, W. Zev Rymer, M.D., Ph.D. 1,2, David J. Reinkensmeyer,

More information

3-DOF Parallel robotics System for Foot Drop therapy using Arduino

3-DOF Parallel robotics System for Foot Drop therapy using Arduino 3-DOF Parallel robotics System for Foot Drop therapy using Arduino Prof. Dr. Bahaa I Kazem, Assist.Prof.Dr Ameer Morad, Kamal Mohammed Hasan Baghdad University, Baghdad, Iraq Abstract This paper discusses

More information

A ROBOTIC SYSTEM FOR UPPER-LIMB EXERCISES TO PROMOTE RECOVERY OF MOTOR FUNCTION FOLLOWING STROKE

A ROBOTIC SYSTEM FOR UPPER-LIMB EXERCISES TO PROMOTE RECOVERY OF MOTOR FUNCTION FOLLOWING STROKE A ROBOTIC SYSTEM FOR UPPER-LIMB EXERCISES TO PROMOTE RECOVERY OF MOTOR FUNCTION FOLLOWING STROKE Peter S. Lum 1,2, Machiel Van der Loos 1,2, Peggy Shor 1, Charles G. Burgar 1,2 1 Rehab R&D Center, VA Palo

More information

A Human Machine Interface Design to Control an Intelligent Rehabilitation Robot System

A Human Machine Interface Design to Control an Intelligent Rehabilitation Robot System 247 Chapter 13 A Human Machine Interface Design to Control an Intelligent Rehabilitation Robot System Erhan Akdoğan Marmara University, Turkey M. Arif Adlı Marmara University, Turkey Ertuğrul Taçgın Marmara

More information

Development of a 6DOF Exoskeleton Robot for Human Upper-Limb Motion Assist

Development of a 6DOF Exoskeleton Robot for Human Upper-Limb Motion Assist Development of a 6DOF Exoskeleton Robot for Human Upper-Limb Motion Assist R. A. R. C. Gopura + and Kazuo Kiguchi ++ Department of Advanced Systems Control Engineering Saga University, Saga, Japan 840-8502

More information

EMG-Driven Human Model for Orthosis Control

EMG-Driven Human Model for Orthosis Control EMG-Driven Human Model for Orthosis Control Christian Fleischer, Günter Hommel Institute for Computer Engineering and Microelectronics Berlin University of Technology, Germany {fleischer, hommel}@cs.tu-berlin.de

More information

Hand of Hope. For hand rehabilitation. Member of Vincent Medical Holdings Limited

Hand of Hope. For hand rehabilitation. Member of Vincent Medical Holdings Limited Hand of Hope For hand rehabilitation Member of Vincent Medical Holdings Limited Over 17 Million people worldwide suffer a stroke each year A stroke is the largest cause of a disability with half of all

More information

Estimation of the Upper Limb Lifting Movement Under Varying Weight and Movement Speed

Estimation of the Upper Limb Lifting Movement Under Varying Weight and Movement Speed 1 Sungyoon Lee, 1 Jaesung Oh, 1 Youngwon Kim, 1 Minsuk Kwon * Jaehyo Kim 1 Department of mechanical & control engineering, Handong University, qlfhlxhl@nate.com * Department of mechanical & control engineering,

More information

STROKE is a leading cause of disability in Canada, MEDARM: a rehabilitation robot with 5DOF at the shoulder complex

STROKE is a leading cause of disability in Canada, MEDARM: a rehabilitation robot with 5DOF at the shoulder complex MEDARM: a rehabilitation robot with 5DOF at the shoulder complex Stephen J. Ball, Ian E. Brown and Stephen H. Scott Abstract A key approach for reducing motor impairment and regaining independence after

More information

ECHORD call1 experiment MAAT

ECHORD call1 experiment MAAT ECHORD call1 experiment MAAT Multimodal interfaces to improve therapeutic outcomes in robot-assisted rehabilitation Loredana Zollo, Antonino Salerno, Eugenia Papaleo, Eugenio Guglielmelli (1) Carlos Pérez,

More information

A ROBOT FOR WRIST REHABILITATION

A ROBOT FOR WRIST REHABILITATION A ROBOT FOR WRIST REHABILITATION Dustin J. Williams 1, Hermano Igo Krebs 1, Neville Hogan 1 1 Mechanical Engineering Dept., Massachusetts Institute of Technology, Cambridge, USA Abstract -In 1991, a novel

More information

MINERVA MEDICA COPYRIGHT

MINERVA MEDICA COPYRIGHT EUR J PHYS REHABIL MED 2008;44:431-5 Robot-assisted therapy for neuromuscular training of sub-acute stroke patients. A feasibility study Aim. Several studies have described the contribution of robotics

More information

sensors ISSN

sensors ISSN Sensors 2012, 12, 16046-16059; doi:10.3390/s121216046 Article OPEN ACCESS sensors ISSN 1424-8220 www.mdpi.com/journal/sensors Implementation of Human-Machine Synchronization Control for Active Rehabilitation

More information

Augmented reflection technology for stroke rehabilitation a clinical feasibility study

Augmented reflection technology for stroke rehabilitation a clinical feasibility study Augmented reflection technology for stroke rehabilitation a clinical feasibility study S Hoermann 1, L Hale 2, S J Winser 2, H Regenbrecht 1 1 Department of Information Science, 2 School of Physiotherapy,

More information

A Musculoskeletal Model-based Assistance-As-Needed Paradigm for Assistive Robotics

A Musculoskeletal Model-based Assistance-As-Needed Paradigm for Assistive Robotics A Musculoskeletal Model-based Assistance-As-Needed Paradigm for Assistive Robotics Marc G. Carmichael Submitted in fulfillment of the requirement for the degree of Doctor of Philosophy 2013 The Faculty

More information

Research Article Robotic Upper Limb Rehabilitation after Acute Stroke by NeReBot: Evaluation of Treatment Costs

Research Article Robotic Upper Limb Rehabilitation after Acute Stroke by NeReBot: Evaluation of Treatment Costs BioMed Research International, Article ID 265634, 5 pages http://dx.doi.org/10.1155/2014/265634 Research Article Robotic Upper Limb Rehabilitation after Acute Stroke by NeReBot: Evaluation of Treatment

More information

Development of Mirror Image Motion System with semg for Shoulder Rehabilitation of Post-stroke Hemiplegic Patients

Development of Mirror Image Motion System with semg for Shoulder Rehabilitation of Post-stroke Hemiplegic Patients INTERNATIONAL JOURNAL OF PRECISION ENGINEERING AND MANUFACTURING Vol. 13, No. 8, pp. 1473-1479 AUGUST 2012 / 1473 DOI: 10.1007/s12541-012-0194-0 Development of Mirror Image Motion System with semg for

More information

Wearable Hand Rehabilitation Robot Capable of Hand Function Assistance in Stroke Survivors*

Wearable Hand Rehabilitation Robot Capable of Hand Function Assistance in Stroke Survivors* The Fourth IEEE RAS/EMBS International Conference on Biomedical Robotics and Biomechatronics Roma, Italy. June 4-7, 0 Wearable Hand Rehabilitation Robot Capable of Hand Function Assistance in Stroke Survivors*

More information

Spatial and Temporal Movement Characteristics after Robotic Training of Arm and Hand: A Case Study of a Person with Incomplete Spinal Cord Injury

Spatial and Temporal Movement Characteristics after Robotic Training of Arm and Hand: A Case Study of a Person with Incomplete Spinal Cord Injury Spatial and Temporal Movement Characteristics after Robotic Training of Arm and Hand: A Case Study of a Person with Incomplete Spinal Cord Injury D.P. Eng Dept of Mechanical Engineering and Materials Science

More information

REHABILITATION IN VIRTUAL REALITY VAST.REHAB IS DESIGNED FOR PHYSICAL, OCCUPATIONAL AND COGNITIVE THERAPY

REHABILITATION IN VIRTUAL REALITY VAST.REHAB IS DESIGNED FOR PHYSICAL, OCCUPATIONAL AND COGNITIVE THERAPY VAST.REHAB IS DESIGNED FOR PHYSICAL, OCCUPATIONAL AND COGNITIVE THERAPY We create a fully-featured virtual reality exercise and diagnostic system with the flexibility to work for everyone from small physiotherapy

More information

SUEFUL-7: A 7DOF Upper-Limb Exoskeleton Robot with Muscle-Model-Oriented EMG-Based Control

SUEFUL-7: A 7DOF Upper-Limb Exoskeleton Robot with Muscle-Model-Oriented EMG-Based Control The 2009 IEEE/RSJ International Conference on Intelligent Robots and Systems October 11-15, 2009 St. Louis, USA SUEFUL-7: A 7DOF Upper-Limb Exoskeleton Robot with Muscle-Model-Oriented EMG-Based Control

More information

Expert System-Based Post-Stroke Robotic Rehabilitation for Hemiparetic Arm

Expert System-Based Post-Stroke Robotic Rehabilitation for Hemiparetic Arm Expert System-Based Post-Stroke Robotic Rehabilitation for Hemiparetic Arm Pradeep Natarajan Department of EECS University of Kansas Outline 2 Introduction Stroke Rehabilitation Robotics Expert Systems

More information

Maximal isokinetic and isometric muscle strength of major muscle groups related to age, body weight, height, and sex in 178 healthy subjects

Maximal isokinetic and isometric muscle strength of major muscle groups related to age, body weight, height, and sex in 178 healthy subjects Maximal isokinetic and isometric muscle strength of major muscle groups related to age, body weight, height, and sex in 178 healthy subjects Test protocol Muscle test procedures. Prior to each test participants

More information

Freebal: dedicated gravity compensation for the upper extremities

Freebal: dedicated gravity compensation for the upper extremities Proceedings of the 2007 IEEE 10th International Conference on Rehabilitation Robotics, June 12-15, Noordwijk, The Netherlands Freebal: dedicated gravity compensation for the upper extremities Amo H.A.

More information

Learning Classifier Systems (LCS/XCSF)

Learning Classifier Systems (LCS/XCSF) Context-Dependent Predictions and Cognitive Arm Control with XCSF Learning Classifier Systems (LCS/XCSF) Laurentius Florentin Gruber Seminar aus Künstlicher Intelligenz WS 2015/16 Professor Johannes Fürnkranz

More information

Verification of Immediate Effect on the Motor Function of a Plegic Upper Limb after Stroke by using UR-System 2.2

Verification of Immediate Effect on the Motor Function of a Plegic Upper Limb after Stroke by using UR-System 2.2 Verification of Immediate Effect on the Motor Function of a Plegic Upper Limb after Stroke by using UR-System. Hiroki Sugiyama,a, Hitomi Hattori,b, Ryosuke Takeichi,c, Yoshifumi Morita,d, Hirofumi Tanabe,e

More information

DESIGN AND IMPLEMENTATION OF A MECHATRONIC SYSTEM FOR LOWER LIMB MEDICAL REHABILITATION

DESIGN AND IMPLEMENTATION OF A MECHATRONIC SYSTEM FOR LOWER LIMB MEDICAL REHABILITATION International Journal of Modern Manufacturing Technologies ISSN 2067 3604, Vol. IV, No. 2 / 2012 17 DESIGN AND IMPLEMENTATION OF A MECHATRONIC SYSTEM FOR LOWER LIMB MEDICAL REHABILITATION Ana-Maria Amancea

More information

Design considerations for an active soft orthotic system for shoulder rehabilitation

Design considerations for an active soft orthotic system for shoulder rehabilitation Design considerations for an active soft orthotic system for shoulder rehabilitation The Harvard community has made this article openly available. Please share how this access benefits you. Your story

More information

Development of an Arm Rehabilitation System with Different Control Approaches

Development of an Arm Rehabilitation System with Different Control Approaches A Multidimensional Decision Making Model for Information Technology Investments in Universities Development of an Arm Rehabilitation System with Different Control Approaches R. Ghazali *1, M.S.M. Aris

More information

Unilateral and Bilateral Rehabilitation of the Upper Limb Following Stroke via an Exoskeleton

Unilateral and Bilateral Rehabilitation of the Upper Limb Following Stroke via an Exoskeleton Unilateral and Bilateral Rehabilitation of the Upper Limb Following Stroke via an Exoskeleton Jacob Rosen, Dejan Milutinović, Levi M. Miller, Matt Simkins, Hyunchul Kim, and Zhi Li Abstract Recent studies

More information

rapael smart rehab solution

rapael smart rehab solution rapael smart rehab solution RAPAEL Smart Rehabilitation concept Intensive Repetitive Task-oriented Learning Schedule Algorithm for rehabilitation Individualized adaptive training RAPAEL Smart Rehabilitation

More information

Control principles in upper-limb prostheses

Control principles in upper-limb prostheses Control principles in upper-limb prostheses electromyographic (EMG) signals generated by muscle contractions electroneurographic (ENG) signals interface with the peripheral nervous system (PNS) interface

More information

IMECE THE RiceWrist: A DISTAL UPPER EXTREMITY REHABILITATION ROBOT FOR STROKE THERAPY

IMECE THE RiceWrist: A DISTAL UPPER EXTREMITY REHABILITATION ROBOT FOR STROKE THERAPY Proceedings of IMECE Proceedings of IMECE26 November 5-1, 26, Chicago, Illinois, USA IMECE26-1613 THE RiceWrist: A DISTAL UPPER EXTREMITY REHABILITATION ROBOT FOR STROKE THERAPY Marcia K. O Malley Alan

More information

Five-Fingered Assistive Hand with Mechanical Compliance of Human Finger

Five-Fingered Assistive Hand with Mechanical Compliance of Human Finger 28 IEEE International Conference on Robotics and Automation Pasadena, CA, USA, May 19-23, 28 Five-Fingered Assistive Hand with Mechanical Compliance of Human Finger Yasuhisa Hasegawa, Yasuyuki Mikami,

More information

A Computational Framework for Quantitative Evaluation of Movement during Rehabilitation

A Computational Framework for Quantitative Evaluation of Movement during Rehabilitation A Computational Framework for Quantitative Evaluation of Movement during Rehabilitation Yinpeng Chen a, Margaret Duff a,b, Nicole Lehrer a, Hari Sundaram a, Jiping He b, Steven L. Wolf c and Thanassis

More information

FUSE TECHNICAL REPORT

FUSE TECHNICAL REPORT FUSE TECHNICAL REPORT 1 / 16 Contents Page 3 Page 4 Page 8 Page 10 Page 13 Page 16 Introduction FUSE Accuracy Validation Testing LBD Risk Score Model Details FUSE Risk Score Implementation Details FUSE

More information

rapael smart rehab solution

rapael smart rehab solution rapael smart rehab solution RAPAEL Smart Rehabilitation concept Intensive Repetitive Task-oriented Learning Schedule Algorithm for rehabilitation Individualized adaptive training RAPAEL Smart Rehabilitation

More information

KINEMATIC ANALYSIS OF ELBOW REHABILITATION EQUIPMENT

KINEMATIC ANALYSIS OF ELBOW REHABILITATION EQUIPMENT Bulletin of the Transilvania University of Braşov Vol. 10 (59) No. 2-2017 Series I: Engineering Sciences KINEMATIC ANALYSIS OF ELBOW REHABILITATION EQUIPMENT G. VETRICE 1 A. DEACONESCU 1 Abstract: The

More information

WE propose the development of a new class of rehabilitation

WE propose the development of a new class of rehabilitation 1 A Dual Input Device for Self-Assisted Control of a Virtual Pendulum Kari A. Danek, R. Brent Gillespie, J. Wayne Aldridge, D.P. Ferris, J.W. Grizzle Abstract We are proposing a new approach to rehabilitation

More information

Copyright Warning & Restrictions

Copyright Warning & Restrictions Copyright Warning & Restrictions The copyright law of the United States (Title 17, United States Code) governs the making of photocopies or other reproductions of copyrighted material. Under certain conditions

More information

Q: What is the relationship between muscle forces and EMG data that we have collected?

Q: What is the relationship between muscle forces and EMG data that we have collected? FAQs ABOUT OPENSIM Q: What is the relationship between muscle forces and EMG data that we have collected? A: Muscle models in OpenSim generate force based on three parameters: activation, muscle fiber

More information

of the joint coordination in reach-to-grasp movements.

of the joint coordination in reach-to-grasp movements. The Joint Coordination in Reach-to-grasp Movements Zhi Li, Kierstin Gray, Jay Ryan Roldan, Dejan Milutinović and Jacob Rosen Abstract Reach-to-grasp movements are widely observed in activities of daily

More information

EMG-Based Control of an Exoskeleton Robot for Human Forearm and Wrist Motion Assist

EMG-Based Control of an Exoskeleton Robot for Human Forearm and Wrist Motion Assist 2008 IEEE International Conference on Robotics and Automation Pasadena, CA, USA, May 19-23, 2008 EMG-Based Control of an Exoskeleton Robot for Human Forearm and Wrist Motion Assist Ranathunga Arachchilage

More information

Thesis Rehabilitation robotics (RIA) Robotics for Bioengineering Forefront research at PRISMA Lab and ICAROS Center

Thesis Rehabilitation robotics (RIA) Robotics for Bioengineering Forefront research at PRISMA Lab and ICAROS Center Thesis Rehabilitation robotics (RIA) RIA-1. Mechanical design of sensorized and under-actuated artificial hands with simulation and/or prototype tests The thesis work involves the study of kinematics of

More information

12 th National Congress on Theoretical and Applied Mechanics September 2013, Saints Constantine and Helena, Varna, Bulgaria

12 th National Congress on Theoretical and Applied Mechanics September 2013, Saints Constantine and Helena, Varna, Bulgaria 12 th National Congress on Theoretical and Applied Mechanics 23-26 September 2013, Saints Constantine and Helena, Varna, Bulgaria DESIGN OF UPPER LIMB EXOSKELETON ACTUATED WITH PNEUMATIC ARTIFICIAL MUSCLES

More information

Chapter 15 Unilateral and Bilateral Rehabilitation of the Upper Limb Following Stroke via an Exoskeleton

Chapter 15 Unilateral and Bilateral Rehabilitation of the Upper Limb Following Stroke via an Exoskeleton Chapter 15 Unilateral and Bilateral Rehabilitation of the Upper Limb Following Stroke via an Exoskeleton Jacob Rosen, Dejan Milutinović, Levi M. Miller, Matt Simkins, Hyunchul Kim, and Zhi Li Abstract

More information

1. EXECUTIVE SUMMARY 2. PRODUCT SUMMARY

1. EXECUTIVE SUMMARY 2. PRODUCT SUMMARY 1. EXECUTIVE SUMMARY Success or failure in the medical industry begins with a diagnosis. This fact is well known and revered by clinicians, therapists, and health care providers. However, despite this

More information

Drawing Pressure Estimation Using Torque Feedback Control Model of A 4-DOF Robotic Arm

Drawing Pressure Estimation Using Torque Feedback Control Model of A 4-DOF Robotic Arm Drawing Estimation Using Torque Feedback Control Model of A 4-DOF Robotic Arm Meng Cheng Lau 1, Chi-Tai Cheng 2, Jacky Baltes 3, and John Anderson 3 1 Center for Artificial Intelligence Technology, Universiti

More information

Estimation of Complex Anatomical Joint Motions Using a Spatial Goniometer

Estimation of Complex Anatomical Joint Motions Using a Spatial Goniometer Estimation of Complex Anatomical Joint Motions Using a Spatial Goniometer V. A. Dung Cai *, Philippe Bidaud *, Vincent Hayward *, and Florian Gosselin * UPMC Univ Paris 06, Institut des Systèmes Intelligents

More information

Relationship between clinical assessments of function and measurements from an upper-limb robotic rehabilitation device in cervical spinal cord injury

Relationship between clinical assessments of function and measurements from an upper-limb robotic rehabilitation device in cervical spinal cord injury TNSRE-211-211.R1 1 Relationship between clinical assessments of function and measurements from an upper-limb robotic rehabilitation device in cervical spinal cord injury José Zariffa, Member, IEEE, Naaz

More information

UNIVERSITY OF CALIFORNIA, IRVINE. WRIST (Wrist Resonator for Independent Stroke Training) THESIS

UNIVERSITY OF CALIFORNIA, IRVINE. WRIST (Wrist Resonator for Independent Stroke Training) THESIS UNIVERSITY OF CALIFORNIA, IRVINE WRIST (Wrist Resonator for Independent Stroke Training) THESIS submitted in partial satisfaction of the requirements for the degree of MASTER OF SCIENCE in Mechanical and

More information

CHAPTER 11 UNIVERSITY OF DENVER

CHAPTER 11 UNIVERSITY OF DENVER CHAPTER 11 UNIVERSITY OF DENVER School of Engineering and Computer Science Department of Engineering 2390 S. York Street Denver, CO 80208 Principal Investigator: Kimberly E. Newman (303)871-3436 kinewman@du.edu

More information

PREDICTIVE MAINTENANCE PARADIGMS IN BIOMEDICINE

PREDICTIVE MAINTENANCE PARADIGMS IN BIOMEDICINE NSF I-UCRC on Intelligent Maintenance Systems; Planning Workshop at the University of Texas; May 15, 2012, Austin, TX PREDICTIVE MAINTENANCE PARADIGMS IN BIOMEDICINE PROF. DRAGAN DJURDJANOVIC PROF. JONATHAN

More information

Design Considerations of a Human Arm Exoskeleton

Design Considerations of a Human Arm Exoskeleton , July 5-7, 2017, London, U.K. Design Considerations of a Human Arm Exoskeleton Dumitru S., Rosca A. S., Ciurezu L. and Didu A. Abstract Design considerations are reported for assistive human arm recovery

More information

SYNCHRONIZATION OF ARM AND HAND ASSISTIVE ROBOTIC DEVICES TO IMPART ACTIVITIES OF DAILY LIVING TASKS

SYNCHRONIZATION OF ARM AND HAND ASSISTIVE ROBOTIC DEVICES TO IMPART ACTIVITIES OF DAILY LIVING TASKS SYNCHRONIZATION OF ARM AND HAND ASSISTIVE ROBOTIC DEVICES TO IMPART ACTIVITIES OF DAILY LIVING TASKS Duygun Erol 1 and Nilanjan Sarkar 2 1 Department of Electrical & Electronics Engineering, Yeditepe University,

More information

INTELLIGENT CONTROL OF A ROBOT MANIPULATOR FOR KNEE REHABILITATION

INTELLIGENT CONTROL OF A ROBOT MANIPULATOR FOR KNEE REHABILITATION Proceedings of 5th International Smposium on Intelligent Manufacturing Sstems, Ma 29-31, 2006: 695-703 Sakara Universit, Department of Industrial Engineering INELLIGEN CONROL OF A ROBO MANIPULAOR FOR KNEE

More information

Learning and Robotic Imitation of Therapist s Motion and Force for Post-disability Rehabilitation

Learning and Robotic Imitation of Therapist s Motion and Force for Post-disability Rehabilitation 2017 IEEE International Conference on Systems, Man, and Cybernetics (SMC) Banff Center, Banff, Canada, October 5-8, 2017 Learning and Robotic Imitation of Therapist s Motion and Force for Post-disability

More information

A Mechanism for Elbow Exoskeleton for Customised Training

A Mechanism for Elbow Exoskeleton for Customised Training A Mechanism for Elbow Exoskeleton for Customised Training Soumya K. Manna, Venketesh N. Dubey, Member, IEEE Abstract It is well proven that repetitive extensive training consisting of active and passive

More information

METHOD AND DEVICE FOR ISOKINETIC QUADRICEPS EXERCISE USING PNEUMATIC MUSCLE ACTUATOR (PMA) RESISTANCE: COMPARISON TO FREE WEIGHT EXERCISE

METHOD AND DEVICE FOR ISOKINETIC QUADRICEPS EXERCISE USING PNEUMATIC MUSCLE ACTUATOR (PMA) RESISTANCE: COMPARISON TO FREE WEIGHT EXERCISE METHOD AND DEVICE FOR ISOKINETIC QUADRICEPS EXERCISE USING PNEUMATIC MUSCLE ACTUATOR (PMA) RESISTANCE: COMPARISON TO FREE WEIGHT EXERCISE M. COWGILL, C. PHILLIPS, D. REYNOLDS, J. SERRES, K. HALL, S. MOHLER

More information

DEVELOPMENT OF WEARABLE MASTER-SLAVE TRAINING DEVICE FOR UPPER LIMB CONSTRUCTED WITH PNEUMATIC ARTIFICIAL MUSCLES

DEVELOPMENT OF WEARABLE MASTER-SLAVE TRAINING DEVICE FOR UPPER LIMB CONSTRUCTED WITH PNEUMATIC ARTIFICIAL MUSCLES DEVELOPMENT OF WEARABLE MASTER-SLAVE TRAINING DEVICE FOR UPPER LIMB CONSTRUCTED WITH PNEUMATIC ARTIFICIAL MUSCLES Daisuke SASAKI*, Toshiro NORITSUGU* and Masahiro TAKAIWA* * Graduate School of Natural

More information

Arduino Based Arm Rehabilitation Assistive Device

Arduino Based Arm Rehabilitation Assistive Device ISSN 2231-8798 2011 UniKLBMI Arduino Based Arm Rehabilitation Assistive Device R. Ambar 1, 3, M. S. Ahmad 2, A. M. Mohd Ali 2, 3, 4 and M. M. Abdul Jamil 2, 3 1 Department of Computer Engineering, Faculty

More information

Erigo User Script 1. Erigo Background Information. 2. Intended use and indications

Erigo User Script 1. Erigo Background Information. 2. Intended use and indications Erigo User Script 1. Erigo Background Information The Erigo was developed in collaboration with the Spinal Cord Injury Center at the Balgrist University Hospital in Zurich, Switzerland and the Orthopaedic

More information

Poster A2

Poster A2 Poster A1 Poster A2 Poster A3 Poster A4 Poster A5 Poster A6 Investigation of a passive inter-limb device on step-tostep transition of human walking J Zhang: Queen s University, Canada Q Li: Queen s University,

More information

DESIGNING AN IN-HOME SCALABLE ROBOTIC ARM EXOSKELETON FOR HAND REHABILITATION THERAPY

DESIGNING AN IN-HOME SCALABLE ROBOTIC ARM EXOSKELETON FOR HAND REHABILITATION THERAPY DESIGNING AN IN-HOME SCALABLE ROBOTIC ARM EXOSKELETON FOR HAND REHABILITATION THERAPY A Thesis Presented to The Academic Faculty by Jonathan Tuck In Partial Fulfillment of the Requirements for the Degree

More information

Robotics in the Rehabilitation of Neurological Conditions

Robotics in the Rehabilitation of Neurological Conditions Robotics in the Rehabilitation of Neurological Conditions Gil J Cerros, MSHS The PRANAYAMA Research Group June 14, 2015 Clinical and Translational Research The PRANAYAMA Research Group fdsfsf Robotics

More information

Multi-joint Mechanics Dr. Ted Milner (KIN 416)

Multi-joint Mechanics Dr. Ted Milner (KIN 416) Multi-joint Mechanics Dr. Ted Milner (KIN 416) Muscle Function and Activation It is not a straightforward matter to predict the activation pattern of a set of muscles when these muscles act on multiple

More information

Design and Dynamic Modeling of Flexible Rehabilitation Mechanical Glove

Design and Dynamic Modeling of Flexible Rehabilitation Mechanical Glove IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Design and Dynamic Modeling of Flexible Rehabilitation Mechanical Glove To cite this article: M X Lin et al 2018 IOP Conf. Ser.:

More information

Verification of Passive Power-Assist Device Using Humanoid Robot: Effect on Bending and Twisting Motion

Verification of Passive Power-Assist Device Using Humanoid Robot: Effect on Bending and Twisting Motion 15 IEEE-RAS 15th International Conference on Humanoid Robots (Humanoids) November 3-5, 15, Seoul, Korea Verification of Passive Power-Assist Device Using Humanoid Robot: Effect on Bending and Twisting

More information

Feasibility of the UR5 Industrial Robot for Robotic Rehabilitation of the Upper Limbs After Stroke

Feasibility of the UR5 Industrial Robot for Robotic Rehabilitation of the Upper Limbs After Stroke Feasibility of the UR5 Industrial Robot for Robotic Rehabilitation of the Upper Limbs After Stroke Erik Kyrkjebø 1, Mads Johan Laastad 2, Øyvind Stavdahl 3 Abstract Robot-assisted therapy is an emerging

More information

1. Introduction. Correspondence should be addressed to Kenneth N. K. Fong;

1. Introduction. Correspondence should be addressed to Kenneth N. K. Fong; BioMed Research International Volume 2016, Article ID 9346374, 9 pages http://dx.doi.org/10.1155/2016/9346374 Research Article Effects of Arm Weight Support Training to Promote Recovery of Upper Limb Function

More information

Northeastern University. Boston, Massachusetts. August, 2013

Northeastern University. Boston, Massachusetts. August, 2013 Design and Control of a 2 Degree of Freedom Upper Limb Robotic Rehabilitation Device A Thesis Presented by Patrick Joseph Murphy to The Department of Mechanical and Industrial Engineering in partial fulfillment

More information

Chapter 20: Muscular Fitness and Assessment

Chapter 20: Muscular Fitness and Assessment Chapter 20: Muscular Fitness and Assessment American College of Sports Medicine. (2010). ACSM's resource manual for guidelines for exercise testing and prescription (6th ed.). New York: Lippincott, Williams

More information

We recommend you cite the published version. The publisher s URL is:

We recommend you cite the published version. The publisher s URL is: Tzemanaki, A., Raabe, D. and Dogramadzi, S. (2011) Development of a novel robotic system for hand rehabilitation. In: 2011 24th International Symposium on Computer-Based Medical Systems (CBMS), Bristol,

More information

Emotion Embodiment in Robot-Assisted Rehabilitation System using Hybrid Automata

Emotion Embodiment in Robot-Assisted Rehabilitation System using Hybrid Automata 2014 IIAI 3rd International Conference on Advanced Applied Informatics Emotion Embodiment in Robot-Assisted Rehabilitation System using Hybrid Automata Shahrul Naim Sidek Department of Mechatronics, Faculty

More information

Real-time computer modeling of weakness following stroke optimizes robotic assistance for movement therapy

Real-time computer modeling of weakness following stroke optimizes robotic assistance for movement therapy Proceedings of the 3rd International IEEE EMBS Conference on Neural Engineering Kohala Coast, Hawaii, USA, May 2-5, 2007 ThE1.2 Real-time computer modeling of weakness following stroke optimizes robotic

More information

Reach and grasp by people with tetraplegia using a neurally controlled robotic arm

Reach and grasp by people with tetraplegia using a neurally controlled robotic arm Leigh R. Hochberg et al. Reach and grasp by people with tetraplegia using a neurally controlled robotic arm Nature, 17 May 2012 Paper overview Ilya Kuzovkin 11 April 2014, Tartu etc How it works?

More information

REHABILITATION PROTOTYPE TO IMPROVES MOBILITY OF ACUTE ROTATOR CUFF INJURY

REHABILITATION PROTOTYPE TO IMPROVES MOBILITY OF ACUTE ROTATOR CUFF INJURY CLAWAR 2018: 21st International Conference on Climbing and Walking Robots and the Support Technologies for Mobile Machines, Panama City, Panama, 10-12 September 2018 REHABILITATION PROTOTYPE TO IMPROVES

More information

CONTROL OF THE BOUNDARY CONDITIONS OF A DYNAMIC KNEE SIMULATOR

CONTROL OF THE BOUNDARY CONDITIONS OF A DYNAMIC KNEE SIMULATOR CONTROL OF THE BOUNDARY CONDITIONS OF A DYNAMIC KNEE SIMULATOR J. Tiré 1, J. Victor 2, P. De Baets 3 and M.A. Verstraete 2 1 Ghent University, Belgium 2 Ghent University, Department of Physical Medicine

More information

Verification of Assistive Effect Generated by Passive Power-Assist Device Using Humanoid Robot

Verification of Assistive Effect Generated by Passive Power-Assist Device Using Humanoid Robot Proceedings of the 14 IEEE/SICE International Symposium on System Integration, Chuo University, Tokyo, Japan, December 13-1, 14 SuP2D.1 Verification of Assistive Effect Generated by Passive Power-Assist

More information

Robotic Training and Kinematic Analysis of Arm and Hand after Incomplete Spinal Cord Injury: A Case Study.

Robotic Training and Kinematic Analysis of Arm and Hand after Incomplete Spinal Cord Injury: A Case Study. Robotic Training and Kinematic Analysis of Arm and Hand after Incomplete Spinal Cord Injury: A Case Study. Z. Kadivar Dept of Physical Medicine and Rehabilitation aylor College of Medicine Houston, U.S.A

More information

Reducing Abnormal Synergies of Forearm, Elbow, and Shoulder Joints in Stroke Patients with Neuro-rehabilitation Robot Treatment and Assessment

Reducing Abnormal Synergies of Forearm, Elbow, and Shoulder Joints in Stroke Patients with Neuro-rehabilitation Robot Treatment and Assessment Journal of Medical and Biological Engineering, 32(2): 139-146 139 Reducing Abnormal Synergies of Forearm, Elbow, and Shoulder Joints in Stroke Patients with Neuro-rehabilitation Robot Treatment and Assessment

More information

WEARABLE TREMOR REDUCTION DEVICE (TRD) FOR HUMAN HANDS AND ARMS

WEARABLE TREMOR REDUCTION DEVICE (TRD) FOR HUMAN HANDS AND ARMS Proceedings of the 2018 Design of Medical Devices Conference DMD2018 April 9-12, 2018, Minneapolis, MN, USA DMD2018-6918 WEARABLE TREMOR REDUCTION DEVICE (TRD) FOR HUMAN HANDS AND ARMS Sreekanth Rudraraju

More information