Assessment of Automated Scoring of Polysomnographic Recordings in a Population with Suspected Sleep-disordered Breathing

Size: px
Start display at page:

Download "Assessment of Automated Scoring of Polysomnographic Recordings in a Population with Suspected Sleep-disordered Breathing"

Transcription

1 Assessment of Automated Scoring of Polysomnographic Recordings in a Population with Suspected Sleep-disordered Breathing Stephen D. Pittman, MSBME, RPSGT 1 ; Mary M. MacDonald, RPSGT 1 ; Robert B. Fogel, MD 1,2 ; Atul Malhotra, MD 1,2 ; Koby Todros 3 ; Baruch Levy 3 ; Amir B. Geva, DSc 3 ; David P. White, MD 1,2 1Division of Sleep Medicine, Brigham and Women s Hospital, Boston, Mass; 2 Harvard Medical School, Boston, Mass; 3 WideMED Ltd., Omer, Israel Study Objectives: To assess the accuracy of an automated system (Morpheus I Sleep Scoring System) for analyzing and quantifying polysomnographic data from a population with sleep-disordered breathing. Setting: Sleep laboratory affiliated with a tertiary care academic medical center. Measurements and Results: 31 diagnostic polysomnograms were blindly analyzed prospectively with the investigational automated system and manually by 2 registered polysomnography technologists (M1 & M2) from the same laboratory. Sleep stages, arousals, periodic limb movements, and respiratory events (apneas and hypopneas) were scored by all 3. Agreement, Cohen κ, and intraclass correlation coefficients were tabulated for each variable and compared between scoring pairs (A-M1, A-M2, M1-M2). The 26,876 epochs (224 hours of recording time) were analyzed. For sleep staging, agreement/κ were A-M1: 78%/0.67, A-M2: 73%/0.61, and M1-M2: 82%/0.73. The mean respiratory disturbance indexes were M1: 20.6 ± 23.0, M2: 22.5 ± 24.5, and A: 23.7 ± 23.4 events per hour of sleep. The respiratory disturbance index concordance between each scoring pair was excellent (intraclass correlation coefficients 0.95 for all pairs), although there was disagreement in the classification of moderate INTRODUCTION POLYSOMNOGRAPHY (PSG) REMAINS THE STANDARD DIAGNOSTIC TEST FOR THE EVALUATION OF SLEEP AND THE DETECTION OF SLEEP PATHOLOGIES such as Disclosure Statement This was an industry supported study by WideMed Ltd. Mr. Pittman has received research support from Respironics Inc., Itamar Medical, WideMed Ltd., and the Alfred E. Mann Foundation; served as a consultant to Itamar Medical; and is currently employed by Respironics Inc. Dr. Geva is the president and CEO of WideMed Ltd. Ms. MacDonald has received support from Cephalon, Respironics Inc., Itamar Medical, and WideMed Ltd. Dr. Malhotra receives research support from Respironics Inc. Dr. White has received research support and consulting fees from Respironics Inc., Itamar Medical, the Alfred E. Mann Foundation, and WideMed Ltd; and has served as a consultant for Aspire Medical. Dr. Fogel is a member of the Visiting Speaker's Bureau for Wyeth Pharmaceuticals. Mr. Todros and Mr. Levy are employed by WideMed Ltd. The manual scoring and statistical analyses for this study were done at the sleep laboratory affiliated with Brigham and Women's Hospital. The automated algorithms used were developed by WideMed Ltd.; however, the actual analysis of the data was performed onsite at the sleep laboratory by the authors with some technical assistance from WideMed Ltd. The authors maintained control of the data throughout the entire study. The paper was written by the authors. Submitted for publication April 2004 Accepted for publication July 2004 Address correspondence to: David P. White MD, Sleep Disorders Research Division of Sleep Medicine, Brigham and Women s Hospital, 75 Francis Street, Boston, MA ; Tel: (617) ; Fax: (617) ; dpwhite@rics.bwh.harvard.edu SLEEP, Vol. 27, No. 7, sleep-disordered breathing (percentage of positive agreement: A-M1, 37.5% and A-M2, 44.4%) defined as a respiratory disturbance index between 15 and 30 events per hour of sleep. For respiratory-event detection, agreement/κ were A-M1 and A-M2: 90%/0.66 and M1-M2: 95%/0.82. The agreement and κ for limb movement detection were A-M1: 93%/0.68, A-M2: 92%/0.66, and M1-M2: 96%/0.77. The scoring of arousals was less reliable (agreement range: 76%-84%, κ range: ) for all pairs. Conclusions: Agreement between manual scorers in a population with moderate sleep-disordered breathing was close to the average pairwise agreement of 87% reported in the Sleep Heart Health Study. The automated classification of sleep stages was also close to this standard. The automated scoring system holds promise as a rapid method to score polysomnographic records, but expert verification of the automated scoring is required. Key Words: automated scoring, obstructive sleep apnea, sleep apnea, sleep-disordered breathing, arousal, periodic limb movement, polysomnography Citation: Pittman SD; MacDonald MM; Fogel RB et al. Assessment of automated scoring of polysomnographic recordings in a population with suspected sleep-disordered breathing. SLEEP 2004;27(7): sleep-disordered breathing (SDB) and periodic limb movement (PLM) disorder. Digital PSG acquisition systems have allowed the diagnostic capacity of sleep disorder centers across the country to increase in recent years, but the manual scoring of sleep stages, brief arousals from sleep, SDB events, and PLMs remains the standard. To date, computer-assisted scoring methods have rarely been rigorously validated and, when evaluated, have rarely performed well. Manual scoring relies on the application of standard rules to identify specific waveforms and quantify their duration. This process is time consuming, expensive, and subject to variability in interpretation. 1 Previous studies have reported a range of agreement in sleep-stage classification among human scorers with scoring pairs from different laboratories tending to agree less than scorers from the same laboratory. 2 One investigation of interscorer variability was conducted as part of the Sleep Heart Health Study following thorough training of the scoring technologists at a central scoring site to maximize the uniform application of scoring rules and improve reproducibility. 3 The average pairwise agreement of sleep staging among 3 scorers on 30 randomly selected PSG records using epoch-by-epoch comparisons was 87%, although the amount of SDB was relatively low in this population (average respiratory disturbance index [RDI] was 6 events per hour using a 4% desaturation requirement). Investigations of automated scoring of adult human sleep to date have considered single components of PSG scoring such as sleep staging, 4-7 detection of respiratory events, 8 arousals from sleep, 9 and PLMs. 10 Recent sleep-stage scoring methods have uti-

2 lized power spectral analysis, 4 a neural network model based on feature extraction, 5 and a Gaussian Observation Hidden Markov Model. 6 Epoch-by-epoch agreement in sleep-stage scoring for these methods and manual scoring varied from 74% to 84.5%, but these comparisons are for normal populations without SDB. Taha and coworkers 8 reported an epoch-by-epoch agreement of 93.1% between manual and automated scoring of respiratory events, including apneas and hypopneas, on 10 PSG records. The method of automated arousal scoring by Pillar and coworkers 9 correlated well with the manual scoring of arousals using standard criteria, 11 but the automated scoring was performed on a measure of peripheral arterial vasomotor tone that is not a standard PSG measure. Thus, this method may have limited clinical utility. The purpose of the present study was to evaluate interscorer agreement in sleep-stage assignment and the detection of defined abnormal events during sleep between 2 expert human scorers and an automated sleep-scoring system. The study was performed on a data set from subjects with suspected SDB. MATERIALS AND METHODS All night, in-laboratory, diagnostic PSG records from adult patients referred to the clinical sleep laboratory of Brigham and Women s Hospital with suspected obstructive sleep apnea were included in this study. These were not consecutive records, but a random sample of records from patients who disclosed on a comprehensive questionnaire between June and December of 2002 that they were interested in being contacted about research studies conducted at the sleep laboratory. The analysis of these records was approved by the Human Research Committee of Brigham and Women s Hospital. Subjects Thirty-one subjects (9 women) with suspected SDB participated in the study. The mean age of the subjects was 44.3 ± 11.3 (range 21-62) years, and the mean body mass index was 33.7 ± 7.0 kg/m 2. The mean Epworth Sleepiness Scale scores were 9.3 ± 4.7. Protocol Recorded signals included electroencephalogram ([EEG] C4- A1, C3-A2, O2-A1 and O1-A2), electrooculogram (EOG), submental and bilateral tibial electromyogram (EMG), electrocardiogram (ECG), airflow (nasal-oral thermistor and nasal pressure [PTAF2, Pro-Tech Services, Woodinville, Wash]), chest and abdominal movement (piezo bands), arterial oxyhemoglobin saturation (Model 930 Pulse-Oximeter, Respironics, Murrysville, Penn), body position, and snoring intensity. All physiologic data were collected and stored using the ALICE3 digital PSG system (Respironics). The EEG and EMG channels were sampled at 100 Hz. The PSGs were then scored manually and independently from the raw data by 2 registered PSG technologists (M1 & M2). Sleep was staged according to standard criteria in 30-second epochs from a high-resolution computer display. 12 The predominant sleep stage was scored for each epoch as wake, stage 1, stage 2, stage 3, stage 4, or rapid eye movement (REM). Arousals from sleep were identified according to American Sleep Disorders Association (ASDA) guidelines. 11 An arousal index was calculated as the number of arousals per hour of sleep. Respiratory events were scored according to recent American Academy of Sleep Medicine (AASM) Medicare guidelines. 13 In particular, an apnea was scored if airflow was absent for 10 seconds, and a hypopnea was scored if thoracoabdominal movement or airflow was reduced by 30% compared to baseline for at least 10 seconds with at least a 4% oxygen desaturation. An RDI was then calculated based on the number of apneas plus hypopneas per hour of sleep. The PLMs were scored according to ASDA criteria. 14 A PLM index (PLMI) was calculated as the number of PLMs per hour of sleep. Unscored copies of all records were then queued for automated batch scoring using the Morpheus I Sleep Scoring System (A) to classify sleep stages and then detect arousals, SDB events, and PLMs according to the same ASDA and AASM criteria. A database containing the scores for each epoch for all PSGs for all scorers (M1, M2, A) was created. Stages 3 and 4 were grouped into delta sleep. Only epochs scored as sleep were analyzed for arousals, respiratory events, and limb movements. Automated Algorithms The Morpheus 1 system performs fully automated PSG data analysis that includes sleep staging, arousal scoring, PLM detection, snoring detection, single-channel ECG analysis, and respiratory-event scoring. These analyses are enabled by utilizing novel algorithms to process EEG, EOG, EMG, ECG, snoring, and respiratory-related signals. The Morpheus I EEG analysis is based on dynamic state modeling of the EEG frequency during sleep. Accordingly, a dynamic, 4 frequency state model with high-frequency, low-energy mixed-frequency, high-energy mixed-frequency, and low-frequency states is applied for modeling the behavior of EEG signals for a given patient. The EEG signals are initially partitioned into quasi-stationary segments by applying a novel adaptive segmentation algorithm. The patient s intrinsic states are then adaptively recognized by utilizing fuzzy clustering 15,16 of features extracted from these quasi-stationary EEG segments. Membership levels in each of the recognized states are attributed to each EEG segment, creating for each state a membership function, which sketches a continuous measurement of intrinsic EEG activity. The membership level in the highfrequency state is dominant in wakefulness, while the membership levels in the low-energy mixed-frequency, high-energy mixed-frequency, and low-frequency states are dominant in sleep stages 1 and rapid eye movement (REM), sleep stage 2, and delta sleep, respectively. The EEG analysis also includes detection of K-complexes, sleep spindles, movement artifacts, and electrode artifacts. The 2 EOG signals are simultaneously portioned into quasi-stationary segments, while features such as variance and cross-correlation are extracted from parallel segments. Segments with enhanced relative energy with respect to their environment, for which the cross-correlation measure decreases below a certain threshold are classified as REM. The EMG signal is also partitioned into quasi-stationary segments, and time-frequency features are extracted from each segment. Segments with enhanced relative energy with respect to their environment are classified as EMG bursts, which are useful for arousal scoring. Zones of reduced EMG energy level are detected by applying hierarchical fuzzy clustering 17 of features extracted from each EMG segment. SLEEP, Vol. 27, No. 7,

3 It is important to mention that artifacts such as ECG tracings and 60-Hz noise are filtered from the EEG, EOG, and EMG by applying adaptive noise-canceling techniques. The information gathered from the EEG, EOG, and EMG signals is fused by the Rechtschaffen and Kales criteria for automatic sleep staging, in the following manner: (1) the membership level in the high-frequency state and detected movements are used for scoring stage wake; (2) the membership level in the low-energy mixed-frequency state is used for scoring stage 1; (3) the membership level in high-energy mixed-frequency state, detected K-complexes, and spindles are used for scoring sleep stage 2; (4) the membership level in low-energy mixed-frequency, detected zones of reduced EMG activities, and detected REM are used for scoring stage REM; and (5) the membership level in low-frequency state and the peak-to-peak amplitudes of the EEG segments are used for scoring stage delta sleep. The automatic arousal-scoring algorithm is based on the ASDA set of rules for arousal scoring. The algorithm utilizes membership levels for each EEG segment, together with scored sleep stages and detected EMG energy bursts. The algorithm also utilizes detected respiratory events, which are related to EEG alternating patterns in a time window of 10 seconds before and after a respiratory event. Both limb channels are adaptively segmented into quasi-stationary segments. Segments with specific duration and with enhanced relative energy with respect to their environment are detected as limb movements. Standard criteria for the classification of PLMs are applied on the detected limb movements in order to score a PLMI. The snoring detection algorithm works like the algorithm for detection of EMG energy bursts. By applying wavelet-based and template-matching algorithms, the Morpheus 1 system performs full 1-lead (lead II) ECG analysis, which includes full-beat segmentation; pathologic beat detection (premature atrial contraction, premature ventricular contraction); heart-rate variability analysis; heart-rate analysis (bradycardia, tachycardia, heart-rate drops, and heart-rate elevations); ST-segment analysis; QT analysis; and arrhythmia detection (supraventricular tachycardia, ventricular tachycardia). We did not validate the ECG analysis in the present study. For respiratory signals, the Morpheus I system utilizes a fuzzylogic decision algorithm to reach more realistic decisions regarding the detection of apnea or hypopnea. The respiratory recordings are first segmented using a fuzzy-logic minimum modelderived distance approach. In addition to the respiratory tracing, the system searches around each suspected event for arousals (5 seconds before and 5 seconds after the suspected event), desaturations (5 seconds before and 70 seconds after the suspected event), and limb movements (5 seconds before and 70 seconds after the suspected event). A fuzzy-logic decision algorithm is then applied to calculate the probability that the suspected event complies with the respiratory-event rules by integrating the information gathered from the different sources. Once an event is detected, adaptive segmentation and spectral analysis are applied to classify the type of the respiratory event. Data Analysis The variables assessed for each scorer included total sleep time, sleep efficiency (total sleep time / time in bed), sleep-onset latency (time from lights out to the first epoch of sleep), REMonset latency (time from lights out to the first epoch of REM), minutes of stage 1 sleep, minutes of stage 2 sleep, minutes of stage delta sleep, minutes of REM sleep, minutes of wake, RDI, Arousal Index, and PLMI. Comparisons of each scoring pair (M1-M2, A-M1, A-M2) included epoch-by-epoch assessments of agreement and Cohen κ 18 for the classification of sleep stages (5 5 matrix: wake, stage 1, stage 2, delta, and REM sleep; 4 4 matrix: wake, stage 1 + stage 2, delta, and REM sleep; and 3 3 matrix: wake, non-rem, and REM sleep) and detection of arousals (3 3 matrix: 0, 1, and 2 arousals per epoch), SDB events (3 3 matrix: 0, 1, and 2 events per epoch) and PLMs (3 3 matrix: 0, 1, and 2 events per epoch). The percentage of agreement for a scoring pair was calculated by summing the main diagonal and dividing by the total number of epochs. The κ represents the nonrandom component of this agreement. Bland- Altman plots were used to evaluate the mean difference between scorers for selected parameters. 19 Concordance was assessed using the intraclass correlation coefficient (ICC). Receiver-operator characteristic curves were constructed for the RDIs calculated for each scoring pair to assess the performance of the automated algorithm across the spectrum of SDB severity (RDI cutoffs of 5, 10, 15, 20, and 30 events per hour for defining obstructive sleep apnea). The area under the receiver-operator characteristic curve was then calculated for each threshold and reported with the standard error and the limits of the 95% confidence interval. Using AASM criteria, 20 normal breathing was defined as an RDI less than 5 events per hour, mild SDB as an RDI of 5 or more but less than 15 events per hour, moderate SDB as an RDI of 15 or more but less than 30 events per hour, and severe SDB as an RDI of 30 or more events per hour of sleep. All results are provided as means ± 1 SD except for the Bland- Altman mean difference, which used ± 2 SD for the limits of agreement. Statistical significance was considered to be present when P <.05. The paired t test was used to assess significance when the means were normally distributed for each scoring pair; otherwise, the nonparametric Wilcoxon signed-rank test was used. RESULTS A total of 26,876 epochs (224.0 hours of recording time) from 31 diagnostic PSG records were analyzed in this study. Epochby-epoch sleep-stage scoring agreement for each scoring pair as shown in Table 1. Agreement between the 2 manual scorers was 82.1% with a κ of Automated scoring identified 57 epochs (0.02%) that did not meet predefined reliability criteria for sleepstage scoring. Thus, 26,819 epochs were used to calculate agreement between the automated scoring system and each manual scorer. Agreement between M1 and A was 77.7% with a κ of Agreement between M2 and A was 73.3% with a κ of Representative hypnograms for each scorer are shown for a single PSG in Figure 1. Agreement between M1 and A improved to 82.6% with a κ of 0.71 when stages 1 and 2 sleep were combined to yield a 4 4 matrix (wake, light sleep, delta sleep, and REM sleep). Agreement between M2 and A was 79.9% with a κ of 0.65 for the same 4 4 matrix. Likewise, agreement between M1 and M2 improved to 88.7% with a κ of 0.80 when stages 1 and 2 sleep were combined. When stages 1, 2, and delta sleep were combined to yield a 3 SLEEP, Vol. 27, No. 7,

4 Table 1 Epoch-by-Epoch Comparisons for Sleep-Stage Scoring for Each Scoring Pair* 3 matrix (wake, non-rem and REM sleep), agreement between M1 and A was 88.0% with a κ of Agreement between M2 and A improved to 88.0% with a κ of 0.74 for the same 3x3 matrix. Likewise, agreement between M1 and M2 was 93.5% with a κ of Table 2 shows the variables commonly used to assess sleep and sleep disorders for each scorer (M1, M2, and A). The mean total sleep time was 8.9 minutes longer for A than for M1 (P <.05) and 12.5 minutes longer than for M2 (P <.01). The total sleep time was highly concordant among all scoring pairs (ICC 0.92), as shown in Table 2. The mean sleep-onset latency was 3.8 minutes shorter for A than for M1 (NS) and 3.9 minutes shorter than for M2 (NS). Concordance was high for calculating the sleeponset latency among all scoring pairs (ICC 0.86), as shown in Figure 2, with the manual scoring pairs demonstrating the best concordance. On the other hand, the mean REM-onset latency was considerably longer for A than for M1 (45.7 minute difference, P<.01) and for M2 (48.1 minute difference, P <.01) because the automated system missed the first REM period in 10 (32%) subjects. In terms of sleep staging, the automated system tended to score less stage 2 sleep (17.2 minute less than M1, P<.01 and 7.8 minutes less than M2, NS) and wakefulness (9.8 minutes less than M1, P <.05 and 13.4 minutes less than M2, P <.01) and more delta sleep (11.9 minutes more than M1, P <.01 and 29.4 minutes more than M2, P <.01) than either manual scorer. Significant differences were observed for the mean duration of stage 1 sleep for all scoring pairs (P <.01). The differences between the REM sleep duration for automated system and both M1 and M2 were not significant. The ICC are provided for all variables in Table 2. The automated system scored fewer arousals than either manual scorer (8.1 arousals per hour less than M1, P <.01, and 14.2 arousals per hour less than M2, P <.01). Concordance was better between M1 and A (ICC = 0.72) than between M2 and A (ICC = 0.58). The agreement for detecting arousals between A and manual scoring seems acceptable (76.2% agreement with M1 and 76.1% agreement with M2), but the κ shown in Table 3 suggest the agreement is only marginally better than chance (κ = 0.28 for A-M1 and 0.30 for A-M2). However, the κ between manual scorers for arousals was only 0.57, suggesting agreement corrected for chance was relatively poor regardless of the scoring pair. The automated system scored slightly more respiratory events than either manual scorer (3.1 events per hour more than M1, P <.01, and 1.2 events per hour more than M2, NS) as shown in Stage Comparison M1-M2 A-M1 A-M2 Wake Wake 4,835 4,037 4,162 PPA+, % Stage Stage Delta REM Stage 1 Stage PPA+, % Stage 2 1,765 1,327 1,765 Delta REM Stage 2 Stage 2 12,233 11,694 11,022 PPA+, % Delta 1,293 1,427 2,080 REM Delta Delta 1,181 2,013 1,146 PPA+, % REM REM REM 3,098 2,686 2,435 PPA+, % Total epochs, no. 26,876 26,819 26,819 Total agreements, no. 22,078 20,838 19,657 Total agreement, % κ (wake, 1, 2, delta, REM) κ (wake, 1+2, delta, REM) κ (wake, non-rem, REM) *Comparisons with automated scoring excluded 57 epochs (0.02%) that were identified as epochs that did not meet reliability criteria for automated scoring. The number of occurrences of the specified stage combinations relative to total number of epochs scored. A refers to automated scoring with the Morpheus I Sleep Scoring System; M 1, manual scoring for Scorer 1; M 2, manual scoring for Scorer 2; PPA+, the percentage of positive agreement for the specified sleep stage [eg, PPA+ associated with wake (M 1 -M 2 ) = 4,835/(4, ) = 80.9%]; REM, rapid eye movement. SLEEP, Vol. 27, No. 7,

5 Figure 1 Representative hypnograms for each scorer are shown for 1 subject. M 1 refers to manual scoring for Scorer 1; M 2, manual scoring for Scorer 2; A, automated scoring with the Morpheus I Sleep Scoring System; REM, rapid eye movement. SLEEP, Vol. 27, No. 7,

6 Table 2 Standard Variables Commonly Used to Quantify Sleep and Sleep Disorders for Each Scorer. Variable Scorer* Intraclass Correlation Coefficient M 1 M 2 A M 1 vs M 2 M 1 vs A M 2 vs A Total sleep time, min ± ± ± υ Sleep efficiency, % 82.7 ± ± ± υ Sleep-onset latency, min 25.9 ± ± ± υ REM-onset latency, min ± ± ± υ Stage wake, min 85.3 ± ± ± υ Stage 1, min 18.6 ± ± ± ΣΝ Γ Stage 2, min ± ± ± Ν Stage 1 & 2, min ± ± ± Σ Γ Stage delta, min 38.2 ± ± ± συ Stage REM, min 59.9 ± ± ± σ Arousal Index, events/h 30.1 ± ± ± ΣΝ Γ PLM Index, events/h 13.1 ± ± ± συ RDI, events/h 20.6 ± ± ± συ *Data for M 1 (manual scoring for Scorer 1), M 2 (manual scoring for Scorer 2), and A (automated scoring with the Morpheus I Sleep Scoring System) are expressed as mean ± 1 SD. A significant difference in the mean for each scoring pair is shown using the upper case Greek letter Σ, Ν, or Γ for the paired t test and the corresponding lower case Greek letter σ, υ, or for the nonparametric Wilcoxon signed-rank test for scoring pairs M 1 - M 2, M 1 -A, and M 2 A, respectively. Intraclass correlation coefficients are also given for each scoring pair to assess concordance. REM refers to rapid eye movement; PLM, periodic limb movement; RDI, respiratory disturbance index. Figure 2 Scatter plots of the sleep-onset latencies (minutes) are shown for each scoring pair. The solid lines represent the linear regression line, and the dashed lines represent the 95% confidence interval. A refers to automated scoring with the Morpheus I Sleep Scoring System; M 1, manual scoring for Scorer 1; M 2, manual scoring for Scorer 2. SLEEP, Vol. 27, No. 7,

7 Table 2. Concordance between both manual scorers and the automated system was excellent (ICC = 0.95), as shown in Figure 3. Table 4 illustrates that the automated algorithm for calculating the RDI has high sensitivity and specificity across the spectrum of SDB severity using receiver-operator characteristic curve analyses. The epoch-by-epoch agreement for detecting respiratory events between manual scoring and A was high (89.7% agreement with both manual scorers) with a κ of 0.66 for both comparisons as shown in Table 3. The pairwise agreement for the classification of SDB (normal, mild, moderate, and severe) is provided in Table 5. The detection of PLMs during sleep yielded a somewhat higher PLMI for the automated system than for either manual scorer (5.7 more limb movements per hour than M1, P <.01 and 2.5 more limb movements per hour than M2, NS) as shown in Table 2. There was also considerable scatter in this variable for each Table 3 Epoch-by-Epoch Agreement and Cohen κ Statistic for Detecting Arousals, Respiratory Events (Apneas and Hypopneas), and Periodic Limb Movements of Sleep for Each Scoring Pair* M 1 vs A M 2 vs A M 1 vs M 2 Parameter Agreement κ Agreement κ Agreement κ Arousals 76.2% % % 0.57 Respiratory Events 89.7% % % 0.82 Limb Movements 93.1% % % 0.77 *For each analysis, a 3 3 matrix was created to identify 0, 1, or 2 events per epoch for each scorer. A refers to automated scoring with the Morpheus I Sleep Scoring System; M 1, manual scoring for Scorer 1; M 2, manual scoring for Scorer 2. N = 20,940 epochs (only epochs scored as sleep by all 3 scorers were included in the analysis) Figure 3 Scatter plots of the respiratory disturbance indexes (RDI) calculated for each scoring pair are shown in A-C. The solid lines represent the linear regression line, and the dashed lines the 95% confidence interval. (D-F) Bland-Altman plots showing the difference in the RDI versus the mean of the RDI calculated for each scoring pair are shown in D-F. The solid lines represent the mean difference, and the dotted lines represent the limits of the agreement (± 2 SD). A refers to automated scoring with the Morpheus I Sleep Scoring System; M 1, manual scoring for Scorer 1; M 2, manual scoring for Scorer 2. SLEEP, Vol. 27, No. 7,

8 scoring pair as shown in Figure 4. However, the 2 manual scorers showed a closer relationship than the automated system did with either manual scorer. Concordance measured by ICC was relatively poor between A and manual scoring (ICC = 0.61 for A vs M1 and ICC = 0.65 for A vs M2). However, epoch-by-epoch agreement for scoring limb movements between the automatic and the manual scorers was better as shown in Table 3. DISCUSSION In the present study, we observed reasonably good agreement between automated PSG scoring algorithms, with that of the current accepted standard (ie, manual scoring according to established criteria). This applies to sleep staging, respiratory events, and PLMs. An automated technique would be a major advance Table 4 Receiver-Operator Characteristic Curve Data for the Evaluation of the Automated Algorithm Across the Spectrum of Sleep-disordered Breathing Severity OSA defined by OSA Prevalence ROC Area Under the Curve M 1 vs M 2 M 1 vs A M 1 RDI 5 events/h (SE: 0) 0.90 (SE: 0.06, 95% CI ) M 1 RDI 10 events/h (SE: 0.03, 95% CI ) 1.0 (SE: 0) M 1 RDI 15 events/h (SE: 0) 0.98 (SE: 0.02, 95% CI ) M 1 RDI 20 events/h (SE: 0) 0.98 (SE: 0.02, 95% CI ) M 1 RDI 30 events/h (SE: 0) 0.97 (SE: 0.03, 95% CI ) Obstructive sleep apnea (OSA) defined by various M1-determined respiratory disturbance index (RDI) cutoffs. A refers to automated scoring with the Morpheus I Sleep Scoring System; M 1, manual scoring for Scorer 1; M 2, manual scoring for Scorer 2; ROC, receiver-operator characteristic; SE: standard error, CI: confidence interval. Figure 4 Scatter plots of the periodic limb movement indexes (PLMI) are shown for each scoring pair. The solid lines represent the linear regression line, and the dashed lines represent the 95% confidence interval. A refers to automated scoring with the Morpheus I Sleep Scoring System; M 1, manual scoring for Scorer 1; M 2, manual scoring for Scorer 2. SLEEP, Vol. 27, No. 7,

9 for the field of sleep medicine because the manual scoring process is subjective, labor intensive, and expensive. Although manual scoring is currently considered the accepted standard for PSG interpretation, we observed relatively comparable concordance between a pair of manual scorers, as was observed between manual and automated scoring. Our manual scoring data are consistent with previous studies assessing interobserver correlations for manual PSG scoring. In a report by Schaltenbrand and coworkers, 5 interobserver agreement was 87.5% in a population with relatively unfragmented sleep. In the Sleep Heart Health Study, after extensive training of professional scorers within a single laboratory, the average pairwise agreement for manual scoring was reportedly 87%. 3 However, the patients in this Sleep Heart Health Study population had relatively mild apnea (RDI = 5.7 events per hour based on Medicare criteria). Overall, the literature supports better agreement between observers when they are within the same laboratory and for patients with minimal sleep fragmentation. 2 Thus, we believe that our agreement between manual scorers is comparable to the existing literature in this area. The automated algorithm that we employed in the present study performed reasonably well based on epoch-by-epoch agreement for sleep staging, respiratory event detection, and limb movements but less well for arousals. However, given the clinical population that we studied, with moderate obstructive sleep apnea and associated sleep fragmentation, we consider the results to be generally acceptable for clinical purposes. For example, the disagreement between the automated-detection algorithm and the manual scoring of sleep was partially driven by discrepancies in scoring the various stages of non-rem sleep. For clinical purposes, we do not believe that this distinction is likely to be impor- Table 5 Pairwise Comparisons for Classification of Sleep-disordered Breathing tant. The ability of the automated algorithm to distinguish non- REM sleep, REM sleep, and wakefulness is quite good. The lack of concordance in arousal detection has been previously reported. 21 This may reflect ambiguity in implementing the ASDA arousal criteria rather than failure of the automated system per se. In addition, arousal frequency and duration are poor predictors of daytime sequelae and therefore may not be critical in most clinical scenarios. 22, 23 On the other hand, the automated analysis did generate occasional errors in important sleep variables. For RDI, as indicated in Figure 3 and Table 4, several subjects would have been misclassified in terms of the presence or absence of sleep apnea. Table 5 also illustrates that there was significant disagreement in the classification of moderate SDB (RDI between 15 and 30 events per hour) between the automated analysis and both manual scorers. The same principal applies to PLMs as well. One potential clinical limitation of the automated analysis is the difficulty in detecting the first REM period in 10 out of the 31 subjects in this study. This resulted in poor agreement between automated scoring and the manual scorers (ICCs in the mid-40s) for calculating the REM-onset latency. As a result of these discrepancies, some manual editing will likely be required to verify the automated analysis before clinical decisions are reached. The difficulty in detecting the first REM period also suggests that the automated sleep scoring cannot be applied to Multiple Sleep Latency Test recordings when narcolepsy is suspected. This study had many strengths, including blinded manual scoring by 2 observers, the use of a clinical sleep population rather than relatively healthy subjects, and the use of automated algorithms that analyze all-important PSG outcome variables rather than just sleep staging or respiratory events. However, it also has SDB Severity M1-M2 A-M1 A-M2 Normal Normal * PPA+, % Mild Moderate Severe Mild Mild PPA+, % Moderate Severe Moderate Moderate PPA+, % Severe Severe Severe PPA+, % Total subjects, no Total agreements, no Total Agreement, % SDB refers to sleep-disordered breathing; A, automated scoring with the Morpheus I Sleep Scoring System; M 1, manual scoring for Scorer 1; M 2, manual scoring for Scorer 2; PPA+, the percentage of positive agreement for the specified level of severity [eg, PPA+ associated with normal sleepdisordered breathing (SDB) (A-M1) = 6/( ) = 66.7%]. * The number of occurrences of the specified SDB severity relative to total # of subjects. SLEEP, Vol. 27, No. 7,

10 the following limitations. First, we were unable to study consecutive patients referred to our sleep laboratory and thus have the potential for enrollment bias. However, our institutional review board will not allow the inclusion of consecutive patients and therefore only willing subjects were allowed to participate. We did take careful measures to ensure that systematic bias did not occur among the subjects that we approached; however, we cannot be confident that unrecognized factors may have affected participation. Second, although the agreement between the manual scorers in the present study compared favorably to past reports of interobserver reliability, there were scoring differences between M1 and M2, especially the scoring of stage 1 and delta sleep. Third, one could argue that the ideal assessment of an automated system would include computer analysis followed by expert verification and validation. Thus, it is possible that the combination of our automated algorithm and manual verification would have been superior to any of the approaches in our present study. In the future, a more definitive study assessing the optimal scoring techniques and the associated cost effectiveness of each of these approaches will need to be undertaken. Fourth, Medicare criteria were the only criteria used for the definition of hypopneas. Thus, we cannot comment on the ability of the algorithm to detect more subtle events such as those described in the Chicago criteria that do not require an oxyhemoglobin desaturation. 20 Fifth, previous reports have suggested greater variability across sleep laboratories in PSG scoring. Thus, our single-center study would need to be reproduced in other centers to be broadly generalizable. Sixth, we only studied an adult population with suspected SDB, thus our results cannot be generalized to other populations such as insomniacs, children, or patients with parasomnias. We conclude that the good epoch-by-epoch agreement for sleep stages, respiratory events, and limb movements lead us to believe that this automated system could be successfully implemented clinically. However, expert verification of the automated scoring is essential. During the initial implementation of the system, this might consist of verifying the scoring of every epoch. With time and expert experience with the automated system, we would expect such verification could become more limited, such as identifying the first REM period and selective validation of the algorithm sensitivity and specificity for detecting respiratory events and PLMs within an individual PSG. REFERENCES 1. Collop NA. Scoring variability between polysomnography technologists in different sleep laboratories. Sleep Med 2002;3: Norman RG, Pal I, Stewart C, Walsleben JA, Rapoport DM. Interobserver agreement among sleep scorers from different centers in a large dataset. Sleep 2000;23: Whitney CW, Gottlieb DJ, Redline S et al. Reliability of scoring respiratory disturbance indices and sleep staging. Sleep 1998;21: Prinz PN, Larsen LH, Moe KE, Dulberg EM, Vitiello MV. C STAGE, automated sleep scoring: development and comparison with human sleep scoring for healthy older men and women. Sleep 1994;17: Schaltenbrand N, Lengelle R, Toussaint M, et al. Sleep stage scoring using the neural network model: comparison between visual and automatic analysis in normal subjects and patients. Sleep 1996;19: Grube G, Flexer A, Dorffner G. Unsupervised continuous sleep analysis. Methods Find Exp Clin Pharmacol 2002;24 Suppl D: Agarwal R, Gotman J. Computer-assisted sleep staging. IEEE Trans Biomed Eng 2001;48: Taha BH, Dempsey JA, Weber SM et al. Automated detection and classification of sleep-disordered breathing from conventional polysomnography data. Sleep 1997;20: Pillar G, Bar A, Shlitner A, Schnall R, Shefy J, Lavie P. Autonomic arousal index: an automated detection based on peripheral arterial tonometry. Sleep 2002;25: Kayed K, Roberts S, Davies WL. Computer detection and analysis of periodic movements in sleep. Sleep 1990;13: EEG arousals: scoring rules and examples: a preliminary report from the Sleep Disorders Atlas Task Force of the American Sleep Disorders Association. Sleep 1992;15: Rechtschaffen A, Kales A, eds. A Manual of Standardized Terminology, Techniques, and Scoring System for Sleep Stages of Human Subjects. Los Angeles: Brain Information Service/ Brain Research Institute, UCLA; Meoli AL, Casey KR, Clark RW, et al. Hypopnea in sleep-disordered breathing in adults. Sleep 2001;24: Recording and scoring leg movements. The Atlas Task Force. Sleep 1993;16: Geva AB, Kerem DH. Brain state identification and forecasting of acute pathology using unsupervised fuzzy clustering of EEG temporal patterns. In: Teodorescu HN, Kandel A, Jain LC, eds. Fuzzy and Neuro-Fuzzy Systems in Medicine: CRC Press; 1999: Gath I, Feuerstein C, Geva AB. Unsupervised classification and adaptive definition of sleep patterns. Pattern Recognition Letters 1994;15: Geva AB. Feature Extraction and State Recognition in Biomedical Signals with Hierarchical Unsupervised Fuzzy Clustering Methods. Medical & Biomedical Engineering & Computing 1998;36: Cohen J. A coefficient of agreement for nominal scales. Educational and Psychol Meas 1960;20: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1: Sleep-related breathing disorders in adults: recommendations for syndrome definition and measurement techniques in clinical research. The Report of an American Academy of Sleep Medicine Task Force. Sleep 1999;22: Drinnan MJ, Murray A, Griffiths CJ, Gibson GJ. Interobserver variability in recognizing arousal in respiratory sleep disorders. Am J Respir Crit Care Med 1998;158: Martin SE, Engleman HM, Kingshott RN, Douglas NJ. Microarousals in patients with sleep apnoea/hypopnoea syndrome. J Sleep Res 1997;6: Kingshott RN, Engleman HM, Deary IJ, Douglas NJ. Does arousal frequency predict daytime function? Eur Respir J 1998;12: SLEEP, Vol. 27, No. 7,

Assessment of a wrist-worn device in the detection of obstructive sleep apnea

Assessment of a wrist-worn device in the detection of obstructive sleep apnea Sleep Medicine 4 (2003) 435 442 Original article Assessment of a wrist-worn device in the detection of obstructive sleep apnea Najib T. Ayas a,b,c, Stephen Pittman a,c, Mary MacDonald c, David P. White

More information

Using a Wrist-Worn Device Based on Peripheral Arterial Tonometry to Diagnose Obstructive Sleep Apnea: In-Laboratory and Ambulatory Validation

Using a Wrist-Worn Device Based on Peripheral Arterial Tonometry to Diagnose Obstructive Sleep Apnea: In-Laboratory and Ambulatory Validation Using a Wrist-Worn Device Based on Peripheral Arterial Tonometry to Diagnose Obstructive Sleep Apnea: In-Laboratory and Ambulatory Validation Stephen D. Pittman, MSBME, RPSGT 1 ; Najib T. Ayas, MD, MPH

More information

Basics of Polysomnography. Chitra Lal, MD, FCCP, FAASM Assistant professor of Medicine, Pulmonary, Critical Care and Sleep, MUSC, Charleston, SC

Basics of Polysomnography. Chitra Lal, MD, FCCP, FAASM Assistant professor of Medicine, Pulmonary, Critical Care and Sleep, MUSC, Charleston, SC Basics of Polysomnography Chitra Lal, MD, FCCP, FAASM Assistant professor of Medicine, Pulmonary, Critical Care and Sleep, MUSC, Charleston, SC Basics of Polysomnography Continuous and simultaneous recording

More information

Diagnostic Accuracy of the Multivariable Apnea Prediction (MAP) Index as a Screening Tool for Obstructive Sleep Apnea

Diagnostic Accuracy of the Multivariable Apnea Prediction (MAP) Index as a Screening Tool for Obstructive Sleep Apnea Original Article Diagnostic Accuracy of the Multivariable Apnea Prediction (MAP) Index as a Screening Tool for Obstructive Sleep Apnea Ahmad Khajeh-Mehrizi 1,2 and Omid Aminian 1 1. Occupational Sleep

More information

Frequency-domain Index of Oxyhemoglobin Saturation from Pulse Oximetry for Obstructive Sleep Apnea Syndrome

Frequency-domain Index of Oxyhemoglobin Saturation from Pulse Oximetry for Obstructive Sleep Apnea Syndrome Journal of Medical and Biological Engineering, 32(5): 343-348 343 Frequency-domain Index of Oxyhemoglobin Saturation from Pulse Oximetry for Obstructive Sleep Apnea Syndrome Liang-Wen Hang 1,2 Chen-Wen

More information

NATIONAL COMPETENCY SKILL STANDARDS FOR PERFORMING POLYSOMNOGRAPHY/SLEEP TECHNOLOGY

NATIONAL COMPETENCY SKILL STANDARDS FOR PERFORMING POLYSOMNOGRAPHY/SLEEP TECHNOLOGY NATIONAL COMPETENCY SKILL STANDARDS FOR PERFORMING POLYSOMNOGRAPHY/SLEEP TECHNOLOGY Polysomnography/Sleep Technology providers practice in accordance with the facility policy and procedure manual which

More information

The AASM Manual for the Scoring of Sleep and Associated Events

The AASM Manual for the Scoring of Sleep and Associated Events The AASM Manual for the Scoring of Sleep and Associated Events Summary of Updates in Version 2.1 July 1, 2014 The American Academy of Sleep Medicine (AASM) is committed to ensuring that The AASM Manual

More information

Coding for Sleep Disorders Jennifer Rose V. Molano, MD

Coding for Sleep Disorders Jennifer Rose V. Molano, MD Practice Coding for Sleep Disorders Jennifer Rose V. Molano, MD Accurate coding is an important function of neurologic practice. This section of is part of an ongoing series that presents helpful coding

More information

Non-contact Screening System with Two Microwave Radars in the Diagnosis of Sleep Apnea-Hypopnea Syndrome

Non-contact Screening System with Two Microwave Radars in the Diagnosis of Sleep Apnea-Hypopnea Syndrome Medinfo2013 Decision Support Systems and Technologies - II Non-contact Screening System with Two Microwave Radars in the Diagnosis of Sleep Apnea-Hypopnea Syndrome 21 August 2013 M. Kagawa 1, K. Ueki 1,

More information

The AASM Manual for the Scoring of Sleep and Associated Events

The AASM Manual for the Scoring of Sleep and Associated Events The AASM Manual for the Scoring of Sleep and Associated Events The 2007 AASM Scoring Manual vs. the AASM Scoring Manual v2.0 October 2012 The American Academy of Sleep Medicine (AASM) is committed to ensuring

More information

Key words: Medicare; obstructive sleep apnea; oximetry; sleep apnea syndromes

Key words: Medicare; obstructive sleep apnea; oximetry; sleep apnea syndromes Choice of Oximeter Affects Apnea- Hypopnea Index* Subooha Zafar, MD; Indu Ayappa, PhD; Robert G. Norman, PhD; Ana C. Krieger, MD, FCCP; Joyce A. Walsleben, PhD; and David M. Rapoport, MD, FCCP Study objectives:

More information

Validation of a Self-Applied Unattended Monitor for Sleep Disordered Breathing

Validation of a Self-Applied Unattended Monitor for Sleep Disordered Breathing Scientific investigations Validation of a Self-Applied Unattended Monitor for Sleep Disordered Breathing Indu Ayappa, Ph.D.; Robert G. Norman, Ph.D.; Vijay Seelall, M.D.; David M. Rapoport, M.D. Division

More information

Summary of Features and Performance

Summary of Features and Performance MICHELE SLEEP SCORING SYSTEM Summary of Features and Performance Suite PE438, Princess Elizabeth Building 1 Morley Ave / Winnipeg, Manitoba / R3L 2P4 Canada phone 1 877 949 3202 / fax 204 943 6295 Table

More information

Autonomic Arousal Index: an Automated Detection Based on Peripheral Arterial Tonometry

Autonomic Arousal Index: an Automated Detection Based on Peripheral Arterial Tonometry INSTRUMENTATION & METHODOLOGY Autonomic Arousal Index: an Automated Detection Based on Peripheral Arterial Tonometry Giora Pillar, 1 Amir Bar, 2 Arie Shlitner, 1 Robert Schnall, 2 Jacob Shefy, 2 and Peretz

More information

FAQ CODING & REIMBURSEMENT. WatchPAT TM Home Sleep Test

FAQ CODING & REIMBURSEMENT. WatchPAT TM Home Sleep Test FAQ CODING & REIMBURSEMENT WatchPAT TM Home Sleep Test TABLE OF CONTENTS PATIENT SELECTION CRITERIA 3 CODING & MODIFIERS 4-6 PLACE OF SERVICE 6 FREQUENCY 7 ACCREDITATION 7 SLEEP MEDICINE GLOSSARY AND ACRONYMS

More information

Portable Computerized Polysomnography in Attended and Unattended Settings*

Portable Computerized Polysomnography in Attended and Unattended Settings* Portable Computerized Polysomnography in Attended and Unattended Settings* Ivanka J. Mykytyn; Dimitar Sajkov, MBBS, PhD; Alister M. Neill, MBBS, FRACP; and R. Douglas McEvoy, MBBS, MD, FRACP Study objective:

More information

Web-Based Home Sleep Testing

Web-Based Home Sleep Testing Editorial Web-Based Home Sleep Testing Authors: Matthew Tarler, Ph.D., Sarah Weimer, Craig Frederick, Michael Papsidero M.D., Hani Kayyali Abstract: Study Objective: To assess the feasibility and accuracy

More information

There has been a substantial increase in the number of studies. The Validity of Wrist Actimetry Assessment of Sleep With and Without Sleep Apnea

There has been a substantial increase in the number of studies. The Validity of Wrist Actimetry Assessment of Sleep With and Without Sleep Apnea Scientific investigations The Validity of Wrist Actimetry Assessment of Sleep With and Without Sleep Apnea David Wang, Ph.D. 1,2 ; Keith K. Wong, MB. BS. 1,2 ; George C. Dungan II, RPSGT 1 ; Peter R. Buchanan,

More information

CLASSIFICATION OF SLEEP STAGES IN INFANTS: A NEURO FUZZY APPROACH

CLASSIFICATION OF SLEEP STAGES IN INFANTS: A NEURO FUZZY APPROACH CLASSIFICATION OF SLEEP STAGES IN INFANTS: A NEURO FUZZY APPROACH J. E. Heiss, C. M. Held, P. A. Estévez, C. A. Perez, C. A. Holzmann, J. P. Pérez Department of Electrical Engineering, Universidad de Chile,

More information

A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation

A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation 1 A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation The following 3 minute polysomnogram (PSG) tracing was recorded in a 74-year-old man with severe ischemic cardiomyopathy

More information

Automated analysis of digital oximetry in the diagnosis of obstructive sleep apnoea

Automated analysis of digital oximetry in the diagnosis of obstructive sleep apnoea 302 Division of Respiratory Medicine, Department of Medicine, University of Calgary, Calgary, Alberta, Canada T2N 4N1 J-C Vázquez W H Tsai W W Flemons A Masuda R Brant E Hajduk W A Whitelaw J E Remmers

More information

There have been many traffic accidents in our country, and the

There have been many traffic accidents in our country, and the http://dx.doi.org/1.5664/jcsm.2584 Reliability of the Watch-PAT 2 in Detecting Sleep Apnea in Highway bus Drivers Melike Yuceege, M.D. 1 ; Hikmet Firat, M.D. 1 ; Ahmet Demir, M.D. 2 ; Sadik Ardic, M.D.

More information

Arousal detection in sleep

Arousal detection in sleep Arousal detection in sleep FW BES, H KUYKENS AND A KUMAR MEDCARE AUTOMATION, OTTHO HELDRINGSTRAAT 27 1066XT AMSTERDAM, THE NETHERLANDS Introduction Arousals are part of normal sleep. They become pathological

More information

DECLARATION OF CONFLICT OF INTEREST

DECLARATION OF CONFLICT OF INTEREST DECLARATION OF CONFLICT OF INTEREST Obstructive sleep apnoea How to identify? Walter McNicholas MD Newman Professor in Medicine, St. Vincent s University Hospital, University College Dublin, Ireland. Potential

More information

Split Night Protocols for Adult Patients - Updated July 2012

Split Night Protocols for Adult Patients - Updated July 2012 Split Night Protocols for Adult Patients - Updated July 2012 SUMMARY: Sleep technologists are team members who work under the direction of a physician practicing sleep disorders medicine. Sleep technologists

More information

* Cedars Sinai Medical Center, Los Angeles, California, U.S.A.

* Cedars Sinai Medical Center, Los Angeles, California, U.S.A. Sleep. 18(2):115-126 1995 American Sleep Disorders Association and Sleep Research Society Home Monitoring-Actimetry Assessment of Accuracy and Analysis Time of a Novel Device to Monitor Sleep and Breathing

More information

Nasal pressure recording in the diagnosis of sleep apnoea hypopnoea syndrome

Nasal pressure recording in the diagnosis of sleep apnoea hypopnoea syndrome 56 Unité de Recherche, Centre de Pneumologie de l Hôpital Laval, Université Laval, Québec, Canada F Sériès I Marc Correspondence to: Dr F Sériès, Centre de Pneumologie, 2725 Chemin Sainte Foy, Sainte Foy

More information

Obstructive sleep apnoea How to identify?

Obstructive sleep apnoea How to identify? Obstructive sleep apnoea How to identify? Walter McNicholas MD Newman Professor in Medicine, St. Vincent s University Hospital, University College Dublin, Ireland. Potential conflict of interest None Obstructive

More information

AASM guidelines, when available. Does this mean if our medical director chooses for us to use an alternative rule that our accreditation is at risk?

AASM guidelines, when available. Does this mean if our medical director chooses for us to use an alternative rule that our accreditation is at risk? GENERAL G.1. I see that the STANDARDS FOR ACCREDITATION state that we are to use the recommended AASM guidelines, when available. Does this mean if our medical director chooses for us to use an alternative

More information

Internet Journal of Medical Update

Internet Journal of Medical Update Internet Journal of Medical Update 2009 July;4(2):24-28 Internet Journal of Medical Update Journal home page: http://www.akspublication.com/ijmu Original Work EEG arousal prediction via hypoxemia indicator

More information

Simple diagnostic tools for the Screening of Sleep Apnea in subjects with high risk of cardiovascular disease

Simple diagnostic tools for the Screening of Sleep Apnea in subjects with high risk of cardiovascular disease Cardiovascular diseases remain the number one cause of death worldwide Simple diagnostic tools for the Screening of Sleep Apnea in subjects with high risk of cardiovascular disease Shaoguang Huang MD Department

More information

EEG Arousals: Scoring Rules and Examples. A Preliminary Report from the Sleep Disorders Atlas Task Force of the American Sleep Disorders Association

EEG Arousals: Scoring Rules and Examples. A Preliminary Report from the Sleep Disorders Atlas Task Force of the American Sleep Disorders Association EEG Arousals: Scoring Rules and Examples A Preliminary Report from the Sleep Disorders Atlas Task Force of the American Sleep Disorders Association Sleep in patients with a number of sleep disorders and

More information

In 1994, the American Sleep Disorders Association

In 1994, the American Sleep Disorders Association Unreliability of Automatic Scoring of MESAM 4 in Assessing Patients With Complicated Obstructive Sleep Apnea Syndrome* Fabio Cirignotta, MD; Susanna Mondini, MD; Roberto Gerardi, MD Barbara Mostacci, MD;

More information

The recommended method for diagnosing sleep

The recommended method for diagnosing sleep reviews Measuring Agreement Between Diagnostic Devices* W. Ward Flemons, MD; and Michael R. Littner, MD, FCCP There is growing interest in using portable monitoring for investigating patients with suspected

More information

Polysomnography Course Session: Sept 2017

Polysomnography Course Session: Sept 2017 Polysomnography Course Session: Sept 2017 General Information Polysomnography course will be held at SLEEP AND ALERTNESS CLINIC Med-West Medical centre 750 Dundas St. W., Suite 2-259 (Conference Room)

More information

Simplest method: Questionnaires. Retrospective: past week, month, year, lifetime Daily: Sleep diary What kinds of questions would you ask?

Simplest method: Questionnaires. Retrospective: past week, month, year, lifetime Daily: Sleep diary What kinds of questions would you ask? Spencer Dawson Simplest method: Questionnaires Retrospective: past week, month, year, lifetime Daily: Sleep diary What kinds of questions would you ask? Did you nap during the day? Bed time and rise time

More information

The STOP-Bang Equivalent Model and Prediction of Severity

The STOP-Bang Equivalent Model and Prediction of Severity DOI:.5664/JCSM.36 The STOP-Bang Equivalent Model and Prediction of Severity of Obstructive Sleep Apnea: Relation to Polysomnographic Measurements of the Apnea/Hypopnea Index Robert J. Farney, M.D. ; Brandon

More information

Effectiveness of Portable Monitoring Devices for Diagnosing Obstructive Sleep Apnea: Update of a Systematic Review

Effectiveness of Portable Monitoring Devices for Diagnosing Obstructive Sleep Apnea: Update of a Systematic Review Effectiveness of Portable Monitoring Devices for Diagnosing Obstructive Sleep Apnea: Update of a Systematic Review Submitted to: Agency for Healthcare Research and Quality 540 Gaither Road Rockville, Maryland

More information

Out of Center Sleep Testing (OCST) - Updated July 2012

Out of Center Sleep Testing (OCST) - Updated July 2012 Out of Center Sleep Testing (OCST) - Updated July 2012 SUMMARY: Out of center sleep testing (OCST) can be used as an alternative to full, attended polysomnography (PSG) for the diagnosis of obstructive

More information

Comparison of Nasal Prong Pressure and Thermistor Measurements for Detecting Respiratory Events during Sleep

Comparison of Nasal Prong Pressure and Thermistor Measurements for Detecting Respiratory Events during Sleep Clinical Investigations Respiration 2004;71:385 390 DOI: 10.1159/000079644 Received: August 25, 2003 Accepted after revision: March 16, 2004 Comparison of Nasal Prong Pressure and Thermistor Measurements

More information

The International Franco - Palestinian Congress in Sleep Medicine

The International Franco - Palestinian Congress in Sleep Medicine The International Franco - Palestinian Congress in Sleep Medicine Temporomandibular Disorders and Sleep Apnea 26 and 27 October, 2017 Notre Dame Hotel, Jerusalem Polysomnography Reports Interpreting the

More information

The Latest Technology from CareFusion

The Latest Technology from CareFusion The Latest Technology from CareFusion Contents 1 Introduction... 2 1.1 Overview... 2 1.2 Scope... 2 2.1 Input Recordings... 2 2.2 Automatic Analysis... 3 2.3 Data Mining... 3 3 Results... 4 3.1 AHI comparison...

More information

FEP Medical Policy Manual

FEP Medical Policy Manual FEP Medical Policy Manual Effective Date: October 15, 2018 Related Policies: 2.01.18 Diagnosis and Medical Management of Obstructive Sleep Apnea Syndrome Polysomnography for Non-Respiratory Sleep Disorders

More information

Introducing the WatchPAT 200 # 1 Home Sleep Study Device

Introducing the WatchPAT 200 # 1 Home Sleep Study Device Introducing the WatchPAT 200 # 1 Home Sleep Study Device Top 10 Medical Innovation for 2010 Cleveland Clinic Fidelis Diagnostics & Itamar Medical Fidelis Diagnostics founded in 2004, is a privately-held

More information

Utility of Technologist Editing of Polysomnography Scoring Performed by a Validated Automatic System

Utility of Technologist Editing of Polysomnography Scoring Performed by a Validated Automatic System Utility of Technologist Editing of Polysomnography Scoring Performed by a Validated Automatic System Magdy Younes 1,2, Wayne Thompson 1, Colleen Leslie 1, Tanya Egan 1, and Eleni Giannouli 1 1 Sleep Disorders

More information

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Testing the Accuracy of ECG Captured by through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Data Analysis a) R-wave Comparison: The mean and standard deviation of R-wave amplitudes

More information

Critical Review Form Diagnostic Test

Critical Review Form Diagnostic Test Critical Review Form Diagnostic Test Diagnosis and Initial Management of Obstructive Sleep Apnea without Polysomnography A Randomized Validation Study Annals of Internal Medicine 2007; 146: 157-166 Objectives:

More information

Validating the Watch-PAT for Diagnosing Obstructive Sleep. Apnea in Adolescents

Validating the Watch-PAT for Diagnosing Obstructive Sleep. Apnea in Adolescents Validating the Watch-PAT for Diagnosing Obstructive Sleep Apnea in Adolescents Ji Ho Choi, MD, PhD 1 ; Bora Lee 2 ; Jae Yong Lee, MD, PhD 1 ; Hyun Jun Kim, MD, PhD 3 1 Department of Otorhinolaryngology-Head

More information

Outlining a simple and robust method for the automatic detection of EEG arousals

Outlining a simple and robust method for the automatic detection of EEG arousals Outlining a simple and robust method for the automatic detection of EEG arousals Isaac Ferna ndez-varela1, Diego A lvarez-este vez2, Elena Herna ndez-pereira1 and Vicente Moret-Bonillo1 1- Universidade

More information

EFFICACY OF MODAFINIL IN 10 TAIWANESE PATIENTS WITH NARCOLEPSY: FINDINGS USING THE MULTIPLE SLEEP LATENCY TEST AND EPWORTH SLEEPINESS SCALE

EFFICACY OF MODAFINIL IN 10 TAIWANESE PATIENTS WITH NARCOLEPSY: FINDINGS USING THE MULTIPLE SLEEP LATENCY TEST AND EPWORTH SLEEPINESS SCALE EFFICACY OF MODAFINIL IN 10 TAIWANESE PATIENTS WITH NARCOLEPSY: FINDINGS USING THE MULTIPLE SLEEP LATENCY TEST AND EPWORTH SLEEPINESS SCALE Shih-Bin Yeh 1 and Carlos Hugh Schenck 2,3 1 Department of Neurology

More information

Obstructive sleep apnea (OSA) is characterized by. Quality of Life in Patients with Obstructive Sleep Apnea*

Obstructive sleep apnea (OSA) is characterized by. Quality of Life in Patients with Obstructive Sleep Apnea* Quality of Life in Patients with Obstructive Sleep Apnea* Effect of Nasal Continuous Positive Airway Pressure A Prospective Study Carolyn D Ambrosio, MD; Teri Bowman, MD; and Vahid Mohsenin, MD Background:

More information

O bstructive sleep apnoea-hypopnoea (OSAH) is a highly

O bstructive sleep apnoea-hypopnoea (OSAH) is a highly 422 SLEEP-DISORDERED BREATHING Predictive value of automated oxygen saturation analysis for the diagnosis and treatment of obstructive sleep apnoea in a home-based setting V Jobin, P Mayer, F Bellemare...

More information

The identification of obstructive apneas and hypopneas in

The identification of obstructive apneas and hypopneas in SCIENTIFIC INVESTIGATIONS pii: jc-00069-15 http://dx.doi.org/10.5664/jcsm.5280 The 2012 AASM Respiratory Event Criteria Increase the Incidence of Hypopneas in an Adult Sleep Center Population Brett Duce,

More information

Differentiating Obstructive from Central and Complex Sleep Apnea Using an Automated Electrocardiogram-Based Method

Differentiating Obstructive from Central and Complex Sleep Apnea Using an Automated Electrocardiogram-Based Method SLEEP APNEA Differentiating Obstructive from Central and Complex Sleep Apnea Using an Automated Electrocardiogram-Based Method Robert Joseph Thomas, MD, MMSc 1 ; Joseph E. Mietus, BS 2 ; Chung-Kang Peng,

More information

The Familial Occurrence of Obstructive Sleep Apnoea Syndrome (OSAS)

The Familial Occurrence of Obstructive Sleep Apnoea Syndrome (OSAS) Global Journal of Respiratory Care, 2014, 1, 17-21 17 The Familial Occurrence of Obstructive Sleep Apnoea Syndrome (OSAS) Piotr Bielicki, Tadeusz Przybylowski, Ryszarda Chazan * Department of Internal

More information

Procedures in the Sleep Laboratory

Procedures in the Sleep Laboratory AAST Technologist Fundamentals Date: May 7, 2017 Focus Conference Location: Orlando, Florida Workshop Procedures in the Sleep Laboratory Laree Fordyce, RST, RPSGT, CCRP Conflict of Interest Disclosures

More information

NORAH Sleep Study External Comment Mathias Basner, MD, PhD, MSc

NORAH Sleep Study External Comment Mathias Basner, MD, PhD, MSc NORAH Sleep Study External Comment Mathias Basner, MD, PhD, MSc University of Pennsylvania Perelman School of Medicine Page 1 > Mathias Basner Disclaimer The University of Pennsylvania and the German Aerospace

More information

In-Patient Sleep Testing/Management Boaz Markewitz, MD

In-Patient Sleep Testing/Management Boaz Markewitz, MD In-Patient Sleep Testing/Management Boaz Markewitz, MD Objectives: Discuss inpatient sleep programs and if they provide a benefit to patients and sleep centers Identify things needed to be considered when

More information

Prediction of sleep-disordered breathing by unattended overnight oximetry

Prediction of sleep-disordered breathing by unattended overnight oximetry J. Sleep Res. (1999) 8, 51 55 Prediction of sleep-disordered breathing by unattended overnight oximetry L. G. OLSON, A. AMBROGETTI ands. G. GYULAY Discipline of Medicine, University of Newcastle and Sleep

More information

The Effect of Altitude Descent on Obstructive Sleep Apnea*

The Effect of Altitude Descent on Obstructive Sleep Apnea* CHEST The Effect of Altitude Descent on Obstructive Sleep Apnea* David Patz, MD, FCCP; Mark Spoon, RPSGT; Richard Corbin, RPSGT; Michael Patz, BA; Louise Dover, RPSGT; Bruce Swihart, MA; and David White,

More information

Methods of Diagnosing Sleep Apnea. The Diagnosis of Sleep Apnea: Questionnaires and Home Studies

Methods of Diagnosing Sleep Apnea. The Diagnosis of Sleep Apnea: Questionnaires and Home Studies Sleep, 19(10):S243-S247 1996 American Sleep Disorders Association and Sleep Research Society Methods of Diagnosing Sleep Apnea J The Diagnosis of Sleep Apnea: Questionnaires and Home Studies W. Ward Flemons

More information

Sleep Bruxism and Sleep-Disordered Breathing

Sleep Bruxism and Sleep-Disordered Breathing Sleep Bruxism and Sleep-Disordered Breathing Author STEVEN D BENDER, DDS*, Associate Editor EDWARD J. SWIFT JR., DMD, MS Sleep bruxism (SB) is a repetitive jaw muscle activity with clenching or grinding

More information

Effect of body mass index on overnight oximetry for the diagnosis of sleep apnea

Effect of body mass index on overnight oximetry for the diagnosis of sleep apnea Respiratory Medicine (2004) 98, 421 427 Effect of body mass index on overnight oximetry for the diagnosis of sleep apnea Hiroshi Nakano*, Togo Ikeda, Makito Hayashi, Etsuko Ohshima, Michiko Itoh, Nahoko

More information

Proceedings 23rd Annual Conference IEEE/EMBS Oct.25-28, 2001, Istanbul, TURKEY

Proceedings 23rd Annual Conference IEEE/EMBS Oct.25-28, 2001, Istanbul, TURKEY AUTOMATED SLEEP STAGE SCORING BY DECISION TREE LEARNING Proceedings 23rd Annual Conference IEEE/EMBS Oct.25-28, 2001, Istanbul, TURKEY Masaaki Hanaoka, Masaki Kobayashi, Haruaki Yamazaki Faculty of Engineering,Yamanashi

More information

(To be filled by the treating physician)

(To be filled by the treating physician) CERTIFICATE OF MEDICAL NECESSITY TO BE ISSUED TO CGHS BENEFICIAREIS BEING PRESCRIBED BILEVEL CONTINUOUS POSITIVE AIRWAY PRESSURE (BI-LEVEL CPAP) / BI-LEVEL VENTILATORY SUPPORT SYSTEM Certification Type

More information

Published Papers Cardio Pulmonary Coupling

Published Papers Cardio Pulmonary Coupling Thomas RJ, Mietus JE, Peng CK, Goldberger AL An electrocardiogram-based technique to assess cardiopulmonary coupling (CPC) during sleep SLEEP 2005; 28:1151-61 O: Evaluate a new automated measure of cardiopulmonary

More information

Validating the Watch-PAT for Diagnosing Obstructive Sleep Apnea in Adolescents

Validating the Watch-PAT for Diagnosing Obstructive Sleep Apnea in Adolescents pii: jc-18-00029 http://dx.doi.org/10.5664/jcsm.7386 SCIENTIFIC INVESTIGATIONS Validating the Watch-PAT for Diagnosing Obstructive Sleep Apnea in Adolescents Ji Ho Choi, MD, PhD 1 ; Bora Lee 2 ; Jae Yong

More information

Biomedical Signal Processing

Biomedical Signal Processing DSP : Biomedical Signal Processing What is it? Biomedical Signal Processing: Application of signal processing methods, such as filtering, Fourier transform, spectral estimation and wavelet transform, to

More information

Polysomnography Artifacts and Updates on AASM Scoring Rules. Robin Lloyd, MD, FAASM, FAAP 2017 Utah Sleep Society Conference

Polysomnography Artifacts and Updates on AASM Scoring Rules. Robin Lloyd, MD, FAASM, FAAP 2017 Utah Sleep Society Conference Polysomnography Artifacts and Updates on AASM Scoring Rules Robin Lloyd, MD, FAASM, FAAP 2017 Utah Sleep Society Conference x Conflict of Interest Disclosures for Speakers 1. I do not have any relationships

More information

Polysomnography (PSG) (Sleep Studies), Sleep Center

Polysomnography (PSG) (Sleep Studies), Sleep Center Policy Number: 1036 Policy History Approve Date: 07/09/2015 Effective Date: 07/09/2015 Preauthorization All Plans Benefit plans vary in coverage and some plans may not provide coverage for certain service(s)

More information

Effect of Manual Editing of Total Recording Time: Implications for Home Sleep Apnea Testing

Effect of Manual Editing of Total Recording Time: Implications for Home Sleep Apnea Testing pii: jc-00225-16 http://dx.doi.org/10.5664/jcsm.6404 SCIENTIFIC INVESTIGATIONS Effect of Manual Editing of Total Recording Time: Implications for Home Sleep Apnea Testing Ying Y. Zhao, MD, MPH 1,2 ; Jia

More information

A Sleep Laboratory Evaluation of an Automatic Positive Airway Pressure System for Treatment of Obstructive Sleep Apnea

A Sleep Laboratory Evaluation of an Automatic Positive Airway Pressure System for Treatment of Obstructive Sleep Apnea A Sleep Laboratory Evaluation of an Automatic Positive Airway Pressure System for Treatment of Obstructive Sleep Apnea Khosrow Behbehani, 1 Fu-Chung Yen, 1 Edgar A. Lucas, 2 and John R. Burk 2 (1) Joint

More information

linkedin.com/in/lizziehillsleeptechservices 1

linkedin.com/in/lizziehillsleeptechservices  1 BSS2015 Hands-On Tech Breakfast SCORING SLEEP USING AASM GUIDELINES: A BRIEF INTRODUCTION Lizzie Hill BSc RPSGT EST Specialist Respiratory Clinical Physiologist, Royal Hospital for Sick Children, Edinburgh

More information

Κλινικό Φροντιστήριο Αναγνώριση και καταγραφή αναπνευστικών επεισοδίων Λυκούργος Κολιλέκας Επιμελητής A ΕΣΥ 7η Πνευμονολογική Κλινική ΝΝΘΑ Η ΣΩΤΗΡΙΑ

Κλινικό Φροντιστήριο Αναγνώριση και καταγραφή αναπνευστικών επεισοδίων Λυκούργος Κολιλέκας Επιμελητής A ΕΣΥ 7η Πνευμονολογική Κλινική ΝΝΘΑ Η ΣΩΤΗΡΙΑ Κλινικό Φροντιστήριο Αναγνώριση και καταγραφή αναπνευστικών επεισοδίων Λυκούργος Κολιλέκας Επιμελητής A ΕΣΥ 7 η Πνευμονολογική Κλινική ΝΝΘΑ Η ΣΩΤΗΡΙΑ SCORING SLEEP -Rechtschaffen and Kales (1968) - AASM

More information

Pulse Rate Variability Analysis to Enhance Oximetry as at-home Alternative for Sleep Apnea Diagnosing

Pulse Rate Variability Analysis to Enhance Oximetry as at-home Alternative for Sleep Apnea Diagnosing Pulse Rate Variability Analysis to Enhance Oximetry as at-home Alternative for Sleep Apnea Diagnosing Gonzalo C. Gutiérrez-Tobal, Daniel Álvarez, Fernando Vaquerizo-Villar, Verónica Barroso-García, Adrián

More information

Periodic Leg Movement, L-Dopa, 5-Hydroxytryptophan, and L-Tryptophan

Periodic Leg Movement, L-Dopa, 5-Hydroxytryptophan, and L-Tryptophan Sleep 10(4):393-397, Raven Press, New York 1987, Association of Professional Sleep Societies Short Report Periodic Leg Movement, L-Dopa, 5-Hydroxytryptophan, and L-Tryptophan C. Guilleminault, S. Mondini,

More information

LEARNING MANUAL OF PSG CHART

LEARNING MANUAL OF PSG CHART LEARNING MANUAL OF PSG CHART POLYSOMNOGRAM, SLEEP STAGE SCORING, INTERPRETATION Sleep Computing Committee, Japanese Society of Sleep Research LEARNING MANUAL OF PSG CHART POLYSOMNOGRAM, SLEEP STAGE SCORING,

More information

Appendix 1. Practice Guidelines for Standards of Adult Sleep Medicine Services

Appendix 1. Practice Guidelines for Standards of Adult Sleep Medicine Services Appendix 1 Practice Guidelines for Standards of Adult Sleep Medicine Services 1 Premises and Procedures Out-patient/Clinic Rooms Sleep bedroom for PSG/PG Monitoring/Analysis/ Scoring room PSG equipment

More information

Medicare CPAP/BIPAP Coverage Criteria

Medicare CPAP/BIPAP Coverage Criteria Medicare CPAP/BIPAP Coverage Criteria For any item to be covered by Medicare, it must 1) be eligible for a defined Medicare benefit category, 2) be reasonable and necessary for the diagnosis or treatment

More information

Efremidis George, Varela Katerina, Spyropoulou Maria, Beroukas Lambros, Nikoloutsou Konstantina, and Georgopoulos Dimitrios

Efremidis George, Varela Katerina, Spyropoulou Maria, Beroukas Lambros, Nikoloutsou Konstantina, and Georgopoulos Dimitrios Sleep Disorders Volume 2012, Article ID 324635, 5 pages doi:10.1155/2012/324635 Clinical Study Clinical Features and Polysomnographic Findings in Greek Male Patients with Obstructive Sleep Apnea Syndrome:

More information

Western Hospital System. PSG in History. SENSORS in the field of SLEEP. PSG in History continued. Remember

Western Hospital System. PSG in History. SENSORS in the field of SLEEP. PSG in History continued. Remember SENSORS in the field of SLEEP Mrs. Gaye Cherry: Scientist in Charge Department of Sleep and Respiratory Medicine Sleep Disorders Unit Western Hospital PSG in History 1875: Discovery of brain-wave activity

More information

FEP Medical Policy Manual

FEP Medical Policy Manual FEP Medical Policy Manual Effective Date: January 15, 2018 Related Policies: 2.01.18 Diagnosis and Medical Management of Obstructive Sleep Apnea Syndrome Diagnosis and Medical Management of Obstructive

More information

Patterns of Sleepiness in Various Disorders of Excessive Daytime Somnolence

Patterns of Sleepiness in Various Disorders of Excessive Daytime Somnolence Sleep, 5:S165S174 1982 Raven Press, New York Patterns of Sleepiness in Various Disorders of Excessive Daytime Somnolence F. Zorick, T. Roehrs, G. Koshorek, J. Sicklesteel, *K. Hartse, R. Wittig, and T.

More information

PEDIATRIC SLEEP GUIDELINES Version 1.0; Effective

PEDIATRIC SLEEP GUIDELINES Version 1.0; Effective MedSolutions, Inc. Clinical Decision Support Tool Diagnostic Strategies This tool addresses common symptoms and symptom complexes. Requests for patients with atypical symptoms or clinical presentations

More information

Evaluation of a 2-Channel Portable Device and a Predictive Model to Screen for Obstructive Sleep Apnea in a Laboratory Environment

Evaluation of a 2-Channel Portable Device and a Predictive Model to Screen for Obstructive Sleep Apnea in a Laboratory Environment Evaluation of a 2-Channel Portable Device and a Predictive Model to Screen for Obstructive Sleep Apnea in a Laboratory Environment Jianyin Zou, Lili Meng MD, Yupu Liu MSc, Xiaoxi Xu, Suru Liu PhD, Jian

More information

Obstructive Sleep Apnea

Obstructive Sleep Apnea Obstructive Sleep Apnea by Barbara Phillips MD MSPH and Matthew T Naughton MD FRACP Epidemiology and risk factors 7 Clinical presentation 13 Medical complications 22 Diagnosis 40 Medical management 50

More information

ALMA MATER STUDIORUM UNIVERSITÀ DI BOLOGNA

ALMA MATER STUDIORUM UNIVERSITÀ DI BOLOGNA ALMA MATER STUDIORUM UNIVERSITÀ DI BOLOGNA DOTTORATO DI RICERCA IN Scienze Mediche e Specialistiche Ciclo XXVII Settore Concorsuale di afferenza: 06/D6 Settore Scientifico disciplinare: MED 26 ANALYSIS

More information

Sleep disordered breathing is a spectrum of diseases that SCIENTIFIC INVESTIGATIONS

Sleep disordered breathing is a spectrum of diseases that SCIENTIFIC INVESTIGATIONS http://dx.doi.org/xxxxxxxxxxxxxxx Clinical Usefulness of Watch-PAT for Assessing the Surgical Results of Obstructive Sleep Apnea Syndrome Chong Yoon Park, M.D.; Joon Hyeong Hong, M.D.; Jae Heon Lee, M.D.;

More information

SPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS

SPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS SPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS Vessela Tzvetanova Krasteva, Irena Ilieva Jekova Centre of Biomedical Engineering Prof. Ivan Daskalov - Bulgarian Academy of Sciences Acad.G.Bonchev

More information

PREDICTIVE VALUE OF AUTOMATED OXYGEN SATURATION ANALYSIS FOR THE DIAGNOSIS AND TREATMENT OF OBSTRUCTIVE SLEEP APNEA IN A HOME-BASED SETTING

PREDICTIVE VALUE OF AUTOMATED OXYGEN SATURATION ANALYSIS FOR THE DIAGNOSIS AND TREATMENT OF OBSTRUCTIVE SLEEP APNEA IN A HOME-BASED SETTING Thorax Online First, published on January 24, 2007 as 10.1136/thx.2006.061234 PREDICTIVE VALUE OF AUTOMATED OXYGEN SATURATION ANALYSIS FOR THE DIAGNOSIS AND TREATMENT OF OBSTRUCTIVE SLEEP APNEA IN A HOME-BASED

More information

The Epworth Sleepiness Scale (ESS) was developed by Johns

The Epworth Sleepiness Scale (ESS) was developed by Johns Clinical Reproducibility of the Epworth Sleepiness Scale Anh Tu Duy Nguyen, M.D. 1 ; Marc A. Baltzan, M.D., M.Sc. 1,2 ; David Small, M.D. 1 ; Norman Wolkove, M.D. 1 ; Simone Guillon, M.D. 3 ; Mark Palayew,

More information

Respiratory Event Detection by a Positive Airway Pressure Device

Respiratory Event Detection by a Positive Airway Pressure Device RESPIRATORY EVENT DETECTION BY A POSITIVE AIRWAY PRESSURE DEVICE http://dx.doi.org/.5665/sleep.1696 Respiratory Event Detection by a Positive Airway Pressure Device Richard B. Berry, MD 1 ; Clete A. Kushida,

More information

INFORMATION SOCIETY TECHNOLOGIES (IST) PROGRAMME

INFORMATION SOCIETY TECHNOLOGIES (IST) PROGRAMME INFORMATION SOCIETY TECHNOLOGIES (IST) PROGRAMME SENSATION 507231 Report on definition of home care needs, security requirements, literacy and educational level, and attitudes of average patients in the

More information

Obstructive sleep apnea (OSA) is common but underdiagnosed

Obstructive sleep apnea (OSA) is common but underdiagnosed The Role of Single-Channel Nasal Airflow Pressure Transducer in the Diagnosis of OSA in the Sleep Laboratory Lydia Makarie Rofail, M.B.B.S., Ph.D. 1,2,3 ; Keith K. H. Wong, M.B.B.S., Ph.D. 1,2,4 ; Gunnar

More information

Key words: adenotonsillectomy; arousal; rapid eye movement sleep; sleep apnea

Key words: adenotonsillectomy; arousal; rapid eye movement sleep; sleep apnea Sleep Characteristics Following Adenotonsillectomy in Children With Obstructive Sleep Apnea Syndrome* Asher Tal, MD; Amir Bar, MD; Alberto Leiberman, MD; and Ariel Tarasiuk, PhD Objective: To compare the

More information

The amount of nightly variability that occurs in sleepdisordered. Nightly variability of sleep-disordered breathing measured over 3 nights

The amount of nightly variability that occurs in sleepdisordered. Nightly variability of sleep-disordered breathing measured over 3 nights Nightly variability of sleep-disordered breathing measured over 3 nights CARL J. STEPNOWSKY, Jr, PHD, WILLIAM C. ORR, PHD, and TERENCE M. DAVIDSON, MD, San Diego, California, and Oklahoma City, Oklahoma

More information

RESEARCH PACKET DENTAL SLEEP MEDICINE

RESEARCH PACKET DENTAL SLEEP MEDICINE RESEARCH PACKET DENTAL SLEEP MEDICINE American Academy of Dental Sleep Medicine Dental Sleep Medicine Research Packet Page 1 Table of Contents Research: Oral Appliance Therapy vs. Continuous Positive Airway

More information

Milena Pavlova, M.D., FAASM Department of Neurology, Brigham and Women's Hospital Assistant Professor of Neurology, Harvard Medical School Medical

Milena Pavlova, M.D., FAASM Department of Neurology, Brigham and Women's Hospital Assistant Professor of Neurology, Harvard Medical School Medical Milena Pavlova, M.D., FAASM Department of Neurology, Brigham and Women's Hospital Assistant Professor of Neurology, Harvard Medical School Medical Director, Faulkner EEG and Sleep Testing Center Course

More information

Sleep 101. Kathleen Feeney RPSGT, RST, CSE Business Development Specialist

Sleep 101. Kathleen Feeney RPSGT, RST, CSE Business Development Specialist Sleep 101 Kathleen Feeney RPSGT, RST, CSE Business Development Specialist 2016 Why is Sleep Important More than one-third of the population has trouble sleeping (Gallup) Obstructive Sleep Apnea Untreated

More information