Respiratory Physiology & Neurobiology

Size: px
Start display at page:

Download "Respiratory Physiology & Neurobiology"

Transcription

1 Respiratory Physiology & Neurobiology 162 (2008) Contents lists available at ScienceDirect Respiratory Physiology & Neurobiology journal homepage: Bench study of auto-cpap devices using a collapsible upper airway model with upstream resistance Minoru Hirose a, Junichi Honda b, Eiji Sato a, Toshihiro Shinbo a, Kenichi Kokubo a, Toshio Ichiwata c, Hirosuke Kobayashi a,d, a Department of Clinical Engineering, School of Allied Health Sciences, Kitasato University, Kanagawa , Japan b Clinical Engineering, Fujinomiya City General Hospital, Shizuoka , Japan c Department of Respiratory Medicine, Dokkyo Medical University Koshigaya Hospital, Saitama , Japan d Division of Respiratory Medicine, Kitasato University Hospital, Sagamihara, Kanagawa , Japan article info abstract Article history: Accepted 28 March 2008 Keywords: Sleep apnea syndrome Mechanical ventilation Auto-CPAP Bench study Upper airway model The aim of this study was to investigate the response of auto-cpap devices to respiratory events (apnea, hypopnea, flow-limitation and snoring) on the same condition using a physiological upper airway model. The hypothesis of this study is that collapsibility of the flow-limiting collapsible segment of the airway is influenced by the upstream airway resistance. Five auto-cpap devices, AutoSet T, AutoSet Spirit TM, Goodnight 420E, PV10i and REMstar Auto were evaluated. Apnea: all the devices increased the auto- CPAP level, while AutoSet T and AutoSet Spirit TM did not respond to apnea for 30 s. Hypopnea: all the devices except the AutoSet T and Goodnight 420E increased pressure. Flow-limitation: all the devices except the PV10i and REMstar Auto increased pressure. Snoring: the snoring sounds disappeared when REMstar Auto and PV10i were used, and the Goodnight 420E lowered the level of snoring. In conclusion, the response of auto-cpap devices to respiratory events differed. Collapsible upper airway model with upstream resistance is useful for the first-step assessment of auto-cpap devices Elsevier B.V. All rights reserved. 1. Introduction Sleep-disordered breathing (SDB) is accompanied by several undesirable states, such as headache at awakening and sleepiness in daytime. Obstructive sleep apnea/hypopnea syndrome (OSAHS) accounts for 90% of patients with SDB (Bradley and Phillipson, 1992). Currently available medical treatments for OSAHS include (1) life style improvement (exercise, weight loss, cessation of smoking, temperance, etc.), (2) application of nasal continuous positive airway pressure (ncpap) during sleep, (3) the use of dental splinting, such as mandible advance devices, and (4) upper airway surgery. Among these treatments, ncpap is the first choice therapy in evidence-based medicine (Hoffstein et al., 1992; Polo et al., 1994; Hudgel, 1996). When applying ncpap to OSAHS patients, it is important to determine optimal treatment pressure for individual patients (Hsu and Lo, 2003; Akashiba et al., 1999). The suitable pressure is determined by titration using the CPAP device, i.e., by Corresponding author at: Department of Clinical Engineering, School of Allied Health Sciences, Kitasato University, Kitasato , Sagamihara, Kanagawa , Japan. Tel.: ; fax: addresses: hirose@kitasato-u.ac.jp (M. Hirose), ei-san@kitasato-u.ac.jp (E. Sato), tshinbo@kitasato-u.ac.jp (T. Shinbo), kokubo@kitasato-u.ac.jp (K. Kokubo), ichiwata@dokkyomed.ac.jp (T. Ichiwata), hiro@kitasato-u.ac.jp (H. Kobayashi). adjusting the pressure manually or adjusting the pressure automatically (auto-cpap) (Meurice et al., 1996; Sharma et al., 1996; Konermann et al., 1998). The auto-cpap devices deliver variable pressure levels in response to the detection of respiratory events (apnea, hypopnea, flow-limitation and snoring) based on their own algorithm. Auto-CPAP is now used not only for titration to determine the fixed level of CPAP, but also for the treatment of OSHAS patients at home. Although many auto-cpap devices are available now, the algorithm used by each device considerably differ each other. Since quality assessments of these auto-cpap devices are usually performed during direct clinical application to individual patient, it is difficult to compare the auto-cpap devices under the same upper airway and breathing condition. The aim of this study was to investigate the response of auto- CPAP devices to several respiratory events using a physiological upper airway model. Recently, bench studies have been proposed (Farré et al., 2002; Abdenbi et al., 2004; Lofaso et al., 2006; Rigau et al., 2006). Among them Abdenbi et al. (2004) developed a physiological collapsible upper airway model, but they did not attach an upstream airway resistance to the collapsible rubber tube, i.e., the resistance between auto-cpap device and the collapsible tube. We first tried to examine auto-cpap devices without the upstream airway resistance, but we could not observe the tube collapse and flow-limitation. If there is no upstream resistance, the /$ see front matter 2008 Elsevier B.V. All rights reserved. doi: /j.resp

2 M. Hirose et al. / Respiratory Physiology & Neurobiology 162 (2008) pressure in the collapsible tube should be atmospheric pressure, and the tube will be collapsed only when surrounding pressure is larger than the atmospheric pressure. However, this situation is considered to be limited in some extremely severe OSHAS patients, since surrounding pressure should not be so higher than the atmospheric pressure. The hypothesis of this study is that collapsibility of the flowlimiting collapsible segment of the airway is influenced by the upstream airway resistance, and based upon this hypothesis we have added the upstream airway resistance to the collapsible tube. We used pressure-controlled ventilation on the lung-side to simulate breathing rather than volume-controlled ventilation as Abdenbi et al. (2004) used, since volume-controlled ventilation creates stable ventilation and flow, irrespectively of the airway resistance, or there might be air-leak somewhere between the collapsible tube and the lung, which is not physiological. We attached the upper airway resistance between the auto- CPAP devices tested and the collapsible rubber tube, since the pressure in the collapsible airway is influenced by its upstream resistance on inspiration and the pressure detected by auto- CPAP device is also influenced by the resistance between the device and the collapsible site, and we verified the responses of several currently available auto-cpap devices to the simulated respiratory events. We found that the response of auto-cpap devices to the respiratory events considerably differed each other. 2. Methods 2.1. Collapsible upper airway model with upstream resistance The cross-area of an upper airway is determined by balance between the pressure in the airway and the pressure in the surrounding tissue (Remmers et al., 1978; Brouillette and Thach, 1979), and a net negative pressure results in the narrowing of the upper airway. The cross-sectional area of the upper airway decreases during inspiration and increases during expiration. Therefore, the upper airway model was designed as an assembly equipped with a collapsible elastic-tube (with a diameter of 18 mm and a length of 48 mm) housed in a Plexiglas box to simulate flow-limitation and snoring by adjusting the pressure surrounding the elastic-tube in the box (Fig. 1). This collapsible upper airway model is similar to the set-up used by Abdenbi et al. (2004), but we further added upstream (mouth-side) resistance of 50 cmh 2 O/(L s) to their model. Fig. 1. Upper airway model. An upper airway model equipped with an elastic-tube was housed in a Plexiglas box. The surround pressure was adjusted by applying a positive pressure to the interior of the Plexiglas box Verification of flow-limitation and snoring in the upper airway simulation Flow-limitation in the upper airway was verified using a test circuit as follows (Fig. 2). A vacuum pump attached to a watersealed equipment was connected downstream of the upper airway model to adjust the downstream negative pressure by altering the height of water in the water-sealed equipment; the upstream pressure was adjusted by the surrounding pressure of the collapsible tube. The flow and pressure before and after the upper airway model were measured using two pressure-flow analyzers (Calibration-Analyzer RT-200, Timeter Instruments Co., St. Louis, MO, USA), and the surrounding pressure of the elastic-tube was varied from 0 to 20 cmh 2 O for flow-limitation and 10 cmh 2 Ofor snoring. Flow-limitation was then created by changing the level of the surrounding pressure with fixed upstream-resistance of 50 cmh 2 O/(L s). This level of upstream-resistance was determined, since snoring sounds were generated by adjusting the surrounding pressure of the elastic-tube at that upstream-resistance level. The pressure vibration was measured directly by the pressure-flow analyzer, and the snoring sounds were recorded by a digital recorder (Voice-Trek DS-10, Olympus Co. Ltd., Tokyo, Japan). The pressure frequency of snoring was checked using both the pressure record and the frequency spectrum of the sounds analyzed using FFT analysis software (FFT Wave, E.N. Software, Tokyo, Japan). Fig. 2. Test circuit for upper airway model. A vacuum pump was connected downstream of the upper airway model, and the flux in the pressure difference before and after the upper airway model was measured to confirm flow-limitation. The negative pressure was adjusted according to the height of water in the water-seal equipment.

3 50 M. Hirose et al. / Respiratory Physiology & Neurobiology 162 (2008) Fig. 3. Test circuit for evaluating algorithms of auto-cpap devices. Each auto-cpap device was evaluated using the test circuit. Spontaneous breathing was simulated as a negative pressure swing downstream of the simulated upper airway model. The negative pressure swing was generated in one chamber of the test lung, to which another chamber was firmly fixed and was ventilated with a ventilator (Puritan Bennett 7200ae, Puritan Bennett Co., Carlsbad, CA, USA) in the pressure-controlled ventilation mode with a presumed tidal volume of 0.5 L and a respiratory rate of 10 breaths/min. In this manner, the positive pressure swing generated by the ventilator was converted to a negative pressure swing downstream (lung-side) of the upper airway model (collapsible tube connected with mouth-side airway resistance) Evaluation of the response of auto-cpap devices to respiratory events Each algorithm of five auto-cpap devices was evaluated: AutoSet T (Resmed Ltd., Sydney, Australia), AutoSet Spirit TM (a newer version of AutoSet T), Goodnight 420E (Tyco Healthcare, USA), PV10i (BREAS, Sweden), and REMstar Auto (Respironics, CA, USA). The auto-cpap devices were evaluated using a test assembly (Fig. 3) in which spontaneous breathing was simulated as a negative pressure swing downstream of the simulated upper airway model, to which the two-chamber Michigan test lung (TTL, Michigan Instruments Inc., Grand Rapids, MI, USA) was connected as was used by Lofaso et al. (2006). The negative pressure swing was generated in one chamber of the test lung, to which another chamber was firmly fixed and was ventilated with a ventilator (Puritan Bennett 7200ae, Puritan Bennett Co., Carlsbad, CA, USA) in the pressure-controlled ventilation mode with a presumed tidal volume of 0.5 L and a respiratory rate of 10 breaths/min. In this manner, the positive pressure swing generated by the ventilator was converted to a negative pressure swing downstream (lung-side) of the upper airway model (collapsible tube connected with mouth-side airway resistance). The Auto-CPAP devices were connected to the mouth-side airway resistance of the upper airway model, and the flow and pressure upstream and downstream of the upper airway model were measured. The time course of the pressure level generated by each auto-cpap device was measured under four kinds of simulated respiratory events, as follows: Fig. 4. Flow pattern of sleep events: (a) control changing to apnea; (b) control changing to hypopnea; (c) control changing to flow-limitation; (d) control changing to snoring. Each flow pattern was confirmed before the pressure response of auto-cpap devices was measured.

4 M. Hirose et al. / Respiratory Physiology & Neurobiology 162 (2008) (a) Apnea: the upstream tube in the upper airway model was closed for over 36 s until pressure change was observed. (b) Hypopnea: the tidal volume of the ventilator was reduced from 0.5 to 0.2 L for 48 s. (c) Flow-limitation: the pressure surrounding the elastic-tube was changed from 0 to 20 cmh 2 O. (d) Snoring: a surrounding pressure of 0 20 cmh 2 O was applied and snoring sounds were generated. Each flow pattern was confirmed (Fig. 4) before the pressure response of auto-cpap devices was measured, and to assess repeatability of the response of auto-cpap device, each test was repeated three times after resetting the device (power ON/OFF). The pressure at the start of measurements was set 4.0 cmh 2 Oin each auto-cpap device. 3. Results 3.1. Verification of flow-limitation and snoring in the upper airway simulation The flow through the upper airway model increased as the pressure difference between the upstream and downstream regions of the upper airway model increased but reached a lower plateau level when the pressure surrounding the elastic-tube was increased (Fig. 5), indicating a flow-limitation. Sounds that resembled snoring were generated when a surrounding pressure of 10 cmh 2 O was applied. The center frequencies of the snoring ranged from 280 to 300 Hz, as confirmed by both the pressure record and the frequency spectrum of the sounds (data not shown). These frequencies correspond to the reported center frequencies of natural snoring (Agrawal et al., 2002) Evaluation of the response of auto-cpap devices to respiratory events Apnea The pressure patterns generated by each auto-cpap device in response to apnea were as follows (Fig. 6). Goodnight 420E: the pressure began to raise 6 s after the start of apnea and rose to 5.0 cmh 2 O at 12 s after the start of apnea. REMstar Auto: the pressure began to raise 12 s after the start of apnea and rose to 5.0 cmh 2 O at 18 s after the start of apnea. PV10i: the pressure gradually began to raise 6 s after the start of apnea and rose to 5.3 cmh 2 O at 30 s after the start of apnea. AutoSet T and AutoSet Spirit TM : Fig. 6. Pressure response during apnea. All of the auto-cpap devices responded to apnea and increased the pressure. However, each device generated the pressure differently. the pressure did not change with apnea at the first time, but it suddenly began to rise 36 s after the start of apnea at the second time and rose to 9.2 cmh 2 O with the AutoSet T and 7.7 cmh 2 O with the AutoSet Spirit TM Hypopnea The pressures generated by each auto-cpap device in response to hypopnea were as follows (Fig. 7). AutoSet T and Goodnight 420E did not respond to hypopnea. PV10i: the pressure gradually began to rise 6 s after the start of hypopnea and rose to 5.3 cmh 2 O after about 30 s. REMstar Auto: the pressure began to rise after 12 s of hypopnea and rose to 5.0 cmh 2 O at 18 s after the start of hypopnea. AutoSet Spirit TM : the pressure began to raise 18 s after the start of hypopnea and rose gradually to 4.7 cmh 2 Oat 48 s after the start of hypopnea Flow-limitation The pressures generated by each auto-cpap device in response to flow-limitation were as follows (Fig. 8). AutoSet T: the pressure began to rise to 6.2 cmh 2 O at a surrounding pressure of 15 cmh 2 O and to about 10 cmh 2 O at a surrounding pressure of 20 cmh 2 O. Goodnight 420E: the pressure began to rise to 4.5 cmh 2 O at a surrounding pressure of 15 cmh 2 O and to 6.5 cmh 2 O at a surrounding pressure of 20 cmh 2 O. PV10i: the pressure did not rise within the surrounding pressure range investigated, i.e., from 0 to 20 cmh 2 O. REMstar Auto: the pressure began to rise to 7.5 cmh 2 O only when Fig. 5. Verification of flow-limitation. Flow-limitation was verified, since the flux rate at a large driving pressure was limited to a lower level as the surrounding pressure around the elastic-tube increased. Fig. 7. Pressure response during hypopnea. The AutoSet T and Goodnight 420E devices did not respond to hypopnea, whereas the other devices responded and increased the pressure.

5 52 M. Hirose et al. / Respiratory Physiology & Neurobiology 162 (2008) Fig. 8. Pressure response during flow-limitation. All the devices except PV10i responded to flow-limitation and increased the pressure. Fig. 9. Pressure response during snoring. Snoring sounds disappeared when the REMstar Auto and PV10i devices were used, and the Goodnight 420E lowered the snoring sound level. The AutoSet Spirit TM and AutoSet T devices did not change the snoring sound level. the surrounding pressure was 15 cmh 2 O and did not rise at a surrounding pressure of 20 cmh 2 O Snoring The pressures generated by each auto-cpap device in response to snoring were as follows (Fig. 9). At a surrounding pressure of 10 cmh 2 O, snoring sounds were generated at a CPAP pressure of 4cmH 2 O, the sound level decreased when the CPAP pressure was raised to 4.5 cmh 2 O, and the snoring sounds disappeared at a pressure of 5 cmh 2 O. The AutoSet T and AutoSet Spirit TM devices did not respond to the simulated snoring. Goodnight 420E: the CPAP pressure gradually increased after 24 s and rose to 4.5 cmh 2 O after 48 s, when the snoring sounds disappeared. PV10i: the CPAP pressure rose gradually 90 s after the start of snoring, and the snoring sounds disappeared about 3 min after the start of snoring. REMstar Auto: the CPAP pressure suddenly rose to 5 cmh 2 O about 1 min after the start of snoring and the snoring sounds disappeared. 4. Discussion 4.1. Examination of auto-cpap devices under the same condition The number of patients who use auto-cpap device is increasing worldwide. New auto-cpap devices are being developed, and many kinds of devices are being introduced into the medical market, including devices aimed at medical care at home. When a patient uses an auto-cpap device, the optimal type of auto-cpap must be determined for each patient, but the comparison of several kinds of auto-cpap devices under the same condition in a clinical test is not feasible. To evaluate the response of different auto-cpap devices to respiratory events, each response must be evaluated under the same bench-test condition. In order to perform the bench testing of auto-cpap device, the upper airway simulation model which can imitate respiratory events (apnea, hypopnea, flow-limitation and snoring) are required. Farré et al. (2002) evaluated the response of auto-cpap devices using an open-loop computer-controlled breathing waveform. This breathing waveform was created by a servo-controlled pump which was able to produce flow pattern with any respiratory events stored in the computer. Rigau et al. (2006) further developed this method using a closed-loop feedback control. Lofaso et al. (2006) also evaluated auto-cpap devices using the similar method. These studies, however, did not use physiological upper airway model, whereas only Abdenbi et al. (2004) evaluated auto-cpap devices using a physiological collapsible upper airway model. However, Abdenbi et al. (2004) did not take upstream resistance (the airway resistance between the auto-cpap device and the collapsible tube) into consideration. The pressure in the collapsible airway is influenced not only by the surrounding pressure but also by its upstream resistance on inspiration, and the pressure detected by auto-cpap device is also influenced by the resistance between the device and the collapsible site. Therefore, we considered that the resistance between the auto-cpap device and the collapsible site should be taken into account. This model has a potential advantage over previous bench methods in investigating the interaction between the applied CPAP and re-opening of upper airway in a more physiological setting. In this study we confined air-leak to the conventional leak of auto-cpap devices, and did not evaluated the effect of mask-leak on the response of auto-cpap devices, since the mask-leak is variable among patients and depends on individual patient and mask fitting. We did not place cardiogenic oscillation between the collapsible tube and the lung component. It should be noted that the influence of the mask-leak as well as the cardiogenic oscillation to the response of auto-cpap devices was out of the scope of this study. In this study we focused on relatively acute response (within 1 min) of auto-cpap devices, since long-term (20 45 min) fixed upper airway condition is unlikely in clinical settings and long term response is considered to be cumulative steps of acute responses Features of the auto-cpap devices Five kinds of auto-cpap devices and their response were evaluated using our collapsible upper airway with upstream resistance, modeling spontaneous breathing and the respiratory events (apnea, hypopnea, flow-limitation, and snoring), and the features of these devices were as follows AutoSet T and AutoSet Spirit TM As designed by the manufacturer, the pressure generated by both AutoSet T and AutoSet Spirit TM did not increase during apnea episode and rose only after the apnea improved, but the pressure levels after apnea at the first time increased higher than those of other auto-cpap devices. It is very likely that these auto-cpap devices were designed not to produce an extremely high pressure during apnea which would be needed to open the closed upperairway, but were designed to maintain sleep quality, by preventing repeated apnea. AutoSet T and AutoSet Spirit TM maintained a sufficient flow, even if the surrounding pressure of the elastic-tube was increased from 0 to 20 cmh 2 O. Although they did not respond to single episodes of hypopnea, their generated pressure increased during episodes of hypopnea with flow-limitation. Although the characteristic responses of AutoSet T and AutoSet Spirit TM to respiratory events were almost the same, the reaction to hypopnea was added to AutoSet Spirit TM. When the AutoSetT and the AutoSet Spirit TM were compared, the AutoSet Spirit TM was con-

6 M. Hirose et al. / Respiratory Physiology & Neurobiology 162 (2008) sidered to be more competent at detecting flow-limitation than the AutoSet T, since the AutoSet Spirit TM was able to react potently to the increased surrounding pressure. Therefore, AutoSet T and AutoSet Spirit TM devices are considered to be suitable for patients with frequent strong airway obstruction or narrowing, such as patients with thick layers of tissue surrounding their upper airway or those who have a large tongue, and who may sleep poorly with a large pressure increase during apnea as is generated with other auto-cpap devices Goodnight 420E Although the Goodnight 420E responded to apnea, it did not respond to hypopnea. Although the device responded to flow-limitation, flow was not maintained when the surrounding pressure of the elastic-tube was more than 15 cmh 2 O. Although the level of snoring sound was lessened, the snoring sound did not disappear completely, since the pressure supplied by the Goodnight 420E was mild. Consequently, the Goodnight 420E is considered to be suitable for patients with mild upper airway obstructions or narrowing PV10i Although the PV10i responded to apnea, hypopnea, and snoring, it did not respond to flow-limitation. The response of PV10i for respiratory events differed from those of other models. Other models raised the pressure linearly, but the PV10i raised the pressure gradually until it reached a suitable pressure. The responses for adjusting the pressure level in other models were determined by the number and the duration of respiratory events. The algorithm of the PV10i, in which a method similar to speech recognition technology is used according to the manufacturer s data book, differs from the algorithm of other devices, and PV10i raised the pressure gradually compared with other models, monitoring the pressure level every 2.5 s. Therefore, the PV10i is suitable for patients who are sensitive to sudden increase in applied pressure and for those who do not have strong upper airway obstruction REMstar Auto The REMstar Auto responded to all the respiratory events: apnea, hypopnea, flow-limitation, and snoring. According to the manufacturer s data sheet, the REMstar Auto has been designed to respond to apnea twice per minutes when 80% reduction in ventilation occurs for 10 s. Therefore, the REMstar Auto is suitable for patients having all kinds of respiratory events, and it may also be used for the titration of CPAP pressure. However, REMstar Auto reacted only when the surrounding pressure was 15 cmh 2 Oinflowlimitation, but did not react when the surrounding pressure was less than or more than 15 cmh 2 O in flow-limitation. The reaction of REMstar Auto to flow-limitation may be unstable for patients with variable level of upper airway collapse Comparison with other bench studies The same devices evaluated in other bench studies were compared. AutoSet T was evaluated by Farré et al. (2002), where the pressure increase in auto-cpap device were observed in repetitive respiratory events and in prolonged flow-limitation for 5 min. Although we evaluated them for a shorter time period, AutoSet T similarly raised its pressure rapidly to apnea, but did not respond to mild hyponea. However, AutoSet T responded to hyponea with snoring in their study, whereas it did not respond to snoring in our study, possibly due to difference in upstream resistance, thereby resulting in reduction in snoring pressure swing. PV10i, Goodnight 420E, AutoSet Spirit TM and REMstar Auto were evaluated by Rigau et al. (2006), where the pressure increase in auto-cpap device were observed for 20 min. PV10i and Goodnight 420E responded to each respiratory event as were observed in our study: PV10i did not respond to flow-limitation and Goodnight 420E did not respond to hypopnea. PV10i, Goodnight 420E and REMstar Auto were also evaluated by Lofaso et al. (2006), but in their study REMstar Auto consistently failed to detect flow-limitation. The reason of these responses, which differ from our observation, is not clear, but the fixed constant flow pattern generated in their study might cause the difference in the responses. PV10i and REMstar Auto exhibited unexpected responses with all their test protocols, i.e., auto-triggering occurred after 5 min of constant pressure in PV10i and slow increase and decrease in pressure recurred every 540 s in REMstar Auto. We did not observe pressure for a long time period of respiratory events, and further studies are required for the response of auto-cpap devices to prolonged flowlimitation. AutoSet T, PV10i and REMstar Auto were evaluated by Abdenbi et al. (2004). The pressure responded similarly to respiratory events. It rose in AutoSet T earlier, whereas pressure raised little by little in PV10i in response to apnea. AutoSet T did not respond to snoring as was observed in our study, but in contrast to their study, it responded to flow-limitation without snoring in our study. The addition of upstream resistance between auto-cpap device and collapsible airway may affect the response of auto-cpap devices to respiratory events, possibly because of the lessen pressure (and flow) swing due to the resistance. Previous studies showed the cumulative short-term response on the long-term trace, and showed the capping of applied pressure at a higher level. Auto-CPAP devices respond to sleep events and their response changes the airway condition. For example, flow-limitation will disappear after the short-term response of the auto-cpap device, and long-term stable flow-limitation is not realistic in the response of the auto-cpap device. We considered that the long-term response is the result of cumulative shortterm response, and we focused on short-term response in this study. Long-term man-machine feedback fluctuation, if it occurs, is an interesting theme to be investigated, and warrant further studies Limitation of this upper airway model The various proprietary responses designed by the manufacturers make it possible to develop highly efficient auto-cpap devices. However, bench studies including our study have disclosed that the auto-cpap devices differed in their responses to each respiratory event. Bench tests are useful to better understand principles of operation of each device, and will probably help medical stuffs to expect the performance in clinical application. However, it should be noted that bench tests could not allow medical stuffs to draw any conclusion about the applicability of the device to each patient before the clinical use. The collapsible airway model is different from practical upper airway, for example effects upon the sympathetic nervous system, and biological complicated effects with the higher respiratory control returned from apnea/hyponea events are neglected. It is very likely that the upper airway resistance and the collapsibility of the flow-limiting segment are variable in each patient during sleep, and the role of variable upstream resistance in respiratory events is an important point to be investigated. The results of bench tests could be used at the first step assessment to formulate hypothesis about the most appropriate way of use of each device in clinical application, and to create a new con-

7 54 M. Hirose et al. / Respiratory Physiology & Neurobiology 162 (2008) cept in developing devices. It is possible that tailor-made algorithm will be designed for each patient with sleep apnea/hypopnea in future. 5. Conclusion The response of auto-cpap devices to respiratory events considerably differed from each other. Collapsible upper airway model, a collapsible tube with resistance between auto-cpap device and the tube, was considered to be useful for the first-step assessment of auto-cpap devices. Acknowledgment The authors thank the manufacturers for lending their auto- CPAP devices used in this study. References Abdenbi, F., Chambille, B., Escourrou, P., Bench testing of auto-adjusting positive airway pressure devices. Eur. Respir. J. 24, Agrawal, S., Stone, P., Mcguinness, K., Morris, L., Camilleri, A.E., Sound frequency analysis and the site of snoring in natural and induced sleep. Clin. Otolaryngol. 27, Akashiba, T., Minemura, H., Yamamoto, H., Itoh, D., Kosaka, N., Saitoh, O., Horie, T., Effects of nasal continuous positive airway pressure on pulmonary haemodynamics and tissue oxygenation in patients with obstructive sleep apnoea. Respirology 4, Bradley, T.D., Phillipson, E.A., Central sleep apnea. Clin. Chest Med. 13, Brouillette, R.T., Thach, B.T., A neuromuscular mechanism maintaining extrathoracic airway patency. J. Appl. Physiol. 46, Farré, R., Montserrat, J.M., Rigau, J., Trepat, X., Pinto, P., Navajas, D., Response of automatic continuous positive airway pressure devices to different sleep breathing patterns. Am. J. Respir. Crit. Care Med. 166, Hoffstein, V., Viner, S., Mateika, S., Conway, J., Treatment of obstructive sleep apnea with nasal continuous positive airway pressure. Patient compliance, perception of benefits, and side effects. Am. Rev. Respir. Dis. 145, Hsu, A.A., Lo, C., Continuous positive airway pressure therapy in sleep apnoea. Respirology 8, Hudgel, D.W., Treatment of obstructive sleep apnea. A review. Chest 109, Konermann, M., Sanner, B.M., Vyleta, M., Laschewski, F., Groetz, J., Sturm, A., Zidek, W., Use of conventional and self-adjusting nasal continuous positive airway pressure for treatment of severe obstructive sleep apnea syndrome: a comparative study. Chest 113, Lofaso, F., Desmarais, G., Leroux, K., Zalc, V., Fodil, R., Isabey, D., Louis, B., Bench evaluation of flow limitation detection by automated continuous positive airway pressure devices. Chest 130, Meurice, J.C., Marc, I., Series, F., Efficacy of autocpap in the treatment of obstructive sleep apnea/hypopnea syndrome. Am. J. Respir. Crit. Care Med. 153, Polo, O., Berthon-Jones, M., Sulliran, C., Douglas, N.J., Management of obstructive sleep apnoea/hypopnoea syndrome. Lancet 344, Remmers, J.E., degroot, W.J., Sauerland, E.K., Anch, A.M., Pathogenesis of upper airway occlusion during sleep. J. Appl. Physiol. 44, Rigau, J., Montserrat, J.M., Wohrle, H., Plattner, D., Schwaibold, M., Navajas, D., Farre, R., Bench model to simulate upper airway obstruction for analyzing automatic continuous positive airway pressure devices. Chest 130, Sharma, S., Wali, S., Pouliot, Z., Peters, M., Neufeld, H., Kryger, M., Treatment of obstructive sleep apnea with a self-titrating continuous positive airway pressure (CPAP) system. Sleep 19,

Auto Servo Ventilation Indications, Basics of Algorithm, and Titration

Auto Servo Ventilation Indications, Basics of Algorithm, and Titration Auto Servo Ventilation Indications, Basics of Algorithm, and Titration 1 ASV Learning Objectives Understand the indications for Auto Servo Ventilation Differentiate obstructive versus central hypopneas

More information

Monitoring and Troubleshooting Adherence to PAP Devices and Understanding Device Downloads

Monitoring and Troubleshooting Adherence to PAP Devices and Understanding Device Downloads Monitoring and Troubleshooting Adherence to PAP Devices and Understanding Device Downloads Christine Won, MD MS Medical Director, Yale Sleep Center Associate Professor of Medicine Yale School of Medicine

More information

Test Bench for Pressure Calibration of Continuous Positive Airway Pressure (CPAP) Device for Sleep Apnea Applications

Test Bench for Pressure Calibration of Continuous Positive Airway Pressure (CPAP) Device for Sleep Apnea Applications Test Bench for Pressure Calibration of Continuous Positive Airway Pressure (CPAP) Device for Sleep Apnea Applications Massimiliana Carello 1 and Daniela Maffiodo 2 1,2 Mechanical and Aerospace Engineering

More information

Snoring detection during auto-nasal continuous positive airway pressure

Snoring detection during auto-nasal continuous positive airway pressure Eur Respir J 2002; 19: 108 112 DOI: 10.1183/09031936.00213302 Printed in UK all rights reserved Copyright #ERS Journals Ltd 2002 European Respiratory Journal ISSN 0903-1936 Snoring detection during auto-nasal

More information

Comparison of three auto-adjusting positive pressure devices in patients with sleep apnoea

Comparison of three auto-adjusting positive pressure devices in patients with sleep apnoea Eur Respir J 2006; 28: 159 164 DOI: 10.1183/09031936.06.00127205 CopyrightßERS Journals Ltd 2006 Comparison of three auto-adjusting positive pressure devices in patients with sleep apnoea G.M. Nolan, S.

More information

Bench testing of auto-adjusting positive airway pressure devices

Bench testing of auto-adjusting positive airway pressure devices Eur Respir J 24; 24: 649-658 DOI: 1.1183/931936.4.13373 Printed in UK - all rights reserved Copyright '< ERS Journals Ltd 24 European Respiratory Journal ISSN 93-1936 Bench testing of auto-adjusting positive

More information

A simple solution for your complex patients

A simple solution for your complex patients A simple solution for your complex patients The market-leading servo ventilation device System One BiPAP autosv Advanced simplifies treating complex sleep-disordered breathing patients Developed for your

More information

Chronic NIV in heart failure patients: ASV, NIV and CPAP

Chronic NIV in heart failure patients: ASV, NIV and CPAP Chronic NIV in heart failure patients: ASV, NIV and CPAP João C. Winck, Marta Drummond, Miguel Gonçalves and Tiago Pinto Sleep disordered breathing (SDB), including OSA and central sleep apnoea (CSA),

More information

How to write bipap settings

How to write bipap settings How to write bipap settings 6-6-2013 Living On O2 for Life If you use a bipap machine, like I do, this post is for you. I've been using a bipap machine since 1993 which is a pretty long time. BiPAP 's

More information

Effective Treatment for Obstructive Sleep Apnoea

Effective Treatment for Obstructive Sleep Apnoea Effective Treatment for Obstructive Sleep Apnoea The Series of Positive Airway Pressure devices from DeVilbiss Healthcare is designed to meet the varied needs of people suffering from Obstructive Sleep

More information

PRODUCT TRAINING TREND II

PRODUCT TRAINING TREND II Product Training TREND II HOFFRICHTER GmbH 1 General Introduction Complete system spectrum for CPAP therapy TREND II TREND II AUTO TREND II BILEVEL TREND II BILEVEL ST20 AquaTREND uni Therapy Humidifier

More information

The intelligent solution. System One Sleep Therapy System uses advanced intelligence to deliver excellent care while making patient management easy

The intelligent solution. System One Sleep Therapy System uses advanced intelligence to deliver excellent care while making patient management easy The intelligent solution System One Sleep Therapy System uses advanced intelligence to deliver excellent care while making patient management easy Proven efficacy The System One REMstar Auto algorithm

More information

OSA and COPD: What happens when the two OVERLAP?

OSA and COPD: What happens when the two OVERLAP? 2011 ISRC Seminar 1 COPD OSA OSA and COPD: What happens when the two OVERLAP? Overlap Syndrome 1 OSA and COPD: What happens when the two OVERLAP? ResMed 10 JAN Global leaders in sleep and respiratory medicine

More information

VPAP Auto Positive Airway Pressure Device Data Management Guide English

VPAP Auto Positive Airway Pressure Device Data Management Guide English VA Auto ositive Airway ressure Device Data Management Guide English The following table shows where data from the VA Auto flow generator can be viewed. Data displayed in ResScan can be downloaded via:

More information

Effective Treatment for Obstructive Sleep Apnoea

Effective Treatment for Obstructive Sleep Apnoea Effective Treatment for Obstructive Sleep Apnoea The Series of Positive Airway Pressure devices from DeVilbiss Healthcare is designed to meet the varied needs of people suffering from Obstructive Sleep

More information

Management of OSA in the Acute Care Environment. Robert S. Campbell, RRT FAARC HRC, Philips Healthcare May, 2018

Management of OSA in the Acute Care Environment. Robert S. Campbell, RRT FAARC HRC, Philips Healthcare May, 2018 Management of OSA in the Acute Care Environment Robert S. Campbell, RRT FAARC HRC, Philips Healthcare May, 2018 1 Learning Objectives Upon completion, the participant should be able to: Understand pathology

More information

A new beginning in therapy for women

A new beginning in therapy for women A new beginning in therapy for women OSA in women Tailored solutions for Her AutoSet for Her algorithm ResMed.com Women and OSA OSA has traditionally been considered to be a male disease. However, recent

More information

Stellar 100 Stellar 150

Stellar 100 Stellar 150 Stellar 100 Stellar 150 Invasive and noninvasive ventilator Data Management Guide English The following table shows where data from the Stellar device can be viewed. Data displayed in ResScan can be downloaded

More information

F. Lofaso*, L. Heyer*, A. Leroy**, H. Lorino*, A. Harf*, D. Isabey*

F. Lofaso*, L. Heyer*, A. Leroy**, H. Lorino*, A. Harf*, D. Isabey* Eur Respir J, 14,, 2 21 DOI: 1.113/313.4.112 Printed in UK - all rights reserved Copyright ERS Journals Ltd 14 European Respiratory Journal ISSN 3-13 TECHNICAL NOTE Do turbines with servo-controlled speed

More information

CPAP titration by an auto-cpap device based on snoring detection: a clinical trial and economic considerations

CPAP titration by an auto-cpap device based on snoring detection: a clinical trial and economic considerations Eur Respir J 199; : 759 7 DOI:.113/09031936.9.0759 Printed in UK - all rights reserved Copyright ERS Journals Ltd 199 European Respiratory Journal ISSN 0903-1936 CPAP titration by an auto-cpap device based

More information

Reasons Providers Use Bilevel

Reasons Providers Use Bilevel Reasons Providers Use Bilevel More comfort, improve therapy compliance Noncompliant OSA (NCOSA) 1 Scripts from lab referrals Central/Complex Sleep Apnea 2 For ventilations needs Restrictive Thoracic Disorders/Neuromuscular

More information

Comparative assessment of several automatic CPAP devices responses: a bench test study

Comparative assessment of several automatic CPAP devices responses: a bench test study ORIGINAL ARTICLE SLEEP MEDICINE Comparative assessment of several automatic CPAP devices responses: a bench test study Valentina Isetta 1,2, Daniel Navajas 1,2,3, Josep M. Montserrat 2,4,5 and Ramon Farré

More information

Obstructive Sleep Apnea Syndrome. Common sleep disorder causes high blood pressure and heart attacks

Obstructive Sleep Apnea Syndrome. Common sleep disorder causes high blood pressure and heart attacks Obstructive Sleep Apnea Syndrome Common sleep disorder causes high blood pressure and heart attacks Message: Sleep apnea is very common. It is estimated that 158 million people worldwide suffer from sleep

More information

BiPAPS/TVAPSCPAPASV???? Lori Davis, B.Sc., R.C.P.T.(P), RPSGT

BiPAPS/TVAPSCPAPASV???? Lori Davis, B.Sc., R.C.P.T.(P), RPSGT BiPAPS/TVAPSCPAPASV???? Lori Davis, B.Sc., R.C.P.T.(P), RPSGT Modes Continuous Positive Airway Pressure (CPAP): One set pressure which is the same on inspiration and expiration Auto-PAP (APAP) - Provides

More information

a new beginning in managing CSA ResMed.com/AirSolutions

a new beginning in managing CSA ResMed.com/AirSolutions a new beginning in managing CSA ResMed.com/AirSolutions As your partner in patient care, ResMed gives you more clinical tools to manage patients with breathing disorders that can sometimes be challenging

More information

Index. sleep.theclinics.com. Note: Page numbers of article titles are in boldface type.

Index. sleep.theclinics.com. Note: Page numbers of article titles are in boldface type. Index Note: Page numbers of article titles are in boldface type. A Acquired central hypoventilation syndrome, NPPV in children with, 475 Acute cardiogenic pulmonary edema, PAP therapy in, 394 395 Adaptive

More information

Mechanical Ventilation Principles and Practices

Mechanical Ventilation Principles and Practices Mechanical Ventilation Principles and Practices Dr LAU Chun Wing Arthur Department of Intensive Care Pamela Youde Nethersole Eastern Hospital 6 October 2009 In this lecture, you will learn Major concepts

More information

Lumis Tx: the all-in-one hospital ventilation solution

Lumis Tx: the all-in-one hospital ventilation solution Lumis Tx: the all-in-one hospital ventilation solution The Lumis Tx is a multi-purpose non-invasive ventilator that treats the full range of respiratory conditions and is suitable for a variety of hospital

More information

1 2 3 Based on the contents detailed below, assess the substantial equivalence to a legally marketed predicate device. In that case, the standards, et

1 2 3 Based on the contents detailed below, assess the substantial equivalence to a legally marketed predicate device. In that case, the standards, et Attachment 8 Application for Treatment Positive Airway Pressure Unit, etc. (1) Scope of application Continuous positive airway pressure units and Continuous-automated positive airway pressure units in

More information

Equivalence of Autoadjusted and Constant Continuous Positive Airway Pressure in Home Treatment of Sleep Apnea*

Equivalence of Autoadjusted and Constant Continuous Positive Airway Pressure in Home Treatment of Sleep Apnea* Original Research SLEEP MEDICINE Equivalence of Autoadjusted and Constant Continuous Positive Airway Pressure in Home Treatment of Sleep Apnea* Yvonne Nussbaumer, MD; Konrad E. Bloch, MD, FCCP; Therese

More information

Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients

Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients Objectives Describe nocturnal ventilation characteristics that may indicate underlying conditions and benefits of bilevel therapy for specific

More information

About VirtuOx. Was marketed exclusively by Phillips Healthcare division, Respironics for 3 years

About VirtuOx. Was marketed exclusively by Phillips Healthcare division, Respironics for 3 years About VirtuOx VirtuOx, Inc. assists physicians and Durable Medical Equipment (DME)( companies diagnose respiratory diseases and qualify patients for home respiratory equipment under the guidelines of CMS

More information

THE NEW HOME FOR C-FLEX IS BUILT ON THE IDEA THAT A BETTER NIGHT S SLEEP COMES FROM PRODUCTS THAT ARE EASY TO USE.

THE NEW HOME FOR C-FLEX IS BUILT ON THE IDEA THAT A BETTER NIGHT S SLEEP COMES FROM PRODUCTS THAT ARE EASY TO USE. THE NEW HOME FOR C-FLEX IS BUILT ON THE IDEA THAT A BETTER NIGHT S SLEEP COMES FROM PRODUCTS THAT ARE EASY TO USE. W W W. M S E R I E S. R E S P I R O N I C S. C O M INTRODUCING THE REMSTAR M SERIES SLEEP

More information

Treatment of obstructive sleep-disordered breathing with positive airway pressure systems

Treatment of obstructive sleep-disordered breathing with positive airway pressure systems Eur Respir Rev 2007; 16: 106, 125 131 DOI: 10.1183/09059180.00010603 CopyrightßERSJ Ltd 2007 Treatment of obstructive sleep-disordered breathing with positive airway pressure systems D. Pevernagie*, J.F.

More information

Sleep Disordered Breathing

Sleep Disordered Breathing Sleep Disordered Breathing SDB SDB Is an Umbrella Term for Many Disorders characterized by a lack of drive to breathe Results n repetitive pauses in breathing with no effort Occurs for a minimum of 10

More information

One intelligent solution

One intelligent solution One intelligent solution System One sleep therapy platform gets even smarter with advanced intelligence for optimum care and easier patient management. Industry-leading enhancements and innovations The

More information

Physiological consequences of prolonged periods of flow limitation in patients with sleep apnea hypopnea syndrome

Physiological consequences of prolonged periods of flow limitation in patients with sleep apnea hypopnea syndrome Respiratory Medicine (2006) 100, 813 817 Physiological consequences of prolonged periods of flow limitation in patients with sleep apnea hypopnea syndrome Gabriel Calero, Ramon Farre, Eugeni Ballester,

More information

Philip L. Smith, MD; Christopher P. O Donnell, PhD; Lawrence Allan, BS; and Alan R. Schwartz, MD

Philip L. Smith, MD; Christopher P. O Donnell, PhD; Lawrence Allan, BS; and Alan R. Schwartz, MD A Physiologic Comparison of Nasal and Oral Positive Airway Pressure* Philip L. Smith, MD; Christopher P. O Donnell, PhD; Lawrence Allan, BS; and Alan R. Schwartz, MD Study objectives: The effectiveness

More information

Natural performance. Introducing the BiPAP A30 - because ease of use and therapy efficacy are key to patient well-being

Natural performance. Introducing the BiPAP A30 - because ease of use and therapy efficacy are key to patient well-being Natural performance Introducing the BiPAP A30 - because ease of use and therapy efficacy are key to patient well-being Because our innovations are inspired by you and your patients, the bi-level ventilator

More information

Tired of being tired?

Tired of being tired? Tired of being tired? Narval CC MRD ResMed.com/Narval Sleepiness and snoring are possible symptoms of sleep apnea. Did you know that one in every four adults has some form of sleep disordered-breathing

More information

Comparison of three ways to determine and deliver pressure during nasal CPAP therapy for Obstructive Sleep Apnoea

Comparison of three ways to determine and deliver pressure during nasal CPAP therapy for Obstructive Sleep Apnoea Thorax Online First, published on October 27, 2005 as 10.1136/thx.2005.046300 Comparison of three ways to determine and deliver pressure during nasal CPAP therapy for Obstructive Sleep Apnoea Sophie D

More information

11/20/2015. Beyond CPAP. No relevant financial conflicts of interest. Kristie R Ross, M.D. November 12, Describe advanced ventilation options

11/20/2015. Beyond CPAP. No relevant financial conflicts of interest. Kristie R Ross, M.D. November 12, Describe advanced ventilation options Beyond CPAP Kristie R Ross, M.D. November 12, 2015 No relevant financial conflicts of interest Sponsored by The Warren Alpert Medical School of Brown University Describe advanced ventilation options Compare

More information

NON-INVASIVE VENTILATION MADE RIDICULOUSLY SIMPLE

NON-INVASIVE VENTILATION MADE RIDICULOUSLY SIMPLE NON-INVASIVE VENTILATION MADE RIDICULOUSLY SIMPLE Jennifer Newitt, MD 3 rd year Pulmonary/Critical Care Fellow Mentor: Patrick Strollo Jr, MD Myth or Fact?!? Myth or Fact?!? Treatment for Obstructive

More information

Respironics Home Bipap Manual

Respironics Home Bipap Manual Respironics Home Bipap Manual Home / Health & Wellness / Daily Living Aids / Sleeping Comfort 27 results found for "philips respironics bipap" All Products & Sellers (27 (for e.g. manual, Philips Home

More information

Pain patient with sleep-disordered breathing

Pain patient with sleep-disordered breathing Pain patient with sleep-disordered breathing Clinical scenario: A 50-year old female with a history of acid reflux, fibromyalgia, spinal stenosis and degenerative disk disease s/p C3/C4 disk fusion, depression

More information

Comparison of two in-laboratory titration methods to determine evective pressure levels in patients with obstructive sleep apnoea

Comparison of two in-laboratory titration methods to determine evective pressure levels in patients with obstructive sleep apnoea Thorax 2000;55:741 745 741 Centre de Recherche, Hôpital Laval, Institut Universitaire de Cardiologie et de Pneumologie de l Université Laval, Sainte-Foy, Québec G1V 4G5, Canada M P Bureau F Sériès Correspondence

More information

Comfort and pressure profiles of two autoadjustable positive airway pressure devices: a technical report

Comfort and pressure profiles of two autoadjustable positive airway pressure devices: a technical report Respiratory Medicine (23) 97, 93 98 Comfort and pressure profiles of two autoadjustable positive airway pressure devices: a technical report Katrien B. Hertegonne, Pascal M. Proot, Romain A. Pauwels, Dirk

More information

RECENT ADVANCES IN THE TREATMENT OF OBSTRUCTIVE JAUNDICE RECENT ADVANCES IN THE MANAGEMENT OF CHRONIC OBSTRUCTIVE

RECENT ADVANCES IN THE TREATMENT OF OBSTRUCTIVE JAUNDICE RECENT ADVANCES IN THE MANAGEMENT OF CHRONIC OBSTRUCTIVE PDF RECENT ADVANCES IN THE TREATMENT OF OBSTRUCTIVE JAUNDICE RECENT ADVANCES IN THE MANAGEMENT OF CHRONIC OBSTRUCTIVE 1 / 5 2 / 5 3 / 5 advances in obstructive pdf Abstract. In addition to the hazards

More information

Heart Failure and Sleep Disordered Breathing (SDB) Unhappy Bedfellows

Heart Failure and Sleep Disordered Breathing (SDB) Unhappy Bedfellows Question Heart Failure and Sleep Disordered Breathing (SDB) Unhappy Bedfellows 1 ResMed 2012 07 2 ResMed 2012 07 Open Airway 3 ResMed 2012 07 Flow Limitation Snore 4 ResMed 2012 07 Apnoea 5 ResMed 2012

More information

Prefabricated Oral Appliances for Obstructive Sleep Apnea

Prefabricated Oral Appliances for Obstructive Sleep Apnea Medical Policy Manual Allied Health, Policy No. 36 Prefabricated Oral Appliances for Obstructive Sleep Apnea Next Review: May 2019 Last Review: April 2018 Effective: May 1, 2018 IMPORTANT REMINDER Medical

More information

Positive Airway Pressure Systems for Sleep Disordered Breathing

Positive Airway Pressure Systems for Sleep Disordered Breathing Positive Airway Pressure Systems for Sleep Disordered Breathing Lori Pickrell, RRT Account Manager Roberts Home Medical Lpickrell@robertshomemedical.com Objectives Upon completion of the session, attendees

More information

Obstructive sleep apnoea How to identify?

Obstructive sleep apnoea How to identify? Obstructive sleep apnoea How to identify? Walter McNicholas MD Newman Professor in Medicine, St. Vincent s University Hospital, University College Dublin, Ireland. Potential conflict of interest None Obstructive

More information

Provider guide. DreamStation CPAP DreamStation CPAP Pro DreamStation Auto CPAP DreamStation BiPAP Pro DreamStation Auto BiPAP

Provider guide. DreamStation CPAP DreamStation CPAP Pro DreamStation Auto CPAP DreamStation BiPAP Pro DreamStation Auto BiPAP Provider guide DreamStation CPAP DreamStation CPAP Pro DreamStation Auto CPAP DreamStation BiPAP Pro DreamStation Auto BiPAP IMPORTANT! Remove this guide before giving the device to the patient. Only medical

More information

DECLARATION OF CONFLICT OF INTEREST

DECLARATION OF CONFLICT OF INTEREST DECLARATION OF CONFLICT OF INTEREST Obstructive sleep apnoea How to identify? Walter McNicholas MD Newman Professor in Medicine, St. Vincent s University Hospital, University College Dublin, Ireland. Potential

More information

PAP Therapy Devices: Delivering the Right Therapy To The Right Patient. Ryan Schmidt, BS, RRT Clinical Specialist Philips Respironics

PAP Therapy Devices: Delivering the Right Therapy To The Right Patient. Ryan Schmidt, BS, RRT Clinical Specialist Philips Respironics PAP Therapy Devices: Delivering the Right Therapy To The Right Patient Ryan Schmidt, BS, RRT Clinical Specialist Philips Respironics Conflict of Interest Disclosure(s) I do not have any potential conflicts

More information

NASAL CONTINUOUS POSITIVE AIRWAY PRESSURE FOR OBSTRUCTIVE SLEEP APNEA IN CHILDREN. Dr. Nguyễn Quỳnh Anh Department of Respiration 1

NASAL CONTINUOUS POSITIVE AIRWAY PRESSURE FOR OBSTRUCTIVE SLEEP APNEA IN CHILDREN. Dr. Nguyễn Quỳnh Anh Department of Respiration 1 1 NASAL CONTINUOUS POSITIVE AIRWAY PRESSURE FOR OBSTRUCTIVE SLEEP APNEA IN CHILDREN Dr. Nguyễn Quỳnh Anh Department of Respiration 1 CONTENTS 2 1. Preface 2. Definition 3. Etiology 4. Symptoms 5. Complications

More information

Ventilator Waveforms: Interpretation

Ventilator Waveforms: Interpretation Ventilator Waveforms: Interpretation Albert L. Rafanan, MD, FPCCP Pulmonary, Critical Care and Sleep Medicine Chong Hua Hospital, Cebu City Types of Waveforms Scalars are waveform representations of pressure,

More information

Sleep Apnea. What is sleep apnea? How does it occur? What are the symptoms?

Sleep Apnea. What is sleep apnea? How does it occur? What are the symptoms? What is sleep apnea? Sleep Apnea Sleep apnea is a serious sleep problem. If you have it, you stop breathing for more than 10 seconds at a time many times while you sleep. Another term for this problem

More information

CPAP in the treatment of sleep apnea. Titration, fixed and intelligent systems. Marc Josep M Montserrat. Barcelona. Spain

CPAP in the treatment of sleep apnea. Titration, fixed and intelligent systems. Marc Josep M Montserrat. Barcelona. Spain CPAP in the treatment of sleep apnea. Titration, fixed and intelligent systems Marc 2015 Josep M Montserrat. Barcelona. Spain Faculty disclosure No 1. INTRODUCTION 2. TITRATION 3. AUTOMATIC CPAP 4. FOLLOW

More information

ResScan Report Interpretation Guide. FOR CLINICAL USE ONLY V1.0 June, 2011

ResScan Report Interpretation Guide. FOR CLINICAL USE ONLY V1.0 June, 2011 ResScan Report Interpretation Guide FOR CLINICAL USE ONLY V1.0 June, 2011 ResMed Software Support Hours 5:30 AM to 5:30 PM (Pacific Time) Monday Friday Phone +1 (800) 424-0737, Option 6 Email TechSupportUSA@resmed.com

More information

Advanced Algorithms for Sleep Disordered Breathing. 1.5 AARC and AAST CEU credits

Advanced Algorithms for Sleep Disordered Breathing. 1.5 AARC and AAST CEU credits Advanced Algorithms for Sleep Disordered Breathing 1.5 AARC and AAST CEU credits Learning Objectives Restate the technology of auto adjusting expiratory pressure, servo ventilation, volume assured pressure

More information

DIGITAL AUTO-TRAK + Bi-FLEX + AUTO Bi-LEVEL THE POWER OF. Sometimes when three come together the results can be quite extraordinary. Unique even.

DIGITAL AUTO-TRAK + Bi-FLEX + AUTO Bi-LEVEL THE POWER OF. Sometimes when three come together the results can be quite extraordinary. Unique even. DIGITAL AUTO-TRAK + Bi-FLEX + AUTO Bi-LEVEL THE POWER OF 3 Sometimes when three come together the results can be quite extraordinary. Unique even. W W W. R E S P I R O N I C S. C O M THE BiPAP M SERIES

More information

Hypoventilation? Obstructive Sleep Apnea? Different Tests, Different Treatment

Hypoventilation? Obstructive Sleep Apnea? Different Tests, Different Treatment Hypoventilation? Obstructive Sleep Apnea? Different Tests, Different Treatment Judith R. Fischer, MSLS, Editor, Ventilator-Assisted Living (fischer.judith@sbcglobal.net) Thanks to Josh Benditt, MD, University

More information

Challenging Cases in Pediatric Polysomnography. Fauziya Hassan, MBBS, MS Assistant Professor Pediatric Pulmonary and Sleep

Challenging Cases in Pediatric Polysomnography. Fauziya Hassan, MBBS, MS Assistant Professor Pediatric Pulmonary and Sleep Challenging Cases in Pediatric Polysomnography Fauziya Hassan, MBBS, MS Assistant Professor Pediatric Pulmonary and Sleep Conflict of Interest None pertaining to this topic Will be using some slides from

More information

WHAT DO YOU WANT FROM A HOME VENTILATION SYSTEM? 8322_RS_HomeNIV_brochure_v14.ind1 1 4/7/06 12:57:35

WHAT DO YOU WANT FROM A HOME VENTILATION SYSTEM? 8322_RS_HomeNIV_brochure_v14.ind1 1 4/7/06 12:57:35 WHAT DO YOU WANT FROM A HOME VENTILATION SYSTEM? 8322_RS_HomeNIV_brochure_v14.ind1 1 4/7/06 12:57:35 D I L E M M A DIFFERENT VENTILATORS DIFFERENT ALGORITHMS TO KNOW YOU VE CHANGED PATIENT LIVES?PATIENT??

More information

During the therapeutic titration of nasal continuous

During the therapeutic titration of nasal continuous Cardiogenic Oscillations on the Airflow Signal During Continuous Positive Airway Pressure as a Marker of Central Apnea* Indu Ayappa, PhD; Robert G. Norman, MS, RRT; and David M. Rapoport, MD, FCCP Therapeutic

More information

TO THE OPERATOR AND PERSON IN CHARGE OF MAINTENANCE AND CARE OF THE UNIT:

TO THE OPERATOR AND PERSON IN CHARGE OF MAINTENANCE AND CARE OF THE UNIT: fabian HFO Quick guide TO THE OPERATOR AND PERSON IN CHARGE OF MAINTENANCE AND CARE OF THE UNIT: This Quick Guide is not a substitute for the Operation Manual. Read the Operation Manual carefully before

More information

Titration protocol reference guide

Titration protocol reference guide PN 1079754_Cover:22037_Cov_Canada 11/22/10 Philips Healthcare is part of Royal Philips Electronics How to reach us www.philips.com/healthcare healthcare@philips.com Asia +49 7031 463 2254 Europe, Middle

More information

CoughAssist E70. More than just a comfortable cough. Flexible therapy that brings more comfort to your patients airway clearance

CoughAssist E70. More than just a comfortable cough. Flexible therapy that brings more comfort to your patients airway clearance CoughAssist E70 More than just a comfortable cough Flexible therapy that brings more comfort to your patients airway clearance Flexible, customisable loosening and clearing therapy An effective cough is

More information

PEDIATRIC PAP TITRATION PROTOCOL

PEDIATRIC PAP TITRATION PROTOCOL PURPOSE In order to provide the highest quality care for our patients, our sleep disorders facility adheres to the AASM Standards of Accreditation. The accompanying policy and procedure on pediatric titrations

More information

BTS sleep Course. Module 10 Therapies I: Mechanical Intervention Devices (Prepared by Debby Nicoll and Debbie Smith)

BTS sleep Course. Module 10 Therapies I: Mechanical Intervention Devices (Prepared by Debby Nicoll and Debbie Smith) BTS sleep Course Module 10 Therapies I: Mechanical Intervention Devices (Prepared by Debby Nicoll and Debbie Smith) S1: Overview of OSA Definition History Prevalence Pathophysiology Causes Consequences

More information

BPAP 25A Training A.Giudice,RPSGT Clinical Education Manager

BPAP 25A Training A.Giudice,RPSGT Clinical Education Manager 1 Solutions in Sleep Therapy BPAP 25A Training A.Giudice,RPSGT Clinical Education Manager 2 To access press and hold the On/Off Button and Ramp Button and at the same time connect the power cord into the

More information

Inspiratory flow-volume curve in snoring patients with and without obstructive sleep apnea

Inspiratory flow-volume curve in snoring patients with and without obstructive sleep apnea Brazilian Journal of Medical and Biological Research (1999) 32: 407-411 Flow-volume curve and obstructive sleep apnea ISSN 0100-879X 407 Inspiratory flow-volume curve in snoring patients with and without

More information

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea Policy Number: Original Effective Date: MM.01.009 11/01/2009 Line(s) of Business: Current Effective Date: HMO; PPO;

More information

Up to 50% of continuous flow oxygen therapy patients experience clinically significant nocturnal desaturation. 1

Up to 50% of continuous flow oxygen therapy patients experience clinically significant nocturnal desaturation. 1 Up to 50% of continuous flow oxygen therapy patients experience clinically significant nocturnal desaturation. 1 Continuous Flow Oxygen Delivery & Sleep A number of theories and studies are published surrounding

More information

The diagnosis of obstructive sleep apnea syndrome. Combined Effects of a Nasal Dilator and Nasal Prongs on Nasal Airflow Resistance*

The diagnosis of obstructive sleep apnea syndrome. Combined Effects of a Nasal Dilator and Nasal Prongs on Nasal Airflow Resistance* Combined Effects of a Nasal Dilator and Nasal Prongs on Nasal Airflow Resistance* Anne Marie Lorino, PhD; Marie Pia d Ortho, MD; Estelle Dahan; Olivier Bignani; Carine Vastel; and Hubert Lorino, PhD Study

More information

Data Management of the Sleep Disordered Breathing Patient

Data Management of the Sleep Disordered Breathing Patient Data Management of the Sleep Disordered Breathing Patient 1 AARC or AAST CEU credit As Allies in Better Sleep and Breathing, we make it our #1 priority to work in harmony with caregivers and patients to

More information

UPDATES IN SLEEP APNEA:

UPDATES IN SLEEP APNEA: UPDATES IN SLEEP APNEA: CPAP,CPAP COMPLIANCE, & ALTERNATIVES MICHELLE ZETOONY, DO, FCCP, FACOI BOARD CERTIFIED PULMONARY, CRITICAL CARE, SLEEP & INTERNAL MEDICINE CLEARWATER, FL DISCLOSURE I have no conflicts

More information

Equivalence of Nasal and Oronasal Masks during Initial CPAP Titration for Obstructive Sleep Apnea Syndrome

Equivalence of Nasal and Oronasal Masks during Initial CPAP Titration for Obstructive Sleep Apnea Syndrome NASAL VERSUS ORO-NASAL MASKS DURING CPAP TITRATION FOR OSA DOI: 1.5665/SLEEP.1134 Equivalence of Nasal and Oronasal s during Initial CPAP Titration for Obstructive Apnea Syndrome Ming Teo, MBBS 2 ; Terence

More information

Key words: AutoSet; continuous positive airway pressure; obstructive sleep apnea; snoring; SOMNOsmart

Key words: AutoSet; continuous positive airway pressure; obstructive sleep apnea; snoring; SOMNOsmart Efficacy of Flow- vs Impedance-Guided Autoadjustable Continuous Positive Airway Pressure* A Randomized Cross-over Trial Dirk A. Pevernagie, MD, PhD; Pascal M. Proot, MD; Katrien B. Hertegonne, MD; Marleen

More information

Alaska Sleep Education Center

Alaska Sleep Education Center Alaska Sleep Education Center The 3 Types of Sleep Apnea Explained: Obstructive, Central, & Mixed Posted by Kevin Phillips on Jan 28, 2015 6:53:00 PM Sleep apnea is a very common sleep disorder, affecting

More information

Goals Session Putting it all Together

Goals Session Putting it all Together Goals Session Putting it all Together Adherence to therapy Case presentations Treatment issues Charles Atwood, MD, FCCP, FAASM University of Pittsburgh Disclosures Advisory - Carecore National; Philips;

More information

EXPLORE NEW POSSIBILITIES

EXPLORE NEW POSSIBILITIES EXPLORE NEW POSSIBILITIES TREATING SNORING AND SLEEP APNOEA HAS CHANGED FOREVER Introducing the Oventus O 2 Vent, a custom made, comfortable oral appliance with a unique airway design for the treatment

More information

Medicare C/D Medical Coverage Policy. Respiratory Assist Devices for Obstructive Sleep Apnea and Breathing Related Sleep Disorders

Medicare C/D Medical Coverage Policy. Respiratory Assist Devices for Obstructive Sleep Apnea and Breathing Related Sleep Disorders Medicare C/D Medical Coverage Policy Respiratory Assist Devices for Obstructive Sleep Apnea and Breathing Related Sleep Disorders Origination: June 26, 2000 Review Date: January 18, 2017 Next Review January,

More information

All About Positive Airway Pressure (PAP) Therapy

All About Positive Airway Pressure (PAP) Therapy All About Positive Airway Pressure (PAP) Therapy Nitipatana Chierakul Division of Respiratory Disease and Tuberculosis, Department of Medicine, Siriraj Medical School Siriraj Sleep Center: Fiscal-year

More information

6. The Upper Airway. Anatomy of the Pharyngeal Airway in Sleep Apneics: Separating Anatomic Factors From Neuromuscular Factors

6. The Upper Airway. Anatomy of the Pharyngeal Airway in Sleep Apneics: Separating Anatomic Factors From Neuromuscular Factors Sleep, 16:S80-S84 1993 American Sleep Disorders Association and Sleep Research Society 6 The Upper Airway (a) Response to Anatomy Anatomy of the Pharyngeal Airway in Sleep Apneics: Separating Anatomic

More information

SLEEP DISORDERED BREATHING The Clinical Conditions

SLEEP DISORDERED BREATHING The Clinical Conditions SLEEP DISORDERED BREATHING The Clinical Conditions Robert G. Hooper, M.D. In the previous portion of this paper, the definitions of the respiratory events that are the hallmarks of problems with breathing

More information

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea Policy Number: Original Effective Date: MM.01.009 11/01/2009 Line(s) of Business: Current Effective Date: HMO; PPO

More information

Mario Kinsella MD FAASM 10/5/2016

Mario Kinsella MD FAASM 10/5/2016 Mario Kinsella MD FAASM 10/5/2016 Repetitive episodes of apnea or reduced airflow Due to upper airway obstruction during sleep Patients often obese Often have hypertension or DM 1 Obstructive apneas, hypopneas,

More information

Continuous Positive Airway Pressure (CPAP)

Continuous Positive Airway Pressure (CPAP) Patient Information Continuous Positive Airway Pressure (CPAP) Author: Pulmonary physiology, sleep and ventilation Produced and designed by the Communications Team Issue date July 2017 Expiry date July

More information

Titration protocol reference guide

Titration protocol reference guide Titration protocol reference guide 2 Notes Description Page Patient types 4 Titration protocol goals 5 CPAP CPAP protocol 6-7 Auto CPAP Auto CPAP protocol 8-9 BiPAP S BiPAP S protocol 10-11 BiPAP Auto

More information

Heated Humidification or Face Mask To Prevent Upper Airway Dryness During Continuous Positive Airway Pressure Therapy*

Heated Humidification or Face Mask To Prevent Upper Airway Dryness During Continuous Positive Airway Pressure Therapy* Heated Humidification or Face Mask To Prevent Upper Airway Dryness During Continuous Positive Airway Pressure Therapy* Maria Teresa Martins de Araújo, MS; Sérgio Barros Vieira, MD; Elisardo Corral Vasquez,

More information

The Use of Matrix Repatterning in the Management Upper Airway Obstruction (Snoring and Sleep Apnea)

The Use of Matrix Repatterning in the Management Upper Airway Obstruction (Snoring and Sleep Apnea) The Use of Matrix Repatterning in the Management Upper Airway Obstruction (Snoring and Sleep Apnea) A Preliminary Investigation Dr. George B. Roth Introduction: Upper airway obstruction (UAO) may be partial,

More information

Clinical update of BiPAP autosv for treatment of Sleep Disordered Breathing

Clinical update of BiPAP autosv for treatment of Sleep Disordered Breathing BiPAP autosv Advanced System One Authors: Dr. Teofilo Lee-Chiong, Medical Liaison, Philips Respironics Cheryl Needham, Senior Clinical Marketing Manager, Philips Respironics Bill Hardy, Senior Scientific

More information

Comparative Efficacy of Two Expiratory Pressure Reduction Systems in the Treatment of Obstructive Sleep Apnea

Comparative Efficacy of Two Expiratory Pressure Reduction Systems in the Treatment of Obstructive Sleep Apnea Sleep Diagnosis and Therapy 2009 V4 N1 Wolfe et al Comparative Efficacy of Two Expiratory Pressure Reduction Systems in the Treatment of Obstructive Sleep Apnea Lisa Wolfe, MD, 1 Clifford A. Massie, PhD,

More information

Pap Settings. A review of fine tuning settings For patient comfort and compliance Wendy Cook BSRT Judy Salisbury RPGST

Pap Settings. A review of fine tuning settings For patient comfort and compliance Wendy Cook BSRT Judy Salisbury RPGST Pap Settings A review of fine tuning settings For patient comfort and compliance Wendy Cook BSRT Judy Salisbury RPGST Conflict of Interest Disclosure x 1. I do not have any relationships with any entities

More information

O bstructive sleep apnoea (OSA) is a common condition

O bstructive sleep apnoea (OSA) is a common condition 226 SLEEP DISORDERED BREATHING Comparison of three ways to determine and deliver pressure during nasal CPAP therapy for obstructive sleep apnoea S D West, D R Jones, J R Stradling... See end of article

More information

Increasing the Functional Residual Capacity May Reverse Obstructive Sleep Apnea

Increasing the Functional Residual Capacity May Reverse Obstructive Sleep Apnea Sleep 11(4):349-353, Raven Press, Ltd., New York 1988 Association of Professional Sleep Societies ncreasing the Functional Residual Capacity May Reverse Obstructive Sleep Apnea F. Series, Y. Cormier, N.

More information

Monitoring: gas exchange, poly(somno)graphy or device in-built software?

Monitoring: gas exchange, poly(somno)graphy or device in-built software? Monitoring: gas exchange, poly(somno)graphy or device in-built software? Alessandro Amaddeo Noninvasive ventilation and Sleep Unit & Inserm U 955 Necker Hospital, Paris, France Inserm Institut national

More information