Long-Term Results With the Medtronic-Hall Valvular Prosthesis
|
|
- Arlene Wright
- 5 years ago
- Views:
Transcription
1 Long-Term Results With the Medtronic-Hall Valvular Prosthesis Cary W. Akins, MD Cardiac Surgical Unit, Massachusetts General Hospital, Boston, Massachusetts Background. Although more than 170,000 Medtronic- Hall mechanical valvular prostheses have been inserted world-wide, long-term results are available on only a small percent of those valves inserted. Methods. A prospective data registry of all Medtronic- Hall cardiac prostheses inserted by one surgeon was used to identify 460 valves inserted during 391 operations from 1983 to 1994: single aortic (n = 210), single mitral (n = 115), or double aortic and mitral (n = 66) replacements, including three tricuspid valve replacements. Follow-up was sought five times in 10 years and was available for 280 (99%) of 283 survivors with only an isolated aortic or mitral Medtronic-Hall valve followed up for at least 1 year (1,246 patient-years). Results. Hospital mortality was 4.6% (18 patients). Of 40 late deaths, eight were valve-related (0.6% per patientyear). The linearized rates of complications for aortic and mitral valve replacements (percent per patient-year) were, respectively: structural deterioration, 0 and 0; nonstructural dysfunction, 0.1 and 2.1; thromboembolism, 1.3 and 2.1; thrombosis, 0 and 0.2; anticoagulant-related bleeding, 1.7 and 1.9; and prosthetic valve endocarditis, 0.6 and 1.0. Actuarial freedom from reoperation at 10 years was 97% for aortic and 88% for mitral valves. Conclusions. The Medtronic-Hall mechanical valvular prosthesis has excellent durability and acceptably low rates of valve-related complications and remains my mechanical prosthetic valve of choice for both aortic and mitral valve replacements. (Ann Thorac Surg 1996;61:806-13) cardiac surgeons choose a prosthesis for valular replacement, they are required to make a compromise between the excellent durability of approved mechanical prostheses and the greater freedom from thromboembolism and diminished requirement for anticoagulation associated with current bioprostheses. The Medtronic-Hall mechanical cardiac valvular prosthesis, one of the latest evolutionary designs of singledisc prostheses, offers a wider opening angle, a larger minor orifice, and translation as well as rotation of the disc on a central, machined strut. These features were developed to offer potentially greater durability, improved hemodynamics, and lower thromboembolic rates. This report documents the 10-year clinical experience with the Medtronic-Hall valve by one surgeon in one institution. The series is divided into two parts: (1) an in-hospital study, which details the preoperative information and operative result for all patients having insertion of a Medtronic-Hall valve, and (2) a long-term follow-up review for those patients having only an isolated aortic or mitral Medtronic-Hall valve for whom follow-up information was available for more than I year after hospitalization. Material and Methods Patients Beginning in October 1983, when the first Medtronic-Hall valve was implanted at the Massachusetts General Hospital, a registry of all patients having implantation of a Accepted for publication Oct 30, Address reprint requests to Dr Akins, Cardiac Surgical Unit, Massachusetts General Hospital, White 503, 32 Fruit St, Boston, MA Medtronic-Hall valve was begun. This prospective registry contains preoperative information on each patient, details of the operation, and a record of postoperative complications. All of the information about the patient's hospitalization for valve insertion was obtained by trained research personnel on a prospective basis. All patients who had a Medtronic-Hall valve implanted by one surgeon are the subject of this report. From October 1983 through December 1994, 460 Medtronic-Hall valvular prostheses were inserted during 391 operations, which constitute the in-hospital series. Valve implantation into only the aortic position (AVR) was performed in 210 operations, the mitral position (MVR) in 115 operations, and combined aortic and mitral positions (DVR) in 66 operations. In addition three valves were inserted into the tricuspid position during DVR. Demographic and clinical presentation characteristics of the patient population, segregated according to type of valve operation, are presented in Table 1. Aortic valve replacement was more commonly performed in men, and MVR and DVR in women. Most patients having MVR or DVR presented with congestive heart failure. Of the 391 total operations performed, 148 (37.9%) were valve reoperations. For the 210 patients having only an AVR, 51 (24.2%) were reoperations: first-time in 45 (21.4%), second-time in 5 (2.4%), and third-time in 1 (0.5%). For the 115 patients having only an MVR, 59 (51.3%) were reoperations: first-time in 42 (36.5%), second-time in 12 (10.4%), and third-time in 5 (4.3%). For the 66 patients having DVR, 38 (57.6%) were reoperations: first-time in 27 (40.9%), second-time in 9 (13.6%), and third-time in 2 (3.0%). Selected findings from cardiac catheterizations, ac by The Society of Thoracic Surgeons /96/$15.00 Published by Elsevier Science Inc SSDI (95)
2 Ann Thorac Surg AKINS ;61: MEDTRONIC-HALL VALVE Table 1. Demographic and Clinical Characteristics AVR MVR DVR Characteristic (210) (115) (66) Sex Male 160 (76.2%) 31 (27.0%) 22 (33.3%) Female 50 (23.8%) 84 (73.0%) 44 (66.7%) Age (mean, y) NYHA class (mean) Symptoms Congestive heart 110 (52.4%) 100 (86.9%) 58 (87.9%) failure Angina pectoris 85 (40.5%) 10 (8.7%) 13 (19.7%) Syncope 25 (11.9%) 2 (1.7%) 4 (6.1%) Endocarditis 23 (10.9%) 7 (6.1%) 5 (7.6%) Stroke 9 (4.3%) 25 (21.7%) 10 (15.1%) AVR - aortic valve replacement; MVR - mitral valve replacement; ation. DVR = double valve replacement; NYHA New York Heart Associ- cording to valve operation, are noted in Table 2. Coronary artery disease was present in 122 patients (31.2%). In one DVR reoperation an aortic bioprosthesis was removed only because it was 11 years old, and in another DVR reoperation a rnitral bioprosthesis was removed only because the patient had endocarditis on the aortic valve extending into the mitral annulus. Operative Techniques Operations were performed during standard cardiopulmonary bypass with moderate hypothermia (25 to 28 C). Venous cannulation was accomplished through the right atrium, or occasionally in addition via the femoral vein. Arterial cannulation was performed through the ascending aorta, although for some composite aortic valve and ascending aortic root replacements perfusion was performed through a femoral artery. Myocardial protection was achieved with either cold, oxygenated crystalloid cardioplegia or cold, oxygenated blood cardioplegia, instilled antegrade either into the aortic root or selectively into the coronary orifices. Valves were secured to the annulus with either figureof-8 sutures or pledgeted mattress sutures, if the annulus was friable. In the aortic position the greater orifice of the valve was aimed toward the greater curvature of the aorta, usually about the middle of the noncoronary sinus. In the mitral position the greater orifice of the valve was directed posteriorly, with the occluder of the prosthesis functioning essentially as an anterior mitral leaflet. In all cases complete freedom of motion of the occluder was obtained, even if it meant rotating the housing to a position somewhat different from that usually selected. The native mitral leaflets were not routinely preserved in this series for several reasons. In the first place 59 (51.3%) of the 115 isolated MVR and 38 (57.6) of the 66 DVR operations were reoperations. Second, virtually all native mitral valves replaced were rheumatically diseased or had active endocarditis and were completely excised to allow implantation of the largest possible prosthesis. (During these study years, I reconstructed 96% of mitral valves that presented with pure, nonrheumatic mitral regurgitation.) Anticoagulant Management Anticoagulant therapy with warfarin was begun as soon as the patient could receive oral medications, usually on the first postoperative day. The dosage of warfarin was adjusted to achieve target prothrombin times that differed by valve position and patient risk factors [1]. The target international normalized ratios sought were as follows: low-risk AVR patients, 2.0 to 2.5; high-risk AVR patients, 2.5 to 3.0; low-risk MVR patients, 2.5 to 3.0; and high-risk MVR patients, 3.0 to 3.5. If the prothrombin time was subtherapeutic on the fifth postoperative day, intravenous administration of low molecular weight dextran, or more recently heparin, was begun, and continued until the prothrombin time became therapeutic. Antiplatelet agents were not routinely administered to any patient. Follow-up During the 10 years of the study, complete follow-up was sought five times, on average every 2 years. Follow-up information about survival and valve-related events was obtained through direct communication with the patients by trained research personnel. If subsequent hospitalization, death, or valve-related events had occurred, the patient's physician or appropriate hospital record department was contacted to document the events and hospitalizations. All follow-up information was organized to conform with the published guidelines for reporting morbidity and mortality after cardiac valvular operations [2]. Specifically, operative mortality was defined as any death that occurred during the hospitalization in which valve implantation was performed, or death within 30 days of the operation if the patient was discharged before that time. Further, paravalvular leak was recorded not only for those patients who required reoperation because of Table 2. Catheterization Findings AVR MVR DVR Finding (210) (115) (66) Ejection fraction (mean) Valve lesion Aortic stenosis 108 (51.4%) (33.3%) Aortic regurgitation 74 (35.3%) (33.3%) Aortic stenosis and 28 (13.3%) (31.8%) regurgitation Mitral stenosis... 9 (7.8%) 17 (25.8%) Mitral regurgitation (52.2%) 31 (47.0%) Mitral stenosis and (40.0%) 17 (25.8%) regurgitation Coronary artery 82 (39.0%) 27 (23.5%) 13 (19.7%) disease AVR = aortic valve replacement; MVR - mitral valve replacement. DVR = double valve replacement;
3 808 AKINS Ann Thorac Surg MEDTRONIC-HALL VALVE 1996;61: the complication, but also for any patient who had a bland periprosthetic leak demonstrated by echocardiography, regardless of symptoms. All early and late valverelated events were counted. The most recent follow-up was performed between June 1993 and March 1994 and was recorded as of June 30, Follow-up was sought for those patients with only an isolated aortic or mitral Medtronic-Hall valve that had been in place for at least 1 year. Patients with multiple valve replacements or intracardiac valvular prostheses of another type were excluded from followup. Patients were included in follow-up if they had coronary artery bypass grafting (CABG), ascending aortic grafts, or suture repair of other intracardiac or aortic defects. Of the 177 AVR patients appropriate for longterm evaluation, only 1 patient was lost to follow-up, yielding complete follow-up in 99%. Average follow-up for AVR patients was 4.3 years, and cumulative follow-up was 765 patient-years. Of the 106 MVR patients appropriate for long-term study, only 2 patients were lost to follow-up, yielding complete follow-up in 98%. Average follow-up for the MVR patients was 4.6 years, and cumulative follow-up was 481 patient-years. Statistical Analysis All early and late episodes or events were included in the calculation of the linearized incidence of complications or events and were expressed as number of events per 100 patient-years of follow-up, segregated by valve position. To describe the long-term results in accordance with the guidelines and to compare those results between aortic and mitral valves, actuarial curves for mortality and late complications were obtained by the life-table method [3]. The significance of differences between the two groups was determined by means of the Mantel-Cox test in the BMDP program PIL [4]. To assess the impact of the valve position, age, and associated coronary artery disease on long-term survival, these factors were inserted into a multiple stepwise logistic regression algorithm, BMDP program PLR [4]. Results Operative Mortality Operative mortality for the total group was 4.6% (18/391). The operative mortality by valve position was 2.9% for AVR (0.9% for 110 isolated AVR), 4.3% for MVR (3.6% for 84 isolated MVR), and 10.6% for DVR (8.9% for 45 isolated DVR). Twelve of the 18 deaths occurred after the 148 reoperative valve replacements, yielding an operative mortality rate for reoperations of 8.1%, compared with 2.5% for first-time single or double valve replacements. Late Mortality Late deaths occurred in 40 (14.3%) of the 280 hospital survivors of AVR or MVR followed up long-term. Linearized mortality was 2.5% per patient-year for hospital survivors of AVR and 4.4% per patient-year for hospital survivors of MVR. The causes of late mortality are noted in Table 3. Table 3. Causes of Late Mortality Cause of Death AVR MVR Total Valve-related Thromboembolism Anticoagulant hemorrhage Prosthetic endocarditis Sudden, unexplained Other cardiac Heart failure Coronary artery disease Noncardiac AVR = aortic valve replacement; MVR = mitral valve replacement. Actuarial survival at 10 years, including operative mortality, was 72.4% % after AVR and 62.6% + 6.6% after MVR (p = not significant). Actuarial survival for AVR and MVR patients, separated according to the presence or absence of CABG, is depicted in Figure 1. At 10 years actuarial survival for the AVR patients was significantly better than for AVR + CABG patients: 79.3% % versus 58.1% ± 11.4%, respectively (p < 0.02). Actuarial survival at 10 years for MVR compared with MVR + CABG patients was 67.3% + 6.9% versus 41.5% % (p = 0.059). However, the mean ages of the patients in the AVR, AVR + CABG, MVR, and MVR + CABG subgroups were , 62 _~ 6, 62 _+ 9, and years, respectively. Therefore, to further delineate the true impact of CABG and age on actuarial survival for AVR and MVR patients, we employed a multiple stepwise logistic regression algorithm into which age, valve position, and CABG were inserted as variables. In univariate testing age was by far the strongest predictor of death (p < ), and CABG was a weaker predictor (p = 0.01). Valve position did not quite attain statistical significance as a predictor of death (p = 0.08). When interactions between effects of valve position and CABG on survival were sought, only in the case of MVR did adding CABG significantly decrease survival (p = 0.02). In the multivariate testing, when the patient population was corrected for age, valve position was not a significant predictor of death, and CABG just failed to attain statistical significance (p = 0.06). Valve-Related Complications Table 4 contains the incidence and linearized rates and Table 5 the 5-year, 8-year, and 10-year actuarial eventfree rates for valve-related complications and consequences for those patients followed up long-term. STRUCTURAL DETERIORATION. Structural deterioration did not occur in any patient in this study. NONSTRUCTURAL DYSFUNCTION. No patient in this study had clinically significant hemolysis that was not secondary to paravalvular leak. Nonstructural dysfunction occurred in only 1 AVR patient, who at 42 months postoperatively had removal of the valve at another hospital, where no abnormality of
4 Ann Thorac Surg AK1NS ;61: MEDTRONIC-HALL VALVE Fig 1. Actuarial survival for aortic (AVR) and mitral valve replacement (MVR) patients segregated according to the presence of coronary artery bypass grafting (CABG). 5o- = Avs ~ \1 4 $ -- Av~, r,u~ '/- 6.2 P<0.02 ~ ~ I~ l l I I I I I I I I l I $ 9 to YEAR the valve was reportedly identified except possibly for some pannus ingrowth. No AVR patient had a periprosthetic leak that was not due to endocarditis. The linearized rate for nonstructural dysfunction after AVR was 0.1% per patient-year, and the 10-year actuarial eventfree rate was 99.1% _+ 0.9%. Bland periprosthetic leak was documented in 10 patients after MVR, for a linearized rate of 2.1% per patientyear and a 10-year actuarial event-free rate of 83.0% + 5.5%. Of the 10 paravalvular leaks, eight occurred in patients after reoperative MVR (five first-time reoperations, three second-time reoperations) at a mean of 44 months postoperatively. Four asymptomatic patients had incidental leaks discovered by echocardiography and were followed up medically. Five patients required reoperation; all of these reoperations were successful. One patient was awaiting reoperation (which was performed Table 4. Incidence and Linearized Rates of Complications and Consequences AVR MVR Linearized Linearized Event Number Rate a Number Rate ~ Late death Nonstructural dysfunction Thromboembolism Thrombosis Anticoagulant bleeding Prosthetic endocarditis Reoperation Physical impairment Valve-related death All valve-related morbidity and mortality a Linearized rate expressed as percent per patient-year AVR = aortic valve replacement; MVR = mitral valve replacement. AVR AVR,CABG MVR MVR,CABG successfully after the closing date of the study). The linearized rate for nonstructural dysfunction requiring reoperation after MVR was 1.2% per patient-year. THROMBOEMBOLISM. Valvular thrombosis did not develop in any AVR patient in this study. Thromboembolism was documented in ten events in 10 patients after AVR: five after isolated AVR, four after AVR + CABG, and one after composite AVR + ascending aortic replacement. A history of atrial fibrillation could be documented in 6 of the patients. Only one event occurred during the valve replacement hospitalization at 5 days after AVR + CABG. The mean interval from operation to thromboembolism for the other 9 patients was 45 months. One late thromboembolic event was fatal. The linearized rate for thromboembolism after AVR was 1.3% per patient-year, and the 10-year actuarial eventfree rate was 87.3% ± 4.4% (Fig 2). Valvular thrombosis occurred in I MVR patient in this study, for a linearized rate of 0.2% per patient-year and a 10-year actuarial event-free rate of 99.0% ~+ 1.0%. The patient presented with gastrointestinal bleeding 6 months after MVR and was treated with fresh frozen plasma and vitamin K. He progressively lost crisp valve sounds and was then referred for reoperation, at which time his valve was successfully replaced with a bioprosthesis. Thromboembolism after MVR was reported on nine occasions in 8 patients, of whom 5 had atrial fibrillation. In all 5 of the patients for whom anticoagulation data were available, the prothrombin time was suboptimal. No event occurred during valve replacement hospitalization. The nine postoperative events were reported at a mean of 19 months postoperatively. One late thromboembolic event was fatal. The linearized rate for thromboembolism, including the valve thrombosis, after MVR was 2.1% per patient-year, and the 10-year actuarial event-free rate was 90.8% ± 3.2% (see Fig 2). ANTICOAGULANT-RELATED HEMORRHAGE. Anticoagulantrelated bleeding occurred in 13 patients after AVR. Gastrointestinal bleeding developed in 1 patient in the hospital 10 days after AVR + CABG. The remaining 12
5 810 AKINS Ann Yhorac Surg MEDTRONIC-HALL VALVE 1996;61: Table 5. Actuarial Event-Free Rates From Complications and Consequences AVR Event Five -Year Eight-Year Ten-Year Five -Year Eight-Year Ten-Year Death 86.1% _+ 3.4% 75.4% ± 5.1% 72.4% 5.7% Nonstructural dysfunction 99.1% +_ 0.9% 99.1% _+ 0.9% 99.1% +_ 0.9% Thromboembolism 95.4% +_ 1.9% 87.3% _+ 4.4% 87.3% +_ 4.4% Thrombosis 100% 100% 100% Anticoagulant bleeding 93.8% ± 2.8% 86.6% _+ 4.2% 80.0% % Prosthetic endocarditis 97.9% ± 1.2% 95.8% % 95.8% _+ 2.4% Reoperation 97.0% ± 1.5% 97.0% _+ 1.5% 97.0% _+ 1.5% Physical impairment 99.1% ± 0.9% 96.2% +_ 3.0% 96.2% % Valve-related death 99.2% _+ 0.8% 97.2% +_ 2.1% 97.2% _+ 2.1% All valve-related morbidity 87.2% +_ 3.0% 76.7% _+ 4.9% 70.8% % and mortality MVR 80.3% % 65.5% % 62.6% % 93.1% % 83.0% % 83.0% % 90.8% % 90.8% % 90.8% % 99.0% % 99.0% + 1.0% 99.0% ± 1.0% 91.2% % 85.6% % 85.6% % 95.2% ± 2.7% 88.2% ± 5.4% 88.2% % 93.1% + 2.8% 88.0% + 4.4% 88.0% % 96.8% % 96.8% + 1.8% 96.8% ± 1.8% 95.5% % 87.4% % 87.4% % 71.6% ± 5.1% 54.2% % 54.2% % AVR = aortic valve replacement; MVR = mitral valve replacement. events occurred at a mean of 45 months postoperatively. One patient died of intracranial hemorrhage secondary to trauma 25 months after AVR + CABG with composite ascending aorta and aortic arch replacement. The linearized rate for anticoagulant-related hemorrhage following AVR was 1.7% per patient-year, and the ten-year actuarial event-free rate was % (Figure 3). After MVR, 9 patients experienced anticoagulantrelated bleeding, including 2 patients who died of intracerebral hemorrhage at 20 and 66 months postoperatively. No event occurred during MVR hospitalization. The mean postoperative interval for the nine events was 32 months. The linearized rate for anticoagulant-related hemorrhage after MVR was 1.9% per patient-year, and the 10-year actuarial event-free rate was 85.6% ± 4.9% (Fig 3). PROSTHETIC ENDOCARDITIS. After AVR, endocarditis developed in 5 patients, early in none. The mean time to the complication was 46 months. All 5 patients required and survived reoperation, one of which was performed after the closing date for follow-up. The linearized rate for endocarditis after AVR was 0.6% per patient-year, and the 10-year actuarial event-free rate was 95.8% ± 2.4%. Prosthetic endocarditis occurred in 5 patients after MVR, early in none, at a mean postoperative interval of 59 months. Two patients survived reoperation, 2 died before operation could be undertaken, and I was treated successfully with antibiotics alone. The linearized rate for endocarditis after MVR was 1.0% per patient-year, and the 10-year actuarial event-free rate was 88.2% %. Valve-Related Consequences REOPERATION. Valve replacement after AVR was required in 5 patients: 4 for endocarditis and 1 for presumed pannus ingrowth. The mean interval to reoperation was 42 months. All patients survived reoperation. The linearized rate for reoperation after AVR was 0.6% per patientyear, and the 10-year actuarial event-free rate was 97.0% ± 1.5% (Fig 4). After MVR, 8 patients required reoperation at a mean postoperative interval of 29 months: 5 for periprosthetic leak 2 for endocarditis, and 1 for valve thrombosis. All patients survived reoperation. The linearized rate for reopera- Fig 2. Actuarial event-free rates for thromboembolism for aortic (AVR) and mitral valve replacement (MVR) patients. (NS = not significant.) 100 ~..~ - -,.,.., % [ PERCENT FREE 10 V~OtS -~ AVR /- 4.4% p:ns "--O--- MVR /- 3.2% 10 AVR 158 IM'W 86 o/ o l l I I I I I l YEAR
6 Ann Thorac Surg AKINS ;61: MEDTRONIC-HALL VALVE,oo % % Fig 3. Actuarial event-free rates for anticoagulantrelated bleeding for aortic (AVR) and mitral valve replacement (MVR) patients. (NS = not significant.) PERCENT FREE ~. AVR /- 7.5% p=ns ---'O'--- MVR /- 4.9% 20-10" ~VR 158 MVR89 0 I I I I I I I I I I l0 YEAR tion after MVR was 1.7% per patient-year, and the 10-year actuarial event-free rate was 88.0% + 4.4% (see Fig 4). PERMANENT PHYSICAL IMPAIRMENT. Permanent physical impairment after AVR occurred in 2 patients, both secondary to thromboembolism. The linearized rate for permanent physical impairment after AVR was 0.3% per patient-year, and the 10-year actuarial event-free rate was 96.2% + 3.0%. After MVR, permanent physical impairment developed in 3 patients, secondary to thromboembolism in all 3. The linearized rate for permanent physical impairment after MVR was 0.6% per patient-year, and the 10-year actuarial event-flee rate was 96.8% + 1.8%. VALVE-RELATED DEATH. After AVR, valve-related death occurred in 2 patients (see Table 3). The linearized rate for valve-related death after AVR was 0.3% per patientyear, and the 10-year actuarial event-free rate was 97.3% + 2.1%. Valve-related death occurred in 6 patients after MVR (see Table 3). The linearized rate for valve-related death after MVR was 1.2% per patient-year, and the 10-year actuarial event-free rate was 87.4% + 5.1%. ALL VALVE-RELATED MORBIDITY AND MORTALITY. The linearized rate for all mortality and valve-related morbidity, including hospital death, after AVR was 4.6% per patientyear, and the 10-year actuarial event-free rate was 70.8% + 7.3% (Fig 5). The linearized rate for all mortality and valve-related morbidity, including hospital death, after MVR was 8.1% per patient-year, and the 10-year actuarial event-free rate was 54.2% _+ 7.0% (see Fig 5). Comment Although more than 170,000 Medtronic-Hall valves have been inserted worldwide, follow-up data are available on only a small percent of valves implanted [5]. This report documents the clinical experience of one surgeon using this valve exclusively as a mechanical replacement for more than 10 years. The functional performance of the Medtronic-Hall % Fig 4. Actuarial event-free rates for reoperation for aortic (AVR) and mitral valve replacement (MVR) patients. b, O l0- AVR MVR I I PERCENT FREE 10 YEARS AVR /- 1.5% p = 0.06 MVR /- 4.4% I I I I I! I I YEAR
7 812 AKINS Ann Thorac Surg MEDTRONIC -HALL VALVE 1996;61: Fig 5. Actuarial event-free rates for all mortality and valve-related morbidity for aortic (AVR) and mitral valve replacement (MVR) patients. 1oo 9o 70.89t A w A v 10- AVR MVR i i PERCENT FREE lo YV.AltS AVR /- 7.3 p = MVR / I I I I I I I I % valve is notable for its low profile, rotatability, excellent gradient relief even in small sizes [6, 7], relatively low regurgitant fraction [8], and low incidence of hemolysis [9]. Its excellent durability, due in large part to the fact that the housing and strut mechanism are completely machined from one solid piece of titanium, led to its choice as the prosthesis for the Jarvik total artificial heart [10]. There has never been a report of the clinical loss of structural integrity of a standard Medtronic-Hall valve, the only model available in the United States. In prospective, randomized comparisons the Medtronic-Hall valve has been found to be at least comparable with the St. Jude Medical valve in terms of hemodynarnic performance and incidence of late valve-related complications in the rnitral position in one study [6] and in both aortic and mitral positions in another study [11]. In a prospective, randomized comparison with the Monostrut prosthesis the Medtronic-Hall valve was found to be somewhat better in terms of thrombogenicity and at least comparable in terms of other late valverelated complications for AVR, MVIL and combined valve replacements [12]. In a recent retrospective review of late results with Starr-Edwards models 1260 and 6120, St. Jude Medical, Omniscience, CarboMedics, and Medtronic-Hall mechanical valves, the Medtronic-Hall valve had the best freedom from combined thromboembolic and anticoagulant-related bleeding complications [13]. As noted in that review, the principal disadvantage of the Medtronic- Hall valve is the careful attention required of the implanting surgeon to ensure that the motion of the occluder is not impinged. Increasing experience with this prosthesis, however, has shown a dramatic decrease in the reported occurrence of that complication [5]. Although it is tempting to impugn the design features of a prosthetic valve when implantation complications occur, in most cases the problem is not related to the prosthesis but to the insertion technique. As with any prosthetic valve, the surgeon must be certain that the device is implanted properly, which means cutting the tails of sutures appropriately short and debriding any native valve remnants that might interfere with occluder func- tion. Rotatability of the housing within the sewing ring is a great advantage in helping to avoid intracardiac structures that might hinder normal disc motion. The orientation of the Medtronic-Hall valve that I prefer in the aortic position is to have the greater orifice of the prosthesis directed toward the greater curve of the ascending aorta, which usually means about the middle of the noncoronary sinus. In the rnitral position the valve should be oriented so that the occluder behaves like an anterior rnitral leaflet, that is, with the greater orifice directed posteriorly. This yields the normal flow of blood: first down the posterolateral wall of the left ventricle to the apex and then up the septum to the left ventricular outflow tract, as we have confirmed with intraoperative transesophageal echocardiography. Bioprostheses cannot mimic this flow [14], nor do bileaflet mechanical valves in our experience. The actuarial event-free rates in the current series demonstrate improved survival and significantly better event-free survival for AVR patients than for MVR patients (see Table 5). Of interest is that the actuarial event-free rates for thromboembolism and anticoagulant-related bleeding at 10 years, although not significantly different, favor the mitral prostheses over aortic prostheses, whereas the linearized rates are lower for aortic valve patients for both complications (see Table 4). Part of the reason for this apparent contradiction may lie with the fact that the majority of events in the mitral patients occurred in the earliest years after operation, whereas events in the aortic patients were spread more broadly through the follow-up period. Linearized calculations assume that the hazard function for any complication is a constant throughout the follow-up period, while actuarial determinations assess the risk over time. Furthermore, the two thromboembolic events that occurred in 1 MVR patient would be counted twice for the linearized rate but only once for the actuarial rate. Patients having MVR do have higher linearized rates for and poorer long-term freedom from nonstructural dysfunction, prosthetic endocarditis, reoperation, and valve-related death than do AVR patients (see Table 4). Of these complications, nonstructural dysfunction re-
8 Ann Thorac Surg AKINS ;61: MEDTRONIC-HALL VALVE quires some additional discussion. The linearized rate for nonstructural dysfunction for mitral valves of 2.1% per patient-year is higher than that reported in most other studies [5]. There are several possible reasons for this. In the first place all paravalvular leaks were counted in this study, not just those that required reoperation or led to death. Four of the ten leaks reported were minor asymptomatic leaks detected at late echocardiography. Counting only the six other leaks, the linearized rate for significant nonstructural dysfunction would be 1.2% per patient-year, a rate comparable with those reported in other published series. Also, of the ten leaks, eight occurred at a mean of 44 months postoperatively in patients whose MVR was a reoperation (five first-time and three second-time reoperations), despite the frequent use of pledgeted sutures. Poor tissue quality may have contributed to the late appearance of these complications. Fortunately all 6 patients who came to reoperation survived without complications. One other feature of this study that is worthy of comment is that complete follow-up was sought five times during the course of the study, or on average every 2 years, not just once at the end of the study interval. We know that the more often patients are interviewed for results, the more complications they will remember, and thus the greater will be the reported rates of complications. The impact of this message is well illustrated by the report of Horstkotte and associates [15] on the St. Jude Medical valve. Patients in their study, with its average follow-up of 114 months, were reexamined at a median interval of 8 months. This intense follow-up protocol yielded linearized thromboembolic rates of 2.7% and 4.4% per patient-year for aortic and mitral valves, respectively, and those rates do not include events that occurred within the first 3 months after valve implantation or more than the first event for patients who suffered multiple events. Similarly the linearized anticoagulantrelated bleeding rates for aortic and mitral valves in that study were 4.1% and 6.4% per patient-year, respectively. Thus, much of the importance of complication rates reported in any series is a function of the intensity with which reports of those complications are sought. The lowest complication rates would be measured when follow-up is sought only once after a long time interval. The impact of concomitant CABG on mortality is obvious, particularly in patients requiring MVR. As many late deaths were due to coronary disease as to valverelated complications (see Table 3). Even after patient subgroups were corrected for age, CABG just failed to be a significant predictor of death (p = 0.06). With a larger patient population the age-corrected impact of coronary disease would probably be significant, as would the differences between AVR and MVR patients. This deleterious impact of coronary disease on the long-term survival of valve replacement patients has been previously examined by Jones and associates [16]. The results in this study corroborate their findings and support the contention that associated coronary artery disease should be strongly considered when choosing a mechanical valve or bioprosthesis, particularly when the patient is more than 60 years of age and requires an MVR. As with Jones and associates' study, the current study found less impact of coronary disease in AVR patients. In summary, the long-term follow-up of patients who received the Medtronic-Hall prosthesis in this study revealed that the valve has superb durability and the incidence of valve-related complications was low and certainly comparable with those reported for other prostheses. These results support the continued use of the Medtronic-Hall valve as my prosthesis of choice when replacement of the aortic or mitral valve with a mechanical prosthesis is indicated. I thank Barbara Akins, BSN, and Annetta Boisselle, BSN, for help in data acquisition and John Newell, Director of the Cardiac Computer Center, Massachusetts General Hospital, for assistance with statistical evaluation. References 1. Butchart EG, Lewis PA, Bethel JA, Brechenridge IM. Adjusting anticoagulation to prosthesis thrombogenicity and patient risk factors. Circulation 1991;84(Suppl 3): Edmunds LH Jr, Clark RE, Cohn LH, Miller DC, Weisel RD. Guidelines for reporting morbidity and mortality after cardiac valvular operations. Ann Thorac Surg 1988;46: Kaplan EL, Meier P. Nonparametric estimation from incomplete observation. J Am Star Assoc 1958;53: Dixon WJ, Brown MB, Engleman L, Jennrich RI. BMDP statistical software manual, release ;2: Akins CW. Review of the global experience with the Medtronic-Hall valve. Eur J Cardiothorac Surg 1992;6(Suppl 1):$ Fiore AC, Naunheim KS, D'Orazio S, et al. Mitral valve replacement: randomized trial of St. Jude and Medtronic- Hall prostheses. Ann Thorac Surg 1992;54: Barner HB, Labovitz AJ, Fiore AC. Prosthetic valves for the small aortic root. J Card Surg 1994;9(Suppl): Walker DK, Scotten LN. A database obtained from in vitro function testing of mechanical heart valves. J Heart Valve Dis 1994;3: Skoularigis J, Essop MR, Skudicky D, Middlemost SJ, Sareli P. Frequency and severity of intravascular hemolysis after left-sided cardiac valve replacement with Medtronic Hall and St. Jude Medical prostheses and influence of prosthetic type, position, size and number. Am J Cardiol 1993;71: DeVries WC, Anderson JL, Joyce LD, et al. Clinical use of the total artificial heart. N Engl J Med 1984;310: Antunes MJ. Clinical performance of St. Jude and Medtronic-Hall prostheses: a randomized comparative study. Ann Thorac Surg 1990;50: Kim YI, Lesaffre E, Scheys 1, Stalpaert G, Flameng WJ, Daenen WJ. The Monostrut versus Medtronic-Hall prosthesis: a prospective randomized study. J Heart Valve Dis 1994;3: Akins CW. Results with mechanical cardiac valvular prostheses. Ann Thorac Surg 1995;60: McKenney PA, Davidoff R, Faxon DP, Shemin RJ. Reversal of left ventricular flow pattern following bioprosthetic mitral valve replacement [Abstract]. Circulation 1991;84(Suppl 2): Horstkotte D, Schulte HG, Bircks W, Strauer BE. Lower intensity anticoagulation therapy results in lower complication rates with the St. Jude Medical prosthesis. J Thorac and Cardiovasc Surg 1994;107: Jones EL, Weintraub WS, Craver JM, Guyton RA, Shen Y. Interaction of age and coronary disease after valve replacement: implications for valve selection. Ann Thorac Surg 1994;58:
16 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 6900
CLINICAL COMMUNIQUé 6 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 69 The Carpentier-Edwards PERIMOUNT Mitral Pericardial Valve, Model 69, was introduced into clinical
More informationThe CarboMedics bileaflet prosthetic heart was introduced
The CarboMedics Valve: Experience With 1,049 Implants José M. Bernal, MD, José M. Rabasa, MD, Francisco Gutierrez-Garcia, MD, Carlos Morales, MD, J. Francisco Nistal, MD, and José M. Revuelta, MD Department
More informationCLINICAL COMMUNIQUE 16 YEAR RESULTS
CLINICAL COMMUNIQUE 6 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 6900 Introduction The Carpentier-Edwards PERIMOUNT Mitral Pericardial Valve, Model 6900, was introduced
More informationTSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD
TSDA Boot Camp September 13-16, 2018 Introduction to Aortic Valve Surgery George L. Hicks, Jr., MD Aortic Valve Pathology and Treatment Valvular Aortic Stenosis in Adults Average Course (Post mortem data)
More informationClinical material and methods. Copyright by ICR Publishers 2003
Fourteen Years Experience with the CarboMedics Valve in Young Adults with Aortic Valve Disease Jan Aagaard 1, Jens Tingleff 2, Per V. Andersen 1, Christel N. Hansen 2 1 Department of Cardio-Thoracic and
More informationThe operative mortality rate after redo valvular operations
Clinical Outcomes of Redo Valvular Operations: A 20-Year Experience Naoto Fukunaga, MD, Yukikatsu Okada, MD, Yasunobu Konishi, MD, Takashi Murashita, MD, Mitsuru Yuzaki, MD, Yu Shomura, MD, Hiroshi Fujiwara,
More informationOutcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease
Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease TIRONE E. DAVID, MD ; SEMIN THORAC CARDIOVASC SURG 19:116-120c 2007 ELSEVIER INC. PRESENTED BY INTERN 許士盟 Mitral valve
More informationDurability and Outcome of Aortic Valve Replacement With Mitral Valve Repair Versus Double Valve Replacement
Durability and Outcome of Aortic Valve Replacement With Mitral Valve Repair Versus Double Valve Replacement Masaki Hamamoto, MD, Ko Bando, MD, Junjiro Kobayashi, MD, Toshihiko Satoh, MD, MPH, Yoshikado
More informationLong-term results (22 years) of the Ross Operation a single institutional experience
Long-term results (22 years) of the Ross Operation a single institutional experience Authors: Costa FDA, Schnorr GM, Veloso M,Calixto A, Colatusso D, Balbi EM, Torres R, Ferreira ADA, Colatusso C Department
More informationHani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz
Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia Decision process for
More informationEffect of Valve Suture Technique on Incidence of Paraprosthetic Regurgitation and 10-Year Survival
Effect of Valve Suture Technique on Incidence of Paraprosthetic Regurgitation and 10-Year Survival Sukumaran K. Nair, FRCS (C Th), Gauraang Bhatnagar, MBBS, Oswaldo Valencia, MD, and Venkatachalam Chandrasekaran,
More informationNineteen-Millimeter Aortic St. Jude Medical Heart Valve Prosthesis: Up to Sixteen Years Follow-up
Nineteen-Millimeter Aortic St. Jude Medical Heart Valve Prosthesis: Up to Sixteen Years Follow-up Dilip Sawant, FRCS, Arun K. Singh, MD, William C. Feng, MD, Arthur A. Bert, MD, and Fred Rotenberg, MD
More informationDepartment of Cardiothoracic Surgery, Heart and Lung Center, Lund University Hospital, Lund, Sweden
Long-Term Outcome of the Mitroflow Pericardial Bioprosthesis in the Elderly after Aortic Valve Replacement Johan Sjögren, Tomas Gudbjartsson, Lars I. Thulin Department of Cardiothoracic Surgery, Heart
More informationHani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia.
Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia. Decision process for Management of any valve Timing Feasibility
More informationThe CarboMedics prosthetic valve (Sulzer CarboMedics,
Long-Term Result of 1144 CarboMedics Mechanical Valve Implantations Chang Hyun Kang, MD, Hyuk Ahn, MD, Kyung Hwan Kim, MD, and Ki-Bong Kim, MD Department of Thoracic and Cardiovascular Surgery, Seoul National
More informationA valve was initiated at the Medical University of
St. Jude Prosthesis for Aortic and Mitral Valve Replacement: A Ten-Year Experience John M. Kratz, MD, Fred A. Crawford, Jr, MD, Robert M. Sade, MD, Arthur J. Crumbley, MD, and Martha R. Stroud, MS Division
More informationTHE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT
THE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT Gideon Cohen, MD Tirone E. David, MD Joan Ivanov, MSc Sue Armstrong, MSc
More informationBicuspid aortic root spared during ascending aorta surgery: an update of long-term results
Short Communication Bicuspid aortic root spared during ascending aorta surgery: an update of long-term results Marco Russo, Guglielmo Saitto, Paolo Nardi, Fabio Bertoldo, Carlo Bassano, Antonio Scafuri,
More informationT sors in the following aspects: the porcine aortic valve
Clinical and Hemodynamic Assessment of the Hancock I1 Bioprosthesis Tirone E. David, MD, Susan Armstrong, MSc, and Zhao Sun, MA Division of Cardiovascular Surgery, The Toronto Hospital and University of
More informationReoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment
Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment W.R.E. Jamieson, MD; L.H. Burr, MD; R.T. Miyagishima, MD; M.T. Janusz, MD; G.J. Fradet, MD; S.V. Lichtenstein, MD; H. Ling, MD Background
More informationBall Valve (Smeloff-Cutter) Aortic Valve Replacement Without Anticoagulation
Ball Valve (Smeloff-Cutter) Aortic Valve Replacement Without Anticoagulation Begonia Gometza, MD, and Carlos M. G. Duran, MD, PhD Department of Cardiovascular Diseases, King Faisal Specialist Hospital
More informationPresenter Disclosure. Patrick O. Myers, M.D. No Relationships to Disclose
Presenter Disclosure Patrick O. Myers, M.D. No Relationships to Disclose Aortic Valve Repair by Cusp Extension for Rheumatic Aortic Insufficiency in Children Long term Results and Impact of Extension Material
More information2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD
2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD David L Saint M.D. Tallahassee Memorial Hospital Southern Medical Group Division of Cardiothoracic
More informationTwenty-year experience with the St Jude Medical mechanical valve prosthesis
Surgery for Acquired Cardiovascular Disease Ikonomidis et al Twenty-year experience with the St Jude Medical mechanical valve prosthesis John S. Ikonomidis, MD, PhD John M. Kratz, MD Arthur J. Crumbley
More informationCarpentier-Edwards Pericardial Valve in the Aortic Position: 25-Years Experience
SURGERY: The Annals of Thoracic Surgery CME Program is located online at http://www.annalsthoracicsurgery.org/cme/ home. To take the CME activity related to this article, you must have either an STS member
More informationClinical event rates with the On-X bileaflet mechanical heart valve: A multicenter experience with follow-up to 12 years
Clinical event rates with the On-X bileaflet mechanical heart valve: A multicenter experience with follow-up to 12 years John B. Chambers, MD, FRCP, FACC, a Jose L. Pomar, MD, PhD, FETCS, b Carlos A. Mestres,
More informationUpdate on Oral Anticoagulation for Mechanical Heart Valves
Update on Oral Anticoagulation for Mechanical Heart Valves Douglas C. Anderson, Pharm.D., D.Ph. Professor and Chair Dept. of Pharmacy Practice Cedarville University School of Pharmacy OHIO SOCIETY OF HEALTH-SYSTEM
More informationMedtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance
Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance W. R. Eric Jamieson, MD, a Friedrich-Christian Riess, MD, b Peter J. Raudkivi, MD, c Jacques Metras, MD, d Edward F. G. Busse,
More informationReconstruction of the intervalvular fibrous body during aortic and
Aortic and mitral valve replacement with reconstruction of the intervalvular fibrous body: An analysis of clinical outcomes Nilto C. De Oliveira, MD Tirone E. David, MD Susan Armstrong, MSc Joan Ivanov,
More informationP have been used for mitral and aortic valve replacement
A -Year Comparison of Mitral Valve Replacement With Carpentier-Edwards and Hancock Porcine Bioprostheses P. Perier, MD, A. Deloche, MD, S. Chauvaud, MD, J. C. Chachques, MD, J. Relland, MD, J. N. Fabiani,
More informationThe use of mitral valve (MV) repair to correct mitral
Outcomes and Long-Term Survival for Patients Undergoing Repair Versus Effect of Age and Concomitant Coronary Artery Bypass Grafting Vinod H. Thourani, MD; William S. Weintraub, MD; Robert A. Guyton, MD;
More informationThe Ross Procedure: Outcomes at 20 Years
The Ross Procedure: Outcomes at 20 Years Tirone David Carolyn David Anna Woo Cedric Manlhiot University of Toronto Conflict of Interest None The Ross Procedure 1990 to 2004 212 patients: 66% 34% Mean age:
More informationValve Disease in Patients With Heart Failure TAVI or Surgery? Miguel Sousa Uva Hospital Cruz Vermelha Lisbon, Portugal
Valve Disease in Patients With Heart Failure TAVI or Surgery? Miguel Sousa Uva Hospital Cruz Vermelha Lisbon, Portugal I have nothing to disclose. Wide Spectrum Stable vs Decompensated NYHA II IV? Ejection
More informationAlthough mitral valve replacement (MVR) is no longer the surgical
Surgery for Acquired Cardiovascular Disease Ruel et al Late incidence and predictors of persistent or recurrent heart failure in patients with mitral prosthetic valves Marc Ruel, MD, MPH a,b Fraser D.
More informationThe St. Jude Medical Biocor Bioprosthesis
The St. Jude Medical Biocor Bioprosthesis Clinical Evidence of Long-term Durability Long-term Biocor Experience A Review and Comparative Assessment Long-term Biocor Stented Tissue Valve Studies Twenty-year
More informationLong-term Experience with the Bjork-Shiley Monostrut Tilting Disc Valve
J Korean Med Sci 2007; 22: 1060-4 ISSN 1011-8934 DOI: 10.3346/jkms.2007.22.6.1060 Copyright The Korean Academy of Medical Sciences Long-term Experience with the Bjork-Shiley Monostrut Tilting Disc Valve
More informationDoes Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement?
Original Article Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Hiroaki Sakamoto, MD, PhD, and Yasunori Watanabe, MD, PhD Background: Recently, some articles
More informationMechanical vs. Bioprosthetic Aortic Valve Replacement: Time to Reconsider? Christian Shults, MD Cardiac Surgeon, Medstar Heart and Vascular Institute
Mechanical vs. Bioprosthetic Aortic Valve Replacement: Time to Reconsider? Christian Shults, MD Cardiac Surgeon, Medstar Heart and Vascular Institute Assistant Professor, Georgetown School of Medicine
More informationInterventional procedures guidance Published: 26 September 2014 nice.org.uk/guidance/ipg504
Transcatheter valve-in-valve e implantation for aortic bioprosthetic valve dysfunction Interventional procedures guidance Published: 26 September 2014 nice.org.uk/guidance/ipg504 Your responsibility This
More informationMitral Valve Repair Versus Replacement in Simultaneous Mitral and Aortic Valve Surgery for Rheumatic Disease
Mitral Valve Repair Versus Replacement in Simultaneous Mitral and Aortic Valve Surgery for Rheumatic Disease Kenji Kuwaki, MD, PhD, Nobuyoshi Kawaharada, MD, PhD, Kiyofumi Morishita, MD, PhD, Tetsuya Koyanagi,
More information164 Ann Thorac Surg 45: , Feb Copyright by The Society of Thoracic Surgeons
Heart Valve Replacement with the Bjork-Shiley Mbnostrut Valve: Early Results of a Multicenter Clinical Investigation Lars I. Thulin, M.D., William H. Bain, F.R.C.S., Hans H. Huysmans, M.D., Gerrit van
More informationSpotlight on valvular heart disease guidelines. Prosthetic heart valves. Bernard Iung Bichat Hospital, Paris Diderot University Paris, France
Spotlight on valvular heart disease guidelines. Prosthetic heart valves. Bernard Iung Bichat Hospital, Paris Diderot University Paris, France Faculty disclosure First name - last name I disclose the following
More information15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses
ORIGINAL CONTRIBUTION 15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses WR Eric Jamieson, MD, Eva Germann, MSc, Michel R Aupart, MD 1, Paul H Neville, MD 1, Michel A Marchand,
More information42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim
42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim Current Guideline for AR s/p TOF Surgery is reasonable in adults with prior repair of
More informationORIGINAL PAPER. The long-term results and changing patterns of biological valves at the mitral position in contemporary practice in Japan
Nagoya J. Med. Sci. 78. 369 ~ 376, 2016 doi:10.18999/nagjms.78.4.369 ORIGINAL PAPER The long-term results and changing patterns of biological valves at the mitral position in contemporary practice in Japan
More informationTAVI- Is Stroke Risk the Achilles Heel of Percutaneous Aortic Valve Repair?
TAVI- Is Stroke Risk the Achilles Heel of Percutaneous Aortic Valve Repair? Elaine E. Tseng, MD and Marlene Grenon, MD Department of Surgery Divisions of Adult Cardiothoracic and Vascular and Endovascular
More informationMechanical Bleeding Complications During Heart Surgery
Mechanical Bleeding Complications During Heart Surgery Arthur C. Beall, Jr., M.D., Kenneth L. Mattox, M.D., Mary Martin, R.N., C.C.P., Bonnie Cromack, C.C.P., and Gary Cornelius, C.C.P. * Potential for
More informationIncidence of prosthesis-patient mismatch in patients receiving mitral Biocor porcine prosthetic valves
INTERVENTION/VALVULAR HEART DISEASE ORIGINAL ARTICLE Cardiology Journal 2016, Vol. 23, No. 2, 178 183 DOI: 10.5603/CJ.a2016.0011 Copyright 2016 Via Medica ISSN 1897 5593 Incidence of prosthesis-patient
More informationIntensity of oral anticoagulation after implantation of St. Jude Medical mitral or multiple valve replacement: lessons learned from GELIA (GELIA 5)
European Heart Journal Supplements () 3 (Supplement Q), Q39 Q43 Intensity of oral anticoagulation after implantation of St. Jude Medical mitral or multiple valve replacement: lessons learned from GELIA
More informationIndication, Timing, Assessment and Update on TAVI
Indication, Timing, Assessment and Update on TAVI Swedish Heart and Vascular Institute Ming Zhang MD PhD Interventional Cardiology Structure Heart Disease Conflict of Interest None Starr- Edwards Mechanical
More informationW e have previously reported the results of a randomised
715 CARDIOVASCULAR MEDICINE Twenty year comparison of a mechanical heart valve with porcine bioprostheses H Oxenham, P Bloomfield, D J Wheatley, R J Lee, J Cunningham, R J Prescott, H C Miller... See end
More informationCIPG Transcatheter Aortic Valve Replacement- When Is Less, More?
CIPG 2013 Transcatheter Aortic Valve Replacement- When Is Less, More? James D. Rossen, M.D. Professor of Medicine and Neurosurgery Director, Cardiac Catheterization Laboratory and Interventional Cardiology
More informationProf. Patrizio LANCELLOTTI, MD, PhD Heart Valve Clinic, University of Liège, CHU Sart Tilman, Liège, BELGIUM
The Patient with Aortic Stenosis and Mitral Regurgitation Prof. Patrizio LANCELLOTTI, MD, PhD Heart Valve Clinic, University of Liège, CHU Sart Tilman, Liège, BELGIUM Aortic Stenosis + Mitral Regurgitation?
More informationOn October 3, 1977, the first St. Jude Medical (SJM)
The St. Jude Medical Cardiac Valve Prosthesis: A 25-Year Experience With Single Valve Replacement Robert W. Emery, MD, Christopher C. Krogh, Kit V. Arom, MD, PhD, Ann M. Emery, RN, Kathy Benyo-Albrecht,
More informationA 20-year experience of 1712 patients with the Biocor porcine bioprosthesis
Acquired Cardiovascular Disease Mykén and Bech-Hansen A 2-year experience of 1712 patients with the Biocor porcine bioprosthesis Pia S. U. Mykén, MD, PhD, a and Odd Bech-Hansen, MD, PhD b Objective: The
More informationExpanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated?
Expanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated? RM Suri, V Sharma, JA Dearani, HM Burkhart, RC Daly, LD Joyce, HV Schaff Division of Cardiovascular Surgery, Mayo Clinic, Rochester,
More informationIschemic mitral valve reconstruction and replacement: Comparison of long-term survival and complications
Surgery for Acquired Cardiovascular Disease Ischemic mitral valve reconstruction and replacement: Comparison of long-term survival and complications Eugene A. Grossi, MD Judith D. Goldberg, ScD Angelo
More informationKinsing Ko, Thom de Kroon, Najim Kaoui, Bart van Putte, Nabil Saouti. St. Antonius Hospital, Nieuwegein, The Netherlands
Minimal Invasive Mitral Valve Surgery After Previous Sternotomy Without Aortic Clamping: Short- and Long Term Results of a Single Surgeon Single Institution Kinsing Ko, Thom de Kroon, Najim Kaoui, Bart
More informationEchocardiographic Evaluation of Mitral Valve Prostheses
Echocardiographic Evaluation of Mitral Valve Prostheses Dennis A. Tighe, M.D., FACC, FACP, FASE Cardiovascular Medicine University of Massachusetts Medical School Worcester, MA www.asecho.org 1 Nishimura
More informationLONG-TERM RESULTS OF HEART VALVE REPLACEMENT WITH THE EDWARDS DUROMEDICS BILEAFLET PROSTHESIS: A PROSPECTIVE TEN-YEAR CLINICAL FOLLOW-UP
LONG-TERM RESULTS OF HEART VALVE REPLACEMENT WITH THE EDWARDS DUROMEDICS BILEAFLET PROSTHESIS: A PROSPECTIVE TEN-YEAR CLINICAL FOLLOW-UP Bruno K. Podesser, MD a Gudrun Khuenl-Brady, MD a Ernst Eigenbauer,
More informationHeart valve replacement with the Bjork-Shiley and St Jude Medical prostheses: A randomized comparison in 178 patients
European Heart Journal (1990) 11, 583-591 Heart valve replacement with the Bjork-Shiley and St Jude Medical prostheses: A randomized comparison in 178 patients S. VOGT, A. HOFFMANN, J. ROTH, P. DUBACH,
More informationPROSTHETIC VALVE BOARD REVIEW
PROSTHETIC VALVE BOARD REVIEW The correct answer D This two chamber view shows a porcine mitral prosthesis with the typical appearance of the struts although the leaflets are not well seen. The valve
More informationClinical material and methods. Copyright by ICR Publishers 2007
16847_JHVD_Biancari_3197_(116-121)_r1:Layout 1 21/3/07 17:07 Page 116 Predicting Immediate and Late Outcome after Surgery for Mitral Valve Regurgitation with EuroSCORE Jouni Heikkinen, Fausto Biancari,
More informationMitral valve infective endocarditis (IE) is the most
Mitral Valve Replacement for Infective Endocarditis With Annular Abscess: Annular Reconstruction Gregory J. Bittle, MD, Murtaza Y. Dawood, MD, and James S. Gammie, MD Mitral valve infective endocarditis
More informationReplacement of the mitral valve in the presence of
Mitral Valve Replacement in Patients with Mitral Annulus Abscess Christopher M. Feindel Replacement of the mitral valve in the presence of an abscess of the mitral annulus presents a major challenge to
More informationManagement of Difficult Aortic Root, Old and New solutions
Management of Difficult Aortic Root, Old and New solutions Hani K. Najm MD, Msc, FRCSC,, FACC, FESC Chairman, Pediatric and Congenital Heart Surgery Cleveland Clinic Conflict of Interest None Difficult
More informationSurgery for Acquired Cardiovascular Disease
Performance of bioprostheses and mechanical prostheses assessed by composites of valve-related complications to 15 years after mitral valve replacement W. R. E. Jamieson, MD, O. von Lipinski, MD, R. T.
More informationThe clinical experience reported in recent Western series has provided
Surgery for Acquired Cardiovascular Disease Yu et al Long-term evaluation of Carpentier-Edwards porcine bioprosthesis for rheumatic heart disease Hsi-Yu Yu, MD a Yi-Lwun Ho, MD b Shu-Hsun Chu, MD c Yih-Sharng
More informationLong term outcomes of posterior leaflet folding valvuloplasty for mitral valve regurgitation
Featured Article Long term outcomes of posterior leaflet folding valvuloplasty for mitral valve regurgitation Igor Gosev 1, Maroun Yammine 1, Marzia Leacche 1, Siobhan McGurk 1, Vladimir Ivkovic 1, Michael
More informationControversy exists regarding which valve type is best
Treatment of Endocarditis With Valve Replacement: The Question of Tissue Versus Mechanical Prosthesis Marc R. Moon, MD, D. Craig Miller, MD, Kathleen A. Moore, BS, Phillip E. Oyer, MD, PhD, R. Scott Mitchell,
More informationExperience with 500 Stentless Aortic Valve Replacements
Experience with 500 Stentless Aortic Valve Replacements Dimitrios C. Iliopoulos, MD Cardiac Surgeon Ass. Professor of Surgery University of Athens, School of Medicine I declare no conflict of interest
More informationEarly and One-year Outcomes of Aortic Root Surgery in Marfan Syndrome Patients
Early and One-year Outcomes of Aortic Root Surgery in Marfan Syndrome Patients A Prospective, Multi-Center, Comparative Study Joseph S. Coselli, Irina V. Volguina, Scott A. LeMaire, Thoralf M. Sundt, Elizabeth
More informationIntra-operative Echocardiography: When to Go Back on Pump
Intra-operative Echocardiography: When to Go Back on Pump GREGORIO G. ROGELIO, MD., F.P.C.C. OUTLINE A. Indications for Intraoperative Echocardiography B. Role of Intraoperative Echocardiography C. Criteria
More informationAutologous Pulmonary Valve Replacement of the Diseased Aortic Valve
Autologous Pulmonary Valve Replacement of the Diseased Aortic Valve By L. GONZALEZ-LAvIN, M.D., M. GEENS. M.D., J. SOMERVILLE, M.D., M.R.C.P., ANm D. N. Ross, M.B., CH.B., F.R.C.S. SUMMARY Living tissue
More informationThe Edge-to-Edge Technique f For Barlow's Disease
The Edge-to-Edge Technique f For Barlow's Disease Ottavio Alfieri, Michele De Bonis, Elisabetta Lapenna, Francesco Maisano, Lucia Torracca, Giovanni La Canna. Department of Cardiac Surgery, San Raffaele
More informationCase. 15-year-old boy with bicuspid AV Severe AR with moderate AS. Ross vs. AVR (or AVP)
Case 15-year-old boy with bicuspid AV Severe AR with moderate AS Ross vs. AVR (or AVP) AMC case 14-year-old boy with bicuspid AV Severe AS with mild AR Body size Bwt: 55 kg, Ht: 154 cm, BSA: 1.53 m 2 Echocardiography
More informationThe stentless bioprosthesis has many salient features that
Aortic Valve Replacement with the Medtronic Freestyle Xenograft Using the Subcoronary Implantation Technique D. Michael Deeb, MD The stentless bioprosthesis has many salient features that make it an attractive
More informationThe St. Jude Medical mechanical valve is a low-profile,
Twenty-Five Year Experience With the St. Jude Medical Mechanical Valve Prosthesis J. Matthew Toole, MD, Martha R. Stroud, MS, John M. Kratz, MD, Arthur J. Crumbley III, MD, Scott M. Bradley, MD, Fred A.
More informationA Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision
A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision Prof. Pino Fundarò, MD Niguarda Hospital Milan, Italy Introduction
More informationReoperations after primary aortic valve replacement
Third-Time Aortic Valve Replacement: Patient s and Operative Outcome Kasra Shaikhrezai, MD, MRCS, Giordano Tasca, MD, FETCS, Mohamed Amrani, PhD, FETCS, Gilles Dreyfus, MD, FETCS, and George Asimakopoulos,
More informationAnticoagulation Therapy and Valve Surgery. Dr Pau Kiew Kong Consultant Cardiothoracic Surgeon
Anticoagulation Therapy and Valve Surgery Dr Pau Kiew Kong Consultant Cardiothoracic Surgeon Outline of lecture 1. Type of Valve Surgery 2. Anticoagulation requirements 3. Mechanical (Metallic) prosthetic
More informationValvular heart disease (VHD) is present in 2.5% of the
2017 Focused Update for Management of Patients With Valvular Heart Disease: Summary of New Recommendations Richard Matiasz, MD; Vera H. Rigolin, MD Valvular heart disease (VHD) is present in 2.5% of the
More informationAspirin or Coumadin as the Drug of Choice
Aspirin or Coumadin as the Drug of Choice for Valve Replacement with Porcine Bioprosthesis L. Nufiez, M.D., M. Gil Aguado, M.D., D. Celemin, M.D., A. Iglesias, M.D., and J. L. Larrea, M.D. ABSTRACT Eight
More information(Ann Thorac Surg 2008;85:845 53)
I Made Adi Parmana The utility of intraoperative TEE has become increasingly more evident as anesthesiologists, cardiologists, and surgeons continue to appreciate its potential application as an invaluable
More informationQuality Outcomes Mitral Valve Repair
Quality Outcomes Mitral Valve Repair Moving Beyond Reoperation Rakesh M. Suri, D.Phil. Professor of Surgery 2015 MFMER 3431548-1 Disclosure Mayo Clinic Division of Cardiovascular Surgery Research funding
More informationEmergency Intraoperative Echocardiography
Emergency Intraoperative Echocardiography Justiaan Swanevelder Department of Anaesthesia, Glenfield Hospital University Hospitals of Leicester NHS Trust, UK Carl Gustav Jung (1875-1961) Your vision will
More informationDepartment of Cardiac Surgery, Trousseau University Hospital, Tours, France
Risk Factors for Valve-Related Complications after Mechanical Heart Valve Replacement in 505 Patients with Long-Term Follow Up Thierry Bourguignon, Eric Bergöend, Alain Mirza, Grégoire Ayegnon, Paul Neville,
More informationRepair or Replacement
Surgical intervention post MitraClip Device: Repair or Replacement Saudi Heart Association, February 21-24 Rüdiger Lange, MD, PhD Nicolo Piazza, MD, FRCPC, FESC German Heart Center, Munich, Germany Division
More informationDurability of Pericardial Versus Porcine Aortic Valves
Journal of the American College of Cardiology Vol. 44, No. 2, 2004 2004 by the American College of Cardiology Foundation ISSN 0735-1097/04/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2004.01.053
More informationIt is controversial whether to perform aortic valve replacement
Management of Patients With Mild Aortic Stenosis Undergoing Coronary Artery Bypass Grafting James W. Tam, MD, Roy G. Masters, MD, Ian G. Burwash, MD, Alain D. Mayhew, BS, and Kwan-Leung Chan, MD University
More informationThe increase in the lifespan of the western population
Outcome After Aortic Valve Replacement in Octogenarians Bruno Chiappini, MD, Nicola Camurri, MD, Antonio Loforte, MD, Luca Di Marco, MD, Roberto Di Bartolomeo, MD, and Giuseppe Marinelli, MD Department
More informationAdult Cardiac Surgery
Adult Cardiac Surgery Mahmoud ABU-ABEELEH Associate Professor Department of Surgery Division of Cardiothoracic Surgery School of Medicine University Of Jordan Adult Cardiac Surgery: Ischemic Heart Disease
More informationAortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients
Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients Augusto D Onofrio, MD, Stefano Auriemma, MD, Paolo Magagna, MD, Alessandro Favaro, MD,
More informationValve Disease. Valve Surgery. Total Volume. In 2016, Cleveland Clinic surgeons performed 3039 valve surgeries.
Valve Surgery Total Volume 1 1 Volume 35 3 5 15 1 5 1 13 1 N = 773 5 79 15 93 1 339 In 1, surgeons performed 339 valve surgeries. surgeons have implanted more than 1, bioprosthetic aortic valves since
More informationPrimary Tissue Valve Degeneration in Glutaraldehvde-Preserved Porcine Biomostheses: Hancock I Vekus Carpentier-Edwards at 4- to 7-Years Follow-up
Primary Tissue Valve Degeneration in Glutaraldehvde-Preserved Porcine Biomostheses: A Hancock I Vekus Edwards at 4- to 7-Years Follow-up Francisco Nistal, M.D., Edurne Artifiano, M.D., and Ignacio Gallo,
More informationEchocardiographic Evaluation of Aortic Valve Prosthesis
Echocardiographic Evaluation of Aortic Valve Prosthesis Amr E Abbas, MD, FACC, FASE, FSCAI, FSVM, RPVI Co-Director, Echocardiography, Director, Interventional Cardiology Research, Beaumont Health System
More informationVALVE REPAIR VERSUS REPLACEMENT FOR MITRAL INSUFFICIENCY: WHEN IS A MECHANICAL VALVE STILL INDICATED?
VALVE REPAIR VERSUS REPLACEMENT FOR MITRAL INSUFFICIENCY: WHEN IS A MECHANICAL VALVE STILL INDICATED? Eugene A. Grossi, MD Aubrey C. Galloway, MD Jeffrey S. Miller, MD Greg H. Ribakove, MD Alfred T. Culliford,
More informationAtrial fibrillation (AF) is associated with increased morbidity
Ablation of Atrial Fibrillation with Concomitant Surgery Edward G. Soltesz, MD, MPH, and A. Marc Gillinov, MD Atrial fibrillation (AF) is associated with increased morbidity and mortality in coronary artery
More informationPATIENT BOOKLET MEDTRONIC MITRAL AND TRICUSPID HEART VALVE REPAIR
PATIENT BOOKLET MEDTRONIC MITRAL AND TRICUSPID HEART VALVE REPAIR ARE MEDTRONIC HEART VALVE REPAIR THERAPIES RIGHT FOR YOU? Prosthetic (artificial) heart valve repair products are used by physicians to
More informationMechanical Tricuspid Valve Replacement Is Not Superior in Patients Younger Than 65 Years Who Need Long-Term Anticoagulation
Mechanical Tricuspid Valve Replacement Is Not Superior in Patients Younger Than 65 Years Who Need Long-Term Anticoagulation Ho Young Hwang, MD, PhD, Kyung-Hwan Kim, MD, PhD, Ki-Bong Kim, MD, PhD, and Hyuk
More information