Overview. Proton Therapy in lung cancer 8/3/2016 IMPLEMENTATION OF PBS PROTON THERAPY TREATMENT FOR FREE BREATHING LUNG CANCER PATIENTS
|
|
- Osborn Strickland
- 5 years ago
- Views:
Transcription
1 IMPLEMENTATION OF PBS PROTON THERAPY TREATMENT FOR FREE BREATHING LUNG CANCER PATIENTS Heng Li, PhD Assistant Professor, Department of Radiation Physics, UT MD Anderson Cancer Center, Houston, TX, 773 8/3/15 Overview Background Understand the motion-induced dose uncertainty in IMPT Implementation of IMPT for selected lung patients Conclusion Proton Therapy in lung cancer Potential to improves therapeutic ratio and allows dose escalation/acceleration By limiting dose to normal tissue Tumor Normal tissue 1
2 Rationale IMPT: Reduce normal tissue dose compared with IMRT in stage IIIB NSCLC (Zhang et al IJROBP 21) IMRT IMPT Motion Induced Dose Uncertainties The motion of the beam could interfere with the motion of target (interplay effect) May result in distortion of the planned dose distribution, local over- and under- dosage One of the major concerns for treating lung cancer with scanning beam proton Understanding Motion Induced Dose Uncertainties Determine the relationship between the delivered dose and the plan dose Statistical quantification of the dose uncertainty with consideration of Fractionation Breathing period Repainting Total delivery time 2
3 Repainting Zenklusen et al. PMB 21 Iso-layer repainting (within each energy) Not necessary helpful repainting could complete within a short time relative to breathing cycle Volumetric repainting (visit all energies, then repeat) To simulate passive scattering beam delivery The total irradiation time would increase considerably Energy change needs to be fast typically ~ 1 to 2 sec; PSI ~ 8 ms E n. E 2 E 1 Proton Beam Target volume Require large number of repainting Scanning motion and target motion are uncorrelated. Motion Induced Dose Uncertainties - Scanning Beam Proton Mean dose difference between 4D dose and delivered dose as a function of number of deliveries (fractions x repainting) 1 1/ =.3 s.8 Mean of D/ D max Number of Deliveries (k) Li et al. Med Phys 212 Delivery Sequence Optimization Minimize the motion-induced dose uncertainty by optimizing delivery sequence Evaluate the efficacy of spot delivery sequence by measurements and compare measurements to analytical model Patient study to validate the technique Reducing Dose Uncertainty for Spot-Scanning Proton Beam Therapy of Moving Tumors by Optimizing the Spot Delivery Sequence Li et. al. IJROBP 215 3
4 T 8/3/216 Delivery Sequence for SSPT Nominal Plan T T5 Nominal Spot Pattern Nominal Target Target at the time of delivery Delivered Spots Optimization of Delivery Sequence Nominal Plan T T5 No rescanning Same delivery time Nominal Spot Pattern Nominal Target Target at the time of delivery Delivered Spots 4D Dynamic Dose T1 T9 Images and deform vectors IMPT treatment plan 4D Dynamic Dose Simulator Y. Li et. al. Med Phys 214 Dose Distribution Breathing Pattern 4
5 Patient Study Optimization of Delivery Sequence Regular sequence (RS) Worst sequence (WS) Optimized sequence (OS) Max dose error in 1 fx 23.4% for WS 14.3% for RS 7.3% for OS Patient Study Li et. al. IJROBP 215 Understanding Motion Induced Dose Uncertainties The delivered dose converges to the 4D dose The dose difference between the delivered dose (4D Dynamic dose) and the planned dose (4D dose) reduces with Fractionation Breathing period Repainting Total delivery time 5
6 Distance [mm] Y[mm] Y[mm] Y[mm] Percent Volume [%] Percent Volume [%] 8/3/216 Summary of Current Techniques (FB) Margin based approach Needed but likely not sufficient Motion assessment 4DCT and Water equivalent thickness (WET) based analysis Optimization based techniques 4D optimization and dynamic dose analysis Robust optimization and analysis Delivery based techniques Scanning direction Rescanning Delivery sequence optimization Patient specific dose evaluation 4D Dynamic dose calculation Repeated CT and adaptive planning Clinical Implementation of Intensity Modulated Proton Therapy for Thoracic Malignancies Chang et. al. IJROBP 214 WET Analysis Example a) Reference b) Target X[mm] X[mm] c) Difference 5 12 d) Histogram X[mm] WET Difference [mm] WET Analysis Example 7 (a) Change in WET (b) Fraction of Surface with change in WET less than 5mm Mean WET change S 5mm passing rate Beam Angles [Degrees] Beam Angles [Degrees] 6
7 D95% (Gy) for ICTV, based CT/5 dose 8/3/216 WET change and Motion Induced Dose Uncertainties WETave (mm) Yu et. al. Med Phys 216 Robust Optimization 5 different plans were needed to decide the best treatment plan for a lung patient: IMRT, PSPT, SFO, MFO and robust MFO. Robust Optimization Original treatment plan for a patient (left, MFO- PTV; right, MFO-RO) Robust optimization in intensity-modulated proton therapy to account for anatomy changes in lung cancer patients Li et. al. Radiotherapy and Oncology 215 7
8 Robust Optimization Dose calculated on verification CT for a patient (left, MFO-PTV; right, MFO-RO) 4D Robust Optimization Interplay effect could be more significant for scanning beam 4D accumulated dose predicts the dynamic dose to patient after multiple fractions and/or repainting and deviates from planned (static) dose generated on single phase 4D robust optimization technique could be adopted to produce a deliverable plan with reduced interplay effect Liu et. al. IJROBP 216 Conclusion Accurate dose calculation is challenging for treating moving targets with proton A (not perfect) system was developed and running Continue developing and improving on the scanning beam proton treatment of moving targets 8
9 Acknowledgments The entire team of Radiation Oncology: Radiation Oncologists Physicists Dosimetrists Physics assistant Therapists Others 9
8/3/2016. Implementation of Pencil Beam Scanning (PBS) Proton Therapy Treatment for Liver Patients. Acknowledgement. Overview
Implementation of Pencil Beam Scanning (PBS) Proton Therapy Treatment for Liver Patients Liyong Lin, PhD, Assistant Professor Department of Radiation Oncology University of Pennsylvania Acknowledgement
More information8/1/2016. Motion Management for Proton Lung SBRT. Outline. Protons and motion. Protons and Motion. Proton lung SBRT Future directions
Motion Management for Proton Lung SBRT AAPM 2016 Outline Protons and Motion Dosimetric effects Remedies and mitigation techniques Proton lung SBRT Future directions Protons and motion Dosimetric perturbation
More informationEvaluation of the systematic error in using 3D dose calculation in scanning beam proton therapy for lung cancer
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, OLUME 15, NUMBER 5, 2014 Evaluation of the systematic error in using 3D dose calculation in scanning beam proton therapy for lung cancer Heng Li, 1a Wei Liu,
More informationSpecifics of treatment planning for active scanning and IMPT
Specifics of treatment planning for active scanning and IMPT SFUD IMPT Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Treatment planning for scanning 1. Single Field,
More informationClinical Aspects of Proton Therapy in Lung Cancer. Joe Y. Chang, MD, PhD Associate Professor
Clinical Aspects of Proton Therapy in Lung Cancer Joe Y. Chang, MD, PhD Associate Professor Clinical Service Chief Thoracic Radiation Oncology Lung Cancer Basic Factors No. 1 cancer killer 161,840 patients
More informationState-of-the-art proton therapy: The physicist s perspective
State-of-the-art proton therapy: Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Overview of presentation 1. State-of-the-art proton delivery 2. Current challenges 3. New
More informationRobust Optimization accounting for Uncertainties
Robust Optimization accounting for Uncertainties Thomas Bortfeld Massachusetts General Hospital, Boston tbortfeld@mgh.harvard.edu Supported by: RaySearch Philips Medical Systems Outline 1. Optimality and
More informationConsiderations when treating lung cancer with passive scatter or active scanning proton therapy
Mini-Review Considerations when treating lung cancer with passive scatter or active scanning proton therapy Sara St. James, Clemens Grassberger, Hsiao-Ming Lu Department of Radiation Oncology, Massachusetts
More informationProton Beam Therapy at Mayo Clinic
Proton Beam Therapy at Mayo Clinic Jon J. Kruse, Ph.D. Mayo Clinic Dept. of Radiation Oncology Rochester, MN History of Proton Therapy at Mayo 2002: Decided to consider particle therapy analysis and education
More informationClinical Education A comprehensive and specific training program. carry out effective treatments from day one
Proton Therapy Clinical Education A comprehensive and specific training program carry out effective treatments from day one Forewarned is forearmed Although over 100,000 patients have been treated in proton
More informationProton Stereotactic Radiotherapy: Clinical Overview. Brian Winey, Ph.D. Physicist, MGH Assistant Professor, HMS
Proton Stereotactic Radiotherapy: Clinical Overview Brian Winey, Ph.D. Physicist, MGH Assistant Professor, HMS Acknowledgements Radiation Oncologists and Physicists at various institutions (MGH, MDACC,
More informationRange Uncertainties in Proton Therapy
Range Uncertainties in Proton Therapy Harald Paganetti PhD Professor of Radiation Oncology, Harvard Medical School Director of Physics Research, Massachusetts General Hospital, Department of Radiation
More informationFuture upcoming technologies and what audit needs to address
Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple
More informationVariable Dose Rate Dynamic Conformal Arc Therapy (DCAT) for SABR Lung: From static fields to dynamic arcs using Monaco 5.10
Variable Dose Rate Dynamic Conformal Arc Therapy (DCAT) for SABR Lung: From static fields to dynamic arcs using Monaco 5.10 Simon Goodall Radiation Oncology Physicist Genesis Care Western Australia Introduction
More informationAdvances in external beam radiotherapy
International Conference on Modern Radiotherapy: Advances and Challenges in Radiation Protection of Patients Advances in external beam radiotherapy New techniques, new benefits and new risks Michael Brada
More informationDosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer
Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer Tan Chek Wee 15 06 2016 National University Cancer Institute, Singapore Clinical Care Education Research
More informationSIMPLE IMPLEMENTATION OF A DIBH TANGENTIAL IMRT TECHNIQUE FOR LEFT-SIDED BREAST CANCER
SIMPLE IMPLEMENTATION OF A DIBH TANGENTIAL IMRT TECHNIQUE FOR LEFT-SIDED BREAST CANCER Dominique Mathieu MD MSc, Nicolas Côté MSc, Andrée-Anne Bernard MD, Noémie Lahaie RTT, Stéphane Bedwani PhD, Jean-François
More informationIMRT - the physician s eye-view. Cinzia Iotti Department of Radiation Oncology S.Maria Nuova Hospital Reggio Emilia
IMRT - the physician s eye-view Cinzia Iotti Department of Radiation Oncology S.Maria Nuova Hospital Reggio Emilia The goals of cancer therapy Local control Survival Functional status Quality of life Causes
More informationDosimetry for Epidemiology Cohorts Who Receive Radiation Therapy
Dosimetry for Epidemiology Cohorts Who Receive Radiation Therapy Wayne Newhauser, PhD Eurados Winter School, Milan, 2016 Introduction About 1 in 2 men and women born today will be diagnosed with some form
More informationDiscuss the general planning concepts used in proton planning. Review the unique handling of CTV / ITV / PTV when treating with protons
Mark Pankuch, PhD Discuss the general planning concepts used in proton planning Review the unique handling of CTV / ITV / PTV when treating with protons Pencil Beam distributions and PBS optimization Cover
More informationClinical Protontherapy Head and Neck cancer
Clinical Protontherapy Head and Neck cancer Coen Rasch Department of Radiation Oncology Academic Medical Center Amsterdam No disclosures 1 Introduction Proton vsphotonin the literature Model basedindicationprotocol
More informationCharacterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning
Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Supervisors Prof. V. Patera PhD R. Van Roermund Candidate Annalisa Patriarca
More informationWho Should Know Radiation Oncology Coding?
Why Should We Learn Radiation Oncology Coding? Terry Wu, Ph.D. Chief Physicist Radiation Oncology Department Willis-Knighton Cancer Center Who Should Know Radiation Oncology Coding? Radiation Oncologist
More informationImage Guided Proton Therapy and Treatment Adaptation
Image Guided Proton Therapy and Treatment Adaptation www.hollandptc.nl d.r.schaart@tudelft.nl Cancer in The Netherlands About 1 in 3 people get cancer in some stage of their life 86.800 new cancer patients
More informationZhengfei Zhu 1,3, Wei Liu 2, Michael Gillin 2, Daniel R Gomez 1, Ritsuko Komaki 1, James D Cox 1, Radhe Mohan 2 and Joe Y Chang 1*
Zhu et al. Radiation Oncology 2014, 9:108 RESEARCH Open Access Assessing the robustness of passive scattering proton therapy with regard to local recurrence in stage III non-small cell lung cancer: a secondary
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationIGRT Solution for the Living Patient and the Dynamic Treatment Problem
IGRT Solution for the Living Patient and the Dynamic Treatment Problem Lei Dong, Ph.D. Associate Professor Dept. of Radiation Physics University of Texas M. D. Anderson Cancer Center Houston, Texas Learning
More informationTwo Week Course Curriculum Catalog
Welcome to The University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, TX 1840 Old Spanish Trail Houston, TX 77054 Two Week Course Curriculum Catalog 1 Welcome to MD Anderson Cancer
More informationA Dosimetric Comparison of Whole-Lung Treatment Techniques. in the Pediatric Population
A Dosimetric Comparison of Whole-Lung Treatment Techniques in the Pediatric Population Corresponding Author: Christina L. Bosarge, B.S., R.T. (R) (T) Indiana University School of Medicine Department of
More informationSBRT I: Overview of Simulation, Planning, and Delivery
Disclosure SBRT I: Overview of Simulation, Planning, and Delivery I have received research funding from NIH, the Golfers Against Cancer (GAC) foundation, and Philips Health System. Jing Cai, PhD Duke University
More informationOn the use of 4DCT derived composite CT images in treatment planning of SBRT for lung tumors
On the use of 4DCT derived composite CT images in treatment planning of SBRT for lung tumors Zhe (Jay) Chen, Ph.D. Department of Therapeutic Radiology Yale University School of Medicine and Yale-New Haven
More informationTwo Week Course Curriculum October 8-20, 2018
Welcome to The University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, TX 1840 Old Spanish Trail Houston, TX 77054 Two Week Course Curriculum October 8-20, 2018 1 Welcome to MD Anderson
More informationPotential conflicts-of-interest. Respiratory Gated and Four-Dimensional Tumor Tracking Radiotherapy. Educational objectives. Overview.
Respiratory Gated and Four-Dimensional Tumor Tracking Radiotherapy Potential conflicts-of-interest I am PI of a sponsored research agreement between Stanford University and Varian Medical Systems P Keall
More informationEvaluation of Whole-Field and Split-Field Intensity Modulation Radiation Therapy (IMRT) Techniques in Head and Neck Cancer
1 Charles Poole April Case Study April 30, 2012 Evaluation of Whole-Field and Split-Field Intensity Modulation Radiation Therapy (IMRT) Techniques in Head and Neck Cancer Abstract: Introduction: This study
More informationSBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session
Stereotactic Body Radiation Therapy Quality Assurance Educational Session J Perks PhD, UC Davis Medical Center, Sacramento CA SBRT fundamentals Extra-cranial treatments Single or small number (2-5) of
More informationImage Guided Stereotactic Radiotherapy of the Lung
Image Guided Stereotactic Radiotherapy of the Lung Jamie Marie Harris, MS DABR Avera McKennan Radiation Oncology September 25, 2015 Stereotactic Body Radiotherapy - Clinical Dose/Fractionation - Normal
More informationUnderstanding Radiation Therapy. For Patients and the Public
Understanding Radiation Therapy For Patients and the Public Introduction to Radiation Oncology Radiation has been an effective tool for treating cancer for more than 100 years. Radiation oncologists are
More informationTHE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS
THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS ICTP SCHOOL ON MEDICAL PHYSICS FOR RADIATION THERAPY DOSIMETRY AND TREATMENT PLANNING FOR BASIC AND ADVANCED APPLICATIONS March
More informationReirradiazione. La radioterapia stereotassica ablativa: torace. Pierluigi Bonomo Firenze
Reirradiazione La radioterapia stereotassica ablativa: torace Pierluigi Bonomo Firenze Background Stage III NSCLC isolated locoregional recurrence in 25% of pts mostly unresectable; low RR with 2 nd line
More informationRole of radiotherapy in the treatment of lymphoma in Lena Specht MD DMSc Professor of Oncology Rigshospitalet, University of Copenhagen Denmark
Role of radiotherapy in the treatment of lymphoma in 2017 Lena Specht MD DMSc Professor of Oncology Rigshospitalet, University of Copenhagen Denmark Disclosures Member of Advisory Board and Principal Investigator,
More informationChanging Paradigms in Radiotherapy
Changing Paradigms in Radiotherapy Marco van Vulpen, MD, PhD Mouldroomdag-2015 Towards the elimination of invasion 1 NIH opinion on the future of oncology Twenty-five years from now,i hope that we won
More informationPitfalls in SBRT Treatment Planning for a Moving Target
Pitfalls in SBRT Treatment Planning for a Moving Target Cynthia F. Chuang, Ph.D. Department of Radiation Oncology University of California-San Francisco I have no conflicts of interests to disclose In
More informationMedical Errors in Radiation Therapy 2014
Medical Errors in Radiation Therapy 2014 Amy L. Carlson Division of Emergency Preparedness and Community Support Bureau of Radiation Control Florida Department of Health Reportable Medical Events State
More informationPartial Breast Irradiation using adaptive MRgRT
Partial Breast Irradiation using adaptive MRgRT Shyama Tetar, radiation-oncologist VUmc Amsterdam 15-12-2017 5 th Vumc SBRT symposium 2017 Current practice Breast conserving treatment (BCT) Breast conserving
More informationTransition to Heterogeneity Corrections. Why have accurate dose algorithms?
Transition to Heterogeneity Corrections Eric E. Klein, M.S., Washington University, St. Louis, MO Craig Stevens, M.D., Ph.D., MD Anderson Cancer Center, Houston, TX Nikos Papinikolou, Ph.D., University
More informationPhysics Treatment Margins. Laurence Court University of Texas MD Anderson Cancer Center
Physics Treatment Margins Laurence Court University of Texas MD Anderson Cancer Center Disclosure Employer: UT MD Anderson Cancer Center Grants from: NCI, CPRIT, Varian, Elekta, Mobius Learning Objectives
More informationIMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor
IMRT/IGRT Patient Treatment: A Community Hospital Experience Charles M. Able, Assistant Professor Disclosures I have no research support or financial interest to disclose. Learning Objectives 1. Review
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More information8/2/2012. Transitioning from 3D IMRT to 4D IMRT and the Role of Image Guidance. Part II: Thoracic. Peter Balter, Ph.D.
8/2/2012 Transitioning from 3D IMRT to 4D IMRT and the Role of Image Guidance Part II: Thoracic Peter Balter, Ph.D. Disclosure Dr. Balter is Physics PI on a trial comparing Cyberknife based SBRT with surgery,
More informationDoes the IMRT technique allow improvement of treatment plans (e.g. lung sparing) for lung cancer patients with small lung volume: a planning study
Does the IMRT technique allow improvement of treatment plans (e.g. lung sparing) for lung cancer patients with small lung volume: a planning study Received: 22.04.2008 Accepted: 4.07.2008 Subject: original
More informationReview of Workflow NRG (RTOG) 1308: Phase III Randomized Trial Comparing Overall Survival after Photon versus Proton Chemoradiation Therapy for
Review of Workflow NRG (RTOG) 1308: Phase III Randomized Trial Comparing Overall Survival after Photon versus Proton Chemoradiation Therapy for Inoperable Stage II-IIIB NSCLC 1 Co-Chairs Study Chair: Zhongxing
More informationTreatment Efficiency and Optimization of Patient Care with IBA ProteusOne
Treatment Efficiency and Optimization of Patient Care with IBA ProteusOne Terry Wu, Ph.D. Chief Physicist, Radiation Oncology Department Willis-Knighton Cancer Center/Proton Therapy Center Shreveport,
More informationA TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,
More informationDepartment of Radiation Oncology SMART. Stereotactic MR-guided adaptive radiotherapy. Marloes Jeulink Omar Bohoudi
SMART Stereotactic MR-guided adaptive radiotherapy Department of Radiation Oncology Marloes Jeulink Omar Bohoudi First clinical treatment May 4th 2016 First clinical treatment May 4 th 2016 MR-guided RT
More informationWhich Planning CT Should be Used for Lung SBRT? Ping Xia, Ph.D. Head of Medical Physics in Radiation Oncology Cleveland Clinic
Which Planning CT Should be Used for Lung SBRT? Ping Xia, Ph.D. Head of Medical Physics in Radiation Oncology Cleveland Clinic Outline Image quality and image dose Free breathing CT, 4DCT, and synthetic
More informationIN VIVO IMAGING Proton Beam Range Verification With PET/CT
IN VIVO IMAGING Proton Beam Range Verification With PET/CT Antje-Christin Knopf 1/3 K Parodi 2, H Paganetti 1, T Bortfeld 1 Siemens Medical Solutions Supports This Project 1 Department of Radiation Oncology,
More information- In potentially operable patients -
Lung Stereotactic Ablative Radiotherapy (SABR) - In potentially operable patients - Frank Lagerwaard VUMC Amsterdam Stereotactic Ablative Radiotherapy (SABR) 2003-2008 4DCT-based target definition Non-gated
More informationFuture Directions in Prostate Cancer: The Case for Protons. John J. Coen, MD Helen & Harry Gray Cancer Center
Future Directions in Prostate Cancer: The Case for Protons John J. Coen, MD Helen & Harry Gray Cancer Center November 14, 2012 Protons and prostate cancer Early proton experience at the MGH The case for
More informationART for Cervical Cancer: Dosimetry and Technical Aspects
ART for Cervical Cancer: Dosimetry and Technical Aspects D.A. Jaffray, Ph.D. Radiation Therapy Physics Princess Margaret Cancer Centre/Techna/Ontario Cancer Institute Professor Departments of Radiation
More informationCBCT of the patient in the treatment position has gained wider applications for setup verification during radiotherapy.
Gülcihan CÖDEL Introduction The aim of this study is to evaluate the changes in bladder doses during the volumetric modulated arc therapy (VMAT) treatment of prostate cancer patients using weekly cone
More informationCURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT
CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT Purpose The purpose of this curriculum outline is to provide a framework for multidisciplinary training for radiation oncologists, medical
More informationA Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer
A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer Tony Kin Ming Lam Radiation Planner Dr Patricia Lindsay, Radiation Physicist Dr John Kim, Radiation Oncologist Dr Kim Ann Ung,
More informationTecniche Radioterapiche U. Ricardi
Tecniche Radioterapiche U. Ricardi UNIVERSITA DEGLI STUDI DI TORINO Should we always rely on stage? T4N0M0 Stage IIIB T2N3M0 Early stage NSCLC The treatment of choice for early-stage NSCLC is anatomic
More informationIMRT FOR CRANIOSPINAL IRRADIATION: CHALLENGES AND RESULTS. A. Miller, L. Kasulaitytė Institute of Oncolygy, Vilnius University
IMRT FOR CRANIOSPINAL IRRADIATION: CHALLENGES AND RESULTS A. Miller, L. Kasulaitytė Institute of Oncolygy, Vilnius University Content 1.Introduction 2.Methods and materials 3.Results and discussion 4.Conclusion
More informationAcknowledgments. Introduction. Managing the Care of Radiotherapy Patients Implanted with Cardiac Devices. Outline
Managing the Care of Radiotherapy Patients Implanted with Cardiac Devices Acknowledgments Members of AAPM TG203 Joann Prisciandaro Moyed Miften, PhD Dimitris Mihailidis, PhD Professor and Chief Physicist
More informationPatient-Specific QA & QA Process. Sasa Mutic, Ph.D. Washington University School of Medicine
Patient-Specific QA & QA Process Sasa Mutic, Ph.D. Washington University School of Medicine Outline of Presentation QA and QC in RT Safety vs. quality Patient specific QM program Learning Objectives Describe
More informationAshley Pyfferoen, MS, CMD. Gundersen Health Systems La Crosse, WI
Ashley Pyfferoen, MS, CMD Gundersen Health Systems La Crosse, WI 3 Radiation Oncologists 3 Physicists 2 Dosimetrists 9 Radiation Therapists o o o o o o o o o Brachial Plexus Anatomy Brachial Plexopathy
More informationMartin Stuschke 1,2*, Andreas Kaiser 1,2, Jehad Abu Jawad 1, Christoph Pöttgen 1, Sabine Levegrün 1 and Jonathan Farr 2,3
Stuschke et al. Radiation Oncology 2013, 8:145 SHORT REPORT Open Access Multi-scenario based robust intensity-modulated proton therapy (IMPT) plans can account for set-up errors more effectively in terms
More informationOriginal Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem
Original Article Treatment Planning Evaluation of Sliding Window and Multiple Static Segments Technique in Intensity Modulated Radiotherapy for Different Beam Directions Teyyiba Kanwal, Muhammad Khalid,
More informationRadiation Damage Comparison between Intensity Modulated Radiotherapy (IMRT) and Field-in-field (FIF) Technique In Breast Cancer Treatments
Radiation Damage Comparison between Intensity Modulated Radiotherapy () and Field-in-field (FIF) Technique In Breast Cancer Treatments Huisi Ai 1 and Hualin Zhang 2 1. Department of Radiation Oncology,
More informationThe Physics of Oesophageal Cancer Radiotherapy
The Physics of Oesophageal Cancer Radiotherapy Dr. Philip Wai Radiotherapy Physics Royal Marsden Hospital 1 Contents Brief clinical introduction Imaging and Target definition Dose prescription & patient
More informationUncertainties and Quality Assurance of Localization and Treatment in Lung SBRT Jing Cai, PhD Duke University Medical Center
Uncertainties and Quality Assurance of Localization and Treatment in Lung SBRT Jing Cai, PhD Duke University Medical Center 2013 AAPM 55 th Annual Meeting, Educational Course, Therapy Track, MOC SAM Program
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationIntroduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology
Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?
More informationImprove dosimetric outcome in stage III non-small-cell lung cancer treatment using spot-scanning proton arc (SPArc) therapy
Li et al. Radiation Oncology (2018) 13:35 https://doi.org/10.1186/s13014-018-0981-6 RESEARCH Open Access Improve dosimetric outcome in stage III non-small-cell lung cancer treatment using spot-scanning
More informationProton Therapy for Prostate Cancer. Andrew K. Lee, MD, MPH Director Proton Therapy Center
Proton Therapy for Prostate Cancer Andrew K. Lee, MD, MPH Director Proton Therapy Center Disclosures No relevant financial disclosures This presentation will not discuss off-label or investigational treatments
More informationNew Radiation Treatment Modalities in the Treatment of Lung Cancer
New Radiation Treatment Modalities in the Treatment of Lung Cancer David Perry, M.D. Chief, Radiation Oncology Medical Director, CyberKnife Radiosurgery Center Medstar Franklin Square Medical Center Definitions
More informationIMRT Planning Basics AAMD Student Webinar
IMRT Planning Basics AAMD Student Webinar March 12, 2014 Karen Chin Snyder, MS Senior Associate Physicist Department of Radiation Oncology Disclosures The presenter has received speaker honoraria from
More informationImplementing SBRT Protocols: A NRG CIRO Perspective. Ying Xiao, Ph.D. What is NRG Oncology?
Implementing SBRT Protocols: A NRG CIRO Perspective Ying Xiao, Ph.D. What is NRG Oncology? One of five new NCI-supported National Clinical Trials Network (NCTN) groups. NCTN officially started March 1,
More informationNIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed
More informationIROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015
IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During
More informationIGRT Protocol Design and Informed Margins. Conflict of Interest. Outline 7/7/2017. DJ Vile, PhD. I have no conflict of interest to disclose
IGRT Protocol Design and Informed Margins DJ Vile, PhD Conflict of Interest I have no conflict of interest to disclose Outline Overview and definitions Quantification of motion Influences on margin selection
More informationApplication of Implanted Markers in Proton Therapy. Course Outline. McLaren Proton Therapy Center Karmanos Cancer Institute McLaren - Flint
Application of Implanted Markers in Proton Therapy Sung Yong Park, Ph.D. McLaren Proton Therapy Center Karmanos Cancer Institute McLaren - Flint AAPM 2016, SAM Therapy Educational Course, 2016.08.04. Course
More informationDisclosures. Proton therapy advantages. Why are comparing therapies difficult? Proton Therapy for Low Risk Prostate Cancer
Proton Therapy for Low Risk Prostate Cancer Disclosures No relevant financial disclosures This presentation will not discuss off-label or investigational treatments Andrew K. Lee, MD, MPH Associate Professor
More informationExperimental measurement with an anthropomorphic phantom of the proton dose distribution in the presence of metal implants
E13195: Slightly better series/treatment with PRONE combination with vertical field. - This field is only possible in case the tumor is not too caudal (problem with energy). - It will increase the integral
More informationA dosimetric comparison between volumetric-modulated arc therapy and dynamic conformal arc therapy in SBRT
JBUON 2019; 24(2): 838-843 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE A dosimetric comparison between volumetric-modulated arc therapy and
More informationFeasibility of 4D IMRT Delivery for Hypofractionated High Dose Partial Prostate Treatments
Feasibility of 4D IMRT Delivery for Hypofractionated High Dose Partial Prostate Treatments R.A. Price Jr., Ph.D., J. Li, Ph.D., A. Pollack, M.D., Ph.D.*, L. Jin, Ph.D., E. Horwitz, M.D., M. Buyyounouski,
More informationTreatment Planning (Protons vs. Photons)
Treatment Planning Treatment Planning (Protons vs. Photons) Acquisition of imaging data Delineation of regions of interest Selection of beam directions Dose calculation Optimization of the plan Hounsfield
More informationHampton University Proton Therapy Institute
Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI
More informationPerturbation of water-equivalent thickness as a surrogate for respiratory motion in proton therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Perturbation of water-equivalent thickness as a surrogate for respiratory motion in proton therapy Jason E. Matney, 1a Peter C. Park,
More informationLearning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy
3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy Sofie Ceberg, PhD Medical Physicist and Lund University Lund, Sweden Learning objectives 1: 3D Dosimetry in the Clinic: Background
More informationMotion gating and tracking techniques: overview and recent developments
Motion gating and tracking techniques: overview and recent developments Gig S Mageras, PhD, FAAPM Department of Medical Physics Memorial Sloan Kettering Cancer Center, New York MSK/gsm 15-Jun-2018 1 Disclosure
More informationIMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)
IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)
More informationAnatomical-Based Adaptive RT: It Begins Here!
Anatomical-Based Adaptive RT: It Begins Here! Kristy K Brock, Ph.D., DABR Physicist, Radiation Medicine Program, Princess Margaret Hospital Associate Professor, Depts of Radiation Oncology & Medical Biophysics,
More informationIROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015
IROC Lung Phantom 3D CRT / IMRT Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 The IROC requests that each institution keep the phantom for no more than 2 weeks. During
More informationThe Evolution of RT Techniques for Gynaecological Cancers in a developing country context
The Evolution of RT Techniques for Gynaecological Cancers in a developing country context Hannah Simonds Stellenbosch University/ Tygerberg Academic Hospital ESMO Africa 2017 I have no disclosures External
More informationThe Evolution of SBRT and Hypofractionation in Thoracic Radiation Oncology
The Evolution of SBRT and Hypofractionation in Thoracic Radiation Oncology (specifically, lung cancer) 2/10/18 Jeffrey Kittel, MD Radiation Oncology, Aurora St. Luke s Medical Center Outline The history
More informationImage Fusion, Contouring, and Margins in SRS
Image Fusion, Contouring, and Margins in SRS Sarah Geneser, Ph.D. Department of Radiation Oncology University of California, San Francisco Overview Review SRS uncertainties due to: image registration contouring
More informationPEDIATRIC ORBITAL TUMORS RADIOTHERAPY PLANNING
PEDIATRIC ORBITAL TUMORS RADIOTHERAPY PLANNING ANATOMY ANATOMY CONT ANATOMY CONT. ANATOMY CONT. EYE OF A CHILD Normal tissue tolerance doses (in conventional #) TD 5/5 TD 50/5 Endpoint Gy Gy Optic nerve
More informationTreatment Planning for Lung. Kristi Hendrickson, PhD, DABR University of Washington Dept. of Radiation Oncology
Treatment Planning for Lung Kristi Hendrickson, PhD, DABR University of Washington Dept. of Radiation Oncology Outline of Presentation Dosimetric planning strategies for SBRT lung Delivery techniques Examples
More information