1 REF TV7-100/2 FRT CMV Real-TM Quant Real Time Kit for use with SmartCycler (Cepheid), RotorGene 3000/6000 (Corbett Research), Applied Biosystems 7300/7500 Real Time PCR Systems (Applera), iq icycler and iq5 (Biorad), MX3000P and MX3005P (Stratagene), Key to symbols used REF List Number Store at 2-8 C/-20 C RUO For Research Use Only Caution! LOT Lot Number VER Version Expiration Date Contains reagents Consult instructions for use Manufacturer NAME CMV Real-TM Quant INTENDED USE Kit CMV Real-TM Quant is a Real-Time test for the Quantitative detection of Cytomegalovirus in human plasma and simultaneous detection of a CMV-specific Internal Control (IC), by dual color detection. PRINCIPLE OF ASSAY kit CMV Real-TM Quant is a Real-Time test for the Quantitative detection of Cytomegalovirus in human plasma. CMV DNA is extracted from plasma, amplified using real time amplification and detected through the use of fluorescent reporter dye probes specific for CMV or CMV IC. CMV Rec IC is a quantitative Internal Control (concentration reported in Data Card) and represents recombinant DNAcontaining-structure which carried through all steps of analysis from nucleic acid extraction to PCR amplification-detection. The presence of quantitative CMV Rec IC allows not only to monitor the extraction procedure and to check possible PCR inhibition but also to verify possible losses of the DNA during extraction procedure thus enabling to calculate precisely the viral load. IC is detected in a channel other than the CMV DNA. Monitoring the fluorescence intensities during Real Time allows the detection and quantification of the accumulating product without having to reopen the reaction tube after the real time amplification. The kit will allow the quantification of 100 samples, including all the necessary reagents to generate the external standard curve of CMV and IC. To generate CMV and IC standard curve for quantification of the amplification products all calibrators should be used and defined as standards with specific concentrations.
2 MATERIALS PROVIDED Part N 1 DNA-Sorb-B : isolation of DNA from clinical specimens; Part N 2 Controls Part N 3 CMV Real-TM Quant : Real Time amplification; Part N 1 DNA-Sorb-B : Lysis Solution, 2 x 15 ml; Washing Solution 1, 2 x 15 ml; Washing Solution 2, 2 x 50 ml; Sorbent, 2 x 1,25 ml; DNA-eluent, 2 x 5 ml. Contains reagents for 100 extractions Part N 2 Controls CMV Rec Pos1 Control, 4 x 0,04 ml; CMV Rec Pos2 Control, 4 x 0,04 ml; CMV Rec IC (Internal Control), 2 x 1,0 ml. Negative Control, 2 x 1,2 ml; Part N 3 CMV Real-TM Quant : PCR-mix-1-TM, 2 x 0,6 ml. PCR-mix-2-TM, 2 x 0,3 ml. TaqF Polymerase, 2 x 0,03 ml. TE-buffer, 2 x 0,5 ml. Quantitation Standard CMV (QS1 CMV/IC, QS2 CMV/IC, QS3 CMV/IC), 2 x 0,1 ml; 1 Standards and controls concentrations are specific for every lot. MATERIALS REQUIRED BUT NOT PROVIDED Zone 1: sample preparation: Biological cabinet Desktop microcentrifuge for eppendorf type tubes (RCF max. 16,000 x g); Eppendorf 5415D or equivalent 60 C ± 2 C dry heat block Vortex mixer Pipettors (capacity 5-40 µl; µl; µl) with aerosol barrier 1,5 ml polypropylene sterile tubes (Sarstedt, QSP, Eppendorf) Disposable gloves, powderless Biohazard waste container Refrigerator Freezer Tube racks 70% Ethanol (freshly prepared mixture of reagent grade 96% ethanol and distilled water) Acetone Zone 2: RT and amplification: Real Time Thermal cycler Reaction tubes or PCR plate Workstation Pipettors (capacity 0,5-10 µl; 5-40 µl) with aerosol barrier Tube racks
3 WARNINGS AND PRECAUTIONS 1. Lysis Solution contains guanidine thiocyanate. Guanidine thiocyanate is harmful if inhaled, or comes in contact with skin or if swallowed. Contact with acid releases toxic gas. (Xn; R: 20/21/22-36/37/38; S: 36/37/39). 2. Wear disposable gloves, laboratory coats and eye protection when handling specimens and reagents. Thoroughly wash hands afterward. 3. Do not pipette by mouth. 4. Do not eat, drink, smoke, apply cosmetics, or handle contact lenses in laboratory work areas. 5. Do not use a kit after its expiration date. 6. Do not mix reagents from different kits. 7. Dispose all specimens and unused reagents in accordance with local regulations. 8. Biosafety Level 2 should be used for materials that contain or are suspected of containing infectious agents. 9. Specimens and controls should be prepared in a laminar flow hood. 10. Heparin has been shown to inhibit PCR. Use of heparinized specimens is not recommended. 11. Avoid repeated thawing and freezing of the reagents, this may reduce the sensitivity of the test. 12. Once the reagents have been thawed, vortex and centrifuge briefly the tubes. 13. Prepare quickly the Reaction mix on ice or in the cooling block. 14. Specimens may be infectious. Use Universal Precautions when performing the assay. 15. Clean and disinfect all spills of specimens or reagents using a disinfectant such as 0,5% sodium hypochlorite, or other suitable disinfectant. Follow by wiping down the surface with 70% ethanol. 16. Avoid contact of specimens and reagents with the skin, eyes and mucous membranes. If these solutions come into contact, rinse immediately with water and seek medical advice immediately. 17. Material Safety Data Sheets (MSDS) are available on request. 18. Use of this product should be limited to personnel trained in the techniques of PCR. 19. Workflow in the laboratory must proceed in a uni-directional manner, beginning in the Extraction Area and moving to the Amplification Area. Do not return samples, equipment and reagents in the area where you performed previous step. Personnel should be using proper anti-contamination safeguards when moving between areas. STORAGE INSTRUCTIONS Part N 1 DNA-Sorb-B must be stored at 2-8 C. Part N 2 Controls must be stored at 2-8 C. Part N 3 CMV Real-TM Quant must be stored at -20 C. The CMV Real-TM Quant kit can be shipped at 2-8 C but should be stored at 2-8 C and -20 C immediately on receipt STABILITY CMV Real-TM Quant Test is stable up to the expiration date indicated on the kit label. The product will maintain performance through the control date printed on the label. Exposure to light, heat or humidity may affect the shelf life of some of the kit components and should be avoided. Repeated thawing and freezing of these reagents should be avoided, as this may reduce the sensitivity. Components stored under conditions other than those stated on the labels may not perform properly and may adversely affect the assay results. SAMPLE COLLECTION, STORAGE AND TRANSPORT CMV Real-TM Quant can analyze DNA extracted with DNA-Sorb-B from: Plasma: o EDTA tubes may be used with the CMV Real-TM Quant. Follow sample tube manufacturer s instructions. o Whole blood collected in EDTA should be separated into plasma and cellular components by centrifugation at x g for 20 min within six hours. The isolated plasma has to be transferred into a sterile polypropylene tube. Plasma may be stored at 2-8 C for an additional 3 days. Alternatively, plasma may be stored at -18 C for up to one month or 1 year when stored at -70 C. o Do not freeze whole blood. o Specimens anti-coagulated with heparin are unsuitable for this test. o Thaw frozen specimens at room temperature before using. o Whole blood must be transported at 2-25 C and processed within 6 hours of collection. Plasma may be transported at 2-8 C or frozen. Liquor (CSF) collected in the sterile Eppendorf tube: o Liquor may be stored at 2-8 C for 1 days. Alternatively, may be stored at -18 C for up to one month or 1 year when stored at -70 C Amniotic liquid collected in the sterile Eppendorf tube: o centrifuge 1,0 ml of sample at x g for 10 min. Discard the supernatant and leave about 200 µl of solution. Vortex the the tube and use 100 µl for DNA extraction. Specimens can be stored at +2-8 C for no longer than 12 hours, or freeze at -20 C to -80 C. Transportation of clinical specimens must comply with country, federal, state and local regulations for the transport of etiologic agents.
4 SPECIMEN AND REAGENT PREPARATION 1. Lysis Solution and Washing Solution (in case of their storage at +2-8 C) should be warmed up to 56 C until disappearance of ice crystals. 2. Prepare required quantity of 1.5 ml polypropylene tubes. 3. Add to each tube 10 µl of Internal Control and 300 µl of Lysis Solution. 4. Vortex specimens for 5 sec. 5. Add 100 µl of Samples to the appropriate tube. 6. Prepare Controls as follows: add 100 µl of C (Negative Control) to labeled Cneg. add 90 µl of C (Negative Control) and 10 µl of CMV Rec Pos1 (or CMV Rec Pos2) to the tube labeled Cpos. 7. Vortex the tubes and centrifuge for 7-10 sec. 8. Vortex vigorously Sorbent and add 25 µl to each tube. 9. Vortex for 5-7 sec and incubate all tubes for 10 min at room temperature. 10. Centrifuge all tubes for 1 min at 10000g and using a micropipette with a plugged aerosol barrier tip, carefully remove and discard supernatant from each tube without disturbing the pellet. Change tips between tubes. 11. Add 300 µl of Washing Solution 1 to each tube. Vortex vigorously and centrifuge for 1 min at 10000g and using a micropipette with a plugged aerosol barrier tip, carefully remove and discard supernatant from each tube without disturbing the pellet. Change tips between tubes. 12. Add 500 µl of Washing Solution 2 to each tube. Vortex vigorously and centrifuge for 1 min at 10000g and using a micropipette with a plugged aerosol barrier tip, carefully remove and discard supernatant from each tube without disturbing the pellet. Change tips between tubes. 13. Repeat step Incubate all tubes with open cap for 5 min at 65 C. 16. Resuspend the pellet in 50 µl of DNA-eluent. Incubate for 10 min at 65 C and vortex periodically. 17. Centrifuge the tubes for 2 min at maximum speed ( g). The supernatant contains DNA ready for amplification. The amplification can be performed on the same day of extraction. PCR PROTOCOL: 1. Thaw one set of reagents, vortex and centrifuge briefly the tubes. 2. Prepare reaction tubes 3. Add for each sample in the new sterile tube 10*(N+1) µl of PCR-mix-1-TM, 5,0*(N+1) µl of PCR-mix-2-TM and 0,5*(N+1) µl of TaqF Polymerase. Vortex the tube. 4. Add 15,0 µl of Reaction Mix into each SmartCycler tube. 5. Add 10,0 µl of extracted DNA sample to appropriate tube with Reaction Mix. 6. Prepare for each run 3 standards and 1 Neg Control*: add 10,0 µl of Quantitation Standards CMV (QS1 CMV/IC, QS2 CMV/IC, QS3 CMV/IC) into 3 labeled tubes; add 10,0 µl of TE-buffer to the tube labeled PCR Negative Control; 7. Insert the tubes in the thermalcycler. *The user can import the experiment with Standard Curves to subsequent experiments with clinical samples. The user can feel certain that the reaction efficiency has not varied between the two runs.
5 Program SmartCycler as follows: 1. Select in the main menu Define Protocols and press button New Protocol. Give a name to the protocol and set the following parameters: 2. Choose Save Protocol 3. Click the Creat Run button in the main menu, then button Dye Set and select FCTC Choose Add/Remove Sites and select in the new windows the protocol and sites for analysis. Click OK. 5. Choose Start Run and Give a name to the experiment. 6. Introduce the concentrations of the Quantitative Standards (reported on the CMV Real-TM Quant Data Card) in the columns Fam Std and Cy3 Std in order to generate Standard curve. 7. Fluorescence is observed in Real Time on the Cy3 channel for CMV DNA and FAM channel for Internal Control. Results Analysis 1. Choose in the menu Analysis settings Manual Thresh Setting and the value 30 for the channels Fam and Cy3. 2. In the table of results (Results Table) appear the values of Ct (Threshold cycle) for CMV DNA and IC DNA. 3. The value of R (correlation coefficient) in the windows Standard Cy3 and Standard FAM must be 0,9 (if < 0,9 a retesting of sample is required).
6 Program Rotor-Gene 3000/6000 as follows: Reaction Volume (µl ): 25 Carousel: 36-well Temperature Profile: Denature: 95 C 20 min Cycling 95 C 20 sec 60 C 35 sec 65 C 25 sec Cycle Repeats 45 times 1. Fluorescence is measured at 65 C on FAM (Green) and JOE (Yellow) channels 2. Press button OK twice 3. Select button Calibrate (Gain Optimisation for RG6000) and function Perform calibration (Optimisation) before 1 st Acquisition 4. Select Next and Start run Program position of the tubes in the carousel of the Rotor-Gene 3000/6000 and introduce the concentrations of the Quantitative Standards (reported on the CMV Real-TM Quant Data Card) in order to generate Standard curve. Results Analysis IC amplification analysis (Cycling A.Fam (Green) 1. Press Analysis then select Quantitation Cycling A.Fam (Cycling A.Green ) Show, set 2 2. Turn off the automatic option Threshold. 3. Press buttons Dynamic Tube, Slope Correct 4. Press Quant Settings and set value to 10% 5. Select Threshold: 0,03 6. In the menu window Quantitation Results appear values of IC DNA copy/specimen. 7. The Coefficient correlation value R in Standard Curve window should be 0,9 (if < 0,9 a retesting of all samples is required). CMV amplification analysis (channel Cycling A.Joe) 1. Press Analysis then select Quantitation Cycling A.Joe (Cycling A.Yellow) Show, set 1 2. Turn off the automatic option Threshold. 3. Press buttons Dynamic Tube, Slope Correct 4. Press Quant Settings and set value to 10% 5. Select Threshold: 0,03 6. In the table of results (Quantitation Results) appear the values of Ct (Threshold cycle). 7. In the menu window Quantitation Results column Calculation concentration appear values of CMV DNA copy/specimen. 8. The Coefficient correlation value R in the Standard Curve window should be 0,9 (if < 0,9 a retesting of all samples is required).
7 Programming the 7300/7500 Real-Time PCR System (Applied): 1. Select in the main menu option New and set the data of new document: select in the window Assay the option Absolute Quantitation, in the window Template the option Blank Document. Press OK 2. In the new window in the Tools menu click button Detector Manager. 3. At the low left side of the window click File and select New. Set in the window New detector probes characteristics: a. Detection of CMV DNA: in the lines Name and Description indicate CMV DNA; in the line Reporter Dye Joe and in Quencher Dye None. Select the Color (for example, red). Click button Create Another. b. The window New detector is opened against. Set the following parameters for Internal Control: in the lines Name and Description indicate CMV IC; in the line Reporter Dye Fam and in Quencher Dye None. Select the Color (for example, blue). Click OK. 4. Close the window Detector manager with probes information. 5. Select window Instrument. 6. Activate Thermal profile and set the following amplification program: Stage Profile Reps 1 95 C 15: C 0:20 60 C 0:20 72 C 0: C 0:20 60 C 0:40* C 0:15 5 *Fluorescence detection on the Fam, Joe channels 7. Indicate reaction volume, 25 µl. Program position of the tubes and introduce the concentrations of the Quantitative Standards (reported on the CMV Real-TM Quant Data Card) in order to generate Standard curve.
8 Programming of iq icycler and iq5 (Biorad) 1. Select in the main menu Define Protocols and click Create a new protocol. Set the following parameters: Cycle Repeats Step Dwell Time Set Point : : : : : :30 FAM, HEX : Select Edit Plate for iq5 or View/Save Data for iqicycler to create the plate for samples and controls. Enter the concentrations of the Quantitative Standards (reported on the CMV Real-TM Quant Data Card) in the Standard Quantity box. Use icon Unknown for all samples. 3. Choose the FAM and HEX channels for all samples. 4. For the iq5 instrument enter the reaction volume Sample volume 25 µl. Select caps type Seal Type and tubes type Vessel Type. Save plate setup. 5. Begin the amplification: For the iq icycler instrument activate the key Run with selected plate and choose in the window of the opened menu, the reaction volume 25 µl, select PCR Quantification Melt Curve and Experimental Plate. Click the button Begin Run and save the experiment settings. For the iq5 instrument activate the button Run and choose in the window of the opened menu Collect Well Factors from Experimental Plate. Click the button Begin Run and save the experiment settings. Results Analysis The results are interpreted with the software of iq icycler or iq5 through the presence of crossing of fluorescence curve with the threshold line. Internal Control (IC) is detected on the FAM channel and CMV DNA on the HEX channel Activate the button PCR Quant for the results analysis Activate the button Log View. Put the threshold line (with the left button of the mouse) at such level where curves of fluorescence are linear (see figure)
9 Programing of MX3000P and MX3005P (Stratagene) 1. Open the program, select Quantitative PCR (Multiple Standards) and click OK 2. At the top left of the window choose Plate Setup 3. In the window Well type set Unknown for the samples and Standard to identify calibrators. 4. In the window Collect fluorescence data select for all samples the channels Fam and Joe. 5. Enter the concentrations of the Quantitative Standards (reported on the CMV Real-TM Quant Data Card) in the Standard Quantity box. 6. At the top left of the window select button Thermal Profile Setup 7. Set the following parameters of amplification: 1 95 C 15: C 0:05 60 C 0:20 72 C 0: C 0:05 60 C 0:40* C 0: Fluorescence is measured at 60 C. To do this, set on the Thermal Profile graph the Endpoints marker. 9. Click Run button, enter a name for the experiment and save it. Results Analysis 1. Soon after amplification is over, choose button Analysis at the top left of the window. 2. Choose button Results 3. At the right angle of the window Area to analyze select Amplification plots. 4. Set automatic Threshold fluorescence. 5. In the window Text report appear for each sample the values of Ct and experimental values of copies DNA CMV and DNA IC. 6. Take care that the value of RSq (correlation coefficient) in the window Standard curve is not lower than 0,9 for both channels.
10 RESULTS INTERPRETATION The Internal Control (IC) is detected on the FAM channel and CMV DNA on the Joe/HEX/Cy3 channel. For each control and patient specimen, calculate the concentration of CMV DNA using the following formula: CMV DNA copies/specimen x coefficient* = copies CMV/mL IC DNA copies/specimen *coefficient is specific for each lot and reported in the CMV Real-TM Quant Data Card provided in the kit. Analytical sensitivity The kit CMV Real-TM Quant allows to detect CMV DNA in 100% of the tests with a sensitivity not less than 200 copies/ml. The detection was carried out on the control standard and its dilutions by negative plasma. Linearity The linearity of the CMV Real-TM Quant assay was tested with the CMV DNA Standard and it s dilution using negative human plasma. Each dilution was analysed three times and the mean CMV DNA titer of each sample was determined. CMV Real-TM Quant is linear from 5 x 10 2 to 5 x 10 5 copies/ml. Test results greater than copies/ml are above the upper limit of quantitation of the test and should be reported as greater than copies/ml. If quantitation results are desired for such samples, the specimen should be diluted 1:10 with negative serum and retested. Test results less than 500 copies/ml are below the lower limit of quantitation of the test and should be reported as less than 500 copies/ml. TROUBLESHOOTING 1. Weak (Ct > 32) or no signal of the IC (Fam (Green) channel). The PCR was inhibited. Make sure that you use a recommended DNA extraction method and follow to the manufacturer s instructions. Re-centrifuge all the tubes before pipetting of the extracted DNA for 2 min at maximum speed ( g) and take carefully supernatant. Don t disturb the pellet, sorbent inhibit reaction The reagents storage conditions didn t comply with the instructions. Check the storage conditions The PCR conditions didn t comply with the instructions. Check the PCR conditions and select for the IC detection the fluorescence channel reported in the protocol. The IC was not added to the sample during the pipetting of reagents. Make attention during the DNA extraction procedure. 2. Weak (Ct > 32) or no signal of the Positive Control. The PCR conditions didn t comply with the instructions. Check the amplification protocol and select the fluorescence channel reported in the manual. 3. Any signal with Negative Control of extraction. Contamination during DNA extraction procedure. All samples results are invalid. Decontaminate all surfaces and instruments with sodium hypochlorite and ethanol. Use only filter tips during the extraction procedure. Change tips between tubes. Repeat the DNA extraction with the new set of reagents. 4. Any signal with Negative Control of PCR (TE-buffer). Contamination during PCR preparation procedure. All samples results are invalid. Decontaminate all surfaces and instruments with sodium hypochlorite and ethanol or special DNA decontamination reagents. Pipette the Positive control at last. Repeat the PCR preparation with the new set of reagents. *icycler and iq5 are trademarks of Bio-Rad Laboratories * Rotor-Gene Technology is a registered trademark of Corbett Research * *Applied Biosystems is trademarks of Applera Corporation * SmartCycler is a registered trademark of Cepheid Sacace Biotechnologies Srl 18 San Carlo str., Caserta, Italy *PCR: The Polymerase Chain Reaction (PCR) process is covered by patents owned by Hoffmann-La Roche and applicable in certain countries. Sacace does not encourage or support the unauthorized or unlicensed use of the PCR process. Use of this kit is recommended for persons that either have a license to perform PCR or are not required to obtain a license
HBV Real-TM Quant Handbook
HBV Real-TM Quant Handbook Real Time Kit for the Quantitative detection of Hepatitis B Virus in human plasma for use with SmartCycler (Cepheid), RotorGene 3000/6000 (Corbett Research), iq icycler and iq5
More informationCMV FEP ALA For nucleic acid amplification of CMV DNA and Fluorescence detection with End Point analysis (FEP) on Aladin (Sacace)
REF TV7-00FRT VER 3.0.06 CMV FEP ALA For nucleic acid amplification of CMV DNA and Fluorescence detection with End Point analysis (FEP) on Aladin (Sacace) Key to symbols used REF List Number Store at 2-8
More informationHIV DNA Qual FRT Real Time Kit
REF TR-V1-D HIV DNA Qual FRT Real Time Kit Key to symbols used List Number Store at 2-8 C/-20 C RUO For Research Use Only Caution! Lot Number VER Version Expiration Date Negative Control Contains reagents
More informationHDV Real-TM. Handbook
HDV Real-TM Handbook for use with RotorGene 3000/6000 (Corbett Research), SmartCycler (Cepheid), iq icycler and iq5 (Biorad), Applied Biosystems 7300/7500 Real Time PCR Systems (Applera), MX3000P and MX3005P
More informationReal Time Kit. Pneumocystis jirovecii (carinii) Real-TM Handbook. for the detection of Pneumocystis jirovecii (carinii) in biological materials
IVD For in Vitro Diagnostic Use Pneumocystis jirovecii (carinii) Real-TM Handbook Real Time Kit for the detection of Pneumocystis jirovecii (carinii) in biological materials REF P2-50FRT VER 07.02.2010
More informationHDV Real-TM Quant Handbook
HDV Real-TM Quant Handbook Real Time PCR Kit for the Quantitative detection of Hepatitis D Virus in human plasma for use with Real Time PCR instruments, like RotorGene 3000/6000/Q (Corbett Research, Qiagen),
More informationPneumocystis jirovecii (carinii) Real-TM
IVD For in Vitro Diagnostic Use Pneumocystis jirovecii (carinii) Real-TM Handbook Real Time PCR Kit for the detection of Pneumocystis jirovecii (carinii) in biological materials REF P2-50FRT REF TP2-50FRT
More informationHandbook. Influenza A H1N1 & H3N2 Real-TM. Real Time Amplification test for the detection of Influenza A H1N1 and H3N2 viruses
Influenza A H1N1 & H3N2 Real-TM Handbook Real Time Amplification test for the detection of Influenza A H1N1 and H3N2 viruses REF V54-50FRT REF TV54-50FRT 50 NAME Influenza A H1N1 & H3N2 Real-TM INTENDED
More informationEBV Real-TM Quant Handbook
IVD For in Vitro Diagnostic Use EBV Real-TM Quant Handbook Real Time PCR Kit for quantitative detection of Epstein Barr Virus (EBV) for use with RotorGene 3000/6000 (Corbett Research), SmartCycler (Cepheid),
More informationPoliovirus Real-TM Handbook
Poliovirus Real-TM Handbook for use with RotorGene 3000/6000 (Corbett Research), SmartCycler (Cepheid), iq icycler and iq5 (Biorad), Applied Biosystems 7300/7500 Real Time PCR Systems (Applera), MX3000P
More informationHPV High Risk Screen Real-TM Quant
For in Vitro Diagnostic Use HPV High Risk Screen Real-TM Quant HANDBOOK Real Time PCR kit for quantitative detection of Human Papillomavirus (16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59) REF V31-100/2FRT
More informationLeukemia BCR-ABL Fusion Gene Real Time RT-PCR Kit
Revision No.: ZJ0003 Issue Date: Aug 7 th, 2008 Leukemia BCR-ABL Fusion Gene Real Time RT-PCR Kit Cat. No.: TR-0126-02 For use with ABI Prism 7000/7300/7500/7900(96 well); Smart Cycler II; icycler iq 4/iQ
More informationPneumocystis Carinii Real Time PCR Kit. For In Vitro Diagnostic Use Only User Manual
Revision No.: ZJ0003 Issue Date: Aug 7 th, 2008 Pneumocystis Carinii Real Time PCR Kit Cat. No.: QD-0082-02 For use with ABI Prism 7000/7300/7500/7900; Smart CyclerII; icycler iq 4/iQ 5; Rotor Gene 2000/3000;
More informationRealLine Mycoplasma genitalium Str-Format
Instructions for use ASSAY KIT FOR THE QUALITATIVE DETECTION OF MYCOPLASMA GENITALIUM DNA BY REAL-TIME PCR METHOD In vitro Diagnostics () VBD4396 96 Tests valid from December 2018 Rev06_1218_EN Page 1
More informationaltona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS
altona DIAGNOSTICS Instructions for Use RealStar CMV PCR Kit 1.2 08/2017 EN RealStar RealStar CMV PCR Kit 1.2 For research use only! (RUO) 021202 INS-021200-EN-S01 48 08 2017 altona Diagnostics GmbH Mörkenstr.
More informationRealLine HIV quantitative Str-Format
Instructions for use DETECTION AND QUANTIFICATION OF THE HUMAN IMMUNODEFICIENCY VIRUS RNA BY REAL TIME PCR Research Use Only (RUO) Attention! Please read the information about quantification process carefully!
More information1. Intended Use New Influenza A virus real time RT-PCR Panel is used for the detection of universal influenza A virus, universal swine Influenza A vir
New Influenza A Virus Real Time RT-PCR Kit User Manual LT028310RRFY - 1 - 1. Intended Use New Influenza A virus real time RT-PCR Panel is used for the detection of universal influenza A virus, universal
More informationFor in vitro Veterinary Diagnostics only. Kylt Rotavirus A. Real-Time RT-PCR Detection.
For in vitro Veterinary Diagnostics only. Kylt Rotavirus A Real-Time RT-PCR Detection www.kylt.eu DIRECTION FOR USE Kylt Rotavirus A Real-Time RT-PCR Detection A. General Kylt Rotavirus A products are
More informationAmpliSens EBV / CMV / HHV6-screen-FRT PCR kit
For Professional Use Only AmpliSens EBV / CMV / HHV6-screen-FRT PCR kit Instruction Manual AmpliSens Ecoli s.r.o., Studenohorska 12 841 03 Bratislava 47 Slovak Republic Tel.: +421 2 6478 9336 Fax: +421
More informationRealLine HBV / HCV / HIV Str-Format
Instructions for use ASSAY KIT FOR THE QUALITATIVE AND DIFFERENTIAL DETECTION OF HEPATITIS B VIRUS DNA, HEPATITIS C VIRUS RNA, AND HUMAN IMMUNODEFICIENCY VIRUS TYPES 1 AND 2 RNA USING THE PCR/RT-PCR METHOD
More informationCMV Real-TM Handbook
For in Vitro Diagnostic Use For Professional Use Only CMV Real-TM Handbook Real Time PCR kit for the qualitative detection of Cytomegalovirus (CMV) REF V7-100FRT 100 NAME CMV Real-TM INTENDED USE CMV Real-TM
More informationRealLine HCV Qualitative Str-Format
Instructions for use REAL TIME PCR DETECTION KIT FOR THE HEPATITIS C VIRUS RNA (HCV) Research Use Only (RUO) (Str-format) VBD0795 96 Tests valid from: October 2018 Rev05_1018_EN Page 1 of 8 Explanation
More informationRealLine HCV Quantitative - Uni-Format
Instructions for use - KIT FOR EXTRACTION OF RNA AND SUBSEQUENT REAL TIME PCR DETECTION AND QUANTIFICATION OF HEPATITIS C VIRUS RNA Attention! Please read the information about quantification process carefully!
More informationRealLine HIV qualitative Str-Format
Instructions for Use REAL TIME PCR DETECTION KIT FOR HUMAN IMMUNODEFICIENCY VIRUS RNA Research Use Only (RUO) RealLine HIV Qualitative (Str-format) VBD0196 96 Tests valid from July 2016 Rev01072016_EN
More informationHIV-1 Viral Load Real Time (RG)
-1 Viral Load Real Time (RG) Real Time RT-PCR type 1 RNA quantification assay MSP Reg. pending Valdense 3616. 11700. Montevideo. Uruguay. phone (598) 2 336 83 01. Fax (598) 2 336 71 60. Info@atgen.com.uy
More informationABIOpure TM Viral (version 2.0)
ABIOpure TM Viral (version 2.0) DNA/RNA Extraction Handbook Cat No: M561VT50 FOR RESEARCH USE ONLY Table of Contents Contents Page Kit Components 3 Precautions 3 Stability & Storage 4 General Description
More informationRealLine HCV quantitative Str-Format
Instructions for use QUANTITATIVE KIT FOR REAL TIME PCR DETECTION FOR RNA OF HEPATITIS C VIRUS Attention! Please read the information about quantification process carefully! Research Use Only (RUO) (Str
More informationAmoyDx TM BRAF V600E Mutation Detection Kit
AmoyDx TM BRAF V600E Mutation Detection Kit Detection of V600E mutation in the BRAF oncogene Instructions For Use Instructions Version: B3.1 Date of Revision: April 2012 Store at -20±2 o C 1/5 Background
More informationHBeAg and HBeAg Ab ELISA Kit
HBeAg and HBeAg Ab ELISA Kit Catalog Number KA0290 96 assays Version: 17 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3
More informationInstructions for Use. RealStar Influenza Screen & Type RT-PCR Kit /2017 EN
Instructions for Use RealStar Influenza Screen & Type RT-PCR Kit 4.0 05/2017 EN RealStar Influenza Screen & Type RT-PCR Kit 4.0 For research use only! (RUO) 164003 INS-164000-EN-S01 96 05 2017 altona
More informationInstructions for Use. RealStar Influenza S&T RT-PCR Kit /2017 EN
Instructions for Use RealStar Influenza S&T RT-PCR Kit 3.0 01/2017 EN RealStar Influenza S&T RT-PCR Kit 3.0 For research use only! (RUO) 163003 INS-163000-EN-S02 96 01 2017 altona Diagnostics GmbH Mörkenstr.
More informationAmpliX HBV Quantitative
Instructions for use REAL TIME PCR DETECTION AND QUANTITATION KIT OF HEPATITIS B VIRUS DNA Research Use Only (RUO) (Lyo-format) VBD0595 96 rcs valid from May 2013 Explanation of symbols used in labeling
More informationMolecularMD. One-Step qrt-pcr BCR-ABL Kit. Product Description and User Manual. For Quantitative RT-PCR Analysis of BCR-ABL. Contact Us.
Contact Us If you have any questions for or comments about MolecularMD, please feel free to contact us. Email Customer Service CustomerService@MolecularMD.com Technical Support TechSupport@MolecularMD.com
More informationHuman Rotavirus B. Non structural protein 5 (NSP5) 150 tests. Quantification of Human Rotavirus B genomes Advanced kit handbook HB10.01.
PCR Max Ltd TM qpcr test Human Rotavirus B Non structural protein 5 (NSP5) 150 tests For general laboratory and research use only 1 Introduction to Human Rotavirus B Rotavirus is a genus of double-stranded
More informationAmpliSens Influenza virus A H5N1-FRT PCR kit
For Professional Use Only AmpliSens Influenza virus A H5N1-FRT PCR kit Instruction Manual AmpliSens Ecoli s.r.o., Studenohorska 12 841 03 Bratislava 47 Slovak Republic Tel.: +421 2 6478 9336 Fax: +421
More information(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR
Real-Time (DNA) Real-time Exicycler 96 Rotor-Gene Q/6000 IU Mix1 Mix2 IU/μl IU/μl IU/μl IU/μl IU/μl IPC NTC C 1 Lot# 2 Freeze & thawing 1 MSDS: Material Safety Data Sheets Real- (TaqMan time ' FAM ' BHQ1
More informationMouse C-Peptide ELISA Kit
Mouse C-Peptide ELISA Kit Cat.No: DEIA4507 Lot. No. (See product label) Size 96T Intended Use The Mouse C-Peptide ELISA kit is for the quantitative determination of c-peptide in mouse serum, plasma, and
More informationManual FTD BCE. Quantitative assay for in vitro diagnostics
Manual 32 reactions (catalog no. FTD-54.1-32) 64 reactions (catalog no. FTD-54.1-64) Quantitative assay for in vitro diagnostics For use with the ABI 7500, ABI 7500 Fast, ViiA 7, Bio-Rad CFX96, LightCycler
More informationPhosphate buffered saline (PBS) for washing the cells TE buffer (nuclease-free) ph 7.5 for use with the PrimePCR Reverse Transcription Control Assay
Catalog # Description 172-5080 SingleShot Cell Lysis Kit, 100 x 50 µl reactions 172-5081 SingleShot Cell Lysis Kit, 500 x 50 µl reactions For research purposes only. Introduction The SingleShot Cell Lysis
More informationProduct # Kit Components
3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Pneumocystis jirovecii PCR Kit Product # 42820 Product Insert Background Information
More informationSputum DNA Collection, Preservation and Isolation Kit 50 Individual Devices
3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Sputum DNA Collection, Preservation and Isolation Kit 50 Individual
More information(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR
Real-Time (DNA) Real-time Exicycler 96 Rotor-Gene Q/6000 IU Mix1 Mix2 IU/μl IU/μl IU/μl IU/μl IU/μl IPC NTC C 1 Lot# 2 Freeze & thawing 1 MSDS: Material Safety Data Sheets (TaqMan Real-time ' FAM ' BHQ1
More informationHuman Rotavirus A. genesig Standard Kit. Non structural protein 5 (NSP5) 150 tests. Primerdesign Ltd. For general laboratory and research use only
TM Primerdesign Ltd Human Rotavirus A Non structural protein 5 (NSP5) genesig Standard Kit 150 tests For general laboratory and research use only 1 Introduction to Human Rotavirus A Rotavirus is a genus
More informationHAV IgG/IgM Rapid Test
HAV IgG/IgM Rapid Test Cat. No.:DTS649 Pkg.Size: Intended use CD HAV IgG/IgM Rapid Test is a solid phase immunochromatographic assay for the rapid, qualitative and differential detection of IgG and IgM
More informationKit Components Product # EP42720 (24 preps) MDx 2X PCR Master Mix 350 µl Cryptococcus neoformans Primer Mix 70 µl Cryptococcus neoformans Positive
3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Cryptococcus neoformans End-Point PCR Kit Product# EP42720 Product
More informationProcine sphingomyelin ELISA Kit
Procine sphingomyelin ELISA Kit For the quantitative in vitro determination of Procine sphingomyelin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH
More informationAmpliSens Influenza virus A-type-FRT
For Professional Use Only AmpliSens Influenza virus A-type-FRT PCR kit Instruction Manual AmpliSens Ecoli s.r.o., Studenohorska 12 841 03 Bratislava 47 Slovak Republic Tel.: +421 2 6478 9336 Fax: +421
More informationE.Z.N.A. SQ Blood DNA Kit II. Table of Contents
E.Z.N.A. SQ Blood DNA Kit II Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Blood Storage and DNA Yield...4 Preparing Reagents...5 100-500 μl Whole Blood Protocol...6
More informationInstructions for use. TSH rat ELISA. Please use only the valid version of the Instructions for Use provided with the kit AR E-8600
Instructions for use TSH rat ELISA AR E-8600 TSH rat ELISA 1. INTRODUCTION 1.1 Intended use The TSH rat ELISA is an enzyme immunoassay for the quantitative measurement of TSH in rat serum. For research
More informationMALDI Sepsityper Kit
Instructions for Use MALDI Sepsityper Kit Kit for identification of microorganisms from positive blood cultures using the MALDI Biotyper system CARE products are designed to support our worldwide customers
More informationHuman Rotavirus C. genesig Advanced Kit. DNA testing. Everything... Everyone... Everywhere... Non structural protein 5 (NSP5) 150 tests
TM Primerdesign Ltd TM Primerdesign Ltd Human Rotavirus C Non structural protein 5 (NSP5) genesig Advanced Kit 150 tests DNA testing Everything... Everyone... Everywhere... For general laboratory and research
More informationartus CMV TM PCR Kit Handbook
artus CMV TM PCR Kit Handbook 24 (catalog no. 4503163) 96 (catalog no. 4503165) Quantitative in vitro Diagnostics For use with the ABI PRISM 7000, 7700, and 7900HT Sequence Detection Systems December 2014
More informationPRODUCT: RNAzol BD for Blood May 2014 Catalog No: RB 192 Storage: Store at room temperature
PRODUCT: RNAzol BD for Blood May 2014 Catalog No: RB 192 Storage: Store at room temperature PRODUCT DESCRIPTION. RNAzol BD is a reagent for isolation of total RNA from whole blood, plasma or serum of human
More informationNorgen s HIV proviral DNA PCR Kit was developed and validated to be used with the following PCR instruments: Qiagen Rotor-Gene Q BioRad icycler
3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com HIV Proviral DNA PCR Kit Product # 33840 Product Insert Background Information
More informationaltona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 always a drop ahead. 11/2012
altona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 11/2012 altona Diagnostics GmbH Moerkenstr. 12 22767 Hamburg Germany phone +49 40 548 06 76-0 fax +49 40 548 06 76-10 e-mail info@altona-diagnostics.com
More informationAbbott RealTime IDH2. For Infomation Only - Not a Controlled Copy /R1 08N /R1 08N82-090
08N82-090 51-608430/R1 Abbott RealTime IDH2 en 08N82-090 51-608430/R1 ONLY ONLY Key to Symbols Used List Number In Vitro Diagnostic Medical Device Lot Number In Vitro Test For In Vitro Diagnostic Use Product
More informationStorage: Logix Smart Zika Virus Master Mix and Logix Smart Zika Virus Positive Control must be stored at -20 ⁰C and can last up to 60 days.
Logix Smart Zika Virus (ZIKV-K-003; ZIKV-PC-003; GEN-NTC-001) Description The Logix Smart Zika Virus Test developed by Co-Diagnostics, Inc. detects ribonucleic acid (RNA) of Zika Virus in a single step
More informationHuman Rotavirus A. genesig Advanced Kit. Non structural protein 5 (NSP5) 150 tests. Primerdesign Ltd. For general laboratory and research use only
TM Primerdesign Ltd Human Rotavirus A Non structural protein 5 (NSP5) genesig Advanced Kit 150 tests For general laboratory and research use only 1 Introduction to Human Rotavirus A Rotavirus is a genus
More informationRat Hemoglobin A1c (HbA1c) Kit Instructions
V.3 Crystal Chem Rat Hemoglobin A1c (HbA1c) Kit Instructions For the quantitative determination of hemoglobin A1c (HbA1c) in rat whole blood Catalog #80300 96 Assays For research use only. Not for use
More informationCotinine (Mouse/Rat) ELISA Kit
Cotinine (Mouse/Rat) ELISA Kit Catalog Number KA2264 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle
More informationTriiodothyronine (T3) ELISA
For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. Triiodothyronine (T3) ELISA Kit is intended for the detection of total T3 in human serum or plasma. For research
More informationDETECTION AND QUANTITATION OF THE HUMAN IMMUNODEFICIENCY VIRUS RNA BY REAL TIME PCR
Lyo-Format Instructions for use Lyo-Format DETECTION AND QUANTITATION OF THE HUMAN IMMUNODEFICIENCY VIRUS RNA BY REAL TIME PCR (Lyo-format) 5645 96 rcs valid from May 2013 Distribué par / Distributed by:
More informationSIV p27 Antigen ELISA Catalog Number:
INTENDED USE The RETRO-TEK SIV p27 Antigen ELISA is for research use only and is not intended for in vitro diagnostic use. The RETRO-TEK SIV p27 Antigen ELISA is an enzyme linked immunoassay used to detect
More informationHepatitis A virus IgM ELISA Kit
Hepatitis A virus IgM ELISA Kit Catalog Number KA0285 96 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle
More informationHuman Cytomegalovirus Virus (CMV) IgG ELISA Kit
Human Cytomegalovirus Virus Catalog No: IRAPKT1410 (CMV) IgG ELISA Kit Lot No: SAMPLE INTENDED USE The CMV IgG ELISA is intended for use in evaluating a patient s serologic status to cytomegalovirus (CMV)
More informationMidi Plant Genomic DNA Purification Kit
Midi Plant Genomic DNA Purification Kit Cat #:DP022MD/ DP022MD-50 Size:10/50 reactions Store at RT For research use only 1 Description: The Midi Plant Genomic DNA Purification Kit provides a rapid, simple
More informationCladosporium_spp genesig Standard Kit
TM Primerdesign Ltd Cladosporium_spp genesig Standard Kit 150 tests For general laboratory and research use only 1 Introduction to Cladosporium_spp Cladosporium is a genus of slow growing fungi, colonies
More informationHuman Cytomegalovirus IgM ELISA Kit
Human Cytomegalovirus IgM Catalog No: IRAPKT2012 ELISA Kit Lot No: SAMPLE INTENDED USE The CMV IgM ELISA is intended for use in the detection of IgM antibodies to Cytomegalovirus (CMV) infection in human
More informationAbbott RealTime HIV-1
E RealTime HIV-1 6L18 51-602146/R2 Abbott RealTime HIV-1 Customer Service: 1-800-553-7042 This package insert must be read carefully prior to use. Package insert instructions must be followed accordingly.
More informationRotavirus A. genesig Standard Kit. DNA testing. Everything... Everyone... Everywhere... Non structural protein 5 (NSP5) 150 tests.
TM Primerdesign Ltd TM Primerdesign Ltd Rotavirus A Non structural protein 5 (NSP5) genesig Standard Kit 150 tests DNA testing Everything... Everyone... Everywhere... For general laboratory and research
More informationUrine - Based XMRV RT-PCR Detection Kit Product # 34700
3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Urine - Based XMRV RT-PCR Detection Kit Product # 34700 Product
More informationTechnical Bulletin. Technical Information for Lyra Direct Strep Assay
Technical Bulletin Technical Information for Lyra Direct Strep Assay Quidel Corporation has verified the performance of the Lyra Direct Strep Assay on Cepheid s SmartCycler II, software version 3.0b. Internal
More informationNorgen s HIV Proviral DNA PCR Kit was developed and validated to be used with the following PCR instruments: Qiagen Rotor-Gene Q BioRad T1000 Cycler
3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com HIV Proviral DNA PCR Kit Product# 33840 Product Insert Intended
More informationGlobal Histone H3 Acetylation Assay Kit
Global Histone H3 Acetylation Assay Kit Catalog Number KA0633 96 assays Version: 06 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle
More informationMicrosart Calibration Reagent
Instructions for Use Microsart Calibration Reagent Prod. No. SMB95-2021 Mycoplasma arginini Prod. No. SMB95-2022 Mycoplasma orale Prod. No. SMB95-2023 Mycoplasma gallisepticum Prod. No. SMB95-2024 Mycoplasma
More informationHuman influenza A virus subtype (H1)
PCRmax Ltd TM qpcr test Human influenza A virus subtype (H1) Haemoglutinin H1 gene 150 tests For general laboratory and research use only 1 Introduction to Human influenza A virus subtype (H1) Influenza,
More informationEpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric)
EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric) Base Catalog # PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE Uses: The EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric)
More informationCladosporium_spp genesig Advanced Kit
TM Primerdesign Ltd Cladosporium_spp genesig Advanced Kit 150 tests For general laboratory and research use only 1 Introduction to Cladosporium_spp Cladosporium is a genus of slow growing fungi, colonies
More informationRotavirus Test Kit. Instructions For Use. Format: Cassette Specimen: Fecal Extract Catalog Number: VEL-001-ROTA
Rotavirus Test Kit Instructions For Use Format: Cassette Specimen: Fecal Extract Catalog Number: VEL-001-ROTA * Please read the instructions carefully before use INTENDED USE Velotest Rotavirus Test is
More informationAvian Influenza A H5N8
TM Primerdesign Ltd Avian Influenza A H5N8 Hemagglutinin (HA) gene & Neuraminidase (NA) gene genesig Standard Kit 150 tests For general laboratory and research use only 1 Introduction to Avian Influenza
More informationAvian influenza A virus subtype (H7)
TM Primerdesign Ltd Avian influenza A virus subtype (H7) Haemoglutinin H7 gene genesig Advanced Kit 150 tests For general laboratory and research use only 1 Introduction to Avian influenza A virus subtype
More informationAvian influenza A virus subtype (H9)
Techne qpcr test Avian influenza A virus subtype (H9) Hemagglutinin (HA) gene 150 tests For general laboratory and research use only 1 Introduction to Avian influenza A virus subtype (H9) Influenza, commonly
More informationTNF-alpha ELISA. For Research Use Only. Not For Use In Diagnostic Procedures.
TNF-alpha ELISA For the quantitative determination of TNF-alpha in serum, plasma, buffered solution or cell culture medium. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number:
More informationAvian influenza A virus subtype (H5)
TM Primerdesign Ltd Avian influenza A virus subtype (H5) Haemoglutinin H5 gene genesig Standard Kit 150 tests For general laboratory and research use only 1 Introduction to Avian influenza A virus subtype
More informationHuman influenza A virus subtype (H3)
PCRmax Ltd TM qpcr test Human influenza A virus subtype (H3) Haemoglutinin H3 gene 150 tests For general laboratory and research use only 1 Introduction to Human influenza A virus subtype (H3) Influenza,
More informationby nexttec TM 1 -Step
Protocol DNA Isolation from Bacteria by nexttec TM 1 -Step - nexttec cleancolumns - Cat. No. 20N.010 Cat. No. 20N.050 Cat. No. 20N.250 Version 1.0 For research only Principle nexttec 1 -Step is the easiest
More informationEPIGENTEK. EpiQuik Global Histone H3 Acetylation Assay Kit. Base Catalog # P-4008 PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE
EpiQuik Global Histone H3 Acetylation Assay Kit Base Catalog # PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE The EpiQuik Global Histone H3 Acetylation Assay Kit is suitable for specifically measuring global
More informationTechnical Bulletin. Technical Information for Quidel Molecular Influenza A+B Assay on the Bio-Rad CFX96 Touch
Technical Bulletin Technical Information for Quidel Molecular Influenza A+B Assay on the Bio-Rad CFX96 Touch Quidel Corporation has verified the performance of the Quidel Molecular Influenza A+B Assay
More informationHSV-1 IgM ELISA. Catalog No (96 Tests) For Research Use Only. Not for use in Diagnostic Procedures.
For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. HSV-1 IgM ELISA Kit is intended for the detection of IgM antibody to HSV-1 in human serum or plasma. SUMMARY AND
More informationFor Information Only - Not a Controlled Copy. Abbott RealTime CMV. en 09N /R1
Abbott RealTime CMV en 09N2-090 5-608290/R 09N2-090 5-608290/R CONTROL LOW + Key to Symbols Used List In Vitro Diagnostic Medical Device Lot In Vitro Test For In Vitro Diagnostic Use Product of USA Global
More informationAvian influenza A virus subtype (H7)
TM Primerdesign Ltd Avian influenza A virus subtype (H7) Haemoglutinin H7 gene genesig Standard Kit 150 tests For general laboratory and research use only 1 Introduction to Avian influenza A virus subtype
More informationAmpliSens Influenza virus A/B-FRT PCR kit
For Professional Use Only AmpliSens Influenza virus A/B-FRT PCR kit Instruction Manual AmpliSens Ecoli s.r.o., Studenohorska 12 841 03 Bratislava 47 Slovak Republic Tel.: +421 2 6478 9336 Fax: +421 2 6478
More informationHuman TSH ELISA Kit. User Manual
Human TSH ELISA Kit User Manual Catalog number: GTX15585 GeneTex Table of Contents A. Product Description... 2 B. Kit Components... 3 C. Additional Required Materials (not included)... 3 D. Reagent Preparation...
More informationXCF TM COMPLETE Exosome and cfdna Isolation Kit (for Serum & Plasma)
XCF TM COMPLETE Exosome and cfdna Isolation Kit (for Serum & Plasma) Cat# XCF100A-1 User Manual Store kit components at +4ºC and +25ºC Version 1 2/2/2017 A limited-use label license covers this product.
More informationCytomegalovirus (CMV) End-Point PCR Kit Product# EP36300
3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Cytomegalovirus (CMV) End-Point PCR Kit Product# EP36300 Product
More informationRat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.
Rat Insulin ELISA For the quantitative determination of insulin in rat serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: Size: 80-INSRT-E01, E10 96 wells, 10
More informationExiPrep TM Viral DNA/ RNA Kit
Ver. 2.0 ExiPrep TM Viral DNA Kit (K-3515) ExiPrep TM Viral RNA Kit (K-3525) (K-3535) Safety Warnings and Precautions is developed and sold for research purposes only. It is not recommended for human or
More informationMouse/Rat THYROXINE (T4) ELISA Catalog No (96 Tests)
For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. Mouse/Rat Thryroxine Kit is intended for the detection of total T4 in mouse/rat serum or plasma. SUMMARY AND EXPLANATION
More informationMorinaga Mouse C-peptide ELISA Kit
Morinaga Mouse C-peptide ELISA Kit For the quantitative determination of C-peptide in mouse serum, plasma, and fluid 96wells For Laboratory Use Only, not for use in diagnostic procedure Please read full
More informationMethodology for the Extraction of Brain Tissue Protein. Learning Objectives:
Proteomics Extraction of Brain Tissue Protein Methodology for the Extraction of Brain Tissue Protein Extraction of the entire protein from the sample requires optimized protocol and many protocols have
More information