1 RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU 2 1 Bucharest Emergency University Hospital, Romania 2 Horia Hulubei National Institute for Physics and Nuclear Engineering, IFIN-HH, Bucharest Reactorului 30, P.O. Box MG-6, Măgurele, Romania Received July 6, 2012 The aspects of radiation injuries in IR were highlighted and the increasing concern about the high skin dose levels in cardiology and other IR procedures, as it is very important a close monitoring in order to prevent stochastic effects and, in time, deterministic effects. The present study reports on investigations that we have performed to allow the calculation of the individual penetrant radiation dose, H p (10) in interventional radiology and also the important results in determining the homogeneity of the radiation flux. In order to determine the uniformity of large area beams with a better precision, thermoluminescent matrices of 8 12cm size, containing arrays of chips, uniformly distributed in each row and column of the matrix, were used. From the measurements we obtained accurate data of the dose received by patients in different areas of the body. Measurement method using TLD's is net superior to the measurement with Dap-meter. Due to the potential for stochastic and deterministic effects existing in interventional radiology, dosimetry information is required to make the interventionalist aware of the potential for such effects. Key words: equivalent dose, TLD, DAP. INTRODUCTION International standardization in dosimetry is essential for good exploitation of the X-radiation based technology. International Atomic Energy Agency (IAEA) and other nuclear activity control bodies initiated different programs regarding ionisation radiation dosimetry, in order to help the international community, including the EU member states, to develop standards and measurement capacities of dosimetry equipments required for applying radiation in medicine. For some time there was a growing awareness that the amount of the equivalent doses to patients after performing medical procedures for diagnostic radiology contributes at growing proportion of the total dose to the population. Interventional radiology procedures (IR) and interventional cardiology procedures (IC) are registering the highest radiation doses received by patients within one year of radiological examinations. Stochastic effects that may result from these investigations are very important, especially for the young population. Rom. Journ. Phys., Vol. 58, Nos. 3 4, P , Bucharest, 2013

2 2 Measurement of the equivalent individual doses 331 To optimize patient exposure to radiation, the responsible doctor must know very well the individual penetrant dose, H p (10), to the radiosensitive organs for each type of procedure performed. A precise determination would be very difficult because of exposure to different areas of the body which may overlap in the investigation. Even when a full protocol of all exposure parameters is used, it can t provide a good estimate of the effective dose for each organ exposed to the radiation field. There are several direct and indirect methods used to determine the patient's skin dose, but one of the most accurate methods we consider that is thermoluminescent dosimetry (TLD), allowing for evaluation of the equivalent dose at various organs. The aim of the study presented in this paper is to measure the individual penetrant dose, Hp(10), from angiographic investigations with the TLD method and to determine the homogeneity of the radiation beams. MATERIALS AND METHODS For the measurement of the individual penetrant radiation dose equivalent, H p (10) values in Sievert, we used the thermoluminescent system, SD-TL type, set up at the Horia Hulubei National Institute of Physics and Nuclear Engineering, IFIN-HH, described in detail in [1], whose characteristics are in full agreement with the requirements of the standard IEC 61066:2006 [2]. It contains the dosimeters based on the use of LiF:Mg, Cu, P detectors, as chips, commercially known as GR- 200A and a TLD reader type READER ANELYSER RA 94. The calibration of the system was done according to the prescriptions of the ICRU Report 39 [3] in units of H p (10), using a standard radiation field from a 137 Cs standard source. To determine the dose in interventional radiology examinations we used a 30 cm, plexiglas phantom calibrated at PTW Freiburg, Germany, to position the detector arrays. The irradiations were done using the Integris H5000 angiography equipment produced by Philips. In order to determine the uniformity of large area beams with a better precision, thermoluminescent matrices of 8 12cm size, containing arrays of chips, uniformly distributed in each row and column of the matrix, were used. The average time for carrying out an investigation by angiography or interventional cardiology is 20 min. During this time, exposure to radiation is done in area of interest and adjacent areas of the patient to be investigated. In order to prevent deterministic effects it is important to measure the surface input dose; to have the control over stochastic effects is essentially to know deeply absorbed dose. In this respect, we measured the H p (10) values at different depths in phantom: 10 cm, 16 cm, 23 cm and 30 cm, as Fig. 1, representing the PTW-phantom, shows.

3 332 DanielaAdam, Ana Stochioiu 3 Fig. 1 Plexiglas Phantom type PTW. TLD matrices were exposed to an X radiation beam characterized by the following parameters: Tube high voltage, T = 110 kv; Intensity, I = 14.5 ma; Dose Area Product, D DAP = 12 Gy cm 2. Several measurement sets were performed: (i) the first set, 96 chips placed at 10 and 30 cm depths and an exposure time, t = 2 min; (ii) the following sets, with 15 and 11 chips, at 16, 23 and 30 cm depths with t = 1 min exposure times. After readings we obtained the individual and mean values per detectors column of dose equivalent H p (10), with the standard deviation, σ n(col) such as presented in Tables 1 5. Table 1 The values of dose equivalent [msv] for the matrix I, (10 cm depth, t = 2 min). No. of a b c d e f g h i j k l chip Mean σ n(col)

4 4 Measurement of the equivalent individual doses 333 Table 2 The values of dose equivalent [msv], for the matrix II, (30 cm depth, t = 2 min) No of a b c d e f g h i j k l chip Mean σ n (col) Table 3 The values of dose equivalent [msv] for the matrix III, (16 cm depth, t = 1 min) No a b c d e Mean σ n (col) Table 4 The values of dose equivalent [msv] for the matrix IV (23 cm depth, t = 1 min) No a b c d Mean σ n (col) Table 5 The values of dose equivalent [msv] for the matrix V (30 cm depths, t = 1 min) No a b c d Aver. value σ n (col)

5 334 DanielaAdam, Ana Stochioiu 5 Measurements are also important in determining the homogeneity of the radiation flux. By plotting the TLD dose equivalent derived from exposures of 1 min and 2 min, we can observe a variation of radiation flux that influences the radiation dose to the patient skin and collected dose at organs at risk, which are near the exposed area. In principle, the values depend of radiation flux homogeneity. Using the data from the upward tables, one represents graphically minimum and maximum dose distribution on TLD matrix surface, exposed to X radiation, such as presented in Figure 2, which provides an intuitive image of uniformity of beam. Fig. 2 - Equivalent dose distribution on TLD matrix surface, exposed to X radiation. As we can see in Fig 2. the radiation flux is not uniform on the TLD surface. Table 6 represents the summarized values of dose equivalent value [msv] of the 5 matrices of detectors, expressed as mean, minimum and maximum values, calculated from all columns, corresponding to an exposition time of 1 min. Table 6 Mean, maximum and minimum values of individual penetrant dose rate p H (10), msv/min Depths (cm) Average value (msv/min) Min value (msv/min) Max value (msv/min) ± ± ± ± ± ± ± ± ± ± ±0.006 Mean: 0.063± ± ±0.005 Mean: 0.051± ± ±0.004 Mean: 0.067±0.004

6 6 Measurement of the equivalent individual doses 335 RESULTS INTERPRETATION As first conclusion from tables, as expected, the central part of the beam has the maximum intensity; the last two columns contain lower values; regarding the matrices lines, the differences are lower. For the 30 cm, the second line values, calculated from a measurement time of 2 min are in agreement with the 1 min results, what means that for these short intervals the generator parameters remained constant. The graph below, Fig.3, represents the average value data from Table 6. Fig. 3 Individual penetrant dose rate p H (10), msv/min, as function of the depth, d. The mathematical fitting of the data, represented by the continuous line in Figure 3, results in a relation of variation of the equivalent dose rate, msv/min, as function of the depth d, in cm, of the type: H p (10)[mSv/min] = (8.089 ± 0.672) (0.589 ± 0.074) d + ( ± 0.018) This relation approximates satisfactory the variation of equivalent dose rate with the depth and allows the calculation of the value for each depth, inclusively at the entrance surface, d = 0. One can deduce that the exposure times of 20 minutes result in significant equivalent dose values. It is very important a close monitoring in order to prevent stochastic effects and, in time, deterministic effects. The aspects of radiation injuries in IR were highlighted and the increasing concern about the high skin dose levels in d 2

7 336 DanielaAdam, Ana Stochioiu 7 cardiology and other IR procedures were mentioned in Publication 85 of the ICRP [3]. The two main causes mentioned for radiation injuries were: a) the use of suboptimal equipment, and b) procedures performed by individuals inadequately trained in radiation protection. CONCLUSIONS Measurements with TLDs give a good estimation of the dose rates for various exposition times and depths. Dose Area Product (DAP), is a multiplication of the dose and the area exposed, often expressed in Gy.cm 2. Total DAP is not always a good estimator of the risk of radiation. Using the method of measurement with TLD one can have a better overview of doses at different distances and times. From the above measurements we can obtain accurate data of the dose received by patients in different areas of the body. Measurement method using TLD's is net superior to the measurement with Dap-meter. A potential for stochastic and deterministic effects exists in interventional radiology. Dosimetry information is required to make the interventionalist aware of the potential for such effects. Also, for the purposes of tracking doses to patients after IR procedures, it is necessary for details of the patient s dosimetry to be available within the patient record. Radiation protection for patients and staff is one of the main issues for IR. Acknowledgement. Thanks for assistance in achieving this article to Mrs. Prof. Dr. Maria Sahagia and Eng. Ion Tudor from Horia Hulubei National Institute for Physics and Nuclear Engineering, IFIN-HH, Bucharest. REFERENCES 1. A. Stochioiu, F. Mihai, F. Scarlat. A new passive dosimetric system with thermoluminiscent LiF:Mg, Cu, P detectors applied in individual radiation monitoring. Journal of Optoelectronics and Advanced Materials. Vol. 8, No 4, pp (2006). 2. International Standard, International Electrotechnical Commission, IEC 61066:2006. Thermoluminescent dosimetrical systems for the survey of environment and personnel. 3. International Commission for Radiation Units, ICRU Report 39. Determination of dose equivalents resulting from external radiation sources, ICRU, 1985.

THERMOLUMINESCENT DOSIMETRIC SYSTEM USED FOR CARACTERIZATION OF THE UNDERGROUND ENVIRONMENT

THERMOLUMINESCENT DOSIMETRIC SYSTEM USED FOR CARACTERIZATION OF THE UNDERGROUND ENVIRONMENT THERMOLUMINESCENT DOSIMETRIC SYSTEM USED FOR CARACTERIZATION OF THE UNDERGROUND ENVIRONMENT A. Stochioiu*, M. Sahagia, R. Margineanu, S. Bercea, I.Tudor Horia Hulubei National Institute of R&D for Physics

More information

RADIATION MONITORING EXPERIMENT USING THERMOLUMINESCENT DOSIMETER FOR THE TR 19 CYCLOTRON AREA IN NUCLEAR RESEARCH INSTITUTE

RADIATION MONITORING EXPERIMENT USING THERMOLUMINESCENT DOSIMETER FOR THE TR 19 CYCLOTRON AREA IN NUCLEAR RESEARCH INSTITUTE RADIATION MONITORING EXPERIMENT USING THERMOLUMINESCENT DOSIMETER FOR THE TR 19 CYCLOTRON AREA IN NUCLEAR RESEARCH INSTITUTE A. STOCHIOIU, L. CRACIUN, F. MIHAI, I. TUDOR Horia Hulubei National Institute

More information

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS Romanian Reports in Physics, Vol. 63, No. 3, P. 700 706, 2011 DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS ZOE GHITULESCU 1, ANA STOCHIOIU 2, MIHAI DUMITRACHE 3, 1 CNCAN- National

More information

DOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE

DOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE DOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE F. MIHAI, A. STOCHIOIU, C. TUCA National Institute for Nuclear Physics and Engineering, P.O.Box

More information

TLD SYSTEM FOR THE MONITORING OF THE ENVIRONMENTAL RADIOACTIVITY

TLD SYSTEM FOR THE MONITORING OF THE ENVIRONMENTAL RADIOACTIVITY ENVIRONMENTAL PHYSICS TLD SYSTEM FOR THE MONITORING OF THE ENVIRONMENTAL RADIOACTIVITY ANA STOCHIOIU *, MARIA SAHAGIA, ION TUDOR Horia Hulubei National Institute of Research and Development for Physics

More information

TLD system for the monitoring of the environmental radioactivity

TLD system for the monitoring of the environmental radioactivity TLD system for the monitoring of the environmental radioactivity Ana Stochioiu*, Maria Sahagia, Ion Tudor Horia Hulubei National Institute of Research and Development for Physics and Nuclear Engineering,

More information

IMAGES THAT SPEAK ABOUT ERRORS IN PERSONAL DOSIMETER HANDLING AND PARTICULAR SOLUTIONS

IMAGES THAT SPEAK ABOUT ERRORS IN PERSONAL DOSIMETER HANDLING AND PARTICULAR SOLUTIONS Romanian Reports in Physics, Vol. 68, No. 4, P. 1554 1565, 2016 IMAGES THAT SPEAK ABOUT ERRORS IN PERSONAL DOSIMETER HANDLING AND PARTICULAR SOLUTIONS F. MIHAI, A. STOCHIOIU Horia Hulubei National Institute

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

Performance of several active personal dosemeters in interventional radiology and cardiology

Performance of several active personal dosemeters in interventional radiology and cardiology Performance of several active personal dosemeters in interventional radiology and cardiology S. Chiriotti 1,*, M. Ginjaume 1, E. Vano 2,3, R. Sanchez 3, J.M. Fernandez 2,3, M.A. Duch 1, J. Sempau 1 1.

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine Radiation physics and radiation protection University of Szeged Department of Nuclear Medicine Radiation doses to the population 1 Radiation doses to the population 2 Sources of radiation 1 Radiation we

More information

Staff Exposure Monitoring in Interventional Radiology

Staff Exposure Monitoring in Interventional Radiology Conference on Physics in Medicine: From Diagnosis to Treatment "Enhancing safety and quality in radiation medicine". KFMC, Riyadh 7-9 November 2017 Staff Exposure Monitoring in Interventional Radiology

More information

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Luciana C. Matsushima 1, Glauco R. Veneziani 1, Roberto K. Sakuraba

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Radiation Monitoring Instruments

Radiation Monitoring Instruments Radiation Monitoring Instruments This set of slides is based on Chapter 4 authored byg. Rajan, J. Izewska of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics: A Handbook for Teachers

More information

Radiation Protection Services in R.Macedonia

Radiation Protection Services in R.Macedonia Radiation Protection Services in R.Macedonia Lidija Nikolovska Institute of Public Health Ionizing Radiation Laboratory Skopje, Republic of Macedonia Background - no NPP or research reactors, - limited

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 CHARACTERIZATION OF TLD-100

More information

Implementation of personnel dosimetry systems using thermoluminescence dosimetry

Implementation of personnel dosimetry systems using thermoluminescence dosimetry Implementation of personnel dosimetry systems using thermoluminescence dosimetry J. C. Rojas 1, S. O. Benavides 1,2, and H. Olaya 3 1 Radiation Safety Committee, Fundación Clínica Valle del Lili, A.A.

More information

X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS

X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS Carlos Salas Nucleoeléctrica Argentina SA - Embalse NPP Environmental Control and Dosimetry Division INTRODUCTION: This paper describes the methodology

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

A NEW USER-FRIENDLY EXTREMITY DOSIMETRY SYSTEM DESIGN, COMPARISON, AND TESTING

A NEW USER-FRIENDLY EXTREMITY DOSIMETRY SYSTEM DESIGN, COMPARISON, AND TESTING A NEW USER-FRIENDLY EXTREMITY DOSIMETRY SYSTEM DESIGN, COMPARISON, AND TESTING H. Stadtmann 1, J. Fellinger 2, K.J. Velbeck 2, C. Schmitzer 1 J.E. Rotunda 2 and E. Michler 1 1 Austrian Research Centre,

More information

The print quality of this copy is not an accurate representation of the original.

The print quality of this copy is not an accurate representation of the original. Cardiovasc Intervent Radiol DOI 10.1007/s00270-010-9945-4 TECHNICAL NOTE Staff Radiation Doses in a Real-Time Display Inside the Angiography Room Roberto Sanchez E. Vano J. M. Fernandez J. J. Gallego Received:

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Assessment of Patient Radiation Dose in Interventional Procedures at Shahid Madani Heart Center in Khorramabad, Iran Mehrdad Gholami 1*, Somayeh Gharloghi

More information

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY M. R. NEMTANU, C. OPROIU, M. BRASOVEANU, M. OANE National Institute for Laser, Plasma and Radiation Physics, Electron Accelerator

More information

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS Bayero Journal of Pure and Applied Sciences, 2(1): 1-5 Received: October, 2008 Accepted: February, 2009 PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS *Sharifat,

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

Dosimetric Characteristics of LiF:Mg,Cu,Na,Si Phosphor

Dosimetric Characteristics of LiF:Mg,Cu,Na,Si Phosphor Dosimetric Characteristics of LiF:Mg,Cu,Na,Si Phosphor Y. M. Nam 1, J. L. Kim 1, S. Y. Chang 1, B. H. Kim 1 and H. J. Kim 2 1 Health Physics Department, Korea Atomic Energy Research Institute, Taejon 305-353

More information

Establishing the Greek National Reference Levels for Interventional Cardiology procedures

Establishing the Greek National Reference Levels for Interventional Cardiology procedures Establishing the Greek National Reference Levels for Interventional Cardiology procedures George Simantirakis*, Christina Koukorava, Maria Kalathaki, Christos Pa

More information

Accreditation of Dosimetry Services

Accreditation of Dosimetry Services Accreditation of Dosimetry Services J. Van Cauteren, E. De Geest, M. Bricoult AV-Controlatom Content History Regulatory requirements Discussion of reference documents/standards Practical implementation

More information

Radiologic Units: What You Need to Know

Radiologic Units: What You Need to Know Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report

More information

International Journal of Radiation Research, January 2017

International Journal of Radiation Research, January 2017 Downloaded from ijrr.com at : +0330 on Tuesday November th Volume 5, No International Journal of Radiation Research, January 7 Radiation dose to the thyroid, eyes and parotid glands of patients undergoing

More information

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3 Radiation Exposure by Routine Radiographic Examinations: Multicenter Study in Japan with Thermoluminescence Dosimetry and Estimation from the Radiographic Data T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1,

More information

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS I. Villanueva (1), I. Amor (1), A. Brosed (), R. Villanueva () M. Ginjaume (3), X. Ortega (3), M. Roig (3) (1)

More information

Assessing the standard dose to standard patients for x-ray investigations

Assessing the standard dose to standard patients for x-ray investigations Assessing the standard dose to standard patients for x-ray investigations A. Almén and W. Leitz Swedish Radiation Protection Authority, SE-171 16 Stockholm, Sweden E-mail: anja.almen@ssi.se Abstract. The

More information

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing

More information

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY E. BADITA 1, C.VANCEA 1,3, I. CALINA 1,3, D. STROE 2, M. DUMITRACHE 2,3,

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 8529-3 First edition 1998-11-15 Reference neutron radiations Part 3: Calibration of area and personal dosimeters and determination of their response as a function of neutron

More information

Protection of the contralateral breast during radiation therapy for breast cancer

Protection of the contralateral breast during radiation therapy for breast cancer Protection of the contralateral breast during radiation therapy for breast cancer Edgardo Garrigó a*, Alejandro Germanier b, Silvia Zunino a a Instituto Privado de Radioterapia, Ob Oro 423 (5000) Córdoba,

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison MELODI 7th Workshop, 9 11 November 2015 Helmholtz Zentrum München Next Generation Radiation Protection Research Out-of-field dosimetry in radiotherapy for input to epidemiological studies Roger Harrison

More information

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose Submitted to Radiation Measurements Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose P. Bilski 1, M. Sadel 1, J. Swakon 1, A. Weber 2 1

More information

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-8 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry in Interventional Radiology Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

3/26/2017. Personal Dosimetry Monitoring and Dose Measurements. Agenda. Dosimetric Terms and Definitions Dose Limits External Dosimetry

3/26/2017. Personal Dosimetry Monitoring and Dose Measurements. Agenda. Dosimetric Terms and Definitions Dose Limits External Dosimetry Speaker David Pellicciarini, CHP, MBA Vice President, Pharmacy Safety, Practice and Technical Operations Cardinal Health Nuclear Pharmacy Services david.pellicciarini@cardinalhealth.com Personal Dosimetry

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * Romanian Reports in Physics, Vol. 66, No. 2, P. 401 410, 2014 A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * M. D. SUDITU 1,2, D. ADAM 1,2, R. POPA 1,2, V. CIOCALTEI

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Amanda Bravim, Leticia Lucente Campos Instituto de Pesquisas Energéticas e Nucleares (IPEN / CNEN - SP) Av. Professor

More information

Personal Dosimetry in Sweden - Challenges and Developments

Personal Dosimetry in Sweden - Challenges and Developments Personal Dosimetry in Sweden - Challenges and Developments - Present status - Unresolved issues - Developments and trends Swedish Facilities Nuclear facilities 10 operating and 3 closed reactors, fuel

More information

Individual Dosimetry Monitoring: What Is Available In The Context Of A Retrospective Evaluation Of Potential Exposure To The Lens?

Individual Dosimetry Monitoring: What Is Available In The Context Of A Retrospective Evaluation Of Potential Exposure To The Lens? Individual Dosimetry Monitoring: What Is Available In The Context Of A Retrospective Evaluation Of Potential Exposure To The Lens? ORAMED 2011 Barcelona Thursday 20 January Sophie Jacob, PhD sophie.jacob@irsn.fr

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

Determination of the conversion factor for the estimation of effective dose in lungs, urography and cardiac procedures

Determination of the conversion factor for the estimation of effective dose in lungs, urography and cardiac procedures Department of Medical Radiation Physics KAROLINSKA INSTITUTET STOCKHOLM UNIVERSITY Determination of the conversion factor for the estimation of effective dose in lungs, urography and cardiac procedures

More information

Calibration of two 90 Sr+ 90 Y dermatological applicators

Calibration of two 90 Sr+ 90 Y dermatological applicators Calibration of two 90 Sr+ 90 Y dermatological applicators Patrícia L. Antonio *, Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP Av. Prof. Lineu Prestes 2242, 05508-000,

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Radiation Protection- Cath lab

Radiation Protection- Cath lab Radiation Protection- Cath lab Dr. Mawya A Khafaji Associate Prof. Medical Physics, Faculty of Medicine, KAU Head of Medical Physics Unit Dept. of Radiology -KAUH Head, Volunteer Office -KAUH Outline:

More information

Patient Dosimetry in Interventional Procedures: what to measure and estimate, why and how?

Patient Dosimetry in Interventional Procedures: what to measure and estimate, why and how? ICTP/IAEA Training Course on Radiation Protection of Patients 16-27 September 2013, Trieste, Italy Patient Dosimetry in Interventional Procedures: what to measure and estimate, why and how? Renato Padovani

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

The Status and Challenges of External Personal Dosimetry

The Status and Challenges of External Personal Dosimetry The Status and Challenges of External Personal Dosimetry Filip Vanhavere SCK-CEN Introduction 20 years ago Pencil dosemeters Film dosemeters Thermoluminescent dosemeters Personal dosimetry: measuring operational

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

REPORT OF THE TECHNICAL MEETING: THE NEW DOSE LIMIT FOR THE LENS OF THE EYE IMPLICATIONS AND IMPLEMENTATION VIENNA, 2-4 OCTOBER 2012

REPORT OF THE TECHNICAL MEETING: THE NEW DOSE LIMIT FOR THE LENS OF THE EYE IMPLICATIONS AND IMPLEMENTATION VIENNA, 2-4 OCTOBER 2012 REPORT OF THE TECHNICAL MEETING: THE NEW DOSE LIMIT FOR THE LENS OF THE EYE IMPLICATIONS AND IMPLEMENTATION VIENNA, 2-4 OCTOBER 2012 Introduction A Technical Meeting on the The new dose limit for the lens

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

The x-rays produced penetrate the body which absorbs, refracts, or reflects the x-ray beam energy depending on the tissue. Bone

The x-rays produced penetrate the body which absorbs, refracts, or reflects the x-ray beam energy depending on the tissue. Bone Authors Sari Cohen, Poh Yan Lim, Merng Koon Wong, Siew Hong Lau, Donna Russell-Larson 1.6.2 Image intensifier Poh Yan Lim, Merng Koon Wong The discovery of x-rays had a profound impact on the diagnosis

More information

Assessment of effective dose in paediatric CT examinations

Assessment of effective dose in paediatric CT examinations Assessment of effective dose in paediatric CT examinations E. Dougeni 1,2 CL. Chapple 1, J. Willis 1, G. Panayiotakis 2 1 Regional Medical Physics Department, Freeman Hospital, Freeman Road, Newcastle

More information

Conception and design of the National Database of the Individual Dose of the Republic of Cuba

Conception and design of the National Database of the Individual Dose of the Republic of Cuba Conception and design of the National Database of the Individual Dose of the Republic of Cuba Miguel Prendes Alonso *, Maryzury Valdés Ramos, Juan Tomás Zerquera, Daniel Molina Pérez, Gladys López Bejerano1,

More information

OCCUPATIONAL EXPOSURE IN CANDU NUCLEAR POWER PLANT: INDIVIDUAL DOSIMETRY PROGRAM AT CERNAVODA NPP

OCCUPATIONAL EXPOSURE IN CANDU NUCLEAR POWER PLANT: INDIVIDUAL DOSIMETRY PROGRAM AT CERNAVODA NPP OCCUPATIONAL EXPOSURE IN CANDU NUCLEAR POWER PLANT: INDIVIDUAL DOSIMETRY PROGRAM AT CERNAVODA NPP Catalina Chitu, cchitu@cne.ro, Vasile Simionov, vsimionov@cne.ro CNE Cernavoda NPP, No. 2, Medgidiei Str.

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

More information

Isoeffective Dose Specification of Normal Liver in Yttrium-90 Microsphere Radioembolization*

Isoeffective Dose Specification of Normal Liver in Yttrium-90 Microsphere Radioembolization* Isoeffective Dose Specification of Normal Liver in Yttrium-90 Microsphere Radioembolization* Barry W. Wessels, Ph.D 1 ; Amilia G. Di Dia, PhD 2 ;Yiran Zheng, PhD 1 Marta Cremonesi, PhD 2 1 University Hospitals

More information

Radiation Protection Dosimetry (2008), Vol. 131, No. 1, pp Advance Access publication 23 August 2008

Radiation Protection Dosimetry (2008), Vol. 131, No. 1, pp Advance Access publication 23 August 2008 Radiation Protection Dosimetry (2008), Vol. 131, No. 1, pp. 62 66 Advance Access publication 23 August 2008 doi:10.1093/rpd/ncn229 AN OVERVIEW OF THE USE OF EXTREMITY DOSEMETERS IN SOME EUROPEAN COUNTRIES

More information

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography Wilson Otto Gomes Batista Instituto Federal da Bahia Rua Emidio dos Santos s/n. Barbalho 40301-015

More information

Chapter 4: Radiation Monitoring Instruments

Chapter 4: Radiation Monitoring Instruments Chapter 4: Radiation Monitoring Instruments Set of 107 slides based on the chapter authored by G. Rajan, J. Izewska of the IAEA publication (ISBN 92-0-107304-6): Review of Radiation Oncology Physics: A

More information

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters 133 Clinical Radiation Oncology Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters Bhaskar Mukherjee 1,2, Carolina Lina 1 Vladimir Mares 3 1Westdeutsches

More information

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging Iran. J. Radiat. Res., 2008; 6 (2): 83-87 Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging B. Azadbakht 1, M.H. Zahmatkesh 2 *, k. Hadad 1, S. Bagheri 2

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 233-3 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 29 Dosimetry Framework, Formalism including Uncertainties Costas Hourdakis Greek Atomic

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Estimation of Organ and Effective Doses for Neonate and Infant Diagnostic Cardiac Catheterizations

Estimation of Organ and Effective Doses for Neonate and Infant Diagnostic Cardiac Catheterizations Medical Physics and Informatics Original Research Kawasaki et al. Dose Estimates for Neonate and Infant Diagnostic Cardiac Catheterization Medical Physics and Informatics Original Research Toshio Kawasaki

More information

Energy dependent response of Al 2 O 3 and its potential application in personal monitoring.

Energy dependent response of Al 2 O 3 and its potential application in personal monitoring. Energy dependent response of Al 2 O 3 and its potential application in personal monitoring. Nelson, V. K. a, b a Sydney University, Faculty of Health Sciences, Discipline of Medical Imaging and Radiation

More information

EURADOS: Overview of actions and initiatives in radiation protection in medicine

EURADOS: Overview of actions and initiatives in radiation protection in medicine EURADOS: Overview of actions and initiatives in radiation protection in medicine Filip Vanhavere Oslo, June 8 th, 2011 EURADOS: Carries out projects and network activities to Advance the scientific understanding

More information

ISEMIR An IAEA Tool for Radiation Protection Optimization in Interventional Cardiology and Industrial Radiography

ISEMIR An IAEA Tool for Radiation Protection Optimization in Interventional Cardiology and Industrial Radiography An IAEA Tool for Radiation Protection Optimization in Interventional Cardiology and Industrial Radiography Information System on Occupational Exposure in Medicine, Industry and Research What is? is an

More information

Basic definitions. Dosimetry, radiation protection. Nuclear measurement techniques. Interaction of the nuclear radiation with the matter

Basic definitions. Dosimetry, radiation protection. Nuclear measurement techniques. Interaction of the nuclear radiation with the matter Dosimetry, radiation protection. Nuclear measurement techniques. properties measurement dosimetry medical applications of the nuclear radiation Basic definitions Nuclear radiation: Produced in the transition

More information

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Clarita Saldarriaga Vargas, Corinne Amalberto, Lara Struelens,

More information

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2. Experiences of the Secondary Standard Dosimetry Laboratory of the CPHR in the implementation and validation of the Postal Quality Audit in Radiotherapy, using TLDs Stefan Gutiérrez Lores 1, Gonzalo Walwyn

More information

"Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY *

Research Note CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY * Iranian Journal of Science & Technology, Transaction A, Vol. 32, No. A3 Printed in the Islamic Republic of Iran, 2008 Shiraz University "Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 :

More information

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Radiation exposure of the Yazd population from medical conventional X-ray examinations Iran. J. Radiat. Res., 2007; 4 (4): 195-200 Radiation exposure of the Yazd population from medical conventional X-ray examinations F. Bouzarjomehri 1*, M.H. Dashti 2, M.H. Zare 1 1 Department of Medical

More information

Recent Achievements in External Radiation Dosimetry A. Delgado CIEMAT Radiation Dosimetry Unit Avda Complutense 22, Madrid, Spain

Recent Achievements in External Radiation Dosimetry A. Delgado CIEMAT Radiation Dosimetry Unit Avda Complutense 22, Madrid, Spain Topical Session TS 1.1.1 External Exposure to Ionizing Radiation Keynote Lecture: Recent Achievements in External Radiation Dosimetry A. Delgado CIEMAT Radiation Dosimetry Unit Avda Complutense 22, 28040

More information

Radiation Dose Risk and Diagnostic Benefit in Imaging Investigations

Radiation Dose Risk and Diagnostic Benefit in Imaging Investigations American Journal of Bioscience and Bioengineering 2015; 3(3-1): 22-26 Published online April 10, 2015 (http://www.sciencepublishinggroup.com/j/bio) doi: 10.11648/j.bio.s.2015030301.14 ISSN: 2328-5885 (Print);

More information

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors G.Gambarini 1, C.Tenconi 1, N.Mantaut 1 M.Carrara 2, M.Borroni 2, E.Pignoli 2 D.Cutajar 3, M.Petasecca 3, I.Fuduli 3, M.Lerch

More information

Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy

Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy MB. Tavakoli 1, M. Saeb 2, H. Emami 3 1 MB Tavakoli, Department of Medical Physics

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

ABSTRACT. Objectives: To investigate the radiation dose received by patients undergoing routine plain x-ray

ABSTRACT. Objectives: To investigate the radiation dose received by patients undergoing routine plain x-ray Radiation Dose Distribution for Patients Undergoing Routine Radiological Scans for Kidney Stone Diagnosis at the University Hospital of the West Indies DC Walker 1, WD Aiken 2, S Shah 3, MK Voutchkov 1,

More information