X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS

Size: px
Start display at page:

Download "X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS"

Transcription

1 X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS Carlos Salas Nucleoeléctrica Argentina SA - Embalse NPP Environmental Control and Dosimetry Division INTRODUCTION: This paper describes the methodology implemented in Embalse NPP for individual X-ray dose assessment in professionals, especially dentists and radiologists, who work in the surrounding areas of the plant This dosimetry is free of charge, and is performed upon request As a contribution to the community, the professionals are temporarily provided with TLD personal dosimeters for thoracic use, as well as TLD ring dosimeters to assess the doses during their job, as well as in their equipments Ring dosimeters are not considered in this paper The implemented methodology is based in several techniques described in the scientific bibliography (see Ref 1) The effective energy in the radiation field is of main importance in this kind of dosimetry, since dosimetry factors depend on that energy The response of the TLD crystals depends on this energy too This is fundamental when assessing the absorbed dose It is really convenient and necessary for the dosimeter to simultaneously determine the effective energy value as well as the corresponding dose value The basic principle for determining effective energy is by using at least two different TLD materials covered by filters of different thickness, in the same dosimeter The different TLD materials give respond differently to energy Therefore, different readouts from each of the crystals are obtained The ratio between both readouts provides a factor that depends on the effective energy, but that is independent from the exposure values irradiated to the dosimeter 1

2 EXPERIMENTAL DETAILS The Personal TLD dosimeter currently in use is Bicron-Harshaw It comprises a model 8807 card holder This holder contains a card model 2211 which groups two TLD 200 crystals and two TLD 100 crystals It has internal filters at each side of the TLD 200 crystals, which diminish energy dependence Using four crystals has the advantage of improving measurement statistics, and at the same time provides the necessary redundancy in case of crystal damage These dosimeters are evaluated by an automatic Bicron-Harshaw 6600 E TLD reader (see figure 1) Figure 1 The periodical calibration of these dosimeters consists in the irradiation of some dosimeters with different X-ray energy beams in the National Atomic Energy Commission (CNEA) Once Embalse NPP gets the corresponding readouts, different ratios are made, and then different graphics are made a) TLD 200/TLD 100 ratio versus X rays energy 2

3 b) Dosimetric factors for each material versus X rays energy Please see Charts 1 and 2 Chart #1 Calibracion de TLD 2211 para Rx Determinación de la energía efectiva 2 TLD 200/ TLD 100 1,5 1 0, ( Kev ) 3

4 Chart # 2 Factores Dosimétricos vs Energía ( msv/ Gu ) 0, , , , , , , , , ( Kev ) TLD 200 TLD 100 DOSIMETER EVALUATION When a professional working in the surroundings requires an evaluation of his personal dosimeter, the corresponding readout is made Once the ratio is determined, it is entried in chart #1 and the effective energy in the X ray field of the professional s working place is thus determined Finally, with the energy value obtained in Chart #1, the dosimetric factors needed for dose assessment in Chart # 2, are obtained Corrective actions due to TLD fading and to variation in the relative response of each crystal are performed in every step of the way Factors showed in Chart #2 allow us to obtain the equivalent personal dose HP(10) in msv for TLD 100 and for TLD 200; those values are averaged and the dose value for each person is therefore obtained The dosimeter referred to in this paper, was used by the National Regulatory Authority (ARN) in several comparisons, always getting satisfactory results The abovementioned comparisons consist in the irradiation of several dosimeters to doses and energies unknown for the participants Some examples of the results obtained are those of year 2006, when the average difference between the 4

5 assessed values and the real values was +14%; the largest difference was close to 15% CONCLUSIONS: This method is rather tedious, and demands lots of work, but it is necessary, considering the dependence on X Rays that TLD dosimeters and the dosimetric factors have It is similar to that performed by labs when using FILM dosimeters Both kinds of dosimeters (TLD and FILM) have advantages and disadvantages which are not being considered in this paper There s another factor that increases the uncertainties of the results It s the Angular dependence that happens in every personal dosimeter Angular dependence occurs because when personal dosimeters used on the body,are not placed facing the beam, they will assess less radiation than when positioned facing it Calibrations used for Charts 1 and 2 refer to latter position This is an unresolved problem International bibliography recommend the usage of two dosimeters instead of one, or the usage of corrective factors (see Bibliography, ref 2) Finally, this methodology for personal dose assessment is not so necessary for gamma ray fields, like Embalse NPP, or in other places using Gamma sources like CO-60 or Cs 137, for TLD effective energy dependence is lower than TLD X Rays dependence BIBLIOGRAPHY: 1HARSHAW- A personnel dosimeter having high x-ray sensitivity TP Radiation Protection Dosimetry, vol 93 Nº3 pp (2001) 5

Radiation Protection Services in R.Macedonia

Radiation Protection Services in R.Macedonia Radiation Protection Services in R.Macedonia Lidija Nikolovska Institute of Public Health Ionizing Radiation Laboratory Skopje, Republic of Macedonia Background - no NPP or research reactors, - limited

More information

First National Intercomparison of Individual Dosimetry of Photon Radiation in Ukraine

First National Intercomparison of Individual Dosimetry of Photon Radiation in Ukraine First National Intercomparison of Individual Dosimetry of Photon Radiation in Ukraine Vadim Chumak National Research Center for Radiation Medicine NAMS Ukraine on behalf of N. Deniachenko b, O. Makarovska

More information

Implementation of personnel dosimetry systems using thermoluminescence dosimetry

Implementation of personnel dosimetry systems using thermoluminescence dosimetry Implementation of personnel dosimetry systems using thermoluminescence dosimetry J. C. Rojas 1, S. O. Benavides 1,2, and H. Olaya 3 1 Radiation Safety Committee, Fundación Clínica Valle del Lili, A.A.

More information

A NEW USER-FRIENDLY EXTREMITY DOSIMETRY SYSTEM DESIGN, COMPARISON, AND TESTING

A NEW USER-FRIENDLY EXTREMITY DOSIMETRY SYSTEM DESIGN, COMPARISON, AND TESTING A NEW USER-FRIENDLY EXTREMITY DOSIMETRY SYSTEM DESIGN, COMPARISON, AND TESTING H. Stadtmann 1, J. Fellinger 2, K.J. Velbeck 2, C. Schmitzer 1 J.E. Rotunda 2 and E. Michler 1 1 Austrian Research Centre,

More information

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS I. Villanueva (1), I. Amor (1), A. Brosed (), R. Villanueva () M. Ginjaume (3), X. Ortega (3), M. Roig (3) (1)

More information

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 CHARACTERIZATION OF TLD-100

More information

Radiologic Units: What You Need to Know

Radiologic Units: What You Need to Know Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

Accreditation of Dosimetry Services

Accreditation of Dosimetry Services Accreditation of Dosimetry Services J. Van Cauteren, E. De Geest, M. Bricoult AV-Controlatom Content History Regulatory requirements Discussion of reference documents/standards Practical implementation

More information

Energy dependent response of Al 2 O 3 and its potential application in personal monitoring.

Energy dependent response of Al 2 O 3 and its potential application in personal monitoring. Energy dependent response of Al 2 O 3 and its potential application in personal monitoring. Nelson, V. K. a, b a Sydney University, Faculty of Health Sciences, Discipline of Medical Imaging and Radiation

More information

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU

More information

Hand Dose in Nuclear Medicine Staff Members

Hand Dose in Nuclear Medicine Staff Members Hand Dose in Nuclear Medicine Staff Members T.M.Taha, Amany.Y.Shahein and R.Hassan* Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority * National Cancer Institute, Cairo

More information

Development of the Nation-Wide Dosimetric Monitoring Network in Ukraine

Development of the Nation-Wide Dosimetric Monitoring Network in Ukraine Development of the Nation-Wide Dosimetric Monitoring Network in Ukraine V.Chumak 1, A.Boguslavskaya 2, A.Musijachenko 1 1 Scientific Center for Radiation Medicine AMS Ukraine, 04050 Melnikova, 53, Kiev,

More information

Performance of several active personal dosemeters in interventional radiology and cardiology

Performance of several active personal dosemeters in interventional radiology and cardiology Performance of several active personal dosemeters in interventional radiology and cardiology S. Chiriotti 1,*, M. Ginjaume 1, E. Vano 2,3, R. Sanchez 3, J.M. Fernandez 2,3, M.A. Duch 1, J. Sempau 1 1.

More information

A Method for Statistical Process Control of Radiation Sterilization Facilities

A Method for Statistical Process Control of Radiation Sterilization Facilities A Method for Statistical Process Control of Radiation Sterilization Facilities 2006 A Method for Statistical Process Control (SPC) of Radiation Sterilization Facilities Scope This document describes a

More information

Model 3500 Manual TLD Reader with WinREMS 3500-W-O-0602 Page 5-1 Operator's Manual

Model 3500 Manual TLD Reader with WinREMS 3500-W-O-0602 Page 5-1 Operator's Manual Model 35 Manual TLD Reader with WinREMS 35-W-O-62 Page 5-1 5. Quality Assurance The procedures in this section help you maintain the accuracy and reliability of your Model 35. The main topics covered are:!

More information

Radiation Monitoring Instruments

Radiation Monitoring Instruments Radiation Monitoring Instruments This set of slides is based on Chapter 4 authored byg. Rajan, J. Izewska of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics: A Handbook for Teachers

More information

OCCUPATIONAL EXPOSURE IN CANDU NUCLEAR POWER PLANT: INDIVIDUAL DOSIMETRY PROGRAM AT CERNAVODA NPP

OCCUPATIONAL EXPOSURE IN CANDU NUCLEAR POWER PLANT: INDIVIDUAL DOSIMETRY PROGRAM AT CERNAVODA NPP OCCUPATIONAL EXPOSURE IN CANDU NUCLEAR POWER PLANT: INDIVIDUAL DOSIMETRY PROGRAM AT CERNAVODA NPP Catalina Chitu, cchitu@cne.ro, Vasile Simionov, vsimionov@cne.ro CNE Cernavoda NPP, No. 2, Medgidiei Str.

More information

Passive dosimeters: summary of users requirements

Passive dosimeters: summary of users requirements Conclusions of the 22 July 2004 RADMON meeting ALICE: - a few 10 th passive dosimeters (probably TLD and RPL); ATLAS: - passive dosimeters only outside the inner detector - a few hundred measurement points;

More information

Variation of Occupational Doses among Subspecialties in Diagnostic Radiology. A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito

Variation of Occupational Doses among Subspecialties in Diagnostic Radiology. A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito Variation of Occupational Doses among Subspecialties in Diagnostic Radiology A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito Biomedical Physics Department, MBC 03 King Faisal Specialist Hospital and Research

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

The need for standardization of dosimetry in experimental radiation biology

The need for standardization of dosimetry in experimental radiation biology The need for standardization of dosimetry in experimental radiation biology Kurt Pedersen University of Wisconsin Medial Radiation Research Center Under the direction of Larry DeWerd, PhD. NCCAAPM Fall

More information

MIXED FIELD PERSONNEL DOSIMETRY, PART I: HIGH TEMPERATURE PEAK CHARACTERISTICS OF THE READER-ANNEALED TLD-600*

MIXED FIELD PERSONNEL DOSIMETRY, PART I: HIGH TEMPERATURE PEAK CHARACTERISTICS OF THE READER-ANNEALED TLD-600* MIXED FIELD PERSONNEL DOSIMETRY, PART I: HIGH TEMPERATURE PEAK CHARACTERISTICS OF THE READER-ANNEALED TLD-600* James C. Liu Stanford Linear Accelerator Center, Bin 48, P.O. Box 4349, Stanford, CA 94309,

More information

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions)

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions) B3-20 Database System for GEX B3 Dosimetry (gamma and/or electron beam versions) and GEX combine to provide a Genesys 20 version of the flexible 21 CFR Part 11 certified database dosimetry system for use

More information

Good manufacturing practices (GMP) utilized on human blood irradiation process

Good manufacturing practices (GMP) utilized on human blood irradiation process Artigos Good manufacturing practices (GMP) utilized on human blood irradiation process Cláudio Boghi Pesquisador do IPEN-USP. cboghi@uol.com.br; cboghi@usp.br Dorlivete M. Shitsuka Professora de Organização

More information

Pille Measurements on ISS (May 2014 March 2015)

Pille Measurements on ISS (May 2014 March 2015) Pille Measurements on ISS (May 2014 March 2015) I. Apáthy 1, A. Hirn 1, S. Deme 1, P. Szántó 1, T. Pázmándi 1 Y. A. Akatov 2, V. V. Arkhangelsky 2, I. Nikolaev 3 1 MTA Centre for Energy Research, Budapest,

More information

Medical Physics 4 I3 Radiation in Medicine

Medical Physics 4 I3 Radiation in Medicine Name: Date: 1. This question is about radiation dosimetry. Medical Physics 4 I3 Radiation in Medicine Define exposure. A patient is injected with a gamma ray emitter. The radiation from the source creates

More information

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Luciana C. Matsushima 1, Glauco R. Veneziani 1, Roberto K. Sakuraba

More information

PERSONNEL MONITORING AND DOSIMETRY POLICIES

PERSONNEL MONITORING AND DOSIMETRY POLICIES PERSONNEL MONITORING AND DOSIMETRY POLICIES All individuals who are required to have their exposure to ionizing radiation monitored must be trained prior to using the source(s) of radiation. The radioactive

More information

The ARN Critical Dosimetry System. Gregori, B.N.; Papadópulos, S.; Cruzate, J.A.; Equillor, H.E. and Kunst, J.J.

The ARN Critical Dosimetry System. Gregori, B.N.; Papadópulos, S.; Cruzate, J.A.; Equillor, H.E. and Kunst, J.J. The ARN Critical Dosimetry System Gregori, B.N.; Papadópulos, S.; Cruzate, J.A.; Equillor, H.E. and Kunst, J.J. Presentado en: 11 th International Congress on the International Radiation Protection Association.

More information

Assessment of radiological hazard and occupational dose to the lens of the eye at the Bruce Power Nuclear Generating Station

Assessment of radiological hazard and occupational dose to the lens of the eye at the Bruce Power Nuclear Generating Station Assessment of radiological hazard and occupational dose to the lens of the eye at the Bruce Power Nuclear Generating Station Andrei Hanu, PhD Senior Scientist Dosimetry Bruce Power Overview Largest operating

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

Sandy Perle. Radiation Litigation: an Expert Witness Perspective

Sandy Perle. Radiation Litigation: an Expert Witness Perspective Sandy Perle Radiation Litigation: an Expert Witness Perspective Hoosier Chapter Health Physics Society Fall Meeting Purdue University, West Lafayette, IN October 2, 2008 Agenda NVLAP and External Dosimetry

More information

Overview of Radiation Monitoring Laboratory Capabilities in Tunisia. Azza Hammou, CNRP, Tunis RMCC-8 Meeting, Amman Jordan JUNE 2013 CONTENT

Overview of Radiation Monitoring Laboratory Capabilities in Tunisia. Azza Hammou, CNRP, Tunis RMCC-8 Meeting, Amman Jordan JUNE 2013 CONTENT Overview of Radiation Monitoring Laboratory Capabilities in Tunisia Azza Hammou, CNRP, Tunis RMCC-8 Meeting, Amman Jordan 17-19 JUNE 2013 CONTENT TUNISIA RADIATION PROTECTION SYSTEM RADITION MONITORING

More information

RADIATION MONITORING DEVICES R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S

RADIATION MONITORING DEVICES R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S RADIATION MONITORING DEVICES 10-526- 1 9 7 R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S DETECTION AND MEASUREMENT OF IONIZING RADIATION Dosimeter Dose-measuring device Two classifications:

More information

COMPARING THE RESPONSES OF TLD 100, TLD 600, TLD 700 AND TLD 400 IN MIXED NEUTRON-GAMMA FIELDS

COMPARING THE RESPONSES OF TLD 100, TLD 600, TLD 700 AND TLD 400 IN MIXED NEUTRON-GAMMA FIELDS 2015 International Nuclear Atlantic Conference - INAC 2015 São Paulo, SP, Brazil, October 4-9, 2015 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-06-9 COMPARING THE RESPONSES OF TLD

More information

Sandy Perle. Radiation Litigation: an Expert Witness Perspective

Sandy Perle. Radiation Litigation: an Expert Witness Perspective Sandy Perle Radiation Litigation: an Expert Witness Perspective Northern California Chapter Health Physics Society Meeting Oakland, CA January 22, 2009 Contact Information Sandy Perle Mirion Technologies,

More information

Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters

Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters Patrícia de Lara Antonio 1*, Rogério M. V. Silva 2, Divanizia do Nascimento Souza 2 and Linda

More information

U.S. DEPARTMENT OF ENERGY LABORATORY ACCREDITATION PROGRAM (DOELAP) FOR PERSONNEL DOSIMETRY SYSTEMS

U.S. DEPARTMENT OF ENERGY LABORATORY ACCREDITATION PROGRAM (DOELAP) FOR PERSONNEL DOSIMETRY SYSTEMS U.S. DEPARTMENT OF ENERGY LABORATORY ACCREDITATION PROGRAM (DOELAP) FOR PERSONNEL DOSIMETRY SYSTEMS F. M. Cummings ^ R. D. Carlson^ R. M. Loesch INTRODUCTION Accreditation of personnel dosimetry systems

More information

IAEA-TECDOC-796. Radiation doses in diagnostic radiology

IAEA-TECDOC-796. Radiation doses in diagnostic radiology IAEA-TECDOC-796 Radiation doses in diagnostic radiology The IAEA does The originating Section FOREWORD It is well recognized that diagnostic radiology is the largest contributor to the collective dose

More information

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH A COMBINATION OF TLD ALBEDO AND SULPHUR ACTIVATION TECHNIQUES FOR FAST NEUTRON PERSONNEL DOSIMETRY

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH A COMBINATION OF TLD ALBEDO AND SULPHUR ACTIVATION TECHNIQUES FOR FAST NEUTRON PERSONNEL DOSIMETRY : ~ : ; EUROPEAN ORGANZATON FOR NUCLEAR RESEARCH 20 1983 TS-RP/ 113/ CF TS DVSONAL REPORT September 1983 A COMBNATON OF TLD ALBEDO AND SULPHUR ACTVATON TECHNQUES FOR FAST NEUTRON PERSONNEL DOSMETRY J.W.N.

More information

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10 GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10 NOTICE: This document is version controlled and was produced as a part of the GEX Information Program which requires that all Series 100

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ Eighth Annual Chapman Phytosanitary Irradiation Forum Dosimetry for phytosanitary applications: dos and don ts Florent KUNTZ June 14 th, 2018 What does a dosimeter measure? Question A dosimeter measures

More information

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Gonzalo Walwyn Salas, Stefan Gutiérrez Lores Center for Radiation Protection and Hygiene,

More information

Calibration of two 90 Sr+ 90 Y dermatological applicators

Calibration of two 90 Sr+ 90 Y dermatological applicators Calibration of two 90 Sr+ 90 Y dermatological applicators Patrícia L. Antonio *, Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP Av. Prof. Lineu Prestes 2242, 05508-000,

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

The Status and Challenges of External Personal Dosimetry

The Status and Challenges of External Personal Dosimetry The Status and Challenges of External Personal Dosimetry Filip Vanhavere SCK-CEN Introduction 20 years ago Pencil dosemeters Film dosemeters Thermoluminescent dosemeters Personal dosimetry: measuring operational

More information

CALIBRATION OF A TLD ALBEDO INDIVIDUAL NEUTRON MONITOR

CALIBRATION OF A TLD ALBEDO INDIVIDUAL NEUTRON MONITOR 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 CALIBRATION OF A TLD

More information

Chapter 4: Radiation Monitoring Instruments

Chapter 4: Radiation Monitoring Instruments Chapter 4: Radiation Monitoring Instruments Set of 107 slides based on the chapter authored by G. Rajan, J. Izewska of the IAEA publication (ISBN 92-0-107304-6): Review of Radiation Oncology Physics: A

More information

High-Dose Standardization Study For γ-rays radiation processing at NIM

High-Dose Standardization Study For γ-rays radiation processing at NIM High-Dose Standardization Study For γ-rays radiation processing at NIM ZHANG Yanli, GAO Junchen, YANG Yuandi. Division of Metrology in Ionizing Radiation and medicine, National Institute of Metrology,

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

RADIATION MONITORING EXPERIMENT USING THERMOLUMINESCENT DOSIMETER FOR THE TR 19 CYCLOTRON AREA IN NUCLEAR RESEARCH INSTITUTE

RADIATION MONITORING EXPERIMENT USING THERMOLUMINESCENT DOSIMETER FOR THE TR 19 CYCLOTRON AREA IN NUCLEAR RESEARCH INSTITUTE RADIATION MONITORING EXPERIMENT USING THERMOLUMINESCENT DOSIMETER FOR THE TR 19 CYCLOTRON AREA IN NUCLEAR RESEARCH INSTITUTE A. STOCHIOIU, L. CRACIUN, F. MIHAI, I. TUDOR Horia Hulubei National Institute

More information

A Novel Dosimetry System for Military Use in Response of Nuclear Emergencies

A Novel Dosimetry System for Military Use in Response of Nuclear Emergencies A Novel Dosimetry System for Military Use in Response of Nuclear Emergencies Mirela Kirr 1, Craig Yoder, PhD 1, Chris Passmore, CHP 1 1 Landauer Inc. 2 Science Road, Glenwood IL 60425-1586 USA Abstract.

More information

DOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE

DOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE DOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE F. MIHAI, A. STOCHIOIU, C. TUCA National Institute for Nuclear Physics and Engineering, P.O.Box

More information

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2. Experiences of the Secondary Standard Dosimetry Laboratory of the CPHR in the implementation and validation of the Postal Quality Audit in Radiotherapy, using TLDs Stefan Gutiérrez Lores 1, Gonzalo Walwyn

More information

DOELAP On-Site Assessment Requirements Checklist Page 1 of 24 Participant:

DOELAP On-Site Assessment Requirements Checklist Page 1 of 24 Participant: DOELAP On-Site Assessment Requirements Checklist Page 1 of 24 General Requirements Y, N, N/A Demonstration of Conformance G.1 Latest version of protocols or procedures G.2 Latest version of dosimeter specifications

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

"Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY *

Research Note CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY * Iranian Journal of Science & Technology, Transaction A, Vol. 32, No. A3 Printed in the Islamic Republic of Iran, 2008 Shiraz University "Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 :

More information

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Clarita Saldarriaga Vargas, Corinne Amalberto, Lara Struelens,

More information

Staff Exposure Monitoring in Interventional Radiology

Staff Exposure Monitoring in Interventional Radiology Conference on Physics in Medicine: From Diagnosis to Treatment "Enhancing safety and quality in radiation medicine". KFMC, Riyadh 7-9 November 2017 Staff Exposure Monitoring in Interventional Radiology

More information

Introduction. Chapter 15 Radiation Protection. Advisory bodies. Regulatory bodies. Main Principles of Radiation Protection

Introduction. Chapter 15 Radiation Protection. Advisory bodies. Regulatory bodies. Main Principles of Radiation Protection Introduction Chapter 15 Radiation Protection Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. F.M. Khan, The Physics of Radiation Therapy, 4th ed., Chapter

More information

RadPro International GmbH i

RadPro International GmbH i RadPro International GmbH i...radiation Protection for the Radiation Professionals... Your provider for complete TLD systems We supply full spectrum of TLD-readers, TLD-materials, dedicated equipment and

More information

Trends in Occupational Exposure in Malaysia

Trends in Occupational Exposure in Malaysia Trends in Occupational Exposure in Malaysia Noriah Mod Ali Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute for Nuclear Technology Research (MINT) Bangi, 43000 KAJANG, Selangor Darul

More information

Conception and design of the National Database of the Individual Dose of the Republic of Cuba

Conception and design of the National Database of the Individual Dose of the Republic of Cuba Conception and design of the National Database of the Individual Dose of the Republic of Cuba Miguel Prendes Alonso *, Maryzury Valdés Ramos, Juan Tomás Zerquera, Daniel Molina Pérez, Gladys López Bejerano1,

More information

Venue: IEEE NSS/MIC/RTSD Conference, Seoul, South Korea, 27 th October 2013 Workshop: NWK3/RD1 Radiation Protection and Dosimetry

Venue: IEEE NSS/MIC/RTSD Conference, Seoul, South Korea, 27 th October 2013 Workshop: NWK3/RD1 Radiation Protection and Dosimetry Venue: IEEE NSS/MIC/RTSD Conference, Seoul, South Korea, 27 th October 2013 Workshop: NWK3/RD1 Radiation Protection and Dosimetry M. Caresana a, A. Sashala Naik a,c, S. Rollet b, M. Ferrarini a,d a Polytechnic

More information

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose Submitted to Radiation Measurements Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose P. Bilski 1, M. Sadel 1, J. Swakon 1, A. Weber 2 1

More information

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS Etieli Camargo da Costa 1, Luiz Antonio Ribeiro da Rosa 2 Delano Valdivino Santos Batista 3 1 Instituto de Radioproteção

More information

The clinical use of OSLD

The clinical use of OSLD The clinical use of OSLD Stephen F. Kry PhD Radiological Physics Center, Imaging and Radiation Oncology Core - Houston IROC Houston The University of Texas MD Anderson Cancer Center, Houston, TX Spring

More information

Radiation Health Protection Manual

Radiation Health Protection Manual Bureau of Medicine and Surgery 7700 Arlington Blvd. CH-1 of 12 Apr 2018 Falls Church, VA 22042 NAVMED P-5055 (FEB 2011) Radiation Health Protection Manual This edition includes CH-1 of 12 Apr 2018. DISTRIBUTION

More information

Chapter 4 RADIATION MONITORING INSTRUMENTS

Chapter 4 RADIATION MONITORING INSTRUMENTS Chapter 4 RADIATION MONITORING INSTRUMENTS G. RAJAN Medical Physics and Safety Section, Bhabha Atomic Research Centre, Mumbai, Maharashtra, India J. I Z E W S K A Division of Human Health, International

More information

Evaluation of Several Dosimeters for Identification of Irradiated Foods Using a 5 MeV Electron Beam

Evaluation of Several Dosimeters for Identification of Irradiated Foods Using a 5 MeV Electron Beam Journal of Health Science, 48(1) 37 41 (2002) 37 Evaluation of Several Dosimeters for Identification of Irradiated Foods Using a 5 MeV Electron Beam Makoto Miyahara,*, a Hitoshi Ito, b Kouji Ueno, c Yutaka

More information

NPL REPORT CIRM 29. Guidelines for the Calibration of Routine Dosimetry Systems for use in Radiation Processing. Peter Sharpe and Arne Miller

NPL REPORT CIRM 29. Guidelines for the Calibration of Routine Dosimetry Systems for use in Radiation Processing. Peter Sharpe and Arne Miller NPL REPORT CIRM 29 Guidelines for the Calibration of Routine Dosimetry Systems for use in Radiation Processing Peter Sharpe and Arne Miller SEPTEMBER 2009 Guidelines for the Calibration of Routine Dosimetry

More information

DEVELOPMENTS IN NEUTRON DOSIMETRY

DEVELOPMENTS IN NEUTRON DOSIMETRY DEVELOPMENTS IN NEUTRON DOSIMETRY C A Perks 1, S Faugoin 1 C N Passmore and M Million 1 1 LANDAUER EUROPE, 33 Avenue du General Leclerc, F-966, Fontenay-aux-Roses CEDEX, France. Telephone: +33 (0) 1 40

More information

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 11137-3 Second edition 2017-06 Sterilization of health care products Radiation Part 3: Guidance on dosimetric aspects of development, validation

More information

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Chien-Hau Chu 1, Uei-Tyng Lin 1, Ngot-Swan Chong 2, Wen-Song Hwang 1,

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

IRRADIATORS. September 11, :00 am-5:00 pm Hock Plaza Auditorium Duke University Medical Center, Erwin Road

IRRADIATORS. September 11, :00 am-5:00 pm Hock Plaza Auditorium Duke University Medical Center, Erwin Road IRRADIATORS Terry Yoshizumi,, PhD September 11, 2006 8:00 am-5:00 pm Hock Plaza Auditorium Duke University Medical Center, Erwin Road Radiation Countermeasures Center of Research Excellence 1 Acknowledgements

More information

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 12749-4 First edition 2015-08-15 Nuclear energy, nuclear technologies, and radiological protection Vocabulary Part 4: Dosimetry for radiation processing

More information

Hampton University Proton Therapy Institute

Hampton University Proton Therapy Institute Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI

More information

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,

More information

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine Radiation physics and radiation protection University of Szeged Department of Nuclear Medicine Radiation doses to the population 1 Radiation doses to the population 2 Sources of radiation 1 Radiation we

More information

Medical dosimetry using Optically Stimulated Luminescence dots and microstar readers

Medical dosimetry using Optically Stimulated Luminescence dots and microstar readers Medical dosimetry using Optically Stimulated Luminescence dots and microstar readers Christopher A Perks *a, Cliff Yahnke b, Marc Million c a LANDAUER EUROPE, Unit 28, Bankside, Station Approach, Kidlington,

More information

Dosimetric Characteristics of LiF:Mg,Cu,Na,Si Phosphor

Dosimetric Characteristics of LiF:Mg,Cu,Na,Si Phosphor Dosimetric Characteristics of LiF:Mg,Cu,Na,Si Phosphor Y. M. Nam 1, J. L. Kim 1, S. Y. Chang 1, B. H. Kim 1 and H. J. Kim 2 1 Health Physics Department, Korea Atomic Energy Research Institute, Taejon 305-353

More information

Radiation Protection in BNCT Patients

Radiation Protection in BNCT Patients Radiation Protection in BNCT Patients J. Kessler 2, H. Blaumann 1, D. Feld 2,, E. Scharnichia 1, I. Levanón 1, C. Fernández 1, G. Facchini 1, M. Casal 3, S. González 2, J. Longhino 1, O. Calzetta 1, B.

More information

QUANTIFICATION OF THE RISK-REFLECTING STOCHASTIC AND DETERMINISTIC RADIATION EFFECTS

QUANTIFICATION OF THE RISK-REFLECTING STOCHASTIC AND DETERMINISTIC RADIATION EFFECTS RAD Conference Proceedings, vol. 2, pp. 104 108, 2017 www.rad-proceedings.org QUANTIFICATION OF THE RISK-REFLECTING STOCHASTIC AND DETERMINISTIC RADIATION EFFECTS Jozef Sabol *, Bedřich Šesták Crisis Department,

More information

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

Y FILMS DOSIMETR Nederland België / Belgique

Y FILMS DOSIMETR Nederland België / Belgique DOSIMETRY FILMS GAFCHROMIC Dosimetry Films The Self-developing Dosimetry Films that Allow You to Go Filmless Do away with the cost and headache of calibrating and maintaining a processor Do away with hazardous

More information

Personal Dosimetry. Radiation Protection Section Department of Medical Physics Colchester Hospital University NHS Foundation Trust

Personal Dosimetry. Radiation Protection Section Department of Medical Physics Colchester Hospital University NHS Foundation Trust Personal Dosimetry Radiation Protection Section Department of Medical Physics Colchester Hospital University NHS Foundation Trust Why do we monitor members of staff? The staff dosimetry service of the

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

Back to Basics - Uncertainty. Presented by: Neill Stanford, CHP Stanford Dosimetry LLC Presented at the 2007 Dosimetry and Records Symposium

Back to Basics - Uncertainty. Presented by: Neill Stanford, CHP Stanford Dosimetry LLC Presented at the 2007 Dosimetry and Records Symposium Back to Basics - Uncertainty Presented by: Neill Stanford, CHP Stanford Dosimetry LLC Presented at the 2007 Dosimetry and Records Symposium Overview From NIST TN 1297 (1994) 2.1 In general, the result

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

Description of the consecutive stages (which took place from December 2002 to July 2003)

Description of the consecutive stages (which took place from December 2002 to July 2003) Radiation Protection Issues in a PET/CT Installation M. Coronado 1, R. Plaza 2, R. Couto 1, MD. Marin 1, C. Huerga 2, J. Coya 1, LM Martin Curto 1, M. Téllez de Cepeda 2 1 Servicio de Medicina Nuclear,

More information

International Journal of Radiation Research, January 2017

International Journal of Radiation Research, January 2017 Downloaded from ijrr.com at : +0330 on Tuesday November th Volume 5, No International Journal of Radiation Research, January 7 Radiation dose to the thyroid, eyes and parotid glands of patients undergoing

More information

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction?

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Poster No.: C-0096 Congress: ECR 2012 Type: Scientific Paper Authors: M. Mazonakis,

More information

DOSIMETRIC STUDY OF MONOCRYSTALLINE SILICON SOLAR CELL

DOSIMETRIC STUDY OF MONOCRYSTALLINE SILICON SOLAR CELL DOSIMETRIC STUDY OF MONOCRYSTALLINE SILICON SOLAR CELL H. DIAB*, A. IBRAHIM**, R. ELMALLAWANY*** *Radiation Dosimetry Laboratory, National Institute for Standards, Egypt, e-mail: hdiab0030037@yahoo.com

More information

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing

More information

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS Bayero Journal of Pure and Applied Sciences, 2(1): 1-5 Received: October, 2008 Accepted: February, 2009 PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS *Sharifat,

More information

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Amanda Bravim, Leticia Lucente Campos Instituto de Pesquisas Energéticas e Nucleares (IPEN / CNEN - SP) Av. Professor

More information