Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

Size: px
Start display at page:

Download "Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter"

Transcription

1 Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter E.K. Chie M.D. 1, S.W. Park M.D. 1, W.S. Kang PhD. 1,2, I.H. Kim M.D. 1,2, S.W. Ha M.D. 1,2, and C.I. Park M.D. 1,2 1 Department of Therapeutic Radiology, Seoul National University College of Medicine Seoul, Korea 2 Institute of Radiation Medicine, Medical Research Center, Seoul National University, Seoul, Korea INTRODUCTION Although radiotherapy is generally used as a local modality, it also can be used as systemic modality in number of occasions. Indications of total body irradiation (TBI) or whole body irradiation are immune suppression prior to bone marrow transplantation (immunologic diseases, aplastic anemia), eradication of malignant cells (leukemia, lymphoma and some solid tumor), and eradication of genetic diseases (Fanconi s anemia, thalasemia major). Though low single dose irradiation of 2-3 Gy has been used for immune suppression of immune diseases and in combination with chemotherapetic agents, such as cyclophosphamides for aplastic anemia, higher doses of fractionated irradiation is required to eradicate malignant cells as well as to overcome graft injection in allogenic bone marrow transplantation. TBI also plays role in providing grafting space in bone marrow, which heightens successful transplantation rate. Single high dose therapy of 9- Gy was initially which was later replaced with fractionated and/or low dose rate therapy to lessen the treatment limiting complications, such as radiation pneumonitis. In cases of fractionated regimen, adjustments could be done in following fraction for more homogenous dose delivery by accurately estimating the exposed dose to organ of interest (1,2). Methods of TBI can be divided into parallel-opposed lateral technique and parallel opposed anteroposterior/posteroranterior (AP/PA) technique, although there are many other different methods. Advantages of parallel-opposed lateral technique are natural lung compensation with arms and more comfortable patient position. Although parallel opposed anteroposterior/posteroranterior (AP/PA) technique does not require additional compensator as patient thickness along long axis is minimized, additional lung block to reduce lung dose and additional irradiation to increase the doses to chest wall are necessary and also requires long source-to patient distance (1). TBI setup represents a very irregular and extended field. It is essential to deliver homogenous radiation dose over whole body, which requires careful setup design to minimize the possible errors. Verification of total body irradiation dosimetry techniques at many institutions in United States revealed that quality control and assurance is essential for corresponding institute, facility, and treatment setup (3). Reported range of error of dose homogeneity with methods incorporating water phantom and entrance dose was ±7%, and with methods using humanoid phantom was ±% (4-6). American Association of Physicists in Medicine (AAPM) recommended error range of ~ +5% for dose homogeneity in case of total body irradiation in the report 17 (7). Frequently used in vivo dosimeters are thermoluminescent dosimeter (TLD) and diode. TLD was used more frequently in multiple dosimetry due to low cost, but with development of multi-channel diode electrometer, use of diodes is increasing (4). Small water phantom was frequently used in confirmation of dose distribution. But, as systemic error in range of ±4% was reported in case of TBI, use of humanoid phantom is recommended (3). Dose distribution and homogeneity for Co- teletherapy has been reported previously from this department (8). But, dose homogeneity and quality assurance program using linear accelerator has not been settled. Recently, some institution in Korea have reported dose profiles in TBI setup (9-11). The objective of this study was to analyze and confirm the accuracy and the homogeneity of the treatment setup, the parallel opposed lateral technique, currently used in Seoul National University Hospital. METHODS and MATERIALS 1. Patient Dosimetry Surface dose data, measured with a thermoluminescent dosimeter in 8 patients among patients, who were given total body irradiation with the parallel-opposed lateral technique between September 1996 to August 1998, at Seoul National University Hospital was analyzed. Age distribution was from 15 to years old with median value of 21. Five patients were male and three patients were female. Six patients had acute leukemia, one had aplastic anemia, and the other had high-grade non-hodgkin s lymphoma. MV x-rays from Clinac 2C (Varian Associates, USA) were used. Seven patients were irradiated with daily 3Gy/fraction for 4 fractions totaling 12Gy, while one patient was treated with daily 3.5Gy/fraction for 3 fractions totaling.5gy. Lowest output rate of MU/min from Clinac 2C was chosen to keep dose rate under.1gy/min (.95+/-.57 Gy/min). Source-to-isocenter distance was 274cm, resulting in field size of 9.6*9.6cm 2 with collimator opening of *cm 2. Parallel-opposed lateral technique was used with patient in knee-chest position. Acrylic plate with thickness of 1cm was placed cm in front of treatment couch to 1

2 function as a beam spoiler. Thickness of head, neck, shoulder, axilla, abdomen (umbilicus plane), hip, thigh, and ankle level was measured. Tissue deficit compensator was constructed to compensate for the missing tissue enabling more homogenous dose distribution. Thickness of the compensator was calculated from equation below. T Pb = (T R T M )*.7/ 2, where T Pb is thickness of tissue deficit compensator made of lead; T R is the thickness of the prescribed point (umbilicus plane for the patient); T M is the thickness of plane of interest; and.7 is compensatory ratio of lead to soft tissue for MV x-rays from Clinac 2C used at Department of Therapeutic Radiology, Seoul National University Hospital. Dose was prescribed to the umbilicus level, which was calculated from monitor unit conversion table (12). TLD was placed parallel to skin surface on both sides of head, neck, axilla, thigh, and ankle level for each treatment session to measure both entrance and exit dose. 2. Phantom Dosimetry Humanoid phantom (Huestis, USA) was used to simulate the patient set-up. Identical method was used to construct tissue deficit compensator for the phantom. Surface and midline doses of head, neck, axilla, abdomen, and hip level were similarly measured. Contrary to patient setup, Humanoid phantom was in erect position as knee-chest position was not possible. Cylinder shaped water phantom with diameter of 8cm and length of 32cm was used to simulate the upper arm. X-ray output used was MU/min resulting in dose rate of.48gy/min. Measurement was repeated for eight times. 3. TLD Dosimetry Chip shaped TLD- (Harshaw Chemie BV, the Netherlands) containing 7.5% Li-6 and 92.5% Li-7 and 28mg of LiF with the dimension of 3.1*3.1*.89 mm 3 and maximum thermoluminescent temperature of 195 C was used. TLD s with reproduction range within 3% was chosen for the experiment. TLD s were heated at C for one hour and at C for another one hour with TLD oven (PTW-Freiberg, Germany) before every use. TLD reading was done with Harshaw/QS TLD system reader (Solon Technologies, USA) at temperature range of ~ C for seconds after preheating at C for seconds. 4. Data Analysis Mean value of measured surface dose and ratio to prescribed dose was calculated from the patient data. Relative dose for points of interest and ratio of surface dose to mid-depth dose as well as mean value of measured surface dose and mid-depth dose were calculated from the phantom data. SAS program (The SAS program for Windows version 6.12, SAS Institute Inc., USA) was used for the statistical analysis. RESULTS 1. Measured Surface Dose in Patient 2

3 Patient Fig.1 Radiation doses measured with TLD chips on the surface of the patients relative to mid-depth dose prescribed to umbilicus level in 3 measurements for case #7 or 4 measurements for all others per treatment (Mean±S.D.). Distribution of surface doses and standard deviations relative to prescribed dose for individual patient is shown in Fig. 1. Mean measured surface dose and standard deviation relative to prescribed dose from every points of every patient was 95.7±7.2%. Measured surface doses and standard deviations relative to prescribed dose for the head, neck, axilla, thigh, and ankle level were 91.3±7.8%, 98.3±7.5%, 95.1±6.3%, 98.3±5.5%, and 95.3±6.3%, respectively in patients (Fig. 2). 2. Measured Dose in Phantom 1) Measured Surface Dose Measured surface doses and standard deviations relative to prescribed dose for the head, neck, axilla, abdomen, and thigh level were 85.±4.%, 86.6±5.8%, 83.9±4.9%, 94.8±2.8%, respectively (Fig. 3). 2) Measured Mid-depth Dose Measured mid-depth doses and standard deviations relative to prescribed dose for the head, neck, axilla, abdomen, and thigh level were 96.6±2.2%, 95.3±3.2%,.4±1.9%,.±3.1%,.5±2.2%, respectively (Fig. 4). Head Neck Axilla Thigh Ankle Fig.2 Radiation doses measured with TLD chips for different levels on the surface of the patients relative to midline dose prescribed to umbilicus level in 8 patients (Mean±S.D.). Head Neck Axilla Umbilicus Hip Fig.3 Radiation doses measured with TLD chips for different levels on the surface of the Humanoid phantom relative to mid-depth dose prescribed to umbilicus level of 8 different measurements (Mean±S.D.). 3) Measured Dose to Mid-depth Dose Ratio The surface to mid-depth dose conversion ratios and standard deviations obtained from the phantom 3 Head Neck Axilla Umbilicus Hip Fig.4 Radiation doses measured with TLD chips for different levels in the mid-depth of the Humanoid phantom relative to mid-depth dose prescribed to umbilicus level of 8 different measurements (Mean±S.D.).

4 study were 1.14±.6, 1.±.9,.96±.5, 1.6±.6,.95±.2 for head, neck, axilla, abdomen, and hip level, respectively. 3. Converted Mid-depth Dose in Patient The mid-depth doses and standard deviations of the head, neck, axilla, thigh, and ankle in patients estimated from the surface to mid-depth dose conversion ratios were 3.4±9.%, 7.8±.5%, 91.1±6.1%, 93.8±4.5%, and 4.5±9.3%, respectively (Fig. 5). DISCUSSION Many different treatment setups have been devised to deliver homogenous dose to whole body including skin surface for respective institution with different therapeutic facilities and treatment conditions. The effects of scattered radiation and inhomogeneity of lateral radiation quality, morphological difference and thickness difference of patient, inhomogeneity and surface irregularity of tissue, and tissue deficit present special problem. Different types of compensation and quality assurance measures had been devised to overcome these problems and deliver homogenous dose to the patient. Despite necessity of tissue deficit compensator and beam spoiler, the parallel-opposed lateral technique, currently used in Seoul National University Hospital, has an advantage of homogenous lung dose delivery without additional compensator due to natural compensation with arms (1,2). As treatment settings of TBI differ from institute to institute, customized methods of quality assurance is mandatory. It has been reported that dose inhomogeneity of over % could result from omitting regular quality assurance (3). Some of the institutions in Korea have reported on the dose distribution with TBI. Most used dose analysis from phantom, omitting the actual patient data analysis, with dose range of ±8% (9-11). 4

5 Head Neck Axilla Thigh Ankle Fig.5 Estimated mid-depth doses of patients from the surface to mid-depth ratio obtained from the Humanoid phantom (Mean±S.D.). Surface dose range of 8.7~-1.7% for the patients from this study was within satisfactory range, but converted mid-depth dose distribution showed error range of 8.9~+7.8%, which failed to fall in recommended dose range. This difference is much bigger, taking the dose range of 1.2~+3.% obtained from water phantom dosimetry to acquire monitor unit conversion table, which is expected dose range, into account. However, middepth dose conversion ratio should be modified and better fitted, considering the thickness difference between the patient and the phantom. Additionally, repetition limitation of 3% of TLD s, allocation error of TLD placement, and output of linear accelerator could, although should be minute, attribute to the widening of the dose range. Allocation error could be a cause of wider dose range for surface dose distribution than mid-depth dose distribution. Table 1. Comparison of Thickness for the Patients and the Humanoid Phantom Patient Humanoid phantom Site Thickness (cm) Ratio* (Mean±S.D.) (Mean±S.D.) Thickness (cm) Ratio Head 16.1±.6.61± Neck.6±1.7.39± Shoulder 43.± ± Axilla 28.9± ± Umbilicus 27.3± Hip 34.1± ± * ratio of the thickness relative to the reference level 5

6 Deviation of dose range from the expected value for points of interest could be attributed to following causes. First, more than % decrease of relative mid-depth dose for axilla to prescribed dose could result from thickness difference between patients and the phantom. As shown in table 1, relative thickness difference for the axilla region to prescribed umbilicus is quite significant. This is largely due to the fact that shape of the phantom follows that of Caucasian. While difference of thickness of head and neck region could be compensated using tissue-deficit compensators, use of this type of compensator is not possible for shoulder and axilla region. This directly leads to difference in absorbed dose. Converting the thickness difference to relative rate to that of prescribed umbilicus level results in % increase in thickness for the phantom. By plotting into the percent depth dose profile for TBI used at this institution, this thickness difference of % could cause dose difference of % (Fig 6). This in turn results in % dose difference in mid-depth dose relative to the prescribed dose. Surface dose used in this study is sum of the entrance dose and the exit dose. The impact on the entrance dose is minimal with thickness difference, while the exit dose is more readily influenced by the difference. This could lead to less decrease of surface dose than the mid-depth dose. It has been reported that calculating the mid-depth dose from the entrance dose and the exit dose could also result in % decrease. Some have attributed lower incidence of radiation pneumonitis with parallel-opposed lateral technique to this decrease in radiation dose (4). For more accurate mid-depth dose calculation, construction of humanoid phantom with skeletal composition of corresponding trait, which in this case would be Korean, seems necessary. Improving the dose estimates by measuring percent depth dose (PDD) or tissue maximum ratio (TMR) profiles for individual thickness and offaxis ratio (OAR) for corresponding points of interest have been reported. But, the resulted range of dose estimates tend to be narrower then those obtained from actual measurement (3-6). Percent depth dose (PDD) Depth(cm) Fig.6 Beam data for TBI setup at Seoul National University Hospital using MV x-ray from Clinac 2C (Varian Associates, USA) with phantom size 25*25* cm 3, SSD 274cm, and field size 4.9*4.9 cm 2, measured with Farmer type chamber Secondly, decrease in surface dose of neck region could not be attributed to overcompensation of tissue-deficit compensator, as mid-depth dose was not decreased. Shank have reported that thickness of more than 2cm is required to obtain adequate scattering effect for 6~18 MV x-rays (13). Decrease in thickness of neck region relative to umbilicus region could lessen the scattering effect of the scattering acrylic plate. This decrease in surface dose could be emphasized in the patients, as relative thickness is smaller than that of the phantom. Furthermore, there was 5% increase in mid-depth dose of neck region, estimated from the ratio of surface dose to mid-depth dose obtained from the phantom data. Penetrating power of high energy x-rays above 6~8 MV from the linear accelerator decreases as the distance from the isocenter to point of interest is lengthened. Erect position of the phantom has longer isocenter-to-point distance for the neck region than the knee-chest position of the patient. This increase in distance could be a cause of decrease in surface to mid-depth dose ratio for the phantom. Thirdly, decrease in surface dose for the head region could be attributed to the identical causes as the neck region. Lastly, as for the ankle region, conversion ratio of the neck region was adopted due to similarity in depth ratio, which results in identical deviation of dose profile. As mentioned in the introduction, AAPM recommended dose variation range of ~+5% to the prescribed point, which is mid-depth dose of umbilicus region, for the TBI treatment setup in their report number 17 (7). Measured surface dose range of 8.7~-1.7% and converted mid-depth dose range of 8.9~+7.8% for the 6

7 patients is somewhat superior to reported dose range of ±% from other institutions (4-6). Further more, accounting the possible increase in surface to mid-depth conversion ratio of the head and neck region, current TBI setup utilized at this institution should be superior to the experiment derived dose range values. Due to factors such as scattered irradiation, simple calculation methods could not derive at actual delivered dose to the points of interest or homogeneity of dose distribution. Due to the difference in the treatment facilities and treatment setups, simple adaptation of calculation methods devised from the literature should be avoided. Current method of TLD measurement for each and every session of TBI employed at this department enables possible dose modification in the following sessions. This result in more accurate and homogenous treatment, which in turn heightens the treatment effect and decrease possible complications. REFERENCES 1. H.Lin and R.E.Drzymala, Total body and hemibody irradiation. In C.A.Perez and L.W.Brady eds. Principles and practices of radiation oncology, 3rd ed. Lippincott-Raven Publishers, Philadelphia (1997) 2. B.Shank, Total body irradiation. In S.A.Leibel and T.L.Phillips TL eds. Textbook of radiation oncology. WB Saunders Company, Philadelphia (1998) 3. T.H.Kirby, W.F.Hanson, and D.A.Cates, Verification of total body photon irradiation dosimetry techniques. Med Phys 15, (1988) 4. J.R.Greig, R.W.Miller, and P.Okunieff, An approach to dose measurement for total body irradiation. Int J Radiat Oncol Biol Phys 36, (1996) 5. P.C.Lee, J.M.Sawicka, and G.P.Glasgow, Patient dosimetry quality assurance program with a commercial diode system. Int J Radiat Oncol Biol Phys 29, (1994) 6. E.B.Podgorsak, C.Pla, M.Evans, and M.Pla, The influence of phantom size on output factor, peak scatter factor, and percentage depth dose in large-field photon irradiation. Med Phys 12, (1985) 7. J.Van Dyk, J.M.Galvin, G.P.Glasgow, et al, The physical aspect of total and half body photon irradiation. A report of Task Group 29 Radiation Therapy Committee, American Association of Physicists in Medicine. AAPM Report No. 17 (1986) 8. W.S.Kang, Dose distribution of total body irradiation with Co-. Korean J Med Phys 2, 9- (1991) 9. D.R.Choi, I.B.Choi, K.M.Kang, K.S.Shin, and C.C.Kim, Total body irradiation technique: basic data measurements and in vivo dosimetry. J Korean Soc Ther Radiol 12, (1994). S.J.Ahn, W.S.Kang, S.J.Park, T.K.Nam, W.K.Chung, and B.S.Nah, The dosimetric data of MV linear accelerator photon beam for total body irradiation. J Korean Soc Ther Radiol 12, (1994) 11. S.J.Park, W.K.Chung, S.J.Ahn, T.K.Nam, and B.S.Nah, Utilization of tissue compensator for uniform dose distribution in total body irradiation. J Korean Soc Ther Radiol 12, (1994) 12. Dept. of Therapeutic Radiology. Seoul National University Hospital. Dosimetric databook for TBI, SRS, and FSRT (unpublished) 13. B. Shank B, Techniques of magna-field irradiation. Int J Radiat Oncol Biol Phys 9, (1983) 7

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * Romanian Reports in Physics, Vol. 66, No. 1, P. 157 165, 2014 EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * M. SPUNEI 1,2, M. MIHAI 3, I. MĂLĂESCU 1 1 West

More information

Dr. Rossella Vidimari. Department of Medical Physics Ospedale Maggiore A.S.U.I.T.S Ospedali Riuniti di Trieste

Dr. Rossella Vidimari. Department of Medical Physics Ospedale Maggiore A.S.U.I.T.S Ospedali Riuniti di Trieste Dr. Rossella Vidimari Department of Medical Physics Ospedale Maggiore A.S.U.I.T.S Ospedali Riuniti di Trieste School on Medical Physics for Radiation Therapy 27 March 7 April 2017 Clinical indications

More information

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-861.Volume 6, Issue Ver. II (May-Jun. 201), PP 2- A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

More information

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS Etieli Camargo da Costa 1, Luiz Antonio Ribeiro da Rosa 2 Delano Valdivino Santos Batista 3 1 Instituto de Radioproteção

More information

JRPR. Prediction of Midline Dose from Entrance and Exit Dose Using OSLD Measurements for Total Body Irradiation. Original Research.

JRPR. Prediction of Midline Dose from Entrance and Exit Dose Using OSLD Measurements for Total Body Irradiation. Original Research. Journal of Radiation Protection and Research 2017;42(2):77-82 pissn 2508-1888 eissn 2466-2461 Prediction of Midline Dose from Entrance and Exit Dose Using OSLD Measurements for Total Body Irradiation Chang

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS Romanian Reports in Physics, Vol. 63, No. 3, P. 700 706, 2011 DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS ZOE GHITULESCU 1, ANA STOCHIOIU 2, MIHAI DUMITRACHE 3, 1 CNCAN- National

More information

First application of hemi-body electron beam irradiation for Kaposi sarcoma at the lower extremities

First application of hemi-body electron beam irradiation for Kaposi sarcoma at the lower extremities JBUON 2018; 23(1): 268-272 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com SHORT COMMUNICATION First application of hemi-body electron beam irradiation for Kaposi

More information

Patient dosimetry for total body irradiation using single-use MOSFET detectors

Patient dosimetry for total body irradiation using single-use MOSFET detectors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Patient dosimetry for total body irradiation using single-use MOSFET detectors Tina Marie Briere, 1 Ramesh Tailor, 1 Naresh Tolani,

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules Journal of the Egyptian Nat. Cancer Inst., Vol. 8, No., June: -, 00 Spinal Cord Doses in Palliative Lung Radiotherapy Schedules HODA AL-BOOZ, FRCR FFRRCSI M.D.* and CAROL PARTON, Ph.D.** The Departments

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction?

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Poster No.: C-0096 Congress: ECR 2012 Type: Scientific Paper Authors: M. Mazonakis,

More information

Extracranial doses in stereotactic and conventional radiotherapy for pituitary adenomas

Extracranial doses in stereotactic and conventional radiotherapy for pituitary adenomas JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 2, SPRING 2006 Extracranial doses in stereotactic and conventional radiotherapy for pituitary adenomas Thomas Samuel Ram, a Paul B. Ravindran,

More information

Protection of the contralateral breast during radiation therapy for breast cancer

Protection of the contralateral breast during radiation therapy for breast cancer Protection of the contralateral breast during radiation therapy for breast cancer Edgardo Garrigó a*, Alejandro Germanier b, Silvia Zunino a a Instituto Privado de Radioterapia, Ob Oro 423 (5000) Córdoba,

More information

The rotary dual technique for total skin irradiation in the treatment of mycosis fungoides a description of the applied method

The rotary dual technique for total skin irradiation in the treatment of mycosis fungoides a description of the applied method Original Paper Received: 2005.06.14 Accepted: 2005.11.14 Published: 2006.02.27 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data Interpretation E Manuscript Preparation

More information

A translational couch technique for total body irradiation

A translational couch technique for total body irradiation JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 4, FALL 2001 A translational couch technique for total body irradiation Mehrdad Sarfaraz, Cedric Yu, D. J. Chen, and Leon Der Radiation Oncology

More information

Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution

Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution NUKLEONIKA 2007;52(4):153 158 ORIGINAL PAPER Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution Tomasz

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

3D heterogeneous dose distributions for total body irradiation patients

3D heterogeneous dose distributions for total body irradiation patients JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 3, summer 2011 3D heterogeneous dose distributions for total body irradiation patients Marie-Claude Lavallée, 1,2a Sylviane Aubin, 1 Marie

More information

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Multilayer Gafchromic film detectors for breast skin dose determination in vivo INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Luciana C. Matsushima 1, Glauco R. Veneziani 1, Roberto K. Sakuraba

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Electron therapy Class 3: Clinical procedures

Electron therapy Class 3: Clinical procedures Electron therapy Class 3: Clinical procedures Laurence Court lecourt@mdanderson.org Reference: Faiz M. Khan, The Physics of Radiation Therapy Slide acknowledgements: Karl Prado, Rebecca Howell, Kent Gifford,

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate

More information

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri

More information

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System Original Article Research in Oncology 2018; Vol. 14, No. 1: 34-38. DOI: 10.21608/resoncol.2018.3026.1047 Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

More information

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG

More information

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Tagreed AL-ALAWI Medical Physicist King Abdullah Medical City- Jeddah Aim 1. Simplify and standardize

More information

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health

More information

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams Chapter 12 Treatment Planning Combination of Beams Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther Outline Combination

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,

More information

Experimental Evaluation of Depth Dose by Exit Surface Diode Dosimeters for Off-Axis Wedged Fields in Radiation Therapy

Experimental Evaluation of Depth Dose by Exit Surface Diode Dosimeters for Off-Axis Wedged Fields in Radiation Therapy Iranian Journal of Medical Physics Vol. 12, No. 4, Autumn 201, 262-270 Received: August 1, 201; Accepted: March 12, 2016 Original Article Experimental Evaluation of Depth Dose by Exit Surface Diode Dosimeters

More information

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Sang Hyun Cho and Geoffrey S. Ibbott Department of Radiation Physics The

More information

Quality assurance in external radiotherapy

Quality assurance in external radiotherapy Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance

More information

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed

More information

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics

More information

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS Romanian Reports in Physics, Vol. 64, No. 3, P. 728 736, 2012 MEDICAL PHYSICS IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS G. VASILE 1,3, MADALINA VASILE 2,3, O.G. DULIU 3 1 Center of

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence SCIENCE & TECHNOLOGY To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence Alburuj R. Rahman*, Jian Z. Wang, Dr. Z. Huang, Dr. J. Montebello Department of

More information

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad

More information

3D Conformal Radiation Therapy for Mucinous Carcinoma of the Breast

3D Conformal Radiation Therapy for Mucinous Carcinoma of the Breast 1 Angela Kempen February Case Study February 22, 2012 3D Conformal Radiation Therapy for Mucinous Carcinoma of the Breast History of Present Illness: JE is a 45 year-old Caucasian female who underwent

More information

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES MELISSA C. MARTIN, M.S., FACR, FAAPM PRESIDENT AAPM - 2017 PRESIDENT - THERAPY PHYSICS INC., GARDENA, CA MELISSA@THERAPYPHYSICS.COM AAPM Spring

More information

Neutron dose evaluation in radiotherapy

Neutron dose evaluation in radiotherapy Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Radiation therapy with a linear accelerator (LINAC) Photoneutron production in accelerator head

More information

Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy

Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy MB. Tavakoli 1, M. Saeb 2, H. Emami 3 1 MB Tavakoli, Department of Medical Physics

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 1, WINTER 2000

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 1, WINTER 2000 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 1, WINTER 2000 Clinical implementation of the AAPM Task Group 36 recommendations on fetal dose from radiotherapy with photon beams: A head

More information

Ali S. Meigooni, Kai Dou, Navid J. Meigooni, Michael Gnaster, Shahid Awan, Sharifeh Dini, and Ellis L. Johnson

Ali S. Meigooni, Kai Dou, Navid J. Meigooni, Michael Gnaster, Shahid Awan, Sharifeh Dini, and Ellis L. Johnson Dosimetric Characteristics of a Newly Designed Grid Block for Megavoltage Photon Radiation and Its Therapeutic Advantage Using A Linear Quadratic Model Ali S. Meigooni, Kai Dou, Navid J. Meigooni, Michael

More information

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose Submitted to Radiation Measurements Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose P. Bilski 1, M. Sadel 1, J. Swakon 1, A. Weber 2 1

More information

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem Original Article Treatment Planning Evaluation of Sliding Window and Multiple Static Segments Technique in Intensity Modulated Radiotherapy for Different Beam Directions Teyyiba Kanwal, Muhammad Khalid,

More information

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D. Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation

More information

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging Iran. J. Radiat. Res., 2008; 6 (2): 83-87 Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging B. Azadbakht 1, M.H. Zahmatkesh 2 *, k. Hadad 1, S. Bagheri 2

More information

The use of the TLD-100 for quality assurance in Total Body Irradiation (TBI)

The use of the TLD-100 for quality assurance in Total Body Irradiation (TBI) The use of the TLD-100 for quality assurance in Total Body Irradiation (TBI) ARNIE VERDE NOLASCO Dept. de Engenharia Nuclear/UFMG Belo Horizonte - MG BRAZIL arnienolasco@hotmail.com LUIZ OLIVEIRA FARIA

More information

5th ADAMAS Workshop at GSI December 15-16, 2016, Darmstadt, Germany

5th ADAMAS Workshop at GSI December 15-16, 2016, Darmstadt, Germany Evaluation of 3D diamond detectors for application in medical radiation dosimetry K. Kanxheri (1,2), L. Servoli (2), C. Zucchetti (5), A. C. Dipilato (5), M. Iacco (5), S. Lagomarsino (3,4), A. Morozzi

More information

Risk of a second cancer after radiotherapy

Risk of a second cancer after radiotherapy Risk of a second cancer after radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Medical radiological procedures worldwide 2.5 billion diagnostic radiological examinations 78%

More information

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters 133 Clinical Radiation Oncology Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters Bhaskar Mukherjee 1,2, Carolina Lina 1 Vladimir Mares 3 1Westdeutsches

More information

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac.

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac. IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-4861.Volume 9, Issue 5 Ver. IV (Sep. - Oct. 2017), PP 46-52 www.iosrjournals.org Quanitative and qualitative analysis of Image Quality and Imaging

More information

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS 6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types

More information

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article. Original Article PMP Progress in Medical Physics 29(1), March 218 https://doi.org/1.14316/pmp.218.29.1.16 pissn 258-4445, eissn 258-4453 Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers

More information

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF Nuclear Science and Techniques 24 (2013) 060102 A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF LI Zhaobin SHI Zeliang ZHANG Qing WANG Yong FU Shen * The 6 th People s Hospital

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

JBUON, (2018) 23 (1) ISSN

JBUON, (2018) 23 (1) ISSN Platoni, Kalliopi and Diamantopoulos, Stefanos and Dilvoi, Maria and Delinikolas, Panagiotis and Kypraiou, Efrosyni and Efstathopoulos, Efstathios and Kouloulias, Vassilios (2018) First application of

More information

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results A.Piermattei 1, L. Azario 1, S. Cilla 2, A.Fidanzio 1, F.Greco 1, M.T.Russo 3, S. Zucca 4 1 Istituto di Fisica e Unità

More information

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC) Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 http://dx.doi.org/10.14316/pmp.2015.26.4.223 Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the

More information

Calibration of two 90 Sr+ 90 Y dermatological applicators

Calibration of two 90 Sr+ 90 Y dermatological applicators Calibration of two 90 Sr+ 90 Y dermatological applicators Patrícia L. Antonio *, Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP Av. Prof. Lineu Prestes 2242, 05508-000,

More information

Slide 1. Slide 2. Slide 3. Introduction of INTRABEAM IORT. Disclosure. Contents. I have nothing to disclose.

Slide 1. Slide 2. Slide 3. Introduction of INTRABEAM IORT. Disclosure. Contents. I have nothing to disclose. Slide 1 AAPM 2017 Introduction of INTRABEAM IORT Hualin Zhang, Ph.D. Dept of Radiation Oncology, Northwestern University, Northwestern Memorial Hospital, Chicago, Illinois August 2, 2017 Slide 2 Disclosure

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET University of Michigan Department of Radiation Oncology Division of Radiation Physics Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET Resident: Rotation staff mentor/ advisor: _Scott Hadley,

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

On the use of bolus for pacemaker dose measurement and reduction in radiation therapy

On the use of bolus for pacemaker dose measurement and reduction in radiation therapy Received: 8 February 2017 Revised: 16 October 2017 Accepted: 23 October 2017 DOI: 10.1002/acm2.12229 RADIATION ONCOLOGY PHYSICS On the use of bolus for pacemaker dose measurement and reduction in radiation

More information

Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center

Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center All students accepted into the Medical Dosimetry Graduate Certificate Program

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

IMRT point dose measurements with a diamond detector

IMRT point dose measurements with a diamond detector IMRT point dose measurements with a diamond detector Erin Barnett, Marc MacKenzie, B. Gino Fallone Department of Physics, University of Alberta, and Department of Medical Physics, Cross Cancer Institute,

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

INTRODUCTION. Material and Methods

INTRODUCTION. Material and Methods Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University

More information

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams

More information

Dosisverifikation mit Delta 4 Discover während der Behandlung

Dosisverifikation mit Delta 4 Discover während der Behandlung Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

Radiochromic film dosimetry in water phantoms

Radiochromic film dosimetry in water phantoms INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MP (CLINICAL ONCOLOGY) PART I EXAMINATION - AUGUST Time : p.m p.m.

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MP (CLINICAL ONCOLOGY) PART I EXAMINATION - AUGUST Time : p.m p.m. POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MP (CLINICAL ONCOLOGY) PART I EXAMINATION - AUGUST 2015 Date :- 24 th August 2015 PAPER I Time :- 2.00 p.m. - 4.15 p.m. If the examiners cannot

More information

Commissioning and Radiobiology of the INTRABEAM System

Commissioning and Radiobiology of the INTRABEAM System Commissioning and Radiobiology of the INTRABEAM System Susha Pillai and Junan Zhang Scheme INTRABEAM System. Physics Commissioning, QA, and Radiation Protection Radiobiology 1 Disclosure OHSU is an INTRABEAM

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

A simple technique for craniospinal radiotherapy in the supine position

A simple technique for craniospinal radiotherapy in the supine position Radiotherapy and Oncology 78 (2006) 217 222 www.thegreenjournal.com Craniospinal radiotherapy A simple technique for craniospinal radiotherapy in the supine position William A. Parker a, *, Carolyn R.

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS SB Crowe, E Whittle, C Jones, T Kairn SMALL FIELD DOSIMETRY The measurement of small field tissue-maximum ratio (TMR) and percentage depth dose (PDD)

More information