Bell in 1903 was first suggested implanting. First Report on Study of Quality Control Program of High Dose Rate (Hdr) Brachytherapy.

Size: px
Start display at page:

Download "Bell in 1903 was first suggested implanting. First Report on Study of Quality Control Program of High Dose Rate (Hdr) Brachytherapy."

Transcription

1 Original Article First Report on Study of Quality Control Program of High Dose Rate (Hdr) Brachytherapy. Adhikari KP*, Dhakal BP**, Prasiko G*** * Sr. Medical Physicist & Radiation Safety Officer, NAMS, Bir Hospital, Kathmandu. ** Research Student, Central Department of Physics, Tribhuvan University, Kirtipur. *** Consultant Radiation Oncologists, Bhaktapur Cancer Center, Bhaktapur. ABSTRACT introduction: To evaluate the accuracy of radiation dose delivered to the cancer patients by using HDR Brachytherapy and to know the safety of radiation level for working personnel and public. MethodE: The observation of sweet spots (maximum dose distribution) on different dates by using well chamber (SI HDR 1000) and electrometer (SI CDX 2000). On the basis of the observed sweet spot, Air Kerma Strength of the source is determined. Beside this includes the conformation of step size, radioactivity of the nuclides, safety measures of the machine. A careful radiation survey has been undertaken around the brachytherapy by using well calibrated radiation survey meter (FAG, FH 40F1). Results: Air Kerma strength in newly installed source exhibit small variation but within the limit. The step size has standard deviation 0.05 with the planned step size. The measurement of radiation level around brachytherapy shows the level is within the criteria. Conclusion: The QA test shows that status of brachytherapy and its components are functioning well. Radiation dose delivered to the cancer patients are within planned dose. Key Words: air kerma strength, brachytherapy, quality control, radiation protection INTRODUCTION Bell in 1903 was first suggested implanting the radioactive source directly into a tumor. Radioactive seeds are carefully placed inside of the cancerous tissue and positioned in a manner that will attack the cancer most efficiently 1. This Greek word for short distance (brachy) is presently known as "Brachytherapy". Depending on the lesion being treated, brachytherapy can be practiced in following ways: Intracavitary, Interstitial, Surface moulds and Intraluminal. Brachytherapy is used in the treatment of various kinds of cancer, including prostate, breast, cervical, and ocular. Brachytherapy is also used to treat coronary artery disease to prevent restenosis after angioplasty. For a safe and accurate dose delivery using brachytherapy many aspects need to be considered carefully. Furthermore the general safety aspects for the patient, the personnel, and Correspondence : Kanchan P. Adhikari kanchanadhikari@gmail.com the environment are important issues. In order to ensure the optimal treatment of patients much effort is required during the commissioning phase of new brachytherapy equipment, and afterwards during its clinical lifetime. The institution must therefore develop a proper QA program for sources and equipment. The overall objective of the study is to know the quality of brachytherapy at Om Hospital in the treatment of cancer and investigate the new method for maintaining the best quality. The specific objectives of the studies are to find out the Air Kerma Strength of the source, status of brachytherapy machine, dose delivered, activity of Ir-192 source used in brachytherpy, emergency procedure and radiation level. Different principles are explored by many scientists and researchers in the world about quality control. It is of extreme importance for the general process of quality assurance that these procedures are well defined and understood by a responsible medical physicist. Unfortunately, till now, in Nepal, there are no radiation protection laws and infrastructures, as well as no radiation regulatory board, to control the use of ionizing Volume 10 Number 2 Jul-Dec

2 radiations. Although there are many radiological societies/associations, hospital departments in Nepal, but no research has been carried out on quality control of brachytherapy till now. Recently, Nepal is part of the International Atomic Energy Agency (IAEA) and this will certainly strengthen and speed up the creation of radiation protection infrastructures This study may help to Radiation Oncologist, scholar, radiation workers and concerned persons to be aware on the efficiency of the machine and to be aware from radiation by providing test results and suggestions. METHODE This study was done at the Department of Radiotherapy, Brachytherapy Unit at Om Hospital, Kathmandu, mainly deals with the observation of sweet spots (maximum dose distribution) on different dates by using well chamber (SI HDR 1000) and electrometer (SI CDX 2000). On the basis of the observed sweet spot, Air Kerma Strength of the source is determined. Beside this includes the conformation of step size, radioactivity of the nuclides, safety measures of the machine. A careful radiation survey has been undertaken around the brachytherapy by using well calibrated radiation survey meter (FAG, FH 40F1). Limitation: The value of atmospheric pressure obtained from metro office located at the Tribhuwan International Airport may not coincide with the atmospheric pressure of Om hospital, hence found slight variation. RESULT Room design In the High Dose Rate (HDR) brachytherapy equipment, it is always best to provide single room accommodation. The layout of room will depend upon local circumstances but it will be necessary to ensure that the room is adequately protected, suitably located, large enough to allow access for patients on beds, enough space for the afterloading equipment and sufficient access for emergencies to be dealt with safely. Also wall thickness, floor, ceiling protection and door protection must meet the IAEA s recommendation. We found all above mention parameters. QA measures of HDR Machine Since individual treatment equipment and installation details in the treatment room may differ, the exact method to be used for safety checks has to be adapted to the local situation. The following is a list of functions and/or items were tested. Table:1 S.N Test Result 1. Communication equipment Video camera & intercom is functioning. 2. Check of treatment without machine is on. No connection to after loader. 3. Applicator attachment: Without/with The treatment starts and dummy source pushed at channel/ treatment start 4. Treatment with/without pressing last man out. Afterloader enabled./interrupted door. 5. Door interlock test/ Treatment interrupts/ Emergency stop 6. Power loss 7. Emergency equipment (forceps, emergency safe, survey meter): 8. Physical parameters Source positioning accuracy: Timer Consistency: Timer linearity: Length of treatment tubes: Door interrupted and source back to safe position/ treatment stop & source back to safe position/ treatment stop & source back to safe position Treatment continue by backup Emergency equipment required during the treatment are available OK The stop watch time and planned time matched. The stop watch time and planned time matched. Accurate 34 Volume 10 Number 2 Jul-Dec 2010

3 Sweet spot test A well calibrated well type chamber is used to perform a number of regular QC tests. The sweet spot was recorded by using well chamber and electrometer. All the instructions given in the manual were followed. The time exposure is 60 second and biasing voltage is -302V. Figure 4. Comparative radiation dose at different source positions The sweet position observed at the 10 th position in the chamber. The variation of the radiation dose observed at different date is shown in (Fig 4). From the result mentioned above it was found that radiation delivered dose by the source was maximum at 10 th position. This means the chambers center is at 10 th position. Calibrated Air Kerma Strength (S k ) test Figure 1. Radiation dose at different positions (mm) Figure 2. Radiation dose at different positions (mm) The Calibrated Air Kerma Strength observed at various date was compared with the respective manufacturer s calibration Air Kerma strength. It was found that the manufacturer s Air Kerma Strength and current Air Kerma Strength variation is large due to the radioactive decay. The comparative Air Kerma Strength study on the manufacturer s data and the observed data found that the difference in manufacturer s AKS with Observed AKS on date 7 th July, 13 th July and 15 th July 2009 was found to be -0.27%, 4.45% and 2.37%. The comparison between manufacturer s AKS and calibrated AKS on the various dates (Fig: 5 ). Figure 5. Comparative study of Air Kerma Strength (S k ) Figure 3. Radiation dose at different positions Volume 10 Number 2 Jul-Dec

4 Figure 6. Radiation dose in different positions at various date Dwell position test Dwell position test is used to find out exact position of the source. The standard deviation obtained is Radiation level test The radiation level was observed on the bottom, back, front and left/right side of the brachytherapy equipment. The observed result on date 8 th march 2009, 7 th July 2009 and 13 th July 2009 (Fig 6). Radioactivity of the source The observed radioactive decay and the supplied radioactive decay data by manufacturer s was observed and standard deviation is observed to be (Fig 7). The fall of the graph was because of the decay of the radioactive source (Irridium192) which half life is 74 days. Figure 5. Comparative study of supplied and observed radioactivity DISCUSSION In the High Dose Rate (HDR) brachytherapy equipment, it is always best to provide single room accommodation. The layout of room will depend on thickness of the wall, shielding of door 3,12. The reason for the maximum peak to be at 10 th position is that the most sensitive part of the well chamber is at the centre of the chamber. The slight decrease of the readings on the 10 th position on various dates was due to the decay of Ir-192 source 7. Data provided in above table as action level, reflect the upper limit in clinical conditions. For an acceptance test the design specifications must be compared. Often the design of the system is such that a much better performance can be obtained under reference conditions, such as positional checks with autoradiography. It is the physicist s task to inspect the performance history of the system 2. All observations Air Kerma Strength falls within 5% tolerance limit, which is acceptable 2. The Air Kerma Strength of the source in the specified position is as expected. The slight variation of the Air Kerma Strength was due to not exact positioning of the source at 10 th position of well chamber or due to the difference of pressure between Om Hospital and Tribhuwan International airport. During the source verification test slight deviation was observed due to the source wire and improper position of the film. The fall of the graph shown in Fig 7 was because of the decay of the radioactive source (Irridium192) which half life is 74 days 1.During survey, slightly high radiation level was found bottom of the head of machine than other positions because source is oriented on the bottom of the head. All observed radiation levels are within the safe limit as recommended by International Commission for Radiological Protection 13. CONCLUSION Brachytherapy among other radiation therapy for the treatment of cancer diseases is proving itself as a frontline therapy process. Attempts were made observing different parameters to find whether the Brachytherapy installed at Om Hospital & Research Centre is within the standard quality or not. Different experiments were conducted to find Air Kerma Strength, confirmation of step size; related procedures were followed for the quality assurance (QA) and safety measure check of the machine. Beside this, for the radiation safety of working personnel, patients and attendants of the hospital, survey of radiation level has been undertaken and found within the limit as described by ICRP. The calibrated Air Kerma strength on three different dates showed that the values were similar to manufacturer's calibrated Air Kerma strength and hence prescribed dose distribution on 36 Volume 10 Number 2 Jul-Dec 2010

5 specified area or volume of tissue of the patient was as planned. The step size is almost same as we planned. Safety measures and quality assurance tests proved that the status of machine is in good working condition and functioning well. The calculated radioactivity data matched with the manufacturers supplied radioactive data. This deviation observed is due to the source wire and improper position of the film. REFERENCES 1. Ervin B. Podgorsak, Review of Radiation Oncology Physics: A Handbook for Teachers and Students, International Atomic Energy Agency, Vienna, Austria, Margaret B., Edith B., Janez B. et al., 'A Practical Guide to Quality Control of Brachytherapy Equipment', Physics for Clinical Radiotherapy, ESTRO Booklet No. 8, Mounierlaan 83/ Brussels, Belgium ATTIX, F.H., Introduction to Radiological Physics and Radiation Dosimetry, John Wiley and Sons, USA, ISBN (1986) p Cuypers, J. and Robert, J. Double-film technique for the determination of the homogeneity of the line sources. Med. Phys. 22, , De Serd, L.A., Jursinic P., Kitchen R. and Thomadsen B., 'Quality Assurance Tool for High Dose Rate Brachytherapy', Med. Phys., 22, , Dewerd, L.A., Ezzell, G.A. & Williamson, J.F., Calibration Principles and Techniques, High Dose Rate Brachytherapy: A Textbook, Library of Congress Cataloging-In-Publication Data. NAG, S., Editor, Manual, 'CDX-2000B Electrometer, REF 90001', Standard Imaging, Inc., 7601 Murphy Drive, Middleton, WI , Manual, 'HDR 1000 Plus Well Chamber, REF 90008', Standard Imaging, Inc., Nath, R., Anderson, L.L., Meli, J.A., Olch, A.J., Stitt, J.A. and Williamson, J.F. Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56. Med. Phys. 24, , Pérez-Calatayud, J., Ballester, F., Granero, D. and Brosed, A. Influence of the non-homogeneity of Ir-192 wires in calibration of well chambers. Phys. Med. Biol. 48, , Thomadsen, B.R. Achieving quality in brachytherapy. Institute of Physics Publishing. Medical Science Series. Bristol and Philadelphia, Williamson, J.F., Ezzell, G.A., Olch, A. and Thomadsen, B.R. Quality assurance for high dose rate brachytherapy. In: High Dose Rate Brachytherapy: A Textbook. Library of Congress Cataloging-in-Publication Data. Nag, S., editor. Futura Publishing Company Inc., Armonk, New York, IAEA, Radiological protection for medical exposure to ionizing radiation 2002 International Atomic Energy Agency, No. RS-G-1.5, Vienna, Volume 10 Number 2 Jul-Dec

Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w

Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w Quality assurance Classical implant systems w Manchester

More information

Brachytherapy Planning and Quality Assurance

Brachytherapy Planning and Quality Assurance Brachytherapy Planning and Quality Assurance Classical implant systems Most common clinical applications and modern dosimetry methods Quality assurance Classical implant systems Manchester (Paterson-Parker)

More information

A Study on Dosimetry of Gynaecological Cancer and Quality Assurance of HDR Brachytherapy in BPKMCH, Nepal

A Study on Dosimetry of Gynaecological Cancer and Quality Assurance of HDR Brachytherapy in BPKMCH, Nepal A Study on Dosimetry of Gynaecological Cancer and Quality Assurance of HDR Brachytherapy in BPKMCH, Nepal SB Chand, PP Chaursia, MP Adhikary, AK Jha, and S Shrestha Dept. of Radiation Oncology, BPKM Cancer

More information

Brachytherapy. What is brachytherapy and how is it used?

Brachytherapy. What is brachytherapy and how is it used? Scan for mobile link. Brachytherapy Brachytherapy places radioactive sources inside the patient on a temporary or permanent basis to damage cancer cells DNA and destroy their ability to divide and grow.

More information

Radioactive sources in brachytherapy

Radioactive sources in brachytherapy Radioactive sources in brachytherapy Janez Burger Institute of Oncology, Department of Radiophysics, Brachytherapy Unit, Ljubljana, Slovenia Background. In modern brachytherapy, a great step forward was

More information

Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy

Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy Volume 11, No 3 International Journal of Radiation Research, July 2013 Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy M.A. Gadhi 1, S.A. Buzdar 2*, M. Afzal 2,

More information

Intravascular Brachytherapy Dosimetry Techniques for Gamma systems

Intravascular Brachytherapy Dosimetry Techniques for Gamma systems Intravascular Brachytherapy Dosimetry Techniques for Gamma systems Shirish K. Jani, Ph.D., FACR Scripps Clinic La Jolla, California Physics Parameters-1: IVBT Gamma or X ray Isotopes Energy [ kev ] Half

More information

Brachytherapy an Overview

Brachytherapy an Overview Brachytherapy an Overview Yakov Pipman, D Sc North Shore LIJ Health System Monterrey, Nov30-Dec1, 2007 Brachytherapy A procedure in therapeutic radiology that involves the irradiation of a target with

More information

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a The Equipment : safety, commissioning, re commissioning calibration : Equipements : sûreté, recette, modification calibration A medical physicist point of view. Albert Lisbona Medical Physics Department

More information

AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites

AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites A. Purpose and Rationale Many individual users contact the AAPM Brachytherapy

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

Regulatory Guidelines and Computational Methods for Safe Release of Radioactive Patients II. Brachytherapy

Regulatory Guidelines and Computational Methods for Safe Release of Radioactive Patients II. Brachytherapy Regulatory Guidelines and Computational Methods for Safe Release of Radioactive Patients II. Brachytherapy Firas Mourtada, Ph.D., DABR Chief of Clinical Physics Helen F. Graham Cancer Center Christiana

More information

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation. Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation Samuel Chiquita 1 1 University of Porto, Porto, Portugal Abstract In

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,

More information

HDR Brachytherapy Dosimetry at the NPL

HDR Brachytherapy Dosimetry at the NPL 1 Introduct HDR Brachytherapy Dosimetry at the NPL Brachytherapy is a special procedure in radiotherapy that utilises the irradiat of a target volume (e.g. malignant cells) with radioactive sources placed

More information

Brachytherapy The use of radioactive sources in close proximity to the target area for radiotherapy

Brachytherapy The use of radioactive sources in close proximity to the target area for radiotherapy Brachytherapy The use of radioactive sources in close proximity to the target area for radiotherapy Interstitial Seven 192-Ir wires Interstitial implant for breast radiotherapy Intracavitary Three 137-Cs

More information

IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects

IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects 2015 AAPM Annual Meeting Anaheim, CA Wes Culberson, PhD, DABR University of Wisconsin

More information

Sang Hun Shin, Wook Jae Yoo, Kyoung Won Jang, Seunghyun Cho, 1 Kum Bae Kim, 2 and Bongsoo Lee 3*

Sang Hun Shin, Wook Jae Yoo, Kyoung Won Jang, Seunghyun Cho, 1 Kum Bae Kim, 2 and Bongsoo Lee 3* Sensors and Materials, Vol. 28, No. 6 (2016) 619 624 MYU Tokyo 619 S & M 1216 Development of an All-in-One Phantom and Scintillator Radiation Sensor for Real-Time Monitoring of Source Position and Dose

More information

Major accidents in radiotherapy

Major accidents in radiotherapy Major accidents in radiotherapy related to treatment units (a) IAEA International Atomic Energy Agency Incorrect decay data (USA) IAEA International Atomic Energy Agency Background A cobalt unit was used

More information

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method Iran. J. Radiat. Res., 2007; 5 (3): 147-152 Dosimetric characteristics of Cs sources used in after loading Selectron system by Monte Carlo method M.B.Tavakoli, D. Shahbazi-Gahrouei *, M. Hosseinpour Department

More information

Lessons learned from accidents in conventional external radiotherapy

Lessons learned from accidents in conventional external radiotherapy Lessons learned from accidents in conventional external radiotherapy Ola Holmberg 1,2, Renate Czarwinski 2 1 Radiation Protection of Patients Unit 2 Radiation Safety and Monitoring Section Division of

More information

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Amendment No. 2. Item No. 2 (Rfx/ Event number ) Amendment 2 Sub: Amendment to the Document 06.12.2018 Ref.: Notice Inviting Bid ref. HITES/PCD/NCI-AIIMS/36/18-19 dated 26.09.2018 read with its Amendment no. 1 dated 19.11.18 The following changes have

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

RADIOTHERAPY DEPARTMENT

RADIOTHERAPY DEPARTMENT EXPERIENCES OF CLINICAL AUDIT IN A FINNISH RADIOTHERAPY DEPARTMENT Simo Hyödynmaa and Tuija Wigren Department of Radiation Therapy, Tampere University Hospital, Tampere, FINLAND PIRKANMAA HOSPITAL DISTRICT

More information

MultiSource. for a higher standard in HDR brachytherapy.

MultiSource. for a higher standard in HDR brachytherapy. MultiSource for a higher standard in HDR brachytherapy www.ibt-bebig.eu MultiSource HDR afterloader Multiple use: The MultiSource radiation sources have two things in common a small size and a favourable

More information

U.S. Nuclear Regulatory Commission

U.S. Nuclear Regulatory Commission Page 1 of 8 Index Site Map FAQ Help Glossary Contact Us A U.S. Nuclear Regulatory Commission Home Who We Are What We Do Nuclear Reactors Nuclear Materials Radioactive Waste Facility Info Finder Public

More information

A commissioning procedure for breast intracavitary electronic brachytherapy systems

A commissioning procedure for breast intracavitary electronic brachytherapy systems JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 3, SUMMER 2008 A commissioning procedure for breast intracavitary electronic brachytherapy systems Jessica Hiatt, Gene Cardarelli, Jaroslaw

More information

Review of TG-186 recommendations

Review of TG-186 recommendations Review of TG-186 recommendations Implementation of advanced brachytherapy dose calculation algorithms beyond TG-43 Rowan M. Thomson Carleton Laboratory for Radiotherapy Physics Carleton University Ottawa

More information

Slide 1. Slide 2. Slide 3. Introduction of INTRABEAM IORT. Disclosure. Contents. I have nothing to disclose.

Slide 1. Slide 2. Slide 3. Introduction of INTRABEAM IORT. Disclosure. Contents. I have nothing to disclose. Slide 1 AAPM 2017 Introduction of INTRABEAM IORT Hualin Zhang, Ph.D. Dept of Radiation Oncology, Northwestern University, Northwestern Memorial Hospital, Chicago, Illinois August 2, 2017 Slide 2 Disclosure

More information

Quality assurance in external radiotherapy

Quality assurance in external radiotherapy Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance

More information

Recent proceedings in Brachytherapy Physics

Recent proceedings in Brachytherapy Physics Recent proceedings in Brachytherapy Physics Frank-André Siebert UKSH, Campus Kiel, Germany Clinic of Radiotherapy Dept. Medical Physics Physical characteristics of brachytherapy (Courtesy Luc Beaulieu,

More information

FEE RULES RADIATION ONCOLOGY FEE SCHEDULE CONTENTS

FEE RULES RADIATION ONCOLOGY FEE SCHEDULE CONTENTS Tel: +27-21-9494060 Fax: +27-21-9494112 E-mail: leon.gouws@cancercare.co.za FEE RULES RADIATION ONCOLOGY FEE SCHEDULE CONTENTS 1. EXTERNAL BEAM RADIATION... 2 2. PLANNING OF TREATMENT... 2 3. DELIVERY

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

2015 Radiology Coding Survival Guide

2015 Radiology Coding Survival Guide 2015 Radiology Coding Survival Guide Chapter 31: Clinical Brachytherapy (77750-77799) Clinical brachytherapy involves applying radioelements into or around a treatment field. CPT guidelines clarify that

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of: Canadian Partnership for Quality Radiotherapy Technical Quality Control Guidelines for Gamma Knife Radiosurgery A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian Organization

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information

Physical aspects of intravascular brachytherapy. of the coronary arteries

Physical aspects of intravascular brachytherapy. of the coronary arteries Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology

More information

Acknowledgements. QA Concerns in MR Brachytherapy. Learning Objectives. Traditional T&O. Traditional Summary. Dose calculation 3/7/2015

Acknowledgements. QA Concerns in MR Brachytherapy. Learning Objectives. Traditional T&O. Traditional Summary. Dose calculation 3/7/2015 Acknowledgements QA Concerns in MR Brachytherapy Robert A. Cormack Dana Farber Cancer Institute & Brigham and Women s Hospital No financial conflicts of interest I may mention use of devices in ways that

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005 Advantages of inflatable multichannel endorectal applicator in the neo-adjuvant treatment of patients with locally advanced

More information

Therapeutic Medical Physics. Stephen J. Amadon Jr., Ph.D., DABR

Therapeutic Medical Physics. Stephen J. Amadon Jr., Ph.D., DABR Therapeutic Medical Physics Stephen J. Amadon Jr., Ph.D., DABR Outline 1. Why physicists are needed in medicine 2. Branches of medical physics 3. Physics in Radiation Oncology 4. Treatment types and Treatment

More information

Brachytherapy Sources, Dosimetry and Quality Assurance

Brachytherapy Sources, Dosimetry and Quality Assurance AAPM Meeting Therapy Physics Review Course Bibliography and Study Questions 19 July 2014 Brachytherapy Sources, Dosimetry and Quality Assurance Jeffrey F. Williamson, Lecturer General References and Fundamentals

More information

MEDICAL PHYSICS IN SUPPORT OF CANCER MANAGEMENT: HUMAN RESOURCE DEVELOPMENT IN RADIATION ONCOLOGY IN THE CONTEXT OF CANCER CONTROL PROGRAMMES

MEDICAL PHYSICS IN SUPPORT OF CANCER MANAGEMENT: HUMAN RESOURCE DEVELOPMENT IN RADIATION ONCOLOGY IN THE CONTEXT OF CANCER CONTROL PROGRAMMES PROSPECTUS MEDICAL PHYSICS IN SUPPORT OF CANCER MANAGEMENT: HUMAN RESOURCE DEVELOPMENT IN RADIATION ONCOLOGY IN THE CONTEXT OF CANCER CONTROL PROGRAMMES RAF/6/031-006 Title: Place: Regional (AFRA) Training

More information

Chapter 13: Brachytherapy: Physical and Clinical Aspects

Chapter 13: Brachytherapy: Physical and Clinical Aspects Chapter 13: Brachytherapy: Physical and Clinical Aspects Set of 163 slides based on the chapter authored by N. Suntharalingam, E.B. Podgorsak, H. Tolli of the IAEA publication (ISBN 92-0-107304-6): Radiation

More information

PGY-1. Resident Review Session Schedule

PGY-1. Resident Review Session Schedule 1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT

More information

Understanding Radiation Therapy. For Patients and the Public

Understanding Radiation Therapy. For Patients and the Public Understanding Radiation Therapy For Patients and the Public Introduction to Radiation Oncology Radiation has been an effective tool for treating cancer for more than 100 years. Radiation oncologists are

More information

Standards for Radiation Oncology

Standards for Radiation Oncology Standards for Radiation Oncology Radiation Oncology is the independent field of medicine which deals with the therapeutic applications of radiant energy and its modifiers as well as the study and management

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Twelfth Annual Warren K. Sinclair Keynote Address

Twelfth Annual Warren K. Sinclair Keynote Address THE INFLUENCE OF NCRP ON RADIATION PROTECTION IN THE U.S.: REGULATION AND GUIDANCE Twelfth Annual Warren K. Sinclair Keynote Address Kenneth R. Kase Annual Meeting of NCRP 16 March 2015 1 OUTLINE Introduction

More information

3D CONFORMATIONAL INTERSTITIAL BRACHYTHERAPY PLANNING FOR SOFT TISSUE SARCOMA

3D CONFORMATIONAL INTERSTITIAL BRACHYTHERAPY PLANNING FOR SOFT TISSUE SARCOMA 3D CONFORMATIONAL INTERSTITIAL BRACHYTHERAPY PLANNING FOR SOFT TISSUE SARCOMA Alina TĂNASE 1,3, M. DUMITRACHE 2,3, O. FLOREA 1 1 Emergency Central Military Hospital Dr. Carol Davila Bucharest, Romania,

More information

NEWS A Publication of Vantage Oncology, Tri-State Radiation Oncology Centers - TROC

NEWS A Publication of Vantage Oncology, Tri-State Radiation Oncology Centers - TROC Oral, Head & Neck Cancer Screening Conducted Dr. Mark Logan, Midwest ENT Surgery 25 Patients Required Routine Follow up w/ PCP On Thursday, April 26th Evansville Cancer Center/ Vantage Oncology hosted

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Definitions. Brachytherapy in treatment of cancer. Implantation Techniques and Methods of Dose Specifications. Importance of Brachytherapy in GYN

Definitions. Brachytherapy in treatment of cancer. Implantation Techniques and Methods of Dose Specifications. Importance of Brachytherapy in GYN Implantation Techniques and Methods of Dose Specifications Brachytherapy Course Lecture V Krishna Reddy, MD, PhD Assistant Professor, Radiation Oncology Brachytherapy in treatment of cancer GYN Cervical

More information

A survey of quality control practices for high dose rate (HDR) and pulsed dose rate (PDR) brachytherapy in the United Kingdom

A survey of quality control practices for high dose rate (HDR) and pulsed dose rate (PDR) brachytherapy in the United Kingdom A survey of quality control practices for high dose rate (HDR) and pulsed dose rate (PDR) brachytherapy in the United Kingdom Type: Physics contributions Abstract: Purpose: A survey of quality control

More information

Advanced film dosimetry for a UK brachytherapy audit

Advanced film dosimetry for a UK brachytherapy audit Advanced film dosimetry for a UK brachytherapy audit Tony Palmer Lead Scientist Medical Physics, Portsmouth Hospitals NHS Trust, UK University of Surrey, UK Institute of Physics and Engineering in Medicine

More information

Transperineal Interstitial Permanent Prostate Brachytherapy (TIPPB) Quality Assurance Guidelines

Transperineal Interstitial Permanent Prostate Brachytherapy (TIPPB) Quality Assurance Guidelines Prostate Brachytherapy QA Page 1 of 1 Transperineal Interstitial Permanent Prostate Brachytherapy (TIPPB) Quality Assurance Guidelines I. Purpose Table of Contents II. III. IV. Background Credentialing

More information

Survey of Radiation Dose Levels in Patients in X-Ray Units of Some Selected Hospitals in Jos Metropolis

Survey of Radiation Dose Levels in Patients in X-Ray Units of Some Selected Hospitals in Jos Metropolis International Journal of Innovative Scientific & Engineering Technologies Research 6(4):1-9, Oct.-.Dec., 2018 SEAHI PUBLICATIONS, 2018 www.seahipaj.org ISSN: 2360-896X Survey of Radiation Dose Levels in

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

ADVANCED TECHNOLOGY CONSORTIUM (ATC) CREDENTIALING PROCEDURES FOR LUNG BRACHYTHERAPY IMPLANT PROTOCOLS

ADVANCED TECHNOLOGY CONSORTIUM (ATC) CREDENTIALING PROCEDURES FOR LUNG BRACHYTHERAPY IMPLANT PROTOCOLS ACOSOG-RTOG Lung Brachytherapy QA Page 1 of 8 ADVANCED TECHNOLOGY CONSORTIUM (ATC) CREDENTIALING PROCEDURES FOR LUNG BRACHYTHERAPY IMPLANT PROTOCOLS FACILITY QUESTIONNAIRE Institutions wishing to enter

More information

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy Turkish Journal of Cancer Volume 37, No. 2, 2007 59 Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy MUHAMMAD BASIM

More information

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors G.Gambarini 1, C.Tenconi 1, N.Mantaut 1 M.Carrara 2, M.Borroni 2, E.Pignoli 2 D.Cutajar 3, M.Petasecca 3, I.Fuduli 3, M.Lerch

More information

RADI 6033 Physics of Radiotherapy II Spring 2017

RADI 6033 Physics of Radiotherapy II Spring 2017 RADI 6033 Physics of Radiotherapy II Spring CLASS DAYS and TIME: Monday and Wednesday (10:30 12:00pm) CLASSROOM: CTRC Building - Grossman Plaza (G254) COURSE DIRECTOR: COURSE LECTURERS: Dennis N. Stanley,,

More information

The determination of timer error and its role in the administration of specified doses

The determination of timer error and its role in the administration of specified doses JURNAL FIZIK MALAYSIA VOLUME 29, NUMBER 1&2 2008 The determination of timer error and its role in the administration of specified doses S. B. Samat a *, C. J. Evans b, T. Kadni c and M. T. Dolah c a School

More information

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Dosimetric characterization of surface applicators for use with the Xoft ebx system Dosimetric characterization of surface applicators for use with the Xoft ebx system R.M. Kennedy University of Wisconsin Medical Research Center Madison, WI April 23, 2010 Introduction Squamous and basal

More information

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed

More information

Inter-society standards for the performance of brachytherapy: a joint report from ABS, ACMP and ACRO

Inter-society standards for the performance of brachytherapy: a joint report from ABS, ACMP and ACRO Critical Reviews in Oncology/Hematology 48 (2003) 1 17 Inter-society standards for the performance of brachytherapy: a joint report from ABS, ACMP and ACRO Subir Nag a,, Ralph Dobelbower b, Glenn Glasgow

More information

Guideline & Reports 医学物理学会教育委員会資料

Guideline & Reports 医学物理学会教育委員会資料 Guideline & Reports 医学物理学会教育委員会資料 2017/11 更新 report title year keyword 1 keyword 2 AAPM TG211 Classification and evaluation strategies of auto-segmentation approaches for PET 2017 PET autosegmentation

More information

Quality management for Breast Brachytherapy.

Quality management for Breast Brachytherapy. Quality management for Breast Brachytherapy. DORIN TODOR, Ph.D. Medical College of Virginia Campus Department of Radiation Oncology New England AAPM Chapter 2012 Summer Meeting, Providence, RI Quality

More information

IAEA activities in Medical Physics

IAEA activities in Medical Physics IAEA activities in Medical Physics Giorgia Loreti Training Officer (Medical Physics) G.Loreti@iaea.org IAEA - International Atomic Energy Agency Department of Nuclear Sciences and Applications Division

More information

How ICD-10 Affects Radiation Oncology. Presented by, Lashelle Bolton CPC, COC, CPC-I, CPMA

How ICD-10 Affects Radiation Oncology. Presented by, Lashelle Bolton CPC, COC, CPC-I, CPMA How ICD-10 Affects Radiation Oncology Presented by, Lashelle Bolton CPC, COC, CPC-I, CPMA ICD-10 ICD-10-CM has added new challenges to the radiation oncology specialty. Approximately 220 ICD-9-CM codes

More information

Method for verifying the air kerma strength of I-125 plaques for the treatment of ocular melanoma

Method for verifying the air kerma strength of I-125 plaques for the treatment of ocular melanoma JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 4, 2014 Method for verifying the air kerma strength of I-125 plaques for the treatment of ocular melanoma L. W. Zimmermann, 1a D. Allan Wilkinson

More information

Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations

Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations Iran. J. Radiat. Res., 21; 7 (4): 217-222 Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations Downloaded from ijrr.com at 6:52 +33

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

3D ANATOMY-BASED PLANNING OPTIMIZATION FOR HDR BRACHYTHERAPY OF CERVIX CANCER

3D ANATOMY-BASED PLANNING OPTIMIZATION FOR HDR BRACHYTHERAPY OF CERVIX CANCER SAUDI JOURNAL OF OBSTETRICS AND GYNECOLOGY VOLUME 11 NO. 2 1430 H - 2009 G 3D ANATOMY-BASED PLANNING OPTIMIZATION FOR HDR BRACHYTHERAPY OF CERVIX CANCER DR YASIR BAHADUR 1, DR CAMELIA CONSTANTINESCU 2,

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

New Radiation Health Technologies Challenges, Opportunities, Limitations

New Radiation Health Technologies Challenges, Opportunities, Limitations New Radiation Health Technologies Challenges, Opportunities, Limitations IRPA-13 Glasgow Fridtjof Nuesslin Klinik für Strahlentherapie und Radiologische Onkologie 2012-IIRPA13 Radiation Health Technologies

More information

IORT with mobile linacs: the Italian experience

IORT with mobile linacs: the Italian experience IORT with mobile linacs: the Italian experience G. Tosi, M. Ciocca lntroduction At the beginning of 1999 a mobile Linac (Novac 7, manufactured by Hitesys, Aprilia, Italy) able to produce electron beams

More information

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-861.Volume 6, Issue Ver. II (May-Jun. 201), PP 2- A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

Brachytherapy Dosimetry in the Clinic Margaret Bidmead Royal Marsden NHS Trust

Brachytherapy Dosimetry in the Clinic Margaret Bidmead Royal Marsden NHS Trust Brachytherapy Dosimetry in the Clinic Margaret Bidmead Royal Marsden NHS Trust Two Curies Bequerel Brachytherapy What is it? Sealed radioactive source(s) placed directly into, or immediately adjacent to

More information

Trends in Occupational Exposure in Malaysia

Trends in Occupational Exposure in Malaysia Trends in Occupational Exposure in Malaysia Noriah Mod Ali Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute for Nuclear Technology Research (MINT) Bangi, 43000 KAJANG, Selangor Darul

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

A Review of UK Guidelines for Prostate Brachytherapy. Sarah Aldridge. Head of Brachytherapy Physics Guy s & St. Thomas NHS Foundation Trust

A Review of UK Guidelines for Prostate Brachytherapy. Sarah Aldridge. Head of Brachytherapy Physics Guy s & St. Thomas NHS Foundation Trust A Review of UK Guidelines for Prostate Brachytherapy Sarah Aldridge Head of Brachytherapy Physics Guy s & St. Thomas NHS Foundation Trust Overview Review of current guidelines for LDR & HDR prostate brachytherapy

More information

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St. Quality Assurance of Ultrasound Imaging in Radiation Therapy Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St. Louis, Missouri Typical Applications of Ultrasound Imaging in Radiation Therapy

More information

Office: Cell Ph:

Office: Cell Ph: Rajeev K. Badkul, M.S., DABR Lead Medical Physicist & Assistant Professor Dept. of Radiation Oncology University Of Kansas Medical Center 3901 Rainbow Blvd., Kansas City, KS,66160 rbadkul@kumc.edu Office:

More information

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY Journal of Bangladesh Academy of Sciences, Vol. 3, No. 1, 7-17, 211 A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY MAHFUZA BEGUM 1, A. S. MOLLAH

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 37-013 GARD Users Manual March 2005 Manual No. 37-013-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective

More information

Current Status of Electronic Brachytherapy Dosimetry

Current Status of Electronic Brachytherapy Dosimetry Current Status of Electronic Brachytherapy Dosimetry 2014 NCCAAPM Fall Meeting La Crosse, WI Wes Culberson, PhD, DABR University of Wisconsin Madison University of Wisconsin Medical Radiation Research

More information

PRINCIPLES and PRACTICE of RADIATION ONCOLOGY. Matthew B. Podgorsak, PhD, FAAPM Department of Radiation Oncology

PRINCIPLES and PRACTICE of RADIATION ONCOLOGY. Matthew B. Podgorsak, PhD, FAAPM Department of Radiation Oncology PRINCIPLES and PRACTICE of RADIATION ONCOLOGY Matthew B. Podgorsak, PhD, FAAPM Department of Radiation Oncology OUTLINE Physical basis Biological basis History of radiation therapy Treatment planning Technology

More information

IPEM Recommendations for the Provision of a Physics Service to Radiotherapy

IPEM Recommendations for the Provision of a Physics Service to Radiotherapy Institute of Physics and Engineering in Medicine IPEM Recommendations for the Provision of a Physics Service to Radiotherapy 1. Introduction This document reflects the impact of the technological, medical

More information

Advances in Treatment of Cancer by Brachytherapy in Kenya, in Particular, Prostate Cancer

Advances in Treatment of Cancer by Brachytherapy in Kenya, in Particular, Prostate Cancer Research Article imedpub Journals www.imedpub.com Journal Of Medical Physics And Applied Sciences ISSN 2574-285X DOI: 1.21767/2574-285X.12 Advances in Treatment of Cancer by Brachytherapy in Kenya, in

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

Canadian Partnership for Quality Radiotherapy. Quality Assurance Guidelines for Canadian Radiation Treatment Programs

Canadian Partnership for Quality Radiotherapy. Quality Assurance Guidelines for Canadian Radiation Treatment Programs Canadian Partnership for Quality Radiotherapy Quality Assurance Guidelines for Canadian Radiation Treatment Programs A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

The Journal of Global Radiology

The Journal of Global Radiology The ENSURING MEDICAL IMAGING ACCESS FOR ALL ORIGINAL RESEARCH A Survey of Structural Design of Diagnostic X-ray Imaging Facilities and Compliance to Shielding Design Goals in a Limited Re Setting Flavious

More information

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging Iran. J. Radiat. Res., 2008; 6 (2): 83-87 Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging B. Azadbakht 1, M.H. Zahmatkesh 2 *, k. Hadad 1, S. Bagheri 2

More information

Commissioning of brachytherapy Treatment Planning Systems

Commissioning of brachytherapy Treatment Planning Systems Commissioning of brachytherapy Treatment Planning Systems Jacco L.G. Steenhuijsen, PhD Catharina Hospital Eindhoven Landelijke brachydag, 17-11-17 Brachytherapy Treatment Planning Systems 2 Brachytherapy

More information