Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry
|
|
- Carol Simon
- 5 years ago
- Views:
Transcription
1 Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology
2 Conflict of interest None Acknowledgement: Jim Ernsberger, SNC.
3 Transitioning from Phantom geometry to patient geometry for VMAT H & N QA: Purpose: The purpose of this study is to a) investigate the accuracy of patient DVH based QA using volumetric measurement guided dose reconstruction (MGDR) for head & neck Volumetric Modulated Arc Therapy (VMAT) delivery by benchmarking against RPC Head and Neck Phantom results, b) investigate the correlation of 3D detector phantom based QA using local gamma and global gamma results when compared against patient DVH based QA. Methods: The RPC head and neck phantom was planned and irradiated using 2 arc VMAT delivery method. The results of point doses from RPC TLDs at eight locations and dose profiles from radiochromic film through center of primary PTV inside the phantom were compared against the treatment planning system (TPS) and the results from the MGDR analysis based on 3D detector phantom QA. Also, the 3D gamma value was calculated using the MGDR analysis which compares the TPS dose to the predicted dose distribution. Five head and neck patients were planned using VMAT delivery method and the QA was performed using both 3D detector phantom based QA and patient DVH based analysis using MGDR. The correlation between gamma pass rates using both global and local gamma criteria was compared against the gamma pass rates using MGDR analysis. Results: The ratio of MGDR calculated dose to primary and secondary planning target volume (PTV) compared with both RPC TLD and Eclipse TPS doses was The ratio of estimated cord dose from MGDR analysis was 1.09 and 1.03 when compared to RPC TLD and TPS doses respectively. The displacement between calculated dose gradient in the region between primary PTV and organ at risk (OAR) in all 3 planes from MGDR was within 3mm and 1mm when compared to RPC film and TPS profiles respectively. The average 3D local gamma pass rate for the five clinical cases using MGDR was 97.5% and 92% when using 3% 3mm and 2%2mm analysis criteria respectively. The average gamma pass rate using global gamma 3%3mm criteria was 99.8%, compared to an average local gamma pass rate of 87.1% using 3D detector phantom based QA with a maximum increase of 18%.
4 Conclusion: Benchmarking the accuracy of patient DVH based QA results to the RPC head and neck phantom established a baseline accuracy and confidence in use of MGDR analysis for VMAT delivery in a realistic patient geometry. For the 5 clinical test cases studied, the low pass rates obtained using the local gamma evaluation criteria in phantom based QA had no significant clinical impact for the patient when evaluated using DVH based QA.
5 Objectives Background Status Quo adequate? Review of Physics behind MGDR Benchmarking accuracy of MGDR & comparison with traditional phantom QA Conclusions
6 Background: Traditional VMAT/RapidArc QA employs detector based geometry using the ubiquitous 3%/3mm dose and DTA threshold for gamma QA pass rates. The traditional QA detects errors (sensitivity) However no information on the size and spatial location of errors in patient geometry. (specificity) Is status quo enough or accurate to detect clinically meaninful errors?
7 VMAT QA Every patient receives a patient specific QA. ACR/ASTRO Practice guideline definition for IMRT/VMAT:. accuracy of dose delivery should be documented by irradiating a phantom containing a calibrated dosimetry system to verify dose delivered is same as dose planned MGDR refers to a solution where errors in QA can be meaningfully correlated to patient specific geometry and structures
8
9 Gamma QA a) Per beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors, B. Nelms et al Med Phys, 38 January 2011:
10
11 Moving from gamma passing rates to patient DVH based QA metrics in pretreatment dose QA, Zhen H. et al., Med. Phys Oct;38(10):
12 Is current QA/methodology adequate?
13 What does conventional QA results mean?
14 How to reconstruct 3Ddose? Commercially available ArcCheck 3DVH Delta 4 DVH anatomy EPID Math Resolutions etc.
15 ArcCHECK based MGDR In BEV mode, the detector geometry is invariant of gantry angle
16 PDP or 3DVH
17 3DVH
18 Physics of MGDR Basic premise is to use the errors measured in phantom geometry to perturb the TPS dose by a correction factor that faithfully represents the ACTUAL dose delivered to the patient.
19 6 step process going from phantom to patient in 3DVH Step 1 AC Phantom measurement MUST make the measurement with PMMA plug
20 Step 2 Synchronize angle with time AC dose is updated every 50 ms ( in its own clock) RT Plan control points are a function of gantry angle NOT time In VMAT gantry angle is not linear with time (variable speed) Need to synchronize CP with Dose(t) SNC does with virtual gantry angle inclinometer
21 Step 2 Virtual Inclinometer
22 Step 3 Calculate RELATIVE dose in uniform PMMA phantom Divide the dynamic RT plan(cp) into sub beams Process the time stamped sub beams to create the fluence Perform convolution type calculation for each sub beam to derive RELATIVE dose
23 Step 4 Convert sub beam dose to absolute dose Use entrance and exit measured absolute dose Convert relative dose to semi empirical absolute dose Called morphing because calibration factor changes along the ray Larger weight given near the entrance diode vs exit diode
24 Step 5 Sum the sub beams in phantom
25 Step 6 Make the jump to patient geometry
26 PDP Model Is it accurate? NOT interested in solid water and film Need validity in actual patient geometry We used RPC H & N Phantom
27 VMAT Head & Neck Treatment Delivery Benchmarking Accuracy of Patient DVH based QA with Radiological Physics Center (RPC) Head and Neck Phantom Results.
28 RPC HEAD & NECK PHANTOM
29 TLD Results comparison Dose units in cgy Location Eclipse Mean RPC TLD(Mean) 3DVH(MEAN) (RPC/3DVH) (RPC/TPS) Primary PTV sup. ant Primary PTV inf. ant Primary PTV sup. post Primary PTV inf. post Secondary PTV sup Secondary PTV inf Organ at risk sup Organ at risk inf
30 RPC Film vs Eclipse TPS
31 MGDR (green) vs Eclipse TPS Right Left Profile
32 RPC Film vs Eclipse TPS
33 MDGR vs TPS Anterior Posterior Profile
34 RPC Film vs Eclipse TPS
35 MGDR vs Eclipse TPS Superior Inferior Profile
36 3DVH analysis
37 Comparison of QA for Patient H & N VMAT plans Patient Arc check QA local gamma pass rate Arc check Global gamma 3D DVH local gamma pass 3D DVH 2% 2mm % 100% 98.20% 93.80% % 99.50% 97.5% 92% % 100% 99.40% 96.20% % 99.80% 99.20% 96.80%
38 ArcCHECK Phantom QA Patient 2
39 3DVH QA Patient 2
40 3DVH analysis
41 ArcCHECK Phantom QA Patient 3
42 3DVH Patient 3
43 CONCLUSIONS For VMAT, phantom geometry based gamma QA metrics have a weak correlation with actual dose delivered to patient. Need to take a step back and think what is the goal of QA? Are you checking the TPS model indirectly or errors that have patient specific clinical impact? If the answer is latter, then MGDR analysis is the likely solution.
44 Is current QA/methodology adequate?
Leila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationVolumetric Modulated Arc Therapy - Patient Specific QA
Volumetric Modulated Arc Therapy - Patient Specific QA Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA VMAT plan QA methods Composite dose measurement Film & ion chamber diode array Mapcheck
More informationIMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)
IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)
More informationARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.
ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationA Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer
A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer Tony Kin Ming Lam Radiation Planner Dr Patricia Lindsay, Radiation Physicist Dr John Kim, Radiation Oncologist Dr Kim Ann Ung,
More informationThe MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates Daniel W. Bailey, 1,2 a Jason D. Spaans, 2
More informationEvaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell
Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating Characteristic (ROC) Analysis by Allison Lorraine Mitchell Graduate Program in Medical Physics Duke University Date: Approved:
More informationAssessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom
Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson
More informationA model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms
1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,
More informationEvaluation of Dosimetry Check software for IMRT patient-specific quality assurance
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,
More informationEvaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc
Original Article PMP Progress in Medical Physics 28(4), December 217 https://doi.org/1.14316/pmp.217.28.4.181 pissn 258-4445, eissn 258-4453 Evaluation of Dynamic Delivery Quality Assurance Process for
More informationThe Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom
The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom Gary Fisher, B.S. David Followill, Ph.D. Geoffrey Ibbott, Ph.D. This investigation
More informationA VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 401 410, 2014 A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * M. D. SUDITU 1,2, D. ADAM 1,2, R. POPA 1,2, V. CIOCALTEI
More informationIMRT QA: Point Dose Measurements or 2D Array?
QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify
More informationDosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar
More informationActivity report from JCOG physics group
2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer
More informationAgenda. 1.1 Why pre-treament dosimetry in IMRT/VMAT? 1.2 Verification systems in IMRT/VMAT: from 0D to 3D (4D) dosimetry to real time dosimetry
Controlli di qualità preclinici in IMRT statica e dinamica Michele Stasi S.C. Fisica Sanitaria, A.O. Ordine Mauriziano D.O. Fisica Sanitaria, IRCCS Candiolo TORINO mstasi@mauriziano.it Agenda 1. Introduction
More informationPre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system
Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system E. Mezzenga, * E. Cagni, A. Botti, M. Orlandi and M. Iori Medical Physics Unit, IRCCS Arcispedale
More informationPatient QA Recommended Publications
Patient QA Recommended Publications Your Most Valuable QA and Dosimetry Tools 3DVH Accuracy Studies Cross-Validation of Two Commercial Methods for Volumetric High-Resolution Dose Reconstruction on a Phantom
More informationAre Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?
Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or
More informationEffective and efficient radiotherapy dosimetry audit: Where to next?
Effective and efficient radiotherapy dosimetry audit: Where to next? Catharine Clark, Eduard Gershkevitsh, Milan Tomsej and Nuria Jornet National Physical Laboratory Royal Surrey County Hospital Radiotherapy
More informationIMRT pre-treatment verification basic approaches
School of Medical Physics for Radiation Therapy: Dosimetry and Treatment Planning for Basic and Advanced Applications 27-march 7-april 2017 IMRT pre-treatment verification basic approaches Eugenia Moretti
More informationFuture upcoming technologies and what audit needs to address
Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple
More informationDELIVERED PATIENT DOSE
SOFTWARE DELIVERED PATIENT DOSE The only software that provides a true picture of patient and dose at each stage of treatment. Adaptivo patient dosimetry software does something other patient dosimetry
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationVerification of Advanced Radiotherapy Techniques
Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December
More informationEVALUATION OF INTENSITY MODULATED RADIATION THERAPY (IMRT) DELIVERY ERROR DUE TO IMRT TREATMENT PLAN COMPLEXITY AND IMPROPERLY MATCHED DOSIMETRY DATA
Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2011 EVALUATION OF INTENSITY MODULATED RADIATION THERAPY (IMRT) DELIVERY
More information8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient
Number of Publications Number of publications 8/3/2016 The Strikes Back! - In-Vivo Dosimetry AAPM, Washington D.C, USA, 2016 Peter Greer 1,2 (1) University of Newcastle, Australia, (2) Calvary Mater Newcastle,
More informationGAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119
Romanian Reports in Physics, Vol. 68, No. 3, P. 1130 1139, 2016 GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119 M. DUMITRACHE 1,3, Alina TĂNASE
More informationGAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119
GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119 M. DUMITRACHE 1,3, Alina TĂNASE 2,3 1 Institute of Oncology Prof. Dr. Al. Trestioreanu, Bucharest,
More informationVMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT
VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT Poster No.: C-0520 Congress: ECR 2011 Type: Scientific Exhibit Authors: Y. Kawasaki, S. Tadokoro,
More informationIROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014
IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 The study groups are requesting that each institution keep the phantom for a period of time
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationEvaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases
Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases CASE STUDY Institution: Odette Cancer Centre Location: Sunnybrook
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More informationAdvanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*
Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark* Purpose: To evaluate an institution s 3D treatment planning process and the institution
More informationStatistical process control analysis for patient-specific IMRT and VMAT QA
Journal of Radiation Research, 2013, 54, 546 552 doi: 10.1093/jrr/rrs112 Advance Access Publication 7 December 2012 Statistical process control analysis for patient-specific IMRT and VMAT QA Taweap SANGHANGTHUM
More informationIntroduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology
Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?
More informationAn anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery
An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,
More informationUnrivaled, End-to-End
PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by
More informationIROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014
IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014 The IROC Houston is requesting that each institution keep the phantom for two weeks. During
More informationQuality assurance and credentialing requirements for sites using inverse planned IMRT Techniques
TROG 08.03 RAVES Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques Introduction Commissioning and quality assurance of planning systems and treatment delivery
More informationQA for Clinical Dosimetry with Emphasis on Clinical Trials
QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions
More informationCan we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?
Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Dimos Baltas Dept. of Medical Physics & Engineering, Strahlenklinik, Klinikum Offenbach GmbH 63069 Offenbach, Germany
More informationA review of RTTQA audit. Karen Venables
A review of RTTQA audit Karen Venables CHART 8 hours on the machine Chart cord CHART QA: Phantoms(1 st use of anatomical phantoms) Designed for treatments in Bronchus and Head and Neck (2D only) Outlines
More informationLung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014
Lung Spine Phantom Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 The study groups are requesting that each institution keep the phantom for no more than 2 week. During this
More informationComparison of 3D anatomical dose verification and 2D phantom dose verification of IMRT/VMAT treatments for nasopharyngeal carcinoma
Lin et al. Radiation Oncology 2014, 9:71 RESEARCH Open Access Comparison of 3D anatomical dose verification and 2D phantom dose verification of IMRT/VMAT treatments for nasopharyngeal carcinoma Hailei
More informationQuantitative evaluation of 3D dosimetry for stereotactic volumetric-modulated arc delivery using COMPASS
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Quantitative evaluation of 3D dosimetry for stereotactic volumetric-modulated arc delivery using COMPASS Subramani Vikraman, 1,6a
More informationTHE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS
THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS ICTP SCHOOL ON MEDICAL PHYSICS FOR RADIATION THERAPY DOSIMETRY AND TREATMENT PLANNING FOR BASIC AND ADVANCED APPLICATIONS March
More informationElectron Beam ET - Reloaded Therapy - Reloaded
8/4/216 Electron Beam ET - Reloaded Therapy - Reloaded Qiuwen Wu, Ph.D. Department of Radiation Oncology Duke University 1 Introduction: e vs. γ vs. p Electron Photon Proton Depth dose and lateral profiles
More informationDefining Target Volumes and Organs at Risk: a common language
Defining Target Volumes and Organs at Risk: a common language Eduardo Rosenblatt Section Head Applied Radiation Biology and Radiotherapy (ARBR) Section Division of Human Health IAEA Objective: To introduce
More informationDosisverifikation mit Delta 4 Discover während der Behandlung
Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover
More informationEPID dosimetry for pretreatment quality assurance with two commercial systems
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 EPID dosimetry for pretreatment quality assurance with two commercial systems Daniel W. Bailey, 1,2a Lalith Kumaraswamy, 1 Mohammad
More informationPre-treatment verification for IMRT / IGRT and Rotational therapy
In OmniPro-I mrt Vivo treatment software verification Pre-treatment verification for IMRT / IGRT and Rotational therapy Introduction Introduction Introduction Introtro Intro Intro The rapid development
More informationM. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course
Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics
More informationMany vendors are beginning to allow couch motion during radiation delivery.
Dynamic Couch Motion Many vendors are beginning to allow couch motion during radiation delivery. Varian developer mode allows institutions to perform research using these types of treatments. Tomotherapy
More information3D-CRT Breast Cancer Planning
3D-CRT Breast Cancer Planning Tips and Tricks Bednář, V. 3D-CRT Obsolete or not? There are more advanced techniques than 3D-CRT, but 3D-CRT has some advantages: Availability and price Forward planning
More informationThe RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff
The RPC s Evaluation of Advanced Technologies AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff 1 http://rpc.mdanderson.org Supported by: NCI grants CA10953 and CA81647, and an
More informationTreatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)
Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Tagreed AL-ALAWI Medical Physicist King Abdullah Medical City- Jeddah Aim 1. Simplify and standardize
More informationPlanIQ. Plan Smarter, Not Harder. Your Most Valuable QA and Dosimetry Tools
PlanIQ Plan Smarter, Not Harder Your Most Valuable QA and Dosimetry Tools SAVE TIME, IMPROVE QUALITY Use PlanIQ as a patient-specific solution to both measure and improve plan quality. Get instant feedback
More informationDosimetric Analysis Report
RT-safe 48, Artotinis str 116 33, Athens Greece RT-safe +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ------ Irradiation system: ------
More informationIROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015
IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationThe objective of this lecture is to integrate our knowledge of the differences between 2D and 3D planning and apply the same to various clinical
The objective of this lecture is to integrate our knowledge of the differences between 2D and 3D planning and apply the same to various clinical sites. The final aim will be to be able to make out these
More informationAAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction
AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall
More informationDosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.87 Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times Mee
More informationSunCHECK Patient Comprehensive Patient QA
The SunCHECK Platform SunCHECK Patient Comprehensive Patient QA SunCHECK Patient // Comprehensive Patient QA Automate Your Patient QA SunCHECK Patient automates all your Patient QA needs from Secondary
More informationCan we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?
Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Dimos Baltas 1,3, Natasa Milickovic 1, Nikolaos Zamboglou 2 1 Dept. of Medical Physics & Engineering, 2 Strahlenklinik,
More information3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients
3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients G Asuni *1, T vanbeek 1, E VanUtyven 1, P McCowan 1,2, and B.M.C. McCurdy 1,2,3 1 Medical Physics Department, CancerCare
More informationFilm-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom
Received: 4 August 2017 Revised: 1 January 2018 Accepted: 6 February 2018 DOI: 10.1002/acm2.12314 RADIATION ONCOLOGY PHYSICS Film-based dose validation of Monte Carlo algorithm for Cyberknife system with
More informationWhat Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff
What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG
More informationCURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT
CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT Purpose The purpose of this curriculum outline is to provide a framework for multidisciplinary training for radiation oncologists, medical
More informationReena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery
Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery Reena Phurailatpam Tejpal Gupta, Rakesh Jalali, Zubin Master, Bhooshan Zade,
More informationNormal tissue dose in pediatric VMAT Piotr Zygmanski
Normal tissue dose in pediatric VMAT Piotr Zygmanski Department of Radiation Oncology Brigham and Women s Hospital Radiation induced effects in children Risk of inducing 2 nd cancers higher than for adults
More informationPlan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall
More informationUse of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments
Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Jean M. Moran, Ph.D. Department of Radiation Therapy, University of Michigan, 1500 E. Medical Center Drive, Ann Arbor
More informationRisk of Secondary Fatal Malignancies from Hi-Art Tomotherapy
Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy Susannah Lazar David Followill, Ph.D. John Gibbons, Ph.D. * Anita Mahajan, M.D. Mohammad Salehpour, Ph.D. Marilyn Stovall, Ph.D. R. Allen White,
More informationRPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy
RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy Guidelines for Planning and Irradiating the RPC Liver Phantom. Revised Dec 2005 Credentialing for this protocol requires four steps:
More information3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter
Optical Scanning 3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter Wuu, C.-S., Hoogcarspel, S.J.,Deh, K., Hsu, W.-Y., Adamovics, J. 2013 Journal of Physics: Conference Series
More informationA TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,
More informationQuality assurance of volumetric modulated arc therapy using Elekta Synergy
Acta Oncologica, 2009; 48: 11931197 ORIGINAL ARTICLE Quality assurance of volumetric modulated arc therapy using Elekta Synergy AKIHIRO HAGA 1, KEIICHI NAKAGAWA 1, KENSHIRO SHIRAISHI 1, SAORI ITOH 1, ATSURO
More informationFeasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry
Baek et al. Radiation Oncology 2014, 9:273 RESEARCH Open Access Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry Tae Seong Baek 1,2, Eun Ji
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More informationThe Physics of Oesophageal Cancer Radiotherapy
The Physics of Oesophageal Cancer Radiotherapy Dr. Philip Wai Radiotherapy Physics Royal Marsden Hospital 1 Contents Brief clinical introduction Imaging and Target definition Dose prescription & patient
More informationBeyond the DVH: Voxelbased Automated Planning and Quality Assurance
1 Beyond the DVH: Voxelbased Automated Planning and Quality Assurance Presented by Chris McIntosh Research Associate Department of Medical Imaging & Physics Princess Margaret Cancer Centre C. McIntosh
More informationSilvia Pella, PhD, DABR Brian Doozan, MS South Florida Radiation Oncology Florida Atlantic University Advanced Radiation Physics Boca Raton, Florida
American Association of Medical Dosimetrists 2015 Silvia Pella, PhD, DABR Brian Doozan, MS South Florida Radiation Oncology Florida Atlantic University Advanced Radiation Physics Boca Raton, Florida Most
More informationdoi: /j.ijrobp
doi:10.1016/j.ijrobp.2005.05.021 Int. J. Radiation Oncology Biol. Phys., Vol. 63, No. 2, pp. 577 583, 2005 Copyright 2005 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/05/$ see front
More informationVerification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film
Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,
More informationSBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session
Stereotactic Body Radiation Therapy Quality Assurance Educational Session J Perks PhD, UC Davis Medical Center, Sacramento CA SBRT fundamentals Extra-cranial treatments Single or small number (2-5) of
More informationph fax
This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY
More informationFollow this and additional works at: Part of the Medicine and Health Sciences Commons
Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2012 EVALUATION OF THE EFFECTIVENESS OF ANISOTROPIC ANALYTICAL ALGORITHM
More informationGamma analysis dependence on specified low-dose thresholds for VMAT QA
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Gamma analysis dependence on specified low-dose thresholds for VMAT QA Ji-Hye Song, 1 Min-Joo Kim, 1 So-Hyun Park, 1,2 Seu-Ran Lee,
More informationDevelopment and evaluation of an end-to-end test for head and neck IMRT with a novel multiple-dosimetric modality phantom
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Development and evaluation of an end-to-end test for head and neck IMRT with a novel multiple-dosimetric modality phantom Viatcheslav
More informationMVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011
MVCT Image Guidance and QA Robert Staton, PhD DABR MD Anderson Cancer Center Orlando ACMP Annual Meeting 2011 Disclosures MDACCO has received grant funding from TomoTherapy, Inc. Overview TomoTherapy MVCT
More informationAutomated Plan Quality Check with Scripting. Rajesh Gutti, Ph.D. Clinical Medical Physicist
Automated Plan Quality Check with Scripting Rajesh Gutti, Ph.D. Clinical Medical Physicist Veera.Gutti@BSWHealth.org Outline Introduction - BSW Automation in Treatment planning Eclipse Scripting API Script
More informationA treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse
Peters et al. Radiation Oncology 2014, 9:153 RESEARCH Open Access A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse Samuel Peters
More informationClinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations
Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.111 pissn 2508-4445, eissn 2508-4453 Clinical Implications of High Definition Multileaf Collimator
More informationGuidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy
Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy VERSION 2.1 April 2015 Table of Contents Abbreviations & Glossary... 3 Executive Summary...
More informationReliability of the gamma index analysis as a verification method of volumetric modulated arc therapy plans
Park et al. Radiation Oncology (2018) 13:175 https://doi.org/10.1186/s13014-018-1123-x RESEARCH Open Access Reliability of the gamma index analysis as a verification method of volumetric modulated arc
More information