To Shield or Not to Shield? Lincoln L. Berland, M.D.
|
|
- Suzanna Clarke
- 6 years ago
- Views:
Transcription
1 To Shield or Not to Shield? Lincoln L. Berland, M.D.
2 Disclosures Consultant to: Nuance, Inc. Page 2
3 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation. Radiation to the breast on CT pulmonary angiography may be times higher than 2-view screening mammogram ( mgy). Bismuth shield is designed to, and is proven to decrease breast exposure on CT. Page 3
4 Concepts However, is there an alternative that can achieve an equivalent reduced radiation exposure with at least equivalent image quality on chest CT? What are the advantages and disadvantages of various strategies? Page 4
5 Our Phantom Study Lincoln L. Berland, M.D., F.A.C.R. Michael Yester, Ph.D.
6 Methods Compared strategies of using Bismuth breast shield with standard radiation exposure factors vs. no shield with reduced ma. Chest region of an anthropomorphic phantom, without and with Bismuth breast shield Phantom alone Phantom with 1 additional 3.5 cm simulated fat layer Phantom with 2 additional 3.5 cm simulated fat layers Page 6
7 Methods mas adjusted to achieve nearly equivalent standard deviations in the mediastinum without and with the shield. CT attenuation numbers recorded from: Mediastinum Superficial soft tissues Page 7
8
9
10 Results: Without and With Shield Effect on radiation exposure of using shield without modifying technique - constant ma. No Shield With Shield ma Mediastinal SD Dose Reduction- Surface Dose Reduction - Mediastinum % 20-25% Page 10
11 Equivalent Radiation Without Shield What reduced ma without shield is calculated to result in noise (based on SD) in mediastinum similar to routine ma with shield? What reduced ma without shield actually achieves such noise? No Shield - Calculated No Shield - Actual ma Mediastinal SD Dose Reduction- Surface and Mediastinum % % Page 11
12 Radiation at Surface In each case, using noise in mediastinum with shield to calculate what lower ma should be used without a shield suggests the need for a higher exposure than is necessary to achieve the same noise level. Therefore, noise is higher with a shield than one would expect based on the actual dose reduction. These higher noise levels with the shield are not visible as streak artifacts and may be the result of the shield attenuation in both AP and PA orientations. Page 12
13 Radiation at Mediastinum For equivalent noise within mediastinum, radiation dose reduction within mediastinum: No fat rings One added ring Two added rings No Shield, lower ma 42% 41% 40% Shield, routine ma 20-25% 18% 15% Page 13
14 Hounsfield Unit Variation Attenuation in mediastinum varied 3 HU between shield, no shield Surface attenuation 120 kvp No fat rings One added ring Two added rings No Shield -5 HU -69 HU -69 HU Shield 27 HU -39 HU -44 HU Page 14
15 Streak Artifacts Shield creates streak artifacts, particularly marked at upper and lower edges. Page 15
16 Why Do We Expect the Shield to Work? Why is the use of shielding an attractive concept? Use of lead shields ingrained in our training and seems intuitive However, fundamental difference: Lead shields are intended to attenuate radiation for areas in patients or for people who are not being examined, but Bismuth shields intended to limit radiation to areas that are being examined Page 16
17 Why Doesn t the Shield Work? Shield attenuates photons both entering (from the front), but also exiting (through the front), wasting radiation Causes beam-hardening, streak artifacts, HU changes Page 17
18 Why Doesn t the Shield Work? Does the shield selectively attenuate low-energy photons? No. Preinstalled filters attenuate low-energy photons before reaching the patient or shield. Spectrum from one manufacturer: mean photon energy of 69.1 kev for a 120-kV, but shields only increase the mean energy to 70.5 kev kev. McCollough CH, Wang J, Berland LL. Bismuth Shields for CT Dose Reduction: Do They Help or Hurt? Journal of the American College of Radiology. Volume 8, Issue 12, Pages , December 2011 Page 18
19 May confuse AEC programs Why Doesn t the Shield Work? If place shield before radiograph, AEC compensates with increased exposure If place shield after radiograph, exposure is too low, decreasing image quality If real-time AEC, then will partially compensate with increased exposure when shield in field, whether you put the shield on before or after scanogram Do you know if your system has real-time AEC? Page 19
20 Why Doesn t the Shield Work? Four other phantom studies all led to same conclusion: Geleijns J, Wang J, McCollough. The use of breast shielding for dose reduction in pediatric CT: arguments against the proposition. Pediatr Radiol 40: , 2010 Vollmar SV, Kalender WA. Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations. Eur Radiol 18: , 2008 Wang J, Duan X, Christner JA, Leng S, Yu L, McCollough CH. Radiation dose reduction to the breast in thoracic CT: Comparison of bismuth shielding, organ-based tube current modulation, and use of a globally decreased tube current. Med Phys 38: , 2011 Wang J, Duan X, Christner JA, Leng S, Grant KL, McCollough CH. Bismuth Shielding, Organ-based Tube Current Modulation, and Global Reduction of Tube Current for Dose Reduction to the Eye at Head CT. Radiology 262: , 2012 Page 20
21 Why Not Reduce Dose and Add Shield? Dose reduction is linear proportional to mas. Theoretical example: Standard, no shield: 180 ma 120 ma, no shield decreases breast dose 33% 180 ma, with shield decreases breast dose 33% Theoretical reduced dose plus shield: 120 ma, with shield, reduces dose to 44% of original 80 ma, no shield also reduces dose to 44% of original, better noise, better sectional dose profile, no artifacts, etc. Page 21
22 Disadvantages of Shield Same or lower noise in mediastinum can be obtained with the same or lower exposure to breasts without shield using a reduced ma as with shield using a baseline ma Similar results with added simulated fat rings Substantially lower radiation to mediastinum and remainder of region with lower ma without the shield Shield alters attenuation values Shield creates streak artifacts In field: deep to shield At upper and lower edges Worse if any motion Page 22
23 Other Disadvantages of Shield Proper use depends on knowing how your manufacturer, model of scanner and version of software perform AEC Have to stop after scout and place the shield Costs money Need to use infection control techniques because comes in contact with patient Page 23
24 Recommendations Chest CT Identify population of patients for whom decreased breast dose desired: E.g. women under 50 yo, without terminal condition Reduce ma by 30%, do not use a shield Page 24
25 Breast Exposure on Abdominal CT Breasts sometimes in field, depending on: Technologist upper range Diaphragm position Breast size Concerns with using breast shield on abd CT: Causes streaks through upper abdomen, particularly at edge of shield Alters attenuation Page 25
26 Use of shield depends on: Breast, diaphragm location Conclusions Abdominal CT Patient comorbidities and age Need for high-quality images through upper liver Other measures to limit breast radiation: Lead apron or shield can protect from overranging if breasts not in field Can keep bra on or retract breasts above field Page 26
27 Recommendations Abdominal CT Train and encourage technologists to limit overranging Keep bra on if not also scanning chest, attempt to retract breasts If possible, reduce ma 30% for upper component of abdominal scan Page 27
28 Challenges Although several studies suggest that lower dose chest CT is diagnostically acceptable, there are no specific standards for appropriate ma for chest CT. Challenge is to design low-dose protocols and procedures that assure use of appropriately lower ma for chest CT in vulnerable women. Use of Bismuth or lead shield for abdominal CT depends on multiple factors Page 28
Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?
Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT
More informationAsk EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT
Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Shielding in pediatric CT Claudio Granata (IRCCS Istituto Giannina Gaslini, IT) Joana Santos (ESTeSC-Coimbra Health School, PT) Elina
More informationTranslating Protocols Across Patient Size: Babies to Bariatric
Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo
More informationAsk EuroSafe Imaging Tips & Tricks. CT Working Group
Ask EuroSafe Imaging Tips & Tricks CT Working Group Organ Based Tube Current Modulation to Reduce Radiation Dose to Superficial Radiosensitive Organs Eileen Kelly (Galway University Hospitals, IE) Matthias
More informationESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine
ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol
More informationAcknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols
Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic
More informationManaging Radiation Risk in Pediatric CT Imaging
Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist
More informationX-Ray & CT Physics / Clinical CT
Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED
More informationAutomated CT Protocol Design Advantages and Pitfalls of Algorithm-Based Technique Selection in Pediatrics. Disclosures 7/22/2014. Learning Objectives
Automated CT Protocol Design Advantages and Pitfalls of Algorithm-Based Technique Selection in Pediatrics Robert MacDougall, M.Sc. Department of Radiology Boston Children s Hospital Disclosures 2 Learning
More informationDianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX
Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Learning Objectives: Limitations for estimating patient dose for CT Methods
More informationAccounting for Imaging Dose
Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some
More informationCT angiography techniques. Boot camp
CT angiography techniques Boot camp Overview Basic concepts Contrast administration arterial opacification Time scan acquisition during the arterial phase Protocol examples Helical non-gated CTA Pulmonary
More informationDual-Energy CT: The Technological Approaches
Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure
More informationPaediatric Dose Reduction and Image Quality
Paediatric Dose Reduction and Image Quality Alan Whiteside The majority of this work was undertaken as part of MSc Thesis of Helen Dixon. Introduction Paediatric CT protocols result in a higher effective
More informationEstimating Iodine Concentration from CT Number Enhancement
Estimating Iodine Concentration from CT Number Enhancement Rosemary Eaton, Andrew Shah, Jane Shekhdar Medical Physics, Mount Vernon Hospital CT Users Group 4 th October 212, Edinburgh Summary Background
More informationRADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography
More informationCURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC
CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,
More informationDual-Energy CT Applications in Radiation Therapy
THE UNIVERSITY OF WISCONSIN MADISON Dual-Energy CT Applications in Radiation Therapy - Jessica Miller 1 Disclosures Funding provided by Siemens Medical 2 Learning objectives General principles of dual
More informationDose Reduction of the Female Breast in Chest CT
Cardiopulmonary Imaging Original Research Seidenfuss et al. Reduction of CT Radiation Dose to Female Breast Cardiopulmonary Imaging Original Research Downloaded from www.ajronline.org by 46.3.197.9 on
More informationDoses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?
Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.
More informationTo Shield or Not to Shield: Application of Bismuth Breast Shields
Cardiopulmonary Imaging Opinion Colletti et al. Bismuth Breast Shields Cardiopulmonary Imaging Opinion FOCUS ON: Patrick M. Colletti 1 Orlando A. Micheli Kai H. Lee Colletti PM, Micheli OA, Lee KH Keywords:
More informationBreast Dose Reduction Options During Thoracic CT: Influence of Breast Thickness
Cardiopulmonary Imaging Original Research Revel et al. Breast Dose Reduction in Thoracic CT Cardiopulmonary Imaging Original Research Marie-Pierre Revel 1 Isabelle Fitton 2 Etienne Audureau 3 Joseph Benzakoun
More informationDual Energy CT Aortography: Can We Reduce Iodine Dose??
Dual Energy CT Aortography: Can We Reduce Iodine Dose?? William P. Shuman MD, FACR FSCBTMR Department of Radiology University of Washington SCBTMR Annual Course Boston, October 10, 2012 Conflict of Interest
More informationModelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography
The British Journal of Radiology, 81 (2008), 499 503 SHORT COMMUNICATION Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography 1 G R IBALL, MSc, BSc, 2
More informationBone Densitometry Radiation dose: what you need to know
Bone Densitometry Radiation dose: what you need to know John Damilakis, PhD Associate Professor and Chairman University of Crete, Iraklion, Crete, GREECE Estimation of bone status using X-rays Assessment
More informationDental ConeBeam Computed Tomography (CBCT) X-ray Systems
Dental ConeBeam Computed Tomography (CBCT) X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL CONEBEAM COMPUTED TOMOGRAPHY (CBCT) X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability.
More informationDual Energy CT of the Liver
34th Annual Course October 2011 Washington, DC Dual Energy CT of the Liver Vassilios Raptopoulos, MD Beth Israel Deaconess Medical Center Harvard Medical School Dual Energy CT (DECT) Different materials
More informationAsk EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose
Ask EuroSafe Imaging Tips & Tricks CT Working Group Optimization of scan length to reduce CT radiation dose Alban Gervaise (Centre Hospitalier Universitaire Nancy, FR) Mika Kortesniemi (HUS Medical Imaging
More informationRadiation Dose Estimations to the Thorax Using Organ-Based Dose Modulation
Cardiopulmonary Imaging Original Research Lungren et al. Organ-Based Dose Modulation Cardiopulmonary Imaging Original Research Matthew P. Lungren 1 Terry T. Yoshizumi 1 Samuel M. Brady 1 Greta Toncheva
More informationHow to Develop CT Protocols for Children
How to Develop CT Protocols for Children Introduction Prior to 2001 the vast majority of CT imaging of children was conducted using the same or similar techniques used for adult imaging. In 2001, several
More information3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)
Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Rebecca Milman Marsh, Ph.D. University of Colorado Department of Radiology Who in the Facility Works
More informationPreparing for Medical Physics Components of the ABR Core Examination
Preparing for Medical Physics Components of the ABR Core Examination The ABR core examination for radiologists contains material on medical physics. This content is based on the medical physics that is
More informationIntroduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT
Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT protocols Tips Leukaemia / lymphoma: ~ 35% acute lymphoblastic
More informationTrue Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens
DEfinitely Siemens True Dual Energy Dr. Stefan Ulzheimer, Siemens Healthcare GmbH International version. Not for distribution in the US. Unrestricted Siemens AG 2015 All rights reserved. The products/features
More informationOrgan-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT
Medical Physics and Informatics Original Research Hoang et al. Radiation Dose Reduction for Neck CT Medical Physics and Informatics Original Research Jenny K. Hoang 1,2 Terry T. Yoshizumi 1 Kingshuk Roy
More informationDual-Energy 101: Principles, Methods and Dose
Dual-Energy 101: Principles, Methods and Dose Juan Carlos Ramirez-Giraldo, Ph.D Staff Scien2st, Collabora2ons Manager SE Region ISCT San Francisco, 2017 Siemens Medical Solu2ons USA, Inc., 2017 Page 1
More informationEffects of shielding the radiosensitive superficial organs of ORNL pediatric phantoms on dose reduction in computed tomography
Original Article Effects of shielding the radiosensitive superficial organs of ORNL pediatric phantoms on dose reduction in computed tomography Parisa Akhlaghi, Hashem Miri Hakimabad, Laleh Rafat Motavalli
More informationDebra Pennington, MD Director of Imaging Dell Children s Medical Center
Debra Pennington, MD Director of Imaging Dell Children s Medical Center 1 Gray (Gy) is 1 J of radiation energy/ 1 kg matter (physical quantity absorbed dose) Diagnostic imaging doses in mgy (.001 Gy)
More information8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION
Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD
More informationPatient Dose Estimates. from CT Examinations. Patient Dose Estimates
Patient Dose Estimates from CT Examinations Patient Dose Estimates from CT Examinations John M. Boone, Ph.D. Professor and Vice Chairman of Radiology Professor of Biomedical Engineering University of California,
More informationComputed tomography Acceptance testing and dose measurements
Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed
More informationSTRUCTURED EDUCATION REQUIREMENTS EFFECTIVE: JANUARY 1, 2016
Computed Tomography The purpose of structured education is to provide the opportunity for individuals to develop mastery of discipline-specific knowledge that, when coupled with selected clinical experiences,
More informationThe Computed Tomography Examination
CONTENT SPECIFICATIONS The Computed Tomography Examination The purpose of The American Registry of Radiologic Technologists (ARRT ) Computed Tomography Examination is to assess the knowledge and cognitive
More informationBICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives
BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives
More informationComparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck
Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Young Hen Lee 1 Eun-tae Park 1 Pyong Kon Cho 1 Hyung Suk Seo 1 Bo-Kyung Je 1 Sang-il Suh 2 Kyung-Sook
More informationMetal Artifact Reduction by Dual Energy CT
Metal Artifact Reduction by Dual Energy CT Poster No.: C-0108 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper T. Johnson, F. Bamberg, A. Dierks, H.-C. Becker, M. F. Reiser; Munich/DE
More informationMedical Diagnostic Imaging
Medical Diagnostic Imaging Laboratories Medical Diagnostic Imaging Lab Name Location Person in Charge Programs Served Courses Served Patient Care and Management (2) Introduction to MDI Radiographic Technique
More informationBICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives
BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives
More informationWhat s New in Breast Imaging. Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia
What s New in Breast Imaging Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia Disclosure Hologic, Inc. Shareholder and research agreement. Volpara Solutions, Ltd. Shareholder
More informationHow do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review
How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review Mannudeep K. Kalra, MD, DNB Massachusetts General Hospital Harvard Medical School Financial Disclosure This presentation
More informationImplementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL
Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau
More informationDose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation
Medical Physics and Informatics Original Research Angel et al. Radiation Dose During Chest CT Medical Physics and Informatics Original Research Erin Angel 1,2 Nazanin Yaghmai 1 Cecilia Matilda Jude 1 John
More information8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center
How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center Mannudeep K. Kalra, MD Webster Center for Quality and Safety Massachusetts General Hospital Harvard Medical School
More informationBICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:
BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:
More informationAutomatic Patient Centering for MDCT: Effect on Radiation Dose
Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans
More informationManaging Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION
Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,
More informationAn Introduction to Dual Energy Computed Tomography
An Introduction to Dual Energy Computed Tomography Michael Riedel University of Texas Health Science Center at San Antonio Introduction The idea of computed tomography (CT) was first introduced in the
More informationSPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School
SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School OUTLINE 1 st Presentation: Dose optimization strategies
More informationReducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters
Medical Physics and Informatics Review Maldjian and Goldman Reducing Radiation Dose in Body CT Medical Physics and Informatics Review FOCUS ON: Pierre D. Maldjian 1 Alice R. Goldman Maldjian PD, Goldman
More informationCT Radiation Risks and Dose Reduction
CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children
More informationCone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood
Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex
More information1 Course Syllabus + Study Guide for Lecture and Laboratory
1 Course Syllabus + Study Guide for Lecture and Laboratory /BERGEN COMMUNITY COLLEGE Division of Health Professions/Radiography Program Fall 2014 A. General Course Information Title: Radiography I Credits:
More informationLens Dose in Routine Head CT: Comparison of Different Optimization Methods With Anthropomorphic Phantoms
Medical Physics and Informatics Original Research Nikupaavo et al. Lens Dose in Head CT Medical Physics and Informatics Original Research Ulla Nikupaavo 1,2 Touko Kaasalainen 1,3 Vappu Reijonen 1 Sanna-Mari
More informationBreast CT and Dosimetry
2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of
More informationDocumentation, Codebook, and Frequencies
Documentation, Codebook, and Frequencies Dual-Energy X-ray Absorptiometry Femur Bone Measurements Examination Survey Years: 2005 to 2006 SAS Transport File: DXXFEM_D.XPT January 2009 NHANES 2005 2006 Data
More informationA Snapshot on Nuclear Cardiac Imaging
Editorial A Snapshot on Nuclear Cardiac Imaging Khalil, M. Department of Physics, Faculty of Science, Helwan University. There is no doubt that nuclear medicine scanning devices are essential tool in the
More informationCT Radiation Safety in Adults: Where are we now? What can be done?
CT Radiation Safety in Adults: Where are we now? What can be done? David A Leswick MD FRCPC University of Saskatchewan Canadian Association of Radiologists Annual General Meeting Friday April 26, 2013
More informationUltrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging
Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a
More informationDosimetric Consideration in Diagnostic Radiology
Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure
More informationRadiation dose reduction in computed tomography: techniques and future perspective
REVIEW Radiation dose reduction in computed tomography: techniques and future perspective Despite universal consensus that computed tomography (CT) overwhelmingly benefits patients when used for appropriate
More informationUniversity of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert
University of Groningen Quantitative CT myocardial perfusion Pelgrim, Gert IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check
More informationThoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters
Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Poster No.: C-2584 Congress: ECR 2015 Type: Scientific
More informationToshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery
Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,
More informationDose Reduction Options in Cardiac CT
Dose Reduction Options in Cardiac CT Doyle P, Ball P*, Donnelly P # Radiological Sciences & Imaging, Forster Green Hospital *Department of Radiology, Ulster Hospital # Department of Cardiology, Ulster
More informationEXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017
EXAMINATION CONTENT SPECIFICATIONS Computed Tomography The purpose of the computed tomography examination is to assess the knowledge and cognitive skills underlying the intelligent performance of the tasks
More informationCardiopulmonary Imaging Original Research
Cardiopulmonary Imaging Original Research Vardhanabhuti et al. Three Algorithms to Assess Standard- and Low-Dose Chest CT Cardiopulmonary Imaging Original Research FOCUS ON: Varut Vardhanabhuti 1,2 Robert
More information354 Korean J Radiol 9(4), August 2008
Review of Failed CT Phantom Image Evaluations in 2005 and 2006 by the CT Accreditation Program of the Korean Institute for Accreditation of Medical Image Hye Jung Park, MD 1 Seung Eun Jung, MD 1, 2 Young
More informationThe effect of breast shielding during lumbar spine radiography
26 research article The effect of breast shielding during lumbar spine radiography Nejc Mekis 1, Dejan Zontar 2, Damijan Skrk 2 1 University of Ljubljana, Faculty of Health Sciences, Medical Imaging and
More informationLiver, Liver Fat, Fibrosis and HCC Screening with DECT
Liver, Liver Fat, Fibrosis and HCC Screening with DECT Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org
More informationOutline. Lifetime Attributable Risk 10 mgy in 100,000 exposed persons (BEIR VII 2006) SPECT/CT and PET/CT Dosimetry
SPECT/CT and PET/CT Dosimetry Lifetime Attributable Risk 10 mgy in 100,000 exposed persons (BEIR VII 2006) All Solid Tumors Leukemia Frederic H. Fahey DSc Male Female Male Female Children s Hospital Boston
More informationRadiation Exposure 1980 to 2006
Radiation Exposure 1980 to 2006 Background 3-6 msv/yr Natural (85% 45%) Radon Cosmic Rays Air travel Living at Altitude Man-made (15% 55%) Medical Imaging** mgy Radiation Therapy cgy Radiation Whole Body
More informationManaging the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University
Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm
More informationSenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski
SenoBright Contrast Enhanced Spectral Mammography Technology Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski White Paper October 2012 SenoBright Contrast Enhanced Spectral Mammography Technology Ann-Katherine
More informationFused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience
Original Article Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Shun-Yu Gao, Xiao-Peng Zhang, Yong Cui, Ying-Shi
More informationContrast-Enhanced Digital Mammography
2015 ARRS Breast Symposium Contrast-Enhanced Digital Mammography John Lewin, M.D. Diversified Radiology of Colorado CEDM - Outline History Technique Literature Review / Cases Clinical Status Inexpensive,
More informationCollapsed Cone Convolution 2D illustration
Collapsed Cone Convolution 2D illustration 8 cones Energy desposition decreases very quickly with distance Energy is absorber in blue pixels only. IGRT1 technologies Paweł Kukołowicz Warsaw, Poland IGRT
More informationBreast Cancer PET/CT Imaging Protocol
Breast Cancer PET/CT Imaging Protocol Scanning Protocol: Patients are scanned from the top of the neck through the pelvis. Arms-up position is used to avoid beam-hardening artifact in the chest and abdomen.
More informationSimon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3
Coronary Artery Bypass Graft Imaging with CT Angiography and Iterative Reconstruction: Quantitave Evaluation of Radiation Dose Reduction and Image Quality Simon Nepveu 1, Irina Boldeanu 1, Yves Provost
More informationPatti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety
Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February 2008. Radiation Safety Sub -headings Background Radiation Effects of Radiation Safe Levels Effective Doses ALARA Principle Radiation
More informationGender differences in CT calcium scoring: A phantom study
Gender differences in CT calcium scoring: A phantom study Nicholas Petrick, Qin Li, Benjamin Berman, Marios A Gavrielides, Rongping Zeng, Berkman Sahiner CDRH/OSEL/DIDSR U.S. Food and Drug Administration
More informationPatient / Organ Dose in CT
Patient / Organ Dose in CT Patient specific and organ dose estimation H.D. Nagel Dr. HD Nagel, Science & Technology for Radiology Buchholz / Germany www.sascrad.com 1 Topics CTDI & patient dose SSDE Organ
More informationCT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings
CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings Poster No.: C-3289 Congress: ECR 2010 Type: Scientific Exhibit Topic: Chest - Your latest results Authors: M. K. Gill,
More informationUSING TEAM AUDITS TO IMPROVE IMAGE QUALITY: Our Institution s Experience
USING TEAM AUDITS TO IMPROVE IMAGE QUALITY: Our Institution s Experience Lincoln L. Berland, MD Deborah D. Flint, PE, BIE, MBA Beth S. Winningham, BSRT, MBA UAB Health System 900-bed Facility and Level
More informationWhole Body CT Protocol Update 2018
Whole Body CT Protocol Update 2018 10 th Nordic Course in Trauma Radiology Gothenburg, Sweden K.SHANMUGANATHAN M.D. Disclosure of Commercial Interest Neither I nor my immediate family members have a financial
More informationMedical Physics and Informatics Original Research
Medical Physics and Informatics Original Research Christner et al. Estimating Effective Dose for CT Medical Physics and Informatics Original Research FOCUS ON: Jodie A. Christner 1 James M. Kofler Cynthia
More informationEffective Utilization of Imaging. John V. Roberts, M.D. Premier Radiology Abdominal Imaging
Effective Utilization of Imaging John V. Roberts, M.D. Premier Radiology Abdominal Imaging Safety Contrast and Radiation What to order Abdomen/Pelvis Brain/Spine Chest Musculoskeletal Ob/Gyn Head and Neck
More informationPitfalls of the Pediatric Chest and Abdomen SPR 2017
Pitfalls of the Pediatric Chest and Abdomen SPR 2017 Richard I. Markowitz, MD, FACR Children s Hospital of Philadelphia Perelman School of Medicine University of Pennsylvania No Disclosures Cognitive Perceptual
More informationWhy is CT Dose of Interest?
Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase
More informationCT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell
CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction
More information