1 St. Jude Medical External Pulse Generator Model 3599 Clinician's Manual

2 CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or one of its subsidiaries. ST. JUDE MEDICAL and the nine-squares symbol are trademarks and service marks of St. Jude Medical, Inc. and its related companies St. Jude Medical, Inc. All Rights Reserved. For a listing of patents for St. Jude Medical neuromodulation products, visit

3 Contents Prescription and Safety Information... 1 Indications for Use... 1 Contraindications... 1 Warnings... 2 Precautions... 4 Adverse Effects... 7 Product Description... 8 Overview of the EPG and EPG Header...10 Package Contents...12 Getting Started Overview of the EPG Procedure...13 Replacing the EPG Batteries...14 Checking the Status of the EPG Batteries...18 Setting Up Communication Between the EPG and the Clinician Programmer and Patient Controller...19 EPG Indicator Light...20 Starting and Stopping Stimulation...20 i

4 Using the EPG Compatibility Guidelines Connecting the Multilead Trial Cable to the EPG Disconnecting the Multilead Trial Cable from the EPG Securing the Leads or Extensions Connecting the EPG Header to the EPG Connecting the Leads or Extensions to the EPG Header Securing the EPG Programming the EPG Caring for the EPG Handling the EPG Cleaning the EPG Storing and Transporting the EPG Service and Ordering Information Technical Support Ordering Information Appendix A: Product Specifications EPG with Header Appendix B: Regulatory Statements Battery Disposal Guidelines ii

5 Statement of FCC Compliance...38 Statement of Compliance With License-Exempt RSS Standard (Canada)...39 Statement of Compliance With Interference-Causing Equipment Standard (Canada)...39 Identification Information for Product Registration...39 Product Classification Statement (CISPR 11, Class B)...40 Wireless Technology Information...40 Radio Transmitter, Cables, Transducers...41 Quality of Service for Wireless Technology...42 Appendix C: Electromagnetic Compatibility Guidelines Radio Frequency Information...54 Appendix D: Symbols and Definitions iii

6

7 Prescription and Safety Information Read this section to gather important prescription and safety information. Indications for Use The St. Jude Medical invisible neuromodulation trial system is indicated as an aid in the management of chronic, intractable pain of the trunk and/or limbs, including unilateral or bilateral pain associated with any of the following: failed back surgery syndrome and intractable low back and leg pain. The St. Jude Medical invisible neuromodulation trial system is intended to be used externally with St. Jude Medical-approved trial leads and extensions to deliver trial stimulation for a maximum of 30 days. Contraindications This neurostimulation system is contraindicated for patients who are unable to use the system. Neurostimulation should not be used on patients who are poor surgical risks, such as those with multiple illnesses or active general infections. 1

8 Warnings The following warnings apply to these components. Other system components. Refer to the individual system component manuals for additional warnings and precautions related to those devices. Diathermy therapy. Do not use short-wave diathermy, microwave diathermy, or therapeutic ultrasound diathermy (all now referred to as diathermy) on patients implanted with a neurostimulation system. Energy from diathermy can be transferred through the implanted system and cause tissue damage at the location of the implanted electrodes, resulting in severe injury or death. Diathermy is further prohibited because it may also damage the neurostimulation system components. This damage could result in loss of therapy, requiring additional surgery for system implantation and replacement. Injury or damage can occur during diathermy treatment whether the neurostimulation system is turned on or off. Explosive or flammable gases. Do not use the stimulator in an environment where explosive or flammable gas fumes or vapors are present, including hyperbaric chambers. The operation of the stimulator could cause them to ignite, causing severe burns, injury or death. Electrosurgery devices. Electrosurgery devices should not be used in close proximity to an implanted neurostimulation device. Contact between an active electrode and an implanted lead or extension can cause direct stimulation of the tissue at the electrode site and cause severe injury to the patient. If use of electrocautery is necessary, first turn off the neurostimulation system. 2

9 Implanted cardiac system. Physicians need to be aware of the risk and possible interaction between a neurostimulation system and an implanted cardiac system, such as a pacemaker or defibrillator. Electrical pulses from a neurostimulation system may interact with the sensing operation of an implanted cardiac system, causing the cardiac system to respond inappropriately. To minimize or prevent the implanted cardiac system from sensing the output of the neurostimulation system, (1) maximize the distance between the implanted systems; (2) verify that the neurostimulation system is not interfering with the functions of the implanted cardiac system; and (3) avoid programming either device in a unipolar mode (using the device s can as an anode) or using neurostimulation system settings that interfere with the functions of the implantable cardiac system. Magnetic resonance imaging (MRI). Patients with an EPG should not be subjected to MRI. Even if the leads are removed the patient should not have an MRI if any part of a lead is still implanted. Lead movement. Patients should avoid bending, twisting, stretching, or lifting objects over five pounds for six to eight weeks postimplantation. Extension of the upper torso or neck may cause lead movement and alter the stimulation field, resulting in overstimulation or ineffective stimulation. Operation of machines, equipment, and vehicles. Patients should not drive or operate heavy machinery or power tools with the stimulator turned on. Postural changes or abrupt movement could cause overstimulation (jolting sensation) that might cause patients to lose control of their vehicle or equipment. Postural changes. Changes in posture or abrupt movements can change the level of stimulation and potentially cause unpleasant sensations. Patients should turn stimulation off or lower the amplitude 3

10 before stretching, lifting their arms over their head, or exercising. If unpleasant sensations occur, turn off stimulation. Pediatric use. Safety and effectiveness of neurostimulation for pediatric use have not been established. Pregnancy and nursing. Safety and effectiveness of neurostimulation for use during pregnancy and nursing have not been established. System components. The use of non-st. Jude Medical components with this system may result in damage to the system and increased risk to the patient. Power supply. Use only the CR2450 batteries supplied with the device. CR2450 batteries have a nominal voltage of 3V and nominal capacity of at least 600 mah. Precautions The following precautions apply to these components. General Precautions Clinician Training. Clinicians should be experienced in implanting neurostimulation devices and should have undergone sufficient surgical and device implantation training. Patient Selection. It is extremely important to select patients appropriately for neurostimulation. Thorough psychiatric screening should be performed. Patients should not be dependent on drugs and should be able to operate the neurostimulation system. Infection. Follow proper infection control procedures and inform patients to avoid showering and 4

11 touching the bandages. Theft detectors and metal screening devices. Certain types of antitheft devices, such as those used at entrances or exits of department stores, libraries, and other public establishments, and airport security screening devices may affect stimulation. Patients who are implanted with nonadjacent multiple leads and patients who are sensitive to low stimulation thresholds may experience a momentary increase in their perceived stimulation, which has been described by some patients as uncomfortable or jolting. Patients should use caution when approaching such a device and should request assistance to bypass the device. If they must proceed through the device, patients should turn off the EPG and proceed with caution, being sure to move through the detector quickly. Implantation of multiple leads. If multiple leads are implanted, leads and extensions should be routed in close proximity. Nonadjacent leads can possibly create a conduit for stray electromagnetic energy that could cause the patient unwanted stimulation. Handling and Implementation Single use only. The EPG header is intended for single use only. Exposure to liquids. Device components should be stored where they will not be exposed to liquids or excessive moisture, which can damage the package materials and components. Care and handling of components. Use extreme care when handling system components prior to use. Excessive heat, traction, bending, and twisting, or the use of sharp instruments may damage components and cause them to fail. 5

12 Package or component damage. Do not use the device if the package or components show signs of damage. Return any suspect components to St. Jude Medical for evaluation. System testing. To ensure correct operation, the system should always be tested before the patient leaves the surgery suite. Device modification. The equipment is not serviceable by the customer. To prevent injury or damage to the system, do not modify the equipment. If needed, return the equipment to St. Jude Medical for service. Component disposal. Dispose of the EPG header and pouch with other medical waste. Return the EPG to St. Jude Medical for safe disposal when necessary. Hospital and Medical Environments Therapeutic radiation. Therapeutic radiation may damage the electronic circuitry of a neurostimulation system, although no testing has been done and no definite information on radiation effects is available. Sources of therapeutic radiation include therapeutic X rays, computerized tomography (CT) scans, cobalt machines, and linear accelerators. If radiation therapy is required, disconnect the EPG from the leads or extensions. Damage to the system may not be immediately detectable. Home and Occupational Environments Electromagnetic interference (EMI). Certain commercial electrical equipment (for example, arc welders, induction furnaces, and resistance welders), communication equipment (for example, microwave transmitters, linear power amplifiers, and high power amateur transmitters), high voltage 6

13 power lines, and radio-frequency identification (RFID) devices may generate sufficient EMI to interfere with the operation of the neurostimulation system if approached too closely. Keep the neurostimulation system at least 10 cm (4 in) away from RFID devices. Mobile phones. The effect of mobile phones on neurostimulation systems is unknown; patients should avoid placing mobile phones directly over the system. Adverse Effects The use of a neurostimulation system involves risks. In addition to those risks commonly associated with surgery, the following risks are also associated with use of a neurostimulation system: Unpleasant sensations or motor disturbances, including involuntary movement, caused by stimulation at high outputs (If either occurs, turn off stimulation immediately.) Undesirable changes in stimulation, which may be related to cellular changes in tissue around the electrodes, changes in electrode position, loose electrical connections, or lead failure Stimulation in unwanted places Radicular chest wall stimulation Lead migration, causing changes in stimulation or reduced pain relief Epidural hemorrhage, hematoma, infection, spinal cord compression, or paralysis from placement of a lead in the epidural space Cerebrospinal fluid (CSF) leakage Paralysis, weakness, clumsiness, numbness, or pain below the level of the implant 7

14 Persistent pain at the electrode site Seroma (mass or swelling) at the implant site Allergic or rejection response to device materials Product Description The St. Jude Medical External Pulse Generator (EPG) is an external trial generator that when connected to trial neurostimulation leads delivers therapeutic stimulation to parts of the body. The EPG communicates wirelessly with system programmers and controllers. The EPG header connects directly to the EPG for a neurostimulation trial. The EPG header can hold one or more leads or extensions with up to 16 electrodes total. NOTE: Patients may experience a sensation of tingling, pins and needles, prickling, or even burning called paresthesia. Paresthesia may be brief or it may last a long time. NOTE: For more information about the neurostimulation system, see the clinician s programming manual for this system. 8

15 NOTE: The EPG is supplied nonsterile and is intended to be used multiple times. The EPG header is also supplied nonsterile, but is single use only. NOTE: In this document, the term "clinician programmer" refers to the St. Jude Medical Clinician Programmer device, "patient controller" refers to the St. Jude Medical Patient Controller device, "clinician programmer app" refers to the St. Jude Medical Clinician Programmer app, and "patient controller app" refers to the St. Jude Medical Patient Controller app. 9

16 Overview of the EPG and EPG Header Refer to the following figures for the EPG and EPG header features: Figure 1. EPG features 1. Connector for multilead trial cable and EPG header 2. Battery compartment panel 3. Indicator light 10

17 Figure 2. EPG header features 1. Button 2. Port 1, electrodes Port 2, electrodes

18 Package Contents In addition to the product documentation, the External Pulse Generator, 16 channel (Model 3599) contains the following items: 1 EPG (Model 3599) 1 EPG header cap 1 battery door In addition to the product documentation, the External Pulse Generator, 2 Port Header (Model 3032) contains the following items: 1 EPG header 2 CR2450 Batteries Cleaning cloths 1 EPG pouch with adhesive 12

19 Getting Started This section provides a general overview of the EPG procedure and instructions for tasks, such as replacing the batteries and setting up communication between the EPG and the clinician programmer and patient controller. Overview of the EPG Procedure The following steps provide a general overview of the EPG procedure. Refer to the following sections for more information about each of these steps. 1. Replace the EPG batteries. 2. Set up communication between the EPG and the clinician programmer and patient controller. 3. Implant the leads or extensions. 4. Connect the leads or extensions to the multilead trial cable. 5. Connect the multilead trial cable to the EPG. 6. Test the system. 7. Disconnect the multilead trial cable from the EPG. 8. Disconnect the multilead trial cable from the leads or extensions. 9. Secure the leads or extensions and bandage the lead entry site. 10. Connect the EPG header to the EPG. 13

20 11. Insert the leads or extensions into the EPG header. 12. Place the EPG into the pouch and secure to the patient. 13. Program the system. Replacing the EPG Batteries The EPG uses two nonrechargeable CR2450 coin cell batteries. St. Jude Medical recommends replacing the batteries before starting a new trial and only using the batteries included in the External Pulse Generator, 2 Port Header. NOTE: Dispose of batteries safely. Do not treat them as household waste. Refer to the guidelines in Appendix B for more information. To replace the batteries, follow these steps: 1. Disconnect the EPG header or multilead trial cable from the EPG if either is connected. Also, if the EPG is communicating with the clinician programmer or patient controller, end the session. Refer to the clinician s programming or user s controller manual for instructions. 2. Press and slide the battery compartment panel outward to open the battery compartment and remove the panel. 14

21 Figure 3. Battery compartment panel 3. Remove the batteries by either: - turning the EPG over so the battery compartment is facing down and lightly tapping the EPG in the palm of your hand, or 15

22 - pressing down on one side of the battery so the opposite side pops up, lifting the battery out, and repeating the process with the remaining battery. 4. Insert a new battery into the battery compartment, paying attention to the polarity directions provided on the inside of the battery compartment. 5. Insert the second battery into the second battery compartment. Figure 4. Inserting batteries into the battery compartment 16

23 6. Place the battery compartment panel flush to the EPG, lining up the tabs with the coordinating indentation. Figure 5. Replacing the battery compartment panel 7. Slide the panel closed flush with the EPG until it snaps into place. NOTE: When you insert the batteries, the indicator light is solid. You must wait until the indicator light is flashing or off to set up communication. NOTE: If you replace the batteries but the generator battery indicator on the clinician programmer app or patient controller app indicates the batteries are low, make sure the batteries are inserted correctly. 17

24 Checking the Status of the EPG Batteries Check the status of the EPG s batteries using the clinician programmer app or the patient controller app. Refer to the clinician s programming or user s controller manual for these directions. NOTE: Battery life depends on usage and stimulation settings, with batteries lasting at least 14 days under nominal or practical high output settings, with 500 ohm impedance and stimulation on for 16 hours per day. (High output is based on actual patient use collected in St. Jude Medical clinical studies, not the maximum device output.) The following information provides general guidelines for the battery status: The battery status icon on the clinician programmer app or patient controller app shows a decreasing fill as the batteries are used. A warning will appear on the clinician programmer app or patient controller app when the batteries are critically low. Stimulation will automatically turn off when the batteries cannot support stimulation. Replace the EPG s batteries to communicate with the EPG, turn stimulation on, change stimulation settings, or retrieve stimulation usage information. NOTE: If the batteries deplete during a trial, instruct the patient to refrain from replacing the 18

25 batteries to avoid disturbing the bandaging and lead position. Setting Up Communication Between the EPG and the Clinician Programmer and Patient Controller Before starting the surgical procedure, set up communication between the EPG and the clinician programmer and patient controller. The following steps outline the procedure for setting up communication. For further instructions, refer to the clinician s programming manual. 1. Place the magnet perpendicular to and centered directly over the front of the EPG for 10 seconds. The indicator light on the EPG will start flashing. 2. Set up communication using the clinician programmer app or the patient controller app following the directions shown on the screen. After setting up communication, the EPG is paired with that clinician programmer and patient controller. You will not need to pair them again unless another patient controller or more than four clinician programmers are paired to the EPG. NOTE: Store paired EPGs and patient controllers together to reduce the number of times you need to set up communication. 19

26 EPG Indicator Light The indicator light on the EPG will flash for the following reasons: When batteries are inserted into the EPG When a communication session with the clinician programmer app or patient controller app is started When the magnet is held for approximately 10 seconds over the EPG (signaling the EPG is ready to pair with the clinician programmer or patient controller) NOTE: The EPG indicator light flashes faster when it is ready to pair than when batteries are inserted or a communication session is started. Starting and Stopping Stimulation Use the clinician programmer app or the patient controller app to start or stop stimulation. You may also use the magnet (Model 1210) to start or stop stimulation if Magnet Mode is programmed to On/Off or Off. To use the magnet, follow these steps: 1. Take the keeper bar off the magnet. 20

27 Figure 6. Magnet and keeper bar 1. Magnet 2. Keeper bar 21

28 2. Hold the magnet perpendicular to and centered directly over the EPG for 2 seconds. Figure 7. Holding the magnet over the EPG 3. Remove the magnet, replace the keeper bar, and store the magnet. CAUTION: Do not use the magnet around magnetically sensitive items to avoid damaging them. 22

29 Using the EPG Read this section for directions for use related to the EPG and EPG header. For directions for use of other system components not covered in this document, see the clinician s manual for the appropriate device. NOTE: Well in advance of the surgical procedure, set up communication between the EPG and the programmer and controller to ensure it is functional. Refer to "Setting Up Communication Between the EPG and the Clinician Programmer or Patient Controller" for instructions. CAUTION: Before the EPG or associated components are used or secured to the patient, determine if a patient may have an allergic reaction to the materials. Compatibility Guidelines The following table lists the trial cable and leads compatible with the EPG. For more information on these components, refer to the applicable clinician's manual. 23

30 Table 1. Trial cable and leads compatible with the EPG Device Model Number Multilead trial cable 3013 Percutaneous trial leads 3046, 3066, 3086 Connecting the Multilead Trial Cable to the EPG The following information outlines the guidelines for connecting a multilead trial cable to the EPG. For further instructions, refer to the clinician s manual for the multilead trial cable. 1. Remove the EPG header cap to expose the EPG stimulator port. 2. Carefully slide the connector end of the multilead trial cable into the EPG stimulator port. CAUTION: The EPG is not sterile. When connecting and testing the EPG with the multilead trial cable, keep the EPG outside of the sterile field. CAUTION: To avoid delivering jolting stimulation to the patient, turn off stimulation before plugging in the multilead trial cable. 24

31 Figure 8. Inserting the connector into the EPG stimulator port 25

32 Disconnecting the Multilead Trial Cable from the EPG The following steps outline the guidelines for disconnecting the multilead trial cable from the EPG. For further instructions, refer to the clinician s manual for the multilead trial cable. 1. Turn off stimulation on the EPG using the clinician programmer app. 2. Grasp the connector end of the multilead trial cable and gently pull the connector from the EPG stimulator port. Securing the Leads or Extensions The following information outlines the suggested procedure for securing the leads or extensions to the patient: Secure the leads or extensions and bandage the lead entry site. For further instructions, refer to the clinician s manual for the applicable leads or extensions. Connecting the EPG Header to the EPG The following information outlines the guidelines to connect the EPG header to the EPG: Align the tab indentation on the EPG header with the tab on the EPG and carefully insert the EPG header into the EPG body until it clicks into place. 26

33 Figure 9. Connecting the EPG header to the EPG 27

34 Connecting the Leads or Extensions to the EPG Header The following steps outline the suggested guidelines to connect a lead or extension to the EPG header: CAUTION: Do not connect a lead or extension with body fluid or saline residue on its contacts because corrosion can occur and cause failure of the system. 1. If any of the lead or extension contacts came in contact with body fluid or saline, thoroughly clean the contacts with sterile deionized water or sterile water for irrigation and dry them completely. CAUTION: Do not bend the lead sharply or it may be damaged. NOTE: The EPG header comes with open ports and attached port plugs. Do not remove the port plugs from the header. If only using one lead or extension, use the lower port labeled 1-8 and insert a port plug in the unused port. 2. Push and hold the button on the EPG header to open the EPG header ports. 3. Using clean gloves, carefully slide the lead or extension into the EPG header until you reach a hard stop and all the contact bands are fully inside the connector assembly and hidden from view. 28

35 Figure 10. Insert the lead or extension fully into the EPG header 1. Fully inserted 2. Not fully inserted 4. Release the button to lock the lead or extension into place. 5. If applicable, repeat the previous steps to insert the remaining lead or extension. Do not disturb the already inserted lead or extension. 29

36 Securing the EPG The following steps outline the suggested guidelines for securing the EPG to the patient: 1. Place the EPG into the provided pouch. 2. Remove the adhesive liner on the pouch flap. 3. Use the flap to secure the leads in place and seal the pouch. NOTE: The EPG meets the IP22 rating when sealed in the pouch. 4. Remove the adhesive liner on the back of the pouch and press the pouch onto the patient s skin until it is secure. NOTE: It is good practice to secure the EPG pouch away from the lead entry site bandaging. NOTE: Leave some slack in the leads or extensions when securing the EPG onto the patient. 5. Use additional medical tape to secure the pouch and leads or extensions further as needed. 30

37 Programming the EPG Use the clinician programmer app to program the system for the trial period. For information on using the clinician programmer app for programming, refer to the clinician s programming manual. Caring for the EPG The EPG is an important part of the neurostimulation system. This section provides tips and other information about caring for the EPG. Handling the EPG Do not expose the EPG to prolonged direct sunlight. Avoid exposing the EPG to sand, dust, and other environmental hazards. Do not put heavy or sharp objects on the EPG. Keep the EPG away from food and beverages, and avoid getting the EPG wet. If the EPG does get wet, do not use it. Contact Technical Support at for help. 31

38 Cleaning the EPG Using two PDI Sani-Cloths, wipe the EPG for 2 minutes front and back, then allow to air dry. Once the EPG is dry, attach the header cap. NOTE: Clean the EPG before attaching the header cap. Clean the EPG by thoroughly wiping off its outer surface using the provided cleaning cloths. Do not wipe over the stimulator port. Do not submerge the EPG or use a cloth that is soaking wet. Never use ammonia, acidic, or alkaline cleaners or organic chemicals such as paint thinner, acetone, propyl alcohol, or kerosene. They may damage surface finishes of the EPG. Storing and Transporting the EPG Store the EPG and batteries in a cool, dry place. Do not store the EPG or batteries in your car or garage. Transport the EPG in a carrying case. If you are carrying the EPG in a briefcase or other carrying case, make sure no objects in the case are pressing on the device. Store paired EPGs and patient controllers together to reduce the number of times you need to set up communication. 32

39 At the end of a trial, disable stimulation, clean the EPG, and attach the header cap before storing the EPG. NOTE: During a trial store the header cap for reuse. Figure 11. EPG header cap 33

40 Service and Ordering Information This section provides information for contacting Technical Support and for ordering replacement parts and accessories. Technical Support For technical questions and support for your St. Jude Medical neuromodulation product, use the following information: (toll-free within North America) For additional assistance, call your local St. Jude Medical representative. Ordering Information To order parts, contact Technical Support. Refer to the following list for model numbers. Table 2. Ordering information for the EPG Model Description 3599 External Pulse Generator, 16 channel 3032 External Pulse Generator, 2 Port Header 34

41 Model Description 1212 Coin cell batteries 1917 Battery door 3013 MLTC 1210 Magnet 1203 Box of cleaning cloths 1216 EPG header cap 1213 Pouch with adhesive (5) 1214 Pouch without adhesive and belt (5) 1218 Carrying case 35

42 Appendix A: Product Specifications EPG with Header The EPG (Model 3599) and EPG header (Model 3032) has the following physical specifications. Table 3. EPG and EPG header specifications Height Width Thickness Weight (with batteries) Volume Power source Lead or extension compatibility Expected service life Storage temperature Operating temperature 79.8 mm (3.14 in) 57.2 mm (2.25 in) 11 mm (0.433 in) 44 g (1.55 oz) cc (2.72 cu in) Nonrechargeable CR2450 coin cell 2 octrode 5 years -25 C - 70 C (-13 F F) 5 C - 40 C (41 F F) 36

43 Storage humidity Operating humidity Storage pressure Operating pressure up to 93% (noncondensing) 15% - 93% (noncondensing) kpa ( psi) kpa ( psi) Table 4. Operating parameters for the EPG Tonic Range Steps Pulse width µs µs Frequency Hz Hz Hz 2 Hz 10 Hz 20 Hz Amplitude ma ma Stimulation output Constant current NOTE: The maximum current depends on the impedance, frequency, and pulse width settings. 37

44 Appendix B: Regulatory Statements Battery Disposal Guidelines Do not dispose of this device or its battery along with general waste. Follow local regulations for the safe disposal of the battery, and return the device to St. Jude Medical at the end of its operating life. Statement of FCC Compliance This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radiofrequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. 38

45 Consult the dealer or an experienced radio/tv technician for help. Operation is subject to the following two conditions: This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation. Modifications not expressly approved by the manufacturer could void the user s authority to operate the equipment under FCC rules. Statement of Compliance With License-Exempt RSS Standard (Canada) This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Statement of Compliance With Interference-Causing Equipment Standard (Canada) This ISM device complies with Canadian ICES-001. Identification Information for Product Registration This device has a label that contains, among other information, a product identifier in the following format: 39

46 Table 5. Registration identification information Identifier Type FCC Registration Number Industry Canada (IC) Registration Number Registration Identifier PX2SJMX599EPG 8454A-SJMX599EPG Product Classification Statement (CISPR 11, Class B) This product is class B equipment, which is intended primarily for use in the domestic environment. Wireless Technology Information The following table summarized the technical details of the Bluetooth low energy technology as it is implemented in the device. Table 6. Bluetooth low energy information Antenna type Antenna dimensions Modulation Bluetooth low energy Embedded ceramic antenna 4 mm x 3 mm x 1.1 mm GFSK 40

47 Magnetic Field Strength (at 10 m distance) Electric Field Strength Output power (EIRP*) Range Center frequency Channel Bandwidth Data-rate Data flow Protocol *EIRP = Equivalent isotropically radiated power A/m maximum V/m maximum 1 mw (0 dbm) typical, 10 mw (+10 dbm) maximum 2 meters typical 2.44 GHz 40 logical channels 2 MHz per channel 5 kbit/second Bi-directional BLE Radio Transmitter, Cables, Transducers The device contains a radio transmitter/receiver with the following parameters. Radio transmitter parameters: Frequency (range): to GHz Bandwidth (-15dB): to GHz 41

48 Channel: 40 logical channels using AFH Modulation: GFSK Radiated output power: 10 mw (+10 dbm) maximum Magnetic field strength: A/m Duty Cycle: Variable, but low (< 5%) Semi-duplex capability The radio receiver in the device is using the same frequency and bandwidth as the transmitter. Cables and transducers: Cables and transducers are not used during normal use of the device nor while programming the device. Quality of Service for Wireless Technology Bluetooth low energy wireless technology enables communication between the EPG and the clinician programmer or patient controller. The requirements for the quality of service (QoS) vary depending on the use environment (operating room, recovery room, and home environment). 42

49 After the clinician programmer or patient controller is paired with a generator, (Bluetooth symbol) is visible on the clinician programmer or patient controller in the upper right-hand corner. When the BLE connection is not active, the symbol is grayed out. The raw BLE data rate is approximately 2.5 KB/sec and the effective data rate is approximately KB/sec. Other requirements include a semi-duplex transmission with a required acknowledge, a transmission latency in each direction (2x), and a receive-to-transmit mode (RX to TX) time. Data is resent if not sent successfully. Each key press may transmit up to 8 data packets, depending on the number of packets that need to be transmitted (i.e., if there is only one packet to transmit, only one packet will be transmitted). Wireless Security Measures The wireless signals are secured through device system design that includes the following: The generator will encrypt its wireless communication. Only one patient controller or clinician programmer may communicate with the generator at the same time. A unique key for each unit that is checked during each transmission. Built-in pairing that specifies valid and legitimate pairing among units. Proprietary authentication in addition to the pairing procedure specified in Bluetooth low energy, which includes an element of proximity. A proprietary algorithm that detects and prevents an unauthorized user from attempting to pair 43

50 with the generator. Troubleshooting for Wireless and Coexistence Issues If you experience issues with the wireless communication between the generator and the clinician programmer or patient controller, try the following: Decrease the distance between the devices Move the devices so they share line of sight Move the devices away from other devices that may be causing interference Wait a few minutes and try connecting again Do not operate other wireless devices (i.e., laptop, tablet, mobile phone, or cordless phone) at the same time NOTE: Wireless communications equipment, such as wireless home network devices, mobile and cordless telephones, and tablets, can affect the device. 44

51 Appendix C: Electromagnetic Compatibility Guidelines The EPG, hereafter the device, is medical equipment and should be used with the following guidance. The device requires special precautions with regard to electromagnetic compatibility (EMC) and should be used in accordance with the information provided in this manual. The device has no essential performance, as defined by IEC Edition 3.1:2012. Performance of the device was maintained during immunity testing. The device is intended for use in the electromagnetic environment specified in the following tables. The user should ensure that it is used in such an environment. CAUTION: The device complies with the limits for medical devices contained in IEC : , IEC : , EN 55011:2009/A1:2010 (CISPR 11:2010), and ETSI EN V1.8.1 ( ). However, the device may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to mitigate this effect by reorienting or relocating the receiving device. CAUTION: To avoid increasing emissions or decreasing immunity from a device or system, use only St. Jude Medical-approved components with this system. Do not use St. Jude Medical components with other non-st. Jude Medical devices or systems. 45

52 Table 7. Guidance and manufacturer s declaration electromagnetic emissions Emissions Test Compliance Electromagnetic Environment Guidance RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC Voltage fluctuations/ flicker emissions IEC Group 1 Class B Not applicable Not applicable The device uses RF energy for its internal and system interface functions. Its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. The device is suitable for use in all establishments including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. 46

53 Table 8. Guidance and manufacturer s declaration electromagnetic immunity Immunity Test IEC Test Level Electrostatic discharge (ESD) IEC Electrical fast transient/burst IEC Surge IEC Voltage dips, short interruptions IEC ±8 kv contact ±15 kv air ±2 kv for power supply lines ±1 kv for input/output lines ±1 kv line to line ±2 kv line to earth <5% UT 40% UT 70% UT Compliance Level ±8 kv contact ±15 kv air Not applicable Not applicable Not applicable Electromagnetic Environment Guidance The relative humidity should be at least 5%. No guidance for battery-powered devices. No guidance for battery-powered devices. No guidance for battery-powered devices. 47

54 Immunity Test IEC Test Level Power frequency (50/60 Hz) magnetic field IEC Compliance Level Electromagnetic Environment Guidance 30 A/m 30 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. 48

55 Immunity Test IEC Test Level Conducted RF IEC Vrms 150 khz to 80 MHz 6 Vrms (ISM/Radio bands between 150 khz to 80 MHz) 80% AM at 1 khz Compliance Level Electromagnetic Environment Guidance 6 Vrms Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d=0.58 P where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). 49

56 Immunity Test IEC Test Level Radiated RF IEC V/m 80 MHz to 2.7 GHz 80% AM at 1 khz Compliance Level Electromagnetic Environment Guidance 20 V/m Recommended separation distance d=0.18 P 80 to 800 MHz d=0.35 P 800 MHz to 2.5 GHz Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b Interference may occur in the vicinity of equipment marked with the following symbol: NOTE: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. 50

57 a Field strengths from fixed transmitters, such as base stations for radio (mobile/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device. b Over the frequency range 150 khz and 80 MHz, field strengths should be less than 3 V/m. The device is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment. 51

58 Table 9. Guidance and manufacturer s declaration proximity fields Proximity Test IEC Test Level IEC MHz: Hz pulse modulation Compliance Level 450 MHz: 28 FM modulation 28 V/m 710 MHz, 745 MHz, 780 MHz: Hz pulse modulation 810 MHz, 870 MHz, 930 MHz: Hz pulse modulation 1720 MHz, 1845 MHz, 1970 MHz: Hz pulse modulation 2450 MHz: Hz pulse modulation 5240 MHz, 5500 MHz, 5785 MHz: Hz pulse modulation Electromagnetic Environment Guidance 27 V/m Recommended separation distance d = 0.3 m 9 V/m 28 V/m 28 V/m 28 V/m 9 V/m WARNING: Do not use portable RF communications equipment (including peripherals 52

59 such as antenna cables and external antennas) closer than 30 cm (12 inches) to any part of the EPG, including cables specified by St. Jude Medical. Otherwise, performance degradation may occur. NOTE: For the frequency bands in the table above, use the specified recommended separation distance. The recommended separation distances in the table below apply to all other frequencies within the specified ranges. Table 10. Recommended separation distances between portable and mobile RF communications equipment and the device Rated Maximum Separation Distance According to Frequency of Transmitter (m) Output Power of 150 khz to 80 MHz 80 to 800 MHz 800 MHz to 2.5 GHz Transmitter (W) d=1.17 P d=0.18 P d=0.35 P NOTE: Wireless communications equipment, such as wireless home network devices, mobile and cordless telephones, and walkie-talkies, can affect the device. Keep the device away from wireless 53

60 communication equipment at least the distance d as listed in the 800 MHz to 2.5 GHz column in the above table. NOTE: For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. Radio Frequency Information The effective radiated power is below the limits as specified in: USA: FCC 47 CFR Part 15:2012 Canada: RSS-210 Issue 8 Europe: ETSI EN V1.9.2, ETSI EN and ETSI EN NOTE: Maintain a reasonable distance between other electronic equipment and the device. CAUTION: The ISM band used by this device has been approved by the Federal Communications 54

61 Commission for unlicensed use. However, there is no guarantee that this device will not receive interference or that any particular transmission from this device will be free from interference. Appendix D: Symbols and Definitions The following symbols may be used in this document and on some of the products and packaging: Symbol Description Caution, consult accompanying documents Consult this document for important safety-related information (This symbol is blue and white on the device.) Consult instructions for use Follow instructions for use on this website Device contains a type BF applied part to protect you from shock. The device is internally powered and is intended for continuous operation. 55

62 Symbol Description Device contains a radio-frequency (RF) transmitter, which may cause RF interference with other devices near this device. Magnetic Resonance (MR) Unsafe, an item poses unacceptable risks to the patient, medical staff, or other persons within an MR environment Keep the device dry Ingress protection rating for a device that is protected from the intrusion of solid foreign objects as small as 12.5 mm in diameter and is protected from vertically dripping water when the device is tilted at an angle up to 15 degrees Single use only Expiration date Date of manufacture Temperature limits for storage conditions 56

63 Symbol Description Humidity limits Pressure limits Catalog number Manufacturer Contents quantity Code that uniquely identifies an inventory item Serial number Lot number Prescription use only 57

64 Symbol Description This product shall not be treated as household waste. Instead it is the user s responsibility to return this product to St. Jude Medical for reprocessing. By ensuring that this product is disposed of properly, you will help prevent potential negative consequences for the environment and human health, which could be caused by inappropriate waste handling of this product. The recycling of materials will help to conserve natural resources. For more information about how to return this product for recycling, please contact St. Jude Medical. Authorized European representative European conformity, affixed in accordance with the relevant provisions of AIMD directive 90/385/EEC and R&TTE directive 1999/5/EC. Hereby, St. Jude Medical declares that this device is in compliance with the essential requirements and other relevant provisions of these directives. This device is listed by Nationally Recognized Testing Laboratory (NRTL) as certified. Australian Communications and Media Authority (ACMA) and New Zealand Radio Spectrum Management (RSM) Regulatory Compliance Mark (RCM) 58

65 Symbol Description This equipment is certified for type certification pursuant of Article of the Japan Radio Law 59

66

67

68 St. Jude Medical 6901 Preston Road Plano, Texas USA St. Jude Medical Coordination Center BVBA The Corporate Village Da Vincilaan 11 Box F Zaventem Belgium sjm.com * * ARTEN A

St. Jude Medical 8-Channel Adapter. Clinician's Manual

St. Jude Medical 8-Channel Adapter. Clinician's Manual St. Jude Medical 8-Channel Adapter Clinician's Manual CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. For U.S. California Only: Proposition 65, a State of California

More information

Smart Control remote guide

Smart Control remote guide Smart Control remote guide Thank you Thank you for choosing Unitron Smart Control remote for your Unitron hearing aids. At Unitron, we care deeply about people with hearing loss. We work closely with hearing

More information

Prodigy Neurostimulation System Programming and Reference. Models 3855, Clinician's Manual

Prodigy Neurostimulation System Programming and Reference. Models 3855, Clinician's Manual Prodigy Neurostimulation System Programming and Reference Models 3855, 3856 Clinician's Manual CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. Unless otherwise

More information

Implantable Pulse Generator. Prodigy IPG Prodigy MRI IPG. Clinician's Manual

Implantable Pulse Generator. Prodigy IPG Prodigy MRI IPG. Clinician's Manual Implantable Pulse Generator Prodigy IPG Prodigy MRI IPG Clinician's Manual CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. Unless otherwise noted, indicates

More information

Phonak PilotOne II. User Guide

Phonak PilotOne II. User Guide Phonak PilotOne II User Guide Contents 1. Welcome 4 2. Description 5 3. Using Phonak PilotOne II 6 3.1 Inserting a new battery 6 3.2 Switching On/Off 7 3.3 Holding PilotOne II 7 3.4 Changing hearing aid

More information

Remote control 2 guide

Remote control 2 guide Remote control 2 guide Thank you Thank you for choosing remote control 2 for your Unitron hearing aids. At Unitron, we care deeply about people with hearing loss. We work closely with hearing healthcare

More information

PATIENT CONTROLLER ST. JUDE MEDICAL UNDERSTANDING IMPORTANT FEATURES ON YOUR

PATIENT CONTROLLER ST. JUDE MEDICAL UNDERSTANDING IMPORTANT FEATURES ON YOUR UNDERSTANDING IMPORTANT FEATURES ON YOUR ST. JUDE MEDICAL PATIENT CONTROLLER THINGS TO KNOW BEFORE AND AFTER AN MRI OR SURGICAL PROCEDURE WHEN YOU HAVE A ST. JUDE MEDICAL PROCLAIM NEUROSTIMULATION SYSTEM

More information

Lead and Extension Insertion Tool. Model Clinician's Manual

Lead and Extension Insertion Tool. Model Clinician's Manual Lead and Extension Insertion Tool Model 1803 Clinician's Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or one of its subsidiaries. ST. JUDE

More information

ProPocket TM. User Guide

ProPocket TM. User Guide ProPocket TM User Guide A1 A Introduction Dear customer, Your hearing instruments are equipped with wireless technology and can therefore be controlled by your ProPocket. These instruction describes how

More information

Mini Remote Microphone OPERATIONS MANUAL

Mini Remote Microphone OPERATIONS MANUAL Mini Remote Microphone OPERATIONS MANUAL Table of Contents Overview..................................... 4 Basic Use..................................... 7 Daily Use.....................................

More information

Implantable Pulse Generator. Proclaim IPG. Clinician's Manual

Implantable Pulse Generator. Proclaim IPG. Clinician's Manual Implantable Pulse Generator Proclaim IPG Clinician's Manual CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. Unless otherwise noted, indicates that the name is

More information

Wireless accessories. Remote Control 3.0. Instructions for use

Wireless accessories. Remote Control 3.0. Instructions for use Wireless accessories Remote Control 3.0 Instructions for use Introduction to this booklet This booklet guides you on how to use and maintain your new Remote Control 3.0. Please read the booklet carefully

More information

Safety manual for hearing instruments

Safety manual for hearing instruments Safety manual for hearing instruments Content Safety information 3 Intended use 3 Explanation of symbols 3 General warnings 4 BTE or RIC or custom models 12 For infants, small children and mentally disabled

More information

Getting started with Wireless Fitting

Getting started with Wireless Fitting Quick Guide Getting started with Wireless Fitting Over the years, hearing aid manufacturers have introduced more cables to the clinic with the launch of new hearing aids. For dispensers, this has become

More information

Pocket Adaptor Kit FOR SPINAL CORD STIMULATION. Tip Cards

Pocket Adaptor Kit FOR SPINAL CORD STIMULATION. Tip Cards Pocket Adaptor Kit FOR SPINAL CORD STIMULATION Tip Cards model 74001 (1 4) model 74002 (2 4) The Pocket Adaptor Tip Cards are to be used for educational purposes only. Please refer to the Pocket Adaptor

More information

User Manual CentrePoint Data Hub

User Manual CentrePoint Data Hub User Manual CentrePoint Data Hub CentrePoint Data Hub Revision: 0 Released: 07//07 User Manual CentrePoint Data Hub CentrePoint Data Hub Revision: 0 Released: 07//07 Contents Overview Getting Started Assigning

More information

DRG THERAPY FOR CHRONIC PAIN ACCURATE CLINICAL STUDY FACT SHEET FOR PATIENTS

DRG THERAPY FOR CHRONIC PAIN ACCURATE CLINICAL STUDY FACT SHEET FOR PATIENTS ACCURATE CLINICAL STUDY DRG THERAPY FOR CHRONIC PAIN FACT SHEET FOR PATIENTS It was the kind of pain where you couldn t push through it now I am back to living life again and not having any pain. Jenifer,

More information

Percutaneous Lead Kit. Models 3143, 3146, 3149, 3153, 3156, 3159, 3183, 3186, Clinician's Manual

Percutaneous Lead Kit. Models 3143, 3146, 3149, 3153, 3156, 3159, 3183, 3186, Clinician's Manual Percutaneous Lead Kit Models 3143, 3146, 3149, 3153, 3156, 3159, 3183, 3186, 3189 Clinician's Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical

More information

Connevans.info. DeafEquipment.co.uk. This product may be purchased from Connevans Limited secure online store at

Connevans.info. DeafEquipment.co.uk. This product may be purchased from Connevans Limited secure online store at Connevans.info Solutions to improve the quality of life Offering you choice Helping you choose This product may be purchased from Connevans Limited secure online store at www.deafequipment.co.uk DeafEquipment.co.uk

More information

Unitron remote control guide

Unitron remote control guide Unitron remote control guide Thank you Thank you for choosing a Unitron remote control for your Unitron hearing aids. At Unitron, we care deeply about people with hearing loss. We work closely with hearing

More information

HeroTracker Sensor for Diskus Instruction Manual

HeroTracker Sensor for Diskus Instruction Manual HeroTracker Sensor for Diskus Instruction Manual Table of Contents Intended use... 1 Product description... 1 Installation... 3 Usage... 5 Transferring to a new inhaler... 6 Precautions and warnings...

More information

Paddle Lead Kit. Models 3214, 3219, 3224, 3228, 3240, 3243, 3244, 3245, 3246, 3262, 3266, 3268, 3283, 3286, Clinician's Manual

Paddle Lead Kit. Models 3214, 3219, 3224, 3228, 3240, 3243, 3244, 3245, 3246, 3262, 3266, 3268, 3283, 3286, Clinician's Manual Paddle Lead Kit Models 3214, 3219, 3224, 3228, 3240, 3243, 3244, 3245, 3246, 3262, 3266, 3268, 3283, 3286, 3288 Clinician's Manual CAUTION: Federal (USA) law restricts this device to sale by or on the

More information

Medtronic, Inc.

Medtronic, Inc. Recharging System Quick Guide for Chronic pain www.medtronic.com Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432-5604 USA Tel: 763-505-5000 Toll-free: 1-800-328-0810 UC201002746a EN 2013 Medtronic,

More information

PilotOne II. User Guide

PilotOne II. User Guide PilotOne II User Guide Contents 1. Welcome 4 2. Description 5 3. Using Phonak PilotOne II 6 3.1 Inserting a new battery 6 3.2 Switching On / Off 7 3.3 Holding correctly 7 3.4 Changing hearing aid volume

More information

TAKE CONTROL OF YOUR FUTURE

TAKE CONTROL OF YOUR FUTURE NEUROSTIMULATION SYSTEMS FOR PAIN THERAPY Brief Summary: Product manuals must be reviewed prior to use for detailed disclosure. Indications Implantable neurostimulation systems - A Medtronic implantable

More information

Out of pain. Into relief.

Out of pain. Into relief. Spinal Cord Stimulation for Chronic Pain USING YOUR TRIAL STIMULATION SYSTEM Out of pain. Into relief. SCREENING TRIAL GUIDE AND PATIENT DIARY You are taking an important step in having a trial for spinal

More information

TH CROS RIC 19 Transmitter. User Guide

TH CROS RIC 19 Transmitter. User Guide TH CROS RIC 19 Transmitter User Guide Content Welcome 4 Your CROS transmitter 5 Components and names 7 Controls 9 Signal tones 9 Batteries 10 Battery size and handling tips 10 Replacing batteries 11 Daily

More information

HeroTracker Sensor Instruction Manual

HeroTracker Sensor Instruction Manual HeroTracker Sensor Instruction Manual Table of Contents Intended use... 1 Product description... 1 Installation... 2 Usage... 5 Transferring to a new inhaler... 6 Precautions and warnings... 7 Technical

More information

PATIENT PROGRAMMER 37642

PATIENT PROGRAMMER 37642 PATIENT PROGRAMMER 37642 Medtronic DBS Therapy user manual Activa PC Model 37601 Activa RC Model 37612 Activa SC Model 37602 Activa SC Model 37603! USA Rx only 2008 Medtronic, Activa, and SoftStart/Stop

More information

Transcutaneous Electrical Nerve Stimulation Device GF-3 / GF-3T

Transcutaneous Electrical Nerve Stimulation Device GF-3 / GF-3T Transcutaneous Electrical Nerve Stimulation Device GF-3 / GF-3T Operation Manual Read Before Using GF-3-INS-LAB-RevA08 TABLE OF CONTENTS INTRODUCTION TO TENS INDICATIONS AND CONTRAINDICATIONS WARNINGS

More information

easypocket Version 2 User Guide Life sounds brilliant.

easypocket Version 2 User Guide  Life sounds brilliant. easypocket Version 2 User Guide www.bestsound-technology.co.uk Life sounds brilliant. Content Your remote control 4 Overview 5 Display 6 Control buttons 8 Key lock switch 10 Programming connection 10 Batteries

More information

Electrodes (for TransAeris System)

Electrodes (for TransAeris System) Surgical Procedures for Implantation of TransLoc Electrodes (for TransAeris System) 1 R L 1 TransAeris System TransLoc Electrode Kit 3 3 TransAeris Patient Kit (FrictionLoc) TransAeris Patient Kit (TransAeris

More information

This user guide is valid for: CE mark applied Wireless models Phonak CROS II-312 Custom Phonak CROS II-13 Custom

This user guide is valid for: CE mark applied Wireless models Phonak CROS II-312 Custom Phonak CROS II-13 Custom User Guide This user guide is valid for: Wireless models Phonak CROS II-312 Custom Phonak CROS II-13 Custom CE mark applied 2015 2015 This user guide only applies to the CROS device. Please see the hearing

More information

Provider s Guide. Table of Contents

Provider s Guide. Table of Contents Table of Contents 2 3 5 6 7 8 9 10 10 11 13 14 15 17 21 21 21 22 23 How to Use This Guide Warnings, Cautions, and Contraindications Intended Use What is? System Contents Symbols RUSleeping Display Instructions

More information

Clinician's Manual. Paddle Lead Kit

Clinician's Manual. Paddle Lead Kit Paddle Lead Kit Models 3214, 3218, 3219, 3220, 3224, 3227, 3228, 3229, 3240, 3243, 3244, 3245, 3246, 3262, 3266, 3268, 3283, 3286, 3288 Clinician's Manual Unless otherwise noted, indicates that the name

More information

Instructions for Use REMOTE CONTROL RC-N

Instructions for Use REMOTE CONTROL RC-N Instructions for Use REMOTE CONTROL RC-N Congratulations on the purchase of your new RC-N remote control. Your remote control allows you to adjust your hearing aid volume and program settings. Please

More information

Video Otoscope MS101. User s Manual UM-AP A

Video Otoscope MS101. User s Manual UM-AP A Video Otoscope MS0 User s Manual UM-AP005-08-A Thank you for choosing the MDSCOPE Video Otoscope, the Video Otoscope choice for health care professionals. The operating and maintenance instructions found

More information

Proclaim DRG Implantable Pulse Generator. Model Clinician's Manual

Proclaim DRG Implantable Pulse Generator. Model Clinician's Manual Proclaim DRG Implantable Pulse Generator Model 3664 Clinician's Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or one of its subsidiaries. ST.

More information

Proclaim DRG Implantable Pulse Generator. Model Clinician's Manual

Proclaim DRG Implantable Pulse Generator. Model Clinician's Manual Proclaim DRG Implantable Pulse Generator Model 3664 Clinician's Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or one of its subsidiaries. ST.

More information

Pain Relief Patch. Operation Manual Read Before Using MODEL # ET

Pain Relief Patch. Operation Manual Read Before Using MODEL # ET Pain Relief Patch Operation Manual Read Before Using Intended Use The ireliev Pain Relief Patch is intended for temporary relief of pain associated with sore and aching muscles in the upper and lower extremities

More information

Phonak CROS B Custom. User Guide

Phonak CROS B Custom. User Guide Phonak CROS B Custom User Guide Your CROS device This user guide is valid for: Wireless models Phonak CROS B-312 Custom Phonak CROS B-13 Custom CE mark applied 2017 2017 If no box is checked and you do

More information

MRI Procedure Information. For St. Jude Medical MR Conditional Neurostimulation Systems. Clinician's Manual

MRI Procedure Information. For St. Jude Medical MR Conditional Neurostimulation Systems. Clinician's Manual MRI Procedure Information For St. Jude Medical MR Conditional Neurostimulation Systems Clinician's Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical

More information

DRG STIMULATION FOR CHRONIC PAIN FOCUS ON YOUR LIFE NOT YOUR PAIN

DRG STIMULATION FOR CHRONIC PAIN FOCUS ON YOUR LIFE NOT YOUR PAIN FOCUS ON YOUR LIFE NOT YOUR PAIN DRG STIMULATION FOR CHRONIC PAIN Focus on your friendships, not on your pain. FOCUS ON YOUR LIFE, NOT ON YOUR PAIN. Chronic pain is personal, and it can be difficult to

More information

RC-A. Remote Control. Instructions for Use

RC-A. Remote Control. Instructions for Use RC-A Remote Control Instructions for Use Introduction to this booklet This booklet guides you on how to use and maintain your new RC-A. Please read the booklet carefully including the warning section.

More information

ihealth PO3 Fingertip Pulse Oximeter OPERATION GUIDE INDEX

ihealth PO3 Fingertip Pulse Oximeter OPERATION GUIDE INDEX ihealth PO3 Fingertip Pulse Oximeter OPERATION GUIDE INDEX INTRODUCTION AND INTENDED USE...2 CONTENTS AND DISPLAY INDICATORS...2 PRODUCT DESCRIPTION...3 SPECIFICATIONS...3 CAUTIONS...3 Cautions...3 USING

More information

This page has been intentionally left blank

This page has been intentionally left blank Instruction Manual Table of Contents Introduction... 1 Important Safeguards... 2 Warranty... 5 Classifications and Markings... 6 Aeroneb Go System Parts... 7 Unit Assembly and Usage... 7 Prior to Assembly...

More information

DRG STIMULATION INTERVENTIONAL PAIN CENTER

DRG STIMULATION INTERVENTIONAL PAIN CENTER DRG STIMULATION Shawn M. Sills, MD at Touchstone Interventional Pain Center has been selected by St Jude Medical to be first in Southern Oregon delivering this new FDA approved therapy to patients. You

More information

Your umic. Thank you. Hearing healthcare professional: Telephone: Serial number of umic: Warranty: Date of purchase:

Your umic. Thank you. Hearing healthcare professional: Telephone: Serial number of umic: Warranty: Date of purchase: umic guide Thank you Thank you for choosing the Unitron umic. At Unitron, we care deeply about people with hearing loss. We work closely with hearing healthcare professionals to make advanced, purpose-driven

More information

1 1 U S E R M A N U A L

1 1 U S E R M A N U A L 1 1 USER MANUAL CONTENTS 2 MEET FIREFLY 2, YOUR NEW BEST FRIEND. The more you get to know Firefly 2, the more he can do for you. Your travel buddy is smart and you will find adventures are smooth under

More information

NEW JERSEY PAINCONSULTANTS OFFERS PATIENTS FIRST IMPLANTABLE NEUROSTIMULATION SYSTEM WITH MOTION SENSOR TECHNOLOGY FOR THE MANAGEMENT OF CHRONIC PAIN

NEW JERSEY PAINCONSULTANTS OFFERS PATIENTS FIRST IMPLANTABLE NEUROSTIMULATION SYSTEM WITH MOTION SENSOR TECHNOLOGY FOR THE MANAGEMENT OF CHRONIC PAIN Contact: Barbara Carey Practice Manager New Jersey Pain Consultants 175 Morristown Rd Basking Ridge, NJ 07920 (908) 630-0175 NEW JERSEY PAINCONSULTANTS OFFERS PATIENTS FIRST IMPLANTABLE NEUROSTIMULATION

More information

BEFORE AND AFTER YOUR MRI SCAN. For patients with St. Jude Medical neurostimulation systems for chronic pain

BEFORE AND AFTER YOUR MRI SCAN. For patients with St. Jude Medical neurostimulation systems for chronic pain BEFORE AND AFTER YOUR MRI SCAN For patients with St. Jude Medical neurostimulation systems for chronic pain Preparing for your MRI Depending on what type of neurostimulation system you have, you may be

More information

Algovita. MRI Procedure Guidelines. Spinal Cord Stimulation System. 201x

Algovita. MRI Procedure Guidelines. Spinal Cord Stimulation System. 201x Algovita Spinal Cord Stimulation System MRI Procedure Guidelines Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System. ONLY 201x Algovita

More information

PAIN LIFE INTER- RUPTED TRANSFORMED BURSTDR STIMULATION FOR CHRONIC PAIN

PAIN LIFE INTER- RUPTED TRANSFORMED BURSTDR STIMULATION FOR CHRONIC PAIN PAIN INTER- RUPTED BURSTDR STIMULATION FOR CHRONIC PAIN LIFE TRANSFORMED p Sam B, 18 months after implant and able to keep things shipshape on the waters of San Francisco Bay. IMAGINE YOUR LIFE TRANSFORMED

More information

The Biomet EBI Bone Healing System. Patient Manual

The Biomet EBI Bone Healing System. Patient Manual The Biomet EBI Bone Healing System Patient Manual Contents Introduction... Page 1 Symbol Description... Page 2 Warnings... Page 3 Battery Warning... Page 5 Indications, Contraindications, Usage and Adverse

More information

Getting to Know Your System Kit

Getting to Know Your System Kit FOR PATIENTS Getting to Know Your System Kit Your Tablet Your Recorder IMPORTANT FEATURES 1 2 3 1 TABLET POWER BUTTON 2 RECORDER PAIR BUTTON 3 RECORDER PAUSE BUTTON IMPORTANT: CHARGE YOUR TABLET AND RECORDER

More information

Toll-Free Customer Service: Instruction Manual. with SAFE-T-Energetics. Toll-Free Customer Service

Toll-Free Customer Service: Instruction Manual. with SAFE-T-Energetics. Toll-Free Customer Service Toll-Free Customer Service: 1-877-970-4970 5301 North Pima Road Scottsdale, AZ 85250 www.regenesisbio.com info@regenesisbio.com K with SAFE-T-Energetics Instruction Manual Toll-Free Customer Service 1-877-970-4970

More information

BiPAP Pro Bi-Flex. Accessing the Provider Mode Screens PROVIDER GUIDE

BiPAP Pro Bi-Flex. Accessing the Provider Mode Screens PROVIDER GUIDE BiPAP Pro Bi-Flex PROVIDER GUIDE IMPORTANT! Remove this guide before giving the device to the patient. Only medical professionals should adjust pressure settings. This guide provides you with instructions

More information

User s Manual. Model No. NTMPADS15.0 Model No. NTMPAD16.0

User s Manual. Model No. NTMPADS15.0 Model No. NTMPAD16.0 User s Manual Model No. NTMPADS15.0 Model No. NTMPAD16.0 Table of Contents Important Precautions...1 Quickstart Guide...2 Questions?... 3 Before you Begin... 3 Install The ifit Sleep App And Pair The Sensor

More information

AWAY Model # RTLAGF-900 REV

AWAY Model # RTLAGF-900 REV AWAY Model # RTLAGF-900 REV2.4.9.15 Contents User manual English 2 General Description 2 Intended Use 3 Included in this Package 4 Instructions for Use 7 Application Duration 7 Electrode Placement 8 Tips

More information

CP Relief Wand. Model CP 1000 (Rx Only) Operation Manual. Mid-America Medical Innovations, LLC Jefferson City, MO

CP Relief Wand. Model CP 1000 (Rx Only) Operation Manual. Mid-America Medical Innovations, LLC Jefferson City, MO CP Relief Wand Model CP 1000 (Rx Only) Operation Manual Mid-America Medical Innovations, LLC Jefferson City, MO RoHS Compliant Attachment 13.6 Page 13.6-1 INDEX I. Introduction General What is pain? How

More information

MRI Procedure Information. For St. Jude Medical MR Conditional Neurostimulation Systems. Clinician's Manual

MRI Procedure Information. For St. Jude Medical MR Conditional Neurostimulation Systems. Clinician's Manual MRI Procedure Information For St. Jude Medical MR Conditional Neurostimulation Systems Clinician's Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical

More information

ONLY. Physician Implant Manual

ONLY. Physician Implant Manual ONLY Physician Implant Manual NEVRO CORP. All questions or concerns about Nevro Corp. products should be forwarded to: Nevro Corp. 1800 Bridge Parkway Redwood City, CA 94065, USA Tel: +1.650.251.0005 Fax:

More information

USING YOUR INTELLIS IMPLANTABLE SPINAL CORD STIMULATION SYSTEM

USING YOUR INTELLIS IMPLANTABLE SPINAL CORD STIMULATION SYSTEM USING YOUR INTELLIS IMPLANTABLE SPINAL CORD STIMULATION SYSTEM This guide provides basic resources and information to help you use your Medtronic Intellis Spinal Cord Stimulation system. Note that the

More information

SIX PACK ABS Item No INSTRUCTION MANUAL. Read entire manual before operating this product. Use only as directed.

SIX PACK ABS Item No INSTRUCTION MANUAL. Read entire manual before operating this product. Use only as directed. SIX PACK ABS Item No. 206098 INSTRUCTION MANUAL Read entire manual before operating this product. Use only as directed. WARNINGS If you are in the care of a physician, consult your physician before using

More information

Contents Safety Instructions... 4 Package Contents... 6 Notes on Measurements... 7

Contents Safety Instructions... 4 Package Contents... 6 Notes on Measurements... 7 Contents Safety Instructions... 4 Keep battery operated equipment out of reach of children... 4 General information... 4 Handling batteries properly... 4 Measuring body fat... 5 Never undertake your own

More information

Medtronic MiniMed Insulin Infusion Pumps

Medtronic MiniMed Insulin Infusion Pumps Medtronic MiniMed Insulin Infusion Pumps Patients should always discuss potential risks and benefits with a physician. Please review the product manual prior to use for detailed instructions and disclosure.

More information

Single Use Curlew TM Multiple Biopsy Forceps

Single Use Curlew TM Multiple Biopsy Forceps Single Use Curlew TM Multiple Biopsy Forceps 13 SPECIMEN WITH METAL STORAGE CYLINDER With In Situ Fixation, Batch or Single Specimen Collection US Patents 5,782,747; 5,980,468; 6,071,248; foreign patents

More information

Wireless Body Analysis Scale

Wireless Body Analysis Scale Wireless Body Analysis Scale FR Balance d'analyse Corporelle Connectée IT Bilancia pesapersone con analisi corporea wireless ES Báscula inalámbrica de análisis corporal DE Drahtlose Körperanalysewaage

More information

Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System.

Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System. Algovita Spinal Cord Stimulation System MRI Procedure Guidelines Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System. ONLY 0300-000175-001

More information

Jazz Ultra Instruments

Jazz Ultra Instruments Jazz Ultra Instruments Instructions for use Next Contents 1. Introduction...p3 2. Symbols...p4 3. Safety 3.1 Device Classification 3.2 Warnings and cautions...p5 4. Cleaning instructions 4.1 Sterilisation...p6

More information

Instruction Manual Aerogen, Inc. Part No. AG-AL1010 Rev. C

Instruction Manual Aerogen, Inc. Part No. AG-AL1010 Rev. C Instruction Manual 2004 Aerogen, Inc. Part No. AG-AL1010 Rev. C Table of Contents Introduction... 3 System description... 4 Warnings... 5 Cautions... 5 Electromagnetic Susceptibility... 5 Symbols... 6

More information

OTOPod ProgLink. Quick Guide

OTOPod ProgLink. Quick Guide OTOPod ProgLink Quick Guide 1 Contents WHAT IS THE PROGLINK?........................... 3 WIRELESS FITTING WITH THE PROGLINK 3 USING THE PROGLINK WITH COMPASS................... 4 Detecting the Hearing

More information

DRG STIMULATION FOR CHRONIC PAIN FOCUS ON YOUR LIFE NOT YOUR PAIN

DRG STIMULATION FOR CHRONIC PAIN FOCUS ON YOUR LIFE NOT YOUR PAIN FOCUS ON YOUR LIFE NOT YOUR PAIN DRG STIMULATION FOR CHRONIC PAIN Focus on your friendships, not on your pain. FOCUS ON YOUR LIFE, NOT ON YOUR PAIN. Chronic pain is personal, and it can be difficult to

More information

Parkinson s disease, Essential Tremor and primary dystonia

Parkinson s disease, Essential Tremor and primary dystonia Parkinson s disease, Essential Tremor and primary dystonia What is Deep Brain Stimulation and how does it work? Deep Brain Stimulation (DBS) uses one or two surgically implanted medical devices, similar

More information

Over view. Wrist Placement.

Over view. Wrist Placement. CARDIO CONNECT Over view. Thank you for purchasing the DKN Cardio Connect. The DKN Cardio Connect provides an approximate heart rate read-out suitable for fitness training purposes. This device can also

More information

PainShield MD User Manual

PainShield MD User Manual PainShield MD User Manual Cat. # PSUM003 Ver. 07 (KEV) Table of Contents GENERAL INFORMATION 4 IMPORTANT GENERAL SAFETY CAUTIONS AND NOTICES 7 CLINICAL INFORMATION 8 GENERAL DESCRIPTION 10 PAINSHIELD

More information

Remote OPERATIONS MANUAL

Remote OPERATIONS MANUAL Remote OPERATIONS MANUAL 2 Introduction Introduction Starkey Hearing Technologies Remote allows for easy and discreet adjustments to your hearing aids. With its long battery life, the Remote is always

More information

Mini Pulse Electronic Stimulator

Mini Pulse Electronic Stimulator Mini Pulse Electronic Stimulator Model: PM-180 Operating Manual IMPORTANT: Please read all instructions before using this product. Retain this manual for future reference. www.santamedical.com IMPORTANT

More information

Neurostimulation Systems

Neurostimulation Systems Neurostimulation Systems Expanding the Array of Pain Control Solutions PROVEN EFFECTIVE FOR SIMPLE PAIN ITREL 3 SYSTEM For simple neuropathic pain Totally implantable for enhanced quality of life Easy

More information

Dr FuelCell Load Measurement Box

Dr FuelCell Load Measurement Box 1 (855) 251-0016 sales@fuelcellstore.com Dr FuelCell Load Measurement Box Instruction Manual www.fuelcellstore.com Instruction Manual for Dr FuelCell TM Load Measurement Box Version 1.0.1 November 2008

More information

SLEEP IMPROVING WRISTBAND. Item No Owner s Guide

SLEEP IMPROVING WRISTBAND. Item No Owner s Guide SLEEP IMPROVING WRISTBAND Item No. 205350 Owner s Guide Thank you for purchasing the Sharper Image Sleep Improving Wristband. Based on ancient Chinese acupuncture principles, this biofeedback device uses

More information

THE REBUILDER SYSTEM. ReBuilder Model 300

THE REBUILDER SYSTEM. ReBuilder Model 300 All warranty claims are to be processed by the manufacturer directly and must be accompanied by a receipt for purchase or copy of a dated order to dispense by a certified physician s office. DO NOT RETURN

More information

Instruction Manual. Copyright 2016 by Mettler Electronics Corp. Anaheim, CA

Instruction Manual. Copyright 2016 by Mettler Electronics Corp. Anaheim, CA Instruction Manual Distributed by: Mettler Electronics Corp. 1333 South Claudina Street Anaheim, CA 92805 U.S.A. 800.854.9305 1.714.533.2221 Fax: 1.714.635.7539 www.mettlerelectronics.com mail@mettlerelectronics.com

More information

SUPER CPAP POWER PACK

SUPER CPAP POWER PACK SUPER CPAP POWER PACK USER S GUIDE MODELS: C-150, C-222, C-444 SUPER CPAP POWER PACK YOUR CPAP MACHINE DOESN T HAVE TO KEEP YOU FROM ENJOYING THE GREAT OUTDOORS... Super CPAP Power Packs, the most powerful

More information

THE REBUILDER SYSTEM. ReBuilder Model 300. ReBuilder Medical Inc. 636 Treeline Drive, Suite A Charles Town, WV

THE REBUILDER SYSTEM. ReBuilder Model 300. ReBuilder Medical Inc. 636 Treeline Drive, Suite A Charles Town, WV THE REBUILDER SYSTEM ReBuilder Model 300 ReBuilder Medical Inc. 636 Treeline Drive, Suite A Charles Town, WV 25414 Phone: 304-725-2202 Fax: 304-725-4915 GET TO KNOW YOUR REBUILDER Confirm all the following

More information

Manufacturer: Phonak AG Laubisrütistrasse 28 CH-8712 Stäfa Switzerland / /A+W Printed in Switzerland

Manufacturer: Phonak AG Laubisrütistrasse 28 CH-8712 Stäfa Switzerland / /A+W Printed in Switzerland Manufacturer: Phonak AG Laubisrütistrasse 28 CH-8712 Stäfa Switzerland www.phonak.com 0459 029-0466-02/2009-08/A+W Printed in Switzerland Phonak AG, all rights reserved CRT hearing systems User Guide Contents

More information

Mini-Mist. Owner s Manual. portable nebulizer. PMM241 Rev.3 12/10

Mini-Mist. Owner s Manual. portable nebulizer. PMM241 Rev.3 12/10 Mini-Mist TM portable nebulizer PMM241 Rev.3 12/10 Dear Prodigy Owner: Thank you for choosing the Prodigy Mini-Mist vibrating mesh nebulizer. Please read this manual carefully as it contains important

More information

INSTALLATION MANUAL. MedRx TINNOMETER. Revolutionary Tinnitus Assessment.

INSTALLATION MANUAL. MedRx TINNOMETER. Revolutionary Tinnitus Assessment. INSTALLATION MANUAL Revolutionary Tinnitus Assessment MedRx TINNOMETER www.medrx-usa.com Contents Getting to Know Your Tinnometer 3 Computer Requirements 4 Tinnometer 5 Transducers and Accessories 5 Software

More information

IMPACT Pro R Instructions and guidance

IMPACT Pro R Instructions and guidance IMPACT Pro R Instructions and guidance Your IMPACT Pro R hearing instrument Features 1 Receiver Unit 2 Click Dome 3 Battery compartment 4 Push Button 5 Microphone Cover 6 Click Mould 7 Grid 8 HF-4 Wax

More information

Deep Brain Stimulation (DBS) For patients with tremor

Deep Brain Stimulation (DBS) For patients with tremor Patient Education Deep Brain Stimulation (DBS) For patients with tremor Please read this handout before reading and signing the form Special Consent for Procedural Treatment. This handout describes how

More information

This user guide is valid for: CE mark applied Wireless models Phonak CROS II-312 Custom Phonak CROS II-13 Custom

This user guide is valid for: CE mark applied Wireless models Phonak CROS II-312 Custom Phonak CROS II-13 Custom User Guide This user guide is valid for: Wireless models Phonak CROS II-312 Custom Phonak CROS II-13 Custom CE mark applied 2015 2015 This user guide only applies to the CROS device. Please see the hearing

More information

Compressor Nebulizer Instruction Manual Part No.: 5055

Compressor Nebulizer Instruction Manual Part No.: 5055 Compressor Nebulizer Instruction Manual Part No.: 5055 DISTRIBUTED BY: SAVE THESE INSTRUCTIONS. CAUTION - U.S. Federal Law restricts this device to sale by or on the order of a physician. 666002-6310 V1.3

More information

Instructions for Use. Welcome!

Instructions for Use. Welcome! Instructions for Use Welcome! The AMJEVITA SureClick autoinjector is a single-use prefilled autoinjector. Consult your doctor if you have any questions about your dose. Your doctor has prescribed AMJEVITA

More information

CAUTION...2 INSTRUCTION...5 IF-4000 INTERFERENTIAL STIMUATOR CONTROLS AND FUCTIONS...7 HOW TO USE YOUR IF-4000 STIMULATOR...14 ELECTRODES CARE...

CAUTION...2 INSTRUCTION...5 IF-4000 INTERFERENTIAL STIMUATOR CONTROLS AND FUCTIONS...7 HOW TO USE YOUR IF-4000 STIMULATOR...14 ELECTRODES CARE... CONTENTS CAUTION...2 INSTRUCTION...5 IF-4000 INTERFERENTIAL STIMUATOR CONTROLS AND FUCTIONS...7 HOW TO USE YOUR IF-4000 STIMULATOR...14 SKIN CARE...17 ELECTRODES CARE...18 BATTERY REPLACEMENT...20 TECHNICAL

More information

Precision GetFit. Digital Body Fat Bathroom Scale. EatSmart Products Seattle, WA

Precision GetFit. Digital Body Fat Bathroom Scale. EatSmart Products Seattle, WA Precision GetFit Digital Body Fat Bathroom Scale EatSmart Products Seattle, WA 98134 866-843-3905 INTRODUCTION Thank you for purchasing the EatSmart Precision GetFit Bathroom Scale. The EatSmart Precision

More information

Comfortmax Thermax Hot/Cold Water Circulation System. Instructions for Use

Comfortmax Thermax Hot/Cold Water Circulation System. Instructions for Use Comfortmax Thermax Hot/Cold Water Circulation System Instructions for Use THIS DEVICE CAN BE COLD ENOUGH TO CAUSE SERIOUS INJURY. SERIOUS ADVERSE REACTIONS AND SAFETY HAZARDS MAY OCCUR WHEN USING THIS

More information

Table of Contents. Goniometer Adapter Overview 9 Using the Adapter 10 Charging 10 Sensor Pairing 11 Smart Sensor Features 12

Table of Contents. Goniometer Adapter Overview 9 Using the Adapter 10 Charging 10 Sensor Pairing 11 Smart Sensor Features 12 TRIGNO TM EMG System Goniometer Adapter User s Guide Copyright Delsys Incorporated Delsys Logo and EMGworks are Registered Trademarks of Delsys Inc. MAN-017-1-1 Table of Contents Important Information

More information

I N S T R U C T I O N

I N S T R U C T I O N I N S T R U C T I O N M ANUAL model V2x Acoustical airway clearance device. 5.04.1 SAFETY SYMBOLS This manual uses the following safety symbols. They denote critical information. Please read them carefully.

More information

Cyrus soundkey User Instructions Contents

Cyrus soundkey User Instructions Contents Contents Important! Read before operating this equipment!...1 FCC Compliance Statement... 1 FCC notice... 1 Notices... 1 Introduction... 1 Compatibility...1 Unpacking... 1 Connecting soundkey to an Android

More information

Connect with EatSmart

Connect with EatSmart Connect with EatSmart EatSmart Products EatSmartScales EatSmart Products EatSmartScales www.eatsmartproducts.com Precision Tracker Digital Bathroom Scale Model ESBS-07 www.eatsmartblog.com 12 ESBS-07 CPESBS07

More information