The effect of buccal lingual slot dimension size on third-order torque response

Size: px
Start display at page:

Download "The effect of buccal lingual slot dimension size on third-order torque response"

Transcription

1 European Journal of Orthodontics, 2017, doi: /ejo/cjw043 Advance Access publication 3 June 2016 Original Article The effect of buccal lingual slot dimension size on third-order torque response Dan L. Romanyk 1, Kenji Au 2, Darren Isfeld 1, Giseon Heo 1, Michael P. Major 1 and Paul W. Major 1 1 School of Dentistry, Faculty of Medicine and Dentistry and 2 Department of Mechanical Engineering, Faculty of Engineering, University of Alberta, Edmonton, Alberta, Canada Correspondence to: Paul W. Major, 5 478, Edmonton Clinic Health Academy (ECHA), University of Alberta, Avenue NW, Edmonton, AB, Canada T6G 1C9. major@ualberta.ca Summary Introduction: The focus of the presented study was to investigate the effect of buccal lingual (B-L) orthodontic bracket slot dimension on third-order torque mechanics. Materials and methods: Three types of orthodontic brackets and two archwire sizes were considered. Ortho Classic H4 (0.026 B-L slot, passive), Ormco Damon Q (0.028 B-L slot, passive), and In-Ovation R (0.028 slot, active) brackets were tested using and beta-titanium archwires. An in vitro orthodontic torque simulator (OTS) was used to rotate archwires relative to a single bracket while recording forces and moments in three directions. For each bracket-archwire combination, a total of n = 47 samples were tested. Repeated measures analysis of variance between brackets was conducted for third-order torque values at 3 increments between 9 and 30 during loading and unloading for each archwire size. Results: Statistically significant differences between H4 and Q brackets were only found for archwires during loading, and archwires during unloading. Conversely, differences between H4 and R brackets were found for both archwires during loading and unloading phases. Finally, when using a archwire the H4 brackets reached the 5 Nmm threshold before R and Q brackets; however, there was little difference found when using a archwire. Conclusions: The concept of using a smaller B-L bracket slot dimension in orthodontic treatment showed it may theoretically allow for more options, primarily using smaller archwires to correct third-order rotational misalignments. However, it is suspected that bracket material limitations and added loading on the door currently prevent this from being clinically applicable. Introduction The study of third-order torque, referring to labial or lingual directed root torque, is an important topic of research in orthodontics. A number of in vitro studies in the literature have investigated various factors in orthodontic treatment that may affect torque expression. Archwire material, bracket material, incisal gingival (I-G) bracket dimension, bracket ligation method, use of pre-torqued archwires, and the effect of rotational misalignments have all been considered in previous research (1 10). All previous work has provided valuable insight to the mechanics of procedures utilizing third-order torque, making it imperative to continue such research as new treatment options arise. One aspect of third-order torque treatment that has yet to be investigated is the effect of altering the buccal lingual (B-L) slot dimension of a passively ligated bracket. Conventionally, passive self-ligating brackets utilize a B-L slot dimension. More recently, brackets have been developed with a reduced B-L dimension, namely as compared to the typical Reducing the B-L dimension may have a number of implications on treatment The Author Published by Oxford University Press on behalf of the European Orthodontic Society. All rights reserved. For permissions, please journals.permissions@oup.com 209

2 210 European Journal of Orthodontics, 2017, Vol. 39, No. 2 mechanics, one of which being the archwire rotation at which a third-order couple is generated. It is suggested that with the reduced B-L dimension, the archwire could contact the bracket door and base prior to engaging the I-G walls based on geometry of the archwireslot combination. This in turn should result in a couple being generated sooner in third-order archwire rotation from a neutral position. A primary suggested advantage of the reduced B-L dimension is that it could allow clinicians the option of using smaller archwires in correcting third-order misalignments. This in theory would then allow for a larger amount of archwire rotation inside the suggested clinically effective 5 20 Nmm range (11 13) than when using larger, and thus stiffer, archwires. While these potential advantages would be optimal in orthodontic treatment, they have yet to be investigated in the literature. The study presented here investigates the third-order mechanics of B-L dimension passive, B-L dimension passive, and B-L dimension active self-ligating brackets using an in vitro orthodontic torque simulator (OTS). Beta-titanium archwire sizes of and were both tested for each type of bracket. Results from this study were used to understand how B-L slot dimension influences third-order torque mechanics as well as how different archwire sizes effect results. Materials and methods Study design An orthodontic torque simulator, which is fully described in previous research (2 4, 7, 10), was used to examine the mechanics of thirdorder torque on all three types of brackets: the Damon Q (Ormco Corporation, Orange, California, USA) with B-L dimension, H4 (Ortho Classic, McMinnville, Oegon, USA) with B-L dimension, and B-L dimension In-Ovation R (GAC, Bohemia, New York, USA) active-ligation brackets. All brackets tested had a I-G dimension. Brackets were mounted on top of individual stainless-steel dowels with LOCTITE E-60HP Hysol Epoxy adhesive (Loctite, Westlake, Ohio, USA) for placement in the OTS, as illustrated in Figure 1. All three types of brackets were ligated using both and rectangular beta-titanium straight wires (G&H Wire Company, Franklin, Indiana, USA). The dowels were mounted onto a six-axis load cell (Nano17, ATI Industrial Automation, Apex, North Carolina, USA) that measures three dimensional forces and moments. Using translational and rotational stages on the OTS, initial loads were minimized to below 0.05 N and Nmm for forces and moments, respectively, in all directions. A sample size of 282, 47 brackets per bracket type per archwire, was examined for this study. This sample size was determined from pilot data using beta-titanium archwires and 10 of each bracket type. In order to detect a third-order torque difference of 4 Nmm between brackets with a Type I error rate of 0.05 and power of 0.9, 47 samples of each bracket per archwire was required (14). The test order of brackets was randomized, with the archwire group being tested first followed by the group. Third-order rotations were introduced in 3 increments from a neutral position leading up to a 33 rotation at the end, and then unloaded in the same increments back to the neutral position. Only rotation up to 30 was considered for analysis as the extra step was put in to account for backlash in the motor system. At each of these positions the force and moment readings were logged. The six-axis load cell takes voltage readings and calculates forces and moments at the load cell, and thus a transformation is needed to obtain the torque readings at the bracket base. The offset values and subsequent load transformations for the bracket to dowel interface in all three planes was found using the same procedure defined in previous research (10). The coordinate system at the base of the bracket is defined in Figure 2. Statistical analysis Preliminary analysis of variance (ANOVA) found each factor term archwire angle, archwire type, and bracket type as well as their interaction terms to be statistically significant with P < As such, repeated-measures ANOVA was conducted to investigate the response of third-order torque, T x, during archwire loading and unloading. Post hoc pairwise tests of each bracket-archwire combination, using a Bonferroni corrected procedure, for T x at every 3 in the range of 9 30 was completed to determine whether or not the brackets were statistically different, P Both loading and unloading data was investigated; however, they were studied separately as the presence of backlash prevents comparison between loading phases. The 9 30 range was selected based on post hoc observation of the data. It was found that during archwire rotation, the earliest that any archwire-bracket combination reached the 5 Nmm clinically relevant threshold was approximately 9. As tooth movement is not expected prior to this threshold, statistical analysis prior to this angle was deemed unnecessary. Additionally, even though archwire rotation up to 33 was utilized, statistical analysis was only considered up to 30. As a result of backlash in the system, measurements between 30 and 33 are not accurate, and as such are not included in the analysis. For these reasons, 9 30 was the selected range for statistical analysis of results. Results The average absolute T x values for each archwire size is plotted in Figure 3. Graphically all three types of brackets follow the same trend for the loading and unloading curves. Additionally, the magnitudes of T x at 30, representing maximum torque expression, and angles at which the 5 Nmm clinical threshold was surpassed during loading for each bracket-archwire combination are reported in Table 1. Figure 1. R (left), Q (middle), and H4 (right) brackets ligated in the OTS for testing.

3 D. L. Romanyk et al. 211 R, Q, and H4 brackets were found to generate 42.6, 41.3, and 37.9 Nmm of T x, respectively, at 30 during the loading phase of archwire rotation using a archwire. When using a archwire, the T x values at 30 were found to be 27.4, 24.0, and 21.2 Nmm for R, Q, and H4 brackets, respectively. With respect to reaching the 5 Nmm threshold when using a archwire, both the Q and H4 brackets surpassed this level between 9 and 12 of archwire rotation, while the R brackets reached this limit in the 6 9 range. Conversely, when using a archwire the H4 brackets reached the 5 Nmm limit first in the range of 9 12 of rotation, while the R and Q brackets were in the range of Figure 2. Right-handed coordinate system defined at the base of the bracket. Comparisons that were found to be statistically different for P 0.05 are tabulated in Tables 2 and 3 for loading and unloading, respectively. Comparisons made between sizes of archwires are not reported either as, in general, it would be expected that a archwire would produce greater torque than a archwire for the same material. As such, since these results are not of great interest in this study and in order to simplify a large dataset, they were not reported. The final sample size of brackets used in this study was 264. Samples were omitted due to data corruption (i.e. data not properly exporting from the software or logging during experiments), software crashes during experiments, or in the case of improper experimental setup before testing (e.g. not fully securing an archwire segment or dowel in the OTS before running a trial). There were no statistically significant differences observed between the Q and R brackets for either size of archwire during loading or unloading phases. During the loading phase, H4 demonstrated greater torque expression than Q between 9 and 15 degrees of rotation with the archwire, but there was no difference for larger rotations. Torque expression for H4 was greater than R at 12 of rotation with the archwire, but less than R for rotation beyond 27. H4 demonstrated less torque expression than R for more than 15 of rotation with the archwire. During the unloading phase, H4 demonstrated less torque expression than R for of wire rotation with archwire. H4 demonstrated less torque expression than Q between 21 and 12 of wire rotation with archwire, and less torque expression than R between 30 and 12 wire rotation with archwire. Figure 3. Plots of absolute average T x during archwire rotation for: (top left) archwire loading; (top right) archwire unloading; (bottom left) archwire loading; (bottom right) archwire unloading.

4 212 European Journal of Orthodontics, 2017, Vol. 39, No. 2 Discussion With respect to the lack of difference between Q and R brackets, previous work (3) found they behaved similarly, especially above approximately 9 of archwire rotation, when using stainless steel archwires. As such, the lack of statistical difference in this study is not unexpected. Table 1. Comparisons of three different bracket types for T x and angle of rotation during the loading phase to reach the 5 Nmm threshold. Bracket type Torque value at 30 ± SD (Nmm) archwires Q 41.3 ± H ± R 42.6 ± archwires Q 24.0 ± H ± R 27.4 ± Angle range for 5 Nmm threshold ( ) From a clinical perspective, it is the unloading curve that is of greatest interest. Being that the clinician deforms the archwire and then ligates it in the bracket, it is the peak torque and unloading response that will truly govern treatment. With that being said, the loading curve is also of interest in that it provides information regarding the response of the bracket-archwire pair without any permanent deformation in the system. Having knowledge of this data without any adverse effects can provide valuable information towards future research and bracket development. As a result, both the loading and unloading phases are of interest in the research of third-order torque mechanics and are to be considered here. When studying the H4-Q comparisons, both passive ligating brackets, the change in archwire size resulting in statistically significant differences between loading and unloading phases is immediately interesting. The H4 brackets generated significantly greater torque values than the Q brackets when using a archwire during loading; however, during unloading the brackets were not significantly different. When using a archwire, there was no difference during loading between these brackets, but during the unloading phase the Q brackets generated significantly greater torque than the H4. Table 2. Pairwise comparison of different bracket types for statistically significant (P 0.05) differences, after Bonferroni correction, during the loading phase. Pair for comparison Degree of rotation ( ) Mean difference, T x (Nmm) Std. error Sig. Table 3. Pairwise comparison of different bracket types for statistically significant (P 0.05) differences, after Bonferroni correction, during the unloading phase. Pair for comparison Degree of rotation ( ) Mean difference, T x (Nmm) Std. error Sig. Interval for difference Lower bound H4 Q; H4 R; H4 R; Interval for difference Lower bound H4-Q; H4-R; H4-R; Upper bound Upper bound

5 D. L. Romanyk et al. 213 The most probable factors leading to these results in the H4-Q comparisons are the reduced B-L dimension of the H4 bracket, and the subsequent deformation of the door on the brackets. As from Table 1, the H4 brackets surpassed the 5 Nmm threshold using archwires between 9 and 12, while the Q brackets were in the range of This is most likely a result of the archwire engaging the B-L walls of the H4 bracket sooner than it engages the slot of the Q. This statistical difference only exists at 9, 12, and 15, with no difference afterwards. Beyond 15 of archwire rotation it is suspected that engagement of the I-G walls results in similar torque expression. From the T x generated using archwires during loading, there is a noticeable change in the H4 behaviour. Initially, average T x values of the H4 bracket are larger than the Q; however, in the range the H4 brackets shift below the other two. It is suspected that the door of the H4 bracket supports much of the archwire rotation as a result of the reduced B-L dimension. Then, in the range it is suspected that the door begins to undergo appreciable plastic deformation resulting in the shift in the T x curve. As a result of the suggested plastic deformation, the slot size of the H4 would increase permanently, and thus causing the lack of any difference between H4 and Q brackets during unloading of the archwire. When using a archwire in the H4-Q comparison, there was no difference during loading, only during unloading. As a result of the larger archwire, it is suggested that the reduced B-L dimension of the H4 had no significant influence on T x as the archwire likely engaged the I-G walls of the bracket at approximately the same time; however, during unloading the Q bracket generated statistically greater T x than the H4 from 21 to 12 of rotation. This difference between loading and unloading phases is likely again due to plastic deformation of the H4 bracket, namely the door. Without this plastic deformation, the brackets should behave the same during loading and unloading. The H4-R comparison resulted in statistically significant differences for both archwire sizes during loading and unloading. With the exception of the 12 comparison during loading with a archwire, the R bracket generated larger T x values than the H4. There are potentially a couple of explanations for this result. Firstly, while the R bracket is designed for active ligation, is it still possible that a archwire remained passive during early archwire rotation. Second, when the archwire engages the door, it is possible that the door on the H4 brackets proved stiffer initially than the active clip of the R brackets. It is also certainly possible that a combination of these factors also lead to the findings at 12. At the upper end of the loading phase, 27 and 30, the R brackets generated significantly greater torque than the H4 when using archwires. For the archwires from 15 to 30, the R brackets produced significantly greater T x than the H4. During unloading, the R brackets generated statistically greater torque than H4 s in the range of and for and archwires, respectively. It is suggested, as previously mentioned, that this difference is a result of the H4 door weakening and plastic deformation at higher loads. This phenomenon would explain the increasing difference at higher loads during loading, as well as the greater range of archwire rotations at which the brackets differ during unloading. In regards to third-order torque mechanics and using a reduced B-L slot dimension of 0.026, the results indicate in theory that this could lead to earlier bracket-archwire engagement using smaller archwires. From Table 1, during loading the H4 brackets reached the 5 Nmm threshold before the R and Q brackets when using a archwire. Conversely, when using a archwire the actively ligating R bracket reached 5 Nmm before both of the passively ligating brackets. The discrepancy between the archwire sizes is likely a result of the larger archwire engaging the I-G walls of the bracket before or at approximately the same time as it engages the B-L walls. While earlier bracket-archwire engagement with a smaller B-L dimension does result in reaching the 5 Nmm threshold sooner, it also means that more torsional load will be carried by the bracket door and the base of the slot. While the base of the bracket is supported by the tooth and will likely undergo minimal deformation, the door is more susceptible to this. As it is suspected that the H4 bracket door underwent appreciable plastic deformation during larger archwire rotation, as deduced from the discrepancy in loading and unloading results, it was not able to withstand the added load. Though, in theory, if the door were capable of withstanding this added load within its elastic region, then the advantage of using a B-L slot dimension could be realized. As unloading is the clinical scenario resulting in tooth movement, it is imperative the bracket integrity remain intact during this phase. Clinically, there are a number of important implications from results of this study. The concept of generating a third-order couple sooner in archwire rotation when using smaller archwires and B-L dimension brackets showed some validity. It is unknown at this time if the unloading curve would also demonstrate greater torque expression with reduced B-L dimension if the maximum archwire rotation were reduced. As there was a change observed in H4 T x values during loading in the range, it is possible that if archwire rotation were reduced below this range then the unloading curve would emulate the loading curve more closely. As previously discussed in detail, added loading on the bracket door appears to introduce issues with respect to plastic deformation. When considering orthodontic treatment, should advances be made in brackets to minimize/eliminate this plastic deformation during archwire rotation, this concept could prove useful in allowing for the use of smaller archwires. As a result, lower third-order torque values at higher amounts of archwire rotation could be achieved through use of a archwire as compared to a This could allow for a larger range of tooth movement occurring inside the optimal 5 20 Nmm range. When using a archwire however, there was much less difference, if any, between the type of brackets tested and when they surpassed the 5 Nmm threshold. In clinical practice, if archwires are to be used by the clinician, the potential advantage of using the smaller B-L dimension slot would no longer be relevant with respect to third-order torque mechanics. Another result that requires discussion is the large standard deviation of the H4, or slot, brackets shown in Figure 3 and Table 1. The standard deviation of the H4 brackets at the larger degrees of archwire rotation approaches values around 8 9 Nmm for both sizes of archwires; however, R and Q brackets remain in the vicinity of 4 6 Nmm standard deviation at these same larger T x values. It is suspected that the reduced slot dimension, and difficulty obtaining the zero position as a result, was a cause of this. When attempting to locate the neutral, or zero, position of an archwire for experiments with the slot dimension, it was found to be particularly difficult in that the archwire frequently engaged the B-L walls for small rotations of the archwire. As such, there were a number of instances where the zero position was thought to be located, but once the experiment began it was found that this position was

6 214 European Journal of Orthodontics, 2017, Vol. 39, No. 2 slightly off by approximately 1 3. Additionally, since the smaller slot size results in more torsional loading being resisted by the door, the larger standard deviation could result from a variable response of the door. If the door were to begin to slide as opposed to simply deform (e.g. bend), or have some combination of this, it could result in different behaviour between samples. Finally, dimensional variations of the bracket slot and/or larger tolerances could also be a contributing factor to increased sample standard deviations. Clinically, the larger standard deviation could result in a larger variation in results between patients depending on the initial rotation of the archwire and its location in the slot. Furthermore, the zero position in this study was located in a controlled laboratory setting, making it easier to accurately position the archwire than in the clinical environment. Thus, it can be suspected that the difficulty in locating/knowing the zero position of the archwire for third-order rotation would translate, and potentially magnify, when moving to clinical treatment. In fact, given variations in bracket and archwire tolerances/dimensions, it can even be suggested that an occasion could arise where the B-L dimension of the archwire and bracket are actually the same. In this instance, a true zero position may not even be realized, having implications towards patient treatment. While the presented study provides valuable information with respect to third-order torque mechanics in orthodontic treatment, there are a number of limitations that must be mentioned. Firstly, only a single bracket type was tested for each slot size considered in this study. Additionally, only two sizes of rectangular archwires, and , were considered. These brackets and wire sizes were chosen due to their availability and commonality, however other options do exist. Future work in this area could include the investigation of other archwire sizes and how they impact third-order torque mechanics. Bracket slot dimensions were not measured prior to testing. As it has been shown that dimensions may vary significantly from the nominal value (15), future work regarding this topic could include measurement before experiments to correlate true slot sizes to third-order torque values. Conventional elastic ligation was also not considered in this study. Finally, the true oral environment, namely periodontal ligament compliance, was also not replicated in this in vitro study; however, previous work has shown that replication of compliance in the periodontium has minimal clinical relevance when investigating third-order torque mechanics (16). A number of future investigations should be conducted to elucidate added topics of interest arising from this study. Firstly, while evidence suggests that plastic deformation of the H4 bracket door is the root of a number of aforementioned discrepancies, the 3D deformation should be thoroughly analyzed. This may be completed using methods previously described in the literature (7), or other similar means. Additionally, a study of H4 bracket slot dimensions and their variation would prove useful in understanding the large standard deviation found here. The use of other archwire materials and sizes, as well as inclusion of elastically ligated brackets, would improve upon the limitations of this study. Finally, only the third-order torque mechanics of reduced B-L dimension brackets was considered here, capturing only one aspect of orthodontic treatment. It would be advantageous to investigate these brackets in other aspects of treatment, namely sliding mechanics, to gain a full understanding of their mechanics. In summary, this study elucidated a number of effects resulting from reducing the B-L slot dimension of orthodontic brackets during thirdorder torque. While there is the potential for a smaller B-L slot dimension to expand third-order treatment possibilities for clinicians, current bracket limitations may prevent this. Future investigation of this topic could include the inclusion of other archwire sizes and/or materials. Acknowledgements The authors would like to acknowledge the financial support of the McIntyre Fund at the University of Alberta in the completion of the research presented here. Conflict of interest The authors declare that there are no conflicts of interest to report. References 1. Gmyrek, H., Bourauel, C., Richter, G. and Harzer W. (2002) Torque capacity of metal and plastic brackets with reference to materials, application, technology, and biomechanics. Journal of Orofacial Orthopedics, 63, Archambault, A, Major, T.W., Carey, J.P., Heo, G., Badawi, H. and Major, P.W. (2010) A comparison of torque expression between stainless steel, titanium molybdenum alloy, and copper nickel titanium wires in metallic self-ligating brackets. The Angle Orthodontist, 80, Major, T.W., Carey, J.P., Nobes, D.S., Heo, G. and Major, P.W. (2011) Mechanical effects of third-order movement in self-ligated brackets by measurement of torque expression. American Journal of Orthodontics and Dentofacial Orthopedics, 139, e31 e Major, T.W, Carey, J.P., Nobes, D.S., Heo, G., Melenka, G.W. and Major P.W. (2013) An investigation into the mechanical characteristics of select self-ligated brackets at a series of clinically relevant maximum torquing angles: loading and unloading curves and bracket deformation. European Journal of Orthodontics, 35, Hirai, M., Nakajima, A., Kawai, N., Tanaka, E., Igarashi, Y., Sakaguchi, M., Sameshima, G.T. and Shimizu, N. (2012) Measurements of the torque moment in various archwire-bracket-ligation combinations. European Journal of Orthodontics, 34, Mittal, N., Xia, Z., Chen, J., Stewart, K.T. and Liu, S.S. (2013) Threedimensional quantification of pretorqued nickel-titanium wires in edgewise and prescription brackets. The Angle Orthodontist, 83, Melenka, G.W., Nobes, D.S., Carey, J.P. and Major, P.W. (2013) Threedimensional deformation comparison of self-ligating brackets. American Journal of Orthodontics and Dentofacial Orthopedics, 143, Sifakakis, I., Pandis, N., Makou, M., Eliades, T., Katsaros, C. and Bourauel, C. (2013) A comparative assessment of torque generated by lingual and conventional brackets. European Journal of Orthodontics, 35, Sifakakis, I., Pandis, N., Makou, M., Eliades, T., Katsaros, C. and Bourauel, C. (2014) Torque efficiency of different archwires in and inch conventional brackets. The Angle Orthodontist, 84, Romanyk, D.L., George, A., Li, Y., Heo, G., Carey, J.P. and Major, P.W. (2016) The influence of second-order bracket-archwire misalignment on loads generated during third-order orthodontic torque procedures. The Angle Orthodontist, 86, Harzer, W., Bourauel, C. and Gmyrek, H. (2004) Torque capacity of metal and polycarbonate brackets with and without a metal slot. European Journal of Orthodontics, 26, Partowi, S., Keilig, L., Reimann, S., Jager, A. and Bourauel, C. (2010) Experimental analysis of torque characteristics of orthodontic wires. Journal of Orofacial Orthopedics, 5, Badawi, H.M., Toogood, R.W., Carey J.P.R., Heo G. and Major, P.W. (2008) Torque expression of self-ligating brackets. American Journal of Orthodontics and Dentofacial Orthopedics, 133, Chow, S.C., Shao, J. and Wang, H. (2003) Sample Size Calculations in Clinical Research. Marcel Dekker Inc., New York, NY. 15. Majow, T.W., Carey, J.P., Nobes, D.S. and Major, P.W. (2010) Orthodontic bracket manufacturing tolerances and dimensional differences between select self-ligating brackets. Journal of Dental Biomechanics, 2010, George, M.G., Romanyk, D.L., George, A., Li, Y., Heo, G., Major, P.W. and Carey, J.P. (2015) Comparison of third order torque simulation with and without a periodontal ligament simulant. American Journal of Orthodontics and Dentofacial Orthopedics, 148,

Torque efficiency of different archwires in and inch conventional brackets

Torque efficiency of different archwires in and inch conventional brackets Original Article Torque efficiency of different archwires in 0.018- and 0.022-inch conventional brackets Iosif Sifakakis a ; Nikolaos Pandis b ; Margarita Makou c ; Theodore Eliades d ; Christos Katsaros

More information

A comparative assessment of torque generated by lingual and conventional brackets

A comparative assessment of torque generated by lingual and conventional brackets European Journal of Orthodontics 35 1 of (2013) 6 375 380 doi:10.1093/ejo/cjs029 Advance Access publication 9 May 2012 The Author 2012. Published by Oxford University Press on behalf of the European Orthodontic

More information

Measurements of the torque moment in various archwire bracket ligation combinations

Measurements of the torque moment in various archwire bracket ligation combinations European Journal of Orthodontics 34 (2012) 374 380 doi:10.1093/ejo/cjr022 Advance Access Publication 13 May 2011 The Author 2011. Published by Oxford University Press on behalf of the European Orthodontic

More information

Characterizing constraining forces in the alignment phase of orthodontic treatment

Characterizing constraining forces in the alignment phase of orthodontic treatment Original Article Characterizing constraining forces in the alignment phase of orthodontic treatment Christopher G. Gibson a ; Feng-Chang Lin b ; Ceib Phillips c ; Alex Edelman d ; Ching-Chang Ko e ABSTRACT

More information

Plus Combines Traditional Twin Bracket Treatment With Self-Ligation Convenience

Plus Combines Traditional Twin Bracket Treatment With Self-Ligation Convenience Lotus Plus Combines Traditional Twin Bracket Treatment With Self-Ligation Convenience Designed by an Orthodontist The initial concept for the Lotus Plus Self-Ligating Bracket System started with a simple

More information

With judicious treatment planning, the clinical

With judicious treatment planning, the clinical CLINICIAN S CORNER Selecting custom torque prescriptions for the straight-wire appliance Earl Johnson San Francisco, Calif Selecting custom torque prescriptions based on the treatment needs of each patient

More information

Dental Materials Journal 2015; 34(3): Keisuke TOCHIGI, Souichiro ODA and Kazuhito ARAI INTRODUCTION

Dental Materials Journal 2015; 34(3): Keisuke TOCHIGI, Souichiro ODA and Kazuhito ARAI INTRODUCTION Dental Materials Journal 2015; 34(3): 388 393 Influences of archwire size and ligation method on the force magnitude delivered by nickel-titanium alloy archwires in a simulation of mandibular right lateral

More information

wiirre e a a n n d d w wiirre e ffo orrms ms

wiirre e a a n n d d w wiirre e ffo orrms ms wire and wire forms wire and wire forms wire and wire forms Leone orthodontic wires are available in a range of alloys and different grades of elasticity and hardness to meet any therapeutic requirements.

More information

Torque capacity of metal and polycarbonate brackets with and without a metal slot

Torque capacity of metal and polycarbonate brackets with and without a metal slot European Journal of Orthodontics 26 (2004) 435 441 European Journal of Orthodontics vol. 26 no. 4 European Orthodontic Society 2004; all rights reserved. Torque capacity of metal and polycarbonate brackets

More information

Introducing the Clarity SL Self-Ligating Appliance System

Introducing the Clarity SL Self-Ligating Appliance System Introducing the Clarity SL Self-Ligating Appliance System Lunch and Learn Agenda Clarity SL Brackets Overview Consultation Benefits Bracket Positioning Benefits Bonding Benefits Leveling & Aligning Benefits

More information

Design of an Orthodontic Torque Simulator for Measurement of Bracket Deformation

Design of an Orthodontic Torque Simulator for Measurement of Bracket Deformation Design of an Orthodontic Torque Simulator for Measurement of Bracket Deformation G W Melenka 1, D S Nobes 1, P W Major 2 and J P Carey 1 1 Mechanical Engineering, Faculty of Engineering, University of

More information

GAC Archwire Collection

GAC Archwire Collection GAC Archwire Collection All the Metal That Matters Copper to Nickel and All Points In-Between Your choice in archwires is a critical decision in your quest for excellence. GAC has been recognized as a

More information

Residual stress analysis of fixed retainer wires after in vitro loading: can mastication-induced stresses produce an unfavorable effect?

Residual stress analysis of fixed retainer wires after in vitro loading: can mastication-induced stresses produce an unfavorable effect? Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich www.zora.uzh.ch Year: 2015 Residual stress analysis of fixed retainer wires after in vitro loading:

More information

Frictional and Biomechanical Forces and Biological Effects of a Trapezoidal Versus Rectangular Bracket

Frictional and Biomechanical Forces and Biological Effects of a Trapezoidal Versus Rectangular Bracket Avestia Publishing Journal of Biomedical Engineering and Biosciences (JBEB) Volume 3, Year 2016 ISSN: TBA DOI: TBA Frictional and Biomechanical Forces and Biological Effects of a Trapezoidal Versus Rectangular

More information

Canine Extrusion Technique with SmartClip Self-Ligating Brackets

Canine Extrusion Technique with SmartClip Self-Ligating Brackets Canine Extrusion Technique with SmartClip Self-Ligating Brackets Dr. Luis Huanca Ghislanzoni Dr. Luis Huanca received his DDS in 2006 and the MS and Specialist in Orthodontics in 2009 from the University

More information

INDICATIONS. Fixed Appliances are indicated when precise tooth movements are required

INDICATIONS. Fixed Appliances are indicated when precise tooth movements are required DEFINITION Fixed Appliances are devices or equipments that are attached to the teeth, cannot be removed by the patient and are capable of causing tooth movement. INDICATIONS Fixed Appliances are indicated

More information

Simply Great Wire! Great prices.

Simply Great Wire! Great prices. 2017 CATALOG Simply Great Wire! Great prices. CONTENTS 2-4 NiTi Wires 5 Stainless Steel Wires 6-7 Archforms 8 Beta Titanium Molybdenum Wires 9 Wire Auxiliaries WWW.HIGHLANDMETALS.COM NiTi Wires SuperElastic

More information

Introduction. Formerly known as Susan David

Introduction. Formerly known as Susan David European Journal of Orthodontics 20 (1998) 589 596 1998 European Orthodontic Society A comparative in vitro study of the frictional characteristics of two types of self-ligating brackets and two types

More information

Advantages. Fantastic. systems. Together with Unitek Lateral Development Archwires, this system puts you in control to

Advantages. Fantastic. systems. Together with Unitek Lateral Development Archwires, this system puts you in control to 3M Self-Ligating Appliances Lateral Development System Treatment Advantages Fantastic Finishes For orthodontic professionals who choose lateral development as a desired treatment outcome, 3M now offers

More information

GREAT LAKES BRACKETS. Treat with Quality. Treat with Confidence.

GREAT LAKES BRACKETS. Treat with Quality. Treat with Confidence. GREAT LAKES BRACKETS Treat with Quality. Treat with Confidence. BRACKET SYSTEMS DESIGNED BY ORTHODONTISTS, FOR ORTHODONTISTS, GREAT LAKES BRACKETS DELIVER. EVERY SINGLE BRACKET IS HAND INSPECTED AND ENGINEERED

More information

3D CONTROL HAS ARRIVED

3D CONTROL HAS ARRIVED TM 3D CONTROL HAS ARRIVED 1 Pitts 21 Self-Ligating System Pitts 21 Self-Ligating System Integrated Hooks An available option on the 3 s, 4 s and 5 s for your convenience at no extra charge. Smooth, Rounded

More information

Orthodontic bracket slot dimensions as measured from entire bracket series

Orthodontic bracket slot dimensions as measured from entire bracket series Original Article Orthodontic bracket slot dimensions as measured from entire bracket series Paul Brown a ; Warren Wagner b ; Hyden Choi c ABSTRACT Objective: To measure the slot dimensions of an entire

More information

SE NiTi Heat Activated NiTi Ultra T Heat Activated NiTi Stainless Steel Beta Titanium. Ultra T Archwires

SE NiTi Heat Activated NiTi Ultra T Heat Activated NiTi Stainless Steel Beta Titanium. Ultra T Archwires S1:1 170421 7390 SOLITAIRE Single Pack s OrthoQuest offers the highest quality wires available in the orthodontic industry. Our wires are designed to exact specifications by trained engineers and manufactured

More information

Clinical tips to increase consistency using a Straight-Wire Appliance

Clinical tips to increase consistency using a Straight-Wire Appliance Clinical tips to increase consistency using a Straight-Wire Appliance Drs. Tom Pitts and Duncan Brown discuss an efficient, gentle approach to orthodontic treatment verything should be made as simple as

More information

Outline the significance of the pre-adjusted Edgewise appliance system and useful bracket variations in orthodontics

Outline the significance of the pre-adjusted Edgewise appliance system and useful bracket variations in orthodontics Outline the significance of the pre-adjusted Edgewise appliance system and useful bracket variations in orthodontics Dr MJ Rowland-Warmann BSc BDS (Manc) MSc Aes.Med. (Lond) MJDF RCS (Eng) GDC: 178642

More information

Thermal Nitanium. Archwires. Reduced Friction. Accelerated Results.

Thermal Nitanium. Archwires. Reduced Friction. Accelerated Results. Accel Low Friction Thermal Nitanium Archwires Reduced Friction. Accelerated Results. A Revolution In Sliding Mechanics Accel Low Friction Thermal Nitanium Archwires are the start of a new age in nickel

More information

Victory Series Active Self-Ligating Brackets. Reliable. and. Effective

Victory Series Active Self-Ligating Brackets. Reliable. and. Effective Victory Series Active Self-Ligating Brackets Reliable and Effective Victory Series Active Self-Ligating Brackets Reliability Victory Series SL Brackets are widely accepted as an industry benchmark for

More information

Delta Force. Bracket System. Putting you in the driver s seat for ultimate control

Delta Force. Bracket System. Putting you in the driver s seat for ultimate control Delta Force Bracket System Putting you in the driver s seat for ultimate control Variable Force Orthodontics The Delta Force Bracket System incorporates an advanced design that allows you to control the

More information

Third-Order Torque and Self-Ligating Orthodontic Bracket Type Effects on Sliding Friction

Third-Order Torque and Self-Ligating Orthodontic Bracket Type Effects on Sliding Friction Original Article Third-Order Torque and Self-Ligating Orthodontic Bracket Type Effects on Sliding Friction Michael Chung a ; Robert J. Nikolai b ; Ki Beom Kim c ; Donald R. Oliver c ABSTRACT Objective:

More information

Bracket Systems. A good bracket is the foundation of a good treatment plan... Go ahead, choose the best. Metal Bracket.

Bracket Systems. A good bracket is the foundation of a good treatment plan... Go ahead, choose the best. Metal Bracket. A good bracket is the foundation of a good treatment plan... Go ahead, choose the best. Metal Bracket Self Ligating System Precimax Iridium Econo Lingual Bracket Aesthetic Bracket Cerabrace Krystal Ice

More information

Advantages. Fantastic. Treatment. Finishes. 3M Self-Ligating Appliances Lateral Development System. Unitek Lateral Development Archwires

Advantages. Fantastic. Treatment. Finishes. 3M Self-Ligating Appliances Lateral Development System. Unitek Lateral Development Archwires 3M Self-Ligating Appliances Lateral Development System Treatment Advantages Fantastic Finishes Adhesive Coated Appliance System Adhesive Coated Appliance System Unitek Lateral Development Archwires For

More information

Comparison of mechanical properties of beta-titanium wires between leveled and unleveled brackets: an in vitro study

Comparison of mechanical properties of beta-titanium wires between leveled and unleveled brackets: an in vitro study Insabralde et al. Progress in Orthodontics 2014, 15:42 RESEARCH ARTICLE Comparison of mechanical properties of beta-titanium wires between leveled and unleveled brackets: an in vitro study Open Access

More information

The Loading And Unloading Properties Of Various Arch Wires As A Function Of Cross Sectional Dimension And Inter Bracket Span Width System

The Loading And Unloading Properties Of Various Arch Wires As A Function Of Cross Sectional Dimension And Inter Bracket Span Width System The Loading And Unloading roperties Of Various Arch Wires As A Function Of Cross Sectional Dimension And Inter Bracket Span Width System Authors: Thomas Mathew MDS Orthodontics, Lecturer, Faculty of Dentistry,

More information

The Tip-Edge appliance and

The Tip-Edge appliance and Figure 1: Internal surfaces of the edgewise archwire slot are modified to create the Tip-Edge archwire slot. Tipping surfaces (T) limit crown tipping during retraction. Uprighting surfaces (U) control

More information

Developing lingual orthodontics

Developing lingual orthodontics New Options for Aesthetic Appliance Yeong Sim, DDS, MSD, PhD Director of Gorman Lingual Orthodontic Seminar Faculty of Samsung Medical Center & KyungHee Medical School Developing lingual orthodontics Dr

More information

Honing Damon System Mechanics for the Ultimate in Efficiency and Excellence Jeff Kozlowski, DDS East Lyme, CT

Honing Damon System Mechanics for the Ultimate in Efficiency and Excellence Jeff Kozlowski, DDS East Lyme, CT Honing Damon System Mechanics for the Ultimate in Efficiency and Excellence Jeff Kozlowski, DDS East Lyme, CT Dr. Kozlowski received his DDS degree and certificate in orthodontics from the State University

More information

Evaluation of Torsional Strength of Ceramic Brackets Produced by Arch Wire Twisting Moment

Evaluation of Torsional Strength of Ceramic Brackets Produced by Arch Wire Twisting Moment Evaluation of Torsional Strength of Produced by Arch Wire Twisting Moment Raghavendra S. Vhanbatte, 1 Rashmi V. Uddanwadiker, 2 Pallavi Patil 3 1 Postgraduate student, Department of Mechanical Engineering,

More information

LINGUAL STRAIGHTWIRE BRACKET SYSTEM Designed with Drs. Kyoto Takemoto and Giuseppe Scuzzo

LINGUAL STRAIGHTWIRE BRACKET SYSTEM Designed with Drs. Kyoto Takemoto and Giuseppe Scuzzo LINGUAL STRAIGHTWIRE BRACKET SYSTEM Designed with Drs. Kyoto Takemoto and Giuseppe Scuzzo THE WORLD S FIRST SELF-LIGATING SQUARE SLOT BRACKET Simplicity I Efficiency I Comfort ALIAS LINGUAL BRACKET SYSTEM

More information

The Rondeau System. Brackets, Bands, Buccal Tubes and Archwires. Orthodontic supplies recommended by Dr. Brock Rondeau

The Rondeau System. Brackets, Bands, Buccal Tubes and Archwires. Orthodontic supplies recommended by Dr. Brock Rondeau The Rondeau System Brackets, Bands, Buccal Tubes and Archwires Orthodontic supplies recommended by Dr. Brock Rondeau Achieve Consistently Accurate Placement The Di-MIM Mini-Twin Bracket s rhomboid shape

More information

Incognito. Appliance System goes Digital. 3Unitek. Now in 3D. The Incognito Appliance System Digital Setup with 3D PDF Documentation

Incognito. Appliance System goes Digital. 3Unitek. Now in 3D. The Incognito Appliance System Digital Setup with 3D PDF Documentation The Digital Setup with 3D PDF Documentation Incognito Appliance System goes Digital Now in 3D Now you can choose between a Manual or Digital Setup for the. High quality, precise 3D images of the Setup

More information

The management of impacted

The management of impacted Using a rigid hook and spring auxiliary slid onto the archwire to direct eruption of impacted teeth BY S. JAY BOWMAN, DMD, MSD, AND ALDO CARANO, DR ODONT, MS, SPEC ORTHOD Figure 1: A 12-year-old female

More information

There has been increased use of self-ligating. Friction between various self-ligating brackets and archwire couples during sliding mechanics

There has been increased use of self-ligating. Friction between various self-ligating brackets and archwire couples during sliding mechanics ORIGINAL ARTICLE Friction between various self-ligating brackets and archwire couples during sliding mechanics Sennay Stefanos, a Antonino G. Secchi, b Guy Coby, c Nipul Tanna, a and Francis K. Mante d

More information

System Orthodontic Treatment Program By Dr. Richard McLaughlin, Dr. John Bennett and Dr. Hugo Trevisi

System Orthodontic Treatment Program By Dr. Richard McLaughlin, Dr. John Bennett and Dr. Hugo Trevisi A Clinical Review of the MBT Versatile+ Appliance System Orthodontic Treatment Program By Dr. Richard McLaughlin, Dr. John Bennett and Dr. Hugo Trevisi Treatment Philosophy of the MBT Appliance System

More information

BRACKET CHOICES SYSTEM CHOICES. Empower 2 reconvertible molar tubes provide versatility without compromise.

BRACKET CHOICES SYSTEM CHOICES. Empower 2 reconvertible molar tubes provide versatility without compromise. IT S ALL ABOUT CHOICE Empower is the industry s most complete self ligating bracket system, bringing you choice like you ve never had before. The Empower family includes bracket and system options that

More information

Premium Heat Activated Arch Wire

Premium Heat Activated Arch Wire Premium Heat Activated Arch Wire TM Premium Heat Activated Arch Wire American Orthodontics is dedicated to continuous improvement, and delivers on this commitment by manufacturing our Tanzo Premium Heat

More information

The Science Behind The System Clinical Abstracts, Volume 2

The Science Behind The System Clinical Abstracts, Volume 2 The Science Behind The System Clinical Abstracts, Volume 2 Visit www.damonsystem.com for Damon product information, clinical procedures, seminar registrations and more! 1717 West Collins Avenue, Orange,

More information

Integrative Orthodontics with the Ribbon Arch By Larry W. White, D.D.S., M.S.D.

Integrative Orthodontics with the Ribbon Arch By Larry W. White, D.D.S., M.S.D. Integrative Orthodontics with the Ribbon Arch By Larry W. White, D.D.S., M.S.D. Abstract The ribbon arch previously had great popularity and utility early in the 20 th century, but lost its appeal as edgewise

More information

The load/deflection characteristics of thermally activated orthodontic archwires

The load/deflection characteristics of thermally activated orthodontic archwires European Journal of Orthodontics 25 (2003) 417 421 2003 European Orthodontic Society The load/deflection characteristics of thermally activated orthodontic archwires Farnaz Parvizi and W. P. Rock School

More information

The stability of orthodontic treatment is a major

The stability of orthodontic treatment is a major CLINICIAN'S CORNER Inadvertent tooth movement with fixed lingual retainers Timothy G. Shaughnessy, a William R. Proffit, b and Said A. Samara c Marietta and Suwanee, Ga, Abu Dhabi and Dubai, United Arab

More information

Is nickel titanium superior to multistranded stainless steel wire in aligning crowded lower anteriors? A comparative in-vivo study.

Is nickel titanium superior to multistranded stainless steel wire in aligning crowded lower anteriors? A comparative in-vivo study. IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861. Volume 9, Issue 3 (Jul.- Aug. 2013), PP 47-51 Is nickel titanium superior to multistranded stainless steel

More information

3M Incognito Appliance System extraction case study.

3M Incognito Appliance System extraction case study. SM 3M Health Care Academy 3M Incognito Appliance System extraction case study. Toru Inami, DDS, Ph.D. Dr. Toru Inami graduated from the Aichigakuin University School of Dentistry in 1976. From 1977 to

More information

The Comparison of Space Closure Rate between Conventional and Passive Self-ligating System Using Elastomeric Chain in Maxilla

The Comparison of Space Closure Rate between Conventional and Passive Self-ligating System Using Elastomeric Chain in Maxilla The Comparison of between Conventional and Passive Self-ligating System Using Elastomeric Chain in Maxilla Sabrina 1, Krisnawati 2 *, Benny Mulyono Soegiharto 2 1. Orthodontic Residency Program, Faculty

More information

H4 Speaker s Bureau. Dr. John Pobanz

H4 Speaker s Bureau. Dr. John Pobanz Join the Movement. H4 Speaker s Bureau Dr. John Pobanz D.D.S. University of Nebraska College of Dentistry 1996 Graduated with Highest Distinction M.S. in Oral Biology (bone physiology emphasis) University

More information

The effect of Gable angle size and spring activation distance of 0.016x0.022 NiTi and TMA sectional T-loop towards force, moment y.

The effect of Gable angle size and spring activation distance of 0.016x0.022 NiTi and TMA sectional T-loop towards force, moment y. Padjadjaran Journal of Dentistry 2009;21(2):78-87. The effect of angle size and spring activation distance of 0.016x0.022 NiTi and TM sectional T-loop towards force, moment y values Norman Wachyudi, Tono

More information

A Collection of High Quality Orthodontic Instruments

A Collection of High Quality Orthodontic Instruments A Collection of High Quality Orthodontic Instruments User friendly ergonomic design Tenvis instruments starts with a great design, the finest materials and a commitment to quality. Meticulously handcrafted

More information

A THESIS SUBMITTED TO THE FACULTY OF THE GRADUATE SCHOOL OF THE UNIVERSITY OF MINNESOTA BY

A THESIS SUBMITTED TO THE FACULTY OF THE GRADUATE SCHOOL OF THE UNIVERSITY OF MINNESOTA BY THE EFFECTIVENESS OF SURESMILE TECHNOLOGY TO ACHIEVE PREDICTED TREATMENT OUTCOME A THESIS SUBMITTED TO THE FACULTY OF THE GRADUATE SCHOOL OF THE UNIVERSITY OF MINNESOTA BY CHRISTOPHER JOHN VAUBEL IN PARTIAL

More information

A Comparison of Frictional Resistance between Active and Passive Self-ligating Brackets with Conventional Bracket Systems

A Comparison of Frictional Resistance between Active and Passive Self-ligating Brackets with Conventional Bracket Systems 10.5005/jp-journals-10015-1102 H ORIGINAL Jyothikiran et RESEARCH al A Comparison of Frictional Resistance between Active and Passive Self-ligating Brackets with Conventional Bracket Systems 1 H Jyothikiran,

More information

New Class of Appliance

New Class of Appliance A New Class of Appliance THE HARNICK SECTIONAL DISTALIZER 42 APRIL 2017 // orthotown.com A seasoned ortho s creation for treating Class 2 patients by David Harnick, DDS Introduction The correction of Class

More information

LINGUAL STRAIGHTWIRE BRACKET SYSTEM Designed with Drs. Kyoto Takemoto and Giuseppe Scuzzo

LINGUAL STRAIGHTWIRE BRACKET SYSTEM Designed with Drs. Kyoto Takemoto and Giuseppe Scuzzo LINGUAL STRAIGHTWIRE BRACKET SYSTEM Designed with Drs. Kyoto Takemoto and Giuseppe Scuzzo THE WORLD S FIRST SELF-LIGATING SQUARE SLOT BRACKET Simplicity I Efficiency I Comfort Twins Digital Auxiliaries

More information

A study of the frictional characteristics of four commercially available self-ligating bracket systems

A study of the frictional characteristics of four commercially available self-ligating bracket systems European Journal of Orthodontics 3 (28) 645 653 doi:1.193/ejo/cjn58 Advance Access publication 28 October 28 The Author 28. Published by Oxford University Press on behalf of the European Orthodontic Society.

More information

THE MBT VERSATILE+ APPLIANCE SYSTEM

THE MBT VERSATILE+ APPLIANCE SYSTEM THE MBT VERSATILE+ APPLIANCE SYSTEM McLaughlin, Bennett, Trevisi The MBT Versatile+ Appliance System THE DEVELOPMENT OF A TREATMENT MECHANICS AND APPLIANCE PHILOSOPHY The first fully programmed preadjusted

More information

Controlled Space Closure with a Statically Determinate Retraction System

Controlled Space Closure with a Statically Determinate Retraction System Original Article Controlled Space Closure with a Statically Determinate Retraction System Kwangchul Choy, DDS, MS, PhD a ; Eung-Kwon Pae, DDS, MSc, PhD b ; Kyung-Ho Kim, DDS, MS, PhD c ; Young Chel Park,

More information

The 20/20 Molar Tube. Ronald M. Roncone, D.D.S., M.S.

The 20/20 Molar Tube. Ronald M. Roncone, D.D.S., M.S. The 20/20 Molar Tube by Ronald M. Roncone, D.D.S., M.S. A) Finish torque STAGE 3: Interactive to Active GOALS 4-6 months B) Finish root uprighting C) Maintain arch form D) Set occlusion with active settling

More information

An Evaluation of Efficiency and Effectiveness of Self-ligating Bracket Systems: A Prospective Clinical Study

An Evaluation of Efficiency and Effectiveness of Self-ligating Bracket Systems: A Prospective Clinical Study JIOS 10.5005/jp-journals-10021-1133 RESEARCH ARTICLE An Evaluation of Efficiency and Effectiveness of Self-ligating Bracket Systems: A Prospective Clinical Study An Evaluation of Efficiency and Effectiveness

More information

Effect of passive self-ligating bracket placement on the posterior teeth on reduction of frictional force in sliding mechanics

Effect of passive self-ligating bracket placement on the posterior teeth on reduction of frictional force in sliding mechanics Original Article THE KOREAN JOURNAL of ORTHODONTICS pissn 2234-7518 eissn 2005-372X http://dx.doi.org/10.4041/kjod.2016.46.2.73 Effect of passive self-ligating bracket placement on the posterior teeth

More information

An In vitro comparative study of intrabracket width as it relates to torque between three different archwires

An In vitro comparative study of intrabracket width as it relates to torque between three different archwires University of Iowa Iowa Research Online Theses and Dissertations Spring 2016 An In vitro comparative study of intrabracket width as it relates to torque between three different archwires Terry Jay Schmitt

More information

Finite-element analysis of NiTi wire deflection during orthodontic levelling treatment

Finite-element analysis of NiTi wire deflection during orthodontic levelling treatment IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Finite-element analysis of NiTi wire deflection during orthodontic levelling treatment To cite this article: M F Razali et al

More information

Enhanced Control in the Transverse Dimension using the Unitek MIA Quad Helix System by Dr. Sven G. Wiezorek

Enhanced Control in the Transverse Dimension using the Unitek MIA Quad Helix System by Dr. Sven G. Wiezorek Enhanced Control in the Transverse Dimension using the Unitek MIA Quad Helix System by Dr. Sven G. Wiezorek Dr. Wiezorek studied dental medicine at Kiel University, Germany from 1987 to 1993. He then finished

More information

Use of a Tip-Edge Stage-1 Wire to Enhance Vertical Control During Straight Wire Treatment: Two Case Reports

Use of a Tip-Edge Stage-1 Wire to Enhance Vertical Control During Straight Wire Treatment: Two Case Reports Case Report Use of a Tip-Edge Stage-1 Wire to Enhance Vertical Control During Straight Wire Treatment: Two Case Reports Helen Taylor, BDS, MScD, DOrth, MOrth, FDSRCS(Eng) a Abstract: Vertical control is

More information

ARCHWIRE. An extensive selection of high quality wire types and arch forms WIRE PRODUCTS. Wires. TruFlex Nickel Titanium Archwire. TruFlex.

ARCHWIRE. An extensive selection of high quality wire types and arch forms WIRE PRODUCTS. Wires. TruFlex Nickel Titanium Archwire. TruFlex. ARCHWIRE An extensive selection of high quality wire types and arch forms Archwire P.76-77, P.79, P.84-85 Thermal Archwire Copper Archwire TRIFECTA Thermal Archwire BetaForce Beta Titanium Archwire P.78-79

More information

The Comprehensive Treatment Management System

The Comprehensive Treatment Management System The Comprehensive Treatment Management System Contact your suremile consultant to learn more! SureSmile is the most powerful Treatment Management System in orthodontics Defining new opportunities that

More information

Mollenhauer Aligning Auxiliary for Bodily Alignment of Blocked-out Lateral Incisors in Preadjusted Edgewise Appliance Therapy

Mollenhauer Aligning Auxiliary for Bodily Alignment of Blocked-out Lateral Incisors in Preadjusted Edgewise Appliance Therapy Original Research 10.5005/jp-journals-10021-1237 Mollenhauer Aligning Auxiliary for Bodily Alignment of Blocked-out Lateral Incisors in Preadjusted Edgewise Appliance Therapy 1 Sreekrishnan B Nair, 2 AT

More information

MemRx Orthodontic Appliances

MemRx Orthodontic Appliances MemRx Orthodontic Appliances Uses and Instructions The MemRx Fundamentals As the need for faster, more efficient treatment of non-compliant patients increases, orthodontic!technology and materials has

More information

History of NiTinol. Ni = Nickel Ti = Titanium N = Naval O = Ordnance L = Laboratory

History of NiTinol. Ni = Nickel Ti = Titanium N = Naval O = Ordnance L = Laboratory History of NiTinol Ni = Nickel Ti = Titanium N = Naval O = Ordnance L = Laboratory NiTinol was discovered in a laboratory of the U.S. Marines in 1962 (US-Naval Ordnance Laboratory). The U.S. Marines wanted

More information

(12) Patent Application Publication (10) Pub. No.: US 2005/ A1

(12) Patent Application Publication (10) Pub. No.: US 2005/ A1 (19) United States US 2005O130094A1 (12) Patent Application Publication (10) Pub. No.: US 2005/0130094A1 Graham (43) Pub. Date: Jun. 16, 2005 (54) ORTHODONTIC ACCESSORY ARCH BAR (52) U.S. Cl.... 433/20

More information

Gentle-Jumper- Non-compliance Class II corrector

Gentle-Jumper- Non-compliance Class II corrector 15 CASE REPORT Gentle-Jumper- Non-compliance Class II corrector Amit Prakash 1,O.P.Mehta 2, Kshitij Gupta 3 Swapnil Pandey 4 Deep Kumar Suryawanshi 4 1 Senior lecturer Bhopal - INDIA 2 Professor Bhopal

More information

Effect of Sodium Fluoride Mouthwash on the Frictional Resistance of Orthodontic Wires

Effect of Sodium Fluoride Mouthwash on the Frictional Resistance of Orthodontic Wires Original Article Effect of Sodium Fluoride Mouthwash on the Frictional Resistance of Orthodontic Wires Allahyar Geramy 1, Tabassom Hooshmand 2, Tahura Etezadi 3 1 Professor, Dental Research Center, Dentistry

More information

Experience with Contemporary Tip-Edge plus Technique A Case Report.

Experience with Contemporary Tip-Edge plus Technique A Case Report. IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861. Volume 13, Issue 3 Ver. I. (Mar. 2014), PP 12-17 Experience with Contemporary Tip-Edge plus Technique A Case

More information

Anterior tooth morphology and its effect on torque

Anterior tooth morphology and its effect on torque European Journal of Orthodontics 27 (2005) 258 262 doi:10.1093/ejo/cji007 The Author 2005. Published by Oxford University Press on behalf of the European Orthodontics Society. All rights reserved. For

More information

We Give Back. Reduce Emergency Visits! ATTACHMENTS. No More Poking Archwires! Crimpable Mini Stops. Tooth Tone Crimpable Mini Stops

We Give Back. Reduce Emergency Visits! ATTACHMENTS. No More Poking Archwires! Crimpable Mini Stops. Tooth Tone Crimpable Mini Stops TTCHMENTS We Give Back Ortho Technology is a proud Founder Level Partner of the merican ssociation of Orthodontists Foundation. Ortho Technology is committed to making a positive difference in the future

More information

Arch Development: Clear Fixed Appliance vs. Aligner Treatment

Arch Development: Clear Fixed Appliance vs. Aligner Treatment Arch Development: Clear Fixed Appliance vs. Aligner Treatment Today, for many of us, adults make up a significant portion of our patient base. Many of these adults have had treatment before and remember

More information

S.H. Age: 15 Years 3 Months Diagnosis: Class I Nonextraction Severe crowding, very flat profile. Background:

S.H. Age: 15 Years 3 Months Diagnosis: Class I Nonextraction Severe crowding, very flat profile. Background: S.H. Age: 15 Years 3 Months Diagnosis: Class I Nonextraction Severe crowding, very flat profile Background: This case was selected to illustrate the long-term impact of treatment planning on the face and

More information

LIMITING FACTORS AFFECTING STAGE I TREATMENT: A BIOMECHANICAL PERSPECTIVE. Christopher G. Gibson. Chapel Hill 2016

LIMITING FACTORS AFFECTING STAGE I TREATMENT: A BIOMECHANICAL PERSPECTIVE. Christopher G. Gibson. Chapel Hill 2016 LIMITING FACTORS AFFECTING STAGE I TREATMENT: A BIOMECHANICAL PERSPECTIVE Christopher G. Gibson A thesis submitted to the faculty at the University of North Carolina at Chapel Hill in partial fulfillment

More information

An Effectiv Rapid Molar Derotation: Keles K

An Effectiv Rapid Molar Derotation: Keles K An Effectiv ective e and Precise Method forf Rapid Molar Derotation: Keles K TPA Ahmet Keles, DDS, DMSc 1 /Sedef Impar, DDS 2 Most of the time, Class II molar relationships occur due to the mesiopalatal

More information

The Ultimate in Crystal Clear Aesthetics

The Ultimate in Crystal Clear Aesthetics The Ultimate in Crystal Clear Aesthetics SEM x500 SEM x30 SEM x25 SEM x30 Patient Aesthetics The macro photograph shows the clarity of the flawless PURE brackets. Made from pure monocrystalline sapphire

More information

Lingual correction of a complex Class III malocclusion: Esthetic treatment without sacrificing quality results.

Lingual correction of a complex Class III malocclusion: Esthetic treatment without sacrificing quality results. SM 3M Health Care Academy Lingual correction of a complex Class III malocclusion: Esthetic treatment without sacrificing quality results. Christopher S. Riolo, DDS, M.S, Ph.D. Dr. Riolo received his DDS

More information

Technique Guide. Orthodontic Bone Anchor (OBA) System. Skeletal implants for the orthodontic movement of the teeth.

Technique Guide. Orthodontic Bone Anchor (OBA) System. Skeletal implants for the orthodontic movement of the teeth. Technique Guide Orthodontic Bone Anchor (OBA) System. Skeletal implants for the orthodontic movement of the teeth. Table of Contents Introduction Orthodontic Bone Anchor (OBA) System 2 Indications and

More information

A New Low-Friction Ligation System

A New Low-Friction Ligation System 2005 JCO, Inc. May not be distributed without permission. www.jco-online.com A New Low-Friction Ligation System ARTURO FORTINI, MD, DDS MASSIMO LUPOLI, MD, DDS VITTORIO CACCIAFESTA, DDS, MS Many factors

More information

The Tip-Edge Concept: Eliminating Unnecessary Anchorage Strain

The Tip-Edge Concept: Eliminating Unnecessary Anchorage Strain Welcome Ron Not Ron? Click here. My Account The Tip-Edge Concept: Eliminating Unnecessary Anchorage Strain VOLUME 26 : NUMBER 03 : PAGES (165-178) 1992 CHRISTOPHER K. KESLING, DDS, MS Tooth movement in

More information

Archwire Insertion and Disengagement Instruments Technique Guide

Archwire Insertion and Disengagement Instruments Technique Guide Clarity SL and SmartClip SL3 Self-Ligating Brackets Archwire Insertion and Disengagement Instruments Technique Guide Recommended Archwire Insertion Instruments Single point torquing key for tight spots

More information

Anchorage system. tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282

Anchorage system. tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282 tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282 266 . innovative comprehensive efficient. Dentaurum Online Shop shop.dentaurum.com

More information

Victory Series Buccal Tubes Designed for Comfort and Performance. Advanced, Precise. Solutions. for Molars

Victory Series Buccal Tubes Designed for Comfort and Performance. Advanced, Precise. Solutions. for Molars Victory Series Buccal Tubes Designed for Comfort and Performance Advanced, Precise Solutions for Molars Your reliable molar appliance solution Victory Series Buccal Tubes provide you with a well-designed,

More information

Unilateral Horizontally Impacted Maxillary Canine and First Premolar Treated with a Double Archwire Technique

Unilateral Horizontally Impacted Maxillary Canine and First Premolar Treated with a Double Archwire Technique Case Report Unilateral Horizontally Impacted Maxillary Canine and First Premolar Treated with a Double Archwire Technique Chien-Lun Peng a ; Yu-Yu Su b ; Sheng-Yang Lee c Abstract: A patient with a unilateral

More information

Frictional resistance in monocrystalline ceramic brackets with conventional and nonconventional elastomeric ligatures

Frictional resistance in monocrystalline ceramic brackets with conventional and nonconventional elastomeric ligatures Galvão et al. Progress in Orthodontics 2013, 14:9 RESEARCH Open Access Frictional resistance in monocrystalline ceramic brackets with conventional and nonconventional elastomeric ligatures Mariana Bulhoes

More information

MATERIAL CHARACTERISTICS OF THE ORTHODONTIC ARCHWIRES

MATERIAL CHARACTERISTICS OF THE ORTHODONTIC ARCHWIRES DAAAM INTERNATIONAL SCIENTIFIC BOOK 20 pp. 30-308 CHAPTER 2 MATERIAL CHARACTERISTICS OF THE ORTHODONTIC ARCHWIRES SZUHANEK, C. Abstract Our aim was to investigate the mechanical properties of coated superelastic

More information

FRICTION PRODUCED BY ESTHETIC BRACKETS WITH VARYING LIGATION CHRISTINE KNOX ABENOJA

FRICTION PRODUCED BY ESTHETIC BRACKETS WITH VARYING LIGATION CHRISTINE KNOX ABENOJA FRICTION PRODUCED BY ESTHETIC BRACKETS WITH VARYING LIGATION by CHRISTINE KNOX ABENOJA ANDRE FERREIRA, COMMITTEE CHAIR LIONEL SADOWSKY JOHN BURGESS MARK LITAKER FIROZ RAHEMTULLA A THESIS Submitted to the

More information

Efficient Bonding Protocol for the Insignia Custom Bracket System

Efficient Bonding Protocol for the Insignia Custom Bracket System Efficient Bonding Protocol for the Insignia Custom Bracket System Abstract The Insignia appliance is reverse-engineered from a digital set-up of the prescribed dental alignment. Each bracket configuration,

More information

Focus on Minor Tooth Movement Systems

Focus on Minor Tooth Movement Systems Accutech Orthodontic Lab Inc (800) 734-7855 accutechortho.com Allesee Orthodontic Appliances (AOA Labs) (800) 262-5221 aoalab.com Appliance Therapy Group (800) 423-3270 ApplianceTherapy.com ClearCorrect

More information

04 Inserting the HYCON TUBE

04 Inserting the HYCON TUBE H y c o n D e v i c e f o r s p a c e c l o s u r e hycon HYCON t u b e placement information TUBE placement information 01 The HYCON TUBE block. The teeth of the front block should be laced with a figure

More information

Sliding mechanics in an extraction case treated with 3M Clarity ADVANCED Ceramic Brackets and the 3M MBT Appliance System.

Sliding mechanics in an extraction case treated with 3M Clarity ADVANCED Ceramic Brackets and the 3M MBT Appliance System. SM 3M Health Care Academy Sliding mechanics in an extraction case treated with 3M Clarity ADVANCED Ceramic Brackets and the 3M MBT Appliance System. Dr. Gianluigi Fiorillo Dr. Gianluigi Fiorillo received

More information