1 Bio-Medical Materials and Engineering 25 (2015) DOI /BME IOS Press The effect of pedicle screw implantation depth and angle on the loading and stiffness of a spinal fusion assembly Alexander Tsouknidas a,b, a Department of Mechanical Engineering, Aristotle University of Thessaloniki, Thessaloniki, Greece b Department of Mechanical Engineering, Frederick University, Nicosia, Cyprus Received 18 February 2014 Accepted 13 April 2015 Abstract. BACKGROUND: The increasing prevalence of spine disorders in industrialized environments has impaired the quality of life in the elder population. In an effort to relieve pain, physicians strive to improve treatment through the consideration of patient specific characteristics during preoperative planning of procedures such as spinal fusion. OBJECTIVE: This study aims at quantifying aspects of spondylodesis to the loading and mobility of the utilized instrumentation, as the use of rigid vs. motion sparing materials as well as implantation angle and depth of the pedicle screws are still subject to controversy among surgeons. METHODS: A fixation assembly was reverse engineered based on µct measurements of the involved instrumentation. Two pedicle screws were connected with a rod, thus representing a mono-segmental fixation device. The pedicle screws were embedded in hexahedral structures simulated by bone properties. Upon validation and verification, the response of the model to a compressive and a torsional load was simulated in ANSYS 14, while altering the implantation depth and insertion angle of the pedicle screws along with the rod material. RESULTS: The mobility of the instrumentation was drastically increased (by up to 390%) when PEEK rods were used in place of traditional Ti ones, a tendency observed at varying extent for all simulated scenarios. Shallow implantation induced a slight stress increase ( 21%) on the implant and a notable distressing of the bony tissue ( 44%), whereas inclined screw positioning was overall beneficial to the developing stress fields in both, with bone profiting a max. stress release of 15% during the application of torsion. CONCLUSIONS: The investigation presented refined insight into the biomechanical response of a spinal fusion device. As expected, rigid fixation seems preferable in fusion oriented instrumentation whereas semi rigid devices should be employed for non-fusion applications. Shallow implantation resulted in a slight posterior offset of the stabilization device, which could be beneficial in the treatment of osteoporotic patients. Keywords: Spine fusion, load transition, stiffness, implantation 1. Introduction Spinal fusion is predominantly used to relieve pain due to abnormal vertebral motion. Next to treatment of degenerative conditions, spine fixation is also considered the preferred approach for the restoration of deformities, i.e. scoliosis [1]. Initial techniques favoured immobilization trough rigid instrumen- * Address for correspondence: Alexander Tsouknidas, Department of Mechanical Engineering, Aristotle University Thessaloniki, 9th floor of Building D, Aristotle Campus, Thessaloniki, Greece. alextso@auth.gr /15/$ IOS Press and the authors. All rights reserved

2 426 A. Tsouknidas / The effect of pedicle screw implantation parameters on a spinal fusion assembly tation. The interspace rigidity provided by these systems, facilitates restoration of extensive congenital and degenerative deformities, but exposes the implants to excessive diurnal loading if no fusion occurs. This renders the fixation device susceptible to rod slippage, screw loosening [2] and eventual failure [3]. Rigid fusion constructs have also been associated to pseudoarthrosis [4] due to stress shielding of interbody joints. Even though, motion preservation techniques have been introduced as a promising alternative, the use of PEEK or Ti rods in posterior fusions devices, still remains a subject to controversy [5]. The use of a semi-rigid rod preserves the integrity of the native disk providing balanced load sharing, thus shielding the instrumentation from failure [6]. Recent studies suggest that motion-sparing stabilization may also promote fusion [7] based on Wolf s law. Clinical practice however, recommends the use of motion sparing systems as adjuncts to fusion treatments, predominantly aiming at relieving intradiscal pressure and off-load intrasegmental facet joints [8]. As the sophistication of fixation systems continues to evolve, their patient specific implementation gains importance. Several studies investigate the interrelation of assembly features such as spacer length [9] and rod material [6] to inter-segmental motion, while Finite Element (FE) models have been successfully employed to addressed the occurring biomechanics of transpedicle screw fixation [10,11]. Certain process parameters however, e.g. implantation angle and depth of the pedicle screws, remain subject to the acting surgeon s discretion, as they have never been investigated in a systematic way. The aim of this study was to quantify the importance of these parameters. The effect of the insertion angle was investigated on the preliminary hypothesis that altering the geometry of the instrumentation would affect the load transfer within the system. Two rod materials (Ti alloy and PEEK) representing a rigid and a motion sparing configuration were simulated during the analysis to evaluate the effect of fixation stiffness on the developing stress field. The implantation depth was considered as the final surgical parameter under investigation. 2. Methods 2.1. Model development A commercial polyaxial pedicle screw (ARMADA Spinal System) with a double lead thread and a provisional locking system was considered as the main component of the stabilization device. This type of implant facilitates spanning of screws over adjacent vertebral levels, at a predefined relative angle within the sagittal plane and their subsequent posterior stabilization through a surgical rod. The simulated configuration followed a similar principle. The as received pedicle screw was reverse engineered by subjecting the implant (pre-operatively) to consecutive micro Computed Tomography (µct) measurements on a Werth TomoScope HV Compact device. Measurements were conducted at a spatial resolution of 10 µm, to digitize its 3D shape and facilitate its analysis in a Finite Element (FE) oriented software. Some features of the implant were considered as redundant for the analysis (i.e. internal immobilization mechanism) as their consideration would increase the complexity and computational effort, without contributing to the bio-realisticity of the model. The implant, some internal details as well as the final 3D model considered during the simulation, is demonstrated in Fig. 1. Each screw was embedded in a cubic geometry imitating vertebral bone. This geometry consisted of three compartments, the first one considering material properties of pedicle, the second one cancellous and the latter one cortical bone. This consideration is important as the pedicles have been reported to carry up to 80% of the load induces by spinal fusion devices [12]. The part of the geometry imitating

3 A. Tsouknidas / The effect of pedicle screw implantation parameters on a spinal fusion assembly 427 Fig. 1. Physical vs. reverse engineered implant geometry (model simplifications). (Colors are visible in the online version of the article; pedicle bone covered 17.8 mm of the pedicle screw [13], the anterior cortical shell was considered to be 1 mm thick [14] and the interposing volume reflected the trabecular structure of the vertebra [15]. The first implantation scenario considered the pedicle screw to be embedded 42 mm deep into the bone matter whereas 10 mm of the screw s body length were revealed during shallow insertion. In either case, the pedicle screw spanned over the cancellous bone part into the anterior cortical shell, thus the bone cube hosting the screw was shortened in the latter implantation scenario. This approach was chosen as pedicle bone and anterior cortical shell (first and third bone compartment) are the main contributors to the implant s anchorage and were therefore considered to maintain a constant length throughout both implantation depth scenarios. This hypothesis was considered critical, as the aim of the study was to

4 428 A. Tsouknidas / The effect of pedicle screw implantation parameters on a spinal fusion assembly Fig. 2. Load magnitude/direction (compression and torsion), boundary condition and relative implantation angle of the pedicle screws considered during the modelled scenarios. (Colors are visible in the online version of the article; /BME ) evaluate the impact of the anterior posterior sift of the stabilization and not its load distributing capacity among bone types. During the simulation, the inferior support was always restricted of any movement and the load applied to the upper body. The response of the system was examined for two typical loading scenarios [16,17]. The first one consisted of a 50 N compressive load, which was applied on the upper fixation device, whereas a second loading scenario examined the biomechanical response of the system to 10 Nm torsion. The centre of rotation was assumed at the intersection of a theoretical line between the dorsal and ventral borders of vertebral body and the widest distance between the lateral borders of the vertebral body. This resulted in an offset of the torque application point, which was considered as a remote load at the upper body of the fixation device. Based on morphological characteristics of the spine available in literature [18], the torque was applied at a 14 mm offset to the sagittal plane, 15 mm one to coronal plane and 12.5 mm to the transverse plane of the outmost implantation point of the upper transpedicle screw (see Fig. 2). The model was simulated in ANSYS 14.0 considering the linear elastic material properties [19] presented in Table Theoretical verification Determining the bio-realisticity of a numerical analysis is of vital importance when assessing the clinical perspective of a trauma. The verification of a FE model is fundamental aspect of the analysis [20], as erroneous predictions may yield catastrophic complications. The verification of the theoretical model was achieved through the generation of a mesh independent FEgrid [21]. Convergence studies, conducted separately on every model entity (pedicle screw, rod etc.),

5 A. Tsouknidas / The effect of pedicle screw implantation parameters on a spinal fusion assembly 429 Table 1 Material properties considered during the simulation Young s modulus [MPa] Poisson s ratio Ref. Cortical bone [25] Cancellous bone [26] Posterior element [27] Titanium [27] Peek [28] indicated the optimum element density in terms of computational efficiency and results accuracy. To avoid element shear locking and hourglassing phenomena during the analysis, second order reduced integration elements were employed throughout the model Model validation There exists a consensus throughout literature that inter-patient variability is a significant limitation of medical FE models. The model was therefore validated against trend values which were experimentally determined. Two pedicle screws were embedded in two PMMA cubes and held together through a Ti rod, imitating a complete (unilateral) fixation assembly. The FE model was then evaluated against several uniaxial compression scenarios within the elastic response of the system. The hierarchy of this validation process emphasizes on the ability of the model to simulate benchmark scenarios [22] fitting the purpose of this study. Based on the foregoing procedure, the developed model is accepted to provide an adequate degree of confidence for a comparative evaluation of the procedure variables involved in spinal fusion. 3. Results The implantation of the pedicle screws at a 10 relative angle, lead to a stress decrease of approximately 10% for the Ti rod compressive scenarios, as illustrated in Fig. 3(a). The use however of PEEK rods significantly increases the mobility of the construct thus changing the loading angle of the pedicle screws which induces a stress increase of up to 15%. The compressive stiffness of the fixation system predominantly depended on the rod material, with the implantation angle having a negligible and the implantation depth a marginable effect on the calculated deformation (see Fig. 3(b)). During the application of torsion, the bone implant interface exhibited notably lower stress values for shallow implantation. This was more pronounced in parallel embedded implants, with the PEEK rods favouring peri-implant stress concentrations, as observed in Fig. 4(a). The range of motion increased drastically for all PEEK based immobilization scenarios in particular for shallow embedded implants where the rotation of the system increased by up to 390%, as shown in Fig. 4(b). This was less pronounced during the use of Ti rods where the stiffness of the system was slightly higher for inclined screw positioning, but comparable for shallow and deep implantation. 4. Discussion There exists a consensus in literature that the anchorage of pedicle screws strongly depends on their implantation depth [23] and thus loosening is favoured by partially embedded screws. Clinical practice

6 430 A. Tsouknidas / The effect of pedicle screw implantation parameters on a spinal fusion assembly Fig. 3. (a) Max. von Mises stress developing on the transpedicle screws and (b) resulting mobility of the system, for all compressive scenarios (implantation depth/angle and rod material dependent). (Colors are visible in the online version of the article; also suggests that optimum anchorage is achieved if the implant spams through the vertebral body into its anterior cortical shell [24]. This has lead physicians to assembly spinal fusion constructs as close to the vertebral column as possible. A similar tendency was observed in this study, as the results indicate that implantation should always be performed as deep as possible in patients with healthy Bone Mineral Density (BMD) values. In these cases the choice of rod material entirely depends on desired rigidity of the immobilized segment. It should however be noted that the use of Ti rods restricts the mobility of the fused segment, while being able to carry a significant portion of the load the instrumented level is subjected to. The investigation, provided refined insight for spinal fusion of osteoporotic patients, as the decreased BMD would render individual bone struts susceptible to fracture at lower stress values than in healthy patients. Therefore shallow implantation might be preferable in osteoporotic patients as the reduced stress concentrations in the bone implant interface could contribute towards the long term stability of the implant. This effect was observed for parallel and inclined screw positioning, with the rod stiffness

7 A. Tsouknidas / The effect of pedicle screw implantation parameters on a spinal fusion assembly 431 Fig. 4. (a) Max. von Mises stress developing on the vertebral bone and (b) resulting rotational mobility of the system, for all torsional scenarios (implantation depth/angle and rod material dependent). (Colors are visible in the online version of the article; having a pronounced effect on the mobility of the system. This is expected to drastically affect the intradiscal pressure within the immobilized segment and therefore inclined positioned implants, connected with PEEK rods, could result in hyper-physiological loading of a degenerated disc, potentially further disturbing the balance of a pathological disc s matrix under prolonged exposure. This should however be taken into account while considering clinical practice, suggesting that a longer pedicle screw could provide both, proper anchorage and stress relief in the bone implant interface. The considerable variations observed in the flexibility of stabilization devices, utilizing different rod materials, suggest that a rigid fixation (e.g. Ti rods) should be used for fusion of a spine segment whereas motion-sparing stabilization is preferable for non-fusion applications [8]. Acknowledgements The author would like to acknowledge that this investigation was funded by the General Secretariat for Research and Technology of Greece under grant agreement PE8(3227) and thank BETA CAE Systems SA for providing him with the CAE pre-processor ANSA, used during surface and volume meshing of the model.

8 432 A. Tsouknidas / The effect of pedicle screw implantation parameters on a spinal fusion assembly References [1] S.S. Rajaee, H.W. Bae, L.E. Kanim and R.B. Delamarter, Spinal fusion in the united states: Analysis of trends from 1998 to 2008, Spine 37(1) (2012), [2] K.J. Cho, S.I. Suk, S.R. Park, J.H. Kim, S.B. Kang, H.S. Kim and S.J. Oh, Risk factors of sagittal decompensation after long posterior instrumentation and fusion for degenerative lumbar scoliosis, Spine 35(17) (2012), [3] M. Wedemeyer, S. Parent, A. Mahar, T. Odell, T. Swimmer and P. Newton, Titanium versus stainless steel for anterior spinal fusions: An analysis of rod stress as a predictor of rod breakage during physiologic loading in a bovine model, Spine 32 (2007), [4] R.C. Huang, F.P. Girardi, M.R. Lim and F.P. Cammisa, Advantages and disadvantages of nonfusion technology in spine surgery, Orthop. Clin. North Am. 36(3) (2005), [5] D.K. Sengupta and H.N. Herkowitz, Pedicle screw-based posterior dynamic stabilization: Literature review, Adv. Orthop (2012), Article ID [6] L. Qi, M. Li, S. Zhang, J. Xue and H. Si, Comparative effectiveness of PEEK rods versus titanium alloy rods in lumbar fusion: A preliminary report, Acta Neurochir. 155(7) (2013), [7] M. Moumene and P. Afshari, Biomechanical comparison of rigid vs. semi-rigid rods in spinal fusion constructs, in: EuroSpine 2011, Milan, Italy, October 2011, [8] C.Y. Barrey, R.K. Ponnappan, J. Song and A.R. Vaccaro, Biomechanical evaluation of pedicle screw-based dynamic stabilization devices for the lumbar spine: A systematic review, SAS Journal 2(4) (2008), [9] C.A. Niosi, Q.A. Zhu, D.C. Wilson, O. Keynan, D.R. Wilson and T.R. Oxland, Biomechanical characterization of the three-dimensional kinematic behaviour of the dynesys dynamic stabilization system: An in vitro study, Eur. Spine J. 15(6) (2008), [10] A. Faizan, V.K. Goel, A. Biyani, S.R. Garfin and C.M. Bono, Adjacent level effects of bi level disc replacement, bi level fusion and disc replacement plus fusion in cervical spine A finite element based study, Clin. Biomech. 27(3) (2012), [11] A. Rohlmann, N.K. Burra, T. Zander and G. Bergmann, Comparison of the effects of bilateral posterior dynamic and rigid fixation devices on the loads in the lumbar spine: A finite element analysis, Eur. Spine J. 16(8) (2007), [12] J.N. Weinstein, B.L. Rydevik and W. Rauschning, Anatomic and technical considerations of pedicle screw fixation, Clin. Orthop. Relat. Res. 284 (1992), [13] S.G. López, E.U. Vega, O. De la Garza Castro and R.E.E. Omaña, Morphometry of pedicle and vertebral body in a mexican population by CT and fluroscopy, Int. J. Morphol. 27(4) (2009), [14] W.T. Edwards, Y. Zheng, L.A. Ferrara and H.A. Yuan, Structural features and thickness of the vertebral cortex in the thoracolumbar spine, Spine 26(2) (2001), [15] A. Tsouknidas, G. Maliaris, S. Savvakis and N. Michailidis, Anisotropic post-yield response of cancellous bone simulated by stress strain curves of bulk equivalent structures, Comput. Methods Biomech. Biomed. Engin. (2013). doi: / [16] M.M. Panjabi, T.R. Oxland, I. Yamamoto and J.J. Crisco, Mechanical behavior of the human lumbar and lumbosacral spine as shown by three-dimensional load displacement curves, J. Bone Joint Surg. Am. 76 (1994), [17] A. Tsouknidas, N. Michailidis, S. Savvakis, K. Anagnostidis, K.D. Bouzakis and G. Kapetanos, A finite element model technique to determine the mechanical response of a lumbar spine segment under complex loads, J. Appl. Biomech. 28(4) (2012), [18] H.J. Wilke, A. Kettler, K.H. Wenger and L.E. Claes, Anatomy of the sheep spine and its comparison to the human spine, Anat. Rec. 247(4) (1997), [19] M.F. Gornet, F.W. Chan, J.C. Coleman, B. Murrell, R.P. Nockels, B.A. Taylor, T.H. Lanman and J.A. Ochoa, Biomechanical assessment of a PEEK rod system for semi-rigid fixation of lumbar fusion constructs, J. Biomech. Eng. 133(8) (2011), [20] M. Viceconti, S. Olsen, L.P. Nolte and K. Burton, Extracting clinically relevant data from finite element simulations, Clin. Biomech. 20 (2005), [21] A. Tsouknidas, S. Savvakis, Y. Asaniotis, K. Anagnostidis, A. Lontos and N. Michailidis, The effect of kyphoplasty parameters on the dynamic load transfer within the lumbar spine considering the response of a bio-realistic spine segment, Clin. Biomech. 28(9,10) (2013), [22] A. Erdemir, S. McLean, W. Herzog and A.J. van den Bogert, Model-based estimation of muscle forces exerted during movements, Clin. Biomech. 22(2) (2007), [23] Z. Wu, S.A. Nassar and X. Yang, Pullout performance of self-tapping medical screws, J. Biomech. Eng. 133(11) (2011), [24] S.R. Kantelhardt, C.H. Bock, J. Larsen, V. Bockermann, W. Schillinger, V. Rohde and A. Giese, Intraosseous ultrasound in the placement of pedicle screws in the lumbar spine, Spine 34(4) (2009),

9 A. Tsouknidas / The effect of pedicle screw implantation parameters on a spinal fusion assembly 433 [25] E.C. Teo and K.K. Lee, An accurately represented finite element model of lumbar motion segment (L2 L3), in: Proc. Int. Conf. Biomedical Engineering, 2002, pp [26] V.K. Goel, Y.E. Kim, H. Lim and J.N. Weinsten, An analytical investigation of the mechanics of spinal instrumentation, Spine 13 (1988), [27] D.J. Maiman, S. Kumaresan, N. Yoganandan and F.A.F.A. Pintar, Biomechanical effect of anterior cervical spine fusion on adjacent segments, Biomed. Mater. Eng. 9 (1999), [28] S. Vadapalli, K. Sairyo, V.K. Goel, M. Robon, A. Biyani, A. Khandha and N.A. Ebraheim, Biomechanical rationale for using polyetheretherketone (PEEK) spacers for lumbar interbody fusion A finite element study, Spine 31(26) (2006), E992 E998.

FINITE ELEMENT BASED IMPLANT OPTIMIZATION AND PREOPERATIVE PREPARATION

FINITE ELEMENT BASED IMPLANT OPTIMIZATION AND PREOPERATIVE PREPARATION FINITE ELEMENT BASED IMPLANT OPTIMIZATION AND PREOPERATIVE PREPARATION 1,4 Alexander Tsouknidas *, 2 Savvas Savvakis, 3 Kleovoulos Anagnostidis, 2 Yiannis Asaniotis, 4 Antonios Londos, 1 Nikolaos Michailidis

More information

Finite element based surgical optimization

Finite element based surgical optimization Finite element based surgical optimization Alexander Tsouknidas 1, Savvas Savvakis 2, Yiannis Asaniotis 2, Nikolaos Michailidis 1 1 Department of Mechanical Engineering, Aristotle University of Thessaloniki,

More information

Multisegmental fusion of the lumbar spine a curse or a blessing?

Multisegmental fusion of the lumbar spine a curse or a blessing? Current Directions in Biomedical Engineering 2015; 1:376 380 S. Bauer* and D. Paulus* Multisegmental fusion of the lumbar spine a curse or a blessing? A MultiBodySimulation (MBS) modeling Abstract: Excessive

More information

An Advanced 3D Multi-Body System Model for the Human Lumbar Spine

An Advanced 3D Multi-Body System Model for the Human Lumbar Spine An Advanced 3D Multi-Body System Model for the Human Lumbar Spine Sousa, S. 1 and Claro, J. C. P. 2 1 University of Minho, Portugal, e-mail: sofia_sts@hotmail.com 2 University of Minho, Portugal, e-mail:

More information

Introduction. Original Article. Abstract

Introduction. Original Article. Abstract J Musculoskel Neuron Interact 2004; 4(2):152-158 Original Article Hylonome Loading simulation of lumbar spine vertebrae during a compression test using the finite elements method and trabecular bone strength

More information

Thoracolumbar Spine Locking Plate (TSLP) System. A low-profile plating system for anterior stabilization of the thoracic and lumbar spine.

Thoracolumbar Spine Locking Plate (TSLP) System. A low-profile plating system for anterior stabilization of the thoracic and lumbar spine. Thoracolumbar Spine Locking Plate (TSLP) System. A low-profile plating system for anterior stabilization of the thoracic and lumbar spine. Technique Guide Instruments and implants approved by the AO Foundation

More information

INTRODUCTION.

INTRODUCTION. online ML Comm www.jkns.or.kr http://dx.doi.org/1.334/jkns.16.59..91 J Korean Neurosurg Soc 59 () : 91-97, 16 Print ISSN 5-3711 On-line ISSN 1598-7876 Copyright 16 The Korean Neurosurgical Society Laboratory

More information

A Finite Element Study of the Stress Redistribution of the Lumbar Spine after Posterior Lumbar Interbody Fusion Surgery

A Finite Element Study of the Stress Redistribution of the Lumbar Spine after Posterior Lumbar Interbody Fusion Surgery A Finite Element Study of the Stress Redistribution of the Lumbar Spine after Posterior Lumbar Interbody Fusion Surgery Hsuan-Teh Hu 1, Kuo-Yuan Huang 2,3, Che-Jung Liu 1, Ching-Sung Kuo 1,4 1 Department

More information

KINEMATIC RESPONSE OF THE L4-L5 FUNCTIONAL SPINAL UNIT AFTER A LATERAL LUMBAR FUSION SURGERY

KINEMATIC RESPONSE OF THE L4-L5 FUNCTIONAL SPINAL UNIT AFTER A LATERAL LUMBAR FUSION SURGERY Proceedings of the 1 st Iberic Conference on Theoretical and Experimental Mechanics and Materials / 11 th National Congress on Experimental Mechanics. Porto/Portugal 4-7 November 2018. Ed. J.F. Silva Gomes.

More information

Idiopathic scoliosis Scoliosis Deformities I 06

Idiopathic scoliosis Scoliosis Deformities I 06 What is Idiopathic scoliosis? 80-90% of all scolioses are idiopathic, the rest are neuromuscular or congenital scolioses with manifest primary diseases responsible for the scoliotic pathogenesis. This

More information

Int J Clin Exp Med 2018;11(2): /ISSN: /IJCEM Yi Yang, Hao Liu, Yueming Song, Tao Li

Int J Clin Exp Med 2018;11(2): /ISSN: /IJCEM Yi Yang, Hao Liu, Yueming Song, Tao Li Int J Clin Exp Med 2018;11(2):1278-1284 www.ijcem.com /ISSN:1940-5901/IJCEM0063093 Case Report Dislocation and screws pull-out after application of an Isobar TTL dynamic stabilisation system at L2/3 in

More information

А тация 69, 47, Ц Д - ( Д). -,. - ANSYS. И,,.,,.,.

А тация 69, 47, Ц Д - ( Д). -,. - ANSYS. И,,.,,.,. А тация 69, 47, 18 31. Ц Д - ( Д). -,. - ANSYS. И,,.,,.,. Abstract Master's dissertation was set out on 69 pages with 47 figures, 18 tables and 31 literature sources. The purpose of this work was to investigate

More information

A morphometric study of the Pedicles of dry human typical lumbar vertebrae

A morphometric study of the Pedicles of dry human typical lumbar vertebrae Original article: A morphometric of the Pedicles of dry human typical lumbar vertebrae Dhaval K. Patil 1 *, Pritha S. Bhuiyan 2 1Resident, Department of Anatomy, Seth G S Medical College, Parel, Mumbai-400012,

More information

Biomechanical Behavior of L3-L5 Vertebrae in Six Cadaveric Spines of Fusion Cage with or without Posterolateral and Bilateral Instrumentation

Biomechanical Behavior of L3-L5 Vertebrae in Six Cadaveric Spines of Fusion Cage with or without Posterolateral and Bilateral Instrumentation Cronicon OPEN ACCESS EC ORTHOPAEDICS Research Article Biomechanical Behavior of L3-L5 Vertebrae in Six Cadaveric Spines of Fusion Cage with or without Posterolateral and Bilateral Takahiro Funato 1 *,

More information

EFSPINE CERVICAL COMBINED SET DISC PROTHESIS ORGANIZER BOX

EFSPINE CERVICAL COMBINED SET DISC PROTHESIS ORGANIZER BOX EFSPINE CERVICAL COMBINED SET INSTRUMENTS CERVICAL CAGE & DISC PROTHESIS ORGANIZER BOX Cervical Thoracic Thoraco - Lumbar Sacral EFSPINE CERVICAL COMBINED SET CERVICAL IMPLANTS INTRODUCTION Cervical Disc

More information

PARADIGM SPINE. Anterior Cervical Fusion Cage. Cervical Interbody Fusion

PARADIGM SPINE. Anterior Cervical Fusion Cage. Cervical Interbody Fusion PARADIGM SPINE Anterior Cervical Fusion Cage Cervical Interbody Fusion DESIGN RATIONALE The OptiStrain TM C* interbody fusion cage follows well established biomechanical principles: The slot design of

More information

USS Variable Axis Screw (VAS) System. For posterior fixation of the lumbar spine.

USS Variable Axis Screw (VAS) System. For posterior fixation of the lumbar spine. USS Variable Axis Screw (VAS) System. For posterior fixation of the lumbar spine. Technique Guide Instruments and implants approved by the AO Foundation Table of Contents Introduction USS Variable Axis

More information

USS Variable Axis Screw

USS Variable Axis Screw USS Variable Axis Screw Polyaxial side-opening pedicle screw Surgical technique Original Instruments and Implants of the Association for the Study of Internal Fixation AO/ASIF USS Variable Axis Screw

More information

Lumbar Dynamic. Reinhold Knebel Product Manager

Lumbar Dynamic. Reinhold Knebel Product Manager Lumbar Dynamic Reinhold Knebel Product Manager The DSS system provides options for dynamic and rigid stabilization to treat hypermobile segments and segments that require fusion. Allows pedicle displacement

More information

Preliminary measurements of lumbar spine kinematics and stiffness

Preliminary measurements of lumbar spine kinematics and stiffness 5 th Australasian Congress on Applied Mechanics, ACAM 2007 10-12 December 2007, Brisbane, Australia Preliminary measurements of lumbar spine kinematics and stiffness L. Jirková 1, Z. Horák 1, R. Sedláek

More information

Lumbo-sacral destruction fixation biomechanics

Lumbo-sacral destruction fixation biomechanics Lumbo-sacral destruction fixation biomechanics Amin Joukhar MSc * Jwalant S. Mehta FRCS Orth ^ David Marks FRCS Orth ^ Prof Vijay Goel PhD * ^ Birmingham Spine Centre, England *Engineering Centre for Orthopaedic

More information

Thoracolumbar Anterior Compression (TAC) System. Allows distraction, compression, and lateral fixation of the lower thoracic and lumbar spine.

Thoracolumbar Anterior Compression (TAC) System. Allows distraction, compression, and lateral fixation of the lower thoracic and lumbar spine. Thoracolumbar Anterior Compression (TAC) System. Allows distraction, compression, and lateral fixation of the lower thoracic and lumbar spine. Technique Guide Instruments and implants approved by the AO

More information

DEVELOPMENT OF L1 TO L5 FINITE ELEMENT MODEL OF LUMBAR SPINE: A BIOMECHANICAL STUDY

DEVELOPMENT OF L1 TO L5 FINITE ELEMENT MODEL OF LUMBAR SPINE: A BIOMECHANICAL STUDY Volume 118 No. 20 2018, 829-838 ISSN: 1311-8080 (printed version); ISSN: 1314-3395 (on-line version) url: http://www.ijpam.eu ijpam.eu DEVELOPMENT OF L1 TO L5 FINITE ELEMENT MODEL OF LUMBAR SPINE: A BIOMECHANICAL

More information

4.5 System. Surgical Technique. This publication is not intended for distribution in the USA.

4.5 System. Surgical Technique. This publication is not intended for distribution in the USA. 4.5 System Surgical Technique This publication is not intended for distribution in the USA. Contents EXPEDIUM 4.5 Spine System 2 Features and Benefits 3 Surgical Technique Extended Tandem Connector 4 Placement

More information

Effect of implant stiffness on spinal growth in the pig spine

Effect of implant stiffness on spinal growth in the pig spine Paediatric & Geriatric Orthopaedics Effect of implant stiffness on spinal growth in the pig spine R Rizza 1 *, XC Liu 2 *, J Thometz 2, C Tassone 2 Page 1 of 5 Abstract Introduction According to the vicious

More information

A U X I L I A R Y C O N N E C T O R S Surgical Technique

A U X I L I A R Y C O N N E C T O R S Surgical Technique A U X I L I A R Y C O N N E C T O R S Surgical Technique AUXILIARY CONNECTORS ISSYS LP Auxiliary Connectors The ISSYS LP auxiliary connectors were designed to provide medial-lateral variability for the

More information

A Fusion of Versatility, Performance and Ease.

A Fusion of Versatility, Performance and Ease. MATRIX Spine System. Degenerative and Deformity. System Guide A Fusion of Versatility, Performance and Ease. Instruments and implants approved by the AO Foundation The MATRIX Spine System is a universal

More information

PACH PLATE 2 OR 4 HOLE FIXATION

PACH PLATE 2 OR 4 HOLE FIXATION 2 OR 4 HOLE FIXATION The Science of Fusion PACH_v6.indd 1 6/03/2015 7:10 am PACH Plate Design Rationale The PACH Plate has been designed to be used as an anterior or lateral fixation plate for the thoracolumbar,

More information

The Effects of L4/5 Fusion on the Adjacent. Segments in the Lumbar Spine

The Effects of L4/5 Fusion on the Adjacent. Segments in the Lumbar Spine The Effects of L4/5 Fusion on the Adjacent Segments in the Lumbar Spine A Thesis submitted to The University of Manchester for the degree of Doctor of Philosophy in the Faculty of Engineering and Physical

More information

Visit our website on www.biotech-medical.com The DLP - Dorso-Lumbar Polyaxial Screw System has been designed to address the pathologies of the thoracolumbar spine. The DLP System contains a wide range

More information

5/19/2017. Interspinous Process Fixation with the Minuteman G3. What is the Minuteman G3. How Does it Work?

5/19/2017. Interspinous Process Fixation with the Minuteman G3. What is the Minuteman G3. How Does it Work? Interspinous Process Fixation with the Minuteman G3 LLOYDINE J. JACOBS, MD CASTELLVI SPINE MEETING MAY 13, 2017 What is the Minuteman G3 The world s first spinous process plating system that is: Minimally

More information

Royal Oak Cervical Plate System

Royal Oak Cervical Plate System Royal Oak Cervical Plate System Manufactured by Nexxt Spine, Inc. Royal Oak Cervical Plate System INTRODUCTION FEATURES AND BENEFITS Table of Contents SURGICAL TECHNIQUE Step 1. Patient Positioning Step

More information

EXCELLA ll. Spinal System

EXCELLA ll. Spinal System EXCELLA ll Spinal System Excella II Spinal System INDICATIONS FOR USE The Innovasis Excella II Spinal System is intended for use in the non-cervical area of the spine. WARNING: The safety and effectiveness

More information

Stress analysis of the disc adjacent to interbody fusion in lumbar spine. Med Eng Phys

Stress analysis of the disc adjacent to interbody fusion in lumbar spine. Med Eng Phys See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/11774617 Stress analysis of the disc adjacent to interbody fusion in lumbar spine. Med Eng

More information

EXTERNAL FIXATION SYSTEM

EXTERNAL FIXATION SYSTEM EXTERNAL FIXATION SYSTEM The Apex Pin Fixation System is a one step procedure reducing insertion time and reducing insertion temperature. There are three types of fixation pins: Self-drilling / Self-tapping

More information

ACP. Anterior Cervical Plate System SURGICAL TECHNIQUE

ACP. Anterior Cervical Plate System SURGICAL TECHNIQUE ACP Anterior Cervical Plate System SURGICAL TECHNIQUE ACP TABLE OF CONTENTS INTRODUCTION 4 INDICATIONS AND CONTRAINDICATIONS 5 WARNINGS AND PRECAUTIONS 6 IMPLANT DESCRIPTION 7 INSTRUMENTS 10 SURGICAL

More information

The Influence of Muscle Forces on the Stress Distribution in the Lumbar Spine

The Influence of Muscle Forces on the Stress Distribution in the Lumbar Spine The Open Spine Journal, 2011, 3, 21-26 21 Open Access The Influence of Muscle Forces on the Stress Distribution in the Lumbar Spine Christian Wong *,1, John Rasmussen 2, Erik Simonsen 3, Lone Hansen 4,

More information

BIOMECHANICS OF THE SPINE AND IMPLANT FIXATION

BIOMECHANICS OF THE SPINE AND IMPLANT FIXATION BIOMECHANICS OF THE SPINE AND IMPLANT FIXATION ORTHOKINETIC TECHNOLOGIES & ORTHOKINETIC TESTING TECHNOLOGIES Lisa A. Ferrara, Ph.D. lisa@orthokintech.com Voice: 910.253.9883 Email: lisa@orthokintech.com

More information

Lumbar Disc Degeneration Is an Equally Important Risk Factor as Lumbar Fusion for Causing Adjacent Segment Disc Disease

Lumbar Disc Degeneration Is an Equally Important Risk Factor as Lumbar Fusion for Causing Adjacent Segment Disc Disease Lumbar Disc Degeneration Is an Equally Important Risk Factor as Lumbar Fusion for Causing Adjacent Segment Disc Disease Raghu N. Natarajan, Gunnar B.J. Andersson Department of Orthopedic Surgery, Rush

More information

Spinologics.com. Scoliosis Surgery Simulation

Spinologics.com. Scoliosis Surgery Simulation Scoliosis Surgery Simulation Finite Element (FE) Models Finite element model: mathematical method widely used in traditional engineering fields (aeronautics, car industry, civil engineering). Principle:

More information

Freedom. Lumbar Disc Polymer-Metal Bond Integrity. CAUTION: Investigational device. Limited by Federal law to investigational use.

Freedom. Lumbar Disc Polymer-Metal Bond Integrity. CAUTION: Investigational device. Limited by Federal law to investigational use. Freedom Lumbar Disc Polymer-Metal Bond Integrity CAUTION: Investigational device. Limited by Federal law to investigational use. 1 White Paper Freedom Lumbar Disc Polymer-Metal Bond Integrity Abstract

More information

Lumbar Facet Joint Replacement

Lumbar Facet Joint Replacement Rome Spine 2011 THE SPINE TODAY International Congress Rome 6-7th December 2011 Lumbar Facet Joint Replacement Prof. Dr. Karin Büttner-Janz Past President International Society for the Advancement of Spine

More information

X-spine Surgical Technique

X-spine Surgical Technique X-spine Surgical Technique The X90 Pedicle Screw System Revolutionary Design and Function This document is intended exclusively for experts in the field, particularly physicians, and is not intended for

More information

Experimental Study. Pain Physician 2015; 18:E1101-E1110 ISSN

Experimental Study. Pain Physician 2015; 18:E1101-E1110 ISSN Pain Physician 2015; 18:E1101-E1110 ISSN 2150-1149 Experimental Study Effect of Augmentation Material Stiffness on Adjacent Vertebrae after Osteoporotic Vertebroplasty Using Finite Element Analysis with

More information

TABLE OF CONTENTS. Vault C Anterior Cervical Discectomy 2 and Fusion (ACDF) System Overview. Implants 3. Instruments 5. Surgical Technique 10

TABLE OF CONTENTS. Vault C Anterior Cervical Discectomy 2 and Fusion (ACDF) System Overview. Implants 3. Instruments 5. Surgical Technique 10 Surgical Technique TABLE OF CONTENTS Vault C Anterior Cervical Discectomy 2 and Fusion (ACDF) System Overview Indications 2 Implants 3 Instruments 5 Surgical Technique 10 1. Preoperative planning 10 2.

More information

TSLP Thoracolumbar Spine Locking Plate

TSLP Thoracolumbar Spine Locking Plate Anterior thoracolumbar spine locking plate TSLP Thoracolumbar Spine Locking Plate Surgical Technique Image intensifier control This description alone does not provide sufficient background for direct use

More information

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics.

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. Chandrasenan.P 1, Vishnu.G 2, Akshay K Nair 3 1M Tech student, Department of Mechanical

More information

Solitaire Anterior Spinal System

Solitaire Anterior Spinal System Surgical Technique Solitaire Anterior Spinal System Independent Stabilization for the Anterior Column Available in Titanium and Contents Introduction... Page 1 Design Features... Page 2 Instruments...

More information

nvt Transforaminal Lumbar Interbody Fusion System

nvt Transforaminal Lumbar Interbody Fusion System nvt Transforaminal Lumbar Interbody Fusion System 1 IMPORTANT INFORMATION FOR PHYSICIANS, SURGEONS, AND/OR STAFF The nv a, nv p, and nv t are an intervertebral body fusion device used in the lumbar spine

More information

It consist of two components: the outer, laminar fibrous container (or annulus), and the inner, semifluid mass (the nucleus pulposus).

It consist of two components: the outer, laminar fibrous container (or annulus), and the inner, semifluid mass (the nucleus pulposus). Lumbar Spine The lumbar vertebrae are the last five vertebrae of the vertebral column. They are particularly large and heavy when compared with the vertebrae of the cervical or thoracicc spine. Their bodies

More information

Zimmer Anterior Buttress Plate System. Surgical Technique

Zimmer Anterior Buttress Plate System. Surgical Technique Zimmer Anterior Buttress Plate System Surgical Technique 2 Zimmer Anterior Buttress Plate System Surgical Technique Zimmer Anterior Buttress Plate System Surgical Technique Description, Indications & Contraindications...

More information

nva Anterior Lumbar Interbody Fusion System

nva Anterior Lumbar Interbody Fusion System nva Anterior Lumbar Interbody Fusion System 1 IMPORTANT INFORMATION FOR PHYSICIANS, SURGEONS, AND/OR STAFF The nv a, nv p, and nv t are an intervertebral body fusion device used in the lumbar spine following

More information

Module: #15 Lumbar Spine Fusion. Author(s): Jenni Buckley, PhD. Date Created: March 27 th, Last Updated:

Module: #15 Lumbar Spine Fusion. Author(s): Jenni Buckley, PhD. Date Created: March 27 th, Last Updated: Module: #15 Lumbar Spine Fusion Author(s): Jenni Buckley, PhD Date Created: March 27 th, 2011 Last Updated: Summary: Students will perform a single level lumbar spine fusion to treat lumbar spinal stenosis.

More information

Study on Mechanical Characteristics of Lumbar Spine for Snatch Action in Weight Lifting Based on Finite Element Method

Study on Mechanical Characteristics of Lumbar Spine for Snatch Action in Weight Lifting Based on Finite Element Method ISSN 1750-9823 (print) International Journal of Sports Science and Engineering Vol. 04 (2010) No. 01, pp. 048-052 Study on Mechanical Characteristics of Lumbar Spine for Snatch Action in Weight Lifting

More information

Biomechanical evaluation of three surgical scenarios of posterior lumbar interbody fusion by finite element analysis

Biomechanical evaluation of three surgical scenarios of posterior lumbar interbody fusion by finite element analysis Xiao et al. BioMedical Engineering OnLine 2012, 11:31 RESEARCH Open Access Biomechanical evaluation of three surgical scenarios of posterior lumbar interbody fusion by finite element analysis Zhitao Xiao

More information

Departement of Neurosurgery A.O.R.N A. Cardarelli- Naples.

Departement of Neurosurgery A.O.R.N A. Cardarelli- Naples. Percutaneous posterior pedicle screw fixation in the treatment of thoracic, lumbar and thoraco-lumbar junction (T12-L1) traumatic and pathological spine fractures. Report of 45 cases. G. Vitale, A. Punzo,

More information

EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE INFLUENCE OF EMBEDDING

EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE INFLUENCE OF EMBEDDING EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE INFLUENCE OF EMBEDDING THICKNESS IN COMPRESSIVE MECHANICAL TESTING OF VERTEBRAL AUGMENTATION AND SPINAL INTERBODY FUSION Márta Kurutz 1, Péter Nédli 1, Péter

More information

Anterior cervical discectomy and fusion is frequently employed

Anterior cervical discectomy and fusion is frequently employed ORIGINAL ARTICLE Biomechanical Evaluation of Relationship of Screw Pullout Strength, Insertional Torque, and Bone Mineral Density in the Cervical Spine Charles Alan Reitman, MD, Lyndon Nguyen, and Guy

More information

Adolescent Idiopathic Scoliosis

Adolescent Idiopathic Scoliosis Adolescent Idiopathic Scoliosis Surgical Treatment Comparisons By: Dr. Alex Rabinovich and Dr. Devin Peterson Options 1. Pedicle Screws versus Hooks 2. Posterior versus Anterior Instrumentation 3. Open

More information

The Most Reliable Spinal Solution

The Most Reliable Spinal Solution MFG : 09.02.20 REV : 1 4CIS VANE introduces you the reliable pedicle screw The Most Reliable Spinal Solution Non-slip threaded joint Ergonomic design Wedged trapezoid thread Superior locking mechanism

More information

Pedicle screw placement accuracy in thoracic and lumbar spinal surgery with a patient-matched targeting guide: A cadaveric study

Pedicle screw placement accuracy in thoracic and lumbar spinal surgery with a patient-matched targeting guide: A cadaveric study Pedicle screw placement accuracy in thoracic and lumbar spinal surgery with a patient-matched targeting guide: A cadaveric study [ based on the homonymous paper from Prof.Lamartina et al. Anticipated publication

More information

nvp Posterior Lumbar Interbody Fusion System

nvp Posterior Lumbar Interbody Fusion System nvp Posterior Lumbar Interbody Fusion System 1 IMPORTANT INFORMATION FOR PHYSICIANS, SURGEONS, AND/OR STAFF The nv a, nv p, and nv t are an intervertebral body fusion device used in the lumbar spine following

More information

Technique Guide. NFlex. Semi-rigid rods for posterior lumbar stabilization.

Technique Guide. NFlex. Semi-rigid rods for posterior lumbar stabilization. Technique Guide NFlex. Semi-rigid rods for posterior lumbar stabilization. Table of Contents Introduction NFlex 2 Indications and Contraindications 4 NFlex Principles 6 Surgical Technique Preoperative

More information

HawkeyeTM Peek. surgical technique

HawkeyeTM Peek. surgical technique HawkeyeTM Peek surgical technique Introduction The ChoiceSpine HAWKEYE Vertebral Body Replacement (VBR) System is intended for use in the thoracolumbar spine (T1 - L5) to replace a collapsed, damaged,

More information

Original Article Clinics in Orthopedic Surgery 2016;8:

Original Article Clinics in Orthopedic Surgery 2016;8: Original Article Clinics in Orthopedic Surgery 2016;8:71-77 http://dx.doi.org/10.4055/cios.2016.8.1.71 More than 5-Year Follow-up Results of Two- Level and Three-Level Posterior Fixations of Thoracolumbar

More information

Topology optimisation of hip prosthesis to reduce stress shielding

Topology optimisation of hip prosthesis to reduce stress shielding Computer Aided Optimum Design in Engineering IX 257 Topology optimisation of hip prosthesis to reduce stress shielding S. Shuib 1, M. I. Z. Ridzwan 1, A. Y. Hassan 1 & M. N. M. Ibrahim 2 1 School of Mechanical

More information

Dorsal Cervical Surgeries and Techniques

Dorsal Cervical Surgeries and Techniques Dorsal Cervical Approaches Dorsal Cervical Surgeries and Techniques Gregory R. Trost, MD Professor and Vice Chair of Neurological Surgery University of Wisconsin-Madison Advantages Straightforward Easily

More information

Screws versus hooks: implant cost and deformity correction in adolescent idiopathic scoliosis

Screws versus hooks: implant cost and deformity correction in adolescent idiopathic scoliosis J Child Orthop (2012) 6:137 143 DOI 10.1007/s11832-012-0400-8 ORIGINAL CLINICAL ARTICLE Screws versus hooks: implant cost and deformity correction in adolescent idiopathic scoliosis Bradley P. Jaquith

More information

OPERATIVE TECHNIQUE. anterior cervical plating system

OPERATIVE TECHNIQUE. anterior cervical plating system OPERATIVE TECHNIQUE 3º anterior cervical plating system Introduction 1 Pre-Operative Technique 2 Oerative Technique 3 Instructions for Use 12 Part Numbers 16 The surgical technique shown is for illustrative

More information

Raymond Wiegand, D.C. Spine Rehabilitation Institute of Missouri

Raymond Wiegand, D.C. Spine Rehabilitation Institute of Missouri 2D Pattern matching of frontal plane radiograph to 3D model identifies structural and functional deficiencies of the spinal pelvic system in consideration of mechanical spine pain (AKA Spine distortion

More information

Ref: Q400-09T1 EBI Spine. September 05/VS02. c/o BIOMET Spain Orthopaedics, S.L.

Ref: Q400-09T1 EBI Spine. September 05/VS02. c/o BIOMET Spain Orthopaedics, S.L. Ref: Q400-09T1 EBI Spine. September 05/VS02 c/o BIOMET Spain Orthopaedics, S.L. www.ebimedical.com EBI Omega 21 TM LP Since its introduction in 1996, and with thousands of patients treated so far, the

More information

O.I.C. PEEK UniLIF Unilateral Lumbar Interbody Fusion

O.I.C. PEEK UniLIF Unilateral Lumbar Interbody Fusion O.I.C. PEEK UniLIF Unilateral Lumbar Interbody Fusion A new implant for an evolving procedure Unique wedge nose Adaptive longer sizing Large internal graft volume Introduction Stryker Spine enters a new

More information

Original Article T1 finite element model of Kümmell s disease shows changes in the vertebral stress distribution

Original Article T1 finite element model of Kümmell s disease shows changes in the vertebral stress distribution Int J Clin Exp Med 2015;8(11):20046-20055 www.ijcem.com /ISSN:1940-5901/IJCEM0014291 Original Article T1 finite element model of Kümmell s disease shows changes in the vertebral stress distribution Yunshan

More information

Influence of Pedicle Screw Design on the Fatigue Fixation Strength in the Human Lumbar Spine

Influence of Pedicle Screw Design on the Fatigue Fixation Strength in the Human Lumbar Spine Influence of Pedicle Screw Design on the Fatigue Fixation Strength in the Human Lumbar Spine Rebecca Kueny, MSc. 1, Reza Danyali 1, Bartosz Nowak, PhD 2, Klaus Pueschel, Dr med. 3, Michael M. Morlock,

More information

Quality of Life. Quality of Motion.

Quality of Life. Quality of Motion. Quality of Life. Quality of Motion. Lateral Bend Vertical Translation Flexion Extension Lateral Translation Axial Rotation Anterior Posterior Translation Motion in all Directions Kinematics is the study

More information

2008 International ANSYS Conference

2008 International ANSYS Conference 2008 International ANSYS Conference Simulation of Mandible Biomechanics after Fixation of Subcondylar Fractures Dana Coombs Analytical Engineer Synthes USA 2008 ANSYS, Inc. All rights reserved. 1 ANSYS,

More information

Comparative Study of Fixation Devices for Intertrochanteric Fractures

Comparative Study of Fixation Devices for Intertrochanteric Fractures Comparative Study of Fixation Devices for Intertrochanteric Fractures C. Sticlaru * A. Davidescu Politehnica University of Timişoara Politehnica University of Timişoara Timişoara, România Timişoara, România

More information

1 Department of Neurosurgery, Ümraniye Training and Research Hospital,

1 Department of Neurosurgery, Ümraniye Training and Research Hospital, Original Article Comparison of rigid and semi-rigid instrumentation under acute load on vertebrae treated with posterior lumbar interbody fusion/ transforaminal lumbar interbody fusion procedures: An experimental

More information

Ultimate Spinal Analysis PA USA-XRAY ( )

Ultimate Spinal Analysis PA USA-XRAY ( ) Page: 1 Spine Atlas Angle 7.24 S Atlas Angle 21.67 S The Atlas Angle is a measurement of the stability of the Atlas. The Atlas Plane Line is compared to true horizontal. Any increase or decrease of this

More information

Am I eligible for the TOPS study? Possibly, if you suffer from one or more of the following conditions:

Am I eligible for the TOPS study? Possibly, if you suffer from one or more of the following conditions: Am I eligible for the TOPS study? Possibly, if you suffer from one or more of the following conditions: Radiating leg pain Greater leg / buttock pain than back pain Severe pain sets in when walking as

More information

C-THRU Anterior Spinal System

C-THRU Anterior Spinal System C-THRU Anterior Spinal System Surgical Technique Manufactured From Contents Introduction... Page 1 Design Features... Page 2 Instruments... Page 3 Surgical Technique... Page 4 Product Information... Page

More information

Kinematic Analysis of Lumbar Spine Undergoing Extension and Dynamic Neural Foramina Cross Section Measurement

Kinematic Analysis of Lumbar Spine Undergoing Extension and Dynamic Neural Foramina Cross Section Measurement Copyright c 2008 ICCES ICCES, vol.7, no.2, pp.57-62 Kinematic Analysis of Lumbar Spine Undergoing Extension and Dynamic Neural Foramina Cross Section Measurement Yongjie Zhang 1, Boyle C. Cheng 2, Changho

More information

Micro-CT imaging of surgical screw tightening in human trabecular bone

Micro-CT imaging of surgical screw tightening in human trabecular bone Micro-CT imaging of surgical screw tightening in human trabecular bone E. Perilli 1, M. Ryan 1, R. Ab-Lazid 1, J.J. Costi 1, K.J. Reynolds 1 1 Medical Device Research Institute, School of Computer Science,

More information

An effort is made to analyse the stresses experienced by the human femur. In order

An effort is made to analyse the stresses experienced by the human femur. In order Finite Element Analysis of Human Femur Bone U N Mughal 1, H A Khawaja 2*, M Moatamedi 1, M Souli 3 1. Narvik University College, Norway 2. UiT-The Arctic University of Norway, Norway 3. University of Lille,

More information

Effect of range of motion (ROM) for pedicle-screw fixation on lumbar spine with rigid and semi-rigid rod materials: A finite element study

Effect of range of motion (ROM) for pedicle-screw fixation on lumbar spine with rigid and semi-rigid rod materials: A finite element study IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Effect of range of motion (ROM) for pedicle-screw fixation on lumbar spine with rigid and semi-rigid rod materials: A finite element

More information

Hiroto Yamaguchi*, Hidetoshi Nojiri**, Kei Miyagawa*, Nozomu Inoue***

Hiroto Yamaguchi*, Hidetoshi Nojiri**, Kei Miyagawa*, Nozomu Inoue*** Three-Dimensional High-Resolution Image Analysis of the Segmental Reduction Effect Obtained with Lateral Lumbar Interbody Fusion for Lumbar Intervertebral Deformity Hiroto Yamaguchi*, Hidetoshi Nojiri**,

More information

Introduction. Our hope is that this Surgical Technique Guide enhances your knowledge and contributes to clinical success for your patients.

Introduction. Our hope is that this Surgical Technique Guide enhances your knowledge and contributes to clinical success for your patients. Surgical Technique Introduction DePuy Spine continues to support the goal of expanding the spine surgeon s options for the treatment of spinal disorders. Collaborating with renowned spine specialists,

More information

Spine and Fusion. Adjacent Segment Disease. 36Ihsan SOLAROGLU M.D., M. Ozerk OKUTAN M.D., Gurdal NUSRAN M.D.

Spine and Fusion. Adjacent Segment Disease. 36Ihsan SOLAROGLU M.D., M. Ozerk OKUTAN M.D., Gurdal NUSRAN M.D. Lumbar Posterior Hybrid Dynamic Stabilisation and Fusion Systems 36Ihsan SOLAROGLU M.D., M. Ozerk OKUTAN M.D., Gurdal NUSRAN M.D. Spine and Fusion It has been more than a century since 1911 when Albee

More information

Y o u r Id e a s En g i n e e r e d t o Li f e

Y o u r Id e a s En g i n e e r e d t o Li f e ISSYS LP Spinal Fixation System Surgical Guide Y o u r Id e a s En g i n e e r e d t o Li f e In t r o d u c t i o n ISSYS LP Sp i n a l Fixation System The foundation of the ISSYS LP Spinal Fixation System

More information

Finite Element Analysis of Radius and Ulna. Eli Pavlatos April 24, 2013

Finite Element Analysis of Radius and Ulna. Eli Pavlatos April 24, 2013 Finite Element Analysis of Radius and Ulna Eli Pavlatos April 24, 2013 Outline Review bone biomechanics bone behavior during fracture Stress analysis concepts properties of bone for models Fracture simulations

More information

Patient Information MIS LLIF. Lateral Lumbar Interbody Fusion Using Minimally Invasive Surgical Techniques

Patient Information MIS LLIF. Lateral Lumbar Interbody Fusion Using Minimally Invasive Surgical Techniques Patient Information MIS LLIF Lateral Lumbar Interbody Fusion Using Minimally Invasive Surgical Techniques Table of Contents Anatomy of Spine...2 General Conditions of the Spine....4 What is Spondylolisthesis....5

More information

SURGICAL TECHNIQUE. Easyspine PEDICLE SCREW SYSTEM

SURGICAL TECHNIQUE. Easyspine PEDICLE SCREW SYSTEM SURGICAL TECHNIQUE Easyspine PEDICLE SCREW SYSTEM Easyspine PEDICLE SCREW SYSTEM Designed by leading spine surgeons, Easyspine features a simplified surgical technique and adaptable implants to accommodate

More information

Biomechanics of Fractures and Fixation

Biomechanics of Fractures and Fixation Biomechanics of Fractures and Fixation Theodore Toan Le, MD Original Author: Gary E. Benedetti, MD; March 2004 New Author: Theodore Toan Le, MD; Revised October 09 Basic Biomechanics Material Properties

More information

3D titanium interbody fusion cages sharx. White Paper

3D titanium interbody fusion cages sharx. White Paper 3D titanium interbody fusion cages sharx (SLM selective laser melted) Goal of the study: Does the sharx intervertebral cage due to innovative material, new design, and lordotic shape solve some problems

More information

SURGICAL TECHNIQUE. SECURIS Pedicle Screw System for Minimally Invasive Surgery. 2 I SECURIS Pedicle Screw System

SURGICAL TECHNIQUE. SECURIS Pedicle Screw System for Minimally Invasive Surgery. 2 I SECURIS Pedicle Screw System Surgical Technique e Guide SECURIS Pedicle Screw System for Minimally Invasive Surgery Securis Pedicle Screw System has been engineered to provide temporary posterior stabilization of the thoracolumbar

More information

pissn: , eissn: Yonsei Med J 55(5): , 2014

pissn: , eissn: Yonsei Med J 55(5): , 2014 Original Article http://dx.doi.org/10.3349/ymj.2014.55.5.1386 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 55(5):1386-1394, 2014 Biomechanical Analysis of Fusion Segment Rigidity Upon Stress at Both

More information

Table of Contents.

Table of Contents. surgical technique The Ambassador TM Anterior Cervical Plate System is a versatile system of implants and instruments with a variety of sizes to provide optimal anatomic compatibility. The integrated cam

More information

VLIFT System Overview. Vertebral Body Replacement System

VLIFT System Overview. Vertebral Body Replacement System VLIFT System Overview Vertebral Body Replacement System VLIFT System System Description The VLIFT Vertebral Body Replacement System consists of a Distractible In Situ (DIS) implant, which enables the surgeon

More information

The Nucleus Pulpous of Intervertebral Disc Effect on Finite Element Modeling of Spine

The Nucleus Pulpous of Intervertebral Disc Effect on Finite Element Modeling of Spine Original Article The Nucleus Pulpous of Intervertebral Disc Effect on Finite Element Modeling of Spine Midiya Khademi 1, Yousef Mohammadi 2, Seifollah Gholampour 3, Nasser Fatouraee 4 1 M.Sc. Department

More information

vertaplan the spine surgeon s software vertaplan System for successful reconstruction of the individual sagittal balance

vertaplan the spine surgeon s software vertaplan System for successful reconstruction of the individual sagittal balance the spine surgeon s software System for successful reconstruction of the individual sagittal balance What do you think of patient-specific reconstruction of the spine geometry? Optimum surgical outcome

More information