1 Original Paper Caries Res 372 DOI: /0000XXXXX Evaluation of Selective Caries Removal by a Fluorescence Feedback-Controlled Er:YAG Laser in vitro J. Eberhard a A.K. Eisenbeiss a A. Braun c J. Hedderich b S. Jepsen a Department of Operative Dentistry and Periodontology and b Department of Medical Statistics, University Hospital Schleswig-Holstein, Kiel and c Department of Periodontology, Operative and Preventive Dentistry, University Hospital Bonn, Bonn, Germany c Received: September 29, 2004 Accepted after revision: February 17, 2005 Key Words Dental caries Fluorescence Laser Treatment Abstract Aim: To establish a fluorescence threshold level that could guide a therapeutic Er:YAG laser through a caries lesion to determine a therapeutic endpoint of caries removal. Materials and Methods: A total of 65 extracted human teeth, 35 with dentine caries and 30 healthy, were used for this study. An Er:YAG laser system that emitted at a wavelength of 2.94 m was used. The laser was equipped with a laser fluorescence feedback system, excitation wavelength 655 nm, to control the irradiation by the Er:YAG laser. The evaluated threshold levels of the fluorescence feedback system were 3, 7, 8, 10, 12, 15 and 20. After treatment the teeth were prepared for histological staining according to the method of Brown and Brenn for the identification of bacteria. The specimens were subjected to a quantitative evaluation of residual bacteria on the treated dentine surface. In addition, the internal fluorescence of dentine and potential fluorescence changes of dentine after laser irradiation were evaluated. Results: About 80% of the irradiated dentine surface showed residual bacteria with threshold levels of 20, 15, 12, and 10. Residual bacteria were not found with threshold levels of 7 and 3. The study revealed a significant increase in dentine fluorescence after laser irradiation. Conclusion: The results of the present in vitro study indicate that a fluorescence threshold level of 7 or 8 units can guide an Er:YAG laser to a complete removal of carious dentine. Copyright 2005 S. Karger AG, Basel By the time a caries process can be detected the lesion is no longer incipient [Gwinnett, 1971] and is generally destined to be restored because of the irreversible nature of the lesion [Angmar-Månsson and ten Bosch, 1987]. The most common procedures for removing infected dental hard tissues are by excavators and burs [Banerjee et al., 2000]. When removing demineralized dentine it is not always easy to know at what point excavation is completed because there is an apparent lack of objective clinical markers for the differentiation between infected and healthy dentine. The clinician tends to rely on the consistency of the tissue, but no rigorous correlation between the consistency of the dentine and the area of infection could be established and changes in hardness and colour are continuous and not abrupt. Therefore, the operative treatment of carious lesions by rotary instruments depends to a significant degree upon the clinical skills of the operator and has often resulted in considerable removal of healthy tooth structure [Banerjee et al., 2000]. Conse- Fax karger@karger.ch S. Karger AG, Basel /05/ $22.00/0 Accessible online at: Jörg Eberhard, DDS, PhD Department of Operative Dentistry and Periodontology University Hospital Schleswig-Holstein, Arnold-Heller-Stasse 16 DE Kiel (Germany) Tel. +49 (431) , Fax +49 (431) , eberhard@konspar.uni-kiel.de CRE372.indd :46:23

2 quently, a clinical method for the distinction between the outer infected dentine and the inner non-infected dentine is required for the rational treatment of dental caries with minimal destruction of healthy tissues. Several approaches for the selective removal of infected dentine have been introduced during the last decades including the use of caries detector dyes and the chemomechanical caries removal technique. The staining of different carious layers was introduced in 1975 by Ogushi and Fusayama [1975]. However the susceptibility to dye staining by acid red appears to be a feature of reduced mineral content rather than being specific for a carious lesion. Sound circumpulpal dentine and the dentinoenamel junction could also be stained, which may result in the unnecessary removal of sound dentine [Kidd et al., 1993; Yip et al., 1994]. The chemomechanical caries removal approach is basically a gel-assisted hand excavation that is also aimed to determine the endpoint of caries excavation. Chemomechanical caries removal may be associated with a risk of leaving caries in the dentinoenamel junction [Cederlund et al., 1999] and is much more timeconsuming than conventional excavation [Tavares et al., 1988], even though other study groups reported successful caries removal with this system [Ericson et al., 1999]. There has been growing interest in using lasers as diagnostic devices that capitalize on measurements that discriminate between benign and malignant cells [Bohorfoush, 2000]. These methods are based on laser-induced fluorescence that is used for a variety of applications like electrophoresis [Paez and Hernandez, 2001] or photodynamic therapy [Sternberg and Dolphin, 1993]. Fluorescence is a phenomenon by which the wavelength of the emitted light is changed into a longer wavelength upon absorption and subsequent re-emission of light. The fluorescence of dental hard tissues has been known for a very long time and more recently laser light was used to induce fluorescence of enamel in a sensitive, non-destructive diagnostic method for the detection of enamel demineralization and dental caries [Bjelkhagen et al., 1982; Sundström et al., 1985; Lussi et al., 2004]. Laser fluorescence was introduced to aid detection of occlusal caries as an adjunct to visual inspection and radiographic examination [Hibst and Gall, 1998]. Red laser diode light is directed to the tooth surface by a specially designed probe tip, and the fluorescent signal is filtered from the incident light and fed back to the detector through the same device. The excitation wavelength of 655 nm (red light) induces a fluorescence signal that has been assigned to a protoporphyrin, which is present as a bacterial breakdown product [König et al., 1998]. The use of laser-induced fluorescence to control a device that is capable of removing dental hard tissues would be a very promising approach for the selective removal of carious dentine. It has been technically realized by the combination of an Er:YAG laser and a modification of laser fluorescence in one dental instrument. Within this system the removal of dental hard tissues by the Er:YAG laser is controlled by the fluorescence signal from the tooth surface induced by the red-infrared diagnostic laser. However, at present there are no data available on the feasibility of this system for selective caries removal. Therefore, it was the aim of this in vitro study to establish a fluorescence threshold level that could guide the therapeutic laser through a caries lesion until no bacteria were detectable by histological staining and hence a therapeutic endpoint of caries removal was reached. Moreover, for the safe use of this system additional information about the internal fluorescence of dentine and on changes of the dentine fluorescence induced by Er:YAG laser irradiation were gathered. Materials and Methods Specimens Several private dental offices provided a total of 65 extracted human teeth, 35 with dentine caries and 30 healthy, that were used for this study. In each office, the teeth were extracted for periodontal, orthodontic or prosthodontic reasons and were obtained from patients who consented to their use for research. The teeth were stored immediately after extraction in buffered saline at 8 C until further processing. Laser Device An Er:YAG laser system (Key III Laser: Kavo, Biberach, Germany) that emitted at a wavelength of 2.94 m with a spot size of 0.63 mm was used. The output settings were 250 mj/pulse and the pulse repetition rate was 4 pulses/s. These parameters were chosen on the basis of preliminary experiments. Irradiation of a focused beam was performed for the present experiments with the non-contact handpiece 2060 (Kavo, Biberach, Germany). The irradiated area was continuously cooled by a water spray system (1 ml/min). The laser was equipped with a laser fluorescence feedback system ( fig. 1 ). The emitted light with a wavelength of 655 nm (red light) is transported through a fibre bundle to the tip of the handpiece and the same tip, but different fibres, samples the fluorescent light. The laser-induced fluorescence of the dentine is measured and used to control the therapeutic irradiation by turning on the Er: YAG laser if the fluorescence value is above a preselected threshold level. If the dentine fluorescence is below this value the Er:YAG laser does not emit. For the present study the evaluated threshold levels of the fluorescence feedback system were 3, 7, 8, 10, 12, 15 and Caries Res 372 Eberhard /Eisenbeiss/Braun/Hedderich/ Jepsen CRE372.indd :47:32

3 Fig. 1. Schematic illustration showing the construction and light guidance of the fluorescence-assisted Er:YAG laser. (1) The diagnostic laser with a wavelength of 655 nm is transported through a fibre bundle to the tip of the handpiece. (2) The tooth substance emits a fluorescence signal that is collected by separate fibres and transported to an internal analysis unit, which calculates the intensity of the fluorescence light in a range from 0 to 100 fluorescence values. A fluorescence signal smaller than a pre-selected threshold value activates the diagnostic laser. A fluorescence signal greater than a pre-selected threshold value activates the therapeutic Er:YAG laser. (3) The Er:YAG laser light is transported through the handpiece and ablates the tooth substance Diagnostic laser (655 nm) Red fluorescence > Threshold level Er:YAG laser 2.94 µm Threshold level Fig. 2. Brown and Brenn staining shows the bacterial deposits within the untreated carious dentine ( A ) and after Er:YAG laser treatment with threshold levels of 15 ( B ), 12 ( C ) and 7 ( D ). The insert ( E ) shows the residual dentine after conventional excavation with a bur. A Original magnification! 40. B E Original magnification! 20. Fluorescence Threshold Level For all experimental procedures the enamel was removed by high-speed burs and water cooling until access to the carious lesion was achieved. On the surface of the carious lesion small incisions were cut with a sharp scalpel blade to divide the lesion into three areas that were treated as follows: (A) no treatment, (B) caries removal with a bur at a speed of 700 rpm without water cooling and (C) laser irradiation. The specimens were assigned randomly to the different groups. Before the experiments the operator received a full training in caries removal using the laser. After calibration of the fluorescence feedback system following the instructions of the manufacturer the handpiece was manually adjusted perpendicular to the treated surface. The handpiece was used in the non-contact irradiation mode with a distance from the tip to the cavity floor of approximately mm. The pilot laser beam was used to adjust the correct distance and as caries removal progressed and the distance of the treated cavity floor to the tip of the handpiece changed, irradiation Caries Treatment by Fluorescence-Guided Er:YAG Laser Caries Res CRE372.indd :47:32

4 was paused and the distance was readjusted. The laser treatment was terminated if the fluorescence feedback system did not indicate any reflective fluorescence from the dentine above the pre-selected threshold level. No caries disclosing agent or dental probe was used to explore the laser-treated dentine surface after laser irradiation. For every pre-selected threshold level 5 specimens were treated. Conventional bur treatment was performed by an experienced dentist in a dry field by means of steel burs of the appropriate size mounted on a micromotor at low speed. The consistency of the tissue was examined with a dental explorer to confirm the completeness of caries removal. Sample Preparation After treatment the teeth were cut through the incisions along the tooth axis in three specimens with a cutting saw (WoCo Apparatebau, Clausthal-Zellerfeld, Germany). The specimens were fixed in phosphate-buffered formalin for 24 h and then washed in Sörensen phosphate-buffered saline for another 24 h. After dehydration by alcohol the specimens were embedded in methacrylate (Sigma-Aldrich, Dreieich, Germany). The polymerized specimens were placed onto microscopic slides and ground to approximately 30- to 60- m-thin slices. Two slices were fabricated from each specimen (n = 240). The slices were stained by the method of Brown and Brenn [1931] for the identification of gram-positive (blue) and gram-negative (red) bacteria. Quantitative Evaluation The samples were investigated under a microscope (Axiophot 2, Carl Zeiss, Jena, Germany) at a total magnification of! (Lens Plan-Neofluar! 63 and ocular E-P1! 10; Zeiss). A test grid was mounted on the focal plane of the eyepiece of the microscope and consisted of a square frame of a definitive size enclosing an area of 4,681 m 2. Five test grids were randomly located on the surface of treated dentine and the number of test squares positive for bacteria within the dentinal tubules was expressed as percentage of the total number of squares covering the treated dentine surface. For each threshold level 5 teeth were used and two individual specimens were investigated quantitatively. Internal Fluorescence of the Dentine Twenty non-carious third molars that had been removed during orthodontic treatments were selected for this experiment. The teeth were divided into halves along the long axis by a cutting saw with water cooling. In 10 teeth, beginning from the occlusal surface to the roof of the pulp chamber the fluorescence of the tooth hard substance was measured with the laser handpiece 2060 (Key III Laser, Kavo) and with the DIAGNOdent device (Kavo). After each measurement about 0.5 mm of dentine was removed with a low-speed bur until the pulp chamber became exposed. A micrometer measuring system under a stereomicroscope at! 16 magnification was used to measure the distance between the prepared tooth surface and the pulp chamber. In a second group (n = 10) this procedure was repeated to measure the internal fluorescence of the dentine in the opposite direction from the pulp chamber to the enamel-dentine border. Fluorescence of Irradiated Dentine Ten non-carious third molars were selected and cut along the tooth long axis as described to determine the fluorescence changes in dentine induced by therapeutic laser irradiation. The dentine was irradiated with the therapeutic Er:YAG laser without the fluorescence-feedback system under a stereomicroscope at! 16 magnification until approximately 0.5 mm of dentine had been removed. The fluorescence of the resulting dentine surface was recorded by both the therapeutic laser system (Key III laser handpiece 2060) and by the diagnostic device (DIAGNOdent) and subsequently 0.5 mm of dentine was removed with a rotary bur. Again, the fluorescence values were recorded and the dentine was irradiated by the Er:YAG laser until another 0.5 mm of dentine was removed. This procedure was continued until the pulp chamber became exposed. Statistical Analysis For each experimental group the fluorescence values were described by means 8 standard deviations (SD). Obvious extreme values or outliers were not observed. Nevertheless, normal distribution was not assumed in the data for statistical testing, because for 10 teeth in each experimental group any test for normal distribution has only small power. Differences between reading points at different steps of the experimental procedure were evaluated with Friedman analysis of variances by ranks and multiple comparisons. Differences between the fluorescence readings of the DIAGNOdent and the laser and between different treatment regimes were compared at each step by Wilcoxon matched-pairs signed-rank test. All calculations were performed with the statistical program SPSS 11. Effects were regarded as statistically significant for p! Results Fluorescent Threshold Level The dentine surface of an untreated carious lesion showed that 100% of the dentinal tubules located in the carious lesion contained bacteria that completely filled the dentine tubules ( fig. 2 A). Following treatment of the carious lesions with a rotary bur and confirmation of caries removal with a dental explorer no bacteria were found in the residual dentine of the cavity floor ( fig. 2 E). The results for the histological evaluation following caries removal by Er:YAG laser are presented in figure 2 A D. Employing the Er:YAG laser with fluorescence threshold levels of 20, 15, 12 and 10, respectively, revealed only a minor reduction of residual bacteria within the dentinal tubules from 94.0 to 88.0% ( table 1 ). In contrast, the fluorescence threshold level of 8 resulted in a significant decrease in residual bacteria within dentinal tubules of the cavity floor to a mean percentage of 6.7%. Laser ablation of carious dentine with the guidance of the fluorescence threshold levels of 7 and 3 resulted in no residual bacteria within the dentine tubules located at the cavity floor ( table 1 ). 4 Caries Res 372 Eberhard /Eisenbeiss/Braun/Hedderich/ Jepsen CRE372.indd :47:39

5 Fluorescence Fluorescence Distance to pulp (mm) Distance to dentinoenamel junction (mm) 3.5 Fig. 3. Variation in the fluorescence readings of the laser handpiece 2060 (circles) and for the DIAGNOdent device (squares) in the direction towards the roof of the pulp chamber (means 8 SD). The fluorescence readings of the DIAGNOdent increase towards the pulp, but without any statistically significance. The experiment was reproduced 10 times on different samples. Fig. 4. Variation in the fluorescence readings of the laser handpiece 2060 (circles) and for the DIAGNOdent device (squares) from the roof of the pulp chamber to the dentinoenamel junction (mean 8 SD). The experiment was reproduced 10 times on different tooth samples. Table 1. Means and SD of the residual amounts of bacteria (%) estimated by a quantitative approach on the laser-treated dentine surface with different threshold levels of the fluorescence feedback system n Residual bacteria, % of tubules mean SD Control Untreated caries Bur treatment Laser threshold value For control purposes untreated carious lesions and carious lesions treated by a bur were used. For each threshold level 5 teeth were investigated and the quantitative evaluation was performed on two separate specimens. PROOF Fluorescence of Dentine The measurements of the dentine fluorescence by the laser handpiece and the DIAGNOdent device from the enamel-dentine surface to the roof of the pulp chamber are illustrated in figure 3. The mean fluorescence values of healthy dentine measured by the DIAGNOdent device were statistically significantly higher than those measured with the laser handpiece for every reading point (p = 0.005). The statistical analysis demonstrated no significant differences between the mean fluorescence values for different measuring points from the dentinoenamel junction to the roof of the pulp chamber for the DIAGNOdent device or for the laser handpiece, even though the values had a tendency to increase 2.0 mm away from the roof of the pulp chamber. In consequence, a mean fluorescence recording for dentine of (mean 8 SD) was measured with the DIAGNOdent device and of with the laser handpiece. In a similar fashion fluorescence values of healthy dentine from the roof of the pulp chamber to the dentinoenamel junction are presented in figure 4. Again, the mean fluorescence values measured with the DIAGNO- Caries Treatment by Fluorescence-Guided Er:YAG Laser Caries Res CRE372.indd :47:39

6 Fluorescence 20 Fluorescence Distance to pulp (mm) Distance to pulp (mm) Fig. 5. Variation in the fluorescence readings of the DIAGNOdent device before (squares) and after (filled squares) laser irradiation at different intervals of the distance towards the roof of the pulp chamber (mean 8 SD). The experiment was reproduced 10 times on different samples. Fig. 6. Variation in the fluorescence readings of the dentine surface after bur treatment (circles) and after laser irradiation (filled circles) at different intervals of the distance towards the roof of the pulp chamber (mean 8 SD). For the experiment the readings were performed on the same samples used for the DIAGNOdent readings. dent device were statistically significantly higher than those obtained with the laser handpiece for every reading point (p = 0.005). There were no statistically significant differences between the mean fluorescence values for individual measuring points from the roof of the pulp chamber to the dentinoenamel junction for both fluorescence devices. Thus, by scanning the dentine from the roof of the pulp chamber to the dentinoenamel junction opposite to the direction of the first series of measurements a mean fluorescence recording for dentine of was measured with the DIAGNOdent device and of with the laser handpiece. Fluorescence of Laser-Irradiated Dentine The Er:YAG laser irradiation significantly increased the fluorescence signal of sound dentine measured by DI- AGNOdent ( fig. 5 ) or the laser handpiece ( fig. 6 ) compared to the dentine treated with a bur, irrespective of the reading point within the dentine (p = 0.005). The mean fluorescence of dentine increased from to after laser irradiation measured by the DIAGNOdent device, and from to measured with the laser handpiece. PROOF Discussion Present diagnostic methods used to indicate the endpoint of caries removal in a clinical situation are quite subjective and may thus lead to either under- or overinstrumentation. Ideally, a diagnostic tool should indicate the presence or absence of bacteria because of the infectious nature of the disease. A defined endpoint of caries removal would therefore coincide with the complete removal of all bacteria-infected dentine. The results of the present in vitro study could demonstrate the capacity of selective caries removal for an Er:YAG laser that is controlled by the fluorescence signal from bacterial breakdown products induced by a red-infrared diagnostic laser. Hibst and Keller [1989] first demonstrated the effective ablation of dental hard tissues by means of the Er: YAG laser. The application of the Er:YAG laser has been used for caries treatment in vitro [Aoki et al., 1998; Armengol et al., 1999] and in clinical situations as well [Keller and Hibst, 1997; Keller et al., 1998]. No differences in the effectiveness of dentine removal by an Er: YAG laser compared to a conventional drill have been 6 Caries Res 372 Eberhard /Eisenbeiss/Braun/Hedderich/ Jepsen CRE372.indd :47:40

7 reported [Hibst and Keller, 1989; Keller and Hibst, 1989; Shigetani et al., 2002]. In general, it is accepted that under adequate water spray and with a careful irradiation technique, cavities without any sign of thermal damage to the surrounding tissues as well as to the dental pulp could be produced with the Er:YAG laser [Hossain et al., 1999; Yamada et al., 2001]. Animal histological studies also showed that the pulp response to the Er:YAG laser appears to be similar to the response from high-speed handpiece application [Sonntag et al., 1996]. In an in situ study no pathological changes were observed 7 days after Er: YAG laser treatment by light transmission electron microscopy and 3 months postoperatively the apposition of tertiary dentine has been demonstrated [Nair et al., 2003]. The experiments aimed at determining a threshold level for the fluorescence-aided caries removal by an Er: YAG laser demonstrated almost no caries removal with threshold levels between 10 and 20, a marked reduction of residual bacteria below 10% with a threshold level of 8, and no detectable residual bacteria for threshold levels of 7 and smaller. It is noticeable that no linear correlation between the bacterial reduction and the pre-selected threshold level could be observed, but rather an abrupt decrease in a narrow range between 7 and 9 units. Thus, the fluorescence feedback system appears to operate by a true yes or no decision with respect to the detection of even small numbers of bacteria within the dentine. In contrast to other methods like explorers aimed to determine the endpoint of caries removal the induced fluorescence of 655 nm is likely related to bacterial components as proposed by König et al. [1998]. The relation between microoganisms or bacterial products and the fluorescence signal has been confirmed in the present study and in a recently published study by Iwami et al. [2003]. In contrast to other fluorescence-aided caries excavation techniques like the FACE method [Lennon et al., 2002; Lennon, 2003], that combines caries detection by fluorescence with caries removal by a rotary instrument, the fluorescence directly regulates the activity of the device used for dentine treatment, which might be an advantage during clinical use. For the safe use of this system, additional information about the internal fluorescence of dentine and on changes of the dentine fluorescence induced by Er:YAG laser irradiation was required. With respect to the determination of a threshold level, substantial variations of the fluorescence in different regions of sound dentin would rule out the possibility of determining a single threshold level or even the use of this system for clinical caries removal. In fact, previous studies showed that the transmission and scattering properties of dentine differ according to the direction of the tubules and this affected the fluorescence measurements [Vaarkamp et al., 1995; Iwami et al., 2003]. Therefore, in the first part of the present study we evaluated the fluorescence characteristics of sound dentine with the laser handpiece and with a laser fluorescence device to investigate the fluorescence readings from the dentinoenamel junction to the roof of the pulp. It could be demonstrated that the fluorescence readings showed only minor variations at different locations of the dentine that were not statistically significantly different. Increased fluorescence values of the dentine approaching the roof of the pulp chamber were also reported by Lussi et al. [2000]. Irrespective of the direction of the readings, towards the pulp or away from the roof of the pulp chamber, no differences in fluorescence signals were observed and it was therefore possible to determine mean fluorescence values for non-carious dentine. Laser fluorescence has been proven as a useful tool for the detection of early lesions on the occlusal surface [Lussi et al., 1999, 2001; Francescut and Lussi, 2003], although the agreement between validated caries and the fluorescence signal is still unsatisfactory when using the device to detect carious dentine under enamel [Heinrich- Weltzien et al., 2003]. However, the carious lesion is not covered by enamel in the current approach and this might improve detection. When comparing the fluorescence readings by the laser handpiece and by laser fluorescence, the latter measurements were about 4 6 times greater, but the two sets closely correlated with each other. This observation was most likely due to different light transmission systems in the laser handpiece and in the DIAG- NOdent device. The light transmission window of the laser handpiece has a larger diameter than that of the cone-shaped tip of the DIAGNOdent and the former picked up light from outside the target surface as well. Therefore, the manufacturer adjusted the sensitivity of the laser handpiece. Another safety issue that had to be addressed before any clinical use was the possible impact on the dentine fluorescence by the Er:YAG laser irradiation itself. Instead of perpendicular measurements of sectioned carious teeth we selected an experimental protocol that closely resembled the clinical direction of caries removal towards the pulp chamber. Interestingly, the results of the present study could demonstrate that the ablation of dentine by the Er:YAG laser significantly increased the fluorescence values of the resulting dentine surface as measured by the laser and the laser fluorescence device. Again, Caries Treatment by Fluorescence-Guided Er:YAG Laser Caries Res CRE372.indd :47:40

8 the laser fluorescence readings were greater by a factor of 5 6 than the laser readings. The observed increase of the fluorescence induced by the laser irradiation was by a factor of 2 3. It could be assumed that laser treatment of the dentine was followed by a dehydration of the dentine that could induce the observed increase in fluorescence. In contrast to the present findings in vitro studies demonstrated that dehydration of enamel was accompanied by decreased fluorescence values [Al-Khateeb et al., 2002; Pretty et al., 2004]. Whether structural changes of the dentine after laser ablation, dehydration or any other thermal effects were the origin of the changes in fluorescence values could not be evaluated in the present study; this question should be addressed by future research. The increase in fluorescence values by the Er:YAG laser itself is of potential concern, because sound dentine could emit fluorescence signals exceeding the established threshold level leading to its removal. However, from the present data it could be calculated that 68 and 98% of fluorescence values read by the laser handpiece from irradiated sound dentine were in the range of and 0 7.1, respectively. From these data it could be concluded that a threshold level of 7 or greater will not have any impact on changes in dentinal fluorescence induced by the laser irradiation itself for most of the cases under the present in vitro situation. In conclusion, the results of the present in vitro study suggest that a fluorescence threshold level of 7 or 8 units could guide an Er:YAG laser to a complete removal of carious dentine. Because ablation by this laser is controlled by the induced fluorescence of bacteria or bacterial products this method may be superior to current methods for caries removal where the endpoint of the treatment is determined by subjective means. Thus, on the basis of these promising results future studies are warranted to validate the present findings under in vivo conditions, before fluorescence-controlled selective caries removal by Er:YAG laser can be recommended for clinical use. References Al-Khateeb S, Exterkate RA, de Josselin de Jong E, Angmar-Mansson B, ten Cate JM: Light-induced fluorescence studies on dehydration of incipient enamel lesions. Caries Res 2002; 36: Angmar-Månsson B, ten Bosch JJ: Optical methods for the detection and quantification of caries. Adv Dent Res 1987; 1: Aoki A, Ishikawa I, Yamada T, Otsuki M, Watanabe H, Tagami J, Ando Y, Yamamoto H: Comparison between Er:YAG laser and conventional technique for root caries treatment in vitro. J Dent Res 1998; 77: Armengol V, Jean A, Rohanizadeh R, Hamel H: Scanning electron microscopic analysis of diseased and healthy dental hard tissues after Er: YAG laser irradiation: In vitro study. J Endod 1999; 25: Banerjee A, Kidd EA, Watson TF: In vitro evaluation of five alternative methods of carious dentine excavation. Caries Res 2000; 34: Bjelkhagen H, Sundström F, Angmar-Månsson B, Ryden H: Early detection of enamel caries by the luminescence excited by visible laser light. Swed Dent J 1982; 6: 1 7. Bohorfoush AG: New diagnostic methods for esophageal carcinoma. Recent Results Cancer Res 2000; 155: Brown JH, Brenn L: Method for the differential staining of Gram-positive and Gram-negative bacteria in tissue sections. Bull John Hopkins Hosp 1931; 48: Cederlund A, Lindskog S, Blomlof J: Efficacy of Carisolv-assisted caries excavation. Int J Periodontics Restorative Dent 1999; 19: Ericson D, Zimmerman M, Raber H, Gotrick B, Bornstein R, Thorell J: Clinical evaluation of efficacy and safety of a new method for chemomechanical removal of caries: A multi-centre study. Caries Res 1999; 33: Francescut P, Lussi A: Correlation between fissure discoloration, Diagnodent measurements, and caries depth: An in vitro study. Pediatr Dent 2003; 25: Gwinnett AJ: A comparison of proximal carious lesions as seen by clinical radiography, contact microradiography, and light microscopy. J Am Dent Assoc 1971; 83: Heinrich-Weltzien R, Kuhnisch J, Oehme T, Ziehe A, Stosser L, Garcia-Godoy F: Comparison of different DIAGNOdent cut-off limits for in vivo detection of occlusal caries. Oper Dent 2003; 28: Hibst R, Gall R: Development of a Diode Laser- Based Fluorescence Caries Detector. Caries Res 1998; 32: Hibst R, Keller U: Experimental studies of the application of the Er:YAG laser on dental hard substances. I. Measurement of the ablation rate. Lasers Surg Med 1989; 9: Hossain M, Nakamura Y, Yamada Y, Kimura Y, Matsumoto N, Matsumoto K: Effects of Er,Cr: YSGG laser irradiation in human enamel and dentine: Ablation and morphological studies. J Clin Laser Med Surg 1999; 17: Iwami Y, Shimizu A, Yamamoto H, Hayashi M, Takeshige F, Ebisu S: In vitro study of caries detection through sound dentine using a laser fluorescence device, DIAGNOdent. Eur J Oral Sci 2003; 111: Keller U, Hibst R: Experimental studies of the application of the Er:YAG laser on dental hard substances. II. Light microscopic and SEM investigations. Lasers Surg Med 1989; 9: Keller U, Hibst R: Effects of Er:YAG laser in caries treatment: A clinical pilot study. Lasers Surg Med 1997; 20: Keller U, Hibst R, Geurtsen W, Schilke R, Heidemann D, Klaiber B, Raab WH: Erbium:YAG laser application in caries therapy: Evaluation of patient perception and acceptance. J Dent 1998; 26: Kidd EA, Joyston-Bechal S, Beighton D: The use of a caries detector dye during cavity preparation: A microbiological assessment. Br Dent J 1993; 174: König K, Flemming G, Hibst R: Laser-induced autofluorescence spectroscopy of dental caries. Cell Mol Biol (Noisy-le-grand) 1998; 44: Lennon ÁM: Fluorescence-aided caries excavation (FACE) compared to conventional method. Oper Dent 2003; 28: Lennon ÁM, Buchalla W, Switalski L, Stookey GK: Residual caries detection using visible fluorescence. Caries Res 2002; 36: Caries Res 372 Eberhard /Eisenbeiss/Braun/Hedderich/ Jepsen CRE372.indd :47:41

9 Lussi A, Francescut P, Achermann F, Reich E, Hotz P, Megert B: The use of the DIAGNOdent during cavity preparation (abstract 56). Caries Res 2000; 34: Lussi A, Hibst R, Paulus R: DIAGNOdent: An optical method for caries detection. J Dent Res 2004; 83(special issue C):C Lussi A, Imwinkelried S, Pitts N, Longbottom C, Reich E: Performance and reproducibility of a laser fluorescence system for detection of occlusal caries in vitro. Caries Res 1999; 33: Lussi A, Megert B, Longbottom C, Reich E, Francescut P: Clinical performance of a laser fluorescence device for detection of occlusal caries lesions. Eur J Oral Sci 2001; 109: Nair PN, Baltensperger MM, Luder HU, Eyrich GK: Pulpal response to Er:YAG laser drilling of dentine in healthy human third molars. Lasers Surg Med 2003; 32: Ogushi K, Fusayama T: Electron microscopic structure of the two layers of carious dentine. J Dent Res 1975; 54: Paez X, Hernandez L: Biomedical applications of capillary electrophoresis with laser-induced fluorescence detection. Biopharm Drug Dispos 2001; 22: Pretty IA, Edgar WM, Higham SM: The effect of dehydration on quantitative light-induced fluorescence analysis of early enamel demineralisation. J Oral Rehabil 2004; 31: Shigetani Y, Tate Y, Okamoto A, Iwaku M, Abu- Bakr N: A study of cavity preparation by Er: YAG laser: Effects on the marginal leakage of composite resin restoration. Dent Mater J 2002; 21: Sonntag KD, Klitzman B, Burkes EJ, Hoke J, Moshonov J: Pulpal response to cavity preparation with the Er:YAG and Mark III free electron lasers. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1996; 81: Sternberg ED, Dolphin D: Second generation photodynamic agents: A review. J Clin Laser Med Surg 1993; 11: Sundström F, Fredriksson K, Montan S, Hafstrom-Bjorkman U, Strom J: Laser-induced fluorescence from sound and carious tooth substance: Spectroscopic studies. Swed Dent J 1985; 9: Tavares M, Soparkar PM, DePaola PF: Evaluation of a chemomechanical method of caries removal in root surface lesions. Quintessence Int 1988; 19: Vaarkamp J, ten Bosch JJ, Verdonschot EH: Propagation of light through human dental enamel and dentine. Caries Res 1995; 29: Yamada Y, Hossain M, Nakamura Y, Suzuki N, Matsumoto K: Comparison between the removal effect of mechanical, Nd:YAG, and Er: YAG laser systems in carious dentine. J Clin Laser Med Surg 2001; 19: Yip HK, Stevenson AG, Beeley JA: The specificity of caries detector dyes in cavity preparation. Br Dent J 1994; 176: Caries Treatment by Fluorescence-Guided Er:YAG Laser Caries Res CRE372.indd :47:41

10 Filename: CRE372 Article-No: 372, Fig.: 6, Tab.: 1 <Issueid>000</Issueid> </S> <F16_Sup>a </F16_Sup> </S> <F16_Sup>a </F16_Sup> <I>A. </I> <S>Braun </S> <F16_Sup>c </F16_Sup> </S> <F16_Sup>b </F16_Sup> <I>S. </I> <S>Jepsen </S> <F16_Sup>c </F16_Sup> <I>J. </I> <S>Eberhard <I>A.K. </I> <S>Eisenbeiss <I>J. </I> <S>Hedderich Abstract Sorry, there is no abstract. Read the first few lines of the text instead! Copyright 2005 S. Karger AG, Basel CRE372.indd :47:41

Diagnodent and the caveats of caries diagnosis by laser fluorescence

Diagnodent and the caveats of caries diagnosis by laser fluorescence Diagnodent and the caveats of caries diagnosis by laser fluorescence Introduction Professor Laurence Walsh School of Dentistry and Centre for Laser Science, The University of Queensland The Diagnodent

More information

Performance of a laser fluorescence device in detecting oclussal caries in vitro

Performance of a laser fluorescence device in detecting oclussal caries in vitro Performance of a laser fluorescence device in detecting oclussal caries in vitro Roxana Oancea1, Ruxandra Sava-Roºianu 1, Atena Gãluºcan 2, Daniela Jumanca 3, Ramona Amina Popovici 1, Angela Codruþa Podariu

More information

Near-UV light detection

Near-UV light detection C L I N I C A L Near-UV light detection Javier Tapia Guadix 1 Near-UV light induced fluorescence has already proven to be very useful as an alternative to classic caries-detector dyes. However its potential

More information

Preprint. Reference. Caries diagnosis using light fluorescence devices: VistaProof and DIAGNOdent. BETRISEY, Emilie, et al.

Preprint. Reference. Caries diagnosis using light fluorescence devices: VistaProof and DIAGNOdent. BETRISEY, Emilie, et al. Preprint Caries diagnosis using light fluorescence devices: VistaProof and DIAGNOdent BETRISEY, Emilie, et al. Reference BETRISEY, Emilie, et al. Caries diagnosis using light fluorescence devices: VistaProof

More information

Assessing Caries Removal by Undergraduate Dental Students Using Quantitative Light- Induced Fluorescence

Assessing Caries Removal by Undergraduate Dental Students Using Quantitative Light- Induced Fluorescence Assessing Caries Removal by Undergraduate Dental Students Using Quantitative Light- Induced Fluorescence Adejumoke A. Adeyemi, Ph.D.; Fadi D. Jarad, Ph.D.; Gleb N. Komarov, Ph.D.; Neil Pender, Ph.D.; Susan

More information

Performance of Laser Fluorescence for Residual Caries Detection in Primary Teeth

Performance of Laser Fluorescence for Residual Caries Detection in Primary Teeth Performance of Laser Fluorescence for Residual Caries Detection in Primary Teeth Taskin Gurbuz a Yucel Yilmaz b Fatih Sengul c Abstract Objectives: The aim of this study was to evaluate the performance

More information

Dental caries are bacterial induced necrosis of tooth structure

Dental caries are bacterial induced necrosis of tooth structure Dental caries are bacterial induced necrosis of tooth structure Treatment should focus on the removal of necrotic tissue with a bacteriocidal approach. Since the invention and application of rotary instruments,

More information

DIAGNOdent Measurements and Correlation with the Depth and Volume of Minimally Invasive Cavity Preparations

DIAGNOdent Measurements and Correlation with the Depth and Volume of Minimally Invasive Cavity Preparations Operative Dentistry, 2006, 31-3, 291-296 Clinical Research DIAGNOdent Measurements and Correlation with the Depth and Volume of Minimally Invasive Cavity Preparations JC Hamilton WA Gregory JB Valentine

More information

Evaluation of Dentin Root Canal Permeability After Instrumentation and Er:YAG Laser Application

Evaluation of Dentin Root Canal Permeability After Instrumentation and Er:YAG Laser Application Lasers in Surgery and Medicine 26:277 281 (2000) Evaluation of Dentin Root Canal Permeability After Instrumentation and Er:YAG Laser Application Jesus D. Pecora, DDS, PhD, 1 * Aldo Brugnera-Júnior, DDS,

More information

Correlation between Laser Fluorescence Readings and Volume of Tooth Preparation in Incipient Occlusal Caries In Vitrojerd_

Correlation between Laser Fluorescence Readings and Volume of Tooth Preparation in Incipient Occlusal Caries In Vitrojerd_ Correlation between Laser Fluorescence Readings and Volume of Tooth Preparation in Incipient Occlusal Caries In Vitrojerd_309 31..39 EDUARDO S. GHANAME, DDS, MS* ANDRÉ V. RITTER, DDS, MS HARALD O. HEYMANN,

More information

Scanning Is Believing

Scanning Is Believing KaVo DIAGNOdent /DIAGNOdent pen Laser Caries Detection Aid Scanning Is Believing Reveal the Hidden... Treat with Confidence NEW DIAGNOdent pen! Why you need DIAGNOdent Why yyou need If you cannot detect

More information

Original Paper. D.F. Côrtes a K.R. Ekstrand b A.R. Elias-Boneta c, R.P. Ellwood d

Original Paper. D.F. Côrtes a K.R. Ekstrand b A.R. Elias-Boneta c, R.P. Ellwood d Original Paper Caries Res 2000;34:443 447 Received: December 9, 1999 Accepted after revision: July 10, 2000 An in vitro Comparison of the Ability of Fibre-Optic Transillumination, Visual Inspection and

More information

71 Photothermal Radiometry and Modulated Luminescence: Applications for Dental Caries Detection

71 Photothermal Radiometry and Modulated Luminescence: Applications for Dental Caries Detection j1047 71 Photothermal Radiometry and Modulated Luminescence: Applications for Dental Caries Detection Jose A. Garcia, Andreas Mandelis, Stephen H. Abrams, and Anna Matvienko 71.1 Introduction The traditional

More information

Collaborators. Dental caries. Fluorescence and photochemical reactions: Lessons from teeth and tissues. 54,000+ patient appointments per year

Collaborators. Dental caries. Fluorescence and photochemical reactions: Lessons from teeth and tissues. 54,000+ patient appointments per year Fluorescence and photochemical reactions: Lessons from teeth and tissues Laurence J. Walsh School of Dentistry, The University of Queensland 2006 Collaborators Dr Jelena Vlacic Dr Fardad Shakabaie Dr Srdjan

More information

MinimalInvasiveTechniquesinCariesDetectionDiagnosisandMangagementAClinicalStudy

MinimalInvasiveTechniquesinCariesDetectionDiagnosisandMangagementAClinicalStudy Global Journal of Medical Research: J Dentistry & Otolaryngology Volume 16 Issue 2 Version 1.0 Year 2016 Type: Double Blind Peer Reviewed International Research Journal Publisher: Global Journals Inc.

More information

Emphasis has been placed on the diagnosis of occlusal

Emphasis has been placed on the diagnosis of occlusal Scientific Article Occlusal Caries Detection: A Comparison of a Laser Fluorescence System and Conventional Methods Cíntia Burin, DDS 1 Cristian Burin, DDS 1 Alessandro Dourado Loguercio, DDS, MS, PhD 2

More information

CLINICAL EVALUATION OF CHEMO-MECHANICAL CARIES REMOVAL USING CARIE-CARE SYSTEM AMONG SCHOOL CHILDREN

CLINICAL EVALUATION OF CHEMO-MECHANICAL CARIES REMOVAL USING CARIE-CARE SYSTEM AMONG SCHOOL CHILDREN Original Article NUJHS Vol. 4, No.3, September 2014, ISSN 2249-7110 CLINICAL EVALUATION OF CHEMO-MECHANICAL CARIES REMOVAL USING CARIE-CARE SYSTEM AMONG SCHOOL CHILDREN 1 2 3 4 Amitha M. Hegde, Preethi

More information

Er:YAG and adhesion in conservative dentistry : clinical overview

Er:YAG and adhesion in conservative dentistry : clinical overview Er:YAG and adhesion in conservative dentistry : clinical overview Carlo Fornaini MD, DDS, IMeLaS 1,2 1: Dental School, Faculty of Medicine and Surgery, University of Parma,Italy 2: Faculty of Dentistry,

More information

The Effect of Distance and Tooth Structure on Laser Fluorescence Caries Detection

The Effect of Distance and Tooth Structure on Laser Fluorescence Caries Detection Ó Operative Dentistry, 2012, 37-2, 150-160 The Effect of Distance and Tooth Structure on Laser Fluorescence Caries Detection K Markowitz RM Stenvall M Graye Clinical Relevance Devices used to aid occlusal

More information

Early lesion detection at 6-7 years and years schoolchildren from Bucharest, a prediction factor of decay evolution

Early lesion detection at 6-7 years and years schoolchildren from Bucharest, a prediction factor of decay evolution Early lesion detection at 6-7 years and 12-13 years schoolchildren from Bucharest, a prediction factor of decay evolution Roxana Ranga 1, Adina Dumitrache 2, Oana Slusanschi 3, Marian Cuculescu 4 Bucharest,

More information

Detection of intraoral lesions using a fluorescence camera

Detection of intraoral lesions using a fluorescence camera Detection of intraoral lesions using a fluorescence camera Michael Thoms Dürr Dental GmbH & Co KG, Höpfigheimer Str. 17, D-74321 Bietigheim-Bissingen University of Erlangen-Nürnberg, Institute of Material

More information

Endodontics Cracked Tooth: How to manage it in daily practice

Endodontics Cracked Tooth: How to manage it in daily practice Calogero Bugea Endodontics Cracked Tooth: How to manage it in daily practice 5 Feb 2016 Tooth Fractures are not rare, surface cracks, or craze lines, are relatively common in teeth. In most of cases they

More information

EMDOLA Education Program: First Year Details. History of lasers: Optics data : The quantum nature of the light:

EMDOLA Education Program: First Year Details. History of lasers: Optics data : The quantum nature of the light: EMDOLA Education Program Module 1: Optics Module 2: Physics of lasers Module 3:Interaction laser-tissues Module 4: Laser Safety Properties of lasers Module 5:Lasers conservative dentistry Laser in caries

More information

Investigation of Dental Caries Using Laser and Light-Induced Autofluorescence Methods

Investigation of Dental Caries Using Laser and Light-Induced Autofluorescence Methods Bulg. J. Phys. 33 (2006) 55 67 Investigation of Dental Caries Using Laser and Light-Induced Autofluorescence Methods E. Borisova 1, Tz. Uzunov 2, L. Avramov 1 1 Institute of Electronics, Bulgarian Academy

More information

Direct composite restorations for large posterior cavities extended range of applications for high-performance materials

Direct composite restorations for large posterior cavities extended range of applications for high-performance materials Direct composite restorations for large posterior cavities extended range of applications for high-performance materials A case study by Ann-Christin Meier, Dr. med. dent., Stapelfeld, Germany When large

More information

New horizons for Er:YAG lasers: QSP mode advantages in the Lightwalker AT

New horizons for Er:YAG lasers: QSP mode advantages in the Lightwalker AT I research_ Er:YAG s New horizons for Er:YAG s: QSP mode advantages in the Lightwalker AT Author_Evgeniy Mironov, DDS; Assist. Prof. Zhasmina Mironova, DDS; Assoc. Prof. Radosveta Vassileva, DDS and Assoc.

More information

Diagnostic Values of Laser Fluorescence Device Compared to Other Techniques in Occlusal Caries Detection

Diagnostic Values of Laser Fluorescence Device Compared to Other Techniques in Occlusal Caries Detection Original Article Diagnostic Values of Laser Fluorescence Device Compared to Other Techniques in Occlusal Caries Detection M. Biria 1, 2, M. Moshfeghi 3, A. Akbarzadeh Baghban 4, A. Chehresaz 5 1 Assistant

More information

KaVo DIAGNOdent. Laser Fluorescence Caries Detection. Ever Play Hide and Seek With Caries?

KaVo DIAGNOdent. Laser Fluorescence Caries Detection. Ever Play Hide and Seek With Caries? KaVo DIAGNOdent Laser Fluorescence Caries Detection Ever Play Hide and Seek With Caries? Must have for microdentistry Only device to quantify caries Best ROI in dentistry DIAGNOdent: Because you can t

More information

Up to date literature at your disposal to discover the benefits of laser dentistry

Up to date literature at your disposal to discover the benefits of laser dentistry Up to date literature at your disposal to discover the benefits of laser dentistry For further information contact info@doctor-smile.com +39-0444-349165 Dipartimento di Scienze e Tecnologie Biofisiche

More information

Diagnosis of Pit and Fissure Caries Using Frequency-Domain Infrared Photothermal Radiometry and Modulated Laser Luminescence

Diagnosis of Pit and Fissure Caries Using Frequency-Domain Infrared Photothermal Radiometry and Modulated Laser Luminescence Original Paper DOI: 10.1159/000080579 Received: August 1, 2003 Accepted after revision: February 24, 2004 Diagnosis of Pit and Fissure Caries Using Frequency-Domain Infrared Photothermal Radiometry and

More information

Chemomechanical Caries Removal for Children

Chemomechanical Caries Removal for Children Stomatologija, Baltic Dental and Maxillofacial Journal, 7:40-4, 2005 Chemomechanical Caries Removal for Children Irena Balčiunienė, Rūta Sabalaitė, Inga Juškienė SUMMARY Pain is still an actual problem

More information

Comparison of two different methods of detecting residual caries

Comparison of two different methods of detecting residual caries Research article ISSN 2234-7658 (print) / ISSN 2234-7666 (online) Comparison of two different methods of detecting residual caries Uzay Koç Vural*, Zeynep Bilge Kütük, Esra Ergin, Filiz Yalçın Çakır, Sevil

More information

Title remineralization therapy. Issue Date Australian Dental Associatio

Title remineralization therapy. Issue Date Australian Dental Associatio NAOSITE: Nagasaki University's Ac Title Author(s) Early detection of white spot lesio remineralization therapy. Iijima, Youichi Citation Australian dental journal, 53(3), p Issue Date 2008-09 URL http://hdl.handle.net/10069/23123

More information

Caries prevention is critical for children, especially. Quantitative Assessment of Enamel Hypomineralization. Permanent Molars of Children in China

Caries prevention is critical for children, especially. Quantitative Assessment of Enamel Hypomineralization. Permanent Molars of Children in China Scientific Article Quantitative Assessment of Enamel Hypomineralization by KaVo DIAGNOdent at Different Sites on First Permanent Molars of Children in China Shao-Min Li, DDS Jing Zou, DDS, MS Zhengyan

More information

Effects of Dental Probing on Occlusal Surfaces A Scanning Electron Microscopy Evaluation

Effects of Dental Probing on Occlusal Surfaces A Scanning Electron Microscopy Evaluation Original Paper DOI: 10.1159/000096104 Received: April 29, 2004 Accepted after revision: March 31, 2006 Effects of Dental Probing on Occlusal Surfaces A Scanning Electron Microscopy Evaluation Jan Kühnisch

More information

Bangladesh Journal of Medical Science Vol. 17 No. 02 April 18

Bangladesh Journal of Medical Science Vol. 17 No. 02 April 18 Bangladesh Journal of Medical Science Vol. 17 No. 02 April 18 Original article: Efficacy of an experimental propylene glycol-based, water-free caries detecting dye in comparison with Snoop using histological

More information

Chapter 5. Investigation of in vitro Dental Erosion by Optical Techniques

Chapter 5. Investigation of in vitro Dental Erosion by Optical Techniques Chapter 5 Investigation of in vitro Dental Erosion by Optical Techniques 5.1 INTRODUCTION 5 Dental erosion is becoming an increasing predicament in dentistry and can be attributed to either exogenous

More information

DENTAL CLINICS OF NORTH AMERICA. Emmiam. Incipient and Hidden Caries. GUEST EDITOR Daniel W. Boston, DMD. October 2005 Volume49 Number4

DENTAL CLINICS OF NORTH AMERICA. Emmiam. Incipient and Hidden Caries. GUEST EDITOR Daniel W. Boston, DMD. October 2005 Volume49 Number4 DENTAL CLINICS OF NORTH AMERICA Incipient and Hidden Caries GUEST EDITOR Daniel W. Boston, DMD October 2005 Volume49 Number4 Emmiam An Imprint of Elsevier, Inc. PHILADELPHIA LONDON TORONTO MONTREAL SYDNEY

More information

The Effect of Mineralizing Fluorine Varnish on the Progression of Initial Caries of Enamel in Temporary Dentition by Laser Fluorescence

The Effect of Mineralizing Fluorine Varnish on the Progression of Initial Caries of Enamel in Temporary Dentition by Laser Fluorescence American Journal of Engineering Research (AJER) e-issn: 2320-0847 p-issn : 2320-0936 Volume-6, Issue-9, pp-39-43 www.ajer.org Research Paper Open Access The Effect of Mineralizing Fluorine Varnish on the

More information

Micro-CT evaluation of cavities prepared with with different Er:YAG handpieces

Micro-CT evaluation of cavities prepared with with different Er:YAG handpieces L. Savatier*, F. Curnier, I. Krejci Division of Cariology and Endodontology University of Geneva, Geneva, Switzerland *General Dentist, London, United Kingdom e-mail: ivo.krejci@unige.ch Micro-CT evaluation

More information

Use of the ICDAS system and two fluorescence-based intraoral devices for examination of occlusal surfaces

Use of the ICDAS system and two fluorescence-based intraoral devices for examination of occlusal surfaces A. Theocharopoulou, M.D. Lagerweij, A.J.P. van Strijp Department of Cariology Endodontology Pedodontology, Academic Centre for Dentistry Amsterdam (ACTA), Amsterdam, The Netherlands e-mail: katerina5the@gmail.com

More information

Department of Oral Surgery, Heinrich Heine University, D Düsseldorf, Germany.

Department of Oral Surgery, Heinrich Heine University, D Düsseldorf, Germany. Lasers Surg Med. 2007 Jun;39(5):428-40 Immunohistochemical characterization of periodontal wound healing following nonsurgical treatment with fluorescence controlled Er:YAG laser radiation in dogs. Schwarz

More information

Dental Depth Profilometric Diagnosis of Pit & Fissure Caries Using Frequency-Domain Infrared Photothermal Radiometry and Modulated Laser Luminescence

Dental Depth Profilometric Diagnosis of Pit & Fissure Caries Using Frequency-Domain Infrared Photothermal Radiometry and Modulated Laser Luminescence Dental Depth Profilometric Diagnosis of Pit & Fissure Caries Using Frequency-Domain Infrared Photothermal Radiometry and Modulated Laser Luminescence Raymond J. Jeon, Calvin Han, Andreas Mandelis* and

More information

A Comparison of Methods for the Detection of Smooth Surface Caries

A Comparison of Methods for the Detection of Smooth Surface Caries A Comparison of Methods for the Detection of Smooth Surface Caries B. Wong 1, K. Sivagurunathan 1,J. D. Silvertown 1, W. M. P. Hellen 2, G. I. Elman 2, S. H. Abrams 1, L. O. Okoye 3, B. T. Amaechi 4 1

More information

MDJ Evaluation the effect of eugenol containing temporary Vol.:9 No.:2 2012

MDJ Evaluation the effect of eugenol containing temporary Vol.:9 No.:2 2012 MDJ Evaluation the effect of eugenol containing temporary fillings on shear bond strength of composite restoration Dr. Farid G. Numan B.D.S., M.Sc Abstract The purpose of this in-vitro study was to evaluate

More information

Morphological changes in hard dental tissue prepared using the Er:YAG laser

Morphological changes in hard dental tissue prepared using the Er:YAG laser I user report _ morphological changes Morphological changes in hard dental tissue prepared using the Er:YAG Author_Drs Snejana Ts. Tsanova & Georgi T. Tomov, Bulgaria Fig. 1a Figs. 1a c_extracted teeth

More information

Anisotropy of Tensile Strengths of Bovine Dentin Regarding Dentinal Tubule Orientation and Location

Anisotropy of Tensile Strengths of Bovine Dentin Regarding Dentinal Tubule Orientation and Location Original paper Dental Materials Journal 21 (1): 32-43, 2002 Anisotropy of Tensile Strengths of Bovine Dentin Regarding Dentinal Tubule Orientation and Location Toshiko INOUE, Hidekazu TAKAHASHI and Fumio

More information

Dental Depth Profilometric Diagnosis of Pit & Fissure Caries using Frequency-Domain Infrared Photothermal Radiometry and Modulated Laser Luminescence

Dental Depth Profilometric Diagnosis of Pit & Fissure Caries using Frequency-Domain Infrared Photothermal Radiometry and Modulated Laser Luminescence Dental Depth Profilometric Diagnosis of Pit & Fissure Caries using Frequency-Domain Infrared Photothermal Radiometry and Modulated Laser Luminescence Raymond J. Jeon a, Andreas Mandelis *a, Victor Sanchez

More information

Comparison of White MTA And Grey MTA in the Apical Sealing Ability of Lased And Unlased Root Canal Walls - A Pilot Study

Comparison of White MTA And Grey MTA in the Apical Sealing Ability of Lased And Unlased Root Canal Walls - A Pilot Study IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861.Volume 15, Issue 6 Ver. XIV (June 2016), PP 100-104 www.iosrjournals.org Comparison of White MTA And Grey MTA

More information

Chapter 15 Detection of Early Caries by Laser-Induced Breakdown Spectroscopy

Chapter 15 Detection of Early Caries by Laser-Induced Breakdown Spectroscopy Chapter 15 Detection of Early Caries by Laser-Induced Breakdown Spectroscopy Yuji Matsuura Abstract To improve sensitivity of dental caries detection by laser-induced breakdown spectroscopy (LIBS) analysis,

More information

powered by technology easy laser guide

powered by technology easy laser guide powered by technology easy laser guide Rev. 1.0 / Oct. 2009 2 Easy Laser Guide The elexxion Easy Laser Guide offers both general information about dental lasers as well as specific information about elexxion

More information

Current Concepts in Caries Management Diagnostic, Treatment and Ethical/Medico-Legal Considerations. Radiographic Caries Diagnosis

Current Concepts in Caries Management Diagnostic, Treatment and Ethical/Medico-Legal Considerations. Radiographic Caries Diagnosis Current Concepts in Caries Management Diagnostic, Treatment and Ethical/Medico-Legal Considerations Richard N. Bohay, DMD, MSc, MRCDC Associate Professor, Schulich School of Medicine & Dentistry Schulich

More information

Medical sciences (2018) 1 16

Medical sciences (2018) 1 16 Medical sciences (2018) 1 16 Profilometric and histological evaluation of the root surface after instrumentation in vitro with a fluorescence-controlled er: yag laser, hand instruments and an ultrasonic

More information

Photoluminescence spectra of sound tooth and those of different carious stages

Photoluminescence spectra of sound tooth and those of different carious stages Journal Biological Series 7(1): 000-000, January 2019 DOI: 10.15413/jbs2019.0121 2019 Academia Publishing Research Paper Photoluminescence spectra of sound tooth and those of different carious stages Accepted

More information

Application of two fluorescence methods for detection and quantification of smooth surface carious lesions. Abdulaziz Saad Aljehani

Application of two fluorescence methods for detection and quantification of smooth surface carious lesions. Abdulaziz Saad Aljehani From the Department of Cariology and Endodontology, Institute of Odontology, Karolinska Institutet, Stockholm, Sweden Application of two fluorescence methods for detection and quantification of smooth

More information

A NOVEL APPROACH FOR TREATING FISSURE CARIES. The problem of hidden caries. Current diagnostic methods

A NOVEL APPROACH FOR TREATING FISSURE CARIES. The problem of hidden caries. Current diagnostic methods A NOVEL APPROACH FOR TREATING FISSURE CARIES Daniel W. Boston, DMD. Associate Professor and Chairman Temple University School of Dentistry Department of Restorative Dentistry 3223 North Broad Street Philadelphia,

More information

Active Care - Shadeguides How to choose the right Laser

Active Care - Shadeguides How to choose the right Laser Giuseppe Chiodera Active Care - Shadeguides How to choose the right Laser 2018-04-18 Just as you, I was once looking to buy a laser for my dental practice but struggling to choose the right one. Today,

More information

The Histology of Dentin

The Histology of Dentin The Histology of Dentin Pauline Hayes Garrett, D.D.S. Department of Endodontics, Prosthodontics, and Operative Dentistry University of Maryland, Baltimore This material was taken from: Essentials of Oral

More information

Relationship between Colors of Carious Dentin and Laser Fluorescence Evaluations in Caries Diagnosis

Relationship between Colors of Carious Dentin and Laser Fluorescence Evaluations in Caries Diagnosis Dental Materials Journal 25(3):584-590, 2006 Relationship between Colors of Carious Dentin and Laser Fluorescence Evaluations in Caries Diagnosis Yukiteru IWAMI, Ayako SHIMIZU, Mikako HAYASHI, Fumio TAKESHIGE

More information

COVER FEATURE SYNOPSIS

COVER FEATURE SYNOPSIS Supplementary Methods for Detection and Quantification of Dental Caries Lena Karlsson, RDH; Sofia Tranæus, DDS, PhD Department of Cariology and Endodontology, Institute of Odontology, Karolinska Institute,

More information

Preclinical Dentistry. I. Dental Caries Non carious lesions: trauma, erosion. abrasion, wedge shaped defects. Lenka Roubalíková

Preclinical Dentistry. I. Dental Caries Non carious lesions: trauma, erosion. abrasion, wedge shaped defects. Lenka Roubalíková Preclinical Dentistry I. Dental Caries Non carious lesions: trauma, erosion. abrasion, wedge shaped defects Lenka Roubalíková Understanding dental caries crown pulp chamber neck root canal root Dental

More information

In Vitro Evaluation of the Spectra Early Caries Detection System

In Vitro Evaluation of the Spectra Early Caries Detection System In Vitro Evaluation of the Spectra Early Caries Detection System Maria Graye, DMD Pedodontist in Private Practice Hamilton, NJ, USA Kenneth Markowitz, DDS, MSD Assistant Professor, Department of Oral Biology

More information

Practice Impact Questionnaire

Practice Impact Questionnaire Practice Impact Questionnaire Your practitioner identifier is: XXXXXXXX It is very important that ONLY YOU complete this questionnaire because your responses will be compared to responses that you provided

More information

Recovery percentage of remineralization according to severity of early caries

Recovery percentage of remineralization according to severity of early caries Research Article Recovery percentage of remineralization according to severity of early caries HEE EUN KIM, RDH, PHD, HO KEUN KWON, DDS, PHD & BAEK IL KIM, DDS, PHD ABSTRACT: Purpose: To analyze the cutoff

More information

LASERS FOR DENTAL APPLICATIONS.

LASERS FOR DENTAL APPLICATIONS. LASERS FOR DENTAL APPLICATIONS. Dr. Zsolt Tóth University of Szeged, Faculty of Dentistry Department of Oral Biology and Experimental Dental Research CONTENT: LASER dental tissue interactions Laser - material

More information

Efficacy of Chemomechanical Caries Removal System as Compared to Conventional Rotary Method of Caries Removal - An In Vitro Comparative Study

Efficacy of Chemomechanical Caries Removal System as Compared to Conventional Rotary Method of Caries Removal - An In Vitro Comparative Study ORIGINAL RESEARCH Efficacy of Chemomechanical Caries Removal System as Compared to Conventional Rotary Method of Caries Removal - An In Vitro Comparative Study Shilpy Dwivedi 1, Vinod Patel 2, Premkishore

More information

In vivo study of different methods for diagnosing pit and fissure caries

In vivo study of different methods for diagnosing pit and fissure caries Journal section: Operative Dentistry and Endodontics Publication Types: Research doi:10.4317/jced.52347 http://dx.doi.org/10.4317/jced.52347 In vivo study of different methods for diagnosing pit and fissure

More information

Efficiency of the Polymer Bur SmartPrep Compared with Conventional Tungsten Carbide Bud Bur in Dentin Caries Excavation

Efficiency of the Polymer Bur SmartPrep Compared with Conventional Tungsten Carbide Bud Bur in Dentin Caries Excavation Operative Dentistry, 2006, 31-2, 256-260 Efficiency of the Polymer Bur SmartPrep Compared with Conventional Tungsten Carbide Bud Bur in Dentin Caries Excavation T Dammaschke TN Rodenberg E Schäfer KHR

More information

Research Article Nd:YAG Lasers Treating of Carious Lesion and Root Canal In Vitro

Research Article Nd:YAG Lasers Treating of Carious Lesion and Root Canal In Vitro Photoenergy Volume 2012, Article ID 584079, 5 pages doi:10.1155/2012/584079 Research Article Nd:YAG Lasers Treating of Carious Lesion and Root Canal In Vitro Danqing Xia, 1, 2 Zenggui Mo, 3 Gang Zhao,

More information

Update in Caries Diagnosis

Update in Caries Diagnosis Update in Caries Diagnosis Jenny Smits, AAS, RDH University of Michigan E-Learning Program 482 Oral Diseases: Prevention & Management University of Michigan Dental Desktop Support 12/16/2009 Introduction

More information

An in vitro experiment in the early diagnosis of caries using four diagnostic methods

An in vitro experiment in the early diagnosis of caries using four diagnostic methods An in vitro experiment in the early diagnosis of caries using four diagnostic methods M. Peneva 1, Rashkova 2, M., Golchev Y 3 Sofia, Bulgaria Summary The visual diagnostics of early caries is an extremely

More information

Restorative Dentistry and it s related to Pulp health. Dr.Ahmed Al-Jobory

Restorative Dentistry and it s related to Pulp health. Dr.Ahmed Al-Jobory Restorative Dentistry and it s related to Pulp health Dr.Ahmed Al-Jobory Pulp Is a viscous connective tissue of collagen fibers and ground substance supporting the vital cellular, vascular, and nerve structures

More information

stabilisation and surface protection

stabilisation and surface protection Guiding the way to caries stabilisation and surface protection Fissure sealing MI restorations Pulp capping Hypersensitivity Protection Caries stabilisation Fuji Triage from GC. Temporary restorations

More information

VistaCam ix HD for diagnostics support in HD quality

VistaCam ix HD for diagnostics support in HD quality VistaCam ix HD for diagnostics support in HD quality The innovative intraoral camera with interchangeable head system COMPRESSED AIR SUCTION IMAGING DENTAL CARE HYGIENE Vista Cam ix HD real HD resolution

More information

Influence of Water Absorption Shift on Ablation Speed of Er:YAG and Er,Cr:YSGG Dental Lasers

Influence of Water Absorption Shift on Ablation Speed of Er:YAG and Er,Cr:YSGG Dental Lasers Absorption (mm -1 ) ISSN 1855-9913 Journal of the Laser and Health Academy Vol. 2013, No.1; www.laserandhealth.com Influence of Water Absorption Shift on Ablation Speed of Er:YAG and Er,Cr:YSGG Dental

More information

Diagnosis versus Detection of Caries

Diagnosis versus Detection of Caries Caries Res 2004;38:192 198 DOI: 10.1159/000077754 Diagnosis versus Detection of Caries B. Nyvad School of Dentistry, Faculty of Health Sciences, University of Aarhus, Aarhus, Denmark Key Words Dental caries,

More information

Impact of Photodynamic Therapy Applied by FotoSan on Periodontal Tissues Clinical Parameters

Impact of Photodynamic Therapy Applied by FotoSan on Periodontal Tissues Clinical Parameters Impact of Photodynamic Therapy Applied by FotoSan on Periodontal Tissues Clinical Parameters REVIEWED PAPER Most patients visiting dental surgeries suffer from various types of periodontopathies. Since

More information

Removal of Human Gingival Melanin Pigmentation by Er:YAG vs Nd:YAG Laser: A Case Report

Removal of Human Gingival Melanin Pigmentation by Er:YAG vs Nd:YAG Laser: A Case Report Removal of Human Gingival Melanin Pigmentation Er:YAG vs Nd:YAG Laser: A Case Report Satoshi Hasumi a, Jun Hasumi a, Koukichi Matsumoto b a Research Fellow, Department of Cariology and Endodontics, School

More information

Periodontal Regeneration

Periodontal Regeneration Periodontal Regeneration Regeneration The most ideal treatment Attempts to recreate the tissues destroyed by periodontitis Cement, bone and ligament Reduces the risk for recession and sensitivity (could

More information

Lec. 3-4 Dr. Saif Alarab Clinical Technique for Class I Amalgam Restorations The outline form

Lec. 3-4 Dr. Saif Alarab Clinical Technique for Class I Amalgam Restorations The outline form Lec. 3-4 Dr. Saif Alarab Clinical Technique for Class I Amalgam Restorations Class I refers to -Restorations on the occlusal surfaces of posterior teeth, - The occlusal two thirds of facial and lingual

More information

Clinical Evaluation of Polyamide Polymer Burs for Selective Carious Dentin Removal

Clinical Evaluation of Polyamide Polymer Burs for Selective Carious Dentin Removal Clinical Evaluation of Polyamide Polymer Burs for Selective Carious Dentin Removal Attiguppe Prabhakar, MDS; N. K. Kiran, MDS Abstract Aim: The aim of this study was to evaluate the carious dentin removal

More information

SOLASE DENTAL DIODE LASER

SOLASE DENTAL DIODE LASER TECHNICAI SPECIFICATIONS Wavelength Power Output Pulse Duration Pulse Repetition Rate Tip Diameter Power Supply Input Battery Pack Size / Weight Two models, 808nm or 976nm 976nm: 0.-6W@pulse mode, 0.-0W@CW

More information

QSP Mode Characteristics of 3rd Generation ASP Powered Er:YAG Dental Lasers

QSP Mode Characteristics of 3rd Generation ASP Powered Er:YAG Dental Lasers ISSN 1855-9913 Journal of the Laser and Health Academy Vol. 2016, No.1; www.laserandhealth.com QSP Mode Characteristics of 3rd Generation ASP Powered Er:YAG Dental Lasers Nejc Lukac 1, Matjaz Lukac 2,

More information

THERMAL CHANGES IN THE HARD DENTAL TISSUE AT DIODE LASER ROOT CANAL TREATMENT

THERMAL CHANGES IN THE HARD DENTAL TISSUE AT DIODE LASER ROOT CANAL TREATMENT 10.1515/amb-2014-0018 THERMAL CHANGES IN THE HARD DENTAL TISSUE AT DIODE LASER ROOT CANAL TREATMENT Ts. Uzunov 1, R. Grozdanova 1, E. Popova 1 and T. Uzunov 2 1 Department of Conservative Dentistry, Section

More information

Measurement of surface hardness of primary carious lesions in extracted human enamel Measurement of Knoop hardness using Cariotester

Measurement of surface hardness of primary carious lesions in extracted human enamel Measurement of Knoop hardness using Cariotester Dental Materials Journal 2015; 34(2): 252 256 Measurement of surface hardness of primary carious lesions in extracted human enamel Measurement of Knoop hardness using Cariotester Akihiko SHIMIZU 1, Takatsugu

More information

FACT SHEET: ABOUT DENTAL CARIES

FACT SHEET: ABOUT DENTAL CARIES FACT SHEET: ABOUT DENTAL CARIES What is dental caries? Dental caries is the technical term used by healthcare and dental professionals to describe the disease more commonly known as tooth decay. 1-2 Dental

More information

Examination and Treatment Protocols for Dental Caries and Inflammatory Periodontal Disease

Examination and Treatment Protocols for Dental Caries and Inflammatory Periodontal Disease Examination and Treatment Protocols for Dental Caries and Inflammatory Periodontal Disease Dental Caries The current understanding of the caries process supports the shift in caries management from a restorative-only

More information

Temperature Measurements During Dental Cutting Using an Internally Cooled Diamond Dental Bur

Temperature Measurements During Dental Cutting Using an Internally Cooled Diamond Dental Bur Journal of Materials Science and Engineering B 2 (10) (2012) 551-559 D DAVID PUBLISHING Temperature Measurements During Dental Cutting Using an Internally Cooled Diamond Dental Bur Uri Ben-Hanan, Michael

More information

LASERS IN PERIODONTAL THERAPY A REVIEW

LASERS IN PERIODONTAL THERAPY A REVIEW LASERS IN PERIODONTAL THERAPY A REVIEW Dr. B.M. Bhusari., Dr. Ridhima.V. Mahajan., Dr. Namrata. J. Suthar& Dr. Shubhangi.R. Rajbhoj* *Department of Periodontics, Y.M.T Dental college, Kharghar, Navi Mumbai,

More information

NON-IONIZING DIAGNOSTIC PROCEDURES

NON-IONIZING DIAGNOSTIC PROCEDURES NON-IONIZING DIAGNOSTIC PROCEDURES UnitedHealthcare Dental Clinical Policy Policy Number: DCP041.01 Effective Date: February 1, 2017 Table of Contents Page INSTRUCTIONS FOR USE...1 BENEFIT CONSIDERATIONS...1

More information

T he survival of an implant involves

T he survival of an implant involves KU ET AL IMPLANT DENTISTRY / VOLUME 25, NUMBER 3 2016 1 Explaining the Red Fluorescence Evident on the Surface of Failed Dental Implants: Case Reports Hye-Min Ku, RDH,* Mi-Kyoung Jun, RDH, MSD,* Jee-Hwan

More information

Air-Abrasive Technology for Caries Diagnosis and Fissure Sealing. A Clinical Study

Air-Abrasive Technology for Caries Diagnosis and Fissure Sealing. A Clinical Study Air-Abrasive Technology for Caries Diagnosis and Fissure Sealing. A Clinical Study Kathleen Fritz 1, Maik Wagner 2, Annerose Borutta 3 1 Dr. Med. Dent. WHO Collaborating Centre on Prevention of Oral Diseases,

More information

Review of possible predictors for pain perception with class 1 5 cavity preparations using Er,Cr:YSGG laser: a retrospective clinical in vivo study

Review of possible predictors for pain perception with class 1 5 cavity preparations using Er,Cr:YSGG laser: a retrospective clinical in vivo study Laser Dent Sci (2017) 1:9 21 DOI 10.1007/s41547-017-0001-4 ORIGINAL ARTICLE Review of possible predictors for pain perception with class 1 5 cavity preparations using Er,Cr:YSGG laser: a retrospective

More information

Gorduysus MO et al. Operating Microscope improves ability to locate and negotiate second canal

Gorduysus MO et al. Operating Microscope improves ability to locate and negotiate second canal ORIGINAL RESEARCH I.J.C.M.R Operating microscope improves ability to locate and negotiate second canal in mandibular ıncisors Mehmet Omer Gorduysus, Melahat Gorduysus, Zeliha Yilmaz * Department of Endodontics,

More information

ECC II Caries Disease Status. Drs Francisco Ramos-Gomez, Man Wai Ng and Jessica Lee

ECC II Caries Disease Status. Drs Francisco Ramos-Gomez, Man Wai Ng and Jessica Lee ECC II Caries Disease Status Drs Francisco Ramos-Gomez, Man Wai Ng and Jessica Lee Strategies for the detection and monitoring of early carious lesions. Detection and Staging Traditional Non- Surgical

More information

Fuji II LC. A Perfect Choice

Fuji II LC. A Perfect Choice A Perfect Choice is a remarkable restorative material The world s first resin-reinforced glass ionomer has remained the benchmark for light cured glass ionomer cements, delivering more than 15 years of

More information

Selected Laser Bibliography and Literature Review Bold type incidates highly recommended reading.

Selected Laser Bibliography and Literature Review Bold type incidates highly recommended reading. Selected Laser Bibliography and Literature Review Bold type incidates highly recommended reading. Academy of Laser Dentistry Website: laserdentistry,org/prof/edu_bibliography.cfm laserdentistry.org/prof/edu_resources.cfm

More information

The influence of erbium:yttrium aluminum garnet laser ablation with variable pulse width on morphology and microleakage of composite restorations

The influence of erbium:yttrium aluminum garnet laser ablation with variable pulse width on morphology and microleakage of composite restorations DOI 10.1007/s10103-009-0736-6 ORIGINAL ARTICLE The influence of erbium:yttrium aluminum garnet laser ablation with variable pulse width on morphology and microleakage of composite restorations Ricardo

More information

Orange/Red Fluorescence of Active Caries by Retrospective Quantitative Light-Induced Fluorescence Image Analysis

Orange/Red Fluorescence of Active Caries by Retrospective Quantitative Light-Induced Fluorescence Image Analysis Original Paper Received: December 15, 2014 Accepted after revision: October 21, 2015 Published online: May 11, 2016 Orange/Red Fluorescence of Active Caries by Retrospective Quantitative Light-Induced

More information