Detection of Atrial Fibrillation and Flutter by a Dual-Chamber Implantable Cardioverter-Defibrillator

Size: px
Start display at page:

Download "Detection of Atrial Fibrillation and Flutter by a Dual-Chamber Implantable Cardioverter-Defibrillator"

Transcription

1 Detection of Atrial Fibrillation and Flutter by a Dual-Chamber Implantable Cardioverter-Defibrillator Charles D. Swerdlow, MD; Wolfgang Schšls, MD; Barbara Dijkman, MD; Werner Jung, MD; Nirav V. Sheth, MS; Walter H. Olson, PhD; Bruce D. Gunderson, MS; for the Worldwide Jewel AF Investigators Background To distinguish prolonged episodes of atrial fibrillation (AF) that require cardioversion from self-terminating episodes that do not, an atrial implantable cardioverter-defibrillator (ICD) must be able to detect AF continuously for extended periods. The ICD should discriminate between atrial tachycardia/flutter (AT), which may be terminated by antitachycardia pacing, and AF, which requires cardioversion. Methods and Results We studied 80 patients with AT/AF and ventricular arrhythmias who were treated with a new atrial/dual-chamber ICD. During a follow-up period lasting 6 2 months, we validated spontaneous, device-defined AT/AF episodes by stored electrograms in all patients. In 58 patients, we performed 80 Holter recordings with telemetered atrial electrograms, both to validate the continuous detection of AT/AF and to determine the sensitivity of the detection of AT/AF. Detection was appropriate in 98% of 132 AF episodes and 88% of 190 AT episodes (98% of 128 AT episodes with an atrial cycle length 300 ms). Intermittent sensing of far-field R waves during sinus tachycardia caused 27 inappropriate AT/AF detections; these detections lasted minutes. AT/AF was detected continuously in 27 of 28 patients who had spontaneous episodes of AT/AF (96%). The device memory recorded 90 appropriate AT/AF episodes lasting 1 hour, for a total of 2697 hours of continuous detection of AT/AF. During Holter monitoring, the sensitivity of the detection of AT/AF (116 hours) was 100%; the specificity of the detection of non-at/af rhythms (1290 hours) was 99.99%. Of 166 appropriate episodes detected as AT, 45% were terminated by antitachycardia pacing. Conclusions A new ICD detects AT/AF accurately and continuously. Therapy may be programmed for long-duration AT/AF, with a low risk of underdetection. Discrimination of AT from AF permits successful pacing therapy for a significant fraction of AT. (Circulation. 2000;101: ) Key Words: atrial fibrillation defibrillators, implantable arrhythmia atrial flutter Sensing and detection requirements for implantable cardioverter-defibrillators (ICDs) that treat atrial fibrillation (AF) and atrial tachycardia/flutter (AT) differ from those for ventricular ICDs and dual-chamber pacemakers. Because AF is usually hemodynamically stable and frequently terminates spontaneously, it may be desirable to delay painful and potentially proarrhythmic atrial defibrillation shocks for periods 24 hours. To achieve this goal, an ICD must sense low and varying-amplitude AF electrograms and detect AF continuously. It must distinguish continuous AF, which may require shocks, from termination and subsequent reinitiation of AF, for which shocks should be withheld. However, atrial sensing of far-field R waves should not cause inappropriate detection of AT/AF. Further, an atrial ICD should discriminate between AT, which may be terminated by antitachycardia pacing, and AF, which requires cardioversion. A new dual-chamber ICD (Medtronic Jewel AF 7250) detects specific atrial and ventricular tachyarrhythmias; has independently programmable therapy for AT, AF, ventricular tachycardia (VT), and ventricular fibrillation (VF); and functions as a DDD pacemaker. We studied patients treated with this ICD to determine if the device could detect AT/AF accurately and continuously and discriminate AT from AF. Methods Patients We studied 80 consecutive patients treated with this ICD in 19 medical centers (Appendix). Patients had both a clinical indication for the implantation of a ventricular ICD and 2 episodes of AF and/or AT in the 3 months before implantation. All patients gave written, informed consent according to a protocol approved by the Human Subjects Committee of the institution at which the devices were implanted. Received June 7, 1999; revision received September 8, 1999; accepted September 23, From Cedars-Sinai Medical Center, Los Angeles, Calif (C.D.S.); University of Heidelberg, Heidelberg, Germany (W.S.); University of Maastricht, Maastricht, the Netherlands (B.D.); University of Bonn, Bonn, Germany (W.J.); and Medtronic Inc, Minneapolis, Minn (N.V.S., W.H.O., B.D.G.). A list of all participating investigators and medical centers is given in the Appendix. Correspondence to Charles D. Swerdlow, MD, 8635 W. Third Street, Ste 1190 W, Los Angeles, CA swerdlow@ucla.edu 2000 American Heart Association, Inc. Circulation is available at 878

2 Swerdlow et al AF Detection by ICD 879 Figure 1. A, P:R marker diagram illustrating AT/AF evidence counter. Upward markers indicate atrial events; downward markers indicate ventricular events. Long markers indicate nontachyarrhythmia events; short markers indicate tachyarrhythmia events. Numbers indicate change in counter ( Count) and value of counter (Count) after each RR interval. The counter creates increments if 2 atrial events exist per RR interval and if an RR interval with 1 atrial event follows an RR interval with 2 atrial events. Otherwise, it creates decrements. B, AT/AF autodiscrimination. AF and AT detection zones may overlap. If median atrial cycle length is in this overlap zone, atrial rhythm is classified as AT if it is regular or AF if it is not. After each sensed atrial event, atrial detection is blanked for 100 ms. AF detection zone extends from this value to programmable AF detection interval (AFDI). AT detection zone extends from programmable lower boundary (ATDI min ) to programmable AT detection interval (ATDI). AT/AF autodiscrimination zone extends from ATDI min to AFDI. Implant Procedure The ICD pulse generator and electrodes were inserted through a single left pectoral incision. A closely spaced (9.1 or 9.5 mm) right atrial bipole was used for atrial sensing. The P-wave amplitude in sinus rhythm was measured. A 2-lead defibrillation system (right atrium and right ventricle) was used in 74 patients, and a 3-lead system (including a coronary sinus electrode) was used in 6 patients. The electrode system was selected by investigator preference. Detection of AT/AF AT/AF was detected by a combination of median atrial cycle length and an AT/AF evidence counter that uses the number of sensed atrial electrograms in consecutive RR intervals. This counter provides high specificity for detecting AT with N:1 atrioventricular conduction (Figure 1A). The counter operates in 2 different modes for preliminary and sustained detection. Preliminary detection occurs when the count reaches 32 and the median cycle length for the last 12 PP intervals is less than the programmed AT or AF detection interval. Preliminary detection initiates an AT/AF episode. The episode timer begins, and the evidence counter switches to the sustained-detection mode in which AT/AF remains detected if the count is 27 and the atrial rate criterion remains fulfilled (Figure 2). Termination of Device-Defined AT/AF Episodes A device-defined atrial arrhythmia episode ends with the detection of 5 consecutive beats of sinus- or atrial-paced rhythm or 3 minutes of unclassified atrial rhythm. Sinus rhythm is defined by a previously described dual-chamber algorithm (PR Logic). 1,2 This algorithm is also used to discriminate between rapidly conducted AT/AF and VT/VF for the purpose of withholding ventricular therapy. Accuracy of this discrimination was not analyzed in the present study. The AT/AF detection algorithm studied in this report was used only for delivering or withholding AT/AF therapy, not to discriminate AT/AF from VT/VF. Far-Field R-Wave Discrimination To ensure the accurate sensing of atrial rate and rhythm, the ICD has minimal postventricular atrial blanking. Minimal blanking and autoadjusting atrial sensitivity may result in inappropriate atrial sensing of far-field R waves. The detection algorithm discriminates far-field R waves from AT/AF using the following 3 criteria: exactly 2 atrial events in each RR interval, a stable and short interval between the R wave on the ventricular channel and the far-field R wave on the atrial channel, and stable alternation in atrial intervals. If all 3 criteria are met, the atrial electrogram after the R wave is classified as a far-field R wave and excluded from the AT/AF evidence count (Figure 3). 2,3 AT/AF Autodiscrimination AF and AT detection zones may overlap to permit antitachycardia pacing for fast, regular atrial arrhythmias (Figure 1B). If the atrial cycle length is in this overlap zone, the atrial rhythm is classified as AT if it is regular and AF if it is not. ICD Programming Automatic therapies for AT/AF were programmed using a randomized, crossover design: on or off for the first 3 months and the opposite setting for the next 3 months. Thus, not all episodes of spontaneous AT/AF were treated. Holter Recordings Digital Holter recordings were performed in 59 patients. Recordings included 1 ECG lead, a telemetered atrial electrogram, and atrioventricular pace/sense markers. In 14 patients, Holter recordings were used to mark the time for the initial detection of AF in a postoperative electrophysiologic study. In 58 patients, a total of 80 Holter recordings were performed during follow-up both to validate the continuous detection of spontaneous AF and to determine the sensitivity of the detection algorithm for AT/AF. Patients who had frequent AT/AF, AF in progress, or inappropriate detection of AT/AF were selected for multiple recordings. Follow-Up Patients were followed for a total of 491 patient-months from the date of implant until the date of study closure or patient death. Five patients died during follow-up, 4 from heart failure and 1 from cancer. The mean duration of follow-up was 6 2 months (median, 6 months; range, 1 to 11 months). ICDs were interrogated at 1 month and 3 months, every 3 months thereafter, and whenever patients reported shocks or palpitations. Data Analysis Appropriate versus inappropriate detection of spontaneous AT/AF was determined from electrograms and intervals stored before therapy. Because the ICD does not store electrograms or intervals for untreated AT/AF episodes, only treated episodes were analyzed for the group as a whole. Continuous versus intermittent detection of ongoing, sustained AT/AF was determined from intervals stored before the termination of device-defined episodes. Termination was judged appropriate when the P:R pattern showed sinus rhythm or DDI pacing. Inappropriate episode termination caused by atrial undersensing was suspected if 50% of intervals before episode termination were shorter than the detection interval corresponding to that episode (AT or AF). A pattern of repetitive detections and terminations was used to identify inappropriate episode termination. Device-defined sustained AT/AF was classified as continuous if episode termination was appropriate and as intermittent if episode termination was inappropriate. Data are presented as means 1SD. P 0.05 was used to reject the null hypothesis. Basic comparisons were made using the paired t-test, unpaired t-test, Mann-Whitney U test, or chi-squared test. Results Patient characteristics are summarized in Table 1. The acute, sinus-rhythm P-wave amplitude at implantation was mv. The programmed minimum value of the autoadjusted atrial sensitivity was mv (nominal, 0.3 mv). It was 0.15 mv in 2 patients, 0.3 mv in 63 patients, 0.45 mv in

3 880 Circulation February 29, 2000 Figure 2. Holter recording shows detection of spontaneous AF. ECG, telemetered atrial electrogram (AEGM), and marker channel are shown. See Figure 1A for explanation of markers. Two lines below marker channel indicate median PP interval for last 12 atrial events and value of AT/AF evidence counter. Top shows onset of AF (arrow), and bottom shows AF detection (arrow). Panels are not continuous; bottom begins 10 s after top ends. Intermediate-height atrial markers (AS) correspond to sensed sinus P waves. Short markers (AR) indicate atrial events in postventricular atrial refractory period for pacing. Short double markers (FS) indicate intervals below programmed AF detection interval of 270 ms. Intermediate-height ventricular markers (VS) correspond to sensed R waves; tall ventricular markers (VP) correspond to ventricular paced events. Sinus rhythm is present at beginning of top panel. Atrial electrogram shows a low-amplitude far-field R wave that is not sensed. AT/AF evidence counter remains at 0 for first 4 QRS complexes because a single atrial event exists in preceding RR interval. There are 3 atrial events in RR interval between fourth and fifth complexes, and counter first creates increments on fifth complex. Absence of postventricular atrial blanking permits sensing of first AF electrogram, which follows sensed ventricular event by only 30 ms. Intermittent undersensing occurs at beginning of bottom panel. Counter creates decrements from 28 to 27 on fourth QRS complex and from 29 to 28 on seventh complex. Detection of AF occurs when count reaches 32 (arrow) and PP median is less than AF detection interval. This starts AT/AF duration timer and begins atrial episode designated by *A. Short triple atrial markers indicate ongoing AT/AF episode. 7 patients, 0.6 mv in 5 patients, and 0.9 mv in 3 patients. The programmed atrial cycle length below which arrhythmias were classified as AT or AF were ms and ms, respectively. Initial Detection of AF The initial detection of AF was analyzed during an electrophysiologic study in 14 patients for 129 induced and 36 spontaneous episodes of AF (Figure 2). The time for initial detection was 18 4 s. The median PP interval was ms. Spontaneous AT/AF During follow-up, 31 patients had episodes of spontaneous AT/AF that were detected and treated: AF only occurred in 13 patients, AT only in 8 patients, and both in 10 patients. Twenty-eight patients had only appropriate detections, 3 had only inappropriate detections, and 2 patients had both. Detection was appropriate in 88% of 190 AT episodes, and 98% of 132 AF episodes. See Table 2 for details. The longest inappropriate episode lasted 9.5 minutes. The 27 inappropriate AT/AF detections resulted in 29 antitachycardia pacing therapies; 28 were asymptomatic, and 1 induced AF, which was terminated by atrial defibrillation. Figure 4 shows a histogram of the median atrial cycle length for treated AT and AF episodes. The median cycle length was 240 ms in 84% of appropriate AF episodes, but it was 250 ms for all 3 inappropriate AF episodes (P 0.01). Of the 128 AT episodes with a median cycle length 300 ms, 98% were appropriate. However, 22 of 24 inappropriate AT episodes (92%) had a median cycle length 300 ms. Overall, the median cycle length was longer for inappropriately detected episodes than for appropriately detected episodes (AT, versus ms, P 0.001; AF, versus ms, P 0.01). Inappropriate Detection of AT/AF Misclassification of far-field R waves caused all inappropriate detections. These occurred most commonly because of conducted premature atrial complexes that changed the pattern of the stable alternation of PP intervals (n 19, 2 patients;

4 Swerdlow et al AF Detection by ICD 881 Figure 3. Holter recording shows correct classification of far-field R waves. ECG, atrial electrogram (AEGM), marker, median PP interval, and AT/AF evidence counter are shown, as in Figure 2. Additional far-field R line identifies RR intervals in which counter does not create increments because atrial electrogram is classified as a far-field R wave (FR). TF on atrial marker channel denotes intervals shorter than programmed AT detection interval of 270 ms. Other abbreviations as in Figure 2. TF intervals are classified correctly as far-field R waves because of stable PR interval, stable RP interval, and stable alternation of PP intervals with short intervals, including R wave. Figure 5A). A second cause was DDI pacing, which resulted in unstable PR intervals (n 6, 1 patient; Figure 5B). Other causes included variations in the intervals between the right ventricular electrogram and the far-field R-wave electrogram (n 1) and first-degree atrioventricular block in sinus tachycardia, resulting in equal PR and R to far-field R intervals (n 1). 3 In all cases, the short duration of inappropriate detection was due to rapid redetection of the sinusrhythm pattern or correct classification of far-field R waves. Inappropriate detection of far-field R waves was corrected in 3 patients by decreasing atrial sensitivity. No parameters were reprogrammed in 2 patients who had no subsequent inappropriate detections. Detection of Ongoing AT/AF Of the 295 appropriate episodes of spontaneous AT/AF, 294 (99.7%) were detected continuously (Figure 6). One episode TABLE 1. n 80. Patient Characteristics Age, y (mean SD) Sex, male/female 63/17 Cardiac disease Coronary artery disease 43 Myocardial or valvular disease 35 No structural disease 2 Atrial arrhythmia AF 50 AT 5 AF and AT 25 Antiarrhythmic drugs at implantation Amiodarone 21 Digoxin 14 Sotalol 5 -Blockers 2 Type 1 4 Verapamil 2 was detected continuously for 10 days. There were 90 AT/AF episodes 1 hour in 19 patients, with a total duration of 2697 hours. One episode of asymptomatic AF was undersensed and intermittently underdetected because the atrial sensitivity had been set to 0.9 mv to prevent oversensing of far-field R waves. Undersensing was corrected by reprogramming atrial sensitivity. No patient had symptomatic, undetected AT/AF. Holter Recordings Of the 80 Holter recordings made during follow-up, 1406 hours had a suitable quality of both telemetered atrial electrograms and surface ECG for analysis (18 7 hours of recording). Tables 3 through 5 summarize all Holter-recorded episodes of AT/AF that satisfied the minimum duration corresponding to the initial detection criterion (32 RR intervals). All 120 AT and 26 AF episodes that satisfied the programmed atrial rate criterion were detected continuously, for a total of 116 hours. Three episodes of AT lasting a total of 3 minutes were not detected because the AT cycle length exceeded the programmed detection interval. Holter-recorded AF electrograms had a wide variation in slew rate and amplitude, as shown in Figures 2, 5, and 6. TABLE 2. Detected and Treated Spontaneous AT/AF AT AF Total episodes Appropriate 166 (88%) 129 (98%) Inappropriate 24 (12%) 3 (2%) Cycle length, ms Duration, min Appropriate episodes ( ) (1 7317) Inappropriate episodes ( )* ( ) Episodes are mean SD (range). *Excluding 1 patient in whom the AT detection interval was 450 ms, the longest inappropriate AT episode was 5 min.

5 882 Circulation February 29, 2000 Figure 4. Histogram of median atrial cycle length for treated AT and AF episodes. Numerical values show upper boundary of each bin. For example, 180-ms bin includes range of values from 160 to 179 ms. See text for details. A total of 1290 hours of Holter recordings were made during which the atrial rhythm was sinus- or atrial-paced. In this period, 6 inappropriate detections of AT/AF occurred, lasting a total of 7 minutes. Expressed in terms of duration of ICD-classified rhythms, sensitivity for the detection of AT/AF that satisfied detection criteria according to Holter analysis was 100%, whereas specificity for the detection of rhythms other than AT/AF was 99.99%. AT/AF Discrimination Of the 166 spontaneous, appropriate, and treated AT episodes, 116 (70%) had a cycle length in the AT/AF overlap zone ( ms), and 50 (30%) had a cycle length more than the AF detection interval ( ms) (Figure 7). No significant difference existed between the success rate of antitachycardia pacing in the overlap zone (49 of 116 episodes, 42%) and the success rate in the slower AT zone (26 of 50 episodes, 52%; P 0.34). Discussion This study demonstrates that a new dual-chamber ICD provides both rapid and continuous detection of AT/AF and discriminates accurately between AF and AT. Goals of AT/AF Detection Antiarrhythmic devices detect AT/AF to perform 3 functions: mode switching to prevent inappropriate tracking of AT/ AF, 4,5 withholding inappropriate ventricular therapy, 1,6 and delivery of atrial therapy. 7,8 The dual-chamber ICD evaluated in the present study performs all 3 functions. It is the first ICD that permits therapy for long-duration AT/AF while withholding therapy from self-terminating AT/AF. It is, thus, the first ICD that must detect low and varying fibrillation electrograms continuously for extended periods. It also discriminates between AT and AF to deliver antitachycardia pacing for AT; thus, it must determine the atrial rate and rhythm accurately. AT/AF Detection in Other Devices Dual-chamber pacemakers and some ICDs do not require accurate determination of atrial rate and rhythm. They have postventricular atrial blanking and refractory periods that prevent atrial oversensing of far-field R waves 9 but Figure 5. Inappropriate detection of AT/AF. Holter recordings have same format as Figure 3. When atrial episode is in progress, Episode line displays a for every ventricular event. A, Inappropriate mode switch (arrow) from DDD to DDI caused by far-field R wave oversensing. B, DDI mode facilitates incorrect classification of far-field R waves by altering PR patterns. Approximately 9 minutes after mode switch to DDI, inappropriate detection of AT occurs (arrow) when AT/AF evidence counter is 32 and median PP interval is less than programmed AT detection interval of 350 ms. C, About 20 s later, 5 consecutive AP-VP or AP-VS cardiac cycles (arrow) result in episode termination. AP indicates atrial-paced; TF, interval in AF zone; and other abbreviations as in Figure 2. result in undersensing of AT/AF, particularly during high ventricular rates and AT with 2:1 atrioventricular conduction. 5 An atrial ICD 10 detects AF intermittently. Thus, continuous AF cannot be distinguished from sequential, self-terminating episodes. Atrial electrograms are analyzed only during the electrocardiographic ST segment, so the precise rate and

6 Swerdlow et al AF Detection by ICD 883 Figure 6. Holter recording illustrates continuous detection of AF. A shows onset of AF and initial detection of AF after 32 ventricular events (arrow). Double atrial markers (FS) change to triple markers (FD) and *A symbol appears on atrial evidence channel to indicate AF episode in progress. B shows rhythms after 5 hours of continuous recording. C shows spontaneous termination of AF after almost 8 hours. VP indicates ventricular-paced events; VS, sensed R waves; and AP, atrial-paced. regularity of the atrial rhythm cannot be determined. AT is not detected, AF is not detected during ventricular-paced rhythm, and a coronary sinus electrode is required for sensing. Wellens et al 8 reported that the sensitivity for the detection of AF was 92%. AT/AF Detection and Discrimination in the Jewel AF Several features combined to permit continuous detection of AF for extended periods. A closely spaced, atrial sensing bipole, 11 autoadjusting sensitivity with a short time constant, and minimal atrial blanking all minimize atrial undersensing. The detection algorithm continues to detect AF, despite some undersensing and brief periods of unclassified atrial rhythm. Determining the atrial rate and rhythm accurately requires that postventricular, atrial blanking be minimized to a value insufficient to blank far-field R waves. Thus, the detection algorithm must prevent the sensing of far-field R waves from causing inappropriate detection of AT/AF. Accuracy of AT/AF Detection The present study describes the initial, multicenter experience with this ICD, during which the importance of minimizing far-field R waves was not recognized uniformly. Nevertheless, the true-positive detection rate was 98% for AF and 98% for AT, with a detection interval 300 ms. All inappropriate detections with a median atrial cycle length 320 ms lasted 5 minutes. However, an AT detection interval 300 ms facilitates inappropriate detection by increasing the probability that inappropriately classified far-field R-wave intervals will reduce the measured atrial cycle length below the AT detection interval. All sustained episodes of AF were detected continuously when atrial sensitivity was 0.6 mv. Holter monitoring with telemetered atrial electrograms confirmed the 100% sensitiv-

7 884 Circulation February 29, 2000 TABLE 3. Appropriately-Detected AT/AF: Total Duration per Holter Recording Holter Recording No. of Episodes Median PP Interval, ms Duration per Holter, min AT AF 55004b c 1* c a 1* a a b b d a 1* b 1* a 1* b b Total (115.9 hr) *Continuous AF throughout Holter. Recordings from the same patient have the same number. Multiple recordings in the same patient are designated by differing letters. ity for continuous detection of AT/AF and the high specificity for the rejection of non-at/af atrial rhythms. Discrimination of AT and AF The Jewel AF classifies the atrial rhythm as AT if the atrial rate is in the AT zone or if the atrial rhythm is regular and the atrial rate is in the AT/AF overlap zone. The success rates for antitachycardia pacing were similar in the 2 zones, which permitted painless termination of some regular atrial arrhythmias in the AF rate zone. Limitations The study had some limitations. (1) The sensitivity for the detection of AT/AF could not be evaluated for the entire population. (2) Continuous detection of AT/AF was inferred for most episodes and validated only for those recorded by Holter monitoring. (3) AT/AF was classified on the basis of rate and regularity of bipolar atrial electrograms. Some episodes classified as AT may have been electrophysiological AF. 12 (4) The study protocol did not require repositioning the TABLE 4. Inappropriately-Detected AT/AF: Individual Episodes Holter Recording Rhythm Median PP Interval, ms Duration, min s 55100a AT a AT a AT a AT a AF a AF Total 7 21 TABLE 5. Undetected AT/AF Episodes With 32 RR Intervals: Individual Episodes* Holter Recording Rhythm Median PP Interval, ms Duration, min s 5527a AT a AT a AT Total 3 03 *AT not detected because median PP interval was more than programmed AF detection interval of 280 ms. atrial electrodes during implantation to avoid sensing far-field R waves or recording the amplitude of far-field R waves. Clinical Implications This study clarifies several principles for the optimal application of atrial ICDs that have minimal atrial blanking periods. (1) The atrial electrode should be positioned to minimize far-field R waves. (2) The AT/AF detection interval should be set slightly greater than the anticipated AT/AF interval. In contrast, optimal programming for DDD pacemakers or dual-chamber ICDs requires setting the modeswitch or AT/AF rate threshold slightly above the maximum sinus rate to compensate for atrial blanking and refractory periods. (3) The AT detection interval should be 300 ms to minimize inappropriate detection due to far-field R waves. It may be 300 ms only if no atrial sensing of far-field R waves occurs. (4) Inappropriate detection of AT/AF caused by atrial oversensing of far-field R waves is transient, and it need not result in inappropriate therapy. Restricting therapy to episodes 10 minutes prevents inappropriate AT/AF therapy and does not delay VT/VF therapy. Attention to the first 3 principles should minimize inappropriate detection, whereas attention to the fourth should minimize inappropriate AT/AF therapy in the event of inappropriate detection. Conclusions AT/AF can be detected rapidly and continuously by a dualchamber, atrial ICD. Thus, shocks may be programmed for long-duration AT/AF without the risk of inappropriate episode termination due to atrial undersensing. Discrimination of AT versus AF on the basis of the regularity of the atrial rhythm permits successful, painless antitachycardia pacing for some rhythms that would otherwise require painful cardioversion. Appendix Participating Investigators and Medical Centers Michael Block, Klinik d Westf Wilhelmsuniversitat, Munster, Germany; Johannes Brachmann, Klinikum d Reprecht-Karls-Universitat, Heidelberg, Germany; George Crossley, North Carolina Baptist Hospital, Winston-Salem, NC; Karel den Dulk, Academisch Ziekenhuis Maastricht, Maastricht, the Netherlands; Jeronimo Farre, Funcacion Jimenez Diaz, Madrid, Spain; Martin Fromer, CHUV Hospital, Lausanne, Switzerland; Charles Haffajee, St Elizabeth s Hospital, Boston, Mass; Ellen Hoffmann, Klinikum Grosshadern, Munchen, Germany; Luc Jordaens, Academisch Ziekenhuis Gent, Gent, Belgium; Werner Jung, Med Universitatsklinik-Bonn, Bonn, Germany; Bruce Lerman, New York Hospital, New York; Randy Lieberman, Alta Bates Medical Center, Berkeley, Calif; Ole-Jorgen Ohm, Haukeland Sykehus, Bergen, Norway; Antonio Pacifico, Methodist Hospital Houston, Houston,

8 Swerdlow et al AF Detection by ICD 885 Figure 7. Discrimination of AT from AF is based on both rate and regularity of P-P intervals, as shown in this Holter recording. Atrial episode is in progress (*A symbol on Atrial Evidence channel and triple atrial markers). At beginning of tracing, rhythm is classified as AF (short triple markers, FD). Between fifth and sixth QRS complexes, electrograms change morphology, and atrial rhythm slows and becomes regular. Rhythm is classified as AT (tall triple markers; TD, arrow) in autodiscrimination zone (170 to 270 ms) when P-P median exceeds lower boundary of AT zone and rhythm remains regular. VS indicates sensed R waves in sinus zone; TS, sensed R waves in ventricular tachycardia zone. Texas; Massimo Santini, San Filippo Neri Hospital, Roma, Italy; Herwig Schmidinger, Allgemeines Krankenhaus Wien, Vienna, Austria; Karl Heinz Seidl, Klinikum Ludwigshafen, Ludwigshafen, Germany; Marshall Stanton, Mayo Clinic, Rochester, Minn; Charles Swerdlow, Cedars-Sinai Medical Center, Los Angeles, Calif. Acknowledgment We wish to thank Lucy Gonzales for manuscript preparation. References 1. Gillberg J, Brown M, Stanton M, Hammill S, Olson W. Clinical testing of a dual chamber atrial tachyarrhythmia detection algorithm. Comput Cardiol. 1996; Brown M, Gillberg J, Stanton M, Hammill S, Olson W. Acute human testing of a dual chamber ventricular tachyarrhythmia detection algorithm at Mayo Clinic. Comput Cardiol. 1996; Wolpert C, Jung W, Scholl C, Spehl S, Cyran J, Luderitz B. Electrical proarrhythmia: induction of inappropriate atrial therapies due to far-field R wave oversensing in a new dual chamber defibrillator. J Cardiovasc Electrophysiol. 1998;9: Bonnet J, Brusseau E, Limousin M, Cazeau S. Mode switch despite undersensing of atrial fibrillation in DDD pacing. Pacing Clin Electrophysiol. 1996;19: Kamalvand K, Tan K, Kotsakis A, Bucknall C, Sulke N. Is mode switching beneficial? A randomized study in patients with paroxysmal atrial tachyarrhythmias. J Am Coll Cardiol. 1997;30: Nair M, Saoudi N, Kroiss D, Letac B. Automatic arrhythmia identification using analysis of the atrioventricular association. Circulation. 1997;95: Lau CP, Tse HF, Lok NS, Lee KL, Ho DS, Sopher M, Murgatroyd F, Camm AJ. Initial clinical experience with an implantable human atrial defibrillator. Pacing Clin Electrophysiol. 1997;20: Wellens HJ, Lau CP, Luderitz B, Akhtar M, Waldo AL, Camm AJ, Timmermans C, Tse HF, Jung W, Jordaens L, Ayers G. Atrioverter: an implantable device for the treatment of atrial fibrillation. Circulation. 1998;98: Olson W. Dual chamber sensing and detection for ICDs. In: Singer I, Barold SS, Camm AJ, eds. Implantable Cardioverter Defibrillator. 2nd ed. Armonk: Futura; 1998: Kim J, Bocek J, White H, Crone B, Alferness C, Adams J. An atrial fibrillation detection algorithm for an implantable atrial defibrillator. Comput Cardiol. 1995; Brouwer J, Nagelkerke D, den Heijer P, Ruiter J, Mulder H, Begemann M, Lie K. Analysis of atrial sensed far-field ventricular signals: a reassessment. Pacing Clin Electrophysiol. 1997;20: Wells JL, Karp R, Kouchoukos N, Maclean W, James T, Waldo A. Characterization of atrial fibrillation in man: studies following open heart surgery. Pacing Clin Electrophysiol. 1978;1:

The incidence of paroxysmal atrial fibrillation (AF) in patients

The incidence of paroxysmal atrial fibrillation (AF) in patients Atrial Therapies Reduce Atrial Arrhythmia Burden in Defibrillator Patients Paul A. Friedman, MD; Barbara Dijkman, MD; Eduardo N. Warman, PhD; H. Amy Xia, PhD; Rahul Mehra, PhD; Marshall S. Stanton, MD;

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators 22 March 2002 Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators F. HINTRINGER, O. PACHINGER Division of Cardiology, Department for Internal Medicine, University

More information

Critical Analysis of Dual-Chamber Implantable Cardioverter-Defibrillator Arrhythmia Detection

Critical Analysis of Dual-Chamber Implantable Cardioverter-Defibrillator Arrhythmia Detection Critical Analysis of Dual-Chamber Implantable Cardioverter-Defibrillator Arrhythmia Detection Results and Technical Considerations Bruce L. Wilkoff, MD; Volker Kühlkamp, MD; Kent Volosin, MD; Kenneth Ellenbogen,

More information

A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support

A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support 100 April 2000 A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support A. SH. REVISHVILI A.N. Bakoulev Institute of Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow,

More information

Tachyarrhythmia Suspicion and Detection

Tachyarrhythmia Suspicion and Detection Tech Corner Tachyarrhythmia Suspicion and Detection NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN THE ARTICLE.

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features Definition of SVT Discrimination Identify which patient populations might benefit from these features Understand the 4 types of SVT discriminators used by St Jude Medical Be aware of programmable parameters

More information

The Nuts and Bolts of ICD Therapy

The Nuts and Bolts of ICD Therapy Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction

More information

Dual-Chamber Implantable Cardioverter-Defibrillator

Dual-Chamber Implantable Cardioverter-Defibrillator February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary

More information

Teaching Rounds in Cardiac Electrophysiology

Teaching Rounds in Cardiac Electrophysiology Teaching Rounds in Cardiac Electrophysiology Sustained Multiple Railroad Tracks on Implantable Cardiac Defibrillator Interval Plots Mechanisms and Management Alex Y. Tan, MD; Kenneth A. Ellenbogen, MD;

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

Initial Clinical Experience With Ambulatory Use of an Implantable Atrial Defibrillator for Conversion of Atrial Fibrillation

Initial Clinical Experience With Ambulatory Use of an Implantable Atrial Defibrillator for Conversion of Atrial Fibrillation Initial Clinical Experience With Ambulatory Use of an Implantable Atrial Defibrillator for Conversion of Atrial Fibrillation Emile G. Daoud, MD; Carl Timmermans, MD; Chris Fellows, MD; Robert Hoyt, MD;

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm April 2000 107 Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm B. MERKELY Semmelweis University, Dept. of Cardiovascular Surgery,

More information

Early Recurrence of Atrial Fibrillation After Ambulatory Shock Conversion

Early Recurrence of Atrial Fibrillation After Ambulatory Shock Conversion Journal of the American College of Cardiology Vol. 40, No. 1, 2002 2002 by the American College of Cardiology Foundation ISSN 0735-1097/02/$22.00 Published by Elsevier Science Inc. PII S0735-1097(02)01912-5

More information

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Programming of Bradycardic Parameters C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Programming of ICD Brady Parameters Conflict of Interest Biotronik

More information

Dual chamber pacing mode in an atrial antitachycardia pacing device without a ventricular lead e A necessary evil

Dual chamber pacing mode in an atrial antitachycardia pacing device without a ventricular lead e A necessary evil indian pacing and electrophysiology journal 15 (2015) 133e137 HOSTED BY Available online at www.sciencedirect.com ScienceDirect journal homepage: www.elsevier.com/locate/ipej Dual chamber pacing mode in

More information

PARAD/PARAD+ : P and R Based Arrhythmia Detection

PARAD/PARAD+ : P and R Based Arrhythmia Detection Tech Corner PARAD/PARAD+ : P and R Based Arrhythmia Detection NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN

More information

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? CASE REPORTS Arrhythmia 2015;16(3):173-177 doi: http://dx.doi.org/10.18501/arrhythmia.2015.029 Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? Eun-Sun Jin, MD, PhD Cardiovascular

More information

Shock Reduction Strategies Michael Geist E. Wolfson MC

Shock Reduction Strategies Michael Geist E. Wolfson MC Shock Reduction Strategies Michael Geist E. Wolfson MC Shock Therapy Thanks, I needed that! Why Do We Need To Reduce Shocks Long-term outcome after ICD and CRT implantation and influence of remote device

More information

Advanced ICD Troubleshooting: Part I REVIEW

Advanced ICD Troubleshooting: Part I REVIEW REVIEW Advanced ICD Troubleshooting: Part I CHARLES D. SWERDLOW* and PAUL A. FRIEDMAN *From the Cedars-Sinai Medical Center, Los Angeles, California, and Mayo Clinic, Rochester, Minnesota (PACE 2005;28:1322

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Implantable cardioverter defibrillators for the treatment of arrhythmias and cardiac resynchronisation therapy for the treatment of heart failure (review

More information

Introduction to a new "X out of Y" Mode-Switching Algorithm in Inos² CLS and Logos Dual-Chamber Pacemakers

Introduction to a new X out of Y Mode-Switching Algorithm in Inos² CLS and Logos Dual-Chamber Pacemakers April 1999 117 Introduction to a new "X out of Y" Mode-Switching Algorithm in Inos² CLS and Logos Dual-Chamber Pacemakers C. W. ISRAEL Department of Cardiology, St.-Josef-Hospital, Ruhr-University, Bochum,

More information

Tech Corner. ATP in the Fast VT zone

Tech Corner. ATP in the Fast VT zone Tech Corner ATP in the Fast VT zone NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN THE ARTICLE. FOR ADDITIONAL

More information

Inappropriate electrical shocks: Tackling the beast

Inappropriate electrical shocks: Tackling the beast ESC Paris 2011 Inappropriate electrical shocks: Tackling the beast Gerhard Hindricks University of Leipzig Heart Center Dept. of Electrophysiology ESC Paris 2011 Inappropriate electrical shocks: Tackling

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

Journal of Arrhythmia

Journal of Arrhythmia Journal of Arrhythmia 29 (2013) 204 210 Contents lists available at ScienceDirect Journal of Arrhythmia journal homepage: www.elsevier.com/locate/joa Original Article Factors involved in correct analysis

More information

Contents. Example: Pacemaker in DDD Mode. I. Introduction II. Website and Relevant Documents III. Background Information IV.

Contents. Example: Pacemaker in DDD Mode. I. Introduction II. Website and Relevant Documents III. Background Information IV. Contents I. Introduction II. Website and Relevant Documents III. Background Information IV. Pacemaker System V. i. Components ii. Operating Modes iii. Parameters Example: Pacemaker in DDD Mode Pacemaker

More information

CONSULTA CRT-P Model C4TR01

CONSULTA CRT-P Model C4TR01 CONSULTA CRT-P Model C4TR01 Physical characteristics Volume a 15 cm 3 Mass H x W x D Surface area of titanium device can Surface area of each titanium nitride LECG electrode Radiopaque ID Materials in

More information

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study ANEESH V. TOLAT, M.D.,* MELISSA WOICIECHOWSKI, M.S.N.,* ROSEMARIE KAHR, R.C.I.S.,* JOSEPH

More information

Device Diagnostics and Stroke Prevention: State of the Art

Device Diagnostics and Stroke Prevention: State of the Art Device Diagnostics and Stroke Prevention: State of the Art Ass. Prof. Dr. Carsten W. Israel Dept. of Cardiology Evangelical Hospital Bielefeld Germany C.W.Israel@em.uni-frankfurt.de Disclosure of Relationships

More information

Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm

Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm Guy Amit, MD, MPH Soroka University Medical Center Ben-Gurion University of the Negev Beer-Sheva, Israel Disclosures Consultant:

More information

Review Article Automatic mode switching in atrial fibrillation

Review Article Automatic mode switching in atrial fibrillation www.ipej.org 186 Review Article Automatic mode switching in atrial fibrillation Giuseppe Stabile, MD, Antonio De Simone, MD, Enrico Romano* Laboratorio di Elettrofisiologia, Casa di Cura San Michele, Maddaloni

More information

Document Version Publisher s PDF, also known as Version of Record (includes final page, issue and volume numbers)

Document Version Publisher s PDF, also known as Version of Record (includes final page, issue and volume numbers) Initiation of ventricular tachycardia by interruption of pacemaker-mediated tachycardia in a patient with a dualchamber implantable cardioverter defibrillator Van Gelder, B.M.; Meijer, A.; Dekker, L.R.C.;

More information

A 'shocking' case rectified

A 'shocking' case rectified A 'shocking' case rectified Journal: Pacing and Clinical Electrophysiology Manuscript ID: PACE-13-00.R1 Wiley - Manuscript type: Device Rounds Date Submitted by the Author: n/a Complete List of Authors:

More information

Advances in Arrhythmia and Electrophysiology

Advances in Arrhythmia and Electrophysiology Advances in Arrhythmia and Electrophysiology Troubleshooting Implanted Cardioverter Defibrillator Sensing Problems I Charles D. Swerdlow, MD; Samuel J. Asirvatham, MD; Kenneth A. Ellenbogen, MD; Paul A.

More information

Biatrial Stimulation and the Prevention of Atrial Fibrillation

Biatrial Stimulation and the Prevention of Atrial Fibrillation 374 September 2001 Biatrial Stimulation and the Prevention of Atrial Fibrillation L. MELCZER Heart Institute, Faculty of Medicine, University of Pécs, Pécs, Hungary I. LORINCZ 1 st Internal Department,

More information

Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations

Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations Journal of Geriatric Cardiology (2018) 15: 193 198 2018 JGC All rights reserved; www.jgc301.com Case Report Open Access Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

ICD Discrimination Algorithms

ICD Discrimination Algorithms ICD Discrimination Algorithms How Do They Work? Ji-Hae Yun, RN Certified Cardiac Device Specialist by IBHRE Asan Medical Center, Seoul Korea EGM of A.Fib with fast ventricular response Case 1 A EGM V EGM

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Recent studies suggest that atrial fibrillation (AF) is not a

Recent studies suggest that atrial fibrillation (AF) is not a Atrial Fibrillation Detection and R-Wave Synchronization by Metrix Implantable Atrial Defibrillator Implications for Long-Term Efficacy and Safety Hung-Fat Tse, MB, BS; Chu-Pak Lau, MD; Jasbir S. Sra,

More information

Sustained tachycardia with wide QRS

Sustained tachycardia with wide QRS Sustained tachycardia with wide QRS Courtesy from Prof. Antonio Américo Friedmann. Electrocardiology Service of University of Faculty of São Paulo. Opinions from colleagues Greetings to everyone, In a

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

Mechanism of Immediate Recurrences of Atrial Fibrillation After Restoration of Sinus Rhythm

Mechanism of Immediate Recurrences of Atrial Fibrillation After Restoration of Sinus Rhythm Mechanism of Immediate Recurrences of Atrial Fibrillation After Restoration of Sinus Rhythm AMAN CHUGH, MEHMET OZAYDIN, CHRISTOPH SCHARF, STEVE W.K. LAI, BURR HALL, PETER CHEUNG, FRANK PELOSI, JR, BRADLEY

More information

December D. EL ALLAF Centre Hospitalier Hutois, Huy, Belgium. P. ATTUEL Centre Chirurgical Val d'or, Saint Cloud, France

December D. EL ALLAF Centre Hospitalier Hutois, Huy, Belgium. P. ATTUEL Centre Chirurgical Val d'or, Saint Cloud, France December 2000 449 European Multicenter Study on the Prevention of Paroxysmal Atrial Fibrillation by Permanent Overdrive Pacing: Atrial Rate Behavior and Patient Tolerance D. EL ALLAF Centre Hospitalier

More information

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular

More information

Ventricular tachycardia Ventricular fibrillation and ICD

Ventricular tachycardia Ventricular fibrillation and ICD EKG Conference Ventricular tachycardia Ventricular fibrillation and ICD Samsung Medical Center CCU D.I. Hur Ji Won 2006.05.20 Ventricular tachyarrhythmia ventricular tachycardia ventricular fibrillation

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

A Closer Look Product Education at a glance

A Closer Look Product Education at a glance A Closer Look Product Education at a glance an ICD or CRT-D When Atrial Information is Not Used SUMMARY Boston Scientific dual-chamber ICDs and multi-chamber CRT-Ds will respond to atrial sensing regardless

More information

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.

More information

Friedman, Rott, Wokhlu, Asirvatham, Hayes 201. Figure 65.7 Shortening of the AV interval during pacing.

Friedman, Rott, Wokhlu, Asirvatham, Hayes 201. Figure 65.7 Shortening of the AV interval during pacing. Friedman, Rott, Wokhlu, Asirvatham, Hayes 201 Figure.7 Shortening of the AV interval during pacing. 202 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization Figure.8 is obtained from

More information

Antiarrhythmic Drugs and Ablation in Patients with ICD and Shocks

Antiarrhythmic Drugs and Ablation in Patients with ICD and Shocks Antiarrhythmic Drugs and Ablation in Patients with ICD and Shocks Alireza Ghorbani Sharif, MD Interventional Electrophysiologist Tehran Arrhythmia Clinic January 2016 Recurrent ICD shocks are associated

More information

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia By Sandeep Joshi, MD and Jonathan S. Steinberg, MD Arrhythmia Service, Division of Cardiology

More information

Continuous ECG telemonitoring with implantable devices: the expected clinical benefits

Continuous ECG telemonitoring with implantable devices: the expected clinical benefits Continuous ECG telemonitoring with implantable devices: the expected clinical benefits C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Declaration of

More information

Introduction. CLINICAL RESEARCH Clinical Trial Design. Mohammad Saeed 1 *, Mehdi Razavi 1, Curtis G. Neason 2, and Simona Petrutiu 2. Aims.

Introduction. CLINICAL RESEARCH Clinical Trial Design. Mohammad Saeed 1 *, Mehdi Razavi 1, Curtis G. Neason 2, and Simona Petrutiu 2. Aims. Europace (2011) 13, 1648 1652 doi:10.1093/europace/eur195 CLINICAL RESEARCH Clinical Trial Design Rationale and design for programming implantable cardioverter defibrillators in patients with primary prevention

More information

First worldwide clinical experience with a new dual chamber implantable cardioverter defibrillator

First worldwide clinical experience with a new dual chamber implantable cardioverter defibrillator Europace (1999) 1, 96 102 First worldwide clinical experience with a new dual chamber implantable cardioverter defibrillator Advantages and complications C. Sticherling, A. Schaumann*, T. Klingenheben

More information

Advanced ICD Troubleshooting: Part II REVIEW

Advanced ICD Troubleshooting: Part II REVIEW REVIEW Advanced ICD Troubleshooting: Part II CHARLES D. SWERDLOW* and PAUL A. FRIEDMAN From the *Cedars-Sinai Medical Center, Los Angeles, California, and Mayo Clinic, Rochester, Minnesota, USA Failure

More information

Journal of the American College of Cardiology Vol. 38, No. 2, by the American College of Cardiology ISSN /01/$20.

Journal of the American College of Cardiology Vol. 38, No. 2, by the American College of Cardiology ISSN /01/$20. Journal of the American College of Cardiology Vol. 38, No. 2, 2001 2001 by the American College of Cardiology ISSN 0735-1097/01/$20.00 Published by Elsevier Science Inc. PII S0735-1097(01)01351-1 Prevalence,

More information

Clinical and Electrocardiographic Characteristics of Patients with Brugada Syndrome: Report of Five Cases of Documented Ventricular Fibrillation

Clinical and Electrocardiographic Characteristics of Patients with Brugada Syndrome: Report of Five Cases of Documented Ventricular Fibrillation J Arrhythmia Vol 25 No 1 2009 Original Article Clinical and Electrocardiographic Characteristics of Patients with Brugada Syndrome: Report of Five Cases of Documented Ventricular Fibrillation Seiji Takashio

More information

Case Description A 25 years old male served in a combat unit for 3 years (GOLANY) Implantation of Dual chamber ICD in 7/2010 due to Severe Non ischemi

Case Description A 25 years old male served in a combat unit for 3 years (GOLANY) Implantation of Dual chamber ICD in 7/2010 due to Severe Non ischemi ICD Case Study tracing your own images Prof Glikson, Sandra Schor-Wider Nadav Hayman Case Description A 25 years old male served in a combat unit for 3 years (GOLANY) Implantation of Dual chamber ICD in

More information

Performance of Atrial Fibrillation Detection in a New Single-Chamber ICD

Performance of Atrial Fibrillation Detection in a New Single-Chamber ICD DEVICES Performance of Atrial Fibrillation Detection in a New Single-Chamber ICD ABHISHEK DESHMUKH, M.D.,* MARK L. BROWN, PH.D., ELISE HIGGINS, B.S., BRIAN SCHOUSEK, B.S., ATHULA ABEYRATNE, PH.D., GIOVANNI

More information

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,

More information

John H. Ip, M.D FACC Associate Professor of Medicine Michigan State University Medical Director, EP Service

John H. Ip, M.D FACC Associate Professor of Medicine Michigan State University Medical Director, EP Service ICD Trouble shooting: Case study John H. Ip, M.D FACC Associate Professor of Medicine Michigan State University Medical Director, EP Service Bullet proof evidence: ICD does prevent sudden death Disclosures

More information

Three case studies have reported that patients have received

Three case studies have reported that patients have received Interactions Between Electronic Article Surveillance Systems and Implantable Cardioverter-Defibrillators William J. Groh, MD; Scott A. Boschee, BS; Erica D. Engelstein, MD; William M. Miles, MD; M. Erick

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

Life-saving shocks are the raisons d être of implantable

Life-saving shocks are the raisons d être of implantable Advances in Arrhythmia and Electrophysiology Minimizing Inappropriate or Unnecessary Implantable Cardioverter-Defibrillator Shocks Appropriate Programming Jayanthi N. Koneru, MBBS; Charles D. Swerdlow,

More information

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART December 2015 This chart encompasses all Medtronic cardiac devices FDA-Approved as MR Conditional and included in the MRI SureScan portfolio. If a device

More information

Diagnostic Value of Single Versus Dual Chamber Electrograms Recorded from an Implantable Defibrillator

Diagnostic Value of Single Versus Dual Chamber Electrograms Recorded from an Implantable Defibrillator Journal of Interventional Cardiac Electrophysiology 9, 49 53, 2003 C 2003 Kluwer Academic Publishers. Manufactured in The Netherlands. Diagnostic Value of Single Versus Dual Chamber Electrograms Recorded

More information

Implantable Cardioverter Defibrillator Shocks: A Troubleshooting Guide

Implantable Cardioverter Defibrillator Shocks: A Troubleshooting Guide TREATMENT UPDATE Implantable Cardioverter Defibrillator Shocks: A Troubleshooting Guide Charles D. Swerdlow, MD, Jianxin Zhang, MS Division of Cardiology, Cedars-Sinai Medical Center, Los Angeles, CA Implantable

More information

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment Karen L. Booth, MD, Lucile Packard Children s Hospital Arrhythmias are common after congenital heart surgery [1]. Postoperative electrolyte

More information

Bradycardia Pacing-Induced Short-Long-Short Sequences at the Onset of Ventricular Tachyarrhythmias

Bradycardia Pacing-Induced Short-Long-Short Sequences at the Onset of Ventricular Tachyarrhythmias Journal of the American College of Cardiology Vol., No., by the American College of Cardiology Foundation ISSN -//$. Published by Elsevier Inc. doi:./j.jacc... Heart Rhythm Disorders Bradycardia -Induced

More information

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Device Interrogation- Pacemakers, ICD and Loop Recorders Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Disclosures Consultant: Medtronic Speaker s Bureau: St. Jude Medical

More information

Preventive Pacing in Atrial Fibrillation

Preventive Pacing in Atrial Fibrillation 358 September 2001 Preventive Pacing in Atrial Fibrillation W. KAINZ Department of Cardiology, Hanusch Hospital, Vienna, Austria Summary Patients with paroxysmal atrial fibrillation are often highly symptomatic,

More information

Pacemaker-ICD/Drug Interaction

Pacemaker-ICD/Drug Interaction Review Article Pacemaker-ICD/Drug Interaction Tsu-Juey Wu When a drug is prescribed for a patient with a permanent pacemaker or an implantable cardioverter defibrillator (ICD), consideration must be given

More information

ICD Diagnostics: Making the most of it

ICD Diagnostics: Making the most of it ICD Basics and Beyond ICD Diagnostics: Making the most of it Dulce Obias-Manno RN, BSN, MHSA, FHRS, CCDS, CEPS Cardiac Arrhythmia Center, Device Clinic Washington Hospital Center, Washington DC Objectives

More information

ICD in a young patient with syncope

ICD in a young patient with syncope ICD in a young patient with syncope Konstantinos P. Letsas, MD, FESC Second Department of Cardiology Evangelismos General Hospital of Athens Athens, Greece Case presentation A 17-year-old apparently healthy

More information

Panagiotis N. Margos MD, Rolf Schomburg MD, Jorg Kynast MD, Ahmed A. Khattab MD, Gert Richardt MD.

Panagiotis N. Margos MD, Rolf Schomburg MD, Jorg Kynast MD, Ahmed A. Khattab MD, Gert Richardt MD. www.ipej.org 64 Case Report Termination of Ventricular Tachycardia with Antitachycardia Pacing after Ineffective Shock Therapy in an ICD Recipient with Hypertrophic Cardiomyopathy Panagiotis N. Margos

More information

Automatic Pacemaker Termination of Two Different Types of Supraventricular Tachycardia

Automatic Pacemaker Termination of Two Different Types of Supraventricular Tachycardia JACC Vol. 6, No. I 201 Automatic Pacemaker Termination of Two Different Types of Supraventricular Tachycardia KAREL DEN DULK, MD,* PEDRO BRUGADA, MD,* BERND WALDECKER, MD,* MALCOLM BEGEMANN, MSc,t TOM

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,800 116,000 120M Open access books available International authors and editors Downloads Our

More information

Tachycardia Devices Indications and Basic Trouble Shooting

Tachycardia Devices Indications and Basic Trouble Shooting Tachycardia Devices Indications and Basic Trouble Shooting Peter A. Brady, MD., FRCP Cardiology Review Course London, March 6 th, 2014 2011 MFMER 3134946-1 Tachycardia Devices ICD Indications Primary and

More information

Epidemiological data from the Framingham heart study indicate that the cumulative incidence

Epidemiological data from the Framingham heart study indicate that the cumulative incidence 106 * Electrophysiology PACING FOR ATRIAL FIBRILLATION c PHARMACOTHERAPY Correspondence to: Professor Chu-Pak Lau, Cardiology Division, University of Hong Kong, Queen Mary Hospital, Hong Kong; cplau@hkucc.hku.hk

More information

Practice Questions.

Practice Questions. IBHRE Prep Practice Questions Question 1 The relative refractory yperiod of the ventricular myocardium corresponds to which of the following phases of the action potential? A. (0) B. (1) C. (2) D. (3)

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Introduction. CLINICAL RESEARCH Syncope and implantable loop recorders

Introduction. CLINICAL RESEARCH Syncope and implantable loop recorders Europace (2013) 15, 1215 1222 doi:10.1093/europace/eut036 CLINICAL RESEARCH Syncope and implantable loop recorders Tachycardia detection performance of implantable loop recorders: results from a large

More information

Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention

Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Dr Ngai-Yin Chan, MBBS, FRCP(Lond), FRCP(Edin), FRCP(Glasg), FACC, FHRS, Consultant Physician, Department of Medicine & Geriatrics, Princess

More information

Troubleshooting ICD. NASPE Training Lancashire & South Cumbria Cardiac Network

Troubleshooting ICD. NASPE Training Lancashire & South Cumbria Cardiac Network Troubleshooting ICD NASPE Training Lancashire & South Cumbria Cardiac Network Bradycardia Pacing by ICD Isolated ICD discharge Multiple discharges Appropriate Inappropriate No ICD Therapy or delay in ICD

More information

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Francesco Zanon, MD, FESC, FHRS Arrhythmia and Electrophysiology Unit, Cardiology Dept. Santa Maria della Misericordia General

More information

Introduction. * Corresponding author. Tel: þ address:

Introduction. * Corresponding author. Tel: þ address: Europace (2006) 8, 1057 1061 doi:10.1093/europace/eul119 Performance of a new single-chamber ICD algorithm: discrimination of supraventricular and ventricular tachycardia based on vector timing and correlation

More information

Journal of Arrhythmia

Journal of Arrhythmia Journal of Arrhythmia 28 (2012) 91 95 Contents lists available at SciVerse ScienceDirect Journal of Arrhythmia journal homepage: www.elsevier.com/locate/joa Review Unresolved matters related to implantable

More information

Is This Thing Working?

Is This Thing Working? Is This *#@!* Thing Working? Pacemaker (and ICD) ECG and Telemetry Pitfalls Wayne O. Adkisson, MD adki0004@umn.edu Disclosures I currently receive research support from Medtronic, Inc. I have been compensated

More information

CONTAK RENEWAL 3 AVT CLINICAL SUMMARY

CONTAK RENEWAL 3 AVT CLINICAL SUMMARY CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY CONTAK RENEWAL 3 AVT Boston Scientific

More information

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Patient Scheduling and Screening Q: Is there any special (other than regular MRI consent) consent form that should be signed by the patient?

More information

a lecture series by SWESEMJR

a lecture series by SWESEMJR Electrolyte disturbances Hypokalaemia Decreased extracellular potassium increases excitability in the myocardial cells and consequently the effect of very severe hypokalaemia is ventricular arrhythmia.

More information