1 Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere
2 Disclosure Speaker s bureau Boehringer Ingelheim, MSD, Sanofi Consultant Fees/Honoraria Biosense Webster, Boehringer-Ingelheim, Meda Pharma, St Jude Medical, Sanofi Research support Biosense Webster, St Jude Medical 2
3 Modern pacemakers Accurate A & V sensing Physiological pacing On demand pacing Rate response Regulation of the rate Antitachycardia pacing Collection of diagnostic information
4 Goals of pacemaker diagnostics To collect more and better information on Function and activity of the device Patient (arrhythmias, underlying disease) Alleviate the burden of the device clinics (faster follow-up with less face-to-face contact) 4
5 Information on function of the device Background information on the device and the lead(s) programmed settings Automatic up-to-date information on device longevity lead(s) performance (impedance etc.) sensing and pacing activity 5
6 Device longevity estimate Based on battery voltage battery impedance percent pacing with the programmed settings Elective Replacement Indicator warning
7 Chronic lead impedance Automatic lifetime lead impedance trends
8 Lead noise discrimination Specific lead noise discrimination provides the ability to monitor and issue alerts for early signs of lead failure
9 Non-sustained lead noise SecureSense RV Lead Noise Discrimination Algorithm compares the near field electrogram to the far field electrogram If noise is visible on the near field electrogram and not in the far field electrogram therapy is withheld If binned events are equal on both electrograms the rhythm is treated as VT/VF and therapy is delivered 9
10 Non-sustained lead noise Lead Noise True VT/VF RV Tip- RV Ring (near field) RV Coil- Can (far field) Rate Correlation Different rates Rates Correlate Failure Type Tip or ring conductor noise None Decision Inhibit therapy Deliver therapy
11 Chronic P and R wave amplitude
12 Information on sensing and pacing Event counter Atrial / ventricular sensing and pacing Biventricular pacing Ventricular premature beats Rate histograms AV conduction histograms 12
13 Event counter Detected events are classified as paced/sensed and further divided into various rate ranges 13
14 Event percentage 36.8%
15 Event and rate histograms
16 AV conduction histograms Useful for Timing optimization Evaluation of appropriate pacing function Verification that device performance matches programmed settings
17 VS-AS Interval Measures from an intrinsic ventricular event (VS) to the next intrinsic atrial event (AS) A very short VS-AS interval (<80 ms) suggests far field R wave sensing Check VS-AS interval information on the histogram screen if there are sensing problems (real or suspected) there are a lot of mode switch episodes there seems to be a lot of high-rate atrial activity there is atrial oversensing a tracing of atrial activity does not match device annotations
18 Information on cardiac arrhythmias Did that patient experience arrhythmias since the last follow-up? Is the number/duration of arrhythmias for this time period congruent with the patient s history or has something changed? What is the arrhythmia diagnosis?
19 Mode switching Fast switching to a non-p wave tracking mode after detection of high rate atrial activity 4/7 Mode Switching
20 Mode switch data Heart rate peak atrial rate ventricular rate during mode switch Episode details number and duration of episodes AF burden
21 Mode switch summary Useful for verification that mode switch settings and behavior are appropriate for the patient making adjustments (if required) to the settings
22 Stored diagnostic functions Diagnostic Function Clinical Application Limitations/pitfalls Event counters Evaluation of PM function and adjustment of programming Inappropriately sensed events affect the accuracy of the counter Cannot identify individual VT or AT episodes P & R wave amplitude histogram Optimization of A & V sensing Unable to determine if low amplitude is due to AF or true undersensing Rate histogram Optimization of rate response Identification of SN dysfunction Short arrhythmia episodes are not evident Cannot identify the type of the arrhythmia Mode switch Track occurrence of Ats followed by mode switch to a non-p wave tracking mode Undersensing may underestimate the # and duration of ATs Oversensing may trigger MS High A or V rate episodes Track occurrence of ATs and VTs Undersensing may underestimate the # and duration of tachycardias Oversensing may trigger AT/VT 22
23 Noise reaction Heart Center Tampere University Hospital 23
24 Continous arrhythmia monitoring Arrhytmia Log Duration of the arrhythmia and maximum atrial and ventricular, average ventricular, and sensor rates during the episode EGM collection triggered by either A or V high rates
25 Intracardiac ECG EGM with annotation markers
26 Rolling arrhythmia trend Arrhythmia burden
27 Ventricular rate
28 Diagnostic counters vs IEGM A total of 828 EGMs (520 automatically and 308 manually triggered episodes) were recorded in 140 patients Trigger Episodes Confirmed by EGM (%) Atrial tachyarrhythmia (ATR) (47.0%) VT 37 8 (21.6%) NSVT (27.7%) ATR = atrial tachy response (rate >160 bpm, duration 8 cycles); ventricular tachycardia (VT) (rate >160 bpm, duration 8 cycles); nonsustained ventricular tachycardia (NSVT) ( 3 consecutive ventricular extrasystoles) It is important to use IEGM to confirm that the episodes were properly diagnosed by the device 28 Raatikainen et al. unpublished
29 Automatically stored IEGM Automatically triggered IEGM showing an atrial tachycardia/fibrillation. Atrial rate is >150 bpm and ventricular rate varies due to variable atrioventricular nodal conduction block. 29
30 Automatically stored IEGM Automatically trigged EGM showing sustained ventricular tachycardia in an asymptomatic patient. 30
31 Automatically stored IEGM False-positive atrial arrhythmia detection caused by noise and myopotential artifact. 31
32 Patient initiated events Patients enrolled (n=199) Completed the protocol (n=157) Symptoms after implantation (n=55) No symptoms (n=84) No symptoms Unintentional magnet use (n=18) Arrhythmia confirmed (n=13) Applied magnet (n=48) Magnet storage overwritten by ATR (n=1) No EGM stored (n=1) No arrhytmia on EGM (n=33) 32 Raatikainen et al. unpublished
33 Patient initiated IEGM storage Magnet-triggered IEGM in a patient with palpitations. The device stored both the atrial and ventricular activity during the episode, but did not reveal any arrhythmia during the symptoms. 33
34 Patient initiated IEGM storage Magnet-triggered EGM in a patient with palpitations. The device stored both the atrial and ventricular activity during the episode and revealed premature ventricular contractions (PVC). 34
35 Clinical value of AF detection 2012;366:
36 Atrial tachycardias during the F-U Subclinical ATs occurred in 633 pts (24.5%) with no AT during the monitoring period Clinical ATs occured in 41 of the 261 patients with subclinical ATs during the monitoring period (15.7%) 71 of the 2319 patients with no ATs during the monitoring period (3.1%) 36
37 Ischemic stroke/ systemic embolism Hazard ratio 2.46 ( ), P=
38 Risk of stroke The population attributable risk of stroke or systemic embolism associated with subclinical ATs was 13% Episodes > 6 minutes vs. no episodes increased risk of ischemic stroke or systemic embolism (HR 1.76 (0.99 to 3.11); P = 0.05) Episodes > 6 hours vs. no episodes HR 2.00; 95% CI, 1.13 to 3.55; P = 0.02) Episodes > 24 hours vs. no episodes HR 1.98; 95% CI, 1.11 to 3.51; P = 0.02) 38
39 Risk according to CHADS 2 score 39
40 Risk of thromboembolic events P= % 5.0% 58 pts No AF 80 pts No AF 24 pts No AF 4 pts No AF 55 pts AF > 5 min 76 pts AF > 5 min 42 pts AF > 5 min 7 pts AF > 5 min 59 pts AF > 24 hr 113 pts AF > 24 hr 45 pts AF > 24 hr 6 pts AF > 24 hr CHADS 2 score 40 Botto et al. JCE 2009
41 Atrial fibrillation Clinically detected AF Subclinical AF can be detected using pacemaker diagnostics The higher the risk score the shorter episodes of AF have clinical impact 41
42 Information on patient activity and underlying disease Data on activity and exercise Exercicise training Heart rate variability Prediction of VT/VF episodes Congestion (fluid) monitoring Prediction of heart failure worsening ST segment (ischemia) monitoring 42
43 Activity and exercise training
44 Activity 30 day daily trend Based on accelerometer input Calibrated automatically or manual Bar graph represents hours/day active
45 Exercice training Exercise = extended activity 30 Day trend Bar graph broken into 3 zones: Mild exercise Target exercise Heavy exercise
46 Congestion monitoring 46
47 ST segment monitoring ST segment deviations are measured using a dedicated unipolar ventricular electogram EGM configuration: RV electrode tip (+) - Can (-) - cathode + anode 47
48 Baseline A qualified reference of the patient s normal IEGM Patients are their own control Collected every 6 hours in resting heart rate zone The baseline obtained by the device is used as a reference for ST Shift analysis 24 hour later 48
49 ST monitoring data ST deviation trend ST histogram ST log
50 Conclusions Device diagnostics help early detection and treatment Provides a powerful diagnostic tool for detection of subclinical atrial fibrillation and other arrhythmias, worsening of HF, myocardial ischemia... Proper use of the diagnostic capabilities of implantable devices improves patient care Potential to alleviate the burden of device clinics Stored IEGM is critical component in confirmation of the arrhythmia diagnosis 50
51 Enhanced monitoring Atrial and ventricular arrhythmias Heart rate variability Mean heart rate Myocardial ischemia % CRT Activity / 24h Thoracic Fluid (intrathoracic impedance) Hemodynamics (intracardiac impedance)
EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs
EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for
More informationThe Nuts and Bolts of ICD Therapy
Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction
More informationDevice Diagnostics and Stroke Prevention: State of the Art
Device Diagnostics and Stroke Prevention: State of the Art Ass. Prof. Dr. Carsten W. Israel Dept. of Cardiology Evangelical Hospital Bielefeld Germany C.W.Israel@em.uni-frankfurt.de Disclosure of Relationships
More informationInappropriate electrical shocks: Tackling the beast
ESC Paris 2011 Inappropriate electrical shocks: Tackling the beast Gerhard Hindricks University of Leipzig Heart Center Dept. of Electrophysiology ESC Paris 2011 Inappropriate electrical shocks: Tackling
More informationSVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features
Definition of SVT Discrimination Identify which patient populations might benefit from these features Understand the 4 types of SVT discriminators used by St Jude Medical Be aware of programmable parameters
More informationTachycardia Devices Indications and Basic Trouble Shooting
Tachycardia Devices Indications and Basic Trouble Shooting Peter A. Brady, MD., FRCP Cardiology Review Course London, March 6 th, 2014 2011 MFMER 3134946-1 Tachycardia Devices ICD Indications Primary and
More informationCONSULTA CRT-P Model C4TR01
CONSULTA CRT-P Model C4TR01 Physical characteristics Volume a 15 cm 3 Mass H x W x D Surface area of titanium device can Surface area of each titanium nitride LECG electrode Radiopaque ID Materials in
More informationDevice Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI
Device Interrogation- Pacemakers, ICD and Loop Recorders Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Disclosures Consultant: Medtronic Speaker s Bureau: St. Jude Medical
More informationCardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up
Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a
More informationICD: Basics, Programming and Trouble-shooting
ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications
More informationINTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS
INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS
More informationICD Diagnostics: Making the most of it
ICD Basics and Beyond ICD Diagnostics: Making the most of it Dulce Obias-Manno RN, BSN, MHSA, FHRS, CCDS, CEPS Cardiac Arrhythmia Center, Device Clinic Washington Hospital Center, Washington DC Objectives
More informationProgramming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany
Programming of Bradycardic Parameters C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Programming of ICD Brady Parameters Conflict of Interest Biotronik
More informationEssentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH
Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing
More informationGDT1000 SENSING ACUTE STUDY
CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY GDT1000 SENSING ACUTE STUDY Boston
More informationContinuous ECG telemonitoring with implantable devices: the expected clinical benefits
Continuous ECG telemonitoring with implantable devices: the expected clinical benefits C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Declaration of
More informationPERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists
PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical
More informationAdvanced ICD Concepts
1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation
More informationIs This Thing Working?
Is This *#@!* Thing Working? Pacemaker (and ICD) ECG and Telemetry Pitfalls Wayne O. Adkisson, MD adki0004@umn.edu Disclosures I currently receive research support from Medtronic, Inc. I have been compensated
More informationMEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring
MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine
More informationINNOVATIONS IN DEVICE THERAPY:
INNOVATIONS IN DEVICE THERAPY: Subcutaneous ICDs, Leadless Pacemakers, CRT Indications David J Wilber MD Loyola University Medical Center Disclosures: ACC Foundation: Consultant; Biosense / Webster: Consultant,
More informationPERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS
PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing
More informationClinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm
April 2000 107 Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm B. MERKELY Semmelweis University, Dept. of Cardiovascular Surgery,
More informationA Closer Look Product Education at a glance
A Closer Look Product Education at a glance an ICD or CRT-D When Atrial Information is Not Used SUMMARY Boston Scientific dual-chamber ICDs and multi-chamber CRT-Ds will respond to atrial sensing regardless
More informationPacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital
Pacing and Device Jargon Made Simple Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Disclosure Dr Jonathan Timperley Honorarium from Bayer Arne Larson (1915
More informationInteractive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators
22 March 2002 Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators F. HINTRINGER, O. PACHINGER Division of Cardiology, Department for Internal Medicine, University
More informationTroubleshooting ICD. NASPE Training Lancashire & South Cumbria Cardiac Network
Troubleshooting ICD NASPE Training Lancashire & South Cumbria Cardiac Network Bradycardia Pacing by ICD Isolated ICD discharge Multiple discharges Appropriate Inappropriate No ICD Therapy or delay in ICD
More informationIntroduction to a new "X out of Y" Mode-Switching Algorithm in Inos² CLS and Logos Dual-Chamber Pacemakers
April 1999 117 Introduction to a new "X out of Y" Mode-Switching Algorithm in Inos² CLS and Logos Dual-Chamber Pacemakers C. W. ISRAEL Department of Cardiology, St.-Josef-Hospital, Ruhr-University, Bochum,
More informationSupplementary Online Content
Supplementary Online Content Tseng ZH, Hayward RM, Clark NM, et al. Sudden death in patients with cardiac implantable electronic devices. JAMA Intern Med. Published online June 22, 2015. doi:10.1001/jamainternmed.2015.2641.
More informationΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ
ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted
More informationThe Relationship Between Daily Atrial Tachyarrhythmia Burden From Implantable Device Diagnostics and Stroke Risk: The TRENDS Study
The Relationship Between Daily Atrial Tachyarrhythmia Burden From Implantable Device Diagnostics and Stroke Risk: The TRENDS Study Taya V. Glotzer, MD; Emile G. Daoud, MD; D. George Wyse, MD, PhD; Daniel
More informationSubclinical AF: Implications of device based episodes
Subclinical AF: Implications of device based episodes Michael R Gold, MD, PhD Medical University of South Carolina Charleston, SC Disclosures: Clinical Trials and Consulting: Medtronic, Boston Scientific
More informationPediatric pacemakers & ICDs:
Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components
More informationAutomatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms
Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable
More informationTeaching Rounds in Cardiac Electrophysiology
Teaching Rounds in Cardiac Electrophysiology Sustained Multiple Railroad Tracks on Implantable Cardiac Defibrillator Interval Plots Mechanisms and Management Alex Y. Tan, MD; Kenneth A. Ellenbogen, MD;
More informationDual-Chamber Implantable Cardioverter-Defibrillator
February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary
More informationRemote monotoring of cardiac rhythm devices: present and future Pacemaker and ICD
Remote monotoring of cardiac rhythm devices: present and future Pacemaker and ICD Philippe Mabo University Hospital, Rennes, France ESC congress, Paris 30 August 2011 U642 Disclosures Biotronik: research
More informationEnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can
EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds
More informationPanorama. Arrhythmia Analysis Frequently Asked Questions
Panorama Arrhythmia Analysis Frequently Asked Questions What ECG vectors are used for Beat Detection? 3-wire lead set 5-wire lead set and 12 lead What ECG vectors are used for Beat Typing? 3-wire lead
More informationImplantable Cardiac Monitors for Atrial Fibrillation (AF) Detection: Ready for Routine Use?
Implantable Cardiac Monitors for Atrial Fibrillation (AF) Detection: Ready for Routine Use? Helmut Pürerfellner, MD, Assoc. Prof. Saint Elisabeth s Sisters Hospital Academic Teaching Center Linz/Austria
More informationPATIENT WITH ARRHYTHMIA IN DENTIST S OFFICE. Małgorzata Kurpesa, MD., PhD. Chair&Department of Cardiology
PATIENT WITH ARRHYTHMIA IN DENTIST S OFFICE Małgorzata Kurpesa, MD., PhD. Chair&Department of Cardiology Medical University of Łódź The heart is made up of four chambers Left Atrium Right Atrium Left Ventricle
More informationSinus rhythm with premature atrial beats 2 and 6 (see Lead II).
Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess
More informationAdvances in Arrhythmia and Electrophysiology
Advances in Arrhythmia and Electrophysiology Troubleshooting Implanted Cardioverter Defibrillator Sensing Problems I Charles D. Swerdlow, MD; Samuel J. Asirvatham, MD; Kenneth A. Ellenbogen, MD; Paul A.
More informationHow to Approach the Patient with CRT and Recurrent Heart Failure
How to Approach the Patient with CRT and Recurrent Heart Failure Byron K. Lee MD Associate Professor of Medicine Electrophysiology and Arrhythmia Section UCSF Update in Electrocardiography and Arrhythmias
More informationPractice Questions.
IBHRE Prep Practice Questions Question 1 The relative refractory yperiod of the ventricular myocardium corresponds to which of the following phases of the action potential? A. (0) B. (1) C. (2) D. (3)
More informationA New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support
100 April 2000 A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support A. SH. REVISHVILI A.N. Bakoulev Institute of Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow,
More informationFriedman, Rott, Wokhlu, Asirvatham, Hayes 201. Figure 65.7 Shortening of the AV interval during pacing.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 201 Figure.7 Shortening of the AV interval during pacing. 202 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization Figure.8 is obtained from
More informationPACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I
1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL
More informationCan Catheter Ablation of AF Reduce the Risk of Stroke? CCCEP 2015 October 31, 2015
Can Catheter Ablation of AF Reduce the Risk of Stroke? CCCEP 2015 October 31, 2015 Atul Verma, MD FRCPC FHRS Director, Heart Rhythm Program Southlake Regional Health Centre Newmarket, Ontario, Canada Chair,
More informationCase Description A 25 years old male served in a combat unit for 3 years (GOLANY) Implantation of Dual chamber ICD in 7/2010 due to Severe Non ischemi
ICD Case Study tracing your own images Prof Glikson, Sandra Schor-Wider Nadav Hayman Case Description A 25 years old male served in a combat unit for 3 years (GOLANY) Implantation of Dual chamber ICD in
More informationA Rare Type of Pacemaker Mediated Tachycardia
A Rare Type of Pacemaker Mediated Tachycardia Serkan ÇAY M.D., Ümit GÜRAY M.D., Cihan DEMİR M.D. Yuksek Ihtisas Heart-Education and Research Hospital, Department of Cardiology, Ankara, Turkey ABSTRACT
More informationPresented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO
Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! What is a
More informationJohn H. Ip, M.D FACC Associate Professor of Medicine Michigan State University Medical Director, EP Service
ICD Trouble shooting: Case study John H. Ip, M.D FACC Associate Professor of Medicine Michigan State University Medical Director, EP Service Bullet proof evidence: ICD does prevent sudden death Disclosures
More informationLATITUDE Patient Management System
LATITUDE Patient Management System Boston Scientific Corporation Overview Key Benefits Benefits in HF-Management The future of RPM 2 Boston Scientific Corporation LATITUDE Patient Management Overview Patient
More informationRN-BC, MS, CCRN, FAHA
Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! 1 A. Atropine
More informationTachyarrhythmia Suspicion and Detection
Tech Corner Tachyarrhythmia Suspicion and Detection NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN THE ARTICLE.
More information11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants
EKG Pop Quiz Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants 1 Disclosures No financial relationships to disclose. EKG #1 75 y/o woman with a dual-chamber pacemaker
More informationAdvanced ICD Troubleshooting: Part II REVIEW
REVIEW Advanced ICD Troubleshooting: Part II CHARLES D. SWERDLOW* and PAUL A. FRIEDMAN From the *Cedars-Sinai Medical Center, Los Angeles, California, and Mayo Clinic, Rochester, Minnesota, USA Failure
More informationPuzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC
Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact
More informationM Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use
DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.
More informationElectrocardiography for Healthcare Professionals
Electrocardiography for Healthcare Professionals Kathryn A. Booth Thomas O Brien Chapter 10: Pacemaker Rhythms and Bundle Branch Block Learning Outcomes 10.1 Describe the various pacemaker rhythms. 10.2
More informationCAPTIVATE SUMMARY CLINICAL SUMMARY. CAPTure Information Via Automatic Threshold Evaluation
CLINICAL SUMMARY CAPTIVATE SUMMARY CAPTure Information Via Automatic Threshold Evaluation CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in
More informationTesting the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device
Testing the Accuracy of ECG Captured by through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Data Analysis a) R-wave Comparison: The mean and standard deviation of R-wave amplitudes
More informationTiming cycles. Objectives Upon completion of this program the participant will be able to:
Timing Cycles Timing cycles Objectives Upon completion of this program the participant will be able to: Identify the basic timing cycles of a single and dual chamber pacemaker. Describe the characteristics
More informationPARAD/PARAD+ : P and R Based Arrhythmia Detection
Tech Corner PARAD/PARAD+ : P and R Based Arrhythmia Detection NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN
More informationRecurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm
Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm Guy Amit, MD, MPH Soroka University Medical Center Ben-Gurion University of the Negev Beer-Sheva, Israel Disclosures Consultant:
More informationCONTAK RENEWAL 3 AVT CLINICAL SUMMARY
CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY CONTAK RENEWAL 3 AVT Boston Scientific
More informationTech Corner. ATP in the Fast VT zone
Tech Corner ATP in the Fast VT zone NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN THE ARTICLE. FOR ADDITIONAL
More informationA Deep Dive: Using Device Diagnostics to Assist with Patient Management. Karen Donatello, RN, RCIS, CCDS
A Deep Dive: Using Device Diagnostics to Assist with Patient Management Karen Donatello, RN, RCIS, CCDS Financial Relationships Device Check Specialist Karen Donatello, RN, RCIS, CCDS Objectives Demonstrate
More informationRiata Lead Summit Jan 20, SJM Perspective Dr. Mark Carlson
Riata Lead Summit Jan 20, 2012 SJM Perspective Dr. Mark Carlson 1 1 Presentation Outline Background, Characterization of Externalized Conductors Modeling and Testing Diagnostics and Therapeutics Recommendations
More informationA Closer Look. LATITUDE NXT Alerts SUMMARY. Alerts. Red Alerts
A Closer Look SUMMARY Boston Scientific s LATITUDE NXT Patient Management System enables a clinician to periodically monitor patient and device information remotely via a Communicator placed in the patient
More informationWe are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors
We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 4,100 116,000 120M Open access books available International authors and editors Downloads Our
More informationWide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?
CASE REPORTS Arrhythmia 2015;16(3):173-177 doi: http://dx.doi.org/10.18501/arrhythmia.2015.029 Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? Eun-Sun Jin, MD, PhD Cardiovascular
More informationA Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study
A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study ANEESH V. TOLAT, M.D.,* MELISSA WOICIECHOWSKI, M.S.N.,* ROSEMARIE KAHR, R.C.I.S.,* JOSEPH
More informationJournal of Arrhythmia
Journal of Arrhythmia 29 (2013) 204 210 Contents lists available at ScienceDirect Journal of Arrhythmia journal homepage: www.elsevier.com/locate/joa Original Article Factors involved in correct analysis
More informationClinical Data Summary: Avoid FFS Study
Atrial Pacing Lead with 1.1 mm Tip-to-Ring Spacing Clinical Data Summary: Avoid FFS Study A Multi-center, Randomized, Prospective Clinical Study Designed to Evaluate the 1699T Lead Clinical Data Summary:
More informationHRC Carole Joyce. Bradford Royal Infirmary. Senior Chief Cardiac Physiologist. Pacing & Invasive Services.
HRC 2010 Carole Joyce Carole.joyce@bradfordhospitals.nhs.uk Pacing & Invasive Services Carole Joyce Introduction Basic Protocol Troubleshooting Case Studies Questions & Answer Carole Joyce ICD Troubleshooting
More informationIntermittent episodes of paced tachycardia: what is the cause?
Intermittent episodes of paced tachycardia: what is the cause? Lindsey Parkinson and Lucy Broadhurst, Rotherham NHS Foundation Trust Contact: Lindsey.Parkinson@rothgen.nhs.uk Introduction A hospital in-patient
More informationEndpoints When Treating VT/VF in Patients with ICDs Programming Wojciech Zareba, MD, PhD
Endpoints When Treating VT/VF in Patients with ICDs Programming Wojciech Zareba, MD, PhD Professor of Cardiology/Medicine Director of the Heart Research Follow Up Program, University of Rochester, Rochester,
More informationCARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART
CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART December 2015 This chart encompasses all Medtronic cardiac devices FDA-Approved as MR Conditional and included in the MRI SureScan portfolio. If a device
More informationAmbulatory Cardiac Monitors and Outpatient Telemetry Corporate Medical Policy
Ambulatory Cardiac Monitors and Outpatient Telemetry Corporate Medical Policy File Name: Ambulatory Event Monitors and Mobile Cardiac Outpatient Telemetry File Code: UM.SPSVC.13 Origination: 10/2015 Last
More informationSolara CRT-P MRI SureScan W1TR06
Solara CRT-P MRI SureScan W1TR06 MR Conditional pacemaker with cardiac resynchronization therapy, SureScan technology, and Bluetooth wireless telemetry (OAE-DDDR) Device Manual The following list includes
More informationSARASOTA MEMORIAL HOSPITAL
SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish
More informationCall Medtronic at 1 (800) to verify the patient s current implanted system
MRI SURESCAN SYSTEMS Patient Scanning Process Transvenous Implantable Cardiac Systems PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT (CRT-D and CRT-P) Systems Verification Verify that patient
More informationVictory on demand, time and time again
Victory on demand, time and time again Victory Pacemaker Family Cardiac Rhythm St. Jude Medical AB Management Division Veddestavägen 19 15900 Valley View Court SE-175 84 Järfälla Sylmar, CA 91342 USA SWEDEN
More informationCardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution
Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Francesco Zanon, MD, FESC, FHRS Arrhythmia and Electrophysiology Unit, Cardiology Dept. Santa Maria della Misericordia General
More informationESC Stockholm Arrhythmias & pacing
ESC Stockholm 2010 Take Home Messages for Practitioners Arrhythmias & pacing Prof. Panos E. Vardas Professor of Cardiology Heraklion University Hospital Crete, Greece Disclosures Small teaching fees from
More informationPercepta CRT-P MRI SureScan W1TR01
Percepta CRT-P MRI SureScan W1TR01 MR Conditional pacemaker with cardiac resynchronization therapy, SureScan technology, and Bluetooth wireless telemetry (OAE-DDDR) Device Manual Caution: Federal law (USA)
More informationAF monitoring and stroke: XPECT & REVEAL LINQ. Helmut Pürerfellner, Linz, Austria
AF monitoring and stroke: XPECT & REVEAL LINQ Helmut Pürerfellner, Linz, Austria 2 Medtronic Confidential 3 Circ Arrhythm Electrophysiol 2010;3:141-47 Reveal XT versus LINQ 5 Parameter Reveal XT Reveal
More informationHow to set up a device clinic for heart failure patients Dr C Sridevi
How to set up a device clinic for heart failure patients Dr C Sridevi Consultant Cardiologist /Electro physiologist CARE Hospitals, CARE Foundation Need to set up a device clinic for heart failure patients
More informationDual chamber pacing mode in an atrial antitachycardia pacing device without a ventricular lead e A necessary evil
indian pacing and electrophysiology journal 15 (2015) 133e137 HOSTED BY Available online at www.sciencedirect.com ScienceDirect journal homepage: www.elsevier.com/locate/ipej Dual chamber pacing mode in
More informationPACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART II
1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART II 2012 Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 ADDITIVES IN DEVICE THERAPY Key Choice / CNEA 1 Adaptive Rate
More informationGE Healthcare. The GE EK-Pro Arrhythmia Detection Algorithm for Patient Monitoring
GE Healthcare The GE EK-Pro Arrhythmia Detection Algorithm for Patient Monitoring Table of Contents Arrhythmia monitoring today 3 The importance of simultaneous, multi-lead arrhythmia monitoring 3 GE EK-Pro
More informationSupplemental Material
Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement
More informationNew generations pacemakers and ICDs: an update
Advances in Cardiac Arrhythmias and Great Innovations in Cardiology XXVII Giornate Cardiologiche Torinesi New generations pacemakers and ICDs: an update Prof. Fiorenzo Gaita, MD Division of Cardiology
More informationLa strategia diagnostica: il monitoraggio ecg prolungato. Michele Brignole
La strategia diagnostica: il monitoraggio ecg prolungato Michele Brignole ECG monitoring and syncope In-hospital monitoring Holter Monitoring External loop recorder Remote (at home) telemetry Implantable
More informationRole of Implantable Loop Recorder in the Evaluation of Syncope
Role of Implantable Loop Recorder in the Evaluation of Syncope June Soo Kim, M.D., Ph.D. Department of Medicine Cardiac & Vascular Center Sungkyunkwan University School of Medicine Definition & Mechanism
More informationAzure XT DR MRI SureScan W2DR01
Azure XT DR MRI SureScan W2DR01 MR Conditional dual chamber pacemaker with SureScan technology and Bluetooth wireless telemetry (OAE-DDDR) Device Manual The following list includes trademarks or registered
More informationMRI and CIEDs. Disclosures
Date Clinical Title MRI and CIEDs C. p. Lao Raul Weiss MD, FAHA, FACC, FHRS and CCDS Director, Electrophysiology Fellowship Program Professor of Medicine The Ohio State University Wexner Medical Center
More informationManual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev
Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value
More information