Transcutaneous Pacing on Zoll Defibrillator Version 4 Related Documents SOP-Zoll defibrillator ; Phillips MRX SOP Dr Randal McRoberts
|
|
- Malcolm Walton
- 6 years ago
- Views:
Transcription
1 Emergency Medical Retrieval Service (EMRS) Standard Operating Procedure Public Distribution Title Transcutaneous Pacing on Zoll Defibrillator Version 4 Related Documents SOP-Zoll defibrillator ; Phillips MRX SOP Author Dr Randal McRoberts Reviewer Dr Stephen Hearns THE PHILLIPS MRX DEVICE (DEFAULT MONITOR FOR EMRS) HAS A SEPARATE SOP WHICH INCLUDES A PACING SECTION. THIS SOP PURELY REFERS TO THE ZOLL DEVICE. Aims This document discusses the transcutaneous pacing facilities using the Zoll defibrillator currently in use with EMRS Background Patients in complete heart block may require temporary pacing transcutaneously prior to undergoing formal permanent pacemaker (PPM) insertion at a definitive care centre. Unless a functional transvenous pacing wire has been inserted at the referring site, the patient may require transcutaneous pacing (TCP) by EMRS during transport to definitive care. The E series Zoll (manual) defibrillator/pacing devices currently used as a backup monitor by EMRS have been cleared for use on Military aircraft and SAS-A & SAS-B helicopters. Application EMRS Team Members SAS Paramedics SOP- Transcutaneous Pacing 1
2 Patients appropriate for retrieval team activation Patients in complete heart block or those patients whose bradycardia demands measures beyond that of drug intervention to achieve adequate cardiac output. Advice to GP/Hospital referrer prior to team arrival: Arrange for transvenous pacing if equipment & expertise available Support heart rate and blood pressure according to the current ALS algorithm for bradycardia Continuous ECG/SpO2/Heart Rate/BP monitoring 12 lead ECG & repeat at 30min intervals Consider brady-arrhythmia may be part of ACS and treat accordingly Look for electrolyte abnormalities and treat accordingly If transcutaneous pacing is available and is required, institute its use with sedation for patient comfort as required Medical management on scene The patient should be fully assessed on arrival. The treatment instituted to this point by the referring centre should be noted. 1. Resuscitate as necessary according to ALS guidelines 2. If the patient has had transvenous or transcutaneous pacing, the following should be checked: The effectiveness of this intervention i.e. that the pacing is achieving both electrical and mechanical capture and that the patient s clinical condition reflects this 12 lead ECG Any bloods/investigations if done already (eg: Troponin; U&E) If transvenous: The position of the pacing wire on the CXR. Ensure no immediate complication of the procedure (eg: Pneumothorax) If transcutaneous: The condition/length of use of the pads (replace if have been used for 8hours of continuous pacing) Position of pads Front & back positioning of pads is optimal for pacing Patient comfort check for need for sedation/analgesia to achieve effective pacing SOP- Transcutaneous Pacing 2
3 Having decided clinically on the need for transcutaneous pacing, the following steps should be taken when setting up the Zoll E series defibrillator/pacer device: a) Dry and prepare patient s chest for application of Stat-Padz AND 3 (or 5) Lead ECG. Both are needed to pace in either Demand or Standby mode. On the rare occasion that asynchronous pacing is used, the ECG will not be applied. b) Explain to the patient that it feels uncomfortable during the pacing period c) Check stat padz multi-function electrodes (MFE) expiry date d) Connect the MFE cable to the Zoll device multi-function cable (white opaque connector on MFE to red connector on Zoll) e) Apply the MFE to the patient as per the instructions on the packaging. The pads should be placed in the anterior-posterior orientation for pacing. If AP positioning not possible then consider apex-sternum positioning (less effective for pacing). Ensure adhesion to skin. f) If the MFE pads are not making good contact with the patient, the messages CHECK PADS and POOR PAD CONTACT are alternately displayed and the energy will not be delivered. g) If a short circuit exists between the MFE pads, the message DEFIB PAD SHORT is displayed. h) The following steps can be applied to optimise the set-up of the device for pacing: SOP- Transcutaneous Pacing 3
4 Step 1 The Pacer (Green) mode is accessed by turning the Selector Switch counterclockwise Milliamps (ma) are the unit of pacing energy No AED capability or ANALYZE button can be used in pacing mode Step 3 Turn the Pacer Output dial to adjust the level of discharged milliamps. If capture is achieved, the ECG will have a wide complex reflecting ventricular contraction following the thin Pacing spike marker Step 2 Pacer markers (PM) indicate the rate set to attempt to capture the ventricle. Default settings of 70 Paces Per Minute (PPM) and 0 ma are displayed upon access of Pacer Mode To increase or decrease pacer marker rate, (PM) turn the Pacer Rate Dial i) Once electrical and mechanical capture has been confirmed, dial the ma up 10% from capture threshold as a safety margin The 4:1 button may be pushed at anytime to allow for review of the underlying rhythm. The Zoll device can perform 3 different types of pacing: 1) Demand Pacing: Most frequent form of ventricular pacing. The Pacing rate is set above patient s rate (or lack thereof) and the Pacer Output dial is turned to increase the ma in attempt to obtain capture and pace the ventricles. 2) Stand-by Pacing: Setting the Pacing rate and Pacer output at a back-up rate less than a patient s intrinsic heart rate. The Pacing rate will initially be set above the patient s heart rate and pacer output (ma) is increased to achieve 100% capture. The Pacing rate is then decreased to desired rate below the patient s intrinsic heart rate. Should the HR drop, the stand-by pacer will initiate impulses and begin to pace. 3) Asynchronised Pacing: Rarely used. This form of pacing is performed when no ECG electrodes can be placed due to burns, trauma or interference. The async on/off softkey button is pressed and async mode is displayed. No Pacing spike or electrical capture will be seen on the screen. Mechanical capture will only be proven, by palpating a pulse if one is achieved SOP- Transcutaneous Pacing 4
5 Triage The normal referral pathways for the referring site should be followed where possible and the local cardiology unit capable of permanent pacemaker insertion should be the destination of choice. In the situation where a patient may have had an acute coronary syndrome, that patient may require PCI and as such, must be triaged to a centre capable of providing this. Bibliography 1. ALSG website algorithm for brady-arrhythmia: 2. Zoll website for instruction manual for E series: SOP- Transcutaneous Pacing 5
Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD
External Pacemaker Set-up and Revised Date: 11/13/2014 Page 1 of 5 POLICY: The transcutaneous external pacemaker will be used as a noninvasive method of delivering electrical stimuli to the myocardium
More informationTitle and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2
Temporary Transvenous Pacing Guideline Classification: Clinical Guideline Lead Author: Dr Peter Woolfson Additional author(s): Dr Alan Fitchet Sister Joanne Hughes, Matron Julie Winstanley Authors Division:
More informationFigure 2. Normal ECG tracing. Table 1.
Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are
More informationPresented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO
Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! What is a
More informationRN-BC, MS, CCRN, FAHA
Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! 1 A. Atropine
More informationCardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 2018
Cardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 08 4 Universal cable may be inserted to the side Defib Short Block to perform 0 J check Universal Cable must be connected to APEX
More informationTranscutaneous Pacing. Approval: Medical Director James Stubblefield, MD. Approval: EMS Director Michael Petrie
Transcutaneous Pacing 6091D Low Frequency/High Risk: Transcutaneous Pacing Applies To: Paramedics Effective Approval: Medical Director James Stubblefield, MD Approval: EMS Director Michael Petrie Expires
More informationEarly Defibrillation. Dr. M. Ravishankar
Early Defibrillation Dr. M. Ravishankar THE CHAIN OF SURVIVAL CONCEPT Early Access Early CPR Early Defibrillation Early Advance Care 1 st Link 2 nd Link 3 rd Link 4 th Link Why early defibrillation VF
More informationPuzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC
Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact
More informationM Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use
DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.
More informationTemporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS
Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS NBG Codes 1st Letter 2nd Letter 3rd Letter A V D Chamber(s) Paced = atrium = ventricle = dual (both atrium and ventricle) Chamber(s) Sensed A = atrium
More informationManual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev
Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value
More informationCardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand:
Cardiac Pacing CHAPTER 10 Learning outcomes To understand: The indications for cardiac pacing in the peri-arrest setting How to perform percussion pacing How to apply non-invasive, transcutaneous electrical
More informationOBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.
Defibrillators OBJECTIVE 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe
More informationPACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I
1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL
More informationSARASOTA MEMORIAL HOSPITAL
SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish
More informationModel 5392 EPG Temporary Pacer
Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short
More informationMODEL EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER
MODEL 53401 EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER In-service Guide 53401 TRAINING LEARNING OBJECTIVES Identify the control knobs and be able to explain how to use them to program device. Distinguish
More informationManual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing
Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Key Points Defibrillators: - know the definition & electrical value of a joule - monophasic vs biphasic types:
More informationFlowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields
Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Identify patient with an ICD at POAC. Notify Cardiac Physiologist that patient is due to have surgery and
More informationTIP CARD. Model Temporary External Pacemaker
TIP CARD Model 5392 Temporary External Pacemaker PRE-USE What to do... What NOT to do... Battery Replace the battery for each new patient. Do NOT reuse battery. Physical Check case for cracks/damage.**
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa INTENSIVE CARE POLICY & PROCEDURES Policy Number: R-25 Subject: Policy: Purpose: Cardioversion - Synchronized An RN may perform a synchronized
More informationPacemakers 12/04. Pacemakers. 1. What is a pacemaker?
Pacemakers 12/04 1-What is a pacemaker? 2- What does intrinsic mean? 3- How exactly do pacemakers work on the heart? 4- What are the parts of a pacemaker? 5- Are there different kinds of pacemakers? 6-
More informationPediatric pacemakers & ICDs:
Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components
More informationPeri-operative management of pacemakers and implantable cardiac defibrillators
Guideline Title: Peri-operative management of patients fitted with Permanent Pacemakers (PPMs) and Implantable Cardioverter Defibrillators (ICDs). Author(s): Jessica Osman (Chief Pacing Physiologist),
More informationAss. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital
Treatment Bradyarrhythmia Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital 1 What is the rhythm? Sinus Bradycardia What s rhythm Sinus Bradycardia Treatment Asymptomatic No treatment
More informationThe heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.
PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in
More information1. What additional information needs to be collected to properly treat this client?
CASE 1 A 45-year-old male presents to the emergency department with a complaint of chest pain for the past two hours. 1. What additional information needs to be collected to properly treat this client?
More informationPanorama. Arrhythmia Analysis Frequently Asked Questions
Panorama Arrhythmia Analysis Frequently Asked Questions What ECG vectors are used for Beat Detection? 3-wire lead set 5-wire lead set and 12 lead What ECG vectors are used for Beat Typing? 3-wire lead
More informationElectrical Interventions
3 Electrical Interventions Quick Contents Rationale for Electricity - p. 74 Paddle Placement - p. 77 Defibrillation - p. 82 Using an AED- p. 85 Manual Defibrillation - p. 86 Cardioversion - p. 87 Transcutaneous
More informationNon-Invasive Transcutaneous Pacing
Non-Invasive Transcutaneous Pacing Application Note Introduction This Application Note provides an overview of non-invasive transcutaneous pacing. It includes a brief history of pacing, describes the electrical
More informationIndications The AED mode is to be used only on patients in cardiopulmonary arrest.
14 October 2008 TO: Dr. Copass FR: Brian W. Chinn B3 A2 1018 RE: SFD LP12 Guidelines: I. AED / Advisory II. Manual A. Defibrillation B. Synchronize Cardioversion C. Noninvasive Pacing 1. Demand 2. Non-Demand
More informationJune 2009 CE. Site code # E-1209
1 June 2009 CE Site code # 107200E-1209 Region X SOP Review ACS, Asystole/PEA, Bradycardia, VF/Pulseless VT, Conscious Sedation, Region X Field Triage Criteria Prepared by: Sharon Hopkins, RN, BSN, EMT-P
More informationPERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS
PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing
More informationProject Title Temporary Pacemaker Training Simulator
Project Title Temporary Pacemaker Training Simulator Project Description Problem: There is no available training device for temporary pacemakers (pacemakers). A training device will have to essentially
More informationAutomated External Defibrillation Principle of Early Defibrillation States that all BLS personnel be trained, equipped and allowed to operate a if
1 2 3 4 5 6 Automated External Defibrillation Principle of Early Defibrillation States that all BLS personnel be trained, equipped and allowed to operate a if they are expected to respond to persons in
More information1 Cardiology Acute Care Day 22 April 2013 Arrhythmia Tutorial Course Material
1 Cardiology Acute Care Day 22 April 2013 Arrhythmia Tutorial Course Material Arrhythmia recognition This tutorial builds on the ECG lecture and provides a framework for approaching any ECG to allow the
More information1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium
١ 1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium ٢ When VF/pulseless VT cardiac arrest is associated with torsades de pointes, providers may administer magnesium sulfate at a dose of 1-2
More informationMINI MASSAGER. Relaxation,Anywhere, Anytime
MINI MASSAGER Relaxation,Anywhere, Anytime Inclu ded in this standard package: Device x 1pc Large pad x 2pcs Small pad x 2pcs Electric wire x 2pcs USB cable & AC adapter x 1set Acupuncture point chart
More informationCOALINGA STATE HOSPITAL. NURSING POLICY AND PROCEDURE MANUAL SECTION Emergency Procedures POLICY NUMBER: 716. Effective Date: March 3, 2007
COALINGA STATE HOSPITAL NURSING POLICY AND PROCEDURE MANUAL SECTION Emergency Procedures POLICY NUMBER: 716 Effective Date: March 3, 2007 SUBJECT: MONITOR/DEFIBRILLATOR 1. PURPOSE: This policy and procedure
More informationT.E.N.S (Transcutaneous Electrical Nerve Stimulation)
You have been prescribed a T.E.N.S machine to help in the management of your pain. You should use it only for the condition for which it is prescribed. Not all pain will respond to the use of T.E.N.S and
More informationChapter 80 Implantable Cardiac Devices
Chapter 80 Implantable Cardiac Devices Episode Overview: 1. List 5 Indications for permanent pacing 2. Pacemaker nomenclature - what do the 5 letters mean a. What are common Pacemaker types? 3. List the
More informationNORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN)
NORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN) Operational policy for optimal management of: Cardiac pacemakers, Implantable cardioverter defibrillators (ICD), Cardiac resynchronisation devices, (CRT-D,
More informationMichigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS
Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS Pediatric Asystole Section 4-1 Pediatric Bradycardia Section 4-2 Pediatric Cardiac Arrest General Section 4-3 Pediatric Narrow Complex Tachycardia
More informationImplantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević
Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular
More informationSTANDARD OPERATING PROCEDURE (SOP)
STANDARD OPERATING PROCEDURE (SOP) MRI Scanning for Patients with Cardiac Implantable Electronic Devices (Pacemakers and Implantable Cardiac Defibrillators) Effective date 05/04/2018 Version number V1.0
More informationNewer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.
Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded
More informationPermanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale
Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Jason Collinson & Stuart Tan Essex Cardiothoracic Centre, Basildon and Thurrock University Hospital. Contact: jason.collinson@nhs.net
More informationMichigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS
Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS Pediatric Asystole Section 4-1 Pediatric Bradycardia Section 4-2 Pediatric Cardiac Arrest General Section 4-3 Pediatric Narrow Complex Tachycardia
More informationMONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES
PURPOSE Guidelines for caring for and monitoring patients with a temporary external epicardial pacemaker. Epicardial pacemaker monitoring is performed on a continuous basis by a nurse to evaluate the effectiveness
More informationECG Monitoring, Defibrillation and Synchronized Cardioversion with Hands-Free Electrodes
Suggestions for Performance Evaluation with the LIFEPAK 20 Defibrillator/Monitor ECG Monitoring, Defibrillation and Synchronized Cardioversion with Hands-Free Electrodes Name: Unit: Reviewer: Date: This
More informationDefibrillator. BiomedGuy
Defibrillator BiomedGuy Medtronic Physiocontrol LifePak 10 Introduction This life-support system is used by paramedic, hospital staff, and other trained authorized healthcare providers. Provides, ECG,
More informationDEFIBRILLATORS. Prof. Yasser Mostafa Kadah
DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic
More informationCORONARY ARTERIES HEART
CARDIAC/ECG MODULE THE HEART CORONARY ARTERIES FIBRILLATING HEART CORONARY ARTERIES HEART PRACTICE RHYTHMS PRACTICE RHYTHMS ELECTRICAL CONDUCTION SA Node (60 100) Primary pacemaker AV Node (40 60) ***Creates
More informationEXTERNAL TEMPORARY PACEMAKERS
Czech Republic EXTERNAL TEMPORARY PACEMAKERS Lukas Peter Engineering and Business Coordinator Pavel Taborsky Chief Executive Officer MEDIATRADE s.r.o. PRAGUE Valasske Mezirici MEDIATRADE s.r.o. More than
More informationAddendum: User Guide AutoPulse Resuscitation System Model 100 with Defibrillator Interface
Addendum: User Guide AutoPulse Resuscitation System Model 100 with Defibrillator Interface 3.5 Periodic Electrocardiogram (ECG) Monitoring and/or Defibrillation When the AutoPulse is used in conjunction
More informationDefibrillators TEC-8300K series. Improving quality of resuscitation
Defibrillators TEC-8300K series Improving quality of resuscitation Transitioning back to life For intensive treatment and advanced life support Pre-hospital Out-of-hospital 12-lead ECG acquisition and
More informationSTATE OF OKLAHOMA 2014 EMERGENCY MEDICAL SERVICES PROTOCOLS
5M VENTRICULAR ASSIST DEVICE (VAD) MANAGEMENT ADULT EMERGENCY MEDICAL DISPATCHER EMERGENCY MEDICAL RESPONDER EMT EMT-INTERMEDIATE 85 ADVANCED EMT PARAMEDIC A Ventricular Assist Device, or VAD, is a mechanical
More informationDual Chamber Temporary External Pacemaker
5392 Dual Chamber Temporary External Pacemaker Technical Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician. The following list includes trademarks or registered
More informationSubcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death
NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE Interventional procedure consultation document Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death A subcutaneous
More informationMini Pulse Electronic Stimulator
Mini Pulse Electronic Stimulator Model: PM-180 Operating Manual IMPORTANT: Please read all instructions before using this product. Retain this manual for future reference. www.santamedical.com IMPORTANT
More informationΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ
ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted
More informationWHY IS FIRST AID IMPORTANT?
Chapter 1 -BACKGROUND INFORMATION WHY IS FIRST AID IMPORTANT? 1. 2. SOME THINGS I SHOULD HAVE IN MY FIRST AID KIT 1. 2. 3. 4. 5. VOCABULARY Place the key points of the first aid term in the box provided
More informationTEMPORARY PACEMAKERS HINTS ( DUAL CHAMBER 5392 PACER)
TEMPORARY PACEMAKERS HINTS ( DUAL CHAMBER 5392 PACER) TYPES OF PACERS 2018 1 TRANSCUTANEOUS: ( VENTRICULAR pacing only) Through the skin via gel electrode pads placed on the chest and back. Disadvantage:
More informationEssentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH
Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing
More informationElectrocardiography for Healthcare Professionals
Electrocardiography for Healthcare Professionals Kathryn A. Booth Thomas O Brien Chapter 10: Pacemaker Rhythms and Bundle Branch Block Learning Outcomes 10.1 Describe the various pacemaker rhythms. 10.2
More informationIcd 10 for pacemaker pocket infection
Search Search pages & people Search Search Search pages & people Search Icd 10 for pacemaker pocket infection Oct 29, 2014. 9.1.2 In generator pocket infection, ICED-LI and ICED-IE? 9.1.3 What is the.
More informationEmergency Department Management of Patients with Implantable Cardioverter Defibrillators
IAEM Clinical Guideline Emergency Department Management of Patients with Implantable Cardioverter Defibrillators Version 1 June 2014 DISCLAIMER IAEM recognises that patients, their situations, Emergency
More informationAWAY Model # RTLAGF-900 REV
AWAY Model # RTLAGF-900 REV2.4.9.15 Contents User manual English 2 General Description 2 Intended Use 3 Included in this Package 4 Instructions for Use 7 Application Duration 7 Electrode Placement 8 Tips
More informationSJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only.
SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000 1 APPROVED MRI SCAN CONDITIONS 1.5 TESLA FULL BODY 2 SETUP SJM MRI ACTIVATOR HANDHELD DEVICE The SJM MRI Activator handheld device must be enabled
More informationAdult Advanced Cardiovascular Life Support 2015 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular
Adult Advanced Cardiovascular Life Support 2015 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care 1 DR. Alireza Abootalebi Assistant Professor Of
More informationICD Implantation Patient Information
Melbourne Heart Rhythm ICD Implantation Patient Information The Heart The heart is a pump responsible for maintaining blood supply to the body. It has four chambers. The two upper chambers (the right atrium
More informationQUICK SETUP GUIDE SECULIFE DFBASE
QUICK SETUP GUIDE SECULIFE DFBASE SECULIFE DFBASE is a microprocessor-based instrument that is used in the testing of defibrillators. It measures the energy output and provides information about the defibrillator
More informationUnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program
Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes
More informationPerioperative Management of Cardiac Implantable Devices
Financial Disclosures Breandan Sullivan MD Assistant Professor Co Director Cardiothoracic and Vascular Surgery ICU Department of Anesthesiology and Critical Care Medicine University of Colorado None Perioperative
More informationCardiac Resynchronisation Therapy Patient Information
Melbourne Heart Rhythm Cardiac Resynchronisation Therapy Patient Information Normal Heart Function The heart is a pump responsible for maintaining blood supply to the body. It has four chambers. The two
More informationNEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS
NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies
More informationDEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION
1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,
More informationPediatric Cardiac Arrest General
Date: November 15, 2012 Page 1 of 5 Pediatric Cardiac Arrest General This protocol should be followed for all pediatric cardiac arrests. If an arrest is of a known traumatic origin refer to the Dead on
More informationPacing Lead Implant Testing. Document Identifier
Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant
More informationNew and Future Trends in EMS. Ron Brown, MD, FACEP Paramedic Lecture Series 2018
New and Future Trends in EMS Ron Brown, MD, FACEP Paramedic Lecture Series 2018 New technologies and protocols DSD Mechanical Compression ITD BiPAP Ultrasound Double Sequential Defibrillation Two defibrillators
More information7 Cedar Street, Suite 202 Sudbury, ON P3E 1A
7 Cedar Street, Suite 202 Sudbury, ON P3E 1A2 705.675.4783 To: Northeast EMS Providers CC: J. Cruickshank, J. Bradbury From: Dr. Jason Prpic, Regional Medical Director Re: Deployment and/or Redeployment
More informationPERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists
PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical
More informationSupplemental Material
Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement
More informationCardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.
Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. The Miracle of Living February 21, 2018 Matthew Ostrom MD,FACC,FHRS Division of
More informationImproving quality of resuscitation. Defibrillators TEC-5600 series
Improving quality of resuscitation Defibrillators TEC-5600 series Transitioning back to Life Improving quality of resuscitation Saving Life Shock Efficiency on Demand Fast shocks and continuous observation
More informationSIX PACK ABS Item No INSTRUCTION MANUAL. Read entire manual before operating this product. Use only as directed.
SIX PACK ABS Item No. 206098 INSTRUCTION MANUAL Read entire manual before operating this product. Use only as directed. WARNINGS If you are in the care of a physician, consult your physician before using
More informationBEDSIDE ECG INTERPRETATION
BEDSIDE ECG INTERPRETATION Presented by: Ryan Dean, RN, MSN, CCRN, CCNS, CFRN Flight Nurse 2017 Based on presentations originally by Gennifer DePaoli, RN Objectives Hospital policies Electrical conduction
More informationArrhythmias. Pulmonary Artery
Arrhythmias Introduction Cardiac arrhythmia is an irregularity of the heart beat that causes the heart to beat too slowly, too fast, or irregularly. There are different types of arrhythmias. Most arrhythmias
More informationINTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network
INTERNAL CARDIOVERSION Lancashire & South Cumbria Cardiac Network DC Cardioversion AF & AFL safe 1,2 efficacious 3,4 (60-94%) 5 SR - Increases exercise tolerance 6 Maintainence SR unlikely in patients
More information1. Normal sinus rhythm 2. SINUS BRADYCARDIA
1. Normal sinus rhythm 2. SINUS BRADYCARDIA No signs and symptoms observe There are severe signs or symptoms o What are the signs and symptom Hypotension
More informationTranscutaneous Electrical Nerve Stimulation Device GF-3 / GF-3T
Transcutaneous Electrical Nerve Stimulation Device GF-3 / GF-3T Operation Manual Read Before Using GF-3-INS-LAB-RevA08 TABLE OF CONTENTS INTRODUCTION TO TENS INDICATIONS AND CONTRAINDICATIONS WARNINGS
More informationPacing Codes and Modes Concepts
Pacing Codes and Modes Concepts Pacing codes and modes concepts Objectives Upon completion of this program the participant will be able to: State what the first four positions of the NBG code represent.
More informationKadlec Regional Medical Center Cardiac Electrophysiology
Definition of electrophysiology study and ablation Kadlec Regional Medical Center Cardiac Electrophysiology Electrophysiology Study and Ablation An electrophysiology, or EP, study is a test of the heart
More informationRevised 3/3/2008 References provided on last slide
/ Medics 12 Lead ECG and STEMI Triage Protocols 5.1 & 4.27 Revised 3/3/2008 References provided on last slide Objectives Purpose of Pre-Hospital 12-Lead ECGs Pre-Hospital 12-Lead ECG Background 12-Lead
More informationTASER and In-Custody Deaths
TASER and In-Custody Deaths Cardiac Effects of Conducted Electrical Weapons Patrick Tchou, MD Heart and Vascular Institute What is a stun gun? A device that delivers rapid electrical impulses that stimulates
More informationFinally, a pacemaker may be either permanent or temporary, which will also factor into your code selections.
2015 Cardiology Survival Guide Chapter 15: Pacemakers With more than 30 codes and many variables to choose from, you must weigh your options carefully when reporting insertion, revision, or removal of
More informationCardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up
Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a
More informationTemporary Pacemakers
07 Part I Anaesthesia Refresher Course 2018 University of Cape Town Dr Owen Porrill Private Practice Honorary lecturer- University of Cape Town Temporary pacemakers are usually inserted as part of the
More information