Introduction. Case Report ECG & EP CASES ABSTRACT
|
|
- Marilynn Turner
- 5 years ago
- Views:
Transcription
1 Successful extraction of an implantable cardioverter-defibrillator lead in a patient with pocket infection via the femoral approach with a basket snare Jin-Bae Kim, MD, PhD. Cardiology Division, Department of Internal Medicine, Kyung Hee University College of Medicine, Seoul, Korea ABSTRACT A 27-year-old man was admitted to our institution with implantable cardioverter-defibrillator (ICD) pocket infection. The ICD (single chamber, dual shock coils) was implanted for the prevention of aborted sudden cardiac death due to idiopathic ventricular fibrillation. Because we could not use locking stylet and countertraction manuever, baseket snare with metal wire should be hanged at the maximal distal site of lead. With repeated simple traction, the ICD lead was completely removed. The femoral approach with snares and kits is not the first choice of lead extraction. It is indicated in cases of lead breakage, cases where other techniques fail to extract leads from the superior veins, and cases where application of excessive force to the mechanical sheaths is to be avoided. Therefore, we recommend the femoral approach with basket snare should be used in some selective cases. Key words: device removal implantable-cardioverter defibrillator infection Introduction The application of implantable cardioverterdefibrillators (ICDs) for prevent of sudden cardiac death is well established. With the increase in patients with electronic devices and leads, removal procedures are also increasing steadily. Lead extraction involving simple traction and countertraction techniques that are performed by using a locking stylet and polypropylene dilator sheath is the only method available in Korea. Received: May 26, 2012 Revision Received: July 05, 2012 Accepted: August 15, 2012 Correspondence: Jin-Bae Kim, MD, PhD, Cardiology Division, Department of Internal Medicine, Kyung Hee University College of Medicine, Seoul, Korea Tel: , Fax: , jinbbai@khu.ac.kr However, this conventional method cannot be performed in some complicated cases. In such cases, another extraction instrument is inevitably required. We report the case of a patient with an infection at the pocket site who had undergone lead extraction with a basket snare via the femoral approach. Case Report A 27-year-old man visited our division for the evaluation of a color changes at the ICD pocket site. He had received ICD implantation 13 months before. An ICD (single chamber, dual coils; Virtuoso DR ó D164AWG; Medtronic Inc., Minneapolis, MN, 44 The Official Journal of Korean Heart Rhythm Society
2 A B Figure 1. Patient's photograph showing the color change at the site of pocket infection and abscess formation (A, B). USA) was implanted for the treatment of aborted sudden cardiac death due to idiopathic ventricular fibrillation. His body temperature, pulse rate, and blood pressure were recorded 36.5 C, 78 beats/min, and 120/80 mmhg, respectively. On physical examination, showed that the ICD pocket site was pale red and had an abscess (size, 4ø5 cm) however, no tenderness was observed (Figures 1A, 1B). The laboratory tests included, biochemical profile, cardiac enzyme evaluation, and a hemoglobin was also obtained. The results showed that the levels of creatine kinase MB isoenzyme, creatine kinase, and creatine kinase isoenzyme index were 1.4 ng/ml (reference range, 0.0~6.9 ng/ml), 43 U/L (reference range, 40~150 U/L), and 3.2% (reference range, 0.0~3.5%), respectively. The patient's troponin I level was found to be elevated to 0.02 ng/ml (reference range, 0.00~0.04 ng/ml), and the high-sensitivity C-reactive protein level was 1.5 mg/dl. The results of other laboratory tests were within the reference limits. After identifying the site of pocket infection, the ICD can and pocket site debris were removed. The ICD lead was extracted with surgical back-up team 3 days after the generator was removed. The ICD lead (Sprint Quattro ó 6944; Medtronic Inc.) was a tined lead, and the diameter of the lead tip and shaft (comprising the shock coil) was 2.7 mm. We performed lead extraction (Figure 2A, 2B) via the femoral approach combined with the lead-vein entry-site approach by using a basket snare. Simple traction applied via the lead vein entry site caused the lead coil to stretch. A basket snare with a metal wire was hung at the maximal distal site of lead. Furthermore, countertraction could not be used. To separate the ICD lead from the apex of the right ventricle (RV). only repeated traction force was applied to the basket snare. After applying simple traction with a basket snare for an extended period, the ICD lead was successfully removed without any evidence of lead fragments remaining in the RV. The extracted lead was found to be encapsulated with fibrous tissue and had an elongated shock coil (Figure 3). Monitoring of hemodynamic parameters and fluoroscopic examination of the cardiac silhouette were performed regulary VOL.13 NO.2 45
3 A B Figure 2. Basket snare (A) and a reversible loop formed by the basket snare around the lead (B). Figure 3. Extracted lead surrounded by the encapsulating fibrous tissue. 46 The Official Journal of Korean Heart Rhythm Society
4 during the procedure. No evidence of major complications such as hemo-pericardium, hemothorax, or myocardial inversion was observed. We planned the insertion of a new lead after the success of ICD lead extraction was confirmed. However, the patient and his guardians did not consent to the procedure. Therefore, we discharged the patient after complete treatment of the infection and achievement of wound closure. Discussion Transvenous pectoral implantation of an active can has become the standard technique for ICD placement. The complication rates involving leads have been increasing with the increase in the number of ICD and cardiac resynchronization therapy defibrillator (CRT-D) implantations. Lead extraction is a high-risk procedure, with a morbidity rate of 1.4~2.5%. 1, 2 ICD lead extraction is known to have a higher complication rate than pacemaker lead extraction. 3 Segments of chronically implanted leads are usually encapsulated with fibrous tissue that adheres to the major veins or ventricular wall. Lead extraction techniques are designed to free the lead from the encapsulating fibrous tissue or to free the encapsulating tissue from the vein or the heart wall. Major complications such as hemopericardium, hemothorax, and death may occur during the dissection of fibrous tissue around implanted leads. 2 To increase the success rate of procedures involving complete lead removal, various extraction techniques have been used, including direct traction using rotational forces; 4 traction with a locking stylet; and countertraction using a dilating sheath, femoral workstation, laser sheath, or electrosurgical sheath. Traction and countertraction performed using a locking stylet and dilating sheath is a conventional technique used in pacemaker lead extraction and ICD lead extraction techinques 5 and has been proven to be effective and safe in clinical trials. However, this conventional kit was not available in Korea when we performed this procedure. Therefore, we performed lead extraction via the femoral approach with a basket snare. Various snares have been used in lead extraction procedures. A snare has to have enough tensile strength to support the forces required for the lead extraction procedure and provide force for aligning the target lead in different directions. Although basket snares were not developed for lead extraction, they can be used to form a reversible loop around the target lead. Forming a reversible loop is a complicated technique that involves a few specialized kits and procedures that require repeated practice to achieve perfection. 6 By using a basket snare with a metal wire, we easily created a reversible loop around the ICD lead, thereby providing more efficient traction to the lead. Currently, the femoral approach involving snares and kits is not considered a first-line procedure for lead extraction. It is indicated in cases of lead breakage, cases where other techniques fail to extract leads from the superior veins, and cases where application of excessive force to the mechanical sheaths is to be avoided. Therefore, we recommend that lead extraction be performed via the femoral approach using basket snares only in select cases. References 1. Smith HJ, Fearnot NE, Byrd CL, Wilkoff BL, Love CJ, Sellers TD. Five-years experience with intravascular lead extraction. U.S. Lead Extraction Database. Pacing Clin Electrophysiol. 1994;17: Byrd CL, Wilkoff BL, Love CJ, Sellers TD, Turk KT, Reeves R, Young R, Crevey B, Kutalek SP, Freedman R, Friedman R, Trantham J, Watts M, Schutzman J, Oren J, Wilson J, Gold F, Fearnot NE, Van Zandt HJ. Intravascular extraction of problematic or infected permanent pacemaker leads: U.S. Extraction Database, MED Institute. Pacing Clin Electrophysiol. VOL.13 NO.2 47
5 1999;22: Hauser RG, Katsiyiannis WT, Gornick CC, Almquist AK, Kallinen LM. Deaths and cardiovascular injuries due to device-assisted implantable cardioverter-defibrillator and pacemaker lead extraction. Europace. 2010;12: Rosenheck S, Weiss A, Leibowitz D, Sharon Z. Noninstrumental pacemaker and defibrillator lead removal: the importance of the rotation forces. Pacing Clin Electrophysiol. 2002;25: Alt E, Neuzner J, Binner L, Göhl K, Res JC, Knabe UH, Zehender M, Reinhardt J. Three-year experience with a stylet for lead extraction: a multicenter study. Pacing Clin Electrophysiol. 1996;19: Klug D, Jarwé M, Messaoudéne SA, Kouakam C, Marquié C, Gay A, Lacroix D, Kacet S. Pacemaker lead extraction with the needle's eye snare for countertraction via a femoral approach. Pacing Clin Electrophysiol. 2002;25: The Official Journal of Korean Heart Rhythm Society
Results of transvenous lead extraction of coronary sinus leads in patients with cardiac 4,703 resynchronization therapy
Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich www.zora.uzh.ch Year: 2013 Results of transvenous lead extraction of coronary sinus leads in patients
More informationRiata lead extraction- a single centre experience
Riata lead extraction- a single centre experience Rebecca L Noad, Keith W Morrice, Vivek N Kodoth, Carol M Wilson, Michael JD Roberts. Royal Victoria Hospital, Belfast, United Kingdom Background of previous
More informationPacemaker/defibrillator lead extraction: a single centre experience
Pacemaker/defibrillator lead extraction: a single centre experience Pier Giorgio GOLZIO Division of Cardiology, Department of Internal Medicine University of Turin, Azienda Ospedaliero Universitaria San
More informationNewer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.
Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded
More informationLead extraction. Dr. Mervat Abo El Maaty Professor of Cardiology Ain shams University 2013
Lead extraction Dr. Mervat Abo El Maaty Professor of Cardiology Ain shams University 2013 Agenda Introduction History of consensus Definitions Complications Indications Lead management environment Extraction
More informationVisualization of Critical Isthmus by Delayed Potential Mapping
Visualization of Critical Isthmus by Delayed Potential Mapping Yong-Seog Oh, MD, PhD Cardiovascular Center, Seoul St. Mary s Hospital, Catholic University of Korea ABSTRACT A 57-year-old man was admitted
More informationSupplemental Material
Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement
More informationReview Article The Lead Extractor's Toolbox: A Review Of Current Endovascular Pacemaker And ICD Lead Extraction Techniques
www.ipej.org 101 Review Article The Lead Extractor's Toolbox: A Review Of Current Endovascular Pacemaker And ICD Lead Extraction Techniques Bracke FA, MD. Department of Cardiology, Catharina Hospital,
More informationCIEDs Infection: Lead Extraction, First or Last option?
CIEDs Infection: Lead Extraction, First or Last option? More CRM Systems & Leads 600,000 new devices annually 1.2 million new leads annually Sources: 1) Medtronic CRDM Product Performance Report, Jan 2009.
More informationIntroduction. CLINICAL RESEARCH Leads and Lead Extraction
Europace (2011) 13, 543 547 doi:10.1093/europace/euq400 CLINICAL RESEARCH Leads and Lead Extraction Percutaneous extraction of cardiac pacemaker and implantable cardioverter defibrillator leads with evolution
More informationThe Riata Implantable Cardioverter Defibrillator Lead: Extraction Experience for Conductor Exteriorization and Electrical Malfunction
The Riata Implantable Cardioverter Defibrillator Lead: Extraction Experience for Conductor Exteriorization and Electrical Malfunction Heath E. Saltzman, MD, Faiz Subzposh, MD, Christine Saari, CRNP, S.
More informationPERFORMANCE MEASURE TECHNICAL SPECIFICATIONS. HRS-3: Implantable Cardioverter-Defibrillator (ICD) Complications Rate
PERFORMANCE MEASURE TECHNICAL SPECIFICATIONS HRS-3: Implantable Cardioverter-Defibrillator (ICD) Complications Rate Measure Title Description Measure Type Data Source Level of Analysis Numerator HRS-3:
More informationSUPPLEMENTAL MATERIAL
SUPPLEMENTAL MATERIAL Table S1. Patient Selection Codes, CIED Generator Procedures Code Type Code Description ICD9 Proc 00.51 Implantation of cardiac resynchronization defibrillator, total system [CRT-D]
More informationThe techniques and tools for percutaneous removal of
Initial Experience With Larger Laser Sheaths for the Removal of Transvenous Pacemaker and Implantable Defibrillator Leads Laurence M. Epstein, MD; Charles L. Byrd, MD; Bruce L. Wilkoff, MD; Charles J.
More informationExtraction for Class II Indications - Strategic Management of Recalled CIEDs - HRS Satellite Symposium
Extraction for Class II Indications - Strategic Management of Recalled CIEDs - HRS Satellite Symposium 2015 Steven P. Kutalek, MD Director, Cardiac Electrophysiology & Pacing Drexel University College
More informationLead Extraction: Challenges in our area
Lead Extraction: Challenges in our area More CRM Systems & Leads 600,000 new devices annually 1.2 million new leads annually Sources: 1) Medtronic CRDM Product Performance Report, Jan 2009. 2) Boston Scientific
More informationRiata implantable cardioverter-defibrillator lead failure: Analysis of explanted leads with a unique insulation defect
Riata implantable cardioverter-defibrillator lead failure: Analysis of explanted leads with a unique insulation defect Robert G. Hauser, MD, FHRS, Deepa McGriff, BS, Linda Kallinen Retel, BS, FHRS From
More informationUnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:
UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Electrophysiology Implant Code Classification Table The
More informationExtraction of Old Pacemaker or Cardioverter Defibrillator Leads by Laser. Sheath Versus Femoral Approach
Extraction of Old Pacemaker or Cardioverter Defibrillator Leads by Laser Sheath Versus Femoral Approach Running title: Bordachar et al.; Laser versus snare for lead extractions Pierre Bordachar, MD, 1
More informationSuperior Vena Cava Echocardiography as a Screening Tool to Predict Cardiovascular Implantable Electronic Device Lead Fibrosis
http://dx.doi.org/10.4250/jcu.2015.23.1.27 pissn 1975-4612/ eissn 2005-9655 Copyright 2015 Korean Society of Echocardiography www.kse-jcu.org ORIGINAL ARTICLE J Cardiovasc Ultrasound 2015;23(1):27-31 Superior
More informationAs the rate of cardiac devices implanted increases rapidly
Extraction of Old Pacemaker or Cardioverter-Defibrillator Leads by Laser Sheath Versus Femoral Approach Pierre Bordachar, MD; Pascal Defaye, MD; Eric Peyrouse, MD; Serge Boveda, MD; Bilel Mokrani, MD;
More informationUnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program
Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes
More informationGirish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie
268 Case Report Cardiac Resynchronization Therapy in a Patient with Persistent Left Superior Vena Cava Draining into the Coronary Sinus and Absent Innominate Vein: A Case Report and Review of Literature
More informationRecent Advances in Pacing and Defibrillation Harish Doppalapudi, MD
Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD Harish Doppalapudi, MD Assistant Professor of Medicine Director, Clinical Cardiac Electrophysiology Training Program University of Alabama
More informationComplications of Lead Extraction: Prevention and treatment. Maria Grazia Bongiorni, MD, FESC
Complications of Lead Extraction: Prevention and treatment Maria Grazia Bongiorni, MD, FESC Director of Cardiovascular Division University Hospital of Pisa (Italy) ourtesy of Dr Eivind Platou Potential
More informationImplantable defibrillator lead extraction with optimized standard extraction techniques
Journal of Geriatric Cardiology (2013) 10: 3 9 2013 JGC All rights reserved; www.jgc301.com Research Article Open Access Implantable defibrillator lead extraction with optimized standard extraction techniques
More informationA number of large, randomized, clinical trials have demonstrated that patients left ventricular dysfunction (ejection fraction 35%) due to either
A number of large, randomized, clinical trials have demonstrated that patients left ventricular dysfunction (ejection fraction 35%) due to either ischemic or non-ischemic cardiomyopathy benefit from ICD
More informationNEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS
NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies
More informationPatient Safety: the optimal lead body design
Patient Safety: the optimal lead body design E. Soldati U.O. Malattie Cardiovascolari II Azienda Ospedaliero Universitaria Pisana Advances in Cardiac Arrhythmias Torino, 25-27 Ottobre 2012 Lead Malfunction
More informationUnexpected challenging case of coronary sinus lead extraction
W J C C World Journal of Clinical Cases Submit a Manuscript: http://www.wjgnet.com/esps/ DOI: 10.12998/wjcc.v5.i2.46 World J Clin Cases 2017 February 16; 5(2): 46-49 ISSN 2307-8960 (online) CASE REPORT
More informationDeaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads
Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads Robert G. Hauser, MD, FHRS, Raed Abdelhadi, MD, Deepa McGriff, BS, Linda Kallinen Retel, BS, FHRS From the
More informationUnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:
Notification and Prior Authorization Program: Electrophysiology Implant Classification Table The following chart contains the codes that require notification or prior authorization as part of UnitedHealthcare
More informationModel 5392 EPG Temporary Pacer
Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short
More informationWide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?
CASE REPORTS Arrhythmia 2015;16(3):173-177 doi: http://dx.doi.org/10.18501/arrhythmia.2015.029 Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? Eun-Sun Jin, MD, PhD Cardiovascular
More informationLead removal in young patients in view of lifelong pacing
Europace (2010) 12, 714 718 doi:10.1093/europace/euq059 CLINICAL RESEARCH Leads and Lead Extraction Lead removal in young patients in view of lifelong pacing Peter A. Zartner*, Walter Wiebe, Nicole Toussaint-Goetz,
More informationTransvenous Pacemaker Procedures
Cardiology: Pacemaker and Defibrillator Coding Presented By: Moderate Sedation 2017 99151 : under age 5, initial 15 minutes by MD performing intervention 99152: age 5 or older, initial 15 minutes by MD
More informationBroken leads with proximal endings in the cardiovascular system: Serious consequences and extraction difficulties
ORIGINAL ARTICLE Cardiology Journal 2013, Vol. 20, No. 2, pp. 161 169 DOI: 10.5603/CJ.2013.0029 Copyright 2013 Via Medica ISSN 1897 5593 Broken leads with proximal endings in the cardiovascular system:
More informationDifferent indications for pacemaker implantation are the following:
Patient Resources: ICD/Pacemaker Overview ICD/Pacemaker Overview What is a pacemaker? A pacemaker is a device that uses low energy electrical pulses to prompt the heart to beat whenever a pause in the
More informationLeadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt
Leadless Pacing Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt The weakest link in Pacemaker system the lead. The more the leads the more the complications Dislodgement Fracture Insulation
More informationHow adequate are the current methods of lead extraction? A review of the efficiency and safety of transvenous lead extraction methods
Europace (2015) 17, 689 700 doi:10.1093/europace/euu378 REVIEW How adequate are the current methods of lead extraction? A review of the efficiency and safety of transvenous lead extraction methods Maurits
More informationRiata Lead Extraction. Thomas D. Callahan, MD, FACC, FHRS
Riata Lead Extraction Thomas D. Callahan, MD, FACC, FHRS Outline Riata lead history Lead design Lead Failure Extraction outcomes Techniques Special considerations Riata Lead Background 8F introduced in
More informationLEAD EXTRACTION GOOD LEAD MANAGEMENT. An update on how to achieve well functioning CIED hardware with a special focus on risk-benefit analysis
LEAD EXTRACTION GOOD LEAD MANAGEMENT An update on how to achieve well functioning CIED hardware with a special focus on risk-benefit analysis ESC, Stockholm 2010-08-30 Stockholm 2010-04-16 Charles Kennergren,
More informationTechnical option of surgical approach for trouble-shooting
JHRS Corner Device and lead trouble-shooting - standard strategy and technical option - Technical option of surgical approach for trouble-shooting Katsuhiko IMAI Department of Cardiovascular surgery, Hiroshima
More information2992 YOSHITAKE T et al. Circ J 2018; 82: ORIGINAL ARTICLE doi: /circj.CJ
2992 YOSHITAKE T et al. Circ J 2018; 82: 2992 2997 ORIGINAL ARTICLE doi: 10.1253/circj.CJ-18-0869 Arrhythmia/Electrophysiology Safety and Efficacy of Transvenous Lead Extraction With a High-Frequency Excimer
More informationLead Extraction Workshop. By Cook Medical
Lead Extraction Workshop By Cook Medical Lead Extraction Lead Extraction is the removal from the body of implanted cardiac leads. Cardiac leads: They are conductor wires that electrically connect the implanted
More informationImplantable cardioverter defibrillator, Inappropriate shock, Lead failure
Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,
More informationCardiac Rhythm Management Coder 2017
Cardiac Rhythm Management Coder 2017 An easy-to-use tool for coding and reimbursement compliance Prepared and Published By: MedLearn Publishing, A Division of Panacea Healthcare Solutions, Inc. 287 East
More informationDevice Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI
Device Interrogation- Pacemakers, ICD and Loop Recorders Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Disclosures Consultant: Medtronic Speaker s Bureau: St. Jude Medical
More informationCase Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement
www.ipej.org 49 Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement Naga VA Kommuri, MD, MRCPCH, Sri Lakshmi S Kollepara, MD,
More informationA Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study
A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study ANEESH V. TOLAT, M.D.,* MELISSA WOICIECHOWSKI, M.S.N.,* ROSEMARIE KAHR, R.C.I.S.,* JOSEPH
More informationIHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012
IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT201203 JANUARY 24, 2012 The IHCP to reimburse implantable cardioverter defibrillators separately from outpatient implantation Effective March 1, 2012, the
More informationStuck dialysis catheters. ANZSIN 2013 Michael Lam & Kendal Redmond
Stuck dialysis catheters ANZSIN 2013 Michael Lam & Kendal Redmond NT 39 yr old CI Maori - ESKD 2 o to cortical necrosis HD August 2002 R IJ tunneled Tesio catheter Oct 2002 Failed L RC AVF Feb 2004 Failed
More informationCardiac Rhythm Device Summit 2016
PROGRAM Friday, June 17, 2016 6:45 a.m. Registration and Continental Breakfast 7:20 a.m. Welcome and Pre-test David Hayes, M.D. Device Therapy: Basics, Guidelines, and Future Technologies Moderator: David
More informationProcurement Support to Ministry of Health, Ukraine. Invitation to Bid for:
Procurement Support to Ministry of Health, Ukraine Invitation to Bid for: PROGRAMME XIX: MEDICAL PRODUCTS FOR HEALTHCARE INSTITUTIONS FOR TREATMENT OF PATIENTS WITH CARDIOVASCULAR AND CEREBROVASCULAR DISEASES
More informationCardiac Rhythm Management Coder 2018
Cardiac Rhythm Management Coder 2018 An easy-to-use tool for coding and reimbursement compliance Prepared and Published By: MedLearn Publishing, A Division of MedLearn Media, Inc. 445 Minnesota Street,
More informationHow Long Should Prophylactic Antibiotics be Prescribed for Permanent Pacemaker Implantations? One Day versus Three Days
Original Article Acta Cardiol Sin 2013;29:341 346 EP & Arrythmia How Long Should Prophylactic Antibiotics be Prescribed for Permanent Pacemaker Implantations? One Day versus Three Days Kuang-Hsing Chiang,
More informationSame-day contralateral implantation of a permanent device after lead extraction for isolated pocket infection
Europace (2014) 16, 252 257 doi:10.1093/europace/eut220 CLINICAL RESEARCH Pacing and resynchronization therapy Same-day contralateral implantation of a permanent device after lead extraction for isolated
More informationMY CONFLICTS OF INTEREST ARE
MY CONFLICTS OF INTEREST ARE Consulting Spectranetics, St Jude Research Support Spectranetics Advisory Board Spectranetics S L I D E 1 When Devices Go Bad!! S L I D E 2 ICD Erosion Secondary to Pocket
More informationHis Bundle Pacing in Bundle Branch Block May 11, 2017
His Bundle Pacing in Bundle Branch Block May 11, 2017 Gopi Dandamudi, MD FHRS System Medical Director, IUH Cardiac EP Program Director, IUH Atrial Fibrillation Center Assistant Professor of Clinical Medicine
More informationImplantation of a biventricular implantable cardioverter-defibrillator guided by an electroanatomic mapping system
Europace (2012) 14, 107 111 doi:10.1093/europace/eur250 CLINICAL RESEARCH Pacing and Resynchronization Therapy Implantation of a biventricular implantable cardioverter-defibrillator guided by an electroanatomic
More informationCRT Implantation Techniques 부천세종병원순환기내과박상원
Cardiac Venous System and CRT Implantation Techniques 부천세종병원순환기내과박상원 Cardiac Resynchronization Therapy (CRT) Goal: Atrial synchronous biventricular pacing Transvenous approach for left ventricular lead
More informationImplantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect
The Ochsner Journal 10:27 31, 2010 f Academic Division of Ochsner Clinic Foundation Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect Anas Bitar, MD, Maria Malaya
More informationCardiac Resynchronisation Therapy Patient Information
Melbourne Heart Rhythm Cardiac Resynchronisation Therapy Patient Information Normal Heart Function The heart is a pump responsible for maintaining blood supply to the body. It has four chambers. The two
More informationImplantation of additional subcutaneous array electrode reduces defibrillation threshold in ICD patients preliminary results
Clinical research Implantation of additional subcutaneous array electrode reduces defibrillation threshold in ICD patients preliminary results Maciej Kempa, Szymon Budrejko, Łukasz Drelich, Tomasz Królak,
More informationMRI ACCESS. Evera MRI SureScan ICD Systems
ACCESS Evera SureScan ICD Systems ACCESS. 36% of ICD patients are likely to have an ordered over 4 years 3 Introducing Evera, featuring SureScan Technology. ICD Patients are not receiving s 1.4% 36% CONTOURED.
More informationL: Cardiovascular. Saskatchewan Association of Licensed Practical Nurses, Competency Profile for LPNs, 3rd Ed. 107
L: Cardiovascular Saskatchewan Association of Licensed Practical Nurses, Competency Profile for LPNs, 3rd Ed. 107 Major Competency Area: L Cardiovascular Competency: L-1 Cardiovascular Nursing Date: January
More informationUnderstanding Atrial Fibrillation A guide for patients
Understanding Atrial Fibrillation A guide for patients Your doctor has determined that you have atrial fibrillation (AF), a common disturbance of the heart s rhythm. This pamphlet will answer many of your
More informationComparison of clinical trials evaluating cardiac resynchronization therapy in mild to moderate heart failure
HOT TOPIC Cardiology Journal 2010, Vol. 17, No. 6, pp. 543 548 Copyright 2010 Via Medica ISSN 1897 5593 Comparison of clinical trials evaluating cardiac resynchronization therapy in mild to moderate heart
More informationPERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS
PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing
More informationProgramming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany
Programming of Bradycardic Parameters C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Programming of ICD Brady Parameters Conflict of Interest Biotronik
More informationA rare case of acute myocardial infarction during extraction of a septally placed implantable cardioverter-defibrillator lead
Accepted Manuscript A rare case of acute myocardial infarction during extraction of a septally placed implantable cardioverter-defibrillator lead Eric Wierda, MD, LLM, Astrid A. Hendriks, MD, Giovanni
More informationCardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.
Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. The Miracle of Living February 21, 2018 Matthew Ostrom MD,FACC,FHRS Division of
More informationAssessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis
Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis Investigator: Keiko Saito, MD Mentor: Yuji Saito, MD, PhD, FACP, FACC Department
More informationKey Words: Balloon Venoplasty of Subclavian Vein, Cardiac Resynchronisation Therapy. Case report
221 Case Report Balloon Venoplasty of Subclavian Vein and Brachiocephalic Junction to Enable Left Ventricular Lead Placement for Cardiac Resynchronisation Therapy Thanh Trung Phan, Simon James, Andrew
More informationAutomatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms
Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable
More informationThe wearable cardioverter defibrillator as a bridge to reimplantation in patients with ICD or CRT-D-related infections
Castro et al. Journal of Cardiothoracic Surgery (2017) 12:99 DOI 10.1186/s13019-017-0669-2 RESEARCH ARTICLE The wearable cardioverter defibrillator as a bridge to reimplantation in patients with ICD or
More informationPACING AT THE BUNDLE OF HIS
PACING AT THE BUNDLE OF HIS Cardiac Rhythm Heart Failure Coding Corner Disclaimer While Medtronic offers opinions about how to code when using our technology according to its FDA-approved labeling, these
More informationPacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report
Case Report Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Taku Nishida MD, Tamio Nakajima MD, PhD, Yutaka Goryo MD, Ken-ichi Ishigami MD, PhD, Hiroyuki Kawata
More informationPrimary prevention ICD recipients: the need for defibrillator back-up after an event-free first battery service-life
Chapter 3 Primary prevention ICD recipients: the need for defibrillator back-up after an event-free first battery service-life Guido H. van Welsenes, MS, Johannes B. van Rees, MD, Joep Thijssen, MD, Serge
More informationPERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists
PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical
More informationMETHODS OBJECTIVES BACKGROUND METHODS RESULTS CONCLUSIONS
Journal of the American College of Cardiology Vol. 33, No. 6, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(99)00074-1 Pacemaker
More informationICD leads survival and troubles in the last 10 years
XVI Symposium on Progress in Clinical Pacing, Rome, Italy - Dec 2 to 7, 2014 ICD leads survival and troubles in the last 10 years Antonio Raviele, MD, FESC, FHRS President ALFA Alliance to Fight Atrial
More informationDisclosures. Optimal Lead Selection: An Extractor s Guide to Lead Choice and Implant Technique. Extraction Experts Have Learned!
1 Optimal Lead Selection: An Extractor s Guide to Lead Choice and Implant Technique Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services Wexner Medical
More informationClinical Study Percutaneous Extraction of Transvenous Permanent Pacemaker/Defibrillator Leads
BioMed Research International, Article ID 949785, 6 pages http://dx.doi.org/10.1155/2014/949785 Clinical Study Percutaneous Extraction of Transvenous Permanent Pacemaker/Defibrillator Leads Stylianos Paraskevaidis,
More informationUpgrade to Resynchronization Therapy. Saeed Oraii MD, Cardiologist Interventional Electrophysiologist Tehran Arrhythmia Clinic May 2016
Upgrade to Resynchronization Therapy Saeed Oraii MD, Cardiologist Interventional Electrophysiologist Tehran Arrhythmia Clinic May 2016 Event Free Survival (%) CRT Cardiac resynchronization therapy (CRT)
More informationPacing Lead Implant Testing. Document Identifier
Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant
More informationImplantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević
Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular
More informationOHTAC Recommendation: Internet- Based Device-Assisted Remote Monitoring of Cardiovascular Implantable Electronic Devices
OHTAC Recommendation: Internet- Based Device-Assisted Remote Monitoring of Cardiovascular Implantable Electronic Devices Ontario Health Technology Advisory Committee January 2012 Issue Background The Ontario
More informationPOLICIES AND PROCEDURE MANUAL
POLICIES AND PROCEDURE MANUAL Policy: MP140 Section: Medical Benefit Policy Subject: Automatic Implantable Cardioverter-Defibrillator I. Policy: Automatic Implantable Cardioverter-Defibrillator II. Purpose/Objective:
More informationCharging Considerations. Agenda 1/30/2013
February 2013 Jean C. Russell, MS, RHIT jrussell@epochhealth.com Richard Cooley, BA, CCS rcooley@epochhealth.com 518-430-1144 2 Agenda Cardiology special considerations 2013 Cardiology Codes Deleted Codes
More informationFebruary Jean C. Russell, MS, RHIT Richard Cooley, BA, CCS
February 2013 Jean C. Russell, MS, RHIT jrussell@epochhealth.com Richard Cooley, BA, CCS rcooley@epochhealth.com 518-430-1144 2 Agenda Cardiology special considerations 2013 Cardiology Codes Deleted Codes
More informationCardiovascular Implantable Electronic Device Leads & Arteriovenous Hemodialysis Access
Cardiovascular Implantable Electronic Device Leads & Arteriovenous Hemodialysis Access Theodore F. Saad, MD Nephrology Associates, PA Christiana Care Health System Newark, Delaware Cardiovascular Implantable
More informationPacemakers, Defibrillators, and Electrophysiology in 2012
Pacemakers, Defibrillators, and Electrophysiology in 2012 Presented by: David B. Dunn, MD, FACS, CIRCC, CPC-H, CCC National Coding Standards Sources of information Centers for Medicare & Medicaid (CMS)
More informationPacemaker and Internal Cardioverter Defibrillator Lead Extraction: A Safe and Effective Surgical Approach
ORIGINAL ARTICLES: SURGERY: The Annals of Thoracic Surgery CME Program is located online at http://cme.ctsnetjournals.org. To take the CME activity related to this article, you must have either an STS
More informationPacing Without Wires: Leadless Cardiac Pacing
REVIEWS AND CONTEMPORARY UPDATES Ochsner Journal 16:238 242, 2016 Ó Academic Division of Ochsner Clinic Foundation Pacing Without Wires: Leadless Cardiac Pacing Michael L. Bernard, MD, PhD Department of
More informationevicore cardiology procedures and services requiring prior authorization
evicore cardiology procedures and services requiring prior authorization Moda Health Commercial Group and Individual Members* *Check EBT to verify member enrollment in evicore program Radiology Advanced
More information12 YEARS PROVEN PERFORMANCE. Sprint Quattro Defibrillation Leads
12 YEARS PROVEN PERFORMANCE Sprint Quattro Defibrillation Leads PROVEN PERFORMANCE. PROVEN RELIABLE. We made intentional choices and took proactive steps from lead design to active product monitoring to
More informationCigna - Prior Authorization Procedure List Cardiology
Cigna - Prior Authorization Procedure List Cardiology Category CPT Code CPT Code Description 33206 Insertion of new or replacement of permanent pacemaker with transvenous electrode(s); atrial 33207 Insertion
More informationFinally, a pacemaker may be either permanent or temporary, which will also factor into your code selections.
2015 Cardiology Survival Guide Chapter 15: Pacemakers With more than 30 codes and many variables to choose from, you must weigh your options carefully when reporting insertion, revision, or removal of
More informationDual-Chamber Implantable Cardioverter-Defibrillator
February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary
More information