Patient-Specific Computational Analysis of Transvenous Defibrillation: A Comparison to Clinical Metrics in Humans

Size: px
Start display at page:

Download "Patient-Specific Computational Analysis of Transvenous Defibrillation: A Comparison to Clinical Metrics in Humans"

Transcription

1 Annals of Biomedical Engineering, Vol. 32, No. 6, June 2004 ( 2004) pp Patient-Specific Computational Analysis of Transvenous Defibrillation: A Comparison to Clinical Metrics in Humans DANIEL MOCANU, 1 JOACHIM KETTENBACH, 2 MICHAEL O. SWEENEY, 3 RON KIKINIS, 2 BRUCE H. KENKNIGHT, 4 and SOLOMON R. EISENBERG 1 1 Department of Biomedical Engineering, Boston University, 44 Cummington Street, Boston, MA; 2 Surgical Planning Laboratory, Brigham and Women s Hospital, 75 Francis Street, Boston, MA; 3 Cardiac Pacing and Implantable Device Therapies, Brigham and Women s Hospital, 75 Francis Street, Boston, MA; and 4 Heart Failure Research, Guidant Corporation, St. Paul, MN (Received 3 March 2003; accepted 29 January 2004) Abstract The goal of this study is to assess the predictive capacity of computational models of transvenous defibrillation by comparing the results of patient-specific simulations to clinical defibrillation thresholds (DFT). Nine patient-specific models of the thorax and in situ electrodes were created from segmented CT images taken after implantation of the cardioverter-defibrillator. The defibrillation field distribution was computed using the finite volume method. The DFTs were extracted from the calculated field distribution using the 95% critical mass criterion. The comparison between simulated and clinical DFT energy resulted in a rms difference of 12.4 J and a 0.05 correlation coefficient (cc). The model-predicted DFTs were well matched to the clinical values in four patients (rms = 1.5 J;cc=0.84). For the remaining five patients the rms difference was 18.4 J with a cc = These results suggest that computational models based soley on the critical mass criterion and a single value of the inexcitability threshold are not able to consistently predict DFTs for individual patients. However, inspection of the weak potential gradient field in all nine patients revealed a relationship between the degree of dispersion of the weak field and the clinical DFT, which may help identify high DFT patients. Keywords Transvenous defibrillation, Critical mass, Patientspecific, Modeling, Finite volume method. INTRODUCTION Ventricular fibrillation (VF) is a severe heart arrhythmia that can lead to sudden cardiac death if not treated promptly. The only effective clinical intervention to extinguish VF is electrical defibrillation. The implantable cardioverter defibrillator (ICD) is an electronic device designed to detect the onset of VF and to shock the heart back into normal sinus rhythm. ICDs have been shown to be very effective in protecting against sudden cardiac death and have become the treatment of choice for patients with drug-resistant arrhythmias. In the standard configuration, the ICD is surgi- Address correspondence to Solomon R. Eisenberg, Sc. D., Department of Biomedical Engineering, Boston University, Boston, MA Electronic mail: sre@bu.edu 775 cally implanted subcutaneously in the patient s chest, with two catheter electrodes inserted in the superior vena cava (SVC) and the right ventricle (RV) (Fig. 1). Typically, the shock energy is delivered via a dual-current pathway from the RV electrode to the SVC electrode and the metallic enclosure (CAN) of the ICD. Determining energy settings during ICD implantation requires repetitive sequences of induction and termination of VF to find the defibrillation threshold (DFT) energy. 15 Since each VF induction has some element of risk, including the possibility of nonconversion and death, the ICD implantation procedure would be substantially improved if an accurate estimate of the patient s DFT energy requirement was availabe prior to implant. Anatomically realistic computational models have been used previously to model and explore the current distribution produced by defibrillation shocks. 1,5,11,17 Jorgenson et al. 10 tested the accuracy of simulated defibrillation fields in six pig thoracic models and found a good correlation between model predicted voltages and the experimental data measured at 52 sites inside the thorax. Panescu et al. 16 also found a good correlation between measured voltages on the thorax surface and finite element model predictions. These results suggest that computational models with realistic geometries can closely simulate the field and current distributions produced by defibrillation shocks. The critical mass criterion 24 has been used in a number of modeling studies 6,13,14 to define successful defibrillation and to extract defibrillation metrics from the calculated field distribution. This criterion states that a minimum electric field E th is necessary throughout a critical mass of the ventricular myocardium to terminate fibrillation activation waves. These investigators have reported model DFT values that fall within one standard deviation of the mean DFTs reported in human clinical studies. The success of these computational studies suggested that similar computational models may be useful in the /04/ /1 C 2004 Biomedical Engineering Society

2 776 MOCANU et al. FIGURE 1. X-ray image showing the ICD with the catheter RV and SVC electrodes. presurgical planning of ICD implantation by providing an estimate of the DFT for individual patients. Thus, the goal of our research was to assess the predictive capacity of patientspecific computational models of transvenous defibrillation by comparing patient-specific simulated and clinical defibrillation metrics. In contrast with previous modeling studies, which have been based on a single average human thorax, we used a set of nine computational models built from segmented CT images of patients with an implanted ICD. The clinical defibrillation parameters were determined for each patient at the time of implantation and then used in the evaluation of the modeling results. We also present an analysis of the distribution of the weak field and its relationship to the DFT. METHODS Clinical DFT Testing Nine patients were recruited for implant (patient demographics shown in Table 1). Clinical DFT testing was performed with an active can emulator (CPI Ventak ECD, model 2185; 150 µf capacitance) and followed a standardized step-down clinical protocol. 19 VF was induced by applying a pulse of 60 Hz alternating current and the defibrillation shock was delivered 10 s later. The defibrillation waveform had a biphasic shape, with 60% tilt in the positive phase and 50% tilt in the negative phase (Fig. 2). The corresponding pulse width of the first and second phases was 60% and 40%, respectively, of the total pulse duration (10 15 ms). Typical trials started at 20 J (stored energy) and decremented until VF was no longer terminated using the following step-down scheme: 20 J 15 J 10 J 8 J 5 J. In cases where a 20 J shock could not defibrillate, the energy was gradually increased in 5 J increments until defibrillation was achieved. The ICD output was programmed at the minimum energy that successfully defibrillated on two occasions, single confirmation, plus a safety margin. The single confirmation could be consecutive, i.e. two consecutive at 5 J, or by way of a single reversal, i.e. 8 J success, 5 J failure, 8 J success. However, for the purposes of the modeling study, the minimum delivered energy that defibrillated on a single occasion was defined as the DFT and used in the comparison with model-predicted values. Finally, in all but one of the patients, the Endotak catheter lead system (Guidant/CPI) was implanted. A similar lead system (Medtronic) was used in the remaining patient. The ICDs were implanted in the left pectoral region with venous access via the subclavian vein (Fig. 1). Fluoroscopic imaging was used to verify the correct lead placement. CT Image Acquisition and Segmentation All patients were scanned post-implant on a spiral X-ray CT system (Somatom Plus 4, Siemens Medical Systems, NJ), with the ICD, SVC, and RV electrodes in place. The imaging protocol was approved by the Internal Review Board at Brigham and Women s Hospital, Boston, and each patient gave informed consent. The scanning TABLE 1. Patient demographics. Heart muscle Heart Patient # Gender Age volume a (cm 3 ) disease Heart arrhythmia Drugs 1 M CAD Primary prevention B-blocker 2 M CAD Spontaneous sustained VT Amiodarone Digitalis B-blocker 3 M CAD VF or cardiac arrest B-blocker 4 M CAD Primary prevention Amiodarone 5 M NDCP Spontaneous syncopal VT 6 M NDCP Spontaneous syncopal VT Digitalis 7 F CAD VF or cardiac arrest Digitalis B-blocker 8 M CAD Spontaneous sustained VT 9 M NDCP Spontaneous sustained VT Amiodarone B-blocker Note. CAD coronary artery disease; NDCP nonischemic dilated cardiomyopathy; VF ventricular fibrillation; VT ventricular tachycardia. a The heart muscle volume (ventricles and atria) was estimated based on the number and volume of the heart muscle voxels in the segmented CT images.

3 Patient-Specific Modeling of Defibrillation 777 FIGURE 2. CPI/Guidant biphasic defibrillation waveform with 60% tilt in the positive phase and 50% tilt in the negative phase. The pulse width of the first phase is 60% of the total pulse duration (typically ms). The DFT voltage is denoted with V th. procedure 12 was customized for high resolution image acquisition (120 kv, 200 ma, matrix size , mm/sec table feed). For better differentiation of soft tissues, a contrast agent was delivered to each patient prior to imaging (Omnipaque 300; Iohexol 300 g I/ml; Nycomed, Norway). To provide images of high contrast and a low noise level, an anisotropic filter was applied prior to the segmentation. Each segmented label was overlaid to the corresponding grayscale images to ensure accurate positioning. If the results of the semiautomated segmentation were not satisfying, manual fine-editing was done with MrX (General Electric Medical Systems, Milwaukee, WI) using a Sun workstation (UltraSPARC, Sun Microsystems Inc., Mountain View, CA). The following tissue types were identified: fat, blood, heart muscle, thoracic wall muscle, lung, bone (Fig. 3). The RV and SVC electrodes together with the ICD were also identified. Image-Based Computational Model For each patient, the 3-D computational model was constructed directly from the segmented images ( FIGURE 4. Voxel-based finite volume mesh. For clarity, only the bone structure and the lungs are shown. pixels) using a structured meshing algorithm in which the segmented voxels in the image data sets were defined as volume elements in the computational domain (Fig. 4). The size of each volume element was roughly 3 3 6mm 3, with slight variations depending on individual patient geometrical features. The total number of elements in the models varied between 350,000 and 450,000. Electrical conductivities 8,20 were assigned to the six tissue types according to Table 2. Conduction Boundary Value Problem In the quasistatic approximation of the electromagnetic field, 9 the electric potential distribution during a defibrillation shock is governed by Laplace s equation subject to Dirichlet boundary conditions on the electrodes (RV, SVC, and CAN) and Neumann homogeneous boundary condition (no flux) on the thoracic surface. The equation was solved numerically by the finite volume method using I-DEAS software package (Integrated Design and Engineering Analysis Software, Structural Dynamics Research Corporation, Milford, OH, USA). Calculation of Defibrillation Metrics Four defibrillation metrics were calculated to interpret and evaluate the solutions: interelectrode impedance Z, TABLE 2. Tissue conductivities. Tissue type Electrical conductivity σ (ms/cm) FIGURE 3. Segmented CT image. Heart muscle 2.5 Thoracic wall muscle 2.5 Blood 8 Lung 0.7 Bone 0.1 Fat 0.5

4 778 MOCANU et al. the DFT energy, the DFT current I th, and the DFT voltage V th. For each patient-specific simulation, the critical mass hypothesis 24 was used to define successful defibrillation with minimum delivered energy. According to this, a successful shock must expose a critical mass of the ventricular myocardium to electric fields equal to or greater than the inexcitability threshold E th. Zhou et al. 23 reported that a minimum electric field of 2.7 V/cm is required for biphasic shocks to successfully defibrillate approximately 77% of the time, and 5.4 V/cm to defibrillate approximately 87% of the time. We used an inexcitability threshold of 3.5 V/cm in our study to slightly exceed an 80% probability of success. In the simulations, a unit voltage was applied between the RV cathode and SVC and CAN anodes. The resulting electric field magnitudes were then scaled so that 95% of the ventricular myocardium was exposed to a minimum electric field magnitude of E th = 3.5 V/cm. The simulated DFT voltage and the DFT current are the corresponding scaled values of the applied voltage and the resulting delivered current. The DFT energy was calculated based on features of the Guidant/CPI biphasic waveform (Fig. 2) by integrating the power over the pulse duration: 1.6τ Vth 2 DFT = 0 Z e 2t/τ dt = 0.48CVth 2 (1) where Z is the patient-specific interelectrode impedance, V th is the DFT voltage, C is the capacitance (150 µf), and τ = ZC is the time constant. Statistical Analysis Correspondence between the clinical and model predicted defibrillation metrics was assessed using the correlation coefficient (cc) relating the predicted to measured values and the root mean square difference (rms) between the predicted and measured values. The paired t test was used to compare clinical and predicted DFT energy, and a value of p < 0.05 was used to establish statistically significant differences between the clinical and model-predicted DFTs. RESULTS Figure 5 shows the clinical DFT testing at the time of ICD implantation for all patients. Successful trials are shown with empty circles (Yes), while defibrillation failures are shown with solid circles (No). A demographic profile of the patient population used in this study is shown in Table 1. The comparison between the model-predicted and clinical interelectrode impedances resulted in a root mean squared difference (rms) of 8.2 and a correlation coefficient (cc) of 0.37 with a p-value = 0.09 (Fig. 6). In all patients but one (patient # 1), the model impedances overestimate the measured impedances with a cc = 0.7. Patient-specific clinical and model-predicted DFT energy values are compared in Fig. 7, where the clinical values correspond to the lowest energy shock that defibrillated on a single occasion. The rms difference for all nine patients was 12.4 J and the cc = Model-predicted DFTs were good estimates of the clinically determined thresholds in patients #1 4, with rms = 1.5 J and cc = A paired t test for patients #1 4 yielded a p-value of 0.41, indicating that there was no statistically significant difference between the clinical and model-predicted DFTs. For the second group of patients (# 5 9), the simulations underestimated the clinical DFT energy with rms = 18.4 J, cc = For patients # 5 9, the paired t test gave p = 0.01, indicating a statistically significant difference between the clinical and predicted DFTs. Similar trends were observed for the DFT voltages and the DFT currents (Figs. 8 and 9). We examined the electric field magnitude distribution associated with the clinical defibrillation threshold for each patient by applying the clinically measured DFT voltage to the corresponding model. The resulting cumulative histograms of the ventricular field magnitudes are shown in Fig. 10. We also examined the distribution of the weak field in each patient (Fig ), where the weak field is defined as the electric field magnitude in the ventricular myocardium less than 3.5 V/cm. By definition, for the critical mass criterion applied in this study, only 5% of the ventricular volume is included in the weak field region. Volume elements in the weak field region are indicated by a black cone (fixed volume) placed in the center of each volume element. The size of the largest cluster (percentage of total weak field) is also shown for each patient. DISCUSSION Our study investigates the degree to which patientspecific computational models can predict defibrillation threshold for individual patients. To our knowledge, this is the first study to directly compare clinical and modelpredicted DFTs on a patient-specific basis. The first step in our simulations was to compute the electric field distribution produced by a defibrillation shock using the finite volume method. We believe that this computational technique, like many other numerical methods (e.g. finite element method, finite difference method), is able to closely simulate the electric field distribution in the heart and thorax provided the model geometry closely approximates the patient s anatomy, and accurate tissue conductivity values are used. The second step of the simulation was to extract the defibrillation thresholds from the calculated cardiac electric fields. To do this, we used the critical mass criterion, a simple phenomenological model of defibrillation. Although debated, 3,18 the critical mass hypothesis for defibrillation is supported by experimental evidence 7,22,24

5 Patient-Specific Modeling of Defibrillation 779 FIGURE 5. Clinical DFT testing for all nine patients. = successful defibrillation attempt; = failed defibrillation attempt. For the modeling purposes, the lowest energy that defibrillated on a single occasion was defined as the DFT energy. and provides a simple means to relate the shock field at the continuum scale to the cellular-based events through which defibrillation is achieved. The interelectrode impedance is independent of the defibrillation assumptions used and is determined by the geometry and tissue properties. In all patients but one (patient FIGURE 6. Comparison between clinical and model-predicted impedances. FIGURE 7. Comparison between clinical and model-predicted DFT energies.

6 780 MOCANU et al. FIGURE 8. Comparison between clinical and model-predicted DFT voltages. #1) the model-predicted impedance overestimates the measured impedance (to a lesser extent for patient #5) (Fig. 6). Overestimates of the measured impedance were also obtained by Jorgenson et al. 10 for their porcine subject-specific finite element models based on similar mesh resolution and tissue conductivity values. This indicates the presence of a systematic error, which may orginate in the tissue conductivity values used or in the reconstruction of the anatomical structure. Model-predicted DFTs (energy, voltage, and current) were well matched to the clinically determined DFTs in four of the nine patients examined (first group: patients #1 4, Fig. 7). It is important to note that the respective clinical DFT energy values for these patients spanned an approximately 2-fold range. This suggests that the goodness of the achieved match was not due to the similarity of the patients examined but is a reflection of the ability of the modeling approach to capture the inherent anatomical variations in this group of patients. The clinical DFTs of the remaining five patients (second group: patients #5 9) FIGURE 9. Comparison between clinical and model-predicted DFT currents. FIGURE 10. Cumulative histograms for the computed electric field distribution. The applied voltage and all electric field magnitudes were scaled to correspond with the clinical DFT voltage. Vertical lines represent the mean field magnitudes (3.4 V/cm and 7.4 V/cm) associated with the two clusters. were not well matched by the simulated DFTs. We could not identify a consistent set of patient characteristics (e.g. heart volume, heart disease, or drug regimen) that were able to differentiate between these two groups (Table 1). The cumulative histograms shown in Fig. 10 were obtained by scaling the myocardial field distribution obtained for each patient model by the clinically measured DFT voltage corresponding to each patient. When viewed relative to the 95% critical mass perspective, these cumulative histograms (Fig. 10) reveal two distinct clusters of patients. The first cluster includes patients #1 4, with 95% of the ventricular myocardium exposed to field magnitudes greater than or equal to 3.4 ± 0.5 V/cm (mean ± standard deviation). The second cluster includes patients #5, 7, 8, and 9, with 95% of the ventricular myocardium exposed to field magnitudes greater than or equal to 7.4 ± 0.4 V/cm. Patient #6 belongs to neither cluster. One interpretation of these results is that a single inexcitability threshold is not able to describe all patients. While the 3.5 V/cm threshold does a very adequate job of characterizing patients #1 4, the data suggest that patients #5, 7, 8, and 9 require a substantially higher inexcitability threshold of 7.4 V/cm. It is unclear what causes the observed behavior. It has been shown that after unsuccessful shocks VF is reinitiated from the regions where the electric field is the weakest (weak field regions). 2,4 For this reason, we examined the spatial distribution of the weak field regions in each patient. In all of our simulations, the weak field was located in the posterolateral left ventricle (Figs ), consistent with what has been shown experimentally. 21 Our results show interesting differences in the patterns of the weak field distribution. Specifically, we found the weak field regions for patients that exhibited the highest clinical DFTs (Fig. 11) to be the most focal and compact (patients #7 9),

7 Patient-Specific Modeling of Defibrillation 781 FIGURE 11. Weak field distribution for patients with high clinical DFT. Patient #7: DFT = 19 J (the ICD pulse generator and the catheter are not shown for this patient); patient #8: DFT = 27 J; patient #9: DFT = 30 J. consisting of a single contiguous region that included 100% of the weak field region. Patients with more moderate clinical DFTs (#4, 6) had weak field regions that were still rather focal, but aggregated into two separate clusters (Fig. 12), with the largest being 90% of the weak field region. Finally, patients with the lowest clinical DFTs (#1 3, 5) exhibited the most scattered distribution of the weak field regions (Fig. 13). For these patients, the largest cluster contained less than 90% of the weak field region, with the exception of patient #3. These results suggest that large and compact weak field regions may favor the reinitiation of fibrillation, and that patients with such weak field distributions require a higher shock energy to defibrillate. The correspondence between the compactness of the weak field distribution and the clinical DFT may provide a predictive means of prospectively identifying high DFT patients prior to implant. Limitations The modeling approach used in this study uses a simple phenomenological model to relate the computed field distributions to defibrillation threshold values. Factors relating to the underlying cellular electrophysiology enter the model only through the inexcitability threshold E th and the critical mass criterion (expressed as a percentage of the ventricular mass). To the extent that the empirically-based model captures normative behavior, the model cannot be expected to predict defibrillation parameters for patients with cellular electrophysiology that differs substantially from the norm with a single E th. Similarly, we did not account for the presence of infarct regions, myocardial ischemia, or the effect of patient drug regimens in our models. The modeling approach is also inherently deterministic, and ignores the known probabilistic nature of defibrillation. Additionally, we have treated the lowest energy that defibrillates for each patient as the corresponding DFT, without taking into account the discrete nature of the clinical search protocol used to find the DFT. The truly lowest defibrillation energy could be somewhat lower than that found by the search protocol. CONCLUSIONS FIGURE 12. Weak field distribution for patients with moderate clinical DFT. Patient #4: DFT = 9.8 J; patient #6: DFT = 14 J. This paper presents comparisons between modelpredicted and clinical defibrillation metrics determined for individual patients with implanted ICDs Model-predicted parameters were extracted from the calculated cardiac electric field distribution using the 95% critical mass criterion and an inexcitability threshold of 3.5 V/cm for biphasic waveforms. The patient-specific simulations produced good estimates of the clinical DFTs in four of the nine patients investigated. Clinical DFTs and model predictions were well correlated for both well-matched and poorly-matched

8 782 MOCANU et al. FIGURE 13. Weak field distribution for patients with low clinical DFT. Patient #1: DFT=5J;patient #2: DFT = 4.8 J; patient #3: DFT = 7.4 J; patient #5: DFT = 7.7 J. groups of patients. Our results thus indicate that computational models based solely on the critical mass criterion and a single value of the inexcitability threshold are not able to consistently predict defibrillation thresholds for individual patients. However, inspection of the weak field distribution in all nine patient-specific models revealed that patients with the most compact weak field region were the most difficult to defibrillate and exhibited the highest clinical DFTs. This correspondence indicates that the weak field analysis may provide a promising approach to identify high DFT patients prior to implant. ACKNOWLEDGMENTS This study was supported in part by a grant from Guidant Corporation., St. Paul, MN, and the Trustees of Boston University. The authors would like to thank Dr. Michael Benser from the Cardiac Rhythm Management Laboratory (Guidant Corporation) for his valuable suggestions and comments. The excellent support of the Scientific Computing and Visualization Group at Boston University is also acknowledged. REFERENCES 1 Camacho, M. A., J. L. Lehr, and S. R. Eisenberg. A threedimensional finite element model of human transthoracic defibrillation: Paddle placement and size. IEEE Trans. Biomed. Eng. 42: , Chattipakorn, N., B. H. KenKnight, J. M. Rogers, R. G. Walker, G. P. Walcott, D. L. Rollins, W. M. Smith, and R. E. Ideker. Locally propagated activation immediately after internal defibrillation. Circulation 97: , Chen, P. S., N. Shibata, E. Dixon, P. D. Wolf, N. D. Danieley, M. B. Sweeney, W. M. Smith, and R. E. Ideker. Activation during ventricular defibrillation in open-chest dogs: Evidence of

9 Patient-Specific Modeling of Defibrillation 783 complete cessation and regeneration of ventricular fibrillation after unsuccessful shocks. J. Clin. Invest. 77: , Chen, P. S., P. D. Wolf, F. J. Claydon, E. Dixon, H. J. Vidaillet, N. D. Danieley, T. C. Pilkington, and R. E. Ideker. The potential gradient field created by epicardial defibrillation electrodes in dogs. Circulation 74: , Claydon, F. J., T. C. Pilkington, A. S. L. Tang, N. M. Morrow, and R. E. Ideker. A volume conductor model of thorax for the study of defibrillation fields. IEEE Trans. Biomed. Eng. 35: , De Jongh, A. L., E. G. Entcheva, J. A. Repogle, B. S. Booker, B. H. KenKnight, and F. J. Claydon. Defibrillation efficacy of different electrode placement in a human thorax model. PACE 22: , Garrey, W. E. The nature of fibrillary contraction of the heart: Its relations to tissue mass and form. Am. J. Physiol. 33: , Geddes, L. A., and L. E. Baker. The specific resistance of biological material: A compendium of data for the biomedical engineer and physiologist. Med. Biol. Eng. 5: , Haus, H. A., and J. Melcher. Electromagnetic Field and Energy Englewood Cliffs NJ: Prentice-Hall, 1988, p Jorgenson, D. B., P. H. Schimpf, I. Shen, G. Johnson, G. H. Bardy, D. R. Haynor, and Y. Kim. Predicting cardiothoracic voltages during high energy shocks: Methodology and comparison of experimental to finite element model data. IEEE Trans. Biomed. Eng. 42: , Karlon, W. J., S. R. Eisenberg, and J. L. Lehr. Effects of paddle placement and size on defibrillation current distribution: A three dimensional finite element model. IEEE Trans. Biomed. Eng. 40: , Kettenbach, J., A. G. Schreyer, S. Okuda, M. O. Sweeney, S. R. Eisenberg, C. F. Westin, F. Jacobson, R. Kikinis, B. H. KenKnight, and F. A. Jolesz. 3-D Modeling of the chest in patients with implanted cardiac defibrillator for further bioelectrical simulation. In: Proceedings of the C.A.R., edited by M. W. Vannier and K. Inmura. New York: Elsevier Science, 1998, pp Kinst, T. F., M. O. Sweeney, J. L. Lehr, and S. R. Eisenberg. Simulated internal defibrillation in humans using an anatomically realistic three-dimensional finite element model of the thorax. J. Cardiovasc. Electrophysiol. 8: , Min, X., and R. Mehra. Finite element analysis of defibrillation fields in a human torso model for ventricular defibrillation. Prog. Biophys. Mol. Biol. 69: , Naccarelli, G. V., and E. P. Veltri. Implantable cardioverterdefibrillators: What s in the future? In: Implantable Cardioverter- Defibrillators, edited by G. V. Naccarelli and G. V. Veltri. Boston: Blackwell Scientific, 1993 pp Panescu, D., J. G. Webster, W. J. Tompkins, and R. A. Stratbucker. Optimization of cardiac defibrillation by threedimensional finite element of the human thorax. IEEE Trans. Biomed. Eng. 42: , Sepulveda, W. J. P., and D. S. Echt. Finite element analysis of cardiac defibrillation current distribution. IEEE Trans. Biomed. Eng. 37: , Shibata, N., P. S. Chen, E. G. Dixon, P. D. Wolf, N. D. Danieley, W. M. Smith, and R. E. Ideker. Epicardial activation following unsuccessful defibrillation shocks in dogs. Am. J. Physiol. 255 H902-H904, Singer, I., and D. Lang. Defibrillation threshold: Clinical utility and therapeutic implications. PACE 15:932, Tacker, W. A., J. Mercer, P. Foley, and S. Cuppy. Resistivity of skeletal muscle, skin, and lung to difibrillation shocks. Paper presented at the AAMI 19th Annual Meeting, Washington, DC, 1984, p Tang, A. S. L., P. D. Wolf, Y. Afework, W. M. Smith, and R. E. Ideker. Three-dimensional potential gradient fields generated by intracardiac catheter and cutaneous patch electrodes. Circulation 85: , Witkowski, F. X., P. A. Penkoske, and R. Plonsey. Mechanism of cardiac defibrillation in open-chest dogs with unipolar DCcoupled simultaneous activation and shock potential recordings. Circulation 82: , Zhou, X., P. J. Daubert, P. D. Wolf, W. M. Smith, and R. E. Ideker. Epicardial mapping of ventricular defibrillation with monophasic and biphasic shocks in dogs. Circ. Res. 72: , Zipes, D. P., J. Fischer, R. M. King, A. Nicoll, and W. W. Jolly. Termination of ventricular fibrillation in dogs by depolarizing a critical amount of myocardium. Am. J. Cardiol. 36:37 44, 1975.

Visualization of Defibrillation Simulation Using Multiple Transthoracic Electrodes

Visualization of Defibrillation Simulation Using Multiple Transthoracic Electrodes Visualization of Defibrillation Simulation Using Multiple Transthoracic Electrodes Joseph Qualls, 1 David J. Russomanno, 1 Amy de Jongh Curry, 2 and Deepika Konakanchi 2 1 Department of Electrical and

More information

Partially Extracted Defibrillator Coils and Pacing Leads Alter Defibrillation Thresholds

Partially Extracted Defibrillator Coils and Pacing Leads Alter Defibrillation Thresholds Partially Extracted Defibrillator Coils and Pacing Leads Alter Defibrillation Thresholds Pilcher T 1, Stinstra J 2, Tate J 2,3, Saarel E 1, Puchalski M 1, Macleod R 2,3 Division of Pediatric Cardiology

More information

Europace (2001) 3, doi: /eupc , available online at on

Europace (2001) 3, doi: /eupc , available online at   on Europace (21) 3, 278 284 doi:1.153/eupc.21.184, available online at http://www.idealibrary.com on ICDs Clinical evaluation of defibrillation efficacy with a new single-capacitor biphasic in patients undergoing

More information

Transvenous Biventricular Defibrillation Halves Energy Requirements in Patients

Transvenous Biventricular Defibrillation Halves Energy Requirements in Patients Transvenous Biventricular Defibrillation Halves Energy Requirements in Patients Christian Butter, MD; Eckhard Meisel, MD; Juergen Tebbenjohanns, MD; Lothar Engelmann, MD; Eckart Fleck, MD; Bernd Schubert,

More information

Shock-induced termination of cardiac arrhythmias

Shock-induced termination of cardiac arrhythmias Shock-induced termination of cardiac arrhythmias Group members: Baltazar Chavez-Diaz, Chen Jiang, Sarah Schwenck, Weide Wang, and Jinglei Zhang Abstract: Cardiac arrhythmias occur when blood flow to the

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

NIH Public Access Author Manuscript Heart Rhythm. Author manuscript; available in PMC 2011 May 30.

NIH Public Access Author Manuscript Heart Rhythm. Author manuscript; available in PMC 2011 May 30. NIH Public Access Author Manuscript Published in final edited form as: Heart Rhythm. 2010 May ; 7(5): 692 698. doi:10.1016/j.hrthm.2010.01.030. Finite element modeling of subcutaneous implantable defibrillator

More information

Shock-induced termination of cardiac arrhythmias

Shock-induced termination of cardiac arrhythmias Shock-induced termination of cardiac arrhythmias Group members: Baltazar Chavez-Diaz, Chen Jiang, Sarah Schwenck, Weide Wang, and Jinglei Zhang Cardiac arrhythmias, also known as irregular heartbeat, occur

More information

Defibrillation threshold testing should no longer be performed: contra

Defibrillation threshold testing should no longer be performed: contra Defibrillation threshold testing should no longer be performed: contra Andreas Goette St. Vincenz-Hospital Paderborn Dept. of Cardiology and Intensive Care Medicine Germany No conflict of interest to disclose

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System Revision

Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System Revision Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System Revision EMILE G. DAOUD, K. CHING MAN, ERED MORADY, and S. ADAM STRIGKBERGER From the Division of

More information

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic

More information

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Dual-Chamber Implantable Cardioverter-Defibrillator

Dual-Chamber Implantable Cardioverter-Defibrillator February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary

More information

TASER and In-Custody Deaths

TASER and In-Custody Deaths TASER and In-Custody Deaths Cardiac Effects of Conducted Electrical Weapons Patrick Tchou, MD Heart and Vascular Institute What is a stun gun? A device that delivers rapid electrical impulses that stimulates

More information

Comparative Efficacy of Triphasic and Biphasic Internal Defibrillation

Comparative Efficacy of Triphasic and Biphasic Internal Defibrillation 224 Vol. 8, No. 4, December 2003 Comparative Efficacy of Triphasic and Biphasic Internal Defibrillation K. MISCHKE, M. ZARSE, K. BREHMER, C. STELLBRINK, P. HANRATH, P. SCHAUERTE Department of Cardiology,

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Point of View. The experimental evidence supporting the critical mass hypothesis was obtained before the era of

Point of View. The experimental evidence supporting the critical mass hypothesis was obtained before the era of Point of View 913 Mechanism of Cardiac Defibrillation A Different Point of View* Peng-Sheng Chen, MD; Patrick D. Wolf, MS; and Raymond E. Ideker, MD, PhD Downloaded from http://ahajournals.org by on October

More information

Journal of the American College of Cardiology Vol. 35, No. 2, by the American College of Cardiology ISSN /00/$20.

Journal of the American College of Cardiology Vol. 35, No. 2, by the American College of Cardiology ISSN /00/$20. Journal of the American College of Cardiology Vol. 35, No. 2, 2000 2000 by the American College of Cardiology ISSN 0735-1097/00/$20.00 Published by Elsevier Science Inc. PII S0735-1097(99)00579-3 High

More information

Single- versus Dual-Coil ICD Leads: Does it Matter?

Single- versus Dual-Coil ICD Leads: Does it Matter? Single- versus Dual-Coil ICD Leads: Does it Matter? C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Conflict of Interest Biotronik Boston-Scientific

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.

More information

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network INTERNAL CARDIOVERSION Lancashire & South Cumbria Cardiac Network DC Cardioversion AF & AFL safe 1,2 efficacious 3,4 (60-94%) 5 SR - Increases exercise tolerance 6 Maintainence SR unlikely in patients

More information

Atrial Defibrillation With a Transvenous Lead A Randomized Comparison of Active Can Shocking Pathways

Atrial Defibrillation With a Transvenous Lead A Randomized Comparison of Active Can Shocking Pathways Journal of the American College of Cardiology Vol. 34, No. 2, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(99)00214-4 Atrial

More information

Effects of myocardial infarction on catheter defibrillation threshold

Effects of myocardial infarction on catheter defibrillation threshold Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1983 Effects of myocardial infarction on catheter defibrillation threshold

More information

Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement

Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement www.ipej.org 49 Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement Naga VA Kommuri, MD, MRCPCH, Sri Lakshmi S Kollepara, MD,

More information

Different indications for pacemaker implantation are the following:

Different indications for pacemaker implantation are the following: Patient Resources: ICD/Pacemaker Overview ICD/Pacemaker Overview What is a pacemaker? A pacemaker is a device that uses low energy electrical pulses to prompt the heart to beat whenever a pause in the

More information

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION 1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,

More information

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing

More information

Efficacy and safety of the reciprocal pulse defibrillator current waveform

Efficacy and safety of the reciprocal pulse defibrillator current waveform Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1984 Efficacy and safety of the reciprocal pulse defibrillator current

More information

Systems Biology Across Scales: A Personal View XXVII. Waves in Biology: Cardiac Arrhythmia. Sitabhra Sinha IMSc Chennai

Systems Biology Across Scales: A Personal View XXVII. Waves in Biology: Cardiac Arrhythmia. Sitabhra Sinha IMSc Chennai Systems Biology Across Scales: A Personal View XXVII. Waves in Biology: Cardiac Arrhythmia Sitabhra Sinha IMSc Chennai The functional importance of biological waves Spiral Waves Cardiac Arrhythmias Arrhythmias:

More information

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes

More information

Biphasic Clinical Summaries

Biphasic Clinical Summaries Biphasic Clinical Summaries Defibrillation of Ventricular Fibrillation and Ventricular Tachycardia Background Physio-Control conducted a multi-centered, prospective, randomized and blinded clinical trial

More information

Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis

Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis Investigator: Keiko Saito, MD Mentor: Yuji Saito, MD, PhD, FACP, FACC Department

More information

Effect of a Passive Endocardial Electrode on Defibrillation Efficacy of a Nonthoracotomy Lead System

Effect of a Passive Endocardial Electrode on Defibrillation Efficacy of a Nonthoracotomy Lead System JACC Vol. 29, No. 4 March 15, 1997:825 30 825 Effect of a Passive Endocardial Electrode on Defibrillation Efficacy of a Nonthoracotomy Lead System PARWIS C. FOTUHI, MD, BRUCE H. KENKNIGHT, MS,* SHARON

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL Table S1. Patient Selection Codes, CIED Generator Procedures Code Type Code Description ICD9 Proc 00.51 Implantation of cardiac resynchronization defibrillator, total system [CRT-D]

More information

Defibrillation testing should be routinely performed at the time of implantable cardioverter-defibrillator implantation Contra

Defibrillation testing should be routinely performed at the time of implantable cardioverter-defibrillator implantation Contra Defibrillation testing should be routinely performed at the time of implantable cardioverter-defibrillator implantation Contra Fernando Arribas Cardiology Service Hospital 12 de Octubre Madrid Spain D.O.I.

More information

An Experimental Study of Transvenous Defibrillation Using a Coronary Sinus Catheter

An Experimental Study of Transvenous Defibrillation Using a Coronary Sinus Catheter An Experimental Study of Transvenous Defibrillation Using a Coronary Sinus Catheter ALAN H. KADISH, M.D., KEITH CHILDS, B.S., and JOSEPH LEVINE, M.D. From the Division of Cardiology, Department of Internal

More information

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II). Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Implantable cardioverter defibrillators for the treatment of arrhythmias and cardiac resynchronisation therapy for the treatment of heart failure (review

More information

Engineering involved in Cardiac Arrest Management Sahar Sorek

Engineering involved in Cardiac Arrest Management Sahar Sorek Engineering involved in Cardiac Arrest Management Sahar Sorek The University of Rhode Island College of Engineering BME 281-0001 Fascinating Facts: 424,000/year out of hospital CA s. More than half of

More information

Biomedical Instrumentation

Biomedical Instrumentation Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Therapeutic and Prosthetic Devices 1 Figure 13.1 Block diagram of an asynchronous

More information

Simulation of T-Wave Alternans and its Relation to the Duration of Ventricular Action Potentials Disturbance

Simulation of T-Wave Alternans and its Relation to the Duration of Ventricular Action Potentials Disturbance The Open Pacing, Electrophysiology & Therapy Journal, 2010, 3, 21-27 21 Open Access Simulation of T-Wave Alternans and its Relation to the Duration of Ventricular Action Potentials Disturbance Dariusz

More information

Ventricular tachycardia Ventricular fibrillation and ICD

Ventricular tachycardia Ventricular fibrillation and ICD EKG Conference Ventricular tachycardia Ventricular fibrillation and ICD Samsung Medical Center CCU D.I. Hur Ji Won 2006.05.20 Ventricular tachyarrhythmia ventricular tachycardia ventricular fibrillation

More information

Tech Corner. ATP in the Fast VT zone

Tech Corner. ATP in the Fast VT zone Tech Corner ATP in the Fast VT zone NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN THE ARTICLE. FOR ADDITIONAL

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

ICD THERAPIES: are they harmful or just high risk markers?

ICD THERAPIES: are they harmful or just high risk markers? ICD THERAPIES: are they harmful or just high risk markers? Konstantinos P. Letsas, MD, PhD, FESC LAB OF CARDIAC ELECTROPHYSIOLOGY EVANGELISMOS GENERAL HOSPITAL ATHENS ICD therapies are common In a meta-analysis

More information

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. Defibrillators OBJECTIVE 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe

More information

Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention

Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Dr Ngai-Yin Chan, MBBS, FRCP(Lond), FRCP(Edin), FRCP(Glasg), FACC, FHRS, Consultant Physician, Department of Medicine & Geriatrics, Princess

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

Effects of Polarity for Monophasic and Biphasic Shocks on Defihrillation Efficacy with an Endocardiai System

Effects of Polarity for Monophasic and Biphasic Shocks on Defihrillation Efficacy with an Endocardiai System Effects of Polarity for Monophasic and Biphasic Shocks on Defihrillation Efficacy with an Endocardiai System MASAHIRO USUI, GREGORY P. WALCOTT, S. ADAM STRICKBERGER,* DENNIS L. ROLLINS, WILLIAM M. SMITH,

More information

Cardiac Electrical Therapies. By Omar AL-Rawajfah, PhD, RN

Cardiac Electrical Therapies. By Omar AL-Rawajfah, PhD, RN Cardiac Electrical Therapies By Omar AL-Rawajfah, PhD, RN Outlines What are cardiac electrical therapies Ablation Defibrillation Cardioversion What are the nursing considerations for each type of therapy

More information

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Case Report Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Taku Nishida MD, Tamio Nakajima MD, PhD, Yutaka Goryo MD, Ken-ichi Ishigami MD, PhD, Hiroyuki Kawata

More information

The Facts about Biphasic Defibrillation

The Facts about Biphasic Defibrillation The Facts about Biphasic Defibrillation Introduction In reference to the SMART Biphasic waveform, Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care, published by the American

More information

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012 IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT201203 JANUARY 24, 2012 The IHCP to reimburse implantable cardioverter defibrillators separately from outpatient implantation Effective March 1, 2012, the

More information

The Nuts and Bolts of ICD Therapy

The Nuts and Bolts of ICD Therapy Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction

More information

Effect of Shock Thning on Efficacy and Safety Cardioverso on for Ventricular Tachycardia

Effect of Shock Thning on Efficacy and Safety Cardioverso on for Ventricular Tachycardia JAM Vol. 24. No. 3 September 1994 : 70 3 -ti 703 ELECTROPHYSIOLOGY Effect of Shock Thning on Efficacy and Safety Cardioverso on for Ventricular Tachycardia of Internal HUAGUI G. LI, MD, PHD, RAYMOND YEE,

More information

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds

More information

Triphasic Waveforms Are Superior to Biphasic Waveforms for Transthoracic Defibrillation Experimental Studies

Triphasic Waveforms Are Superior to Biphasic Waveforms for Transthoracic Defibrillation Experimental Studies Journal of the American College of Cardiology Vol. 42, No. 3, 2003 2003 by the American College of Cardiology Foundation ISSN 0735-1097/03/$30.00 Published by Elsevier Inc. doi:10.1016/s0735-1097(03)00656-9

More information

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study ANEESH V. TOLAT, M.D.,* MELISSA WOICIECHOWSKI, M.S.N.,* ROSEMARIE KAHR, R.C.I.S.,* JOSEPH

More information

DEVELOPMENT OF SCREENING TOOL TO IDENTIFY POTENTIAL IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) RECEIVER

DEVELOPMENT OF SCREENING TOOL TO IDENTIFY POTENTIAL IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) RECEIVER DEVELOPMENT OF SCREENING TOOL TO IDENTIFY POTENTIAL IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) RECEIVER Tan Wei Kiat a, Megalla Packrisamy a, M. B Malarvili b,a Faculty of Biosciences and Medical Engineering,

More information

Review of Pacemakers and ICD Therapy: Overview and Patient Management

Review of Pacemakers and ICD Therapy: Overview and Patient Management Review of Pacemakers and ICD Therapy: Overview and Patient Management Pacing Systems Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services OSU Division

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Atrial fibrillation or atrial flutter (AF) may occur after

Atrial fibrillation or atrial flutter (AF) may occur after Relation of Atrial Refractoriness to Upper and Lower Limits of Vulnerability for Atrial Fibrillation/Flutter Following Implantable Ventricular Defibrillator Shocks Amos Katz, MD; Robert J. Sweeney, PhD;

More information

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia By Sandeep Joshi, MD and Jonathan S. Steinberg, MD Arrhythmia Service, Division of Cardiology

More information

3/17/2014. WS # 3 ICD Registry Case Scenarios with Structural Abnormalities. Objectives. Denise Pond BSN, RN

3/17/2014. WS # 3 ICD Registry Case Scenarios with Structural Abnormalities. Objectives. Denise Pond BSN, RN WS # 3 ICD Registry Case Scenarios with Structural Abnormalities Denise Pond BSN, RN The following relationships exist related to this presentation: No Disclosures Objectives Discuss specific coding instructions

More information

SPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS

SPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS SPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS Vessela Tzvetanova Krasteva, Irena Ilieva Jekova Centre of Biomedical Engineering Prof. Ivan Daskalov - Bulgarian Academy of Sciences Acad.G.Bonchev

More information

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie 268 Case Report Cardiac Resynchronization Therapy in a Patient with Persistent Left Superior Vena Cava Draining into the Coronary Sinus and Absent Innominate Vein: A Case Report and Review of Literature

More information

Managing Metallic Artifacts in MRI

Managing Metallic Artifacts in MRI Managing Metallic Artifacts in MRI William Faulkner, BS, RT (R) (MR) (CT), FSMRT Background Magnetism is a fundamental property of matter. As such, all substances have some type of magnetic property. Magnetic

More information

Comparison of the defibrillation threshold and the

Comparison of the defibrillation threshold and the LABORATORY INVESTIGATION ELECTROPHYSIOLOGY Comparison of the defibrillation threshold and the upper limit of ventricular vulnerability PENG-SHENG CHEN, M.D., NITARO SHIBATA, M.D., ELLEN G. DIXON, M.S.,

More information

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Notification and Prior Authorization Program: Electrophysiology Implant Classification Table The following chart contains the codes that require notification or prior authorization as part of UnitedHealthcare

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Tseng ZH, Hayward RM, Clark NM, et al. Sudden death in patients with cardiac implantable electronic devices. JAMA Intern Med. Published online June 22, 2015. doi:10.1001/jamainternmed.2015.2641.

More information

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators 22 March 2002 Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators F. HINTRINGER, O. PACHINGER Division of Cardiology, Department for Internal Medicine, University

More information

Shock Reduction Strategies Michael Geist E. Wolfson MC

Shock Reduction Strategies Michael Geist E. Wolfson MC Shock Reduction Strategies Michael Geist E. Wolfson MC Shock Therapy Thanks, I needed that! Why Do We Need To Reduce Shocks Long-term outcome after ICD and CRT implantation and influence of remote device

More information

MEDICAL POLICY Cardioverter Defibrillators

MEDICAL POLICY Cardioverter Defibrillators POLICY........ PG-0224 EFFECTIVE......06/01/09 LAST REVIEW... 01/27/17 MEDICAL POLICY Cardioverter Defibrillators GUIDELINES This policy does not certify benefits or authorization of benefits, which is

More information

Effects of Pericardial Effusates of Various Conductivities on Body-Surface Potentials in Dogs

Effects of Pericardial Effusates of Various Conductivities on Body-Surface Potentials in Dogs 788 Effects of Pericardial Effusates of Various Conductivities on Body-Surface Potentials in Dogs Documentation of the Eccentric Spheres Model Daniel A. Kramer, Robert L. Hamlin, and Herman R. Weed From

More information

Electrophysiologist-Implanted, Nonthoracotomy-Implantable Cardioverter/Defibrillators

Electrophysiologist-Implanted, Nonthoracotomy-Implantable Cardioverter/Defibrillators 2503 Electrophysiological Laboratory, Electrophysiologist-Implanted, Nonthoracotomy-Implantable Cardioverter/Defibrillators Adam P. Fitzpatrick, MD, MRCP; Michael D. Lesh, MD; Laurence M. Epstein, MD;

More information

CARDIOINSIGHT TM NONINVASIVE 3D MAPPING SYSTEM CLINICAL EVIDENCE SUMMARY

CARDIOINSIGHT TM NONINVASIVE 3D MAPPING SYSTEM CLINICAL EVIDENCE SUMMARY CARDIOINSIGHT TM NONINVASIVE 3D MAPPING SYSTEM CLINICAL EVIDENCE SUMMARY April 2017 SUPPORTING EVIDENCE RHYTHM AF VT PUBLICATIONS Driver Domains in Persistent Atrial Fibrillation (Haissaiguerre, et al)

More information

Experience with pectoral versus abdominal implantation of a small defibrillator

Experience with pectoral versus abdominal implantation of a small defibrillator European Heart Journal (1998) 19, 185 198 Article No. hj9816 Experience with pectoral versus abdominal implantation of a small defibrillator A multicenter comparison in 778 patients E. Hoffmann, G. Steinbeck,

More information

Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD

Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD Harish Doppalapudi, MD Assistant Professor of Medicine Director, Clinical Cardiac Electrophysiology Training Program University of Alabama

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Quantitative Electrophysiology

Quantitative Electrophysiology ECE 795: Quantitative Electrophysiology Notes for Lecture #10 Wednesday, November 22, 2006 14. FUNDAMENTALS OF FUNCTIONAL ELECTRICAL STIMULATION (FES) We will look at: Design issues for FES Subthreshold

More information

Spatiotemporal Cardiac Activation Sites Localization Using ECG Precordial Leads

Spatiotemporal Cardiac Activation Sites Localization Using ECG Precordial Leads Spatiotemporal Cardiac Activation Sites Localization Using ECG Precordial Leads Jaime R. De La Cruz BSEE a, Joseph H.Pierluissi PhD a, Ubaldo Robles BSEE a, Zainul Abedin MD b a Electrical and Computer

More information

Fibrillation in Dogs

Fibrillation in Dogs 325 Effects of Monophasic and Biphasic Shocks on Action Potentials During Ventricular Fibrillation in Dogs Xiaohong Zhou, Patrick D. Wolf, Dennis L. Rollins, Yohannes Afework, William M. Smith, Raymond

More information

A multifilamented electrode in the middle cardiac vein reduces energy requirements for defibrillation in the pig

A multifilamented electrode in the middle cardiac vein reduces energy requirements for defibrillation in the pig Heart 2000;84:425 430 425 Angeion Corporation, Minneapolis, Minnesota, USA J F Urban D E Euler M J Kallok Section of Paediatrics, National Heart and Lung Institute, Imperial College School of Medicine,

More information

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection Proceedings of the 2012 International Conference on Industrial Engineering and Operations Management Istanbul, Turkey, July 3 6, 2012 Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter

More information

Cigna - Prior Authorization Procedure List Cardiology

Cigna - Prior Authorization Procedure List Cardiology Cigna - Prior Authorization Procedure List Cardiology Category CPT Code CPT Code Description 33206 Insertion of new or replacement of permanent pacemaker with transvenous electrode(s); atrial 33207 Insertion

More information

1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium

1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium ١ 1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium ٢ When VF/pulseless VT cardiac arrest is associated with torsades de pointes, providers may administer magnesium sulfate at a dose of 1-2

More information

Novel Approaches to VT Management Glenn M Polin MD

Novel Approaches to VT Management Glenn M Polin MD Novel Approaches to VT Management Glenn M Polin MD Medical Director, Electrophysiology Laboratory John Ochsner Heart and Vascular Institute New Orleans, LA Disclosures Pfizer Speaker Bureau Bristol Myers

More information

A Dynamic model of Pulmonary Vein Electrophysiology. Harry Green 2 nd year Ph.D. student University of Exeter

A Dynamic model of Pulmonary Vein Electrophysiology. Harry Green 2 nd year Ph.D. student University of Exeter A Dynamic model of Pulmonary Vein Electrophysiology Harry Green 2 nd year Ph.D. student University of Exeter Background to the Project Cardiac disease is the leading cause of death in most developed countries

More information

IBHRExam Prep Implanted CRM Device System Radiography

IBHRExam Prep Implanted CRM Device System Radiography Implanted CRM Device System Radiography IBHRExam Prep www.pacericd.com 2 Where do they go? 3 Anatomy diagram 4 Anatomy 5 Pulse generator configurations www.pacericd.com 6 Guidant pacemaker pulse generator

More information

The Electrocardiogram

The Electrocardiogram The Electrocardiogram Chapters 11 and 13 AUTUMN WEDAN AND NATASHA MCDOUGAL The Normal Electrocardiogram P-wave Generated when the atria depolarizes QRS-Complex Ventricles depolarizing before a contraction

More information

Non-Invasive Ablation of Ventricular Tachycardia

Non-Invasive Ablation of Ventricular Tachycardia Non-Invasive Ablation of Ventricular Tachycardia Dr Shaemala Anpalakhan Newcastle upon Tyne Hospitals NHS Foundation Trust Freeman Road, Newcastle Upon Tyne, NE7 7DN Contact: shaemala@doctors.org.uk Introduction

More information