Blood Pressure Determination Using Analysis of Biosignals

Size: px
Start display at page:

Download "Blood Pressure Determination Using Analysis of Biosignals"

Transcription

1 Blood Pressure Determination Using Analysis of Biosignals RADIM ČÍŽ 1, MILAN CHMELAŘ 2 1 Department of Telecommunications Brno University of Technology Purkyňova 118, Brno CZECH REPUBLIC cizr@feec.vutbr.cz, 2 Department of Biomedical Engineering Brno University of Technology Kolejní 4, Brno CZECH REPUBLIC chmelar@feec.vutbr.cz, Abstract: - The paper deals with an analysis of biosignals for the purpose of non-invasive blood pressure monitoring without the necessity of using a pneumatic cuff. The method of simultaneous recording of electrocardiogram (ECG) and photoplethysmogram (PPG) and evaluations of their mutual time shift, which corresponds to the pulse transit time (PTT), are described in the paper. Results of original and improved methods of blood pressure determination using calibration measurement and subsequent analysis of pulse transit time values are also presented. Key-Words: - Biosignal Processing, Blood Pressure, Pulse Transit Time, Photoplethysmography, Electrocardiography 1 Introduction The circulatory system is, in principle, a hydraulic system. Using the analogy between electrical and hydraulic circuits, we can draw the equivalent electrical scheme of a part of artery. The serial RLC circuit, which is an equivalent circuit that describes the properties of a part of artery, is a filter. Its parameters depend on the pressure in the corresponding part of the hydraulic system (i.e. artery). With changes in the parameters of this filter its transfer function also change. This entails a change in not only the pulse shape [1] but also in the delay of this pulse. This means that we can monitor changes in the blood pressure in artery in dependence on the shape or pulse wave velocity. The analysis of the pulse wave velocity in arteries includes many variables such as blood viscosity, elasticity of arterial wall, attenuation of the wave amplitude and the presence of reflected waves due to artery branching. Therefore the real velocity is only 3-12 m.s -1. In connection with measuring the velocity of pulse wave propagation there is a question of which signals could be used for the determination of pulse transit time (PTT) in a given position. One from two possibilities is to compare the ECG signal and the signal of photoplethysmograph (PPG) [2]. The R- wave peak of ECG, which corresponds to the systole of ventricles, is sharp and therefore it is a suitable reference point. Moreover, the peak on ECG signal can be easily detected by some automatic method. The second possibility is to measure the delay between two corresponding points of two PPG signals obtained by sensors placed on distant positions of the same artery. From the technical point of view, the ECG signal sensing is easy and precise. The problem is, however, that we get exact time of the ventricle systole, but the expulsion of blood from the left ventricle to the artery takes a certain time, which need not be constant. This fact can introduce mistakes in final assessment [3]. Obtaining the PPG signal using the photoplethysmographic sensor is also simple, but this method is very sensitive to any movements of the subject being scanned, which cause motion artifacts in the PPG waveform. In case of need, another method of pulse wave sensing could be used instead of photoplethysmography. For example, the sequential detection of artery sectional area described by Dr. Říha [4] uses ultrasound scanning. ISBN:

2 2 Methodology and Arrangement When we want to relate the pulse wave velocity to changes in blood pressure, we must measure the blood pressure at the same time as the pulse wave velocity. Non-invasive methods for blood pressure measurement are mostly discontinuous so that fast changes in blood pressure cannot be recorded. It is the source of further mistakes in the results of measurement. Only volume-clamp method, invented by prof. J. Penaz in 1967 and implemented in Finapres device now, enable non-invasive and continuous blood pressure measurement. For the measurement of mutual relationship between pulse wave velocity (PWV) and blood pressure we have used ECG as reference signal, which was detected by electrodes located on the chest of an experimental person. The blood pressure was measured by Finapres and its sensor was placed on the middle finger of the person. The PPG signal was obtained by reflexive photoplethysmographic sensor on the forefinger. The PPG and ECG signals were amplified and stored together with the synchronized signal from Finapres using the special data-acquisition device [5] for further processing by the MATLAB software. The experimental person was sitting in an armchair, with the left hand placed (together with the sensors) on the arm rest of the armchair and with a weight held in the right hand. We used this isometric load in order to continually increase blood pressure. Male and female persons aged about 30, were subjects of measurement. 2.1 Initial measurement As can be seen in Figs 1 to 3, measuring on the experimental person proved the supposed correlation of beat-to-beat blood pressure values and corresponding pulse wave velocities. But the variance of measured values suggests the impossibility of calculating blood pressure with a sufficient precision using a simple linear function of pulse wave velocity. Fig. 1 Correlation between the systolic BP and pulse wave velocity, corr. coeficient = 0.87 Fig. 2 Correlation between the diastolic BP and pulse wave velocity, corr. coeficient = 0.79 Fig. 3 Correlation between the mean BP and pulse wave velocity, corr. coeficient = The First Experiment and Results In the first phase of the research, the relation between the systolic blood pressure value, which was measured by Finapres, and the reciprocal value of the pulse transit time PTT between the R-wave of ECG signal and the corresponding peak of PPG signal was sought. The reciprocal value of PTT could be simply used instead of pulse wave velocity value because the distance between sensors is constant during whole measurement with an experimental person. It also introduced adequate clarity of the depiction of this dependence in following graphs. For each load a short section of 3 signals was recorded (see Fig. 7). Values of systolic (SBP) and diastolic (DBP) blood pressures from Finapres signal were averaged as well as values of the pulse transit time which were read from ECG and PPG signals. The results are given in Table 1. They are processed graphically in Figs 4 to 6. From the results obtained it is clear that there is an evident correlation between systolic blood pressure and reciprocal value of PTT only for the experimental person No.1 (compare the progression of blue and green points in Fig. 4). For the other two persons the results are inconsistent (see Fig. 5 and Fig. 6). ISBN:

3 Person No.1 No.2 No.3 Table 1 Results of the first experiment SBP DBP 1/PTT [s -1 ] 1/PTT [s -1 ] In an analysis of the obtained results it was discovered that in cases of persons without evident correlation between the blood pressure values and reciprocal values of PTT, there are random time shifts between PPG and Finapres signal (see ΔT differences in Fig. 7). The time shift was nearly constant in all records acquired during measurement with person No.1. Therefore we carried out yet another assessment. We measured the pulse transit time PTT between the R-wave of ECG signal and the corresponding peak of the Finapres signal. In this case the relation between PTT and blood pressure values was always in agreement with the assumption that the transit time decreases with increasing blood pressure. The better correlation between blue points (systolic blood pressure) and red points (1/PTT ) is clearly proved in Figs 4 to 6. to the place of branching strikes against the wall between the branches, and these results in turbulence. The turbulence quantity depends on the Reynolds number at the point of artery branching. This turbulence causes a random instability in the distribution of the blood flow into branches [6] and this is reflected in the velocities of pulse wave propagation. In addition, this phenomenon depends on the anatomy of the arterial system of the given person. This may cause significant deviations between signals from Finapres and from the photoplethysmographic sensor. In some cases these difference may be very small, as can be seen from results of person No.1. Fig. 5 Progression of values person No.2 Fig. 6 Progression of values person No.3 Fig. 4 Progression of values person No.1 The problem originates from the location of the photoplethysmographic sensor and the Finapres sensor on the different branches of artery. It disappeared when the place for blood pressure measurement and the place for detection of pulse transit time were the same. We suppose these random time shifts between PPG and Finapres signal are caused by phenomena that occur at points where the artery becomes branched. Blood flowing 3 Improvement During a further phase of our research we did a multiple correlation analysis of the parameters obtained from the signals measured. We were searching for parameters with maximum correlation with blood pressure values measured simultaneously by Finapres. Subsequently obtained coefficients of approximating equation were used for the calculation of systolic and diastolic beat-to-beat blood pressure. Finapres and photoplethysmographic sensors were placed on adjacent fingers of an experimental person during the measurement. ISBN:

4 Fig. 7 Recorded signals of person No.2 without load The peak of the R-wave of ECG signal, which was measured at the same time, was used as the reference point for the determination of pulse transit time. Compared to the previous method, it is not the corresponding point of R-wave but the intersection point of the tangent line passing through the inflection point in the rising edge of pulse and the maximum or minimum level of the pulse that was sought in the PPG waveform. inflection points of rising edges in PPG and their values equal the difference quotients of the respective tangent lines. Person No.4 No.5 No.6 Table 2 Results of improved method Measured SBP/DBP Calculated SBP/DBP ΔSBP/ ΔDBP Note 97.1/ /50.5 0/0 calibration 99.0/ / / min. after calib / / /-0.1 with isometric load 153.6/ /84.8 0/-0.1 calibration 139.6/ / /3.3 5 min. after calib / / /0.4 with isometric load 114.6/ /80.9 0/0 calibration 112.5/ / / min. after calib / / /-2.8 with isometric load As depicted in Fig. 8, PPG pulses are flatter and the position of their peaks sometimes varies because of motion artifacts. Therefore the Matlab program for the analysis of biosignals calculates the first differences of sampled PPG and then finds simply detectable local maximums among them. The time positions of those maximums correspond to the Fig. 8 Improved determination of pulse transit time For subsequent calculation of systolic and diastolic blood pressure values, two differently ISBN:

5 defined pulse transit times were used as input parameters. PTT 1 is the difference between the time position of a the intersection point of the tangent line with the maximum level of the PPG pulse while in case of PTT 2 the intersection point of the tangent line with the minimum level was used as a point of reference. One measurement lasted 1 minute. The calculated and real values of beat-to-beat blood pressure were averaged for each measurement (see Fig. 9 and Fig. 10). A comparison of these averaged values showed that the results of the method were accurate enough for the most of experimental persons (see Table 2). Thus the averaging of beatto-beat blood pressure values during 1 minute measurement minimizes the effect of time deviations caused by the artery branching described in the previous chapter. It also suppresses the influence of deviations in an interval which lasts from the moment of R-wave of ECG signal to the time of real heart ventricle systole. Fig. 9 Beat-to-beat systolic blood pressure values for the person No. 4 during the calibration 4 Conclusion In this paper, the method of measuring the pulse transit time and its using for calculation of blood pressure values have been introduced. During the short-time measurement time shift variations between the PPG and the Finapres signals in the record have been registered. It is supposed that the variations are influenced by artery branching. The wrong position of PPG and Finapres sensors at different branches of the artery may cause deviations in the expected correlation between pulse wave velocity and blood pressure. The pulse wave velocity is also influenced by many other effects [7], which cannot be totally suppressed at the time of measurement. Another question is the reliability of using the ECG as a reference signal. According to some authors [3], the peak of the R-wave is not a reliable reference point for the measurement of pulse wave propagation delay, although it is very suitable from the technical point of view. However, the finding of the optimal coefficients of approximating equation using the multiple correlation analysis and subsequent averaging of calculated beat-to-beat values enable a more precise determination of blood pressure. The method described in this paper can be used for non-invasive long-term (up to 42 h) monitoring of blood pressure. The fact that it is not necessary to use a pneumatic cuff is the main advantage of the method. It is because inflating the cuff disturbs the monitored person, especially in their sleep, and it can cause undesirable deviation of blood pressure. Acknowledgement: This work was supported by the Ministry of Education, Youth, and Sports of Czech Republic under project No. 2B06111 and research programme MSM Fig. 10 Beat-to-beat systolic blood pressure values for the person No. 4 five min. after the calibration References: [1] J. M. Padilla, E. J. Berjano, J. Siz, L. Fcila, P. Daz, S. Merce Assessment of relationships between blood pressure, pulse wave velocity and digital volume pulse, Computers in Cardiology, 2006; 33: [2] M. Magiuire, T. Ward, C. Markham, D. O Shea, L. Kevin, A comparative study in the use of brachial photoplethysmography and the QRS complex as timing references in determination of pulse transit time, in Proceedings of the 23rd Annual International ISBN:

6 Conference of the IEEE Engineering in Medicine and Biology Society, October 25-28, 2001, Turkey. [3] R. A. Payne, C. N. Symeonides, D. J. Webb, S. R. Maxwell, Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure, J.Appl. Physiol. 100: , [4] K. Říha, I. Potúček The Sequential Detection of Artery Sectional Area Using Optical Flow Technique. In Proceedings of The 8th WSEAS International Conference on Circuits, Systems, Electronics, Control & Signal Processing. Puerto De La Cruz (Spain), pp ISBN [5] M. Chmelař,R. Číž, O. Krajsa, J. Kouřil Biosignal Data Acquisition and its Postprocessing. In Advances in Biology, Bioengineering and Environment. WSEAS Press, Athens (Greece), pp ISBN: [6] D. A. McDonald, Blood flow in arteries, Southampton: The Camelot Press Ltd, 1974, ISBN [7] B. Choi, S. Kim, D. Jung, G. Jeon, Effect of factitious breathing on pulse transit time and heart rate. In IFBME Proceedings, Vol.14/2. ISBN:

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND D. Špulák 1, R. Čmejla 1, V. Fabián 2 Czech Technical University in Prague, Faculty of Electrical Engineering, 1 Department of Circuit Theory 2

More information

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time Soo-young Ye, Gi-Ryon Kim, Dong-Keun Jung, Seong-wan Baik, and Gye-rok Jeon Abstract In this paper, algorithm estimating the blood

More information

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (  The story so far. Medical Electronics Dr Neil Townsend Michaelmas Term 2001 (wwwrobotsoxacuk/~neil/teaching/lectures/med_elec) Blood Pressure has been measured (in some way) for centuries However, clinical interpretation

More information

Heart Abnormality Detection Technique using PPG Signal

Heart Abnormality Detection Technique using PPG Signal Heart Abnormality Detection Technique using PPG Signal L.F. Umadi, S.N.A.M. Azam and K.A. Sidek Department of Electrical and Computer Engineering, Faculty of Engineering, International Islamic University

More information

Blood Pressure Estimation Using Photoplethysmography (PPG)

Blood Pressure Estimation Using Photoplethysmography (PPG) Blood Pressure Estimation Using Photoplethysmography (PPG) 1 Siddhi Sham Karande, BE E&TC, VIIT Pune. 2 Kiran Rajendrasingh Thakur, BE E&TC, VIIT Pune. 3 Sneha Dattatraya Waghmare, BE E&TC, VIIT Pune 4

More information

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 64 CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 4.1 GENERAL This chapter presents the methodologies that are usually adopted for the measurement of blood pressure, heart rate and pulse

More information

NONINVASIVE CONTINUOUS BLOOD PRESSURE DETERMINATION. D. Barschdorff*, M. Erig*, E. Trowitzsch**

NONINVASIVE CONTINUOUS BLOOD PRESSURE DETERMINATION. D. Barschdorff*, M. Erig*, E. Trowitzsch** NONINVASIVE CONTINUOUS BLOOD PRESSURE DETERMINATION D. Barschdorff*, M. Erig*, E. Trowitzsch** *Institute of Electrical Measurement University of Paderborn, D3398 Paderborn, Germany **Sleep Laboratories

More information

Results of geometric ballistocardiography

Results of geometric ballistocardiography Results of geometric ballistocardiography D. Jezbera, J. Kříž, J. Rajsner, F. Studnička, P. Šeba Abstract Results of application of differential geometry on ballistocardiographic signals is presented.

More information

EVALUATION OF NONINVASIVE PULSE TRANSIT TIME METHODOLOGIES FOR DIAGNOSIS OF HYPERTENSION DANIEL JOHN BADGER. A thesis submitted to the

EVALUATION OF NONINVASIVE PULSE TRANSIT TIME METHODOLOGIES FOR DIAGNOSIS OF HYPERTENSION DANIEL JOHN BADGER. A thesis submitted to the EVALUATION OF NONINVASIVE PULSE TRANSIT TIME METHODOLOGIES FOR DIAGNOSIS OF HYPERTENSION by DANIEL JOHN BADGER A thesis submitted to the Graduate School-New Brunswick Rutgers, The State University of New

More information

Blood Pressure Laboratory

Blood Pressure Laboratory Introduction The blood that circulates throughout the body maintains a flow and pressure. The nervous system can change the flow and pressure based on the particular needs at a given time. For example,

More information

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis Journal of Physics: Conference Series PAPER OPEN ACCESS Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis To cite this article: Valerie Tan et al 2018 J. Phys.: Conf. Ser. 1049 012062

More information

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes Dual channel photoplethysmography studies of cardio-vascular response to the body position changes R. Erts, I. Kukulis, J. Spigulis, L. Kere University of Latvia, Raina Blvd. 9, Riga, LV-, Latvia Renars.erts@lu.lv

More information

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive

More information

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination JIRI JILEK 1, MILAN STORK 2 1 Carditech, Culver City, California, USA jilekj@usa.net 2 Dept. of Applied Electronics and

More information

Biomedical Instrumentation E. Blood Pressure

Biomedical Instrumentation E. Blood Pressure Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body

More information

Blood Pressure (BP) measurement becomes a vital parameter now days due to continuous increase in chronic heart

Blood Pressure (BP) measurement becomes a vital parameter now days due to continuous increase in chronic heart ISSN: 0975-766X CODEN: IJPTFI Available Online through Research Article www.ijptonline.com FAST AND ACCURATE METHOD FOR CONTINUOUS BP MEASUREMENT Harkirat Kaur*, Dr. Sunil Agrawal, Mr. Sarvjit Singh Dept.

More information

PCA Enhanced Kalman Filter for ECG Denoising

PCA Enhanced Kalman Filter for ECG Denoising IOSR Journal of Electronics & Communication Engineering (IOSR-JECE) ISSN(e) : 2278-1684 ISSN(p) : 2320-334X, PP 06-13 www.iosrjournals.org PCA Enhanced Kalman Filter for ECG Denoising Febina Ikbal 1, Prof.M.Mathurakani

More information

Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable Device

Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable Device 2016 3 rd International Conference on Engineering Technology and Application (ICETA 2016) ISBN: 978-1-60595-383-0 Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable

More information

Measurement of Hemodynamic Parameters: Design of Methods and Hardware

Measurement of Hemodynamic Parameters: Design of Methods and Hardware Measurement of Hemodynamic Parameters: Design of Methods and Hardware Jan Havlík xhavlikj@fel.cvut.cz Lenka Lhotská Dep. of Cybernetics lhotska@fel.cvut.cz Vratislav Fabián Dep. of Physics fabiav1@fel.cvut.cz

More information

Identification of Ischemic Lesions Based on Difference Integral Maps, Comparison of Several ECG Intervals

Identification of Ischemic Lesions Based on Difference Integral Maps, Comparison of Several ECG Intervals 1.2478/v148-9-21-7 MEASUREMENT SCIENCE REVIEW, Volume 9, No. 5, 29 Identification of Ischemic Lesions Based on Difference Integral Maps, Comparison of Several ECG Intervals J. Švehlíková, M. Kania 1, M.

More information

LATEST TRENDS on SYSTEMS (Volume I) New Seismocardiographic Measuring System with Separate QRS Detection. M. Stork 1, Z. Trefny 2

LATEST TRENDS on SYSTEMS (Volume I) New Seismocardiographic Measuring System with Separate QRS Detection. M. Stork 1, Z. Trefny 2 New Seismocardiographic Measuring System with Separate QRS Detection M. Stork 1, Z. Trefny 2 1 Department of Applied Electronics and Telecommunications University of West Bohemia, P.O. Box 314, 30614 Plzen,

More information

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation J. JILEK*, M. STORK** *Carditech, Culver City, California, USA **Department of Applied Electronics

More information

Accuracy of Beat-to-Beat Heart Rate Estimation Using the PulseOn Optical Heart Rate Monitor

Accuracy of Beat-to-Beat Heart Rate Estimation Using the PulseOn Optical Heart Rate Monitor Accuracy of Beat-to-Beat Heart Rate Estimation Using the PulseOn Optical Heart Rate Monitor Abstract Wrist photoplethysmography allows unobtrusive monitoring of the heart rate (HR). Even if it is frequently

More information

1 Non-invasive measurement of arterial pressure

1 Non-invasive measurement of arterial pressure Non-invasive measurement of arterial pressure I. Background A. Circulatory systems Human circulation takes place in a closed system that consists of two subsystems, pulmonary circulation and systemic circulation,

More information

Heart Rate Analysis for Human Factors: Development and Validation of an Open Source Toolkit for Noisy Naturalistic Heart Rate Data

Heart Rate Analysis for Human Factors: Development and Validation of an Open Source Toolkit for Noisy Naturalistic Heart Rate Data Full paper for 6 th Humanist Conference, The Hague, Netherlands, 13-14 June 2018 Heart Rate Analysis for Human Factors: Development and Validation of an Open Source Toolkit Van Gent, P., Delft University

More information

Science in Sport. 204a ECG demonstration (Graph) Read. The Electrocardiogram. ECG Any 12 bit EASYSENSE. Sensors: Loggers: Logging time: 10 seconds

Science in Sport. 204a ECG demonstration (Graph) Read. The Electrocardiogram. ECG Any 12 bit EASYSENSE. Sensors: Loggers: Logging time: 10 seconds Sensors: Loggers: ECG Any 12 bit EASYSENSE Science in Sport Logging time: 10 seconds 204a ECG demonstration (Graph) Read Regular medical check ups are essential part of the life of a professional sports

More information

A bioimpedance-based cardiovascular measurement system

A bioimpedance-based cardiovascular measurement system A bioimpedance-based cardiovascular measurement system Roman Kusche 1[0000-0003-2925-7638], Sebastian Hauschild 1, and Martin Ryschka 1 1 Laboratory of Medical Electronics, Luebeck University of Applied

More information

A REVIEW ON NON-INVASIVE HYPERTENSION MONITORING SYSTEM BY USING PHOTOPLETHYSMOGRAPHY METHOD

A REVIEW ON NON-INVASIVE HYPERTENSION MONITORING SYSTEM BY USING PHOTOPLETHYSMOGRAPHY METHOD A REVIEW ON NON-INVASIVE HYPERTENSION MONITORING SYSTEM BY USING PHOTOPLETHYSMOGRAPHY METHOD Zahari Taha, Lum Shirley * and Mohd Azraai Mohd Razman Innovative Manufacturing, Mechatronics and Sports (imams)

More information

Electronic System for Measuring and Optimization of Pacemaker Pulses

Electronic System for Measuring and Optimization of Pacemaker Pulses Electronic System for Measuring and Optimization of Pacemaker Pulses MILAN STORK 1, VLASTIMIL VANCURA 2 1 Department of Applied Electronics and Telecommunications, RICE University of West Bohemia, Czech

More information

WIRELESS PATIENT HEALTH MONITORING SYSTEM

WIRELESS PATIENT HEALTH MONITORING SYSTEM WIRELESS PATIENT HEALTH MONITORING SYSTEM PADMAVATHI KAVURU 1, B.CHINNA RAO 2 & P.M FRANCIS 3 1 (VLSI&ES) 2 Electronics & Communication Engg, Gokul Institute of Technology & Sciences (JNTUK): Bobbili,

More information

LABVIEW based expert system for Detection of heart abnormalities

LABVIEW based expert system for Detection of heart abnormalities LABVIEW based expert system for Detection of heart abnormalities Saket Jain Piyush Kumar Monica Subashini.M School of Electrical Engineering VIT University, Vellore - 632014, Tamil Nadu, India Email address:

More information

A New Approach for Blood Pressure Monitoring based on ECG and PPG Signals by using Artificial Neural Networks

A New Approach for Blood Pressure Monitoring based on ECG and PPG Signals by using Artificial Neural Networks A New Approach for Blood Pressure Monitoring based on ECG and PPG Signals by using Artificial Neural Networks Younessi Heravi Mohammad Amin Department of Basic Sciences, School of Medicine, North Khorasan

More information

Estimation of Haemoglobin Using Non-Invasive Technique

Estimation of Haemoglobin Using Non-Invasive Technique Estimation of Haemoglobin Using Non-Invasive Technique Tatiparti Padma #1, Pinjala Jahnavi *2 #Department of ECE, GRIET, Hyderabad, Telangana *Department of ECE, GRIET, Hyderabad, Telangana 1tatipartipadma@gmail.com,

More information

ECG SENSOR ML84M USER S GUIDE. CENTRE FOR MICROCOMPUTER APPLICATIONS

ECG SENSOR ML84M USER S GUIDE. CENTRE FOR MICROCOMPUTER APPLICATIONS ECG SENSOR ML84M USER S GUIDE CENTRE FOR MICROCOMPUTER APPLICATIONS http://www.cma-science.nl Short description The ECG sensor measures electrical potentials produced by the heart (Electro- cardiogram).

More information

Simulation of the Blood Pressure Estimation using the Artery Compliance Model and Pulsation Waveform Model

Simulation of the Blood Pressure Estimation using the Artery Compliance Model and Pulsation Waveform Model Simulation of the Blood Pressure Estimation using the Artery Compliance Model and Pulsation Waveform Model Ahyoung Jeon*, Sooyoung Ye**, Gyerok Jeon* *Department of Biomedical-engineering Pusan National

More information

Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease

Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease Lab #10 Table of Contents: Expected Learning Outcomes...... 83 Introduction....... 84 Activity 1: Collecting ECG Data..... 85 Activity

More information

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection Proceedings of the 2012 International Conference on Industrial Engineering and Operations Management Istanbul, Turkey, July 3 6, 2012 Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter

More information

Lab 4: Introduction to Physiological Measurements - Cardiovascular

Lab 4: Introduction to Physiological Measurements - Cardiovascular Lab 4: Introduction to Physiological Measurements - Cardiovascular INTRODUCTION: This lab will demonstrate cardiovascular measurements by creating an ECG with instruments used in previous labs. Students

More information

Application of PPG Signal Processing to Patients undergoing Angioplasty Procedure

Application of PPG Signal Processing to Patients undergoing Angioplasty Procedure Application of PPG Signal Processing to Patients undergoing Angio Procedure D Narayana Dutt 1, Nikhil C 2, Akshatha P S 3, Chaithra V K 4, Animita Das 5 Department of Medical Electronics, Dayananda Sagar

More information

BP-600 Noninvasive Blood Pressure Sensor

BP-600 Noninvasive Blood Pressure Sensor Technical Note BP-600 Overview A person's cardiac output, peripheral vascular resistance, blood pressure, and other cardiovascular parameters change in response to the activities and events taking place

More information

LabVIEW based Non-Invasive prototype device for cardiac diagnosis using Nadi Shastra

LabVIEW based Non-Invasive prototype device for cardiac diagnosis using Nadi Shastra Original Article LabVIEW based Non-Invasive prototype device for cardiac diagnosis using Nadi Shastra Parth Shah 1, A.V. Leio 1, Siddaram Mahajan* 1, Shaik Jameel 2 1 Department of Biomedical Engineering,

More information

Keun-Ho Kim and Sang-Suk Lee* Department of Oriental Biomedical Engineering, Sangji University, Wonju , Korea

Keun-Ho Kim and Sang-Suk Lee* Department of Oriental Biomedical Engineering, Sangji University, Wonju , Korea Journal of Magnetics 20(1), 47-51 (2015) ISSN (Print) 1226-1750 ISSN (Online) 2233-6656 http://dx.doi.org/10.4283/jmag.2015.20.1.047 Peripheral Blood Flow Velocity and Peripheral Pulse Wave Velocity Measured

More information

IJRIM Volume 1, Issue 2 (June, 2011) (ISSN ) ECG FEATURE EXTRACTION FOR CLASSIFICATION OF ARRHYTHMIA. Abstract

IJRIM Volume 1, Issue 2 (June, 2011) (ISSN ) ECG FEATURE EXTRACTION FOR CLASSIFICATION OF ARRHYTHMIA. Abstract ECG FEATURE EXTRACTION FOR CLASSIFICATION OF ARRHYTHMIA Er. Ankita Mittal* Er. Saurabh Mittal ** Er. Tajinder Kaur*** Abstract Artificial Neural Networks (ANN) can be viewed as a collection of identical

More information

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work Electrical Activity of the Human Heart Oguz Poroy, PhD Assistant Professor Department of Biomedical Engineering The University of Iowa Outline Basic Facts about the Heart Heart Chambers and Heart s The

More information

Influence of 50-Hz electric and magnetic fields on human blood pressure

Influence of 50-Hz electric and magnetic fields on human blood pressure Radiat Environ Biophys (1996) 35: 199 204 Springer-Verlag 1996 ORIGINAL PAPER L. Korpinen J. Partanen Influence of 50-Hz electric and magnetic fields on human blood pressure Received: 6 June 1994 / Accepted

More information

Author's copy - produced at Lamar University

Author's copy - produced at Lamar University BLOOD PRESSURE ESTIMATION USING PHOTOPLETHYSMOGRAPHY WITH TELEMEDICINE APPLICATION (Centered, double spaced if more than one line, all capitals) (2 inches from title to A Thesis ) A Dissertation Presented

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms

Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms J. Jilek*, M. Stork** *Carditech, Culver City, California, USA **Department of Applied Electronics and Telecommunications University

More information

Cardiovascular Diseases Detecting via Pulse Analysis

Cardiovascular Diseases Detecting via Pulse Analysis Engineering, 2013, 5, 176-180 http://dx.doi.org/10.4236/eng.2013.510b038 Published Online October 2013 (http://www.scirp.org/journal/eng) Cardiovascular Diseases Detecting via Pulse Analysis Jingjing Xia,

More information

HUMAN ANATOMY AND PHYSIOLOGY

HUMAN ANATOMY AND PHYSIOLOGY HUMAN ANATOMY AND PHYSIOLOGY NAME Detection of heart sounds. Clean the ear pieces of the stethoscope before using. The ear pieces should be pointing slightly forward when inserted into the ears because

More information

The Investigation of the Effect of Aging through Photoplethysmogram Signal Analysis of Erectile Dysfunction Subjects

The Investigation of the Effect of Aging through Photoplethysmogram Signal Analysis of Erectile Dysfunction Subjects The Investigation of the Effect of Aging through Photoplethysmogram Signal Analysis of Erectile Dysfunction Subjects Y. K. QAWQZEH 1, M. B. I. REAZ 1, O. MASKON 2, KALAIVANI CHELLAPPAN 1, M. T. ISLAM 3,

More information

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model Gye-rok Jeon, Jae-hee Jung, In-cheol Kim, Ah-young Jeon, Sang-hwa Yoon, Jung-man Son, Jae-hyung Kim, Soo-young Ye,

More information

QUANTIFICATION OF EMOTIONAL FEATURES OF PHOTOPLETHYSOMOGRAPHIC WAVEFORMS USING BOX-COUNTING METHOD OF FRACTAL DIMENSION

QUANTIFICATION OF EMOTIONAL FEATURES OF PHOTOPLETHYSOMOGRAPHIC WAVEFORMS USING BOX-COUNTING METHOD OF FRACTAL DIMENSION QUANTIFICATION OF EMOTIONAL FEATURES OF PHOTOPLETHYSOMOGRAPHIC WAVEFORMS USING BOX-COUNTING METHOD OF FRACTAL DIMENSION Andrews Samraj*, Nasir G. Noma*, Shohel Sayeed* and Nikos E. Mastorakis** *Faculty

More information

Study and Design of a Shannon-Energy-Envelope based Phonocardiogram Peak Spacing Analysis for Estimating Arrhythmic Heart-Beat

Study and Design of a Shannon-Energy-Envelope based Phonocardiogram Peak Spacing Analysis for Estimating Arrhythmic Heart-Beat International Journal of Scientific and Research Publications, Volume 4, Issue 9, September 2014 1 Study and Design of a Shannon-Energy-Envelope based Phonocardiogram Peak Spacing Analysis for Estimating

More information

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure #6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure Objectives: Observe slide of artery and vein cross-section Auscultate heart sounds using a stethoscope Measure

More information

Human Anatomy and Physiology II Laboratory Cardiovascular Physiology

Human Anatomy and Physiology II Laboratory Cardiovascular Physiology Human Anatomy and Physiology II Laboratory Cardiovascular Physiology 1 This lab involves two exercises: 1) Conduction System of the Heart and Electrocardiography and 2) Human Cardiovascular Physiology:

More information

Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique

Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique www.jbpe.org Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique Original 1 Department of Biomedical Engineering, Amirkabir University of technology, Tehran, Iran Abbaspour

More information

5.2. MEASUREMENT OF BLOOD PRESSURE

5.2. MEASUREMENT OF BLOOD PRESSURE 5.2. MEASUREMENT OF BLOOD PRESSURE Purpose of experiment To examine a person s blood pressure. Tasks of experiment To record the pressure changes in the vascular system: using the traditional method measure

More information

Systolic Blood Pressure Estimation Using Video Processing

Systolic Blood Pressure Estimation Using Video Processing 1 Systolic Blood Pressure Estimation Using Video Processing Dao Lam, Gung-Yu Pan, Patrick Hsu, Albert Xu and Chung-Kuan Cheng Department of Computer Science and Engineering, University of California, San

More information

DETECTION OF HEART ABNORMALITIES USING LABVIEW

DETECTION OF HEART ABNORMALITIES USING LABVIEW IASET: International Journal of Electronics and Communication Engineering (IJECE) ISSN (P): 2278-9901; ISSN (E): 2278-991X Vol. 5, Issue 4, Jun Jul 2016; 15-22 IASET DETECTION OF HEART ABNORMALITIES USING

More information

SYNCHRONIZED MEASUREMENTS OF MAXIMUM BLOOD FLOW VELOCITIES IN CAROTID, BRACHIAL AND FEMORAL ARTERIES, AND ECG IN HUMAN POSTURE CHANGES

SYNCHRONIZED MEASUREMENTS OF MAXIMUM BLOOD FLOW VELOCITIES IN CAROTID, BRACHIAL AND FEMORAL ARTERIES, AND ECG IN HUMAN POSTURE CHANGES SYNCHRONIZED MEASUREMENTS OF MAXIMUM BLOOD FLOW VELOCITIES IN CAROTID, BRACHIAL AND FEMORAL ARTERIES, AND ECG IN HUMAN POSTURE CHANGES Y.Hirao*, **, T.Kuroda*, D.Zhang*, Y.Kinouchi*, H.Yamaguchi*** and

More information

JAMES R. KNIGHT

JAMES R. KNIGHT JAMES R. KNIGHT knightjr@ah.org James is currently the supervisor of clinical engineering at Sonora Regional Medical Center. He is the chairman of the CMIA Training & Education Committee. He also develops

More information

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS LAB 8: CARDIOVASCULAR PHYSIOLOGY PART 1. HEART SOUNDS AND PULSE DETERMINATIONS Introduction Two distinct sounds can be heard during each cardiac cycle. These sounds are commonly described as lub and dup

More information

Monitoring Cardiac Health Using a Laser Doppler Approach to Measure Pulse Wave Velocity

Monitoring Cardiac Health Using a Laser Doppler Approach to Measure Pulse Wave Velocity Monitoring Cardiac Health Using a Laser Doppler Approach to Measure Pulse Wave Velocity Madison Boston - Team Leader Michal Adamski - Communicator Haley Knapp - BSAC Katie Barlow - BWIG Lizzie Krasteva

More information

PROCESSING THE ABDOMINAL FETAL ECG USING A NEW METHOD. Zentralinstitut fur Biomedizinische Technik der Universitat Erlangen-Nurnberg, FRG

PROCESSING THE ABDOMINAL FETAL ECG USING A NEW METHOD. Zentralinstitut fur Biomedizinische Technik der Universitat Erlangen-Nurnberg, FRG PROCESSING THE ABDOMINAL FETAL ECG USING A NEW METHOD J Nagel, M Schaldach Zentralinstitut fur Biomedizinische Technik der Universitat Erlangen-Nurnberg, FRG INTRODUCTION The excellent natural shielding

More information

An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG

An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG Introduction An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG data can give important information about the health of the heart and can help physicians to diagnose

More information

Optical sensor for Indian Siddha Diagnosis System

Optical sensor for Indian Siddha Diagnosis System Available online at www.sciencedirect.com Procedia Engineering 38 (2012 ) 1126 1131 International conference on modeling, optimization and computing Optical sensor for Indian Siddha Diagnosis System N.Deepa*,

More information

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 10, April 2013

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 10, April 2013 ECG Processing &Arrhythmia Detection: An Attempt M.R. Mhetre 1, Advait Vaishampayan 2, Madhav Raskar 3 Instrumentation Engineering Department 1, 2, 3, Vishwakarma Institute of Technology, Pune, India Abstract

More information

PHONOCARDIOGRAPHY (PCG)

PHONOCARDIOGRAPHY (PCG) PHONOCARDIOGRAPHY (PCG) The technique of listening to sounds produced by the organs and vessels of the body is called auscultation. The areas at which the heart sounds are heard better are called auscultation

More information

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method To cite

More information

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03 Introduction Cardiac Imaging Modalities Arnaud Bistoquet 12/19/03 Coronary heart disease: the vessels that supply oxygen-carrying blood to the heart, become narrowed and unable to carry a normal amount

More information

12.2 Monitoring the Human Circulatory System

12.2 Monitoring the Human Circulatory System 12.2 Monitoring the Human Circulatory System Video 1: 3D Animation of Heart Pumping Blood blood flow through the heart... Video 2: Hank Reviews Everything on the Heart https://www.youtube.com/watch?v=x9zz6tcxari

More information

: Biomedical Signal Processing

: Biomedical Signal Processing : Biomedical Signal Processing 0. Introduction: Biomedical signal processing refers to the applications of signal processing methods, such as Fourier transform, spectral estimation and wavelet transform,

More information

Biomedical Signal Processing

Biomedical Signal Processing DSP : Biomedical Signal Processing What is it? Biomedical Signal Processing: Application of signal processing methods, such as filtering, Fourier transform, spectral estimation and wavelet transform, to

More information

Design of the HRV Analysis System Based on AD8232

Design of the HRV Analysis System Based on AD8232 207 3rd International Symposium on Mechatronics and Industrial Informatics (ISMII 207) ISB: 978--60595-50-8 Design of the HRV Analysis System Based on AD8232 Xiaoqiang Ji,a, Chunyu ing,b, Chunhua Zhao

More information

Development of Non-Restraint Blood Pressure Measurement Method Utilizing PWV

Development of Non-Restraint Blood Pressure Measurement Method Utilizing PWV Development of Non-Restraint Blood Pressure Measurement Method Utilizing PWV Yasunari Asakura School of Integrative and Global Majors, University of Tsukuba, Tsukuba, Japan Email: asakura@ golem.kz.tsukuba.ac.jp

More information

TEPZZ _ 849 A_T EP A1 (19) (11) EP A1. (12) EUROPEAN PATENT APPLICATION published in accordance with Art.

TEPZZ _ 849 A_T EP A1 (19) (11) EP A1. (12) EUROPEAN PATENT APPLICATION published in accordance with Art. (19) TEPZZ _ 849 A_T (11) EP 3 138 493 A1 (12) EUROPEAN PATENT APPLICATION published in accordance with Art. 3(4) EPC (43) Date of publication: 08.03.17 Bulletin 17/ (21) Application number: 786271. (22)

More information

DEVELOPMENT OF NON-INVASIVE STEERING-TYPE BLOOD PRESSURE SENSOR FOR DRIVER STATE DETECTION. Received December 2017; revised April 2018

DEVELOPMENT OF NON-INVASIVE STEERING-TYPE BLOOD PRESSURE SENSOR FOR DRIVER STATE DETECTION. Received December 2017; revised April 2018 International Journal of Innovative Computing, Information and Control ICIC International c 2018 ISSN 1349-4198 Volume 14, Number 4, August 2018 pp. 1301 1310 DEVELOPMENT OF NON-INVASIVE STEERING-TYPE

More information

MR Advance Techniques. Cardiac Imaging. Class IV

MR Advance Techniques. Cardiac Imaging. Class IV MR Advance Techniques Cardiac Imaging Class IV Heart The heart is a muscular organ responsible for pumping blood through the blood vessels by repeated, rhythmic contractions. Layers of the heart Endocardium

More information

Special Lecture 10/28/2012

Special Lecture 10/28/2012 Special Lecture 10/28/2012 HYPERTENSION Dr. HN Mayrovitz Special Lecture 10/28/2012 Arterial Blood Pressure (ABP) - Definitions ABP Review Indirect Oscillographic Method Resistance (R), Compliance (C)

More information

Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference

Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference ORIGINAL RESEARCH Physiological Reports ISSN 2051-817X Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

Cardiovascular Physiology

Cardiovascular Physiology Cardiovascular Physiology The mammalian heart is a pump that pushes blood around the body and is made of four chambers: right and left atria and right and left ventricles. The two atria act as collecting

More information

Design of Context-Aware Exercise Measurement SoC Based on Electromyogram and Electrocardiogram

Design of Context-Aware Exercise Measurement SoC Based on Electromyogram and Electrocardiogram Design of Context-Aware Exercise Measurement SoC Based on Electromyogram and Electrocardiogram Seongsoo Lee Abstract Exercise measurement is an important application of smart healthcare devices. Conventional

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Hemon, M. C. & Phillips, J. P. (2015). Comparison of foot finding methods for deriving instantaneous pulse rates from

More information

V. Marozas 1, S. Daukantas 1, A. Petrėnas 1, A. Sološenko 1, D. Stankevičius 1, B. Paliakaitė 1, D. Jegelevičius 1, A. Lukoševičius 1, E.

V. Marozas 1, S. Daukantas 1, A. Petrėnas 1, A. Sološenko 1, D. Stankevičius 1, B. Paliakaitė 1, D. Jegelevičius 1, A. Lukoševičius 1, E. V. Marozas 1, S. Daukantas 1, A. Petrėnas 1, A. Sološenko 1, D. Stankevičius 1, B. Paliakaitė 1, D. Jegelevičius 1, A. Lukoševičius 1, E. Kaldoudi 2 1 Biomedical Engineering Institute, Kaunas University

More information

MULTILEAD SIGNAL PREPROCESSING BY LINEAR TRANSFORMATION

MULTILEAD SIGNAL PREPROCESSING BY LINEAR TRANSFORMATION MULTILEAD SIGNAL PREPROCESSING BY LINEAR TRANSFORMATION TO DERIVE AN ECG LEAD WHERE THE ATYPICAL BEATS ARE ENHANCED Chavdar Lev Levkov Signa Cor Laboratory, Sofia, Bulgaria, info@signacor.com ECG signal

More information

PULSE WAVE VARIABILITY WITHIN TWO SHORT-TERM MEASUREMENTS. David Korpas, Jan Halek

PULSE WAVE VARIABILITY WITHIN TWO SHORT-TERM MEASUREMENTS. David Korpas, Jan Halek Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2006, 150(2):339 344. D. Korpas, J. Halek 339 PULSE WAVE VARIABILITY WITHIN TWO SHORT-TERM MEASUREMENTS David Korpas, Jan Halek Department of Medical

More information

The Heart 1 of 34 Boardworks Ltd 2012

The Heart 1 of 34 Boardworks Ltd 2012 The Heart 1 of 34 Boardworks Ltd 2012 2 of 34 Boardworks Ltd 2012 What does the heart do? 3 of 34 Boardworks Ltd 2012 The heart is a muscular organ located in the thorax. It pumps blood continuously around

More information

EGR1200 Freshman Mini Project Biomedical Signal Analysis

EGR1200 Freshman Mini Project Biomedical Signal Analysis EGR1200 Freshman Mini Project Biomedical Signal Analysis Instructors Meltem Izzetoglu, PhD Electrical & Computer Engineering And Andrea Welker, PhD Civil and Environmental Engineering Fall 2018 Engineering

More information

Temporal Analysis and Remote Monitoring of ECG Signal

Temporal Analysis and Remote Monitoring of ECG Signal Temporal Analysis and Remote Monitoring of ECG Signal Amruta Mhatre Assistant Professor, EXTC Dept. Fr.C.R.I.T. Vashi Amruta.pabarekar@gmail.com Sadhana Pai Associate Professor, EXTC Dept. Fr.C.R.I.T.

More information

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin Ashita Diwate 1, Yogesh Kumar 2, Dr. Sanjeev Kumar 3, Prof. T.

More information

Permanent City Research Online URL:

Permanent City Research Online URL: Kyriacou, P. A., Pal, S. K., Langford, R. & Jones, DP (2006). Electro-optical techniques for the investigation of oesophageal photoplethysmographic signals and blood oxygen saturation in burns. Measurement

More information

A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS

A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS R. Mukkamala, R. G. Mark, R. J. Cohen Haard-MIT Division of Health Sciences and Technology, Cambridge, MA, USA Abstract We

More information

Chapter 3 Biological measurement 3.1 Nerve conduction

Chapter 3 Biological measurement 3.1 Nerve conduction Chapter 3 Biological measurement 3.1 Nerve conduction Learning objectives: What is in a nerve fibre? How does a nerve fibre transmit an electrical impulse? What do we mean by action potential? Nerve cells

More information

II. PROCEDURE DESCRIPTION A. Normal Waveform from an Electrocardiogram Figure 1 shows two cycles of a normal ECG waveform.

II. PROCEDURE DESCRIPTION A. Normal Waveform from an Electrocardiogram Figure 1 shows two cycles of a normal ECG waveform. Cardiac Monitor with Mobile Application and Alert System Miguel A. Goenaga-Jimenez, Ph.D. 1, Abigail C. Teron, BS. 1, Pedro A. Rivera 1 1 Universidad del Turabo, Puerto Rico, mgoenaga1@suagm.edu, abigailteron@gmail.com,

More information

ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS

ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS V.S. Murthy 1, Sripad Ramamoorthy 1, Narayanan Srinivasan 2, Sriram Rajagopal 2, M. Mukunda Rao 3 1 Department of Physics, Indian Institute

More information

Non-invasive examination

Non-invasive examination Non-invasive examination Segmental pressure and Ankle-Brachial Index (ABI) The segmental blood pressure (SBP) examination is a simple, noninvasive method for diagnosing and localizing arterial disease.

More information

Biomedical Instrumentation

Biomedical Instrumentation Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Therapeutic and Prosthetic Devices 1 Figure 13.1 Block diagram of an asynchronous

More information

ECE ECE PRINCIPLES OF BIOMEDICAL SYSTEMS & DEVICES LAB 1 - ELECTROCARDIOGRAM

ECE ECE PRINCIPLES OF BIOMEDICAL SYSTEMS & DEVICES LAB 1 - ELECTROCARDIOGRAM ECE0909.404.01 ECE 0909.504.03 PRINCIPLES OF BIOMEDICAL SYSTEMS & DEVICES LAB 1 - ELECTROCARDIOGRAM The purpose of this laboratory is to introduce you to electrocardiogram, its acquisition and interpretation.

More information