The Exeter Universal stem

Size: px
Start display at page:

Download "The Exeter Universal stem"

Transcription

1 The Exeter Universal stem A MINIMUM TEN-YEAR REVIEW FROM AN INDEPENDENT CENTRE S. Hook, E. Moulder, P. J. Yates, B. J. Burston, E. Whitley, G. C. Bannister From the Avon Orthopaedic Centre, Bristol, England S. Hook, MRCS, Specialist Registrar E. Moulder, MRCS, Specialist Registrar B. J. Burston, MB ChB(Hons), MRCS, Specialist Registrar E. Whitley, PhD, MSc, BSc(Hons), Medical Statistician G. C. Bannister, MD, MCh(Orth), FRCS, FRCS Ed(Orth), Consultant Orthopaedic Surgeon Avon Orthopaedic Centre Southmead Hospital, Westbury-on-Trym, Bristol BS10 5NB, UK. P. J. Yates, MBBS(Hons), BSc (Hons), MRCS, FRCS(Trauma & Orth), Orthopaedic Consultant and Senior Lecturer University of Western Australia, Fremantle Hospital, Alma Road, Perth 6160, Western Australia. Correspondence should be sent to Mr P. J. Yates; piersyates@hotmail.com 2006 British Editorial Society of Bone and Joint Surgery doi: / x.88b $2.00 J Bone Joint Surg [Br] 2006;88-B: Received 30 June 2006; Accepted after revision 15 August 2006 We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up. The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips. All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure. Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p < 0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p < 0.001). The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis. Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results. The original Exeter total hip replacement (THR) was implanted using a first-generation cementing technique between 1970 and It was the first true double-tapered, polished stem and was designed to be inserted through the posterior approach. A 30 mm head was used initially to reduce the perceived increased risk of dislocation, relative to the mm Charnley head. The taper was designed to improve compression of the cement during insertion of the stem. The sole reason for having a polished stem was that the British Standard 3531 (1962) required that stems manufactured from the stainless-steel alloy EN58J should be polished. The results were excellent, although there was subsidence at the cement-stem interface. There was almost no loosening at the cement-bone interface and good preservation of the proximal medial bone stock. 1,2 Because of reports of a few cases of fracture of the neck of the prosthesis, in 1976 the metal was changed to 316L with a matt surface finish. Unfortunately, the importance of the surface finish was not appreciated at the time and there was some delay before the inferior results with the matt finish became apparent. 3-6 High-nitrogen stainless steel was used from 1983 (Rex 734, Orthinox, Howmedica International Ltd, London, United Kingdom), and the implant became polished again in In 1988, the distal taper was incorporated into a modular prosthesis named the Exeter Universal (Howmedica International Ltd, London, United Kingdom) (Fig. 1), which featured a more complete taper, especially in the anteroposterior proximal geometry. The Scandinavian arthroplasty registries reported good results with the Exeter hip system, but did not distinguish between failure of the acetabular and the femoral component and gave no full account of the radiological assessment. 3 The only detailed reports of the Exeter system have come from the original centre. 1, THE JOURNAL OF BONE AND JOINT SURGERY

2 THE EXETER UNIVERSAL STEM 1585 Fig. 1a Fig. 1b Photographs of the Exeter Universal stem. Fowler et al 1 reviewed the older designs of femoral component implanted using first-generation cementing techniques and Williams et al 7 described the use of the third-generation Exeter Universal stem, but at a mean of only 8.9 years. The other Exeter series, unfortunately, described the use of a mixture of implants. 8,9 Our aim was to establish whether published results for the Exeter Universal stem were reproducible in another centre and to identify the features associated with the survival of this prosthesis. Fig. 2 Diagram of the defects of the cement mantle zones of Gruen et al. 13 Patients and Methods Between August 1988 and May 1993, 123 consecutive patients (142 hips) underwent a primary THR with the Exeter Universal stem and a variety of acetabular components. The size of the femoral head varied between 22 mm and 32 mm depending on the type and size of the acetabular component. Surgery was performed either by the senior author (GCB) or by trainees under his supervision. All the hips were implanted using the posterolateral approach. The insertions of gluteus maximus and psoas were left intact and the short external rotators were not repaired. The femoral components were cemented using a secondgeneration cementing technique. The femoral canal was prepared by thorough curettage, non-pressurised lavage and drying. The intramedullary canal was occluded 16 cm distal to the tip of the greater trochanter using a Hardinge restrictor (Corin, Cirencester, United Kingdom). Palacos R bone cement with gentamicin (Schering-Plough Ltd, Welwyn Garden City, United Kingdom) was mixed in an open bowl and inserted prograde in dough form using a disposable cement syringe (Centerpulse Ltd, Zurich, Switzerland), a venting tube and proximal digital pressurisation. The Exeter Universal stem was used in conjunction with the Exeter cemented metal-backed acetabular component in 32 hips (Howmedica International Ltd), the cemented Elite in 23 hips (De Puy International Ltd, Leeds, United Kingdom) or the uncemented Harris-Galante I acetabular component in 33 hips (Zimmer Ltd, Swindon, United Kingdom). The Exeter acetabular prosthesis had an inside diameter of 26 mm and was abandoned in March 1991 in favour of the Harris-Galante and Elite components because of poor instrumentation and a number of early dislocations. The acetabulum for the cemented components was prepared with sequential hemispherical reamers without preserving the subchondral bone. Most of the earlier Exeter acetabular components in the series were undersized to allow for thick cement mantles. The uncemented Harris- Galante acetabular component was generally used in younger patients (< 65 years) and was fixed by two screws in the superior aspect of the ilium. Clinical assessment. The patients and their records were reviewed and the clinical outcome was measured using the Oxford hip score. 10 Radiological assessment. The radiological outcome was assessed using sequential, calibrated anteroposterior radiographs of the pelvis and lateral views of the affected hips by two authors (BJB, PJY). Radiological variables. We used well-recognised assessment techniques to standardise the radiological and clinical evaluation to allow easier comparison with other prostheses as recommended by Johnston et al. 11 The cementing technique as described by Barrack, Mulroy and Harris, 12 was assessed in all the 14 zones of Gruen, McNeice and Amstutz 13 (Fig. 2). Grade A is defined as complete filling of the medullary cavity by cement, grade B as the presence of slight radio- VOL. 88-B, No. 12, DECEMBER 2006

3 1586 S. HOOK, E. MOULDER, P. J. YATES, B. J. BURSTON, E. WHITLEY, G. C. BANNISTER Fig. 4 Graph showing subsidence over 132 months. Fig. 3 Diagram of osteolysis in the zones of Gruen et al. 13 lucency at the interface between the bone and cement, grade C as radiolucency involving 50% to 99% of the cement-bone interface, or a defective or incomplete mantle of cement of any size, with metal against bone, and grade D as radiolucency involving 100% of the cementbone interface in any projection, or a failure to fill the canal with cement such that the distal tip of the prosthesis is not covered. The thickness of the cement mantle was also measured in all 14 zones of the femoral and the three zones of the acetabular components. 14 Alignment and subsidence were measured directly from calibrated sequential films. The radiological landmarks used for measuring subsidence were the greater trochanter, the proximolateral cement mantle and the prosthesis. Radiolucent lines at the cement-stem and cement-bone interface in the post-operative and ten-year radiographs were defined using the zones described by Gruen et al 13 and extended by Johnston et al 11 (Fig. 3). Proximomedial bone resorption was defined using the criteria described by Engh, Bobyn and Glassman. 15 In the first grade only the most proximal medial edge of the cut femoral neck is rounded off slightly, in the second there is rounding off of the proximal medial femoral neck combined with the loss of medial cortical density to the level of the lesser trochanter, in the third there is resorption of both the medial and the anterior cortex at the level of the lesser trochanter as well as the medial cortex below the lesser trochanter and in the fourth the resorption extends into the diaphysis. Cortical hypertrophy and fracture of the cement mantle were also measured. Ectopic ossification was defined using the criteria of Brooker et al. 16 In class I, islands of bone are present within the soft tissues about the hip, in class II, there are bone spurs from the pelvis or proximal end of the femur, leaving at least 1 cm between opposing bone surfaces, in class III, there is less than 1 cm between the bone spurs and in class IV, there is apparent bony ankylosis. Filling of the canal was measured at the mid-point of the lesser trochanter and divided into metal (implant), total cement and cancellous bone. Wear of the acetabular component was measured by the technique described by Latimer and Lachiewicz. 17 Statistical analysis. Kaplan-Meier survival curves were produced, with 95% confidence intervals. 18 A p-value of < 0.05 was considered to be statistically significant. Results Of the 123 patients, 49 (54 hips) (40%) had died at ten years, a typical mortality figure at ten years for these patients. 7,19 Only two patients (two hips) were unable to attend for radiological review, but the clinical outcome was obtained by telephone interview. The fate of every stem is known. Therefore, the 74 patients with 88 THRs who survived for ten or more years were reviewed. Their clinical details and pre-operative diagnoses are given in Table I. The four patients who underwent revision before ten years were also included in the review. Clinical assessment. The mean follow-up was 12.7 years (10 to 17). The mean Oxford hip score at follow-up was 24 (12 to 49), and increased with pain in the hip. Many with poor scores had impaired function as a result of other arthritic joints or medical conditions and found it difficult to complete the questionnaire accurately. Revision. The rate of revision of the femoral component for aseptic loosening was 1.1% (1 stem), and that for revision for any cause was 2.2% (2 stems). A total of 19 acetabular components (21.6%) were revised, and in the THE JOURNAL OF BONE AND JOINT SURGERY

4 THE EXETER UNIVERSAL STEM 1587 Table I. Details of the 88 hips (74 patients) surviving for more than ten years Hips Patients Number Gender (%) Male 26 (35) Female 48 (65) Mean age at surgery in yrs (range) 61 (18 to 84) Diagnosis (%) Osteoarthritis 55 (63) Inflammatory arthritis 5 (6) Developmental dysplasia of the hip 18 (20) Avascular necrosis 8 (9) Other 2 (2) Number lost to follow-up 0 0 Number with radiological follow-up Number with clinical follow-up Follow-up in yrs (range) 12.7 (10 to 17) Types of acetabular component * Harris-Galante I 33 Elite 23 Exeter 32 * Harris-Galante I, Zimmer Ltd, Swindon, United Kingdom; Elite, DePuy International Ltd, Leeds, United Kingdom; Exeter, Howmedica International Ltd, London, United Kingdom. process of gaining surgical access, five stems were removed and fresh prostheses implanted into the same cement mantle. Further, we intend to revise ten acetabular components in ten patients for migration and/or wear of the Exeter component. We also intend to revise one stem with excessive subsidence associated with pain in the thigh. A high proportion of Exeter acetabular components has been, or will be revised (37.5%; 12 of 32), along with 15% (5 of 33) of the Harris-Galante components and 4% (1 of 23) of the Elite components. One patient with deep infection underwent revision of both components. The deceased patients records were assessed for problems with their implants and, when possible, radiographs were analysed and subsidence measured. There were no revisions of the stem in these patients. Radiological assessment. The radiological assessment of cementing on post-operative radiographs in all 14 zones showed that Barrack 12 grade A was seen in 63 hips (72%), grade B in none, grade C in 21 hips (24%) and grade D in four hips (4%). Alignment, subsidence and lucent lines. All stems were inserted within 5 of neutral in the anteroposterior plane. Only five were in between 3 and 5 of varus. Three stems had a change in alignment of more than 3 over ten years, two moved into valgus and one into varus. These were associated with a primary diagnosis of atrophic osteoarthritis Table II. Proximomedial bone resorption according to the grading of Engh et al 15 in the 88 hips Engh grade Number (%) 0 23 (26) 1 23 (26) 2 29 (33) 3 12 (14) 4 1 (1) and all had subsided more than the average distance to 3 mm, 4 mm and 10 mm, respectively. No stems had measurable migration at the cement-bone interface. The mean vertical subsidence at the stem-cement interface was 1.52 mm (0 to 8.3) over 12.7 years. The mean subsidence was 0.12 mm (0 to 0.65) per year, slowing from 0.5 mm (0 to 0.9) in the first year to 0.03 mm (0 to 0.9) per year after four years (Fig. 4). Subsidence was significantly related to the presence of defects in the cement mantle (p = 0.01). In grade-a hips, the mean rate of subsidence was 1.24 mm (0 to 5.26) over ten years, whereas in those with defects (grade C or grade D) the mean subsidence was 2.06 mm (0.42 to 8.0) over ten years. Of the 88 stems followed for ten years, nine had radiolucent lines present at the cement-bone interface in one or more zones. Eight of these were progressive and one had stabilised by three years after operation. None of the lucent lines involved 50% or more of the cement-bone interface. Survivorship. Survivorship curves for revision for aseptic loosening, osteolysis and re-operation for any cause are shown in Figures 5 and 6. One further stem was considered to be a failure, but is yet to be revised. This was in a 41-year-old man with inflammatory arthritis whose stem had subsided by 8 mm over 12 years. Post-operatively, he had a defect of the cement mantle at the tip of the stem (Barrack grade D 12 ). Proximal bone preservation. The results of resorptive bone remodelling in the region of the proximomedial cortex are shown in Table II. Zero or first-degree resorption was seen in 46 hips (52%), second-degree in 29 hips (33%) and third-degree in 12 hips (14%). Only one hip showed fourth-degree resorption (1%). This was not significantly related to the size of the stem (p = 0.32) or the thickness of the mantle at the calcar (p = 0.65) Cortical hypertrophy and fractures of the cement mantle. Cortical hypertophy was apparent in seven stems. It occurred exclusively in Gruen zones 3, 5 and 6, but was not related to excessive subsidence, loosening, thickness of the mantle, Barrack grade, fracture of the mantle, age or poor clinical outcome. Fracture of the cement mantle was seen in five stems. This was associated with a mean excessive subsidence of 4.7 mm (2 to 8), defects of the mantle, male gender and poor bone quality (four had atrophic osteoarthritis, one avascular necrosis (AVN)). VOL. 88-B, No. 12, DECEMBER 2006

5 1588 S. HOOK, E. MOULDER, P. J. YATES, B. J. BURSTON, E. WHITLEY, G. C. BANNISTER Fig. 5 Survival curve with revision of the stem for osteolysis or loosening as the end-point. Fig. 6 Survival curve with revision of any component for any cause as the endpoint. Wear of the acetabular component and osteolysis. Wear of the acetabular component was significantly associated with osteolysis that was virtually entirely confined to the periarticular zones (Fig. 3). The mean wear in cases without significant lysis was 1.9 mm (0 to 6) and 3.6 mm (0.5 to 11) in the lytic group (t-test, p < 0.001). Other significant risk factors were young age (analysis of variance, p = 0.01) and male gender (t-test, p < 0.001). Significant localised endosteal bone lysis occurred in 22 hips (25%). Ectopic ossification. Grades of ectopic ossification were particularly low in comparison with previous studies Discussion This is the first series which has reported the results for the Exeter Universal stem with a minimum follow-up of more than ten years. Compared with the series reported from Exeter, 7 our follow-up was more complete, longer and more detailed. Our revision rate for the femoral component was 1.1% (1 stem) for osteolysis and loosening, but our follow-up was a mean of four years longer. The femoral component had a very low incidence of lucent lines at the cement-bone interface, in both our study and that from Exeter. 7 Thus our second-generation cementing technique with a longer follow-up was comparable with the thirdgeneration cementing techniques at Exeter. Both techniques resulted in fewer lucent lines than with the first-generation technique reported by Fowler et al 1 and the results of all three series were comparable with the best in the literature. The Exeter stem migrates within the cement mantle without disrupting the cement-bone interface. Subsidence increased with Barrack grading both in our series and that from Exeter, 7 indicating that cementing technique does have some influence on the behaviour of the stem. However, in this type of stem this increased subsidence is probably not important, indicating the forgiving nature of the stem to surgical technique, and supporting its applicability to general use. Three of the stems changed alignment significantly over the ten-year period. These were associated with a primary diagnosis of atrophic osteoarthritis and all subsided much more than average. Although the study from Exeter 7 failed to report this phenomenon, it has been described by Fowler et al 1 (18% of stems changed alignment), by Ling 2 and also by RSA studies. 24 This characteristic does not seem to occur with other polished tapered stems 25,26 even when there is significant distal migration. We feel that this phenomenon is due to the rounded shape of the implant at the shoulder (zone 1), which gives less resistance to the rotation of the stem around the calcar. We do not think that this change in alignment is an important problem in itself, but the associated excessive distal migration may be more of an issue, and suggestive of inadequate support of the construct. The rounded shape of the Exeter Universal stem also appears to reduce rotational stability when compared with squarer stems, 25,27 which may be important in long-term survival as the stems move into retroversion. However, the magnitude and timescale of these changes suggest that these modifications of alignment and rotation are unlikely to have any impact on the clinical performance and survival of the implant. Lucent lines at the cement-bone interface were present in nine hips, of which eight were progressive. They were not associated with poor cementing techniques or excessive subsidence, but six had occurred in what may be described as biologically and possibly structurally deficient bone (two AVN, four atrophic osteoarthritis). Other series have noted that both AVN and atrophic arthritis are risk factors for failure through loosening in both cemented and uncemented stems. 28,29 The small numbers of cases involved in our series make it difficult to draw further conclusions. The cement mantles in our series were extremely good. Using the most critical analytical method in all the 14 zones, 12 Barrack grade C and grade D were not associated THE JOURNAL OF BONE AND JOINT SURGERY

6 THE EXETER UNIVERSAL STEM 1589 with failure, unlike other types of stem, but osteolysis was significantly associated with thinner or absent cement at the greater trochanter (zone 1) (p = 0.04), but not reduced thickness in zone 7. Significant, localised endosteal bone lysis occurred in 22 of our hips (25%) and in only 0.5% in the Exeter series. 7 Most osteolysis in our series occurred in the peri-articular zones (Fig. 3) and was strongly associated with wear of polyethylene. On the basis of these findings we suggest that in the presence of wear particles, the Exeter stem reduces the likelihood of distal osteolysis by minimising access to the cement-bone interface with the polished surface and tapered stem. When there is inadequate cement at the greater trochanter, peri-articular osteolysis occurs frequently. Complete cement mantles prevent access of fluid and debris 30 and hence late loosening at the cement-bone interface caused by migratory debris Incomplete cement mantles can occur from oversizing of the stems. Chiu et al 34 from their experience with the Exeter stem in Chinese patients with small femora, showed that there was early loosening in a population in which oversizing of the stem was common, with a resultant incomplete cement mantle and high rates of failure. These incomplete mantles can be avoided by downsizing the implant from the last broach used as long as there are adequate smaller sizes available to allow this. Scherrlinck et al 35 confirmed that cement mantles were less likely to be deficient when the stems were downsized from the broach, although they felt that support for the larger stems was good because of excellent penetration of the cancellous bone and the more secure support afforded by the cortical bone. We have previously shown that downsizing actually reduces subsidence of the stem with polished tapers, 27 and now downsize most of our stems. However, we have also shown that adequate mantles can routinely be achieved with larger stems as long as care is taken to remove enough cancellous bone and to align the stem properly. 36 There was a low rate of fracture of the cement mantle (five cases) and cortical hypertrophy (seven cases) in our series in keeping with others using polished tapered stems. The hypertrophy was always minor and was not statistically significantly related to adverse outcome, poor mantles or excessive subsidence. Even in the presence of excessive load or poor mantles, this type of stem optimises bone loading through the even double taper and debonding. Hence, signs of over- and underload are unusual. The fractures on the other hand appeared to be related to poor supporting bone and poor mantles, leading to excessive subsidence. Nevertheless, even in this undesirable circumstance, the stems did not fail clinically. The access for wear debris which fractures of the mantle normally allow, was sealed off by the wedging effect of the polished taper, and distal osteolysis did not occur. There was a noticeable number of hips with resorption of the medial calcar. This contrasts with the original series, but with a different system, in which Fowler et al 1 noted that preservation of the cut proximal femoral neck appeared to be uniquely satisfactory. In the subsequent series with the Universal stem from Exeter, there was more calcar resorption, but it was still minimal. 7 These findings are likely to be related to the increased rigidity of the Universal stem compared with the original design used in the series of Fowler et al. 1 A narrower implant with a larger proximal cement mantle is logical and likely to reduce proximal stress. This did not reach statistical significance in this study, but did so in our other series using the similar collarless polished tapered stems (p = 0.02). 27 The use of the Exeter cemented metal-backed acetabular component was unsatisfactory with 34% revised so far, and most of the others radiologically loose. Only one of the cemented Elite stems was revised for loosening, suggesting that the critical variable was the design of the component or the instrumentation and not the technique of cementing. This finding is consistent with that of other series using cemented metal-backed components reported in the literature. 37 However, we are aware that the results of the Exeter metal-backed acetabular component were satisfactory in Exeter, although these findings have never specifically been reported in the literature. The cement mantles of non-flanged metal-backed cemented acetabular components have failed in our hands. The liners of uncemented components (Harris- Galante I) wear, but flanged cemented components without metal backing are the most reliable devices over the medium to long term. One advantage of the use of the Exeter and other polished tapered stems with failure of the acetabular component is the ease of revision. If the component needs to be exchanged, then access can be dramatically improved by taking out the stem and then replacing the same stem, or a new one, into the original mantle. The principal benefit of hip replacement is relief from pain. The Oxford hip score, like all composite hip scores, is biased by co-morbidity in the ageing population. Our mean follow-up was four years longer than that of the Exeter study. 7 The clinical results were comparable, with respect to the Oxford hip score and rate of revision of the stem. Our mean Oxford hip score at follow-up was 24 (12 to 49) and in the series from Exeter it was (12 to 48). 7 The Exeter Universal stem is performing at least as well as any published cemented implant, but unlike the results for the Charnley stem over ten years, the Universal stem is identical to the current Exeter stem within the cement mantle. 3 It is relatively forgiving of the surgical technique and the results of polished tapered stems are comparable with those of first-, second- and third-generation cementing techniques. This suggests that the cemented collarless polished tapered stem is suitable for general use and represents a reference standard against which other femoral components should be compared. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. VOL. 88-B, No. 12, DECEMBER 2006

7 1590 S. HOOK, E. MOULDER, P. J. YATES, B. J. BURSTON, E. WHITLEY, G. C. BANNISTER References 1. Fowler JL, Gie GA, Lee AJ, Ling RS. Experience with the Exeter total hip replacement since Orthop Clin North Am 1988;19: Ling RS. The history and development of the Exeter hip. Stryker, UK, Malchau H, Herberts P, Garellick G, Soderman P, Eisler T. Prognosis of total hip replacements. Procs 69th American Academy of Orthopedic Surgeons annual meeting, Middleton RG, Howie DW, Costi K, Sharpe P. Effects of design changes on cemented tapered femoral stem fixation. Clin Orthop 1998;355: Howie DW, Middleton RG, Costi K. Loosening of matt and polished cemented femoral stems. J Bone Joint Surg [Br] 1998;80-B: Rockborn P, Olsson SS. Loosening and bone resorption in Exeter hip arthroplasties: review at a minimum of five years. J Bone Joint Surg [Br] 1993;75-B: Williams HD, Browne G, Gie G, et al. The Exeter universal cemented femoral component at 8 to 12 years: a study of the first 325 hips. J Bone Joint Surg [Br] 2002;84-B: Chiu KH, Shen WY, Cheung KW, Tsui HF. Primary Exeter total hip arthroplasty in patients with small femurs: a minimal of 10 years follow-up. J Arthroplasty 2005;20: Franklin J, Robertsson O, Gestsson J, Lohmander LS, Ingvarsson T. Revision and complication rates in 654 Exeter total hip replacements, with a maximum follow-up of 20 years. BMC Musculoskeletal Disord 2003;4: Dawson J, Fitzpatrick R, Carr A, Murray D. Questionnaire on the perceptions of patients about total hip replacement. J Bone Joint Surg [Br] 1996;78-B: Johnston RC, Fitzgerald RH Jr, Harris WM, et al. Clinical and radiological evaluation of total hip replacement: a standard system of terminology reporting results. J Bone Joint Surg [Am] 1990;72-A: Barrack RL, Mulroy RD, Jr, Harris WH. Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty: a 12-year radiographic review. J Bone Joint Surg [Br] 1992;74-B: Gruen TA, McNeice GM, Amstutz HC. Modes of failure of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop 1979;141: DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop 1976;121: Engh CA, Bobyn JJ, Glassman AH. Porous-coated hip replacement: the factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg [Br] 1987;69- B: Brooker AF, Bowerman JW, Robinson RA, Riley LH Jr. Ectopic ossification following total hip replacement: incidence and a method of classification. J Bone Joint Surg [Am] 1973;55-A: Latimer HA, Lachiewicz PF. Porous-coated acetabular components with screw fixation: five to ten-year results. J Bone Joint Surg [Am] 1996;78-A: Dobbs HS. Survivorship of total hip replacements. J Bone Joint Surg [Br] 1980;62-B: Issack P, Botero HG, Hiebert RN, et al. Sixteen-year follow-up of the cemented spectron femoral stem for hip arthroplasty. J Arthroplasty 2003;18: Madey SM, Callaghan JJ, Olejniczak JP, Goetz DD, Johnston RC. Charnley total hip arthroplasty with use of improved techniques of cementing: the results after a minimum of fifteen years follow-up.j Bone Joint Surg [Am] 1997;79-A: Weiderhielm LRA, Mikhail WE, Nelissen RG, Bauer TW. Cemented collarless (Exeter-CPT) versus cementless collarless (PCA) femoral components: a 2- to 14-year follow-up evaluation. J Arthroplasty 1995;10: Oishi CS, Walker RH, Colwell CW Jr. The femoral component in total hip arthroplasty: six to eight-year follow-up of one hundred consecutive patients after use of a third-generation cementing technique. J Bone Joint Surg [Am] 1994;76-A: Clohisy JC, Harris WH. Primary hybrid total hip replacement, performed with insertion of the acetabular component without cement and a precoat femoral component with cement. J Bone Joint Surg [Am] 1999;81-A: Alfaro-Adrian J, Gill HS, Murray DW. Should total hip arthroplasty femoral components be designed to subside?: a radiostereometric analysis study of the Charnley Elite and Exeter stems. J Arthroplasty 2001;16: Glyn-Jones S, Gill HS, Beard DJ, McLardy-Smith P, Murray DW. Influence of stem geometry on the stability of polished tapered cemented femoral stems. J Bone Joint Surg [Br] 2005;87-B: Yates P, Gobel D, Bannister G. Collarless polished tapered stem: clinical and radiological follow-up over 5 years. J Arthroplasty 2002;17: Yates P, Burston B, Bannister G. Collarless Polished Tapered stem (CPT): clinical and radiological follow-up at a minimum of 10 years. J Bone Joint Surg [Am] 2006: in press. 28. Kobayashi A, Donnelly W, Scott G, Freeman MA. Early radiological observations may predict the long-term survival of femoral hip prostheses. J Bone Joint Surg [Br] 1997;79-A: Clarke HJ, Jinnah RH, Brooker AF, Michaelson JD. Total replacement of the hip for avascular necrosis in sickle cell disease. J Bone Joint Surg [Br] 1989;71- B: Crawford RW, Evans M, Ling RS, Murray DW. Fluid flow around model femoral components of differing surface finishes: in vitro investigations. Acta Orthop Scand 1999;70: Breusch SJ, Lukoschek M, Kreutzer J, Brocai D, Gruen TA. Dependency of cement mantle thickness on femoral stem design and centralizer. J Arthroplasty 2001;16: James SP, Schmalzried TP, McGarry J, Harris WH. Extensive porostity at the cement-femoral-prosthesis interface: a preliminary study. J Biomed Mater Res 1993;27: Jasty M, Maloney WJ, Bragdon CR, et al. The initiation of failure in cemented femoral components of hip arthroplasties. J Bone Joint Surg [Br] 1991;73-B: Chiu KH, Shen WY, Tsui HF, Chan KM. Experience with primary exeter total hip arthroplasty in patients with small femurs: review at average follow-up period of 6 years. J Arthroplasty 1997;12: Scheerlinck T, de May J, Deklerck R, Noble PC. CT analysis of defects of the cement mantle and alignment of the stem: in vitro comparison of Charnley-Kerboul femoral hip implants inserted line-to-line and undersized in paired femora. J Bone Joint Surg [Br] 2006;88-B: Yates PJ, Quraishi NA, Kop A, et al. Fractures of modern high nitrogen stainless steel cemented stems: cause, mechanism and avoidance in 14 cases. J Arthroplasty 2006: in press. 37. Ritter MA, Thong AE. The role of cemented sockets in 2004: is there one? J Arthroplasty 2004;19(Suppl 1):92-4. THE JOURNAL OF BONE AND JOINT SURGERY

Collarless polished tapered stem

Collarless polished tapered stem Collarless polished tapered stem CLINICAL AND RADIOLOGICAL RESULTS AT A MINIMUM OF TEN YEARS FOLLOW-UP P. J. Yates, B. J. Burston, E. Whitley, G. C. Bannister From the Avon Orthopaedic Centre, Bristol,

More information

Initial Evaluation on Subsidence of Cemented Collarless Polished Tapered Stem Applied to the Patients with Narrow Femoral Medullar Canal

Initial Evaluation on Subsidence of Cemented Collarless Polished Tapered Stem Applied to the Patients with Narrow Femoral Medullar Canal The Open Orthopaedics Journal, 2010, 4, 147-151 147 Open Access Initial Evaluation on Subsidence of Cemented Collarless Polished Tapered Stem Applied to the Patients with Narrow Femoral Medullar Canal

More information

Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists

Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists : 31 35 doi 10.1308/1478708051450 Audit Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists Jac Ciampolini, Matthew JW Hubble Princess Elizabeth Orthopaedic Centre,

More information

Optimum implant geometry

Optimum implant geometry Surgical Technique Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

CAUTION: Ceramic liners are not approved for use in the United States.

CAUTION: Ceramic liners are not approved for use in the United States. Total Hip Prostheses, Self-Centering Hip Prostheses and Hemi-Hip Prostheses IMPORTANT: This essential product information sheet does not include all of the information necessary for selection and use of

More information

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS INTRODUCTION The Summit Tapered Hip System s comprehensive set of implants and instruments

More information

The proven, simple solution. CPT 12/14 Hip System

The proven, simple solution. CPT 12/14 Hip System The proven, simple solution CPT 12/14 Hip System Primary CPT Hip The proven, simple solution 5mm The collarless, polished, doubletaper design concept used in the CPT 12/14 Hip System has proven itself

More information

Ten-year survival of the cemented MS-30 femoral stem Increased revision rate in male patients

Ten-year survival of the cemented MS-30 femoral stem Increased revision rate in male patients Acta Orthop. Belg., 9, 5, 6-5 ORIGINAL STUDY Ten-year survival of the cemented MS-3 femoral stem Increased revision rate in male patients Daniela WITTE, Michael KLIMM, Dominik Parsch, Michael CLARIUS,

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Continuing the Tradition. VerSys Heritage Hip System

Continuing the Tradition. VerSys Heritage Hip System Continuing the Tradition VerSys Heritage Hip System Heritage Following the Tradition The low-friction hip prosthesis developed by Sir John Charnley has more than a 20-year history of outstanding results.

More information

Ten-year survival of the MS-30 matt-surfaced cemented stem

Ten-year survival of the MS-30 matt-surfaced cemented stem Ten-year survival of the MS-30 matt-surfaced cemented stem B. J. Berli, D. Schäfer, E. W. Morscher From the Orthopaedic Department, University of Basel, Basel, Switzerland The Morscher-Spotorno (MS-30)

More information

Long-term Outcome of Polished Stems in Total Hip Arthroplasty

Long-term Outcome of Polished Stems in Total Hip Arthroplasty ORIGINAL ARTICLE Hip Pelvis 27(2): 83-89, 2015 http://dx.doi.org/10.5371/hp.2015.27.2.83 Print ISSN 2287-3260 Online ISSN 2287-3279 Long-term Outcome of Polished Stems in Total Hip Arthroplasty Jin-Young

More information

Twenty-five-Year Results After Charnley Total Hip Arthroplasty in Patients Less than Fifty Years Old A CONCISE FOLLOW-UP OF A PREVIOUS REPORT*

Twenty-five-Year Results After Charnley Total Hip Arthroplasty in Patients Less than Fifty Years Old A CONCISE FOLLOW-UP OF A PREVIOUS REPORT* 1066 COPYRIGHT 2003 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED Twenty-five-Year Results After Charnley Total Hip Arthroplasty in Patients Less than Fifty Years Old A CONCISE FOLLOW-UP OF A

More information

Featuring. Technology. Product Rationale

Featuring. Technology. Product Rationale Featuring Technology Product Rationale 2 Optimum implant geometry Extending proven TRI-LOCK heritage The original TRI-LOCK Stem was introduced in 1981. This implant was the first proximally coated tapered-wedge

More information

Bone Preservation Stem

Bone Preservation Stem TRI-LOCK Bone Preservation Stem Featuring GRIPTION Coating Surgical Technique Implant Geometry Extending the TRI-LOCK Stem heritage The original TRI-LOCK Stem was introduced in 1981. This implant was

More information

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5.

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5. Surgical Technique Contents Templating and Pre Operative Planning 2 Preparation of the Acetabulum 4 Trial Sizing and Impaction of the Shell 5 Cup Positioning 6 Joint Stability 7 Trial sizing and Impaction

More information

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing Journal of Orthopaedic Surgery 2001, 9(1): 45 50 Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing KY Chiu, TP Ng, WM Tang and P Lam Department of Orthopaedic Surgery, The University

More information

The shape and size of femoral components in revision total hip arthroplasty among Chinese patients

The shape and size of femoral components in revision total hip arthroplasty among Chinese patients Journal of Orthopaedic Surgery 2003: 11(1): 53 58 The shape and size of femoral components in revision total hip arthroplasty among Chinese patients KY Chiu, TP Ng, WM Tang Department of Orthopaedic Surgery,

More information

TaperFit. Cemented Total Hip Replacement Surgical technique

TaperFit. Cemented Total Hip Replacement Surgical technique TaperFit Cemented Total Hip Replacement Surgical technique TaperFit Contents Operative summary 4 Pre-operative templating 5 Surgical exposure 5 Femoral neck resection 5 Acetabular preparation 5 Cenator

More information

28 Surgical Technique

28 Surgical Technique Surgical Technique 10 12 14 16 18 20 22 24 28 26 Technique described by James L. Guyton, MD Campbell Clinic Memphis, Tennessee James W. Harkess, MD Campbell Clinic Memphis, Tennessee David G. LaVelle,

More information

DESIGN RATIONALE AND SURGICAL TECHNIQUE

DESIGN RATIONALE AND SURGICAL TECHNIQUE DESIGN RATIONALE AND SURGICAL TECHNIQUE ANCHOR PEG GLENOID DESIGN RATIONALE In total shoulder arthroplasty, most cases of clinical and radiographic loosening involve failure of the fixation of the glenoid

More information

Optimizing function Maximizing survivorship Accelerating recovery

Optimizing function Maximizing survivorship Accelerating recovery Surgical Technique Optimizing Function Maximizing Survivorship Accelerating Recovery The company believes in an approach to patient treatment that places equal importance on: Optimizing function Maximizing

More information

21st Century Fracture Management ETS. Surgical Protocol

21st Century Fracture Management ETS. Surgical Protocol 21st Century Fracture Management ETS Surgical Protocol ETS Operative Technique Step 1 Confirm that a cemented hemiarthroplasty is indicated. An X-ray template of the ETS is provided. This should be used

More information

Bone Bangalore

Bone Bangalore Dr Suresh Annamalai MBBS, MRCS(Edn), FRCS( Tr & Orth)(Edn), FEBOT(European Board), Young Hip and Knee Fellowship(Harrogate, UK) Consultant Arthroplasty and Arthroscopic Surgeon Manipal Hospital, Whitefield,

More information

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY E C H E L O N P R I M A R Y H I P S Y S T E M P R O D U C T R A T I O N A L E ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY Echelon Primary Total Hip System HIGH OFFSET STANDARD OFFSET Cementless

More information

Absolut TM Cemented Stem. Surgical Technique

Absolut TM Cemented Stem. Surgical Technique Absolut TM Cemented Stem Surgical Technique Contents ABSOLUT Cemented Stem 2 Absolut Confidence 2 Absolut Reproducibility 2 Absolut Choice 2 Pre-Operative Planning 3 Suggested Templating Method 3 Surgical

More information

Charnley Total Hip Arthroplasty with Use of Improved Cementing Techniques A MINIMUM TWENTY-YEAR FOLLOW-UP STUDY

Charnley Total Hip Arthroplasty with Use of Improved Cementing Techniques A MINIMUM TWENTY-YEAR FOLLOW-UP STUDY 1840 COPYRIGHT 2001 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED Charnley Total Hip Arthroplasty with Use of Improved Cementing Techniques A MINIMUM TWENTY-YEAR FOLLOW-UP STUDY BY AIMEE S. KLAPACH,

More information

IMPROVED CEMENTING TECHNIQUES AND FEMORAL COMPONENT LOOSENING IN YOUNG PATIENTS WITH HIP ARTHROPLASTY

IMPROVED CEMENTING TECHNIQUES AND FEMORAL COMPONENT LOOSENING IN YOUNG PATIENTS WITH HIP ARTHROPLASTY IMPROVED CEMENTING TECHNIQUES AND FEMORAL COMPONENT LOOSENING IN YOUNG PATIENTS WITH HIP ARTHROPLASTY A 12-YEAR RADIOGRAPHIC REVIEW ROBERT L. BARRACK, RICHARD D. MULROY JR, WILLIAM H. HARRIS From the Massachusetts

More information

Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem

Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem Original Article Clinics in Orthopedic Surgery 2009;1:105-109 doi:10.4055/cios.2009.1.2.105 Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem Kyoung Ho Moon, MD, Joon Soon

More information

PRODUCT RATIONALE & SURGICAL TECHNIQUE

PRODUCT RATIONALE & SURGICAL TECHNIQUE This publication is not intended for distribution in the USA. PRODUCT RATIONALE & SURGICAL TECHNIQUE THE PRODUCT OF LONG-TERM CLINICAL EXPERIENCE The TRILOC cemented UHMWPE cup is a direct descendant of

More information

Tri-Lock Bone Preservation Stem

Tri-Lock Bone Preservation Stem Tri-Lock Bone Preservation Stem Clinical Results of the Tri-lock BPS Femoral Stem at One Year Follow-Up William L. Healy, M.D. Lahey Clinic Medical Center Burlington, Massachusetts Introduction Cementless

More information

The Basis CL cemented femoral stem: results after 8.9 years follow-up

The Basis CL cemented femoral stem: results after 8.9 years follow-up Hip Int 2013; 23 ( 2) : 147-153 DOI: 10.5301/HIP.2013.10877 Original Article The Basis CL cemented femoral stem: results after 8.9 years follow-up Sabine Mai, Matthias Golla, Werner E. Siebert Vitos Orthopaedic

More information

AML Hip System. Design Rationale/ Surgical Technique

AML Hip System. Design Rationale/ Surgical Technique AML Hip System Design Rationale/ Surgical Technique Design Rationale Evolution In 1977, DePuy Synthes Companies introduced the original cementless total hip. The AML Hip launched in order to solve one

More information

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter Design Philosophy Cementless Stems of 12/14 Taper Stem Length Neck shaft Angle STEM-N3 STEM-N4 STEM-N5 STEM-N6 STEM-N7 STEM-N8 STEM-N9 123 mm 125 mm 130 mm 135 mm 140 mm 145 mm 150 mm 132 Cement Polished

More information

This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE DePuy Synthes DURALOC Surgical Technique CONTENTS Templating and Pre Operative Planning 2 Preparation of the Acetabulum

More information

AGGRESSIVE GRANULOMATOUS LESIONS AFTER HIP ARTHROPLASTY

AGGRESSIVE GRANULOMATOUS LESIONS AFTER HIP ARTHROPLASTY AGGRESSIVE GRANULOMATOUS LESIONS AFTER HIP ARTHROPLASTY KAJ TALLROTH, ANTTI ESKOLA, SEPPO SANTAVIRTA, YRJ T. KONTTINEN, T. SAM LINDHOLM From the Orthopaedic Hospitalofthe Invalid Foundation, Helsinki We

More information

Revision hip arthroplasty with S-ROM prosthesis: a study of clinical outcomes and implant stability

Revision hip arthroplasty with S-ROM prosthesis: a study of clinical outcomes and implant stability J Orthopaed Traumatol () 7:1 1 DOI 1.17/s1195--15- ORIGINAL M. El-Deen S. Zahid D.T. Miller A. Nargol R. Logishetty Revision hip arthroplasty with S-ROM prosthesis: a study of clinical outcomes and implant

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending the proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

5A* ODEP. Conserving bone. Preserving soft tissue. Restoring biomechanics. rating

5A* ODEP. Conserving bone. Preserving soft tissue. Restoring biomechanics. rating MiniHip is designed to give patients the advantages of a traditional hip replacement whilst also preserving the natural anatomy. MiniHip potentially allows for future revisions and so provides a pre-primary

More information

ORIGINAL PAPER. Department of Orthopedic Surgery, Hamamatsu Medical Center ABSTRACT

ORIGINAL PAPER. Department of Orthopedic Surgery, Hamamatsu Medical Center ABSTRACT Nagoya J. Med. Sci. 71. 145 ~ 150, 2009 ORIGINAL PAPER AUTOGENOUS BULK STRUCTURAL BONE GRAFTING FOR RECONSTRUCTION OF THE ACETABLUM IN PRIMARY TOTAL HIP ARTHROPLASTY: AVERAGE 12-YEAR FOLLOW-UP TETSUO MASUI,

More information

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA 94065 info@stinsonortho.com www.stinsonortho.com Table of Contents Introduction 3 Features 4 Surgical Technique 5 Preoperative

More information

Femoral Revision in Hip Resurfacing Compared With Large-Bearing Metal-on-Metal Hip Arthroplasty

Femoral Revision in Hip Resurfacing Compared With Large-Bearing Metal-on-Metal Hip Arthroplasty The Journal of Arthroplasty Vol. 26 No. 8 2011 Femoral Revision in Hip Resurfacing Compared With Large-Bearing Metal-on-Metal Hip Arthroplasty Simon J.W. Garrett, BM, MRCS, FRCS(Tr&Orth),*y Ben J.R.F.

More information

Surgical Technique. Hip System

Surgical Technique. Hip System Surgical Technique Hip System INDICATIONS FOR USE The TaperSet Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System.

More information

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Diagnosis: Ceramic head fracture In the 1970 s, Boutin implemented ceramic in modern total hip arthroplasty (THA).

More information

Results of Conversion Total Hip Prosthesis Performed Following Painful Hemiarthroplasty

Results of Conversion Total Hip Prosthesis Performed Following Painful Hemiarthroplasty M Nomura, S The Journal et al. of International Medical Research Endovascular 2000; 28: Embolization 307 312 of Unruptured Results of Conversion Total Hip Prosthesis Performed Following Painful Hemiarthroplasty

More information

EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION

EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION SINCE 1983 PREOPERATIVE PLANNING EXPLANTATION OPTIONS the cement from inside the cement canal until the bone/ cement bond

More information

Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year

Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year H. Pandit, C. Jenkins, D. J. Beard, J. Gallagher, A. J. Price, C. A. F. Dodd, J. W. Goodfellow, D. W. Murray From

More information

Revision of the cemented femoral stem using a cement-in-cement technique

Revision of the cemented femoral stem using a cement-in-cement technique Revision of the cemented femoral stem using a cement-in-cement technique A FIVE- TO 15-YEAR REVIEW W. W. Duncan, M. J. W. Hubble, J. R. Howell, S. L. Whitehouse, A. J. Timperley, G. A. Gie From Princess

More information

Femoral stem loosening rate of less than 0.16% per year at years 1. Exeter Hip

Femoral stem loosening rate of less than 0.16% per year at years 1. Exeter Hip Exeter Clinical Results Femoral stem loosening rate of less than 0.16% per year at 20 25 years 1 Exeter Hip Survival of stem 2 CUMULATIVE ESTIMATE OF SURVIVAL (%) AT 18YRS - TAKEN FROM 20-25 YR RESULTS

More information

Fixation and loosening of the cemented Müller straight stem

Fixation and loosening of the cemented Müller straight stem Fixation and loosening of the cemented Müller straight stem A LONG-TERM CLINICAL AND RADIOLOGICAL REVIEW M. Clauss, M. Luem, P. E. Ochsner, T. Ilchmann From Kantonsspital Liestal, Liestal, Switzerland

More information

A further enhanced classic. Wagner SL Revision Hip Stem

A further enhanced classic. Wagner SL Revision Hip Stem A further enhanced classic Wagner SL Revision Hip Stem The original Wagner SL Revision Stem offers a time-proven solution in the treatment of revision hips. While its underlying anchorage philosophy and

More information

SUMMIT and DURALOC. Clinical Summary

SUMMIT and DURALOC. Clinical Summary SUMMIT and DURALOC Clinical Summary Ten-year results of a press-fit, porous-coated acetabular component Grobler G.P. Learmonth I.D. Bernstein B.P. Dower B.J. (2005) The Journal of Bone and Joint Surgery;

More information

Hedrocel trabecular metal monoblock acetabular cups : mid-term results

Hedrocel trabecular metal monoblock acetabular cups : mid-term results Acta Orthop. Belg., 2006, 72, 326-331 ORIGINAL STUDY Hedrocel trabecular metal monoblock acetabular cups : mid-term results Michiel MULIER, Bart RYS, Lieven MOKE From the University Hospital Pellenberg,

More information

Bone Conserving Hip Replacement Surgical technique

Bone Conserving Hip Replacement Surgical technique MiniHip Bone Conserving Hip Replacement Surgical technique MiniHip Contents Operative summary 4 Overview 5 Pre-operative templating 6 Operative technique 7 Neck resection 7 Femoral canal preparation 7

More information

PLR. Proximal Loading Revision Hip System

PLR. Proximal Loading Revision Hip System PLR Proximal Loading Revision Hip System The PLR splined revision stem is designed to recreate the natural stresses in the revised femur, where proximal bone may be compromised. PLR Hip System Design Considerations

More information

Progeny Hip Stem. Surgical Protocol and Product Specifications

Progeny Hip Stem. Surgical Protocol and Product Specifications Progeny Hip Stem Surgical Protocol and Product Specifications Progeny Hip Stem Introduction With emphasis on maximum stability and ease of use, the StelKast ProgenyTM Hip System provides the surgeon with

More information

Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination.

Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination. ISPUB.COM The Internet Journal of Orthopedic Surgery Volume 17 Number 1 Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination. Roberts, Garlick

More information

Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision

Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision Clin Orthop Relat Res (2009) 467:1501 1506 DOI 10.1007/s11999-008-0499-6 ORIGINAL ARTICLE Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision Byung-Woo Min MD, Kwang-Soon Song

More information

MINIMUM TEN-YEAR FOLLOW-UP OF COMPUTED TOMOGRAPHY BASED, CUSTOM CEMENTLESS STEM AFTER INTERTROCHANTERIC OSTEOTOMY FOR DYSPLASTIC HIPS

MINIMUM TEN-YEAR FOLLOW-UP OF COMPUTED TOMOGRAPHY BASED, CUSTOM CEMENTLESS STEM AFTER INTERTROCHANTERIC OSTEOTOMY FOR DYSPLASTIC HIPS (65) MINIMUM TEN-YEAR FOLLOW-UP OF COMPUTED TOMOGRAPHY BASED, CUSTOM CEMENTLESS STEM AFTER INTERTROCHANTERIC OSTEOTOMY FOR DYSPLASTIC HIPS YOSHIHIDE MASUDA Department of Orthopaedic Surgery, Nara Prefecture

More information

The science of simplicity

The science of simplicity 1 Design Rationale The science of simplicity The advanced features of the Corail stem, and its bone-preserving surgical technique, have made it a great choice for minimally invasive hip surgery. More than

More information

This publication is not intended for distribution in the USA.

This publication is not intended for distribution in the USA. This publication is not intended for distribution in the USA. Extraction of a CORAIL Stem Surgical Technique Introduction Given the excellent long-term results of the CORAIL stem, 1,2,3 its extraction

More information

Surgical Technique r5.indd 1 12/8/10 10:36 AM

Surgical Technique r5.indd 1 12/8/10 10:36 AM Surgical Technique The science of simplicity With more than 700,000 implantations and two and a half decades of clinical success, the Corail Total Hip System now has the most extensive experience with

More information

LIMITED INFLUENCE OF PROSTHETIC POSITION ON ASEPTIC LOOSENING OF THE SOUTER-STRATHCLYDE TOTAL ELBOW PROSTHESIS

LIMITED INFLUENCE OF PROSTHETIC POSITION ON ASEPTIC LOOSENING OF THE SOUTER-STRATHCLYDE TOTAL ELBOW PROSTHESIS 6 LIMITED INFLUENCE OF PROSTHETIC POSITION ON ASEPTIC LOOSENING OF THE SOUTER-STRATHCLYDE TOTAL ELBOW PROSTHESIS J.C.T. van der Lugt R.B. Geskus P.M. Rozing Acta Orthop. 2005;76(5):654-61 Chapter 6 Abstract

More information

simplicity THE SCIENCE OF DESIGN RATIONALE

simplicity THE SCIENCE OF DESIGN RATIONALE simplicity THE SCIENCE OF DESIGN RATIONALE T H E S C I E N C E O F S I M P L I C I T Y The advanced features of the Corail stem, and its bone-preserving surgical technique, have made it a great choice

More information

Design Rationale. ECHELON Primary Hip System

Design Rationale. ECHELON Primary Hip System Design Rationale ECHELON Primary Hip System ECHELON Primary Total Hip System Addressing clinical issues of cementless hip arthroplasty Cementless total hip arthroplasty has provided a proven method of

More information

The Norwegian Arthroplasty Register: 11 years and 73,000 arthroplasties

The Norwegian Arthroplasty Register: 11 years and 73,000 arthroplasties Acta Orthopaedica Scandinavica ISSN: 1-647 (Print) (Online) Journal homepage: http://www.tandfonline.com/loi/iort19 The Norwegian Arthroplasty Register: 11 years and 73, arthroplasties Leif I Havelin,

More information

AUTOBLOQUANTE AUTOBLOQUANTE. Product Rationale Surgical Technique

AUTOBLOQUANTE AUTOBLOQUANTE. Product Rationale Surgical Technique AUTOBLOQUANTE AUTOBLOQUANTE Product Rationale Surgical Technique AUTOBLOQUANTE The Product of Long-Term Clinical Experience The AUTOBLOQUANTE femoral component is a direct descendant of the original straight

More information

TaperFit. Cemented Femoral Stem Product rationale

TaperFit. Cemented Femoral Stem Product rationale TaperFit Cemented Femoral Stem Product rationale TaperFit Design History Function TaperFit cemented femoral stems have been in use for over 17 years and are clinically proven to produce excellent results

More information

Revision. Hip Stem. Surgical Protocol

Revision. Hip Stem. Surgical Protocol U2 TM Revision Hip Stem Surgical Protocol U2 Revision Hip Stem Table of Contents Introduction... 1 Preoperative Planning... 2 Femoral Preparation... 3 Trial Reduction... 5 Implant Insertion... 6 Ordering

More information

CPCS. Training Manual. Collarless Polished Cemented Stem

CPCS. Training Manual. Collarless Polished Cemented Stem SMITH & NEPHEW ORTHOPAEDICS ~ TRUST. INNOVATION. PERFORMANCE. CPCS Collarless Polished Cemented Stem Training Manual This document is for internal use by Smith & Nephew and sales force only and is not

More information

REVISING THE DEFICIENT PROXIMAL FEMUR

REVISING THE DEFICIENT PROXIMAL FEMUR REVISING THE DEFICIENT PROXIMAL FEMUR by David Mattingly, Boston, MA Joseph McCarthy, Boston, MA Benjamin E. Bierbaum, Boston, MA Hugh P. Chandler, Boston, MA Roderick H. Turner, Boston, MA Hugh U. Cameron,

More information

111. ACETABULAR FIXATION

111. ACETABULAR FIXATION CLINICAL ORTHOPAEDICS AND RELATED RESEARCH Number 369, pp. 157-164 0 1999 Lippincott Williams & Wilkins, Inc. 111. ACETABULAR FIXATION IN PRIMARY TOTAL HIP ARTHROPLASTY Fixation, Polyethylene Wear, and

More information

S U R G I C A L T E C H N I Q U E

S U R G I C A L T E C H N I Q U E SURGICAL TECHNIQUE RECOVERY FUNCTION SURVIVORSHIP DePuy believes in an approach to total hip replacement that places equal importance on recovery, function and survivorship. The DePuy PROXIMA Hip System

More information

PINNACLE Acetabular Cup System

PINNACLE Acetabular Cup System PINNACLE Acetabular Cup System Clinical Summary A Prospective, Randomized Study of Cross-Linked and Non-Cross-Linked Polyethylene for Total Hip Arthroplasty at 10-Year Follow-Up Engh CA Jr., Hopper RH

More information

A radiographic comparison of femoral offset after cemented and cementless total hip arthroplasty

A radiographic comparison of femoral offset after cemented and cementless total hip arthroplasty Hip Int 2014; 00 ( 00 ): 000-000 DOI: 10.5301/hipint.5000160 ORIGINAL ARTICLE A radiographic comparison of femoral offset after cemented and cementless total hip arthroplasty James R. Berstock, Adrian

More information

Cementless Tapered Femoral Stem Surgical technique

Cementless Tapered Femoral Stem Surgical technique Cementless Tapered Femoral Stem Surgical technique Contents Operative summary 4 Pre-operative planning 5 Femoral neck osteotomy 5 Femoral canal preparation 5 Intra-medullary (IM) reamer 6 Sequential rasping

More information

SURGICAL TECHNIQUE. Alpine Cementless Hip Stem

SURGICAL TECHNIQUE. Alpine Cementless Hip Stem SURGICAL TECHNIQUE Alpine Cementless Hip Stem The following technique is a general guide for the instrumentation of the Alpine Cementless Hip Stem. It is expected that the surgeon is already familiar with

More information

HELIOS h i p s y s t e m

HELIOS h i p s y s t e m HELIOS h i p s y s t e m Design The Helios stem is a highly polished, High Nitrogen Stainless Steel (ISO5832-9) dual tapered cemented stem. The design of the stem is based on the clinically lly successful

More information

Two Years Outcomes Of Total Hip Replacement Using A Short Stem With Femoral Neck Anchoring

Two Years Outcomes Of Total Hip Replacement Using A Short Stem With Femoral Neck Anchoring Two Years Outcomes Of Total Hip Replacement Using A Short Stem With Femoral Neck Anchoring Elias I. 1, Krieger M 1, Rinaldi G. 2, Christian L. 1, Mancini N. 2 1 Klinik Rotes Kreutz, Frankfurt Am Main,

More information

Approach Patients with Confidence

Approach Patients with Confidence Approach Patients with Confidence The is the first stem specifically designed to be utilized with tissue sparing approaches, such as the anterior approach, as well as traditional approaches. The implant

More information

Cementless Tapered Femoral Stem Surgical technique

Cementless Tapered Femoral Stem Surgical technique Cementless Tapered Femoral Stem Surgical technique Contents Operative summary 4 Pre-operative planning 5 Femoral neck osteotomy 5 Femoral canal preparation 5 Intra-medullary (IM) reamer 6 Sequential rasping

More information

ACTIS. TOTAL HIP SYSTEM Approach Active Patients with Confidence DESIGN RATIONALE

ACTIS. TOTAL HIP SYSTEM Approach Active Patients with Confidence DESIGN RATIONALE ACTIS TOTAL HIP SYSTEM Approach Active Patients with Confidence DESIGN RATIONALE APPROACH ACTIVE PATIENTS WITH CONFIDENCE The ACTIS Total Hip System is the first DePuy Synthes Companies of Johnson & Johnson

More information

Management Of Acetabular Deficiency In Total Hip Arthroplasty: A Series Of 15 Cases

Management Of Acetabular Deficiency In Total Hip Arthroplasty: A Series Of 15 Cases ISPUB.COM The Internet Journal of Orthopedic Surgery Volume 21 Number 2 Management Of Acetabular Deficiency In Total Hip Arthroplasty: A Series Of 15 Cases G Khanna, R Sharma, D Singh, T A Chandy Citation

More information

CASE REPORT COMPLETE BONE REMODELING AFTER CALCAR RECONSTRUCTION WITH METAL WIRE MESH AND IMPACTION BONE GRAFTING: A CASE REPORT

CASE REPORT COMPLETE BONE REMODELING AFTER CALCAR RECONSTRUCTION WITH METAL WIRE MESH AND IMPACTION BONE GRAFTING: A CASE REPORT Nagoya J. Med. Sci. 75. 287 ~ 293, 2013 CASE REPORT COMPLETE BONE REMODELING AFTER CALCAR RECONSTRUCTION WITH METAL WIRE MESH AND IMPACTION BONE GRAFTING: A CASE REPORT Toshiki Iwase, M.D., Ph.D., Atsushi

More information

AGGRESSIVE GRANULOMATOUS LESIONS IN CEMENTLESS TOTAL HIP ARTHROPLASTY

AGGRESSIVE GRANULOMATOUS LESIONS IN CEMENTLESS TOTAL HIP ARTHROPLASTY AGGRESSIVE GRANULOMATOUS LESIONS IN CEMENTLESS TOTAL HIP ARTHROPLASTY SEPPO SANTAVIRTA, VEIJO HOIKKA, ANTTI ESKOLA, YRJO T. KONTTINEN, TIMO PAAVILAINEN, KAJ TALLROTH From the Orthopaedic Hospitalofthe

More information

Metha Short Hip Stem System

Metha Short Hip Stem System Metha Short Hip Stem System Accuracy That Stands Alone Aesculap Orthopaedics Metha Short Hip Stem System Designed For Anatomic Accuracy The Metha Short Hip Stem is designed for anatomic accuracy to restore

More information

Clinical and radiological results of the stemmed Mc Minn cup in hip revision surgery

Clinical and radiological results of the stemmed Mc Minn cup in hip revision surgery Acta Orthop. Belg., 2010, 76, 58-62 ORIGINAL STUDY Clinical and radiological results of the stemmed Mc Minn cup in hip revision surgery Pax WILLEMSE, Rene M. CASTELEIN, Paul L. P. A. BOM, Aart VERBURG,

More information

Acetabular Cup System. Clinical Summary

Acetabular Cup System. Clinical Summary Acetabular Cup System Clinical Summary A Prospective, Randomized Study of Cross-Linked and Non-Cross-Linked Polyethylene for Total Hip Arthroplasty at 10-Year Follow-Up Engh CA Jr., Hopper RH Jr., Huynh

More information

The Journal of Arthroplasty

The Journal of Arthroplasty The Journal of Arthroplasty 29 (2014) 2256 2261 Contents lists available at ScienceDirect The Journal of Arthroplasty journal homepage: www.arthroplastyjournal.org The 27 to 29-Year Outcomes of the PCA

More information

Avenir Femoral Hip System

Avenir Femoral Hip System Avenir Femoral Hip System Avenir Hip System Intuitive and Reproducible During its decade of experience in the bone compaction stem market, the Avenir Hip System has helped surgeons experience flexible,

More information

SURGICAL TECHNIQUE. Alpine Cemented Hip Stem

SURGICAL TECHNIQUE. Alpine Cemented Hip Stem SURGICAL TECHNIQUE Alpine Cemented Hip Stem The following technique is a general guide for the instrumentation of the Alpine Cemented Hip Stem. It is expected that the surgeon is already familiar with

More information

operative technique Kent Hip

operative technique Kent Hip operative technique Kent Hip The Kent Hip Operative Technique The Kent Hip was developed by Mr Cliff Stossel, FRCS in Maidstone, Kent, UK and first implanted in 1986. It was designed to deal with problems

More information

9800 Metric Blvd. Austin, Texas

9800 Metric Blvd. Austin, Texas rev. A Encore Orthopedics, Inc. 1998 www.encoremed.com 9800 Metric Blvd. Austin, Texas 78758 512-832-9500 1 contents How to use the Foundation Hip 1 2 Plan your approach Select your hardware Preparing

More information

Hip Fracture Management. Product overview

Hip Fracture Management. Product overview Hip Fracture Management Product overview Hip Fracture Management Intracapsular fractures Total hip replacement (THR) can offer clinical benefits over traditional internal fixation or hemiarthroplasty solutions

More information

Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty

Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty The Journal of Arthroplasty Vol. 23 No. 2 2008 Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty 7135 Primary Total Hip Arthroplasties after Developmental Dysplasia of the

More information

Plain Film CT. Principal Modality (2): Case Report # [] Date accepted: 15 March 2014

Plain Film CT. Principal Modality (2): Case Report # [] Date accepted: 15 March 2014 Radiological Category: Musculoskeletal Principal Modality (1): Principal Modality (2): Plain Film CT Case Report # [] Submitted by: Dr. Jason E. Lally, M.D. Faculty reviewer: Dr. Naga Ramesh Chinapuvvula,

More information

Section of Modular Hip Prostheses cemented. TMC-3 Modular Hip Prosthesis, cemented. TMC-3 Modular Hüftprothese, zementiert

Section of Modular Hip Prostheses cemented. TMC-3 Modular Hip Prosthesis, cemented. TMC-3 Modular Hüftprothese, zementiert Section of Modular Hip Prostheses cemented TMC-3 Modular Hip Prosthesis, cemented TMC-3 Modular Hüftprothese, zementiert Prothèse de hanche modulaire TMC-3, cimentée Indication : The TMC-3 Modular hip

More information

The effect of a lateral flare feature on implant stability

The effect of a lateral flare feature on implant stability International Orthopaedics DOI 10.1007/s00264-002-0355-3 Original Paper The effect of a lateral flare feature on implant stability Alex Leali( ) Joseph Fetto Harvey Insler David Elfenbein A. Leali J. Fetto

More information

The Treatment of Pelvic Discontinuity During Acetabular Revision

The Treatment of Pelvic Discontinuity During Acetabular Revision The Journal of Arthroplasty Vol. 20 No. 4 Suppl. 2 2005 The Treatment of Pelvic Discontinuity During Acetabular Revision Scott M. Sporer, MD, MS,*y Michael O Rourke, MD,z and Wayne G. Paprosky, MD, FACS*y

More information

Klinische und radiologische Ergebnisse einer «isoelastischen» Pfanne

Klinische und radiologische Ergebnisse einer «isoelastischen» Pfanne Klinische und radiologische Ergebnisse einer «isoelastischen» Pfanne Martin Beck, Prof. Dr.med. Klinik für Orthopädie und Unfallchirurgie Luzerner Kantonsspital resorption of acetabular bone can be observed

More information