STANDARD OPERATING PROCEDURE: RIGAKU RU 300 XRD POWDER SAMPLES

Size: px
Start display at page:

Download "STANDARD OPERATING PROCEDURE: RIGAKU RU 300 XRD POWDER SAMPLES"

Transcription

1 STANDARD OPERATING PROCEDURE: RIGAKU RU 300 XRD POWDER SAMPLES Purpose of this instrument: The Rigaku RU 300 XRD is used for phase analysis of polycrystalline films, characterizing single crystal, and epitaxial films, and powders. Location: 402 White Hall Primary Staff Contact: Harley Hart (412) cell Secondary Staff Contact: Weiqiang Ding (304) cell The Shared Research Facilities are operated for the benefit of all researchers. If you encounter any problems with this piece of equipment, please contact the staff member listed above immediately. There is never a penalty for asking questions. If the equipment is not behaving exactly the way it should, contact a staff member. WARNING: This system generates x rays. A radiation dosimeter MUST be worn at all times while using the instrument. START UP 1. Ensure the radiation dosimeter is attached to the person. 2. Sign in on the FOM and the system log book located on the XRD computer bench. Figure 1: Screenshot of FOM login page 3. Enter the water recirculator room and verify that the water recirculator reservoir is full as shown in Figure 2. This is indicated by the yellow FULL light being illuminated. 1

2 TANK WATER LEVEL INIDICATOR Figure 2: Image of indicator light on water recirculator 4. Verify that the flow meters on the front of the water recirculator are all above the minimum set points: X ray Target: 3.2 gpm TMP (Turbo Pump): 16 gph X ray Tube: 1.6 gpm If the readings of the flow meters are below the minimum set points, please contact a Shared Research Facilities staff member for assistance. 5. Locate the inlet and outlet pressure gauges behind the water recirculator. Read the values of each gauge and calculate the differential pressure. Record this value in the log book. WARNING: If the differential pressure is less than 16 psi, please contact a Shared Research Facilities staff member for assistance. Do not operate the system as damage to the tool can occur. 6. Verify that the ion gauge display on the Vacuum Controller panel has a minimum of two green lights illuminated as shown in Figure 3. Contact a Shared Facilities staff for assistance if there are fewer than two green lights illuminated. 7. Turn on the digital volt meter and record the resistance in the log book. If the resistance of the is above 150 mv, please contact a Shared Facilities staff for assistance. 2

3 Vacuum Pressure Indicator Lights Figure 3: Ion Gauge Control panel with vacuum level indicator lights illuminated 8. Check and set both shutter switches to the EXT position as shown in Figure 4. This ensures that the shutters are controlled by the external computer. Figure 4: Image of Shutter Control Panel with switches in EXT position 9. Verify that the voltage knob is set to 20 kv and the amperage knob is set to 10 ma on XG Controller panel see Figure 5 below. 10. Press the white T REV button and wait 10 seconds for the anode to spin up to speed see Figure 5 below. 3

4 TREV Button Current Control Knob Voltage Control Knob Figure 5: Image of XG Controller Panel 11. Press the white ON button to turn on the X ray generator. Verify that the X ray FOCUS NORMAL light is illuminated and that the voltage and the current gauges are reading 20 kv / 10 ma, respectively see Figure On the tool, check that the aluminum filter is in place in front of the high angle detector see Figure 6. Aluminum Filter Figure 6: Image of High Angle detector with filter in place WARNING: This is needed to prevent saturation and damage to the detector during the alignment process. 13. Turn on the corresponding detectors for the X ray scans to be conducted by pressing the toggle POWER button to ON and then the HV switch to see Figure 7 and 8 respectively. 4

5 For High Angle Scans: High Angle Detector and High Angle Reference Detector HV Switch POWER Button Figure 7: Image of High Angle Detectors with power and HV in the ON positions For Low Angle Scans: Low Angle Detector ONLY Figure 8: Image of Low Angle Detector with power and HV in the ON positions 14. Verify that the RED SAFETY light on the window is illuminated when the doors are closed before using the tool. If the doors are not closed properly before operating the tool, the system safety interlock will fail and the shutters will not open. Contact a Shared Research Facilities staff if the RED SAFETY light is off while the door is closed or the X ray generator is operating 15. Wait a minimum of ten minutes before conducting any scans to allow the optics to warm. This is to protect the lifetime of the detectors. 5

6 POWDER SAMPLE PREP 1. On a zero reflective piece of fused silica, apply a very thin layer of vacuum grease see Figures 9 and 10. Figure 9 and 10: Application of a thin layer of vacuum grease to the zero reflective fused silica pie ce. 2. Over a piece of weighing paper, slowly pour the powder over the fused silica. Then turn the fused silica upside down to remove any excess powder. Repeat this process until a thin layer of powder has adhered to the fused silica see Figures 11 and 12. Figure 11 and 12: Application of powder to zero reflective fused silica piece. 6

7 3. Remove excess material from edge of fused silica. The sample is now ready to be loaded into the system see Figures 13 and 14. Figure 13 and 14: Removal of sample excess from edge of zero reflective fused silica piece and final result. 4. Attach adhesive to the goniometer using the red sticky tabs and firmly and evenly press the sample dwon on top of the goniometer see figure 15 below. 5. Adjust the goniometer positions to zero using the goniometer screws. If they do not move easily, STOP and ensure that the set positions are not tightened never force the screws to move. Figure 15: Sample attached to the goniometer 7

8 POWDER SAMPLE ALIGNMENT 1. On the computer desktop open the High Angle lab view program see Figure 16 below. The Home screen is divided into three sections: Figure 16: High Angle home screen. Inputs: Allows the user to input scan type, range, increments for system scans Move Angle: The user can select one of the four drives (Theta, 2 Theta, Phi, Chi) and move the desired drive to a particular angle. This function is used during alignment and for returning drives to the home position when shutting the system down after use. Zero and Reset: This is used during alignment for entering the offset values. 2. Select the White Arrow on the toolbar to activate the system software. The arrow will change to a Black Arrow. 3. Move Θ and 2Θ to zero. This is done by selecting Θ in the Move Angle section, enter a value of zero, and then select the Move icon. Once the Θ drive has stopped moving, select Θ in the Zero and Reset section and select the Zero icon. Repeat this process for 2Θ. 4. Run a 2Θ scan with the following settings: Scan Type: 2Θ Final Angle: 2 Data Interval: 0.1 Θ Initial: 0 2Θ Initial: 2 Count Time: 1.0 s Phi Initial: 0 Chi Initial: 0 8

9 Select SCAN and then select REPLACE in the processing window. If you do not see a peak, check that the detectors are on and then repeat the scan with a date interval of 4 to 4. If this does not correct the problem, then contact a Shared Research Facilities staff member. 5. Once the scan is complete, open ORIGIN 7.5 software (this can be found on the Desktop). 6. In Origin, select File New Project. Then press CTRL+K and select the tempxr.dat file which contains the scan data. 7. Highlight the values in the C(Y) column which are your data values and press the plot icon. 8. Select Analysis Fit Gaussian. Record the Height and Center Offset values. NOTE: If the location of center is less than then proceed to step# 9 9. Return to the High Angle program. In the Move Angle section, select 2Θ from the drop down menu, enter the offset from center value, and then select MOVE. NOTE: The center offset value needs to be a multiple of If the offset is <1, then add 1 to the value and select MOVE. This will move the drive to the offset value plus one degree. Then enter the offset value and select MOVE. This will move the drive to the desired value. For example, if the offset is 0.075, first move to angle and then MOVE to This is necessary to prevent backlash in the system. 10. Repeat steps 4 until the Center Offset is less than Once the 2Θ offset is complete, enter the radiation enclosure. 12. Verify that the shutter is closed before placing your hands in the potential path of the X ray. If the shutter is open, set the shutter toggle switches to CLOSE. 13. Mount the sample goniometer to the goniometer chuck see Figure 17 below. Figure 17: Image of goniometer being attached to chuck. 14. Manually adjust the z position of the goniometer downwards by turning the z knob clockwise until the knob fails to turn see Figure 18. 9

10 Figure 18: Image of adjustment of z position. 15. Leave the radiation enclosure and enter the following Input values in the software program and select SCAN: Scan Type: Θ Final Angle: 2 Data Interval: 0.0 Θ Initial: 0 2Θ Initial: 0 Count Time: 1.0 s Phi Initial: 0 Chi Initial: Move the pop up Scan Graph window to the monitor inside the radiation enclosure. 17. Enter the radiation enclosure. Leave the entry door into the radiation enclosure open as this prevents the shutters from being controlled by the external computer. Next, change the shutter toggle switch position to OPEN. 18. Manually adjust the z position upwards so that the sample is halfway into the path beam. This is done by turning the z knob (see Figure 19) counter clockwise until the counts are half of the counts observed in the previous 2Θ (see Figure 20). 19. Reset the shutter toggle switch position to EXT and exit the radiation enclosure. 10

11 Figure 19: Using the z knob to adjust the position of the sample to half way into the x ray beam Figure 20: Screenshot of adjustment of z position 20. Move the pop up window back to the outside computer screen and select STOP. 21. Run a Θ scan by entering the following values in the input section: Scan Type: Θ Final Angle: 4 Data Interval: 0.1 Θ Initial: 4 2Θ Initial: 0 Count Time: 1.0 s Phi Initial: 0 Chi Initial: 90 Then select SCAN and REPLACE in the processing window. 22. Once the scan is complete, open ORIGIN 7.5 software (this can be found on the Desktop). 11

12 23. In Origin, select File New Worksheet. Then press CTRL+K and select the tempxr.dat file which contains the scan data. 24. Highlight the values in the C(Y) column which are your data values and press the plot icon. 25. Select Analysis Fit Gaussian. Record the Height and Center. 26. Return to the High Angle program. In the Move Angle section, select Θ from the drop down menu, enter the offset from center value and select MOVE. NOTE: The center offset value needs to be a multiple of If the offset is <1, then add 1 to the value and select MOVE. This will move the drive to the offset value plus one degree. Then enter the offset value and select MOVE. This will move the drive to the desired value. For example, if the offset is 0.075, first move to angle and then MOVE to This is necessary to prevent backlash in the system. 27. Repeat steps until the Center Offset is less than Repeat steps until the Height (counts) observed are half of the 2Θ value. 29. Once this is complete, the alignment is complete and it is time to increase the power of the generator to the desired settings. NOTE: The maximum power of the tool is 50 kv and 200 ma. 30. Increase the voltage and amperage using the following increments NOTE: Observe the DVM while slowly increasing the power so as to not allow the DVM to read above 200 mv. These settings are recommendations and are contingent upon the pressure not increasing above 200 mv. If the pressure nears 200mV, do not increase the voltage or amperage further until the pressure recovers. Also always increase the voltage before the amperage: 20 kv / 10mA 30 kv / 100 ma wait a minimum of 10 minutes 30kV / 100 ma40 kv / 160 ma wait a minimum of 10 minutes 40kV / 160 ma 40 kv / 200 ma wait a minimum of 10 minutes 40kV / 200 ma 50 kv / 200 ma wait a minimum of 10 minutes 31. Remove the aluminum filter from the detector and verify that the shutters switches are in the EXT position. 32. Enter the following values in the INPUTS sections: Scan Type: Θ 2Θ Final Angle: User Defined (<100 ) Data Interval: User Defined (> ) Θ Initial: User Defined (<50 ) 2Θ Initial: 2 times the value of Θ Initial Count Time: User Defined Phi Initial: 0 12

13 Chi Initial: 90 Then select SCAN and then Replace in the Processing window. 33. Once the scan is complete, open ORIGIN 7.5 software (this can be found on the Desktop). 34. In Origin, select File New Worksheet. Then press CTRL+K and select the tempxr.dat file which contains the scan data. 35. Highlight the values in the C(Y) column which are your data values and press the plot icon. Save DATA in appropriate USER FOLDER. 36. Proceed to Shut Down section below to return the tool to idle state. 13

14 SHUT DOWN 1. Move all the drives back to zero one at a time by selecting the drive from the drop down menu in the MOVE ANGLE section, enter a value of zero, and select MOVE. 2. On the X ray generator, reduce the power by following the sequence below: 50 kv / 200mA 40 kv / 200 ma wait a minimum of 10 seconds 40kV / 200 ma 40 kv / 160 ma wait a minimum of 10 seconds 40kV / 160 ma 30 kv / 100 ma wait a minimum of 10 seconds 30kV / 100 ma 20 kv / 10 ma these are the minimum values 3. Turn off the X ray generator by pressing the RED OFF button and listening for the rotating anode to spindown. 4. Remove sample holder goniometer from the tool. Gently remove sample from goniometer and clean adhesive from the sample holder using IPA and an alpha wipe. Dispose of all contaminated alpha wipes in the white contaminated wipes container. 5. Clean up the work area by doing the following: a. Clean the zero reflective piece of fused silica with IPA if it was used b. Place all tools into the appropriate storage containers c. Properly dispose any used materials i.e. alpha wipes, glass slides, etc. d. Return radiation monitoring badge to storage location 6. Log off the session on the FOM. Sign out in the log book. Report any problems in the comment section on the FOM and the log book. WARNING: If the powder has fallen off of your sample onto the system, a Shared Research Facilities staff member MUST be notified so that it can be properly cleaned. 14

15 EMERGENCY PROCEDURES If no one is available and the machine is not acting as expected, the user should do the following: Press STOP in Labview software. Set the voltage to 20kV and 5mA. Do not leave the machine running in an abnormal state. If the machine cannot be placed in the default state, immediately contact: Primary Staff Contact: Harley Hart (412) or Secondary Staff Contact: Dr. Weiqiang Ding (304) or The user should then complete a Tool Issue Report form and submit this form to room 409 of White Hall. The user should also add comment using the FOM software as to tool status. Then, if possible, the user should stay by the tool while trying to contact a Shared Facilities staff member. If it becomes necessary to leave the tool then the user should leave a large, legible note at the XRD stating the tool is DOWN. If a dangerous situation is evident (smoke, fire, sparks, etc.), ONLY if it is safe to do so, the user should turn off the power the x ray source by turning pressing the RED off button on the XG Controller. The user should turn off the pumps to the system by moving the pump toggle switches to the OFF positions. The user should then move the switch the breaker box at the back of the tool into the OFF position. The user should then notify all other lab persons within the White Hall 402 to evacuate and leave the lab immediately. The user should then contact proper emergency personnel from a safe place. Harley Hart Weiqiang Ding (412) (cell) (304) (cell) 15

Bruker D8 Discover User Manual (Version: July 2017)

Bruker D8 Discover User Manual (Version: July 2017) University of Minnesota College of Science and Engineering Characterization Facility Eulerian cradle Reference 2θ (two-theta): Detector angle (GADDS angle # 1) Bruker D8 Discover User Manual (Version:

More information

XRDUG Session II Edward Laitila 2/7/2009

XRDUG Session II Edward Laitila 2/7/2009 XRDUG Session II Edward Laitila 2/7/2009 DMSNT-Software and Computer Basic Information: XRDUG SESSION II DMS-NT Software: All bold words in this handout and contained in the operation manual refer to actual

More information

Commonwealth of Pennsylvania PA Test Method No. 423 Department of Transportation October Pages LABORATORY TESTING SECTION. Method of Test for

Commonwealth of Pennsylvania PA Test Method No. 423 Department of Transportation October Pages LABORATORY TESTING SECTION. Method of Test for Commonwealth of Pennsylvania PA Test Method No. 423 Department of Transportation 10 Pages 1. SCOPE LABORATORY TESTING SECTION Method of Test for RETRO-DIRECTIVE REFLECTIVITY OF REFLECTIVE MATERIALS 1.1

More information

Alignment and Calibration of the APEX Instrument. March 30, Chemical Crystallography Laboratory

Alignment and Calibration of the APEX Instrument. March 30, Chemical Crystallography Laboratory Alignment and Calibration of the APEX Instrument Page 1 of 8 Alignment and Calibration of the APEX Instrument March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Alignment and

More information

ProVari Owner s Manual

ProVari Owner s Manual ProVari Owner s Manual Thank you for purchasing a ProVari Electronic Cigarette! Please read this manual before using your e cigarette. www.provape.com 1 An electronic cigarette, otherwise known as a personal

More information

RADIATION MONITORING DEVICES R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S

RADIATION MONITORING DEVICES R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S RADIATION MONITORING DEVICES 10-526- 1 9 7 R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S DETECTION AND MEASUREMENT OF IONIZING RADIATION Dosimeter Dose-measuring device Two classifications:

More information

THE RIGAKU JOURNAL VOL. 23 / 2006, A1-A10 CONSIDERATIONS REGARDING THE ALIGNMENT OF DIFFRACTOMETERS FOR RESIDUAL STRESS ANALYSIS

THE RIGAKU JOURNAL VOL. 23 / 2006, A1-A10 CONSIDERATIONS REGARDING THE ALIGNMENT OF DIFFRACTOMETERS FOR RESIDUAL STRESS ANALYSIS THE RIGAKU JOURNAL VOL. 23 / 2006, A1-A10 CONSIDERATIONS REGARDING THE ALIGNMENT OF DIFFRACTOMETERS FOR RESIDUAL STRESS ANALYSIS THOMAS R. WATKINS, O. BURL CAVIN, CAMDEN R. HUBBARD, BETH MATLOCK, AND ROGER

More information

Management of Central Venous Access Devices. Blood Glucose Monitoring

Management of Central Venous Access Devices. Blood Glucose Monitoring Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including

More information

FT-302 Force Transducer

FT-302 Force Transducer Technical Note FT-302 LabScribe is a trademark of 2015 Overview The FT-302 is a high-sensitivity dual-range research grade force transducer designed to measure forces in the 0.005 to 10 gram and 0 to 100

More information

MiniFlex Diffractometer. Start-Up Manual

MiniFlex Diffractometer. Start-Up Manual MiniFlex Diffractometer Start-Up Manual Rigaku/MSC 9009 New Trails Drive The Woodlands, TX 77381 Tel: 281-363-1033 Fax: 281-364-3628 Web: www.rigakumsc.com Email: info@rigakumsc.com Table of Contents

More information

OWNER S MANUAL. BODYCRAFT 7699 Green Meadows Dr., Lewis Center, Ohio Tel:

OWNER S MANUAL. BODYCRAFT 7699 Green Meadows Dr., Lewis Center, Ohio Tel: OWNER S MANUAL BODYCRAFT 7699 Green Meadows Dr., Lewis Center, Ohio 43035 Tel: 800.990.5556 1 www.bodycraft.com SAFETY INSTRUCTION WARNING: To reduce the risk of serious injury, read the following safety

More information

User Manual Important: First read the Medication Guide that comes inside your FORTEO carton.

User Manual Important: First read the Medication Guide that comes inside your FORTEO carton. FORTEO (for-tay-o) teriparatide (rdna origin) injection 1 User Manual Important: First read the Medication Guide that comes inside your FORTEO carton. Before you use your new FORTEO delivery device, please

More information

Motorised Treadmill Model No: JI1625

Motorised Treadmill Model No: JI1625 BODYWORX Motorised Treadmill Model No: JI1625 Retain this owner s manual for future reference Read and follow all instructions in this owner s manual Version A 1 TABLE OF CONTENTS OVERVIEW DRAWING 2 ASSEMBLY

More information

Instructions For Use. SOLUTION FOR INJECTION IN PRE-FILLED PEN 250 micrograms

Instructions For Use. SOLUTION FOR INJECTION IN PRE-FILLED PEN 250 micrograms Instructions For Use SOLUTION FOR INJECTION IN PRE-FILLED PEN 250 micrograms Contents 1. How to use your Ovitrelle pre-filled pen 2. Before you start using your Ovitrelle pre-filled pen 3. Getting your

More information

Accu-Chek Inform II: Point of Care Glucose Testing. Sharp Healthcare 2014

Accu-Chek Inform II: Point of Care Glucose Testing. Sharp Healthcare 2014 Accu-Chek Inform II: Point of Care Glucose Testing Sharp Healthcare 2014 OBJECTIVES At the completion of this module the participant will be able to: Learn the proper technique of performing a finger stick

More information

SP850 OWNER S MANUAL

SP850 OWNER S MANUAL SP850 OWNER S MANUAL Product may vary slightly from the item pictured due to model upgrades Read all instructions carefully before using this product. Retain this owner s manual for future reference. NOTE:

More information

Glucose Meter. User Guide. Veterinary Monitoring System. For dog and cat use only

Glucose Meter. User Guide. Veterinary Monitoring System. For dog and cat use only Glucose Meter User Guide Veterinary Monitoring System For dog and cat use only Gpet instruction Manual 31/5/09 18:06 Page 2 Gpet instruction Manual 31/5/09 18:06 Page 3 TABLE OF CONTENTS Your g-pet system

More information

Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient.

Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient. 1 2 Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient. All such tests are considered lab tests and are

More information

Dental Intraoral X-ray Systems

Dental Intraoral X-ray Systems Dental Intraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 2.0 4732.#### DENTAL INTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Commented [JC(1]: Based on part 4732.0880. Subpart 1. Applicability.

More information

Medescan Nebuliser Med - S600A Instructions Manual

Medescan Nebuliser Med - S600A Instructions Manual Medescan Nebuliser Med - S600A Instructions Manual Please read this guidebook carefully before operating this unit Your Nebuliser is intended for use in the treatment of asthma, COPD and other respiratory

More information

CoaguChek S System. Quick Reference Guide. This is a CLIA-waived System. Coagulation Testing with Fresh Whole Blood

CoaguChek S System. Quick Reference Guide. This is a CLIA-waived System. Coagulation Testing with Fresh Whole Blood CoaguChek S System This is a CLIA-waived System Quick Reference Guide Coagulation Testing with Fresh Whole Blood May not print or view at 100% All unmarked prints black. Coag Swoosh prints gradients of

More information

TR1120 OWNER S MANUAL

TR1120 OWNER S MANUAL TR1120 OWNER S MANUAL IMPORTANT: For maximum effectiveness and safety, please read this owner s manual before using your BodyCraft treadmill. Retain this owner s manual for future reference. 1. It is the

More information

IC031 Magnetic Belt Drive Performance Indoor Cycle Bike

IC031 Magnetic Belt Drive Performance Indoor Cycle Bike IC031 Magnetic Belt Drive Performance Indoor Cycle Bike USER MANUAL IMPORTANT: Read all instructions carefully before using this product. Retain owner s manual for future reference. For customer service,

More information

Product Specifications and Usage Information P4405, P4410, & P4420 Genesys 30 Dosimeter Holder System

Product Specifications and Usage Information P4405, P4410, & P4420 Genesys 30 Dosimeter Holder System DESCRIPTION The Thermo Genesys 30 Spectrophotometer (G30) is supplied with specially designed dosimeter holders to fit a variety of dosimeters, including the B3 DoseStix and B3 WINdose dosimeters. The

More information

Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation

Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation Background The arterial system functions as a pressure reservoir. Blood enters via the heart and exits through the capillaries.

More information

Living with Newton's Laws

Living with Newton's Laws Task #1 - Newton s 1 st Law - This is a pain in the neck Let's suppose you are in your car, waiting at a stop light. Like any good driver, you have your seat belt buckled. (It's the law.) Suddenly, a car

More information

Kappa Apex3 instruction manual. Procedure. Version 1.0 CuKα / MoKα Duo Edition

Kappa Apex3 instruction manual. Procedure. Version 1.0 CuKα / MoKα Duo Edition Kappa Apex instruction manual Version 1.0 CuKα / MoKα Duo Edition Procedure Kappa Apex instruction manual... 1 Introduction:... 2 Assumptions:... 2 Adjusting Temperature:... New Project... Centering your

More information

Skeletal muscles are composed of hundreds to thousands of individual cells,

Skeletal muscles are composed of hundreds to thousands of individual cells, 2 E X E R C I S E Skeletal Muscle Physiology O B J E C T I V E S 1. To define these terms used in describing muscle physiology: multiple motor unit summation, maximal stimulus, treppe, wave summation,

More information

TWIST STEPPER NO. 045

TWIST STEPPER NO. 045 TWIST STEPPER NO. 045 IMPORTANT: Read all instructions carefully before using this product. Retain owner s manual for future reference. For customer service, please contact: support@sunnyhealthfitness.com

More information

Lilly. Disposable Insulin Delivery Device User Manual

Lilly. Disposable Insulin Delivery Device User Manual 1 PV 3734 AMP Lilly Disposable Insulin Delivery Device User Manual Instructions for Use Read and follow all of these instructions carefully. If you do not follow these instructions completely, you may

More information

Sanako Lab 100 STS USER GUIDE

Sanako Lab 100 STS USER GUIDE Sanako Lab 100 STS USER GUIDE Copyright 2002-2015 SANAKO Corporation. All rights reserved. Microsoft is a registered trademark. Microsoft Windows XP, Windows Vista and Windows 7 are trademarks of Microsoft

More information

BlueBayCT - Warfarin User Guide

BlueBayCT - Warfarin User Guide BlueBayCT - Warfarin User Guide December 2012 Help Desk 0845 5211241 Contents Getting Started... 1 Before you start... 1 About this guide... 1 Conventions... 1 Notes... 1 Warfarin Management... 2 New INR/Warfarin

More information

Manual. HTR 2.0 Ergometer Hometrainer. T +31 (0) e w

Manual. HTR 2.0 Ergometer Hometrainer. T +31 (0) e w HTR 2.0 Ergometer Hometrainer Manual For questions or missing parts please contact Fitness Benelux: Fitness Benelux Uw partner in fitness twekkelerweg 263 7553 lz Hengelo the netherlands T +31 (0)74-7600219

More information

RaySafe i3 INSTALLATION & SERVICE MANUAL

RaySafe i3 INSTALLATION & SERVICE MANUAL RaySafe i3 INSTALLATION & SERVICE MANUAL 2017.06 Unfors RaySafe 5001104-1.1 All rights are reserved. Reproduction or transmission in whole or in part, in any form or by any means, electronic, mechanical

More information

Allergy Basics. This handout describes the process for adding and removing allergies from a patient s chart.

Allergy Basics. This handout describes the process for adding and removing allergies from a patient s chart. Allergy Basics This handout describes the process for adding and removing allergies from a patient s chart. Accessing Allergy Information Page 1 Recording No Known Medication Allergies Page 2 Recording

More information

Adventure 4 Plus Service Manual

Adventure 4 Plus Service Manual Adventure 4 Plus Service Manual 1 TABLE OF CONTENTS CHAPTER 1: SERIAL NUMBER LOCATION...3 CHAPTER 2: PREVENTATIVE MAINTENANCE 2.1 Preventative Maintenance. 4 2.2 Tension and Centering the Running Belt....6

More information

Contour Diabetes app User Guide

Contour Diabetes app User Guide Contour Diabetes app User Guide Contents iii Contents Chapter 1: Introduction...5 About the CONTOUR DIABETES app...6 System and Device Requirements...6 Intended Use...6 Chapter 2: Getting Started...7

More information

CS User and Installation Guide

CS User and Installation Guide CS 3600 User and Installation Guide Notice No part of this publication may be reproduced, stored in a retrieval system, translated to another language, or transmitted in any form by any means, electronic,

More information

Adventure 3 Plus Service Manual

Adventure 3 Plus Service Manual Adventure 3 Plus Service Manual 1 TABLE OF CONTENTS CHAPTER 1: SERIAL NUMBER LOCATION...3 CHAPTER 2: PREVENTATIVE MAINTENANCE 2.1 Preventative Maintenance. 4 2.2 Tension and Centering the Running Belt....6

More information

Point of Care Testing. Bedside Glucose Meter

Point of Care Testing. Bedside Glucose Meter Point of Care Testing Bedside Glucose Meter 11/2017 Regulatory Overview Bedside glucose is a point of care lab test and YOU are performing this lab test under the Lab s CLIA license. All lab tests are

More information

User Guide for Lifegear RUNNER App

User Guide for Lifegear RUNNER App User Guide for Lifegear RUNNER App http://www.lifegear.com.tw Contents LifeGear RUNNER App Installation 2 Connect Your Mobile Device to the Specific Bluetooth Fitness Equipment 4 Entering Your Profile

More information

BEMC/BSMD O.Tsai, 04/13/2006 Latest Update 01/30/2014

BEMC/BSMD O.Tsai, 04/13/2006 Latest Update 01/30/2014 1 BEMC/BSMD O.Tsai, 04/13/2006 Latest Update 01/30/2014 Detector Operator Manual, Run 14 You will be operating three detectors: (BTOW) Barrel Electromagnetic Calorimeter (BSMD) Barrel Shower Maximum Detector

More information

Tips for Training New Customers

Tips for Training New Customers Tips for Training New Customers Good Habits: Test Strips Recap vial immediately and tightly after removing a single test strip. Never remove the desiccant from the vial. Do not lay test strips out ahead

More information

Activity # 4. Measurement of Enzyme Activity

Activity # 4. Measurement of Enzyme Activity Activity # 4. Measurement of Enzyme Activity Adapted from Experimental Molecular Biology Lab Manual, Stephen H. Munroe, Marquette University Week 1 Protein Extraction, Concentration Measurement & Instrument

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 37-013 GARD Users Manual March 2005 Manual No. 37-013-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective

More information

Instructions for Use Enbrel (en-brel) (etanercept) for injection, for subcutaneous use Multiple-dose Vial

Instructions for Use Enbrel (en-brel) (etanercept) for injection, for subcutaneous use Multiple-dose Vial Instructions for Use Enbrel (en-brel) (etanercept) for injection, for subcutaneous use Multiple-dose Vial How do I prepare and give an injection with Enbrel multiple-dose vial? A multiple-dose vial contains

More information

Adventure 5 Plus Service Manual

Adventure 5 Plus Service Manual Adventure 5 Plus Service Manual 1 TABLE OF CONTENTS CHAPTER 1: SERIAL NUMBER LOCATION...3 CHAPTER 2: PREVENTATIVE MAINTENANCE 2.1 Preventative Maintenance. 4 2.2 Tension and Centering the Running Belt....6

More information

Dental Extraoral X-ray Systems

Dental Extraoral X-ray Systems Dental Extraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL EXTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability. A registrant s x-ray system used for dental

More information

Evolve 3 & 5 Service Manual

Evolve 3 & 5 Service Manual Evolve 3 & 5 Service Manual 1 Product Browse 2 Contents CHAPTER 1: SERIAL NUMBER LOCATION... 5 CHAPTER 2: CONSOLE INSTRUCTIONS 2.1 Console Overview... 6 2.1.1 Evolve 3 Console Overview... 6 2.1.2 Evolve

More information

SANAKO Lab 100 STS USER GUIDE

SANAKO Lab 100 STS USER GUIDE SANAKO Lab 100 STS USER GUIDE Copyright 2008 SANAKO Corporation. All rights reserved. Microsoft is a registered trademark. Microsoft Windows 2000 and Windows XP are trademarks of Microsoft Corporation.

More information

TIMS Lab 1: Phase-Locked Loop

TIMS Lab 1: Phase-Locked Loop TIMS Lab 1: Phase-Locked Loop Modules: VCO, Multiplier, Quadrature Utilities, Wideband True RMS Meter, Tuneable LPF, Digital Utilities, Noise Generator, Speech, Headphones 0 Pre-Laboratory Reading Phase-locked

More information

How to use your Pergoveris pre-filled pen

How to use your Pergoveris pre-filled pen How to use your Pergoveris pre-filled pen Information in this user guide is based on the European product information for Pergoveris. It is not country specific and may vary from country to country. Please

More information

Monitor Instructions for Models: CHB-R6 CHB-UV6

Monitor Instructions for Models: CHB-R6 CHB-UV6 Monitor Instructions for Models: CHB-R6 CHB-UV6 Power supply: output DC 9v/ 1000mA transformer adaptor TABLE OF CONTENTS Function Buttons ----------------------------------------------------------------------------

More information

General Single Ion Calibration. Pete 14-May-09

General Single Ion Calibration. Pete 14-May-09 General Single Ion Calibration Pete 14-May-09 Purpose of SI calibration Measure the instrument response of a single ion. Necessary for understanding of error in instrument (counting statistics) Calculation

More information

TABLE OF CONTENTS. 2.0 USING MANAGER PREFERENCE 2.1 Using manager preference mode 2.2 Selecting US or metric 2.3 Using CSafe

TABLE OF CONTENTS. 2.0 USING MANAGER PREFERENCE 2.1 Using manager preference mode 2.2 Selecting US or metric 2.3 Using CSafe E L L I P T I C A L T O T A L B O D O W N E R S M A N U A L TABLE OF CONTENTS 1.0 IMPORTANT SAFET INSTRUCTIONS Read and save these instructions Setting up the Elliptical total body 1.1 Installation requirements

More information

Thermo Finnigan. 6-2 LCQ Advantage Getting Started-

Thermo Finnigan. 6-2 LCQ Advantage Getting Started- To set up the LCQ Advantage to acquire data in the APCI/MS mode, you do the following:.install the APCI probe assembly.set up to infuse your analyte into the MS detector using a syringe pump connected

More information

RESULTS REPORTING MANUAL. Hospital Births Newborn Screening Program June 2016

RESULTS REPORTING MANUAL. Hospital Births Newborn Screening Program June 2016 RESULTS REPORTING MANUAL Hospital Births Newborn Screening Program June 2016 CONTENTS GETTING STARTED... 1 Summary... 1 Logging In... 1 Access For New Hires... 2 Reporting Parental Refusals... 3 Adding

More information

Direct Reading (Pocket) Dosimeter

Direct Reading (Pocket) Dosimeter Direct Reading (Pocket) Dosimeter What is a Pocket Dosimeter???? Direct Read Pocket Dosimeters are rugged, precision instruments about the size of a pocket fountain pen, which are used to measure accumulative

More information

Splitting of the characteristic K-alpha doublet radiation of molybdenum (Item No.: P )

Splitting of the characteristic K-alpha doublet radiation of molybdenum (Item No.: P ) Splitting of the characteristic K-alpha doublet radiation of molybdenum (Item No.: P2540701) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne Physik

More information

Cigarette Smoking Machine CSM-SCSM. User s Manual

Cigarette Smoking Machine CSM-SCSM. User s Manual Cigarette Smoking Machine CSM-SCSM User s Manual CH TECHNOLOGIES (USA), Inc. Copyright 2011 CONTENTS 1. Introduction... 3 2. CSM-SCSM Description... 3 3. CSM-SCSM Specifications...6 4. Preparation of CSM-SCSM

More information

PRE-FILLED PEN 300 IU/0.5 ml N F

PRE-FILLED PEN 300 IU/0.5 ml N F PRE-FILLED PEN 300 IU/0.5 ml N1990207F PRE-FILLED PEN 300 IU/0.5 ml Instructions for use 300 IU/0.5 ml (22 micrograms/0.5 ml) Contents 1. How to use the GONAL-f pre-filled pen 2. How to use your GONAL-f

More information

PEN USER MANUAL Byetta 5 micrograms solution for injection in pre-filled pen (exenatide)

PEN USER MANUAL Byetta 5 micrograms solution for injection in pre-filled pen (exenatide) PEN USER MANUAL Byetta 5 micrograms solution for injection in pre-filled pen (exenatide) Section 1. WHAT YOU NEED TO KNOW ABOUT YOUR BYETTA PEN Read this section completely before you begin. Then, move

More information

8 Performing Medical Procedures

8 Performing Medical Procedures 8 The Terason usmart3200t Ultrasound System can aid in performing medical procedures such as biopsies. Depending on whether you purchased the additional equipment required for these procedures, you may

More information

WEIGHT BENCH User s Manual

WEIGHT BENCH User s Manual Model No. 831.15945.0 Serial No. Write the serial number in the space above for reference. WEIGHT BENCH User s Manual Serial Number Decal Assembly Operation Maintenance Part List and Drawing Sears, Roebuck

More information

Technical Bulletin. Technical Information for Quidel Molecular Influenza A+B Assay on the Bio-Rad CFX96 Touch

Technical Bulletin. Technical Information for Quidel Molecular Influenza A+B Assay on the Bio-Rad CFX96 Touch Technical Bulletin Technical Information for Quidel Molecular Influenza A+B Assay on the Bio-Rad CFX96 Touch Quidel Corporation has verified the performance of the Quidel Molecular Influenza A+B Assay

More information

SP450 OWNER S MANUAL

SP450 OWNER S MANUAL OWNER S MANUAL Product may vary slightly from the item pictured due to model upgrades Read all instructions carefully before using this product. Retain this owner s manual for future reference. NOTE: This

More information

BLINDED MODE USERS GUIDE SUPPLEMENT

BLINDED MODE USERS GUIDE SUPPLEMENT BLINDED MODE USERS GUIDE SUPPLEMENT BLINDED MODE user s guide SUPPLEMENT WARNING: The SEVEN PLUS when set to Blinded Mode using the Data Manager Software will not provide real-time continuous glucose readings,

More information

CIRCUIT ZONE ELLIPTICAL

CIRCUIT ZONE ELLIPTICAL CIRCUIT ZONE ELLIPTICAL SF-E386 USER MANUAL IMPORTANT! Please retain owner s manual for maintenance and adjustment instructions. Your satisfaction is very important to us, PLEASE DO NOT RETURN UNTIL YOU

More information

TWISTING STAIR STEPPER WITH BANDS NO. 068 USER MANUAL

TWISTING STAIR STEPPER WITH BANDS NO. 068 USER MANUAL TWISTING STAIR STEPPER WITH BANDS NO. 068 USER MANUAL IMPORTANT: Read all instructions carefully before using this product. Retain owner s manual for future reference. For customer service, please contact:

More information

#20 Separation of Alcohols Using Gas Chromatography

#20 Separation of Alcohols Using Gas Chromatography #20 Separation of Alcohols Using Gas Chromatography Purpose: The purpose of this experiment is to determine the composition of an unknown alcohol mixture by using gas chromatography. Introduction: A gas

More information

Operational Safety Procedure

Operational Safety Procedure Operational Safety Procedure For Bruker DAVINCI Location : room 2407 Emergency Response In case of emergency a) cease all operations that will put you at risk to exposure b) de-energize instrumentation

More information

Graphical User Interface for Windows for MCM Module

Graphical User Interface for Windows for MCM Module Page 1 of 8 Installation:- Please ensure that you have the latest updated GUI and download the GUIv.x.x.x.x.ZIP from our website. Extract the folder to a temporary location (These files can be deleted

More information

OKLAHOMA STATE UNIVERSITY X-RAY MACHINE PROGRAM AND PI HANDBOOK For easy navigation, please use Adobe s bookmark feature.

OKLAHOMA STATE UNIVERSITY X-RAY MACHINE PROGRAM AND PI HANDBOOK For easy navigation, please use Adobe s bookmark feature. OKLAHOMA STATE UNIVERSITY X-RAY MACHINE PROGRAM AND PI HANDBOOK For easy navigation, please use Adobe s bookmark feature. Table of Contents Overview... 2 Section 1: X-ray Program Oversight... 2 1.a: Institutional

More information

PedAlert 100. PedAlert 120

PedAlert 100. PedAlert 120 PedAlert 100 & PedAlert 120 Operator s Manual 1996 PLANET LLC, Patent #6,174,294, B1 Notice: The manufacturer reserves the right to make changes to the device and Operator s Manual without obligation to

More information

USER MANUAL Rev. 1.15

USER MANUAL Rev. 1.15 QUANTUM QSP 7 TM BEACH CHAIR SHOULDER POSITIONER (BCP) USER MANUAL Rev. 1.15 Please read all the following instructions carefully to ensure proper functionality and to avoid injury or damage to your device.

More information

ELECTRONIC CIGARETTE KIT USER MANUAL

ELECTRONIC CIGARETTE KIT USER MANUAL ELECTRONIC CIGARETTE KIT USER MANUAL Mouthpiece Sliding Top Cap CONGRATULATIONS ON PURCHASING YOUR TOTALLY WICKED OPTIMUS ELECTRONIC CIGARETTE KIT. Airflow Control Ring LED Indicator Light OLED Screen

More information

TWISTING STAIR STEPPER WITH BANDS

TWISTING STAIR STEPPER WITH BANDS TWISTING STAIR STEPPER WITH BANDS NO. 068 USER MANUAL IMPORTANT! Please retain owner s manual for maintenance and adjustment instructions. Your satisfaction is very important to us, PLEASE DO NOT RETURN

More information

Gambro Renal Products Technical Assistance Services

Gambro Renal Products Technical Assistance Services Gambro Renal Products Technical Assistance Services http://tech.usa-gambro.com 800/525-2623 303/232-6800 Technical Update PH,TR,11060 12/2008 TROUBLESHOOTING Equipment Type: Phoenix Dialysis System Subject:

More information

LECTURE 13. Dr. Teresa D. Golden University of North Texas Department of Chemistry

LECTURE 13. Dr. Teresa D. Golden University of North Texas Department of Chemistry LECTURE 13 Dr. Teresa D. Golden University of North Texas Department of Chemistry Goniometer circle - centered at the sample, with the x-ray source and detector on the circumference of the circle. Focusing

More information

USER MANUAL CAUTION. MADE IN CHINA 2015 Body Action Enterprise Co., Ltd. All rights reserved.

USER MANUAL CAUTION. MADE IN CHINA 2015 Body Action Enterprise Co., Ltd. All rights reserved. USER MANUAL CAUTION MADE IN CHINA 05 Body Action Enterprise Co., Ltd. All rights reserved. READ ALL PRECAUTIONS AND INSTRUCTIONS IN THIS MANUAL BEFORE USING THIS EQUIPMENT AND KEEP THIS FOR FUTURE REFERENCE

More information

Fraser Health Authority Accu-chek Inform II: Training Script

Fraser Health Authority Accu-chek Inform II: Training Script 1. Introduction Ensure that each participant Signs the attendance Sheet (please PRINT first & last name, and middle initial legibly) Refer to; For Reference Only: o Quick Reference Guide o What s new document

More information

SpiderX. Portable Residual Stress X-Ray Diffractometer.

SpiderX. Portable Residual Stress X-Ray Diffractometer. Portable Residual Stress X-Ray Diffractometer www.gnr.it About SpiderX GNR Analytical Instrument offers equipment based on X-Ray Diffraction to measure residual stress state and retained austenite content.

More information

How to Use ENBREL : Vial Adapter Method

How to Use ENBREL : Vial Adapter Method How to Use ENBREL : Vial Adapter Method SETTING UP FOR AN INJECTION Select a clean, well-lit, flat working surface, such as a table. Take the ENBREL dose tray out of the refrigerator and place it on your

More information

ATLANTIS WebOrder. ATLANTIS ISUS User guide

ATLANTIS WebOrder. ATLANTIS ISUS User guide ATLANTIS WebOrder ATLANTIS ISUS User guide Contents ATLANTIS WebOrder Entering an ATLANTIS ISUS order 3 ATLANTIS ISUS implant suprastructures 4 ATLANTIS ISUS Bar 5 ATLANTIS ISUS Bridge 7 ATLANTIS ISUS

More information

SureFlow. Oxygen Flow Station PROVIDER TECHNICAL MANUAL

SureFlow. Oxygen Flow Station PROVIDER TECHNICAL MANUAL SureFlow Oxygen Flow Station PROVIDER TECHNICAL MANUAL Table of Contents Definition of Product Symbols... 3 General Information... 3 Warning and Caution Statements...3 Introduction to the SureFlow Oxygen

More information

Sleep Apnea Therapy Software Clinician Manual

Sleep Apnea Therapy Software Clinician Manual Sleep Apnea Therapy Software Clinician Manual Page ii Sleep Apnea Therapy Software Clinician Manual Notices Revised Notice Trademark Copyright Sleep Apnea Therapy Software Clinician Manual 103391 Rev A

More information

Breezhaler. Open to see clear capsule. How do I use my Breezhaler? Care of my Breezhaler

Breezhaler. Open to see clear capsule. How do I use my Breezhaler? Care of my Breezhaler Breezhaler 3 Pull cap off Press buttons once and release Breathe in rapidly and steadily. Hold. Open. Place capsule in chamber. Close (click). Breathe out 4 Open to see clear capsule. If not all clear

More information

For use only with INSTRUCTIONS FOR USE

For use only with INSTRUCTIONS FOR USE TM For use only with INSTRUCTIONS FOR USE IMPORTANT NOTICE: Please read this safety information first. 1. Follistim Pen is a precision device. It is very important that you read and follow all directions

More information

GETTING STARTED WITH MINIMED 670G CONTINUOUS GLUCOSE MONITORING

GETTING STARTED WITH MINIMED 670G CONTINUOUS GLUCOSE MONITORING GETTING STARTED WITH MINIMED 670G CONTINUOUS GLUCOSE MONITORING Contents Continuous Glucose Monitoring Getting started with continuous glucose monitoring Section 1: Section 2: Section 3: Section 4: Section

More information

THE TOTALLY WICKED TORNADO NC CONGRATULATIONS ON PURCHASING YOUR TOTALLY WICKED TORNADO NC ELECTRONIC CIGARETTE KIT

THE TOTALLY WICKED TORNADO NC CONGRATULATIONS ON PURCHASING YOUR TOTALLY WICKED TORNADO NC ELECTRONIC CIGARETTE KIT USER MANUAL THE TOTALLY WICKED TORNADO NC CONGRATULATIONS ON PURCHASING YOUR TOTALLY WICKED TORNADO NC ELECTRONIC CIGARETTE KIT 3 QUICK START GUIDE For those familiar with e-cigarettes, the steps below

More information

Exercise 6: Muscle Physiology II Twitch & Summation

Exercise 6: Muscle Physiology II Twitch & Summation Exercise 6: Muscle Physiology II Twitch & Summation Text Reading: Silverthorn, 5 th ed. 412 419, 425 427; 6 th ed. pg. 410 420 In this exercise, we will investigate the physiology of contraction in the

More information

Vital Signs Monitor Series (VSM) Outline Train the Trainer

Vital Signs Monitor Series (VSM) Outline Train the Trainer Vital Signs Monitor 52000 Series (VSM) Outline Train the Trainer A. Light Weight and Durable 1. Weighs approximately 6 pounds 2. Made of heavy plastic B. Battery 1. Sealed lead acid battery 6 V 2. Battery

More information

IMPORTANT INFORMATION - PLEASE READ. Table of Contents. Introduction. General Information

IMPORTANT INFORMATION - PLEASE READ. Table of Contents. Introduction. General Information IMPORTANT INFORMATION - PLEASE READ Introduction Congratulations! Welcome to the world of the ELLIPTICAL CROSS TRAINER. The ELLIPTICAL CROSS TRAINER is one of the finest and most comprehensive pieces of

More information

Warfarin Help Documentation

Warfarin Help Documentation Warfarin Help Documentation Table Of Contents Warfarin Management... 1 iii Warfarin Management Warfarin Management The Warfarin Management module is a powerful tool for monitoring INR results and advising

More information

Exercise 2: Effects of Cold Temperature Aim: To record changes in heart rate after the heart is bathed in cold Ringer s solution.

Exercise 2: Effects of Cold Temperature Aim: To record changes in heart rate after the heart is bathed in cold Ringer s solution. Experiment AM-3: Heart Muscle Warning: The heart preparation used in this experiment is functional for a limited period of time. If the muscle is bathed periodically in Ringer s solution, it will work

More information

NPTEL NPTEL ONLINE COURSE. NPTEL Online Certification Course (NOC) NPTEL. Theory and Practice of Non Destructive Testing

NPTEL NPTEL ONLINE COURSE. NPTEL Online Certification Course (NOC) NPTEL. Theory and Practice of Non Destructive Testing NPTEL NPTEL ONLINE COURSE NPTEL Online Certification Course (NOC) NPTEL Theory and Practice of Non Destructive Testing Dr. Ranjit Bauri Dept. of Metallurgical & Materials Engineering IIT Madras, Chennai

More information

Quick guide. ReSound LiNX 3D 77

Quick guide. ReSound LiNX 3D 77 Quick guide ReSound LiNX 3D 77 My hearing aid information Product: ReSound LiNX 3D Model: 77 Serial number (R): Serial number (L): Battery size: 13 Program Button (A): ACTION RESULT Programs: (if set up

More information