1 This copy is for personal use only. To order printed copies, contact Original Research n Pediatric Imaging Feasibility and Diagnostic Accuracy of Supersonic Shear- Wave Elastography for the Assessment of Liver Stiffness and Liver Fibrosis in Children: A Pilot Study of 96 Patients 1 Stéphanie Franchi-Abella, MD Lucie Corno, MD Emmanuel Gonzales, MD, PhD Guillemette Antoni, MD Monique Fabre, MD Béatrice Ducot, PhD Danièle Pariente, MD Jean-Luc Geisson, PhD Mickael Tanter, PhD Jean-Michel Corréas, MD, PhD 1 From the Departments of Pediatric Radiology (S.F.A., L.C., D.P.), Pediatric Hepatology (E.G.), and Epidemiology (G.A., B.D.), Hôpital Bicêtre, Hôpitaux Universitaires Paris Sud, Assistance Publique Hôpitaux de Paris, Université Paris Sud, 78 rue du Général Leclerc, Le Kremlin Bicêtre, France; Departments of Pathology (M.F.) and Adult Radiology (J.M.C.), Hôpital Necker, Assistance Publique Hôpitaux de Paris, Université Paris V, Assistance Publique Hôpitaux de Paris, Paris, France; and Institut Langevin, ESPCI ParisTech, CNRS UMR7587, INSERM U979, Paris, France (J.L.G., M.T.). Received December 12, 2014; revision requested January 12, 2015; revision received April 21; accepted April 30; final version accepted May 21. Address correspondence to S.F.A. ( stephanie. franchi@bct.aphp.fr). q RSNA, 2015 Purpose: Materials and Methods: Results: Conclusion: To evaluate the feasibility of using supersonic shear-wave elastography (SSWE) in children and normal values of liver stiffness with the use of control patients of different ages (from neonates to teenagers) and the diagnostic accuracy of supersonic shear wave elastography for assessing liver fibrosis by using the histologic scoring system as the reference method in patients with liver disease, with a special concern for early stages of fibrosis. The institutional review board approved this prospective study. Informed consent was obtained from parents and children older than 7 years. First, 51 healthy children (from neonate to 15 years) were analyzed as the control group, and univariate and multivariate comparisons were performed to study the effect of age, transducer, breathing condition, probe, and position on elasticity values. Next, 45 children (from 1 month to 17.2 years old) who underwent liver biopsy were analyzed. SSWE measurements were obtained in the same region of the liver as the biopsy specimens. Biopsy specimens were reviewed in a blinded maer by a pathologist with the use of METAVIR criteria. The areas under the receiver operating characteristics curve (AUCs) were calculated for patients with fibrosis stage F0 versus those with stage F1 F2, F2 or higher, F3 or higher, and F4 or higher. A successful rate of SSWE measurement was 100% in 96 patients, including neonates. Liver stiffness values were significantly higher when an SC6-1 probe (Aixplorer; SuperSonic Imagine SA, Aix-enProvence, France) was used than when an SL15-4 probe (Aixplorer) was used (mean 6 standard deviation, 6.94 kpa vs 5.96 kpa ; P =.006). There was no influence of sex, the location of measurement, or respiratory status on liver elasticity values (P =.41.93), although the power to detect such a difference was low. According to the degree of liver fibrosis at liver biopsy, 88.5% 96.8% of patients were correctly classified, with AUCs of (95% confidence interval [CI]: 0.8, 1.0). The AUC for patients with stage F0 versus stage F1 F2 was 0.93 (95% CI: 0.87, 0.99). SSWE allows accurate assessment of liver fibrosis, even in children with early stage (F1 F2) disease, and the choice of transducer influences liver stiffness values. q RSNA, 2015 Online supplemental material is available for this article. 554 radiology.rsna.org n Radiology: Volume 278: Number 2 February 2016

2 Although liver biopsy is still considered the reference standard for diagnosing and staging liver fibrosis in children, its limitations (eg, invasiveness and interobserver and sampling variability) are well known (1,2). The inapplicability of repeating biopsy in clinical practice prevents easy longitudinal evaluation. Among the noninvasive methods of measuring liver stiffness that is supposed to mainly result from fibrosis, quantitative elastography methods are based on the estimation of shear-wave velocity (measured in m/sec) in the liver because it is related to the stiffness of the tissue (Young modulus, measured in kpa). Several clinical ultrasonographic Advances in Knowledge Liver stiffness measurements obtained with supersonic shearwave elastography (SSWE) had a success rate of 100% in 96 children, including neonates; sex, the location of measurement, and respiratory status (eg, apnea versus free-breathing) do not seem to influence liver stiffness measurements (P =.41 to P =.93). Liver stiffness values were significantly higher when a curvilinear 6-1 MHz transducer was used compared with a linear 15-4 MHz transducer; for the SL15-4 probe (Aixplorer; SuperSonic Imagine SA, Aix-enProvence, France), liver stiffness values in healthy children were 5.96 kpa , and for the SC6-1 probe (Aixplorer), they were 6.94 kpa for (P =.006). SSWE had high diagnostic accuracy for differentiating lower stages of fibrosis (F0 and F1) from higher stages (F2 F4), with an area under the receiver operating characteristics curve (AUC) of 0.98 (95% confidence interval [CI]: 0.95, 1.0), and liver with no fibrosis from mild and moderate fibrosis (stage F1 F2), with an AUC of 0.93 (95% CI: 0.87, 0.99). systems are currently available for quantitative elastography measurements in clinical practice, such as transient elastography, acoustic radiation force impulse (ARFI), point shear-wave elastography, and supersonic shear wave elastography (SSWE). Whereas a large number of studies consider the diagnostic accuracy of transient elastography and ARFI for detecting liver fibrosis with histologic correlations in adults, only six pediatric series that used transient elastography or ARFI and showed good diagnostic accuracy for the detection of mild and severe liver fibrosis are published (Table E1 [online]) (3 9). However, transient elastography has some limitations in young children, with some authors reporting technical failure in 17% of children younger than 2 years old and others reporting technical failure in 10% of children of any age, and the possibilities of obtaining intercostal measurements in pediatric liver recipients who underwent left lobe grafting are limited (6,10,11). There are few studies on SSWE, and only one concerns children (10 21). Most of them focus on the assessment of substantial fibrosis and cirrhosis, which are usually diagnosed at physical examination, and few are on the detection of early stage liver fibrosis, despite its role as a predictor of disease outcomes and its impact on therapy and follow-up (7,12). The purpose of the present study was to evaluate the feasibility of performing SSWE and determine normal values of liver stiffness in children by using control patients of different ages (from neonates to teenagers) and assess its diagnostic accuracy in assessing Implications for Patient Care SSWE can be used to noninvasively detect and stage liver fibrosis in children (including infants) with diffuse liver disease. Liver stiffness values should be interpreted taking into account the probe used to obtain measurements and underlying pathologic conditions. liver fibrosis by using the histologic scoring system as the reference method in patients with liver disease, with special concern paid to those with early stages of liver fibrosis. Materials and Methods The patent holder for real-time shear wave elastography is Institut Langevin Ondes et Images (Paris, France), where author M.T. is an employee. M.T. is a cofounder and shareholder of SuperSonic Imagine (Aix-en-Provence, France). J.L.G. is a consultant for SuperSonic Imagine (Aix-en-Provence, France). S.F.A. gave lectures about applications of ultrafast imaging in children for SuperSonic Imagine. SuperSonic Imagine did not provide any type of support for this study. Data and information submitted for publication were controlled by authors who do not have any financial interest in the Institut Langevin Ondes et Images or SuperSonic Imagine (S.F.A., L.C., G.A., E.G., B.D., M.F., D.P., J.M.C.). Study Population The study protocol was approved by the institutional review board. Informed consent was obtained from both parents and children older than 7 years. Two groups of patients were Published online before print /radiol Content codes: Radiology 2016; 278: Abbreviations: ARFI = acoustic radiation force impulse AUC = area under the receiver operating characteristics curve SSWE = supersonic shear-wave elastography Author contributions: Guarantors of integrity of entire study, S.F.A., L.C., D.P., J.L.G.; study concepts/study design or data acquisition or data analysis/interpretation, all authors; manuscript drafting or manuscript revision for important intellectual content, all authors; approval of final version of submitted manuscript, all authors; agrees to ensure any questions related to the work are appropriately resolved, all authors; literature research, S.F.A., E.G., M.F., J.L.G., J.M.C., M.T.; clinical studies, S.F.A., L.C., E.G., M.F., D.P.; statistical analysis, G.A., B.D., J.L.G.; and manuscript editing, S.F.A., E.G., M.F., D.P., J.L.G., M.T. Conflicts of interest are listed at the end of this article. Radiology: Volume 278: Number 2 February 2016 n radiology.rsna.org 555

3 prospectively included: a control group of children with no medical history suggestive of liver disorder who were referred for abdominal or renal ultrasonography (US) (group 1) and a second group of children with liver disease who underwent liver biopsy within 4 months from the time SSWE was performed (group 2). In group 1, 51 children (26 girls and 25 boys) were included and arbitrarily divided into the following five subgroups according to age: premature newborns, less than 1 year old, 1 5 years old, 5 10 years old, and more than 10 years old. The patients were referred for abdominal US because of the presence of urogenital disorders (n = 17); abdominal pain (n = 12); premature birth (n = 7); neurologic disorders (n = 5); genital disorders (n = 4); abdominal trauma with no liver injury (n = 3); sickle cell disease (n = 2); and bronchiolitis, joint pain, and fever (n = 1). Forty-seven children were included in group 2, and two were excluded because biopsy specimens did not allow reliable histologic evaluation of fibrosis. Thus, the data from 45 patients were analyzed. There were 26 girls and 19 boys, with a median age of 2 years (range, 1 month to 17 years). Indications for liver biopsy were liver transplantation in 18 patients (mean age, 7 years [standard deviation]; range, 1 17 years), 10 of whom had an interval of less than 45 days after transplantation and eight of whom had an interval of more than 45 days after transplantation; biliary atresia in 11 patients (mean age, 0.5 year 6 0.4; range, years); cholestasis of unknown origin in eight patients (mean age, 0.6 year 6 0.5; range, years); and other causes (portal obstruction with cavernoma in three, progressive familial intrahepatic cholestasis type 2 in two, sclerosing cholangitis in two, and acute leukemia in one) in eight patients (mean age, 6.7 years 6 3.3; range, years). The median time between SSWE and liver biopsy was 2 days (range, days). Twenty-four children underwent SSWE and liver biopsy on the same day. Two patients underwent repeated Figure 1 Figure 1: SSWE measurements in a 2-year-old boy from the control group. US images obtained with a SC6-1 transducer show a stiffness color map (top), which is homogeneous, with blue areas that correspond with low values of liver stiffness on the color scale, and SSWE measurements in the regions of interest (bottom), which confirm that liver elasticity values are normal. liver biopsies (two and three, respectively); only the first biopsy was considered for this study. Liver Stiffness Measurements All patients were evaluated by one radiologist (S.F.A., with 12 years of experience, or L.C., with 2 years of experience) who was trained to perform SSWE. SSWE was performed in the US room or the incubator for premature newborns. A complete evaluation of the liver was performed by using B-mode and Doppler imaging before obtaining SSWE measurements. Children underwent imaging in the supine position with the right arm raised to widen the intercostal spaces. Liver stiffness measurements were obtained with SSWE (Aixplorer; SuperSonic Imagine SA, Aix-en-Provence, France). SSWE is based on the velocity estimation of a shear wave, which is generated by the radiation force of a focused beam (13). SSWE measurements were obtained by using two different transducers: a high frequency linear transducer SL15-4 (4 15 MHz) and a low frequency convex abdominal transducer SC6-1 (1 6 MHz). The transducer was positioned in the intercostal space and epigastric area to assess the right and left (or left graft) sides of the liver, respectively, and a large liver area was selected with B-mode imaging, avoiding rib attenuation and vessels. An SSWE sequence measurement was then launched by using a stiffness color map with a fixed size ( cm with an SL15-4 transducer and cm with an SC6-1 transducer) (Figs 1, 2). Care was taken to not include large 556 radiology.rsna.org n Radiology: Volume 278: Number 2 February 2016

4 Figure 2 Figure 2: SSWE measurements obtained with a SL15-4 transducer in a 4-month-old baby with biliary atresia and a METAVIR score of F4 at liver biopsy. US images show the stiffness color map (top), which is heterogeneous, with yellow and red areas that correspond with high elasticity values on the color scale, and SSWE measurements in the regions of interest (bottom), which are kpa. blood vessels or biliary structures in the regions of interest. Measurements were classified as failed when no or weak signals were obtained in the SSWE box for all acquisitions. Group 1: feasibility study in healthy children. We evaluated the number of technical failures in obtaining liver stiffness measurements and studied the influence of the location of the probe (inside or outside the intercostal space), breathing condition (eg, free breathing and apnea), and transducer (ie, SL15-4 and SC6-1) on liver stiffness measurements by obtaining combinations of measurement conditions in each patient according to his or her ability to breathhold and age. Combinations of measurement conditions consisted of one transducer with one respiratory condition and one probe location. Because of technical limitations related to the size of the children, the SL15-4 transducer was used in all patients except those who were older than 10 years, and the SC6-1 transducer was used in all patients except neonates. For example, because neonates caot breath hold and are too small for the SC6-1 transducer, they only underwent two series of liver stiffness measurements: one with free breathing, an SL15-4 probe, and an intercostal transducer position and the other with free breathing, an SL15-4 probe, and an epigastric transducer position. The combinations of conditions of liver stiffness measurements according to age, the number of patients per age group, and the number of measurements per patient and per age group are summarized in Table E2 (online). For each combination of measurement conditions (type of transducer, breathing condition, and transducer position), at least three consecutive different stiffness color maps were obtained with two to three regions of interest selected, resulting in 10 liver stiffness measurements. A total of 1740 liver stiffness measurements were obtained, which corresponded to 174 combinations of measurement conditions (Table E2 [online]). All SSWE measurements were obtained in less than 10 minutes. Group 2: children who underwent liver biopsy. The choice of transducer was determined on the basis of the patient s age. When possible, 10 measurements were obtained in the intercostal position (right lobe). When an intercostal position was not possible, an epigastric position (left lobe graft liver) was used. If possible, SSWE was performed during a short breath hold (, 3 seconds). Otherwise, the acquisitions were made during normal, gentle breathing. Liver Biopsy and Histologic Studies (Group 2) Liver biopsy was performed percutaneously with a 16- or 18-gauge needle (Hepafix; Braun, Melsungen, Germany) in 30 patients, surgically in 11 patients, in an explanted liver after transplantation in four patients, and with transjugular access with a transjugular liver biopsy set (Laboratories 100; Cook Medical, Bloomington, Ind) in two patients. Biopsy specimens were fixed in 10% buffered formalin and embedded in paraffin. The following stains for Radiology: Volume 278: Number 2 February 2016 n radiology.rsna.org 557

5 coective tissues were routinely used: Masson trichrome, Picrosirius red, and reticulin (Gordon Sweet). The length of each specimen was recorded. At least 10 portal tracts were required for fibrosis staging. The mean length of biopsy specimens was 11.2 mm (range, 6 24 mm). All needle biopsy specimens except two had at least 10 portal tracts (mean, 15.1 portal tracts per biopsy specimen). The two patients with fibrotic liver microbiopsy specimens were excluded from the study. Specimens were analyzed by a single pathologist (M.F., with 30 years of experience) who was blinded to SSWE results but not to patients clinical and biologic data. Liver fibrosis was evaluated semiquantitatively according to the METAVIR scoring system, in which F0 = absent, F1 = stellate enlargement of portal tract without septa formation, F2 = enlargement of portal tract with rare septa formation and intact architecture, F3 = numerous septa but no obvious cirrhosis, and F4 = cirrhosis. Statistical Analysis Liver elasticity in group 1. The mean of the 10 measurements obtained with one combination of conditions of measurements in each patient was calculated and used for statistical analysis. The mean values and standard deviation of elasticity according to age, sex, type of transducer, transducer location, and breathing pattern were calculated. A multivariate linear regression model that included patient age and sex, transducer type, and type of respiration was used to study the effect of age, sex, transducer type and type of respiration on elasticity. Comparisons between age classes, sex, transducer type, and respiration were made by Wald tests. Relationship between liver elasticity and degree of fibrosis. Children in group 1 (except premature babies) were added to patients with a score of F0. To compare values obtained with the same transducer in children who presented with liver disorders, the elasticity values used were chosen according to patients age. For infants younger than 1 year, the mean of the elasticity Table 1 Mean Elasticity Values and P Values For Differences between Sex, Location of the Probe, Respiratory Conditions, and Type of Probe in the Control Group (Group 1) Characteristic No. of Measurements* Mean Elasticity P Value Sex.41 Male Female Transducer position.81 Intercostal Epigastric Respiratory condition.93 Free breathing Apnea Type of transducer.006 SL SC Note. A total of 174 combinations (one patient with one probe plus one respiratory condition plus one approach), with 10 measurements obtained per combination, were performed, for a total 1740 measurements. The mean of the 10 measurements was calculated and used for statistical analysis. P values refer to the difference between consecutive characteristics and were determined with the Wald test. * Data are the number of groups of 10 measurements obtained. Data are the mean (in kpa) plus or minus standard deviation. values were obtained with an SL15-4 transducer, and for children over 1 year old, the mean of the values was obtained with an SC6-1 transducer. The receiver operating characteristics curve was used to determine the optimal cut-off value for the presence of any (F0), mild to moderate (F1 F2), substantial ( F2), and severe ( F3) fibrosis and cirrhosis (F4). The diagnostic performance of the method was expressed as sensitivity, specificity, and correctly classified patients. Cut-off values that gave the best proportion of correctly classified patients were selected. The standard error for the area under the receiver operating characteristics curve (AUC) and asymptotic normal confidence intervals were calculated by using an algorithm suggested by DeLong, DeLong, and Clarke-Pearson (14). Correlation between elasticity and the degree of fibrosis was also evaluated with the Spearman coefficient test. For all analyses, P,.05 was considered to indicate a significant difference. Statistical analysis was performed with SAS version 9.3 (StataCorp, College Station, Tex) and Stata version 12.1 (StataCorp) software. Results Control Group 1 Measurements were possible in all patients. In all cases, elastograms were homogeneous and dark blue by using a 50-kPa stiffness scale (Fig 1). Mean liver elasticity value was 6.58 kpa in the whole group considering both transducers. A significant difference was found for mean elasticity between the two transducers: 5.96 kpa for SL15-4 and 6.94 kpa for SC6-1 (P =.006) (Table 1). There was no influence of sex, the location of the measurement, or respiratory status on mean liver elasticity values (Table 1). Considering all measurements, there was no significant difference between age groups (P =.11). The mean elasticity value exhibited a trend to increase with age when using the SL15-4 transducer (P =.05) (Fig 3, Table 2). Assessment of Liver Fibrosis (Group 2) The median values, range, and standard deviation of SSWE measurements obtained for each fibrosis stage (METAVIR score) are shown in Table 3. Sensitivity, specificity, AUC, positive and 558 radiology.rsna.org n Radiology: Volume 278: Number 2 February 2016

6 negative likelihood ratio, and the percentage of correctly classified patients are given according to the optimal cutoff values of SSWE for each METAVIR stage selected to favor the proportion of correctly classified patients (Table 4 ). AUCs were above 0.93 regardless of the transducer used and the degree of fibrosis, with 88.5% 96.8% of patients correctly classified. Cut-off values varied depending on the transducer used Figure 3 and were higher with the SC6-1 except in the case of severe fibrosis. In seven of the 45 evaluations, there was discordance between elasticity values and histologic type. Liver fibrosis was overestimated by SSWE measurement in five patients: a transplant recipient who presented with acute rejection, a patient with avascular liver related to hepatic vein obstruction, a patient with ischemic-necrotic lesions, a patient with biliary atresia, and a patient with acute leukemia and hepatic involvement. In two patients, liver fibrosis was underestimated by the measurement of SSWE, one of whom received a liver transplant and experienced acute rejection and the other presented with sclerosing cholangitis. When considering all transducers, the Spearman correlation coefficient between the SSWE values and liver fibrosis histologic score was high in the whole population (r = 0.83; P,.0001), but it varied depending on the type of pathologic condition present; it was low for biopsy performed less than 45 days after liver transplantation (r = 0.60; P =.05), moderate for biopsy performed more than 45 days after liver transplantation (r = 0.79; P =.006) and biliary atresia (r = 0.77; P =.005), and high for other type of cholestasis (r = 0.87; P =.005) and other diseases (r = 0.89; P =.003). Figure 3: Box plots show SSWE measurements for each age group. The bottom and top of the box are the first and third quartiles, and the band inside the box is the second quartile (the median). The top and bottom of the whiskers represent the highest and lowest data within 1.5 interquartile range of the upper and lower quartiles, respectively. Discussion In this study, we showed that SSWE was feasible in very young patients, including neonates, and that its diagnostic accuracy was good for assessing even early stages of fibrosis (F1 F2) that are usually missed at physical examination and with other noninvasive techniques. Another important finding is that the underlying pathologic condition and the transducer used to obtain measurements should be taken Table 2 Mean Elasticity Values according to Age in the Control Group (n = 51) Age Group, Mean, and P Value No. of Patients Age Mean Elasticity (kpa) Mean Range All Probes SL15-4 SC6-1 P Value Age group Preterm newborn* weeks weeks ,1 year months months years years years years years years years years years Mean P value Note. Unless otherwise indicated, data are mean plus or minus standard deviation. P values refer to the difference between measurements obtained with the transducers according to age group and were determined with the Wald test. * Gestational age. Radiology: Volume 278: Number 2 February 2016 n radiology.rsna.org 559

7 into account when interpreting SSWE measurements. SSWE appears to be useful in infants and children, especially when compared with transient elastography, the use of which appears to be limited in children less than 5 years old and in pediatric liver recipients with left lateral grafting and no intercostal access, with as much as 17% of measurements reported to be invalid in previous studies (6,10,11). In the control group, liver elasticity values seemed to be influenced by the transducer (SL15-4 vs SC6-1). Mean elasticity values were 5.96 kpa for the SL15-4 and 6.94 kpa for the SC6-1 (P =.006). Similar results that showed a significant difference in healthy liver stiffness measurements depending on the probe used have been reported in adults and children by using transient elastography with small (S+), medium (M+), and extra-large (XL+) probes (11,15,16). However, two pediatric studies that used ARFI did not demonstrate differences between the measurements obtained with two transducers (4C1 and 9L4) except in neonates (1 month old) (17,18). This may be related to the use of a velocity scale rather than an elasticity scale. Therefore, tissue stiffness is displayed with a compressed dynamic because the speed of the shear wave is proportional to the square root of the elastic shear modulus, c s = (E/3r), which may hide tiny differences between measurements. The influence of the transducer on measurements is related to the relationship between the viscoelastic properties of the tissues and the frequency of the transducer, as well as the broadband characteristics of the mechanical excitation generated with the ultrasonic radiation force of the shear wave elastography mode. In a few other pediatric studies that reported on transient elastography or ARFI techniques and used different transducers, there is no analysis of the results according to the transducer used (9,10). These findings are important because they show that the stiffness value should be interpreted according to the transducer frequency and that same transducer should be Table 3 Median, Range, Standard Deviation, and No. of Patients for Each Fibrosis Stage Using METAVIR Score on Liver Biopsy in Group 2 Characteristic METAVIR Stage F0 F1 F2 F3 F4 Median (kpa) Range (kpa) Standard deviation No. of patients Table 4 Clinical Performance of SSWE by Using Optimal Measurement Cut-Off Values according to the Probe Used Probe All SL15-4 SC6-1 Characteristic METAVIR Score F0 vs F1 F2 No. biopsy specimens 48 vs vs 9 33 vs 18 Cut-off value (kpa) AUC % CI 0.86, , , 0.99 Sensitivity (%) Specificity (%) Correctly classified (%) METAVIR Score F2 No. biopsy specimens 67 vs vs vs 12 Cut-off value (kpa) AUC % CI 0.95, , , 1.0 Sensitivity (%) Specificity (%) Correctly classified (%) METAVIR Score F3 No. biopsy specimens 78 vs vs 6 51 vs 8 Cut-off value (kpa) ,602 AUC % CI 0.94, , , 1.0 Sensitivity (%) Specificity (%) Correctly classified (%) METAVIR Score F4 No. biopsy specimens 82 vs 7 26 vs 4 59 vs 3 Cut-off value (kpa) AUC % CI 0.95, 1.0 Sensitivity (%) 85.7 Specificity (%) 97.7 Correctly classified (%) 96.6 Note. Patients in group 1 (excluding preterm babies) were considered to have a METAVIR score of F0 and were added to the subset of those with a score of F0 in group 2 to add statistical power to the study. CI = confidence interval. 560 radiology.rsna.org n Radiology: Volume 278: Number 2 February 2016

8 used for follow-up evaluation to allow comparison. SSWE allowed for noninvasive assessment of liver fibrosis, with 88.5% 96.8% of infants and children correctly classified according to their stage of fibrosis, even those who presented with mild or moderate fibrosis (stage F1 F2). Although a diagnosis of substantial fibrosis (stage F2 or greater) has been regarded as an important target in most studies of adults with viral hepatitis infection, the detection of early stages (F1 and F2) of fibrosis is important in children because minor liver fibrosis is often a predictor of disease outcomes and can affect therapy and follow-up (6,7,19,20). In our study, the detection of low stages of fibrosis (stage F1 F2) when compared with healthy or stage F0 livers was very good, with AUC of 0.90 for the SC6-1 transducer and 0.99 for the SL15-4 transducer and 88.9% of patients appropriately classified with the SC6-1 transducer and 95.8% of patients appropriately classified with the SL15-4 transducer. When comparing our results with those in the pediatric literature using different US elastographic techniques, only one study that evaluated the accuracy of transient elastography for identifying fibrosis in a homogeneous group of teenagers with nonalcoholic steatohepatitis had better results for diagnosing any, substantial, or severe fibrosis (Table E1 [online]) (3,5,6,8,9,21). Cut-off values for staging liver fibrosis vary among studies that used SSWE. Ferraioli et al (22) Tutar et al (21), and our study reported cut-off values of 7.1 kpa versus 10.4 kpa versus 12, 1 kpa for stage F2, respectively. These variations may be related to the type of liver disease. For example, steatosis significantly increased mean shear wave elastography values and was considered a confounding factor in assessing fibrosis in patients with nonalcoholic steatohepatitis (21,22). Other underlying disorders influence the diagnostic accuracy of liver stiffness measurements, such as inflammation, venous obstruction, and cholestasis. Thus, the choice of cut-off values should be assessed according to the type of pathologic condition present (6,23). Our study has several limitations. First, we did not perform any biologic liver tests in the control group to assess for liver disease. Second, the choice of the METAVIR scoring system can be questioned because it is mainly used in patients with chronic viral hepatitis C infection, but it is widely used in pediatric patients with liver disease. However, it is based on architectural modifications and is not perfectly adapted to the various types of pediatric liver conditions. The use of computer-assisted digital analysis of Sirius red stained histologic sections in future studies will help quantify liver fibrosis of any cause at histologic analysis (24). We did not assess interobserver reproducibility. In conclusion, our study showed the potential of SSWE techniques to detect liver fibrosis, especially early stages that are challenging for clinicians and with other elastographic techniques, in the pediatric population. Further studies in larger homogeneous groups of patients with pathologic conditions and with correlation to liver biopsy with quantitative morphometric evaluation of fibrosis are required to precisely establish the place of this technique in treating patients. Disclosures of Conflicts of Interest: S.F.A. Activities related to the present article: disclosed no relevant activities. Activities not related to the present article: payment for lectures from SuperSonic Imagine. Other activities: disclosed no relevant activities. L.C. disclosed no relevant activities. E.G. disclosed no relevant activities. G.A. disclosed no relevant activities. M.F. disclosed no relevant activities. B.D. disclosed no relevant activities. D.P. disclosed no relevant activities. J.L.G. Activities related to the present article: consultant for SuperSonic Imagine. Activities not related to the present article: consultant for SuperSonic Imagine. Other activities: disclosed no relevant activities. J.M.C. disclosed no relevant activities. M.T. Activities related to the present article: disclosed no relevant activities. Activities not related to the present article: co-founder, shareholder, and consultant for SuperSonic Imagine. Other activities: disclosed no relevant activities. References 1. Bedossa P, Dargère D, Paradis V. Sampling variability of liver fibrosis in chronic hepatitis C. Hepatology 2003;38(6): Cohen MB, A-Kader HH, Lambers D, Heubi JE. Complications of percutaneous liver bi- opsy in children. Gastroenterology 1992; 102(2): de Lédinghen V, Le Bail B, Rebouissoux L, et al. Liver stiffness measurement in children using FibroScan: feasibility study and comparison with Fibrotest, aspartate transaminase to platelets ratio index, and liver biopsy. J Pediatr Gastroenterol Nutr 2007;45(4): Eiler J, Kleinholderma U, Albers D, et al. Standard value of ultrasound elastography using acoustic radiation force impulse imaging (ARFI) in healthy liver tissue of children and adolescents. Ultraschall Med 2012;33(5): Noruegas MJ, Matos H, Gonçalves I, Cipriano MA, Sanches C. Acoustic radiation force impulse-imaging in the assessment of liver fibrosis in children. Pediatr Radiol 2012; 42(2): Fitzpatrick E, Quaglia A, Vimalesvaran S, Basso MS, Dhawan A. Transient elastography is a useful noninvasive tool for the evaluation of fibrosis in paediatric chronic liver disease. J Pediatr Gastroenterol Nutr 2013;56(1): Nobili V, Monti L, Alisi A, Lo Zupone C, Pietrobattista A, Tomà P. Transient elastography for assessment of fibrosis in paediatric liver disease. Pediatr Radiol 2011;41(10): Iacobellis A, Marcellini M, Andriulli A, et al. Non invasive evaluation of liver fibrosis in paediatric patients with nonalcoholic steatohepatitis. World J Gastroenterol 2006;12(48): Hanquinet S, Rougemont AL, Courvoisier D, et al. Acoustic radiation force impulse (ARFI) elastography for the noninvasive diagnosis of liver fibrosis in children. Pediatr Radiol 2013;43(5): Engelma G, Gebhardt C, Weing D, et al. Feasibility study and control values of transient elastography in healthy children. Eur J Pediatr 2012;171(2): Goldschmidt I, Streckenbach C, Dingema C, et al. Application and limitations of transient liver elastography in children. J Pediatr Gastroenterol Nutr 2013;57(1): Martínez SM, Crespo G, Navasa M, Forns X. Noninvasive assessment of liver fibrosis. Hepatology 2011;53(1): Bercoff J, Tanter M, Fink M. Supersonic shear imaging: a new technique for soft tissue elasticity mapping. IEEE Trans Ultrason Ferroelectr Freq Control 2004;51(4): DeLong ER, DeLong DM, Clarke-Pearson DL. Comparing the areas under two or Radiology: Volume 278: Number 2 February 2016 n radiology.rsna.org 561

9 more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 1988;44(3): Ferraioli G, Lissandrin R, Zicchetti M, Filice C. Assessment of liver stiffness with transient elastography by using S and M probes in healthy children. Eur J Pediatr 2012;171(9):1415; author reply Myers RP, Pomier-Layrargues G, Kirsch R, et al. Feasibility and diagnostic performance of the FibroScan XL probe for liver stiffness measurement in overweight and obese patients. Hepatology 2012;55(1): Hanquinet S, Courvoisier D, Kanavaki A, Dhouib A, Anooshiravani M. Acoustic radiation force impulse imaging-normal values of liver stiffness in healthy children. Pediatr Radiol 2013;43(5): Fontanilla T, Cañas T, Macia A, et al. Normal values of liver shear wave velocity in healthy children assessed by acoustic radiation force impulse imaging using a convex probe and a linear probe. Ultrasound Med Biol 2014;40(3): Pereira TN, Walsh MJ, Lewindon PJ, Ramm GA. Paediatric cholestatic liver disease: Diagnosis, assessment of disease progression and mechanisms of fibrogenesis. World J Gastrointest Pathophysiol 2010;1(2): Castera L. Noninvasive methods to assess liver disease in patients with hepatitis B or C. Gastroenterology 2012;142(6): , e Tutar O, Beşer OF, Adaletli I, et al. Shear wave elastography in the evaluation of liver fibrosis in children. J Pediatr Gastroenterol Nutr 2014;58(6): Ferraioli G, Tinelli C, Dal Bello B, et al. Accuracy of real-time shear wave elastography for assessing liver fibrosis in chronic hepatitis C: a pilot study. Hepatology 2012;56(6): Gae-Carrié N, Ziol M, de Ledinghen V, et al. Accuracy of liver stiffness measurement for the diagnosis of cirrhosis in patients with chronic liver diseases. Hepatology 2006; 44(6): Germani G, Burroughs AK, Dhillon AP. The relationship between liver disease stage and liver fibrosis: a tangled web. Histopathology 2010;57(6): radiology.rsna.org n Radiology: Volume 278: Number 2 February 2016

PROPOSTA DI UN NUOVO ALGORIMO PER LA DIAGNOSI ECOGRAFICA DELLE MALATTIE CRONICHE DEL FEGATO

PROPOSTA DI UN NUOVO ALGORIMO PER LA DIAGNOSI ECOGRAFICA DELLE MALATTIE CRONICHE DEL FEGATO PROPOSTA DI UN NUOVO ALGORIMO PER LA DIAGNOSI ECOGRAFICA DELLE MALATTIE CRONICHE DEL FEGATO A. Giorgio Direttore del servizio di Ecografia Interventistica Istituto Clinico S.Rita -IRCCS -Atripalda (Avellino)

More information

JPGN Journal of Pediatric Gastroenterology and Nutrition Publish Ahead of Print

JPGN Journal of Pediatric Gastroenterology and Nutrition Publish Ahead of Print JPGN Journal of Pediatric Gastroenterology and Nutrition Publish Ahead of Print DOI : 10.1097/MPG.0000000000000489 Points to be considered when Applying FibroScan S Probe for Children with Biliary Atresia

More information

Original papers Med Ultrason 2013, Vol. 15, no. 4,

Original papers Med Ultrason 2013, Vol. 15, no. 4, Original papers Med Ultrason 2013, Vol. 15, no. 4, 268-272 DOI: How many measurements are needed for liver stiffness assessment by 2D-Shear Wave Elastography (2D-SWE) and which value should be used: the

More information

Shear Wave Elastography In Characterization Of Liver Tumours

Shear Wave Elastography In Characterization Of Liver Tumours Shear Wave Elastography In Characterization Of Liver Tumours Poster No.: C-1921 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. L. Choong 1, B. J. J. Abdullah 1, G. Kumar 2, C. H. Yeong 1, K. L.

More information

Research Elastography: Liver

Research Elastography: Liver Research Elastography: Liver Giovanna Ferraioli EFSUMB Ultrasound Learning Center Ultrasound Unit - Infectious Diseases Dept. Fondazione IRCCS Policlinico S. Matteo Medical School University of Pavia,

More information

Transient elastography in chronic viral liver diseases

Transient elastography in chronic viral liver diseases 4 th AISF POST-MEETING COURSE Roma, 26 Febbraio 2011 Transient elastography in chronic viral liver diseases CRISTINA RIGAMONTI, M.D., Ph.D. Transient elastography (TE): a rapid, non-invasive technique

More information

Evaluation of Liver Fibrosis using ShearWave Elastography on Aixplorer

Evaluation of Liver Fibrosis using ShearWave Elastography on Aixplorer Evaluation of Liver Fibrosis using ShearWave Elastography on Aixplorer SSID02892-01 November 2012 Evaluation of Liver Fibrosis using SWE on Aixplorer Introduction From a clinical standpoint, staging liver

More information

Motion effects on the measurement of stiffness on ultrasound shear wave elastography: a moving liver fibrosis phantom study

Motion effects on the measurement of stiffness on ultrasound shear wave elastography: a moving liver fibrosis phantom study Original papers Med Ultrason 2018, Vol. 20, no. 1, 14-20 DOI: 10.11152/mu-1138 Motion effects on the measurement of stiffness on ultrasound shear wave elastography: a moving liver fibrosis phantom study

More information

Evaluation of liver and spleen stiffness using a ultrasound guided method: Accuracy of ARFI(R) measurements in liver disease patients

Evaluation of liver and spleen stiffness using a ultrasound guided method: Accuracy of ARFI(R) measurements in liver disease patients Evaluation of liver and spleen stiffness using a ultrasound guided method: Accuracy of ARFI(R) measurements in liver disease patients Poster No.: C-3242 Congress: ECR 2010 Type: Topic: Authors: Keywords:

More information

The role of non-invasivemethods in evaluating liver fibrosis of patients with non-alcoholic steatohepatitis

The role of non-invasivemethods in evaluating liver fibrosis of patients with non-alcoholic steatohepatitis The role of non-invasivemethods in evaluating liver fibrosis of patients with non-alcoholic steatohepatitis Objectives: Liver biopsy is the gold standard for diagnosing the extent of fibrosis in NAFLD/NASH;

More information

Gastrointestinal Imaging Original Research

Gastrointestinal Imaging Original Research Gastrointestinal Imaging Original Research Beland et al. Shear-Wave Elastography in Liver Fibrosis Gastrointestinal Imaging Original Research Michael D. Beland 1 Sanford F. Brown 1 Jason T. Machan 2 Ross

More information

Ultrasound elastography in abdominal imaging

Ultrasound elastography in abdominal imaging Ultrasound elastography in abdominal imaging Olivier ucidarme Pitié Salpêtrière - UPMC Paris France Ultraschall Med. 013 Apr;34():169-84 Ultraschall Med. 013 Jun;34(3):38-53. US elastography Ultraschall

More information

Comparative study between two point Shear Wave Elastographic. Acoustic Radiation Force Impulse (ARFI) elastography

Comparative study between two point Shear Wave Elastographic. Acoustic Radiation Force Impulse (ARFI) elastography Original papers Med Ultrason 2014, Vol. 16, no. 4, 309-314 DOI: Comparative study between two point Shear Wave Elastographic techniques: Acoustic Radiation Force Impulse (ARFI) elastography and ElastPQ

More information

Evaluation of liver and spleen stiffness using a ultrasound guided method: Accuracy of ARFI(R) measurements in liver disease patients

Evaluation of liver and spleen stiffness using a ultrasound guided method: Accuracy of ARFI(R) measurements in liver disease patients Evaluation of liver and spleen stiffness using a ultrasound guided method: Accuracy of ARFI(R) measurements in liver disease patients Poster No.: C-3242 Congress: ECR 2010 Type: Topic: Authors: Keywords:

More information

Byung Ihn Choi, M.D. Department of Radiology Seoul National University Hospital

Byung Ihn Choi, M.D. Department of Radiology Seoul National University Hospital Byung Ihn Choi, M.D. Department of Radiology Seoul National University Hospital CEUS & US Elastography : Contents CEUS Introduction Contrast agents & imaging Clinical application US Video WS Summary US

More information

Improved Hepatic Fibrosis Grading Using Point Shear Wave Elastography and Machine Learning

Improved Hepatic Fibrosis Grading Using Point Shear Wave Elastography and Machine Learning Improved Hepatic Fibrosis Grading Using Point Shear Wave Elastography and Machine Learning Presenter: Hersh Sagreiya 1, M.D. Authors: Alireza Akhbardeh 1, Ph.D., Isabelle Durot 1, M.D., Carlo Filice 2,

More information

Transient elastography the state of the art

Transient elastography the state of the art Transient elastography the state of the art Laurent CASTERA, MD PhD Department of Hepatology, Hôpital Beaujon, Clichy Université Paris-7, France White Nights of Hepatology, St Petersburg, Russia, june

More information

Original Article. Abstract. Background

Original Article. Abstract. Background E2 Original Article THIEME Evaluation of Acoustic Radiation Force Impulse Imaging (ARFI) for the Determination of Liver Stiffness Using Transient Elastography as a Reference in Children Authors A. Sagir

More information

The role of ARFI and APRI in diagnosis of liver fibrosis on patients with common chronic liver diseases

The role of ARFI and APRI in diagnosis of liver fibrosis on patients with common chronic liver diseases RESEARCH ARTICLE The role of ARFI and APRI in diagnosis of liver fibrosis on patients with common chronic liver diseases Objective: This study aimed to investigate the value of liver fibrosis assessment

More information

NON INVASIVE ASSESSMENT OF LIVER FIBROSIS : FIBROSCAN

NON INVASIVE ASSESSMENT OF LIVER FIBROSIS : FIBROSCAN NON INVASIVE ASSESSMENT OF LIVER FIBROSIS : FIBROSCAN M. Beaugrand Service d Hépatologied Hopital J. Verdier BONDY 93143 et Université Paris XIII MAINZ 21.09.2008 ASSESSMENT OF FIBROSIS : WHY? Management

More information

Liver elastography ultrasound cpt code

Liver elastography ultrasound cpt code Liver elastography ultrasound cpt code In a multi-center study, Barbero-Villares et al (2012) evaluated the presence of significant liver fibrosis by transient elastography (FibroScan) in inflammatory.

More information

Non-Invasive Testing for Liver Fibrosis

Non-Invasive Testing for Liver Fibrosis NORTHWEST AIDS EDUCATION AND TRAINING CENTER Non-Invasive Testing for Liver Fibrosis John Scott, MD, MSc Associate Professor, University of Washington Associate Clinic Director, Hep/Liver Clinic, Harborview

More information

Invasive. Sampling error. Interobserver variability. Nondynamic evaluation of

Invasive. Sampling error. Interobserver variability. Nondynamic evaluation of How to assess liver fibrosis Serum markers or FibroScan vs. liver biopsy? Laurent CASTERA & Pierre BEDOSSA Hôpital Beaujon, AP-HP, Clichy Université Paris-VII France 4 th Paris Hepatitis Conference, Paris,

More information

iu22 Liver Shear Wave ElastPQ

iu22 Liver Shear Wave ElastPQ iu22 Liver Shear Wave ElastPQ Clinical Case Study Lucy Wang Clinical Application Specialist ASEAN Case Study: History: 58-year-old male patient, hepatitis B virus (HBV) carrier, with non clinical symptoms

More information

Preliminary clinical experience with Shear Wave Dispersion Imaging for liver viscosity

Preliminary clinical experience with Shear Wave Dispersion Imaging for liver viscosity Preliminary clinical experience with Shear Wave Dispersion Imaging for liver viscosity Dr. Katsutoshi Sugimoto Department of Gastroenterology and Hepatology, Tokyo Medical University, Japan Introduction

More information

NONINVASIVE IMAGING METHODS FOR ASSESSMENT OF LIVER DAMAGE IN NASH

NONINVASIVE IMAGING METHODS FOR ASSESSMENT OF LIVER DAMAGE IN NASH NONINVASIVE IMAGING METHODS FOR ASSESSMENT OF LIVER DAMAGE IN NASH Bachir Taouli, MD Director of Body MRI and Cancer Imaging Program Department of Radiology / Translational and Molecular Imaging Institute

More information

Shear-wave elastography for breast masses: local shear wave speed (m/sec) versus Young modulus (kpa)

Shear-wave elastography for breast masses: local shear wave speed (m/sec) versus Young modulus (kpa) Shear-wave elastography for breast masses: local shear wave speed (m/sec) versus Young modulus (kpa) Ji Hyun Youk, Eun Ju Son, Ah Young Park, Jeong-Ah Kim Department of Radiology, Gangnam Severance Hospital,

More information

DISCLOSURES. This activity is jointly provided by Northwest Portland Area Indian Health Board and Cardea

DISCLOSURES. This activity is jointly provided by Northwest Portland Area Indian Health Board and Cardea DISCLOSURES This activity is jointly provided by Northwest Portland Area Indian Health Board and Cardea Cardea Services is approved as a provider of continuing nursing education by Montana Nurses Association,

More information

RSNA-QIBA SWS Profile What the Clinical Workflow Would look like and a Review of Biological Confounders. Manish Dhyani, MD Anthony Samir, MD MPH

RSNA-QIBA SWS Profile What the Clinical Workflow Would look like and a Review of Biological Confounders. Manish Dhyani, MD Anthony Samir, MD MPH RSNA-QIBA SWS Profile What the Clinical Workflow Would look like and a Review of Biological Confounders Manish Dhyani, MD Anthony Samir, MD MPH Background Liver fibrosis is the final common pathway for

More information

Clinical Medicine Review

Clinical Medicine Review Dighe M. et al. Clinical Medicine Review, 2018 Page 1 of 15 Clinical Medicine Review Published: July 09, 2018 Citation: Dighe M. et al. (2018) Review of ultrasound elastography techniques in the liver.

More information

Transient elastography in chronic liver diseases of other etiologies

Transient elastography in chronic liver diseases of other etiologies 4 Post Meeting A.I.S.F. Unmet Clinical Needs in Hepatology: New and upcoming diagnostic tools" Transient elastography in chronic liver diseases of other etiologies Dr. Vincenza Calvaruso Gastroenterologia

More information

Medical Review Approaches to the Diagnosis of Liver Fibrosis

Medical Review Approaches to the Diagnosis of Liver Fibrosis Medical Review Approaches to the Diagnosis of Liver Fibrosis Hiroko Iijima Department of Hepatobiliary and Pancreatic Disease Ultrasound Imaging Center, Hyogo College of Medicine Hiroko Iijima Department

More information

In vivo non invasive quantitative assessment of passive diastolic stiffness of infarcted myocardium using Shear Wave Imaging

In vivo non invasive quantitative assessment of passive diastolic stiffness of infarcted myocardium using Shear Wave Imaging Munich, August 28th 2012 European Society of Cardiology In vivo non invasive quantitative assessment of passive diastolic stiffness of infarcted myocardium using Shear Wave Imaging Mathieu Pernot 1, Wei-Ning

More information

Elastography in the. technically difficult patient. EPIQ ultrasound system. Ultrasound

Elastography in the. technically difficult patient. EPIQ ultrasound system. Ultrasound Ultrasound Elastography in the technically difficult patient EPIQ ultrasound system Chairman Department of Diagnostic Radiology Allegheny General Hospital Pittsburgh, PA, USA You can offer more information

More information

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association Noninvasive Techniques for the Evaluation and Monitoring of Page 1 of 38 Medical Policy An independent licensee of the Blue Cross Blue Shield Association Title: Noninvasive Techniques for the Evaluation

More information

Quantitative assessment of the elasticity values of liver with shear wave ultrasonographic elastography

Quantitative assessment of the elasticity values of liver with shear wave ultrasonographic elastography Indian J Med Res 137, May 2013, pp 911-915 Quantitative assessment of the elasticity values of liver with shear wave ultrasonographic elastography Kemal Arda, Nazan Ciledag *, Bilgin Kadri Arıbas, Elif

More information

POWERED BY. Innovation in liver disease management

POWERED BY. Innovation in liver disease management POWERED BY VCTE Innovation in liver disease management An intelligent solution to aid clinical diagnosis, FibroScan uses state of the art fibrosis and steatosis quantification with the most advanced non-invasive

More information

Noninvasive Diagnosis and Staging of Liver Disease. Naveen Gara, MD

Noninvasive Diagnosis and Staging of Liver Disease. Naveen Gara, MD Noninvasive Diagnosis and Staging of Liver Disease Naveen Gara, MD Outline Brief overview of the anatomy of liver Liver-related lab tests Chronic liver disease progression Estimation of liver fibrosis

More information

HEPATIC ELASTOGRAPHY AND FIB-4 SCORE VERSUS LIVER BIOPSY FOR ASSESSMENT OF LIVER FIBROSIS IN CHRONIC HCV PATIENTS

HEPATIC ELASTOGRAPHY AND FIB-4 SCORE VERSUS LIVER BIOPSY FOR ASSESSMENT OF LIVER FIBROSIS IN CHRONIC HCV PATIENTS HEPATIC ELASTOGRAPHY AND FIB-4 SCORE VERSUS LIVER BIOPSY FOR ASSESSMENT OF LIVER FIBROSIS IN CHRONIC HCV PATIENTS Essam A. Wahab*, Talaat Fathy** Sameh Saber*** and Hanan S. Ahmad**** Internal Medicine*,

More information

Does elastography change the indication to biopsy? IBDC

Does elastography change the indication to biopsy? IBDC Does elastography change the indication to biopsy? A LEXANDRA A THANASIOU, M D DEPARTMENT OF RADIOLOGY CURIE INSTITUTE PARIS, FRANCE IBDC Ultrasound Detected Cancers Physician-performed ultrasound increases

More information

ShearWave elastography in lymph nodes

ShearWave elastography in lymph nodes ShearWave elastography in lymph nodes Poster No.: B-0158 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Paper F. Houari, O. Lucidarme, J. Gabarre, F. Charlotte, C. Pellot- Barakat, M. Lefort,

More information

Non-Invasive Assessment of Liver Fibrosis. Patricia Slev, PhD University of Utah Department of Pathology

Non-Invasive Assessment of Liver Fibrosis. Patricia Slev, PhD University of Utah Department of Pathology Non-Invasive Assessment of Liver Fibrosis Patricia Slev, PhD University of Utah Department of Pathology Disclosure Patricia Slev has no relevant financial relationships to disclose. 2 Chronic Liver Disease

More information

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association Noninvasive Techniques for the Evaluation and Page 1 of 34 Medical Policy An independent licensee of the Blue Cross Blue Shield Association Title: Noninvasive Techniques for the Evaluation and Professional

More information

ULTRASOUND IN CHRONIC LIVER DISEASE

ULTRASOUND IN CHRONIC LIVER DISEASE ULTRASOUND IN CHRONIC LIVER DISEASE Prof. Ioan Sporea, MD, PhD Head of Department of Gastroenterology and Hepatology University of Medicine and Pharmacy, Timişoara, Romania WFUMB Center of Education (COE),

More information

Transient Elastography and Sonography for Prediction of Liver Fibrosis in Infants With Biliary Atresia

Transient Elastography and Sonography for Prediction of Liver Fibrosis in Infants With Biliary Atresia ORIGINAL RESEARCH Transient Elastography and Sonography for Prediction of Liver Fibrosis in Infants With Biliary Atresia Na-Young Shin, MD, Myung-Joon Kim, MD, Mi-Jung Lee, MD, Seok Joo Han, MD, Hong Koh,

More information

ABSTRACT INTRODUCTION. Jing-Hua Liu 1, Yu Zou 1, Wei Chang 2, Jun Wu 3, Yu Zou 4, Yu-Chen Xie 5, Yong-Ping Lu 6 and Jia Wei 1 *

ABSTRACT INTRODUCTION. Jing-Hua Liu 1, Yu Zou 1, Wei Chang 2, Jun Wu 3, Yu Zou 4, Yu-Chen Xie 5, Yong-Ping Lu 6 and Jia Wei 1 * REVISTA DE INVESTIGACIÓN CLÍNICA Contents available at PubMed www.clinicalandtranslationalinvestigation.com PERMANYER Rev Inves Clin. 2017;69:254-261 ORIGINAL ARTICLE Assessment of Liver Fibrosis Using

More information

Simplifying liver assessment in internal medicine

Simplifying liver assessment in internal medicine Ultrasound Customer story Simplifying liver assessment in internal medicine Philips Affiniti ultrasound for elastography and contrast-enhanced ultrasound (CEUS) Where Sonography Institute, Uster, Switzerland

More information

CONFOUNDING FACTORS affecting the performance of US elastography

CONFOUNDING FACTORS affecting the performance of US elastography Clinical Ultrasound in Hepatology: Training for Hepatologists UCL Institute for Liver and Digestive Health, Royal Free Hospital, London, UK June 2018 CONFOUNDING FACTORS affecting the performance of US

More information

Noninvasive Techniques for the Evaluation and Monitoring of Patients With Chronic Liver Disease

Noninvasive Techniques for the Evaluation and Monitoring of Patients With Chronic Liver Disease Noninvasive Techniques for the Evaluation and Monitoring of Patients With Chronic Liver Disease Policy Number: 2.04.41 Last Review: 2/2019 Origination: 2/2017 Next Review: 2/2020 Policy Blue Cross and

More information

Acoustic Radiation Force Based Imaging: An Overview

Acoustic Radiation Force Based Imaging: An Overview Acoustic Radiation Force Based Imaging: An Overview Kathy Nightingale, Michael Wang, Stephen Rosenzweig, Veronica Rotemberg, Samantha Lipman, Ned Rouze, Mark Palmeri Department of Biomedical Engineering

More information

shear wave elastography

shear wave elastography Ultrasound ElastQ Imaging Introducing next-generation shear wave elastography Philips ElastQ Imaging Vijay Shamdasani, PhD and Hua Xie, PhD, Philips Ultrasound Tissue stiffness has long provided important

More information

Medical technologies guidance Published: 23 September 2015 nice.org.uk/guidance/mtg27

Medical technologies guidance Published: 23 September 2015 nice.org.uk/guidance/mtg27 Virtual Touch Quantification to diagnose and monitor liver fibrosis in chronic hepatitis B and C Medical technologies guidance Published: 23 September 2015 nice.org.uk/guidance/mtg27 NICE 2017. All rights

More information

Clinical Policy Title: Liver elastography

Clinical Policy Title: Liver elastography Clinical Policy Title: Liver elastography Clinical Policy Number: CCP.1118 Effective Date: October 1, 2014 Initial Review Date: June 18, 2014 Most Recent Review Date: August 7, 2018 Next Review Date: August

More information

NONINVASIVE TECHNIQUES FOR EVALUATION AND MONITORING OF CHRONIC LIVER DISEASE

NONINVASIVE TECHNIQUES FOR EVALUATION AND MONITORING OF CHRONIC LIVER DISEASE LIVER DISEASE Non-Discrimination Statement and Multi-Language Interpreter Services information are located at the end of this document. Coverage for services, procedures, medical devices and drugs are

More information

Medical Policy Non-invasive Tests for Hepatic Fibrosis

Medical Policy Non-invasive Tests for Hepatic Fibrosis Medical Policy Non-invasive Tests for Hepatic Fibrosis Subject: Non-invasive Tests for Hepatic Fibrosis Background: Accurately diagnosing liver fibrosis and inflammatory activity are the most important

More information

MP Noninvasive Techniques for the Evaluation and Monitoring of Patients With Chronic Liver Disease

MP Noninvasive Techniques for the Evaluation and Monitoring of Patients With Chronic Liver Disease Medical Policy BCBSA Ref. Policy: 2.04.41 Last Review: 11/15/2018 Effective Date: 11/15/2018 Section: Medicine Related Policies 9.01.502 Experimental / Investigational Services DISCLAIMER Our medical policies

More information

Clinical Criteria for Hepatitis C (HCV) Therapy

Clinical Criteria for Hepatitis C (HCV) Therapy Clinical Criteria for Hepatitis C (HCV) Therapy Pre-Treatment Evaluation o Must have chronic hepatitis C and HCV genotype and sub-genotype documented; o Patients who have prior exposure to DAA therapy

More information

Noninvasive Techniques for the Evaluation and Monitoring of Patients With Chronic Liver Disease

Noninvasive Techniques for the Evaluation and Monitoring of Patients With Chronic Liver Disease Noninvasive Techniques for the Evaluation and Monitoring of Patients With Chronic Liver Disease Policy Number: 2.04.41 Last Review: 2/2018 Origination: 2/2017 Next Review: 2/2019 Policy Blue Cross and

More information

Diagnostic Accuracy of SuperSonic Shear Imaging for Staging of Liver Fibrosis

Diagnostic Accuracy of SuperSonic Shear Imaging for Staging of Liver Fibrosis ORIGINAL RESEARCH Diagnostic Accuracy of SuperSonic Shear Imaging for Staging of Liver Fibrosis A Meta-analysis Jin-Chun Feng, MD, Jun Li, MD, Xiang-Wei Wu, MD, PhD, Xin-Yu Peng, MD, PhD Received March

More information

As a result of the growing obesity epidemic,

As a result of the growing obesity epidemic, AMERICAN ASSOCIATION FOR THE STUDY OFLIVERD I S E ASES HEPATOLOGY, VOL. 63, NO. 6, 2016 STEATOHEPATITIS/METABOLIC LIVER DISEASE Liver Stiffness in Nonalcoholic Fatty Liver Disease: A Comparison of Supersonic

More information

MR Elastography of Liver

MR Elastography of Liver MR Elastography of Liver Sudhakar K. Venkatesh, MD, FRCR Professor of Radiology Mayo Clinic College of Medicine Consultant, Abdominal Division Radiology, Mayo Clinic Rochester, MN, USA 19 th May 2018 2018

More information

A Comparative Study of Shear-Wave Elastography and Strain Elastography on a Breast Phantom for Diagnosis of Tumor and Cyst

A Comparative Study of Shear-Wave Elastography and Strain Elastography on a Breast Phantom for Diagnosis of Tumor and Cyst ` Volume VOLUME 2 ISSUE No 3 A Comparative Study of Shear-Wave Elastography and Strain Elastography on a Breast Phantom for Diagnosis of Tumor and Cyst Mahdi Al-Qahtani 1, Eraj Humayun Mirza 2, Mubarak

More information

Ultrasound elastography of liver: How Radiologist can help.

Ultrasound elastography of liver: How Radiologist can help. 116 Review Article Ultrasound elastography of liver: How Radiologist can help. Maharjan S 1, Panta OB 2, Lohani B 1, Pathak YR 2, Ansari MA 1 1 Department of radiology and Imaging, Tribhuvan University

More information

Liver Stiffness Measurements Using Acoustic Radiation Force Impulse in Recipients of Living-Donor and Deceased- Donor Orthotopic Liver Transplant

Liver Stiffness Measurements Using Acoustic Radiation Force Impulse in Recipients of Living-Donor and Deceased- Donor Orthotopic Liver Transplant ArtIcle Liver Stiffness Measurements Using Acoustic Radiation Force Impulse in Recipients of Living-Donor and Deceased- Donor Orthotopic Liver Transplant Kemal Murat Haberal, 1 Hale Turnaoğlu, 1 Adnan

More information

Real-time elastography as a noninvasive assessment of liver fibrosis in chronic hepatitis C Egyptian patients: a prospective study

Real-time elastography as a noninvasive assessment of liver fibrosis in chronic hepatitis C Egyptian patients: a prospective study ORIGINAL ARTICLE Annals of Gastroenterology (2016) 29, 1-5 Real-time elastography as a noninvasive assessment of liver fibrosis in chronic hepatitis C Egyptian patients: a prospective study Lamiaa Mobarak

More information

1. Introduction. Correspondence should be addressed to Daniela Cabibi;

1. Introduction. Correspondence should be addressed to Daniela Cabibi; Analytical Cellular Pathology Volume 2015, Article ID 431750, 7 pages http://dx.doi.org/10.1155/2015/431750 Research Article Comparison of Histochemical Stainings in Evaluation of Liver Fibrosis and Correlation

More information

Pediatric Hepatobiliary, Pancreatic & Splenic US

Pediatric Hepatobiliary, Pancreatic & Splenic US Pediatric Hepatobiliary, Pancreatic & Splenic US Susan J. Back, MD Department of Radiology, The Children s Hospital of Philadelphia No Disclosures Objectives Normal Abnormal: cases and US advances Objectives

More information

LOGIQ E9 Shear Wave Elastography

LOGIQ E9 Shear Wave Elastography LOGIQ E9 Shear Wave Elastography Introduction Tissue stiffness is often related to underlying disease. For millennia, physicians have used palpation as a diagnostic tool to detect various ailments such

More information

*APHP: Hôpital Beaujon Professor P Bedossa, Dr F Degos, Professor P Marcellin, Professor

*APHP: Hôpital Beaujon Professor P Bedossa, Dr F Degos, Professor P Marcellin, Professor APPENDIX 1 LIST OF COINVESTIGATORS *APHP: Hôpital Beaujon Professor P Bedossa, Dr F Degos, Professor P Marcellin, Professor M Vidaud; Hôpital Cochin-Necker: Professor S Pol, Professor Ph Sogni, Professor

More information

pulse See the brain as never before. GE Healthcare s i g n a page 51

pulse See the brain as never before. GE Healthcare s i g n a page 51 GE Healthcare pulse s i g n a t h e m ag a z i n e o f m r s p r i n g 2 0 1 0 See the brain as never before. Non-contrast enhanced perfusion assessment brings neuro imaging one step closer to the future

More information

Assessment of Liver Stiffness by Transient Elastography in Diabetics with Fatty Liver A Single Center Cross Sectional observational Study

Assessment of Liver Stiffness by Transient Elastography in Diabetics with Fatty Liver A Single Center Cross Sectional observational Study IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861.Volume 16, Issue 6 Ver. IV (June. 2017), PP 49-53 www.iosrjournals.org Assessment of Liver Stiffness by Transient

More information

LOGIQ E9 Shear Wave Elastography

LOGIQ E9 Shear Wave Elastography LOGIQ E9 Shear Wave Elastography Introduction Tissue stiffness is often related to underlying disease. For millennia, physicians have used palpation as a diagnostic tool to detect various ailments such

More information

Functional liver imaging Bernard Van Beers

Functional liver imaging Bernard Van Beers Functional liver imaging Bernard Van Beers Department of Radiology Beaujon University Hospital Paris Nord Laboratory of Imaging biomarkers UMR 1149 Inserm - University Paris Diderot France Functional imaging

More information

Clinical Criteria for Hepatitis C (HCV) Therapy

Clinical Criteria for Hepatitis C (HCV) Therapy Clinical Criteria for Hepatitis C (HCV) Therapy Pre-Treatment Evaluation o Must have chronic hepatitis C and HCV genotype and sub-genotype documented; o Patients who have prior exposure to DAA therapy

More information

Quantitative Assessment of Normal Soft-Tissue Elasticity Using Shear-Wave Ultrasound Elastography

Quantitative Assessment of Normal Soft-Tissue Elasticity Using Shear-Wave Ultrasound Elastography Special Article Original Research Arda et al. Soft-Tissue Elasticity in Ultrasound Elastography Special Article Original Research JOURNAL CLUB Kemal Arda 1 Nazan Ciledag 1 Elif Aktas 1 Bilgin Kadri Arıbas

More information

How useful are ARFI elastography cut-off values proposed by metaanalysis for predicting the significant fibrosis and compensated liver cirrhosis?

How useful are ARFI elastography cut-off values proposed by metaanalysis for predicting the significant fibrosis and compensated liver cirrhosis? Original papers Med Ultrason 2015, Vol. 17, no. 2, 200-205 DOI: 10.11152/mu.2013.2066.172.arfi How useful are ARFI elastography cut-off values proposed by metaanalysis for predicting the significant fibrosis

More information

Hepatic Imaging: What Every Practitioner Should Know

Hepatic Imaging: What Every Practitioner Should Know Hepatic Imaging: What Every Practitioner Should Know Shuchi K. Rodgers, MD Section Chief, Abdominal Imaging Director of Ultrasound Department of Radiology Einstein Medical Center rodgerss@einstein.edu

More information

Update on Non-Alcoholic Fatty Liver Disease. Timothy R. Morgan, MD Chief, Hepatology, VA Long Beach Professor of Medicine, UCI

Update on Non-Alcoholic Fatty Liver Disease. Timothy R. Morgan, MD Chief, Hepatology, VA Long Beach Professor of Medicine, UCI Update on Non-Alcoholic Fatty Liver Disease Timothy R. Morgan, MD Chief, Hepatology, VA Long Beach Professor of Medicine, UCI February 3, 2018 Disclosure Clinical trials: Genfit Speaker s Bureau: none

More information

Two Dimension (2D) elasticity maps of coagulation of blood using SuperSonic Shearwave Imaging

Two Dimension (2D) elasticity maps of coagulation of blood using SuperSonic Shearwave Imaging Two Dimension (2D) elasticity maps of coagulation of blood using SuperSonic Shearwave Imaging Miguel Bernal, Jean-Luc Gennisson, Patrice Flaud, Mickaël Tanter To cite this version: Miguel Bernal, Jean-Luc

More information

Acoustic radiation force impulse (ARFI): a review of an useful technique to assess liver stiffness

Acoustic radiation force impulse (ARFI): a review of an useful technique to assess liver stiffness Acoustic radiation force impulse (ARFI): a review of an useful technique to assess liver stiffness Poster No.: C-2166 Congress: ECR 2013 Type: Educational Exhibit Authors: M. Elorz, A. Benito, L. Garcia

More information

Shear Wave Elastography in diagnostics of supraspinatus tendon.

Shear Wave Elastography in diagnostics of supraspinatus tendon. Shear Wave Elastography in diagnostics of supraspinatus tendon. Poster No.: C-2168 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit V. Saltykova; Moscow/RU Musculoskeletal joint, Musculoskeletal

More information

Ultrasound Committee Activities

Ultrasound Committee Activities Ultrasound Committee Activities Anthony E. Samir MD MPH Clinical Director MGH/MIT Center for Ultrasound Research & Translation PINTAD 2014 Disclosures Grant Support Harvard Catalyst: Pilot Grant (NIBIB

More information

Performance of 2D-SWE.GE for predicting different stages of liver fibrosis, using Transient Elastography as the reference method

Performance of 2D-SWE.GE for predicting different stages of liver fibrosis, using Transient Elastography as the reference method Original papers Med Ultrason 2017, Vol. 19, no. 2, 143-149 DOI: 10.11152/mu-910 Performance of 2D-SWE.GE for predicting different stages of liver fibrosis, using Transient Elastography as the reference

More information

Elastography. White Paper

Elastography. White Paper Elastography White Paper Strain Image in DC-8 Diagnostic Ultrasound System Shuangshuang Li, Rui Fan Relationship between the stiffness of tumor and it s malignance has far been known since the ancient

More information

Noninvasive Assessment of Liver Fibrosis in Egyptian Children with Chronic Liver Diseases.

Noninvasive Assessment of Liver Fibrosis in Egyptian Children with Chronic Liver Diseases. Curr Pediatr Res 2016; 20 (1&2): 57-63 ISSN 0971-9032 www.currentpediatrics.com Noninvasive Assessment of Liver Fibrosis in Egyptian Children with Chronic Liver Diseases. Tawhida Abdel Ghaffar 1, Azza

More information

Elastography to Assess the Stage of Liver Fibrosis in Children: Concepts, Opportunities, and Challenges

Elastography to Assess the Stage of Liver Fibrosis in Children: Concepts, Opportunities, and Challenges REVIEW Elastography to Assess the Stage of Liver Fibrosis in Children: Concepts, Opportunities, and Challenges Suraj D. Serai, Ph.D.,* Andrew T. Trout, M.D.,* and Claude B. Sirlin, M.D. WHAT IS ELASTOGRAPHY?

More information

Noninvasive Evaluation of Liver Fibrosis Using Real-time Tissue Elastography and Transient Elastography (FibroScan)

Noninvasive Evaluation of Liver Fibrosis Using Real-time Tissue Elastography and Transient Elastography (FibroScan) ORIGINAL RESEARCH Noninvasive Evaluation of Liver Fibrosis Using Real-time Tissue Elastography and Transient Elastography (FibroScan) Fankun Meng, MD, Ying Zheng, MD, Qi Zhang, MD, Xiaojie Mu, MD, Xiaoluan

More information

Quantitative Imaging Biomarkers in Ultrasound Elasticity Imaging. Timothy J Hall Medical Physics Department University of Wisconsin AIUM 2012

Quantitative Imaging Biomarkers in Ultrasound Elasticity Imaging. Timothy J Hall Medical Physics Department University of Wisconsin AIUM 2012 Quantitative Imaging Biomarkers in Ultrasound Elasticity Imaging Timothy J Hall Medical Physics Department University of Wisconsin AIUM 2012 Proposed Biomarkers Shear wave speed for quantifying liver fibrosis

More information

Clinical Criteria for Hepatitis C (HCV) Therapy

Clinical Criteria for Hepatitis C (HCV) Therapy Clinical Criteria for Hepatitis C (HCV) Therapy Pre-Treatment Evaluation o Must have chronic hepatitis C and HCV genotype and sub-genotype documented; o Patients who have prior exposure to DAA therapy

More information

Liver Shear-Wave Velocity and Serum Fibrosis Markers to Diagnose Hepatic Fibrosis in Patients with Chronic Viral Hepatitis B

Liver Shear-Wave Velocity and Serum Fibrosis Markers to Diagnose Hepatic Fibrosis in Patients with Chronic Viral Hepatitis B Original Article Gastrointestinal Imaging http://dx.doi.org/10.3348/kjr.2016.17.3.396 pissn 1229-6929 eissn 2005-8330 Korean J Radiol 2016;17(3):396-404 Liver Shear-Wave Velocity and Serum Fibrosis Markers

More information

Novel multiparametric magnetic resonance elastography (MRE) protocol accurately predicts NAS score for NASH diagnosis

Novel multiparametric magnetic resonance elastography (MRE) protocol accurately predicts NAS score for NASH diagnosis Novel multiparametric magnetic resonance elastography (MRE) protocol accurately predicts NAS score for NASH diagnosis Alina M. Allen, Meng Yin, Sudhakar K. Venkatesh, Taofic Mounajjed, Todd A. Kellogg,

More information

Acoustic radiation force impulse elastography in distinguishing hepatic haemangiomata from metastases: preliminary observations

Acoustic radiation force impulse elastography in distinguishing hepatic haemangiomata from metastases: preliminary observations The British Journal of Radiology, 84 (2011), 939 943 SHORT COMMUNICATION Acoustic radiation force impulse elastography in distinguishing hepatic haemangiomata from metastases: preliminary observations

More information

NASH : Diagnosis and investigation. VII Workshop international, Curitiba, Brazil 29/08/2014

NASH : Diagnosis and investigation. VII Workshop international, Curitiba, Brazil 29/08/2014 NASH : Diagnosis and investigation VII Workshop international, Curitiba, Brazil 29/08/2014 Vlad Ratziu, Université Pierre et Marie Curie, Hôpital Pitié Salpêtrière, Paris, France Usual diagnostic circumstances

More information

Assessment of Liver Fibrosis and Cirrhosis by Aspartate Aminotransferase-to-Platelet Ratio Index in Children With Biliary Atresia

Assessment of Liver Fibrosis and Cirrhosis by Aspartate Aminotransferase-to-Platelet Ratio Index in Children With Biliary Atresia ORIGINAL ARTICLE: HEPATOLOGY AND NUTRITION Assessment of Liver Fibrosis and Cirrhosis by Aspartate Aminotransferase-to-Platelet Ratio Index in Children With Biliary Atresia Sang Yong Kim, y Jae Yeon Seok,

More information

Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification

Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification Whitepaper Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification ACUSON S2000 Ultrasound System Answers for life. Page 1 Tissue Strain Analytics: Virtual Touch Tissue Imaging and Quantification

More information

The Future is Here Now!

The Future is Here Now! HCV Treatment: Assessing the Patient Prior to Treatment. How Will This Change in the Future? JORGE L. HERRERA M.D., FACG UNIVERSITY OF SOUTH ALABAMA COLLEGE OF MEDICINE, MOBILE, AL The Future is Here Now!

More information

Original papers Med Ultrason 2013, Vol. 15, no. 2, Ioan Sporea, Iulia Raţiu, Simona Bota, Roxana Şirli, Ana Jurchiş

Original papers Med Ultrason 2013, Vol. 15, no. 2, Ioan Sporea, Iulia Raţiu, Simona Bota, Roxana Şirli, Ana Jurchiş Original papers Med Ultrason 2013, Vol. 15, no. 2, 111-115 DOI: 10.11152/mu.2013.2066.152.is1ir2 Are different cut-off values of liver stiffness assessed by Transient Elastography according to the etiology

More information

Real-time elastography of parotid gland masses: the value of strain ratio for the differentiation of benign from malignant tumors

Real-time elastography of parotid gland masses: the value of strain ratio for the differentiation of benign from malignant tumors Realtime elastography of parotid gland masses: the value of strain ratio for the differentiation of benign from malignant tumors Poster No.: C09 Congress: ECR 05 Type: Scientific Exhibit Authors: M. M.

More information