Multi-Channel Film Dosimetry and Gamma Map Verification

Similar documents
NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy

Evaluation of triple channel correction acquisition method for radiochromic film dosimetry

Evaluation of a single-scan protocol for radiochromic film dosimetry

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Advanced film dosimetry for a UK brachytherapy audit

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

Y FILMS DOSIMETR Nederland België / Belgique

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions)

Verification of micro-beam irradiation

Abstract The thickness of papillary dermis is an important parameter in optical model of the skin [1, ]. Any change in the thickness produces differen

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA

Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ

Small Field Dosimetry & Radiochromic film

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Pre-treatment verification for IMRT / IGRT and Rotational therapy

Dosimetric Characterization of a Prototype Nanotechnology Microbeam Radiation Therapy Device using Gafchromic EBT2 Film

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

Experimental measurement with an anthropomorphic phantom of the proton dose distribution in the presence of metal implants

B3 RADIOCHROMIC FILM DOSIMETRY

Discuss the general planning concepts used in proton planning. Review the unique handling of CTV / ITV / PTV when treating with protons

GEX Recommended Procedure Eff. Date: 09/21/10 Rev.: D Pg. 1 of 7

Info424, UW ischool 11/6/2007

Activity report from JCOG physics group

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter

Prostate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10

Calibration Standards

Digital film dosimetry in radiotherapy and the development of analytical applications software

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

IsoAcoustics Technology Explained!

1- Cochlear Impedance Telemetry

Norland Densitometry A Tradition of Excellence

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance

Overview. Proton Therapy in lung cancer 8/3/2016 IMPLEMENTATION OF PBS PROTON THERAPY TREATMENT FOR FREE BREATHING LUNG CANCER PATIENTS

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Computed tomography Acceptance testing and dose measurements

Radiochromic film dosimetry in water phantoms

Example 1: How would you evaluate this transducer? What is the most common problem found in ultrasound QC testing? Uniformity is Subjective

Liver Fat Quantification

CORESTA RECOMMENDED METHOD N 25

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value

IMAGE-GUIDED RADIATION THERAPY

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

3D Dosimetry with Polymer Gel

IMPT with Carbon Ions

Implementation Guide for the DoseControl Dosimetry System

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter

Agilent 1220 Infinity LC System Performance Specifications

Absolute Identification is Surprisingly Faster with More Closely Spaced Stimuli

WFC3-IR Thermal Vacuum Testing: IR Channel Dark Current

Applications in brachytherapy

Verification of Advanced Radiotherapy Techniques

Confident and Conclusive Diagnosis with Flexible Solutions

Application(s) of Alanine

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

The New 6495 Triple Quadrupole LC/MS

Dave van Woensel 08 June Installation effects in ultrasonic gas flow meters Calibration results of a Z-configuration measurement skid

Characterization of the effect of MRI on Gafchromic film dosimetry

EARLY STAGE DIAGNOSIS OF LUNG CANCER USING CT-SCAN IMAGES BASED ON CELLULAR LEARNING AUTOMATE

A new geometric and mechanical verification device for medical LINACs

Application of Time Reversal Technique for the Inspection of Composite Structures

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

David Hunter Cell Color Measurement on Budget. Audience

Volumetric Modulated Arc Therapy - Patient Specific QA

A guide to using multi-criteria optimization (MCO) for IMRT planning in RayStation

Aixplorer MultiWave Ultrasound System An Innovation in Breast Ultrasound Imaging

Development and evaluation of an end-to-end test for head and neck IMRT with a novel multiple-dosimetric modality phantom

LION. Application Note PRECISION. Linear Position Measurement with Eddy-Current Sensors. LA March, Applicable Equipment: Applications:

Treatment Planning for Skull Base Tumors PTCOG 52, June 2013

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

Factors Impacting the Evaluation of Printer Profile Accuracy

Description of new EyeSuite visual field and trend analysis functions

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

MEM BASED BRAIN IMAGE SEGMENTATION AND CLASSIFICATION USING SVM

Introducing the future of DXA. Powerful images. Clear answers. Horizon DXA System

Surgical Probes अౠ.6,7,1 / ?K

Other Major Component Inspection II

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. YY Translated English of Chinese Standard: YY

Available Online through Research Article

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

Noise-Robust Speech Recognition Technologies in Mobile Environments

Leila E. A. Nichol Royal Surrey County Hospital

BMMC (UG SDE) IV SEMESTER

Transcription:

Multi-Channel Film Dosimetry and Gamma Map Verification Micke A International Specialty Products Ashland proprietary technology, patents pending

Single Channel Film Dosimetry Calibration Curve X=R R ave = R ave (D) D R = D R (R ave ) Color channels X=RGB D X = D( X ave ) correlates average response of film-scanner system Robust method any X value delivers dose D X (X) D=D X

Single Channel Film Dosimetry Problem: Specific pixel does not behaves like average Disturbance ΔX generates ΔD X X + ΔX D( X ) + ΔD X Film uniformity variations Scanner non-linearities Newton rings, noise, finger prints curling, Any X value delivers dose D X (X) Each channel specific ΔD X No indication of big ΔD X What dose D X is best?

Multi-Channel Film Dosimetry RGB Calibration Curves Dose induced color C C(D) = {R(D),G(D),B(D)} Dose exposure generates only certain colors C Not all C deliver dose value Observed color C scan is superposed with disturbance ΔC C scan = C(D) + ΔC Solution: Optimize dose D value, i.e. minimize ΔC C scan - C(D) min D

Triple Channel Film Dosimetry Model: Scanned optical density d X,scan d X,scan ( D ) = d X,D ( D ) * Δd d X = -log( X ) for X = RGB d X,D is calibration function (average behavior)! disturbance Δd independent of dose + X (wave length)! but Δd = Δd( thickness, scanner, noise, artifacts ) Solution: Minimized function vs. disturbance Δd: (Δd) = ( D R - D B ) 2 + ( D B - D G ) 2 + ( D G - D R ) 2 min Δd

Triple Channel Film Dosimetry Example Signal split into dose dependent and dose independent part D RGB + Δd Dose map (D dependent part) Disturbance Δd map (D independent part) includes film uniformity variations, noise etc.

Triple Channel Correction What happen to the Marker Dye? Marker Dye needed at lower dosage: Δd d B,scan (0) / d B,D (0) otherwise scanner noise dominates Δd (sufficient low dosage ~ 100 200 cgy) At higher dosage Red channel is saturating and has function like Marker Dye (sufficient high dosage > 400 cgy)

Triple Channel Dosimetry Film Consistency Film consistent with Calibration Patches Film has same dose response for X=RGB i.e. same dose values D X are calculable Offset between D X measures calibration consistency Example: Profiles original calibration patch and 90 rotated scan

Triple Channel Dosimetry Consistency Map Dose map measurement result Disturbance map removed error Consistency map remaining error ideal case: noise only Film Dose Map Disturbance Map Consistency Map (dark = +, light = -, contrast maximized) Example: dominated by scanner cogging

Consistency Map Comparison Triple Channel Dosimetry Single Channel Dosimetry same patch, same contrast scaling, dark = +, light = -

Triple Channel Dosimetry Calibration Consistency Calibration inconsistent with calibration patches also inconsistent with application film! Minimize inconsistency (numeric value) is uniformity corrected Calibration optimization (e.g. dose range) Detect abnormal patches 90 rotation, curling, Newton rings, top sheets anomaly Examples of calibration patch consistency map

Triple Channel Dosimetry Dose Map Consistency Dose map error estimation known before comparison Recognize mismatch of Calibration and Dose map e.g. batch mismatch Detect abnormal scans 90 rotation, curling, Newton rings, top sheets anomaly Example dose consistency map (iso-map) peak error ~2%

Dose, Gy Triple Channel Dosimetry Lateral Scanner Non-Linearity Scanner signal changes with lateral position (sensor direction) EBT film polarization amplifies lateral effect Non-dose-dependent part of lateral effect is compensated Mitigation only (partial compensation) 2.00 1.80 1.60 1.40 1.20 1.00 0.80 0.60 0.40 0.20 0.00 Red channel dosimetry Triple channel dosimetry 0 50 100 150 200 250 300 350 Position, x 0.353 mm

scan direction Lateral Scanner Non-Linearity Normalized Blank Scan Lateral effect increases with dose Compensates only weakest occurrence of lateral effect Adding disturbances Non-uniformity of blank film Noise of blank scan Worsens consistency for exposed areas 1.00 0.98 0.96 0.94 0.92 0.90 0.88 0.86 0 500 Calibration patch consistency comparison dose <cgy> Consistency <cgy> Consistency <%> None Blank scan None Blank scan 202.0 8.8 11.1 4.3 5.5 151.5 6.8 8.9 4.5 5.9! DO NOT USE! 101.0 5.9 8.4 5.9 8.3 50.5 5.9 7.9 11.6 15.6 0.0 4.8 0.4 Infinity Infinity

Lateral Scanner Non-Linearity Single vs. Triple Channel Passing rate Gamma map Criterion Calibration Composite Single Channel Triple Channel 90% 96% Calibration Calibration 2% / 2mm 87% 96% 98% 98% 94% 97% 1% / 1mm 58% 71% 64% 69% 81% 85% 73% 78% = centered = right edge absolute dosimetry (no dose re-scaling), <5% (<12cG) lowest dose ignored, Composite 1.00 0.98 0.96 0.94 0.92 0.90 0.88 0.86 0 500 Composite

Triple Channel Film Dosimetry Dose to Plan Comparison Triple Channel Passing rate R 97.15 % Single Channel Passing rate R - 87.48 % Gamma Map 2%/2mm - IMRT example (part of FilmQA Pro installation)

Triple Channel Film Dosimetry Dose to Plan Comparison Dose map error can dominate comparison 1% achievable (vs. 3% with single channel method) Comparison Criteria 3%/3mm, 2%/2mm Triple channel: 1% << 3%/2%, i.e. majority < tolerance Single channel: 3% ~ 3% test, i.e. 50% > tolerance Passing rates improves more than dose accuracy

Gamma Map Comparison plan pixel and overlaid pixels of registered dose map same standard deviation low sample number fails, high sample number passes Use dose average across plan pixel e.g. Projection of dose map to plan coordinate system Filtering cannot fix this problem!

Single vs. Triple Channel Noise Dependence Triple Channel Passing rate Single Channel Passing rate Criterion 0% noise 0.5% noise 1% noise 2% noise Criterion 0% noise 0.5% noise 1% noise 2% noise 3% / 3mm 97% 96% 95% 95% 2% / 2mm 94% 96% 94% 87% 1% / 1mm 68% 70% 58% 41% 3% / 3mm 94% 95% 95% 94% 2% / 2mm 89% 92% 94% 91% 1% / 1mm 57% 65% 61% 48% 2% / 2 mm Gamma map comparison at 0%, 2% noise White noise with various standard deviation added to EBT3 film scan, Gamma map with projection

Multi Channel Calibration Calibration = average system response x = x( D ) (single channel calibration, x = RGB) Calculate dose map using single channel calibration x + multi channel dose map Correlate multi-channel calibration X X( D ) = A + B x( D ) (2 point rescaling) X( D ) = A + B x( C D ) (3 point rescaling) X( D ) = A + B x( D C ) X = RGB Requires only 1-2 exposures Enforces perfect consistency at reference Single scan Calibration + Application compensates for Ambient conditions: temperature, humidity Inter-scan scanner variations, Post exposure time, film aging

Triple Channel Film Dosimetry Dynamic Dose Range Same dose mapping method for all channels Enables EBT s full dynamic range Factor >1000 Lateral effect increases substantially! Example: Calibration range 0 30 Gy, Dose map 26 Gy peak passing rate 95.96 % @ 2%/2mm

Triple Channel Film Dosimetry Features and Advantages Separate Dose and Dose-independent effects Compensates for film thickness variation Noise reduction without dose change Mitigates scanner distortions Background compensation, double exposure unnecessary Enables entire film dose range Ebt2 dynamic range ratio >1000 (1 cgy - >40 Gy) Significant improvement of dose map accuracy 1% achievable (vs. 3% with single channel method) Indication of inconsistency between film and calibration, calibration inconsistencies

Lit.: Micke, Lewis, Yu - Multi-channel Film Dosimetry with Non-Uniformity Correction, Medical Physics, 38 (2011) 5, pp. 2523.