Uncertainties and Quality Assurance of Localization and Treatment in Lung SBRT Jing Cai, PhD Duke University Medical Center

Similar documents
SBRT I: Overview of Simulation, Planning, and Delivery

Which Planning CT Should be Used for Lung SBRT? Ping Xia, Ph.D. Head of Medical Physics in Radiation Oncology Cleveland Clinic

Pitfalls in SBRT Treatment Planning for a Moving Target

Required target margins for image-guided lung SBRT: Assessment of target position intrafraction and correction residuals

Motion gating and tracking techniques: overview and recent developments

In-Room Radiographic Imaging for Localization

8/3/2016. Outline. Site Specific IGRT Considerations for Clinical Imaging Protocols. Krishni Wijesooriya, PhD University of Virginia

Image Guided Stereotactic Radiotherapy of the Lung

In-Room Radiographic Imaging for Localization

Stereotaxy. Outlines. Establishing SBRT Program: Physics & Dosimetry. SBRT - Simulation. Body Localizer. Sim. Sim. Sim. Stereotaxy?

IGRT Protocol Design and Informed Margins. Conflict of Interest. Outline 7/7/2017. DJ Vile, PhD. I have no conflict of interest to disclose

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

Potential systematic uncertainties in IGRT when FBCT reference images are used for pancreatic tumors

IGRT Solution for the Living Patient and the Dynamic Treatment Problem

Credentialing for the Use of IGRT in Clinical Trials

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Regional Variation of Interfraction Tumor Breathing Motion in Lung Stereotactic Body Radiation Therapy (SBRT)

RTTs role in lung SABR

1 : : Medical Physics, Città della Salute e della Scienza, Torino, Italy

Overview of Advanced Techniques in Radiation Therapy

Innovative RT SBRT. The variables with REQ in superscript are required.

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

Margins in SBRT. Mischa Hoogeman

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient

On the use of 4DCT derived composite CT images in treatment planning of SBRT for lung tumors

Lung SBRT in a patient with poor pulmonary function

8/2/2012. Transitioning from 3D IMRT to 4D IMRT and the Role of Image Guidance. Part II: Thoracic. Peter Balter, Ph.D.

Inter-fractional Positioning Study for Breast Cancer with Proton Therapy using a 3D Surface Imaging System

Comparison of Interfacility Implementation of Essential SBRT Components. Keith Neiderer B.S. CMD RT(T) VCU Health System

Asynchronization. (aka MLC interplay effect with tumor motion)

1. Murphy, Martin J., et al. "The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75.

A feasibility study of intrafractional tumor motion estimation based on 4D CBCT using diaphragm as surrogate

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

UNIVERSITY OF WISCONSIN-LA CROSSE Graduate Studies

Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer

Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife of-the-art and New Innovations

Image Registration for Radiation Therapy Applications: Part 2: In-room Volumetric Imaging

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014

Managing the imaging dose during image-guided radiation therapy

4D PET: promises and limitations

Implementing SBRT Protocols: A NRG CIRO Perspective. Ying Xiao, Ph.D. What is NRG Oncology?

Acknowledgements. Methods and Experience in Imageguided. SRS/SBRT: Technology, Geometry, and the Therapeutic Ratio. Conflict of Interest

The Effects of DIBH on Liver Dose during Right-Breast Treatments Introduction

SRS Uncertainty: Linac and CyberKnife Uncertainties

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy

Overview. Proton Therapy in lung cancer 8/3/2016 IMPLEMENTATION OF PBS PROTON THERAPY TREATMENT FOR FREE BREATHING LUNG CANCER PATIENTS

SBRT of Lung & Liver lesions using Novalis IGRT System. Patrick Silgen, M.S., DABR Park Nicollet Methodist Hospital

Objectives. Image-Guided RT: Process. kv CBCT Novel Clinical Applications and Future Directions. Standard Clinical Uses. Novel Applications.

Anatomical-Based Adaptive RT: It Begins Here!

8/1/2016. Motion Management for Proton Lung SBRT. Outline. Protons and motion. Protons and Motion. Proton lung SBRT Future directions

SBRT TREATMENT PLANNING: TIPS + TRICKS. Rachel A. Hackett CMD, RT(T)

Clinical evaluation of interfractional variations for whole breast radiotherapy using 3-dimensional surface imaging

Adequacy of inhale/exhale breathhold CT based ITV margins and image-guided registration for free-breathing pancreas and liver SBRT

Volumetric-based image guidance is superior to marker-based alignments for stereotactic body radiotherapy of prostate cancer

Jean Pouliot, PhD Professor and Vice Chair, Department of Radiation Oncology, Director of Physics Division

Feasibility of 4D IMRT Delivery for Hypofractionated High Dose Partial Prostate Treatments

Fiducial-Free Lung Tracking and Treatment with the CyberKnife System: A Non-Invasive Approach

Variable Dose Rate Dynamic Conformal Arc Therapy (DCAT) for SABR Lung: From static fields to dynamic arcs using Monaco 5.10

Deep Inspiration Breath-Hold Techniques: a review of voluntary and controlled breath-hold maneuvers for left-side breast radiation therapy

Accuracy and inter-observer variability of 3D versus 4D cone-beam CT based image-guidance in SBRT for lung tumors

8/1/2017. Clinical Indications and Applications of Realtime MRI-Guided Radiotherapy

ART for Cervical Cancer: Dosimetry and Technical Aspects

Treatment Planning & IGRT Credentialing for NRG SBRT Trials

The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study Abstract: Introduction: Case Description: Conclusion: Introduction

I. Equipments for external beam radiotherapy

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017

Questions may be submitted anytime during the presentation.

Implementation of advanced RT Techniques

Reena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery

Dosimetric effects of positioning shifts using 6D-frameless stereotactic Brainlab system in hypofractionated intracranial radiotherapy

Accuray Radixact at University of WI: The next chapter in TomoTherapy innovation

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

Advances in external beam radiotherapy

ASTRO econtouring for Lymphoma. Stephanie Terezakis, MD

Clinical Aspects of Proton Therapy in Lung Cancer. Joe Y. Chang, MD, PhD Associate Professor

8/3/2016. Implementation of Pencil Beam Scanning (PBS) Proton Therapy Treatment for Liver Patients. Acknowledgement. Overview

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Dosimetric consequences of tumor volume changes after kilovoltage cone-beam computed tomography for non-operative lung cancer during adaptive

Clinical Implications Of Dose Summation And Adaptation

4D Radiotherapy in early ca Lung. Prof. Manoj Gupta Dept of Radiotherapy & oncology I.G.Medical College Shimla

Intra- and inter-fractional liver and lung tumor motions treated with SBRT under active breathing control

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Feasibility study on effect and stability of adaptive radiotherapy on kilovoltage cone beam CT

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Image Guided Proton Therapy and Treatment Adaptation

CBCT of the patient in the treatment position has gained wider applications for setup verification during radiotherapy.

It s All About Margins. Maaike Milder, Ph.D. Accuray Symposium April 21 st 2018

Comparison of CT images with average intensity projection, free breathing, and mid-ventilation for dose calculation in lung cancer

MRI Applications in Radiation Oncology:

Radiation treatment planning in lung cancer

From position verification and correction to adaptive RT Adaptive RT and dose accumulation

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Nano-X A Smarter and Smaller Cancer Radiotherapy System

Work partially supported by VisionRT

Changing Paradigms in Radiotherapy

Transcription:

Uncertainties and Quality Assurance of Localization and Treatment in Lung SBRT Jing Cai, PhD Duke University Medical Center 2013 AAPM 55 th Annual Meeting, Educational Course, Therapy Track, MOC SAM Program

Disclosure I have received research funding from NIH, the Golfers Against Cancer (GAC) foundation, and Philips Health System.

Acknowledgements Duke Radiation Oncology Fang-Fang Yin, PhD Chris R. Kelsey, MD David S. Yoo, MD, PhD James Bowsher, PhD Lei Ren, PhD Jim Chang, PhD Yun Yang, PhD Yunfeng Cui, PhD Zhiheng Wang, PhD Irina Vergalasova, BS Suzanne Catalano, BS, CMD Rhonda May, BS, CMD Duke Medical Physics Program Paul Segars, PhD You Zhang, BS Cindy Qin, MS Kate Turner, MS Lynn Cancer Institute at Boca Raton Regional Hospital Charles Shang, PhD Duke Radiation Safety Fan Zhang, MS

Imaging in Lung SBRT

The Process: Image Guidance X-Ray (ExacTrac) RPM CBCT Fluoroscopy (X-Ray/CBCT) Cine-MV, X-Ray X-Ray/CBCT

Free-breath 3D CBCT Match Planning CBCT Post-treatment images CT with after prior target correction CBCT to contours Wang et al Ref J 2007

Uncertainties in lung SBRT IGRT Tumor volume in CBCT Soft-tissue contrast Inter-observer variations Reproducibility of tumor location at breath-hold Internal-external motion correlation Changes of tumor size and motion Changes of anatomy Shifts and rotations in matching.

Which CT for CBCT Matching? 4DCT-AIP v.s. CBCT

Which CT for CBCT Matching? 3D FB-CT v.s. CBCT

CBCT Matching: Tiny Tumor Tumor Size ~ 5 mm; Tumor Motion ~ 20 mm

CBCT Matching: Large Anatomical Change Pleural effusion at Sim Largely disappeared at 1 fx Re-simed, Re-planned

CBCT ITV Uncertainty ITV at different Inspiration/Expiration (I/E) Ratio 1.0 0.52 0.35 0.26 0.21 4D CBCT Vergalasova, et al, Med Phys. 2011

CBCT ITV Uncertainty Small Tumor A Large Tumor B FB ITV 4D ITV FB ITV 4D ITV 3.5mm 2.5mm Tumor Free-Breathing ITV (cm 3 ) 4D ITV (cm 3 ) Volume Underestimation (%) A 1.78 2.97 40.1 B 35.62 46.98 24.2 Vergalasova et al, Med Phys 2011

Target Matching Uncertainty Turner et al 2013 AAPM

Image Registration Uncertainty: Inter-observer Variation Cui et al, Red J, 2011; 81:305-312.

MVCT for Lung SBRT IGRT Siker et al, Red J, 2006 Cui et al, Red J, 2011; 81:305-312.

Question: Which one of the following answers represents the best estimate of the inter-observer variation in image registration in lung SBRT? 20% 1. 1 mm 20% 20% 20% 20% 2. 2 mm 3. 3 mm 4. 5 mm 5. >5 mm 10

Correct Answer: 2. 2 mm Discussion Reference: Cui Y, Galvin JM, Straube WL, Bosch WR, Purdy JA, Li XA, Xiao Y, Multi-system verification of registrations for image-guided radiotherapy in clinical trials. Int J Radiat Oncol Biol Phys, 2011; 81:305-312.

Degree of Corrections Rotational Shifts in Lung SBRT 3.0 2.0 104 Lung SBRT Cases pitch roll 1.0 0.0-1.0-2.0 Shang et al, 5 th NC IMRT/IGRT Absolute Corrections of Pitch/Roll Symposium, 2012 Pitch Roll Absolute Mean 0.87 0.60 Net Average of Pitch & Roll Variance 0.33 0.30 Pitch Roll Cases >0.5 69.2% 50.0% Average 0.10-0.10 Pitch or Roll >0.5 89.4% SD 1.07 0.65 Pitch or Roll >1.0 51.0% - 3.0

Dosimetric Effects of Rotations Roll Yaw Pitch Catalano et al, 5 th NC IMRT/IGRT Symposium, 2012 95.6% of all differences were <1% or <1Gy. Overall small dosimetric effects of uncorrected rotations.

Dosimetric Effects of Rotations Large inter-subject variations at large rotation angles. Up to 4% reduction in PTV coverage, 6 Gy increase in cord D0.35cc, and 4 Gy in Esophagus D0.35cc observed.

Cine MV: tumor motion during TX Tumor motion during 5-fx lung SBRT Cine MV Zhang et al, RPO 2013 4D-CT

Intra-fractional Mean Tumor Position Shift 409 Patients 427 Tumors 1593 Fractions Shah C, et al, PRO, 2012 Intra-fractional variation: AP: 0.0 ± 1.7 mm ML: 0.6 ± 2.2 mm SI: 1.0 ± 2.0 mm 3D: 3.1 ± 2.0 mm 3D vector variation: > 2mm in 67.8% > 5mm in 14.3% Depending on immobilization (Range: 2.3 3.3 mm) Body Frame < Alpha Cradle < Body Fix < Wing Board

Change of Tumor Size During Lung SBRT 40 lung SBRT patients Initial tumor size: 0.7-7.3 cm Change of tumor diameter: Range: -34.2% to 33.0% Mean: -7.9 ± 11.45% Qin et al, Red J, 2013

ExacTrac ExacTrac 6D v.s. CBCT 6D ML: 1.06 mm AP: 1.43 mm SI: 1.43 mm Chang et al, Radiother Oncol, 2010 Pitch: 1.22 Row: 0.64 Yaw: 1.66 Small but maybe clinically significant discrepancies between ExacTrac X ray 6D and CBCT 6D match

Cyberknife Targeting error 0.1 0.3 mm Correlation error: 0.3 2.5 mm Prediction error: 1.5 ± 0.8 mm Total error 0.7 5.0 mm Synchrony Respiratory Tracking System (RTS) Pepin et al, Med Phys. 2011

Question: Which one of the following answers represents the best estimate of the mean intra-fractional 3D tumor position shift in lung SBRT? 20% 1. 1 mm 20% 20% 20% 20% 2. 2 mm 3. 3 mm 4. 5 mm 5. >5 mm 10

Correct Answer: 3. 3 mm Discussion Reference: Shah C, Kestin LL, Hope AJ, Bissonnette JP, Guckenberger M, Xiao Y, Sonke JJ, Belderbos J, Yan D, Grills IS. Required target margins for image-guided lung SBRT: Assessment of target position intrafraction and correction residuals. Prac Radiat Onco. 3(1), 67-73. 409 Patients 427 Tumors 1593 Fractions 3D: 3.1 ± 2.0 mm

Onboard DTS Imaging kv MV 4D - DTS Free-breathing Reference DTS Phase-matched Reference DTS On-board Acquired DTS Courtesy from Dr. Ren of Duke University Better Match

MRI for Image Guidance

On-Board SPECT Courtesy from Dr. Bowsher of Duke University 4-min scans 7, 10 mm hot spots SPECT on robotic arm Molecular targeting Multi-Pinhole collimation

Summary Uncertainties exist in each step of image guidance of lung SBRT Understanding root causes and characteristics of these uncertainties is important for successful implementation of lung SBRT Next generation of on board imaging techniques has the potential to minimize uncertainties of image guidance of lung SBRT