Simulations of the blood flow in the arterio-venous fistula for haemodialysis

Similar documents
Numerical investigations of the unsteady blood flow in the end-to-side arteriovenous fistula for hemodialysis

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

Arteriovenous Graft Modeling and Hemodynamic Interpretation

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis

A numerical study on the effect of hematocrit on hemodynamic characteristics in arteriovenous graft

Supera for the Juxta-anastomotic AVF Stenosis

Numerical analysis of blood flow in human abdominal aorta

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION

Blood flow in vessels with artificial or pathological geometrical changes

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique

Analysis of the effects of plaque deposits on the blood flow through human artery

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics

Blood Flow Simulation toward Actual Application at Hospital

Introduction to the Native Arteriovenous Fistula: A primer for medical students and radiology residents

Ultrasound and the dialysis patient

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images

CPM Specifications Document Aortofemoral Normal:

Edinburgh Imaging Academy online distance learning courses

Comparison of Stent Designs using Computational Fluid Dynamics

Extraction and FSI modeling of Left coronary artery structure from patient s CTA images

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt

Medical device design using Computational Fluid Dynamics (CFD)

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

A Review: Hemodynamics of Cerebral Aneurysm with Mathematical Modeling

4/30/2014. In Vitro Study of AVF Outflow Vein Wall Shear Stress and Pulse Pressure

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm

3D Modeling of Plaque Progression in the Human Coronary Artery

FOR many decades, cardiovascular disease has been one of

Computational Fluid Dynamics Analysis of Blood Flow in Human Aorta

BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries

Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery

ANALYSIS OF FLUID-STRUCTURE INTERACTION IN ABDOMINAL AORTIC ANEURYSM WITH HIGH AND NORMAL BLOOD PRESSURE

Journal of Biomechanics

Flow Analysis with Stent Placement in the Cerebral Aneurysm

Physiological flow analysis in significant human coronary artery stenoses

Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid Artery with Stenosis

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

EVALUATION OF ABDOMINAL AORTIC ANEURYSM WALL STESS BASED ON FLOW INDUCED LOAD

HD Scanning: Velocities and Volume Flow

VFI Technology to Change the Way You Work

A Computational Fluid Dynamics Study on Bidirectional Glenn Shunt Flow with an Additional Pulsatile Flow Through a modified Blalock-Taussig Shunt

The effect of the LVAD connecting point on the hemodynamics in the aortic arch BMTE 09.16

A Two-layered Model for the Analysis of Arterial Rheology

Effects of surface geometry and non-newtonian viscosity on the flow field in arterial stenoses

Goals. Access flow and renal artery stenosis evaluation by Doppler ultrasound. Reimbursement. WHY use of Doppler Ultrasound

Computational medical imaging and hemodynamics framework for functional analysis and assessment of cardiovascular structures

Saccular Aneurysm Formation in Curved and Bifurcating Arteries

Investigation of blood flow rheology using second-grade viscoelastic model (Phan-Thien Tanner) within carotid artery

Post-conditioning. P a g e 1. To my Thesis Committee,

Particle Hemodynamics Analysis after Coronary Angioplasty

Vascular reconstruction: CFD predictions of bypass graft haemodynamics

Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass

The Study of Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction

Asymmetric flows of non-newtonian fluids in symmetric stenosed artery

Sonographic Evaluation of an Immature Brescia-Cimino Fistula

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY

Flow in re-stenosed artery after angioplasty

Numerical investigation of the blood flow through the middle cerebral artery

4D model of hemodynamics in the abdominal aorta

Mathematical and Computational study of blood flow through diseased artery

AVF Prevalence. Local elastase to aid fistula maturation. I have nothing to disclose

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis

KDOQI Guidelines. Overview. Predicting Successful Fistula Maturation Warren Gasper MD UCSF Vascular Surgery Fellow 2011 UCSF Vascular Symposium

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study

Identification of Atherosclerotic Lesion-Prone Sites through Patient-Specific Simulation of Low-Density Lipoprotein Accumulation

Evaluation of hemodialysis arteriovenous fistula before and after surgery: Teaching points

Computational design of Intracranial Stent using 3D visualization system

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction

What is the mechanism of the audible carotid bruit? How does one calculate the velocity of blood flow?

COMPUTATIONAL FLUID DYNAMICS SIMULATION OF EARLY DIAGNOSIS OF DEEP VEIN THROMBOSIS

NUMERICAL STUDY OF PULSATILE BLOOD FLOW IN THE CORONARY SYSTEM WITH THE RCA BYPASS GRAFT

ANGIOGRAPHIC FEATURES OF

Predicting flow in aortic dissection: comparison of computational model with PC-MRI velocity measurements

Numerical Analysis of Blood Flow in Human Arteries and Medical Devices

Noninvasive Fractional Flow Reserve from Coronary CT Angiography

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes

The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries

Studying the Flow Dynamics Within Endografts in Abdominal Aortic Aneurysms

Measure #330: Adult Kidney Disease: Catheter Use for Greater Than or Equal to 90 Days National Quality Strategy Domain: Patient Safety

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation

Introduction. Introduction 2/18/2015 ASDIN Vascular Access complications: High associated morbi-mortality. Worsened quality of life

EFFECT OF BRANCHINGS ON BLOOD FLOW IN THE SYSTEM OF HUMAN CORONARY ARTERIES. Benchawan Wiwatanapataphee. Yong Hong Wu. Thanongchai Siriapisith

Physician s Vascular Interpretation Examination Content Outline

Transcription:

Acta of Bioengineering and Biomechanics Vol. 16, No. 1, 2014 Original paper DOI: 10.5277/abb140109 Simulations of the blood flow in the arterio-venous fistula for haemodialysis DANIEL JODKO*, DAMIAN OBIDOWSKI, PIOTR REOROWICZ, KRZYSZTOF JÓŹWIK Łódź University of Technology, Institute of Turbomachinery, Łódź, Poland. The Ciminio-Brescia arterio-venous fistula is a preferred vascular access for haemodialysis, but it is often associated with the development of vascular complications, due to changes in hemodynamic conditions. Computational fluid dynamics methods were involved to carry out seven simulations of the blood flow through the fistula for the patient specific (geometrical) case and various boundary conditions. The geometrical data, obtained from the angio-computed tomography, were used to create a 3-dimensional CAD model of the fistula. The blood flow patterns, blood velocity and the wall shear stress, thought to play a key role in the development of typical complications (stenoses, thromboses, aneurysms, etc.), have been analyzed in this study. The blood flow is reversed locally downstream the anastomosis (where the artery is connected to the vein) and downstream the stenosis in the cannulated vein. Blood velocity reaches abnormal value in the anastomosis during the systolic phase of the cardiac cycle (2.66 m/s). The wall shear stress changes in this place during a single cycle of the heart operation from 27.9 to 71.3 Pa (average 41.5 Pa). The results are compared with data found in the literature. Key words: a-v fistula, blood flow, computational fluid dynamics, vascular access 1. Introduction Patients who suffer from the end-stage renal disease need kidney transplants or dialysis. In haemodialysis, blood is withdrawn from the patient and circulates through a special filter that removes products of metabolism. The Cimino-Brescia arterio-venous (a-v) fistula is a surgically created connection between the artery and the vein and it is considered the best vascular access for haemodialysis. As the connection is made, high pressure blood from the artery flows directly to the vein and extends its diameter, which ensures an adequate flow rate (5.0 13.3 cm 3 /s) [1], [2]. This means five to ten times greater flow rate than the normal one. Moreover, the limp vein wall changes and becomes thicker and more elastic, which enables multiple cannulation of the vein. This process lasts 6 8 weeks and is called maturation (arterialisation) of the fistula. The fistula is usually created (if possible) between the cephalic vein and the radial artery at the wrist or the forearm and there are generally three types of such connections, namely: side-to-side, end-to-side, end-to-end. A significant number of complications (stenoses, thromboses, aneurysms) of fistulae is related to the geometry of the anastomosis (the place where the artery connects with the vein) and the local hemodynamics [3]. Local flow conditions, in particular the wall shear stress (WSS), are thought to affect sensitive endothelial cells on the inner vessel wall, which leads to intimal hyperplasia and further complications [1], [3]. This paper shows results (blood velocity patterns, WSS) obtained in blood flow simulations performed with the ANSYS CFX code. * Corresponding author: Daniel Jodko, Łódź University of Technology, Institute of Turbomachinery, Łódź, Poland. Tel: +48 42 631 24 64, fax: +48 42 636 13 83, e-mail: danieljodko@edu.p.lodz.pl Received: March 3rd, 2013 Accepted for publication: July 1st, 2013

70 D. JODKO et al. 2. Methods and materials 2.1. Geometrical model of the fistula The aim of this study was to carry out simulations of the blood flow not only near the anastomosis but also in more distal veins receiving blood from the fistula for the patient specific case. Therefore, a set of DICOM images of the matured radiocephalic end-to-end a-v fistula (obtained from the angio-computed tomography angio-ct) was used as the basis of geometrical data. These images were opened in the RSR2 medical images viewer featuring a possibility of reading coordinates of the points and special measuring functions. The course of the vessels in space could be found and the main dimensions of vessels could be measured. A full 3D CAD model of the fistula was created in Solid Works 2010 (Fig. 1). In fact, this model represents a volume of the blood flowing through the fistula. Thus, external walls of the model are internal walls of the blood vessels. Fig. 1. Model of the fistula observed in the RSR2 (left) and a 3D CAD model created in Solid Works (right); red arrow inlet, blue arrows outlets, black arrows direction of the blood flow through the fistula [5] Fig. 2. Mesh of the model: (a) cannulated vein with prismatic elements in the vicinity of the wall, (b) bifurcation of blood vessels [5]

Simulations of the blood flow in the arterio-venous fistula for haemodialysis 71 In this study, walls of all vessels were assumed rigid and non-deformable, which is a typical assumption used in other papers concerning simulations of blood flows [3], [4]. A geometrical model of the fistula was meshed and tetrahedral elements were built. Additionally, prismatic elements were introduced near walls of all vessels to increase the accuracy of calculations in the vicinity of the wall. Samples of the mesh are shown in Fig. 2. 2.2. Boundary conditions As boundary conditions, a typical blood flow velocity waveform [2] at the inlet cross-section (feeding the brachial artery) and a blood static pressure waveform [4] at the outlet cross-sections (receiving veins) were used (see Fig. 3). The pick value of pressure is offset by approximately 0.1 s from the maximum value of velocity. Seven cases for one cardiac cycle (t = 0.84 s) were defined and 7 stationary simulations (A-G) of the blood flow through the fistula were performed. considered a non-newtonian fluid because its viscosity depends not only on the kind of liquid but also on the blood velocity gradient (flow parameter). For this reason, numerous rheological models of blood can be found in the literature. For the needs of this simulation, a modification [4] of the Power Law was employed where the dynamic viscosity (η) is defined as η = 0.55471 η = η0 n 1 η = 0.00345 for for for 10 9 < 10 9, 327, < 327, (1) where η 0 = 0.0035 Pa s, n = 0.6 and is the shear stress rate. A representation of different blood viscosity models is shown in Fig. 4. 2.4. Governing equations To obtain a solution of the problem described, the ANSYS CFX code (version 13.0) was used. According to the ANSYS CFX Theory Guide, the problem is solved by solving equations of mass, momentum, and energy conservation (Navier Stokes equations). The continuity equation is expressed as ρ + ( ρu ) = 0. (2) The momentum equation takes the following form Fig. 3. Changes in time-dependent velocity at the inlet cross-section and time-dependent pressure at the outlet cross-sections during a single cardiac cycle 2.3. Fluid model From the viewpoint of flow, density, viscosity and heat transfer (neglected in this study) are the fundamental blood parameters. However, blood is very difficult to describe by these parameters because all of them depend on personal features (e.g., hematocrit, age, sex, diet, life-style, etc.). Blood density values fall within the range 1030 1070 kg/m 3 [6]. In this study, the constant value of blood density equal to 1045 kg/m 3 was assumed. Blood is considered a non-newtonian fluid because its ( ρu ) + ( ρu U ) = p + τ + The total energy equation is noted as ( ρh tot ) p + ( ρuh tot = ( λ T ) + ( Uτ ) + US m + S E, ) S m. (3) (4) where: ρ density, U velocity, p pressure, τ stress tensor, S m sum of body forces, h tot = h + 0.5 U 2 the total enthalpy, S E energy source, λ thermal conductivity. In the present investigation, the thermal energy equation is not taken into consideration as the blood

72 D. JODKO et al. Fig. 4. Apparent blood viscosity vs. strain for different blood models [7], [8] flow was assumed to be isothermal. Thus, energy dissipation and heat conductivity are neglected. Blood flows observed in real anastomoses are usually turbulent, therefore the Shear Stress Transport (SST) model of turbulence was used in this study. 3. Results For the seven cases investigated in this study, the results that allow one to analyse velocity, flow patterns and the WSS were obtained. The results from the blood flow patterns in the fistula (anastomosis and veins) are shown in Fig. 5a. The maximum value of velocity of blood in the whole domain was observed in the anastomosis and it varied from 1.28 to 2.66 m/s (average 1.66 m/s) during one cardiac cycle (Fig. 5b). The blood velocity reaches high values also in the cannulated vein stenosis but this is a narrowing of the vessel, thus an increase in the blood velocity value was expected (Fig. 7a). A turbulent flow with large vortices and back flows are observed in the aneurysm just downstream of the anastomosis (Fig. 6a). Similar phenomena, depicted in Fig. 7a, occur downstream of the cannulated a b Fig. 5. (a) visualization of blood flow patterns near the anastomosis (black arrow) extracted from simulations using the systolic velocity of blood at the inlet (case C); (b) changes in the maximum value of blood velocity in the anastomosis versus time during a single cardiac cycle

Simulations of the blood flow in the arterio-venous fistula for haemodialysis 73 vein stenosis where blood velocity rapidly decreases and the blood flow is locally reversed. The WSS was investigated on the walls of the radial artery, the anastomosis and veins. The maximum value of the WSS, depicted in Fig. 6a, was observed in the anastomosis and it changes within a single cardiac cycle from 27.9 to 71.3 Pa (average 41.5 Pa). The second location of the area of high values of the WSS was found near the vein stenosis. Figure 7b shows a distribution of the WSS near the stenosis obtained for the case C for which the blood systolic velocity was used as the boundary condition at the inlet. The WSS distribution is irregular and asymmetric around this place and its maximum values change during a single cycle of the heart operation from 14.5 Pa to 36.4 Pa (average 21.5 Pa). The flow rate at the inlet of the fistula investigated was 18.3 cm 3 /s and this value should provide adequate haemodialysis [2]. From the viewpoint of haemodialysis efficiency, the most important parameter describing the blood flow through the fistula is a flow rate measured in the cannulated vein. Thus, on the basis of the blood velocity values obtained for the outlet of the punctured vein in the simulations performed and knowing the diameter (7.5 mm) of the outlet, an average value of the blood flow rate was determined. It equals 4.3 cm 3 /s. 4. Discussion The objective of this study was to evaluate and quantify the blood flow through the 3-dimensional model of end-to-end fistula used in haemodialysis on the basis of the patient s specific vessel spatial shape. A possibility of modelling the complicated geometry of the blood vessels involved, taking into account the a b Fig. 6. (a) visualization of the WSS near the anastomosis (black arrow) extracted from simulations using the systolic velocity of blood at the inlet (case C); (b) changes in the maximum value of the WSS in the anastomosis versus time during a single cardiac cycle a b Fig. 7. Visualization of the blood flow streamlines (a) and the WSS (b) near the cannulated vein stenosis extracted from simulations using the systolic velocity of blood at the inlet (case C)

74 D. JODKO et al. feeding artery (radial artery), the anastomosis of veins receiving blood was confirmed. The flow in the anastomosis and the stenosis of the cannulated cephalic vein was turbulent and flow patterns that occur in those places were strictly connected with their geometrical shape. The anastomoses and stenoses are considered to be places of occurrence of clot risk because the blood velocity rapidly decreases and the blood flow can be even reversed locally around them. The clot formed, for instance, in the vortex centre, can be transported by the blood flow and may clog smaller blood vessels in the brain or the lung and cause instantaneous death of the patient. Narrowings cause also a pressure drop that usually leads to further thrombosis and, consequently, to a reduction of the diameter of the vessel. This may be the reason of decrease in the flow rate and can make the fistula completely useless for haemodialysis. The velocity fields investigated in this study provide one with some knowledge about the possible values of blood velocity in the fistula with an emphasis on the anastomosis. The average maximum value of blood velocity during a single cardiac cycle obtained in this investigation is 2.66 m/s. Thus, it confirms the information found in the literature about extremely high values of blood velocity that can occur in anastomoses [1], [3]. However, one should remember that rigidity of the blood vessel walls was assumed in this study. This assumption means that the energy is not accumulated during the contraction phase of the heart. If the deformability of blood vessel walls had been taken into account, we could have obtained more accurate results. Nevertheless, wall deformation would play a more important role in the non-stationary analysis and, for the steady state, this assumption of wall rigidity is adequate. The oscillating wall shear stress and its abnormal values are thought to be a fundamental factor influencing the endothelium and to play a key role in the development of intimal hyperplasia. This means a biological effect of a dramatic change in hemodynamic conditions after creation of the fistula and always leads to its dysfunction. The average maximum value of the WSS obtained for the anastomosis is 41.5 Pa. This order of magnitude is consistent with other results obtained from the in vitro tests or the blood flow simulations through the fistula models that have been found in the literature [1], [3]. Thus, the steady state solution may be used whenever limited computational resources are achievable or whenever short response time is a key factor. This may be important when clinicians want to consider numerous variants of the fistulas that may be developed in the particular patient. CFD methods may be helpful in choosing the best geometry and location of the fistula and may support its maintenance. However, further investigations should be devoted to various fistula models with an assumption of time-dependent pressure and blood velocity and deformability of vessel walls. Additionally, a blood velocity waveform and time-dependent pressure should be measured for the selected crosssections of the real fistula and be employed as boundary conditions. The results should be compared with the experimental data obtained from ultrasonic imaging (colour Doppler sonography). Finally, the fistula examined should be investigated after one month or longer time to check if any complications are formed. Thus, a method should be developed to provide a deeper insight into the correlation between the results obtained in simulations and measurements but also to provide one with some new knowledge on the formation of common vascular complications. References [1] SIVANESAN S., HOW T.V., BLACK R.A., BAKRAN A., Flow patterns in the radiocephalic arteriovenous fistula: an in vitro study, J. Biomech., 1999, Vol. 32, 915 925. [2] WIESE P., NONNAST-DANIEL B., Colour Doppler ultrasound in dialysis access, Nephrol. Dial. Transplant., 2004, Vol. 19, 1956 1963. [3] NIEMANN A.K., UDESEN J., THRYSOE S., Can sites prone to flow induced vascular complications in a-v fistulas be assessed using computational fluid dynamics?, J. Biomech., 2010, Vol. 43, 2002 2009. [4] JÓŹWIK K., OBIDOWSKI D., Numerical simulations of the blood flow through vertebral arteries, J. Biomech., 2010, Vol. 43, 177 185. [5] JODKO D., Simulation of the blood flow in the fistula for haemodialysis, Łódź University of Technology, Master Thesis (in Polish), 2011. [6] BOCHENEK A., REICHER M., Human Anatomy, Vol. III, Warsaw, PZWL, 1974. [7] JOHNSTON B.M., JOHNSTON P.R., CORNEY S., KILPATRICK D., Non-Newtonian blood flow in human right coronary arteries: steady state simulations, J. Biomech., 2004, Vol. 37, 709 720. [8] SHIN S., DO-YOUNG KEUM, Measurement of blood viscosity using mass-detecting sensor, J. Biomech., 2002, Vol. 17, 383 388.